Chapter 1
Chapter 1
Chapter
1
Measurements and
Instrumentation
MEASUREMENTS
Measurement is essentially the act, or the result, of a quantitative
comparison between a given quantity and a quantity of the same kind
chosen as a unit. The result of measurement is expressed by a number
representing the ratio of the unknown quantity to the adopted unit of
measurement. The physical embodiment of the unit of measurement as well
as that of its submultiples or multiple values is called a standard. The device
used for comparing the unknown quantity with the unit of measurement or a
standard quantity is called a measuring instrument.
1-1
Measurements and Instrumentation
Depending upon the methods used for comparing the unknown quantity with
the unit of measurement, electrical measuring instruments may be classified
as direct measuring and comparison instruments.
1-2
Measurements and Instrumentation
1-3
Measurements and Instrumentation
MEASUREMENT TERMS
In electrical measurements and instrumentation the knowledge of the terms
and quantities usually used for such purposes is essential. Hence the terms
commonly used in electrical measurements and instrumentation are defined
and explained in a simple form to facilitate easy identification and reference.
1-4
Measurements and Instrumentation
standard signal such as 20— 100 kpa; 10—50 mA etc.) is another form of
measured signal — sometimes called the transmitted signal
2- Range-Related Terms. Terms such as range and span are widely used
to describe the region over which a quantity is to be measured. The
following definitions will clarify the relationship between these terms.
(i) Range. The region between the limits within which a quantity
is measured, received or transmitted, expressed by stating the lower
and upper limits, is called the range such as 0 – 50 A, 50 – 500 V, –
20°C – 80°C
(ii) Span. The algebraic difference between the upper and lower
range values is called the span
(iii) Over Range. Any excess value of the input signal above its
upper range value or below its lower range value is called the over
range of a system or element.
1-5
Measurements and Instrumentation
(iv) Low Range Limit, The lowest value of quantity that a device
can be adjusted to measure is known as low range limit.
(v) Upper Range Limit. The highest value of the measured variable
that a device can be adjusted to measure is known as upper range limit.
(vi) Lower Range Value. The lowest value of the measured variable
a device is adjusted to measure is called the lower range value.
(vii) Upper Range Value. The highest value of the measured variable
a device is adjusted to measure is called the upper range value.
(ii) Pen Travel. The length of the path covered by the pen in moving
from one end of the chart scale to the other is called the pen travel.
The path may be an arc or a straight line.
1-6
Measurements and Instrumentation
details, which can be distinguished one from the other, is called the
resolution.
change in input which causes it after the steady state has been
reached is called the sensitivity.
(iii) Error. The algebraic difference between the indicated value and
+ ve error denotes that the indicated value is higher than true value.
1-7
Measurements and Instrumentation
It is helpful to use the word error in this concept only and to describe it as an
instrumental error meaning the difference between the average of a series of
up and down readings, as indicated by the instrument output, and the
corresponding true values of input.
(v) Span Error. The difference between the actual span and the
ideal span is called the span error and it is usually expressed as a
percent of ideal span.
(vi) Correction. The algebraic difference between the true value and
It is usually determined by subtracting the value of the dead band from the
maximum measured separation between up-scale going and down-scale
going indications of the measured variable (during a full range traverse,
unless otherwise specified) after transients have decayed. This measurement
is some times called hysteresis error.
1-8
Measurements and Instrumentation
(viii) Dead Band. It is the range through which an input can be varied
1-9
Measurements and Instrumentation
5- Dynamic Terms. These terms relate to the question, “how good are
measurements under dynamic conditions"?
1-10
Measurements and Instrumentation
Fig. 1.2
(iv) Signal to Noise Ratio. Ratio of signal amplitude may be peak or rms.
For non-sinusoidal signals, peak values should be used.
(vi) Ramp Response. The total (transient plus steady-state) time response
resulting from a sudden increase in the rate of change from zero to some
finite value of the input stimulus is called the ramp response.
(vii) Step Response. The total (transient plus steady-state) time response
resulting from a sudden change from one constant level of input to another
is called the step response.
1-11
Measurements and Instrumentation
(x) Rise Time. The time required for the output of a system (other than
first order) to make the change from a small specified percentage (say 5 or
10%) of the steady-state increment to a large specified percentage (usually
90 to 95 percent), either before or in the absence of overshoot, is called the
rise time (fig. 1.3).
Fig. 1.3
1-12
Measurements and Instrumentation
INSTRUMENTATION IN AUTOMATION
after The instrumentation system plays an important role in automation. An
automatic control error system (or automation) requires a comparator (or an
error detector), which measures the difference between the actual and desired
performance and actuates the control elements.
1-13
Measurements and Instrumentation
with in accurate tolerance of the desired level of liquid even though the
output flow rate through the value Vi is varied. The float (feed back path
element) senses the liquid level and positions the slider arm B on a
potentiometer. The slider arm A of another potentiometer is positioned
corresponding to the desired liquid level h (the reference input). When the
liquid level rises or falls, the potentiometers (error detector) give an error
voltage (error or actuating signal) proportional to the change in liquid level.
The error voltage actuates the motor through a power amplifier (control
elements) which in turn conditions the plant (i.e. decreases or increases the
opening of the valve V2) in order to restore the desired liquid level Thus the
control system automatically attempts to correct any deviation between the
actual and desired liquid levels in the tank.
1-14
Measurements and Instrumentation
1-15
Measurements and Instrumentation
Many a times certain operations are to be performed on the signal before its
further transmission so that interfering sources are removed so that the signal
may not get distorted. The process may be linear such as amplification,
attenuation, integration, differentiation, addition and subtraction or non-linear
such as modulation , detection, sampling filtering, chopping and clipping etc.
The process is called the signal conditioning. So signal conditioner follows
the primary sensing element or transducer, as the case may be.
1-16
Measurements and Instrumentation
while preserving the original nature of the signal. For example, an electronic
amplifier converts a small low voltage input signal into high voltage output
signal. Thus voltage amplifier acts as a variable manipulation element. It is
not necessary that this element follows the variable conversion element, as
shown in fig. 1.6. In many cases it may precede the variable conversion
element.
The final stage in a measuring system is called the terminating stage. In case
a control device is to be employed for final measurement stage, then it
becomes necessary to apply some feed back to the input signal to accomplish
the control objectives.
1-17
Measurements and Instrumentation
1-18
Measurements and Instrumentation
Fig. 1.7
1-19
Measurements and Instrumentation
room, and may vary continuously with time at any position within the room.
An analog sensor, such as a bulb thermometer or a thermocouple, will
continuously respond to such temperature changes. Such a continuous signal
is shown in Fig. 1.8, where the signal magnitude is analogous to the
measured variable (temperature) and the signal is continuous in both
magnitude and time.
Digital sensors provide a signal that is a direct digital representation of the
measurand. Digital sensors are basically binary (“on” or “off ”) devices.
Essentially, a digital signal exists at only discrete values of time (or space).
And within that discrete period, the signal can represent only a discrete
number of magnitude values. A common variation is the discrete sampled
signal representation, which represents a sensor output in a form that is
discrete both in time or space and in magnitude.
1-20
Measurements and Instrumentation
limit for the different values that a digital device can discern. For example, a
2-bit device can express 22 or 4 different numbers, 00, 01, 10, and 11,
corresponding to the values of 0, 1, 2, and 3, respectively. Thus, the
resolution in a magnitude discerned by a digital sensor is inherently limited to
1 part in 2M.
1-21
Measurements and Instrumentation
1-22
Measurements and Instrumentation
Hysteresis Some sensors do not return to the same output value when the
input stimulus is cycled up or down. The width of the expected error in terms
of the measured quantity is defined as the hysteresis. Typical units are kelvin
or percent of FSO.
Nonlinearity (often called Linearity) The maximum deviation from a linear
transfer function over the specified dynamic range. There are several
measures of this error. The most common compares the actual transfer
function with the “best straight line,” which lies midway between the two
parallel lines that encompass the entire transfer function over the specified
dynamic range of the device. This choice of comparison method is popular
because it makes most sensors look the best. Other reference lines may be
used, so the user should be careful to compare using the same reference.
Noise All sensors produce some output noise in addition to the output signal.
In some cases, the noise of the sensor is less than the noise of the next
element in the electronics, or less than the fluctuations in the physical signal,
in which case it is not important. Many other cases exist in which the noise of
the sensor limits the performance of the system based on the sensor. Noise is
generally distributed across the frequency spectrum. Many common noise
sources produce a white noise distribution, which is to say that the spectral
noise density is the same at all frequencies. Johnson noise in a resistor is a
good example of such a noise distribution. For white noise, the spectral noise
density is characterized in units of volts/Root (Hz). A distribution of this
nature adds noise to a measurement with amplitude proportional to the square
root of the measurement bandwidth. Since there is an inverse relationship
between the bandwidth and measurement time, it can be said that the noise
decreases with the square root of the measurement time.
Resolution The resolution of a sensor is defined as the minimum detectable
signal fluctuation. Since fluctuations are temporal phenomena, there is some
relationship between the timescale for the fluctuation and the minimum
detectable amplitude. Therefore, the definition of resolution must include
some information about the nature of the measurement being carried out.
Many sensors are limited by noise with a white spectral distribution. In these
1-23
Measurements and Instrumentation
This expression may be used to predict the behavior of the sensor, and
contains information about the sensitivity and the offset at the output of the
sensor.
Sensitivity The sensitivity of the sensor is given by the derivative of the
voltage with respect to acceleration at the initial operating point. For this
device, the sensitivity is 167 mV/g.
Dynamic Range The stated dynamic range for the ADXL322 is ±2g. For
signals outside this range, the signal will continue to rise or fall, but the
1-24
Measurements and Instrumentation
1-25