XXXXXX - 1 1-: Burns
XXXXXX - 1 1-: Burns
XXXXXX - 1 1-: Burns
-------xxxxxx-----•-1
( 1 5)
. INCONEL 600 SHEATH
--- - - -�--------"LY" CABLE LENGTH
FULLY SUPPORTED SENSING ELEMENT
ENCLOSED IN ULTRA HIGH PURITY CERAMIC
1-
------- TEMP. LIMIT -50"C TO +200ºC(WITH STANDARD CABLE)------ �
.__ ______ "L" LENGTH :t.125 ---------
NOTES:
1. SEE SHEET #2. FOR ORDERING INFORMATION.
2. CAUTION:
RESISTANCE STABILITY IS HIGHLY DEPENDENT UPON -TOLERANCES- ■■•
■ ■"
••··■ Bu R Ns ENGINEERING
1--------�1------.------------.-------------------1
HANDLING. THEREFORE, IT IS NOT INCLUDED IN THE UNLESS OTHERWISE SPECIFIED
STANDARD WARRANTY. THIS SENSOR IS INTENDED TO ALL DIMENSION IN INCHES !!! ...
BE USED IN A LABORATORY SETTING AND SHOULD NOT FRACTIONS =±1/16
BE EXPOSED TO SEVERE THERMAL SHOCK, PHYSICAL ONE PLACE .X = ±.050 SCALE N.T.S. MOUNTING AND OUTLINE DRAWING,
SHOCK OR VIBRATION. CAREFUL HANDLING WILL ASSURE lWO PLACE .XX = ±.010 DFTM TSH 9-25-03 BUNRS SECONDARY REFERENCE
ACCURACY AND LONG LIFE. THREE PLACE.XXX = ±.005__,
- E ±_ _0_º _30_' CHKD JPZ 9-25-03 TEMPERATURE STANDARD, 25.5 OHM
,___A L-L AN_G _LES_AR_
3. BURNS MODEL 19439 IS DESIGNED FOR LABORATORY USE SHEATH AND LEAD LENGTHS APPD JPZ 9-25-03
WHERE WIDE TEMPERATURE RANGES (-200 ºC TO 500 ºC) PER BURNS PIN 17026
ARE MONITORED. UN LESS OTHERWISE NOTED: DIMENSIONS IN INCHES SIZE DRAWING NUMBER REV
SHEET 1
ALL SURFACES 12 REF: 19453 A 19439- A
,e; OF 4
ALL FINISHES IN MICRO INCHES
ORDERING INFORMATION:
19439 SECONDARY STANDARD, 25.5 OHMS
CALIBRATION
-O O= NO CALIBRATION
-1 1 = TABLE FROM OTO 500ºC IN 1ºC INCREMENTS
-2 2 = TABLE FROM -200 TO 500ºC IN 1ºC INCREMENTS
-3 3 = TABLE FROM -200 TO 300ºC IN 1ºC INCREMENTS
(FOR PROBES < 9" IN LENGTH)
TEMPERATURE SCALE
TIME CONSTANT: 9 SECONDS TYPICAL FOR 63.2% RESPONSE TO STEP CHANGE IN IMMERSION TEMPERATURE IN WATER@ 3 FT/SECOND.
LONG TERM STABILITY: STABILITY AT 0ºC AFTER EXPOSURE TO VARIOUS UPPER TEMPERATURE LIMITS. STABILITY ESTIMATES ASSUME PERIODIC
USAGE AT THE MAXIMIM USE TEMPERATURE DURING THE COURSE OF ONE YEAR.
200ºC .01ºC
300ºC .02ºC
400ºC .05ºC
500ºC .10ºC
PROBE ACCURACY: INCLUDES CALIBRATION UNCERTAINTY AND SHORT TERM REPEATABILITY - DOES NOT INCLUDE SENSOR DRIFT.
TEMP. UNCERTAINTY
-196ºC ±.010ºC
o ±.007ºC
200 ±.024ºC
420 ±.033ºC
A
DRAWING NUMBER
19439-
REV
A
BURNS