2. WORKING PRINCIPLES
2.1 Measuring Principles
The measuring principle of electromagnetic flow meter is based on the electromagnetic
induction law of Farady. The sensor is mainly composed of measuring tube with isolate lining, a
pair of electrodes installed by penetration of the measuring tube wall, a pair of coils and iron
core to produce working magnetic field. When the conductive fluid flows through the measuring
tube of the sensor, the voltage signal in direct proportion to the average flow velocity of the fluid
will be inducted on the electrodes. The signal is amplified and treated by the transmitter to
realize various display functions.
3. PRODUCT CLASSIFICATION
3.1Product components
Electromagnetic flow meter is composed of sensor and converter. The remote-type also
needs a special double-layer shielded cable to connect the converter and the sensor.
3.2Product models
WELTER electromagnetic flow meter has two forms: compact-type and remote-type. There
are seven kinds of electrode materials and four kinds of lining materials available for sensor.
4. TECHNICAL SPECIFICATION
4.1The flow meter is compliant to Standard “JB/T 9248-1999 Electromagnetic Flow Meter”.
4.2Maximum flow speed: 15m/s
4.3DN size: 3, 6, 10, 15, 20, 25, 32, 40, 50, 65, 80, 100, 150, 200, 250, 300, 350, 400, 450,
500, 600, 700, 800, 900, 1000, 1200, 1400, 1600, 1800, 2000
4.4Accuracy: 0.5%, 0.2%
4.5Nominal Pressure: 4.0MPa(DN0-150)
1.6MPa(DN200-600)
1.0MPa(DN700-1200)
0.6MPa(DN400-2000)
Or other specified by order
4.6Materials
Electrode Forms and Materials:
Electrode design has four forms: standard form, scraper form, detachable form and
grounding electrode.
Electrode material is selectable from stainless steel containing Mo, stainless steel coated with
carbonized tungsten, Hastelloy B, Hastelloy C, Titanium, Tantalum and platinum-iridium alloy.
Flange material: carbon steel
Grounding ring: stainless steel
Inlet protection ring: carbon steel, stainless steel
4.7Enclosure
IP65:
IP68: only available for remote-type sensor with neoprene or polyurethane liner, not
including flameproof model.
4.8Flameproof Standard
DN15-DN600 compact-type: md II BT4
DN15-DN1600 remote-type: sensor and converter both installed in dangerous area: md II
BT4.
DN15-DN1600 remote-type: sensor installed in dangerous area and converter installed in
safe area: md II BT4
4.9Connection Cable
Special cable is needed to connect the sensor and the converter for remote-type flowmeter.
The cable length should not exceed 100 meters. A 10-meter cable is supplied free of charge and
the rest is to be ordered.
4.10General Specification of Converter
Power supply: AC 85-265V, 45-63Hz, ≤20W; DC 11-40V
Converter display and operation: four keys are available to set all parameters. External
handheld or PC can be used to do the configuration. High resolution LCD display with
backlight, empty pipe detection and self-diagnostic function are equipped in the converter.
Digital communication: RS485, RS232, MODBUS,REMOTE
Output Signals:
- Current output: fully-isolated, 4-20mA/0-10mA
load resistance: 0-10mA: 0-1.5KΩ; 4-20mA: 0-750Ω.
- Frequency output: bidirectional flow output. Frequency output is proportional to the flow
percentage of the full range. The converter provides fully isolated transistor open collector
frequency output ranged from 1 to 5000 Hz. The external DC power supply should not exceed
35V and maximum collector current is 250mA.
- Pulse output: bidirectional flow output. The converter can output up to 5000cp/s pulse
series, which is dedicated to external totalization. Pulse factor is defined as volume or mass per
pulse. It can be set to 0.001L/p, 0.01L/p, 0.1L/p, 1L/p, 2L/p, 5L/p, 10L/p, 100L/p, 1m3/p, 10
m3/p, 100 m3/p or 1000 m3/p. Pulse width is selectable from auto, 10ms, 20ms, 50ms, 100ms,
150ms, 200ms, 250ms, 300ms, 350ms and 400ms. Photo-coupler isolated transistor open
collector circuit is used for pulse output. The external DC power supply should not exceed 35V
and maximum collector current is 250mA.
- Flow direction indication: The converter is capable of measuring both forward and reverse
flow and recognizing its direction. The converter outputs 0V low level for forward flow, while
+12V high level for reverse flow.
- Alarm output: Two channels of photo-coupler isolated open collector circuit are used for
alarm signal output. There are two alarm outputs: high limit alarm and low limit alarm. The
external DC power supply should not exceed 35V and maximum collector current is 250mA.
- Damping constant: Damping time is selectable from 0.2 to 100s.
5. APPEARANCE AND INSTALLATION
Dimensions of Converter, shown as Fig 3.
+28 .7
Alarm Indicator
Flow Rate
m3/h
Flow Unit
Flow Velocity
P +100.0 % Flow Percentage
+0012345 .78 m3 Electrode Resistence
∑
6 Flow Range Ratio
Forward Total Flow
Alarm Display
Output
100% 20mA
Down-going
Zero Flow
change point
Up-going Change 4mA
Point
Fig. 5 Illustration of Auto-Range-Change
6.5.4 Damping
Long damping constant can improve the stability of display and output and is suitable to
flow control application; while short damping constant has short response time and is suitable to
the totalization of pulse flow. Damping time is selectable from 0.2s to 100s.
6.5.5 Flow Dir.
If the displayed direction sign is not agreed to the actual flow direction, change this item to
the opposite option.
6.5.6 Flow Zero
To conduct zero adjustment, the fluid in the sensor pipe must be held still. The flow zero is
displayed by flow speed and the unit is m/s. The display of flow zero is shown below:
FS=○○.○○○m/s
±○○○○○
On the LCD, the top line displays the measured zero point while the bottom line shows the
adjustment value. If the FS is not equal to 00.000m/s, adjust the sign and value on bottom line
until FS back to nil. Remind again: to adjust the flow zero, the sensor pipe must be filled and the
fluid must be kept still. The flow zero adjustment value is an important constant of the meter and
should be printed on the calibration sheet and label. The value should include the sign and
amount by unit of m/s.
6.5.7 L.F. Cutoff and Cutoff Enble
Low flow cutoff is set in percentage relative to flow range. If Cutoff is enabled and flow is
lower than the set value, the display of flow rate, speed and percentage and signal outputs are
forced to nil. If the item is disabled, no action is taken.
6.5.8 Rate-Of-Chng and Limit Time
‘Rate-of-change’ limit technique is used to eliminate application-related high electrical
noise contained in the process flow signal.
To check electrical noise, two parameters are defined: ‘Rate-of-change’ limit and ‘Control
limit time’. If the sampled flow value exceeds the set rate-of-change limit value based on the
averaged flow rate value up until the sampled time, the system will reject that sampled value and
instead the averaged value including the rate-of-change limit value in place of the rejected
sampled value will be output. However, if the limit-exceeding sampled value continues for the
same flow direction for more than the preset control limit time, that data will be used as output
signal. Fig 6 illustrates the effect of noise-suppressing by rate-of-change limit.
The value of rate-of-change limit can be set from 0 to 30% of flow range and limit time
ranges from 0 to 20 seconds. If either of the two parameters is set to nil, the function is disabled.
The rate-of-change limit function is not suitable for short period measurement and flow
meter calibration.
7. INSTALLATION
The installation of flowmeter is shown as Fig. 7.
8. WIRING
8.1 GROUDING
T- RS485-B
T+ RS485-A
P+ Frequency/pulse output +
I+ Current output +
PUL J
+
+
Internal D
E
-
Circuit PCO
10. PACKAGING
The package includes:
The electromagnetic flowmeter ordered;
Instruction Manual;
12. OPERATION
Before operation, the following inspection should be done to check if:
a. There is any damage caused by transportation or installation;
b. The power used is same as the label on the flowmeter;
c. The wiring is correct.
After inspection, turn the valve on to fill the pipe up and make sure there is no leakage and
the gas inside the pipe is eliminated. Switch on the power supply and the flowmeter is ready to
use after 10 minutes warm-up.
If there is any problem, please refer to the Section 9 for troubleshooting. If still not working
properly, contact the manufacturer immediately.