Calculation Algorithms TPS
Calculation Algorithms TPS
From: http://www.merriam-webster.com/dictionary/
Where, in the planning system, do we
find algorithms?
• MU calculations
• Isodose distributions
• DVH generation
• IMRT optimization
• DRR generation
• Brachytherapy calculations
• Any process that occurs when the user does not
dictate each step (e.g. generating a 3D image from a
series of slices, placing a margin around a structure,
etc.)
In the beginning …
• Calculations were performed
strictly based on empirical
(directly measured) data in
tabular format.
r ( x, y , z )
6 D ( x, y , z , , , E )
Monte Carlo accuracy - Modeling
• Highly dependent on
the modeling of the
components of the
accelerator head
http://health-7.com/imgs/20/7507.jpg
• Finite photon source size
Monte Carlo input Open fluence distribution
Fluence modulation
Head scatter sources
flattening filter
collimators
wedges
Monitor back scatter
Collimator leakage, including
MLC interleaf leakage
shape of MLC leaf ends
• Beam spectra
• Spectral changes
• Electron contamination
Monte Carlo Codes
• EGS4
• EGSnrc
• GEANT
• PENELOPE
• MCNP
• MCNPx
• FLUKA
Different types of calculation
algorithms
• Semi-empirically based
• Model based
• Direct Monte Carlo
• Hybrid
Give me 30 minutes and I can confuse
anyone. I don’t need to prepare.
High density
heterogeneity
Dr r Ar r 'd 3 r '
Convolution-
Tp ( r r ) A r r ' r r 'd 3 r '
Superposition
where r r ' r r 'is the radiologic distance from the dose
Equation
deposition site to the primary photon site and r r is the radiologic
distance from the source to the photon site
Data and Clinical Examples
Treatment Planning and
Measurements
Region of non-
equilibrium
1.4 1.4
Plot = Di/Dh
Das et al., 2008 (11)
Lung Case
1. SBRT case –
10 fields
2. 1 cm volume
3. 6 MV
4. AAA and Pencil
beam
5. Heterogeneity
corrections
“on”
AAA
100% White
95%
90%
80%
Calculation verification
program predicts 11.3%
higher dose than AAA with
same MUs
Pencil Beam
100% White
95%
90%
80%
Calculation verification
program agrees well with
Pencil beam (.01%)
Comparison of isodose curves
AAA Pencil Beam
Pencil Beam
MU
sum = 208
AAA – edge of heterogeneity PBC– edge of heterogeneity
www.pythian.com
An Irish blessing for the Medical
Dosimetrist …
May your first optimization meet your dose constraint
If not, may you meet with a little dose paint
May your calc’d and your plan MUs always agree
May you need not replan for a re-drawn GTV
May your MD not change his mind post approval
And your user rights never suffer removal
May your IMRT QA turn out right
And you get to go home while there is still light
Questions?