Datasheet AMB8059 en

Download as pdf or txt
Download as pdf or txt
You are on page 1of 8

www.narda-sts.

com

NARDA EMF MONITORS

AMB-8059

Continuous, remote ITU-T


K
com .83
plian
monitoring and logging t

of electromagnetic fields

Interchangeable probes from 10 Hz to 40 GHz for


low frequency & high frequency applications
Multi-band probes for telecommunications
monitoring
Simultaneous monitoring of electric and
magnetic fields

Fully autonomous operation:


• Solar panel power supply
• Built-in 3G modem
• Built-in Wi-Fi
• Automatic data transfer
• Daily reports, warnings & alarm messages via SMS
• On-board GPS

Easy integration into test environments and


Web Based Applications
Low weight, robust design, compact size for
indoor and outdoor operations

Drive test capability of AMB-8059/00 model


according to ITU-K.113

Area Monitor AMB-8059/03


with Solar Panel

INTRODUCTION
Narda EMF Monitors are equipped with exclusive, state-of-the-art sensors having high sensitivity, accuracy
and reliability. Their robust, uncluttered construction is perfect for long-term outdoor installation. The
AMB-8059 handles applications from a few Hertz through to long wave and on up to high frequency
microwave radiation using a selection of interchangeable probes for electric and magnetic fields.

AMB8059_BEN_71201 1/8 Subject to change without notice


www.narda-sts.com

MINIMUM OUTLAY, MAXIMUM RESULT


An EMF monitoring system is made up from a series of EMF monitors installed
wherever the EMF presence needs to be assessed continuously or by long
term observation. The EMF monitors store the data and report them using
conventional mobile data communication at set time intervals to a central unit,
e.g. PC or data server. The system size can range from a single location up to
countrywide coverage. Narda EMF monitors combine all the features that are
essential for this purpose: autonomy, outdoor usability, mobility, robustness,
and low operating costs.
You can be certain to find the ideal solution for every area of application
with Narda. And you can depend on its reliability, thanks to our decades of
experience coupled with cutting edge technology, backed up by our own
certified calibration laboratory.

THE AMB SERIES


Its broadband application is the optimum solution for technical
superiority from a tight budget. Four models are available:

Unit designation AMB-8059/03 AMB-8059/02 AMB-8059/01 AMB-8059/00


Solar panel (24/7) &
back-up battery  
Internal 3G modem  
Wi-Fi    
Ethernet port  
USB   
RS232  
Optical link   
GPS sensor    
Battery life 6 – 12 months
(Li-Ion)  
Remote capabilities    
Permanent outdoor
installation ○ ○
Indoor installation ○ ○
Drive test measurement ●
(1)

(1) Car mounting kit required, see optional accessories


○ Suggested ● Necessary

AMB8059_BEN_71201 2/8 Subject to change without notice


www.narda-sts.com

COMPLETE PROGRAM FOR ALL REQUIREMENTS

Narda offers a wide range of different isotropic probes. These include


quad-band probes for separating mobile telephone services as well as
wideband measurement from 100 kHz to 40 GHz. Special probes are
available for low frequency magnetic or electric fields from 10 Hz to 5
kHz. This means that emissions from high-voltage cables and transformer
stations can be recorded. Further, it is possible to combine up to two
probes, e.g. an electric and a magnetic field probe in the so-called “dual
probe configuration”.

Magnetic field probe Electric field probe


Dual probe configuration
(without radome)
THE APPLICATIONS - NARDA AREA MONITOR PROBES
100 kHz 100 kHz 10 Hz 300 kHz 300 kHz 0.1 to 3000 MHz 10 Hz
0.1-3000 MHz 0.1-7000 MHz
Frequency range to to to to to 0.1to 862 MHz to
GSM, UMTS GSM, UMTS
3 GHz 7 GHz 5 kHz 18 GHz 40 GHz 933-3000 MHz 5 kHz
Field type
E E E E E E E E H
(isotropic sensors)
Band type Single Single Single Single Single Tri Quad Quad Single
Probe designation EP-1B-01 EP-1B-03 EP-1B-04 EP-1B-05 EP-1B-06 EP-3B-01 EP-4B-01 EP-4B-02 HP-1B-01
Mobile communications ● ● ● ● ● ● ●
Radio / TV broadcasting ● ● ● ● ● ● ●
Directional radio ○ ● ● ○ ○ ○
Satellite
communications ● ●
Radar ● ●
Industry ● ● ● ● ● ● ●
Railroads ● ●
Power lines ● ●
Transformers ● ●
● more important ○ variable importance

SIMULTANEOUS MONITORING OF ELECTRIC AND MAGNETIC FIELDS


Possible dual probe configuration:
HP-1B-01 HP-1B-01 HP-1B-01 HP-1B-01 HP-1B-01
Probe combination + + + + +
EP-1B-01 EP-1B-03 EP-1B-04 EP-1B-05 EP-1B-06
H 10 Hz to 5 kHz 10 Hz to 5 kHz 10 Hz to 5 kHz 10 Hz to 5 kHz 10 Hz to 5 kHz
Frequency range / field type
E 100 kHz to 3 GHz 100 kHz to 7 GHz 10 Hz to 5 kHz 300 kHz to 18 GHz 300 kHz to 40 GHz

AMB8059_BEN_71201 3/8 Subject to change without notice


www.narda-sts.com

SPECIFICATIONS

EP-1B-01 Electric Field Probe*


Field Probe* Frequency range 100 kHz to 3 GHz
Measurement range 0.2 to 200 V/m (dynamic range > 60 dB)
Measurement resolution 0.01 V/m
Overload 600 V/m
Flatness @ 20 V/m 1 to 200 MHz ± 0.8 dB; 150 kHz to 3 GHz ± 1.5 dB
Linearity ± 0.5 dB (0.5 to 100 V/m)
Anisotropy @ 6 V/m ± 0.8 dB @ 50 MHz (typical 0.6 dB)
H-Field rejection > 20 dB
Size and weight 450 mm length, 55 mm Ø, 180 g

EP-1B-03 Electric Field Probe*


Field Probe* Frequency range 100 kHz to 7 GHz
Measurement range 0.2 V/m to 200 V/m (dynamic range > 60 dB)
Measurement resolution 0.01 V/m
Overload 600 V/m
Flatness @ 20 V/m 3 MHz to 200 MHz: ±0.8 dB; 0.15 MHz to 3 GHz: ±1.5 dB ; 0.1 MHz to 6 GHz: ±2 dB
Linearity ± 0.5 dB (0.5 to 100 V/m)
Anisotropy @ 6 V/m ± 0.8 dB @ 50 MHz (typical 0.6 dB)
H-Field rejection > 20 dB
Size and weight 450 mm x 55 mm Ø, 180 g

EP-1B-04 Electric Field Probe*


Frequency range 10 Hz to 5 kHz
Measurement range 5 V/m to 20 kV/m (dynamic range > 72 dB)
Measurement resolution 0.1 V/m
Overload > 30 kV/m
Flatness @ 100 V/m (40 Hz – 1 kHz) 1 dB (typical 0.5 dB)
Anisotropy @ 100 V/m 0.5 dB @ 50 Hz
H-Field rejection > 20 dB
Size and weight 77 mm x 53 mm Ø, 110 g

EP-1B-05 Electric Field Probe*


Frequency range 300 kHz to 18 GHz
Measurement range 0.5 V/m to 800 V/m (dynamic range > 64 dB)
Measurement resolution 0.01 V/m
Overload 1200 V/m
Flatness @ 6 V/m 1 MHz to 1 GHz ± 1.5 dB; 1 GHz to 12 GHz ± 3.0 dB; 12 GHz to 18 GHz ± 4.0 dB
Linearity ± 0.5 dB (± 0.3 typical) (1.2 V/m to 200 V/m) @ 200 MHz
Anisotropy @ 200 MHz ±0.8 dB (typical 0.5 dB @ 930 and 1800 MHz)
H field rejection > 20 dB
Size and weight 450 mm x 55 mm Ø, 180 g

AMB8059_BEN_71201 4/8 Subject to change without notice


www.narda-sts.com

EP-1B-06 Electric Field Probe*


Frequency range 300 kHz to 40 GHz
Measurement range 0.5 V/m to 800 V/m (dynamic range > 64 dB)
Measurement resolution 0.01 V/m
Overload 1200 V/m
1 MHz to 1 GHz ±1.5 dB
1 GHz to 12 GHz ±3.0 dB
Flatness @ 6 V/m
12 GHz to 23 GHz ±4.0 dB
23 GHz to 40 GHz ±5.0 dB
Linearity ± 0.5 dB (± 0.3 typical) (1.2 V/m to 200 V/m) @ 200 MHz
Anisotropy @ 200 MHz ± 0.8 dB (typical 0.5 dB @ 930 and 1800 MHz)
H field rejection > 20 dB
Size and weight 450 mm x 55 mm Ø, 180 g

EP-3B-01 Tri-Band Electric Field Probe*


Frequency range Wideband: 0.1 to 3000 MHz Low pass: 0.1 to 862 MHz High pass: 933 to 3000 MHz
Measurement resolution 0.01 V/m
Measurement range 0.2 to 200 V/m (dynamic range > 60 dB)
Overload 600 V/m
1 to 200 MHz ± 0.8 dB 1 to 200 MHz ± 0.8 dB
Flatness @ 20 V/m 933 to 3000 MHz ± 1.5dB
150 kHz to 3 GHz ±1.5 dB 150 kHz to 862 MHz ± 1.5 dB
Linearity ± 0.5 dB (0.5 to 100 V/m)
Anisotropy @ 6 V/m ± 0.8 dB @ 50 MHz (typical 0.6 dB) ± 0.8dB @1 GHz(typical 0.6 dB)
933 MHz to 3 GHz > 23 dB 0,1 to 862 MHz > 23 dB
Out of band attenuation Not applicable
(ref. to 50 MHz) (ref. to 1 GHz)
H field rejection > 20 dB
Size and weight 450 mm x 55 mm Ø, 180 g

EP-4B-01 Quad-Band Electric Field Probe*


Wideband EGSM 900 EGSM 1800 UMTS
Frequency range
0.1 to 3000 MHz 925 to 960 MHz 1805 to 1880 MHz 2110 to 2170 MHz
Meas. range 0.2 to 200 V/m 0.03 to 30 V/m 0.03 to 30 V/m 0.03 to 30 V/m
Meas. resolution 0.01 V/m
CW damage level 300 V/m
1 to 200 MHz ± 0.8 dB 925 to 960 MHz 1805 to 1880 MHz 2110 to 2170 MHz
Flatness @ 6 V/m
150 kHz to 3 GHz ± 1.5 dB +0.5/-2.5 dB +0.5/-2.5 dB +0.5/-2.5 dB
Linearity ± 0.5 dB (0.5 to 100 V/m) ± 0.5 dB (0.06 to 20 V/m) ± 0.5 dB (0.06 to 20 V/m) ± 0.5 dB (0.06 to 20 V/m)
± 0.8 dB @ 50 MHz, ± 0.8 dB@ 942.5 MHz, ± 0.8 dB@ 1842.5 MHz, ± 0.8 dB@ 2140 MHz,
Anisotropy
3 V/m (typical 0.6 dB) 3 V/m (typical 0.6 dB) 3 V/m (typical 0.6 dB) 3 V/m (typical 0.6 dB)
Rejection Rejection Rejection
Out of band attenuation Not applicable to 1842 MHz(GSM): 25 dB to 942 MHz(GSM): 15 dB to 942 MHz(GSM): 17dB
to 2140 MHz(UMTS): 25 dB to 2140 MHz(UMTS): 13 dB to 1842 MHz(GSM): 10 dB
40 °C – 50 °C = ± 100kHz
Centre frequency drift Not applicable
-20 °C – 40 °C = ± 100 kHz/°C
H field rejection > 20 dB
Size and weight 450 mm x 55 mm Ø, 210 g

(*) All probes include on board A/D conversion, calibration factors on E2PROM, and temperature sensor

AMB8059_BEN_71201 5/8 Subject to change without notice


www.narda-sts.com

EP-4B-02 Quad-Band Electric Field Probe*


Wideband EGSM 900 EGSM 1800 UMTS
Frequency range
0.1 to 7000 MHz 925 to 960 MHz 1805 to 1880 MHz 2110 to 2170 MHz
Meas. range 0.2 to 200 V/m 0.03 to 30 V/m 0.03 to 30 V/m 0.03 to 30 V/m
Meas. resolution 0.01 V/m
Dynamic range >60 dB
3 to 200 MHz ± 1.5 dB
925 to 960 MHz 1805 to 1880 MHz 2110 to 2170 MHz
Flatness @ 6 V/m 150 kHz to 3 GHz ± 2 dB
+0.5 / -2.5 dB +0.5 / -2.5 dB +0.5 / -2.5 dB
0.1 MHz to 7 GHz ± 3 dB
Linearity ± 0.5 dB (0.5 to 100 V/m) ± 0.5 dB (0.1 to 20 V/m) ± 0.5 dB (0.1 to 20 V/m) ± 0.5 dB (0.1 to 20 V/m)
± 0.8 dB@ 50 MHz, ± 0.8 dB@ 942.5 MHz, ± 0.8 dB@ 1842.5 MHz, ± 0.8 dB@ 2140 MHz,
Anisotropy
3 V/m (typical 0.6 dB) 3 V/m (typical 0.6 dB) 3 V/m (typical 0.6 dB) 3 V/m (typical 0.6 dB)
Rejection Rejection Rejection
Out of band attenuation Not applicable to 1842 MHz(GSM): 25 dB to 942 MHz(GSM): 15 dB to 942 MHz(GSM): 17dB
to 2140 MHz(UMTS): 25 dB to 2140 MHz(UMTS): 13 dB to 1842 MHz(GSM): 10 dB
40 °C – 60 °C = ± 100 kHz
Centre frequency drift Not applicable
-20 °C – 40 °C = - 100 kHz / °C
H field rejection > 20 dB
Size and weight 450 mm x 55 mm Ø, 210 g

HP-1B-01 Magnetic Field Probe*


Frequency range 10 Hz to 5 kHz
Measurement range and overload 50 nT to 200 μT (dynamic range >72 dB); overload: > 1 mT
Measurement resolution 1 nT
Flatness 40 Hz to 1 kHz, 1 dB (typical 0.6 dB)
Linearity ± 0.5 dB (200 nT to 100 μT)
Anisotropy 0.3 dB @ 50 Hz, 3 μT
E field rejection > 20 dB
Size and weight 83 mm x 53 mm Ø, 110 g

(*) All probes include on board A/D conversion, calibration factors on E2PROM, and temperature sensor

AMB-8059 Multi-band EMF Area Monitor


Technical Specifications
Frequency range Depending on probe (see probe specifications)
Dynamic range Depending on probe (see probe specifications)
Resolution Depending on probe (see probe specifications)
Sensitivity Depending on probe (see probe specifications)
Linearity Depending on probe (see probe specifications)
Accuracy Depending on probe (see probe specifications)
Measurement Units V/m, kV/m, nT, μT, mT. The unit shown depends on the probe connected
Field measured Total field, average and Peak (MAX)
Sampling 1 measurement every 1 s

AMB8059_BEN_71201 6/8 Subject to change without notice


www.narda-sts.com

Measurement / acquisition functions


Memorization interval Programmable from 30 seconds to 15 minutes
Memory Over 128 MB
Max data storage capacity
Over 364 days with 1 acquisition every minute
(before old data are replaced by new ones)
Manual
Automatic managed by the unit at predefined timings (1), (3)
Data download
Automatic by PC (2), (3)
Automatic creation of a .TXT and .BMP file after download
AVG, RMS, maximum peak; daily report via SMS (3)
Functions
Display and marking of data acquired during modem transmission (3)
Two programmable field strength thresholds (warning and alarm) with automatic notice both of
Field strength alarm
exceeding the limit and returning within the limits (3)
Clock Real time internal clock
Messages SMS which can be sent to up to 10 mobile phones simultaneously (3)
Sensor Display of model and calibration date
Battery management Every record includes Battery Voltage and Charge Current value
Temperature management Every record includes Internal Temperature value
Humidity management Every record includes Internal Humidity value
GPS coordinates Programmable record
General Specifications
GSM module Penta-band (800, 850, 900, 1900, 2100 MHz) UMTS/3G (3)
Wi-Fi Wi-Fi 802.11 b/g
Field probes Interchangeable, several models available, single and dual probe operation
Interfaces RS232, USB, Ethernet, Micro SD Card, Wi-Fi, Optical (6), 3G modem (3)
Protection Sensor to notify case opening
Protective case opening, internal overheat, internal humidity, low battery, battery overload (model
Other alarms
AMB-8059/01 and AMB-8059/03 only), probe malfunction, field over limit.
AMB-8059/00 - AMB-8059/02: Non rechargeable primary battery, lithium SAFT LSH20 3.6 V,
Internal battery
13 A/h AMB-8059/01 - AMB-8059/03: Lead, 4 V, 2.5 A/h, rechargeable
1 mA with 3G and Wi-Fi module off
500 mA max when 3G module is transmitting and Wi-Fi module off (3)
Consumption
120 mA max when Wi-Fi module is transmitting and 3G off (3)
6 mA optical link data query every 1 second; Wi-Fi and 3G off
External power DC, 5 V, 1 A max (AMB-8059/01 and AMB-8059/03 only)
AMB-8059/02: about 8 months @ 1min GSM module transmission per day and single probe operat-
ing mode (autonomy depends on probe and setting) (4), (5)
Operating time @ 1 sec. rate AMB-8059/03: > 80 days in total darkness @ 1min GSM module transmission per day and single
probe operating mode (autonomy depends on probe and setting) (4), (5)
For best performance install solar panels in direct sunlight.
Recharging time 24 hours with external power unit (AMB-8059/01 and AMB-8059/03 only)
Auto test Automatic
Compliance European directives 89/336 and 73/23, CEI 211-6, CEI 211-7, ITU-T K.83, ITU-T K.113(7)
Ambient temperature -20 °C / +55 °C
Dimensions (WxDxH) 112 x 112 x 730 mm
AMB-8059/00 and AMB-8059/02: 1.2 kg (unit only); 6.5 kg (total weight including supports and
Weight base) AMB-8059/01 and AMB-8059/03: 2.4 kg (unit only); 7.7 kg (total weight including supports
and base)
Environmental protection IP55
Country of origin Italy

Notes:
(1): To the controller PC or to the user’s FTP server depending on the preferred communication mode.
(2): Directly from the station or from the user’s FTP server depending on the preferred communication mode.
(3): AMB-8059/02 and AMB-8059/03 only are equipped with 3G modem.
(4): AMB-8059/00 and AMB-8059/01 power autonomy is longer as they are not equipped with 3G modem module.
(5): Specifications depending on battery age, ambient temperature and GSM field coverage.
(6): Optical link only for the model AMB-8059/03, AMB-8059/01, AMB-8059/00
(7): Only AMB-8059/00 with option car mounting kit for drive test solution

AMB8059_BEN_71201 7/8 Subject to change without notice


www.narda-sts.com

ORDERING INFORMATION

AMB-8059
Remote stations
Area Monitor station powered by internal primary Li-Ion battery AMB-8059/00
Area Monitor station powered by solar panel and back-up battery AMB-8059/01
Area Monitor remote station with 3G internal modem, powered by internal primary Li-Ion battery AMB-8059/02
Area Monitor remote station with 3G internal modem, powered by solar panel and back up battery AMB-8059/03
Field probes
Electric field probe 0.1 to 3000 MHz; 0.2 to 200 V/m EP-1B-01
Electric field probe 0.1 to 7000 MHz; 0.2 to 200 V/m EP-1B-03
Electric field probe 10 Hz to 5 kHz; 5 V/m to 20 kV/m EP-1B-04
Electric field probe 300 kHz to 18 GHz; 0.5 V/m to 800 V/m EP-1B-05
Electric field probe 300 kHz to 40 GHz; 0.5 V/m to 800 V/m EP-1B-06
Tri-band electric field probe 0.1 to 3000 MHz / 0.1 to 862 MHz / 933 to 3000 MHz; 0.2 to 200 V/m EP-3B-01
Quad-band electric field probe 0.1 to 3000 MHz; 0.2 to 200 V/m / 925 to 960 MHz /
EP-4B-01
1805 to 1880 MHz / 2110 to 2170 MHz, 0.03 to 30 V/m
Quad-band electric field probe 0.1 to 7000 MHz; 0.2 to 200 V/m / 925 to 960 MHz /
EP-4B-02
1805 to 1880 MHz / 2110 to 2170 MHz, 0.03 to 30 V/m
Magnetic field probe 10 Hz to 5 kHz; 50 nT to 200 μT HP-1B-01
Optional accessories
8059/mast - Metallic T-shaped base and Fiberglass mast (includes kit of screws, ties and 3 ballast bags) 650.800.085
8059/CMK - Car Mounting Kit for drive test solution (AMB-8059/00 only) 650.800.300
8059/CE - Cover extension for AMB-8059 dual probe configuration 231.830.084
O/E optical converter USB 650.000.176
Cable, FO Duplex RP-02 with cable clamp, 10 m (only models AMB-8059/03 and AMB-8059/01 with optical link) 650.000.289
Cable, FO Duplex RP-02 with cable clamp, 20 m (only models AMB-8059/03 and AMB-8059/01 with optical link) 650.000.290
Cable, FO Duplex RP-02 with cable clamp, 40 m (only models AMB-8059/03 and AMB-8059/01 with optical link) 650.000.291
Cable, FO Duplex RP-02, 10 m (only models AMB-8059/00 with optical link) 650.000.196
Cable, FO Duplex RP-02, 20 m (only models AMB-8059/00 with optical link) 650.000.257
Cable, FO Duplex RP-02, 40 m (only models AMB-8059/00 with optical link) 650.000.275
Includes in delivery
• Primary Li-ion battery (AMB-8059/00 and AMB-8059/02 only)
• Power supply / Battery Charger (AMB-8059/01 and AMB-8059/03 only)
• Assembled Solar Unit (AMB-8059/01 and AMB-8059/03 only)
• Ethernet cable (AMB-8059/01 and AMB-8059/03 only)
• 10 m optical cable and O/E converter USB (AMB-8059/00 only)
• USB cable, 1.8 m, USB(A)/USB(B)
• Swivel joint for installation on AMB-8059-MAST
• Operating Manual, Test & Calibration Certificates
• PC Software 8059-NSTS
• PC Software EMF GPS logger (used only by model AMB-8059/00 Car Mounting Kit option)

Narda Safety Test Solutions GmbH Narda Safety Test Solutions Srl Narda Safety Test Solutions
Sandwiesenstrasse 7 Via Leonardo da Vinci, 21/23 435 Moreland Road
72793 Pfullingen, Germany 20090 Segrate (Milano) - Italy Hauppauge, NY 11788, USA
Phone: +49 7121 9732-0 Phone: +39 02 26 998 71 Phone: +1 631 231-1700
Fax: +49 7121 9732-790 Fax: +39 02 26 998 700 Fax: +1 631 231-1711
[email protected] [email protected] [email protected]
www.narda-sts.com www.narda-sts.it www.narda-sts.com

® Names and Logo are registered trademarks of Narda Safety Test Solutions GmbH and L3 Communications Holdings Inc. – Trade names are trademarks of the owners.

AMB8059_BEN_71201 8/8 Subject to change without notice