Portable Power Factor Meters: Features
Portable Power Factor Meters: Features
2039
Model 2039 is a revolutionary series of portable wattmeters in
which a power factor is used to operate the DC indicator. This is
done using a transducer which provides a DC current
proportionate to the voltage-current phase by means of a circuit.
Model 2039 can be used in single-phase and balanced three-
phase circuits. The electronic transducer, combined with a high
sensitivity moving coil type indicator based on YOKOGAWA's
proprietary taut-band suspension system, provides a level of
performance not possible with conventional power factor meters.
Features
● For both single-phase and three-phase (balanced circuit).
● Excellent current characteristic: 20~200% of rated current (short
time period)
● Wide range of applicable voltages: 60~300 V AC
● Phase angle scale included
● Taut-band suspension system eliminates friction and provides
strong resistance to shock impact.
2039 02
Specifications
Operating principle : Rectifier type Insulation test :
Class : JIS C 1102, equivalent to Class 3.0 Between electrical circuit and the case DC 500V/More than 10MΩ
Operating position : Horizontal Between current circuit and voltage circuit DC 500V/More than 5MΩ
Rated frequency : 50/60 Hz Voltage test :
Scale length : Approximately 135 mm (deflection angle: 85°) Between electrical circuit and the case AC 2000 V for 5 seconds
Scale : Lead 0-0.3 to 1.0 to 0.3-0 lag (with phase angle scale) Between current circuit and voltage circuit AC 1500 V for 5 seconds
Effective measurement range: Lead 0.5 to 1.0 to 0.5 lag External dimensions and weight :
Consumed power Voltage circuit (120 V) : Approximately 0.14 VA Approximately 260 × 180 × 141 mm, approximately 2.9 kg
Current circuit (5 A) : Approximately 2.4 VA Standard accessory : Instruction Manual (1)
Operating temperature and humidity ranges : 0 to 40°C, 25 to 80%RH Optional accessories (sold separately) : 2292 01 Carrying case (page 10)
Storage temperature and humidity ranges : -10 to 50°C, 25 to 80%RH
Notes
1. For measurements exceeding the rated current (25 A), externally connect a current transformer
Model Rated current Rated voltage 2241-2243 (page 9) to the 5 A terminal.
01 0.2/1 A 120 V 2. For measurement with a voltage exceeding 300 V, externally connect a voltage transformer 2261
2039 02 1/5 A (Can be used between or 2262 to the 120 V terminal. (page 9)
03 5/25 A 60 and 300 V.) 3. 2039 is not JIS-approved.
Characteristics
Voltage characteristic Current characteristic
Lead Lead
5 A/120 V 5 A/120 V cos∅= 1.0 Lag cos∅= 0.5
1° 1°
cos∅= 1.0 Lead cos∅= 0.5
0 0
Lead cos∅= 0.5
1° 1°
Lag cos∅= 0.5
Lag Lag
60 120 180 240 300V 60 2 4 5A 6 8 10 12A
• JIS specification: Error within 100% of class index within ±15% of rated voltage. • JIS specification: Error within 100% of class index within 20% to 120% of rated current.
Connection diagrams
Single-phase circuit Balanced three-phase circuit
Source Load
R
Source Load S
T
± 1A 5A P2 ± 1A 5A P2
P1 P3 P1 P3
COS COS