Direct Bilirubin: Instrument

This document provides settings and parameters for validating a direct bilirubin assay on an ARCHITECT C8000 instrument. The reagent is prepared by mixing the contents of one Reagent B-T vial into a Reagent A-D bottle. The instrument is set to a wavelength of 548/700 nm, absorbance range of 0-15 mg/dL, and 3 replicates. A linear calibration method is used with a single calibrator level and validity checks including absorbance ranges and calibration factor tolerance.

Uploaded by

Enrique Duarte
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
91 views1 page

Direct Bilirubin: Instrument

This document provides settings and parameters for validating a direct bilirubin assay on an ARCHITECT C8000 instrument. The reagent is prepared by mixing the contents of one Reagent B-T vial into a Reagent A-D bottle. The instrument is set to a wavelength of 548/700 nm, absorbance range of 0-15 mg/dL, and 3 replicates. A linear calibration method is used with a single calibrator level and validity checks including absorbance ranges and calibration factor tolerance.

Uploaded by

Enrique Duarte
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 1

DIRECT BILIRUBIN

Spectrophotometric
DIAZOTIZED SULFANILIC

The application parameters comprised here constitute a guide to facilitate the validation of our reagents by the
instrument. It is advisable to validate the use when there is any change in software or reagent versions.

Instrument: ARCHITECT C8000


Reagent preparation
Working Reagent: Transfer the contents of one Reagent B-T vial into a Reagent A-D bottle. Mix thoroughly.
Stable for 20 days at 2-8 ºC.
Instrument settings

General Conventional Units


Assay BIL-D Type Photometric Results Unit mg/dL
Number ... Version 1 Assay Defaults Low Linearity 0.05
High Linearity 15
Reaction Definition
Reaction Mode End Up Read Times Gender and Age specific Ranges
Wavelength 548 / 700 Main 16 -- 18 GENDER AGE (Y) NORMAL
Last Required Read 18 Flex __ -- __ Male 0-130 0 – 0.2
Absorbance Range … / … Color Correction __ -- __ Female 0-130 0 – 0.2
Sample Blank Type None Blank Read Time __ -- __ Either 0-130 0 – 0.2

Reagent / Sample R1 R2 Reults Units mg/dL


Reagent BIL-D Reagent Volume 200 Decimal Places 1 (Range 0-4)
Diluent Saline Water Volume __ __ Correlation Factor 1.0000
Diluent Disp. Mode Type 0 Dispense Mode Type 0 Type 0 Intercept 0.0000

Diluent Name Sample Water Dilution Factor


Standard 20 1:1.00

Validity Checks
Reaction definition None
Rate Linearity % ___
Calibration
Method Linear

Calibrators and Volumes


Calibrator Set Standard Calibrator Level Concentration Volume
Blank 0.0 20
Std. 1 (*) 20
Std. 2 -- --
Std. 3 -- --
Std. 4 -- --
Replicates 3 (Range 1-3) Std. 5 -- --

Intervals Validity Checks


Calibration Interval Full Interval 999 hours Blank Absorbance range: 0.0000 -- 0.0000
Calibration Type Adjust Type None Span Absorbance Range: 0.0000 -- 0.0000
Expected cal Factor : 0.00
Expected Cal factor tolerance: %0
(*) Enter the standard value
… Values to be entered by the user
Version 0707