Isolera Spektra TN Web Us
Isolera Spektra TN Web Us
Isolera Spektra TN Web Us
User Manual
Safety and Document Conventions
The Isolera system should only be operated and maintained by trained individuals. Please read this manual
carefully before working with the system. To guarantee safe and effective operation it is absolutely necessary to
follow all of the instructions in this user manual. All information is believed to be complete and accurate at the
time of publication, but is subject to change.
Pay close attention to text with a NOTE or WARNING heading. A NOTE gives information that is crucial to the
operation being described. A WARNING indicates important information that will prevent harm to personnel
and/or damage to the system.
NOTE
Note text appears in the manual to provide additional or crucial information about
the current topic. It may indicate a situation in which the system may not perform
as expected unless specific guidelines are followed.
WARNING
The Warning note indicates a hazard that could result in serious injury to the user
and/or damage to the equipment unless good laboratory practices (GLP) and/or
manufacturer's guidelines are followed.
Trademark Acknowledgement
The following trademarks are owned by Biotage® AB: Biotage, Biotage ZIP, Advancer, Advancer Kilobatch,
Advancer 350, AFFINILUTE, Endeavor, EVOLUTE, ExploraSep, Firefly design ( ), FLASH+, FlashMaster,
FlashVac, Horizion, HPFC, HP-SIL, HP-Sphere, Initiator, Initiator+, Isolera, Isolera Four, Isolera LS, Isolera One,
Isolera Prime, Isolera Spektra, Isolera Spektra Four, Isolera Spektra LS, Isolera Spektra One, ISOLUTE, IST,
IST design ( ), KILOPREP, KP-C18-HS, KP-C18-WP, KP-C4-WP, KP-NH, KP-Sil, MIP4SPE, MIP Rule of 6,
MIP[4]Process, MIP[4]Proteins, PathFinder, PRESSURE+ 48, PRESSURE+ 96, RapidTrace, RapidTrace +,
Resolux, Samplet, SIM, SNAP, SNAP Ultra, SP1, SP4, SPE Dry 96, SPE Dry 96 Dual, SP Wave, Syro Wave,
TurboVap, VacMaster, V-10, ZIF, ZIF-SIM, ZIP-Sphere, and 1-Point Support.
Other product and company names mentioned herein may be trademarks or registered trademarks and/or
service marks of their respective owners, and are used only for explanation and to the owners' benefit, without
intent to infringe.
Copyright
Biotage Sweden AB reserves the right to make changes to the information contained herein without prior notice.
No part of this manual may be reproduced or transmitted in any form or by any means, electronic or
mechanical, for any purpose, without the expressed written permission of Biotage Sweden AB.
© Copyright 2007–2012 Biotage Sweden AB. All rights reserved.
Page ii
Table of Contents
1: Description and Specifications - - - - - - - - - - - - - - - - - - - - - - - - --- - --- --- - 1-1
General System Description - - - - - - - - - - - - - - - - - - - - - - - - - - 1-1
Basic System Components - - - - - - - - - - - - - - - - - - - - - - - - - 1-2
System Configurations - - - - - - - - - - - - - - - - - - - - - - - - - - - 1-2
Programmed Functions - - - - - - - - - - - - - - - - - - - - - - - - - - - 1-3
User Documentation - - - - - - - - - - - - - - - - - - - - - - - - - - - - 1-3
Accessories - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 1-3
Software Description - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 1-7
Main Menu - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 1-7
Data Administration Mode - - - - - - - - - - - - - - - - - - - - - - - - - 1-7
System Mode - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 1-8
Chemistry Mode - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 1-9
Method Tab (Chemistry Mode)- - - - - - - - - - - - - - - - - - - - - - - - 1-11
Status Tab (Chemistry Mode) - - - - - - - - - - - - - - - - - - - - - - - - 1-11
Results Tab (Chemistry Mode)- - - - - - - - - - - - - - - - - - - - - - - - 1-14
Solvents Tab (Chemistry Mode) - - - - - - - - - - - - - - - - - - - - - - - 1-15
Isolera Remote Viewer - - - - - - - - - - - - - - - - - - - - - - - - - - - 1-16
Specifications - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 1-17
General System Specifications- - - - - - - - - - - - - - - - - - - - - - - - 1-17
Operational Performance Specifications - - - - - - - - - - - - - - - - - - - - 1-18
Solvent Specifications - - - - - - - - - - - - - - - - - - - - - - - - - - - 1-19
Chemical Resistance of the Peristaltic Pump Tube (Isolera LS) - - - - - - - - - - 1-20
2: Data Administration - - - - - - - - - - - - - - --- ---- --- ---- --- --- - --- --- - 2-1
Log into the Data Administration Mode - - - - - - - - - - - - - - - - - - - - - - 2-1
Administrate the Cartridge List - - - - - - - - - - - - - - - - - - - - - - - - - 2-2
Add a Cartridge Type - - - - - - - - - - - - - - - - - - - - - - - - - - - 2-2
Delete an Unused Cartridge Type - - - - - - - - - - - - - - - - - - - - - - 2-2
Edit a Cartridge Type- - - - - - - - - - - - - - - - - - - - - - - - - - - - 2-3
Select Cartridge Types to List - - - - - - - - - - - - - - - - - - - - - - - - 2-3
Administrate the Method List - - - - - - - - - - - - - - - - - - - - - - - - - - 2-3
Delete Methods - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 2-3
Export Methods - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 2-4
Import Methods - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 2-4
Set Default Methods - - - - - - - - - - - - - - - - - - - - - - - - - - - - 2-4
Administrate the Rack List - - - - - - - - - - - - - - - - - - - - - - - - - - - 2-5
Add a Rack Type - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 2-5
Delete an Unused Rack Type - - - - - - - - - - - - - - - - - - - - - - - - 2-6
Edit a Rack Type - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 2-6
Select Rack Types to List - - - - - - - - - - - - - - - - - - - - - - - - - - 2-6
Administrate the Result List - - - - - - - - - - - - - - - - - - - - - - - - - - - 2-6
Export and Delete Result Records - - - - - - - - - - - - - - - - - - - - - - 2-7
Clean Up the System’s Database- - - - - - - - - - - - - - - - - - - - - - - 2-7
Page iii
Administrate the Solvent List - - - - - - - - - - - - - - - - - - - - - - - - - - 2-8
Add a Solvent - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 2-8
Delete an Unused Solvent - - - - - - - - - - - - - - - - - - - - - - - - - 2-8
Edit a Solvent - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 2-9
Select Solvents to List - - - - - - - - - - - - - - - - - - - - - - - - - - - 2-9
Administrate the User List - - - - - - - - - - - - - - - - - - - - - - - - - - - 2-9
Add a User Account - - - - - - - - - - - - - - - - - - - - - - - - - - - - 2-9
Change the Password or Privilege - - - - - - - - - - - - - - - - - - - - - - 2-10
Delete a User Account - - - - - - - - - - - - - - - - - - - - - - - - - - - 2-10
Reset the Number of Performed Runs - - - - - - - - - - - - - - - - - - - - 2-10
3: System Settings - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - --- - --- --- - 3-1
Log into the System Mode - - - - - - - - - - - - - - - - - - - - - - - - - - - 3-1
Change Detector Settings - - - - - - - - - - - - - - - - - - - - - - - - - - - - 3-2
Enable or Disable the Internal Detector - - - - - - - - - - - - - - - - - - - - 3-2
Connect and Enable an External Detector - - - - - - - - - - - - - - - - - - - 3-2
Disable an External Detector - - - - - - - - - - - - - - - - - - - - - - - - 3-3
Enable or Disable the Leak Detector - - - - - - - - - - - - - - - - - - - - - 3-3
Change General Settings - - - - - - - - - - - - - - - - - - - - - - - - - - - - 3-4
Set the Date and Time - - - - - - - - - - - - - - - - - - - - - - - - - - - 3-4
Enable or Disable Mouse Pointer - - - - - - - - - - - - - - - - - - - - - - - 3-4
Set the Pressure Unit - - - - - - - - - - - - - - - - - - - - - - - - - - - 3-4
Set the Language Used in the Chemistry Mode- - - - - - - - - - - - - - - - - 3-4
Enable the λ-All Detection Mode and TLC to Step Gradient Editor - - - - - - - - - 3-5
Configure a Network Connection - - - - - - - - - - - - - - - - - - - - - - - - - 3-5
Change Report Settings - - - - - - - - - - - - - - - - - - - - - - - - - - - - 3-6
Set Up a Printer and Auto Print of Reports - - - - - - - - - - - - - - - - - - 3-6
Set Up File Sharing and Auto Save of Reports - - - - - - - - - - - - - - - - - 3-7
Set Up Auto Send of Reports - - - - - - - - - - - - - - - - - - - - - - - - 3-8
Change Runtime Settings - - - - - - - - - - - - - - - - - - - - - - - - - - - - 3-8
Enable or Disable Audible Alarm - - - - - - - - - - - - - - - - - - - - - - - 3-8
Enable or Disable Automatic Rack Allocation- - - - - - - - - - - - - - - - - - 3-8
Enable or Disable Peak Mode - - - - - - - - - - - - - - - - - - - - - - - - 3-9
Enable or Disable Auto Extend- - - - - - - - - - - - - - - - - - - - - - - - 3-9
Enable or Disable Monitoring of Reservoirs - - - - - - - - - - - - - - - - - - 3-9
Enable or Disable Run Requirement Estimation - - - - - - - - - - - - - - - - 3-10
Specify How Flushes Are Performed - - - - - - - - - - - - - - - - - - - - - 3-10
Configure Collection and Fractionation - - - - - - - - - - - - - - - - - - - - 3-11
Maintenance - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 3-11
Export the System Configuration and Logs - - - - - - - - - - - - - - - - - - 3-11
Restore the System Configuration - - - - - - - - - - - - - - - - - - - - - - 3-11
Back Up and Restore the System’s Database - - - - - - - - - - - - - - - - - 3-12
Calibrate the Touch Screen - - - - - - - - - - - - - - - - - - - - - - - - - 3-12
Calibrate the Fraction Collector - - - - - - - - - - - - - - - - - - - - - - - 3-13
Calibrate the Internal Detector - - - - - - - - - - - - - - - - - - - - - - - 3-13
Perform an Intensity Scan of the Internal Detector - - - - - - - - - - - - - - - 3-13
Clean the Collect Valve - - - - - - - - - - - - - - - - - - - - - - - - - - - 3-13
Release Stuck Check Valves- - - - - - - - - - - - - - - - - - - - - - - - - 3-14
View and Reset the Usage Statistics - - - - - - - - - - - - - - - - - - - - - 3-14
Page iv
4: Operation - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - --- --- - 4-1
Start Up the System - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 4-1
Shut Down the System - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 4-1
Control the UV Lamp - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 4-2
Create, Open, and Edit Methods - - - - - - - - - - - - - - - - - - - - - - - - - 4-2
Create or Open a Method - - - - - - - - - - - - - - - - - - - - - - - - - - 4-2
Calculate Gradient from TLC Data - - - - - - - - - - - - - - - - - - - - - - 4-10
Create a Method Using the Method Wizard - - - - - - - - - - - - - - - - - - 4-13
Gradient Optimization - - - - - - - - - - - - - - - - - - - - - - - - - - - 4-13
Create a Method Using a Web Browser - - - - - - - - - - - - - - - - - - - - 4-14
Prepare and Run a Purification - - - - - - - - - - - - - - - - - - - - - - - - - - 4-15
Prepare the System for a Run - - - - - - - - - - - - - - - - - - - - - - - - 4-15
Run a Purification - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 4-19
Change the Processing Order - - - - - - - - - - - - - - - - - - - - - - - - - - 4-22
Monitor and Control a Purification - - - - - - - - - - - - - - - - - - - - - - - - 4-23
Monitor the Purification in Progress - - - - - - - - - - - - - - - - - - - - - 4-23
Bypass the Automatic UV Lamp Warm-up Period - - - - - - - - - - - - - - - - 4-23
End Initial Waste and Start Collecting Fractions - - - - - - - - - - - - - - - - 4-23
Collect Through the Waste Channel - - - - - - - - - - - - - - - - - - - - - 4-24
Start and End an Isocratic Segment - - - - - - - - - - - - - - - - - - - - - 4-24
Control a Manual Collection - - - - - - - - - - - - - - - - - - - - - - - - - 4-24
Edit and Manually Extend a Purification - - - - - - - - - - - - - - - - - - - - 4-24
Pause, End, or Abort a Purification - - - - - - - - - - - - - - - - - - - - - - 4-25
Unload a Purification - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 4-26
Flush a Cartridge with Air - - - - - - - - - - - - - - - - - - - - - - - - - - 4-26
Purge a Cartridge - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 4-27
Access Result Records - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 4-28
Search for Records - - - - - - - - - - - - - - - - - - - - - - - - - - - - 4-28
View, Print, and Save Reports on a USB Memory Device or the Network - - - - - - 4-29
Save Records on a USB Memory Device or the Network - - - - - - - - - - - - - 4-31
View System Status and Results from Your Office - - - - - - - - - - - - - - - - - 4-31
Solvent and Waste Handling- - - - - - - - - - - - - - - - - - - - - - - - - - - 4-32
Assign Solvents to the Solvent Inlets - - - - - - - - - - - - - - - - - - - - - 4-32
Set the Reservoir Volumes - - - - - - - - - - - - - - - - - - - - - - - - - 4-32
Prime the System - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 4-33
Access the Isolera Remote Viewer - - - - - - - - - - - - - - - - - - - - - - - - 4-35
Page v
Clean the Pump Check Valves - - - - - - - - - - - - - - - - - - - - - - - - 5-4
Release Stuck Check Valves- - - - - - - - - - - - - - - - - - - - - - - - - 5-5
Clean the Collect Valve - - - - - - - - - - - - - - - - - - - - - - - - - - - 5-5
Clean or Replace the Sample Loading Pump Tubing- - - - - - - - - - - - - - - - - 5-5
Clean the Sample Loading Pump Tubing - - - - - - - - - - - - - - - - - - - 5-5
Replace the Sample Loading Pump Tubing - - - - - - - - - - - - - - - - - - 5-6
Sonicate Solvent Inlet Filters - - - - - - - - - - - - - - - - - - - - - - - - - - 5-6
Leaks - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 5-7
Shut Down the System - - - - - - - - - - - - - - - - - - - - - - - - - - - 5-7
Internal Leakage - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 5-7
External Leakage - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 5-8
Assembling Tubes Correctly - - - - - - - - - - - - - - - - - - - - - - - - - 5-9
Replace the Fuses - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 5-10
Replace the Needle - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 5-10
Replace the Detector Lamp(s) - - - - - - - - - - - - - - - - - - - - - - - - - - 5-11
6: Troubleshooting - - - - - - - - - - - - - --- --- ---- --- ---- --- --- - --- --- - 6-1
Contact Biotage 1-Point Support - - - - - - - - - - - - - - - - - - - - - - - - - 6-1
Accessories and Spare Parts- - - - - - - - - - - - - - - - - - - - - - - - - - - 6-1
Fraction Collector-Related Problems - - - - - - - - - - - - - - - - - - - - - - - 6-1
Pump-Related Problems - - - - - - - - - - - - - - - - - - - - - - - - - - - - 6-2
Internal Detector-Related Problems - - - - - - - - - - - - - - - - - - - - - - - 6-3
Gradient Problems - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 6-4
Leak Detected- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 6-5
Overpressure Detected - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 6-5
7: Contact Information - - - - - - - --- - --- --- ---- --- ---- --- --- - --- --- - 7-1
Manufacturer - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 7-1
Biotage Sweden AB - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 7-1
Sales Offices and Distributors - - - - - - - - - - - - - - - - - - - - - - - - - - 7-1
Technical Support - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 7-1
Page vi
Chapter
Description and
1 Specifications
A G C
H
D
Sample Loading Pump
on Isolera LS
E
K
I
F Solvent Tray
Collection Trays: 1 1 or 2† 1 or 2† 2
Solvent Inlets: 2 4 4 4
Cartridge Positions: 1 1 4 1
Waste Channels: 1 1 4 1
Flow Rate: 1 to 100 ml/min 1 to 200 ml/min 1 to 200 ml/min 50 to 500 ml/min
Internal Detector: Fixed or variable‡ Fixed, variable, or Fixed, variable, or Variable or UV-
UV-VIS‡ UV-VIS‡ VIS‡
* An Isolera Prime system can be upgraded to and Isolera One or Isolera Four system.
An Isolera One system can be upgraded to an Isolera Four system.
†
One collection tray is standard, but by expanding the fraction collector, it is possible to use two
collection trays (see Figure 1-1) at the same time. The maximum collection volume (with no rack
change) is then 9600 ml instead of 4800 ml (with the 480 ml bottle rack). These systems are called
EXP systems (Isolera One EXP and Isolera Four EXP).
‡
The internal detector is available in three models: 1) 200-400 nm (variable UV detector), 2) 200-800
nm (UV-VIS), and 3) 254 nm (fixed UV detector). Note that the UV-VIS detector is not available with
Isolera Prime and the fixed UV detector is not available with Isolera LS.
Biotage Leak Detector is available for safe, unattended operation. The sample loading pump, on
Isolera LS, can be used for injection of liquid samples into cartridges. This is useful when using
large cartridges and sample volumes.
Online Help
To view context-sensitive help, press the Help button found in the right-hand panel. The following
buttons are available in the help window:
: View the safety requirements for the Isolera system. You must observe all these safety
requirements when installing and operating the Isolera system. Failure to install or use the system
in a manner specified by Biotage may result in personal injury and/or equipment damage.
: View instructions for tasks that can be performed in the present software mode (the Chemistry,
Data Administration, or System mode).
1.1.5 Accessories
Cartridges
The Isolera system is designed to operate with Biotage SNAP™, Biotage SNAP Ultra™ and Biotage
ZIP™ cartridges, which have easy-to-use Luer-lok™ connections. FLASH+® and ISOLUTE®
cartridges can also be used but may require tube adapters and a separate ring stand.
All SNAP and SNAP Ultra disposable cartridge components are made from materials meeting USP
Class VI specifications, which minimize organic extractables providing cleaner fractions.
SNAP cartridges are available packed with silica (KP-Sil), high-performance silica (HP-Sil), C18 (KP-
C18-HS), or NH (KP-NH) media. SNAP 750g and 1500g are available packed with silica (KP-Sil), C18
(KP-C18-HS), or NH (KP-NH). Biotage ZIP cartridges are all packed with silica (KP-Sil).
SNAP Ultra cartridges deliver double the purification capacity of other flash cartridges by utilizing
Biotage HP-Sphere™, small particle, 25 μm spherical silica with a 40% increase in surface area. The
higher surface area provides twice the loading capacity of lower surface area silicas. This improved
load capacity means that a smaller SNAP Ultra cartridge can be used to replace a more expensive and
larger competitive cartridges.
Note that SNAP and SNAP Ultra 10g to 30g and Biotage ZIP 5g to 45g cannot be used on Isolera LS,
and SNAP 1500g is not recommended for use on Isolera Prime, Isolera One, or Isolera Four. SNAP
750g, Flash 75S, Flash 75M, and Flash 75L are not recommended for use on Isolera Prime.
Liquid Loading
Liquid samples can be:
• Injected into the SNAP or SNAP Ultra cartridge inlet Luer fitting using a syringe or, when using
an Isolera LS system, the sample loading pump.
• Loaded onto the frit using a pipette (not possible when using SNAP 750g or 1500g).
Samples should be dissolved at the highest concentration in the weakest solvent. Use of a strong,
polar solvent will compromise purification efficiency.
18 x 150 mm 18 mm x 150 mm 27 ml 28 4
25 x 150 mm 25 mm x 150 mm 45 or 53 ml 15 4
* Only included in the preconfigured list for Isolera Prime, Isolera One, and Isolera Four. Not
recommended on Isolera LS.
†
Only included in the preconfigured list for Isolera LS.
Order Accessories
To order accessories, please contact your local representative. A condensed accessory list is
available on page 5-1.
NOTE
Only users with system owner privilege can log into the Data
Administration and System modes.
Figure 1-5. Main Menu and Data Administration Mode (Isolera LS Shown)
Task field
The system can be idle, paused ( ), or processing ( ). When the system is processing, the
following tasks can be in progress:
• UV ON: The system turns on the detector lamp(s).
• Eq. Flush: Before an equilibration is started, the system runs an equilibration flush to empty
the solvent inlets of solvents used in the previous purification and fills them with new solvents.
WARNING
Keep your hands out of range of the collection arm until it has stopped
moving (with the collection arm in the inner right corner).
NOTE
Only users with system owner privilege can log into the Data
Administration and System modes.
NOTE
You do not have to save your method to be able to run it.
• Monitor and edit a purification in progress. While a purification is running, the programmed
gradient and a dynamic chromatogram are displayed. The chromatogram can be magnified,
dragged to the desired position, etc. If a Spektra license has been installed on your system,
the absorbance spectrum for the whole detector range can be viewed in the gradient view by
pressing the Show λ button. The gradient is then displayed in the chromatogram.
• Start and stop collecting through the waste channel (the Bypass Tray button).
• Stop – end (the enabled flushes and a purge will be performed) or abort – a purification.
• Clear a cartridge position in the software (the Remove button).
WARNING
When pressing the Stop button, keep your hands out of range of the
collection arm until it has stopped moving (with the collection arm in the
inner right corner).
NOTE
If you want to end or abort a purification in progress, press the Stop
button.
Cartridge Status
The status of a cartridge can be one of the following:
• Free: No purification is assigned to the cartridge.
• Queued Equilibration/Gradient Run*: The equilibration or gradient run is queued and will
be performed when the system has completed the task(s) in progress.
• Equilibrating: The system is running the equilibration.
• Equilibration Finished: The equilibration is completed. Press the Load Sample button to
load the liquid sample and then the Gradient button to start or queue up* the gradient run.
• Running Gradient: The system is running the gradient.
• Finished: The purification is completed.
• Manually Extended: The purification was manually extended after it was completed. Press
the Gradient button to start or queue up* the extended run.
* Purifications can be queued up when using an Isolera Four system.
NOTE
When using an Isolera Four system, queued equilibrations take priority
over queued gradient runs.
Cartridge status
Color Legends
Fractions
Fractions that are already collected can be located by matching their colors on the chromatogram
with the vessel colors in the tray overview:
= The vessel is allocated but has not been used.
, , , and = The vessel contains a fraction.
= The vessel contains flush liquid.
If fractions are collected through the waste channel (i.e. the Bypass Tray mode is enabled), the
fractions are colored in magenta in the chromatogram and numbered W1, W2, and so on.
Light Absorbance
Light absorbance and the Start Threshold (if defined) are displayed in the chromatogram using the
following colors:
= Absorption measured by the internal detector, channel 1.
= Absorption measured by the internal detector, channel 2. (Not available if using the fixed
detector.)
= Absorption measured by the external detector (when connected).
= The defined threshold in mAU.
The defined wavelengths and the measured absorption in mAU are displayed underneath the
chromatogram.
If a Spektra license has been installed on your system, the absorbance spectrum for the whole
detector range ( ) can be viewed in the gradient view by pressing the Show λ button.
At the Report Setup tab, you can select the fractions to be displayed in the reports. The colors
used in the tray overview are:
= The vessel has not been used for the viewed purification.
= The vessel is deselected.
= The vessel is selected and contains flush liquid.
, , , and =The vessel is selected and contains a fraction. (The vessel color corresponds
with the fraction color in the chromatogram.)
If a Spektra license has been installed on your system, you can view a 2D and 3D absorbance
spectrum for the whole detector range by pressing the λ... button. Here you can also modify how
the 3D spectrum is presented in the archive report.
To create a method with the same purification parameters as the ones used in the selected record,
press the Create button at the Result Selection or Report Setup tab, or if you want help
optimizing the gradient, press the Optimize button.
1.3 Specifications
1.3.1 General System Specifications
Power
Input Voltage 100-240 VAC, 50/60 Hz
Certification
Certificates CE, UL, CAN/CSA
EN 1127-1:2007
EN 61326-1:2006
UL 61010-1:2004
The Isolera LS system is delivered with two kinds of peristaltic pump tubes. Ensure that the used peristaltic pump
tube is compatible with the solvents and sample to be used; see section 1.3.4 on page 1-20.
At least three centimeters (one inch) of space should be maintained between the
system’s rear panel and other objects to allow proper ventilation.
Environmental
Location A laboratory fume hood with capacity to handle leakage of solvents. If using an Isolera
LS Funnel Rack Kit, a walk-in fume hood must be used.
Interfaces
LAN port, two USB ports (2.0), mouse port, VGA port, external detector port, and leak detector port
Cartridge Positions
Isolera Prime, One, and 1
LS
Solvents
Solvent Supply A maximum of 2 x 4 liter (Isolera Prime) or 4 x 4-liter (Isolera One, Four, and LS) non-
glass reservoirs on top of the system (the solvent tray). Larger reservoirs can be placed
on the side of the system. If using glass reservoirs, up to 2.5-liter reservoirs can be
placed on the solvent tray.
Fraction Collector
Fraction Volume Programmable from 1 to 9999 ml (Isolera Prime, One, and Four) or
5 to 99999 ml (Isolera LS), depending on rack type
Max Fraction Number 192 fractions with no rack change (384 with two collection trays), 13x100 mm racks
Max Total Volume 4800 ml with no rack change (9600 ml with two collection trays), 480 ml bottle racks
If using Isolera LS with a funnel rack, the maximum total volume is 320 liter
Rack Types 13x100 mm*, 16x100 mm*, 18x130 mm (for 17.5 x 130 mm test tubes)*,
16x150 mm*, 18x150 mm, 25x150 mm, 120 ml, 240 ml, 480 ml, and funnel rack†
* Not recommended on Isolera LS.
† Cannot be used with Isolera One or Isolera Four.
Internal Detector
Wavelength 200 nm to 400 nm (variable), 200 to 800 nm (UV-VIS), or 254 nm (fixed)
The UV-VIS detector is not available with Isolera Prime and the fixed UV detector is not
available with Isolera LS. (Isolera Prime can be upgraded to an Isolera One or Four system.)
Pump
Flow Rate Range 1 to 100 ml/min (Isolera Prime), 1 to 200 ml/min (Isolera One and Four), or 50 to 500
ml/min (Isolera LS), in 1 ml/min increments
Gradient
Gradient Precision ±3%(Isolera Prime) or ±2%(Isolera One, Four, and LS)
Display
Screen 10.4" LCD; TFT active matrix; 800 x 600 resolution
WARNING
Many solvents are considered to be hazardous to humans and the
environment, so take appropriate safety precautions when using them.
Comply with Material Safety Data Sheets (MSDS) and any other applicable
regulations for the safe use, handling, transporting, storage, and disposal
of these solvents.
NOTE
The peristaltic pump tube has limited use. To avoid the risk of cross-
contamination and decreased pump performance, it is recommended that
the peristaltic pump tube is discarded after each run. If you still want to
reuse the peristaltic pump tube, clean the tube as soon as possible as
described in section 5.8.1 on page 5-5.
Limited: The peristaltic pump tube must be discarded after each run.
The chemical resistance was tested in room temperature as follows: The peristaltic tube was
1) weighed, 2) submerged in solvent for three hours, 3) dried and visually inspected, 4) submerged for
another three hours, 5) dried, visually inspected, and weighed, and 6) dried for 24 hours and weighed.
WARNING
Before performing any procedures in this chapter, please read and observe
the safety requirements in the “Isolera Installation and Safety” document
(P/N 411828). Failure to follow those requirements may result in personal
injury and/or equipment damage.
NOTE
The first time you log into the Data Administration mode, log in using the
user account “System Owner” and the password “1234”. Before you log
out, it is strongly recommended that the password is changed (see
“Administrate the User List” on page 2-9).
NOTE
The preconfigured Biotage cartridges cannot be edited or deleted.
SNAP and SNAP Ultra 10g to 30g and Biotage ZIP 5g to 45g cannot be used
on Isolera LS. SNAP 1500g is not recommended for use on Isolera Prime,
Isolera One, or Isolera Four. SNAP 750g, Flash 75S, Flash 75M, and Flash
75L are not recommended for use on Isolera Prime.
NOTE
The preconfigured Biotage methods cannot be exported or deleted.
NOTE
If you import a method with a solvent, cartridge, rack, user, or project that
is not available on the system, it will also be imported. The decision is
based on the name of the solvent, cartridge, etc., and will not consider
their settings.
You can either import a single method or all methods available on a USB memory device or in a
share folder on your network (if file share has been set up; see page 3-7). The method(s) to import
must be available at “USB”:\biotage\isolera\methods or “share folder”\biotage\isolera\methods.
1. If you want to import the method(s) to a certain user, select the user at the Methods tab.
2. To import method(s) available on a USB memory device:
a. Connect the memory device to the USB port located underneath the touch screen.
b. Press Import.... The Import Methods from USB dialog opens.
3. To import method(s) available in the specified share folder, press Import.... The Import
Methods from Network dialog opens.
4. To import a single method, select the method and then select Selected Method in the
Method(s) to Import list.
5. To import all methods, select All in the Method(s) to Import list.
6. Select import destination. You can either save the method(s) to the user selected in step 1,
To Selected User, or to the user defined in the method(s), To Original Users.
7. To import the method(s), press OK.
NOTE
The preconfigured Biotage racks cannot be edited or deleted.
The funnel rack kit from Biotage can only be used with Isolera LS.
• The x- and y-coordinates for the first vessel position of each rack on the collection area;
see Figure 2-4.
NOTE
To scroll a list down or up, press or .
NOTE
The preconfigured solvents cannot be edited or deleted.
Figure 2-7. Solvents Tab (System with Four Solvent Inlets Shown)
NOTE
To scroll a list down or up, press or .
WARNING
All the results and methods of the user will also be deleted.
1. Select the user account that you want to delete at the Users tab.
2. Press Delete. The Confirm Delete dialog opens.
3. To confirm delete, press Yes.
WARNING
Before performing any procedures in this chapter, please read and observe
the safety requirements in the “Isolera Installation and Safety” document
(P/N 411828). Failure to follow those requirements may result in personal
injury and/or equipment damage.
NOTE
The first time you log into the System mode, log in using the user account
“System Owner” and the password “1234”. It is strongly recommended
that the password for “System Owner” is changed; see “Administrate the
User List” on page 2-9.
Figure 3-1. Log into the System Mode (Isolera LS with Spektra License Shown)
NOTE
The external detector and the tube assembly must be able to handle a
maximum dispense rate of 200 ml/min (when using an Isolera Prime,
Isolera One, or Isolera Four system) or 500 ml/min (when using an
Isolera LS system) and a maximum pressure of 10 bar.
As some external detectors have flow rate limitations (e.g. ELSD), the flow
stream may need to be split so that the majority of the flow is directed
through the internal detector and a proportion diverted to the external
detector, or the dispense rate (i.e. the speed at which liquid is pushed into
the tubing of the external detector) must be limited (see step 7 below).
Please contact the external detector manufacturer for flow rate
specifications.
Pin Connection
2 Ground
2. Connect the internal and external flow cells in series or by using a splitter, or move the internal
detector's inlet and outlet tubing (see Figure 3-3) to the external flow cell. Ensure that you use
as short extra tubing as possible. For more information, please refer to the manufacturer of the
external detector. Tip! If the flow cells are connected in series and the pressure exceeds 10 bar
(i.e. the run is aborted), retry using a splitter.
3. Select the Detector tab in the right-hand panel.
4. Select the Enabled text box in the External Detector field and select Yes.
5. Select the Signal Range text box and enter the maximum voltage of the signal from the
external detector. The valid range is 1 mV to 5000 mV. During the run, the system will convert
0 mV to 0 mAU and the maximum voltage to 6000 mAU.
6. Select the Extended Flush Volume text box and enter the amount of solvent that has to be
added to the flushes due to the flow cell volume and/or the extra tubing associated with the
external detector.
A
A Flow Cell Outlet Tube
B Flow Cell
7. Select the Max Dispense Rate text box and enter the maximum speed at which liquid can be
pushed into the tubing of the external detector during the dispense stroke. The valid range is
1 to 200 ml/min when using an Isolera Prime, Isolera One, or Isolera Four system and 50 to
500 ml/min when using an Isolera LS system. If this parameter is turned off, the system's
maximum dispense rate will be used (i.e. 200 ml/min or 500 ml/min).
3.3.5 Enable the λ-All Detection Mode and TLC to Step Gradient Editor
To enable the λ-all detection mode and the TLC to Step Gradient editor, you need a Spektra license.
To purchase a Spektra license, please contact your local representative. Note that the license is not
available for Isolera Prime systems.
To install a Spektra license for the λ-all detection mode and the TLC to Step Gradient editor:
1. Save the Spektra license file in a “biotage/isolera/” directory on a USB memory device.
2. Connect the memory device to the USB port located underneath the touch screen.
3. Select the General tab in the right-hand panel.
4. Press Install License in the Licenses field.
5. If the license is valid, License OK appears in the Advanced Features text box.
6. Remove the USB memory device.
7. Restart the system:
a. Press Main Menu and then Shut Down.
b. When the message It is now safe to turn off the system appears on the screen, turn off
the system. The power switch is located underneath the touch screen.
c. Turn on the system.
NOTE
If connecting a network printer, only printers with postscript
support can be used with the system.
5. Select the Printer Model text box. The Printer Setup Wizard opens.
6. Read and follow the instructions that appear on the screen. Tip! If the Isolera system does not
have a matching printer driver for the connected printer, select Generic in the wizard’s
manufacturer list and check if there is a printer protocol available that is supported by your
printer (check the printer user manual).
7. When you have completed the printer setup using the wizard, select the Paper Size text box
and then select the desired paper format.
8. To test the connection, press Test in the Printer field. If a test page is printed, the connection
is working properly.
To connect a network printer with postscript support to the system:
1. Configure the network connection; see instructions on page 3-5.
2. Select the Reporting tab in the right-hand panel.
3. Select the Printer Type text box and select Network.
4. Enter the correct Network Printer Name and Network Printer Port by selecting the
corresponding text box. Contact your IT administrator for more information.
5. Select the Paper Size text box and then select the desired paper format.
6. To test the connection, press Test in the Printer field. If a test page is printed, the connection
is working properly.
• Automatic Vessel: The system allocates vessels in sequence, starting with the first vessel
in rack A, the second in rack A, and so forth. With this rack allocation method, racks can be
shared by several runs but the collection tray cannot be shared by several users.
When enabling automatic rack allocation on an Isolera Four system, the following rules also apply:
• All racks on the collection tray must be of the same type. The rack type to be used is
decided by the first purification that is queued up.
• Allocation only when needed. No rack or vessel allocation is performed when a
purification is queued up. When the run is started racks or vessels are allocated one at a time
as they are needed.
• Always replace all racks when asked to. If the system runs out of vessels, you will be
asked to replace all racks in the collection area at the same time (even if some of the racks
have been used by previous runs).
To enable or disable automatic rack allocation:
1. Select the Runtime tab in the right-hand panel.
2. Select the Rack Allocation Mode text box.
3. To enable automatic rack allocation, select Automatic or Automatic Vessel.
4. To disable automatic rack allocation, select Manual.
6. In the Chemistry mode, enter the capacities and current fluid levels for the solvent and/or
waste reservoirs; see “Set the Reservoir Volumes” on page 4-32. (For the system to be able to
maintain a running calculation of fluid levels in each reservoir, the capacities and current fluid
levels for the reservoirs must be entered each time you empty a waste reservoir or replenish a
solvent.)
To disable solvent and/or waste monitoring:
1. Select the Runtime tab in the right-hand panel.
2. Select the Solvent Monitoring and/or Waste Monitoring text box and select No.
NOTE
If the automatic line flush is disabled, ensure to prime the cartridge flow
path before each run.
4. To enable or disable air flush, select the Air Flush text box and select To Waste or Off. This
feature is not available with Isolera Prime.
5. To change the solvent or solvent mix to be used for the line, detector, and/or air flush, select
the corresponding text box and select the desired solvent or solvent mix:
• Method – The system will flush with the weakest or strongest solvent used in the run.
• System – The system will flush with the weakest or strongest solvent mounted on the system.
• Gradient – The system will flush with the solvent mix used in the start or the end of the
gradient. (Not available for air flush.)
• Channel x – The system will flush with the solvent mounted on the specified channel (1, 2,
3, or 4); see “Assign Solvents to the Solvent Inlets” on page 4-32. Note that Isolera One,
Four, and LS have four solvent inlets while Isolera Prime has two.
3.7 Maintenance
3.7.1 Export the System Configuration and Logs
If requested by 1-Point Support™ to send in your system’s log files, follow the instructions below.
The system configuration and logs can be saved on a USB memory device connected to the USB
port at the front of the system or in a share folder on your network (if file share has been set up;
see page 3-7).
1. Select the Maintenance tab in the right-hand panel.
2. If you want to save the system configuration and logs on a USB memory device:
a. Connect the memory device to the USB port located underneath the touch screen.
b. Press Export in the System Configuration and Logs field.
3. If you want to save the system configuration and logs in the specified share folder, press
Export in the System Configuration and Logs field.
4. The system configuration and logs are saved as a zip file at \biotage\isolera\logs\. Send the zip
file to 1-Point Support; see “Contact Information” on page 7-1.
NOTE
Only restore the system configuration when instructed to do so by Biotage.
If you have problem with the restore, please contact 1-Point Support.
To restore the system configuration e.g. after an update of the system's operating system:
1. Select the Maintenance tab in the right-hand panel.
2. Connect the USB memory device containing the zip file with the desired system configuration to
the USB port located underneath the touch screen. Note that the zip file must be saved at
\biotage\isolera\logs\ on the USB memory device.
3. Press Restore Configuration in the System Configuration and Logs field.
4. In the Select Zip File dialog, select the zip file and press OK. The Restore from USB Device
dialog opens.
5. To confirm restore, press Restore.
6. When the Restore Successful dialog appears, press OK. The Restart Required dialog opens.
7. Read the instructions in the dialog on how to restart the system and then press OK.
8. Restart the system.
9. If you have a USB printer connected to the system, set up the printer as described in “Set Up a
Printer and Auto Print of Reports” on page 3-6.
10. Verify that the calibration of the touch screen and the fraction collector is working properly. If
the system needs to be recalibrated; see “Calibrate the Touch Screen” on page 3-12 and
“Calibrate the Fraction Collector” on page 3-13.
NOTE
Only perform an intensity and absorbance calibration when you have
installed a new detector lamp. If you have installed a new detector flow
cell, perform an absorbance calibration. (The type of calibration is selected
in the wizard.)
WARNING
Before performing any procedures in this chapter, please read and observe
the safety requirements in the “Isolera Installation and Safety” document
(P/N 411828). Failure to follow those requirements may result in personal
injury and/or equipment damage.
WARNING
To avoid being struck by the collection arm, keep your hands out of range
of the collection arm while the homing routine runs in step 2 below.
2. Turn on the system. The power switch is located underneath the touch screen. The collection
arm moves through its homing routine and the system boots to the system’s main menu.
3. To set up, control, monitor, and review a purification, press Chemistry. The system is opened
in the Chemistry mode.
If you want to administrate cartridge types, methods, rack types, results, solvents, and user
accounts, see “Log into the Data Administration Mode” on page 2-1.
If you want to change detector, network, reporting, and runtime settings, set the system clock, set
the language used in the Chemistry mode, enable the λ-all detection mode and the TLC to Step
Gradient editor (requires a Spektra license), calibrate the touch screen, fraction collector, and
internal detector, perform an intensity scan of the internal detector, release stuck check valves,
clean the collect valve, back up and restore the database, export the system configuration and logs,
restore the system configuration, view and reset usage statistics, etc, see “Log into the System
Mode” on page 3-1.
For more information about the software modes, see “Software Description” on page 1-7.
NOTE
The UV lamp should be turned on 7.5 minutes before starting a purification
so the UV lamp can reach operating temperature.
NOTE
If you want to use the method wizard, see page 4-13. The wizard will
guide you step-by-step through the setup of a purification with a gradient
including up to three steps using two solvents.
Figure 4-1. New Method and Open Method Dialogs (Isolera LS Shown)
NOTE
If you want to calculate the gradient from TLC data, see page 4-10.
Isolera One, Four, and LS have four solvent inlets (A, B, C, and D) while
Isolera Prime has two (A and B).
8. To increase or decrease the length of a gradient segment or change the solvent mix:
• Using the gradient graph: Drag the segment or node to the desired position (with 0.5 CV
and 5% resolution). To fine tune (with 0.1 CV and 1% resolution), select a segment or
node (see Figure 4-2) and use the and buttons to increase or decrease the length of a
segment and the and buttons to increase or decrease the solvent percentage. The
fine tuning buttons are only available when the gradient table is not shown.
• Using the gradient table: Select the value you want to change and enter the desired value
using the keypad.
Selected segment
Selected node
As a single-piston pump only delivers liquid during the dispense stroke, the maximum flow rate
when using an Isolera Prime system is equal to (Max Fill Rate x Max Dispense Rate)/(Max Fill
Rate + Max Dispense Rate). Examples: If the fill rates for the used solvents are 50 ml/min and
80 ml/min and you use a external detector with the maximum dispense rate of 100 ml/min, the
highest flow rate that can be used is (50x100)/(50+100)=33 ml/min. If the external detector is
disabled, the highest flow rate that can be used is (50x200)/(50+200)=40 ml/min.
WARNING
If small test tubes are to be used, ensure to adjust the flow rate so that no
splashing will occur when fractions are collected.
4. Select the Rack Type text box and select the rack type to be used; see Table 4-1 on page 4-7
for guidance. If you do not want to use the default max fraction volume for the vessel type
and/or the default dispensation order (S-pattern), select the corresponding text box and
change the setting. The software prevents you from entering a fraction volume that will
overflow the vessels.
5. Select the solvents to be used by selecting the corresponding text box in the Solvents field.
(They should be the solvents determined by the TLC procedure.) To show all solvents, press
Show All. To only show the solvents connected to the system, press Show Connected. Note
that the solvent percentages are specified at the Gradient tab.
NOTE
For reversed-phase purification with methanol and water, it is strongly
recommended to premix the water with 5% of methanol and degass either
through vacuum or sonication. Degassing of protic solvent blends
decreases out-gassing of entrapped air during gradient elution, which will
impact gradient performance and flow rates.
If automatic rack allocation is enabled, all racks on the collection tray must
be of the same type. When using an Isolera Four system, the rack type to
be used is decided by the first purification that is queued up. Any
purification queued up after that must use the same rack type. To change
rack type, wait until all present runs are finished and then remove them at
the Status tab.
Rack Selection
Based on 13 CV gradient in Collect All mode (with slope and valley fractionation disabled), the
following apply in the rack selection table on page 4-7:
Requires multiple racks of this type, but no rack change during the run. See the number of
racks required in the table.
Rack change required during the run. See the number of racks required in the table.
Not recommended.
Note that:
• Isolera Prime can only be used with one collection tray.
• The maximum collection volume for Isolera LS can be increased from 9.6 liters up to 320 liters
by using a funnel rack kit from Biotage. The funnel rack kit comes with two racks (16 positions
each) and a cart with wheels that holds the Isolera LS system and the collection vessels. For
more information, please contact your local representative.
411829-L, Operation
18x130 mm† 4 504 x 4 2 2/3 4/5 7
16x150 mm† 4 770 x 4 2/2 3/3 5
18x150 mm 4 756 x 4 2/2 3/3 5
25x150 mm 4 795 x 4‡ 2/2 3/3 5
120 ml 4 720 x 4 2/2 3/3 5
240 ml 1 4320 2/2 2
480 ml 1 4800 2/2 2
April 2012
25x150 mm 4 795 x 4‡ 2/2 3/3 5 9/10 9
120 ml 4 720 x 4 2/2 3/3 5 10/11 10
240 ml 1 4320 2/2 2 3 4 6
480 ml 1 4800 2/2 2 3 3 6
Page 4-7
4
5. Specify the collection and fractionation settings to be used in the Parameters field. Refer to
Table 4-2 below for guidance. For more information; see the “Collection and Fractionation”
appendix on page A-1.
Parameter Description
Start The shortest wavelength to be included in the λ-all signal. The range is 200 to 400 nm
Wavelength (dual variable UV detector) or 200 to 800 nm (UV-VIS detector). Note that this
parameter is only available with the λ-all detection mode.
End Wavelength The longest wavelength to be included in the λ-all signal. The range is 200 to 400 nm
(dual variable UV detector) or 200 to 800 nm (UV-VIS detector). Note that this
parameter is only available with the λ-all detection mode.
Collect All Light absorbance of sample is monitored, but collection is based on the gradient’s total
volume. When using the detection mode UV1+UV2 or UV1, valley fractionation is
enabled by default in the System mode. If disabled, fractionation is based on the max
fraction volume for the vessel type.
Start Threshold Used to collect samples with an absorbance exceeding the set absorbance threshold.
Initial Waste Delay collection until a specified volume has been delivered to the waste.
Slope Collection is based on the start slope of light peaks. Refer to the “Collection and
Fractionation” appendix on page A-1 for details. Note that slope collection is only
possible with the UV1 and UV1+UV2 detection modes.
NOTE
You do not have to save your method to be able to run it.
5. Select the Rf Product text box and enter the Rf value of the product of interest.
6. Enter the Rf value of the impurity closest below and above the product in the Rf Impurity 1
and Rf Impurity 2 text boxes. A value for Rf Impurity 1 is required.
7. Select the Cartridge text box and select a cartridge type with a loading capacity exceeding
your crude sample mass.
8. To save the data, press OK. The TLC to Linear Gradient editor closes and the software
calculates the purification gradient based on the TLC data and displays it at the Gradient tab.
9. Set up the rest of the method parameters; see “Specify the Method Parameters at the
Parameters Tab” on page 4-4 and “Specify Collection Parameters at the Collection Tab” on
page 4-8.
10. To save the method and reuse it in the future, see “Save and/or Run the Method” on page 4-9.
11. To run a purification using the (saved or unsaved) method, see “Prepare and Run a Purification”
on page 4-15.
Selected segment
8. If desired, select the compound of interest by pressing on its label and press Set Peak of
Interest.
9. When satisfied with the gradient, select the Cartridge Selection tab.
10. Select the Sample Mass text box and enter the crude sample mass.
11. Select the Cartridge text box and select a cartridge type with a loading capacity exceeding
your crude sample mass. If a peak of interest has been set (see step 8), the load capacity is
calculated according to the smallest CV between the peak of interest and the peak before and
after. If no peak of interest has been set, the load capacity for each cartridge type is calculated
according to the smallest CV.
12. To close the TLC to Step Gradient editor and open a method containing the purification gradient
at the Method tab, press OK.
13. Set up the rest of the method parameters; see “Specify the Method Parameters at the
Parameters Tab” on page 4-4 and “Specify Collection Parameters at the Collection Tab” on
page 4-8.
14. To save the method and reuse it in the future, see “Save and/or Run the Method” on page 4-9.
15. To run a purification using the (saved or unsaved) method, see “Prepare and Run a Purification”
on page 4-15.
3. Read and follow the instructions that appear on the screen. (See Table 4-1 on page 4-7 for
guidance in selecting rack type.)
4. If you want to save the method and reuse it in the future:
a. Press Save to Editor.
b. Enter a unique method name by selecting the Method Name text box at the
Parameters tab.
c. Press Save. If your user account is password-protected, the Password dialog opens.
Enter your password, using the keypad, and press OK.
5. To run a purification using the method, press Run....
6. Prepare the system as described on page 4-15.
7. Proceed with step 5 in “Run a Purification” on page 4-19.
NOTE
Gradients that are optimized on an Isolera Prime, Isolera One, or Isolera
Four system cannot be run on an Isolera LS system due to the different
system volumes used for the calculation, and vice versa.
Press to toggle
between
showing ( )
and hiding the
original gradient
Original
gradient
Optimized step
gradient
4. To show the actual peaks, adjust the peak sensitivity and/or the threshold using the sliders (see
Figure 4-9).
5. Select/press the peak that you want to collect. The suggested step gradient is displayed.
6. When you are pleased with the result, press Save to Editor. The method is opened at the
Method tab and can be edited and/or saved.
NOTE
The UV lamp should be turned on approximately 7.5 minutes before
operation so it can warm up and stabilize.
All solvent inlets must be primed with solvent to achieve the specified
pump performance.
Prepare the Sample Loading Pump for a Run (Only Isolera LS)
NOTE
The peristaltic pump tube has limited use. To avoid the risk of cross-
contamination and decreased pump performance, it is recommended that
the peristaltic pump tube is discarded after each run.
The system is delivered with two kinds of peristaltic pump tubes (PharMed
and Fluran). Ensure that the used peristaltic pump tube is compatible with
the solvents and sample to be used; see section 1.3.4 on page 1-20.
Samples to be loaded using the sample load pump must be fully dissolved.
Particulates in the sample may cut or weaken the peristaltic pump tube
causing a sample leak.
To empty the pump tubing of liquid and mount a new peristaltic pump tube:
1. Place the pump’s outlet line into a waste reservoir.
2. Select the Solvents tab in the right-hand panel (in the software) and then the Sample
Loading Pump tab.
3. Press Start to start the pump and draw air into the inlet line.
4. When the pump tubing is empty, press Stop.
5. Remove the inlet and outlet tubing by removing the two screws; see Figure 4-11.
Remove screws
6. Remove the pump by turning it counterclockwise and then pulling it out; see Figure 4-12.
7. Replace the tube inside the pump. Ensure that the new tube is compatible with the solvents and
sample to be used in the next run; see section 1.3.4 on page 1-20.
8. Remount the pump and the inlet and outlet tubing. If necessary, replace the inlet and outlet
tubing.
To reuse the peristaltic pump tube used in the previous run (not recommended – see the note
above), flush and fill the tubing with a solvent suitable for the next purification:
1. Place the pump’s inlet line into the solvent of your choice.
2. Place the pump’s outlet line into a waste reservoir.
3. Select the Solvents tab in the right-hand panel (in the software) and then the Sample
Loading Pump tab.
4. To start the pump, press Start. Adjust the flow rate using the Speed slider.
5. When you are done, press Stop.
WARNING
If you at any time during a purification need to pause the system, press
the Pause button in the right-hand panel (in the software). The collection
arm returns to its home position (the inner right corner) and the system is
paused. Note that the system is pressurized when it is paused. To resume
operation, press the Resume button.
NOTE
When using an Isolera Four system with automatic rack allocation
enabled, several runs (using the displayed method) can be queued up
simultaneously by selecting two or more cartridge positions in the Run
Parameters dialog.
6. In the Estimation dialog (if present), ensure that a sufficient quantity of the correct solvent is
present in each solvent reservoir, the waste reservoir has sufficient capacity for the run, and the
sufficient number of vessels have been allocated.
7. To start or queue up* the purification, press Equilibrate/Gradient (see Figure 4-15). The
Status tab is selected. Before the run is started, the system performs a UV Zero using the
gradient's initial solvent mix. Note that a gradient run is not started until the UV lamp is
sufficiently warmed up.
8. When the equilibration step (if included) is finished:
a. Load your sample; see “Sample Loading Methods for SNAP and SNAP Ultra Cartridges” on
page 1-4. If a liquid sample is used, press Load Sample... at the Status tab to enable
sample injection. The Load Sample dialog opens. To load a sample using the sample
loading pump (only available on Isolera LS), read and follow the instructions in the dialog.
When you have finished loading the sample, press Close and connect the cartridge inlet
tubing to the cartridge’s inlet (see Figure 4-10). See Table 4-3 on page 4-21 for cartridge
loading capacity.
b. Start or queue up* the gradient run by pressing Gradient at the Status tab.
* Purifications can be queued up when using an Isolera Four system.
NOTE
When using an Isolera Four system, queued equilibrations take priority
over queued gradient runs.
Task field
NOTE
When using an Isolera Four system with automatic rack allocation enabled,
you have to replace all racks in the collection area at the same time when
asked to (even if some of the racks have been used by previous runs).
Auto-Extend Mode
If the system is collecting fractions and remaining light absorption is detected when the system
nears the end of a purification, the system enters the Auto-Extend mode (if enabled in the system
mode). This extends the gradient purification stage of the run with 25% of the total gradient length
using the final conditions in the method.
NOTE
Queued equilibrations take priority over queued gradient runs.
5. If you extended a finished run or edited a gradient run that was not started or queued up*,
press Gradient to start (or queue up*) the run. Note that if you extend a run on an Isolera
Four system and UV Zero has been performed on another cartridge in between, the system will
perform a UV Zero.
NOTE
When using an Isolera Four system, queued equilibrations take priority
over queued gradient runs.
WARNING
Keep your hands out of range of the collection arm until it has stopped
moving (with the collection arm in the inner right corner).
NOTE
It is not possible to flush a cartridge with air when using an Isolera Prime
system.
Empty the cartridge of remaining solvents using the Air Flush feature. If you want the system to
automatically perform an air flush at the end of a purification, enable the Air Flush option (see
“Specify How Flushes Are Performed” on page 3-10).
4.9.2 View, Print, and Save Reports on a USB Memory Device or the
Network
Two result reports are available for each purification: 1) an archive report with purification details, a
chromatogram, a 3D absorbance spectrum for the whole detector range (requires a Spektra
license), and TLC data (if entered in the TLC to Step Gradient editor) and 2) a fraction report. These
reports can be printed to a network printer with postscript support or a local USB printer and saved
as PDF files on a USB memory device or in a share folder on your network.
1. Select the Results tab in the right-hand panel.
2. Select the desired record at the Result Selection tab. (If the record is not listed, see “Search
for Records” above.) An archive report with purification details, a chromatogram, a 3D
absorbance spectrum for the whole detector range (requires a Spektra license), and TLC data
(if entered in the TLC to Step Gradient editor) for the selected record is available at the
Archive Report tab and a fraction report is available at the Fraction Report tab.
3. To view a 2D and 3D absorbance spectrum for the whole detector range (requires a Spektra
license):
a. Press λ... at the Report Setup tab. The Spectrum dialog opens.
b. To enable or disable baseline correction, press Correct Baseline. When enabled ( ), the
maximum solvent absorbance spectrum is subtracted from the signal presented by the
detector.
Drag this line to another position to show the absorbance spectrum at that time.
Drag one or both of these lines to other wavelengths to show the chromatogram at
those wavelengths.
c. To show a 3D absorbance spectrum instead, press 3D. Drag the 3D spectrum to change
the perspective, see the examples below in Figure 4-24. Note that the way it looks here is
the way it will look in the archive report.
4. To modify the fraction and archive reports, select the Report Setup tab. The following buttons
can be used to modify the reports:
• Gradient: When enabled ( ), the programmed gradient is shown in the chromatogram.
• Reset Zoom: Reset the zoom to default.
• Zoom: When enabled ( ), you can select a region of the chromatogram to magnify
(zoom in). This is useful when the chromatogram displays many fractions close together.
• Drag: When enabled ( ), you can drag the chromatogram to the desired position.
• Select: When enabled ( ), you can select and deselect fractions in the chromatogram.
(Fractions can at any time be selected and deselected in the tray overview.)
• Peak Mode: When enabled ( ), you can select a peak by selecting one of its fractions.
All fractions of a peak have the same color.
• Select All: Select all fractions.
• Deselect All: Deselect all fractions.
• λ...: Modify how the 3D absorbance spectrum is presented in the archive report. See step 3
above.
• Length Unit: Change the horizontal axis (the x-axis) unit by pressing the button
repeatedly until the desired unit appears; CV, minutes, or milliliter.
If there was one or more rack changes during the run, or the run was performed on a system
with two collection trays with allocated racks on both trays, toggle between the racks by
pressing the and buttons.
5. To print the archive or fraction report to the (local or network) printer connected to the system,
press Print at the corresponding tab. (If the button is disabled, no printer has been installed.
See “Set Up a Printer and Auto Print of Reports” on page 3-6.)
6. To save the selected report as a PDF file on a USB memory device:
a. Connect a USB memory device to the USB port located underneath the touch screen.
b. Press Save. The report is saved at “USB”:\biotage\isolera\results.
7. To save the selected report as a PDF file in a share folder on your network, press Save. The
report is saved at “share folder”\biotage\isolera\results\. (If the button is disabled, no file
sharing has been set up. See “Set Up File Sharing and Auto Save of Reports” on page 3-7.)
Color Legend
The colors used at the Report Setup tab are:
= The vessel has not been used for the viewed purification.
= The vessel is deselected.
= The vessel is selected and contains flush liquid.
, , , and = The vessel is selected and contains a fraction. (The vessel color corresponds
with the fraction color in the chromatogram.)
If fractions are collected through the waste channel (i.e. the Bypass Tray mode is enabled), the
fractions are colored in magenta in the chromatogram and numbered W1, W2, and so on.
WARNING
Never prime the Isolera system without a cartridge mounted on the
system or without the cartridge’s inlet and outlet tubing coupled together.
To avoid leakage, check all tubes and fittings before priming the system.
NOTE
All solvent inlets must be primed with solvent to achieve the specified
pump performance.
5. If using an Isolera Four system, select the Channel text box and select the cartridge position
and waste reservoir to be used.
6. Select the Path text box and select the prime path. If you do not want to prime the cartridge,
press Bypass.
7. If you selected the path With Cartridge, select the Cartridge Type text box and select the
cartridge type mounted on the system.
8. Select the Volume text box and enter the total prime volume.
If you want to fill a solvent inlet with a new solvent, we recommend that you prime with at least
40 ml (when using an Isolera Prime, Isolera One, or Isolera Four system) or 50 ml (when using
an Isolera LS system) of that solvent. If you also want to fill the liquid lines from the pump to
the (selected*) waste valve with the same solvent, we recommend that you prime with at least
60 ml when using an Isolera Prime, Isolera One, or Isolera Four system or 100 ml when using
an Isolera LS system.
9. Select the Flowrate text box and enter the flow rate. The maximum flow rate that can be used
depends on:
• the maximum fill rate(s) defined for the used solvent(s) in the Data Administration mode,
and
• the maximum dispense rate of the system (200 ml/min for Isolera Prime, One and Four,
and 500 ml/min for Isolera LS) or, if using an external detector, the maximum dispense
rate defined for the external detector in the System mode.
As a single-piston pump only delivers liquid during the dispense stroke, the maximum flow rate
when using an Isolera Prime system is equal to (Max Fill Rate x Max Dispense Rate)/(Max Fill
Rate + Max Dispense Rate). Examples: If the fill rates for the used solvents are 50 ml/min and
80 ml/min and you use a external detector with the maximum dispense rate of 100 ml/min, the
highest flow rate that can be used is (50x100)/(50+100)=33 ml/min. If the external detector is
disabled, the highest flow rate that can be used is (50x200)/(50+200)=40 ml/min.
10. Enter the percentage of each solvent connected to the system that shall be used in the prime.
If using an Isolera LS system and a solvent inlet line is empty or filled with air bubbles, you
should prime it separately. (If several solvents are listed and you only want to use one solvent,
enter “100” for the solvent that you want to use and enter “0” or press Clear for the other
solvent(s).)
11. To start priming, press Start.
* When using an Isolera Four system.
5 Maintenance
WARNING
Before performing any procedures in this chapter, please read and observe
the safety requirements in the “Isolera Installation and Safety” document
(P/N 411828). Failure to follow those requirements may result in personal
injury and/or equipment damage. If the system has been damaged and
does not function properly, shut it down and contact Biotage 1-Point
Support immediately.
5.2 Accessories
To order cartridges, please see the Biotage Product Catalog. The catalog can be downloaded at
www.biotage.com. For a complete list of accessories, please contact you local representative.
Test tube rack, 18 x 130 mm (for 17.5 x 130 mm test tubes) 4 412593
Not recommended on Isolera LS.
Cartridge holder for SNAP and SNAP Ultra 50g to 120g 1 411923
Cartridge holder for SNAP and SNAP Ultra 340g to 400g 1 411924
Leak detector for systems with one collection tray (Prime, One, and Four 1 412019
systems)
Leak detector for systems with two collection trays (EXP and LS systems) 1 412062
Cartridge holder for SNAP and SNAP Ultra 10g to 12g 1 411922
Cartridge holder for SNAP and SNAP Ultra 25g to 30g 1 411776
Sample injection valve (3-way) to SNAP or SNAP Ultra cartridge adapter 1 411081
5.2.3 Isolera LS
Funnel rack kit including two racks, portable cart, bottle board, positioning 1 FNRK-032
shafts (for the bottles), and leak detector
WARNING
When cleaning the touch screen, use only non-ammonia based window
cleaner and do not apply the liquid directly to the screen as this could
introduce liquid into the screen electronics section, which could damage
electronic components.
NOTE
Avoid harsh cleaners and chemicals, and getting moisture inside the system.
Keep the detector flow cell clean and protect it from dust and chemical spills. Particular attention
should be paid to preventing the flow cell from leaking.
A dirty cell has decreased transmissivity, which causes increased noise level, decreased response,
and difficulties performing both auto and manual UV Zero functions.
NOTE
A
The retaining latch should close with
little effort. If the retaining latch is
difficult to close, the flow cell is probably
B misaligned.
D B Flow Cell
C
C Flow Cell Inlet Tube
Figure 5-1. Remove the Detector Flow Cell (Isolera Prime, One, and Four Shown)
3. Disconnect the inlet and outlet tubing from the flow cell and visually inspect the cell for
contamination.
4. Flush the flow cell with a series of miscible solvents. Select the solvents based on the
contamination. It is possible to use both organic and inorganic solvents and diluted solutions of
acids (e.g. H2SO4 or HNO3 diluted with distilled water in a ratio of 1:20 to 1:10), unless they
attack stainless steel, PTFE, or fused silica windows.
5. Visually re-examine cell windows for visible contamination. If contamination is still present,
repeat step 4. If you are not able to remove the contamination, replace the flow cell (P/N 09869
for Isolera Prime. One, and Four systems and P/N 412324 for Isolera LS systems) and perform
an absorbance calibration (see “Calibrate the Internal Detector” on page 3-13).
6. Reconnect the tubing to the flow cell.
7. Carefully insert the flow cell and close the retaining latch.
9. Select the Flowrate text box and set the flow rate to 100 ml/min (Isolera Prime), 200 ml/min
(Isolera One and Isolera Four) or 500 ml/min (Isolera LS).
10. Enter 100 percentage for the solvent inlet line used for methanol. Enter “0” or press Clear for
the other solvents that are connected.
11. To start flushing the pump check valves, press Start.
3. Select the Solvents tab in the right-hand panel (in the software) and then the Sample
Loading Pump tab.
4. To start the pump, press Start. Adjust the flow rate using the Speed slider.
5. After rinsing with solvent, continue flushing the sample loading pump with air for a minimum of
one minute.
6. When you are done, press Stop.
5.10 Leaks
WARNING
Follow all generally-accepted lab safety procedures and applicable laws
and regulations.
Always follow local and national safety regulations and the solvent
manufacturer’s safety, handling, storage, and disposal recommendations;
see solvent manufacturer’s MSDS sheets.
Personnel working with or near the Isolera system must wear protective
clothing, safety gear, and eye protection that have been approved by
applicable local and national safety regulations.
Drip sheet
Figure 5-4. Drainage of Leakage from the Detector Flow Cell (Isolera LS Shown)
Open hatch
Leak tray
5. Check all tubes and connections for leaks. Ensure that the tubes are assembled correctly; see
“Assembling Tubes Correctly” on page 5-9. Use caution when tightening fittings to prevent
stripped threads or crushed ferrules. Replace damaged tubing; a list of spare parts is available
at www.biotage.com.
6. Ensure that the drain tube for the solvent tray is not damaged and is safely connected to the
Drain port at the rear of the system. The other end shall be inserted into a separate waste
reservoir.
Drain port
Drain tube
WARNING
To avoid being struck by the collection arm, keep your hands out of range
of the collection arm while the homing routine runs in step 7 below.
7. Once you have located and taken care of the leakage, reconnect the system and turn it on. The
collection arm moves through its homing routine and the system boots to the system’s main
menu.
8. Check all tubes and connections for leaks using the prime function; see “Prime the System” on
page 4-33. Prime with water or another suitable solvent.
9. If the problem persists, press End Prime, shut down the system (see page 4-1), disconnect
the power cord from the power outlet, and contact Biotage 1-Point Support (see contact details
on page 7-1).
WARNING
Shut down the system before replacing any tubing.
All external tubing on the Isolera systems except for the tubing inside the collection arm can be
replaced by the user; a list of spare parts is available at www.biotage.com.
If using an Isolera system with flangeless tubing, please notice that proper sealing is only achieved
with the ferrule oriented as shown in Figure 5-7, with the tapered portion of the ferrule facing
toward the nut. The ferrule should be placed near the end of the tube as shown below.
If using super flangeless tubing, ensure the ferrules are fitted onto the tube ends as described in
the instructions supplied with the ferrules.
When using (super) flangeless tubing, use caution when finger tightening fittings to prevent
stripped threads or crushed ferrules.
5. Grasp the fuse holder with your fingers and pull it out of the system.
Fuse
Fuse holder
6. Replace the two fuses with new fuses of the same type and rate, 4.0 TA/250 V (P/N 411916).
7. Put the fuse holder back in place.
4. Mount the needle on the collection arm. Ensure that the needle is touching the needle guide.
Mount the
needle here
The needle
should touch the
needle guide
Locking screw
5. Turn the UV lamp into a horizontal position with the lamp socket facing you; see Figure 5-13.
6. Remove the two screws on the lamp socket (see Figure 5-13) using a Torx 10 screwdriver.
Lamp socket
7. Pull out the UV lamp and disconnect it from the power cable; see Figure 5-14.
NOTE
Do no touch the UV lamp glass.
8. Mount a new UV lamp (P/N 09830 for all internal Isolera detectors except UV-VIS and P/N
412970 for UV-VIS). Ensure the power cable is positioned so that it does not get in contact with
the fans at the left side of the UV lamp.
9. If using a UV-VIS detector, replace the tungsten lamp (P/N 412970):
a. Loosen the screw holding the tungsten lamp in position using the wrench delivered with
the system.
b. Pull out the tungsten lamp and disconnect it from the power cable; see Figure 5-15.
c. Mount a new tungsten lamp. Ensure the lamp is fully inserted and the power cable is
positioned so that it does not get in contact with the fans at the left side of the UV lamp.
Press here
and disconnect
the lamp
Locking screw
10. When you are done, put the service panel back in place.
11. Reconnect the system and turn it on.
12. When the main menu appears, log into the System mode (see page 3-1).
13. Ensure that the detector flow cell is clean (see page 5-3) and then perform an intensity
calibration:
a. Select the Maintenance tab in the right-hand panel.
b. Press Calibrate in the UV Detector field. Read and follow the instructions that appear on
the screen.
6 Troubleshooting
WARNING
Before performing any procedures in this chapter, please read and observe
the safety requirements in the “Isolera Installation and Safety” document
(P/N 411828). Failure to follow those requirements may result in personal
injury and/or equipment damage. If the system has been damaged and
does not function properly, shut it down and contact Biotage 1-Point
Support immediately.
1. The collection arm a. Misaligned rack(s). Ensure that the rack(s) is/are aligned correctly.
does not position
correctly over each
collection vessel.
b. Wrong rack type Ensure that you select the correct rack type when
selected in the setting up a method; see “Create, Open, and Edit
software. Methods” on page 4-2.
c. Incorrect rack Ensure that the correct rack parameters have been
parameters. (Only if entered for the used rack type; see “Administrate
using a user-defined the Rack List” on page 2-5.
rack type.)
d. The collection arm is Remove any obstacles that obstruct or restrict arm
obstructed. movement.
2. Dripping needle Dirty collect valve. Clean the collect valve with methanol or a similar
and/or inconsistent solvent using the Collect Valve Clean Wizard; see
dispensing page 3-13.
volumes.
c. One or more solvent Sonicate or replace the solvent inlet filters; see
inlet filters are page 5-6.
plugged. Re-circulating the solvent is not recommended.
Particulate-free solvent is required.
If the problem persists, the pump piston seals or the
check valve seals may be worn. Please contact Biotage
1-Point Support; see contact details on page 7-1.
2. Low or inconsistent a. One or more pump 1. Flush the pump check valves with methanol or a
flow delivery volume check valves are not similar solvent using the Check Valves Release
and/or functioning properly. Wizard; see page 3-14.
superimposed 2. Check for leaks in the tubing or fittings; see “Leaks”
periodic detector on page 5-7.
signals. 3. If the problem persists, the pump piston seals or
the check valve seals may be worn. Please contact
Biotage 1-Point Support; see contact details on
page 7-1.
b. One or more pump 1. Flush the pump check valves with methanol or a
check valves are not similar solvent using the Check Valves Release
functioning properly Wizard; see page 3-14.
due to particulates 2. If the problem persists, please contact Biotage
in the solvent. 1-Point Support (see page 7-1).
b. Stuck pump check Flush the pump check valves with methanol or a similar
valve(s). solvent using the Check Valves Release Wizard; see
page 3-14.
1. Noise. a. Dirty flow cell. Clean the flow cell; see page 5-3.
b. Flow cell incorrectly Open the retaining latch and turn the flow cell upside
inserted. down.
2. Drifting baseline. a. Dirty flow cell. Clean the flow cell; see page 5-3.
3. Missing peaks when Solvent(s) are high- If expected peaks do not show when you have the
using baseline absorbing over a wide Baseline Correction option turned on and are using
correction. range of wavelengths. solvents that are high-absorbing over a wide range of
wavelengths (e.g. acetone, and toluene), try
performing the run without baseline correction.
4. The UV Detector a. Dirty flow cell. Clean the flow cell; see page 5-3.
Failure dialog
appears during auto b. Highly absorbing Choose less absorbing solvent(s).
or manual UV Zero. solvent(s).
c. Sample in the flow Retry performing the UV Zero when there is no sample
cell. in the flow cell.
If the problem persists, please contact Biotage 1-Point
Support (see page 7-1).
Problem Solutions
Low composition When generating solvent mixtures where the composition contains less than 10%
gradient is not correct. of one solvent:
1. Ensure that the solvent with the lowest percentage is pumped first, i.e. is
mounted on a solvent inlet with a lower number than the solvent with the
highest percentage. Use for example inlet S1 for the solvent with the lowest
percentage and S2 for the other one.
2. If the problem persists, premix the solvent using the desired final % strong
solvent in the weak solvent and use this as solvent B. Program the gradient
from 0–100% B using the pre-mixed solvent B. See “Create, Open, and Edit
Methods” on page 4-2.
The baseline drift is Two factors contribute to altering the gradient as observed by the detector on the
different from the output side of the cartridge:
programmed gradient. 1. As it takes at least 1 CV for the solvent to pass through the cartridge and
reach the detector, the initial front of the gradient will always be delayed by at
least 1 CV compared to the programmed gradient. A longer gradient delay
may be due to interactions with the silica, where strong solvent is being
selectively retained on the cartridge.
2. The gradient as observed by the detector will at times decline and plateau
before the programmed gradient does. This is due to the detector reaching
saturation for the wavelengths absorbed by the strong solvent. Any increase
in the concentration of strong solvent will not be registered as no light is
penetrating the solvent at those particular wavelengths.
If using a system with a Spektra license installed, try performing the run with the
Baseline Correction option turned on.
Extremely jagged As the system only reports absorption in whole numbers (in mAU), a peak
peaks at absorption recorded at a low level will suffer from quantification noise. This problem may be
levels below 50 mAU. solved by increasing the path length of the UV cell. This increases the sensitivity
at the cost of a lower saturation concentration. For more information, please
contact 1-Point Support.
WARNING
To avoid being struck by the collection arm, keep your
hands out of range of the collection arm while the
homing routine runs in step f below.
f. Turn on the system. The collection arm moves through its homing routine
and the system boots to the system’s main menu.
2. The flow rate is too high a. Press Purge... in the alarm dialog.
for the purification or b. When the pressure has been reduced to an acceptable level, press End in
prime. the Purge dialog.
c. Press Pause in the alarm dialog.
d. If the overpressure occurred during a prime: a) Press End Prime at the
Prime tab and start a new prime with a lower flow rate.
e. If the overpressure occurred during an equilibration or a gradient run:
a) Press Edit... at the Status tab. The Edit Method dialog opens.
b) Lower the flow rate and press OK.
c) To resume the run, press Resume in the right-hand panel.
WARNING
To avoid being struck by the collection arm, keep your
hands out of range of the collection arm while the
homing routine runs in step g below.
g. Turn on the system. The collection arm moves through its homing routine
and the system boots to the system’s main menu.
h. Prime the system using e.g. methanol; see page 4-33.
7 Contact Information
Manufacturer
PartnerTech Åtvidaberg AB for Biotage Sweden AB
Biotage Sweden AB
Box 8
SE-751 03 Uppsala
Sweden
Visiting address: Vimpelgatan 5
Telephone: +46(0)18-56 59 00
Fax: +46 18 59 19 22
E-mail: [email protected]
Website: www.biotage.com
Technical Support
Call, fax, or e-mail Biotage 1-Point Support, or request support online (www.biotage.com).
When contacting 1-Point Support, please have the following information ready:
• Company name and contact information.
• Serial number of your Isolera system (see the product label at the rear of the system).
• A brief description of the symptoms or technical problems you are experiencing.
This appendix describes the details of the different collection and fractionation methods that are
available for the Isolera system. It also contains information on how gradients are automatically
extended.
4. Manual: The collection is started and stopped by the user. Fractionation only occurs on
volume.
• External detector: The system uses the signal from the external detector for collection only.
Fractionation Method
Collection
Method
Manual Volume Valley Shoulder
Manual X X
Collect All X X X
Threshold X X X
Slope X X X X
Manual Collection
When using the Manual collection method, collection is started and stopped manually and
fractionation occurs on volume, which is when a test tube or bottle is full (based on the specified
max fraction volume for the vessel type). You can also at any time switch to a new collection vessel
manually by pressing New Fraction at the Status tab.
Collect All
The Collect All collection method fractionates using the same principles as the Manual method as
well as when a valley is detected in the light absorption curve, if the Valley Slope parameter is
enabled (in the System mode).
Threshold
The Threshold collection method fractionates using the same basic principles as the Collect All
method. When the collection is started and stopped depends on the signal(s) from the detector.
When a dual detector is used with the UV1+UV2 detection mode, collection is started when one of
the signals used for collection and fractionation reaches the specified Start Threshold level.
Fractionations occur when the other signal reaches the Start Threshold level and then when one of
the signals decreases below the Start Threshold level. When both signals decrease below the Start
Threshold level, the collection is stopped.
Slope
The Slope collection method fractionates using the same basic principles as the Collect All method.
When the collection is started depends on the calculated slope that is based on the signal from the
detector. The collection is stopped when the signal is below the absorption level detected when the
collection was started.
When a dual detector is used with the UV1+UV2 detection mode, collection is started when the
calculated slope of one of the signals used for collection and fractionation reaches the specified
Start Slope setting. Fractionations occur when the other signal’s calculated slope reaches the Start
Slope setting and then when one of the signals decreases below the absorption level detected when
the collection was started. When both signals decrease below the level detected when the collection
was started, the collection is stopped. The Slope collection method also enables the possibility to
fractionate when shoulders are detected in the light absorption curve.
It is possible to combine the Collect All with the Threshold and/or the Slope collection method. This
means that everything will be collected and fractionation will occur on the Start Threshold level
and/or the Start Slope setting. Fractionation will also occur on detected valleys as well as on
shoulders, if enabled.
Note that the Slope collection mode is not available with the λ-all detection mode or with an external
detector.
Parameter Description
Detection Mode The basic configuration for collection and fractionation. The range is [UV1 |
UV1+UV2 | λ-all | Manual | External]. The λ-all detection mode is only available
on systems with a Spektra license installed.
Baseline Correction Used to remove the interference of the light absorbance of the used solvents from
the absorbance signal during the chromatography. (Only available on systems
with a Spektra license installed.)
If this option is turned on, the gradient run is preceded by a short solvent light
absorbance detection phase. During this phase, the light absorbance of the used
solvents is measured at all wavelengths that will be used for collection and
fractionation (UV1, UV1+UV2, or λ-all). The measurement results in a baseline
containing the maximum absorbance of the solvent system at each wavelength
used, a solvent absorbance spectrum. During the gradient run, the maximum
solvent absorbance spectrum is subtracted from the signal presented by the
detector. Only absorbance values above zero are reported, anything below that is
reported as zero. In effect this automates the procedure of manually excluding
parts of the spectrum from the summation. In a normal use case, peaks that have
absorbance overlapping that of the solvent are detected because of absorbance
outside of the excluded range.
Start Wavelength The shortest wavelength to be included in the λ-all signal. The range is 200 to 400
nm (dual variable UV detector) or 200 to 800 nm (UV-VIS detector). (Only
available with the λ-all detection mode.)
End Wavelength The longest wavelength to be included in the λ-all signal. The range is 200 to 400
nm (dual variable UV detector) or 200 to 800 nm (UV-VIS detector). (Only
available with the λ-all detection mode.)
Collect All Used to collect the entire run. The range is [On | Off].
Start Threshold Used to collect samples with a light absorbance exceeding the set absorbance
threshold. The range is [Off | 0 to 6000 mAU].
Initial Waste Delay collection until a specified volume has been delivered to the waste.
The range is [0 to 10000 ml]. If the gradient is defined in CV, then the initial waste
is also defined in CV.
Slope Mode Used to collect samples based on the start slope of light peaks. The range is [Off |
Low Slope | Medium Slope | Custom Slope]. (Only available with the UV1 and
UV1+UV2 detection modes.)
Parameter Description
Start Slope The value of the collect start slope. The range is [Off | 1 to 5000 mAU/CV]. (Only
available with the UV1 and UV1+UV2 detection modes.)
Shoulder Slope The value of the fractionation shoulder slope. The range is [Off | 1 to 5000 mAU/CV].
(Only available with the UV1 and UV1+UV2 detection modes.)
System Parameters
The system parameters are defined for all methods and can be changed (in the System mode) only
by a user with system owner privilege. The following system parameters are available:
Parameter Description
Start Slope Enable The level where the parameter Start Slope is enabled. Below this level the Start
Slope parameter is ignored. The range is [0 to 6000 mAU]. The default value is
30 mAU.
Shoulder/Valley Slope The level where the parameters Shoulder Slope and Valley Slope are disabled.
Disable Above this level the Shoulder Slope and Valley Slope parameters are ignored. The
range is [0 to 6000 mAU]. The default value is 5500 mAU.
Valley Slope The value of the fractionation valley slope. The range is [Off | 1 to 5000 mAU/CV].
The default value is 150 mAU/CV.
Manual If the current calculated slope and/or absorption level is greater than the predefined
system settings for auto extend, the gradient is extended.*
Collect All If the current calculated slope and/or absorption level is greater than the predefined
system settings for auto extend, the gradient is extended.*
Threshold If the current absorption level is greater than the threshold defined in the method
and/or the current calculated slope is greater than the predefined slope for auto
extend*, the gradient is extended.
Slope If the system is collecting, the current calculated slope is greater than the slope
defined in the method, and/or the current absorption level is above the predefined
threshold for auto extend*, the gradient is extended.
* Auto extend can be enabled or disabled in the system mode. The predefined system settings for auto
extend are: Threshold = 100 mAU and Slope = 180 mAU/CV.
If the method in use is a combination of two or more of the methods above, the gradient is extended
if one or more of the methods’ criteria are meet. If a dual variable or UV-VIS detector is used with
the UV1+UV2 detection mode, both detector signals used for collection and fractionation are
checked for a possible auto extend of the gradient.
Japan Japan
Phone: +81 3 5627 3123 Phone: +81 3 5627 3123
Fax: +81 3 5627 3121 Fax: +81 3 5627 3121
[email protected] [email protected]
Part Number: 411829-L