Lab Quality Manual
Lab Quality Manual
Lab Quality Manual
DEQ91-LAB-0006-QMP
Version 5.0 – November 2004
Laboratory Division
1712 SW 11th Avenue
Portland, OR 97201
Phone: (503) 229-5983
Fax: (503) 229-6924
www.deq.state.or.us
This Page Intentionally Left Blank
Laboratory Quality Manual Oregon Department of Environmental Quality
DEQ91-LAB-0006-QMP November 2004
Version 5.0 Page iii
Concurrences:
Ron Doughten, Land Quality and Agency Quality Assurance Officer Date
Chris Redman, Water Quality and Laboratory Quality Assurance Officer Date
Laboratory Quality Manual Oregon Department of Environmental Quality
DEQ91-LAB-0006-QMP November 2004
Version 5.0 Page iv
Table of Contents
Laboratory Quality Manual ............................................................................................................. i
Concurrences:.............................................................................................................................. iii
Table of Contents.......................................................................................................................... v
Acronyms and Abbreviations ...................................................................................................... vii
Glossary ......................................................................................................................................viii
1 Quality System ..................................................................................................................1
1.1 Introduction:...................................................................................................................1
1.2 DEQ's Quality Management Policy ...............................................................................2
2 Laboratory Management Structure....................................................................................4
2.1 Laboratory Quality Assurance Team .............................................................................4
2.2 Administration................................................................................................................5
2.3 Air Quality Monitoring ....................................................................................................5
2.4 Watershed Assessment ................................................................................................5
2.5 Inorganic Laboratory .....................................................................................................5
2.6 Organic Laboratory........................................................................................................6
2.7 Technical Services ........................................................................................................6
2.8 Personnel ......................................................................................................................6
3 Standard Operating Procedures......................................................................................12
3.1 Sampling .....................................................................................................................12
3.2 Sample Handling .........................................................................................................13
3.2.1 Sample Identification ...............................................................................................13
3.2.2 Sample Receipt .......................................................................................................13
3.2.3 Sample Integrity.......................................................................................................14
3.3 Test Methods and Method Validation..........................................................................15
3.3.1 Analytical SOPs.......................................................................................................15
3.3.2 Demonstrations of Capability...................................................................................16
3.3.3 Data Validation ........................................................................................................16
3.3.4 Estimation of Uncertainty of Measurement..............................................................17
3.3.5 Data Control ............................................................................................................17
4 Measurement Traceability ...............................................................................................19
4.1 Documentation ............................................................................................................19
4.2 Purchasing Services and Supplies..............................................................................19
4.2.1 Request for Purchase..............................................................................................20
4.2.2 Purchase .................................................................................................................20
Laboratory Quality Manual Oregon Department of Environmental quality
DEQ91-LAB-0006-QMP November 2004
Version 5.0 Table of Contents cont. Page vi
Glossary1
Accuracy: the degree of agreement between an observed value and an
accepted reference value. Accuracy includes a combination
of random error (precision) and systematic error (bias)
components which are due to sampling and analytical
operations; a data quality indicator. (QAMS)
Analyst: the designated individual who performs the "hands-on"
analytical methods and associated techniques and who is the
one responsible for applying required laboratory practices and
other pertinent quality controls to meet the required level of
quality. (NELAC)
Audit: a systematic evaluation to determine the conformance to
quantitative and qualitative specifications of some operational
function or activity. (EPA-QAD)
Batch: environmental samples that are prepared and/or analyzed
together with the same process and personnel, using the
same lot(s) of reagents. A preparation batch is composed of
one to 20 environmental samples of the same NELAC-defined
matrix, meeting the above mentioned criteria and with a
maximum time between the start of processing of the first and
last sample in the batch to be 24 hours. An analytical batch is
composed of prepared environmental samples (extracts,
digestates or concentrates) which are analyzed together as a
group. An analytical batch can include prepared samples
originating from various environmental matrices and can
exceed 20 samples. (NELAC Quality Systems Committee)
Chain of Custody Form: record that documents the possession of the samples from the
time of collection to receipt in the laboratory. This record
generally includes: the number and types of containers; the
mode of collection; collector; time of collection; preservation;
and requested analyses. (NELAC)
Clean Air Act: the enabling legislation in 42 U.S.C. 7401 et seq., Public Law
91-604, 84 Stat. 1676 Pub. L. 95-95, 91 Stat., 685 and Pub.
L. 95-190, 91 Stat., 1399, as amended, empowering EPA and
its delegates to promulgate air quality standards, monitor and
to enforce them. (NELAC/DEQ)
Comprehensive the enabling legislation in 42 U.S.C. 9601-9675 et seq., as
Environmental amended by the Superfund Amendments and Reauthorization
Response, Act of 1986 (SARA), 42 U.S.C. 9601et seq., to eliminate the
Compensation and health and environmental threats posed by hazardous waste
Liability Act sites. (NELAC)
(CERCLA/Superfund):
1
Sources in ( ) are presented at the end of the glossary.
Laboratory Quality Manual Oregon Department of Environmental quality
DEQ91-LAB-0006-QMP November 2004
Version 5.0 Glossary cont. Page ix
1 Quality System
1.1 Introduction:
The Oregon Department of Environmental Quality (DEQ) was formally established in 1969 as a
product of a 1938 citizen initiative known as the Water Purification and Prevention of Pollution
Bill. In 1970 the Environmental Protection Agency (EPA) was created thereby making available
federal funds for environmental programs. The DEQ receives a significant portion of its funding
through the EPA; consequently many of the DEQ’s programs are designed to meet EPA
requirements. To qualify for this federal funding the EPA has made specific requirements of the
recipients, such as having a Quality System, which must be documented in the organizations
Quality Management Plan (QMP). The scope of the QMP is defined by EPA’s memorandum
5360 A1.
The QMP is a policy statement describing how an EPA organization shall comply with the
requirements of EPA Order 5360.1 CHG 2. Quality systems encompass the management and
technical activities necessary to plan, implement, and assess the effectiveness of QA and QC
operations applied to environmental programs. The QMP provides the blueprint for how an
individual EPA Program Office, Region, and National Laboratory or Center EPA Quality Manual
for Environmental Programs will plan, implement, and assess its quality system for the
environmental work to be performed as part of its mission. QMPs are reviewed and approved by
the OEI. Approval is valid for a period of up to five years.2
The DEQ first wrote its Quality Management Plan in 1982 to meet the standards at the time.
The DEQ’s latest QMP can be found on Q-Net and is intended to meet EPA’s current (R-2)
standard. The QMP describes the DEQ’s policies, objectives, principles, authority responsibility
and implementation of the agency’s Quality Management system. EPA has signed and
approved the DEQ’s QMP allowing the EPA programs to grant funds and accept data from the
DEQ. EPA requires the QMP be signed by the agency’s senior management. DEQ’s Executive
Management Team (EMT), which includes the director, the deputy director, and division
administrators have signed the QMP giving their concurrence. This commitment by top
management is essential for the success of the agency’s Quality System.
The QMP describes the use of other Quality Systems documents including this manual. The
Laboratory Quality Manual (LQM) shall describe the policies and, where appropriate,
procedures for all personnel within the Laboratory Division to follow. The LQM is not new to the
Laboratory. Its history is similar to that of the QMP and was formerly titled the DEQ Laboratory
Quality Assurance Manual. The policies and procedures in the LQM shall guide laboratory
personnel in collecting, producing, maintaining, and reporting data of known quality. The LQM
is continuously evolving with the discovery of omissions or undocumented policies, changes in
the agency’s Quality System, and adoption of new quality standards. The Laboratory Quality
Assurance Officer is responsible for the review and revision of the LQM. Laboratory Division
personnel must keep abreast of changes in quality policy and procedures, and therefore, must
read the changes to the LQM with each revision.
The QMP also describes the agency’s commitment to use National Environmental Laboratory
Accreditation Conference (NELAC) standards to evaluate a laboratory’s ability to generate
quality data. This LQM is written to meet NELAC standards and will be reviewed and revised as
required by the adoption of revised NELAC standards. NELAC standards may be revised at the
annual meeting, which may require annual modifications to the Laboratory Quality Manual,
however as the standards become more refined fewer and fewer changes shall come about.
NELAC standards are adopted two years after they are approved at the annual meeting, thus
2
EPA Quality Manual for Environmental Programs 5360 A1, 5/5/2000 US EPA, pg 2-8.
Laboratory Quality Manual Oregon Department of Environmental Quality
DEQ91-LAB-0006-QMP November 2004
Version 5.0 Quality System cont. Page 2 of 64
allowing laboratories and accrediting authorities the opportunity to make the appropriate
adjustments in their systems. NELAC 2002 was used for the revision of this DEQ LQM.
3
“QUALITY MANAGEMENT PLAN FOR THE OREGON DEPARTMENT OF ENVIRONMENTAL
QUALITY”, introduction and chapter 1.1 DEQ’s Management Policy.
Laboratory Quality Manual Oregon Department of Environmental Quality
DEQ91-LAB-0006-QMP November 2004
Version 5.0 Quality System cont. Page 3 of 64
• the implementation of proper procedures for sample collection, storage, preservation, sample
tracking, analysis, and reporting (Chapter 3 Standard Operating Procedures);
• where applicable data is traceable to acceptable reference standards (Chapter 4 Measurement
Traceability);
• the degree of precision, accuracy, and bias of the analyses is known and documented (Chapter 5
Assuring the Quality of Analytical Data);
• that analytical equipment is properly used, calibrated, and maintained (Chapter 6 Equipment);
• all items influencing the quality of data are properly documented (Chapter 7 Quality Assurance);
• data is reported in useful and comparable formats (Chapter 8 Reporting Results).
Laboratory Quality Manual Oregon Department of Environmental quality
DEQ91-LAB-0006-QMP November 2004
Version 5.0 Page 4 of 64
develop systems to help prevent quality related problems. Laboratory personnel are
encouraged to discuss Quality Management issues with the QAT members. When the QAT
recognizes the need to change or adopt new quality polices or practices, the QAT shall prepare
an action plan to ensure that the changes are implemented and monitored. The QAT meeting
minutes shall include the perceived need of a plan, the procedure for initiating the plan, and
application of controls to ensure the preventative measure is effective.
2.2 Administration
The administrative section of the laboratory provides support for the laboratory in the areas of
Human Resources/Accounting, special projects, and Quality Management.
An executive support specialist offers administrative support to all sections of the laboratory in
addition to the administrator (e.g. travel and time accounting) and serves as a liaison to PSU
facilities.
Special projects include updating monitoring strategies, assisting with laboratory and monitoring
operational analysis, legislative coordination for the Lab Division (including grant applications
and reports), and assistance to the Administrator on unanticipated requests from outside
stakeholders or colleagues.
The oversight of quality assurance issues is delegated to three QAOs. There is a QAO for each
DEQ program: Air Quality, Land Quality, and Water Quality. The QAOs provide technical
assistance in the development and implementation of QA project plans; audits monitoring
network; performs assessments of self-monitoring activities under air, NPDES, and RCRA
permits; participates in Oregon Laboratory Accreditation Program (ORELAP) activities; reports
to programs documenting project data quality; and ensures corrective action procedures are
followed when data quality criteria are not met. Additionally one Quality Assurance Officer shall
be assigned the responsibility of ensuring that NELAC standards are implemented at the DEQ
laboratory and is recognized in this document as the Laboratory Quality Assurance Officer
(LQAO).
2.8 Personnel
The DEQ laboratory hires employees with the necessary training and experience through the
recruiting procedures required by the State of Oregon. The minimum qualifications of agency
positions are defined by Oregon State Division of Administrative Services (DAS). The DEQ
must adhere to strict and consistent processing of all recruitments. A prospective employee
must complete a State application form where he/she lists his/her qualifications and certify the
information he/she gives as true and complete. All applications and recruitment materials are
kept in the Human Resource office in confidential files. These files are available for internal
review upon request. In order for an applicant to advance to the interview stage of a
recruitment, his/her application is first reviewed by the agency’s Human Resources department
to insure he/she meets the minimum qualifications for the classification as determined by DAS
(refer to Table 1). An applicant is required to meet the established minimum qualifications in
order to proceed in the recruiting process. The DEQ laboratory management shall assume that
any person who has been selected for the opportunity to interview for a particular position has
successfully met or exceeded the minimum qualifications required for that position.
Many of the skilled laboratory positions require special training. Management shall ensure
personnel receive the special training required for these positions. This training shall be
documented and submitted to the LQAO. During an internal audit the QAO shall review these
records to ensure personnel have had the training to perform their duties as required by
programs, NELAC standards, and laboratory policy.
Laboratory Quality Manual Oregon Department of Environmental quality
DEQ91-LAB-0006-QMP November 2004
Version 5.0 Laboratory Management Structure cont. Page 7 of 64
3.1 Sampling
The DEQ laboratory primarily conducts analytical tests on samples collected by the different
regions of the agency and by the Laboratory sections of Watershed Assessment (WA) and Air
Quality Monitoring (AQM). WA has prepared a Mode of Operations Manual (MOM) to cover
sampling procedures and analytical work performed in the field. AQM staff have prepared
SOPs for each of the sampling techniques they use. Sampling procedures shall instruct
personnel on how to collect representative samples, record data, and when applicable submit
properly preserved samples to the laboratory for further analysis. Data recorded during
sampling should include the sampling procedure used, the identification of sampling equipment
and personnel, environmental conditions (if relevant), the time sample collection started and
stopped, the date, and the sampling location.
The LQAO shall maintain the Field Sampling Reference Guide, which instructs personnel
outside the laboratory on how to collect and submit samples to the DEQ laboratory. The
Technical Services section shall make these documents available on Q-Net.
The laboratory must have on file copies of Quality Assurance Project Plans (QAPPs) as
required by the agency’s QMP. The QAPP should describe the methods used to collect
samples, ensuring the quality of the data and, if appropriate, the statistics used to develop the
sampling procedures. The QAPP shall provide sampling information or the appropriate
template for writing a Sampling & Analysis Plan (SAP). SAPs are a subgroup of the larger more
encompassing QAPP. There are cases where the QAPP shall describe the elements within the
Laboratory Quality Manual Oregon Department of Environmental quality
DEQ91-LAB-0006-QMP November 2004
Version 5.0 Standard Operating Procedures cont. Page 13 of 64
SAP and thus a SAP may not be necessary for all samples. SAPs are also controlled
documents (refer to the agency’s QMP for more information). The SAP should primarily
describe the schedule for collecting samples, the location of the sample, and the requested test
methods. The laboratory should receive the SAP and/or QAPP before sampling begins so the
different sections of the laboratory can make the necessary arrangements to perform the work.
If the project coordinator requires a deviation from the QAPP, a SAP may describe the deviation
without revising the QAPP. Technical Services shall devise a system for ensuring that all
appropriate personnel are advised of SAPs and that they shall have access to SAPs.
Upon receipt of samples, the sample tracker shall fill out a "Sample Receipt Checklist" that will
identify the number of shipping containers, the internal temperature of those containers, the
condition of the samples, the status of preservation checks, and whether or not the samples
were shipped and received with a Chain-of-Custody form. Any abnormalities or departures from
acceptable conditions as described in the QAPP or SOPs shall be noted on the checklist. This
checklist may be used in lieu of the Nonconformance forms in Appendix C, however the
procedures in section 7.5.3 below must be followed to ensure the proper information is collected
and documented. The sample tracker must consult with the project coordinator and document
all correspondence to determine the course of action, i.e. to resample, void, or proceed with
sample analysis. The project coordinator’s instructions must be determined prior to proceeding
with the analyses. The sample tracker’s Sample Receiving procedure shall be provided to
project coordinators during the development of their QAPP so they may add any specific needs
for handling samples. The project coordinator shall either accept the sample trackers procedure
or address procedures in the QAPP for avoiding deterioration, contamination, loss or damage to
the sample during storage, handling, preparation and testing. Without explicit sample handling
procedures described in the QAPP, the laboratory shall assume that there is no deviation from
the tracker’s Sample Receiving SOP. Should the sample tracker discover special sample
handling instructions at the time of receipt, the sample tracker shall communicate the sample
handling deviation to laboratory managers using the proper Nonconformance documentation
procedures (refer to Appendix C: Nonconformance Report).
The laboratory uses a special Legal Chain Of Custody procedure for handling samples collected
for the use of enforcement investigations. This procedure documents the location of the sample
and who has removed it from its locked location to perform his/her work. The sample tracker
shall record the “Legal” status of the sample in LIMS. The tracker, Technical Services manager,
and the assistant to the laboratory director shall be the only personnel to have a key to the
locked “Legal” storage areas.
b) Test method SOPs shall describe the process for the disposal of digestates,
leachates and extracts or other sample preparation products.
c) The Technical Services section shall maintain the SOP for the disposal of samples.
Samples identified as Hazardous Waste shall be collected for disposal at an
appropriate facility, refer to the laboratory’s Chemical Waste Management SOP
(DEQ04-LAB-0014-SOP).
4) After logging in samples the tracker shall notify appropriate personnel of the receipt of
samples with short holding times (for methods with holding times of less than or equal to
48 hours refer to the sample tracker’s Sample Receiving SOP). Chemists shall begin
these test methods within the cited holding times.
Personnel who find that any of the above sample integrity policies has been compromised shall
initiate the corrective action procedure described in this document and begin the completion of
the Nonconformance Report. In most cases the data shall be reported as an estimate with an
appropriate comment attached to the result.
recorded and tracked in LIMS by creating a unique “Standard Parameter” for the new
procedure.
During the development of the QAPP, the QAO shall advise the project coordinator if test
methods appear to be inappropriate or out of date. If a decision is made to continue with an
undocumented procedure, the DEQ laboratory will then develop its own procedure. The QAO
shall assign the task of creating the new SOP to the appropriate laboratory section. The section
manager shall delegate the responsibility for developing the method and writing the SOP to
appropriate personnel.
The analytical sections of the laboratory shall play the most significant role in the development
and documentation of new test methods; however, they must maintain open communication
with the QAO and management. The development of new methods shall be a planned activity,
which will require routine meetings between analytical staff, management, and the QAO. The
meeting minutes shall become part of the SOP control document file. The meetings shall cover
the process of developing the method and its progress. QAT members shall put method
development on their meeting agenda to share the progress of their new methods with other
sections of the laboratory. New methods will be based on the laboratory’s best available
technologies and preferably on test methods published by reputable organizations or instrument
manufacturers. Communication between project coordinators and QAO is essential to ensure
laboratory procedures conform to program requirements.
The Laboratory will not provide detailed data quality assessment services to the Agency by
default. Generally, Project Coordinators sending samples to the Laboratory shall be responsible
for the statistical validation of data for their projects. Project Coordinators submitting samples to
the lab should first identify their Data Quality Objectives (e.g. verifying a null hypothesis) and
establish the appropriate the statistical approach for assessing the data set. Laboratory
assistance with data quality assessments, can be arranged between the project coordinator and
a QAO or the Technical Services section manager. Data quality concerns should be reported to
the QAOs so that DQOs can be refined in future QAPPs. Project Coordinators are encouraged
to request QC data with the analytical report, so they can more readily evaluate the
effectiveness of the project.
The Technical Services section shall develop a SOP for ensuring the data integrity of electronic
data. This SOP shall describe the process used to maintain, document, and store data ensuring
that it can not be altered, is transmitted correctly, and processing does not produce
misinformation. Data shall be stored such that security can be monitored and access controlled.
Laboratory Quality Manual Oregon Department of Environmental quality
DEQ91-LAB-0006-QMP November 2004
Version 5.0 Page 19 of 64
4 Measurement Traceability
The laboratory shall ensure analytical measurements are traceable to acceptable reference
standards.
4.1 Documentation
In order to ensure data is of known quality, laboratory staff must be able to document the source
and tolerance levels of reference standards, and reagents. To the best of its ability, the
laboratory shall use National Institute of Standards and Technology (NIST) traceable standards
and contract services that use NIST traceable standards. The laboratory must be able to show
a paper trail from the test result of a sample back to the NIST standard for the analyte and all
measurement devices used in the analysis. Analysts performing analytical work shall follow the
laboratory’s documentation procedures ensuring that analytical results are easily linked to
calibration data, which is linked to reference material certificates described in Purchasing
Services and Supplies section below.
There are occasions when certificates are not available in which case the analyst must validate
the suitability of the reference. The analyst should obtain approval from his/her section
manager prior to testing the suitability of a product, since the Laboratory should be able to
purchase materials with certificates. This suitability testing procedure must be documented and
approved by the Quality Assurance Team (QAT) or test method signatories. As a final measure
for verifying the suitability of uncertified materials, the analyst must evaluate the result of a
Laboratory Control Sample (LCS) using the material in question. Results of the suitability
testing shall be documented as other test method results, thus the first occurrence of the
product identification number in the analytical records will be that of the suitability test.
• Glassware, Chemicals, Reagent Water, Gases, and Reagents
The laboratory will purchase supplies of the highest quality needed to ensure minimal
interference or contamination with a procedure. As appropriate Chemists and Technicians will:
• use “Class A” volumetric glassware for the preparation and dilution of reagents,
standards, and samples;
• ensure non-volumetric glassware is of an appropriate quality;
• ensure compressed gases are of known purity and guaranteed by the supplier;
• ensure chemicals are dated on receipt, stored according to chemical properties, and
discarded when shelf life is exceeded (for chemicals where shelf life is defined);
• ensure solvents employed in organic analyses are “HPLC” or “Pesticide grade” and
stored in ventilated explosion-proof cabinets; and,
• ensure analytical reagents or solvents are never stored with samples awaiting analysis.
The analyst shall document reagent preparation. Documentation shall include the source of the
reagent, the mass or volume used, and dilution information. The test method SOP should
contain specific instructions for reagent preparation and documentation. Reagents shall be
prepared from “Analytical Reagent” grade (AR) or higher purity chemicals as required by the
method, and shall be stored as recommended in the method, or by the chemical manufacturer.
not directly related to data quality. Whereas the quality of chemical reagents, concentration of
standards, instrumentation, and instrument service contracts do have an impact on the quality of
data reported. In order to ensure the integrity of data, laboratory staff must follow these
procedures for purchasing services and supplies.
It is the responsibility of the Procurement & Contract Specialist (PCS) to purchase, receive,
label appropriate supplies, and to dispose of unused obsolete equipment and supplies by
following the State Surplus procedures and completing the Property Disposition Request (PDR)
forms. The Technical Services section manager shall ensure the PCS is properly trained and
instructed to apply the following Request for Purchase SOP to all chemicals, measuring
devices, and service contracts.
4.2.2 Purchase
All purchases have a formal purchase approval process. Expenses over an amount set by the
Oregon State Division of Administrative Services (DAS) must be approved by DAS. Less
expensive orders may be approved by the DEQ’s accounting office or laboratory personnel.
The triggers, which identify who is to grant final approval of a purchase, change with time and
the PCS shall stay informed of the current policy.
Many purchases are performed using credit cards. Such purchases must be approved by the
credit card holder and the section manager. Since these purchases may occur without prior
consent, personnel should take caution and management shall inform personnel of the current
credit card policy. Personnel may be held responsible for inappropriate purchases.
Laboratory Quality Manual Oregon Department of Environmental quality
DEQ91-LAB-0006-QMP November 2004
Version 5.0 Measurement Traceability cont. Page 21 of 64
To ensure products directly related to data quality are tracked; personnel must copy the
Requisition by printing it from the database or by filing out the paper form. Products that shall
be tracked are often of sufficient value to trigger the requirement to complete a Requisition;
however, laboratory staff may use petty cash for inexpensive supplies. Personnel must be
mindful of the requirement that certain products must be tracked and the PCS must have a copy
of the Requisition to ensure data for such products are entered into the system.
4.2.3 Receipt
Upon receipt of any product the laboratory receptionist shall notify the PCS of its arrival. The
PCS shall identify the product and locate a copy of the Requisition form and notify the staff
member who requested the purchase. The purchaser shall inspect the product for consistency
with the order and possible shipping damage. The purchaser shall also verify the quality of any
chemical received and verify that the appropriate certificate of analysis was sent. If the
purchaser finds no problems, he/she shall sign the receipt and return it to the PCS. The
recipient of contracted services shall request copies of certificates from the contractor to
maintain the laboratory’s traceability requirements. If there is a problem with the order the PCS
shall take appropriate steps to reverse the order or the analyst may attempt to verify the
suitability of the product (refer to section 4 Measurement Traceability).
4.2.4 Labeling
If the product is acceptable the PCS shall label the product with a unique identification number.
The identification number will be logged into the Requisition database and transcribed to the
Requisition form, receipt, and accompanying quality certificates. The PCS shall file the receipt
and Requisition form in the appropriate binder. The PCS shall forward quality certificates to the
LQAO, who shall file the certificate in the Quality Document file. The requisition binders and
Quality Document files shall be retained for at least five years.
4.2.5 Storage
The laboratory has very little storage space; personnel are encouraged to make accommodation
for purchases prior to receiving supplies. Chemical reagents and standards must be handled
such that their composition shall not be jeopardized. Some chemicals must be preserved and
should come with special instructions; chemists shall inform the PCS of special handling
procedures when completing their Requisition form. When the section manager signs the
Requisition form, he/she should also look for special handling instructions of which the PCS
should be aware.
Often Material Safety Data Sheet (MSDS) are packaged with chemicals. The purchaser should
forward MSDSs to the laboratory receptionist who will file them in the MSDS notebooks stored
in the centralized Safety area.
should be disposed of through the State Surplus procedures. Personnel should contact the
PCS for the current procedure, which shall include completing the PDR form and moving the
equipment to the holding area so that it can not be inappropriately put back into service.
Chemical supplies have a shelf life and shall not be used beyond the recommended holding
time unless they are tested and proven to still be suitable for use.
Laboratory Quality Manual Oregon Department of Environmental quality
DEQ91-LAB-0006-QMP November 2004
Version 5.0 Page 23 of 64
6 Equipment
The laboratory shall ensure analytical equipment is properly used, calibrated, and maintained
The laboratory is equipped with state-of-the-art analytical instrumentation for analysis of
environmental samples. Appendix E lists the major analytical instrumentation used in the DEQ
Laboratory (this list does not include field instrumentation). Instruments are maintained in
proper operating condition through service contracts on major equipment, and by following
maintenance and calibration schedules. No equipment shall be used without first verifying its
accuracy. Records of this initial verification and the ongoing maintenance performed on
equipment shall be recorded in controlled maintenance/calibration logs (refer to Analytical
Records section 7.2 below). Some equipment such as volumetric glassware may only require
the purchase receipt for documentation of its accuracy. Should the laboratory use equipment
that is not owned by the laboratory, the laboratory shall maintain and calibrate said instrument
as prescribed in this section of the LQM. Records and maintenance logs shall be retained by
the laboratory, even though the equipment may not be retained.
Calibration procedures must be included in the test method SOPs. Calibration requirements are
divided into three parts:
1) requirements for analytical support equipment,
2) requirements for standardizing reagents used for calculating concentration, and
3) requirements for instrument calibration, which is further divided into
a) initial instrument calibration and
b) continuing instrument calibration verification
available. The acceptability for use or continued use shall depend on the needs of the
intended analysis or application.
• Mechanical volumetric dispensing devices including burettes (except Class A glassware)
shall be checked for accuracy on at least a quarterly use basis. If these devices are not
used during the quarter it is not necessary to check their accuracy. Glass micro-liter
syringes are to be considered in the same manner as Class A glassware, but must come
with a certificate attesting to established accuracy or the accuracy must be initially
demonstrated and documented by the laboratory.
• For chemical tests the temperature, cycle time, and pressure of each run of autoclaves must
be documented by the use of appropriate chemical indicators or temperature recorders and
pressure gauges.
also describe how raw data records are to be retained. Data records shall contain sufficient
information to permit the reconstruction of the initial instrument calibration, including:
• calibration date;
• test method;
• instrument;
• analysis date;
• analyte name (or analyst’s initials or signature);
• concentration and response of each standard;
• unique equation or coefficient used to convert instrument responses to analyte
concentration; and
• calibration assessment factor (i.e., r2 or other criteria).
Sample results shall be quantitated from the initial instrument calibration and may not be
calculated from any continuing instrument calibration verification unless otherwise specified in
the QAPP.
With the exception of those test methods that allow for a single point calibration at least three
calibration standards shall be used in the initial calibration not including the calibration blank or
a zero standard. The concentrations of the standards shall be distributed throughout the
calibration range. The lowest calibration standard shall be equal to or less than program
required reporting levels when appropriate and the analytical method reporting limit. If specific
programs are more prescriptive in establishing the initial calibration, those requirements shall be
followed. During the development of the laboratory method, the lowest calibration standard
(MRL) shall be determined by using a dilution of the stock standard that is one to five times
greater than the Method Detection Limit (refer to Appendix B: 40 CFR Part 136, App. B). If
blank samples tend to yield detectable contamination, the lowest calibration standard shall be
set to five to ten times greater than the MDL. Refer to the Laboratory's procedures for
establishing the Method Detection Limit and Method Reporting Limit for additional details.
For analytical methods that have been approved to use a single point calibration, the following
requirements must be satisfied:
1) the linear dynamic range (LDR) of the instrument must be established (and verified
annually) by analyzing a series of standards, beginning at the MRL and extending to the
highest concentration that will be reported without sample dilution;
2) a new calibration (with a calibration blank and single point calibration standard) must be
created for every analytical batch;
3) the calibration curve must be verified immediately following the initial calibration with two
second source quality control samples. One QCS must be analyzed at the MRL and the
second QCS must be analyzed at a second concentration that will not exceed 90% of
the single point standard concentration;
4) Sample concentrations within an analytical batch that exceed the single point calibration
concentration must be handled with one of the following procedures:
a) analysis of a reference material at or above the sample value that meets established
acceptance criteria for validating the linearity;
b) sample dilution such that the result falls below the single point calibration
concentration; or
Laboratory Quality Manual Oregon Department of Environmental quality
DEQ91-LAB-0006-QMP November 2004
Version 5.0 Equipment cont. Page 27 of 64
c) use of an appropriate data qualifier and comment explaining the qualifier flag.
For analytical methods that do not require an initial calibration with every analytical batch, a new
initial calibration shall be prepared whenever there is sufficient change in instrument setup or
reagents used to cause a quantitative change in instrument response, or when QC failures
initiate the Corrective Action procedure from which it is determined the instrument should be
recalibrated. Immediately following the initial calibration a sample called an Initial Calibration
Verification (ICV) shall be run to verify the quality of the calibration. The ICV will be prepared
from a second source standard when possible and meet the same quality as the primary
calibration standard. The SOP shall cite the control limits set on the ICV.
If it is necessary to report data generated from calibrations that do not meet control limits or the
ICV fails to fall within the acceptable range and it is not possible to rerun samples, the analytical
results shall be reported as estimates with comments explaining the problem.
safety procedures necessary for the work performed at the laboratory. Documentation that
personnel have read the CHP shall be kept in the LQAO’s administrative file.
7 Quality Assurance
The laboratory shall provide a system to ensure all events influencing data quality are
documented.
All instrument logs, sample preparation logs, standard/reagent logs, bench work sheets, and
data output records from electronic instruments used to generate analytical data must be
retained for at least five years. To ensure the laboratory meets this requirement all chemists
and field technicians must read and sign memos of attestation for document and record control
procedures.
The document control coordinator shall assign control numbers to logs and notebooks prior to
use. The document control coordinator shall create control numbers in a database and enter
pertinent data describing the content, version, author, and date of the document or records log.
Once the logs and/or notebooks are full, the chemists and technicians must return the records
to the document control coordinator for archiving. Analytical records shall be archived for a
period of five years since the last entry in the log. Refer to the agency’s Archiving procedures,
for storage and retrieval of archived records.
adjusts the status code of the result, which restricts what can be done to the data. Chemists or
technicians enter data into LIMS and may make corrections to data up until they submit their
work for review. Senior chemists review data and may send the data back to the chemist for
rework. LIMS users can not alter data during the review process. Once the results are reported
and approved for all tests performed on a set of samples, Technical Services shall print an
analytical report. The DEQ laboratory shall release the analytical report after the QA Officer and
Laboratory Administrator review and sign the document. The release process is completed
when Technical Services sends a copy of the report (either electronically or paper) to the
primary recipient. Up until that point the report may be edited without an erratum. However,
after the report is sent to the primary recipient it must not be altered. Subsequent changes in an
analytical report must be made through an erratum. Should an error in the report be detected
following the release of the report an erratum must be prepared (refer to DEQ03-LAB-0002-
SOP for this procedure).
As with other controlled documents the colored title page and the signatures on the analytical
report identifies the report as the official controlled copy. Reprinted electronic reports and
photocopies of the analytical report are not controlled. Technical Services shall ensure that
PDF copies of the controlled report are available through electronic means. Data users may
receive PDF copies of the report through e-mail or retrieve them out of Q-Net. Although the
PDF is not the official copy, Technical Services shall ensure the electronic copy of a report is
equally maintained. Technical Services shall store electronic copies of the original reports and
subsequent errata in a secure server location. The process of maintaining an official controlled
document on paper and a secure electronic copy offers a backup system for system failures in
the controlled document procedure.
During data review, lead chemists/technicians, senior chemists/technicians, managers, QAO,
and the Division Administrator may question the validity of data and initiate an audit. Technical
Services is responsible for tracking the audit request, findings, and corrective action. The
auditor shall identify what prompted the question in a memo addressed to personnel involved in
the corrective action and the QAO. The title of the memo shall contain the sampling event
number.
Documentation of this audit must be retained per NELAC 2002 5.4.12.2.4 d, and 5.4.12.2.5.f.
Although data may be altered during specific steps of the review process all changes are
tracked through the LIMS audit trail.
7.4.2 Complaints
It is not in the scope of the LQM to address all viable complaints that come to the laboratory.
The LQM shall focus on issues that relate to data quality. The laboratory shall respond to
inquiries of data anomalies and complaints of report format or content, response time, and
laboratory policy using the Nonconformance Investigations and Corrective Action procedures
described in section 7.5.3. Complaints that are related to personnel conduct, and agency policy
are handled by laboratory management and the DEQ Human Resources Division.
Agency personnel tend to contact the sample tracker, whereas the receptionist usually receives
calls from the public. The receptionist and the sample tracker shall attempt to direct the caller to
the appropriate employee; however complaints are often difficult to decipher and will be referred
to the Technical Services manager or the laboratory Administrator as the default. All laboratory
staff members shall ensure the complaints they receive are discussed with the appropriate
section manager. If an employee feels uncomfortable with bringing a complaint to his/her
supervisor he/she may contact a QAO, the Technical Services section manager, or the
Laboratory Division Administrator.
Laboratory personnel should not attempt to resolve complaints without informing management.
Section managers shall assess whether the root causes of the complaint puts the integrity of
laboratory work into question. If the section manager determines there is a QA problem, the
manager shall ensure the Internal Audit procedures below are initiated immediately.
A complaint does not necessarily mean nonconforming work has occurred; however, complaints
must be investigated to determine whether or not an error has occurred. The laboratory must
also control records of complaints received. Thus, whenever a complaint is received, the
section manager shall file a Nonconformance report with the Technical Services section. The
Technical Services section manager shall assign a Nonconformance investigation to unbiased
personnel who will determine the validity of the complaint and assess whether the event that
generated the complaint violated laboratory policy or failed to follow procedures. The QAO
team shall participate in the investigation when necessary to guarantee an unbiased evaluation.
Technical Services shall track complaints received by recording pertinent information about the
complaint such as: description, date, staff assigned to investigate, action taken, and date
resolved.
When audit findings cast doubt on the laboratory’s operational effectiveness or on the
correctness or validity of the laboratory's work, the laboratory shall take timely corrective action.
The LQAO shall retain documentation that the project coordinator was notified in writing of
findings showing the laboratory results may have been affected.
The audit team shall use the Oregon Laboratory Accreditation Program (ORELAP) database to
record assessment findings, from which an Audit report will be printed and submitted to the
section managers. The section manager shall ensure that a corrective action plan for his/her
section will be written within 30 days of receiving the report. The LQAO and section manager
shall negotiate reasonable time frames for completion of corrective action procedures. The
LQAO shall record the scheduled completion date of the corrective action procedure in the
database. The LQAO will then monitor and routinely report on the status of the corrective
action. Subsequent internal audits shall verify and record the implementation and effectiveness
of the corrective action taken.
The LQAO shall review NELAC required administrative files. These files should contain
documentation of training requirements, training received, Demonstration of Capability, and
personnel conduct with respect to compliance to the laboratory data integrity policy. Discovery
of potential personnel issues shall be handled in a confidential manner until such time as a
follow up evaluation, full investigation including union representation if requested, or other
appropriate actions have been completed and the issues clarified. All investigations that result
in finding of inappropriate activity shall be documented and shall include corrective actions
taken and all appropriate notifications to clients. Any disciplinary actions taken shall be
documented and stored in the Human Resources Division’s personnel files which are not
accessible to the general public. All documentation of these investigations and actions taken
shall be maintained for at least five years.
Cause analysis should be conducted through data review and QA audits. The data review
process may reveal a Nonconformance which is of a technical nature and a suite of corrective
action procedures to follow (refer to Technical Corrective Action section 7.5.3.5).
Section managers may delegate the responsibility of reviewing data and assigning corrective
action tasks to whom ever they choose in their staff. Since it is the managers’ prerogative to
assign investigations and corrective action tasks as they see fit, each section shall write
procedures for Data Control, which will describe who is responsible for implementing corrective
action procedures for data quality problems.
must attach a comment to the result in LIMS and report the value as an estimate or void
the result. Typically chemists shall void results only due to analytical failures and not to
sample or batch QC limit failures. However, the failure to meet multiple QC limits may
lead to the action of voiding the result.
a) Sample QC measures may include but are not limited to:
i) Matrix Spike
ii) Matrix Spike Duplicate
iii) Duplicate analysis
b) Batch QC measures may include but are not limited to:
i) Method Blank
ii) Laboratory Control Sample
iii) Continuing Calibration Verification
iv) Calibration coefficients
As noted previously, if a chemist believes he/she has identified a questionable
procedure or feels his/her SOP does not meet the policy or procedures detailed in the
LQM, he/she is obligated to bring his/her concerns forward. He/she may contact a QAO,
a QAT member, his/her section supervisor, or the Laboratory Administrator.
2) Senior chemist/Lead Chemist/Lead Monitoring Specialists
a) The lead or senior chemist/technician shall validate data by reviewing sample
history, comparing intra-sample contaminants, and investigating data anomalies.
The lead chemists shall confer with chemists to evaluate any question he/she may
have about the data and may send the results back to the chemist for rework. As
above this process is controlled through the DAR approval process. The chemist
shall verify the transcription of data from the original source to LIMS, and if
necessary and possible, retest the sample.
3) The section manager shall review the decision process, which the chemist/technician
used to qualify the data and either approve the outcome or send the results back for
rework. The procedure for reworking results through LIMS is described in the DAR
SOP. This procedure will ensure NELAC 2002 5.4.12 Control of Records and 5.4.13
Internal Audits are followed.
4) The QAO shall review analytical data packets for completeness and sign analytical
reports certifying the controlled document is in compliance with the laboratory’s quality
policy. The QAO shall re-work analytical reports that do not comply with these policies.
If a quality control measure fails, all samples associated with the failed quality control
measure shall be reported with the appropriate data qualifier(s).
In addition to verifying precision and accuracy of analytical data, QC procedures
described in test method SOPs shall document the process for obtaining QC data. An
independent QAO shall review and sign all controlled documents, for which he/she is not
the author to help ensure that QC collection is documented. During review of the
controlled documents, problems may arise that warrant the creation of a
Nonconformance report. The QAO should solicit an investigation from personnel not
related to the issues.
comply with this goal. It is the QAT’s responsibility to assess and implement the laboratory’s
preventative measures (refer to section 2.1 Laboratory Quality Assurance Team).
Laboratory Quality Manual Oregon Department of Environmental quality
DEQ91-LAB-0006-QMP November 2004
Version 5.0 Page 38 of 64
8 Reporting Results
The laboratory shall report data in useful and comparable formats.
The Technical Services section of the laboratory has the responsibility of creating analytical
reports. Technical Services shall ensure that all projects entered into LIMS and assigned a
sampling event number shall be reported in a standard format, creating an official report record.
The report shall accurately, clearly, unambiguously, and objectively depict the results for
requested measurements. The intent of the report design is to minimize the possibility of
misunderstanding or misuse of data.
Technical Services shall ensure the integrity of the official report by applying the document
control procedures as described in section 7.3 (Data Reports and Verification Records). Even
though project coordinators may only use data stored in an electronic database, Technical
Services shall create an official paper report for document control.
Analytical reports shall be generated from LIMS and converted to a PDF, which shall be printed
on a color printer. Each page of the report shall have the title of the report, the sampling event
name and number prominently displayed at the top of the report. The report footer shall contain
the name of the PDF file, date and time the report was printed, and the page number with the
total number of pages in the PDF. The analytical report shall contain the following sections:
• Title page w/
1) Primary Report Recipient,
2) Report Date,
3) Laboratory Demographics, and
4) Approval Signatures.
• Narrative page w/
1) Sampling Event narrative when present,
2) List of Report Recipients,
3) List of Sample Collectors, and
4) List of Analytical Laboratories involved in the analyses.
• Sampling Event Summary page w/
1) Project ID linking the Sampling Event to the QAPP,
2) List of Sampling Sites w/
a) Item number,
b) Site ID,
c) Sample Description,
d) Matrix,
e) Sample Date,
f) Sample Time, and
g) Endnote references
• Analytical Results w/
1) Sample identification w/
a) Item Number,
b) Site ID,
c) Sample Description,
d) Sample Date, and
e) Sample Time
2) Analytical Parameter w/
a) Method reference,
b) Lower Reporting Limit,
c) Result,
Laboratory Quality Manual Oregon Department of Environmental quality
DEQ91-LAB-0006-QMP November 2004
Version 5.0 Reporting Results cont. Page 39 of 64
d) Unit,
e) Date and Time of sample preparation for test methods with holding time less than 72
hours, and
f) Endnote references
• Endnotes
• Electronic scanned sample collection form
The Technical Services section shall attach the following original forms and reports to the official
analytical report, although they may not be included in the electronic data packet.
• The original sample collection form
• QC reports
• Subcontracted analytical reports
Project coordinators may make special requests to scan documents which could then be
attached to the electronic document.
Technical Services shall transcribe pertinent data to LIMS from the sample collection form and
subcontracted analytical reports, which shall appear in the official report as well.
The analyst entering data into LIMS shall report results associated with failed QC as estimates
and enter a comment explaining his/her decision to qualify data as an estimate. Through the
Data Analysis Report (DAR) approval process management shall scrutinize the data and
identify which comments will be reported. LIMS will generate endnotes from the approved
comments and insert them in the appropriate section of the report.
The Technical Services section shall develop a system for tracking and reporting references to
sampling methods. As noted previously Watershed Assessment and Air Quality Monitoring
have documented procedures for collecting samples. References to these procedures shall be
included in analytical reports where appropriate.
The QAO shall complete the sampling event narrative portion of the analytical report. This
section shall be used to clarify QC measures and to offer opinions or interpretations of the data
by the QAO.
Project coordinators shall identify in the QAPP or SAP the report recipients and how they will
receive the analytical report. Typically agency personnel will receive an e-mail notice with a link
to the PDF. Reports may be emailed as a PDF, or photocopied and mailed. LIMS allows the
sample tracker to modify the list of report recipients at the time of sample entry. Staff may add
or delete personnel to and from the list of people receiving the analytical report. The project
coordinator shall be notified of changes to the recipient list. The laboratory shall only send
analytical reports to personnel listed as recipients. Laboratory personnel shall instruct the public
to request copies of an analytical report from the project coordinator, thus ensuring the project
coordinator is made aware of the potential use of their data.
Laboratory Quality Manual Oregon Department of Environmental quality
DEQ91-LAB-0006-QMP November 2004
Version 5.0 Page 40 of 64
APPENDICES
Laboratory Quality Manual Oregon Department of Environmental quality
DEQ91-LAB-0006-QMP November 2004
Version 5.0 Page 41 of 64
Date: Page of
Laboratory Name: Oregon Department of Environmental Quality
Laboratory Address: 1712 SW 11th Avenue, Portland OR 97201
Analyst(s) Name(s):
Matrix:
(Examples: laboratory pure water, soil, air, solid, biological tissue)
Method:
Method number, SOP#, Rev#, and Analyte, or Class of Analytes or Measured Parameters
(Examples: barium by 200.7, trace metals by 6010, benzene by 8021, etc.)
We, the undersigned, CERTIFY that:
1. The analysts identified above, using the cited test method(s), which is in use at this facility for the
analyses of samples under the National Environmental Laboratory Accreditation Program, have met the
Demonstration of Capability.
2. The test method(s) was performed by the analyst(s) identified on this certification.
3. A copy of the test method(s) and the laboratory-specific SOPs are available for all personnel on-site.
4. The data associated with the demonstration capability are true, accurate, complete and self-explanatory
(1).
5. All raw data (including a copy of this certification form) necessary to reconstruct and validate these
analyses have been retained at the facility, and that the associated information is well organized and
available for review by authorized assessors.
S = (S ) 2 1/2
where: Xi for i = 1 to n, are the analytical results in the final method reporting units obtained from the
n sample aliquots and Σ refers to the sum of the X values from i = 1 to n.
6)
a) Compute the MDL as follows:
Laboratory Quality Manual Oregon Department of Environmental quality
DEQ91-LAB-0006-QMP November 2004
Version 5.0 : 40 CFR Part 136, App. B cont. Page 44 of 64
MDL = t(n-1, 1-α = 0.99) (S)
where:
MDL = the method detection limit
t(n-1, 1-α = 0.99) = the students' t value appropriate for a 99% confidence level and a standard
deviation estimate with n-1 degrees of freedom. See 1.
S = standard deviation of the replicate analyses.
b) The 95% confidence interval estimates for the MDL derived in 6a are computed according to the
following equations derived from percentiles of the chi square over degrees of freedom distribution
(χ²/df).
LCL = 0.64 MDL
UCL = 2.20 MDL
where:
LCL and UCL are the lower and upper 95% confidence limits respectively based on seven
aliquots.
7) Optional iterative procedure to verify the reasonableness of the estimate of the MDL and subsequent
MDL determinations.
a) If this is the initial attempt to compute MDL based on the estimate of MDL formulated in Step 1, take
the MDL as calculated in Step 6, spike the matrix at this calculated MDL and proceed through the
procedure starting with Step 4.
b) If this is the second or later iteration of the MDL calculation, use S² from the current MDL calculation
and S² from the previous MDL calculation to compute the F-ratio. The F-ratio is calculated by
substituting the larger S² into the numerator S²A and the other into the denominator S²B. The
computed F-ratio is then compared with the F-ratio found in the table which is 3.05 as follows: if
S²A/S²B < 3.05, then compute the pooled standard deviation by the following equation:
[ILLUSTRATION GOES HERE]
ER31AU93.075
1/2
⎡ 6 S 2A + 6 S 2B ⎤
S pooled = ⎢ ⎥
⎣ 12 ⎦ 1
if S²A/S²B > 3.05, re-spike at the most recent calculated MDL and process the
samples through the procedure starting with Step 4. If the most recent calculated MDL does not permit
qualitative identification when samples are spiked at that level, report the MDL as a concentration between
the current and previous MDL which permits qualitative identification.
c) Use the Spooled as calculated in 7b to compute the final MDL according to the following equation:
MDL = 2.681 (Spooled)
where 2.681 is equal to t(12, 1-α = .99)
d) The 95% confidence limits for MDL derived in 7c are computed according to the following equations
derived from percentiles of the chi squared over degrees of freedom distribution.
LCL = 0.72 MDL
UCL = 1.65 MDL
where LCL and UCL are the lower and upper 95% confidence limits respectively
based on 14 aliquots.
Laboratory Quality Manual Oregon Department of Environmental quality
DEQ91-LAB-0006-QMP November 2004
Version 5.0 : 40 CFR Part 136, App. B cont. Page 45 of 64
TABLES OF STUDENTS' t VALUES AT THE 99 PERCENT CONFIDENCE LEVEL
Students' t: Table I.
Number of replicates Degrees of freedom t(n-1, 1-α = 0.99)
(n-1)
7 6 3.143
8 7 2.998
9 8 2.896
10 9 2.821
11 10 2.764
16 15 2.602
21 20 2.528
26 25 2.485
31 30 2.457
61 60 2.390
∞ ∞ 2.326
Reporting
The analytical method used must be specifically identified by number or title and the MDL for each analyte
expressed in the appropriate method reporting units. If the analytical method permits options which affect
the method detection limit, these conditions must be specified with the MDL value. The sample matrix
used to determine the MDL must also be identified with MDL value. Report the mean analyte level with the
MDL and indicate if the MDL procedure was iterated. If a laboratory standard or a sample that contained a
known amount analyte was used for this determination, also report the mean recovery.
If the level of analyte in the sample was below the determined MDL or exceeds 10 times the MDL of the
analyte in reagent water, do not report a value for the MDL.
Laboratory Quality Manual Oregon Department of Environmental quality
DEQ91-LAB-0006-QMP November 2004
Version 5.0 Page 46 of 64
Description:
Data impacted:
Sampling Event(s) Item(s) Container(s) Test(s)/Contaminant
Root cause:
CA taken:
To From Date
Case Number Case Name
The review process has revealed the possible problems noted below. Please check bottle numbers, calculations,
dilution factors, and data entry for errors. Rerun samples if necessary to resolve the problem(s). If needed, reanalyze
another aliquot from another sample container to check for contamination, container or site mix-ups, etc. Complete
this report and return it when finished. This rework is a priority task.
New Result Data Correction
Item Number Date Bottle Number Test Result
Site Name
Bottle Number
Alternate Bottle
N b
Standard Parameter Initial Result Verified? Yes No
Initial Test Result Reason Code for Change
Expected Test Result Comments:
Problem Code
pH SU SU Yes No
Comments:
pH SU SU Yes No
Comments:
pH SU SU Yes No
Comments:
pH SU SU Yes No
Comments:
Laboratory Quality Manual Oregon Department of Environmental quality
DEQ91-LAB-0006-QMP November 2004
Version 5.0 Nonconformance Report cont. Page 49 of 64
Corrective Action: Analyze a confirmation sample if the Laboratory Alkalinity, Conductivity, or Turbidity
disagrees with the Field results by more than 20% RPD or if the pH value disagrees with the Field results
by more than +/-0.5 units. When one or both samples are < 5 times the MRL compare the results using +/-
1 MRL for the average. If one or both of the Turbidity samples are < 10 NTU, then compare the results
using +/- 1 MRL for the average + 1 NTU. Another sample container may be required to check for the
possibility of site/sample confusion of container handling or cleaning problems. If the discrepancy is
verified, have the analysis assigned and run all the samples for the entire sampling event.
The Relative Percent Difference (RPD) is calculated using the following equation:
Xs − Xd
RPD = × 100%
[Xs + Xd ] / 2
Where: Xs = field result and
Xd = lab result
The units for Xs must equal those of Xd
Example 1: Given two duplicate results of 18 and 23:
23 − 18 5
RPD = × 100% = × 100% = 24%.
[18 + 23] / 2 20.5
Therefore, the Quality Control Check would fail. The Laboratory RPD results are over the expectable RPD
of 20%.
Example 2: Given two duplicate results of 5 and 7:
The Quality Control Check calculation = (5 + 7)/2 = 6. Therefore, the Acceptance window is 6 ± 1 MRL.
Therefore, the Quality Control Check would pass. The Laboratory results are within the calculated range of
5 to 7.
Example 3 (for Turbidity): Given two duplicate results of 5 and 9, the Quality Control Check calculation =
(5 + 9)/2 = 7. The acceptance window is 7 ± 2. Therefore, the Quality Control Check would pass. The
Laboratory results are within the calculated range 5 to 9.
Laboratory Quality Manual Oregon Department of Environmental quality
DEQ91-LAB-0006-QMP November 2004
Version 5.0 Page 50 of 64
• Microwave Digestor
• CEM MDS-2100
• Autoanalyzer:
• Lachat Flow Injection Ion Analyzer: Nitrate + Nitrite, Nitrite, Chloride, TK-N, and Sulfate
modules.
• Alpchem: Fluoride & Ammonia.
• Ion Chromatographs:
• Dionex Model 120, autosampler and Peaknet software
• Dionex 600 upgraded with CDM detector, gradient pump, SRS suppresser, AS40
autosampler, and Peaknet software.
• X-Ray Fluorescence Spectrometer:
• Fisons KEVEX EDX771
• Amperometric Titrator:
• Wallace & Tiernan Titrator
• Spectrophotometers:
• Bausch & Lomb: Spectronic 21
• Perkin Elmer Model Lambda 20 Spectrophotometer
• Hach DR/700
• Inductively Coupled Plasma (ICP):
• Simultaneous Perkin Elmer Optima 3000 DV
• Inductively Coupled Plasma Mass Spectrometry (ICP/MS):
• Thermo Elemental Model VG PQ Excel
• Analytical Balances:
• Sartorius A200S
• Mettler H15
• H6 w/filter weighing chamber
• Mettler H15, H18
• Sartorius 1712MP8
• Ohaus Balance, E4000D
• Microbalances:
• Cahn C30 - Range 1 µg to 3gm
• ATI Cahn C-44
• Optical Microscope:
• Zeiss Standard 18, Polarizing Trinocular with 35-mm camera
• Spencer Binocular stereo scope
• Olympus CH2 Phase Contrast microscope
Laboratory Quality Manual Oregon Department of Environmental quality
DEQ91-LAB-0006-QMP November 2004
Version 5.0 Page 53 of 64
Appendix F: QC definitions
Field Field Duplicate FD 10% Field Duplicate: discrete samples taken from the same field location and processed and analyzed
samples independently by the laboratory. The original sample is identified by space and time. The field duplicate is
collected collected at the same location and within a reasonable lapse of time. (DEQ)
during a
sampling
expedition
Field Laboratory LRB per SOP Changed from Lab Stored Blank.
Retained Blank
Laboratory Retained Blank: a sample of analyte-free matrix that remains in the laboratory and is used as
comparison with the blanks carried to the field. (DEQ)
Field Manual Precision MP 10% sample Changed from Co-located Sampler.
sites
Used in air sampling. A secondary sample collected from a location. Similar to a Field Duplicate. Multiple
sampling devices run simultaneously within close proximity. (DEQ)
Laboratory Quality Manual Oregon Department of Environmental quality
DEQ91-LAB-0006-QMP November 2004
Version 5.0 Appendix F: QC definitions cont. Page 55 of 64
Field Transport Blank TNPB per QAPP Transport Blank: a sample of analyte-free media which has been carried to the field and returned to the
laboratory. (DEQ)
Operations Automated AA Used in air sampling. Generally a gas from a secondary source analyzed on-site by a secondary auditor.
Accuracy Very similar to a 2nd source QC in many respects. (DEQ)
Operations Blind Sample BLND Blind Sample: a sub-sample for analysis with a composition known to the submitter. The analyst/laboratory
may know the identity of the sample but not its composition. It is used to test the analyst’s or laboratory’s
proficiency in the execution of the measurement process. (NELAC)
Operations Certified CRM Certified Reference Material (CRM): a reference material one or more of whose property values are certified
Reference by a technically valid procedure, accompanied by or traceable to a certificate or other documentation which
Material is issued by a certifying body. (ISO Guide 30 - 2.2)
Operations Inter-Lab Split SPLT Changed from Split: Samples split with an external laboratory.
Sample
Laboratory audit or Split sample: verification of field and/or laboratory performance through the collection
and analysis of field duplicate samples by an alternate laboratory. (DEQ)
Operations Manual Accuracy MA Used in air sampling. A secondary auditor collects audit samples on equipment using equipment with known
properties, essentially the collection of an audit sample. (DEQ)
Operations Proficiency Test PT Proficiency Test Sample (PT): a sample, the composition of which is unknown to the analyst and is provided
Sample to test whether the analyst/laboratory can produce analytical results within specified acceptance criteria.
(QAMS)
Operations Reference RM Reference Material: a material or substance one or more properties of which are sufficiently well established
Material to be used for the calibration of an apparatus, the assessment of a measurement method, or for assigning
values to materials. (ISO Guide 30-2.1)
Laboratory Quality Manual Oregon Department of Environmental quality
DEQ91-LAB-0006-QMP November 2004
Version 5.0 Appendix F: QC definitions cont. Page 56 of 64
Preparation Dilution DR as required Dilution: additional measurement made from a diluted sample aliquot. Used with the undiluted sample or
other dilutions to establish analytical precision, evaluate matrix interferences, and/or bring the analyte
concentration to within the instrument's calibration range. (DEQ)
Preparation Laboratory LCON per SOP Confirmation: verification of the identity of a component through the use of an approach with a different
Confirmation scientific principle from the original method. These may include, but are not limited to:
Second column confirmation
Alternate wavelength
Derivatization
Mass spectral interpretation
Alternative detectors or
Additional cleanup procedures. (NELAC)"
Preparation Laboratory LCS 1/ analytical Laboratory Control Sample (also known as laboratory fortified blank, spiked blank, or QC check sample): a
Control Sample batch sample matrix, free from the analytes of interest, spiked with verified known amounts of analytes or a
material containing known and verified amounts of analytes. It is generally used to establish intra-laboratory
or analyst specific precision and bias or to assess the performance of all or a portion of the measurement
system. (NELAC)
The LCS requirements for analytical QC may be satisfied through the use of a Quality Control Sample
(QCS).
Preparation Laboratory LCSD Laboratory Control Sample Duplicate: a sample matrix, free from the analytes of interest, spiked with verified
Control Sample known amounts of analytes or a material containing known and verified amounts of analytes. It is used with
Duplicate the LCS to establish intra-laboratory or analyst specific precision and bias when more traditional methods are
unavailable. (DEQ)
Preparation Laboratory LD 10% / Changed from Analytical Replicate.
Duplicate preparation
Laboratory Duplicate: aliquots of a sample taken from the same container under laboratory conditions and
batch
processed and analyzed independently. (NELAC)
Preparation Matrix Spike MS Matrix Spike (spiked sample or fortified sample): a sample prepared by adding a known mass of target
analyte to a specified amount of matrix sample for which an independent estimate of Target analyte
concentration is available. Matrix spikes are used, for example, to determine the effect of the matrix on a
method's recovery efficiency. (QAMS)
Laboratory Quality Manual Oregon Department of Environmental quality
DEQ91-LAB-0006-QMP November 2004
Version 5.0 Appendix F: QC definitions cont. Page 57 of 64
NON METALS
BOD5 5210 B Winkler-Azide Modification
CBOD5 5210 B Winkler-Azide Mod./Inhib
Dissolved Oxygen 4500-O C 360.2 Winkler-Azide Modification
Dissolved Oxygen 4500-O G 360.1 Membrane Electrode
Bromide 4110 B 300 Ion Chromatography
-
Chloride 4500-Cl C 325.1 Auto Ferricyanide
Chloride 4110 B 300.0 Ion Chromatography
Chlorine 4500-Cl D 330.1 Amperometric Titration
Chlorine 4500-Cl G 330.5 DPD Colorimetric
Fluoride 4500-F E 340.3 Auto Complexone
Ammonia-Nitrogen 4500-NH3 H 350.1 Auto Phenate
Total Kjeldahl Nitrogen 351.2 S-Auto Block Digestion
Nitrate-Nitrogen 4500-NO3 F 353.2 Auto Cadmium Reduction
Nitrite-Nitrogen 4500-NO3 F 353.2 Auto colorimetric
Nitrite-Nitrogen 4110 B 300.0 Ion Chromatography
Ortho Phosphate-P 4500-P E 365.2 Colorimetric Ascorbic Acid
Ortho Phosphate-P 4110 B 300.0 Ion Chromatography
Total Phosphate-P 4500-P E,5 365.2 Colorimetric Ascorbic Acid
Sulfate 4500-SO42 F 375.2 375.2 Auto Methylthymol Blue
Laboratory Quality Manual Oregon Department of Environmental quality
DEQ91-LAB-0006-QMP November 2004
Version 5.0 Appendix G: Analytical Methods cont. Page 61 of 64
PHYSICAL PARAMETERS
Acidity 2310 B 305.1 Titration
Alkalinity 2320 B 310.1 Titration
Color 2120 C 110.3 Colorimetric Pt/Co
Conductivity 2510 B 120.1 Wheatstone Bridge
Flash Point 1010 Pensky-Martens closed-cup
+
pH 4500-H 150.1 150.1 Electrometric
Filterable (TDS) 2540 C 160.1 Gravimetric, 180oC
Nonfilterable (TSS) 2540 D 160.2 Gravimetric, 103-105oC
Total (TS) 2540 B 160.3 Gravimetric, 103-105oC
Volatile 160.4 Gravimetric, Ignition @ 550oC
Temperature 2550 B 170.1
Turbidity 2130 B 180.1 Nephelometric
ORGANICS
Adipates 625 525.2 8270 GC/MS
BTEX 602 5020A GC/PID
Chlorinated Pest/PCB 608 508 8081 Solvent Extr., GC/ECD
PCB as DCBP 508A Solvent Extr., GC/ECD
Chlorinated Herbicides 515.1 8150A Solvent Extr., GC/MS
EDB, DBCP 504.1 GC/ECD
Laboratory Quality Manual Oregon Department of Environmental quality
DEQ91-LAB-0006-QMP November 2004
Version 5.0 Appendix G: Analytical Methods cont. Page 62 of 64
Gas
Diesel
Oil
Oil and Grease 1664ix Gravimetric (Hexane)
Particulate Fall Out (PFO) Gravimetric
x
Percent Fat Hexane Extraction/Gravimetric
Laboratory Quality Manual Oregon Department of Environmental quality
DEQ91-LAB-0006-QMP November 2004
Version 5.0 Appendix G: Analytical Methods cont. Page 63 of 64
i
Approved Waste Water Methods promulgated in 40 CFR Part 136 appear in “Bold”.
ii
Approved Drinking Water Methods promulgated in 40 CFR Part 141 appear in “Italic”.
iii
Standard Methods for the Examination of Water and Wastewater, 18th Edition, APHA, AWWA, WPCF, 1994.
iv
Methods for Chemical Analysis of Water and Wastes, EPA-600/4-79-020, Revised 3/83.
v
300.0: Methods for the Determination of Inorganic Substances in Environmental Samples, EPA-600/R-93-100, August 1993.
vi
200.7 & 200.9: Methods for the Determination of Metals in Environmental Samples - Supplement I, EPA-600/R-94-111, May 1994.
vii
Five hundred series: Methods for the Determination of Organic Compounds in Drinking Water, EPA-600/4-88-039, December 1988. Methods for the Determination
of Organic Compounds in Drinking Water - Supplement I, EPA-600/4-90-020, July 1990. Methods for the Determination of Organic Compounds in Drinking
Water - Supplement II, EPA-600/R-92-129, August 1992.
viii
Test Methods for Evaluating Solid Waste: SW-846, Third Edition, USEPA, November 1986.
ix
Chuck Clarke, Regional Administrator for EPA, Region 10 has sent a letter of approval for an alternate test procedure. EPA will recognize results from Method 1664, Hexane
Extractables for all Oil and Grease analyses performed on samples that originate in Region 10.
x
DEQ Method Revised 2/3/98.
xi
DEQ Method developed 3/12/1987
xii
DEQ Method Developed 8/29/1989
Laboratory Quality Manual Oregon Department of Environmental Quality
DEQ91-LAB-0006-QMP November 2004
Version 5.0 Appendix H: Organizational Chart Page 64 of 64
Division
Administrator:
Mary Abrams
Special Projects: Executive Support Specialist: Air Quality QAO: Land Quality QAO: Water Quality QAO:
Rick Hafele Sylvia Herrley Paul McKay Ron Doughten Chris Redman
Air Quality Monitoring Watershed Assessment Technical Services Manager: Inorganic Laboratory Manager: Organic & General Chemistry
Manager: Manager: Dan Hickman RaeAnn Haynes Laboratory Manager:
Jeff Smith Greg Pettit Eugene Foster