Tethys UV300 Manual
Tethys UV300 Manual
Contents
2 Installation …………………………………………….…….………………… 10
3 Operating …………………………………………….……...………………… 23
4 Parameters …….................................…………….……...………………… 64
4.1 Aluminium
4.1.1 Principle …………………………………………………………. 64
4.1.2 Settings ….………………………………………………………. 65
4.1.3 Test ………………………………………………………………. 70
4.1.4 Calibration …………………………………………………...…. 72
4.1.5 Maintenance and troubleshooting …..………………………. 74
4.1.6 Specifications ……...………………………..…………………. 76
4.3 Chlorophyll A
4.3.1 Principle …………………………………………………………. 90
4.3.2 Settings ….………………………………………………………. 91
4.3.3 Test ………………………………………………………………. 94
4.3.4 Calibration …………………………………………….……...…. 96
4.3.5 Maintenance and troubleshooting …..…………….….………. 98
4.3.6 Specifications ……...………………………..……….…………. 99
4.4 Color
4.4.1 Principle …………………………………………………………. 100
4.4.2 Settings ….………………………………………………………. 101
4.4.3 Test ………………………………………………………………. 104
4.4.4 Calibration …………………………………………….……...…. 106
4.4.5 Maintenance and troubleshooting …..…………….….………. 108
4.4.6 Specifications ……...………………………..……….…………. 109
4.5 Conductivity
4.5.1 Principle …………………………………………………………. 110
4.5.2 Settings ….………………………………………………………. 111
4.5.3 Test ………………………………………………………………. 114
4.5.4 Calibration …………………………………………….……...…. 115
4.5.5 Maintenance and troubleshooting …..…………….….………. 117
4.5.6 Specifications ……...………………………..……….…………. 118
4.8 Iron
4.8.1 Principle …………………………………………………………. 137
4.8.2 Settings ….………………………………………………………. 138
4.8.3 Test ………………………………………………………………. 143
4.8.4 Calibration …………………………………………….……...…. 145
4.8.5 Maintenance and troubleshooting …..…………….….………. 147
4.8.6 Specifications ……...………………………..……….…………. 149
4.11 ORP
4.11.1 Principle …………………………………………………………. 174
4.11.2 Settings ….………………………………………………………. 175
4.11.3 Test ………………………………………………………………. 178
4.11.4 Calibration …………………………………………….……...…. 179
4.11.5 Maintenance and troubleshooting …..…………….….………. 181
4.11.6 Specifications ……...………………………..……….…………. 181
4.12 pH
4.12.1 Principle …………………………………………………………. 182
4.12.2 Settings ….………………………………………………………. 183
4.12.3 Test ………………………………………………………………. 186
4.12.4 Calibration …………………………………………………...…. 187
4.12.5 Maintenance and troubleshooting …..………………………. 189
4.12.6 Specifications ……...………………………..…………………. 189
4.14 Temperature
4.14.1 Principle …………………………………………………………. 203
4.14.2 Settings ….………………………………………………………. 204
4.14.3 Test ………………………………………………………………. 207
4.14.4 Calibration …………………………………………….……...…. 208
4.14.5 Maintenance and troubleshooting …..…………….….………. 210
4.14.6 Specifications ……...………………………..……….…………. 211
4.17 UV254
4.17.1 Principle …………………………………………………………. 229
4.17.2 Settings ….………………………………………………………. 230
4.17.3 Test …………………………………………………………….... 234
4.17.4 Calibration ……………………………………………….…...…. 236
4.17.5 Maintenance and troubleshooting …..…………….....………. 239
4.17.6 Specifications ……...………………………..…………….……. 240
---
1 Quick start
Step 1
Unpack the analyser, check that nothing has been damaged during the transportation and fix it
on a wall. The dimensions are given on chapter 2.
Step 2
Connect the sample inlet and outlet with plastic tubing, ID 6.4 mm (1/4”) OD 9.6 mm (3/8”).
Refer to chapter 2 for inlet and outlet location if necessary.
Assemble the fitting elements as below:
With the peristaltic sampling pump, the inlet pressure must be limited to 1 Bar (equiv.
to 1000 hPa or 15 psi)
Step 3
Depending on the options, refer to chapter 2 to connect the 4-20 mA output or the RS232 cable.
If a Wi-Fi link has been chosen as an option, check that the antenna is properly fixed on the
right side of the enclosure.
Step 4
Connect the power cable to a grounded power socket.
To avoid electric shock, the analyser must be connected to the earth via the power
socket.
Step 5
Check that the date and time on the bottom right of the screen are correct. If not, refer to
chapter 3 for setting date and time.
Step 6
Refer to chapter 3 to configure the optional interface (4-20 mA or RS232 or Ethernet or Wi-Fi).
Step 7
The default mode is the stopped mode (no automatic measurements, no automatic cleaning).
Set the measuring mode (continuous, periodic or extern) depending of the application (refer to
chapter no 3).
The analyser is factory calibrated and is now ready to start the measurements.
For checking and recalibration, refer to chapter 4 for each individual parameter.
2 Installation
2.1 Fixing
Make 4 holes in the panel or in the wall for 8 mm diameter screws according to the pattern
below:
Refer to the drawing and table below for the hydraulic connections.
Flow diagrams are given on the chapter 2.3.
3 1
1 Sample inlet
2 Sample outlet
The sample inlet and outlet must be connected with plastic tubing, ID 6.4 mm (1/4”) OD 9.6 mm
(3/8”).
Reagents or cleaning solution must be connected with plastic tubing, ID 3.2 mm (1/8”) OD 6.4
mm (1/4”)
2.3.1 Flow diagram for direct absorbance or fluorescence (UV254, NO3, Colour, turbidity,
PAH, chlorophyll A)
4 3
2
H2SO4 5%
1 Flow cell
2 3-way solenoid valve
3 Cleaning pump
4 Optional sampling pump
2.3.2 Flow diagram for colorimetric method (Cl2, NO2, Al, Fe, etc...)
1 5
Reagent 1
6
Reagent 2
7
4
2
H2SO4 5%
1 Flow cell
2 3-way solenoid valve
3 Cleaning pump
4 Optional sampling pump
5 Mixing pump
6 Reagent 1 pump
7 Reagent 2 pump
All the cables must be installed using the glands as show below.
The module OUT4-20-400 delivers an active 4-20 mA signal for PLC, recorders or SCADA
system. Each OUT4-20-400 module is affected to one measuring channel.
Refer to chapter 3 to set up the 4-20 mA module position and scale values for each channel.
Screw terminal J2
The Wi-Fi module follows the IEEE 802.11b standard and is internally connected to a SMA
connector on the right side of the analyser.
The 2.4 GHz antenna must be fixed on this SMA connector as represented below.
Refer to chapter 3 (communications) to enable the Wi-Fi communication and to set up the
network name (SSID) and the WEP key.
The Ethernet module provide a RJ-45 jack that meets the ISO 8877 requirements for
10/100BASE-T. See the following figure and table for pin orientation and pin assignments.
Refer to chapter 3 (communications) to select the Ethernet port and to set up the IP address
and mask.
The RS232 port has a Sub-D 9 pins male connector and is located on the screen board
DSP400TFT, position J15.
Refer to chapter 3 (communications) to select the communication mode and the baud rate.
The format is always 8 bit, no parity, 1 bit stop.
To connect this port to a laptop (directly or through or a USB adaptor), use a null cable (pin 2 to
pin 3, pin 3 to pin 2, pin 5 to pin 5).
The RS485 port has a 5-pin screw terminal located on the screen board DSP400TFT, position
J38 as shown below:
The module MPH400 accepts the direct connection of a pH electrode on the screw terminal B1.
For automatic temperature compensation, a Pt100* or Pt1000* sensor of the electrode must be
connected on B2.
The module MCOND400 accepts the direct connection of a conductivity electrode on the screw
terminal B1. Conductivity constant of K=0.01, K=0.1, K=1 and K=10 are accepted, refer to
chapter 4 to setup the right constant.
For automatic temperature compensation, a Pt100* or Pt1000* sensor of the electrode must be
connected on B2. Refer to chapter 4 to adjust the temperature coefficient.
The module MIN4-20-400 accepts the connection of a 4-20 mA output from any probe or
external measuring system. The floating input is insulated by optocouplers.
A 15V DC output (insulated, max 20 mA) is available on the connector B2. This output must be
used for any probe with electrical contact to the water sample (example: chlorine probe).
The module MRELAY400 gives on B1 a normally closed (NC) and a normally open (NO) relay
contacts for different functions. Refer to chapter 3 to select the right function, threshold, etc...
Screw terminal B1
Pin 1 NO contact
Pin 2 Common
Pin 3 NC contact
Note: pin 1 on left side
Contact rating:
5A @ 277VAC/100VDC resistive
The module MIL accepts two logical signals from a PLC to control the measurement cycles.
The START input is used to start a measurement cycle for all the channels when a pulse (6 to
48V DC) is received. The minimum pulse width is 200 mS. The external measuring mode must
be selected.
The INHIBIT input is used to stop and inhibit any further automatic measurement (periodic mode
or continuous mode) as long as this input is maintained high (6 to 48 VDC).
Note: as these two input signals refer to different measuring mode, it has no sense to use both
of them on a same application.
To avoid electric shock, the analyser must be connected to the earth via the power
socket. Check that the power socked is earthed.
Disconnect the power cord or mains before any wiring, connection or servicing inside
the analyser.
If all the conditions above are satisfied, plug the power cord into a grounded power socket.
If the power cord needs to be changed or removed, the mains must be connected to the screen
board DSP400TFT on the screw terminal J1 as follows:
Line
Earth
Neutral
2.4.12 USB
A USB key (memory stick) can be connected on the USB connector type A represented on the
drawing of the section 2.4 for measurements download, configuration download or upload, or for
upgrading the analyser software. Refer to the Operating section for using the USB key.
Disconnect the USB key as soon as the operation if finished as USB connection is
not watertight and as it may block the analyser's normal operation.
3 Operating
When powered on, the analyser displays the last process screen selected after a fast hardware
check.
Then the measurements start automatically and the process values and graphs are updated as
soon as necessary (except in stopped mode that is the default mode after delivery, see on the
general parameters screen to change the mode).
To change the screen, press on one of the following button as shown below.
No password is required by default but it's recommended the user create one, refer to the
general parameters screens.
Process
graph screen
Process values
screen
Zero screen
The value process screen is one of the two running screen of the analyser. It is display a few
seconds after power on and remains displayed all the time unless another screen has been
selected.
This screen displays the very last measurement and the recorded measurements graph for the
selected channel on the right.
3 6
9 7 8 10 7 11
The label is automatically set when the channel is created. It can be modified in the channel
parameters screen.
The unit is automatically set when the channel is created. It can be modified in the channel
parameters screen.
3 Temperature
Display the auxiliary temperature measurement of the channel. It concerns only the channels
with automatic temperature compensation.
4 Channel value
This value is the very last measurement done for each channel. The number of digit can be
modified in the channel parameters screen. By default, negative values resulting from normal
measuring fluctuations or from a wrong zero are displayed. They can be eliminated and replace
by zero by changing a setting on the channel parameters screen.
This line correspond to the high alarm threshod. This threshold is also used to define the default
Y scale.
6 Cursor move
This two keys move the cursor line right or left. The date, time and value at the current
intersection between the graph and the cursor is displayed.
8 Time scale
This two keys select larger or smaller time scale from the following: 30 days, 5 days, 24 hours, 1
hour, 10 minutes.
This field displays the current date and time. To change the date or the date format, refer to the
clock screen on the general settings.
Depending on the measuring mode, this field displays information about the measurements
schedule.
If an inhibit signal is send to the logical input module MIL (for any mode except the stopped
mode), or if a cleaning cycle is engaged, this field displays “HELD”.
a) For continuous mode, this field displays “Measuring...” followed by the label of the
channel being currently acquired
- During settling down time (if any), this field displays “Settling down...”
- If a measuring cycle command has been received from the logical input module MIL
or through MODBUS communication, a measuring cycle starts and the display is one
of the following:
- During settling down time (if any), this field displays “Settling down...”
Depending on the cleaning method, this field displays information about the cleaning schedule.
- Between cleaning cycle, the time remaining for the next cleaning cycle is displayed,
example: “Next cleaning: 01:03:24” means that the next cleaning cycle will occur in 1
hours 3 min and 24 sec.
- During purging by acid, the time remaining is displayed, example: “Purging... 00:00:15”
means that purging will continue during 15 sec.
- During the waiting time, the time remaining is displayed, example: “Waiting... 00:00:05”
means that waiting time will continue during 5 sec.
- During the recovery time, the time remaining is displayed, example: “Recovery...
00:00:20” means that recovery will continue during 20 sec.
The value process screen is one of the two running screen of the analyser. It is display a few
seconds after power on and remains displayed all the time unless another screen has been
selected.
This screen displays the very last measurement for all the active channels.
5 6 7
The label is automatically set when the channel is created. It can be modified in the channel
parameters screen.
The unit is automatically set when the channel is created. It can be modified in the channel
parameters screen.
3 Temperature
Display the auxiliary temperature measurement of the channel. It concerns only the channels
with automatic temperature compensation.
4 Channel value
This value is the very last measurement done for each channel. The number of digit can be
modified in the channel parameters screen. By default, negative values resulting from normal
measuring fluctuations or from a wrong zero are displayed. They can be eliminated and replace
by zero by changing a setting on the channel parameters screen.
This field displays the current date and time. To change the date or the date format, refer to the
clock screen on the general settings.
The zero screen enables to start a cleaning/zeroing cycle at any time, additionally to the
automatic cleaning/zeroing cycle that occur normally once per day.
The date and time of the ten last zeroing cycle are displayed on this window, followed by the
new offset applied compared to the last zero. This value should normally stays close to zero.
The two button enable to scan up and down the last ten cleaning/zeroing cycles.
This line describes the current operation of the engaged cleaning/zeroing cycle.
The typical cycle is:
This button starts a cleaning/zeroing cycle. If a cycle is already engaged by error, it may be
stopped by pressing on this button.
This screen allows to select different screens to check and change the general configuration of
the analyser:
- Measuring mode
- Relays functions
- Time and date update
- Password
- Language
- RS232 communication
- RS485 communication
- Multiplex mode (sequential only)
- Zeroing
- Logical input module
- USB screen copy
- Reagent and cleaning solution level monitoring
This screen enable to select the right measuring mode. Most of the applications uses the
periodic mode.
1 Measuring mode
- Continuous mode
The measurements are continuously updated at the highest possible rate. The display and the
4-20 mA output (if any) are updated every second. This mode apply only for internal turbidity
and electrode-based measurements (pH, conductivity, DO, 4-20 mA input, external turbidity).
- Periodic mode
The measurements are done at a specific period of time define in the “Measuring period” field.
The display and the 4-20 mA output (if any) are updated at the end of the measurement cycle.
The measurements are stored in the internal memory at each cycle.
If the measuring cycle time is higher than the defined period, a new measurement will start
immediately after the previous one.
The measurement can be held at anytime if an high level voltage has been applied on the inhibit
input of the logical module MIL (if any).
Even in periodic mode, internal turbidity and electrode-based measurements stay in continuous
mode.
- External mode
The measurement cycle is started or by a pulse (>0.3 sec) send to the start input of the logical
module MIL, or by writing a value of 1 in the start measurement register by a MODBUS
command.
2 Measuring period
For periodic mode only, the measuring cycle occurs with a period of time defined in this filed.
For periodic mode only, the sampling pump is running before every measuring cycle during a
time defined in this field.
In periodic mode only, it’s possible to add a settling down time defined in this filed after the
sampling and before the measurement to allow suspended solid to go down.
This mode concerns only a few specific applications, in all other cases, let the default value of 0.
3.5.2 Relays
This screen allows to define the position of the optional relay modules.
Up to 6 relay modules can be installed.
Pressing on this button enable to check or select the function of the specified realy.
Refer to the next screen for the functions.
This field allows to enter the relay module position among the 12 available sockets M1 to M12.
On this example screen, relay no 1 is on position M9 and relay no 2 on position M10.
This following screen allows to define the function of each relay module selected on the
previous screen.
In duplicated multiplex mode, the n first relays are systematically affected to the multiplexing of
the n streams. No changes can be done.
Activated when the measuring channel overpass the high alarm threshold.
Boundaries apply for electrode-based measurements according to the schematic below.
Activated when the measuring channel is under the low alarm threshold. Boundaries apply for
electrode-based measurements according to the schematic below.
- Fault
Any error detected during a measuring cycle activates this relay until cleared by a new
measurement
- Ready
Relay activated when the analyser is ready to measure, deactivated during measuring cycle or
cleaning cycle
In periodic mode only, it’s possible to synchronize a second external sampling pump through a
relay if this function is activated. On time and off time as define in the next screen.
External turbidity and dissolved oxygen (if declared) are exceptionally measured when the on
time is finished and after a delay specified in the "waiting time" field (next screen).
The sampling chronogram is shown below:
When the multiplex mode (duplicated channel mode) has been selected, the concerned relays
are automatically selected for this function.
- Waste collect
This relay function enable to drive an external solenoid valve to collect the sample to prevent
pollution after that the reagent has been added and during a limited purge time. This function
concern only the colorimetric methods.
- Purging
This function concerns only NH4 and H2S measurement. It enable to drive a solenoid valve to
drain the stripping pot at the end of the measuring cycle.
- Periodic
This function is used to drive an external filter cleaning cycle every n measurements. n is define
on the next screen.
3.5.3 Clock
Note: any engaged measuring or cleaning cycle will be aborted in case of time change.
3.5.4 Password
The following screen enable to enter a password to access to any parameter screen access.
The initial password is 0 (no password entering). The code 3333 can be used in case of
password lose.
3.5.5 Language
This screen allows setting or changing the communication settings for the serial ports RS232.
1 Communication mode
1) MODBUS
The MODBUS communication mode allows reading the last measurements or the error codes
as well as starting a new measuring or cleaning cycle. The MODBUS protocol is a standard and
simple protocol implemented on most of the SCADA systems. If necessary, the full protocol can
be downloaded at www.modbus.org .
Basic frames using the function no 3 as shown below allows to read the measurements:
Slave
number
3 Address of first byte NB of bytes CRC16
Slave ……
3 NB of read bytes 1st byte 2nd byte CRC16
number
The measurement values follows the floating point standard IEEE754, format single 32 bit (4
bytes), higher first, lower last. Invalid codes are 1 byte long.
2) HTTP/HTML5
This protocol allows through a Wi-Fi or an Ethernet interface to drive the analyser through an
internet browser.
- Push the switch SW1 bellow the RS232 connector in direction of the "SW1" label
- Connect the module directly to a PC with an emulation terminal software using the RS232
cable
- Start the terminal emulation software at 2400 baud, 8 bit, no parity, 1 stop bit, no flux control
Save
Reset
- Connect the Acksys WL-DONGLE to the RS232 port of the UV300 analyser
- Double check that an Wi-Fi antenna is connected on the WL-DONGLE, "Ant 2" side
- Push the switch below the RS232 connector of the Ethernet board in the “Admin” position .
The orange LED must blink twice per second.
- Connect the module directly to a PC with an emulation terminal software using the RS232
cable
- Start the terminal emulation software at 2400 baud, 8 bit, no parity, 1 stop bit, no flux control
Note: the IP address and net mask must be chosen by the network administrator
- Push back the switch below the RS232 connector of the Ethernet board. The orange LED
must stop to blink.
- Connect a cable on the RJ45 connector to the network. The green LED “Link Ether” must be
ON immediately.
The baud rate must be fixed strictly in accordance with the system connected on the RS232 port
and generally at the higher speed compatible with the media capacity.
Each device on a MODBUS network has a specific number. This number can be entered in this
field. The default value is 1.
This screen allows to configure the RS485 probes the first time they are used.
Only one probe must be connected during this configuration phase.
1 Probe initialisation
If the operation failed, disconnect the probe, reconnect it and press again on
"RESET" key because the probe take in account the new baud rate only on power
on.
2 3
1 Stream selection
The ON/OFF keys allow to select the proper number of stream for the sequential mode.
In this mode, the measurements are done sequentially on a same existing channel.
The stream number is stored in the measurements records table.
2 Circulation time
This field allows to fix the circulation time of each stream for a good renew, function of the
sample flow.
The measurement of any stream can be inhibited by a continuous level on the logical input
module MIL, if for example there is no sample temporary. The logical module position must be
entered in this field.
1 Cleaning/zeroing period
This field contains the period of time between two cleaning cycles.
The recommended period is 24 hours.
2 Purging time
A too short cleaning time may cause a bad zeroing with the auto-zeroing function.
Keep the default value of 60 seconds for a good zeroing.
3 Waiting time
This field contains the waiting time between the end of the cleaning solution pumping and the
zeroing to enable any bubbles produced by the cleaning pump to go up.
The default value is 5 seconds. It is recommended to keep this default value unless authorized
by the manufacturer.
This field contains the recovery time before starting again the measurements on the sample
after a cleaning operation. The default value is 30 seconds. It is recommended to keep this
default value unless authorized by the manufacturer.
A too short recovery time may cause erratic measurements during the few seconds
that follow a cleaning cycle.
Keep the default value of 30 seconds for a good sample recovery.
5 Auto zero
If acid cleaning is available, an auto-zero can be performed automatically on the acid cleaning
solution if selected.
If the auto-zero is selected, the cleaning solution must be prepared with pure water
and analysis grade acid to get a correct zeroing.
If activated, the acid cleaning will start automatically according to the defined period.
It may be remove only in case of very specific conditions like the absence of
cleaning solution.
This button show the actual state of the START input. ON means that a level above 6V DC is
applied on the input.
This button show the actual state of the INHIBIT input. ON means that a level above 6V DC is
applied on the input.
The last 5000 measurements are recorded in the cyclic internal memory and are transferred to
the USB key as a text file when plugged on the USB port. The file name is MXXXXYYYY.txt with
XXXX being the board serial number and YYYY an incremental number starting at 0001.
Each line of measurement starts by the date and time as shown in the example below:
;Cl2;TURB;pH;COND;TEMP;;;;;;;;;;;
;mg/l;NTU; ;uS;C;;;;;;;;;;
01/04/13 11:04:04; 0.55; 0.32; 7.45; 35.19; 18.46; 0.00; 0.00; 0.00; 0.00; 0.00; 0.00; 0.00; 0.00; 0.00; 0.00; 0.00
01/04/13 11:14:04; 0.54; 0.38; 7.47; 37.78; 18.51; 0.00; 0.00; 0.00; 0.00; 0.00; 0.00; 0.00; 0.00; 0.00; 0.00; 0.00
01/04/13 11:24:04; 0.53; 0.36; 7.46; 36.54; 18.66; 0.00; 0.00; 0.00; 0.00; 0.00; 0.00; 0.00; 0.00; 0.00; 0.00; 0.00
The channel measurements are separated by a semicolon and can be imported to Excel® as
show below:
A diagnostic file is also transferred at the same time to the USB key as a text file.
It contains the main configuration parameters and the most important signals that may be useful
for troubleshooting.
The file name is DXXXXYYYY.txt with XXXX being the board serial number and YYYY an
incremental number starting at 0001.
b. Screen copy
The current screen is copied on the USB key as a bmp format file (Windows® compatible) if the
screen copy function has been activated before as shown on the screen below.
This function is automatically deactivated after one hour or in case of power off.
The complete configuration can be backup and restored as explained on section 3.6.9.
d. Software update
The internal software of the analyser can be updated from a USB key by following this
procedure:
Once the software update is finished (about one minute), you must reload the configuration
from the USB key (refer to section 3.6.9).
This screen enable to check the level of the cleaning solution and reagent, and to indicate that a
refill has been done.
3 4
1 Level
This field displays the estimated level based on the previous refill, the pump flow rate, the tank
capacity and the time of use.
2 Refill button
This button must be pressed at each refill. It reposition the level at 100%.
3 Pump flow
4 Tank capacity
This screen allows to test the RS232 port. Each time this screen is called, it send the sequence
"QWERTY0123456789".
Any character received during that this screen is displayed is written after the label "Received:".
It recommended to use a terminal emulator communication software to test the RS232 port.
This screen allows to list all the RS485 probes declared on measuring channels.
The probe model is displayed and followed by "ok" if recognised, otherwise a red message "No
probe" is displayed if the probe is not connected or not recognised.
To test the port itself, press on "TEST COM": each time this screen is called, it send the
sequence "QWERTY0123456789".
Any character received during that this screen is displayed is written after the label "Received:".
It recommended to use a terminal emulator communication software to test the RS485 port.
To determine the address and configuration of any probe connected on the RS485 port, press
on "SCAN". All the 254 possible addresses are tested to check all the connected probes. The
status of the probe is displayed: linked to an measuring channel or unlinked.
This screen displays the diagnostic MODBUS counters defined by the MODBUS protocol.
The counters can be all reset to zero by pressing the key RESET.
Counter 1 displays the total number of valid frames received, whatever they are intended or not
for the analyser.
Counter 2 displays the number of frames received with a CRC error, whatever they are
intended or not for the analyser.
Counter 3 displays the number exception answers preceded. Exception answer is produced in
case of unrecognized command or wrong parameter on a received frame.
Counter 4 displays the number of valid frames received specifically intended for the analyser.
Counter 5 displays the number of valid broadcast frames received. Broadcast frame are not
usual.
Counter 8 displays the number of characters received with a format error, whatever they are
intended or not for the analyser.
Counter 9 displays the number of frames proceeds for which an answer has been sent to the
host.
This screen enable to check individually the 12V DC outputs that drives internal pumps and
solenoid valves.
This screen enable to check individually the relays modules. The relay module position is
indicated on the right column.
To activate a relay, press on the key OFF. To deactivated it, press again on the key.
This screen displays the software release and the board number.
This screen allow to declare a parameter. It is normally used only by the manufacturer.
Press any key to declare a parameter if the corresponding hardware is installed.
On this example, the turbidity channel has his photo detector on position no 2 (from left on the 8
photo detector sockets J8 to J15) and his source on position no 2 (from left on the 6 source
sockets J16 to J21)
1 Clear measurement
This button store in the internal EEPROM all the analyser configuration. This function is
normally used only by the manufacturer.
This button restore form the internal EEPROM all the analyser configuration defined when the
analyser was delivered. This function can be used in case of doubts about the current
configuration.
This button save in a .cfg file on the USB key the analyser configuration. This function is
recommended before doing a software update as the configuration is erased.
This button load from the USB key all the analyser configuration previously saved by the
function above. The operator must choose the right configuration file in case of several
configuration files stored in the USB key.
6 Full re-initialisation
This button erase all the configuration (channel, parameters, general settings...).
It may use exceptionally in case of change of board.
The configuration must be re-introduced manually or from a configuration file saved on the USB
key.
This value can be checked with an ammeter connected directly on the 4-20 mA output
(disconnect previously any cable on the output).
3 Calibration
This button enable to calibrate the 4-20 mA output if the value slightly differ from the theoretical
ones.
First enter the value read on an ammeter for 4-20 mA (example 3.96 mA), then enter the value
read for 20 mA (example 19.95 mA).
Then the 4-20mA output is recalibrated. It can be tested with the function below.
This button allows to scan the 4-20 mA output mA by mA to check the output or the connected
system.
This field allows to enter the measured value corresponding to the full scale of 20 mA.
This field allows to enter the measured value corresponding to the bottom of the scale at 4 mA.
Generally, this value stays at 0.
3.6.11 Temperature
This screen gives the temperature reading from the internal temperature probe.
This temperature do not apply for the external probes (pH, conductivity,...).
This screen displays the internal ADC (Analogue Digital Converter) values for the 8 first
channels.
The field offset enable to introduce an offset in order that the values are centred to zero
(photodiodes disconnected). This offset has no influence on the measurements but improve the
touch screen position accuracy.
The FLASH key launches flashes on all the connected light sources during a few seconds for
testing purpose only.
4 Parameters
The next section describes parameter by parameter the measuring principle, settings, testing,
calibration, troubleshooting and specifications.
4.1 Aluminium
4.1.1 Principle
The measuring principle is based on the pyrocatechol violet colorimetric method. A small
quantity of reagent and buffer are injected on the sample. After a fixed reaction time, the
developed colour is measured by the light absorption according to the Beer-Lambert law:
[C ] = k ⋅ log I 0
I1
4.1.2 Settings
First screen:
1 Label
This field allows changing the channel label displayed with the measurement. The label size is
limited to 6 characters.
2 Unit
This field allows changing the channel unit displayed with the measurement. The unit size is
limited to 6 characters.
The high alarm level can be changed by pressing on this field. A default value is preset when
the channel is created. The high alarm level is primary used to fix the high limit represented by a
red line on the recorded measurement graph. The limit is used as the minimum graph scale.
The secondary use is to drive a high alarm relay if the relay button has been activated and if a
relay module has been installed on the selected module socket. In this case, the relay will be
activated as soon as the measurement overpass the high alarm value and will stay activated
until the measurement goes down below this value after subtracting the boundary value.
Refer to installation section for the relay wiring. Both functions NO (Normally Open) and NC
(Normally Closed) are available.
The low alarm level can be changed by pressing on this field. A default value is preset when the
channel is created. The low alarm level is primary used to fix the low limit represented by a
green line on the recorded measurement graph.
The secondary use is to drive a low alarm relay if the relay button has been activated and if a
relay module has been installed on the selected module socket. In this case, the relay will be
activated as soon as the measurement underpass the low alarm value and will stay activated
until the measurement goes down below this value after adding the boundary value.
Refer to installation section for the relay wiring. Both functions NO (Normally Open) and NC
(Normally Closed) are available.
5 Number of decimals
This field allows changing the number of decimal displayed for the measurement value to adjust
the display to significant decimals only.
Note: this choice does not affect the measurements stored in the internal memory or the
MODBUS registers.
Negatives values are normal for most of the measurements as they result from normal
fluctuations if the measurement is close to zero.
By default, the display of negative values is allowed as negative values may also result from a
wrong zeroing. Consequently, it may be useful to be aware of such problem when seeing
strongly negative values.
But as negative values have no theoretical meaning, it’s possible to replace them by zero by
selecting "NO" for this function.
Note that this function do not apply to the recorded measurements or to the MODBUS registers.
Second screen:
1 Scale factor
This factor is normally determined in factory to transform the basic measurement to a scaled
value in the final unit. It is has no reason to be changed except if optical elements are replaced.
2 Averaging
This field displays the number of elemental acquisitions averaged to form one measurement.
The default value is 50.
3 Sampling time
This field displays the sampling time specifically for this measurement.
A minimum time of 10 sec is recommended to allow a good renew of the sample.
It might be longer depending on the length and height of tubing to take the sample.
4 Reagent time
This field displays the reagent time injection. The default value is 1 sec.
Shorter time may produce instable measurements or limiting the measuring range.
Longer time are not necessary and generate a waste of reagent.
This field displays the colorimetric reaction time. The default value is 120 sec and this delay is in
accordance with the standard method.
Shorter times may reduce the measurement stability, especially for low temperatures.
6 Waste time
This field displays the purge time after the measurement. The default value is 5 sec.
The waste can be collected during this time if a relay module declared with the "waste" function
is installed.
7 Temperature coefficient
The colorimetric reaction might be slightly affected by the temperature. This coefficient is used
for the automatic temperature compensation.
Third screen:
1 Compensation
This button allows to compensate the sample turbidity. It's mandatory for turbid sample,
otherwise the measuring value should be underestimated or even negative.
2 LED selection
This field display the LED chosen for the measurement. Never change it, unless the colorimetric
method is changed.
4.1.3 Test
The check screen displays the last measurement performed, whatever manual or automatic,
with the reaction curve.
5 6 7 8 9 10
1 Reaction curve
This curve corresponds to the last reaction curve, or to the current reaction during a measuring
cycle. Intermediate measurements are performed every 10 sec. The final measurement is the
average of the last 5 intermediate measurements represented by the green line.
2 Final value
This value in black corresponds to the final value of the last measuring cycle. It is display in grey
during a measuring cycle and correspond to the intermediate measurements.
3 Reference signal
This value corresponds to the reference signal before the reagent injection.
Typical values are between 1000 and 3000. It may be lower in case of turbid sample.
4 Current signal
This value correspond to the current signal level during the measuring cycle, or to the final
signal value when finished.
6 Stop button
This button run the sampling pump for 10 seconds, for example to purge the circuit or the check
the sampling pump.
This button run the reagent pump for 10 seconds, for example to fill the reagent circuit during
the first installation.
This button display the LED signal. It should normally be between 1000 and 3000 on pure
water. The screen is shown below.
10 Test button
Signal screen:
4.1.4 Calibration
This window displays the 10 last calibrations done with the date, time and the new calibration
factor.
These two buttons allow to scroll up and down the 10 last calibration records.
3 Calibration factor
This field allows changing manually the calibration factor of the channel.
This factor is normally changed automatically while doing a calibration procedure by pressing on
the “ADJUST” button.
4 Offset
5 Calibration procedure
6 Linearization curve
As the Beer-Lambert law is not linear for high absorbance values, a linearization curve is
entered to compensate automatically this non-linearity. This button display the linearization
curve and enable to enter or check the linearization values.
The Y-axis corresponds to the rough measurements entered on the M1 to M10 fields while the
X-axis corresponds to the standard or final measurement, entered on the S1 to S10 fields.
The M1-M10 and S1-S10 values must be strictly increasing. It’s recommended to put 0.0 and
0.0 as starting values for S1 and M1. The final value is linearly extrapolated between these
points. Unused points at the end of the table must strictly remains at 0.0 both for S and M.
- Prepare a 0.1 g/l aluminium standard. For this weight 0.233 g of aluminum sulfate
octadecahydrate Al2(SO4)3.18H2O and dilute in 1litre of distilled water to obtain a mere solution
of 0.1g/l Al.
Then take 10 ml of this mere solution and complete to 1 litre of distilled water to get a 1 mg/l Al
standard.
- Do a manual measurement on this standard using the check screen (as describe above).
- Then go to the calibration screen and press on the “ADJUST” button. Validate the last
measured value and then enter the standard value on the keypad (1 mg/l).
The calibration is finished. A new calibration factor has been calculated and has been recorded
inside the calibration history displayed on the check screen.
This new calibration factor will be taken in account for all the further measurements.
Maintenance
The maintenance is limited to refilling of the two reagent bottles and the 5% H2SO4 cleaning
solution.
Troubleshooting
Symptoms Origin
Negative value - Very high turbidity on the sample that is not compensated
Value is too low - No water in the flow cell (clogging, bad inlet connection, fault in the
peristaltic pump)
Unstable value - Deposit or dirty on the flow cell, check the light level on the check
screen with pure water
Measuring errors
2 The light level is too - Check that there is no bubble inside the flow cell, if yes check
high the fittings
3 The light level is too - Sample concentration over range, check again with pure
low water
4.1.6 Specifications
Interferences
4.2.1 Principle
The measuring principle is based on the DPD colorimetric method (US-EPA 330.5). A small
quantity of reagent and buffer are injected on the sample. After a fixed reaction time, the
developed colour is measured by the light absorption according to the Beer-Lambert law:
[C ] = k ⋅ log I 0
I1
4.2.2 Settings
First screen:
1 Label
This field allows changing the channel label displayed with the measurement. The label size is
limited to 6 characters.
2 Unit
This field allows changing the channel unit displayed with the measurement. The unit size is
limited to 6 characters.
The high alarm level can be changed by pressing on this field. A default value is preset when
the channel is created. The high alarm level is primary used to fix the high limit represented by a
red line on the recorded measurement graph. The limit is used as the minimum graph scale.
The secondary use is to drive a high alarm relay if the relay button has been activated and if a
relay module has been installed on the selected module socket. In this case, the relay will be
activated as soon as the measurement overpass the high alarm value and will stay activated
until the measurement goes down below this value after subtracting the boundary value.
Refer to installation section for the relay wiring. Both functions NO (Normally Open) and NC
(Normally Closed) are available.
The low alarm level can be changed by pressing on this field. A default value is preset when the
channel is created. The low alarm level is primary used to fix the low limit represented by a
green line on the recorded measurement graph.
The secondary use is to drive a low alarm relay if the relay button has been activated and if a
relay module has been installed on the selected module socket. In this case, the relay will be
activated as soon as the measurement underpass the low alarm value and will stay activated
until the measurement goes down below this value after adding the boundary value.
Refer to installation section for the relay wiring. Both functions NO (Normally Open) and NC
(Normally Closed) are available.
5 Number of decimals
This field allows changing the number of decimal displayed for the measurement value to adjust
the display to significant decimals only.
Note: this choice does not affect the measurements stored in the internal memory or the
MODBUS registers.
Negatives values are normal for most of the measurements as they result from normal
fluctuations if the measurement is close to zero.
By default, the display of negative values is allowed as negative values may also result from a
wrong zeroing. Consequently, it may be useful to be aware of such problem when seeing
strongly negative values.
But as negative values have no theoretical meaning, it’s possible to replace them by zero by
selecting "NO" for this function.
Note that this function do not apply to the recorded measurements or to the MODBUS registers.
Second screen:
1 Scale factor
This factor is normally determined in factory to transform the basic measurement to a scaled
value in the final unit. It is has no reason to be changed except if optical elements are replaced.
2 Averaging
This field displays the number of elemental acquisitions averaged to form one measurement.
The default value is 50.
3 Sampling time
This field displays the sampling time specifically for this measurement.
A minimum time of 10 sec is recommended to allow a good renew of the sample.
It might be longer depending on the length and height of tubing to take the sample.
4 Reagent time
This field displays the reagent time injection. The default value is 1 sec.
Shorter time may produce instable measurements or limiting the measuring range.
Longer time are not necessary and generate a waste of reagent.
This field displays the colorimetric reaction time. The default value is 120 sec and this delay is in
accordance with the standard method.
Shorter times may reduce the measurement stability, especially for low temperatures.
6 Waste time
This field displays the purge time after the measurement. The default value is 5 sec.
The waste can be collected during this time if a relay module declared with the "waste" function
is installed.
7 Temperature coefficient
The colorimetric reaction might be slightly affected by the temperature. This coefficient is used
for the automatic temperature compensation.
Third screen:
1 Compensation
This button allows to compensate the sample turbidity. It's mandatory for turbid sample,
otherwise the measuring value should be underestimated or even negative.
2 LED selection
This field display the LED chosen for the measurement. Never change it, unless the colorimetric
method is changed.
4.2.3 Test
The check screen displays the last measurement performed, whatever manual or automatic,
with the reaction curve.
5 6 7 8 9 10
1 Reaction curve
This curve corresponds to the last reaction curve, or to the current reaction during a measuring
cycle. Intermediate measurements are performed every 10 sec. The final measurement is the
average of the last 5 intermediate measurements.
2 Final value
This value in black corresponds to the final value of the last measuring cycle. It is display in grey
during a measuring cycle and correspond to the intermediate measurements.
3 Reference signal
This value corresponds to the reference signal before the reagent injection.
Typical values are between 1000 and 3000. It may be lower in case of turbid sample.
4 Current signal
This value correspond to the current signal level during the measuring cycle, or to the final
signal value when finished.
6 Stop button
This button run the sampling pump for 10 seconds, for example to purge the circuit or the check
the sampling pump.
This button run the reagent pump for 10 seconds, for example to fill the reagent circuit during
the first installation.
This button display the LED signal. It should normally be between 1000 and 3000 on pure
water. The screen is shown below.
10 Test button
Signal screen:
4.2.4 Calibration
This window displays the 10 last calibrations done with the date, time and the new calibration
factor.
These two buttons allow to scroll up and down the 10 last calibration records.
3 Calibration factor
This field allows changing manually the calibration factor of the channel.
This factor is normally changed automatically while doing a calibration procedure by pressing on
the “ADJUST” button.
4 Offset
5 Calibration procedure
6 Linearization curve
As the Beer-Lambert law is not linear for high absorbance values, a linearization curve is
entered to compensate automatically this non-linearity. This button display the linearization
curve and enable to enter or check the linearization values.
The Y-axis corresponds to the rough measurements entered on the M1 to M10 fields while the
X-axis corresponds to the standard or final measurement, entered on the S1 to S10 fields.
The M1-M10 and S1-S10 values must be strictly increasing. It’s recommended to put 0.0 and
0.0 as starting values for S1 and M1. The final value is linearly extrapolated between these
points. Unused points at the end of the table must strictly remains at 0.0 both for S and M.
- Prepare a 1 mg/l free chlorine standard. For this weight 0.891 g of potassium permanganate
(KMnO4) and dilute in 1litre of distilled water to obtain a mere solution of 1g/l free Cl2.
Then take 1 ml of this mere solution and complete to 1 litre of distilled water.
- Do a manual measurement on this standard using the check screen (as describe above).
- Then go to the calibration screen and press on the “ADJUST” button. Validate the last
measured value and then enter the standard value on the keypad (1 mg/l).
The calibration is finished. A new calibration factor has been calculated and has been recorded
inside the calibration history displayed on the check screen.
This new calibration factor will be taken in account for all the further measurements.
Maintenance
The maintenance is limited to refilling of the two reagent bottles and the 5% H2SO4 cleaning
solution.
Troubleshooting
Symptoms Origin
Negative value - Very high turbidity on the sample that is not compensated
Value is too low - No water in the flow cell (clogging, bad inlet connection, fault in the
peristaltic pump)
Unstable value - Deposit or dirty on the flow cell, check the light level on the check
screen with pure water
Measuring errors
2 The light level is too - Check that there is no bubble inside the flow cell, if yes check
high the fittings
3 The light level is too - Sample concentration over range, check again with pure
low water
4.2.6 Specifications
Interferences
4.3 Chlorophyll A
4.3.1 Principle
The measuring principle is based on the red light fluorescence that is a typical characteristic of
the chlorophyll A
The light is generated by a pulsed xenon lamp to reach an ultra long life time.
The excitation filter limits the bandwidth of the excitation light, while the emission filter selects
the specific fluorescent light.
A second detector (optional) is used to compensate the eventual absorbance of the sample due
to turbidity.
4.3.2 Settings
First screen:
1 Label
This field allows changing the channel label displayed with the measurement. The label size is
limited to 6 characters.
2 Unit
This field allows changing the channel unit displayed with the measurement. The unit size is
limited to 6 characters.
The high alarm level can be changed by pressing on this field. A default value is preset when
the channel is created. The high alarm level is primary used to fix the high limit represented by a
red line on the recorded measurement graph. The limit is used as the minimum graph scale.
The secondary use is to drive a high alarm relay if the relay button has been activated and if a
relay module has been installed on the selected module socket. In this case, the relay will be
activated as soon as the measurement overpass the high alarm value and will stay activated
until the measurement goes down below this value after subtracting the boundary value.
Refer to installation section for the relay wiring. Both functions NO (Normally Open) and NC
(Normally Closed) are available.
The low alarm level can be changed by pressing on this field. A default value is preset when the
channel is created. The low alarm level is primary used to fix the low limit represented by a
green line on the recorded measurement graph.
The secondary use is to drive a low alarm relay if the relay button has been activated and if a
relay module has been installed on the selected module socket. In this case, the relay will be
activated as soon as the measurement underpass the low alarm value and will stay activated
until the measurement goes down below this value after adding the boundary value.
Refer to installation section for the relay wiring. Both functions NO (Normally Open) and NC
(Normally Closed) are available.
5 Number of decimals
This field allows changing the number of decimal displayed for the measurement value to adjust
the display to significant decimals only.
Note: this choice does not affect the measurements stored in the internal memory or the
MODBUS registers.
Negatives values are normal for most of the measurements as they result from normal
fluctuations if the measurement is close to zero.
By default, the display of negative values is allowed as negative values may also result from a
wrong zeroing. Consequently, it may be useful to be aware of such problem when seeing
strongly negative values.
But as negative values have no theoretical meaning, it’s possible to replace them by zero by
selecting "NO" for this function.
Note that this function do not apply to the recorded measurements or to the MODBUS registers.
Second screen:
1 Scale factor
This factor is normally determined in factory to transform the basic measurement to a scaled
value in the final unit. It is has no reason to be changed except if optical elements are replaced.
2 Averaging
This field displays the number of elemental acquisitions averaged to form one measurement.
The default value is 200.
3 Temperature coefficient
The fluorescence is slightly affected by the temperature. This coefficient is used for the
automatic temperature compensation.
4 Temperature adjustment
The value act as an offset on the temperature measurement to adjust the temperature. This
adjustment is not critical because the eventual error is taken in account during the calibration.
4.3.3 Test
The check screen displays the measurement on the sample that is inside the flow cell. Press on
the play key to update the measurement.
2 3 4 5 6 7
1 Measured value
This button run the sampling pump for 10 seconds to renew the sample.
This button injects the cleaning solution. It can be used to check the zero.
4 Stop button
This button display the lamp signal. It should normally be below 300 on pure water. The screen
is shown below.
6 Zero button
Put first pure water inside the flow cell before pressing this button.
7 Start measurement
Signal screen:
4.3.4 Calibration
This window displays the 10 last calibrations done with the date, time and the new calibration
factor.
These two buttons allow to scroll up and down the 10 last calibration records.
3 Calibration factor
This field allows changing manually the calibration factor of the channel.
This factor is normally changed automatically while doing a calibration procedure by pressing on
the “ADJUST” button.
4 Offset
5 Calibration procedure
6 Linearization curve
As the Beer-Lambert law is not linear for high absorbance values, a linearization curve is
entered to compensate automatically this non-linearity. This button display the linearization
curve and enable to enter or check the linearization values.
The Y-axis corresponds to the rough measurements entered on the M1 to M10 fields while the
X-axis corresponds to the standard or final measurement, entered on the S1 to S10 fields.
The M1-M10 and S1-S10 values must be strictly increasing. It’s recommended to put 0.0 and
0.0 as starting values for S1 and M1. The final value is linearly extrapolated between these
points. Unused points at the end of the table must strictly remains at 0.0 both for S and M.
- Dilute it first in 10 ml of alcohol (70% or more), then complete to 1 litre of distilled water. This
standard solution has a short life time, use only the same day !
- Do a manual measurement on this standard using the check screen (as describe above).
- Go on the calibration screen and press on the “ADJUST” button of the check screen. Validate
the previous measurement and then enter the standard value on the keypad.
The calibration is finished. A new calibration factor has been calculated and has been recorded
inside the calibration history displayed on the check screen. This new calibration factor will be
taken in account for all the further measurements.
Maintenance
Troubleshooting
Symptoms Origin
Negative value - Bad zero or auto-zero: the cleaning solution has been prepared with
polluted water or fault in the acid cleaning system (no flow, pump
failure)
Value is too low - No water in the flow cell (clogging, bad inlet connection, fault in the
peristaltic pump)
Unstable value - Deposit or dirty on the flow cell, check the light level on the check
screen
Measuring errors
3 The light level is too - Check that the analyser door is closed to avoid the
high on the photomultiplier to receive ambient light
photomultiplier
- Check that there is no bubble inside the flow cell, if yes check
the fittings
4 The light level is too - Check that the analyser door is closed to avoid the photo
high on the reference detector to receive ambient light
photo detector
- Check that there is no bubble inside the flow cell, if yes check
the fittings
6 The light level is too - Very turbid sample, recheck with pure water
low on the reference
photo detector - Deposit or dirty on the flow cell
4.3.6 Specifications
Specifications
4.4 Colour
4.4.1 Principle
The measuring principle is based on the light absorption around 350 nm according to the Beer-
Lambert law:
[C ] = k ⋅ log I 0
I1
The UV light is generated by a pulsed xenon lamp to reach an ultra long life time.
The light detection is performed by two photodiode detectors through interferential filters; one
centred around 350 nm while the other is used as reference to compensate turbidity or deposit
inside the flow cell.
4.4.2 Settings
First screen:
1 Label
This field allows changing the channel label displayed with the measurement. The label size is
limited to 6 characters.
2 Unit
This field allows changing the channel unit displayed with the measurement. The unit size is
limited to 6 characters.
The high alarm level can be changed by pressing on this field. A default value is preset when
the channel is created. The high alarm level is primary used to fix the high limit represented by a
red line on the recorded measurement graph. The limit is used as the minimum graph scale.
The secondary use is to drive a high alarm relay if the relay button has been activated and if a
relay module has been installed on the selected module socket. In this case, the relay will be
activated as soon as the measurement overpass the high alarm value and will stay activated
until the measurement goes down below this value after subtracting the boundary value.
Refer to installation section for the relay wiring. Both functions NO (Normally Open) and NC
(Normally Closed) are available.
The low alarm level can be changed by pressing on this field. A default value is preset when the
channel is created. The low alarm level is primary used to fix the low limit represented by a
green line on the recorded measurement graph.
The secondary use is to drive a low alarm relay if the relay button has been activated and if a
relay module has been installed on the selected module socket. In this case, the relay will be
activated as soon as the measurement underpass the low alarm value and will stay activated
until the measurement goes down below this value after adding the boundary value.
Refer to installation section for the relay wiring. Both functions NO (Normally Open) and NC
(Normally Closed) are available.
5 Number of decimals
This field allows changing the number of decimal displayed for the measurement value to adjust
the display to significant decimals only.
Note: this choice does not affect the measurements stored in the internal memory or the
MODBUS registers.
Negatives values are normal for most of the measurements as they result from normal
fluctuations if the measurement is close to zero.
By default, the display of negative values is allowed as negative values may also result from a
wrong zeroing. Consequently, it may be useful to be aware of such problem when seeing
strongly negative values.
But as negative values have no theoretical meaning, it’s possible to replace them by zero by
selecting "NO" for this function.
Note that this function do not apply to the recorded measurements or to the MODBUS registers.
Second screen:
1 Scale factor
This factor is normally determined in factory to transform the basic measurement to a scaled
value in the final unit. It is has no reason to be changed except if optical elements are replaced.
2 Averaging
This field displays the number of elemental acquisitions averaged to form one measurement.
The default value is 200.
4.4.3 Test
The check screen displays the measurement that responds directly to what is inside the flow
cell.
2 3 4 5 6 7
1 Measured value
This button run the sampling pump for 10 seconds to renew the sample.
This button injects the cleaning solution. It can be used to check the zero.
4 Stop button
This button display the lamp signal. It should normally be between 1000 and 2500 on pure
water. The screen is shown below.
6 Zero button
Put first pure water inside the flow cell before pressing this button.
7 Start measurement
This button updates the measurement directly with what is inside the flow cell.
Signal screen:
4.4.4 Calibration
This window displays the 10 last calibrations done with the date, time and the new calibration
factor.
These two buttons allow to scroll up and down the 10 last calibration records.
3 Calibration factor
This field allows changing manually the calibration factor of the channel.
This factor is normally changed automatically while doing a calibration procedure by pressing on
the “ADJUST” button.
4 Offset
5 Calibration procedure
6 Linearization curve
As the Beer-Lambert law is not linear for high absorbance values, a linearization curve is
entered to compensate automatically this non-linearity. This button display the linearization
curve and enable to enter or check the linearization values.
The Y-axis corresponds to the rough measurements entered on the M1 to M10 fields while the
X-axis corresponds to the standard or final measurement, entered on the S1 to S10 fields.
The M1-M10 and S1-S10 values must be strictly increasing. It’s recommended to put 0.0 and
0.0 as starting values for S1 and M1. The final value is linearly extrapolated between these
points. Unused points at the end of the table must strictly remains at 0.0 both for S and M.
- Take a Pt-Co colour standard representative of the measuring range, but not lower than 100
Pt-Co
- Do a manual measurement on this standard using the check screen (as describe above).
- Go on the calibration screen and press on the “ADJUST” button of the check screen. Validate
the previous measurement and then enter the standard value on the keypad.
The calibration is finished. A new calibration factor has been calculated and has been recorded
inside the calibration history displayed on the check screen.
This new calibration factor will be taken in account for all the further measurements.
Maintenance
Troubleshooting
Symptoms Origin
Negative value - Bad zero or auto-zero: the cleaning solution has been prepared with
polluted water or fault in the acid cleaning system (no flow, pump
failure)
Value is too low - No water in the flow cell (clogging, bad inlet connection, fault in the
peristaltic pump)
Unstable value - Deposit or dirty on the flow cell, check the light level on the check
screen
Measuring errors
3 The light level is too - Check that there is no bubble inside the flow cell, if yes check
high on the peak the fittings
photo detector PH1
- Failure on the photo detector board or cable (repair or
replace)
4 The light level is too - Check that there is no bubble inside the flow cell, if yes check
high on the reference the fittings
photo detector PH2
- Failure on the photo detector board or cable (repair or
replace)
5 The light level is too - Over range or very high turbidity. Recheck on pure water
low on the peak
photo detector PH1 - Dirty on the flow cell windows. Check if the automatic
cleaning system is working properly
6 The light level is too - Over range or very high turbidity. Recheck on pure water
low on the reference
photo detector PH2 - Dirty on the flow cell windows. Check if the automatic
cleaning system is working properly
4.4.6 Specifications
Specifications
4.5 Conductivity
4.5.1 Principle
The conductivity is the reciprocal of resistance and is determined by the ratio current/voltage
according to Ohm’s law.
1 l amps
C= =
R l volt
d
The cell constant is defined by: k=
A
The conductivity of a solution is proportional to its ion concentration, as long as there is no ionic
interaction.
4.5.2 Settings
First screen:
1 Label
This field allows changing the channel label displayed with the measurement. The label size is
limited to 6 characters.
2 Unit
This field allows changing the channel unit displayed with the measurement. The unit size is
limited to 6 characters.
The high alarm level can be changed by pressing on this field. A default value is preset when
the channel is created. The high alarm level is primary used to fix the high limit represented by a
red line on the recorded measurement graph. The limit is used as the minimum graph scale.
The secondary use is to drive a high alarm relay if the relay button has been activated and if a
relay module has been installed on the selected module socket. In this case, the relay will be
activated as soon as the measurement overpass the high alarm value and will stay activated
until the measurement goes down below this value after subtracting the boundary value.
Refer to installation section for the relay wiring. Both functions NO (Normally Open) and NC
(Normally Closed) are available.
The low alarm level can be changed by pressing on this field. A default value is preset when the
channel is created. The low alarm level is primary used to fix the low limit represented by a
green line on the recorded measurement graph.
The secondary use is to drive a low alarm relay if the relay button has been activated and if a
relay module has been installed on the selected module socket. In this case, the relay will be
activated as soon as the measurement underpass the low alarm value and will stay activated
until the measurement goes down below this value after adding the boundary value.
Refer to installation section for the relay wiring. Both functions NO (Normally Open) and NC
(Normally Closed) are available.
5 Number of decimals
This field allows changing the number of decimal displayed for the measurement value to adjust
the display to significant decimals only.
Note: this choice does not affect the measurements stored in the internal memory or the
MODBUS registers.
6 Boundaries
The purpose of boundary is avoiding too frequent changes of state of the alarm relays what can
harm process devices like pumps or electric-valves. The boundary value is subtracted from the
high level value before leaving the high alarm state and reciprocally added to the low alarm level
before leaving the low alarm state. More details including a schematic are provided in the relay
section.
Negatives values are normal for most of the measurements as they result from normal
fluctuations if the measurement is close to zero.
By default, the display of negative values is allowed as negative values may also result from a
wrong zeroing. Consequently, it may be useful to be aware of such problem when seeing
strongly negative values.
But as negative values have no theoretical meaning, it’s possible to replace them by zero by
selecting "NO" for this function.
Note that this function do not apply to the recorded measurements or to the MODBUS registers.
Second screen:
1 Scale factor
This factor is normally determined in factory to transform the basic measurement to a scaled
value in the final unit. It is has no reason to be changed except if elements are replaced.
2 Temperature coefficient
The conductivity of a solution is affected by the temperature. This coefficient is used for the
temperature compensation. The default value is 2% per °C. It may be adjusted depending of the
solution.
3 Temperature adjustment
The temperature of the sample given by the internal temperature probe can be adjusted by this
field if it differs from the real temperature read on a accurately calibrated thermometer.
This adjustment is not really important as a difference of temperature is taken in account during
the calibration. The adjustment is supposed to stay within +/- 4 °C.
4.5.3 Test
2
4 3
1 Read value
2 Probe temperature
This value is the sample temperature given by the probe. If the probe temperature sensor is not
connected, this value remains fixed at 25.0 °C.
3 Zero button
Put first pure water on the electrode and wait for stabilization (few minutes) before
pressing this button.
4 Signal test
This button is used to display the two analogue-digital converters of the conductivity module for
testing purpose only and is reserved to the manufacturer.
4.5.4 Calibration
This window displays the 10 last calibrations done with the date, time and the new calibration
factor.
These two buttons allow to scroll up and down the 10 last calibration records.
3 Calibration factor
This field allows changing manually the calibration factor of the channel.
This factor is normally changed automatically while doing a calibration procedure by pressing on
the “ADJUST” button.
4 Offset
5 Calibration procedure
6 Linearization curve
For more accuracy, a linearization curve can be entered. This button display the linearization
graph and enable to enter or check the linearization values.
The Y-axis corresponds to the rough measurements entered on the M1 to M10 fields while the
X-axis corresponds to the standard or final measurement, entered on the S1 to S10 fields.
The M1-M10 and S1-S10 values must be strictly increasing. It’s highly recommended to put 0.0
and 0.0 as starting values for S1 and M1. The final value is linearly extrapolated between these
points. Unused points at the end of the table must strictly remains at 0.0 both for S and M.
The conductivity electrode must be checked on a regular base depending on the condition of
use, and recalibrated if necessary.
2. Do manual measurements using the check screen, wait for the stabilisation of the
conductivity value (a few minutes if the electrode is new or dry), then press on the ZERO key.
3. Then put the electrode in a standard solution for conductivity (example 210 µS for 100 mg/l
NaCl at 25°C)
4. Do manual measurement on this solution using the check screen, wait for the stabilisation of
the conductivity
5. Go on the calibration screen, press on the ADJUST key and enter the standard value on the
keypad.
The calibration is finished. A new offset and calibration factor has been calculated and has
been recorded on the calibration history displayed on the calibration screen.
These new offset and factor will be taken in account for all the further measurements.
Maintenance
The conductivity probe must be cleaned on a regular base depending of the application (daily,
weekly or monthly).
Troubleshooting
Symptoms Origin
Unstable value - Electrical interference on the probe cable, put the probe cable away
from power cables
Measuring errors
4.6.1 Principle
In presence of oxygen, the fluorescence decreases that decreases the signal on the red
detector.
The probe has a RS485 interface under Modbus protocol and must be connected to the
dedicated RS485 port (J38) on the DSP400TFT board (for DSP400TFT board release E and
above).
4.6.2 Settings
First screen:
1 Label
This field allows changing the channel label displayed with the measurement. The label size is
limited to 6 characters.
2 Unit
This field allows changing the channel unit displayed with the measurement. The unit size is
limited to 6 characters.
The high alarm level can be changed by pressing on this field. A default value is preset when
the channel is created. The high alarm level is primary used to fix the high limit represented by a
red line on the recorded measurement graph. The limit is used as the minimum graph scale.
The secondary use is to drive a high alarm relay if the relay button has been activated and if a
relay module has been installed on the selected module socket. In this case, the relay will be
activated as soon as the measurement overpass the high alarm value and will stay activated
until the measurement goes down below this value after subtracting the boundary value.
Refer to installation section for the relay wiring. Both functions NO (Normally Open) and NC
(Normally Closed) are available.
The low alarm level can be changed by pressing on this field. A default value is preset when the
channel is created. The low alarm level is primary used to fix the low limit represented by a
green line on the recorded measurement graph.
The secondary use is to drive a low alarm relay if the relay button has been activated and if a
relay module has been installed on the selected module socket. In this case, the relay will be
activated as soon as the measurement underpass the low alarm value and will stay activated
until the measurement goes down below this value after adding the boundary value.
Refer to installation section for the relay wiring. Both functions NO (Normally Open) and NC
(Normally Closed) are available.
5 Number of decimals
This field allows changing the number of decimal displayed for the measurement value to adjust
the display to significant decimals only.
Note: this choice does not affect the measurements stored in the internal memory or the
MODBUS registers.
6 Boundaries
The purpose of boundary is avoiding too frequent changes of state of the alarm relays what can
harm process devices like pumps or electric-valves. The boundary value is subtracted from the
high level value before leaving the high alarm state and reciprocally added to the low alarm level
before leaving the low alarm state. More details including a schematic are provided in the relay
section.
7 Modbus address
This field is the Modbus address of the probe that has been fixed inside the probe during the
initialisation sequence.
This value must not be changed, unless the RS485 scan screen indicates that the probe has not
the proper address.
4.6.3 Test
1 Read value
2 Probe temperature
This value is the sample temperature given by the probe. It is used mainly for the calibration on
air.
4.6.4 Calibration
3 4 5
This window displays the 10 last calibrations done with the date, time and the new calibration
factor.
These two buttons allow to scroll up and down the 10 last calibration records.
3 Zero
This field allows to recalibrate the probe on zero solution. See after on the recommendations
how to prepare a zero solution.
4 Calibration procedure
The DO probe is factory calibrated but the full scale may be recalibrated if necessary (for
example once a year).
Before doing the calibration, allow the probe values to stabilize on air, both for the oxygen
concentration and the temperature by checking the values on the check screen.
The DO probe is factory calibrated and is normally stable for long period of time (one year).
It may be recalibrated if necessary.
Zero calibration
A zero-oxygen solution can be prepared by dissolving 10 g of sodium sulfite into 300 ml of pure
water and eventually adding a shake of cobalt chloride that accelerates the reaction. Allow a few
minutes of reaction time before using this solution.
- Do manual measurements using the check screen (as describe above), wait that the value
stabilizes (it may takes a few minutes). Then go to the calibration screen (flask icon) and press
on the ZERO key.
- Do manual measurements using the check screen (as describe above), wait that both the
value and the temperature stabilizes (it may takes a few minutes).
- Then go to the calibration screen (flask icon) and press on the 100% key.
Maintenance
The DO probe must be cleaned on a regular base depending of the application (daily, weekly or
monthly). It may be cleaned automatically by sending pressurised air.
Troubleshooting
Symptoms Origin
Measuring errors
4.7.1 Principle
The measuring principle is based on the UV light fluorescence that is a typical characteristic of
the mono or poly aromatics hydrocarbons (PAH)
The UV light is generated by a pulsed xenon lamp to reach an ultra long life time.
The excitation filter limits the bandwidth of the excitation light, while the emission filter selects
the specific fluorescent light.
A second detector (optional) is used to compensate the eventual absorbance of the sample due
to organic matter or suspended solids (turbidity).
4.7.2 Settings
First screen:
1 Label
This field allows changing the channel label displayed with the measurement. The label size is
limited to 6 characters.
2 Unit
This field allows changing the channel unit displayed with the measurement. The unit size is
limited to 6 characters.
The high alarm level can be changed by pressing on this field. A default value is preset when
the channel is created. The high alarm level is primary used to fix the high limit represented by a
red line on the recorded measurement graph. The limit is used as the minimum graph scale.
The secondary use is to drive a high alarm relay if the relay button has been activated and if a
relay module has been installed on the selected module socket. In this case, the relay will be
activated as soon as the measurement overpass the high alarm value and will stay activated
until the measurement goes down below this value after subtracting the boundary value.
Refer to installation section for the relay wiring. Both functions NO (Normally Open) and NC
(Normally Closed) are available.
The low alarm level can be changed by pressing on this field. A default value is preset when the
channel is created. The low alarm level is primary used to fix the low limit represented by a
green line on the recorded measurement graph.
The secondary use is to drive a low alarm relay if the relay button has been activated and if a
relay module has been installed on the selected module socket. In this case, the relay will be
activated as soon as the measurement underpass the low alarm value and will stay activated
until the measurement goes down below this value after adding the boundary value.
Refer to installation section for the relay wiring. Both functions NO (Normally Open) and NC
(Normally Closed) are available.
5 Number of decimals
This field allows changing the number of decimal displayed for the measurement value to adjust
the display to significant decimals only.
Note: this choice does not affect the measurements stored in the internal memory or the
MODBUS registers.
6 Display mode
This button allows selecting the concentration display mode as PAH or as OIL.
The measurements previously stored in the internal memory keep their original values.
Negatives values are normal for most of the measurements as they result from normal
fluctuations if the measurement is close to zero.
By default, the display of negative values is allowed as negative values may also result from a
wrong zeroing. Consequently, it may be useful to be aware of such problem when seeing
strongly negative values.
But as negative values have no theoretical meaning, it’s possible to replace them by zero by
selecting "NO" for this function.
Note that this function do not apply to the recorded measurements or to the MODBUS registers.
Second screen:
1 Scale factor
This factor is normally determined in factory to transform the basic measurement to a scaled
value in the final unit. It is has no reason to be changed except if optical elements are replaced.
2 Averaging
This field displays the number of elemental acquisitions averaged to form one measurement.
The default value is 200.
3 Temperature coefficient
The fluorescence is slightly affected by the temperature. This coefficient is used for the
automatic temperature compensation.
4 Temperature adjustment
The value act as an offset on the temperature measurement to adjust the temperature. This
adjustment is not critical because the eventual error is taken in account during the calibration.
4.7.3 Test
The check screen displays the measurement that corresponds directly to what is inside the flow
cell.
2 3 4 5 6 7
1 Measured value
This button run the sampling pump for 10 seconds to renew the sample
This button injects the cleaning solution. It can be used to check the zero.
4 Stop button
This button display the lamp signal. The PM (photomultiplier) value should normally be below
300 on pure water, while the PH (reference photodiode) should be between 1000 and 2500. The
screen is shown below.
6 Zero button
Put first pure water inside the flow cell before pressing this button.
7 Start measurement
This button update the measured value directly with what is inside the flow cell.
Signal screen:
4.7.4 Calibration
This window displays the 10 last calibrations done with the date, time and the new calibration
factor.
These two buttons allow to scroll up and down the 10 last calibration records.
3 Calibration factor
This field allows changing manually the calibration factor of the channel.
This factor is normally changed automatically while doing a calibration procedure by pressing on
the “ADJUST” button.
4 Offset
5 Calibration procedure
6 Linearization curve
As the Beer-Lambert law is not linear for high absorbance values, a linearization curve is
entered to compensate automatically this non-linearity. This button display the linearization
curve and enable to enter or check the linearization values.
The Y-axis corresponds to the rough measurements entered on the M1 to M10 fields while the
X-axis corresponds to the standard or final measurement, entered on the S1 to S10 fields.
The M1-M10 and S1-S10 values must be strictly increasing. It’s recommended to put 0.0 and
0.0 as starting values for S1 and M1. The final value is linearly extrapolated between these
points. Unused points at the end of the table must strictly remains at 0.0 both for S and M.
The analyser is factory calibrated on phenol. Depending on the aromatic ratio of the sample, a
recalibration based on a laboratory measurement is normally necessary.
- When the laboratory result is obtained, do a manual measurement on the second bottle using
the check screen describe above.
- Go on the calibration screen (flask icon) and press on the “ADJUST” button. Enter first the
measured value, then the value determined by the laboratory.
The calibration is finished. A new calibration factor has been calculated and has been recorded
inside the calibration history displayed on the check screen. This new calibration factor will be
taken in account for all the further measurements.
Maintenance
Troubleshooting
Symptoms Origin
Negative value - Bad zero or auto-zero: the cleaning solution has been prepared with
polluted water or fault in the acid cleaning system (no flow, pump
failure)
Value is too low - No water in the flow cell (clogging, bad inlet connection, fault in the
peristaltic pump)
Unstable value - Deposit or dirty on the flow cell, check the light level on the check
screen
Measuring errors
3 The light level is too - Check that the analyser door is closed to avoid the
high on the photomultiplier to receive ambient light
photomultiplier
- Check that there is no bubble inside the flow cell, if yes check
the fittings
replace)
4 The light level is too - Check that the analyser door is closed to avoid the photo
high on the reference detector to receive ambient light
photo detector
- Check that there is no bubble inside the flow cell, if yes check
the fittings
6 The light level is too - Very turbid sample, recheck with pure water
low on the reference
photo detector - Deposit or dirty on the flow cell
4.7.6 Specifications
Specifications
4.8 Iron
4.8.1 Principle
The measuring principle is based on the phenanthroline colorimetric method. A small quantity of
reagent and buffer are injected on the sample. After a fixed reaction time, the developed colour
is measured by the light absorption according to the Beer-Lambert law:
[C ] = k ⋅ log I 0
I1
4.8.2 Settings
First screen:
1 Label
This field allows changing the channel label displayed with the measurement. The label size is
limited to 6 characters.
2 Unit
This field allows changing the channel unit displayed with the measurement. The unit size is
limited to 6 characters.
The high alarm level can be changed by pressing on this field. A default value is preset when
the channel is created. The high alarm level is primary used to fix the high limit represented by a
red line on the recorded measurement graph. The limit is used as the minimum graph scale.
The secondary use is to drive a high alarm relay if the relay button has been activated and if a
relay module has been installed on the selected module socket. In this case, the relay will be
activated as soon as the measurement overpass the high alarm value and will stay activated
until the measurement goes down below this value after subtracting the boundary value.
Refer to installation section for the relay wiring. Both functions NO (Normally Open) and NC
(Normally Closed) are available.
The low alarm level can be changed by pressing on this field. A default value is preset when the
channel is created. The low alarm level is primary used to fix the low limit represented by a
green line on the recorded measurement graph.
The secondary use is to drive a low alarm relay if the relay button has been activated and if a
relay module has been installed on the selected module socket. In this case, the relay will be
activated as soon as the measurement underpass the low alarm value and will stay activated
until the measurement goes down below this value after adding the boundary value.
Refer to installation section for the relay wiring. Both functions NO (Normally Open) and NC
(Normally Closed) are available.
5 Number of decimals
This field allows changing the number of decimal displayed for the measurement value to adjust
the display to significant decimals only.
Note: this choice does not affect the measurements stored in the internal memory or the
MODBUS registers.
Negatives values are normal for most of the measurements as they result from normal
fluctuations if the measurement is close to zero.
By default, the display of negative values is allowed as negative values may also result from a
wrong zeroing. Consequently, it may be useful to be aware of such problem when seeing
strongly negative values.
But as negative values have no theoretical meaning, it’s possible to replace them by zero by
selecting "NO" for this function.
Note that this function do not apply to the recorded measurements or to the MODBUS registers.
Second screen:
1 Scale factor
This factor is normally determined in factory to transform the basic measurement to a scaled
value in the final unit. It is has no reason to be changed except if optical elements are replaced.
2 Averaging
This field displays the number of elemental acquisitions averaged to form one measurement.
The default value is 50.
3 Sampling time
This field displays the sampling time specifically for this measurement.
A minimum time of 10 sec is recommended to allow a good renew of the sample.
It might be longer depending on the length and height of tubing to take the sample.
4 Reagent time
This field displays the reagent time injection. The default value is 1 sec.
Shorter time may produce instable measurements or limiting the measuring range.
Longer time are not necessary and generate a waste of reagent.
This field displays the colorimetric reaction time. The default value is 120 sec and this delay is in
accordance with the standard method.
Shorter times may reduce the measurement stability, especially for low temperatures.
6 Waste time
This field displays the purge time after the measurement. The default value is 5 sec.
The waste can be collected during this time if a relay module declared with the "waste" function
is installed.
7 Temperature coefficient
The colorimetric reaction might be slightly affected by the temperature. This coefficient is used
for the automatic temperature compensation.
Third screen:
1 Compensation
This button allows to compensate the sample turbidity. It's mandatory for turbid sample,
otherwise the measuring value should be underestimated or even negative.
2 LED selection
This field display the LED chosen for the measurement. Never change it, unless the colorimetric
method is changed.
4.2.3 Test
The check screen displays the last measurement performed, whatever manual or automatic,
with the reaction curve.
5 6 7 8 9 10
1 Reaction curve
This curve corresponds to the last reaction curve, or to the current reaction during a measuring
cycle. Intermediate measurements are performed every 10 sec. The final measurement is the
average of the last 5 intermediate measurements.
2 Final value
This value in black corresponds to the final value of the last measuring cycle. It is display in grey
during a measuring cycle and correspond to the intermediate measurements.
3 Reference signal
This value corresponds to the reference signal before the reagent injection.
Typical values are between 1000 and 3000. It may be lower in case of turbid sample.
4 Current signal
This value correspond to the current signal level during the measuring cycle, or to the final
signal value when finished.
6 Stop button
This button run the sampling pump for 10 seconds, for example to purge the circuit or the check
the sampling pump.
This button run the reagent pump for 10 seconds, for example to fill the reagent circuit during
the first installation.
This button display the LED signal. It should normally be between 1000 and 3000 on pure
water. The screen is shown below.
10 Test button
Signal screen:
4.2.4 Calibration
This window displays the 10 last calibrations done with the date, time and the new calibration
factor.
These two buttons allow to scroll up and down the 10 last calibration records.
3 Calibration factor
This field allows changing manually the calibration factor of the channel.
This factor is normally changed automatically while doing a calibration procedure by pressing on
the “ADJUST” button.
4 Offset
5 Calibration procedure
6 Linearization curve
As the Beer-Lambert law is not linear for high absorbance values, a linearization curve is
entered to compensate automatically this non-linearity. This button display the linearization
curve and enable to enter or check the linearization values.
The Y-axis corresponds to the rough measurements entered on the M1 to M10 fields while the
X-axis corresponds to the standard or final measurement, entered on the S1 to S10 fields.
The M1-M10 and S1-S10 values must be strictly increasing. It’s recommended to put 0.0 and
0.0 as starting values for S1 and M1. The final value is linearly extrapolated between these
points. Unused points at the end of the table must strictly remains at 0.0 both for S and M.
For this add slowly 20 ml of concentrated sulphuric acid (H2SO4) in 50 ml of distilled water.
Weight 1.404 g of iron sulphate hex hydrate (NH4)2Fe(SO4)2·6H2O and dilute in this solution.
Then complete to 1litre of distilled water to obtain a mere solution of 0.2 g/l Fe.
Then take 1 ml of this mere solution and complete to 1 litre of distilled water.
- Do a manual measurement on this standard using the check screen (as describe above).
- Then go to the calibration screen and press on the “ADJUST” button. Validate the last
measured value and then enter the standard value on the keypad (0.2 mg/l).
The calibration is finished. A new calibration factor has been calculated and has been recorded
inside the calibration history displayed on the check screen.
This new calibration factor will be taken in account for all the further measurements.
Maintenance
The maintenance is limited to refilling of the two reagent bottles and the 5% H2SO4 cleaning
solution.
Troubleshooting
Symptoms Origin
Negative value - Very high turbidity on the sample that is not compensated
Value is too low - No water in the flow cell (clogging, bad inlet connection, fault in the
peristaltic pump)
pinched or clogged)
Unstable value - Deposit or dirty on the flow cell, check the light level on the check
screen with pure water
Measuring errors
2 The light level is too - Check that there is no bubble inside the flow cell, if yes check
high the fittings
3 The light level is too - Sample concentration over range, check again with pure
low water
4.8.6 Specifications
Interferences
Known as interfering: strong oxidizing agents, cyanide, nitrite, phosphates, chromium, zinc in
concentrations exceeding 10 times that of iron, cobalt and copper in excess of 5 mg/L, nickel
in excess of 2 mg/L.
4.9 Nitrate
4.9.1 Principle
The measuring principle is based on the UV light absorption around 215 nm according to the
Beer-Lambert law:
[C ] = k ⋅ log I 0
I1
The UV light is generated by a pulsed xenon lamp to reach an ultra long life time.
The light detection is performed by two photodiode detectors through interferential filters; one
centred around 215 nm while the other is used as reference to compensate turbidity, deposit
inside the flow cell and organic matter.
4.9.2 Settings
First screen:
1 Label
This field allows changing the channel label displayed with the measurement. The label size is
limited to 6 characters.
2 Unit
This field allows changing the channel unit displayed with the measurement. The unit size is
limited to 6 characters.
The high alarm level can be changed by pressing on this field. A default value is preset when
the channel is created. The high alarm level is primary used to fix the high limit represented by a
red line on the recorded measurement graph. The limit is used as the minimum graph scale.
The secondary use is to drive a high alarm relay if the relay button has been activated and if a
relay module has been installed on the selected module socket. In this case, the relay will be
activated as soon as the measurement overpass the high alarm value and will stay activated
until the measurement goes down below this value after subtracting the boundary value.
Refer to installation section for the relay wiring. Both functions NO (Normally Open) and NC
(Normally Closed) are available.
The low alarm level can be changed by pressing on this field. A default value is preset when the
channel is created. The low alarm level is primary used to fix the low limit represented by a
green line on the recorded measurement graph.
The secondary use is to drive a low alarm relay if the relay button has been activated and if a
relay module has been installed on the selected module socket. In this case, the relay will be
activated as soon as the measurement underpass the low alarm value and will stay activated
until the measurement goes down below this value after adding the boundary value.
Refer to installation section for the relay wiring. Both functions NO (Normally Open) and NC
(Normally Closed) are available.
5 Number of decimals
This field allows changing the number of decimal displayed for the measurement value to adjust
the display to significant decimals only.
Note: this choice does not affect the measurements stored in the internal memory or the
MODBUS registers.
6 Mode
This button allows changing the display mode: NO3 or N of NO3 according to the molar weight.
Negatives values are normal for most of the measurements as they result from normal
fluctuations if the measurement is close to zero.
By default, the display of negative values is allowed as negative values may also result from a
wrong zeroing. Consequently, it may be useful to be aware of such problem when seeing
strongly negative values.
But as negative values have no theoretical meaning, it’s possible to replace them by zero by
selecting "NO" for this function.
Note that this function do not apply to the recorded measurements or to the MODBUS registers.
Second screen:
1 Scale factor
This factor is normally determined in factory to transform the basic measurement to a scaled
value in the final unit. It is has no reason to be changed except if optical elements are replaced.
2 Averaging
This field displays the number of elemental acquisitions averaged to form one measurement.
The default value is 200.
4.9.3 Test
The check screen displays the measurement that corresponds directly to what is inside the flow
cell.
2 3 4 5 6 7
1 Measured value
This button run the sampling pump for 10 seconds to renew the sample.
This button injects the cleaning solution. It can be used to check the zero.
4 Stop button
This button display the lamp signal. The PH1 (peak photo detector) value should normally be
between 1000 and 2500 on pure water, PH2 (reference photodiode) should be between 1000
and 2500 also on pure water. PH1 decrease according to the nitrate concentration and the
turbidity of the sample. PH2 is affected only by the turbidity of the sample. The screen is shown
below.
6 Zero button
Put first pure water inside the flow cell before pressing this button.
7 Start measurement
This button update the measured value directly with what is inside the flow cell.
Signal screen:
4.9.4 Calibration
This window displays the 10 last calibrations done with the date, time and the new calibration
factor.
These two buttons allow to scroll up and down the 10 last calibration records.
3 Calibration factor
This field allows changing manually the calibration factor of the channel.
This factor is normally changed automatically while doing a calibration procedure by pressing on
the “ADJUST” button.
4 Offset
5 Calibration procedure
6 Linearization curve
As the Beer-Lambert law is not linear for high absorbance values, a linearization curve is
entered to compensate automatically this non-linearity. This button display the linearization
curve and enable to enter or check the linearization values.
The Y-axis corresponds to the rough measurements entered on the M1 to M10 fields while the
X-axis corresponds to the standard or final measurement, entered on the S1 to S10 fields.
The M1-M10 and S1-S10 values must be strictly increasing. It’s recommended to put 0.0 and
0.0 as starting values for S1 and M1. The final value is linearly extrapolated between these
points. Unused points at the end of the table must strictly remains at 0.0 both for S and M.
- Prepare a 50 mg/l NO3 standard. For this weight 1.629 g of dried KNO3 and dilute in 1litre of
distilled water to obtain a mere solution of 1g/l. Then take 50 ml of this mere solution and
complete to 1 litre of distilled water.
- Do a manual measurement on this standard using the check screen (as describe above).
- Go on the calibration screen and press on the “ADJUST” button of the check screen. Validate
the previous measurement and then enter the standard value on the keypad.
The calibration is finished. A new calibration factor has been calculated and has been recorded
inside the calibration history displayed on the check screen.
This new calibration factor will be taken in account for all the further measurements.
Maintenance
Troubleshooting
Symptoms Origin
Negative value - Bad zero or auto-zero: the cleaning solution has been prepared with
polluted water or fault in the acid cleaning system (no flow, pump
failure)
Value is too low - No water in the flow cell (clogging, bad inlet connection, fault in the
peristaltic pump)
Unstable value - Deposit or dirty on the flow cell, check the light level on the check
screen
Measuring errors
3 The light level is too - Check that there is no bubble inside the flow cell, if yes check
high on the peak the fittings
photo detector PH1
- Failure on the photo detector board or cable (repair or
replace)
4 The light level is too - Check that there is no bubble inside the flow cell, if yes check
high on the reference the fittings
photo detector PH2
- Failure on the photo detector board or cable (repair or
replace)
5 The light level is too - Over range or very high turbidity. Recheck on pure water
low on the peak
photo detector PH1 - Dirty on the flow cell windows. Check if the automatic
cleaning system is working properly
6 The light level is too - Over range or very high turbidity. Recheck on pure water
low on the reference
photo detector PH2 - Dirty on the flow cell windows. Check if the automatic
cleaning system is working properly
4.9.6 Specifications
Specifications
4.10.1 Principle
The measuring principle is based on the Azo dye (US-EPA 353.3) colorimetric method. A small
quantity of reagent and buffer are injected on the sample. After a fixed reaction time, the
developed colour is measured by the light absorption according to the Beer-Lambert law:
[C ] = k ⋅ log I 0
I1
4.10.2 Settings
First screen:
1 Label
This field allows changing the channel label displayed with the measurement. The label size is
limited to 6 characters.
2 Unit
This field allows changing the channel unit displayed with the measurement. The unit size is
limited to 6 characters.
The high alarm level can be changed by pressing on this field. A default value is preset when
the channel is created. The high alarm level is primary used to fix the high limit represented by a
red line on the recorded measurement graph. The limit is used as the minimum graph scale.
The secondary use is to drive a high alarm relay if the relay button has been activated and if a
relay module has been installed on the selected module socket. In this case, the relay will be
activated as soon as the measurement overpass the high alarm value and will stay activated
until the measurement goes down below this value after subtracting the boundary value.
Refer to installation section for the relay wiring. Both functions NO (Normally Open) and NC
(Normally Closed) are available.
The low alarm level can be changed by pressing on this field. A default value is preset when the
channel is created. The low alarm level is primary used to fix the low limit represented by a
green line on the recorded measurement graph.
The secondary use is to drive a low alarm relay if the relay button has been activated and if a
relay module has been installed on the selected module socket. In this case, the relay will be
activated as soon as the measurement underpass the low alarm value and will stay activated
until the measurement goes down below this value after adding the boundary value.
Refer to installation section for the relay wiring. Both functions NO (Normally Open) and NC
(Normally Closed) are available.
5 Number of decimals
This field allows changing the number of decimal displayed for the measurement value to adjust
the display to significant decimals only.
Note: this choice does not affect the measurements stored in the internal memory or the
MODBUS registers.
Negatives values are normal for most of the measurements as they result from normal
fluctuations if the measurement is close to zero.
By default, the display of negative values is allowed as negative values may also result from a
wrong zeroing. Consequently, it may be useful to be aware of such problem when seeing
strongly negative values.
But as negative values have no theoretical meaning, it’s possible to replace them by zero by
selecting "NO" for this function.
Note that this function do not apply to the recorded measurements or to the MODBUS registers.
Second screen:
1 Scale factor
This factor is normally determined in factory to transform the basic measurement to a scaled
value in the final unit. It is has no reason to be changed except if optical elements are replaced.
2 Averaging
This field displays the number of elemental acquisitions averaged to form one measurement.
The default value is 50.
3 Sampling time
This field displays the sampling time specifically for this measurement.
A minimum time of 10 sec is recommended to allow a good renew of the sample.
It might be longer depending on the length and height of tubing to take the sample.
4 Reagent time
This field displays the reagent time injection. The default value is 1 sec.
Shorter time may produce instable measurements or limiting the measuring range.
Longer time are not necessary and generate a waste of reagent.
This field displays the colorimetric reaction time. The default value is 120 sec and this delay is in
accordance with the standard method.
Shorter times may reduce the measurement stability, especially for low temperatures.
6 Waste time
This field displays the purge time after the measurement. The default value is 5 sec.
The waste can be collected during this time if a relay module declared with the "waste" function
is installed.
7 Temperature coefficient
The colorimetric reaction might be slightly affected by the temperature. This coefficient is used
for the automatic temperature compensation.
Third screen:
1 Compensation
This button allows to compensate the sample turbidity. It's mandatory for turbid sample,
otherwise the measuring value should be underestimated or even negative.
2 LED selection
This field display the LED chosen for the measurement. Never change it, unless the colorimetric
method is changed.
4.10.3 Test
The check screen displays the last measurement performed, whatever manual or automatic,
with the reaction curve.
5 6 7 8 9 10
1 Reaction curve
This curve corresponds to the last reaction curve, or to the current reaction during a measuring
cycle. Intermediate measurements are performed every 10 sec. The final measurement is the
average of the last 5 intermediate measurements.
2 Final value
This value in black corresponds to the final value of the last measuring cycle. It is display in grey
during a measuring cycle and correspond to the intermediate measurements.
3 Reference signal
This value corresponds to the reference signal before the reagent injection.
Typical values are between 1000 and 3000. It may be lower in case of turbid sample.
4 Current signal
This value correspond to the current signal level during the measuring cycle, or to the final
signal value when finished.
6 Stop button
This button run the sampling pump for 10 seconds, for example to purge the circuit or the check
the sampling pump.
This button run the reagent pump for 10 seconds, for example to fill the reagent circuit during
the first installation.
This button display the LED signal. It should normally be between 1000 and 3000 on pure
water. The screen is shown below.
10 Test button
Signal screen:
4.10.4 Calibration
This window displays the 10 last calibrations done with the date, time and the new calibration
factor.
These two buttons allow to scroll up and down the 10 last calibration records.
3 Calibration factor
This field allows changing manually the calibration factor of the channel.
This factor is normally changed automatically while doing a calibration procedure by pressing on
the “ADJUST” button.
4 Offset
5 Calibration procedure
6 Linearization curve
As the Beer-Lambert law is not linear for high absorbance values, a linearization curve is
entered to compensate automatically this non-linearity. This button display the linearization
curve and enable to enter or check the linearization values.
The Y-axis corresponds to the rough measurements entered on the M1 to M10 fields while the
X-axis corresponds to the standard or final measurement, entered on the S1 to S10 fields.
The M1-M10 and S1-S10 values must be strictly increasing. It’s recommended to put 0.0 and
0.0 as starting values for S1 and M1. The final value is linearly extrapolated between these
points. Unused points at the end of the table must strictly remains at 0.0 both for S and M.
For this weight 150 mg of sodium nitrite (NaNO2) and dilute in 1litre of distilled water to obtain a
mere solution of 0.1g/l NO2
Then take 10 ml of this mere solution and complete to 1 litre of distilled water to get a 1 mg/l
NO2 standard.
- Do a manual measurement on this standard using the check screen (as describe above).
- Then go to the calibration screen and press on the “ADJUST” button. Validate the last
measured value and then enter the standard value on the keypad (1 mg/l).
The calibration is finished. A new calibration factor has been calculated and has been recorded
inside the calibration history displayed on the check screen.
This new calibration factor will be taken in account for all the further measurements.
Maintenance
The maintenance is limited to refilling of the two reagent bottles and the 5% H2SO4 cleaning
solution.
Troubleshooting
Symptoms Origin
Negative value - Very high turbidity on the sample that is not compensated
Value is too low - No water in the flow cell (clogging, bad inlet connection, fault in the
peristaltic pump)
pinched or clogged)
Unstable value - Deposit or dirty on the flow cell, check the light level on the check
screen with pure water
Measuring errors
2 The light level is too - Check that there is no bubble inside the flow cell, if yes check
high the fittings
3 The light level is too - Sample concentration over range, check again with pure
low water
4.10.6 Specifications
Interferences
Known as interfering because of precipitation: Sb3+, Au3+, Bi3+, Fe3+, Pb2+, Hg2+, Ag+,
chloroplatinate (PtCl6), and metavanadate (VO3).
4.11.1 Principle
The reduction potential is a measure of the tendency of the solution to either gain or lose
electrons when it is subject to change by introduction of a new species.
A solution with a higher (more positive) reduction potential than the new species will have a
tendency to gain electrons from the new species (i.e. to be reduced by oxidizing the new
species) and a solution with a lower (more negative) reduction potential will have a tendency to
lose electrons to the new species (i.e. to be oxidized by reducing the new species).
The method of measurement is based on the potential between two half-cells (reference and
sensing) containing a conductor immersed in an appropriate electrolyte solution and ended by a
conductive glass membrane. These two half-cells are generally combined in a single body
electrode.
4.11.2 Settings
First screen:
1 Label
This field allows changing the channel label displayed with the measurement. The label size is
limited to 6 characters.
2 Unit
This field allows changing the channel unit displayed with the measurement. The unit size is
limited to 6 characters.
The high alarm level can be changed by pressing on this field. A default value is preset when
the channel is created. The high alarm level is primary used to fix the high limit represented by a
red line on the recorded measurement graph. The limit is used as the minimum graph scale.
The secondary use is to drive a high alarm relay if the relay button has been activated and if a
relay module has been installed on the selected module socket. In this case, the relay will be
activated as soon as the measurement overpass the high alarm value and will stay activated
until the measurement goes down below this value after subtracting the boundary value.
Refer to installation section for the relay wiring. Both functions NO (Normally Open) and NC
(Normally Closed) are available.
The low alarm level can be changed by pressing on this field. A default value is preset when the
channel is created. The low alarm level is primary used to fix the low limit represented by a
green line on the recorded measurement graph.
The secondary use is to drive a low alarm relay if the relay button has been activated and if a
relay module has been installed on the selected module socket. In this case, the relay will be
activated as soon as the measurement underpass the low alarm value and will stay activated
until the measurement goes down below this value after adding the boundary value.
Refer to installation section for the relay wiring. Both functions NO (Normally Open) and NC
(Normally Closed) are available.
5 Number of decimals
This field allows changing the number of decimal displayed for the measurement value to adjust
the display to significant decimals only.
Note: this choice does not affect the measurements stored in the internal memory or the
MODBUS registers.
Negatives values are normal for most of the measurements as they result from normal
fluctuations if the measurement is close to zero.
By default, the display of negative values is allowed as negative values may also result from a
wrong zeroing. Consequently, it may be useful to be aware of such problem when seeing
strongly negative values.
But as negative values have no theoretical meaning, it’s possible to replace them by zero by
selecting "NO" for this function.
Note that this function do not apply to the recorded measurements or to the MODBUS registers.
Second screen:
1 Scale factor
This factor is normally determined in factory to transform the basic measurement to a scaled
value in the final unit. It is has no reason to be changed except if elements are replaced.
2 Temperature adjustment
The temperature of the sample given by the internal temperature probe can be adjusted by this
field if it differs from the real temperature read on a accurately calibrated thermometer.
This adjustment is not really important as a difference of temperature is taken in account during
the calibration. The adjustment is supposed to stay within +/- 4 °C.
4.11.3 Test
1 Read value
2 Signal test
This button is used to display the two analogue-digital converters of the conductivity module for
testing purpose only and is reserved to the manufacturer.
4.11.4 Calibration
This window displays the 10 last calibrations done with the date, time and the new calibration
factor.
These two buttons allow to scroll up and down the 10 last calibration records.
3 Calibration factor
This field allows changing manually the calibration factor of the channel.
This factor is normally changed automatically while doing the second point calibration.
4 Offset
5 Calibration procedure
The ORP electrode must be checked on a regular base depending on the condition of use, and
recalibrated if necessary.
- Put the electrode in a zero buffer solution (or alternatively replace the electrode by a strap)
- Do manual measurement on this buffer using the check screen (as describe above), wait for
the stabilisation of the ORP value, then press on the ZERO key.
- Do a manual measurement on this standard using the check screen (as describe above) and
wait for the stabilisation of the ORP value
- Then go to the calibration screen and press on the “ADJUST” button. Validate the last
measured value and then enter the standard value on the keypad.
The calibration is finished. New offset and calibration factor has been calculated and has been
recorded in the calibration history. These new offset and calibration factor will be taken in
account for all the further measurements.
Maintenance
The pH probe must be cleaned on a regular base (generally weekly) depending of the
application. It must be replace about every 6 months (depending of the application).
Troubleshooting
Symptoms Origin
Unstable value - Electrical interference on the probe cable, put the probe cable away
from power cables
4.12 pH
4.12.1 Principle
The pH of a solution is the negative logarithm of the molar concentration of dissolved hydronium
ions (H3O+).
The method of measurement is based on the potential between two half-cells (reference and
sensing) containing a conductor immersed in an appropriate electrolyte solution and ended by a
conductive glass membrane. These two half-cells are generally combined in a single body
electrode.
The voltage between the half-cell follows the Nernst equation:
V = V0 + 2,3
RT [
H +m
⋅ log +
]
nF [H ref ]
With V0: voltage of the reference half-cell, R: ideal gas constant, T: absolute temperature, n
:valence, F: Faraday constant, H+m: concentration of H+ in the measured solution, H+ref:
reference concentration of H+.
At 25ºC, one pH unit corresponds to 59.16 mV. At pH 7, the output is 0V (for all temperatures).
4.12.2 Settings
First screen:
1 Label
This field allows changing the channel label displayed with the measurement. The label size is
limited to 6 characters.
2 Unit
This field allows changing the channel unit displayed with the measurement. The unit size is
limited to 6 characters.
The high alarm level can be changed by pressing on this field. A default value is preset when
the channel is created. The high alarm level is primary used to fix the high limit represented by a
red line on the recorded measurement graph. The limit is used as the minimum graph scale.
The secondary use is to drive a high alarm relay if the relay button has been activated and if a
relay module has been installed on the selected module socket. In this case, the relay will be
activated as soon as the measurement overpass the high alarm value and will stay activated
until the measurement goes down below this value after subtracting the boundary value.
Refer to installation section for the relay wiring. Both functions NO (Normally Open) and NC
(Normally Closed) are available.
The low alarm level can be changed by pressing on this field. A default value is preset when the
channel is created. The low alarm level is primary used to fix the low limit represented by a
green line on the recorded measurement graph.
The secondary use is to drive a low alarm relay if the relay button has been activated and if a
relay module has been installed on the selected module socket. In this case, the relay will be
activated as soon as the measurement underpass the low alarm value and will stay activated
until the measurement goes down below this value after adding the boundary value.
Refer to installation section for the relay wiring. Both functions NO (Normally Open) and NC
(Normally Closed) are available.
5 Number of decimals
This field allows changing the number of decimal displayed for the measurement value to adjust
the display to significant decimals only.
Note: this choice does not affect the measurements stored in the internal memory or the
MODBUS registers.
6 Boundaries
The purpose of boundary is avoiding too frequent changes of state of the alarm relays what can
harm process devices like pumps or electric-valves. The boundary value is subtracted from the
high level value before leaving the high alarm state and reciprocally added to the low alarm level
before leaving the low alarm state. More details including a schematic are provided in the relay
section.
Negatives values are normal for most of the measurements as they result from normal
fluctuations if the measurement is close to zero.
By default, the display of negative values is allowed as negative values may also result from a
wrong zeroing. Consequently, it may be useful to be aware of such problem when seeing
strongly negative values.
But as negative values have no theoretical meaning, it’s possible to replace them by zero by
selecting "NO" for this function.
Note that this function do not apply to the recorded measurements or to the MODBUS registers.
Second screen:
1 Scale factor
This factor is normally determined in factory to transform the basic measurement to a scaled
value in the final unit. It is has no reason to be changed except if elements are replaced.
2 Temperature adjustment
The temperature of the sample given by the internal temperature probe can be adjusted by this
field if it differs from the real temperature read on a accurately calibrated thermometer.
This adjustment is not really important as a difference of temperature is taken in account during
the calibration. The adjustment is supposed to stay within +/- 4 °C.
4.12.3 Test
1 Read value
2 Probe temperature
This value is the sample temperature given by the probe. If the probe temperature sensor is not
connected, this value remains fixed at 25.0 °C.
3 Signal test
This button is used to display the two analogue-digital converters of the conductivity module for
testing purpose only and is reserved to the manufacturer.
4.12.4 Calibration
This window displays the 10 last calibrations done with the date, time and the new calibration
factor.
These two buttons allow to scroll up and down the 10 last calibration records.
3 Calibration factor
This field allows changing manually the calibration factor of the channel.
This factor is normally changed automatically while doing the second point calibration.
4 Offset
5 pH 7 calibration point
The current pH value is displayed. After stabilisation, press on the key "SET TO 7.0".
After this, the displayed value must be 7.0.
6 pH 4 calibration point
The current pH value is displayed. After stabilisation, press on the key "SET TO 4.0".
After this, the displayed value must be 4.0.
The pH electrode must be checked on a regular base depending on the condition of use, and
recalibrated if necessary.
- Press on the "pH=7" key of the calibration screen. Wait for the stabilisation of the pH value,
then press on the "SET TO 7.0" key.
- Press on the "pH=7" key of the calibration screen. Wait for the stabilisation of the pH value,
then press on the "SET TO 7.0" key.
The calibration is finished. New offset and calibration factor has been calculated and has been
recorded in the calibration history. These new offset and calibration factor will be taken in
account for all the further measurements.
Maintenance
The pH probe must be cleaned on a regular base (generally weekly) depending of the
application. It must be replace about every 6 months (depending of the application).
Troubleshooting
Symptoms Origin
- pH probe disconnected
Unstable value - Electrical interference on the probe cable, put the probe cable away
from power cables
Measuring errors
6.12.6 pH specifications
Measuring range: 0 to 14
4.13 Phosphate
4.13.1 Principle
The measuring principle is based on the vanadomolybdophosphoric acid colorimetric method for
the high range or on the ascorbic acid colorimetric method for the low range.
A small quantity of reagent is injected on the sample. After a fixed reaction time, the developed
colour is measured by the light absorption according to the Beer-Lambert law:
[C ] = k ⋅ log I 0
I1
4.13.2 Settings
First screen:
1 Label
This field allows changing the channel label displayed with the measurement. The label size is
limited to 6 characters.
2 Unit
This field allows changing the channel unit displayed with the measurement. The unit size is
limited to 6 characters.
The high alarm level can be changed by pressing on this field. A default value is preset when
the channel is created. The high alarm level is primary used to fix the high limit represented by a
red line on the recorded measurement graph. The limit is used as the minimum graph scale.
The secondary use is to drive a high alarm relay if the relay button has been activated and if a
relay module has been installed on the selected module socket. In this case, the relay will be
activated as soon as the measurement overpass the high alarm value and will stay activated
until the measurement goes down below this value after subtracting the boundary value.
Refer to installation section for the relay wiring. Both functions NO (Normally Open) and NC
(Normally Closed) are available.
The low alarm level can be changed by pressing on this field. A default value is preset when the
channel is created. The low alarm level is primary used to fix the low limit represented by a
green line on the recorded measurement graph.
The secondary use is to drive a low alarm relay if the relay button has been activated and if a
relay module has been installed on the selected module socket. In this case, the relay will be
activated as soon as the measurement underpass the low alarm value and will stay activated
until the measurement goes down below this value after adding the boundary value.
Refer to installation section for the relay wiring. Both functions NO (Normally Open) and NC
(Normally Closed) are available.
5 Number of decimals
This field allows changing the number of decimal displayed for the measurement value to adjust
the display to significant decimals only.
Note: this choice does not affect the measurements stored in the internal memory or the
MODBUS registers.
Negatives values are normal for most of the measurements as they result from normal
fluctuations if the measurement is close to zero.
By default, the display of negative values is allowed as negative values may also result from a
wrong zeroing. Consequently, it may be useful to be aware of such problem when seeing
strongly negative values.
But as negative values have no theoretical meaning, it’s possible to replace them by zero by
selecting "NO" for this function.
Note that this function do not apply to the recorded measurements or to the MODBUS registers.
Second screen:
1 Scale factor
This factor is normally determined in factory to transform the basic measurement to a scaled
value in the final unit. It is has no reason to be changed except if optical elements are replaced.
2 Averaging
This field displays the number of elemental acquisitions averaged to form one measurement.
The default value is 50.
3 Sampling time
This field displays the sampling time specifically for this measurement.
A minimum time of 10 sec is recommended to allow a good renew of the sample.
It might be longer depending on the length and height of tubing to take the sample.
4 Reagent time
This field displays the reagent time injection. The default value is 1 sec.
Shorter time may produce instable measurements or limiting the measuring range.
Longer time are not necessary and generate a waste of reagent.
This field displays the colorimetric reaction time. The default value is 120 sec and this delay is in
accordance with the standard method.
Shorter times may reduce the measurement stability, especially for low temperatures.
6 Waste time
This field displays the purge time after the measurement. The default value is 5 sec.
The waste can be collected during this time if a relay module declared with the "waste" function
is installed.
7 Temperature coefficient
The colorimetric reaction might be slightly affected by the temperature. This coefficient is used
for the automatic temperature compensation.
Third screen:
1 Compensation
This button allows to compensate the sample turbidity. It's mandatory for turbid sample,
otherwise the measuring value should be underestimated or even negative.
2 LED selection
This field display the LED chosen for the measurement. Never change it, unless the colorimetric
method is changed.
4.13.3 Test
The check screen displays the last measurement performed, whatever manual or automatic,
with the reaction curve.
5 6 7 8 9 10
1 Reaction curve
This curve corresponds to the last reaction curve, or to the current reaction during a measuring
cycle. Intermediate measurements are performed every 10 sec. The final measurement is the
average of the last 5 intermediate measurements.
2 Final value
This value in black corresponds to the final value of the last measuring cycle. It is display in grey
during a measuring cycle and correspond to the intermediate measurements.
3 Reference signal
This value corresponds to the reference signal before the reagent injection.
Typical values are between 1000 and 3000. It may be lower in case of turbid sample.
4 Current signal
This value correspond to the current signal level during the measuring cycle, or to the final
signal value when finished.
6 Stop button
This button run the sampling pump for 10 seconds, for example to purge the circuit or the check
the sampling pump.
This button run the reagent pump for 10 seconds, for example to fill the reagent circuit during
the first installation.
This button display the LED signal. It should normally be between 1000 and 3000 on pure
water. The screen is shown below.
10 Test button
Signal screen:
4.13.4 Calibration
This window displays the 10 last calibrations done with the date, time and the new calibration
factor.
These two buttons allow to scroll up and down the 10 last calibration records.
3 Calibration factor
This field allows changing manually the calibration factor of the channel.
This factor is normally changed automatically while doing a calibration procedure by pressing on
the “ADJUST” button.
4 Offset
5 Calibration procedure
6 Linearization curve
As the Beer-Lambert law is not linear for high absorbance values, a linearization curve is
entered to compensate automatically this non-linearity. This button display the linearization
curve and enable to enter or check the linearization values.
The Y-axis corresponds to the rough measurements entered on the M1 to M10 fields while the
X-axis corresponds to the standard or final measurement, entered on the S1 to S10 fields.
The M1-M10 and S1-S10 values must be strictly increasing. It’s recommended to put 0.0 and
0.0 as starting values for S1 and M1. The final value is linearly extrapolated between these
points. Unused points at the end of the table must strictly remains at 0.0 both for S and M.
- Prepare a 5 mg/l P-PO4 standard. For this weight 2.195 g of anhydrous monopotassium
phosphate (KH2PO4) and dilute in 1litre of distilled water to obtain a mere solution of 500 mg/l P-
PO4.
Then take 10 ml of this mere solution and complete to 1 litre of distilled water.
- Do a manual measurement on this standard using the check screen (as describe above).
- Then go to the calibration screen and press on the “ADJUST” button. Validate the last
measured value and then enter the standard value on the keypad (1 mg/l).
The calibration is finished. A new calibration factor has been calculated and has been recorded
inside the calibration history displayed on the check screen.
This new calibration factor will be taken in account for all the further measurements.
Maintenance
The maintenance is limited to refilling of the two reagent bottles and the 5% H2SO4 cleaning
solution.
Troubleshooting
Symptoms Origin
Negative value - Very high turbidity on the sample that is not compensated
Value is too low - No water in the flow cell (clogging, bad inlet connection, fault in the
peristaltic pump)
Unstable value - Deposit or dirty on the flow cell, check the light level on the check
screen with pure water
Measuring errors
2 The light level is too - Check that there is no bubble inside the flow cell, if yes check
high the fittings
3 The light level is too - Sample concentration over range, check again with pure
low water
4.13.6 Specifications
Al3+, Fe3+, Mg2+, Ca2+, Ba2+, Sr2+, Li+, Na+, K+, NH4+, Cd2+, Mn2+,Pb2+, Hg+, Hg2+,
Sn2+, Cu2+, Ni2+, Ag+, U4+, Zr4+, AsO3–, Br–, CO32–, ClO4–, CN–, IO3–,SiO44–, NO3–,
NO2–, SO42–, O32–, pyrophosphate, molybdate, tetraborate, selenate, benzoate, citrate,
oxalate, lactate, tartrate, formate, and salicylate.
Sulfide (Na2S) and silicate do not interfere at concentrations of 1.0 and 10 mg/L.
4.14 Temperature
4.14.1 Principle
The table below shows the Pt 100 resistance in ohm for temperature between 0°C and 100°C.
°C 0 1 2 3 4 5 6 7 8 9
0 100.00 100.39 100.78 101.17 101.56 101.95 102.34 102.73 103.12 103.51
10 103.90 104.29 104.68 105.07 105.46 105.85 106.24 106.63 107.02 107.40
20 107.79 108.18 108.57 108.96 109.35 109.73 110.12 110.51 110.90 111.28
30 111.67 112.06 112.45 112.83 113.22 113.61 113.99 114.38 114.77 115.15
40 115.54 115.93 116.31 116.70 117.08 117.47 117.85 118.24 118.62 119.01
50 119.40 119.78 120.16 120.55 120.93 121.32 121.70 122.09 122.47 122.86
60 123.24 123.62 124.01 124.39 124.77 125.16 125.54 125.92 126.31 126.69
70 127.07 127.45 127.84 128.22 128.60 128.98 129.37 129.75 130.13 130.51
80 130.89 131.27 131.66 132.04 132.42 132.80 133.18 133.56 133.94 134.32
90 134.70 135.08 135.46 135.84 136.22 136.60 136.98 137.36 137.74 138.12
4.14.2 Settings
First screen:
1 Label
This field allows changing the channel label displayed with the measurement. The label size is
limited to 6 characters.
2 Unit
This field allows changing the channel unit displayed with the measurement. The unit size is
limited to 6 characters.
The high alarm level can be changed by pressing on this field. A default value is preset when
the channel is created. The high alarm level is primary used to fix the high limit represented by a
red line on the recorded measurement graph. The limit is used as the minimum graph scale.
The secondary use is to drive a high alarm relay if the relay button has been activated and if a
relay module has been installed on the selected module socket. In this case, the relay will be
activated as soon as the measurement overpass the high alarm value and will stay activated
until the measurement goes down below this value after subtracting the boundary value.
Refer to installation section for the relay wiring. Both functions NO (Normally Open) and NC
(Normally Closed) are available.
The low alarm level can be changed by pressing on this field. A default value is preset when the
channel is created. The low alarm level is primary used to fix the low limit represented by a
green line on the recorded measurement graph.
The secondary use is to drive a low alarm relay if the relay button has been activated and if a
relay module has been installed on the selected module socket. In this case, the relay will be
activated as soon as the measurement underpass the low alarm value and will stay activated
until the measurement goes down below this value after adding the boundary value.
Refer to installation section for the relay wiring. Both functions NO (Normally Open) and NC
(Normally Closed) are available.
5 Number of decimals
This field allows changing the number of decimal displayed for the measurement value to adjust
the display to significant decimals only.
Note: this choice does not affect the measurements stored in the internal memory or the
MODBUS registers.
6 Boundaries
The purpose of boundary is avoiding too frequent changes of state of the alarm relays what can
harm process devices like pumps or electric-valves. The boundary value is subtracted from the
high level value before leaving the high alarm state and reciprocally added to the low alarm level
before leaving the low alarm state. More details including a schematic are provided in the relay
section.
Negatives values are normal for most of the measurements as they result from normal
fluctuations if the measurement is close to zero.
By default, the display of negative values is allowed as negative values may also result from a
wrong zeroing. Consequently, it may be useful to be aware of such problem when seeing
strongly negative values.
But as negative values have no theoretical meaning, it’s possible to replace them by zero by
selecting "NO" for this function.
Note that this function do not apply to the recorded measurements or to the MODBUS registers.
Second screen:
1 Scale factor
This factor is normally determined in factory to transform the basic measurement to a scaled
value in the final unit. It is has no reason to be changed except if elements are replaced.
2 Temperature adjustment
The temperature of the sample given by the internal temperature probe can be adjusted by this
field if it differs from the real temperature read on a accurately calibrated thermometer.
4.14.3 Test
3 2
1 Read value
2 Zero button
Put the electrode in a Dewar filled with ice and water and wait for stabilization before
pressing this button.
3 Signal test
This button is used to display the two analogue-digital converters of the conductivity module for
testing purpose only and is reserved to the manufacturer.
4.14.4 Calibration
This window displays the 10 last calibrations done with the date, time and the new calibration
factor.
These two buttons allow to scroll up and down the 10 last calibration records.
3 Calibration factor
This field allows changing manually the calibration factor of the channel.
This factor is normally changed automatically while doing a calibration procedure by pressing on
the “ADJUST” button.
4 Offset
5 Calibration procedure
6 Linearization curve
For more accuracy, a linearization curve can be entered. This button display the linearization
graph and enable to enter or check the linearization values.
The temperature probe is factory calibrated with an accuracy of +/- 1°C. It can be recalibrated if
a higher accuracy is requested.
- Do manual measurement using the check screen (as describe above), wait for the stabilisation
of the temperature value, and then press on the ZERO key.
- Put the temperature probe in a Dewar filled of water at a temperature between 50°C and 80°C
with a reference thermometer inside
- Do manual measurement using the check screen, wait for the stabilisation of the temperature
value
- Go to the calibration screen, then press on the ADJUST key. Validate the last measurement
then enter the value given by the reference thermometer on the keypad.
The calibration is finished. New offset and calibration factors has been calculated and has been
recorded inside the zero and calibration history displayed on the check screen.
These new calibration factors will be taken in account for all the further measurements.
Maintenance
No maintenance is required.
Troubleshooting
Symptoms Origin
Measuring errors
2 Probe short-circuited - Check the probe with an ohmmeter, the value must be in a
100 to 120 ohm range, if not replace the probe
4.14.6 Specifications
Specifications
Calibrated range: 0 - 50 °C
Accuracy: +/- 1 °C
4.15.1 Principle
The measuring principle is based on the infra-red light absorbance at 880 nm over a defined
optical path.
Two models are available, both are designed for waste water:
The detection system is placed at the end of an immersed probe in a robust PVC case.
An automatic cleaning of the windows can be done by sending compressed air on a regular
base.
1 LED source
2 Photo detector
4.15.2 Settings
First screen:
1 Label
This field allows changing the channel label displayed with the measurement. The label size is
limited to 6 characters.
2 Unit
This field allows changing the channel unit displayed with the measurement. The unit size is
limited to 6 characters.
The high alarm level can be changed by pressing on this field. A default value is preset when
the channel is created. The high alarm level is primary used to fix the high limit represented by a
red line on the recorded measurement graph. The limit is used as the minimum graph scale.
The secondary use is to drive a high alarm relay if the relay button has been activated and if a
relay module has been installed on the selected module socket. In this case, the relay will be
activated as soon as the measurement overpass the high alarm value and will stay activated
until the measurement goes down below this value after subtracting the boundary value.
Refer to installation section for the relay wiring. Both functions NO (Normally Open) and NC
(Normally Closed) are available.
The low alarm level can be changed by pressing on this field. A default value is preset when the
channel is created. The low alarm level is primary used to fix the low limit represented by a
green line on the recorded measurement graph.
The secondary use is to drive a low alarm relay if the relay button has been activated and if a
relay module has been installed on the selected module socket. In this case, the relay will be
activated as soon as the measurement underpass the low alarm value and will stay activated
until the measurement goes down below this value after adding the boundary value.
Refer to installation section for the relay wiring. Both functions NO (Normally Open) and NC
(Normally Closed) are available.
5 Number of decimals
This field allows changing the number of decimal displayed for the measurement value to adjust
the display to significant decimals only.
Note: this choice does not affect the measurements stored in the internal memory or the
MODBUS registers.
6 Modbus address
This field is the Modbus address of the probe that has been fixed inside the probe during the
initialisation sequence.
This value must not be changed, unless the RS485 scan screen indicates that the probe has not
the proper address.
4.15.3 Test
1 Measured value
4.15.4 Calibration
4 5
This window displays the 10 last calibrations done with the date, time and the new calibration
factor.
These two buttons allow to scroll up and down the 10 last calibration records.
3 Calibration factor
This field allows changing manually the calibration factor of the channel.
This factor is normally changed automatically while doing a calibration procedure by pressing on
the “ADJUST” button.
4 Zero
This button enable to dot the zero. Be sure that the probe is immersed on pure water before
pressing on this button.
5 Calibration procedure
- Do a manual measurement on the same sample using the check screen (as describe above).
- Go on the calibration screen and press on the “ADJUST” button of the check screen. Validate
or enter the measurement and then enter the laboratory value on the keypad.
The calibration is finished. A new calibration factor has been calculated and has been recorded
inside the calibration history displayed on the check screen.
This new calibration factor will be taken in account for all the further measurements.
Maintenance
There is no special maintenance on this probe unless checking that the windows are clean.
Troubleshooting
Symptoms Origin
- Wrong zero
Measuring errors
4.15.6 Specifications
4.16.1 Principle
The light beam is generated by a pulsed laser diode while the light detection is performed by a
photodiode detector placed at 90° from the light beam as represented below.
Two models of laser diode are available: visible red at 655 nm or infrared at 850 nm.
The infra red model meet the IS07027 criteria. However, for the safety reasons and to ease the
troubleshooting, the visible red model is recommended.
Never observe directly the laser diode if dismounted. The infrared light might
damage the eye.
4.16.2 Settings
First screen:
1 Label
This field allows changing the channel label displayed with the measurement. The label size is
limited to 6 characters.
2 Unit
This field allows changing the channel unit displayed with the measurement. The unit size is
limited to 6 characters.
The high alarm level can be changed by pressing on this field. A default value is preset when
the channel is created. The high alarm level is primary used to fix the high limit represented by a
red line on the recorded measurement graph. The limit is used as the minimum graph scale.
The secondary use is to drive a high alarm relay if the relay button has been activated and if a
relay module has been installed on the selected module socket. In this case, the relay will be
activated as soon as the measurement overpass the high alarm value and will stay activated
until the measurement goes down below this value after subtracting the boundary value.
Refer to installation section for the relay wiring. Both functions NO (Normally Open) and NC
(Normally Closed) are available.
The low alarm level can be changed by pressing on this field. A default value is preset when the
channel is created. The low alarm level is primary used to fix the low limit represented by a
green line on the recorded measurement graph.
The secondary use is to drive a low alarm relay if the relay button has been activated and if a
relay module has been installed on the selected module socket. In this case, the relay will be
activated as soon as the measurement underpass the low alarm value and will stay activated
until the measurement goes down below this value after adding the boundary value.
Refer to installation section for the relay wiring. Both functions NO (Normally Open) and NC
(Normally Closed) are available.
5 Number of decimals
This field allows changing the number of decimal displayed for the measurement value to adjust
the display to significant decimals only.
Note: this choice does not affect the measurements stored in the internal memory or the
MODBUS registers.
Negatives values are normal for most of the measurements as they result from normal
fluctuations if the measurement is close to zero.
By default, the display of negative values is allowed as negative values may also result from a
wrong zeroing. Consequently, it may be useful to be aware of such problem when seeing
strongly negative values.
But as negative values have no theoretical meaning, it’s possible to replace them by zero by
selecting "NO" for this function.
Note that this function do not apply to the recorded measurements or to the MODBUS registers.
Second screen:
1 Scale factor
This factor is normally determined in factory to transform the basic measurement to a scaled
value in the final unit. It is has no reason to be changed except if optical elements are replaced.
2 Averaging
This field displays the number of elemental acquisitions averaged to form one measurement.
The default value is 50.
4.16.3 Test
2 3 4 5 6 7
1 Measured value
This button run the sampling pump for 10 seconds to renew the sample.
This button injects the cleaning solution and may be used to check the zero.
4 Stop button
This button display the laser diode signal. It should normally be below 300 on pure water. The
screen is shown below.
6 Zero button
Put first pure water inside the flow cell before pressing this button.
Signal screen:
4.16.4 Calibration
This window displays the 10 last calibrations done with the date, time and the new calibration
factor.
These two buttons allow to scroll up and down the 10 last calibration records.
3 Calibration factor
This field allows changing manually the calibration factor of the channel.
This factor is normally changed automatically while doing a calibration procedure by pressing on
the “ADJUST” button.
4 Offset
5 Calibration procedure
6 Linearization curve
As the Beer-Lambert law is not linear for high absorbance values, a linearization curve is
entered to compensate automatically this non-linearity. This button display the linearization
curve and enable to enter or check the linearization values.
The Y-axis corresponds to the rough measurements entered on the M1 to M10 fields while the
X-axis corresponds to the standard or final measurement, entered on the S1 to S10 fields.
The M1-M10 and S1-S10 values must be strictly increasing. It’s recommended to put 0.0 and
0.0 as starting values for S1 and M1. The final value is linearly extrapolated between these
points. Unused points at the end of the table must strictly remains at 0.0 both for S and M.
- Take a turbidity standard (formazine for example) representative of the measuring range
- Do a manual measurement on this standard using the check screen (as describe above).
- Go on the calibration screen and press on the “ADJUST” button of the check screen. Validate
the previous measurement and then enter the standard value on the keypad.
The calibration is finished. A new calibration factor has been calculated and has been recorded
inside the calibration history displayed on the check screen.
This new calibration factor will be taken in account for all the further measurements.
Maintenance
Troubleshooting
Symptoms Origin
Negative value - Bad zero or auto-zero: the cleaning solution has been prepared with
polluted water or fault in the acid cleaning system (no flow, pump
failure)
Value is too low - No water in the flow cell (clogging, bad inlet connection, fault in the
peristaltic pump)
Value is too high - Bubbles in the sample (check peristaltic pump inlet fitting).
Unstable value - Deposit or dirty on the flow cell, check the light level on the check
screen
Measuring errors
2 The light level is too - Check that there is no bubble inside the flow cell, if yes check
high the fittings
4.16.6 Specifications
Specifications
4.17 UV254
4.17.1 Principle
The measuring principle is based on the UV light absorption from organic matter at 254 nm
according to the Beer-Lambert law:
[C ] = k ⋅ log I 0
I1
The UV light is generated by a pulsed xenon lamp to reach an ultra long life time.
The light detection is performed by two photodiode detectors through interferential filters; one
centred at 254 nm while the other is used as reference to compensate turbidity, deposit inside
the flow cell and organic matter.
4.17.2 Settings
First screen:
1 Label
This field allows changing the channel label displayed with the measurement. The label size is
limited to 6 characters.
2 Unit
This field allows changing the channel unit displayed with the measurement. The unit size is
limited to 6 characters.
The high alarm level can be changed by pressing on this field. A default value is preset when
the channel is created. The high alarm level is primary used to fix the high limit represented by a
red line on the recorded measurement graph. The limit is used as the minimum graph scale.
The secondary use is to drive a high alarm relay if the relay button has been activated and if a
relay module has been installed on the selected module socket. In this case, the relay will be
activated as soon as the measurement overpass the high alarm value and will stay activated
until the measurement goes down below this value after subtracting the boundary value.
Refer to installation section for the relay wiring. Both functions NO (Normally Open) and NC
(Normally Closed) are available.
The low alarm level can be changed by pressing on this field. A default value is preset when the
channel is created. The low alarm level is primary used to fix the low limit represented by a
green line on the recorded measurement graph.
The secondary use is to drive a low alarm relay if the relay button has been activated and if a
relay module has been installed on the selected module socket. In this case, the relay will be
activated as soon as the measurement underpass the low alarm value and will stay activated
until the measurement goes down below this value after adding the boundary value.
Refer to installation section for the relay wiring. Both functions NO (Normally Open) and NC
(Normally Closed) are available.
5 Number of decimals
This field allows changing the number of decimal displayed for the measurement value to adjust
the display to significant decimals only.
Note: this choice does not affect the measurements stored in the internal memory or the
MODBUS registers.
6 Mode
This button allows to choose between the Abs/m mode or the COD mode.
The Abs/m mode is the absolute measuring mode, giving directly the absorbance of the sample
at 254 nm for a one meter optical path, compensated at 350 nm to remove the turbidity effect.
The COD mode is based on an experimental correlation between the UV absorbance at 254 nm
and the COD (Chemical Oxygen Demand).
This correlation is excellent for river water and drinking water and generally good for food
industries and paper mills.
Precautions must be taken for chemical industries effluent as the UV254/COD ratio
may depends on the process, each molecule having a specific UV245/COD ratio.
Mixing different kind of effluent might give inaccurate or inconsistent results.
The linearization curve is applied only in COD mode as the absorbance is inherently
linear.
Negatives values are normal for most of the measurements as they result from normal
fluctuations if the measurement is close to zero.
By default, the display of negative values is allowed as negative values may also result from a
wrong zeroing. Consequently, it may be useful to be aware of such problem when seeing
strongly negative values.
But as negative values have no theoretical meaning, it’s possible to replace them by zero by
selecting "NO" for this function.
Note that this function do not apply to the recorded measurements or to the MODBUS registers.
Second screen:
1 Scale factor
The default scale factor is fixed according to the flow cell optical path to transform the
absorbance to absorbance by meter:
2 Averaging
This field displays the number of elemental acquisitions averaged to form one measurement.
The default value is 200.
This button must be selected depending on the flow cell optical path that represents the
distance between the two quartz cylinders visible inside the flow cell.
No change must be done unless the flow cell is changed. The scale factor is automatically set
according the flow cell size.
The size of the flow cell is factory selected depending of the application:
- 10 mm: drinking water, low turbidity river water and low turbidity effluents
- 3 mm: turbid river water and turbid effluents
- 1 mm: rough waste water, inlet of municipal waste water
4 Background factor
This field allows to add partially the turbidity measurement made on another channel to improve
the COD accuracy, considering that particles may contains organic matter that cannot been
measured directly by UV absorption. This factor is experimentally determined according to
laboratory measurements.
This field displays the turbidity channel number if the turbidity compensation is applied.
This number doesn't matter if the turbidity factor remains at zero.
.
4.17.3 Test
The check screen displays the current value of the sample inside the flow cell.
2 3 4 5 6 7
1 Measured value
Immediate measurement obtained with the sample inside the flow cell. Can be renewed by
pressing on the play button.
This button run the sampling pump for 10 seconds to renew the sample.
This button injects the cleaning solution. It can be used to check the zero.
4 Stop button
This button display the lamp signal. The PH1 (peak photo detector) value should normally be
between 1000 and 2500 on pure water, PH2 (reference photodiode) should be between 1000
and 2500 also on pure water.
PH1 decrease according to the organic matter concentration and the turbidity of the sample.
PH2 is affected only by the turbidity of the sample. The screen is shown below.
6 Zero button
Put first pure water inside the flow cell before pressing this button.
7 Start measurement
This button update the measured value directly with what is inside the flow cell.
Signal screen:
4.17.4 Calibration
This window displays the 10 last calibrations done with the date, time and the new calibration
factor.
These two buttons allow to scroll up and down the 10 last calibration records.
3 Calibration factor
This field allows changing manually the calibration factor of the channel.
This factor is normally changed automatically while doing a calibration procedure by pressing on
the “ADJUST” button.
4 Offset
5 Calibration procedure
Enter the measured value obtained on the sample given to the laboratory
Then enter the laboratory COD value.
6 Linearization curve
As the Beer-Lambert law is not linear for high absorbance values, a linearization curve is
entered to compensate automatically this non-linearity. This button display the linearization
curve and enable to enter or check the linearization values.
The Y-axis corresponds to the rough measurements entered on the M1 to M10 fields while the
X-axis corresponds to the standard or final measurement, entered on the S1 to S10 fields.
The M1-M10 and S1-S10 values must be strictly increasing. It’s recommended to put 0.0 and
0.0 as starting values for S1 and M1. The final value is linearly extrapolated between these
points. Unused points at the end of the table must strictly remains at 0.0 both for S and M.
The analyser gives the result in absorbance by meter by default. No calibration is required in
this measuring mode regarding the accuracy of flow cell optical path (+/- 0.1 mm).
If the measuring mode is in COD and after than the UV254-COD relation has been tested, the
analyser must be calibrated according to a COD laboratory measurement as each kind of
effluent has a specific UV254/COD ratio.
The default calibration in COD mode corresponds to river water with a calibration factor of 0.5.
The calibration factor for municipal wastewater is around 10.
Note that many saturated organic compounds like glucose or alcohol do not have UV
absorption.
- Do a manual measurement on this sample using the check screen (It's recommended to check
first the zero).
- When the laboratory measurement is know, go on the calibration screen and press on the
“ADJUST” button of the check screen. Enter the measurement given by the analyser and then
enter the laboratory value on the keypad.
The calibration is finished. A new calibration factor has been calculated and has been recorded
inside the calibration history displayed on the check screen.
This new calibration factor will be taken in account for all the further measurements.
Maintenance
Troubleshooting
Symptoms Origin
Negative value - Bad zero or auto-zero: the cleaning solution has been prepared with
polluted water or fault in the acid cleaning system (no flow, pump
failure)
Value is too low - No water in the flow cell (clogging, bad inlet connection, fault in the
peristaltic pump)
Unstable value - Deposit or dirty on the flow cell, check the light level on the check
screen
Measuring errors
3 The light level is too - Check that there is no bubble inside the flow cell, if yes check
high on the peak the fittings
photo detector PH1
- Failure on the photo detector board or cable (repair or
replace)
4 The light level is too - Check that there is no bubble inside the flow cell, if yes check
high on the reference the fittings
photo detector PH2
- Failure on the photo detector board or cable (repair or
replace)
5 The light level is too - Over range or very high turbidity. Recheck on pure water
low on the peak
photo detector PH1 - Dirty on the flow cell windows. Check if the automatic
cleaning system is working properly
6 The light level is too - Over range or very high turbidity. Recheck on pure water
low on the reference
photo detector PH2 - Dirty on the flow cell windows. Check if the automatic
cleaning system is working properly
4.17.6 Specifications
Flow cell 1 mm 3 mm 10 mm
Accuracy +/- 2 Abs/m or +/- 5%* +/- 0.6 Abs/m or +/- 5%* +/- 0.2 Abs/m or +/- 5%*
* whichever is greater
The period of replacement of the pump tubing depends on the use. A weekly checking is
recommended. The typical lifetime with Pharmed® tubing is estimated to about 1000 hours of
continuous operation.
General troubleshooting:
The screen remains totally black after - Check the power socket
connecting the power cord.
- Check J1 connector (mains input, high voltage!)
AND
- Check J2 connector on the DSP400TFT (mains
The red LED D11 on the DSP400TFT input for the power supply, high voltage!)
board is OFF.
- Check J3 connector (12V DC output from the
power supply)
The screen is lighted but nothing is - Check the screen connector J6 on the top right of
displayed or unstable display. the DSP400TFT board
6 General Specifications
Sample temperature: 0 ºC to 80 ºC
Sample pressure: 4 bar (4000 hPa or 60 psi) without sampling pump or 1 bar (1000 hPa or
15 psi) with the sampling pump
Display: Colour TFT LCD, Size: 5.7”, resolution: 320 x 240 pixels
LED backlight
Resistive touch screen
Memory: 5000 records (up to 16 measurement channels) with date and time
USB port For USB memory keys, any format (FAT16, FAT32)
Standard USB connector type A with IP68 protective cap
- Recorded measurement downloads (compatible with Excel®)
- Complete configuration backup/restore (proprietary format)
- Screen copy in BMP format (compatible with Windows®)
- Software update
Operating limits: 0 to 50 ºC
-----