Alaris Signature PDF
Alaris Signature PDF
HOLD
ml/hr
OPT PRI HLD SEC KVO
PRI SEC
OPTIONS
1 2 3
4 5 6
7 8 9
. 0 Clear Enter
A B
RUN RUN
HOLD POWER POWER HOLD
ml/hr ml/hr
KVO OPT HLD PRI SEC PRI SEC HLD OPT KVO
PRI SEC
A B OPTIONS
1 2 3
4 5 6
7 8 9
• 0 Clear Enter
Signature Edition® GOLD Infusion System, Models 7130/7131 and 7230/7231 iii
Technical Service Manual
TABLE OF CONTENTS
Chapter 6 - Troubleshooting
6.1 Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-1
6.2 Diagnostics Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-11
6.2.1 Entering Diagnostics Mode .......................................................................... 6-11
6.2.2 Setting Preventive Maintenance Interval . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-12
6.2.3 Viewing Alarm or Event History (Event Log) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-13
6.2.4 Setting Time (and Date) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-16
6.2.5 Viewing Battery Status ............................................................................... 6-16
6.2.6 Changing Rated Capacity of Battery ................................................................ 6-17
6.2.7 Viewing DC Voltages ................................................................................. 6-18
6.2.8 Setting ID Number .................................................................................... 6-18
6.2.9 Viewing Battery and Total Run Times . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-19
6.2.10 Viewing Self-Check Timer ........................................................................... 6-19
List of Figures
1-1 Battery Refresh Cycle ................................................................................ 1-9
1-2 Lower LCD Display Layout . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-11
1-3 Resistance Graph .................................................................................... 1-14
1-4 Pressure and Resistance Graph .................................................................... 1-15
2-1 Map of Configuration Screens (Software versions 4.06 and higher) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-22A
2-2 Map of Configuration Screens (Software version 2.78) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-23B
3-1 Rate Accuracy Test Setup ........................................................................... 3-5
3-2 Pressure Test Setup ................................................................................. 3-8
4-1 Main Block Diagram .................................................................................. 4-2
4-2 COMBO IC Block Diagram . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-5
4-3 Electrical Partitioning ................................................................................. 4-6
4-4 Battery Manager Block Diagram . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-7
4-5 Battery Monitor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-12
4-6 Main Power Supply ................................................................................... 4-13
4-7 System Watchdog .................................................................................... 4-14
4-8 System Reset/Power On ............................................................................. 4-15
4-9 Motor Drive Circuit, Phase 1(A) ..................................................................... 4-19
4-10 Motor and Mechanism Sensors Block Diagram .................................................... 4-19
4-11 Air-in-Line Detector Block Diagram ................................................................. 4-20
4-12 Pressure Sensor Interface Block Diagram . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-20
Signature Edition® GOLD Infusion System, Models 7130/7131 and 7230/7231 vii
Technical Service Manual
TABLE OF CONTENTS
List of Tables
1-1 Specifications ......................................................................................... 1-3
1-2 Battery Trip Points .................................................................................... 1-9
1-3 Abbreviations, Acronyms, Symbols ................................................................ 1-17
2-1 Record of Configured Instruments .................................................................. 2-20A
2-2 Drug List (7130/7230 only) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-6B
2-3 Record of Configured Instruments .................................................................. 2-20B
2-4 Drug List Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-22B
3-1 PM Inspections ....................................................................................... 3-11
4-1 Motor Control Signals ................................................................................ 4-18
5-1 Test Equipment ....................................................................................... 5-1
viii Signature Edition® GOLD Infusion System, Models 7130/7131 and 7230/7231
Technical Service Manual
TABLE OF CONTENTS
T H I S PA G E
I N T E N T I O N A L LY
LEFT BLANK
Signature Edition® GOLD Infusion System, Models 7130/7131 and 7230/7231 1-1
Technical Service Manual
GENERAL INFORMATION
1-2 Signature Edition® GOLD Infusion System, Models 7130/7131 and 7230/7231
Technical Service Manual
GENERAL INFORMATION
Administration Sets: Use only ALARIS Medical Systems® 72 Series administration sets. All
disposable IV set and IV set accessory models are defined on a separate
card.
Signature Edition® GOLD Infusion System, Models 7130/7131 and 7230/7231 1-3
Technical Service Manual
GENERAL INFORMATION
NOTE: When the occlusion alarm pressure limit is set to the maximum
threshold setting, the maximum infusion pressure generated into a hard
occlusion at 25 mL/h is 11.6±3.9 psi. Testing performed using IV set
Model 72003, at 68±4°F (20±2°C).
Critical Volume: Maximum incremental volume in case of single point failure will not
exceed 1.0 mL at 999.9 mL/h.
Dimensions (Nominal): 7130/7131 7230/7231
Width 7.6 in/19.3 cm 10.7 in/26.7 cm
Height 8.6 in/21.8 cm 8.6 in/21.8 cm
Depth § 5.0 in/12.7 cm 5.0 in/12.7 cm
Weight §§ 6.6 lbs/3.0 kg 8.4 lbs/3.8 kg
Power Cord 10 ft/3 m 10 ft/3 m
§ without pole clamp
§§ without power cord
1-4 Signature Edition® GOLD Infusion System, Models 7130/7131 and 7230/7231
Technical Service Manual
GENERAL INFORMATION
Maximum 2 75 2 7
1 mL/h
Typical 0.6 30 0.6 4
Maximum 1 25 1 3
25 mL/h
Typical 0.1 1 1.0 1
Memory: 4.06 and higher: Maintains infusion perimeters indefinitely until the
instrument is reprogrammed.
Signature Edition® GOLD Infusion System, Models 7130/7131 and 7230/7231 1-5
Technical Service Manual
GENERAL INFORMATION
Rate Accuracy: For rates greater than 1 mL/h, up to 999.9 mL/h: ±5%, 95% of the time
with 95% confidence, under the conditions listed below.
For rates equal to or less than 1 mL/h: ±6.5%, 95% of the time with 95%
confidence, under the conditions listed below.
Infusion rate range: Head height: 24 ± 1 in.
0.1 to 999.9 mL/h (61 ± 2.5 cm)
Environment temperature: Back pressure: 0 psi.
68±8°F (20±4°C)
Test solution: distilled water Set Model: 72003
Needle: 18 gauge Minimum collection volume: 6 mL
CAUTION
Variations of head height, back pressure, time, monitoring mode option,
instrument tilt, or any combination of these, may affect rate accuracy.
Factors that can influence head height and back pressure are: IV set
configuration, IV solution viscosity and IV solution temperature. Back
pressure may also be affected by type of catheter.
NOTE: The Signature Edition® GOLD Infusion System has been assessed and complies with the following
Technical Standards: IEC 60601-1 / BS 5724, including amendments A1 and A2; IEC 60601-2-24; CISPR
11, Group 1, Class B Emissions; IEC 60601-1-2.
1-6 Signature Edition® GOLD Infusion System, Models 7130/7131 and 7230/7231
Technical Service Manual
GENERAL INFORMATION
Signature Edition® GOLD Infusion System, Models 7130/7131 and 7230/7231 1-7
Technical Service Manual
GENERAL INFORMATION
1.6 BATTERY MANAGEMENT SYSTEM The charger will suspend top-up if the
(Continued) battery temperature exceeds 37°C. The
time spent to cool down to below 37°C is
1.6.2 Battery and Charging Process in addition to the 180 minutes top-up
(Continued) charge time. If top-up cool down time
The battery charge circuit charges the exceeds 5 1/2 hours, the instrument will
battery with a constant current of 1 ampere go into float charge mode.
whenever the Battery Manager turns the • Float Charge: The float charge phase
charger on. The Battery Manager regulates begins at the end of the top-up phase
average charge current by turning the and helps maintain a fully charged
charger on and off with the appropriate duty battery. This phase charges at an
ratio. The battery charge cycle consists of average rate of 40 mA (1A for 0.2
four modes; fast charge, top-up charge, float seconds every 5 seconds). The fan
charge, and hot charge. remains on or turns on when battery
• Fast Charge: Fast charge is initiated temperature exceeds 22°C.
whenever the battery is less than 36°C, • Hot Charge: The hot charge mode occurs
and has been discharged by more than when the instrument determines that the
200 Ampere-seconds through actual use battery is >36°C (normally due to
or self discharge. Leaving the instrument ambient temperature being >27°C) to
unplugged for a day would cause about allow a charge after waiting 3 hours for it
200 Ampere-seconds of self discharge. to cool down. Hot charge mode charges
The charge current is a continuous at an average rate of 180 mA (1A for 0.9
1 Ampere. The end of a fast charge is seconds every 5 seconds) for a total
detected when the temperature of the charge time of 18 hours. If the battery
battery rises 7°C above its temperature temperature exceeds 43°C, the charging
at start of charge and is at least 30°C, or is turned off until the temperature falls
when the battery voltage declines by below 43°C.
192mV below its peak value, or total
Note that the cool down time is in
charge time exceeds 3.2 hours. Refer to
addition to the 18-hour charge time. The
“Battery Charge Regulator” section in the
float charge cycle begins at the end of
“Principles of Operation” chapter for
the hot charge cycle.
further details.
• Top-Up Charge: The top-up charge 1.6.3 Refresh Cycle
phase begins at the end of the fast
A battery refresh cycle performs a full
charge phase and finishes adding the
charge, discharge, and recharge to
last few percent of charge to the battery
condition and measure the capacity of the
and balances individual cell charges.
battery. This refresh results in a new
This phase charges at an average rate of
"Measured Capacity" in the battery
180 mA (1 A for 0.9 seconds every 5
diagnostics and can be used to judge the
seconds) for 180 minutes. At that time,
condition of the battery.
the instrument will go into float charge
mode.
1-8 Signature Edition® GOLD Infusion System, Models 7130/7131 and 7230/7231
Technical Service Manual
GENERAL INFORMATION
Top-Up* Top-Up*
Fast Fast
Cool* Cool*
Discharge Float
11.45 Volts
Signature Edition® GOLD Infusion System, Models 7130/7131 and 7230/7231 1-9
Technical Service Manual
GENERAL INFORMATION
1-10 Signature Edition® GOLD Infusion System, Models 7130/7131 and 7230/7231
Technical Service Manual
GENERAL INFORMATION
GOLD MNTR
disposed of properly, according to local
regulations.
The Battery Manager provides a "relative • A special circuit removes all load from
time" clock which the main processor can battery when voltage falls too low,
set and read. This clock consists of a 32-bit preventing damage from over discharge
counter which is incremented once a due to long-term storage.
second under all conditions. The main NiCad batteries can be stored indefinitely
processor uses this counter as a means of with no load but will self-discharge from a
determining elapsed time even when power charged state in about 100 days. This does
has been turned off. The clock is used to not damage the battery as it would if it were
compensate for normal battery capacity a lead acid-type battery. Connect
degradation over time. instrument to AC to recharge batteries.
Signature Edition® GOLD Infusion System, Models 7130/7131 and 7230/7231 1-11
Technical Service Manual
GENERAL INFORMATION
1.6 BATTERY MANAGEMENT SYSTEM An ideal charge cycle starts with a fully
(Continued) discharged battery charged at C/10 (C is the
rated capacity in Ah) constant current for 15
1.6.8 Battery Maintenance (Continued) hours while at room temperature. For
instance, a 1.8Ah battery would be charged
NOTE: If under load (inside instrument), the for 15 hours at 180mA constant current with
maximum storage time would be about 130 a room temperature of 23°C.
days before electrolyte extrusion occurs past
the battery seal. The ideal discharge starts with a fully
charged battery under a C/5 constant
If the battery exhibits short run times, a
current load at room temperature,
reconditioning procedure can be used:
discharging to a cell voltage of 0.9V. The
1. Disconnect battery and AC rated capacity is then calculated as the time
2. Press On/Off switch and verify lower to discharge divided by 5. Again, a 1.8Ah
LCD goes blank. cell would be discharged at 360mA constant
current and not reach 0.9V for at least 5
3. Reconnect battery. hours. Note that a given battery type has
4. Reconnect instrument to AC. A refresh different capacities based on the load. For
cycle will be initiated. instance, a battery rated at 1.8Ah at a
360mA load may have only 1.6Ah at a
5. Repeat this procedure at 24-hour 1600mA load.
intervals, 2 or 3 times, to increase
capacity or use a battery conditioner. As can be seen from the preceding ideal
conditions, there are many conditions which
1.7 NICAD BATTERY CAPACITY can affect the battery capacity. The
INFORMATION following conditions have the most practical
impact on battery capacity delivered in this
All batteries have specific conditions under instrument.
which they are guaranteed to meet their
published specifications. Deviations from • Temperature During Charge:
these conditions typically result in a As the effective ambient temperature of
reduction of available capacity. the battery increases, the amount of
Manufacturers of NiCad batteries rate charge that the battery will accept is
capacities, usually expressed in Ah, based decreased. At an ambient temperature of
on a specified "ideal" charge and discharge 35°C, an enclosed battery will
condition, as well as the use of a "new" temporarily accept only about 90% of the
battery. Battery manufacturers do not use charge it would otherwise accept at
the calendar year for battery date codes. 23°C. Because the batteries are internal
The battery date code year starts on to the instrument case and the instrument
September 1. The first two digits of the date itself generates heat, the batteries will be
code are for the year and the last two digits exposed to temperatures above room
are for the week of the year (for example, temperature. Some of the ways the
date code 0411 = the year 2004 and the instrument limits the temperatures that
eleventh week from September 1). the battery sees include forcing air
across it (an internal fan) and turning off
the charger when the battery temperature
gets too high.
1-12 Signature Edition® GOLD Infusion System, Models 7130/7131 and 7230/7231
Technical Service Manual
GENERAL INFORMATION
1.7 NICAD BATTERY CAPACITY in the first 2 hours after fast charge. The
INFORMATION (Continued) next charge phase, top-up, is designed to
finish the charge and to bring all
• Cycle Life and Aging:
individual cells to the fully charged state,
As batteries get older and go through
essentially rematching them. Refer to
many charge/discharge cycles, batteries
“Battery and Charging Process” section
"wear out," in that the chemicals and
in this chapter for “Fast Charge” and
materials used to construct the cell break
“Top-up Charge” information. If the
down. The way the instrument deals with
top–up charge is not completed, then the
this is to assume that a battery will
cell mismatch is not reduced and the
continually reduce capacity at a rate
cumulative capacity reduction occurs.
equivalent to 30% over 4 years, and
Top-up is a 3-hour charge, but the
continually reduce capacity at a rate
elapsed time to complete it may be over
equivalent to 30% per 200 full charge
5 hours, as the charger is turned on and
cycles. These calculated values are
off to keep the battery cool during that
used to reduce the displayed runtime on
time.
the battery gauge.
• Battery Alarm Voltage:
• Partial Discharge/Charge:
The battery alarm voltage is the voltage
When a battery is partially discharged,
at which the instrument stops operating
then charged for less than the full time,
and generates an alarm indicating the
differences between individual cell
instrument needs to be connected to AC
capacities result in cells completing
power. As noted in the ideal discharge
charge at different times. If the full
condition, the end of discharge is
charge sequence is not then completed,
determined by 0.9V/cell.
the cell "mismatch" becomes
progressively greater. This is viewed by Under perfect conditions, a battery of 10
the user as low apparent runtimes and cells connected in series would reach the
premature low battery warning and end of discharge at 9.0V; however, cells
alarms. are not perfectly matched so some will
reach 0.9V before others. The problem
The problem is cumulative in that the
occurs when a cell in series with other
mismatch increases for every partial
cells can go below 0.9V and actually can
cycle. The lowered capacity is not
go into cell reversal, which permanently
permanent, but may require 2-3 full
damages the particular cell. On the other
charge cycles to recover. The way the
hand, increasing the alarm voltage to
instrument deals with this is to reduce the
compensate for imperfectly matched cells
displayed runtime based on a limited
results in reduced runtimes with available
history of partial cycles.
capacity.
• Charge Rate:
The ideal charge rate requires 15 hours
to get to full charge, which is undesirable
from the user's perspective. The
instrument provides a multiphase charge
cycle which results in about 80% capacity
Signature Edition® GOLD Infusion System, Models 7130/7131 and 7230/7231 1-13
Technical Service Manual
GENERAL INFORMATION
1.7 NICAD BATTERY CAPACITY processing of the pressure and flow data
INFORMATION (Continued) can then produce the fluid impedance value.
Such measurements can be made
The user sees this as premature low
continuously at short intervals and be
battery warnings and alarms. The way
independent of the selected rate. Head
the instrument deals with this is to
height and resulting pressure variations,
increase the alarm voltage to guarantee
likewise, will not affect the measurement.
the battery is not damaged and to reduce
the assumed capacity to below that When a complete occlusion occurs, the
printed on the battery. The battery gauge resistive part of the fluid impedance is very
is intended to show the minimum run large (theoretically infinite). Elevated
time left on the battery taking all these resistances due to clotting, clogged filters,
factors into account. partial occlusions or infiltrations can be
measured.
NOTE: In the event ALARIS Medical Systems
provides different battery packs in the future, The Dynamic Monitoring® System provides
the replacement battery may have a different a means to measure the mechanical
rated capacity. Therefore, the Battery properties of the downstream flow path.
Manager of the instrument needs to know that Figure 1-3. Resistance Graph
a new battery has been added and the rated
capacity has changed. Refer to the
“Removing Battery” section of “Corrective
PRESSURE (result)
1-14 Signature Edition® GOLD Infusion System, Models 7130/7131 and 7230/7231
Technical Service Manual
GENERAL INFORMATION
1.8 DYNAMIC MONITORING® SYSTEM Figure 1-4. Pressure and Resistance Graph
(DMS) (Continued)
600
PRESSURE mmHg
Features include:
• Alarm setting is based on dynamic 300
system impedance (pressure changes,
IN PRESSURE
not absolute pressure).
0 TIME
• Detection of complete occlusions.
OCCLUSION
• Reduced nuisance alarms by minimizing
artifact effects such as head height and
patient movement. 100%
mmHg/Liter/Hr
FLUID OHMS
• AutoRestartPlus feature allows 75%
IN PRESSURE
instrument to automatically continue 50%
operation if an occlusion is cleared within
self-check period (40 seconds). A
warning tone and “Checking Line” 0 TIME
message will occur for up to 40 seconds.
OCCLUSION
The feature can be turned off (set
restarts to zero), or number of restarts Pressure rise is quite small with site
may be set from 1 to 9. complication, especially with low flow rates.
Signature Edition® GOLD Infusion System, Models 7130/7131 and 7230/7231 1-15
Technical Service Manual
GENERAL INFORMATION
1-16 Signature Edition® GOLD Infusion System, Models 7130/7131 and 7230/7231
Technical Service Manual
GENERAL INFORMATION
Various abbreviations, acronyms and symbols are used throughout this manual. The following are those that are
not commonly known or easily recognized.
T H I S PA G E
I N T E N T I O N A L LY
LEFT BLANK
CAUTION
Keep latch closed when instrument is not in use.
2.1 INTRODUCTION
This chapter describes the initial setup and
configuration of an instrument with software
versions 4.06 and 4.08.
CAUTION
2.2 NEW INSTRUMENT CHECKOUT
Should an instrument be jarred severely or
dropped, remove it from use immediately. It Refer to the instrument's Directions For Use
should be thoroughly tested and inspected by (DFU) for instructions regarding unpacking
qualified service personnel to ensure proper and setting up the instrument for the first
function prior to reuse. time.
When turning on the instrument, verify the
instrument beeps and that all display LED
segments flash. This confirms that the
instrument has performed its self test and is
operating correctly. During operation, the
instrument continually performs a self test,
and will alarm and display a message if it
detects an internal malfunction.
Contact ALARIS Medical Systems
authorized service personnel if the
instrument has physical damage, fails to
satisfactorily pass startup sequence, fails a
self test or continues to alarm.
For new instrument checkout refer to the
applicable Signature Edition® GOLD
Infusion System DFU.
Signature Edition® GOLD Infusion System, Models 7130/7131 and 7230/7231 2A-1
Technical Service Manual
CHECKOUT AND CONFIGURATION
Software Versions 4.06 and 4.08
2A-2 Signature Edition® GOLD Infusion System, Models 7130/7131 and 7230/7231
Technical Service Manual
CHECKOUT AND CONFIGURATION
Software Versions 4.06 and 4.08
A B OPTIONS
Signature Edition® GOLD Infusion System, Models 7130/7131 and 7230/7231 2A-3
Technical Service Manual
CHECKOUT AND CONFIGURATION
Software Versions 4.06 and 4.08
2A-4 Signature Edition® GOLD Infusion System, Models 7130/7131 and 7230/7231
Technical Service Manual
CHECKOUT AND CONFIGURATION
Software Versions 4.06 and 4.08
(Continued)
Set To Defaults C
1. From C2 page, press Air In Line soft Regional Settings 2
key. Air In Line
Profiles
A B OPTIONS
page
Set To Defaults C
Regional Settings 2
Air In Line
Profiles 1. From C2 page, press Profiles soft key.
page
A B OPTIONS
Profiles C
Off 2
2. Press Threshold soft key to select for
editing. Press again to cycle through 50
mcL, 100 mcL, 200 mcL and 500 mcL.
undo ok
NOTE: Use a 50 microliters setting on
microbore tubing. The other three settings
may be used on macrobore tubing.
2. Press Off or On soft key to select for
A B OPTIONS
editing. Press again to cycle between
Air In Line C On and Off.
Threshold= 100 mcL 2 NOTES:
Reset= Off
Accumulator= On • The data set for Guardrails® Safety
Software must be loaded to set Profiles
undo ok to On. For the Guardrails® Safety
Software to function, Profiles and Dose
Rate must be On.
3. Press Reset soft key to select for editing. • When Profiles is turned on, the
Press again to cycle between On and configuration mode will have limited
Off. access. Access will be limited to
Regional Settings, Profiles, Computer
4. Press Accumulator soft key to select for Link and Optional Features, due to the
editing. Press again to cycle between data set overriding the configuration
On and Off. settings as part of the Guardrails®
5. Press ok to accept change and return to Safety Software.
beginning of C2 page.
Signature Edition® GOLD Infusion System, Models 7130/7131 and 7230/7231 2A-5
Technical Service Manual
CHECKOUT AND CONFIGURATION
Software Versions 4.06 and 4.08
NOTES:
• The maximum rate setting applies to all 5. Press ok to accept change and return to
infusion modes. beginning of C3 page.
• Setting the Maximum Rate below the
2.4.7 Setting Computer Link
preset KVO Rate will lower the KVO Rate.
The KVO Rate will not exceed the The Computer Link feature allows a
Maximum Rate. hospital/facility computer to interact with the
instrument and programs the level of
1. Advance to C3 page. computer control available. If the feature is
off, the computer cannot communicate with
the instrument.
A B OPTIONS
Monitor Mode allows computer to only
Maximum Rate C receive information from instrument.
Computer Link 3 Enabling of monitoring mode automatically
Optional Modes places it in options menu.
Optional Features
Off Mode does not allow any
page communication between instrument and a
computer.
A B OPTIONS
2A-6 Signature Edition® GOLD Infusion System, Models 7130/7131 and 7230/7231
Technical Service Manual
CHECKOUT AND CONFIGURATION
Software Versions 4.06 and 4.08
A B OPTIONS NOTES:
• Dose Rate must be on in order to use
Mode = Mntr Off C Guardrails® Safety Software drug selection
Baud Rate = 9600 3 capability.
Parity = Even
• Instruments configured in European
Regional Settings (with profiles OFF) do
undo ok not have a drug list. When Dose Rate
Calculator is chosen (from OPTIONS
menu), followed by Enter New Program,
the instrument defaults to generic drug
2. Press Mode soft key to select for editing. dose rate calculation.
Press again to cycle through Off and
Mntr.
A B OPTIONS
Signature Edition® GOLD Infusion System, Models 7130/7131 and 7230/7231 2A-7
Technical Service Manual
CHECKOUT AND CONFIGURATION
Software Versions 4.06 and 4.08
3. Press Dose Rate soft key to select for When VTBI (volume to be infused) is on, a
editing. Press again to cycle between On volume to be infused must be entered;
and Off. otherwise, the last remaining VTBI or last
entered VTBI, depending on the channel's
NOTE: This feature allows dose last usage, will be in effect. When VTBI is
parameters to be programmed so that the off, there is no VTBI line capability.
instrument then calculates the volumetric
When the Multi-Dose Alert feature is on, the
rate.
option is available to set an alert at the end
4. Press Multi-Step soft key to select for of every dose when in the Multi-Dose mode.
editing. Press again to cycle between On
and Off.
A B OPTIONS
A B OPTIONS A B OPTIONS
6. Press ok to accept changes and return 2. Press Panel Lock soft key to select for
to beginning of C3 page. editing. Press again to cycle between
On and Off.
2A-8 Signature Edition® GOLD Infusion System, Models 7130/7131 and 7230/7231
Technical Service Manual
CHECKOUT AND CONFIGURATION
Software Versions 4.06 and 4.08
KVO Rate = C
2.4.9 Setting Optional Features (Continued) 5.0 mL/hr 4
3. Verify VTBI is On.
4. Press Multi-Dose Alert soft key to select 3. Press soft key next to rate value once to
for editing. Press again to cycle between select for editing.
On and Off.
5. Press ok to accept changes and return to 1 2 3
beginning of C3 page. 4 5 6
7 8 9
2.4.10 Setting KVO Rate
. 0 Clear Enter
Signature Edition® GOLD Infusion System, Models 7130/7131 and 7230/7231 2A-9
Technical Service Manual
CHECKOUT AND CONFIGURATION
Software Versions 4.06 and 4.08
The Resistance Mode provides two ranges The Checking Line message and tone are
and user-alert limit adjustment, to meet presented whenever a resistance or
neonatal to adult patient needs. pressure measurement exceeds its alarm
threshold.
Many monitoring features and functions are
configurable, by qualified technical • In resistance monitoring modes, the
personnel, to meet specific needs by Checking Line period is caused by a
allowing customization. measured resistance of 100% or a
pressure level exceeding a configured
Monitoring Mode Options threshold. This pressure threshold is
IV lines, catheters, and applications create separate from the Pressure mode
various levels of resistance to flow. threshold.
Monitoring mode options are available to
meet each clinical application, as follows:
2A-10 Signature Edition® GOLD Infusion System, Models 7130/7131 and 7230/7231
Technical Service Manual
CHECKOUT AND CONFIGURATION
Software Versions 4.06 and 4.08
Signature Edition® GOLD Infusion System, Models 7130/7131 and 7230/7231 2A-11
Technical Service Manual
CHECKOUT AND CONFIGURATION
Software Versions 4.06 and 4.08
2A-12 Signature Edition® GOLD Infusion System, Models 7130/7131 and 7230/7231
Technical Service Manual
CHECKOUT AND CONFIGURATION
Software Versions 4.06 and 4.08
KVO Rate C
Undo ok Monitoring Options 4
Audio
Configuration Name
Signature Edition® GOLD Infusion System, Models 7130/7131 and 7230/7231 2A-13
Technical Service Manual
CHECKOUT AND CONFIGURATION
Software Versions 4.06 and 4.08
NOTES:
• If the configuration has been changed
but not the name, the instrument will 1. Advance to C5 page.
display the option to rename before 2. Press Resistance Options soft key.
turning off or proceeding to the Teach
Mode. Refer to “Pop-Up Displays”
section for further explanation.
A B OPTIONS
2A-14 Signature Edition® GOLD Infusion System, Models 7130/7131 and 7230/7231
Technical Service Manual
CHECKOUT AND CONFIGURATION
Software Versions 4.06 and 4.08
Signature Edition® GOLD Infusion System, Models 7130/7131 and 7230/7231 2A-15
Technical Service Manual
CHECKOUT AND CONFIGURATION
Software Versions 4.06 and 4.08
2A-16 Signature Edition® GOLD Infusion System, Models 7130/7131 and 7230/7231
Technical Service Manual
CHECKOUT AND CONFIGURATION
Software Versions 4.06 and 4.08
1. Advance to C5 page
2. Press Manual Baseline soft key to select
for editing.
A B OPTIONS
Manual Baseline= C
C
On 5
undo ok
Signature Edition® GOLD Infusion System, Models 7130/7131 and 7230/7231 2A-17
Technical Service Manual
CHECKOUT AND CONFIGURATION
Software Versions 4.06 and 4.08
Teach Status C
Teach Status C
6
6 Learn/Teach Failed
Awaiting Start Software Mismatch
2D15 GOLD
cancel repeat
cancel start
2A-18 Signature Edition® GOLD Infusion System, Models 7130/7131 and 7230/7231
Technical Service Manual
CHECKOUT AND CONFIGURATION
Software Versions 4.06 and 4.08
A B OPTIONS
Configuration has C
been changed
Press POWER key
to keep name GOLD
cancel rename
A B OPTIONS
Configuration has C
been changed
Press teach key
to keep name GOLD
cancel teach rename
Signature Edition® GOLD Infusion System, Models 7130/7131 and 7230/7231 2A-19
Technical Service Manual
CHECKOUT AND CONFIGURATION
Software Versions 4.06 and 4.08
#1 #2 #3 #4 #5
§ Instrument ID/Serial #
Config Name
(Instr Label)
Location
Regional Settings
Air-In-Line:
Threshold µl µl µl µl µl
Reset
Accumulator
Profiles
Computer Link:
Mode
Baud Rate
Parity
Optional Modes:
Loading Dose
Dose Rate
Multi-Step
Multi-Dose
Optional Features
Panel Lock
VTBI
Multi-Dose Alert
2A-20 Signature Edition® GOLD Infusion System, Models 7130/7131 and 7230/7231
Technical Service Manual
CHECKOUT AND CONFIGURATION
Software Versions 4.06 and 4.08
#1 #2 #3 #4 #5
Monitoring Options:
Mode
Restarts
Trends
Audio:
Volume
Trans. Tone
Resistance Options:
Resist Display
Resist Alarm
Default Alarm % % % % %
Alarm mmHg mmHg mmHg mmHg mmHg
Pressure Options:
Pressure Display
Pressure Alarm
Maximum Pressure mmHg mmHg mmHg mmHg mmHg
Default Alarm mmHg mmHg mmHg mmHg mmHg
Manual Baseline
§Instrument ID/Serial # is accessed through Diagnostic Mode. Refer to the “Entering Diagnostic
Mode” section in the “Troubleshooting” chapter.
Signature Edition® GOLD Infusion System, Models 7130/7131 and 7230/7231 2A-21
Technical Service Manual
CHECKOUT AND CONFIGURATION
Software Versions 4.06 and 4.08
A B OPTIONS A B OPTIONS
page undo ok
A B OPTIONS A B OPTIONS
A B OPTIONS
A B OPTIONS
A B OPTIONS A B OPTIONS
undo ok ok
2A-22 Signature Edition® GOLD Infusion System, Models 7130/7131 and 7230/7231
Technical Service Manual
CHECKOUT AND CONFIGURATION
Software Versions 4.06 and 4.08
A B OPTIONS
A B OPTIONS
undo ok undo ok
A B OPTIONS A B OPTIONS
Loading Dose On C C
Dose Rate= On 3 Mode= Pressure 4
Multi-Step= Off Restarts= 3
Multi-Dose= Off Trends= On
undo ok Undo ok
A B OPTIONS
A B OPTIONS
undo ok Undo ok
A B OPTIONS A B OPTIONS
Signature Edition® GOLD Infusion System, Models 7130/7131 and 7230/7231 2A-23
Technical Service Manual
CHECKOUT AND CONFIGURATION
Software Versions 4.06 and 4.08
A B OPTIONS A B OPTIONS
page page
A B OPTIONS A B OPTIONS
A B OPTIONS A B OPTIONS
A B OPTIONS
Manual Baseline= C
C
On 5
undo ok
2A-24 Signature Edition® GOLD Infusion System, Models 7130/7131 and 7230/7231
Technical Service Manual
2B CHECKOUT & CONFIGURATION
Chapter 2B — CHECKOUT AND CONFIGURATION
Software Version 2.78
Signature Edition® GOLD Infusion System, Models 7130/7131 and 7230/7231 2B-1
Technical Service Manual
CHECKOUT AND CONFIGURATION
Software Version 2.78
WARNINGS
• When an instrument’s configuration is ml/hr
changed, the configuration name OPT PRI HLD SEC KVO
2B-2 Signature Edition® GOLD Infusion System, Models 7130/7131 and 7230/7231
Technical Service Manual
CHECKOUT AND CONFIGURATION
Software Version 2.78
A B OPTIONS
Set All C
CONFIGURATION MODE C configuration 2
ID No. : 001234567 1 items to their
SW Rev.: 02.78 default values
Config.: BFFF IVAC cancel ok
page
Signature Edition® GOLD Infusion System, Models 7130/7131 and 7230/7231 2B-3
Technical Service Manual
CHECKOUT AND CONFIGURATION
Software Version 2.78
Set To Defaults C
Language 2
undo ok Air In Line
Dose Rate Drugs
page
2. Press soft key to select for editing. Press
again to cycle through language choices.
1. From C2 page, press Air In Line soft
NOTE: Dutch and German are choices in key.
the language options, but labels and DFUs
are not available in these languages for 2. Press Threshold soft key to select for
this software version. editing. Press again to cycle through 50
3. Press ok to accept change and return to mcL, 100 mcL, 200 mcL and 500 mcL.
beginning of C2 page.
2B-4 Signature Edition® GOLD Infusion System, Models 7130/7131 and 7230/7231
Technical Service Manual
CHECKOUT AND CONFIGURATION
Software Version 2.78
A B OPTIONS
Signature Edition® GOLD Infusion System, Models 7130/7131 and 7230/7231 2B-5
Technical Service Manual
CHECKOUT AND CONFIGURATION
Software Version 2.78
2B-6 Signature Edition® GOLD Infusion System, Models 7130/7131 and 7230/7231
Technical Service Manual
CHECKOUT AND CONFIGURATION
Software Version 2.78
Maximum Rate C
Computer Link 3
5. Press Drug? soft key to cycle between Optional Modes
“s”, “e” and “blank”. Use “s” for Optional Features
commonly used drugs.
page
6. Press a soft key next to a drug to cycle
between “s”, “e” and “blank”. Press
Page to continue viewing drug list.
Refer to Table 2-2 “Drug List” to view full 2. Press Maximum Rate soft key.
list of drugs.
A B OPTIONS
Summary C
14 drugs selected 2
EXT D LIST selected ok
Drug? Selected
review ok
3. Press the soft key to select the value for
editing.
Signature Edition® GOLD Infusion System, Models 7130/7131 and 7230/7231 2B-7
Technical Service Manual
CHECKOUT AND CONFIGURATION
Software Version 2.78
Maximum Rate C
2.4.6 Setting Maximum Rate (Continued) Computer Link 3
4. Use numeric keypad to enter maximum Optional Modes
rate. Press Enter. Optional Features
page
1 2 3
4 5 6
7 8 9
. 0 Clear Enter
A B OPTIONS
5. Press ok to accept change and return to
beginning of C3 page. Mode = Off C
Baud Rate = 9600 3
2.4.7 Setting Computer Link Parity = None
The Computer Link feature allows a
hospital/facility computer to interact with the undo ok
instrument and programs the level of
computer control available. The computer
cannot start or stop the instrument, set the
rate, or make any change in status. If the 1. On C3 page, press Computer Link soft
feature is off, the computer cannot key.
communicate with the instrument. 2. Press Mode soft key to select for editing.
Control Mode allows computer to send Press again to cycle through Off, Mntr
information to instrument’s display. Off and Ctrl Mntr Off.
Monitor Mode allows computer to only 3. Press Baud Rate soft key to select for
receive information from instrument. editing. Press again to cycle through
Enabling of monitoring mode automatically choices (300, 600, 1200, 1800, 2400,
places it in options menu. 4800 and 9600).
Off Mode does not allow any 4. Press Parity soft key to select for editing.
communication between instrument and a Press again to cycle through Even, Odd,
computer. and None.
5. Press ok to accept changes and return to
beginning of C3 page.
2B-8 Signature Edition® GOLD Infusion System, Models 7130/7131 and 7230/7231
Technical Service Manual
CHECKOUT AND CONFIGURATION
Software Version 2.78
A B OPTIONS A B OPTIONS
undo ok undo ok
Signature Edition® GOLD Infusion System, Models 7130/7131 and 7230/7231 2B-9
Technical Service Manual
CHECKOUT AND CONFIGURATION
Software Version 2.78
undo ok
undo ok
2B-10 Signature Edition® GOLD Infusion System, Models 7130/7131 and 7230/7231
Technical Service Manual
CHECKOUT AND CONFIGURATION
Software Version 2.78
Signature Edition® GOLD Infusion System, Models 7130/7131 and 7230/7231 2B-11
Technical Service Manual
CHECKOUT AND CONFIGURATION
Software Version 2.78
2B-12 Signature Edition® GOLD Infusion System, Models 7130/7131 and 7230/7231
Technical Service Manual
CHECKOUT AND CONFIGURATION
Software Version 2.78
2.4 CONFIGURATION PROCEDURE 2. Press Mode soft key to select for editing.
(Continued) Press again to cycle between Hi Resist,
Resist, and Pressure.
2.4.11 Setting Dynamic Monitoring®
3. Press Restarts soft key. Use numeric
System Features (Continued)
keypad to enter number of restarts, from
Trend Graphs "0" (which turns feature off) to "9", then
press Enter.
Trend graphs display downstream pressure
or flow resistance over time (up to 12 4. Press Trends soft key to select for
hours). editing. Press again to cycle between
On and Off.
In Pressure, Resistance, and High
Resistance monitoring modes, the trend • In Pressure mode, graph displays in
graph displays the flow pattern over time. mmHg.
Downstream occlusions, which occur in the
• Trend data is lost when:
Pressure or Resistance modes, will be
indicated by a vertical tick mark at the top of ♦ graph information is cleared
the trend screen. ♦ instrument is off for more than
6 hours
Trend graphs of fifteen minutes, one hour,
four hours, and twelve hours are available • Trend data is maintained when:
during normal operation. ♦ rate is changed
♦ instrument is off for less than
A B OPTIONS
6 hours
KVO Rate C • Trend data is scaled (by 3 times or by
Monitoring Options 4 1/3, as applicable) when cycling
Audio between Resistance and Hi
Configuration Name Resistance modes.
page 5. Press ok to accept changes and return to
beginning of C4 page.
6. Precision Flow: Selecting Pressure
1. On C4 page, press Monitoring Options Mode and running below 50 mL/h will
soft key. place instrument into Precision Flow
Mode.
A B OPTIONS
Signature Edition® GOLD Infusion System, Models 7130/7131 and 7230/7231 2B-13
Technical Service Manual
CHECKOUT AND CONFIGURATION
Software Version 2.78
2B-14 Signature Edition® GOLD Infusion System, Models 7130/7131 and 7230/7231
Technical Service Manual
CHECKOUT AND CONFIGURATION
Software Version 2.78
Signature Edition® GOLD Infusion System, Models 7130/7131 and 7230/7231 2B-15
Technical Service Manual
CHECKOUT AND CONFIGURATION
Software Version 2.78
Configuration has C
Teach Status C been changed.
5 Press POWER key
Awaiting Start
to keep name IVAC.
BFFF IVAC
cancel rename
cancel start
2B-16 Signature Edition® GOLD Infusion System, Models 7130/7131 and 7230/7231
Technical Service Manual
CHECKOUT AND CONFIGURATION
Software Version 2.78
Signature Edition® GOLD Infusion System, Models 7130/7131 and 7230/7231 2B-17
Technical Service Manual
CHECKOUT AND CONFIGURATION
Software Version 2.78
2.7 PRESSURE OPTIONS the first time RUN/HOLD is pressed; this will
be the maximum baseline. Subsequent
Pressure Display: Provides a bar graph
presses of RUN/HOLD will only lower (not
and indication of pressure in the line while
raise) the baseline.
monitoring a single channel.
Manual Baseline: Sets a fixed baseline
Pressure Alarm: Allows the choice of a
and overrides the Auto Zero level until the
fixed pressure mode or an adjustable
instrument is turned off, the latch is opened,
pressure mode.
the set is reloaded, or the pressure baseline
• Adjustable Pressure: Provides an early function is performed again. This allows the
warning of increases in downstream instrument to display the actual pressure
pressure. Designed to monitor IV required for an IV to infuse. To get this real
line/site based on pressure, and provide time readout, Manual Baseline must be
user- adjustable pressure alarm limits. On.
Used for the smallest of catheter sizes
To activate this mode with the instrument on
and special applications; such as infusion
hold, or at startup, press the Options key
through transducers, into dialysis
and then select Set Pressure Baseline and
systems, and for the highest resistance
press ok.
catheters (including some neonatal PICC
catheters). When RUN/HOLD is pressed, the bar graph
• Fixed Pressure: Designed to monitor IV will show the alarm point and the actual
line/site pressure and alarm, based on a pressure in the line will be displayed below
fixed pressure limit of 600 mmHg. Used the bar graph.
when adjustable pressure is not wanted.
Used for the smallest of catheter sizes A B OPTIONS
2B-18 Signature Edition® GOLD Infusion System, Models 7130/7131 and 7230/7231
Technical Service Manual
CHECKOUT AND CONFIGURATION
Software Version 2.78
Signature Edition® GOLD Infusion System, Models 7130/7131 and 7230/7231 2B-19
Technical Service Manual
CHECKOUT AND CONFIGURATION
Software Version 2.78
#1 #2 #3 #4 #5
§ Instrument ID/Serial #
Config Name
(Instr Label)
Location
Language
Air-In-Line:
Threshold µl µl µl µl µl
Reset
Accumulator
Computer Link:
Mode
Baud Rate
Parity
Optional Modes:
Loading Dose
Dose Rate
Multi-Step
Multi-Dose
Optional Features
Panel Lock
VTBI
Multi-Dose Alert
2B-20 Signature Edition® GOLD Infusion System, Models 7130/7131 and 7230/7231
Technical Service Manual
CHECKOUT AND CONFIGURATION
Software Version 2.78
#1 #2 #3 #4 #5
Monitoring Options:
Mode
Restarts
Trends
Audio:
Volume
Trans. Tone
Resistance Options:
Resist Display
Resist Alarm
Default Alarm % % % % %
Alarm mmHg mmHg mmHg mmHg mmHg
Pressure Options:
Pressure Display
Pressure Alarm
Alarm mmHg mmHg mmHg mmHg mmHg
Manual Baseline
§Instrument ID/Serial # is accessed through Diagnostic Mode. Refer to the “Entering Diagnostic
Mode” section in the “Troubleshooting” chapter.
Signature Edition® GOLD Infusion System, Models 7130/7131 and 7230/7231 2B-21
Technical Service Manual
CHECKOUT AND CONFIGURATION
Software Version 2.78
DRUG NAME/OPTION TRADE NAME NOT SELECTED SHORT LIST EXTENDED LIST
Extended List Access
DRUG?
Alfentanil Alfenta
Alprostadil Prostin (PGE-1)
Alteplase Activase
AminoPhylline
Amrinone Inocar
Atracurium Tracurium
Bretylium Bretylol
Cimetidine Tagamet
Diltiazem Cardizem
Dobutamine Dobutrex
Dopamine Intropin
Esmolol Brevibloc
Heparin
Isoproterenol Isuprel
Labetalol Normodyne /Trandate
Lidocaine Xylocaine
Magnesium Sulfate
Methylprednisolone
Milrinone
Morphine
Nytroglycerin Tridil
Nitroprusside Nipride
Norepinephrine Levophed
Ocytocin
Phenlyphrine Neo-Synephrine
Potassium Chlor
Procainamide
Propofol Dipravan
Ranitidine Zantac
Streptokinase Streptase
Succinylcholine Anectine
Theophylline
Urokinase Abbokinase
Vecuronium Norcuron
2B-22 Signature Edition® GOLD Infusion System, Models 7130/7131 and 7230/7231
Technical Service Manual
CHECKOUT AND CONFIGURATION
Software Version 2.78
A B OPTIONS
A B OPTIONS
A B OPTIONS
A B OPTIONS
A B OPTIONS
A B OPTIONS
cancel ok page
A B OPTIONS
A B OPTIONS
undo ok ok
Signature Edition® GOLD Infusion System, Models 7130/7131 and 7230/7231 2B-23
Technical Service Manual
CHECKOUT AND CONFIGURATION
Software Version 2.78
A B OPTIONS
A B OPTIONS
Mode = Off C
Baud Rate = 9600 3 KVO Rate = C
Parity = None 5.0 mL/hr 4
undo ok
undo ok
A B OPTIONS
A B OPTIONS
Loading Dose On C
Dose Rate = On Monitoring Options C
3 Mode= Pressure
Multi-Step = Off
4
Restarts= 3
Multi-Dose = Off
Trends= On
undo ok
undo ok
A B OPTIONS
A B OPTIONS
Panel Lock= On C
VTBI= On 3 Audio C
Multi-Dose Alert=Off Volumes = Low Med Hi 4
Trans. Tones = On
undo ok
undo ok
A B OPTIONS
A B OPTIONS
KVO Rate C
Config. Name= IVAC C
Monitoring Options 4
Audio
ABCDEFGHIJKLM 4
NOPQRSTUVWXYZ
Configuration Name
0123456789_
page
undo enter ok
2B-24 Signature Edition® GOLD Infusion System, Models 7130/7131 and 7230/7231
Technical Service Manual
CHECKOUT AND CONFIGURATION
Software Version 2.78
A B OPTIONS
A B OPTIONS
A B OPTIONS
A B OPTIONS
A B OPTIONS
Teach Status C
5
Awaiting Start
BFFF IVAC
cancel start
Signature Edition® GOLD Infusion System, Models 7130/7131 and 7230/7231 2B-25
Technical Service Manual
CHECKOUT AND CONFIGURATION
Software Version 2.78
T H I S PA G E
I N T E N T I O N A L LY
LEFT BLANK
2B-26 Signature Edition® GOLD Infusion System, Models 7130/7131 and 7230/7231
Technical Service Manual
3 PREVENTIVE MAINTENANCE
Chapter 3 — PREVENTIVE MAINTENANCE
Signature Edition® GOLD Infusion System, Models 7130/7131 and 7230/7231 3-1
Technical Service Manual
PREVENTIVE MAINTENANCE
3.2 PREVENTIVE MAINTENANCE check that each switch performs its proper
INSPECTIONS (Continued) function. Refer to "Testing Switches" section
of the “Troubleshooting” chapter.
3.2.1 Regular Inspection (Continued)
Mechanism
Case
Clean any surfaces where solution or
Examine the instrument for overall obstructions have accumulated. Verify:
condition. The case should be clean and
free from IV solution residue, especially • Mechanism seal is not torn or worn.
near moving parts. Also check for dried • Cam followers are not broken or cracked
solution deposits on accessible areas of air- and are free of foreign matter.
in-line sensor, pressure transducer, and
latch mechanism. Check that labels and • Proper operation of latching mechanism.
markings are legible. Cam followers should retract and extend
smoothly.
Mounting Bracket
• Air-in-line arm moves smoothly from
Pole mounting bracket should be secure opened to closed position.
and functioning. If the instrument is
• Fluid Control Actuator rotates 180°.
mounted on a pole or stand, examine the
condition of the mount. Also, examine the
3.2.2 Functional Test
pole and stand.
1. Turn instrument on without set installed.
Power Cord Assembly Verify it "beeps" and red alarm light
Examine the power cord assembly for: flashes.
• Signs of damage, cuts or deformities in 2. Set infusion rate to 460 mL/h and VTBI to
the cord. If damaged, replace the entire 100 mL.
cord. 3. Press RUN/HOLD switch with latch
• Integrity of hospital grade power plug. closed, and rate and VTBI ≠ 0 to cause
Attempt to wiggle blades to ensure they "set out" and "air in line" messages.
are secure. If any damage is suspected, 4. Open latch.
replace entire cord.
5. Install primed administration set with
• Appropriate tension and connection. latch open.
Applicable only if IV pole has electrical
receptacles for accessories. 6. Verify instrument displays "air in line" and
"latch open" messages.
• Strain reliefs. Examine strain reliefs at
both ends of line cord. Be sure they hold 7. Close latch and verify display returns to
cord securely. setup page.
3-2 Signature Edition® GOLD Infusion System, Models 7130/7131 and 7230/7231
Technical Service Manual
PREVENTIVE MAINTENANCE
Signature Edition® GOLD Infusion System, Models 7130/7131 and 7230/7231 3-3
Technical Service Manual
PREVENTIVE MAINTENANCE
3-4 Signature Edition® GOLD Infusion System, Models 7130/7131 and 7230/7231
Technical Service Manual
PREVENTIVE MAINTENANCE
Burette
(50 mL)
Burette 30 + 2 inches
Clamp
Equipment
Stand
Tubing 80VCS
Three- Rate Cal Set
Way
Stopcock
Instrument Stand
Table or
Bench IV Tubing
Used Fluid
Receptacle
Signature Edition® GOLD Infusion System, Models 7130/7131 and 7230/7231 3-5
Technical Service Manual
PREVENTIVE MAINTENANCE
3-6 Signature Edition® GOLD Infusion System, Models 7130/7131 and 7230/7231
Technical Service Manual
PREVENTIVE MAINTENANCE
Signature Edition® GOLD Infusion System, Models 7130/7131 and 7230/7231 3-7
Technical Service Manual
PREVENTIVE MAINTENANCE
70ISS Pressure
Calibration Set
Pressure
Gauge
0 775
PSI
mmHg
3-8 Signature Edition® GOLD Infusion System, Models 7130/7131 and 7230/7231
Technical Service Manual
PREVENTIVE MAINTENANCE
Signature Edition® GOLD Infusion System, Models 7130/7131 and 7230/7231 3-9
Technical Service Manual
PREVENTIVE MAINTENANCE
3-10 Signature Edition® GOLD Infusion System, Models 7130/7131 and 7230/7231
Technical Service Manual
PREVENTIVE MAINTENANCE
Signature Edition® GOLD Infusion System, Models 7130/7131 and 7230/7231 3-11
Technical Service Manual
PREVENTIVE MAINTENANCE
T H I S PA G E
I N T E N T I O N A L LY
LEFT BLANK
3-12 Signature Edition® GOLD Infusion System, Models 7130/7131 and 7230/7231
Technical Service Manual
4 PRINCIPLES OF OPERATION
Chapter 4 — PRINCIPLES OF OPERATION
4.1 INTRODUCTION
This chapter describes the mechanical and
electrical systems that comprise the Signature
Edition® GOLD Infusion System.
Signature Edition® GOLD Infusion System, Models 7130/7131 and 7230/7231 4-1
Technical Service Manual
PRINCIPLES OF OPERATION
4.2 GENERAL INFORMATION (Continued) the camshaft, and the administration set
section compressed by the cam followers.
The instrument contains a peristaltic
pumping mechanism and support circuitry to
4.3 OVERVIEW
ensure controlled flow. The peristaltic
mechanism consists of a linear array of 12 The instrument contains one main PCB and
cam followers which travel perpendicular to several modules that interface to it. The
the administration set. These cam followers interfacing modules are as follows:
act like "fingers" kneading the membrane.
• LED module
When the fluid-filled disposable is placed
against the array of cam followers, the • Graphic LCD module (MAIN)
coordinated, sinusoidal motion of the cams • Lower LCD module
causes a peristaltic wave of fluid • Battery
displacement in the instrumenting segment
• AC off-line switcher
of the disposable.
• Keypad
The instrument will alarm at signs of • Nurse Call / RS-232 board
internal problems and at preset thresholds
for external problems (for example, when • Motor
battery charge falls below a critical level, or • Air-in-line sensors
instrument output pressure exceeds a • Pressure module
programmed limit). All alarms provide visual • Motor rotation sensor and mechanism
and auditory signals to alert the operator. latch detector (optocouplers)
Accuracy of fluid delivery is a function of the • ECD board
microprocessor-controlled rotation cycle of
WATCHDOG
L
CPU KERNEL
BATTERY
MOTOR/ MECH
32 Mhz CPU / EPROM MANAGER
CONTROL
Y1 EEPROM / RAM TAMP
SW
LED
AIL KEY
MAIN PAD
ECD COMBO LCD
IC
AUDIO MAIN SPKR
RS232
AMP SPKR TEST
PRESS
SENSOR ADC VBKUP
VCC
BKUP SPKR
ADREF + SPKR TEST
1.0F
5.5V
L SUPER CAP
4-2 Signature Edition® GOLD Infusion System, Models 7130/7131 and 7230/7231
Technical Service Manual
PRINCIPLES OF OPERATION
Signature Edition® GOLD Infusion System, Models 7130/7131 and 7230/7231 4-3
Technical Service Manual
PRINCIPLES OF OPERATION
4.4 MAIN PCB (Continued) device through the serial data registers
within the COMBO chip.
4.4.1 CPU Kernel (Continued)
This device holds 256, 16-bit words. It is
The kernel data communications function used to store data that will not be destroyed
supports RS-232 level serial if power is lost to the instrument. The
communications up to 19200 baud. The EEPROM will store configuration mode,
UART function is embedded in the COMBO calibration (LCD contrast) settings, and
IC, while the RS-232 interface is based on certain diagnostic information e.g.,
an industry standard RS-232 level converter instrument ID number, PM interval, PM
chip. on/off, battery run time, and total instrument
run time.
4.4.2 COMBO IC
The COMBO IC, U10 (U14), is a 160-pin NOTE: The Guardrails® CQI log and Event
Log are NOT stored in the EEPROM.
PQFP device which supports a variety of
kernel functions, primary audio support,
4.4.4 RAM
digital I/O and other functions. The COMBO
latches the address bus and outputs the The RAM is used to store user set
latched addresses as A19-A16, and A7-A0. parameters, for example, volume to be
The COMBO lC has a 16-bit CRC generator infused (VTBI), mode, rate, and PRI/SEC,
which is used to periodically test the Flash as well as the logs. The instrument
EEPROM data. The RAMTEST circuit provides read/write memory integrity by
provides redundant storage and error using redundant storage and automatic
detection of RAM data. The local serial comparison. Data written to RAM is stored
interface control logic is used to interface in both of the RAM devices, main RAM, U8
with serial accessed peripherals; such as (U9), and phantom RAM, U4 (U5). The
the A/D converter, EEPROM, LED processor reads data directly from the main
Module(s), and Battery Manager. The RAM. The phantom RAM data only goes to
device also generates the Main LCD the COMBO IC. The COMBO IC compares
interface control signals. The UART (Intel the data from the two RAMs on each read.
8251 equivalent) and three 16-bit counters If the data does not match, a bit will be set
(Intel 8254 equivalent) are also provided in a register within the COMBO IC.
inside the COMBO chip. Six pulse width
modulators for motor control and LCD
backlight and contrast are also within the
custom IC.
4.4.3 EEPROM
The EEPROM is used to store all
configuration and diagnostic settings. The
EEPROM, U9 (U11) is accessed using the
serial control unit within the COMBO IC.
Data is written and read back from the
4-4 Signature Edition® GOLD Infusion System, Models 7130/7131 and 7230/7231
Technical Service Manual
PRINCIPLES OF OPERATION
IN-07-00 PD7-0
IN-17-10 RAM Test PH_EN
IN-33-30 RAMCS
Input Circuit
Ports
OUT-27-20
Output Ports OUT-36-34
TXD
UART
Signature Edition® GOLD Infusion System, Models 7130/7131 and 7230/7231 4-5
Technical Service Manual
PRINCIPLES OF OPERATION
MAIN
EEPROM
2K bit PCB
Combo IC Serial I/O
Motor Current Logic
Kernel Serial Intf Controller
80C188 CPU RAM Test Logic
Flash EPROM ROM CRC Logic Battery
128KB RAM LCD Interface Logic
Motor Control Motor Control Microprocessor Backup
32MHz Xtal Audio Tone Generator
PWM Stepper Interface PWM Stepper Interface Watchdog
Master Timebase Battery Charge Control
Audio
Motor Current Sense Motor Current Sense
Motor Timing Logic Battery Monitor Buzzer
Audio Sensor Interface Battery Gauge Interface
A/D Converter Rotation Flag Interface Rotation Flag Interface
Amplifier UART Power On/Off Control
12 Bit Resolution Latch Sensor Interface Latch Sensor Interface
8 Input Multiplexor
Main
Speaker RS232 ESD Board Battery Charger
Pressure 4 Hour Charge Rate
RS232 Connector
Speaker Connector Sensor NiCad/NIMH
ESD Protection Interface Capable
Tamper
Switch Optional Nursecall/ECD Ckt
Channel A Channel B
4-6 Signature Edition® GOLD Infusion System, Models 7130/7131 and 7230/7231
Technical Service Manual
PRINCIPLES OF OPERATION
4.4 MAIN PCB (Continued) The Battery Manager has the following
functions:
4.4.6 RS-232 Interface (Continued)
• Instrument on/off
NOTE: The RS-232 board in Models • Battery charge control
7131/7231 does not have the Nurse Call
option. • Battery gauge
• Battery warning and alarm
4.5 POWER SYSTEM
• Real time clock
4.5.1 Battery Manager
• Displays configuration name
Refer to “Battery Management System” in
• Instrument icon display
“General Information” chapter.
• Inter-processor communications
The Battery Manager is a custom
programmed microcontroller with 4K of • Processor self test
ROM and 1K of RAM memory. The Battery • Error detection (battery, temperature
Manager has two system time bases, a input, current integrator, power on/off,
32 kHz crystal and a 4 MHz ceramic and watchdog faults)
resonator. In normal operation, the Battery
Manager operates at 4 MHz. Under low • In-circuit test
power condition (instrument is off and AC is
unplugged) the Battery Manager switches to
the lower frequency to save power. The
battery manager is turned on for three
seconds every minute when the instrument
is off and connected to AC power.
Charger
Lower
LCD Display
Voltage
Current Analog Battery
Circuits Watchdog
Temperature Manager
Main CPU
AC Power
Sensor On/Off
Switch
Signature Edition® GOLD Infusion System, Models 7130/7131 and 7230/7231 4-7
Technical Service Manual
PRINCIPLES OF OPERATION
4.5 POWER SYSTEM (Continued) The Battery Manager controls the average
current into the battery by varying the duty
4.5.2 AC Off-Line Switcher cycle of the charge control. There are four
Refer to Figures 4-5 "Battery Monitor" and possible phases for the average current
4-6 "Main Power Supply". charge cycle.
4-8 Signature Edition® GOLD Infusion System, Models 7130/7131 and 7230/7231
Technical Service Manual
PRINCIPLES OF OPERATION
Signature Edition® GOLD Infusion System, Models 7130/7131 and 7230/7231 4-9
Technical Service Manual
PRINCIPLES OF OPERATION
4.5 POWER SYSTEM (Continued) When MEASURE is low, Q27 (Q43) will
open and remove BATT_PLUS from input to
4.5.7 System Power Source Select (Continued) U29 (U37). Also MEASURE, being low, will
The system power source, open Q26 (Q45) so that VMEAS is removed
DC_DC_SOURCE, is used to drive three as well.
switching power supplies. It is controlled by
the Battery Manager through the signal 4.5.9 VMEAS
PWR_ON. When the instrument is on, the Refer to figures 4-5 "Battery Monitor" and
DC_DC_SOURCE is normally supplied by 4-6 "Main Power Supply".
either the battery or the AC off-line switcher
if the instrument is plugged into AC. The VMEAS is the supply used to power the
Battery Manager disconnects the AC source REF 4.1V reference circuit, the voltage
during the battery refresh cycle. monitor circuit, and the current monitor
circuit. The Battery Manager turns VMEAS
When the Battery Manager asserts off by setting the MEASURE signal low
PWR_ON high, Q24 (Q41) will turn on. If when the instrument is off and AC is
PWR_AC is high, Q7 (Q25) will turn on, and unplugged to reduce the load on the battery.
DC_DC_SOURCE will be supplied by the In this condition, the Battery Manager turns
DC_INPUT. VMEAS on once a minute to check the
battery's voltage and temperature.
If PWR_AC is high, Q8 (Q21) is on and
DC_DC_SOURCE will be supplied by the The 12V Zener CR31 (CR41) is placed
battery. between the gate and source of the FET,
Q26 (Q45), to limit the gate to source
4.5.8 Battery Voltage Monitor voltage. The FET can see up to 24V but the
Refer to figures 4-5 "Battery Monitor" and Vgs (Voltage gate to source) of the FET is
4-6 "Main Power Supply". only rated to 20V.
The battery voltage can range from 10 to 4.5.10 Voltage Reference 4.1V
18V. The Battery Manager monitors the
Refer to figures 4-5 "Battery Monitor" and
battery voltage through its internal A/D
4-6 "Main Power Supply".
converter. The valid input range of the A/D
converter is 0 to 4.1V. The battery voltage The 4.1V reference is used in the voltage
must be reduced to meet the input monitor circuit, the battery temperature
requirements. The 4.1V reference and the sensor circuit, the ambient temperature
voltage subtracter-multiplier amplifier circuit sensor circuit, and is the reference voltage
U29-5 (U37-5) scale the battery voltage and for the A/D converter in the Battery
maintain an accuracy of ±15 counts Manager. The reference voltage is
(1 count = lmV). 4.096V ±2%. VMEAS, which can be
between 7.7 to 24V, turns on the precision
The Battery Manager uses the battery
reference Zener U27 (U47).
voltage for its charging, battery gauge, error
detection, and battery alarm and warning
features.
4-10 Signature Edition® GOLD Infusion System, Models 7130/7131 and 7230/7231
Technical Service Manual
PRINCIPLES OF OPERATION
Signature Edition® GOLD Infusion System, Models 7130/7131 and 7230/7231 4-11
Technical Service Manual
PRINCIPLES OF OPERATION
4.5 POWER SYSTEM (Continued) internal pass element shorts the output
side of the power inductor to ground,
4.5.13 System Switching Supplies then releases it, generating about 29V at
(Continued) the cathode of a diode. This voltage is
The regulator has an internal pass sensed by a resistor pair and the loop
element which turns on current to the controlled to generate a 1.23V reference
output inductor until the output voltage, signal level.
as sensed through the sense resistors, • VNEG supply U16 (U30): A buck-boost
reaches the internal 1.23V reference supply which generates a negative
voltage. It then turns off and the inductive supply (-15V ±1V) from a positive supply
flyback voltage created is clamped by a (9V-24V). The supply can provide about
catch diode. 100mA. The IC ground pin is
• VMOTOR supply U21 (U26): A step-up bootstrapped to the negative output
("boost") supply which provides a higher voltage. Then, referencing the feedback
output voltage (29V ±1.5V) than the input signal to ground potential allows the chip
supply (9V-24V). The supply can provide to sense the negative voltage and
up to 1A peak and has a soft start feature therefore regulate it.
to limit inrush current upon starting. The
4-12 Signature Edition® GOLD Infusion System, Models 7130/7131 and 7230/7231
Technical Service Manual
PRINCIPLES OF OPERATION
85-264VAC 12V
Voltage +
50/60 Hz 1.8AH 1F
Reg 5.5V
Input
U39 (U43)
L L
VAO_SOURCE 7.6-24V
Voltage
Reg
SHUT_DOWN* VBKUP_SRC_EN* U31 (U44) VBKUP_SRC_TST*
Signature Edition® GOLD Infusion System, Models 7130/7131 and 7230/7231 4-13
Technical Service Manual
PRINCIPLES OF OPERATION
10 Hz
U36 (U46)
13 (5)
+5V Q
Q*
PWR_SW_VAO* (A/B*) R
CLK
R245 (R323) U40 (U39)
Q*
C49 (C67) CL*
L U36 (U46)
5 (13) WD_OUT*
Q
9 (1)
CLK NCALL_WD*
12 (4)
Q* Q35 (Q49)
WD_OUT
4-14 Signature Edition® GOLD Infusion System, Models 7130/7131 and 7230/7231
Technical Service Manual
PRINCIPLES OF OPERATION
R61 INV
PWR_SW* (A/B*) PWR_SW_VAO* (A/B*)
On/Off INT1*
Switch(s) 100 100K (INT2*) RST_CPU*
R62
1 uF
PWR-ON
Keypad U40 (U39)
L
9
Q
CLK
Power Latch
Q*
R
U40 (U39) reset by 8 Msec pulse from U36-13
(U46-5) when power switch pressed to remove U36 (U46)
power from instrument.
13 (5)
Q
CLK
WD Alarm Enables U36-13 (U46-5)
R
WD_OUT*
Signature Edition® GOLD Infusion System, Models 7130/7131 and 7230/7231 4-15
Technical Service Manual
PRINCIPLES OF OPERATION
4.5 POWER SYSTEM (Continued) 4.5.20 Lower LCD Display Backlight Drive
The backlight for the Lower LCD display
4.5.18 Power Switch (Continued)
contains 6 LEDs in series. The signal
The signals are pulled up to 5V, by VBKUP AOD_BKLT is tied to the anode of the first
supply, through two pull-up resistors. LED in the series. Each led has about a 2V
(VBKUP is the supply for the backup audio.) forward drop. So AOD_BKLT will be about
These signals are decoupled from the 12-13V when the LEDs are turned on.
signals that drive the Battery Manager logic,
PWR_SW* (PWR_SW_A/B*), through two The Lower LCD display backlight drive
100K resistors. The PWR_SW* circuit controls the amount of current
supplied to the LEDs. The circuit contains
(PWR_SW_A/B*) signals drive interrupt two current sources "or'd" together. One
inputs in the Battery Manager, U34 (U40), current source is powered from DC_INPUT,
and are used by the error reset and power the voltage supply from the AC off-line
on circuits. Upon recognition of the power switcher. The second source is powered
switch, the Battery Manager controls when from VMOTOR which is generated when the
the Main Processor and the rest of the instrument is on.
circuit will receive power.
The circuit has four modes of operation:
4.5.19 System Reset/Power On • The instrument is off and unplugged. In
(refer to Figure 4-8) this mode, the backlight driver is off and
System Reset is controlled by the Battery the two power supplies that drive the
Manager. Either the lack of a reset inactive circuit are both off.
signal from the Battery Manager, U34 (U40), • The instrument is off but plugged into AC.
signal R61 or the lack of the PWR_ON In this mode, the current to the LEDs is
signal being asserted will initiate a limited by a resistor. VMOTOR will not
RST_CPU*. be on, so only the DC_INPUT path
Power on is also controlled by the Battery generates the current for the LEDs and
Manager. Under normal conditions the the current to the LEDs will be about
Battery Manager uses its output R62 to 4 mA.
toggle the power latch, U40-9 (U39-9), to • The instrument is on and plugged into
turn the instrument on/off. The Battery AC. In this mode, both current sources
Manager turns PWR_ON high when the are on, driving the LEDs for maximum
instrument is on and turns PWR_ON low brightness, and the current to the LEDs
when the instrument is off. should be around 6 mA.
If the instrument is on and a watchdog error • The instrument is on and not plugged
is active, WD_OUT* is asserted low. The into AC. In this mode, only VMOTOR is
second one shot, U36-13 (U46-5), will be on, so the driver generates the current to
enabled. When the power switch is pressed, turn on the LEDs and the current to the
an 8 ms pulse is generated at the Q output LEDs should be around 2 mA.
of U36-13 (U46-5). The pulse resets the
power latch, U40-9 (U39-9).
4-16 Signature Edition® GOLD Infusion System, Models 7130/7131 and 7230/7231
Technical Service Manual
PRINCIPLES OF OPERATION
4.6 MOTOR DRIVE/SENSORS comparator will turn the high side switch
back on.
4.6.1 Motor Drive (refer to Figure 4-9)
The comparator circuit has been
The stepper motor drive circuit consists of a designed with a fixed turn on delay of 50
dual H bridge to provide voltage to each microseconds. This is a result of the RC
winding of the hybrid stepper motor, and a network on the output of the sense
voltage comparator to control the duration comparator stage. A second comparator
that voltage is applied to each motor stage will sense when the output of the
winding. first stage rises above 3.3 volts. The turn
on delay results in a maximum chopping
The sequence of operation for a single
frequency of 20 kHz. The lowest
phase [Phase 1 (A)] of the motor is as
chopping frequency is a function of the
follows:
motor current, at a maximum motor
1. Phase A, MTR_PH_1 (MTRA_PH_1) is current of 240 milliampere, the chopping
active, which causes the high side switch frequency is 14 kHz. Inserted between
and low side switch (diagonally opposite the sense resistor and the comparator
the high side switch), to close. This input is an RC network needed for
presents the motor supply voltage across filtering of the short circuit current caused
the motor winding. The MTR_1A by the distributed capacitance of the
(MTRA_IA) and MTR_1B (MTRA_1B) motor winding.
signals are used to drive the motor.
4. The chopping action results in a steady
2. Current will begin to increase at a rate state current in the winding for a given
determined by the ratio of the motor phase duration. When the phase is
voltage to the inductance of the motor; reversed the opposite high side and low
for example, about 0.5 amps/millisecond, side switch will turn on. This forces
and flow in the direction indicated by the current to flow in the opposite direction.
arrow shown. The current will increase in The reference, MTR_l1 (MTRA_I_1), is
the sense resistor at the same rate and controlled to ramp the motor current
result in a voltage sensed by the exponentially, minimizing step oscillation
comparator. and reducing mechanical noise.
3. Once the sense resistor voltage MTR_I1
(MTR A/B_I_1) rises above the reference
voltage at the comparator inputs, U14
(U27), the comparator output, U14-1
(U27-7), will switch low forcing the high
side switch to open. The low side switch
will always remain closed until a phase
change occurs. With the supply voltage
now removed from the coil, the coil
current and the sense resistor current will
decay. Once the sense resistor voltage
drops below the reference voltage, the
Signature Edition® GOLD Infusion System, Models 7130/7131 and 7230/7231 4-17
Technical Service Manual
PRINCIPLES OF OPERATION
4-18 Signature Edition® GOLD Infusion System, Models 7130/7131 and 7230/7231
Technical Service Manual
PRINCIPLES OF OPERATION
MTR_1A MTR_1B
(MTRA_1A) (MTRA_1B)
MTR_PH_1
(MTRA_PH_1)
VIA U13-7 (U25-1)
Q19 (Q26) Q16 (Q24)
Q3 (19)
PHASE A* PHASE A
MTR_SNS*
U12 (U17)
Sense Resistor
Sense Comparators
MOTOR_CURRENT_1_A/B MOTOR_COIL_1A_A/B
COMBO IC Stepper
PWM Output MOTOR_CURRENT_2_A/B
MOTOR_COIL_1B_A/B Motor
Coil 1
Constant
MOTOR_PHASE_1_A/B
Current MOTOR_COIL_2A_A/B
Motor Stepper
MOTOR_PHASE_1_A/B
Driver Motor
MOTOR_COIL_2B_A/B
Coil 2
MOTOR_EN_1_A/B
MOTOR_EN_1_A/B
Motor MOTOR_CURRENT_SENSE A
COMBO IC Current
Digital Output Sense
Mechanism COMBO IC
MOTOR_FLAG_EN MOTOR_FLAG_A/B
Rotation Digital Output
Sensor
Mechanism
LATCH_FLAG_EN MOTOR_LATCH_A/B
Latch
Sensor
Signature Edition® GOLD Infusion System, Models 7130/7131 and 7230/7231 4-19
Technical Service Manual
PRINCIPLES OF OPERATION
COMBO IC AIL_GATE
Digital Output AIL_TEST
Pressure
COMBO IC PRES_TST_A/B PRES_A/B Inverting
PRES_SIG_A/B
12-Bit A/D
Transducer
Digital Output Buffer Converter
Assembly
4-20 Signature Edition® GOLD Infusion System, Models 7130/7131 and 7230/7231
Technical Service Manual
PRINCIPLES OF OPERATION
Signature Edition® GOLD Infusion System, Models 7130/7131 and 7230/7231 4-21
Technical Service Manual
PRINCIPLES OF OPERATION
4.7 USER INTERFACE (Continued) verified at power up. The speaker audio
volume settings are approximately:
4.7.1 Main Speaker Driver
(refer to Figure 4-14) Low = min 45 dB
Med = 65 dB
The main speaker driver is based on an Hi = approx. 70 dB
LM386 low voltage speaker drive chip, U32
(U29), driven by an exponentially weighted 4.7.2 Backup Audio Buzzer and Test Circuit
4-bit control signal (AMP 0 to 3) modulated (refer to Figure 4-15)
at the desired frequency (200 Hz - 4 kHz) by
A backup audio generation capability is
the COMBO IC U10 (U14). The speaker is
provided to allow the instrument to generate
pulsed at 3 to 50 mA with a 50% duty cycle
an audible alarm in the event that the main
(max. 100 mA). The driver has a fixed 26dB
speaker is unable to do so. It is supplied
gain which provides up to 3VPP AC coupled
through VBKUP, a 1.0 Farad "supercap"
into an 8Ω speaker. The speaker is tested
C146 (C179) energy storage device
by monitoring the speaker current with a
charged by a 5V linear regulator, U32 (U44),
0.511Ω resistor.
on the VAO SOURCE supply. The buzzer is
The voltage across the resistor is amplified, a self oscillating audio generator and
rectified, and compared to low and high speaker module which produces a 3 to
threshold values by a window comparator, 4 kHz tone when energized. A logic circuit,
U38 (U16). The speaker is tested when an U30 (U41), powered by VBKUP arms the
alarm or error occurs, while the test circuit is circuit using the BKUP_ALARM_ARM signal
LCD_C31/2
COMBO IC
LCD LCD_EN
Interface LCD_WR
LED
LCD_RS
Backlit
Graphic
5-Bit PWM to LCD
BKLT_LED LED_CATH
0-200mA Module
COMBO IC LED Drive
PWM
Output LCD_CONT 8-Bit PWM to
VEE
-7V to -11V
Converter
BCLK BATT_IRPT
SEQ23-0
COMBO IC
Lower Digital Input
Battery COMM-1 Display
COMBO IC Processor
Serial SDI_BATT
Control COMBO IC
Unit SDO KYBD_IRPT
Serial
SDI_LED
Control
LED_CS_A/B LED
Unit
Module KEY_COL_4-0
SCLK Front
and Rear
WD_OUT KEY_ROW_7-0
Keypads
4-22 Signature Edition® GOLD Infusion System, Models 7130/7131 and 7230/7231
Technical Service Manual
PRINCIPLES OF OPERATION
Rear Case
COMBO U32 (U29)
IC AMP 0 to 3
+
SPKR_POS
3 to 50 mamp SUM AMP
U10 (U14) +
- Main Speaker
2KHz
-
SPKR_NEG
L VTHRES HI
+ MAIN_SPKR_TST
+ U38 (U16)
OP AMP Window
R58 (R351) Comparator
-
0.511 OHM
U37 (U32)
Gain = 37
L L
+
VTHRES LOW
-
Signature Edition® GOLD Infusion System, Models 7130/7131 and 7230/7231 4-23
Technical Service Manual
PRINCIPLES OF OPERATION
VAO_SOURCE
5V REG
VBKUP_SRC_EN* U31 (U44)
+
C146 (C179) Super cap takes up to 30 minutes
Super Cap to charge and lasts from 11 minutes
1.0F to 2 hours.
5.5V
L
PWRSW_VBKUP
U30 (U41)
+ SPKR2
3 to 4 Khz
-
4.9 LOWER LCD DISPLAY was generated to find and avoid problems.
Screen updates occur every 100 mSec.
The Lower LCD Display is visible whether
The backlight intensity is controlled by the
the instrument is on or off, AC or battery
Main Processor to limit power use while
operation. It shows the current battery
running on battery. The LCD contrast is
charge available along with the current
controlled by the main CPU and can be
audio level selected, communications
adjusted from factory default through the
interface status, panel lock status, and
diagnostics mode. Refer to "Changing Main
battery refresh status. It is driven by the
LCD Contrast" section in “Troubleshooting”
Battery Manager and is back lit when the
chapter.
instrument is on or plugged into AC.
Figure 4-16. Lower LCD Display
4.10 MAIN LCD MODULE
The Main LCD Module provides user
GOLD
information on a 128x64 dot matrix display
with LED backlight. The Main Processor MNTR
generates all text and symbols in bit
mapped form, then compares data read
back from the display memory to that which
4-24 Signature Edition® GOLD Infusion System, Models 7130/7131 and 7230/7231
Technical Service Manual
PRINCIPLES OF OPERATION
4.10 MAIN LCD MODULE (Continued) 4.13 ECD BOARD (refer to Figure 4-17)
The ECD board (optional for 7130/7230)
4.10.1 Main LCD Backlight
provides empty container detection using
The graphic LCD backlight is an array of standard ALARIS® Model 180 Flow
LEDs driven by an adjustable constant Sensors.
current source controlled by a PWM signal
from the COMBO IC. The LED current can The flow sensor interface is a separate
be adjusted stepwise linearly over a board assembly which drives a standard
0-200 mA range. The current source ALARIS® Model 180 Flow Sensor, performs
consists of a low on resistance FET, 0.511Ω ambient light rejection on the resulting
sense resistor, and op amp to set the sense signals, and provides digital output signals
resistor voltage based on the filtered PWM for a detected drop and sensor attached
input voltage BKLT_LED. The backlight detection. The board consists of 2 separate
intensity is not user adjustable. ambient light discriminator loops and drop
detectors for independent Model 180 Flow
4.10.2 Graphic LCD Contrast Sensors along with common timing and
control logic. The board assembly is
The graphic LCD contrast is controlled by common to both single and dual channel
varying the VNEG supply to the module instruments, where the "B" channel is not
over a -7V to -11V range. The drive circuit available externally and is ignored in the
inverts and scales the filtered PWM software of the single channel instruments.
LCD_CONT signal from the COMBO IC to Note that the "A" channel is used for both
cover this range. Nominally, the contrast the "A" channel in the dual channel
voltage is -9V, but can be adjusted through instrument as well as the single channel
Diagnostics Mode in software. instrument, even though the mechanisms
are on opposite sides of the case.
4.11 NURSE CALL CIRCUIT
(7130/7230 ONLY) • Common Timing Logic:
The common timing and control logic
An optional nurse call circuit is located on generates the necessary discriminator
the RS-232/Nurse Call board and provides a timing signals to drive the flow sensor
35V @ 1A rated relay contact on system LED and sample signals representing
alarms through pins 6 and 9 of the RS-232 room ambient light and LED driven light
connector. outputs from the flow sensor. The main
system CLK_32MHz is used as the
4.12 PANEL LOCK SWITCH timebase and decade counter U3 that
The instrument can be protected from generates non-overlapping "A" and "B"
unauthorized changes with the Panel Lock channel drive and sense signals. This
Switch. A lock symbol is shown in the reduces the peak LED current load and
Lower LCD display whenever the feature is flow sensor crosstalk during normal
active. When activated, access to all front operation. Transistors Q5 and Q6 provide
panel keys is restricted (except channel logically inverted control signals for the
select and split screen viewing key). LED drive circuits.
Signature Edition® GOLD Infusion System, Models 7130/7131 and 7230/7231 4-25
Technical Service Manual
PRINCIPLES OF OPERATION
4.13 ECD BOARD (Continued) To prevent a drop from skewing the LED
drive signal, a long time constant filter,
• Ambient Light Discriminator Loop:
consisting of CR1/CR2, R31/R13, and
The flow sensor drive current is set to
C9/C10, is enabled on a negative going
maintain a 2.8V level normally at
output signal DROP_A/B.
DROP_A/B. This level is determined by
the difference in input signal from a • Drop Detector:
Model 180 Flow Sensor when the emitter The drop detector circuit generates a
is undriven and driven. Analog digital pulse when a valid drop event is
multiplexor U4 normally grounds the detected. A drop event occurs when a
output of the sense capacitor C18/23 so fluid drop passes between the emitter
that room ambient light signal voltage is and detector of the Model 180 Flow
set across it. When the LED drive is Sensor and appears as a generally
turned on, the output of the sense negative going short duration pulse on
capacitor, representing the driven signal the DROP_A/B signal. Detector
voltage less the ambient signal voltage, comparators U8B/U7B generate a 0 to
is transferred to sample capacitor 5V pulse when a drop "signature" of
C28/C26. This signal is amplified by appropriate length and duration occurs.
U8/U7 with a gain of about 23 and is the One shots U9A/B generate a nominal
DROP_A/B signal. 20ms pulse which indicates a valid drop
(ECD-SIG-A/ECD-SIG-B). The main
The DROP_A/B signal is sensed by
system processor processes the pulse
integrator U1/U6 and C9/C10 to generate
stream and determines whether the
an appropriate LED drive level to
drops are occurring properly for the
maintain DROP_A/B at about 2.8V. The
current instrument rate and operating
integrator output signal is controlled by
mode (for example, primary/secondary).
drive enable FET Q1/Q2 to drive the LED
constant current sink U1/Q4 and relate EMC filtering is provided by filters C4-8 to
components. The constant current limit EMI into or out of the flow sensor
source generates a 0-200mA sink current connection pins.
with a 0V-1V input signal.
Drop_A/B 2.8 V
+5 V
4-26 Signature Edition® GOLD Infusion System, Models 7130/7131 and 7230/7231
Technical Service Manual
PRINCIPLES OF OPERATION
Signature Edition® GOLD Infusion System, Models 7130/7131 and 7230/7231 4-27
Technical Service Manual
PRINCIPLES OF OPERATION
T H I S PA G E
I N T E N T I O N A L LY
LEFT BLANK
4-28 Signature Edition® GOLD Infusion System, Models 7130/7131 and 7230/7231
Technical Service Manual
5 CORRECTIVE MAINTENANCE
Chapter 5 — CORRECTIVE MAINTENANCE
5.1 INTRODUCTION
WARNING
The instrument case should only be opened by This chapter contains procedures required
qualified personnel using proper grounding to properly disassemble, repair and replace
techniques. Disconnect instrument from AC parts as well as to test, calibrate, and
prior to performing maintenance. Hazardous reassemble a Signature Edition® GOLD
voltages are present when AC power is Infusion System.
connected, regardless of the ON/OFF Switch A thorough familiarization with the function
setting. and operation of the mechanical assemblies
and electrical circuits of the pump will enable
CAUTION repair and calibration to be accomplished
CMOS devices are sensitive to static electrical more efficiently (refer to “Principles of
charges and may be damaged during repair if Operation” chapter).
the repair activity is not performed in an The circuit boards used in the instrument
electrostatic discharge (ESD) protected are fitted with surface mount devices and
environment, using approved ESD protective are deemed nonfield repairable. Therefore,
procedures, including personnel grounding, or ALARIS Medical Systems requires that all
the instrument could be damaged. circuit boards be returned to an authorized
ALARIS Medical Systems Service Center
Table 5-1. Test Equipment
NAME MANUFACTURER MODEL NUMBER APPLICATION
Used to test AC wiring and pump
Electrical Safety Tester BIO-TEK 260*
grounding.
IV Infusion Set IVAC 80 VCS Rate accuracy test
Nurse Call Cable IVAC 136111 Nurse call option
Pressure Cal Set IVAC 70ISS Pressure verification and calibration
Dresser Industries: PTE1/901M1*
Pressure Gauge 203-406-3115
Pressure verification and calibration
(-400 to +750mmHg) (Heise) BIO-TEK:
802-655-4040 DPM III*
Burette 50mL, 0.1mL Fisher Scientific Class A or B*
Rate calibration and verification
increment Keynex 113 Sec A*
Connects between 2 instruments to
RS-232 (9-pin, Null modem) IVAC 133450*
download configuration.
Silicon Tubing IVAC 303109* Pressure calibration setup
T-Fitting IVAC 313815* Pressure calibration setup
Alexander Batteries Model 2006*
NiCad Battery Optomizer Test and condition batteries (optional)
800-577-2539 2003*
Hand-Held Vacuum Any Any Main board assembly cleaning
* or equivalent
Signature Edition® GOLD Infusion System, Models 7130/7131 and 7230/7231 5-1
Technical Service Manual
CORRECTIVE MAINTENANCE
Lower LCD
Main Board LED
Main LCD
Front Case Mechanism (including AIL, LATCH and Pressure Sensor)
Main Keypad
5-2 Signature Edition® GOLD Infusion System, Models 7130/7131 and 7230/7231
Technical Service Manual
CORRECTIVE MAINTENANCE
Signature Edition® GOLD Infusion System, Models 7130/7131 and 7230/7231 5-3
Technical Service Manual
CORRECTIVE MAINTENANCE
5.2 DISASSEMBLY/REASSEMBLY
(Continued)
Battery
Battery Cover
5-4 Signature Edition® GOLD Infusion System, Models 7130/7131 and 7230/7231
Technical Service Manual
CORRECTIVE MAINTENANCE
Signature Edition® GOLD Infusion System, Models 7130/7131 and 7230/7231 5-5
Technical Service Manual
CORRECTIVE MAINTENANCE
5.2 DISASSEMBLY/REASSEMBLY
(Continued)
Handle Cap
Snap Fit Points
Battery Compartment
5-6 Signature Edition® GOLD Infusion System, Models 7130/7131 and 7230/7231
Technical Service Manual
CORRECTIVE MAINTENANCE
5.2 DISASSEMBLY/REASSEMBLY
(Continued)
Remaining RS 232
Case Screws
Signature Edition® GOLD Infusion System, Models 7130/7131 and 7230/7231 5-7
Technical Service Manual
CORRECTIVE MAINTENANCE
5.2 DISASSEMBLY/REASSEMBLY
(Continued)
Channel B Channel A
Locking Bar
Ground Wire
Unlocked position
Flex cable
Locked position
PCB
5-8 Signature Edition® GOLD Infusion System, Models 7130/7131 and 7230/7231
Technical Service Manual
CORRECTIVE MAINTENANCE
5.2 DISASSEMBLY/REASSEMBLY
(Continued)
7 6
5 2
3
Torque Sequence
4
1
Signature Edition® GOLD Infusion System, Models 7130/7131 and 7230/7231 5-9
Technical Service Manual
CORRECTIVE MAINTENANCE
5.2 DISASSEMBLY/REASSEMBLY
(Continued)
8 7
6 9
1 4
Torque Sequence
1 Case Screw
inside Battery
Compartment
3 5 2
5-10 Signature Edition® GOLD Infusion System, Models 7130/7131 and 7230/7231
Technical Service Manual
CORRECTIVE MAINTENANCE
Pole Clamp
RS 232
Signature Edition® GOLD Infusion System, Models 7130/7131 and 7230/7231 5-11
Technical Service Manual
CORRECTIVE MAINTENANCE
5.2 DISASSEMBLY/REASSEMBLY
(Continued)
Power Supply
Board Assembly
Cover
5-12 Signature Edition® GOLD Infusion System, Models 7130/7131 and 7230/7231
Technical Service Manual
CORRECTIVE MAINTENANCE
ECD Board
Signature Edition® GOLD Infusion System, Models 7130/7131 and 7230/7231 5-13
Technical Service Manual
CORRECTIVE MAINTENANCE
RS-232 Connector
(cover in place)
RS-232 Connector
(cover open)
5-14 Signature Edition® GOLD Infusion System, Models 7130/7131 and 7230/7231
Technical Service Manual
CORRECTIVE MAINTENANCE
Speaker Connector
RS-232 Board
Cable
Panel Lock Key Pad
Locking Bar Fan Connector
Signature Edition® GOLD Infusion System, Models 7130/7131 and 7230/7231 5-15
Technical Service Manual
CORRECTIVE MAINTENANCE
Line Filter
(Behind RS-232 Board)
5-16 Signature Edition® GOLD Infusion System, Models 7130/7131 and 7230/7231
Technical Service Manual
CORRECTIVE MAINTENANCE
5.2 DISASSEMBLY/REASSEMBLY
(Continued)
Speaker Connector
Fan Connector
Signature Edition® GOLD Infusion System, Models 7130/7131 and 7230/7231 5-17
Technical Service Manual
CORRECTIVE MAINTENANCE
3. Remove tip from end of pole clamp knob 9. Place retaining ring over end of pole
by pressing spiral pin out of opening in clamp mounting pin, on top of flat
end of brass thread. washer. Press downward on retaining
ring, compressing flat washer and wave
4. Unscrew knob from pole clamp. washer downward until retaining ring
snaps into groove on mounting pin.
During Reassembly: 10. Verify retaining ring is correctly installed.
Rotate retaining ring inside groove to
1. Screw knob into pole clamp assembly. ensure correct installation.
2. Attach pole clamp tip to end of thread on
knob and press spiral roll pin into
opening, through thread and tip.
5-18 Signature Edition® GOLD Infusion System, Models 7130/7131 and 7230/7231
Technical Service Manual
CORRECTIVE MAINTENANCE
5.2 DISASSEMBLY/REASSEMBLY
(Continued)
Brass Knob
Stainless Steel Knob (old version)
(new version)
Spiral Pin
Compression Spring
Shown For
Reference Only Tip
Apply grease to
mounting pin Pole Clamp
Flat Washer
Wave
Washer
Retaining
Ring Heat Sink
(reference)
Signature Edition® GOLD Infusion System, Models 7130/7131 and 7230/7231 5-19
Technical Service Manual
CORRECTIVE MAINTENANCE
5.2 DISASSEMBLY/REASSEMBLY
(Continued)
Incorrectly Installed
Exposed Groove
Retaining Ring
Wide Gap
Top View
Mounting Pin
Correctly Installed
Narrow Gap
Top View
5-20 Signature Edition® GOLD Infusion System, Models 7130/7131 and 7230/7231
Technical Service Manual
CORRECTIVE MAINTENANCE
Cable Clamp
(Ground Lead)
Cable Clamp
(Ground Lead)
Signature Edition® GOLD Infusion System, Models 7130/7131 and 7230/7231 5-21
Technical Service Manual
CORRECTIVE MAINTENANCE
5.2 DISASSEMBLY/REASSEMBLY 4. Cut tie #5 for ChA motor and AlL receiver
(Continued) (white).
5. Disconnect ChA and ChB transducer
5.2.8 Disconnecting Cables
wires (blue) from the board.
(Continued)
6. Disconnect ChA and ChB ground wires
Front Case
(black) going to the mechanism
1. Cut tie #l for ChB motor and AlL receiver assembly.
(black).
7. Disconnect flex cables for front panel,
2. Cut tie #2 for ChB motor, Latch, Rotation LED modules, main LCD, RS-232 board
and AlL receiver (black). assembly and lower LCD as needed. Lift
up on locking bar before attempting to
3. Cut tie #4 for ChA motor, Latch, Rotation
remove flex cable.
and AlL receiver (white).
8. Continue disassembly as required.
Back 4-Bar
Latch Cable
Motor Cable
Ground Wires
Tie #2
XDCR Cable
Rotation Cable
Motor Cable
Tie #1
5-22 Signature Edition® GOLD Infusion System, Models 7130/7131 and 7230/7231
Technical Service Manual
CORRECTIVE MAINTENANCE
5.2 DISASSEMBLY/REASSEMBLY
(Continued)
Channel B Channel A
XDCR Cable
Latch Cable
Back 4-Bar
Motor Cable
Tie #4
Rotation Cable
Tie #5
Motor Cable
Signature Edition® GOLD Infusion System, Models 7130/7131 and 7230/7231 5-23
Technical Service Manual
CORRECTIVE MAINTENANCE
1. Disconnect all wire connections to main 1. Check Rate Cal number in Diagnostic
board. Refer to “Disconnecting Cables” Mode.
section. D4 screen should have a Rate Cal # of
0% (refer to “Preventive Maintenance”
NOTES: chapter for “Rate Calibration Procedure”,
• When removing flex cables, carefully if needed, and “Troubleshooting” chapter
lift locking bar to remove cable from for procedure to change).
connector.
2. Perform Soft Pressure Calibration after
• Pay particular attention to wire routing. 30 minutes warm-up.
Wires should be routed back to initial
scheme and similar to the example
shown. This will prevent them from
getting pinched and jamming the
mechanism.
2. Remove screw from main board using
Phillips screw driver. Slide main board to
top, then lift and remove.
5-24 Signature Edition® GOLD Infusion System, Models 7130/7131 and 7230/7231
Technical Service Manual
CORRECTIVE MAINTENANCE
5.2 DISASSEMBLY/REASSEMBLY
(Continued)
Signature Edition® GOLD Infusion System, Models 7130/7131 and 7230/7231 5-25
Technical Service Manual
CORRECTIVE MAINTENANCE
5.2 DISASSEMBLY/REASSEMBLY
(Continued)
Snap Fittings
LED Channel B
LED Channel A
5-26 Signature Edition® GOLD Infusion System, Models 7130/7131 and 7230/7231
Technical Service Manual
CORRECTIVE MAINTENANCE
Latch (opened
half way)
Channel B Channel A
Snap fittings
Mechanism
Snap fittings
Signature Edition® GOLD Infusion System, Models 7130/7131 and 7230/7231 5-27
Technical Service Manual
CORRECTIVE MAINTENANCE
5.2 DISASSEMBLY/REASSEMBLY
(Continued) 3. Slide on AlL gear and align AlL gear tab.
4. Press AlL gear into place.
5.2.12 Removing AIL Transmitter (Arm)
(Figures 5-23 through 5-25) NOTE: When AIL gear is pressed onto
AIL transmitter, AIL transmitter may shift
NOTE: Observe placement of cable ties (for slightly as gear locks into place.
reinstallation). Cut and remove them from
cables. 5. Use needle-nose pliers to lock gear tab
into place.
1. Use needle-nose pliers to unlock AlL
gear tab (Figure 5-23). 6. Turn mechanism latch to far left (closed
position) and check AlL transmitter
2. If replacing AlL receiver, replace before
alignment. AlL transmitter must be in
continuing with AlL transmitter
closed position (horizontal) (Figure 5-25).
removal/replacement.
7. If AlL transmitter is not properly aligned:
3. Remove AlL gear and AlL transmitter
unlock, realign and relock gear tab until
from mechanism.
transmitter is properly aligned.
4. Move mechanism latch to right (open
position). NOTE: AIL transmitter may need to be
aligned with inner edge of mechanism
During Reassembly: seal before locking gear into place.
1. Slide AlL transmitter through faceplate. 8. Move mechanism latch to middle
Rotate AlL transmitter to open (vertical) position. Install mechanism into front
position (Figure 5-24). case.
2. Thread AlL transmitter wire through AlL 9. Route and connect cables. Install cable
gear. tie straps on cables as before.
CAUTION
Grease is applied to a new AlL
transmitter for lubrication purposes. Do
not remove grease when installing a new
AlL transmitter into the mechanism. Do
not apply additional grease to the AlL
gear.
5-28 Signature Edition® GOLD Infusion System, Models 7130/7131 and 7230/7231
Technical Service Manual
CORRECTIVE MAINTENANCE
5.2 DISASSEMBLY/REASSEMBLY
(Continued)
Figure 5-23. Unlocking AIL Gear Tab Figure 5-24. AIL Transmitter in
Open Position
Tab
AIL
Transmitter
Inner Edge of
Slot Mechanism
Seal
AIL
Receiver
Figure 5-25. AIL Transmitter in
Closed Position
AIL
Transmitter
Signature Edition® GOLD Infusion System, Models 7130/7131 and 7230/7231 5-29
Technical Service Manual
CORRECTIVE MAINTENANCE
5.2.13 Removing AIL Receiver (Button) 3. Push AlL receiver in and rotate 90o
(Figures 5-26 and 5-27) counter-clockwise to lock into place.
4 Tuck mechanism seal under AlL receiver.
NOTE: Observe placement of cable ties (for
reinstallation). Cut and remove existing 5. Proceed with AlL transmitter
straps from cables. replacement, if needed, and continue
following steps for AlL receiver
1. Turn mechanism latch to far right. AlL
replacement.
transmitter will be in open position.
6. Close and open mechanism latch several
2. Push AlL receiver in and rotate about 90o
times to ensure AlL transmitter and
clockwise. AlL receiver should pop out.
receiver open and close properly.
3. Remove AlL receiver wire from
7. Route AlL receiver's wire under
faceplate's hook and remove receiver
faceplate's hook (Figure 5-27).
from mechanism.
8. With mechanism latch opened halfway,
NOTE: Leave AIL spring in place for new install mechanism into front case.
AIL receiver.
9. Route and connect cables. Install cable
tie straps onto cables.
During Reassembly:
1. Thread AlL receiver wire through
opening.
2. Install AlL receiver with flat edge of
assembly positioned as shown in Figure
5-26.
Figure 5-26. Flat Edge of AIL Receiver Figure 5-27. Faceplate Hook
Flat
Edge
Faceplate
Hook
Mechanism
Seal
5-30 Signature Edition® GOLD Infusion System, Models 7130/7131 and 7230/7231
Technical Service Manual
CORRECTIVE MAINTENANCE
5.2 DISASSEMBLY/REASSEMBLY
(Continued)
During Reassembly:
Align seal clip and snap it in place.
Latch (opened
half way)
Signature Edition® GOLD Infusion System, Models 7130/7131 and 7230/7231 5-31
Technical Service Manual
CORRECTIVE MAINTENANCE
5-32 Signature Edition® GOLD Infusion System, Models 7130/7131 and 7230/7231
Technical Service Manual
CORRECTIVE MAINTENANCE
5.2 DISASSEMBLY/REASSEMBLY
(Continued)
During Reassembly:
5.2.16 Removing Keypad Assembly
1. Fold back a small portion of protective
backing along right side of keypad.
CAUTION 2. Insert keypad flex cables into slot and
Remove the keypad only if it is defective. guide them through.
Removing a good keypad will ruin it.
3. Align right edge of keypad and affix.
Once aligned, position keypad and
1. Take a corner of keypad and peel it away remove remaining protective backing.
from front case. 4. Press around perimeter of keypad to
2. Remove any remaining residue from assure adhesion to front case and
surface before applying new keypad. prevention of fluid ingress.
5. Remove any protective covering from
front of keypad.
Signature Edition® GOLD Infusion System, Models 7130/7131 and 7230/7231 5-33
Technical Service Manual
CORRECTIVE MAINTENANCE
5-34 Signature Edition® GOLD Infusion System, Models 7130/7131 and 7230/7231
Technical Service Manual
CORRECTIVE MAINTENANCE
5.2 DISASSEMBLY/REASSEMBLY
(Continued)
Back 4-Bar
Latch Cable
Motor Cable
Ground Wires
Tie #2
XDCR Cable
Rotation Cable
Motor Cable
Tie #1
Signature Edition® GOLD Infusion System, Models 7130/7131 and 7230/7231 5-35
Technical Service Manual
CORRECTIVE MAINTENANCE
5.2 DISASSEMBLY/REASSEMBLY
(Continued)
Channel B Channel A
XDCR Cable
Latch Cable
Back 4-Bar
Motor Cable
Tie #4
Rotation Cable
Tie #5
Motor Cable
5-36 Signature Edition® GOLD Infusion System, Models 7130/7131 and 7230/7231
Technical Service Manual
CORRECTIVE MAINTENANCE
5.2 DISASSEMBLY/REASSEMBLY
(Continued)
Tamper Switch
Cable
Cable Clamp
(Ground Lead)
Cable Clamp
(Ground Lead)
Speaker Fan
Battery/Power Cable RS-232 Flex
Connector
Signature Edition® GOLD Infusion System, Models 7130/7131 and 7230/7231 5-37
Technical Service Manual
CORRECTIVE MAINTENANCE
Additional testing and calibration may be Spring Installation Tool Kit; ALARIS
required after certain repairs are completed. P/N 145098 (includes tool and
These tests are in addition to the Preventive instructions for use)
Maintenance tests. See Table 5-2, "Level of Perform a visual check of the mechanism
Testing Guidelines", for more information. prior to installing it into the front case. Each
time the instrument is opened, verify the
5.3.1 Power-On Self Test mechanism springs are properly positioned.
The power-on self test deals with 1. Verify all 12 followers are in place and
determining the proper operating condition that mechanism is flush to front case.
of the fully assembled instrument. The
instrument contains extensive self-testing 2. Rotate latch handle and check for full,
software. The self test is a final test. uninterrupted range of motion.
3. Verify seal is correctly in place.
The self test is initiated each time the
instrument is turned on. The instrument
continually tests itself during operation, as WARNINGS
well. An alarm or error message will appear
• Do not attempt to reinstall springs or
if there is a problem.
repair a mechanism assembly by any
When turning on the instrument, verify that means not specified in the following
the instrument beeps and that all display procedure. Springs can be damaged if
segments and LEDs flash. This confirms improperly installed, which can reduce
that the instrument has performed its self rate accuracy or lead to an "uncontrolled
tests and is operating correctly. flow" condition.
• Repair of a mechanism assembly
requires fixturing not available for field
use. Replace (do not repair) a
mechanism assembly that is found to
have an improperly installed spring after
completing the following procedure.
5-38 Signature Edition® GOLD Infusion System, Models 7130/7131 and 7230/7231
Technical Service Manual
CORRECTIVE MAINTENANCE
c. Verify whether or not spring is properly 6. If, after inspecting mechanism spring
installed onto back guide (white connections per steps 4 and 5, it has
colored part with a metal insert at been determined that mechanism needs
spring attachment). Continue with replacement:
next step. a. Replace mechanism assembly.
5. Use spring installation tool to ensure b. Return mechanism to ALARIS Medical
proper installation of spring to back guide Systems (refer to “General Contact
attachment point. Information” at beginning of this
a. Firmly press groove of installation tool manual).
over rib just below spring connection
point on back guide.
Signature Edition® GOLD Infusion System, Models 7130/7131 and 7230/7231 5-39
Technical Service Manual
CORRECTIVE MAINTENANCE
5-40 Signature Edition® GOLD Infusion System, Models 7130/7131 and 7230/7231
Technical Service Manual
CORRECTIVE MAINTENANCE
Use firm
pressure
to seat
spring
Spring connection to Back Guide
CAUTION
The spring installation tool is for use only at
the spring attachment to the back guide.
Signature Edition® GOLD Infusion System, Models 7130/7131 and 7230/7231 5-41
Technical Service Manual
CORRECTIVE MAINTENANCE
0 775
PSI
mmHg
Instrument
Stand Pressure
Gauge
Reservoir
5-42 Signature Edition® GOLD Infusion System, Models 7130/7131 and 7230/7231
Technical Service Manual
CORRECTIVE MAINTENANCE
Signature Edition® GOLD Infusion System, Models 7130/7131 and 7230/7231 5-43
Technical Service Manual
CORRECTIVE MAINTENANCE
5.3 TEST AND CALIBRATION (Continued) 3. Move AccuSlide® Flow Regulator clamp
up to open position.
5.3.6 Test Run Mode (Figure 5-37)
4. Fill lower portion of tubing with RTV. Do
The test run mode enables the instrument to not let RTV enter dome area.
run without fluid after being repaired. To run
5. Allow to dry (48 hours). Move
the instrument without fluid, perform the
AccuSlide® Flow Regulator clamp down
following:
until it "clicks" into closed position, and
1. Cut AccuSlide® Flow Regulator segment install test set into instrument.
out of standard set. Leave 2 inches of
6. Select rate and run for desired time
tubing at top and bottom.
period, for example 100 mL/h for 15
2. Cut membrane on backside of minutes.
AccuSlide® Flow Regulator clamp at
7. Remove test set.
bottom of pumping segment. Use knife
or other appropriate tool and cut an 'X'
into membrane to relieve pressure that
will build up during testing.
2” Segment
AccuSlide® Flow
Regulator
clamp (open)
AccuSlide® Flow
Place “X” cut here Regulator clamp
(on backside) (closed)
Dome (on
backside)
2” Segment
of RTV
5-44 Signature Edition® GOLD Infusion System, Models 7130/7131 and 7230/7231
Technical Service Manual
CORRECTIVE MAINTENANCE
NOTE: When performing the Hard Pressure a. Press lower left soft key 4 times to see
Calibration, take care not to press downward Count = value.
on the transducer. The pot is a one-turn, b. If Count = value is between 875 and
surface-mount pot with a rivet in the middle.
1275, go to start of Pressure Cal
Use a small screwdriver or scribe to push
sideways on the two slots in the pot. If the
screen. Verify Sensor = reading is
Count = value is 0 or 4095, this could between -80 and +30 mmHg after set
indicate a bad transducer and necessitate is removed (repeat for other channel if
replacement of the mechanism. necessary). If not, perform soft
pressure calibration.
1. Ensure warm-up of 1 hour minimum.
c. If Count = value is less than 875 or
2. Leave instrument on and connected to
more than 1275, perform hard cal for
AC power and battery.
this channel.
3. Follow procedure to disassemble
7. Adjust transducer pot for a Count = value
instrument.
of 1175±25 counts.
NOTE: There is one case screw inside the 8. Once transducer is adjusted for proper
battery compartment. reading, press ok to accept value.
4. Do not disconnect any cables. Adjust 9. Close instrument by setting bottom of
pole clamp so knob is facing downward. front case onto rear case, using case
Lay front case on desk/bench top. bosses for alignment. For a dual channel
5. Locate pot under motor and on backside instrument, route long ground wire into
of transducer (Figure 5-38). corner of case boss. Ensure ground
wires do not rub against mechanism.
Figure 5-38. Transducer Pot
10. Check for proper routing of cables.
11. Close case by following pattern and
tightening procedure outlined earlier in
this chapter.
12. Wait 30 minutes for transducer to warm
up again and check that compensated
value is still in range, as noted earlier.
13. Perform soft pressure cal procedure
(after 30-minute warm-up) as follows:
Motor a. Install 70ISS Press Cal Set.
Transducer
POT
Signature Edition® GOLD Infusion System, Models 7130/7131 and 7230/7231 5-45
Technical Service Manual
CORRECTIVE MAINTENANCE
5-46 Signature Edition® GOLD Infusion System, Models 7130/7131 and 7230/7231
Technical Service Manual
CORRECTIVE MAINTENANCE
This table provides minimum test requirements. If instrument operation is at all doubtful, perform
a complete Preventive Maintenance Inspection.
NOTE: The number in parentheses is the chapter that describes the test/action.
Tests to Perform →
(complete) (5)
(springs) (5)
Repair/Replacement of ↓
Battery •
Case • • • • • X • • X • X
Keyboard Assy. • • • X •
LCD Module, Lower • • • • X •
LCD Module, Main • • • • X •
LED Module • • • • X •
Main Board • • • X • • • A/R • • • • X
Mechanism • • • • • • • A/R • • • X
Panel Lock Keypad • • • X •
Power Switcher X • • • •
Software • • • X • • • A/R • • • • X
Speaker • • • •
Other repairs • X • • • • A/R X • X
Miscellaneous: ↓↓
Miscellaneous:
Instrument Dropped • • • • • • • • • X
New Instrument Checkout • • • • • • • X X X
Rate Accuracy failure
after Rate Cal.
N/A • •
No Fault Found
(instrument opened) • X • • • • X • X
No Fault Found
(instrument not opened) • • • X
Signature Edition® GOLD Infusion System, Models 7130/7131 and 7230/7231 5-47
Technical Service Manual
CORRECTIVE MAINTENANCE
T H I S PA G E
I N T E N T I O N A L LY
LEFT BLANK
5-48 Signature Edition® GOLD Infusion System, Models 7130/7131 and 7230/7231
Technical Service Manual
6 TROUBLESHOOTING
Chapter 6 — TROUBLESHOOTING
6.1 INTRODUCTION
This chapter contains descriptions of
WARNING possible operating errors and technical
Use extreme caution in servicing the problems that may be encountered during
instrument when connected to AC power. use of the instrument. Refer to this chapter
Hazardous voltages are present when AC before attempting to service the Signature
power is connected, regardless of the setting Edition® GOLD Infusion System.
of the power switch.
To facilitate troubleshooting, the instrument
will alarm and display prompts that direct
attention to the problem. Alarms that relate
CAUTION to operating problems are listed in the
Before disconnecting the battery, applicable Signature Edition® GOLD Infusion
disassembling, troubleshooting or testing the System Directions for Use (DFU).
instrument, record the alarm history. Refer to
“Viewing Alarm or Event History (Event Log)”
section of this chapter.
PROBLEM REMEDY
Accuracy Verification 1. Try new 80VCS Rate Cal set and repeat test a second time.
Fails 2. Check for mechanical damage and proper loading of administration
sets.
3. Perform Rate Calibration.
4. Replace mechanism.
5. Return to factory.
Air-in-Line 1. Air in set. Remove air or press RESET to move air through line.
2. Change AIL setting.
3. Clean AIL transducer/receiver.
4. Check connector.
5. Replace AIL transmitter.
6. Replace AIL receiver.
7. Return to factory.
Signature Edition® GOLD Infusion System, Models 7130/7131 and 7230/7231 6-1
Technical Service Manual
TROUBLESHOOTING
PROBLEM REMEDY
Backup Speaker Activated 1. Discharge capacitor C146 (C179) by placing jumper across capacitor
while handling main PCB terminals.
Batt Refresh A charge/discharge/recharge cycle was initiated. This can occur:
1. When battery is disconnected and reconnected to AC power.
2. By entering 0.0Ah in rated capacity, connecting to AC, then change rated
capacity back to 1.3Ah.
6-2 Signature Edition® GOLD Infusion System, Models 7130/7131 and 7230/7231
Technical Service Manual
TROUBLESHOOTING
PROBLEM REMEDY
LCD Contrast is Dark or Light 1. Change contrast setting to 80, tilt unit to help view or follow information in
Diagnostics Mode.
2. Replace main LCD.
3. Replace main PCB.
4. Return to factory.
Mechanical Leak Test Fails 1. Check mechanism for damage and proper mounting into snap fittings.
2. Replace mechanism.
3. Return to factory.
No Power 1. Wait 15 seconds after connecting to AC power.
2. Check/replace battery.
3. Replace power supply board (AC off line switcher).
4. Replace keypad.
5. Replace main PCB.
6. Return to factory.
Press Restart In alarm history, this indicates infusion reached occlusion alarm point in
pressure or resistance modes, but cleared within 40 seconds.
Pressure Calibration (Soft) 1. Check setup and warmup time of at least 30 minutes.
Fails 2. Retry with new 70ISS set.
3. Perform pressure calibration (hard).
4. Replace mechanism.
5. Replace main PCB.
6. Return to factory.
Pressure Verification Fails 1. Check for damage and proper loading of 70ISS set.
2. Perform pressure calibration (soft).
3. Perform pressure calibration (hard).
4. Replace mechanism.
5. Return to factory.
Program Lost Normal condition if battery disconnected or in a low state of charge.
Reprogram settings and continue operation.
Resis Restart In alarm history, this indicates infusion reached 100% resistance or alarm point,
but cleared within 40 seconds.
Self Check Due 1. Install set, wait one minute and remove set.
2. Check latch sensor for proper operation.
3. Check pressure calibration (soft).
4. Perform Hard Pressure Cal.
5. Replace mechanism.
6. Return to factory.
Signature Edition® GOLD Infusion System, Models 7130/7131 and 7230/7231 6-3
Technical Service Manual
TROUBLESHOOTING
PROBLEM REMEDY
Set Out Alarms 1. Try another set.
2. Lower instrument closer to patient level.
3. Perform pressure calibration (soft).
4. Perform pressure calibration (hard).
5. Replace mechanism.
6. Return to factory.
Set Sensor Check Fails 1. Try a second set.
2. Perform pressure verification.
SW APPL Software error. Turn instrument off and back on. If problem reoccurs, return
instrument to factory or replace main PCB.
SW Shutdown Battery voltage reached 10.2 volts, or five minutes elapsed since low battery
alarm, and battery manager told CPU to turn off instrument.
Charge or replace battery as needed.
6-4 Signature Edition® GOLD Infusion System, Models 7130/7131 and 7230/7231
Technical Service Manual
TROUBLESHOOTING
Signature Edition® GOLD Infusion System, Models 7130/7131 and 7230/7231 6-5
Technical Service Manual
TROUBLESHOOTING
6-6 Signature Edition® GOLD Infusion System, Models 7130/7131 and 7230/7231
Technical Service Manual
TROUBLESHOOTING
Signature Edition® GOLD Infusion System, Models 7130/7131 and 7230/7231 6-7
Technical Service Manual
TROUBLESHOOTING
6-8 Signature Edition® GOLD Infusion System, Models 7130/7131 and 7230/7231
Technical Service Manual
TROUBLESHOOTING
WD MTR OFF ERROR Watchdog couldn’t disable motor (start-up). 1. Replace main PCB.
2. Return to factory.
Signature Edition® GOLD Infusion System, Models 7130/7131 and 7230/7231 6-9
Technical Service Manual
TROUBLESHOOTING
The following error codes relate to codes displayed in the lower LCD. These codes are not stored in
memory. The steps in the “Remedy” column are listed in the sequence in which the specified actions
are to be taken.
6-10 Signature Edition® GOLD Infusion System, Models 7130/7131 and 7230/7231
Technical Service Manual
TROUBLESHOOTING
Signature Edition® GOLD Infusion System, Models 7130/7131 and 7230/7231 6-11
Technical Service Manual
TROUBLESHOOTING
A B OPTIONS 7 8 9
. 0 Clear Enter
PM Setup D
History 2 6. Press ok to accept change and return to
Set Time
beginning of D2 page.
Battery Status
page A B OPTIONS
PM Reminder= On D
PM Interval = 52 wks 2
PM Due= 52 wks
2. Press PM Setup soft key.
Reset PM Due
3. Press and release PM Reminder soft
undo ok
key to cycle between On and Off.
A B OPTIONS
PM Reminder= On D
PM Interval = 52 wks 2
PM Due= 52 wks
Reset PM Due
undo ok
6-12 Signature Edition® GOLD Infusion System, Models 7130/7131 and 7230/7231
Technical Service Manual
TROUBLESHOOTING
6.2.3 Viewing Alarm or Event History • History Log will keep 1500 entries in
(Event Log) memory. Entry 0001 will be most
recent and entry 1500 will be oldest.
No More Events will be displayed if
A B OPTIONS less than 1500 events are recorded.
1. From D2 page, press History soft key. NOTE: If AC and battery power are
disconnected from the instrument, event
• For mode/state changes, first line will
history may be lost.
read Misc. Inst. (Miscellaneous
Instruction). Key Inst. will be
displayed to indicate a key/switch
activation. Alarms and errors may be
channel-specific or instrument related,
which will also appear on first line.
• Second line will give month, day, year,
hour and minute that entry/event took
place.
• Third line will show event/entry that
occurred.
Signature Edition® GOLD Infusion System, Models 7130/7131 and 7230/7231 6-13
Technical Service Manual
TROUBLESHOOTING
6-14 Signature Edition® GOLD Infusion System, Models 7130/7131 and 7230/7231
Technical Service Manual
TROUBLESHOOTING
Display Examples
A B OPTIONS A B OPTIONS
A B OPTIONS A B OPTIONS
Signature Edition® GOLD Infusion System, Models 7130/7131 and 7230/7231 6-15
Technical Service Manual
TROUBLESHOOTING
6.2 DIAGNOSTICS MODE (Continued) • Year setting is four digits and will
accept year entries of 2000 and
6.2.4 Setting Time (and Date) beyond. Default date and time is
Jan 1, 1970 00:00
A B OPTIONS • Software clock may lose up to 3
minutes per month. Reset time as
PM Setup D part of Preventive Maintenance or
History 2 more often, as desired.
Set Time
Battery Status • When clock is reset for current
date/time, previous logs will not be
page adjusted retroactively, but will remain
the same.
• Clock will NOT automatically adjust
1. On D2 page, press Set Time soft key. for time changes, such as, daylight
savings.
2. Press Month soft key to highlight. Press
Month soft key as needed to get desired 6. Press ok to accept change and return to
month. beginning of D2 page. Press Undo soft
key to reinstate time/date.
A B OPTIONS
NOTE: Information stored in the Flash
Month Jun D EEPROM will not be lost when Version
Date 5 2 4.08 software is flashed into the
Year 2004 instrument. This includes Rate
Calibration, Pressure Calibration, ID
Time 15 hrs 49 min number (serial number), time/date and
Undo ok PM data.
6-16 Signature Edition® GOLD Infusion System, Models 7130/7131 and 7230/7231
Technical Service Manual
TROUBLESHOOTING
Signature Edition® GOLD Infusion System, Models 7130/7131 and 7230/7231 6-17
Technical Service Manual
TROUBLESHOOTING
OPTIONS
An ID number may be useful for asset
A B tracking, maintenance records, etc.
Batt. Mgr, 3.06 D
Voltage : 14.4volts 2 A B OPTIONS
Meas Cap : 1.8 AH
Rated Cap: 1.3 AH DC Voltages D
Set ID Number 3
undo ok Run Times
Self-Check Timer
page
4. Press ok to accept change and return to
beginning of D2 page.
6-18 Signature Edition® GOLD Infusion System, Models 7130/7131 and 7230/7231
Technical Service Manual
TROUBLESHOOTING
DC Voltages D
Set ID Number 3
Run Times
Self-Check Timer
page
Signature Edition® GOLD Infusion System, Models 7130/7131 and 7230/7231 6-19
Technical Service Manual
TROUBLESHOOTING
6.2 DIAGNOSTICS MODE (Continued) Please Eject the Set," this 12-week interval
has elapsed. If that occurs:
6.2.10 Viewing Self-Check Timer (Continued)
1. Turn instrument on and install set.
2. Verify interval is set to 12 weeks.
2. Wait one minute.
A B OPTIONS 3. Remove set and watch screen for a
message that self-check is in progress.
Self-Check D
Interval = 12 weeks 3 4. When test is done, screen will go back to
parameter page, and instrument
operation can continue as desired.
A B OPTIONS
6-20 Signature Edition® GOLD Infusion System, Models 7130/7131 and 7230/7231
Technical Service Manual
TROUBLESHOOTING
M Flag = Light D
Latch = Open 4
AIL = Air A 3. To change Rate Cal # (%), press Rate
ECD = Not Installed soft key two times.
return • Instrument will beep but display will
not change.
4. Press Cal # soft key to highlight and
select for editing.
3. Press return soft key to return to D4
page. 5. Change Cal # based on results of
calibration rate run (refer to “Rate
6.2.12 Viewing/Changing Rate Calibration Calibration Procedure” section in
Information “Preventive Maintenance” chapter).
Use the following procedure to view or
change the Rate Cal # setting.
1. Advance to D4 page.
Signature Edition® GOLD Infusion System, Models 7130/7131 and 7230/7231 6-21
Technical Service Manual
TROUBLESHOOTING
page
Rate D
Calibration 4
Cal # = 0.0 %
Rate= 174.0 mcL/rev
2. Press Test Main LCD soft key.
undo +/- ok • Test consists of Main LCD alternating
between normal and reverse video
every 2 seconds.
• All LEDs also flash on and off at a
6. Press Enter then ok to accept change or 2-second rate.
press undo to return to old cal number.
A B OPTIONS
NOTES:
• Information stored in the EEPROM will
MAIN D
not be lost when Version 4.08 software
LCD 5
is flashed into the instrument. This TEST
includes Rate calibration, Pressure
Calibration, ID number (serial number),
return
time/date and PM data.
• Once the Rate Cal # is set to 0.0 and
accepted, the instrument will need to
be run for at least 2 seconds before
3. Press return soft key to return to D5
the Rate Cal # can be changed to a
non-zero value. If not, the instrument
page.
will display “Do Rate Accuracy Test at
0%”. The instrument will not allow one 6.2.14 Testing Aux (Lower) LCD
non-zero value to be changed to
another non-zero value. A B OPTIONS
6-22 Signature Edition® GOLD Infusion System, Models 7130/7131 and 7230/7231
Technical Service Manual
TROUBLESHOOTING
A B OPTIONS
Switch Pressed = D
Left Soft Key 2 5
A B OPTIONS
AUXILIARY D
LCD 5 return
TEST
return
3. Press return soft key to return to D5
page.
2. Press return soft key to return to D5 6.2.16 Changing Main LCD Contrast
page.
A B OPTIONS
Test Main LCD D
Test Aux. LCD 5
Test Main LCD D Test Switches
Test Aux. LCD 5 Main LCD Contrast
Test Switches
page
Main LCD Contrast
page
Signature Edition® GOLD Infusion System, Models 7130/7131 and 7230/7231 6-23
Technical Service Manual
TROUBLESHOOTING
4 5 6
Cal A Pressure D
7 8 9
DAC A Settings 6
. 0 Clear Enter
Cal B Pressure
DAC B Settings
3. Use numeric keypad to enter desired
LCD contrast (contrast range is 1 to 256). page
Press Enter.
6-24 Signature Edition® GOLD Infusion System, Models 7130/7131 and 7230/7231
Technical Service Manual
TROUBLESHOOTING
6.2 DIAGNOSTICS MODE (Continued) c. Install a standard set and close latch.
Verify reading is over 170.
6.2.17 Calibrating Channel Pressure
d. Verify Sensor = reading is in -80 to
(Continued)
+30 mmHg range. If instrument will
3. Both "0" and "500" should display a not soft cal, perform Hard Pressure
number and read Pass. Sensor= should Cal Procedure (see “Corrective
have a value between -80 and +30 Maintenance” chapter).
(mmHgs), with no set installed. If Xs
appear or sensor reading is out of range, NOTE: Information stored in the
instrument will require soft pressure EEPROM will not be lost when Version
calibration, as follows (for setup, 4.08 software is flashed into the
reference “Preventive Maintenance” instrument. This includes Rate
Calibration, Pressure calibration, ID
chapter):
number (serial number), time/date and
PM data.
A B OPTIONS
Signature Edition® GOLD Infusion System, Models 7130/7131 and 7230/7231 6-25
Technical Service Manual
TROUBLESHOOTING
Calibrate Pressure D
Calibrate Pressure D
0 mmHg=1175 Pass 0 mmHg=1175 Pass 6
6
500 mmHg=2400 Pass A 500 mmHg=2400 Pass A
Sensor=+001 Pass Offset=-10 mmHg
ok ok
6-26 Signature Edition® GOLD Infusion System, Models 7130/7131 and 7230/7231
Technical Service Manual
TROUBLESHOOTING
Signature Edition® GOLD Infusion System, Models 7130/7131 and 7230/7231 6-27
Technical Service Manual
TROUBLESHOOTING
page
Pressure System D
Boot Rev: 01.02 7
A B OPTIONS
page
Diagnostics Mode D
Diagnostics 1
Communications
Active
2. Press Auto Zero soft key to highlight
settings. Press Auto Zero soft key again
to change between On and Off.
A B OPTIONS
A B OPTIONS
A B OPTIONS
3. Press ok to accept change and return to
main D7 page. Press undo to return to PM Reminder= On D
previous setting. PM Interval = 52 wks 2
PM Due= 52 wks
NOTE: The Boot Rev: xxxx refers to the Reset PM Due
flash EPROM boot program and cannot
be changed. undo ok
6-28 Signature Edition® GOLD Infusion System, Models 7130/7131 and 7230/7231
Technical Service Manual
TROUBLESHOOTING
A B OPTIONS
A B OPTIONS
DC VOLTAGES D
ADREF = 4.100 v 3
0001 Misc. Chan. A D
JUN 5, 2001 15:48 VBACKUP= 5.000 v
2
CHANNEL OFF VCC= 5.000 v
return
return page
OPTIONS
A B
A B OPTIONS
Instrument D
Identification 3
Month Jun D
Date 5 Number =
2
12345678
Year 2004
Time 15 hrs 49 min undo ok
Undo ok
A B OPTIONS
A B OPTIONS
Run Times D
Battery : 20 hrs 3
Batt. Mgr, 3.06 D
Total : 189 hrs
Voltage : 14.4volts 2
Clear Battery Time
Meas Cap : 1.8 AH
Rated Cap: 1.3 AH undo ok
undo ok
A B OPTIONS
A B OPTIONS Self-Check D
Interval = 12 weeks 3
DC Voltages D
Set ID Number 3
Run Times
Self-Check Timer undo ok
page
Signature Edition® GOLD Infusion System, Models 7130/7131 and 7230/7231 6-29
Technical Service Manual
TROUBLESHOOTING
A B OPTIONS
A B OPTIONS
MAIN D
Test A Sensors D LCD 5
Cal A Rate 4 TEST
Test B Sensors
Cal B Rate return
page
A B OPTIONS
A B OPTIONS AUXILIARY D
LCD 5
M Flag = Light D TEST
Latch = Open 4
AIL = Air A
ECD = Not Installed return
return
A B OPTIONS
A B OPTIONS
A B OPTIONS
Main LCD Contrast = D
Test Main LCD D 127 5
Test Aux. LCD 5
Test Switches
Main LCD Contrast ok
undo
page
6-30 Signature Edition® GOLD Infusion System, Models 7130/7131 and 7230/7231
Technical Service Manual
TROUBLESHOOTING
A B OPTIONS
Cal A Pressure D
DAC A Settings 6
Cal B Pressure
DAC B Settings
page
A B OPTIONS
Calibrate Pressure D
0 mmHg=1175 Pass 6
500 mmHg=2400 Pass A
Sensor=+001 Pass
ok
A B OPTIONS
Pressure System D
Boot Rev: 01.02 7
page
A B OPTIONS
undo ok
Signature Edition® GOLD Infusion System, Models 7130/7131 and 7230/7231 6-31
Technical Service Manual
TROUBLESHOOTING
T H I S PA G E
I N T E N T I O N A L LY
LEFT BLANK
6-32 Signature Edition® GOLD Infusion System, Models 7130/7131 and 7230/7231
Technical Service Manual
7 ILLUSTRATED PARTS
BREAKDOWN
Chapter 7 — ILLUSTRATED PARTS BREAKDOWN
7.1 INTRODUCTION
The illustrated parts breakdown for the
instrument is divided into major assemblies
and individual parts.
7.2 ILLUSTRATIONS
The exploded views serve as visual aids for
identifying the parts of each assembly. If a
part/assembly is identified with an item
number (appearing in a bubble), that number
corresponds to the item number on the parts
list. If a part/assembly is not identified with
an item number, it is part of a higher
assembly or kit.
Signature Edition® GOLD Infusion System, Models 7130/7131 and 7230/7231 7-1
Technical Service Manual
ILLUSTRATED PARTS BREAKDOWN
7-2 Signature Edition® GOLD Infusion System, Models 7130/7131 and 7230/7231
Technical Service Manual
ILLUSTRATED PARTS BREAKDOWN
NOTE: If model is not specified, the part can be used on all versions.
* 7130X = 7130B, 7130D, 7130E 120V Version; 7131X = 7131A, 7131B 220V Version;
7230X = 7230B, 7230D, 7230E 120V Version.
Signature Edition® GOLD Infusion System, Models 7130/7131 and 7230/7231 7-3
Technical Service Manual
ILLUSTRATED PARTS BREAKDOWN
7-4 Signature Edition® GOLD Infusion System, Models 7130/7131 and 7230/7231
Technical Service Manual
ILLUSTRATED PARTS BREAKDOWN
406
410
409
Signature Edition® GOLD Infusion System, Models 7130/7131 and 7230/7231 7-5
Technical Service Manual
ILLUSTRATED PARTS BREAKDOWN
90
7 PL 464 435
2 PL 305 300
437
463
399
414
250 413
403
420
7-6 Signature Edition® GOLD Infusion System, Models 7130/7131 and 7230/7231
Technical Service Manual
ILLUSTRATED PARTS BREAKDOWN
90
435
7 PL 464
300
437 2 PL
2 PL 305
463
414
399
413
403
250
420
Signature Edition® GOLD Infusion System, Models 7130/7131 and 7230/7231 7-7
Technical Service Manual
ILLUSTRATED PARTS BREAKDOWN
NOTE: If model is not specified, the part can be used on all versions.
7-8 Signature Edition® GOLD Infusion System, Models 7130/7131 and 7230/7231
Technical Service Manual
ILLUSTRATED PARTS BREAKDOWN
100
463
0 700
250
0 550
0 400
201
0 800
416 2 PL
Signature Edition® GOLD Infusion System, Models 7130/7131 and 7230/7231 7-9
Technical Service Manual
ILLUSTRATED PARTS BREAKDOWN
100 2 PL
463
250
0 550
0 700
0 800
201 2 PL
416 2 PL
0 400
7-10 Signature Edition® GOLD Infusion System, Models 7130/7131 and 7230/7231
Technical Service Manual
ILLUSTRATED PARTS BREAKDOWN
147507-100 AIL Repair Kit (kit includes AIL Button, AIL Arm and installation instructions)
203 136836 AIL Button (Receiver) 1
204 141734-100 AIL Arm (Transmitter) 1
383 136255 Mechanism Seal 1
392 141144 Transducer Cable 1
397 133228 AIL Spring
404 136721 Seal Clip 1
405 136656-000 AIL Gear
383
404
392
White Lead
204
Black Lead
405
397
Note black lead threads
203 through faceplate hook
Signature Edition® GOLD Infusion System, Models 7130/7131 and 7230/7231 7-11
Technical Service Manual
ILLUSTRATED PARTS BREAKDOWN
7-12 Signature Edition® GOLD Infusion System, Models 7130/7131 and 7230/7231
Technical Service Manual
ILLUSTRATED PARTS BREAKDOWN
0 500
2 PL 462
4 PL 465 437
300
427 2 PL
2 PL 417
0 600
2 PL 461 423
0 610 424
477
430
140 2 PL
386
469
466
463
365
2 PL
0 350
401 408 466
479
416 0 410 Optional
2 PL
Signature Edition® GOLD Infusion System, Models 7130/7131 and 7230/7231 7-13
Technical Service Manual
ILLUSTRATED PARTS BREAKDOWN
0 500
462 2 PL
4 PL 465 437
300
427 2 PL
2 PL 417
0 600
2 PL 461 423
0 61 0
424
477 430
140 2 PL
386
469
466
463
365
2 PL
0 350
401 408 466
479
416
0 410 Optional
2 PL
7-14 Signature Edition® GOLD Infusion System, Models 7130/7131 and 7230/7231
Technical Service Manual
ILLUSTRATED PARTS BREAKDOWN
147950-100 Battery Pole Clamp and Knob Kit (includes all items below) 1
421 144334 C-Clamp Assembly Kit (new version) 1
Kit consists of Pole Clamp, Retaining Ring and Washer
425 140408-000 Pole Clamp Lock 1
426 142793-000 Pole Clamp Knob w/Lead Screw (newer version) 1
428 305317 Wave Washer 1
431 139900 Pole Clamp Tip 1
432 305414 Spiral Roll Pin, .078X.3125,SS 1
471 305436 Compression Spring 1
483 320154 Retaining Ring 1
484 320155 Flat Washer 1
Old version brass lead screw Brass knob lead screw, P/N: 140380 (limited to existing inventory)
New version steel lead screw Steel knob lead screw, P/N: 142793-000
Old C-clamp (no sleeve) Pole clamp and knob kit, P/N: 147950-100
Signature Edition® GOLD Infusion System, Models 7130/7131 and 7230/7231 7-15
Technical Service Manual
ILLUSTRATED PARTS BREAKDOWN
426
426 432
432 431
471
421
484
425
428
483
7-16 Signature Edition® GOLD Infusion System, Models 7130/7131 and 7230/7231
Technical Service Manual
ILLUSTRATED PARTS BREAKDOWN
Signature Edition® GOLD Infusion System, Models 7130/7131 and 7230/7231 7-17
Technical Service Manual
ILLUSTRATED PARTS BREAKDOWN
NOTE: If model is not specified, the part can be used on all versions.
7-18 Signature Edition® GOLD Infusion System, Models 7130/7131 and 7230/7231
Technical Service Manual
ILLUSTRATED PARTS BREAKDOWN
610
625 611
630
608 603 606
654 604
627
613
623 605
(Ground post)
626
Section A-A
(Some items not shown for clarity)
Signature Edition® GOLD Infusion System, Models 7130/7131 and 7230/7231 7-19
Technical Service Manual
ILLUSTRATED PARTS BREAKDOWN
T H I S PA G E
I N T E N T I O N A L LY
LEFT BLANK
7-20 Signature Edition® GOLD Infusion System, Models 7130/7131 and 7230/7231
Technical Service Manual
ALARIS Medical Systems, Inc.
10221 Wateridge Circle
San Diego, California 92121 U.S.A.
Mail:
P.O. Box 85335
San Diego, California 92186-5335
U.S.A.
Authorized EU Representative
ALARIS Medical UK Ltd.
The Crescent, Jays Close
Basingstoke, Hampshire
RG22 4BS, UK
ALARIS®, ALARIS Medical Systems®, AccuSlide®, Dynamic Monitoring®, Guardrails®, and Signature Edition® are trademarks and registered
trademarks of ALARIS Medical Systems, Inc.
10010812 ©1997, 2004 ALARIS Medical Systems, Inc. All rights reserved.