Advanced Analysis Method For Energy Material
Advanced Analysis Method For Energy Material
[1] Explain Bragg's law using schematic diagram and mathematical equation.
On the previous diagram, the “reflected” rays combine to form a diffracted beam if they differ
in phase by a whole number of wavelengths, that is, if the path difference AB-AD = n where
[2] Explain what kind of information is available from X-ray diffraction (XRD) and how to
Information:
X-ray crystallography relies on the dual wave/particle nature of x-rays to discover
information about the structure of crystalline materials.
X-ray fluorescence spectrometry relies on characteristic secondary radiation emitted
by materials when excited by a high energy x-ray source and is used primarily to
determine amounts of particular elements in materials.
X-ray radiography is used for creating images of light-opaque materials relies on the
relationship between density of materials and absorption of x-rays. Applications
include a variety of medical and industrial applications.
Uses:
Identification of single-phase materials – minerals, chemical compounds or other
engineered materials.
Identification of multiple phases in microcrystalline mixtures (i.e., rocks).
Determination of crystal structure of identified materials.
Recognition of amorphous materials in partially crystalline mixtures.
Methods for identification and structural analysis of clay minerals.
[3] If your sample contains a lot of nanoparticle but don't know the size, then describe
X-ray Powder Diffraction: Obtain the values from a table for the wavelength used by XRD
machine.
1) Pick the largest peak and look in the appropriate section of the search manual
2) Pick the second peak, look in the first several columns of section for a match.
3) Once you have found a decent match for two peaks, check for other peaks.
4) Relative intensity of peaks is given in the subscript and provides an approximate guide-
[4] The following is the list of the keyword often appeared in the research paper. Describe
briefly it.
Lattice: Crystalline solids consist of regular arrays of atoms, ion, or molecules with interatomic
Reciprocal Space: The reciprocal lattice exists in reciprocal space (also known as momentum
space or less commonly as K-space). The reciprocal lattice represents the Fourier transform of
direct lattice.
Real Space: Direct lattice is usually a periodic spatial function in real-space and is what one
ICSD: Inorganic Crystal Structure Data Base is for fully identified inorganic crystal structures.
PDF: Power Diffraction File contains over 300,000 diffraction patterns. Modern computer
programs can help you determine what phases are present in sample by quickly comparing
[5] Compare the powder diffraction method with the single crystal diffraction method.
Powder Diffraction
Unlike single-crystal XRD, powder XRD looks at a Single-crystal XRD allows for absolute
large sample of polycrystalline material and structure determination. With single-
therefore is considered a bulk characterization crystal XRD data, the exact atomic
technique. The powder pattern is considered a positions can be observed, and thus
"fingerprint" for a given material; it provides bond lengths and angles can be
information about the phase (polymorph) and determined. This technique provides the
crystallinity of the material. Typically, powder structure within a single crystal, which
XRD is used to study minerals, zeolites, metal- does not necessarily represent the bulk
organic frameworks (MOFs), and other extended of the material. Therefore, additional
solids. Powder XRD can also be used to establish bulk characterization methods must be
bulk purity of molecular species. utilized to prove the identity and purity
of a compound.
All the symmetry-equivalent reflections have the When the space-group symmetry is
same d spacing with the result that individual unknown, reflections can be measured
intensities cannot be measured. and compared with other possible
symmetry equivalent reflections.
The three-dimensional data are “compressed” The information is distributed in three-
into one-dimension. dimensional space.
Although structure determination by powder X- For structures with less than 100 atoms
ray diffraction may not be as accurate as that by in the asymmetric unit, single-crystal
single-crystal X-ray diffraction, improvements will techniques are always guaranteed to be
likely be achieved through future advancements successful with accurate parameters of
in instrumentation and algorithm development. cell dimensions.
[6] Define the following Rietveld method related keyword and describe it.
Rwp: Weighted profile R-factor, follows directly from the square root of the quantity minimized,
2 2
scaled by the weighted intensities 𝑤𝑖 : R 2 w y C,i y O,i / w y O,i
wp i i
i i
RB: R-Bragg R I
B
K
O,K
I C , K / I O, K . IK is the intensity of Bragg K reflection at the end
K
RP: R Profile R y
p
i
O ,i
y C ,i
/ y
i
O ,i
Fractional Coordinate: are expressed as fractions of the unit cell in each of the three directions
Profile Parameter: the parameters of the diffraction profile. The parameters specific to each
phase, which vary during the refinement are structural and non-structural (parameters of the
half width (U, V, W), asymmetry, 2θ zero, preferred orientation and coefficients of the
background radiation.
[7] Define the structural parameters in Rietveld method and describe it briefly.
Atomic positions: Crystal structure is considered known when atom positions are known very
atoms in the diffracting planes. X-ray diffraction intensities ––> atom positions.
Unit cell parameters: Indexing the XRD pattern, a [Å], b [Å], c [Å], β [˚], V [Å3]. The diffraction
peak position is a product of the atomic bond distances in the crystal. X-ray diffraction
Structure factor: (Fkhl) quantifies the amplitude of light scattered by a crystal 𝐹𝑘ℎ𝑙 =
𝑗=1 𝑁𝑗 𝑓𝑗 𝑒𝑥𝑝[2𝜋𝑖(ℎ𝑥𝑗 + 𝑘𝑦𝑗 + 𝑙𝑧𝑗 )]. Fhkl sums the result of scattering from all of the atoms in
∑𝑚
the unit cell to form a diffraction peak from the (hkl) planes of atoms.
Scattering factor: (fj) quantifies the efficiency of scattering from a group of electrons in an
atom. The number of electrons around an atom defines how strongly it will scatter the incident
X-ray beam.
Occupancy factors: (sof) indicates what fraction of a site is occupied by a specific atom. The
sof is Nj in the structure factor equation. sof = 1, every equivalent position for xyz is occupied
Wyckoff Site notation: it is a shortcut to indicate the type of site that the atoms occupy in
the crystal, and if x, y or z can be changed or they must be fixed to preserve the symmetry
of the crystal.
Thermal vibration parameters (isotropic and anisotropic): Thermal motion of the atom
changes the scattering factor. The amount of vibration is quantified by the Debye-Waller
temperature factor.
[8] Why you need the quantitative method using X-ray diffraction?
determination of crystal structure or crystallite size and shape, with quantifiable numerical
precision from the experimental data itself. Though “standard” patterns and structural data
are used as a starting point, the most successful quantitative analysis usually involves modeling
the diffraction pattern such that the calculated pattern(s) duplicates the experimental one.
Single line method: The procedure consists in measurements of a given peak intensity for a
sample and for the standard. There are many methods, which utilize single-line intensities for
the Quantitative Phase Analysis (QPA), and these methods can be classified into one category,
Multi line method: This method uses several lines per element in diffraction spectra for
calculations, and high accurate and precise concentration per element can be obtained.
methods simultaneously refine the unit-cell parameters and extract the best estimations of
the Bragg peak intensities from a complete diffractogram. Methods in this category are based
[10] Compare the strong and weak points of the Single line method, Multi line method
http://www.ccp14.ac.uk/.
Fullprof: program has been mainly developed for Rietveld analysis (structure profile
refinement) of neutron (constant wavelength, time of flight, nuclear and magnetic scattering)
or X-ray powder diffraction data collected at constant or variable step in scattering angle
2theta. The program can be also used as a Profile Matching (or pattern decomposition using
Le Bail method) tool, without the knowledge of the structure. Single crystal refinement can
also be performed alone or in combination with powder data. Time of flight (TOF) neutron
data analysis is also available. Energy dispersive X-ray data can also be treated but only for
profile matching.
GSAS: is a set of programs for the processing and analysis of both single crystal and powder
diffraction data obtained with x-rays or neutrons. It is capable of handling all of these types
of data simultaneously for a given structural problem. In addition, it can handle powder
diffraction data from a mixture of phases refining structural parameters for each phase. GSAS
is designed to be easy to use and features a menu driven editor (EXPEDT) equipped with help
listings and error trapping features. EXPEDT is used to prepare all input for the main calculation
programs in GSAS. The entire GSAS system is written in FORTRAN and uses some system
callable features.
EXPO: can solve crystal structures by powder diffraction data by using reciprocal as well as
direct space methods. It is able to carry out the full pathway of the solution process: indexing,
space group determination, estimation of the integrated intensities, ab-initio and non ab-initio
structure solution, Rietveld refinement. With respect to the previous version, EXPO2014 has
been enriched with new powerful computing tools for enhancing the structure solution
process.
FOX: program was made for the ab initio crystal structure solution from diffraction data
(mostly powder diffraction data). Its most interesting features for ab initio structure
determination are:
an automatic correction for special positions and shared atoms between polyhedra,
the ability to use simultaneously multiple powder patterns (X-rays, neutrons), as well as
smart global optimization algorithms which can get out of false minima.
a graphical interface (see the screenshots) with a 3D crystal structure view, with live
X-ray absorption spectroscopy (XAS) is a widely used technique for determining the local
synchrotron radiation facilities, which provide intense and tunable X-ray beams. Samples can
be in the gas-phase, solution, or as solids. XAS data is obtained by tuning the photon energy,
[13] What is the difference between X-ray absorption and X-ray diffraction for application?
X-ray diffraction is a technique in which diffraction of X rays with the crystal lattice are
observed. X rays which are diffracted at Bragg's angle are diffracted and combine to give a
pattern which shows about the planes in the crystal. X ray diffractogram is resulted and the
position and intensity of the peaks are observed. In short, XRD is used to determine crystal
structure and degree of crystallinity of a material. Probes average, long range structure (>10
While X-ray absorption is a technique where the absorption of X-rays in the material and then
emission of other x rays which are resulted due to interaction of incident x rays with the
electrons are observed. The emitted rays have a specific energy depending on the elements
present in the material and are analyzed by spectrometer. By this method elemental
composition of the material is determined. Probes local structural environment (<1 nm).
[14] Compare or explain the X-ray absorption process using photoelectric effect.
In this process, an x-ray photon is absorbed by an electron in a tightly bound quantum core
level (such as the 1s or 2p level) of an atom. In order for a particular electronic core level to
participate in the absorption, the binding energy of this core level must be less than the
energy of the incident x-ray. In this case, the x-ray is destroyed (i.e., absorbed) and any energy
in excess of the electronic binding energy is given to a photo-electron that is ejected from
the atom.
Following an absorption event, the atom is said to be in an excited state, with one of the core
electron levels left empty (a so-called core hole), and a photo-electron. The excited state will
There are two main mechanisms for the decay of the excited atomic state following an x-ray
absorption event. The first of these is x-ray fluorescence, in which a higher energy electron
core-level electron fills the deeper core hole, ejecting an x-ray of well-defined energy. The
second process for de-excitation of the core hole is the Auger Effect, in which an electron
drops from a higher electron level and a second electron is emitted into the continuum (and
K: K-edge is the binding energy of the K-shell (1s) electron of an atom. In order to
photoelectric absorption of the photons occurs, the photons must have more energy than the
binding energy of the K-shell electrons (K-edge).
LI, LII, LIII: LI edge corresponds to the excitation of a 2s electron which requires more energy
than a 2p electron. The 2p electron excitation is split into two edges namely LII and LIII, as a
2p electron gets excited, an open shell 2p5 electronic configuration forms, consequently, spin–
orbit coupling of such a system occurs. A 2p5 excited states corresponds to two terms,
2P1/2 which has higher energy and give rise to the LII edge, and 2P3/2 which has lower energy
and give rise to the LIII edge. Due to degeneracy, the LIII edge has twice the edge jump of the
LII and LI edges.
Destructive interference: the interference of two waves of equal frequency and opposite
phase, resulting in their cancellation where the negative displacement of one always coincides
with the positive displacement of the other.
Constructive interference: the interference of two or more waves of equal frequency and
phase, resulting in their mutual reinforcement and producing a single amplitude equal to the
sum of the amplitudes of the individual waves.
deviate from a straight trajectory by one or more paths due to localized non-uniformities in
the medium through which they pass. The types of non-uniformities which can cause
scattering are known as scatterers or scattering centers It is very common that scattering
centers are grouped together; in such cases, radiation may scatter many times, in what is
known as multiple scattering.
[16] What is the benefit of Laboratory EXAFS compared to Synchrotron Radiation EXAFS?
Reliable structural information with the EXAFS technique can only be obtained from high
quality experimental data. In order to get a sufficient signal to noise ratio high intensity
Disadvantages of the use of synchrotron radiation are: limited availability of beamtime, high
cost and sometimes unreliable operation. Laboratory EXAFS spectrometer has been developed
for rapid and ease collection of high-quality EXAFS data. It enables unlimited use of instrument,
When a group of spins is placed in a magnetic field, each spin aligns in one of the two
possible orientations. At room temperature, the number of spins in the lower energy level,
N+, slightly out numbers the number in the upper level, N-. Boltzmann statistics tells us that
𝑁 − ⁄𝑁 + = 𝑒𝑥𝑝(−𝐸⁄𝑘𝑇) where E is the energy difference between the spin states; k is
Boltzmann's constant, 1.3805 x 10-23 J/Kelvin; and T is the temperature in Kelvin. As the
temperature decreases, so does the ratio N- /N+. As the temperature increases, the ratio
approaches one.
The signal in NMR spectroscopy results from the difference between the energy absorbed by
the spins which make a transition from the lower energy state to the higher energy state, and
the energy emitted by the spins which simultaneously make a transition from the higher
energy state to the lower energy state. The signal is thus proportional to the population
A spin packet is a group of spins experiencing the same magnetic field strength. In this
example, the spins within each grid section represent a spin packet.
At any instant in time, the magnetic field due to the spins in each spin packet can be
represented by a magnetization vector. The size of each vector is proportional to (N+ - N-).
The vector sum of the magnetization vectors from all of the spin packets is the net
magnetization.
The simplest NMR experiment is the continuous wave (CW) experiment. There are two ways
of performing this experiment. In the first, a constant frequency, which is continuously on,
probes the energy levels while the magnetic field is varied. The energy of this frequency is
represented by the blue line in the energy level diagram. The CW experiment can also be
performed with a constant magnetic field and a frequency which is varied. The magnitude of
the constant magnetic field is represented by the position of the vertical blue line in the
If a signal is weak, many scans must be averaged to enhance the signal/noise ratio.
Signal/noise ratio rises as the square root square root of number of scans. Only FT
A scan can be performed much more quickly. CW-NMR: 5 minutes; FT-NMR: 5 seconds.
Spin: Spin is a fundamental property of nature like electrical charge or mass. Spin comes in
multiples of 1/2 and can be + or -. Protons, electrons, and neutrons possess spin. Individual
unpaired electrons, protons, and neutrons each possesses a spin of 1/2. In the deuterium
atom (2H), with one unpaired electron, one unpaired proton, and one unpaired neutron, the
gyromagnetic ratio: In physics, the gyromagnetic ratio of a particle or system is the ratio of
its magnetic moment to its angular momentum, and it is often denoted by the symbol . Its
SI unit is the radian per second per tesla (rad⋅s− 1⋅T− 1) or, equivalently, the coulomb per
kilogram (C⋅kg− 1). When placed in a magnetic field of strength B, a particle with a net spin
can absorb a photon of frequency . The frequency depends on the gyromagnetic ratio, of
the particle. = 𝛾 ∙ 𝐵. For hydrogen, = 42.58 MHz / T.
Energy level: When collection of identical nuclei is placed in a strong external magnetic field
B, an interaction will occur between each individual nuclear magnetic vector and this field.
Therefore, two groups of nuclei appear according to the direction of their spin vector, whose
energies correspond to low energy state E1 and to high energy state E2. These orientations
are called ‘spin up’ (magnet image N-S-N-S) and ‘spin down’ (N-N-S-S).
Transition: A particle can undergo a transition between the two energy states by the
absorption of a photon. A particle in the lower energy state absorbs a photon and ends up
in the upper energy state. The energy of this photon must exactly match the energy difference
between the two states. The energy, E, of a photon is related to its frequency, (Larmor
Rotating frame: A rotating frame of reference rotates about the Z axis at the Larmor frequency.
This rotating coordinate system can be distinguished from the laboratory system by primes
on the X and Y axes, X'Y'. A magnetization vector rotating at the Larmor frequency in the
laboratory frame appears stationary in a frame of reference rotating about the Z axis.
Pulsed magnetic field: A coil of wire placed around the X axis will provide a magnetic field
along the X axis when a direct current is passed through the coil. An alternating current will
produce a magnetic field which alternates in direction. In magnetic resonance, the magnetic
field created by the coil passing an alternating current at the Larmor frequency is called the
B1 magnetic field. When the alternating current through the coil is turned on and off, it creates
a pulsed B1 magnetic field along the X' axis. The spins respond to this pulse in such a way as
to cause the net magnetization vector to rotate about the direction of the applied B1 field.
Spin relaxation: In NMR spectroscopy spin relaxation describes how net magnetization vector
returns to its equilibrium value after the nuclear spin system is exposed to energy of a
frequency equal to the difference between the spin states. Motions in solution which result in
time varying magnetic fields cause spin relaxation. Time varying fields at the Larmor frequency
cause transitions between the spin states and hence a change in MZ.
Free induction decay: In FT NMR, free induction decay (FID) is the observable NMR signal
generated by non-equilibrium nuclear spin magnetization precessing about the magnetic field
applying a pulse of resonant radio-frequency close to the Larmor frequency of the nuclear
spins.
It is possible to change the net magnetization by exposing the nuclear spin system to energy
of a frequency equal to the energy difference between the spin states. If enough energy is
put into the system, it is possible to saturate the spin system and make MZ=0.
The time constant which describes how MZ returns to its equilibrium value is called the spin
lattice relaxation time (T1). The equation governing this behavior as a function of the time t
If the net magnetization is placed in the XY plane it will rotate about the Z axis at Larmor
frequency. The time constant which describes the return to equilibrium of the transverse
magnetization, MXY, is called the spin-spin relaxation time, T2. 𝑀𝑋𝑌 = 𝑀𝑋𝑌𝑜 𝑒 −𝑡⁄𝑇2
[23] Define the Deshielding effect and the chemical shift and explain it.
In some cases, such as the benzene molecule, the circulation of the electrons in the aromatic
orbitals create a magnetic field at the hydrogen nuclei which enhances the Bo field. This
phenomenon is called deshielding. In this example, the Bo field is applied perpendicular to
the plane of the molecule. The ring current is traveling clockwise if you look down at the
plane.
The electron density around each nucleus in a molecule varies according to the types of nuclei
and bonds in the molecule. The opposing field and therefore the effective field at each nucleus
The chemical shift of a nucleus is the difference between the resonance frequency of the
nucleus and a standard, relative to the standard. This quantity is reported in ppm and given
Those nuclei experiencing different environment or having different chemical shifts are
nonequivalent. Nuclei which are close to one another exert an influence on each other's
effective magnetic field. This effect shows up in the NMR spectrum when the nuclei are
nonequivalent. If the distance between non-equivalent nuclei is less than or equal to three
bond lengths, this effect is observable. This effect is called spin-spin coupling or J coupling.
Consider the following example. There are two nuclei, A and B, three bonds away from one
another in a molecule. The magnetic field at nucleus A will be either greater than Bo or less
A Fourier transform is an operation which converts functions from time to frequency domains.
An inverse Fourier transform (IFT) converts from the frequency domain to the time domain.
A magnetization vector, starting at +x, is rotating about the Z axis in a clockwise direction.
The plot of Mx as a function of time is a cosine wave. Fourier transforming this gives peaks at
both + and - because the FT cannot distinguish between a + and a - rotation of the
A plot of My as a function of time is a -sine function. Fourier transforming this gives peaks at
+ and - because the FT cannot distinguish between a positive vector rotating at at + and
The solution is to input both the Mx and My into the FT. The FT is designed to handle two
orthogonal input functions called the real and imaginary components. Detecting just the Mx
or My component for input into the FT is called linear detection. Detection of both Mx and
[26] What is the effect on NMR experiment when the molecular motion is restricted?
Spatial proximity and/or a chemical bond between two atoms can give rise to interactions
between nuclei. In media with no or little mobility (e.g. crystals, powders, large membrane
found in solid-state NMR are the chemical shift anisotropy (CSA) and the internuclear dipolar
coupling. Many more such interactions exist, such as the anisotropic J-coupling in NMR.
Anisotropic interactions between nuclei, which are usually averaged by Brownian motion in
[27] How can you remove the effect on the NMR experiment when the molecular motion
Several methods have been developed and considered in order to minimize large anisotropic
NMR interactions between nuclei and increase signal in rare spin NMR spectra:
depending on the rotor size and type of experiment being conducted) at the magic angle
θm (ca. 54.74°, where cos2θm=1/3) with respect to the direction of the magnetic field, the
normally broad lines become narrower, increasing the resolution for better identification
Advanced Pulse Techniques: Pulse sequences are used to excite signals and to looks for
“free induction decay” and “spin echoes” that are observed in an NMR spectrometer. They
Cross polarization enhances signal from dilute spins potentially by a factor of 𝐼 ⁄𝑆 ,
Since abundant spins are strongly dipolar coupled, they are therefore subject to large
fluctuating magnetic fields resulting from motion. This induces rapid spin-lattice
relaxation at the abundant nuclei. The end result is that one does not have to wait for
Polarization is transferred during the spin locking period, (the contact time) and a /2 pulse
The key to obtaining efficient cross polarization is setting the Hartmann-Hahn match properly.
[29] What happen when the solid sample rotates in MAS NMR?
By spinning the sample (usually at a frequency of 1 to 130 kHz, depending on the rotor size
and type of experiment being conducted) at the magic angle θm (ca. 54.74°, where cos2θm=1/3)
with respect to the direction of the magnetic field, the normally broad lines become narrower,
increasing the resolution for better identification and analysis of the spectrum. The rate of
MAS must be greater than or equal to the magnitude of the anisotropic interaction to average
it to zero.
[30] What is the Spinning rate and resulting effect on the NMR spectrum?
By spinning the sample (usually at a frequency of 1 to 130 kHz, depending on the rotor size
and type of experiment being conducted) at the magic angle θm (ca. 54.74°, where cos2θm=1/3)
with respect to the direction of the magnetic field, the normally broad lines become narrower,
increasing the resolution for better identification and analysis of the spectrum. The rate of
MAS must be greater than or equal to the magnitude of the anisotropic interaction to average
it to zero.
[31] Please explain the real image and reciprocal space image in EM using Fourier
transform.
A real space shift results in multiplication with a phase factor in reciprocal space:
1) A Fourier transform FT (Fourier analysis) that creates the diffraction pattern of the object
in the back focal plane. The diffraction pattern is the Fourier transform of the scattered
electron wave: in turn the primary image is the Fourier transform of the diffraction pattern,
and
2) An inverse FT (Fourier synthesis) that makes the interference of the diffracted beams back
The diffraction pattern from a thin specimen can be considered as the Fourier transform (FT)
of the specimen.
The “reciprocal lattice vectors” a∗, b∗ and c∗ generate the “reciprocal lattice”:
[32] What is the difference between ED and XRD?
Both, ED and XRD, are caused by constructive interference of scattered waves, and the same
fundamental laws (e.g., Bragg law, extinction rules) can be applied for the interpretation of
the resulting diffraction patterns. In both cases, diffraction patterns of powders and of single
crystals appear.
1. The wavelength of electrons (e.g., 1.97 pm for 300 keV electrons) is much shorter than
that of X-rays (about 100 pm). Therefore, the radius of the Ewald sphere is much larger
2. The diffraction angles are very small in ED: 0 < Θ < 2° (cf., XRD: 0 < Θ < 180º)
3. Electrons are scattered by the positive potential inside the electron cloud (Coulomb
interaction), while X-rays interact with the electron cloud. As the result, the interaction of
This has the advantage that the diffracted electron beams have a high intensity and exposure
times are in the order of a few seconds. ED patterns can directly be observed on the viewing
screen of the electron microscope. Thus, orienting a crystal along a direction can be easily
diffraction patterns can be obtained from very small crystals selected with a diffracted aperture
(Selected Area Electron Diffraction SAED) and by a focused electron beam even from nm-
The disadvantage of the strong interaction is that multiple scattering plays an important role,
and the intensities of the reflections are much influenced by this dynamical effect. This makes
structure determination from ED more difficult and less reliable than that from XRD data. A
[33] Is possible to use ED for the structural determination of powder sample and justify
your answer.
Electron crystallography (EC) is another most valuable technique for structure determination.
With the electron as a probe, EC alone can also be used for structure determination, especially
for crystals that are too small to be studied by SCXRD or too complex for PXRD. As electrons
interact much more strongly with matter than X-rays do, both electron diffraction (ED) patterns
and high-resolution transmission electron microscopy (HRTEM) images can be obtained from
nanosized crystals. Nano- and micrometer-sized crystals can be treated as single crystals in
EC, whereas they are considered to be powders in X-ray diffraction. The reflections in ED
patterns are completely separated, and the structure factors, including amplitudes and phases,
However, collecting a complete set of ED patterns or recording a good HRTEM image requires
strong interactions between electrons and materials can also lead to dynamical effects and
beam damage. These difficulties make structure determination from ED patterns and HRTEM
images not straightforward. All in all, EC and X-ray crystallography are remarkably
complementary.
[34] When you combine the EM and EDX, what kind of information is available?
used in conjunction with scanning electron microscopy (SEM). he EDS technique detects x-
rays emitted from the sample during bombardment by an electron beam to characterize the
can be analyzed.
When the sample is bombarded by the SEM's electron beam, electrons are ejected from the
atoms comprising the sample's surface. The resulting electron vacancies are filled by electrons
from a higher state, and an x-ray is emitted to balance the energy difference between the
two electrons' states. The x-ray energy is characteristic of the element from which it was
emitted. The EDS x-ray detector measures the relative abundance of emitted x-rays versus
their energy. The spectrum of x-ray energy versus counts is evaluated to determine the