0% found this document useful (0 votes)
218 views

Msa 1 10

The document reports the results of a measurement system analysis of a part using three appraisers (A, B, C) and two measurement tools (Apprisar A and B). Each appraiser took 10 measurements of the part across 10 trials. The average and range for each set of measurements was calculated. The repeatability, reproducibility, part variation, and total variation were determined using statistical analysis methods. The measurement system was found to have a %R&R of 24.31%, indicating potential need for improvement since it is greater than 10% but less than 30%.

Uploaded by

Mahendra
Copyright
© © All Rights Reserved
Available Formats
Download as XLSX, PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
218 views

Msa 1 10

The document reports the results of a measurement system analysis of a part using three appraisers (A, B, C) and two measurement tools (Apprisar A and B). Each appraiser took 10 measurements of the part across 10 trials. The average and range for each set of measurements was calculated. The repeatability, reproducibility, part variation, and total variation were determined using statistical analysis methods. The measurement system was found to have a %R&R of 24.31%, indicating potential need for improvement since it is greater than 10% but less than 30%.

Uploaded by

Mahendra
Copyright
© © All Rights Reserved
Available Formats
Download as XLSX, PDF, TXT or read online on Scribd
You are on page 1/ 31

PART NAME W/S PARTITION RR DR R/L DATE 26.5.

10
PART NO. 84665/75 M75J00 CHARACTERISTIC Gap
MODEL YN-4 EQUIPMENT USED Feeler Gauge (FG-07)
Apprisar A
Trials 1 2 3 4 5 6 7 8 9 10
1 0.85 0.75 0.68 0.48 0.85 0.72 0.82 0.89 0.52 0.90
2 0.79 0.71 0.65 0.45 0.79 0.68 0.86 0.90 0.55 0.95
3 0.78 0.79 0.69 0.46 0.80 0.64 0.80 0.95 0.60 0.90
Avg. (Xi) 0.81 0.75 0.67 0.46 0.81 0.68 0.83 0.91 0.56 0.92

Range (Ri) 0.07 0.08 0.04 0.03 0.06 0.08 0.06 0.06 0.08 0.05

Apprisar B
Trials 1 2 3 4 5 6 7 8 9 10
1 0.82 0.74 0.65 0.46 0.85 0.67 0.90 0.95 0.55 0.95
2 0.76 0.79 0.70 0.49 0.89 0.72 0.85 1.00 0.60 0.90
3 0.79 0.76 0.64 0.52 0.85 0.68 0.85 0.94 0.55 0.95
Avg. (Xi) 0.79 0.76 0.66 0.49 0.86 0.69 0.87 0.96 0.57 0.93
Range (Ri) 0.06 0.05 0.06 0.06 0.04 0.05 0.05 0.06 0.05 0.05

Apprisar C
Trials 1 2 3 4 5 6 7 8 9 10
1 0.79 0.74 0.65 0.48 0.85 0.70 0.95 0.95 0.56 0.95
2 0.82 0.69 0.68 0.49 0.83 0.68 0.85 0.85 0.59 0.95
3 0.75 0.65 0.69 0.54 0.81 0.65 0.85 0.94 0.62 0.95
Avg. (Xi) 0.79 0.69 0.67 0.50 0.83 0.68 0.88 0.91 0.59 0.95

Range (Ri) 0.07 0.09 0.04 0.06 0.04 0.05 0.10 0.10 0.06 0.00

Part Avg.
(XP) 0.79 0.74 0.67 0.49 0.84 0.68 0.86 0.93 0.57 0.93 RP =
Xp = Average of individual parts
X
Averge of all Xpi parts 0.75 RP =

Average (X bar) Chart for Appraiser A


1.00

0.90

0.80

0.70

0.60

0.50

0.40

0.30

0.20

0.10

0.00
1 2 3 4 5 6 7 8 9 10

Range (R) Chart for Appraiser A


0.16

0.14
Range (R) Chart for Appraiser A
0.16

0.14

0.12

0.10

0.08

0.06

0.04

0.02

0.00
1 2 3 4 5 6 7 8 9 10

PREPARED BY:
MEASUREMENT SYSTEM ANALYSIS
VARIABLE STUDY (Average-Range Method)
NOMINAL DIMENSION 0 APPRAISER A
UPPER TOLERANCE 1 APPRAISER B
uge (FG-07) LOWER TOLERANCE 0 APPRAISER C

Constant Tables (for n=10, r=3)


Actual
No. Of A2 D3 D4 K1 K2 Parts K3
K1 : Based on no. of Trials Trials (r)
(n)
K2 : Based on no. of Appriser No. of
K3 : Based on Sample Size used Parts (n)
in trials
2 1.880 0 3.267 0.8862 0.7071

0.74 0.5908 0.5231


3 1.020 0 2.575

0.061 10 3 1.020 0 2.575 0.591 0.523 10 0.315

Value of K3 with ref. To n (Sample Size)


n 2 3 4 5 6 7 8 9 10
K3 0.7071 0.5231 0.4467 0.403 0.3742 0.3534 0.3375 0.3249 0.3146

MEASUREMENT UNIT ANALYSIS


0.76

0.053 1 Repeatability - Equipment Variation (EV)


EV = R x K1 => 0.0345
2 Reproducibility - Appraiser Variation (AV)
AV = {|(xDIFF x K2) - (EV /nr)|}
2 2 1/2
=> 0.0077
n = number of parts
r = number of trials
3 Repeatability & Reproducibility (R & R)
0.75 R & R (GRR)= {(EV2 + AV2)}1/2 => 0.0353

0.061 4 Part Variation (PV)


PV = Rp X K3 => 0.1409
5 Total Variation (TV)
0.45
TV = {(R&R2 + PV2)}1/2 => 0.1452
Range of Part Avg

Average (X bar) Chart for Appraiser B


1.20

1.00

0.80

0.60

0.40

0.20

0.00
10 1 2 3 4 5 6 7 8 9 10

Range (R) Chart for Appraiser B


0.16

0.14
Range (R) Chart for Appraiser B
0.16

0.14

0.12

0.10

0.08

0.06

0.04

0.02

0.00
10 1 2 3 4 5 6 7 8 9 10

CHECKEED BY:
ER A Umesh Kumar APPRAISEE M. NEGI
R B Laxmikant %R & R 24.31 %
R C Naresh Prtap ndc 6.00

0.0583 Sr.No. ACCEPTABILITY CRITERIA


0.0190
1 If % R & R < 10%, system is acceptable
0.1502
0.0000
If 10% < % R & R < 30%, system may be acceptable
0.7497 2 based on importance of application, cost of gage, cost of
repair, etc.,
0.8092

0.6902
3 If % R & R > 30%, system needs improvement
TOLERANCE 1

TVTOLERANCE 0.17
4 NDC >= 5

COMPARISION TO COMPARISION TO
TOTAL VARIATION TOLERANCE

% EV = 100 (EV/TV) => 23.73 % % EV = 100 (EV/TV tol) => 20.68 %

% AV = 100 (AV/TV) => 5.3 % % AV = 100 (AV/TVtol) => 4.62 %


% R&R = 100 (R&R/TV) => 24.31 % % R&R = 100 (R&R/Tol) => 21.19 %

% PV = 100 (PV/TV) => 97 % % PV = 100 (PV/Tol) => 84.52 %


No.of distinct categories No.of distinct categories
NDC= 1.41(PV/GRR) => 6 NDC= 1.41(PV/GRR) => 6

Average (X bar) Chart for Appraiser C


1.00

0.90

0.80

0.70

0.60

0.50

0.40

0.30

0.20

0.10

0.00
10
1 2 3 4 5 6 7 8 9 10

Range (R) Chart for Appraiser C


0.16

0.14
Range (R) Chart for Appraiser C
0.16

0.14

0.12

0.10

0.08

0.06

0.04

0.02

0.00
10 1 2 3 4 5 6 7 8 9 10

APPROVED BY:
PART NAME W/S PARTITION RR DR R/L DATE 26.5.10
PART NO. 84665/75 M75J00 CHARACTERISTIC Gap
MODEL YN-4 EQUIPMENT USED Feeler Gauge (FG-07)
Apprisar A
Trials 1 2 3 4 5 6 7 8 9 10
1 0.85 0.75 0.68 0.48 0.85 0.72 0.82 0.89 0.52 0.90
2 0.79 0.71 0.65 0.45 0.79 0.68 0.86 0.90 0.55 0.95
3 0.78 0.79 0.69 0.46 0.80 0.64 0.80 0.95 0.60 0.90
Avg. (Xi) 0.81 0.75 0.67 0.46 0.81 0.68 0.83 0.91 0.56 0.92

Range (Ri) 0.07 0.08 0.04 0.03 0.06 0.08 0.06 0.06 0.08 0.05

Apprisar B
Trials 1 2 3 4 5 6 7 8 9 10
1 0.82 0.74 0.65 0.46 0.85 0.67 0.90 0.95 0.55 0.95
2 0.76 0.79 0.70 0.49 0.89 0.72 0.85 1.00 0.60 0.90
3 0.79 0.76 0.64 0.52 0.85 0.68 0.85 0.94 0.55 0.95
Avg. (Xi) 0.79 0.76 0.66 0.49 0.86 0.69 0.87 0.96 0.57 0.93
Range (Ri) 0.06 0.05 0.06 0.06 0.04 0.05 0.05 0.06 0.05 0.05

Apprisar C
Trials 1 2 3 4 5 6 7 8 9 10
1 0.79 0.74 0.65 0.48 0.85 0.70 0.95 0.95 0.56 0.95
2 0.82 0.69 0.68 0.49 0.83 0.68 0.85 0.85 0.59 0.95
3 0.75 0.65 0.69 0.54 0.81 0.65 0.85 0.94 0.62 0.95
Avg. (Xi) 0.79 0.69 0.67 0.50 0.83 0.68 0.88 0.91 0.59 0.95

Range (Ri) 0.07 0.09 0.04 0.06 0.04 0.05 0.10 0.10 0.06 0.00

Part Avg.
(XP) 0.79 0.74 0.67 0.49 0.84 0.68 0.86 0.93 0.57 0.93 RP =
RP =

Average (X bar) Chart for Appraiser A


1.00

0.90

0.80

0.70

0.60

0.50

0.40

0.30

0.20

0.10

0.00
1 2 3 4 5 6 7 8 9 10

Range (R) Chart for Appraiser A


0.16

0.14
Range (R) Chart for Appraiser A
0.16

0.14

0.12

0.10

0.08

0.06

0.04

0.02

0.00
1 2 3 4 5 6 7 8 9 10

PREPARED BY:
MEASUREMENT SYSTEM ANALYSIS
VARIABLE STUDY (Average-Range Method)
NOMINAL DIMENSION 0 APPRAIS
UPPER TOLERANCE 1 APPRAIS
r Gauge (FG-07) LOWER TOLERANCE 0 APPRAIS

Constant Tables (for n=10, r=3)


Actual
No. Of A2 D3 D4 K1 K2 Parts
K1 : Based on no. of Trials Trials (r)
(n)
K2 : Based on no. of Appriser No. of
K3 : Based on Sample Size Parts (n)
used in trials
2 1.880 0 3.267 0.8862 0.7071

0.74 0.5908 0.5231


3 1.020 0 2.575

0.061 10 3 1.020 0 2.575 0.591 0.523 10

Value of K3 with ref. To n (Sample Size)


n 2 3 4 5 6 7 8 9
K3 0.7071 0.5231 0.4467 0.403 0.3742 0.3534 0.3375 0.3249

MEASUREMENT UNIT ANALYSIS


0.76

0.053 1 Repeatability - Equipment Variation (EV)


EV = R x K1 => 0.034
2 Reproducibility - Appraiser Variation (AV)
AV = {|(xDIFF x K2) - (EV /nr)|}
2 2 1/2
=> 0.007
n = number of parts
r = number of trials
3 Repeatability & Reproducibility (R & R)
0.75 R & R (GRR)= {(EV2 + AV2)}1/2 => 0.035

0.061 4 Part Variation (PV)


PV = Rp X K3 => 0.140
5 Total Variation (TV)
0.45
TV = {(R&R2 + PV2)}1/2 => 0.145
Range of Part Avg

Average (X bar) Chart for Appraiser B


1.20

1.00

0.80

0.60

0.40

0.20

0.00
9 10 1 2 3 4 5 6 7 8 9

Range (R) Chart for Appraiser B


0.16

0.14
Range (R) Chart for Appraiser B
0.16

0.14

0.12

0.10

0.08

0.06

0.04

0.02

0.00
9 10 1 2 3 4 5 6 7 8 9

CHECKEED BY:
LYSIS PLANT II
hod)
APPRAISER A Umesh Kumar APPRAISEE M. NEGI
APPRAISER B Laxmikant %R & R 24.31 %
APPRAISER C Naresh Prtap ndc 6.00

0.0583 Sr.No. ACCEPTABILITY CRITER


0.0190
K3 1 If % R & R < 10%, system is accep
0.1502
0.0000
If 10% < % R & R < 30%, system m
0.7497 2 based on importance of application, cost
repair, etc.,
0.315 0.8092

0.6902
3 If % R & R > 30%, system needs im
TOLERANCE 1

10
4 NDC >= 5
0.3146

COMPARISION TO
TOTAL VARIATION CONCLUSION

0.0345 % EV = 100 (EV/TV) => 23.73 %

0.0077 % AV = 100 (AV/TV) => 5.3 %


0.0353 % R&R = 100 (R&R/TV) => 24.31 %

0.1409 % PV = 100 (PV/TV) => 97 %


No.of distinct categories
0.1452 NDC= 1.41(PV/GRR) => 6

Average (X bar) Chart for Appraiser C


1.00

0.90

0.80

0.70

0.60

0.50

0.40

0.30

0.20

0.10

0.00
9 10
1 2 3 4 5 6 7 8

Range (R) Chart for Appraiser C


0.16

0.14
Range (R) Chart for Appraiser C
0.16

0.14

0.12

0.10

0.08

0.06

0.04

0.02

0.00
9 10 1 2 3 4 5 6 7 8

APPROVED BY:
PLANT III

M. NEGI
24.31 %
6.00

ACCEPTABILITY CRITERIA

If % R & R < 10%, system is acceptable

If 10% < % R & R < 30%, system may be acceptable


based on importance of application, cost of gage, cost of
repair, etc.,

If % R & R > 30%, system needs improvement

NDC >= 5

CONCLUSION
(X bar) Chart for Appraiser C

4 5 6 7 8 9 10

ge (R) Chart for Appraiser C


ge (R) Chart for Appraiser C

4 5 6 7 8 9 10
M
PART NAME W/S PARTITION RR DR R/L DATE 26.5.10
PART NO. 84665/75 M75J00 CHARACTERISTIC Gap
MODEL YN-4 EQUIPMENT USED Feeler Gauge (FG-07)
Apprisar A
Trials 1 2 3 4 5 6 7 8 9 10
1 0.85 0.75 0.68 0.48 0.85 0.72 0.82 0.89 0.52 0.90
2 0.79 0.71 0.65 0.45 0.79 0.68 0.86 0.90 0.55 0.95
3 0.78 0.79 0.69 0.46 0.80 0.64 0.80 0.95 0.60 0.90
Avg. (Xi)

Range (Ri)

Apprisar B
Trials 1 2 3 4 5 6 7 8 9 10
1 0.82 0.74 0.65 0.46 0.85 0.67 0.90 0.95 0.55 0.95
2 0.76 0.79 0.70 0.49 0.89 0.72 0.85 1.00 0.60 0.90
3 0.79 0.76 0.64 0.52 0.85 0.68 0.85 0.94 0.55 0.95
Avg. (Xi)

Range (Ri)

Apprisar C
Trials 1 2 3 4 5 6 7 8 9 10
1 0.79 0.74 0.65 0.48 0.85 0.70 0.95 0.95 0.56 0.95
2 0.82 0.69 0.68 0.49 0.83 0.68 0.85 0.85 0.59 0.95
3 0.75 0.65 0.69 0.54 0.81 0.65 0.85 0.94 0.62 0.95
Avg. (Xi)

Range (Ri)

Part Avg.
(XP) RP =
RP = Range of Part Avg

Average (X bar) Chart for Appraiser A


1.00 1.00

0.90 0.90

0.80 0.80

0.70 0.70

0.60 0.60

0.50 0.50

0.40 0.40

0.30 0.30

0.20 0.20

0.10 0.10

0.00 0.00
1 2 3 4 5 6 7 8 9 10 11 1

Range (R) Chart for Appraiser A


1.00 1.00

0.90 0.90

0.80 0.80
Range (R) Chart for Appraiser A
1.00 1.00

0.90 0.90

0.80 0.80

0.70 0.70

0.60 0.60

0.50 0.50

0.40 0.40

0.30 0.30

0.20 0.20

0.10 0.10

0.00 0.00
1 2 3 4 5 6 7 8 9 10 11 1

PREPARED BY: CHECKE


MEASUREMENT SYSTEM ANALYSIS
VARIABLE STUDY (Average-Range Method)
NOMINAL DIMENSION 0 APPRAISER A
UPPER TOLERANCE 1 APPRAISER B
LOWER TOLERANCE 0 APPRAISER C

Constant Tables (for n=10, r=3)


Actual
No. Of A2 D3 D4 K1 K2 Parts K3
K1 : Based on no. of Trials Trials (r)
(n)
K2 : Based on no. of Appriser No. of
K3 : Based on Sample Size used Parts (n)
in trials
2 1.880 0 3.267 0.8862 0.7071
0.5908 0.5231
3 1.020 0 2.575
10 3 1.020 0 2.575 0.591 0.523 10 0.315

Value of K3 with ref. To n (Sample Size) TOLERANC

n 2 3 4 5 6 7 8 9 10
0.00 K3 0.7071 0.5231 0.4467 0.403 0.3742 0.3534 0.3375 0.3249 0.3146

0.00

MEASUREMENT UNIT ANALYSIS

1 Repeatability - Equipment Variation (EV)


EV = R x K1 =>
2 Reproducibility - Appraiser Variation (AV)
AV = {|(xDIFF x K2) - (EV /nr)|}
2 2 1/2
=>
n = number of parts
r = number of trials
3 Repeatability & Reproducibility (R & R)
R & R (GRR)= {(EV2 + AV2)}1/2 =>

4 Part Variation (PV)


PV = Rp X K3 =>
5 Total Variation (TV)
TV = {(R&R2 + PV2)}1/2 =>

Average (X bar) Chart for Appraiser B


1.00 1.00

0.90 0.90

0.80 0.80

0.70 0.70

0.60 0.60

0.50 0.50

0.40 0.40

0.30 0.30
0.20 0.20
0.10
0.10
0.00
0.00
1 2 3 4 5 6 7 8 9 10 11
1

Range (R) Chart for Appraiser B


1.00 1.00
0.90 0.90
0.80 0.80
Range (R) Chart for Appraiser B
1.00 1.00
0.90 0.90
0.80 0.80
0.70 0.70
0.60 0.60
0.50 0.50
0.40 0.40
0.30 0.30
0.20 0.20
0.10 0.10
0.00 0.00
1 2 3 4 5 6 7 8 9 10 11 1

CKEED BY: APPROV


PLANT III

Umesh Kumar APPRAISEE M. NEGI


Laxmikant %R & R 0.00
Naresh Prtap ndc 0.00

Sr.No. ACCEPTABILITY CRITERIA

1 If % R & R < 10%, system is acceptable

If 10% < % R & R < 30%, system may be acceptable


2 based on importance of application, cost of gage, cost of
repair, etc.,

3 If % R & R > 30%, system needs improvement


TOLERANCE 1

4 NDC >= 5

COMPARISION TO
TOTAL VARIATION CONCLUSION

% EV = 100 (EV/TV) =>

% AV = 100 (AV/TV) =>


% R&R = 100 (R&R/TV) =>

% PV = 100 (PV/TV) =>


No.of distinct categories
NDC= 1.41(PV/GRR) =>

Average (X bar) Chart for Appraiser C


1.00

0.90

0.80

0.70

0.60

0.50

0.40

0.30

0.20

0.10

0.00
1 2 3 4 5 6 7 8 9 10 11

Range (R) Chart for Appraiser C


1.00
0.90
0.80
Range (R) Chart for Appraiser C
1.00
0.90
0.80
0.70
0.60
0.50
0.40
0.30
0.20
0.10
0.00
1 2 3 4 5 6 7 8 9 10 11

APPROVED BY:
Sample 1 Sample 2 Sample 3 Sample 4 Sample 5 Sample 6 Sample 7 Sample 8 Sample 9 Sample 10
Trial 1 Row 1
Apprisar
Trial 2 Row 2
A
Trial 3 Row 3
Average (XAi ) Row 4
Range (RAi) Row 5
Trial 1 Row 6
Apprisar
Trial 2 Row 7
B
Trial 3 Row 8
Average (XBi ) Row 9
Range (RBi) Row 10
Trial 1 Row 11
Apprisar
Trial 2 Row 12
C
Trial 3 Row 13
Average (XCi ) Row 14
Range (RCi) Row 15
Part Avg individual
Xpi Row 16 sample

R Row 17

X Sxpi / n Row 18

Rpi Max (Xpi) - Min (Xpi) Row 19

Xdiff Row 20

UCLR D4 x R Row 21

LCLR D3 x R Row 22

UCLX X + A2R Row 23

LCLX X - A2R Row 24

You might also like