Through this interface, Auto-Chemistry Analyzer can apply sample information registered in LIS
workstation manually or automatically, and transmit test result and QC result to LIS workstation
manually or automatically. The data transmitted through this interface all follow ASTM standard
protocol format.
Bidirectional:“Bidirectional” must be selected when applying LIS data. This setting selection is not
Data Transmission Mode:Including real time mode and batch mode. Those two modes are available
Baud Rate:Variation times of carrier parameter per unit time. 4800Bps, 9600Bps or 19200Bps can be
Parity:N, M, E, O, S. Default as N.
Timeout Retry Time:Time interval used for overtime retry in communication. The communication will
stop when there is no effective communication during overtime retry time. The
input range of overtime retry time is 1-15.
Timeout Retry Count:Count used for overtime retry in communication. The communication will stop
when there is no effective communication during overtime retry count. The input range
2
of overtime retry count is 1-6.
[STX] FN Text first char. ...... Text last char. ETX CH CL [CR] [LF]
[STX]
The ASCII code 2, indicating the beginning of a frame transmission.
FN
The frame number modulo 8. Frames of a single Transmission Phase are consecutively
Text
Data content.
[ETX]
ASCII code 3, indicating the end of the text block of an end frame.
CH,CL
3
Represent the high nibble and the low nibble of the 8 bit checksum, respectively. CH and CL
are represented as two digits of hex numbers. The checksum is the modulo 256 of the sum of ASCII
values of the frame characters starting with and including "FN" and completing with [ETX].
[STX]1Test[ETX]
[STX]1Test[ETX]D4[CR][LF]
3. Delimiters
The Presentation Layer uses some characters as special delimiter codes. These characters may not
4
Message Header Record H
Patient Information Record P
Test Order Record O
Result Record R
Message Terminator Record L
Request Information Record Q
Mandatory:
Y: Field is mandatory and have to be considered.
N: Field is not mandatory. The field have to be considered by the LIS, but it is transmitted only if
information is available at Liaison. When the information is not available an empty field is transmitted.
The Message Header Record to LIS is the first record of each message.
Example:
Description:
Example:
5
Field No. E1394 ASTM Field Name Field Contents Max Length Mandatory
1 8.1.1 Record Type ID ‘P’ 1 Y
2 8.1.2 Sequence Number 6 Y
6 8.1.6 Patient Name 30 N
9 8.1.9 Patient Sex ‘M’, ‘F’ or ‘U’ 1 N
15^1 Age(1-200) 3
8.1.15 Special Field 1 N
15^2 Age Unit (‘Y’,’M’,’D’) 1
Description:
This record holds information regarding analysis request and sample information to LIS.
Example::
<STX><FN>O|1|^1^N001^10^N^1|||R|20081124151012|||||||||1||||||||||O
<CR><ETX><CheckSum><CR><LF>
<STX><FN>O|1|CA2201320078^^N001^10^N^1|||R|20081124151012|||||||||1||||||||||O<CR><ETX
><CheckSum><CR><LF>
When transmitting Control result:
<STX><FN>O|1|^0^C001^1^N^1|LANDOX-1^2333^|||20110104172413|||||||||1||||||||||O<CR><ET
X><CheckSum><CR><LF>
Field No. E1394 ASTM Field Name Valid Contents Max Length Mandatory
1 9.4.1 Record Type ID ‘O’ 1 Y
2 9.4.2 Sequence Number 6 Y
3^1 Specimen ID ( Sample ID(ID Mode) 22
3^2 <Sample ID>^ Sample No.(S. No. Mode) 6
3^3 <Sample No>^ Rack No. 4
3^4 9.4.3 <Rack ID>^ Position No. 2 Y
3^5 <Position No>^ ‘Y’ or ‘N’ 1
<Diluent>^ Sample replicate number
3^6 3
<Replicate Number>) (1-100)
4^1 Instrument Qc Name 20
9.4.4 N
4^2 Specimen ID ( Qc Lot No. 15
6
<Qc Name>^
4^3 <Qc Lot No.>^ Module 1
<Module>)
6 9.4.6 Priority ‘S’ or ’R’ 1 Y
Requested/Ordered
7 9.4.7 ‘YYYYMMDDHHMMSS’ 14 Y
Date and Time
0 - Serum
1 - Urine
2 - Plasma
16 9.4.16 Specimen Descriptor 3 - CSF 1 Y
4 - Gastric juice
5 - Ascites
6 - Other
26 9.4.26 Report Types ‘O’, ‘F’, ‘I’, ‘C’ or ‘X’ 1 Y
The Result Record to LIS contains the result of a single analytical determination.
Example:
<CR><ETX><CheckSum><CR><LF>
<CR><ETX><CheckSum><CR><LF>
Field No. E 1394 ASTM Field Name Field Contents Max Length Mandatory
1 10.1.1 Record Type ID R 1 Y
2 10.1.2 Sequence Number 5 Y
3 10.1.3 Universal Test ID 10 Y
4^1 Data result 10 N
Sample replicate number
4^2 3
(1-100)
10.1.4 Data or Measurement Value
1-Out Cell N
4^3 2-In Cell 1
(Control)
5 10.1.5 Units 10 N
(1) If send sample
result, first range =
6 10.1.6 Reference Ranges 30 N
normal range
second range =
7
critical range
separated by
repeat delimiter ('\')
(2) If send QC result,
first value=target
mean second
value=target SD.
separated by
repeat delimiter
('^').
(1) If send sample
result, ‘L’,’H’,’N’
or ’A’
Result Abnormal
7 10.1.7 (2) If send QC result, 1 N
Flags
QC
regulation(‘’,’1-2s’,
’1-3s’)
(1) If send sample
result, ‘’
(2) If send QC
8 10.1.8 Nature of Abnormality Testing result, ’+ 1SD’, ’- N
1SD’, ’+ 2SD’, ’-
2SD’, ’+3SD’, ’-
3SD’
9 10.1.9 Result Status ‘F’ or ‘C’ 1 Y
13 10.1.13 Date/Time Test Completed ‘YYYYMMDDHHMMSS’ 14 N
The Request Information Record to LIS is used to request sample program information from LIS. A
query is typically performed when a sample bar code has been read on the instrument and no sample
program information is found locally. All queries are made by sample ID.
Example::
<STX><FN>Q|1|^1^N001^1^N| |ALL||||||||O<CR><CTX><CheckNum><CR><LF>
<STX><FN>Q|1|AC2201023321^^N001^1^N| |ALL||||||||O<CR><CTX><CheckNum><CR><LF>
Field No. E1394 ASTM Field Name Field Contents Max Length Mandatory
1 10.1.1 Record Type ID ‘Q’ 1 Y
2 10.1.2 Sequence Number 6 Y
8
3^1 Starting Range ID Number Sample ID(ID Mode) 22
3^2 (<Sample ID>^ Sample No.(S. No. Mode) 6
3^3 <Sample No>^ Rack No. 4
10.1.3 Y
3^4 <Rack ID>^ Position No. 2
<Position No>^
3^5 ‘Y’ or ‘N’ 1
<Diluent>)
5 10.1.5 Universal Test ID ‘ALL’ Y
Request Information Status
13 10.1.13 ‘A’ or ‘O’ 1 Y
Codes
Description:
The Message Terminator Record to LIS is used with the Message Header Record to encapsulate
related patient information.
Example:
<STX><FN>L|1|N<CR><ETX><CheckNum><CR><LF>
Field No. E1394 ASTM Field Name Field Contents Max Length Mandatory
1 13.1.1 Record Type ID ‘L’ 1 Y
2 13.1.2 Sequence Number 6 Y
3 13.1.3 Termination Code ‘N’ 1 Y
3)Message Transmitted to Auto-Chemistry Analyzer from LIS Work Station
N: Field is not mandatory. The field have to be considered by the LIS, but it is
transmitted only if information is available at Liaison. When the information is
not available an empty field is transmitted.
Description:
The Message Header Record to the analyzer is the first record of each message.
Example:
This record is used to transfer patient sex and age to the analyzer.
9
Example:
<STX><FN>P|1||123456||ZhangDongdong|||M||||||40^Y<CR><ETX>[Checksum] <CR><LF>
Field No. E1394 ASTM Field Name Field Contents Max Length Mandatory
1 8.1.1 Record Type ID ‘P’ 1 Y
2 8.1.2 Sequence Number 6 Y
Laboratory Assigned Sample ID or Sample
4 8.1.4 22 N
Patient ID No.
6 8.1.6 Patient Name 30 N
9 8.1.9 Patient Sex ‘M’, ‘F’ or ‘U’ 1 N
15^1 Age(1-200) 3
8.1.15 Special Field 1 N
15^2 Age Unit (‘Y’,’M’,’D’) 1
Description:
This record holds information regarding analysis request and sample information to the analyzer.
Example:
<STX><FN>O|1|^1^N001^10^N^1||^^^ALT|R|20081124151012|||||||||1||||||||||O<CR><ETX><Chec
kNum><CR><LF>
Sample ID Mode:
<STX><FN>O|1|AB2201022318^^N001^10^N^1||^^^ALT|R|20081124151012|||||||||1||||||||||O<C
R><ETX><CheckNum><CR><LF>
Field No. E1394 ASTM Field Name Valid Contents Max Length Mandatory
1 9.4.1 Record Type ID ‘O’ 1 Y
2 9.4.2 Sequence Number 6 Y
3^1 Specimen ID ( Sample ID (ID Mode) 22
3^2 <Sample ID>^ Sample No. (S. No. Mode) 6
3^3 <Sample No>^ Rack ID 4
3^4 9.4.3 <Rack ID>^ Position No. 2 Y
3^5 <Position No>^ ‘Y’ or ‘N’ 1
<Diluent>^ Sample replicate number
3^6 3
<Replicate Number>) (1-100)
5 9.4.5 Universal Test ID 10 Y
6 9.4.6 Priority ‘S’ or ’R’ 1 N
Requested/Ordered
7 9.4.7 ‘YYYYMMDDHHMMSS’ 14 N
Date and Time
1 - Serum
2 - Urine
3 - Plasma
16 9.4.16 Specimen Descriptor 1 N
4 - CSF
5 - Gastric juice
6 - Ascites
10
7 - Other
26 9.4.26 Report Types ‘O’, ‘F’, ‘I’, ‘C’ or ‘X’ 1 N
Description
The Message Terminator Record to the analyzer is used with the Message Header Record to
encapsulate related patient information.
Example:
<STX><FN>L|1|N<CR><ETX><CheckNum><CR><LF>
Field No. E1394 ASTM Field Name Field Contents Max Length Mandatory
1 13.1.1 Record Type ID ‘L’ 1 Y
2 13.1.2 Sequence Number 6 Y
3 13.1.3 Termination Code ‘N’ or ’I’ 1 Y
Notice:
When the double direction communication apply the testing from the Analyzer To LIS system, If when
the LIS system didn’t find the sample that should be applied, must send:
ENQ
EOT
Examples
Analyzer [ENQ]
HOST [ACK]
[ETX]24[CR][LF]
HOST [ACK]
HOST [ACK]
[CR][ETX]55[CR][LF]
HOST [ACK]
[ETX]32[CR][LF]
HOST [ACK]
11
Analyzer [STX] 5R|2|^^^AST|1|U/L|^\^|N||F||||20090119123027[CR]
[ETX]0D[CR][LF]
HOST [ACK]
[ETX]F6[CR][LF]
HOST [ACK]
HOST [ACK]
Analyzer [EOT]
2. Apply from analyzer to LIS
Single:
Analyzer [ENQ]
HOST [ACK]
[ETX]25[CR][LF]
HOST [ACK]
HOST [ACK]
HOST [ACK]
Analyzer [EOT]
Batch:
Analyzer [ENQ]
HOST [ACK]
[ETX]25[CR][LF]
HOST [ACK]
HOST [ACK]
12
HOST [ACK]
HOST [ACK]
HOST [ACK]
HOST [ACK]
Analyzer [EOT]
3.Test order from LIS to analyzer
Sample ID mode:
Single:
HOST [ENQ]
Analyzer [ACK]
Analyzer [ACK]
Analyzer [ACK]
|1||||||||||O[CR][ETX]51[CR][LF]
Analyzer [ACK]
|1||||||||||O[CR][ETX]0F[CR][LF]
Analyzer [ACK]
|1||||||||||O[CR][ETX]3C[CR][LF]
Analyzer [ACK]
Analyzer [ACK]
HOST [EOT]
Batch:
13
HOST [ENQ]
Analyzer [ACK]
Analyzer [ACK]
Analyzer [ACK]
|1||||||||||O[CR][ETX]F9[CR][LF]
Analyzer [ACK]
|1||||||||||O[CR][ETX]02[CR][LF]
Analyzer [ACK]
Analyzer [ACK]
Analyzer [ACK]
Analyzer [ACK]
|1||||||||||O[CR][ETX]2E[CR][LF]
Analyzer [ACK]
|1||||||||||O[CR][ETX]1E[CR][LF]
Analyzer [ACK]
Analyzer [ACK]
Analyzer [ACK]
Analyzer [ACK]
14
HOST [STX] 3O|1|3977777^4^1^3^N||^^^AST|R|20090119100534||||||||
|1||||||||||O[CR][ETX]53[CR][LF]
Analyzer [ACK]
|1||||||||||O[CR][ETX]11[CR][LF]
Analyzer [ACK]
|1||||||||||O[CR][ETX]3E[CR][LF]
Analyzer [ACK]
Analyzer [ACK]
HOST [EOT]
15