Basic 100

This document provides specifications for a Mobile X-Ray Unit Basic 100-30. Key details include: - It operates on single phase 230V voltage and 50/60Hz frequency. - It has a maximum power output of 7.5kW for low power mode and 30kW for high power mode. - It can generate X-rays between 40-100kV and exposure times range from 3ms to 1.3s for low power mode and 1ms to 0.5s for high power mode. - The monoblock unit has a focal spot size of 0.6mm or 1.3mm, rotating anode, and inherent minimum filtration of 0.7mm aluminum equivalent

Uploaded by

Hectorl02
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
100% found this document useful (1 vote)
721 views4 pages

Basic 100

This document provides specifications for a Mobile X-Ray Unit Basic 100-30. Key details include: - It operates on single phase 230V voltage and 50/60Hz frequency. - It has a maximum power output of 7.5kW for low power mode and 30kW for high power mode. - It can generate X-rays between 40-100kV and exposure times range from 3ms to 1.3s for low power mode and 1ms to 0.5s for high power mode. - The monoblock unit has a focal spot size of 0.6mm or 1.3mm, rotating anode, and inherent minimum filtration of 0.7mm aluminum equivalent

Uploaded by

Hectorl02
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 4

'

Mobile X-Ray Unit Basic 100 - 30 '


Specifications
* Single Phase Voltage 230 Vac (± 10%) ''
* Frequency 50/60 Hz
*
*
Max. Current Absorbation
Line, Compensation
12A
Automatic
'
* Line Resistance <2.5Ω
* Swing Arm '
LP (Low Power) I HP (High Power) ''
* Max. Power 7.5 kW I 30 kW ''
*
*
kV Range
mAs Range
40+ M 25 kV (step of 1 kV)
0.5+200 mAs in 25 Values '
* max. mA 425mA ''
* Exposure Time 3ms + 1.3s(LP) I 1 m s-0.5s (HP)
(Selected by the Processor According to '
the mAs)
* H.V. Driving Inverter Driven by IGBT
*
*
Operating Frequency
Ripple
100kHz
<_3% to maximum power
''
* Rise Time <1 ms
''
Operating Mode and Fuctionality
* Users Interface : Polycarbonate flat keyboard with alphanumeric

''
LED Display
* Operating Modes : Two points techniques ( kV, mAS )
40 programmable anatomic tehnique (20 for LP and

''
20 for HP)
* X - Ray Control : Distance control with double - click and extensible
cable (> 4 m)
* Safety Filament Current ''
mA min & mA max
Maximum exposure time

''
Temperature maximum X-Ray tube head
: Count thermal units X-Ray tube head
: Max. kV, Min. kV, MaxAkV, max. I.
: Anode Rotation
: Microprocessor self - test
Monoblock
* Focal Spot Size : 0.6 (Small Focus) 1.3 mm (Large Focus) ''
* Anode Type : Rotating Anode ( 3.000 rpm )
* Nominal Anodic Power : 11 kW for small filament 32 kW for large filament
* Anode Diameter : 064 mm
Anode Material : RTM
* Anode Angle : 15°
* Radiation Field : a 70 cm 36 cm
: a 100 cm 50 cm
* Inherent Minimum Filteration : 0.7mm Al eq. / Total 1.8 mm Al at 80 kV
* ·Max. Anode Heat Content : 80 kJ 10 7kHU

* Max. Continous heat dissipation 300 W


* Nominal X-ray tube Voltage 130kV
* Max. Filament Current 5.4 Amp

Collimator
* Multi Level Blades : Parallel and Perpendicular
* Blade Movement : Manual
* Size of Radiated Field : With focus film distance of 1 m 43 x 43cm.
* Focus Film Distance Indication : With Retracticle tape measure (2m).
.. ..
...... * Inherent Collimator Filtration : 2 mm AI.
.. .. .. .. .. ..
... ... ...

---------
------------
INTERMEDICAL.S.r.l
IMD Group
24050 Grassobbio (Bg) Italy
Via E. Fermi 26