TE30 Electricity Meter Tester and Power Quality Analyzer Presentation en
TE30 Electricity Meter Tester and Power Quality Analyzer Presentation en
TE30 Electricity Meter Tester and Power Quality Analyzer Presentation en
www.calmet.com.pl Analyzer
Testing whole measurement system!
Accuracy of all kind of CT/PT burden,
Electricity Meters ε[%] ratio, phase shift
Electromechanical error
Meters
Connection
errors
Electronic (static)
Meters
4-Quadrants
Meters
1 3 4 6 7 9 10 12 1 3 4 6 7 9 10 12
L1 L1
L2 L2
L3 L3
N N
Electricity meter (CT) testing – direct connection Electricity meter testing (CT) – connection with clamps
1 2 3 4 5 6 7 8 9 11 1 2 3 4 5 6 7 8 9 11
P1 P2 P1 P2 P1 P2 P1 P2 P1 P2 P1 P2
L1 L1
L2 L2
L3 L3
N N
TE30 Electricity Meter Tester and Power Quality 5
www.calmet.com.pl Analyzer
RMS measurement results and TE30 Analyzer configuration:
3-Phase Star or Voltage direct connection Automatic or Costant range used Actual Date and
Delta connection Current mesured by flexible clamps for measurements Time
← TABLE
CHART →
TE30 Electricity Meter Tester and Power Quality 8
www.calmet.com.pl Analyzer
Energy meter Register testing on site and laboratory
function of energy measurement with method of
setting time periods for verification of energy meter
Register directly in percent [%],
function of energy measuremnt for
power P, P+, P-, Q, Q+, Q-, S,
function of energy measurement for
the first (fundamental) harmonic of
active power PH1 and reactive power
QH1
Connection diagram
CT Ratio error and phase error testing
1 2 3 4 5 6 7 8 9 11
function of computing
S1 S2 S1 S2 S1 S2
transformer ratio error
L1
P1
P
P1
P
P1
P
directly in percent [%]
2 2 2
L2
L3
N
function of computing phase
shift error [°]
TE30 Electricity Meter Tester and Power Quality 10
www.calmet.com.pl Analyzer
CT, PT Transformers testing (LV i MV, voltage and current, simultaneously in
three phases) directly on site: CT / PT burden testing
1 2 3 4 5 6 7 8 9 11
S1 S2 S1 S2
Why the tranformer burden
S1 S2
L1
P1
P
P1
P
P1
P
(load) is so important?!
2 2 2
L2
L3
N
TE30 Electricity Meter Tester and Power Quality 11
www.calmet.com.pl Analyzer
CT Transformer testing: burden testing
CT – current transformer can operate with
ε – ratio error [%]
stated accuracy only between 25% - 100% of
burden (load). In case of too long, or too thin
Limit of
wire dimension or too small load, the result,
accuracy secondary current can be out of accuracy limits
[%] transformer power rating Sn
25% 100%
Normal work
area
Sn=5VA 1 2 3 4 5 6 7 8 9 11
Ip Is Connection wires
L=10m
Example (power loss in cables):
Us1 Us2 mm2
ρCU ⋅ l 0,0175Ω m ⋅ 2 ⋅10m
RP = = = 0,35Ω
S 1mm2
Current transformer PP = I 2 2 ⋅ RP = 52 A ⋅ 0,35Ω = 8, 75VA
TE30 Electricity Meter Tester and Power Quality 12
www.calmet.com.pl Analyzer
Function of power quality analyser + recording
analyzing of measurement
results for EN 50160
compatibility or individual
requirements of user
recording of power network parameters in the SD Flash 4-32GB memory, which gives
(8÷64)x106 sets of network parameters or long-term registration of power quality
TE30 Electricity Meter Tester and Power Quality 13
www.calmet.com.pl Analyzer
Function of power quality analyser + recording
Contact:
Calmet Ltd., Kukulcza 18 Street, 65-472 Zielona Gora, Poland EU
www.calmet.com.pl e-mail: [email protected]
Tel.: +48 68 324 04 56 Fax: +48 68 324 04 57