Biomedical Instrumentation: Reference
Biomedical Instrumentation: Reference
1 3
Reference
BIOMEDICAL
Khandpur R.S, Handbook of Bio-Medical
INSTRUMENTATION Instrumentation, 3rd Edition, McGraw Hill
Education 2014.
Arumugam,M. Biomedical Instrumentation.
Anuradha Agencies, 2nd Edition, 2002.
Richard C. Fries. Handbook of Medical
Device Design. Marcel Dekker Inc.
Publications, 2001.
Dr. K. Adalarasu
Email id : [email protected]
KA – BEI – Unit II – July, 2018, Sastra University KA – BEI – Unit II – July, 2018, Sastra University
2 4
1
7/25/2018
5 7
Instrumentation System
Detected quantity may be converted to a
mechanical or electrical form of energy
KA – BEI – Unit II – July, 2018, Sastra University KA – BEI – Unit II – July, 2018, Sastra University
6 8
KA – BEI – Unit II – July, 2018, Sastra University KA – BEI – Unit II – July, 2018, Sastra University
2
7/25/2018
9 11
KA – BEI – Unit II – July, 2018, Sastra University KA – BEI – Unit II – July, 2018, Sastra University
10 12
3
7/25/2018
13 15
KA – BEI – Unit II – July, 2018, Sastra University KA – BEI – Unit II – July, 2018, Sastra University
14
Components of Biomedical System
KA – BEI – Unit II – July, 2018, Sastra University KA – BEI – Unit II – July, 2018, Sastra University
4
7/25/2018
17 19
KA – BEI – Unit II – July, 2018, Sastra University KA – BEI – Unit II – July, 2018, Sastra University
18 20
KA – BEI – Unit II – July, 2018, Sastra University KA – BEI – Unit II – July, 2018, Sastra University
5
7/25/2018
21 23
Electrode-Electrolyte Interface Half-Cell Potential
Oxidation reaction causes atom to lose Half-cell potentials can be important when using electrodes
for low frequency or dc measurements
electron
Reduction reaction causes atom to gain
electron
Oxidation is dominant
When the current flow is from electrode to
electrolyte
Reduction dominate
When the current flow is in the opposite.
Half-cell Potentials for Materials and Reactions Encountered in
Biopotential Measurement
KA – BEI – Unit II – July, 2018, Sastra University KA – BEI – Unit II – July, 2018, Sastra University
22 24
Half-Cell Potential Half-Cell Potential
When the metal comes in contact with Electrochemists have adopted the Half Cell
solution potential for hydrogen electrode to be zero.
Electrolyte surrounding the metal is at different Half-Cell potential for any metal electrode is
electric potential from rest of the solution measured with respect to the hydrogen
A second electrode is required to find half cell electrode
potential- hydrogen Reason for Half Cell Potential
Half-Cell potential is determined by Charge Separation at Interface
Oxidation or reduction reactions at the electrode-
Metal involved electrolyte interface lead to a double-charge layer
Concentration of its ion in solution Similar to that which exists along electrically active
Temperature biological cell membranes
KA – BEI – Unit II – July, 2018, Sastra University KA – BEI – Unit II – July, 2018, Sastra University
6
7/25/2018
25 27
KA – BEI – Unit II – July, 2018, Sastra University KA – BEI – Unit II – July, 2018, Sastra University
26 28
KA – BEI – Unit II – July, 2018, Sastra University KA – BEI – Unit II – July, 2018, Sastra University
7
7/25/2018
29 31
the skin Rs
Sweat glands
and ducts
Gel
100 m
Ese EP
Stratum Corneum
Epidermis Ce Re CP RP
KA – BEI – Unit II – July, 2018, Sastra University KA – BEI – Unit II – July, 2018, Sastra University
Electrodes
Use for
Polarizable Electrodes recording
No charge crosses the electrode-electrolyte interface when
a current is applied.
8
7/25/2018
33 35
Properties of Bioelectrodes Bio Potential Electrodes
Good conductors Micro Electrodes
Bio electric potential near or within a single cell
Low impedance
Metal Type—Tip must be tungsten or stainless steel
They should not polarize when a current flows
Micro pipette---It is a glass micropipet with size of 1
through them micron, It is filled with electrolyte
They should establish a good contact with the body
and not cause motion.
Skin surface electrode
They should not cause itching swelling or discomfort Measure ECG,EEG,EMG
to the patient
The metal should not be toxic
Needle electrode
Mechanically rugged
Penetrate the skin to record EEG
Easy to clean.
KA – BEI – Unit II – July, 2018, Sastra University KA – BEI – Unit II – July, 2018, Sastra University
36
Unipolar and Bipolar Electrode
For noninvasive recordings
Proper skin preparation
Which normally involves cleansing the skin with
alcohol or the application of a small amount of
an electrolyte paste
Electrodes Helps to minimize the impedance of the skin–
electrode interface and improve the quality of the
recorded signal considerably
KA – BEI – Unit II – July, 2018, Sastra University KA – BEI – Unit II – July, 2018, Sastra University
9
7/25/2018
37 39
Unipolar and Bipolar Electrode Microelectrodes
Most common type of needle electrode is the Used to measure bio-potential signals at the
concentric bipolar electrode
cellular level
Measure potential difference across cell
membrane
Due to small dimensions (mm), impedance
levels are high
So amplifier needs very high input
impedance
Intracellular
Extracellular
KA – BEI – Unit II – July, 2018, Sastra University KA – BEI – Unit II – July, 2018, Sastra University
Microelectrodes
Bipolar electrode Requirements
Made from thin metallic wires encased inside a larger Small enough to be placed into cell
canula or hypodermic needle
Strong enough to penetrate cell membrane
Two wires serve as the recording and reference
electrodes Typical tip diameter: 0.05 – 10 microns
Unipolar needle electrode Types
Made of a thin wire that is mostly insulated by a thin layer Solid metal -> Tungsten microelectrodes
of Teflon except about 0.3 mm near the distal tip
Supported metal (metal contained within/outside
This electrode requires a second unipolar reference
glass needle)
electrode to form a closed electrical circuit
Glass micropipette -> with Ag-AgCl electrode
metal
KA – BEI – Unit II – July, 2018, Sastra University KA – BEI – Unit II – July, 2018, Sastra University
10
7/25/2018
41 43
42 44
Microns!
KA – BEI – Unit II – July, 2018, Sastra University KA – BEI – Unit II – July, 2018, Sastra University
11
7/25/2018
45 47
46 48
Fill with
intracellular fluid
or 3M KCl
KA – BEI – Unit II – July, 2018, Sastra University KA – BEI – Unit II – July, 2018, Sastra University
12
7/25/2018
49 51
KA – BEI – Unit II – July, 2018, Sastra University KA – BEI – Unit II – July, 2018, Sastra University
50 52
Electrical Properties of Microelectrodes Electrical Properties of Microelectrodes
Metal microelectrode Glass Micropipette Microelectrode
Frequency dependent impedance: 10 – 100 MΩ
High-pass filtering effect
Good for measuring action potentials
Glass micropipet microelectrode
Frequency dependent impedance: 1 – 100 MΩ
Low-pass filtering effect
Good for measuring resting membrane potential
KA – BEI – Unit II – July, 2018, Sastra University KA – BEI – Unit II – July, 2018, Sastra University
13
7/25/2018
55
Metal Plate Electrodes
Large surface: Ancient, therefore still used,
ECG
Metal disk with stainless steel; platinum or
Body-Surface Recording gold coated
EMG, EEG
Electrodes Smaller diameters
Motion artifacts
Disposable foam-pad: Cheap!
KA – BEI – Unit II – July, 2018, Sastra University KA – BEI – Unit II – July, 2018, Sastra University
54 56
Electrode metal
Electrolyte
14
7/25/2018
57 59
Body-Surface Recording Electrodes Floating electrodes
Suction electrodes Insulating
Metal disk
package
No straps or
adhesives required Double-sided
Adhesive-tape
Precordial (chest) ring Electrolyte gel
in recess
ECG (a) (b)
Reusable
Can only be used Snap coated with Ag-AgCl External snap
Gel-coated sponge
for short periods Plastic cup Plastic disk Disposable
KA – BEI – Unit II – July, 2018, Sastra University KA – BEI – Unit II – July, 2018, Sastra University
58 60
15
7/25/2018
61 63
Flexible Electrodes Internal Electrodes
No electrode gel is used and the interface is the
electrode-electrolyte interface
Percutaneous electrode
Electrode or lead wire crosses the skin
Needle electrode: insulated needle electrode, coaxial
needle electrode, bipolar coaxial electrode
(a) Carbon-filled silicone Wire electrode: fine-wire electrode, coiled fine-wire
rubber electrode. electrode
(b) Flexible thin-film neonatal
electrode. EMG, ECG during surgery, fetal ECG (suction electrode,
(c) Cross-sectional view of helical electrode)
the thin-film electrode in (b).
KA – BEI – Unit II – July, 2018, Sastra University KA – BEI – Unit II – July, 2018, Sastra University
64
Internal Electrodes
Internal electrode
Implanted with radio-telemetry connection
KA – BEI – Unit II – July, 2018, Sastra University KA – BEI – Unit II – July, 2018, Sastra University
16
7/25/2018
65 67
Needle and Wire Electrode Needle and Wire Electrode
KA – BEI – Unit II – July, 2018, Sastra University KA – BEI – Unit II – July, 2018, Sastra University
66
Fetal ECG Electrodes
Amplifiers
17
7/25/2018
Amplifiers
69 71
Preamplifiers
Signals available from the transducers are Amplifiers employed
often very small in magnitude AC/DC universal amplifier with special features
such as
Amplifiers boost the level of the input signal to
Capacity neutralization
match the requirements of the
Current injection
recording/display system or Low leakage current
To match the range of the analog-to-digital Low dc drift suitable for intracellular
convertor measurements through high resistance fluid-filled
Thus increasing the resolution and sensitivity electrodes
of the measurement or to make extracellular recordings through metal
microelectrodes for EMG, EEG, EOG, etc
KA – BEI – Unit II – July, 2018, Sastra University KA – BEI – Unit II – July, 2018, Sastra University
70 72
Preamplifiers Preamplifiers
Preamplifiers offer a wide range of input ECG amplifier with full 12 lead selection
sensitivities to cover virtually all signal and patient isolation
sources Transducer amplifier suited for bridge
Calibrated zero suppression to expand measurements on strain gauges
desired portions of an input signal Strain gauge based blood pressure transducers
Selectable low pass filtering facilities Force transducers
To reject noise or unwanted signal components Resistance temperature devices
are available on these amplifiers Direct low level dc input signals
KA – BEI – Unit II – July, 2018, Sastra University KA – BEI – Unit II – July, 2018, Sastra University
18
7/25/2018
73 75
Preamplifiers Preamplifiers
DC amplifier used in conjunction with standard Chopper input dc amplifiers
thermistor probes for the accurate measurement Chopper stabilized dc amplifiers
of temperature within the range of medical DC bridge amplifiers
applications
KA – BEI – Unit II – July, 2018, Sastra University KA – BEI – Unit II – July, 2018, Sastra University
74 76
Preamplifiers Differential Amplifier
Differential Amplifier Three input terminals out
Which will reject any common mode signal that appears One is arranged at the reference potential
simultaneously at both amplifier input terminals Other two are live terminals
Amplifies only the voltage difference that appears across Differential amplifier is employed
its input terminals When it is necessary to measure the voltage difference
Most of the amplifiers used for measuring bioelectric between two points
signals are of the differential type Both of them varying in amplitude at different rates and in
Ac coupled amplifiers different patterns
Used only for special medical applications such as Heart-generated voltages picked
electrocardiograph machine Electrodes on the arms and legs
Carrier amplifiers Brain-generated voltages picked
DC amplifiers Electrodes on the scalp
KA – BEI – Unit II – July, 2018, Sastra University KA – BEI – Unit II – July, 2018, Sastra University
19
7/25/2018
77 79
Differential Amplifier Differential Amplifier
Its excellence lies in its ability to reject common- CMRR of the preamplifiers should be as high as
mode interference signals possible
So that only the wanted signals find a way through the
amplifier
All unwanted signals get rejected in the preamplifier stage
Typical differential Very high CMRR can be achieved with the use of an
amplifier configuration active long-tail
To minimize the effects of changes occurring in the
electrode impedances
Preamplifier having a high input impedance
KA – BEI – Unit II – July, 2018, Sastra University KA – BEI – Unit II – July, 2018, Sastra University
78 80
Differential Amplifier Differential Amplifier
Because of this signal
If the collector current of T1 increases
Collector current of T2 will decrease by the same amount
Collector voltage of T1 will decrease while that of T2 will
increase
Difference voltage between the two output terminals
that is proportional to the gain of the transistors
CMRR is an important specification referred to the
differential amplifier and is normally expressed as
decibels
KA – BEI – Unit II – July, 2018, Sastra University KA – BEI – Unit II – July, 2018, Sastra University
20
7/25/2018
81 83
KA – BEI – Unit II – July, 2018, Sastra University KA – BEI – Unit II – July, 2018, Sastra University
82 84
Differential Amplifier Chopper Amplifier
High input impedance is very necessary in order to obtain a Useful device in the field of medical electronics
high CMRR
Solution to the problem of achieving adequate low
Electrode skin resistance should be low and as nearly equal
frequency response
as possible
CMMR - typically between 60 dB and 90 dB While avoiding the drift problem inherent in direct
coupled amplifiers
Use of a chopping device
Which converts a slowly varying direct current to an
alternating form
Amplitude proportional to the input direct current
Phase dependent on the polarity of the original
signal
KA – BEI – Unit II – July, 2018, Sastra University KA – BEI – Unit II – July, 2018, Sastra University
21
7/25/2018
85 87
Chopper Amplifier Chopper Amplifier
Alternating voltage is then amplified by a Amplifying this chopped signal in an ac amplifier
conventional ac amplifier (A1)
Whose output is rectified back to get an amplified direct Demodulating the output of the ac amplifier
current Low frequency components are derived from the
Chopper amplifier is an excellent device for signals input signal by passing it through the low-pass filter
of narrow bandwidth and reduces the drift problem Chopping signal is generated by the oscillator
Single-ended chopper stabilized amplifier Filtered output is then further amplified in a second
Achieves its ultra low dc offset voltage and bias current stage of dc amplification (A2)
Chopping the low frequency components of the input High frequency signals
signal
Which are filtered out at the input of the chopper channel
Coupled directly into the second stage amplifier
KA – BEI – Unit II – July, 2018, Sastra University KA – BEI – Unit II – July, 2018, Sastra University
86 88
Chopper Amplifier Chopper Amplifier
Technique is to reduce the dc offsets and drift of the
second amplifier by a factor equal to the gain of the
chopper channel
AC amplifier introduces no offsets
Due to imperfect chopping
Minor offsets and bias currents
Extremely small
Amplifier modules
Chopper channel
Including switches and switch-driving oscillator built on
the module
Only the dc power is supplied externally
KA – BEI – Unit II – July, 2018, Sastra University KA – BEI – Unit II – July, 2018, Sastra University
22
7/25/2018
89 91
Chopper Amplifier Isolation Amplifiers
Shielding of feedback components is desirable in chopper
amplifiers
It is particularly necessary in electrically noisy environments
Voltage drift in chopper-stabilised amplifiers is 0.1 µV/ °C
Current drift as 0.5 pA/°C
Insensitivity to component changes due to ageing, temperature
change, power supply variation or other environmental factors
Application
Strain gauge pressure transducers
Temperature sensors such as thermistors
Strain gauge myographs
Thermocouple
KA – BEI – Unit II – July, 2018, Sastra University KA – BEI – Unit II – July, 2018, Sastra University
Isolation Amplifiers
23