A 1 C
A 1 C
A1C-3
Tina-quant Hemoglobin A1c Gen.3 - Hemolysate and Whole Blood Application
Order information
Analyzer(s) on which cobas c pack(s) can be used
05336163 190 Tina-quant Hemoglobin A1c Gen.3 (150 tests) System ID 07 7455 3 Roche/Hitachi cobas c 311, cobas c 501/502
04528417 190 Calibrator f.a.s. HbA1c (3 x 2 mL) Code 674
05479207 190 PreciControl HbA1c norm (4 × 1 mL) Code 208
05912504 190 PreciControl HbA1c path (4 × 1 mL) Code 209
04528182 190 Hemolyzing Reagent Gen.2 (51 mL)* System ID 07 6873 1
HbA1c Hemolyzing Reagent for Tina‑quant HbA1c For Hemolysate
11488457 122
(1000 mL) Application only
* The value encoded in the instrument settings is 45 mL to account for the dead volume of the bottles.
A1C-3
Tina-quant Hemoglobin A1c Gen.3 - Hemolysate and Whole Blood Application
A1C-3
Tina-quant Hemoglobin A1c Gen.3 - Hemolysate and Whole Blood Application
A1C-3
Tina-quant Hemoglobin A1c Gen.3 - Hemolysate and Whole Blood Application
R1 120 μL - Hb
R3 24 μL - Calibrators S1-S2: C.f.a.s. HbA1c
Sample volumes Sample Sample dilution Calibration mode Linear
Sample Diluent HbA1c
(Hemolyzing Calibrators S1-S6: C.f.a.s. HbA1c
reagent)
Calibration mode Spline
Normal 5 μL - -
Calibration frequency Hb and HbA1c: full calibration is
Decreased 5 μL - - recommended
Increased 5 μL - - - after 29 days during shelf life
cobas c 501/502 test definition HbA1c (A1-H3) - after reagent lot change
Assay type 2‑Point End - as required following quality control
procedures
Reaction time / Assay points 10 / 34-70
Always calibrate both assays (Hb
Wavelength (sub/main) 660 / 340 nm
and HbA1c) in parallel. Automatic
Reaction direction Increase calibration at QC failure should be
Unit mmol/L (g/dL) deactivated.
Reagent pipetting Diluent (H2O) Calibration interval may be extended based on acceptable verification of
calibration by the laboratory.
R1 120 μL - Traceability: This method has been standardized against the approved
R3 24 μL - IFCC reference method for the measurement of HbA1c in human blood 14,15
and can be transferred to results traceable to DCCT/NGSP by calculation.
Sample volumes Sample Sample dilution Note for Whole Blood and Hemolysate Application
Sample Diluent Enter the assigned lot-specific and application-specific value of the
(Hemolyzing calibrator. Use the appropriate C.f.a.s. HbA1c calibrator only.
reagent) The cobas c Hemolyzing Reagent Gen.2 pack, 51 mL,
Normal 5 μL - - Cat. No. 04528182 190, needs to be available on the analyzer otherwise
the calibration cannot be performed.
Decreased 5 μL - - Quality control for Whole Blood and Hemolysate Application
Increased 5 μL - - For quality control, use control materials as listed in the "Order information"
Ratio definition for HbA1c (mmol/mol (IFCC) or % (DCCT/NGSP)) cal section.
culation In addition, other suitable control material can be used.
The control intervals and limits should be adapted to each laboratory’s
Protocol 1 (mmol/mol HbA1c acc. to IFCC): individual requirements. Values obtained should fall within the defined
Abbreviated ratio name RHI3 limits. Each laboratory should establish corrective measures to be taken if
values fall outside the defined limits.
Equation (A1-H3/HB-H3) x 1000
Follow the applicable government regulations and local guidelines for
Unit mmol/mol quality control.
Protocol 2 (% HbA1c acc. to DCCT/NGSP): Calculation for Whole Blood and Hemolysate Application
Abbreviated ratio name RHD3 (861) Hb, HbA1c
Roche/Hitachi cobas c systems automatically calculate the analyte
Equation (A1-H3/HB-H3) x 91.5 + 2.15 concentration of each sample.
Unit % HbA1c ratio calculation:
Protocol 2 is already implemented in the application (ACN 861). The For calculation of the mmol/mol HbA1c value (IFCC) and the percent
mmol/mol HbA1c values according to Protocol 1 (IFCC) must be manually HbA1c value (DCCT/NGSP), refer to the Test principle and Ratio
calculated according to the above equation. If requested a calculated test definition for mmol/mol HbA1c and % HbA1c calculation sections in this
with the formula in protocol 1 can be programmed under Utility > calculated method sheet.
test on the Roche/Hitachi cobas c 311 analyzer and on the Roche/Hitachi Limitations – interference for Whole Blood and Hemolysate
cobas c 501/502 analyzers. Please use the following settings: Application12,13,16,17,18,19,20,21,22,23
Sample Type Supernt. 1. For diagnostic purposes, mmol/mol HbA1c values (IFCC) and % HbA1c
values (DCCT/NGSP) should be used in conjunction with information from
Unit of Measure mM/M other diagnostic procedures and clinical evaluations.
Report Name HbA1c Gen.3 IFCC 2. The test is designed only for accurate and precise measurement of
mmol/mol HbA1c (IFCC) and % HbA1c (DCCT/NGSP). The individual
Item RHI3 results for total Hb and HbA1c concentration should not be reported.
Formula (A1-H3/HB-H3) x 1000 3. As a matter of principle, care must be taken when interpreting any HbA1c
The ratio for HbA1c (mmol/mol HbA1c acc. to IFCC and % HbA1c acc. to result from patients with Hb variants. Abnormal hemoglobins might affect
DCCT/NGSP) will be automatically calculated after result output of both the half life of the red cells or the in vivo glycation rates. In these cases
tests. It is recommended to report % HbA1c values (DCCT/NGSP) to one even analytically correct results do not reflect the same level of glycemic
decimal place and mmol/mol HbA1c values (IFCC) without decimal places, control that would be expected in patients with normal hemoglobin.21
which can be entered in the editable field “expected values”. Whenever it is suspected that the presence of an Hb variant (e.g. HbSS,
HbCC or HbSC) affects the correlation between the HbA1c value and
Calibration for Whole Blood and Hemolysate Application glycemic control, HbA1c must not be used for the diagnosis of diabetes
mellitus.
A1C-3
Tina-quant Hemoglobin A1c Gen.3 - Hemolysate and Whole Blood Application
4. Any cause of shortened erythrocyte survival or decrease in mean Limit of Detection = 0.62 mmol/L (1.00 g/dL)
erythrocyte age will reduce exposure of erythrocytes to glucose with a HbA1c:
consequent decrease in mmol/mol HbA1c values (IFCC) and % HbA1c
values (DCCT/NGSP), even though the time-averaged blood glucose level Limit of Blank = 0.12 mmol/L (0.19 g/dL)
may be elevated. Causes of shortened erythrocyte lifetime might be Limit of Detection = 0.18 mmol/L (0.29 g/dL)
hemolytic anemia or other hemolytic diseases, homozygous sickle cell trait, The Limit of Blank and Limit of Detection were determined in accordance
pregnancy, recent significant or chronic blood loss, etc. Similarly, recent with the CLSI (Clinical and Laboratory Standards Institute) EP17‑A
blood transfusions can alter the mmol/mol HbA1c values (IFCC) and requirements.
% HbA1c values (DCCT/NGSP).Caution should be used when interpreting
the HbA1c results from patients with these conditions. HbA1c must not be The Limit of Blank is the 95th percentile value from n ≥ 60 measurements of
used for the diagnosis of diabetes mellitus in the presence of such analyte‑free samples over several independent series. The Limit of Blank
conditions. corresponds to the concentration below which analyte‑free samples are
found with a probability of 95 %.
5. Glycated HbF is not detected by the assay as it does not contain the
glycated β‑chain that characterizes HbA1c. However, HbF is measured in The Limit of Detection is determined based on the Limit of Blank and the
the total Hb assay and as a consequence, specimens containing high standard deviation of low concentration samples.
amounts of HbF (> 10 %) may result in lower than expected mmol/mol The Limit of Detection corresponds to the sample concentration which leads
HbA1c values (IFCC) and % HbA1c values (DCCT/NGSP).13,23 with a probability of 95 % to a measurement result above the Limit of Blank.
6. mmol/mol HbA1c values (IFCC) and % HbA1c values (DCCT/NGSP) are Expected values
not suitable for the diagnosis of gestational diabetes.24 Protocol 1 (mmol/mol HbA1c acc. to IFCC): 29‑42 mmol/mol HbA1c27
7. In very rare cases of rapidly evolving type 1 diabetes the increase of the Protocol 2 (% HbA1c acc. to DCCT/NGSP): 4.8‑5.9 % HbA1c27
HbA1c values might be delayed compared to the acute increase in glucose
concentrations. In these conditions diabetes mellitus must be diagnosed This reference range was obtained by measuring 474 well-characterized
based on plasma glucose concentrations and/or the typical clinical healthy individuals without diabetes mellitus. HbA1c levels higher than the
symptoms.24 upper end of this reference range are an indication of hyperglycemia during
the preceding 2 to 3 months or longer. According to the recommendations
Criterion: Recovery within ± 10 % of initial value. of the American Diabetes Association values above 48 mmol/mol HbA1c
Icterus:20 No significant interference up to an I index of 60 for conjugated (IFCC) or 6.5 % HbA1c (DCCT/NGSP) are suitable for the diagnosis of
and unconjugated bilirubin (approximate conjugated and unconjugated diabetes mellitus.24,28 Patients with HbA1c values in the range of
bilirubin concentration: 1026 µmol/L or 60 mg/dL). 39‑46 mmol/mol HbA1c (IFCC) or 5.7‑6.4 % HbA1c (DCCT/NGSP) may be
Lipemia (Intralipid):20 No significant interference up to an Intralipid at risk of developing diabetes.24,28
concentration of 600 mg/dL. There is poor correlation between triglycerides HbA1c levels may reach 195 mmol/mol (IFCC) or 20 % (DCCT/NGSP) or
concentration and turbidity. higher in poorly controlled diabetes. Therapeutic action is suggested at
Glycemia: No significant interference up to a glucose level of 55.5 mmol/L levels above 64 mmol/mol HbA1c (IFCC) or 8 % HbA1c (DCCT/NGSP).
(1000 mg/dL). A fasting sample is not required. Diabetes patients with HbA1c levels below 53 mmol/mol (IFCC) or 7 %
(DCCT/NGSP) meet the goal of the American Diabetes Association.19,18
Rheumatoid factors: No significant interference from rheumatoid factors up
to a concentration of 750 IU/mL. HbA1c levels below the established reference range may indicate recent
episodes of hypoglycemia, the presence of Hb variants, or shortened
Drugs: No interference was found at therapeutic concentrations using lifetime of erythrocytes.
common drug panels.25,26
Each laboratory should investigate the transferability of the expected values
Other: No cross reactions with HbA0, HbA1a, HbA1b, acetylated to its own patient population and if necessary determine its own reference
hemoglobin, carbamylated hemoglobin, glycated albumin and labile HbA1c ranges.
were found for the anti-HbA1c antibodies used in this kit.
For diagnostic purposes, the results should always be assessed in Specific performance data
conjunction with the patient’s medical history, clinical examination and other Representative performance data on the analyzers are given below.
findings. Results obtained in individual laboratories may differ.
ACTION REQUIRED Precision
Special Wash Programming: The use of special wash steps is mandatory Precision was determined using human samples and controls in
when certain test combinations are run together on Roche/Hitachi accordance with the CLSI (Clinical and Laboratory Standards Institute) EP5
cobas c systems. The latest version of the carry‑over evasion list can be requirements with repeatability and intermediate precision (2 aliquots per
found with the NaOHD-SMS-SmpCln1+2-SCCS Method Sheets. For further run, 2 runs per day, 21 days). The following results were obtained (data
instructions refer to the operator’s manual. cobas c 502 analyzer: All based on DCCT/NGSP values):
special wash programming necessary for avoiding carry‑over is available
via the cobas link, manual input is required in certain cases. Whole Blood Application:
Where required, special wash/carry‑over evasion programming must Repeatability Mean SD CV
be implemented prior to reporting results with this test.
% HbA1c % %
Limits and ranges
Measuring range PreciControl HbA1c norm 5.3 0.07 1.3
Hemoglobin: 2.48‑24.8 mmol/L (4‑40 g/dL). PreciControl HbA1c path 9.9 0.11 1.1
HbA1c: 0.186‑1.61 mmol/L (0.3‑2.6 g/dL) Human sample 1 4.4 0.07 1.6
This corresponds to a measuring range of 23‑196 mmol/mol HbA1c (IFCC) Human sample 2 5.6 0.09 1.6
and 4.2‑20.1 % HbA1c (DCCT/NGSP) at a typical hemoglobin
concentration of 8.2 mmol/L (13.2 g/dL). Human sample 3 8.0 0.08 1.0
In rare cases of “>Test” flags which might occur with the use of the whole Human sample 4 10.6 0.11 1.1
blood application, remix the whole blood sample and repeat the analysis
with the same settings. Intermediate precision Mean SD CV
It is recommended to switch the auto rerun function off. % HbA1c % %
Lower limits of measurement
PreciControl HbA1c norm 5.3 0.08 1.4
Limit of Blank and Limit of Detection
Hemoglobin: PreciControl HbA1c path 9.9 0.15 1.5
Limit of Blank = 0.31 mmol/L (0.50 g/dL) Human sample 1 4.4 0.09 1.9
A1C-3
Tina-quant Hemoglobin A1c Gen.3 - Hemolysate and Whole Blood Application
Human sample 2 5.6 0.11 2.0 % HbA1c (DCCT/NGSP) values for human blood samples obtained on a
Roche/Hitachi cobas c 501 analyzer using the Tina‑quant Hemoglobin A1c
Human sample 3 8.0 0.11 1.4 Gen.3 reagent with the whole blood application (y) were compared with
Human sample 4 10.6 0.16 1.5 those determined using the Tina‑quant Hemoglobin A1c Gen.2 reagent with
the hemolysate application on a COBAS INTEGRA 800 analyzer (x).
Hemolysate Application:
Sample size (n) = 80
Repeatability Mean SD CV
Mean difference -0.09 % HbA1c
% HbA1c % %
Lower 95 % confidence interval of -0.46 % HbA1c
PreciControl HbA1c norm 5.1 0.07 1.3 differences
PreciControl HbA1c path 10.2 0.10 1.0 Upper 95 % confidence interval of 0.28 % HbA1c
Human sample 1 4.3 0.06 1.4 differences
Human sample 2 5.6 0.07 1.2 The sample concentrations were between 4.7 % and 9.8 % (DCCT/NGSP
Human sample 3 8.2 0.08 1.0 values).
Human sample 4 10.9 0.11 1.0 Hemolysate Application:
% HbA1c (DCCT/NGSP) values for human blood samples obtained on a
Intermediate precision Mean SD CV Roche/Hitachi cobas c 501 analyzer using the Tina‑quant Hemoglobin A1c
Gen.3 reagent with the whole blood application (y) were compared with
% HbA1c % % those determined using the Tina‑quant Hemoglobin A1c Gen.2 reagent with
PreciControl HbA1c norm 5.1 0.11 2.2 the hemolysate application on a COBAS INTEGRA 800 analyzer (x).
PreciControl HbA1c path 10.2 0.21 2.0 Sample size (n) = 111
Human sample 1 4.3 0.10 2.3 Mean difference -0.19 % HbA1c
Human sample 2 5.6 0.09 1.6 Lower 95 % confidence interval of -0.52 % HbA1c
Human sample 3 8.2 0.16 1.9 differences
Human sample 4 10.9 0.22 2.0 Upper 95 % confidence interval of 0.14 % HbA1c
differences
Method comparison
The sample concentrations were between 4.6 % and 9.9 % (DCCT/NGSP
Evaluation of method comparison data is according to NGSP certification
criteria. The mean difference between the two methods and the 95 % values).
confidence intervals of the differences in the range from 4‑10 % % HbA1c (DCCT/NGSP) values for human blood samples obtained on a
(DCCT/NGSP) are given. 95 % of the differences between the values Roche/Hitachi cobas c 501 analyzer using the Tina‑quant Hemoglobin A1c
obtained for individual samples with both methods fall within the range Gen.3 reagent with the hemolysate application (y) were compared with
defined by the lower and upper 95 % confidence intervals of the those determined using the Tina‑quant Hemoglobin A1c Gen.2 reagent with
differences. the whole blood application (x).
Whole Blood Application:
Sample size (n) = 84
% HbA1c (DCCT/NGSP) values for human blood samples obtained on a
Roche/Hitachi cobas c 501 analyzer using the Tina‑quant Hemoglobin A1c Mean difference -0.06 % HbA1c
Gen.3 reagent with the whole blood application (y) were compared with
those determined using the corresponding reagent with the hemolysate Lower 95 % confidence interval of -0.53 % HbA1c
application on a COBAS INTEGRA 800 analyzer (x). differences
Upper 95 % confidence interval of 0.41 % HbA1c
Sample size (n) = 80
differences
Mean difference 0.07 % HbA1c
The sample concentrations were between 5.5 % and 9.9 % (DCCT/NGSP
Lower 95 % confidence interval of -0.27 % HbA1c values).
differences
% HbA1c (DCCT/NGSP) values for human blood samples obtained on a
Upper 95 % confidence interval of 0.42 % HbA1c Roche/Hitachi cobas c 501 analyzer using the Tina‑quant Hemoglobin A1c
differences Gen.3 reagent with the hemolysate application (y) were compared with
those determined using the Tina‑quant Hemoglobin A1c Gen.2 reagent with
The sample concentrations were between 4.7 % and 9.8 % (DCCT/NGSP the hemolysate application on a COBAS INTEGRA 800 analyzer (x).
values).
Sample size (n) = 111
% HbA1c (DCCT/NGSP) values for human blood samples obtained on a
Roche/Hitachi cobas c 501 analyzer using the Tina‑quant Hemoglobin A1c Mean difference -0.35 % HbA1c
Gen.3 reagent with the whole blood application (y) were compared with
those determined using the Tina‑quant Hemoglobin A1c Gen.2 reagent with Lower 95 % confidence interval of -0.68 % HbA1c
the whole blood application (x). differences
Upper 95 % confidence interval of -0.02 % HbA1c
Sample size (n) = 82
differences
Mean difference 0.07 % HbA1c
The sample concentrations were between 4.7 % and 9.9 % (DCCT/NGSP
Lower 95 % confidence interval of -0.50 % HbA1c values).
differences
% HbA1c (DCCT/NGSP) values for human blood samples obtained on a
Upper 95 % confidence interval of 0.65 % HbA1c Roche/Hitachi cobas c 501 analyzer using the Tina‑quant Hemoglobin A1c
differences Gen.3 reagent with the hemolysate application (y) were compared with
those determined using the Tina‑quant Hemoglobin A1c Gen.2 reagent with
The sample concentrations were between 5.0 % and 9.9 % (DCCT/NGSP the hemolysate application (x).
values).
A1C-3
Tina-quant Hemoglobin A1c Gen.3 - Hemolysate and Whole Blood Application
Sample size (n) = 113 13 Chang J, Hoke C, Ettinger B, et al. Evaluation and Interference Study
of Hemoglobin A1c Measured by Turbidimetric Inhibition Immunoassay.
Mean difference -0.10 % HbA1c Am J Clin Pathol 1998;109(3):274-278.
Lower 95 % confidence interval of -0.49 % HbA1c 14 Kobold U, Jeppsson JO, Duelffer T, et al. Candidate reference methods
differences for hemoglobin A1c based on peptide mapping. Clin Chem
Upper 95 % confidence interval of 0.31 % HbA1c 1997;43:1944-1951.
differences 15 Jeppsson JO, Kobold U, Finke A, et al. Approved IFCC reference
method for the measurement of HbA1c in human blood. Clin Chem Lab
The sample concentrations were between 4.8 % and 9.7 % (DCCT/NGSP Med 2002;40:78-89.
values).
16 Martina WV, Martijn EG, van der Molen M, et al. β-N-terminal
Analytical specificity for Whole Blood and Hemolysate Application glycohemoglobins in subjects with common hemoglobinopathies:
relation with fructosamine and mean erythrocyte age. Clin Chem
Hb derivatives Labile HbA1c (pre-HbA1c), acetylated Hb, and 1993;39:2259-2265.
carbamylated Hb do not affect the assay results. 17 Weykamp CW, Penders TJ, Muskiet FAJ, et al. Influence of
Hb variants Specimens containing high amounts of HbF hemoglobin variants and derivatives on glycohemoglobin
(> 10 %) may yield lower than expected HbA1c determinations, as investigated by 102 laboratories using 16 methods.
Clin Chem 1993;39:1717-1723.
results.
18 American Diabetes Association. Standards of Medical Care for patients
Please note with diabetes mellitus. Diabetes Care [Suppl.] 1995;18(1):8-15.
According to the consensus statement of the American Diabetes 19 Sacks BW, Bruns DE, Goldstein DE, et al. Guidelines and
Association (ADA), the European Association for the Study of Diabetes recommendations for laboratory analysis in the diagnosis and
(EASD), the International Federation of Clinical Chemistry and Laboratory management of diabetes mellitus. Clin Chem 2002;48:436-472.
Medicine (IFCC) and International Diabetes Federation (IDF) HbA1c results
should be reported in parallel, both in mmol/mol (IFCC) and % 20 Glick MR, Ryder KW, Jackson SA. Graphical Comparisons of
(DCCT/NGSP) values.29 In addition an HbA1c derived estimated average Interferences in Clinical Chemistry Instrumentation.
glucose concentration can be reported which can be calculated according Clin Chem 1986;32:470-475.
to the equations given in the Summary section of this method sheet. Former 21 Miedema K. Influence of hemoglobin variants on the determination of
% HbA1c (IFCC) values must not be used due to the risk of mix up / glycated hemoglobin. Klin Lab 1993;39:1029-1032.
misinterpretation with the % HbA1c (DCCT/NGSP) values.
22 Niederau C, Coe A, Katayama Y. Interference of Non-glucose Adducts
References on the Determination of Glycated Hemoglobins. Klin Lab
1 Goldstein DE, Little RR, Lorenz RA, et al. Tests of glycemia in 1993;39:1015-1023.
diabetes. Diabetes Care 1995;18:896-909. 23 Rohlfing C, Connolly J, England J, et al. Effect of Elevated Fetal
2 Goldstein DE, Little RR. More than you ever wanted to know (but need Hemoglobin on HbA1c Measurements: Four Common Assay Methods
to know) about glycohemoglobin testing. Diabetes Care compared to the IFCC Reference Method. Poster Abstract AACC
1994;17:938-939. Annual Meeting 2006, Chicago. Clin Chem 2006;52(6) Suppl A 108.
3 The Diabetes Control and Complications Trial Research Group. The 24 International Expert Committee Report on the Role of the A1C Assay in
effect of intensive treatment of diabetes on the development and the Diagnosis of Diabetes. Diabetes Care 2009;32(7):1327-1334.
progression of long-term complications in insulin-dependent diabetes 25 Breuer J. Report on the Symposium “Drug effects in Clinical Chemistry
mellitus. N Engl J Med 1993;329:977-986. Methods”. Eur J Clin Chem Clin Biochem 1996;34:385-386.
4 UK Prospective Diabetes Study (UKPDS) group. Intensive blood 26 Sonntag O, Scholer A. Drug interference in clinical chemistry:
glucose control with sulfonylureas or insulin compared with recommendation of drugs and their concentrations to be used in drug
conventional treatment and risk of complications in patients with type 2 interference studies. Ann Clin Biochem 2001;38:376-385.
diabetes (UKPDS 33). Lancet 1998;352:837-853.
27 Junge,W, Wilke B, Halabi A, et al. Determination of reference levels in
5 Finke A, Kobold U, Hoelzel W, et al. Preparation of a candidate primary adults for hemoglobin A1c (HbA1c). Poster presentation
reference material for the international standardization of HbA1c EUROMEDLAB, Barcelona 2003.
determinations. Clin Chem Lab Med 1998;36(5):299-308.
28 Diagnosis and Classification of Diabetes Mellitus. Diabetes Care
6 Goldstein DE, Little RR, Wiedmeyer HM, et al. Glycated hemoglobin: 2010;33(1):62-69.
methodologies and clinical applications. Clin Chem 1986;32:B64-B70.
29 Consensus statement on the worldwide standardization of the
7 Nathan DM, Kuenen J, Borg R, et al. Translating the A1C assay into hemoglobin A1c measurement. American Diabetes Association,
estimated average glucose values. Diabetes Care 2008;31:1473-1478. European Association for the Study of Diabetes, International
8 Bunn HF, Gabbay KH, Gallop PM. The glycosylation of hemoglobin: Federation of Clinical Chemistry and Laboratory Medicine and
relevance to diabetes mellitus. Science 1978;200:21-27. International Diabetes Federation Consensus Committee. Diabetes
Care 2007;30:2399-2400.
9 Zander R, Lang W, Wolf HU. Alkaline haematin D-575, a new tool for
the determination of haemoglobin as an alternative to the A point (period/stop) is always used in this Method Sheet as the decimal
cyanhaemiglobin method. I. Description of the method. Clin Chim Acta separator to mark the border between the integral and the fractional parts of
1984;136:83-93. a decimal numeral. Separators for thousands are not used.
10 Wolf HU, Lang W, Zander R. Alkaline haematin D-575, a new tool for Symbols
the determination of haemoglobin as an alternative to the Roche Diagnostics uses the following symbols and signs in addition to
cyanhaemiglobin method. II. Standardization of the method using pure those listed in the ISO 15223‑1 standard (for USA: see
chlorohaemin. Clin Chim Acta 1984;136:95-104. https://usdiagnostics.roche.com for definition of symbols used):
11 Little RR, Wiedmeyer HM, England JD, et al. Interlaboratory
standardization of measurements of glycohemoglobins. Clin Chem Contents of kit
1992;38:2472-2478. Volume after reconstitution or mixing
12 Frank EL, Moulton L, Little RR, et al. Effects of hemoglobin C and S
traits on seven glycated hemoglobin methods. Clin Chem GTIN Global Trade Item Number
2000;46(6):864-867.
COBAS, COBAS C, COBAS INTEGRA, PRECICONTROL and TINA‑QUANT are trademarks of Roche.
All other product names and trademarks are the property of their respective owners.
A1C-3
Tina-quant Hemoglobin A1c Gen.3 - Hemolysate and Whole Blood Application