Konelab™ / T Series TEST PROCEDURE
Refer to the Reference Manual and Application Notes separate for serum/plasma and urine for
CREATININE (Enzymatic) an automated procedure on your Konelab analyzer. See the latest applications from the web
page. Any application which has not been validated by Thermo Fisher Scientific Oy cannot be
performance guaranteed and therefore must be evaluated by the user.
981845 4 x 60 ml
Materials provided
Reagents as described above.
THIS PACKAGE INSERT IS APPLICABLE FOR USE
OUTSIDE THE US. ANY REFERENCE TO THE Materials required but not provided
WashFluid, code: 981842.
KONELAB SYSTEMS ALSO REFERS TO THE T Calibrator and controls as indicated below.
SERIES. Calibration
Use sCal, code 981831 according to the instructions provided for your Konelab analyzer. Allow
INTENDED USE the calibrator to remain on the analyzer for a maximum of one hour.
For the in vitro quantitative determination of creatinine concentration in human serum, plasma
or urine on Konelab analyzers using enzymatic method. All test results must be interpreted with Traceability:
regard to the clinical context. Refer to the package insert of sCal.
The response (A) is converted to result units by a calculation factor.
SUMMARY (1, 2)
Creatine is synthesized in the kidneys, liver, and pancreas. Creatine is then transported in blood Quality Control
to other organs, such as muscles and the brain. About 1% to 2% of muscle creatine is converted Use quality control samples at least once a day and after each calibration and every time a new
to creatinine daily. Because the amount of endogenous creatinine produced is proportional to bottle of reagent is used. It is recommended to use two level controls.
muscle mass, the production varies with age and sex. Given the same glomerular filtration rate, Available controls:
men have a higher creatinine concentration than women, muscular individuals have higher levels Serum/plasma:
than less muscular individuals, and younger persons higher levels than the elderly. Depending on Nortrol, code: 981043
the individual’s meat intake, diet may influence the value by about 10 %. On the whole, however, Abtrol, code: 981044
dietary fluctuation of creatinine intake cause only minor variation in daily creatinine excretion of the Urine:
same person. High creatinine values are found in acute and chronic kidney insufficiency and uTrol, code: 981821
dehydration. uTrol High, code: 981822
Both chemical and enzymatic methods are used to measure creatinine in body fluids. Under The Control intervals and limits must be adapted to the individual laboratory requirements.
routine conditions, creatinine can be enzymatically determined with only a few interferences. The results of the quality control samples should fall within the limits pre-set by the laboratory.
PRINCIPLE OF THE PROCEDURE CALCULATION OF RESULTS
Creatinine is determined by enzymatic colorimetric assay as follows: The results are calculated automatically by the Konelab analyzer using a calibration curve.
Calibration Curve (example)
Creatininase
Creatinine + H2O ⎯-⎯-⎯-⎯⎯-⎯> Creatine Creatinine (Enzymatic)
Creatinase 0.120
Creatine + H2O ⎯-⎯-⎯-⎯-⎯> Sarcosine + Urea
0.100
0.080
Abs
Sarcosine oxidase
Sarcosine + H2O + O2 ⎯-⎯⎯⎯-⎯-⎯-⎯⎯> Glycine + HCHO + H2O2 0.060
0.040
Peroxidase 0.020
H2O2 + 4-aminophenazone + HTIB* ⎯-⎯⎯⎯-⎯> Quinone imine chromogen + H2O + HI 0.000
* HTIB = 2,4,6-triiodo-3-hydroxybenzoic acid 0 50 100 150 200 250
µmol/l
Creatinine is converted to sarcosine with the aid of creatininase and creatinase. Sarcosine is then
converted to glycine, formaldehyde and hydrogen peroxide in the presence of oxygen by
sarcosine oxidase. The liberated hydrogen peroxide reacts with 4-aminophenazone and HTIB to Konelab 20/20XT/30/60. The calibration curve is lot dependent.
form a quinone imine chromogen in a reaction catalyzed by peroxidase. The color intensity is
directly proportional to the concentration of creatinine present and can be measured LIMITATIONS OF THE PROCEDURE
photometrically at 540nm.
Interference
REAGENT INFORMATION Criterion: Recovery within ± 10% of initial values.
Reagent A 4 x 40 ml
Reagent B 4 x 20 ml Serum/plasma:
Unconjugated Bilirubin: No interference found up to 400 µmol/l (23.4 mg/dl). Higher bilirubine
Concentrations concentrations cause erroneously low creatinine values.
Reagent A Conjugated Bilirubin: No interference found up to 300 µmol/l (17.5 mg/dl). Higher bilirubine
TAPS** buffer, pH 8.1 30 mmol/l concentrations cause erroneously low creatinine values.
Creatinase (microorganisms) > 333 μkat/l Hemoglobin / Hemolysis: No interference found up to 10 g/l of hemoglobin in hemolysate.
Sarcosine oxidase (microorganisms) > 133 μkat/l Lipemia: No interference found up to 10 g/l of Intralipid™ (trademark of
Ascorbate oxidase (microorganisms) > 33 μkat/l Fresenius Kabi AB) or 12 mmol/l (1062mg/dl) of triglycerides.
HTIB 5.9 mmol/l There is a poor correlation between turbidity and concentration of
Detergents triglycerides.
Preservative Monoclonal immunoglobulins (or parts of them) may interfere with the assay. Results from
patiens suspected of having such antibodies should be carefully evaluated.
Reagent B: For other interfering substances, please refer to the reference 4.
TAPS** buffer, pH 8.0 50 mmol/l
Creatininase (microorganisms) > 500 μkat/l Urine:
Peroxidase (horseradish) > 16.7 μkat/l Conjugated Bilirubin: No interference found up to 1000 µmol/l (58 mg/dl).
4-aminophenazone 2.0 mmol/l Hemoglobin / Hemolysis: No interference found up to 10 g/l of hemoglobin in hemolysate.
Potassium hexacyanoferrate (II) 163 μmol/l Glucose: No interference found up to 139 mmol/l (2500 mg/dl).
Detergent Ascorbic acid: No interference found up to 5.7 mmol/l (100 mg/dl).
Preservative Calcium dobesilate and α-methyldopa cause erroneously low creatinine values.
For other interfering substances, please refer to the reference 4.
The results obtained in individual laboratories may differ from the performance data given.
Storage (3)
The serum and plasma samples can be stored for 7 days at 20…25 °C or at 4…8 °C or for
3 months at –20 °C.
The urine sample can be stored for 2 days at 20…25 °C, for 6 days at 4…8 °C or for 6 months
at –20 °C.
Page 2 D04282_03_Insert_Creatinine Enzymatic_MU
Kreatininkonzentration als Frauen, muskulöse Individuen höhere Werte als weniger muskulöse
Detection limit Individuen und junge Menschen höhere Werte als ältere. Abhängig vom Fleischkonsum des
Serum/plasma: Einzelnen kann der Wert durch die Ernährung um etwa 10 % beeinflusst werden. Insgesamt
2 μmol/l (0.02 mg/dl) jedoch bewirken Schwankungen in der Kreatininaufnahme durch die Nahrung nur geringe
The detection limit represents the lowest measurable concentration/activity that can be Variationen in der täglichen Kreatininausscheidung derselben Person. Hohe Kreatininwerte
distinguished from zero. It is calculated as the concentration of zero sample +3 SD (within run, finden sich bei akuter und chronischer Niereninsuffizienz und Dehydratation.
n = 24). Für die Messung von Kreatinin in Körperflüssigkeiten werden sowohl chemische als auch
Urine: enzymatische Methoden eingesetzt. Unter Routinebedingungen kann Kreatinin mit nur wenigen
0.002 mmol/l (0.02 mg/dl) Interferenzen enzymatisch bestimmt werden.
The detection limit represents the lowest measurable concentration/activity that can be
distinguished from zero. It is calculated as the concentration of zero sample +3 SD (within run, TESTPRINZIP
n = 24) Auf dem enzymatischen Weg wird Kreatinin mit dem folgenden kolorimetrischen Test
bestimmt:
Imprecision
Serum/plasma:
Kreatininase
Mean Mean Mean Kreatinin + H2O ⎯-⎯-⎯-⎯⎯-⎯> Kreatin
38 μmol/l 155 μmol/l 484 μmol/l
SD CV% SD CV% SD CV% Kreatinase
Within run 0.56 1.4 0.69 0.4 1.85 0.4 Kreatin + H2O ⎯-⎯-⎯-⎯-⎯> Sarkosin + Harnstoff
Between run 0.16 0.4 0.71 0.5 1.30 0.3
Total 0.84 2.1 2.32 1.5 6.76 1.4 Sarkosinoxidase
Sarkosin + H2O + O2 ⎯-⎯⎯⎯-⎯-⎯-⎯⎯> Glycin + HCHO + H2O2
A precision study was performed according to guidelines in the CLSI (former NCCLS)
Document EP5-A varying calibrations and operators using one Konelab 60 for 20 days, with the
number of measurements being n = 80. Peroxidase
H2O2 + 4-Aminophenazon + HTIB* ⎯-⎯⎯⎯-⎯> Chinonimin-Chromogen + H2O + HI
Urine:
Mean Mean * HTIB = 2,4,6-Trijod-3-hydroxybenzoesäure
6.7 mmol/l 14.6 mmol/l
SD CV% SD CV% Kreatinin wird mit Hilfe von Kreatininase und Kreatinase zu Sarkosin umgewandelt. Sarkosin
Within run 0.06 1.0 0.13 0.9 setzt sich in der Gegenwart von Sauerstoff und Sarkosinoxidase wiederum in Glycin,
Between run 0.05 0.8 0.17 1.2 Formaldehyd und Wasserstoffperoxid um. Das freigesetzte Wasserstoffperoxid reagiert mit 4-
Total 0.24 3.5 0.51 3.4 Aminophenazon und HTIB in einer durch die Peroxidase katalysierten Reaktion zum Chinonimin-
Chromogen. Die Farbintensität ist direkt proportional zur Konzentration des vorliegenden
A precision study was performed according to guidelines in the CLSI (former NCCLS) Kreatinins und kann bei 540 nm fotometrisch gemessen werden.
Document EP5-A varying calibrations and operators using one Konelab 60 for 20 days, with the
number of measurements being n = 80. REAGENZDATEN
Reagenz A 4 x 40 ml
Method comparison Reagenz B 4 x 20 ml
Serum/plasma:
A comparison study was performed according to guidelines in the CLSI (former NCCLS) Konzentrationen
Document EP9-A using a commercially available enzymatic method as a reference. Reagenz A
Linear regression (result unit μmol/l): TAPS**-Puffer, pH 8.1 30 mmol/l
y = 0.97 x – 0.3 Kreatinase (Mikroorganismen) > 333 ìkat/l
r = 0.999 Sarkosinoxidase (Mikroorganismen) > 133 ìkat/l
n = 110 Ascorbatoxidase (Mikroorganismen) > 33 ìkat/l
The sample concentrations were between 38 and 1951 μmol/l. HTIB 5.9 mmol/l
Detergenzien
Urine: Konservierungsmittel
A comparison study was performed according to guidelines in the CLSI (former NCCLS)
Document EP9-A using a commercially available enzymatic method as a reference. Reagenz B:
Linear regression (result unit mmol/l): TAPS**-Puffer, pH 8.0 50 mmol/l
y = 1.10 x + 0.02 Kreatininase (Mikroorganismen) > 500 μkat/l
r = 0.999 Peroxidase (Meerrettich) > 16.7 ìkat/l
n = 119 4-Aminophenazon 2.0 mmol/l
The sample concentrations were between 1.4 and 33.2 mmol/l. Kaliumhexacyanoferrat (II) 163 ìmol/l
Detergenz
BIBLIOGRAPHY Konservierungsmittel
1) Burtis, CA and Ashwood, E R (ed.), Tietz Fundamentals of Clinical Chemistry, 5th edition,
W B Saunders Company, Philadelphia, 2001, pp. 419-422. ** TAPS = 3-[[Tris(hydroxymethyl)methyl]amino]propansulfonsäure
2) Thomas, L. (ed.), Clinical Laboratory Diagnostics; Use and Assessment of Clinical
Laboratory Results, 1st edition, TH-Books Verlagsgesellschaft mbH, Frankfurt/Main, Sicherheitsmaßnahmen
Germany, 1998, pp. 366-370. Nur zur In-vitro-Diagnostik. Die üblichen Sicherheitsmaßnahmen beim Umgang mit
3) Guder WG, Narayanan S, Wisser H, Zawta B. List of Analytes; Preanalytical variables. Laborreagenzien befolgen.
Brochure in: Samples: From Patient to the Laboratory. GIT Verlag GmbH, Darmstadt, 1996.
4) Young, D.S., Effects of Drugs on Clinical Laboratory Tests, Fifth Edition, AACC Press, Vorbereitung
Washington, D.C., 2000, pp. 3-240 - 3-261. Die Reagenzien sind gebrauchsfertig.
5) Mazzachi BC, Peake MJ, Ehrhardt V. Reference range and method comparison studies for Hinweis: Darauf achten, dass sich im Flaschenhals oder an der Reagenzoberfläche keine
enzymatic and Jaffé creatinine assays in plasma and serum and early morning urine. Clin Luftblasen befinden, wenn die Phiolen bzw. Glasfläschchen mit dem Reagenz in das Konelab-
Lab. 2000; 46: 53-55. Analysengerät eingelegt werden.
Sicherheitsmaßnahmen
Proben humanen Ursprungs sind als potenziell infektiös zu betrachten und dementsprechend
zu behandeln und zu entsorgen.
Lagerung (3)
DE Serum- und Plasmaproben können 7 Tage lang bei 20…25 °C, bei 4…8 °C oder 3 Monate
Konelab™ / T Series lang bei -20 °C gelagert werden.
Urinproben können 2 Tage lang bei 20…25 °C, 6 Tage lang bei 4…8 °C oder 6 Monate lang
CREATININE (Enzymatic) bei -20 °C gelagert werden.
KREATININ (Enzymatisch) TESTDURCHFÜHRUNG
Angaben zur Automatisierung mit dem Konelab-Analysengerät dem Referenzhandbuch und
981845 4 x 60 ml den Anwendungshinweisen für Serum/Plasma und Urin entnehmen. Die neuesten Anwendungen
stehen auf der Website bereit. Bei Verwendung von Anwendungen, die nicht durch Thermo
DIESE PACKUNGSBEILAGE IST FÜR DEN Fisher Scientific Oy validiert wurden, kann keine Garantie für die angegebenen Leistungsdaten
übernommen werden. Für die Validierung derartiger Anwendungen ist der Anwender daher selbst
GEBRAUCH AUSSERHALB DER USA VORGESEHEN. verantwortlich.
JEDER VERWEIS AUF KONELAB-SYSTEME Lieferumfang
BEINHALTET AUCH DIE T SERIES. Reagenzien wie oben beschrieben.
Erforderliches, jedoch nicht im Lieferumfang enthaltenes Material
ANWENDUNGSBEREICH WashFluid, Bestellnr.: 981842.
Zur quantitativen In-vitro-Bestimmung der Kreatininkonzentration in Humanserum, -plasma Kalibrator und Kontrollen wie nachstehend angegeben.
Template: D01162_4
oder -urin mit Konelab-Analysengeräten und einer enzymatischen Methode. Alle Testergebnisse
müssen mit Bezug zum klinischen Zusammenhang interpretiert werden. Kalibrierung
sCal, Bestellnr. 981831, gemäß den Anweisungen zum Konelab-Analysengerät verwenden.
ZUSAMMENFASSUNG (1, 2) Den Kalibrator für höchstens eine Stunde im Analysengerät lassen.
Kreatin wird in den Nieren, der Leber und der Pankreas gebildet. Kreatin wird anschließend im
Blut zu anderen Organen wie den Muskeln und dem Gehirn transportiert. Etwa 1 bis 2 % Rückverfolgbarkeit:
Muskelkreatin wird täglich zu Kreatinin umgewandelt. Da die Menge des produzierten endogenen Siehe Packungsbeilage von sCal.
Kreatinins direkt von der Muskelmasse abhängt, variiert die Produktion nach Alter und Das Messergebnis (E) wird mit einem Berechnungsfaktor in Ergebniseinheiten umgerechnet.
Geschlecht. Bei der gleichen glomerulären Filtrationsrate haben Männer eine höhere
Page 3 D04282_03_Insert_Creatinine Enzymatic_MU
Qualitätskontrolle Impräzision
Mindestens einmal täglich sowie nach jeder Kalibrierung und bei jeder Verwendung eines Serum/Plasma:
neuen Reagenzfläschchens eine Qualitätskontrolle durchführen. Es sollten zwei Kontrollniveaus
verwendet werden. Mittelwert Mittelwert Mittelwert
Erhältliche Kontrollen: 38 μmol/l 155 μmol/l 484 μmol/l
Serum/Plasma: SD % VK SD % VK SD % VK
Nortrol, Bestellnr.: 981043 In der Serie 0.56 1.4 0.69 0.4 1.85 0.4
Abtrol, Bestellnr.: 981044 Von Serie zu Serie 0.16 0.4 0.71 0.5 1.30 0.3
Urin: Gesamt 0.84 2.1 2.32 1.5 6.76 1.4
uTrol, Bestellnr.: 981821
uTrol High, Bestellnr.: 981822 Es wurde über 20 Tage eine Präzisionsstudie gemäß CLSI-Richtlinie (vorm. NCCLS), EP5-A,
Die Intervalle und Grenzen der Kontrolle müssen an die Anforderungen der einzelnen mit dem Analysengerät Konelab 60 und variierenden Kalibrierungen und Bedienern durchgeführt
Laboratorien angepasst werden. Die Ergebnisse der Qualitätskontrollen sollten innerhalb der vom (n=80).
Labor vorgegebenen Grenzwerte liegen.
Urin:
BERECHNUNG DER ERGEBNISSE
Die Ergebnisse werden vom Konelab-Analysengerät mithilfe einer Bezugskurve automatisch Mittelwert Mittelwert
berechnet. 6.7 mmol/l 14.6 mmol/l
SD % VK SD % VK
Bezugskurve (Beispiel) In der Serie 0.06 1.0 0.13 0.9
Von Serie zu Serie 0.05 0.8 0.17 1.2
Kreatinin (enzymatisch) Gesamt 0.24 3.5 0.51 3.4
Es wurde über 20 Tage eine Präzisionsstudie gemäß CLSI-Richtlinie (vorm. NCCLS), EP5-A,
0.120 mit dem Analysengerät Konelab 60 und variierenden Kalibrierungen und Bedienern durchgeführt
0.100 (n=80).
0.080 Vergleich der Methoden
Ext
0.060 Serum/Plasma:
Es wurde eine Vergleichsstudie gemäß CLSI-Richtlinie (vorm. NCCLS) EP9-A mit einer
0.040 handelsüblichen enzymatischen Methode als Referenz durchgeführt.
0.020 Lineare Regression (Ergebnisse in µmol/l):
0.000
y = 0.97 x – 0.3
r = 0.999
0 50 100 150 200 250 n = 110
Die Probenkonzentrationen lagen zwischen 38 und 1951 μmol/l.
µmol/l
Urin:
Es wurde eine Vergleichsstudie gemäß CLSI-Richtlinie (vorm. NCCLS) EP9-A mit einer
Konelab 20/20XT/30/60. Die Bezugskurve ist chargenabhängig. handelsüblichen enzymatischen Methode als Referenz durchgeführt.
Lineare Regression (Ergebnisse in mmol/l):
GRENZEN DES VERFAHRENS y = 1.10 x + 0.02
r = 0.999
Störfaktoren n = 119
Kriterium: Wiederfindung innerhalb von ± 10 % der Ausgangswerte. Die Probenkonzentrationen lagen zwischen 1.4 und 33.2 mmol/l.
Serum/Plasma: LITERATURVERWEISE
Nicht konjugiertes Bilirubin: Keine Interferenzen bis zu 400 µmol/l (23.4 mg/dl). Höhere 1) Burtis, CA and Ashwood, E R (ed.), Tietz Fundamentals of Clinical Chemistry, 5. Ausgabe,
Bilirubinkonzentrationen führen zu falsch niedrigen W B Saunders Company, Philadelphia, 2001, Seiten 419-422.
Kreatininwerten. 2) Thomas, L. (Ed.), Clinical Laboratory Diagnostics; Use and Assessment of Clinical
Konjugiertes Bilirubin: Keine Interferenzen bis zu 300 µmol/l (17.5 mg/dl). Höhere Laboratory Results, 1. Ausgabe, TH-Books Verlagsgesellschaft mbH, Frankfurt/Main, 1998,
Bilirubinkonzentrationen führen zu fehlerhaft niedrigen Seiten 366-370.
Kreatininwerten. 3) Guder WG, Narayanan S, Wisser H, Zawta B. List of Analytes; Preanalytical variables.
Hämoglobin/Hämolyse: Keine Interferenzen bis zu 10 g/l Hämoglobin im Hämolysat. Broschüre in: Samples: From Patient to the Laboratory. GIT Verlag GmbH, Darmstadt, 1996.
Lipämie: Keine Interferenzen bis zu 10 g/l Intralipid™ (Marke von 4) Young, D. S.: Effects of Drugs on Clinical Laboratory Tests, 5. Ausgabe, AACC Press,
Fresenius Kabi AB) oder 12 mmol/l (1062 mg/dl) Triglyceride Washington, D.C., 2000, Seiten 3-240 - 3-261.
festgestellt. Die Korrelation zwischen der Trübung und der 5) Mazzachi BC, Peake MJ, Ehrhardt V. Reference range and method comparison studies for
Triglyceridkonzentration ist schlecht. enzymatic and Jaffé creatinine assays in plasma and serum and early morning urine. Clin
Monoklonale Immunoglobuline (oder deren Bestandteile) können die Testergebnisse Lab. 2000; 46: 53-55.
verfälschen. Die Ergebnisse bei Patienten mit Verdacht auf derartige Antikörper müssen mit
größter Sorgfalt ausgewertet werden. HERSTELLER
Siehe Literaturhinweis 4 für weitere Substanzen, die Störfaktoren darstellen können. Thermo Fisher Scientific Oy
Clinical Diagnostics Finland
Urin: Ratastie 2, P.O. Box 100, FI-01621 Vantaa, Finnland
Konjugiertes Bilirubin: Keine Interferenzen bis zu 1000 µmol/l (58 mg/dl). Tel. +358 9 329 100, Fax +358 9 3291 0300
Hämoglobin/Hämolyse: Keine Interferenzen bis zu 10 g/l Hämoglobin im Hämolysat. www.thermo.com/konelab
Glucose: Keine Interferenzen bis zu 139 mmol/l (2500 mg/dl).
Ascorbinsäure: Keine Interferenzen bis zu 5.7 mmol/l (100 mg/dl). Datum der Überarbeitung (JJJJ-MM-TT)
Calciumdobesilat und á-Methyldopa führen zu falsch niedrigen Kreatinwerten. 2007-08-07
Siehe Literaturhinweis 4 für weitere Substanzen, die Störfaktoren darstellen können.
Änderungen gegenüber der vorherigen Fassung
REFERENZBEREICHE (5) Name des Unternehmens aktualisiert.
Serum/Plasma:
Männer: 59 – 104 μmol/l (0.67 – 1.17 mg/dl)
Frauen: 45 – 84 μmol/l (0.51 – 0.95 mg/dl)
Urine :
PRINCIPE DE LA PROCÉDURE uTrol, code : 981821
La créatinine est dosée par une méthode enzymatique colorimétrique de la manière suivante : uTrol High, code : 981822
Les intervalles et les limites de contrôle doivent être adaptés aux besoins de chaque
laboratoire. Les résultats des échantillons de contrôle de qualité doivent se situer dans la plage de
Créatininase tolérance prédéfinie par le laboratoire.
Créatinine + H2O ⎯-⎯-⎯-⎯⎯-⎯> Créatine
CALCUL DES RÉSULTATS
Créatinase Les résultats sont calculés automatiquement par l’analyseur Konelab à l’aide d’une courbe de
Créatine + H2O ⎯-⎯-⎯-⎯-⎯> Sarcosine + Urée calibrage.
Abs
La sarcosine est ensuite convertie en glycine, formaldéhyde et peroxyde d’hydrogène en
présence d’oxygène par la sarcosine oxydase. Le peroxyde d’hydrogène libéré réagit avec la 0.060
4-aminophénazone et le HTIB pour donner naissance à un chromogène quinone imine dans une 0.040
réaction catalysée par la peroxydase. L’intensité de la coloration est directement proportionnelle à
la concentration en créatinine présente et peut être mesurée par photométrie à 540 nm. 0.020
Utiliser les échantillons de contrôle de qualité au moins une fois par jour, après chaque
calibrage et chaque fois que l’on entame un nouveau flacon de réactif. Il est conseillé d’utiliser
deux niveaux de contrôles.
Contrôles disponibles :
Sérum/plasma :
Nortrol, code : 981043
Abtrol, code : 981044
Page 5 D04282_03_Insert_Creatinine Enzymatic_MU
Imprécision
Sérum/plasma :
Urine :
Moyenne Moyenne
6.7 mmol/l 14.6 mmol/l
ET CV % ET CV %
Répétabilité 0.06 1.0 0.13 0.9
Reproductibilité 0.05 0.8 0.17 1.2
Total 0.24 3.5 0.51 3.4
Comparaison de méthodes
Sérum/plasma :
Une étude comparative a été réalisée conformément aux directives du document CLSI
(précédemment NCCLS) EP9-A en utilisant comme référence une méthode enzymatique
disponible dans le commerce.
Régression linéaire (unités du résultat : µmol/l) :
y = 0.97 x – 0.3
r = 0.999
n = 110
Les concentrations des échantillons se situaient entre 38 et 1951 µmol/l.
Urine :
Une étude comparative a été réalisée conformément aux directives du document CLSI
(précédemment NCCLS) EP9-A en utilisant comme référence une méthode enzymatique
disponible dans le commerce.
Régression linéaire (unités du résultat : mmol/l) :
y = 1.10 x + 0.02
r = 0.999
n = 119
Les concentrations des échantillons se situaient entre 1.4 et 33.2 mmol/l.
BIBLIOGRAPHIE
1) Burtis, CA and Ashwood, E R (ed.), Tietz Fundamentals of Clinical Chemistry, 5th edition,
W B Saunders Company, Philadelphia, 2001, pp. 419-422.
2) Thomas, L. (ed.), Clinical Laboratory Diagnostics; Use and Assessment of Clinical
Laboratory Results, 1st edition, TH-Books Verlagsgesellschaft mbH, Frankfurt/Main,
Germany, 1998, pp. 366-370.
3) Guder WG, Narayanan S, Wisser H, Zawta B. List of Analytes; Preanalytical variables.
Brochure in : Samples: From Patient to the Laboratory. GIT Verlag GmbH, Darmstadt, 1996.
4) Young, DS. Effects of Drugs on Clinical Laboratory Tests, cinquième édition, AACC Press,
Washington, D.C., 2000, pp. 3-240 - 3-261.
5) Mazzachi BC, Peake MJ, Ehrhardt V. Reference range and method comparison studies for
enzymatic and Jaffé creatinine assays in plasma and serum and early morning urine. Clin
Lab. 2000; 46: 53-55.
FABRICANT
Thermo Fisher Scientific Oy
Clinical Diagnostics Finland
Ratastie 2, P.O. Box 100, FI-01621 Vantaa, Finlande
Tél. +358 9 329 100, télécopie +358 9 3291 0300
www.thermo.com/konelab
CS
Konelab™ / Série T POSTUP TESTU
Údaje o automatizovaném postupu práce na analyzátoru Konelab naleznete v Referenčním
CREATININE (Enzymatic) manuálu a aplikačních poznámkách, části věnovaná stanovením ze séra/plazmy a moči. Aktuální
aplikace naleznete na webových stránkách. Nelze zaručit provedení žádné aplikace, která nebyla
KREATININ (enzymaticky) validována společnosti Thermo Fisher Scientific Oy. Taková aplikace proto musí být hodnocena
uživatelem.
981845 4 x 60 ml
Dodávané materiály
Reagencia uvedená výše.
TENTO PŘÍBALOVÁ INFORMACE JE URČENA PRO
POUŽITÍ MIMO ÚZEMÍ USA. KAŽDÁ ZMÍNKA O Potřebné materiály, které se dodávají zvlášť
WashFluid, kód: 981842.
SYSTÉMECH KONELAB SE ROVNĚŽ TÝKÁ SÉRIE T. Kalibrátor a kontrolní materiály uvedené dále.
POUŽITÍ Kalibrace
Pro kvantitativní stanovení koncentrace kreatininu in vitro v lidském séru, plazmě nebo v moči Používejte kalibrátor sCal, kód 981831 podle návodu, který vám byl dodán s analyzátorem
na analyzátorech Konelab pomocí enzymatické metody. Všechny výsledky testů musejí být Konelab. Kalibrátor používejte v analyzátoru nejvýše po dobu jedné hodiny.
interpretovány s ohledem na klinický kontext.
Identifikovatelnost:
SHRNUTÍ (1, 2) Viz příbalový leták kalibrátoru sCal.
Kreatin je syntetizován v ledvinách, játrech a slinivce břišní. Poté je kreatin transportován krví Reakce (A) je přepočtena pomocí faktoru výpočtu na výsledné jednotky.
do dalších orgánů, jako jsou svaly a mozek. Asi 1 % až 2 % svalového kreatinu je denně
přeměňováno na kreatinin. Vzhledem k tomu, že množství vyprodukovaného endogenního Řízení jakosti
kreatininu je přímo úměrné svalové hmotě, liší se jeho produkce podle věku a pohlaví. Při stejné Používejte vzorky pro řízení jakosti alespoň jednou denně, po každé kalibraci a vždy, když
rychlosti glomerulární filtrace mají muži vyšší koncentraci kreatininu než ženy, osvalení jedinci použijete novou lahvičku reagens. Doporučuje se použít kontrolní materiály se dvěma hladinami.
mají vyšší hladinu než osoby méně osvalené a mladší osoby mají vyšší hladinu než senioři. V Dodávané kontrolní materiály:
závislosti na příjmu masa potravou konkrétním jedincem může strava ovlivnit hodnotu v rozsahu Sérum/plazma:
asi 10 %. Celkem však výkyvy příjmu kreatininu potravou způsobují jen malé odchylky denní Nortrol, kód: 981043
exkrece kreatininu dané osoby. Vysoké hodnoty kreatininu se vyskytují při akutní a chronické Abtrol, kód: 981044
ledvinové insuficienci a dehydrataci. Moč:
Pro měření kreatininu v tělních tekutinách se používají chemické a enzymatické metody. uTrol, kód: 981821
Za běžných podmínek lze kreatinin stanovit enzymaticky jen s několika interferencemi. uTrol High, kód: 981822
Intervaly a limity kontrolních materiálů musejí být upraveny podle požadavků jednotlivých
PRINCIP POSTUPU laboratoří. Výsledky vzorků pro řízení jakosti by měly vyhovovat limitním hodnotám
Kreatinin se stanovuje enzymatickým kolorimetrickým testem dále uvedeným způsobem: přednastaveným laboratoří.
VÝPOČET VÝSLEDKŮ
Kreatinináza Výsledky jsou automaticky vypočteny analyzátorem Konelab pomocí kalibrační křivky.
Kreatinin + H2O ⎯-⎯-⎯-⎯⎯-⎯> Kreatin
Kalibrační křivka (příklad)
Kreatináza
Kreatin + H2O ⎯-⎯-⎯-⎯-⎯> Sarkosin + Močovina KREATININ (enzymaticky)
Sarkosinoxidáza 0.120
Sarkosin + H2O + O2 ⎯-⎯⎯⎯-⎯-⎯-⎯⎯⎯ Glycin + HCHO + H2O2 0.100
0.080
Abs
Peroxidáza
0.060
H2O2 + 4-aminofenazon + HTIB* ⎯-⎯⎯⎯-⎯⎯ Chinoniminový chromogen + H2O + HI
0.040
* HTIB = kyselina 2,4,6-trijodo-3-hydroxybenzoová 0.020
λιγότερο μυώδη άτομα, ενώ οι νεότεροι έχουν υψηλότερα επίπεδα από τους μεγαλύτερους.
Detekční limit Ανάλογα με την πρόσληψη κρέατος του ατόμου, η δίαιτα μπορεί να επηρεάσει την τιμή περίπου
Sérum/plazma: κατά 10%. Παρόλα αυτά, η διατροφική διακύμανση της πρόσληψης της κρεατινίνης συνολικά
2 μmol/l (0.02 mg/dl) προκαλεί ελάχιστη μεταβολή στην ημερήσια απέκκριση κρεατινίνης του ατόμου αυτού. Υψηλές
Detekční limit představuje nejnižší měřitelnou koncentraci/aktivitu, kterou lze odlišit od nulové τιμές κρεατινίνης βρίσκονται σε οξεία και χρόνια νεφρική ανεπάρκεια και αφυδάτωση.
hodnoty. Vypočítává se jako koncentrace nulového vzorku +3 SD (v rámci série, n = 24). Τόσο οι χημικές όσο και οι ενζυμικές μέθοδοι χρησιμοποιούνται για να μετρήσουν την
Moč: κρεατινίνη στα σωματικά υγρά. Σε συνθήκες ρουτίνας, η κρεατινίνη μπορεί να καθοριστεί
0.002 mmol/l (0.02 mg/dl) ενζυμικά με σχετικά μικρές αλληλεπιδράσεις.
Detekční limit představuje nejnižší měřitelnou koncentraci/aktivitu, kterou lze odlišit od nulové
hodnoty. Vypočítává se jako koncentrace nulového vzorku +3 SD (v rámci série, n = 24). ΑΡΧΗ ΤΗΣ ΔΙΑΔΙΚΑΣΙΑΣ
Η κρεατινίνη καθορίζεται από ενζυμική χρωματομετρική μέθοδο ως ακολούθως:
Nepřesnost
Sérum/plazma:
Κρεατινική κινάση
Střední hodnota Střední hodnota Střední hodnota Κρεατινίνη + H2O ⎯-⎯-⎯-⎯⎯-⎯> Κρεατινίνη
38 µmol/l 155 µmol/l 484 µmol/l
SD % CV SD % CV SD % CV Κρεατινάση
V rámci série 0.56 1.4 0.69 0.4 1.85 0.4 Κρεατίνη + H2O ⎯-⎯-⎯-⎯-⎯> Σακροζίνη + Ουρία
Mezi sériemi 0.16 0.4 0.71 0.5 1.30 0.3
Celkem 0.84 2.1 2.32 1.5 6.76 1.4 Οξειδάση της Σακροζίνης
Σακροζίνη + H2O + O2 ⎯-⎯⎯⎯-⎯-⎯-⎯⎯> Γλυκίνη + HCHO + H2O2
Byla provedena studie přesnosti na analyzátoru Konelab 60 podle pokynů dokumentu CLSI
(dříve NCCLS) EP5-A pomocí různých kalibrací a obsluhy trvající 20 dní, v níž byl počet měření
n = 80. Υπεροξειδάση
H2O2 + 4-αμινοφαιναζόνη + HTIB* ⎯-⎯⎯⎯-⎯> Χρωμογόνο της ιμινοκινόνης + H2O + HI
Moč:
Střední hodnota Střední hodnota * HTIB = 2,4,6-τρι ίωδο-3-υδροξυβεζοϊκό οξύ
6.7 mmol/l 14.6 mmol/l
SD % CV SD % CV Η κρεατινίνη μετατρέπεται σε σακροζίνη με τη βοήθεια της κρεατινικής κινάσης και κρεατινάσης.
V rámci série 0.06 1.0 0.13 0.9 Η σαρκοζίνη στη συνέχεια μετέπεται σε γλυκίνη, φορμαλδεϋδη, και υπεροξείδιο υδρογόνου
Mezi sériemi 0.05 0.8 0.17 1.2 παρουσία οξυγόνου από την οξειδάση της σαρκοζίνης. Το απελευθερωμένο υπεροξείδιο του
Celkem 0.24 3.5 0.51 3.4 υδρογόνου αντιδρά με 4 αμινοφαιναζόνη και ΗΙΤΒ για να σχηματίσει χρωμογόνο της ιμινοκινόνης
σε μια αντίδραση που καταλύεται από τη υπεροξειδάση. Η ένταση του χρώματος είναι άμεσα
Byla provedena studie přesnosti na analyzátoru Konelab 60 podle pokynů dokumentu CLSI ανάλογη προς τη συγκέντρωση της υπάρχουσας κρεατινίνης και μπορεί να μετρηθεί φωτομετρικά
(dříve NCCLS) EP5-A pomocí různých kalibrací a obsluhy trvající 20 dní, v níž byl počet měření στα 540nm.
n = 80.
ΠΛΗΡΟΦΟΡΙΕΣ ΑΝΤΙΔΡΑΣΤΗΡΙΩΝ
Srovnání metod Αντιδραστήριο Α 4 x 40 ml
Sérum/plazma: Αντιδραστήριο Β 4 x 20 ml
Byla provedena srovnávací studie podle pokynů uvedených v dokumentu CLSI (dříve NCCLS)
EP9-A, při níž byla použita komerčně dostupná enzymatická metoda jako metoda referenční. Συγκεντρώσεις
Lineární regrese (jednotky výsledků µmol/l): Αντιδραστήριο Α
y = 0.97 x – 0.3 Ρυθμιστικό διάλυμα TΑΡS**, pH 8.1 30 mmol/l
r = 0.999 Κρεατινάση (μικρο οργανισμοί) > 333 μkat/l
n = 110 Οξειδάση Σαρκοζίνης (μικρο οργανισμοί) > 133 μkat/l
Koncentrace vzorků se pohybovaly v rozmezí 38 až 1 951 µmol/l. Ασκορβική Οξειδάση (μικρο οργανισμοί) > 33 μkat/l
HTIB 5.9 mmol/l
Moč: Απορρυπαντικά
Byla provedena srovnávací studie podle pokynů uvedených v dokumentu CLSI (dříve NCCLS) Συντηρητικό
EP9-A, při níž byla použita komerčně dostupná enzymatická metoda jako metoda referenční.
Lineární regrese (jednotky výsledku mmol/l): Αντιδραστήριο Β:
y = 1.10 x + 0.02 Ρυθμιστικό διάλυμα TΑΡS**, pH 8.0 50 mmol/l
r = 0.999 Κρεατινική κινάση (μικρο οργανισμοί) > 500 μkat/l
n = 119 Υπεροξειδάση (χρένο) > 16.7 μkat/l
Koncentrace vzorků se pohybovaly v rozmezí 1.4 až 33.2 mmol/l. 4- αμινοφαιναζόνη 2.0 mmol/l
Potassium hexacyanoferrate (II) 163 μmol/l
SEZNAM LITERATURY Απορρυπαντικό
1) Burtis, CA and Ashwood, E R (ed.), Tietz Fundamentals of Clinical Chemistry, 5th edition, W Συντηρητικό
Saunders Company, Philadelphia, 2001, pp. 419-422.
2) Thomas, L. (ed.), Clinical Laboratory Diagnostics; Use and Assessment of Clinical ** TAPS = 3-[[τρις(υδροξυμέθυλοl)μέθυλ]άμινο]προπανεσουλφανικό οξύ
Laboratory Results, 1st edition, TH-Books Verlagsgesellschaft mbH, Frankfurt/Main,
Germany, 1998, pp. 366-370. Προφυλάξεις
3) Guder WG, Narayanan S, Wisser H, Zawta B. List of Analytes; Preanalytical variables. Για in vitro διαγνωστική χρήση μόνο. Εφαρμόζετε τις κανονικές προφυλάξεις που απαιτούνται
Brochure in: Samples: From Patient to the Laboratory. GIT Verlag GmbH, Darmstadt, 1996. για το χειρισμό όλων των εργαστηριακών αντιδραστηρίων.
4) Young, D.S., Effects of Drugs on Clinical Laboratory Tests, Fifth Edition, AACC Press,
Washington, D.C., 2000, pp. 3-240 - 3-261. Προετοιμασία
5) Mazzachi BC, Peake MJ, Ehrhardt V. Reference range and method comparison studies for Τα αντιδραστήρια είναι έτοιμα προς χρήση.
enzymatic and Jaffé creatinine assays in plasma and serum and early morning urine. Clin Σημείωση: Ελέγξτε ότι δεν υπάρχουν φυσαλίδες στο λαιμό ή στην επιφάνεια του φιαλιδίου
Lab. 2000; 46: 53-55. αντιδραστηρίου όταν εισάγετε τα φιαλίδια ή τα δοχεία αντιδραστηρίων στην αναλυτή Konelab.
αποτελέσματα της εξέτασης πρέπει να ερμηνεύονται σε σχέση με τις κλινικές συνθήκες. Χρησιμοποιήστε sCal, κωδικός 981831 σύμφωνα με τις οδηγίες που παρέχονται στον αναλυτή
Konelab. Αφήστε το βαθμονομητή σε ηρεμία στον αναλυτή, το πολύ μια ώρα.
ΠΕΡΙΛΗΠΤΙΚΑ (1, 2)
Η κρεατίνη συντίθεται στους νεφρούς, το ήπαρ και το πάγκρεας. Μετά, η κρεατίνη μεταφέρεται Ανιχνευσιμότητα
με το αίμα σε άλλα όργανα, όπως τους μύες και τον εγκέφαλο. Περίπου το 1% έως 2% της Ανατρέξτε στο ένθετο της συσκευασίας του sCal.
κρεατίνης των μυών μετατρέπεται ημερησίως σε κρεατινίνη. Επειδή το ποσό της ενδογενούς Η απόκριση (A) μετατρέπεται σε μονάδες αποτελέσματος από έναν παράγοντα υπολογισμού.
κρεατινίνης που παράγεται είναι ανάλογο της μυϊκής μάζας, η παραγωγή ποικίλει με την ηλικία και
το φύλο. Με δεδομένο την ιδία συχνότητα σπειραματικής διήθησης, οι άνδρες έχουν υψηλότερη
συγκέντρωση κρεατινίνης από τις γυναίκες, τα μυώδη άτομα έχουν υψηλότερα επίπεδα από τα
Page 8 D04282_03_Insert_Creatinine Enzymatic_MU
Ορός/πλάσμα:
0.060 Διεξήχθη συγκριτική μελέτη σύμφωνα με τις οδηγίες στο Έγγραφο CLSI (πρώην NCCLS) EP9-
0.040 A με χρήση μιας εμπορικά διαθέσιμης ενζυμικής μεθόδου ως αναφορά.
0.020 Γραμμική παλινδρόμηση (μονάδα αποτελεσμάτων µmol/l):
y = 0.97 x – 0.3
0.000 r = 0.999
0 50 100 150 200 250 n = 110
Οι συγκεντρώσεις δείγματος ήταν μεταξύ 38 και 1951 µmol/l.
µmol/l
Ούρα:
Διεξήχθη συγκριτική μελέτη σύμφωνα με τις οδηγίες στο Έγγραφο CLSI (πρώην NCCLS) EP9-
Konelab 20/20XT/30/60. Η καμπύλη βαθμονόμησης εξαρτάται από την παρτίδα. A με χρήση μιας εμπορικά διαθέσιμης ενζυμικής μεθόδου ως αναφορά.
Γραμμική συμμεταβολή (μονάδα αποτελεσμάτων mmol/l):
ΠΕΡΙΟΡΙΣΜΟΙ ΤΗΣ ΔΙΑΔΙΚΑΣΙΑΣ y = 1.10 x +0.02
r = 0.999
Παρεμβολές n = 119
Κριτήριο: Ανάκτηση εντός ±10% των αρχικών τιμών. Οι συγκεντρώσεις δείγματος ήταν μεταξύ 1.4 και 33.2 mmol/l.
Ορός/πλάσμα:
Ασύζευκτη χολερυθρίνη: Δε βρέθηκε παρεμβολή έως 400 μmol/l (23.4 mg/dl). Υψηλότερες ΒΙΒΛΙΟΓΡΑΦΙΑ
συγκεντρώσεις χολερυθρίνης μπορεί να προκαλέσουν λανθασμένα 1) Burtis, CA and Ashwood, E R (ed.), Tietz Fundamentals of Clinical Chemistry, 5th edition, W
χαμηλές τιμές κρεατινίνης, B Saunders Company, Philadelphia, 2001, pp. 419-422.
Συζευγμένη χολερυθρίνη: Δε βρέθηκε παρεμβολή έως 300 μmol/l (17.5 mg/dl). Υψηλότερες 2) Thomas, L. (ed.), Clinical Laboratory Diagnostics; Use and Assessment of Clinical
συγκεντρώσεις χολερυθρίνης μπορεί να προκαλέσουν λανθασμένα Laboratory Results, 1st edition, TH-Books Verlagsgesellschaft mbH, Frankfurt/Main,
χαμηλές τιμές κρεατινίνης, Germany, 1998, pp. 366-370.
Αιμοσφαιρίνη/ Αιμόλυση: Δεν βρέθηκε παρεμβολή μέχρι 10 g/l αιμοσφαιρίνης. 3) Guder WG, Narayanan S, Wisser H, Zawta B. List of Analytes; Preanalytical variables.
Λιπιδιαιμία: Δεν βρέθηκε παρεμβολή έως 10 g/l του Intralipid™ (εμπορικό σήμα Brochure in: Samples: From Patient to the Laboratory. GIT Verlag GmbH, Darmstadt, 1996.
της Fresenius Kabi AB) ή 12 mmol/l (1.062mg/dl) τριγλυκεριδίων. 4) Young, D.S., Effects of Drugs on Clinical Laboratory Tests, Fifth Edition, AACC Press,
Υπάρχει μικρή συσχέτιση ανάμεσα στη θολερότητα και τη Washington, D.C., 2000, pp. 3-240 - 3-261.
συγκέντρωση τριγλυκεριδίων. 5) Mazzachi BC, Peake MJ, Ehrhardt V. Reference range and method comparison studies for
Οι μονοκλωνικές ανοσοσφαιρίνες (ή μέρη αυτών) ενδέχεται να παρεμβληθούν στην ανάλυση. enzymatic and Jaffé creatinine assays in plasma and serum and early morning urine. Clin
Τα αποτελέσματα των ασθενών, για τους οποίους υπάρχουν υποψίες ότι διαθέτουν τέτοια Lab. 2000; 46: 53-55.
αντισώματα, πρέπει να αξιολογούνται προσεκτικά (4).
Για άλλες παρεμβαλλόμενες ουσίες, παρακαλούμε να ανατρέξετε στην παραπομπή 4. ΚΑΤΑΣΚΕΥΑΣΤΗΣ
Thermo Fisher Scientific Oy
Ούρα: Clinical Diagnostics Finland
Συζευγμένη χολερυθρίνη: Δε βρέθηκε παρεμβολή έως 1000 μmol/l (58 mg/dl). Ratastie 2, P.O. Box 100, FI-01621 Vantaa, Finland
Αιμοσφαιρίνη/ Αιμόλυση: Δεν βρέθηκε παρεμβολή μέχρι 10 g/l αιμοσφαιρίνης. Τηλ. +358 9 329 100, Fax +358 9 3291 0300
Γλυκόζη: Δε βρέθηκε παρεμβολή έως 139 μmol/l (2.500 mg/dl). www.thermo.com/konelab
Ασκορβικό οξύ: Δε βρέθηκε παρεμβολή έως 5.7 μmol/l (100 mg/dl).
Το δοβεσιλικό ασβέστιο και το α-μεθυλντόπα προκαλεί λανθασμένες χαμηλές τιμές κρεατινίνης. Ημερομηνία αναθεώρησης (εεε-μμ-ηη)
Για άλλες παρεμβαλλόμενες ουσίες, παρακαλούμε να ανατρέξετε στην παραπομπή 4. 2007-08-07
Abs
la sarcosina oxidasa convierte la sarcosina en glicina, formaldehído y agua oxigenada en
presencia de oxígeno. El peróxido de hidrógeno liberado reacciona con la 4-aminofenazona y 0.060
con el HTIB y forma un cromógeno de quinona imina en una reacción catalizada por la 0.040
peroxidasa. La intensidad del color es directamente proporcional a la concentración de creatinina
presente y puede medirse fotométricamente a 540 nm. 0.020
0.000
INFORMACIÓN SOBRE LOS REACTIVOS 0 50 100 150 200 250
Reactivo A 4 x 40 ml
Reactivo B 4 x 20 ml µmol/l
Concentraciones
Reactivo A
Tampón TAPS**, pH 8.1 30 mmol/l Konelab 20/20XT/30/60. La curva de calibración depende del lote.
Creatinasa (microorganismos) > 333 μkat/l LIMITACIONES DEL PROCEDIMIENTO
Sarcosina oxidasa (microorganismos) > 133 μkat/l
Ascorbato oxidasa (microorganismos) > 33 μkat/l Interferencias
HTIB 5.9 mmol/l
Detergentes Criterio: Recuperación en ± 10% de los valores iniciales.
Conservante
Suero/plasma:
Reactivo B: Bilirrubina no conjugada: sin interferencias encontradas hasta 400 µmol/l (23.4 mg/dl). Las
Tampón TAPS**, pH 8.0 50 mmol/l concentraciones de bilirrubina superiores producen valores de
Creatininasa (microorganismos) > 500 μkat/l creatinina erróneamente bajos.
Peroxidase (rábano picante) > 16.7 μkat/l Bilirrubina conjugada: sin interferencias encontradas hasta 300 µmol/l (17.5 mg/dl). Las
4-aminofenazona 2.0 mmol/l concentraciones de bilirrubina superiores producen valores de
Hexacianoferrato de potasio (II) 163 μmol/l creatinina erróneamente bajos.
Detergente Hemoglobina/hemólisis: sin interferencias encontradas hasta 10 g/l de hemoglobina en el
Conservante hemolizado.
Lipemia: sin interferencias encontradas hasta 10 g/l de Intralipid™ (marca
** TAPS = ácido 3-[[tris(hidroximetil)metil]amino]propanosulfónico comercial de Fresenius Kabi AB) o 12 mmol/l (1062 mg/dl) de
triglicéridos. Hay una escasa correlación entre turbidez y
Precauciones concentración de triglicéridos.
Sólo para uso diagnóstico in vitro. Adopte las medidas de precaución habituales para En el ensayo pueden interferir inmunoglobulinas monoclonales (o partes de ellas). Deben
manipular reactivos de laboratorio. evaluarse cuidadosamente los resultados de pacientes sospechosos de tener tales anticuerpos.
Si desea obtener información sobre la interferencia de otras sustancias, consulte la
Preparación referencia 4.
Los reactivos están listos para su uso.
Nota: Compruebe que no hayan burbujas en el cuello de la botella ni en la superficie del Orina:
reactivo cuando inserte viales o recipientes en el analizador Konelab. Bilirrubina conjugada: sin interferencias encontradas hasta 1000 µmol/l (58 mg/dl).
Hemoglobina/hemólisis: sin interferencias encontradas hasta 10 g/l de hemoglobina en el
Almacenamiento y estabilidad hemolizado.
Los reactivos sin abrir son estables a 2…8 °C hasta la fecha de caducidad impresa en la Glucosa: sin interferencias encontradas hasta 139 mmol/l (2500 mg/dl).
etiqueta. Consulte las notas de aplicación del analizador Konelab para determinar la estabilidad Ácido ascórbico: sin interferencias encontradas hasta 5.7 mmol/l (100 mg/dl).
de los reactivos una vez abiertos y puestos en el analizador. El dobesilato de calcio y la α-metildopa producen valores de creatinina erróneamente bajos.
Si desea obtener información sobre la interferencia de otras sustancias, consulte la
RECOGIDA DE MUESTRAS referencia 4.
Controles disponibles:
Suero/plasma:
Nortrol, código: 981043
Abtrol, código: 981044
Page 10 D04282_03_Insert_Creatinine Enzymatic_MU
Se realizó un estudio de la precisión según las directrices del Documento EP5-A del CLSI Sarkosiin oksidaas
(antes NCCLS), variando las calibraciones y los operadores y utilizando un analizador Konelab Sarkosiin + H2O + O2 ⎯-⎯⎯⎯-⎯-⎯-⎯⎯> Glütsiin + HCHO + H2O2
60 durante 20 días, siendo el número de medidas n = 80.
Orina: Perooksidaas
Media Media H2O2 + 4-aminofenasoon + HTIB* ⎯-⎯⎯⎯-⎯> Kinonemiin kromogeen + H2O + HI
6.7 mmol/l 14.6 mmol/l
DE CV (%) DE CV (%) * HTIB = 2,4,6-triiodo-3-hüdroksübensoehape
Intraserie 0.06 1.0 0.13 0.9
Interserie 0.05 0.8 0.17 1.2 Kreatiniin teisendatakse kreatininaasi ja kreatinaasi abil sarkosiiniks. Sarkosiin teisendatakse
Total 0.24 3.5 0.51 3.4 seejärel sarkosiin oksidaasi poolt hapniku juuresolekul glütsiiniks, formaldehüüdiks ja
vesinikperoksiidik oxygen. Vabanenud vesinikperoksiid reageerib 4-aminofenasooni ja HTIB-ga,
Se realizó un estudio de la precisión según las directrices del Documento EP5-A del CLSI moodustades kinonimiin kromogeeni reaktsiooni käigus, mille katalüsaator on peroksidaas.
(antes NCCLS), variando las calibraciones y los operadores y utilizando el analizador Konelab 60 Värvuse intensiivsus on otseselt proportsionaalne kreatiniini kontsentratsiooniga ja seda on
durante 20 días, siendo el número de medidas n = 80. võimalik fotomeetriliselt mõõta lainepikkusel 540 nm.
Hoiatused
Inimpäritolu proovid tuleb käsitsemisel ja kõrvaldamisel lugeda võimalikeks nakkusallikateks.
Säilitamine (3)
Seerumi ja plasma proove võib temperatuuril 20…25 °C või 4…8 °C säilitada kuni 7 päeva,
temperatuuril –20 °C kuni 3 kuud.
ET Uriiniproovi võib temperatuuril 20…25 °C säilitada 2 päeva, temperatuuril 4…8 °C kuni 6
Konelab™ / T seeria päeva, temperatuuril –20 °C kuni 6 kuud.
Kaliibrimine
SIHTOTSTARVE sCal, 981831, mida kasutatakse järgides Konelab analüsaatori juhendit. Kaliibrimisreaktiivi pole
Kreatiniini kontsentratsiooni kvantitatiivseks in vitro määramiseks inimese vereseerumis, soovitav hoida analüsaatoris üle ühe tunni.
plasmas või -uriinis Konelabi analüsaatorite abil ensümaatilisel meetodil. Katsetulemuste
tõlgendamisel tuleb alati arvesse võtta kliinilist tausta. Jälgitavus:
Juhinduda sCal pakendi infolehest.
KOKKUVÕTE (1, 2) Vastus (A) teisendatakse ühikuliseks tulemuseks vastava arvutusteguri abil.
Kreatiini sünteesimine toimub neerus, maksas ja kõhunäärmes. Veri transpordib kreatini
teistesse organitesse, nagu lihased ja aju. Ligikaudu 1% kuni 2% lihaste kreatiinist muundub Kvaliteedikontroll
igapäevaselt kreatiniiniks. Kuna endogeenses reaktsioonis tekkiva kreatiniini hulk on võrdelises Kvaliteedikontrolliproove kasutatakse vähemalt kord päevas, pärast iga kaliibrimist ja iga kord,
sõltuvuses lihase massiga, oleneb produktsioon vanusest ja soost. Sarnase glomeerula kui kasutatakse uut reaktiivipudelit. Soovitav on kasutada kahetasemelisi kontrolle.
filtratsioonikiiruse juures on meeste kreatiniini kontsentratsioon suurem kui naistel, suurema Saadaval on järgmised kontrollproovid:
lihasmassiga isikutel suurem, kui väiksema lihasmassiga isikutel, noortel suurem kui vanuritel. Seerum/plasma:
Template: D01162_4
Olenevalt isiku poolt tarbitavast lihakogusest võib väärtust dieediga kuni 10% võrra mõjutada. Nortrol, koodiga: 981043
Üldiselt mõjutab kreatiniini varieerumine toidus isiku kreatiniinieritust siiski vaid vähesel määral. Abtrol, koodiga: 981044
Kõrgeid kreatiniiniväärtusi on avastatud akuutse ja kroonilise neerupuudulikkuse ja Uriin:
dehüdratsiooni korral. uTrol, kood: 981821
Kreatiniini mõõdetakse kehavedelikes nii keemilisel kui ensümaatilisel meetodil. Igapäevastes uTrol High, kood: 981822
tingimustes on kreatiniini võimalik ensümaatiliselt määrata vaid vähese sekkumise teel. Kontrollide vahemikud ja piirid tuleb kohaldada individuaalsete laborite vajadustele.
Kvaliteedikontrolliproovide tulemused peavad jääma labori poolt seatud piiridesse.
Page 11 D04282_03_Insert_Creatinine Enzymatic_MU
Uriin:
TULEMUSTE ARVUTAMINE Keskmine Keskmine
Tulemused arvutab Konelab analüsaator kaliibrimiskõvera järgi. 6.7 mmol/l 14.6 mmol/l
SD CV% SD CV%
Kaliibrimiskõver (näide) Katseseeriasisene 0.06 1.0 0.13 0.9
Eri partiid 0.05 0.8 0.17 1.2
KREATINIIN (ensümaatiline) Summaarne 0.24 3.5 0.51 3.4
Vastavalt CLSI (endine NCCLS) dokumendi EP5-A suunistele viidi Konelab 60 peal 20 päeva
0.120 jooksul läbi täpsusuuring kaliibrimisväärtusi ja operaatoreid vahetades, mõõtmiste arv uuringus oli
0.100 n = 80.
0.080
Abs
Meetodite võrdlus
0.060 Seerum/plasma:
0.040
CLSI (endine NCCLS) dokumendi EP9-A suuniseid järgides viidi läbi võrdlev uuring, kasutades
etalonina müügilolevat ensümaatilist meetodit.
0.020 Lineaarne regressioon (tulemuse ühik µmol/l):
0.000 y = 0.97 x – 0.3
r = 0.999
0 50 100 150 200 250
n = 110
µmol/l Proovide kontsentratsioonid jäid vahemikku 38 – 1951 µmol/l.
Uriin:
CLSI (endine NCCLS) dokumendi EP9-A suuniseid järgides viidi läbi võrdlev uuring, kasutades
Konelab 20/20XT/30/60. Kalibreerimiskõver oleneb partiist. etalonina müügilolevat ensümaatilist meetodit.
Lineaarne regressioon (tulemuse ühik mmol/l):
PROTSEDUURI PIIRANGUD y = 1.10 x + 0.02
r = 0.999
Segavad mõjud n = 119
Tingimus: saagis algväärtuse suhtes ±10%. Proovikontsentratsioonid jäid vahemikku 1.4 ja 33.2 mmol/l.
Seerum/plasma: KIRJANDUS
Konjugeerimata bilirubiin: kontsentratsioonini kuni 400 µmol/l (23.4 mg/dl) segavat mõju ei 1) Burtis, CA and Ashwood, E R (ed.), Tietz Fundamentals of Clinical Chemistry, 5th edition,
täheldatud. Bilirubiini kõrgema kontsentratsiooni korral võib analüüs W B Saunders Company, Philadelphia, 2001, pp. 419-422.
anda kreatiniini tegelikust madalama väärtuse. 2) Thomas, L. (ed.), Clinical Laboratory Diagnostics; Use and Assessment of Clinical
Konjugeeritud bilirubiin: kontsentratsioonini kuni 300 µmol/l (17.5 mg/dl) segavat mõju ei Laboratory Results, 1st edition, TH-Books Verlagsgesellschaft mbH, Frankfurt/Main,
täheldatud. Bilirubiini kõrgema kontsentratsiooni korral võib analüüs Germany, 1998, pp. 366-370.
anda kreatiniini tegelikust madalama väärtuse. 3) Guder WG, Narayanan S, Wisser H, Zawta B. List of Analytes; Preanalytical variables.
Hemoglobiin/hemolüüs: hemoglobiini kontsentratsioonini kuni 10 g/l segavat mõju ei Brochure in: Samples: From Patient to the Laboratory. GIT Verlag GmbH, Darmstadt, 1996.
täheldatud. 4) Young, D.S., Effects of Drugs on Clinical Laboratory Tests, Fifth Edition, AACC Press,
Lipeemia: Intralipid™ (Fresenius Kabi AB kaubamärk) sisalduseni kuni 10 g/l ja Washington, D.C., 2000, pp. 3-240 - 3-261.
triglütseriidi sisalduseni kuni 12 mmol/l (1062 mg/dl) segavat mõju ei 5) Mazzachi BC, Peake MJ, Ehrhardt V. Reference range and method comparison studies for
täheldatud. Korrelatsioon hägususe ja triglütseriidide sisalduse vahel enzymatic and Jaffé creatinine assays in plasma and serum and early morning urine. Clin
oli väike. Lab. 2000; 46: 53-55.
Analüüsi võivad mõjutada monokloonsed immunoglobuliinid (või nende osad).
Nimetatud antikehadega patsientide puhul saadud tulemusi tuleb hoolikalt hinnata. TOOTJA
Muid segavaid aineid on käsitletud viites 4. Thermo Fisher Scientific Oy
Clinical Diagnostics Finland
Uriin: Ratastie 2, P.O. Box 100, FI-01621 Vantaa, Finland
Konjugeeritud bilirubiin: kontsentratsioonini kuni 1000 µmol/l (58 mg/dl) ei ole segavat mõju Tel. +358 9 329 100, Faks +358 9 3291 0300
täheldatud. www.thermo.com/konelab
Hemoglobiin/hemolüüs: hemoglobiini kontsentratsioonini kuni 10 g/l ei ole hemolüsaadis
segavat mõju täheldatud. Teksti läbivaatamise kuupäev (aaaa-kk-pp)
Glükoos: kontsentratsioonini kuni 139 mmol/l (2500 mg/dl) ei ole segavat mõju 2007-08-07
täheldatud.
Askorbiinhape kontsentratsioonini kuni 5.7 mmol/l (100 mg/dl) ei ole segavat mõju Muudatused võrreldes eelmise versiooniga
täheldatud. Ettevõtte nimi uuendatud.
Kaltsium dobesilate ja á-metüüldopa põhjustavad kreatiniini tegelikust madalama väärtuse.
Muid segavaid aineid on käsitletud viites 4.
Teisendamistegur: 981845 4 x 60 ml
µmol/l x 0.0113 Æ mg/dl
mmol/l x 11.3 Æ mg/dl EZ A TÁJÉKOZTATÓ AZ EGYESÜLT ÁLLAMOKON
MÕÕTEPIIRKOND KÍVÜLI HASZNÁLATRA VONATKOZIK. A KONELAB
Seerum/plasma:
10 – 2500 µmol/l (0.11 – 28 mg/dl)
RENDSZEREKRE TETT MINDEN UTALÁS A T
Pikendatud mõõtepiirkond pärast sekundaarset lahjendust: 10 – 10000 µmol/l (0.11 – SOROZATRA IS VONATKOZIK.
113 mg/dl)
Uriin:
0.2 – 40 mmol/l (2.3 – 452 mg/dl) RENDELTETÉS
Pikendatud mõõtepiirkond pärast sekundaarset lahjendust: 0.01 – 200 mmol/l (0.11 – Emberi szérum-, plazma, illetve vizelet kreatinin koncentrációjának in vitro kvantitatív
2260 mg/dl) enzimatikus, Konelab analizátorokban történő meghatározásához. Minden vizsgálati eredményt a
jooksul läbi täpsusuuring kaliibrimisväärtusi ja operaatoreid vahetades, mõõtmiste arv uuringus oli Szarkozin + H2O + O2 ⎯-⎯⎯⎯-⎯-⎯-⎯⎯> Glicin + HCHO + H2O2
n = 80.
Peroxidáz
H2O2 + 4-aminofenazon + HTIB* ⎯-⎯⎯⎯-⎯> Kinonimin kromogén + H2O + HI
* HTIB = 2,4,6-trijód-3-hidroxibenzoesav
Page 12 D04282_03_Insert_Creatinine Enzymatic_MU
Abs
Koncentrációk
A reagens 0.060
TAPS** puffer, pH 8.1 30 mmol/l 0.040
Kreatináz (mikroorganizmus) > 333 μkat/l 0.020
Szarkozin-oxidáz (mikroorganizmus) > 133 μkat/l
Aszkorbát-oxidáz (mikroorganizmus) > 33 μkat/l 0.000
HTIB 5.9 mmol/l 0 50 100 150 200 250
Detergensek
Tartósítószer µmol/l
B reagens:
TAPS** puffer, pH 8.0 50 mmol/l Konelab 20/20XT/30/60. A kalibrációs görbe tételszámtól függ.
Kreatinináz (mikroorganizmus) > 500 μkat/l
Peroxidáz (torma) > 16.7 μkat/l AZ ELJÁRÁS KORLÁTAI
4-aminofenazon 2.0 mmol/l
Kálium-hexacianoferrát (II) 163 μmol/l Interferencia
Detergens Kritérium: Visszanyerés a kezdeti értékekhez képest ± 10%
Tartósítószer
Szérum/plazma:
** TAPS = 3-[[trisz(hidroximetil)-metil]-amino]-propánszulfonsav Nem konjugált bilirubin: Nem észleltek interferenciát 400 µmol/l (23.4 mg/dl) értékig.
A agasabb bilirubin-koncentrációk tévesen alacsony kreatinin-
Óvintézkedések értékeket okoznak.
Kizárólag in vitro diagnosztikus használatra. Tartsa be a laboratóriumi reagensek kezelésére Konjugált bilirubin: Nem észleltek interferenciát 300 µmol/l (17.5 mg/dl) értékig.
vonatkozó szokásos előírásokat. A agasabb bilirubin-koncentrációk tévesen alacsony kreatinin-
értékeket okoznak.
Előkészítés Hemoglobin / hemolízis: Nem észleltek interferenciát a hemolizátumban 10 g/l
A reagensek használatra készek. hemoglobinszintig.
Megjegyzés: Ellenőrizze, hogy nincs-e buborék az üveg nyakánál vagy a reagens felszínén, Lipémia: Nem mutatható ki interferencia 10 g/l Intralipid™ (a Fresenius Kabi
amikor reagenst tartalmazó üvegeket, illetve edényeket tesz a Konelab analizátorba. AB védjegye) vagy 12 mmol/l (1062mg/dl) triglicerid értékig. Gyenge
összefüggés van a zavarosság és a trigliceridkoncentráció között.
Tárolás és stabilitás Monoklonális immunglobulinok (vagy azok részei) zavarhatják a vizsgálatot. Amennyiben
Felbontatlan üvegben tárolt reagensek 2…8 °C között a címkén feltüntetett lejárati időpontig előfordulhat a betegnél ilyen antitestek jelenléte, akkor az eredményeket óvatosan kell értékelni.
stabilak. A reagensek elemzőkészüléken belüli stabilitásához olvassa el a Konelab analizátorhoz Egyéb interferenciát okozó anyagok tekintetében lásd a 4. referenciát.
tartozó Alkalmazási tudnivalókat.
Vizelet:
MINTAVÉTEL Konjugált bilirubin: Nem észleltek interferenciát 1000 µmol/l (58 mg/dl) értékig.
Hemoglobin / hemolízis: Nem észleltek interferenciát a hemolizátumban 10 g/l
A minta típusa hemoglobinszintig.
Csak az alább megadott minták vizsgálata történt meg, és ezek találtattak elfogadhatónak: Glukóz: Nem észleltek interferenciát 139 mmol/l (2500 mg/dl) értékig.
Szérum, Li-heparinos plazma, Na-EDTA-s plazma és vizelet. Aszkorbinsav: Nem észleltek interferenciát 5.7 mmol/l (100 mg/dl) értékig.
A kalcium-dobezilαt ιs az alfa-metildopa tιvesen alacsony kreatinin-ιrtιkeket okoznak.
ÜGYELJEN RÁ, hogy kizárólag megfelelő kémcsöveket, illetve gyűjtőedényeket használjon Egyéb interferenciát okozó anyagok tekintetében lásd a 4. referenciát.
mintavételre és előkészítésre. A vizeletmintát adalékanyagok nélkül kell levenni.
VÁRHATÓ ÉRTÉKEK (5)
Óvintézkedések Szérum/plazma:
Az emberi mintákat úgy kell kezelni és kidobni, mintha azok fertőzőek lennének. Férfi: 59 – 104 μmol/l (0.67 – 1.17 mg/dl)
Nυ: 45 – 84 μmol/l (0.51 – 0.95 mg/dl)
Tárolás (3)
A szérum- és plazmaminták 20…25 °C-on, illetve 4…8 °C-on 7 napig, -20 °C-on pedig Vizelet (1. reggeli vizelet):
3 hónapig tárolhatók. Férfi: 3.54 – 24.6 mmol/l (40 – 278 mg/dl)
A vizeletminta 20…25 °C-on 2 napig, 4…8 °C-on 6 napig, –20 °C-on pedig 6 hónapig tárolható. Nő: 2.55 – 20.0 mmol/l (29 – 226 mg/dl)
A VIZSGÁLATI ELJÁRÁS A feltüntetett értékek kizárólag tájékoztató jellegűek. Minden laboratórium esetében ajánlott a
Automatizált eljárás kivitelezéséhez olvassa el a Konelab analizátorhoz tartozó Használati normál tartomány megerősítése, illetve a kiszolgált populációra jellemző referenciaintervallum
kézikönyvet és a külön szérumra/plazmára és vizeletre vonatkozó Alkalmazási tudnivalókat. A kialakítása.
legfrissebb alkalmazásokat az interneten találja meg. Nem garantálható semmilyen olyan
alkalmazás eredménye, amelyet a Thermo Fisher Scientific Oy nem hagyott jóvá, ezért ezeket a Átváltási tényező:
felhasználónak kell értékelnie. µmol/l x 0.0113 Æ mg/dl
mmol/l x 11.3 Æ mg/dl
Szolgáltatott anyagok
A fent leírt reagensek. MÉRÉSI TARTOMÁNY
Szérum/plazma:
Szükséges, de nem szolgáltatott anyagok 10 – 2500 μmol/l (0.11 – 28 mg/dl)
Mosófolyadék, kód: 981842. Kiterjesztett mιrιsi tartomαny mαsodlagos hνgνtαs utαn: 10 – 10000 μmol/l (0.11 – 113 mg/dl)
Az alább leírt kalibrátorok és kontrollok. Vizelet:
0.2 – 40 mmol/l (2.3 – 452 mg/dl)
Kalibrálás Kiterjesztett mérési tartomány másodlagos hígítás után: 0.01 – 200 mmol/l (0.11 – 2260 mg/dl)
Használjon sCal oldatot (kód: 981831) a Konelab analizátorhoz tartozó tájékoztató szerint.
Legfeljebb egy órára hagyja az analizátorban a kalibrátort. TELJESÍTMÉNY JELLEMZŐK
Az egyes laboratóriumokban nyert értékek különbözhetnek a megadott teljesítmény adatoktól.
Nyomonkövethetőség:
Olvassa el az sCal oldat csomagjában található tájékoztatót. Detektálási határérték
A válasz (A) egy számítási tényező segítségével eredményegységekre átszámítva jelenik meg. Szérum/plazma:
2 μmol/l (0.02 mg/dl)
Minőségellenőrzés
A CLSI (régen NCCLS) EP5-A dokumentumának előírásai alapján pontossági mérés készült
20 napon keresztül egy Konelab 60 analizátorral, változó kalibrálásokkal és felhasználókkal, 80-
as mérésszámmal (n=80).
Template: D01162_4
Page 13 D04282_03_Insert_Creatinine Enzymatic_MU
Vizelet: La creatinina viene convertita in sarcosina con l’aiuto della creatininasi e creatinasi. La
Középérték Középérték sarcosina è quindi convertita in glicina, formaldeide e perossido di idrogeno in presenza di
6.7 mmol/l 14.6 mmol/l ossigeno mediante sarcosina ossidasi. Il perossido di idrogeno liberato reagisce con 4-
SD CV% SD CV% aminofenazone e HTIB per formare un chinone immino cromogeno in una reazione catalizzata
Mérés közben 0.06 1.0 0.13 0.9 mediante perossidasi. L’intensità cromatica è direttamente proporzionale alla concentrazione di
Futtatások közt 0.05 0.8 0.17 1.2 creatinina presente e può esser misurata fotometricamente a 540nm.
Teljes 0.24 3.5 0.51 3.4 INFORMAZIONI SUI REAGENTI
Reagente A 4 x 40 ml
A CLSI (régen NCCLS) EP5-A dokumentumának előírásai alapján pontossági mérés készült Reagente B 4 x 20 ml
20 napon keresztül egy Konelab 60 analizátorral, változó kalibrálásokkal és felhasználókkal, 80-
as mérésszámmal (n=80). Concentrazioni
Reagente A
Módszerösszehasonlítás Tampone TAPS**, pH 1.8 30 mmol/l
Szérum/plazma: Creatinasi (microorganismi) > 333 μkat/l
Összehasonlítási tanulmány készült a CLSI (régen NCCLS) EP9-A dokumentumának előírásai Sarcosina ossidasi (microorganismi) > 133 μkat/l
alapján, referenciaként egy másik forgalomban lévő enzimatikus módszert használva. Ascorbato ossidasi (microorganismi) > 33 µkat/l
Lineαris regressziσ (eredmιny mιrtιkegysιge μmol/l): HTIB 5.9 mmol/l
y = 0.97 x – 0.3 Detergenti
r = 0.999 Conservante
n = 110
A mintakoncentrációk 38 ιs 1951 μmol/l között voltak. Reagente B:
Tampone TAPS**, pH 8.0 50 mmol/l
Vizelet: Creatininasi (microorganismi) > 500 μkat/l
Összehasonlítási tanulmány készült a CLSI (régen NCCLS) EP9-A dokumentumának előírásai Perossidasi (Horseradish) > 16.7 μkat/l
alapján, referenciaként egy másik forgalomban lévő enzimatikus módszert használva. 4-aminofenazone 2.0 mmol/l
Lineáris regresszió (eredmény egység mmol/l): Potassio esacianoferrato (II) 163 μmol/l
y = 1.10 x +0.02 Detergente
r = 0.999 Conservante
n = 119
A mintakoncentrációk 1.4 és 33.2 mmol/l között voltak. ** TAPS = acido 3-[[tris(idrossimetil)metil]amino]propansulfonico
BIBLIOGRÁFIA Precauzioni
1) Burtis, CA and Ashwood, E R (ed.), Tietz Fundamentals of Clinical Chemistry, 5th edition, W Solo per uso diagnostico in vitro. Rispettare le normali precauzioni previste per l'utilizzo di tutti i
B Saunders Company, Philadelphia, 2001, pp. 419-422. reagenti di laboratorio.
2) Thomas, L. (ed.), Clinical Laboratory Diagnostics; Use and Assessment of Clinical
Laboratory Results, 1st edition, TH-Books Verlagsgesellschaft mbH, Frankfurt/Main, Preparazione
Germany, 1998, pp. 366-370. I reagenti sono pronti all'uso.
3) Guder WG, Narayanan S, Wisser H, Zawta B. List of Analytes; Preanalytical variables. Nota: Controllare che non siano presenti bolle sul collo del flacone o sulla superficie del
Brochure in: Samples: From Patient to the Laboratory. GIT Verlag GmbH, Darmstadt, 1996. reagente durante l'inserimento di vial o recipienti di reagente nell'analizzatore Konelab.
4) Young, D.S., Effects of Drugs on Clinical Laboratory Tests, Fifth Edition, AACC Press,
Washington, D.C., 2000, pp. 3-240 - 3-261. Conservazione e stabilità
5) Mazzachi BC, Peake MJ, Ehrhardt V. Reference range and method comparison studies for I reagenti in flaconi intatti sono stabili fino alla data di scadenza indicata sull'etichetta, se
enzymatic and Jaffé creatinine assays in plasma and serum and early morning urine. Clin conservati a una temperatura di 2…8 °C. Per la stabilità dei reagenti sullo strumento, vedere le
Lab. 2000; 46: 53-55. note applicative dell'analizzatore Konelab in uso.
GYÁRTÓ RACCOLTA DEL CAMPIONE
Thermo Fisher Scientific Oy
Clinical Diagnostics Finland Tipo di campione
Ratastie 2, P.O. Box 100, FI-01621 Vantaa, Finland Solo i campioni sotto forniti sono stati testati e giudicati accettabili:
Tel. +358 9 329 100, Fax +358 9 3291 0300 Siero, plasma Li-eparina, plasma Na-EDTA e urina.
www.thermo.com/konelab
NOTA per la raccolta e la preparazione dei campioni, utilizzare unicamente provette o
Ellenőrzés időpontja (éééé-hh-nn) contenitori di raccolta idonei. Raccogliere l’urina senza conservanti.
2007-08-07
Precauzioni
Változtatások az előző változathoz képest I campioni umani devono essere maneggiati e smaltiti come campioni potenzialmente infetti.
A cég neve frissítésre került.
Conservazione (3)
I campioni di siero e di plasma possono essere conservati per 7 giorni a 20…25 °C o a 4…8 °C
o per 3 mesi a –20 °C.
I campioni di urina possono essere conservati per 2 giorni a 20…25 °C, per 6 giorni a 4…8 °C
o per 6 mesi a –20 °C.
PROCEDURA ANALITICA
IT Per le procedure automatiche dell'analizzatore Konelab per siero/plasma e urina consultare il
Konelab™ / Serie T manuale d'uso e le note applicative. Vedere le più recenti applicazioni sulla pagina web. Tutte le
applicazioni non esplicitamente approvate da Thermo Fisher Scientific Oy, non possono essere
CREATININE (Enzymatic) garantite in termini di prestazioni e dovranno pertanto essere valutate dall'utilizzatore.
CREATININA (metodo enzimatico) Materiali inclusi nel kit
I reagenti sopra descritti.
981845 4 x 60 ml
Materiali necessari ma non inclusi nel kit
WashFluid, codice: 981842.
IL PRESENTE INSERTO NELL’IMBALLO PUO’ Calibratore e controlli indicati di seguito.
ESSERE APPLICATO AL DI FUORI DEGLI STATI UNITI.
Calibrazione
EVENTUALI RIFERIMENTI A KONELAB SYSTEMS SI Usare Calibratore sCal codice 981831 secondo le istruzioni fornite con l'analizzatore Konelab in
RIFERISCONO ANCHE ALLA SERIE T. uso. Lasciare riposare il calibratore sull’analizzatore per un tempo massimo di un’ora.
Tracciabilità:
USO CONSIGLIATO Fare riferimento all'inserto nell'imballo del Calibratore sCal.
Prodotto impiegato per la determinazione quantitativa in vitro della concentrazione di creatinina La risposta (A) è convertita nelle unità di misura del risultato per mezzo di un fattore di calcolo.
nel siero, nel plasma o nell'urina umani su analizzatori Konelab con metodo enzimatico. Tutti i
Sarcosina ossidasi
Sarcosina + H2O + O2 ⎯-⎯⎯⎯-⎯-⎯-⎯⎯> Glicina + HCHO + H2O2
Perossidasi
H2O2 + 4-aminofenazone + HTIB* ⎯-⎯⎯⎯-⎯> Chinone immino cromogeno + H2O + HI
Urina:
Curva di calibrazione (esempio) Media Media
6.7 mmol/l 14.6 mmol/l
CREATININA (metodo enzimatico) DS CV% DS CV%
Entro la serie 0.06 1.0 0.13 0.9
Fra la serie 0.05 0.8 0.17 1.2
0.120 Totale 0.24 3.5 0.51 3.4
0.100
Lo studio di precisione è stato eseguito secondo il protocollo CLSI (ex NCCLS) Documento
0.080
Abs
Lo studio di precisione è stato eseguito secondo il protocollo CLSI (ex NCCLS) Documento Kreatinazė
EP5-A su un analizzatore Konelab 60 per un periodo di 20 giorni, includendo varie calibrazioni e Kreatinas + H2O ⎯-⎯-⎯-⎯-⎯> Sarkozinas + Karbamidas
operatori, per un numero di misurazioni pari a n = 80.
Template: D01162_4
Sarkozino oksidazė
Sarkozinas + H2O + O2 ⎯-⎯⎯⎯-⎯-⎯-⎯⎯> Glicinas + HCHO + H2O2
Peroksidazė
H2O2 + 4-aminofenazonas + HTIB* ⎯-⎯⎯⎯-⎯> Kvinono imino chromogenas + H2O + HI
Kreatininas virsta sarkozinu kreatininazės ir kreatinazės pagalba. Jei yra deguonies, tai PROCEDŪROS RIBOTUMAI
sarkozino oksidazė toliau verčia sarkoziną i gliciną, formaldehidą ir vandenilio peroksidą.
Išlaisvintas vandenilio peroksidas reaguoja su 4-aminofenazonu ir HTIB, reakciją katalizuoja Interferencija
peroksidazė ir susidaro kvinono imino chromogenas. Spalvos intensyvumas yra tiesiogiai Kriterijus: Pradinių verčių atstatymas ± 10 % ribose.
proporcingas esančio kreatinino koncentracijai ir gali būti nustatomas fotometriškai prie 540 nm.
Serumas / plazma:
INFORMACIJA APIE REAGENTUS Nekonjuguotasis bilirubinas: Iki 400 µmol/l (23.4 mg/dl) interferencija nepastebėta. Aukštesnės
A reagentas 4 x 40 ml bilirubino koncentracijos sukelia klaidingai žemas kreatinino
B reagentas 4 x 20 ml vertes.
Konjuguotasis bilirubinas: Iki 300 µmol/l (17.5 mg/dl) interferencija nepastebėta. Aukštesnės
Koncentracijos bilirubino koncentracijos sukelia klaidingai žemas kreatinino
A reagentas vertes.
TAPS buferis, pH 8.1 30 mmol/l Hemoglobinas/Hemolizė: Iki 10 g/l hemoglobino hemolizate interferencija nepastebėta.
Creatinase (microorganisms) > 333 μkat/l Lipemija : Iki 10 g/l Intralipid™ (Fresenius Kabi AB prekės ženklas) arba
Sarcosine oxidase (microorganisms) > 133 μkat/l 12 mmol/l (1062 mg/dl) trigliceridų interferencija nepastebėta.
Ascorbate oxidase (microorganisms) > 33 μkat/l Koreliacija tarp drumstimosi ir trigliceridų koncentracijos labai
HTIB 5.9 mmol/l nedidelė.
Detergents Reakcijai gali trukdyti heterofiliniai antikūniai ir monokloniniai imunoglobulinai (ar jų dalys).
Preservative Pacientų, įtariamai turinčių tokių antikūnių, rezultatus reikia atsargiai įvertinti.
Daugiau apie interferencijos medžiagas skaitykite 4 nuorodoje.
B reagentas:
TAPS** buffer, pH 8.0 50 mmol/l Šlapimas:
Creatininase (microorganisms) > 500 μkat/l Konjuguotasis bilirubinas: Iki 1000 µmol/l (58 mg/dl) interferencija nepastebėta.
Peroxidase (horseradish) > 16.7 μkat/l Hemoglobinas/Hemolizė: Iki 10 g/l hemoglobino hemolizate interferencija nepastebėta.
4-aminophenazone 2.0 mmol/l Gliukozė: Iki 139 mmol/l (2500 mg/dl) interferencija nepastebėta.
Potassium hexacyanoferrate (II) 163 μmol/l Askorbo rūgštis: Iki 5.7 mmol/l (100 mg/dl) interferencija nepastebėta.
Detergentas Kalcio dobesilatas ir α-metildopa duoda klaidingai žemas kreatinino vertes.
Preservative Daugiau apie interferencijos medžiagas skaitykite 4 nuorodoje.
** TAPS = 3-[[tri(hidroksimetil)metil]amino]propansulfonrūgštis NUMATOMOS VERTĖS (5)
Serumas / plazma:
Atsargumo priemonės Vyras: 59 – 104 μmol/l (0.67 – 1.17 mg/dl)
Tik in vitro diagnostiniam naudojimui. Laikykitės įprastų atsargumo priemonių, kurios būtinos Moteris: 45 – 84 μmol/l (0.51 – 0.95 mg/dl)
dirbant su laboratorijos reagentais.
Šlapimas (pirmas rytinis šlapimas):
Paruošimas Vyras: 3.54 – 24.6 mmol/l (40 – 278 mg/dl)
Reagentai paruošti naudojimui. Moteris: 2.55 – 20.0 mmol/l (29 – 226 mg/dl)
Pastaba: Kai dedate reagento buteliukus ar indus į Konelab analizatorių, patikrinkite ar prie
buteliuko kaklelio arba reagento paviršiuje nėra burbuliukų. Šios vertės turi būti naudojamos tik kaip orientyras. Rekomenduojama, kad kiekviena
laboratorija patikrintų šį diapazoną arba išvestų atskaitos intervalą populiacijai, kuriai ji tarnauja.
Saugojimas ir pastovumas
Reagentai uždarytuose buteliukuose yra pastovūs prie 2…8 °C temperatūros iki galiojimo
datos, užrašytos ant etiketės. Savo Konelab analizatoriaus, Pastabose apie pritaikymą, skaitykite MATAVIMO SRITIS
apie prietaise patalpintų reagentų pastovumą. Perskaièiavimo koeficientas:
µmol/l x 0.0113 Æ mg/dl
MĖGINIŲ SURINKIMAS mmol/l x 11.3 Æ mg/dl
Tik žemiau išvardyti mėginiai buvo patikrinti ir pripažinti tinkamais:
Serumas, ličio heparinas plazma, Na-EDTA plazma ir šlapimas Serumas / plazma:
10 – 2500 μmol/l (0.11 – 28 mg/dl)
PASTABA: mėginių surinkimui ir paruošimui naudokite tik tinkamus mėgintuvėlius arba Praplėsta matavimo sritis antrą kartą atskiedus: 10 – 10000 μmol/l (0.11 – 113 mg/dl)
saugojimo talpas. Šlapimo mėginius paimkite be priedų. Šlapimas:
0.2 – 40 mmol/l (2.3 – 452 mg/dl)
Atsargumo priemonės Praplėsta matavimo sritis antrą kartą atskiedus: 0.01 – 200 mmol/l (0.11 – 2260 mg/dl)
Su žmonių mėginiais reikia dirbti ir juos šalinti taip, lyg jie būtų potencialiai užkrečiami.
EKSPLOATACIJOS CHARAKTERISTIKOS
Laikymas (3) Individualiose laboratorijose gauti rezultatai gali skirtis nuo pateiktų duomenų.
Serumo ir plazmos mėginius galima laikyti 7 dienas esant 20…25 °C temperatūrai arba
4…8 °C arba 3 mėnesius esant –20 °C. Aptikimo ribos
Šlapimo mėginį galima laikyti 2 dienas esant 20…25 °C, 6 dienas prie 4…8 °C arba 6 2 µmol/l (0.02 mg/dl)
mėnesius esant –20 °C. Aptikimo riba žymi žemiausią išmatuojamą koncentraciją/ aktyvumą, kurį galima atskirti nuo
nulio. Ji apskaičiuojama kaip nulinio mėginio koncentracija + 3 SD (serijoje, n=24).
TYRIMO TVARKA Šlapimas:
Apie automatizuotą serumo/plazmos ir šlapimo procedūrą naudojant Konelab analizatorių 0.002 mmol/l (0.02 mg/dl)
skaitykite Nuorodų vadove ir Pastabose apie Pritaikymą. Apie vėliausius taikymus skaitykite Aptikimo riba žymi žemiausią išmatuojamą koncentraciją/ aktyvumą, kurį galima atskirti nuo
gamintojo tinklalapyje. Jei prietaisas naudojamas procedūroms, kurių nepatvirtino Thermo Fisher nulio. Ji apskaičiuojama kaip nulinio mėginio koncentracija + 3 SD (serijoje, n=24)
Scientific Oy, negalima garantuoti tų procedūrų sėkmės ir tokį pritaikymą turi įvertinti vartotojas.
Netikslumas
Skiriamos medžiagos Serumas / plazma:
Prieš tai aprašyti reagentai.
Vidutinis Vidutinis Vidutinis
Būtinos, bet neskiriamos medžiagos 38 μmol/l 155 μmol/l 484 μmol/l
WashFluid, code: 981842. 1 sn CV% 1 sn CV% 1 sn CV%
Toliau aprašytos kontrolinės medžiagos ir kalibratorius. Serijoje 0.56 1.4 0.69 0.4 1.85 0.4
Kas seriją 0.16 0.4 0.71 0.5 1.30 0.3
Kalibravimas
Viso 0.84 2.1 2.32 1.5 6.76 1.4
Naudokite sCal, kodas 981831 pagal Konelab analizatoriaus instrukcijas. Kalibratorių laikykite
analizatoriuje daugiausiai vieną valandą.
Tikslumo studija atlikta pagal gaires, išdėstytas CLSI (anksčiau NCCLS) Document EP5-A,
Sietis: keičiant kalibravimus ir operatorius ir naudojant vieną Konelab 60 20 dienų, matavimų skaičius
Žr. sCal pakuotės informacinį lapelį. buvo n= 80.
Atsakas (dA/min.) paverčiamas rezultato vienetais, pritaikant skaičiavimo koeficientą.
Šlapimas:
Kokybės valdymas Vidutinis Vidutinis
Naudokite kokybės valdymo mėginius bent kartą per dieną ir po kiekvieno kalibravimo, taip pat 6.7 mmol/l 14.6 mmol/l
kiekvieną kartą, kai naudojate naują reagento buteliuką. Rekomenduojama taikyti dviejų lygių 1 sn CV% 1 sn CV%
KREATININAS (fermentinis)
0.120
0.100
0.080
Abs
0.060
Template: D01162_4
0.040
0.020
0.000
0 50 100 150 200 250
µmol/l
Šlapimas: Reaģents B:
Palyginimų studija atlikta, kaip gaires naudojant NCCLS dokumentą EP9-A ir, kaip etaloną TAPS** buferšķīdums, pH 8.0 50 mmol/l
naudojant komerciškai prieinamą nefelometrinį metodą. Kreatinīns (mikroorganismu) > 500 μkat/l
Linijinė regresija (rezultatas mmol/l): Peroksidāze (mārrutks) > 16.7 μkat/l
y = 1.10 x + 0.02 4-aminofenazons 2.0 mmol/l
r = 0.999 Kālija heksacianoferāts (II) 163 μmol/l/
n = 119 Mazgāšanas līdzeklis
Mėginio koncentracijos buvo nuo 1.4 iki 33.2 mmol/l. Konservants
** TAPS = 3-[[tris(hidroksimetil)metil]amino]propānsulfoskābe
BIBLIOGRAFIJA
1) Burtis, CA ir Ashwood, E R (red.), Tietz klinikinės chemijos pagrindai, 5 leidimas, W B Piesardzības pasākumi
Saunders kompanija, Filadelfija, 2001, psl. 419-422. Lietot tikai “in vitro “diagnostikas vajadzībām. Nodrošiniet parastos piesardzības pasākumus,
2) Thomas, L. (ed.), Clinical Laboratory Diagnostics; (red.), Klinikinė laboratorijų diagnostika; kas jāievēro attiecībā uz rīkošanos ar visiem laboratorijas reaģentiem.
Klinikinių laboratorijos rezultatų naudojimas ir įvertinimas, 1 leidimas, TH-Books
Verlagsgesellschaft mbH, Frankfurt/Main, Vokietija 1998, psl. 366-370. Sagatavošana
3) Guder WG, Narayanan S, Wisser H, Zawta B. Analitų sąrašas; Preanalitiniai kintamieji; Reaģenti ir gatavi lietošanai.
Preanalytical variables. Mėginių: brošiūra: iš pacientas laboratorijai. GIT Verlag GmbH, Piezīme: Ievietojot reaģenta pudelītes vai traukus Konelab analizatorā, pārliecinieties, ka
Darmstatas. pudelītes kakliņā vai uz virsmas nav reaģenta burbuļi.
4) Young, D.S., Vaistų įtaka klinikiams laboratorijų tyrimams, penktasis leidimas, AACC Press,
Wašingtonas, 2000, 3-240 - 3-261. Uzglabāšana un stabilitāte
5) Mazzachi BC, Peake MJ, Ehrhardt V. Reference range and method comparison studies for Reaģenti neatvērtās pudelītēs ir stabili pie T 2…8 °C līdz uzglabāšanas datuma beigām, kas
enzymatic and Jaffé creatinine assays in plasma and serum and early morning urine. Clin uzdrukāts uz etiķetes. Vadieties pēc sava Konelab analizatora pielietojuma piezīmēm, lai iegūtu
Lab. 2000; 46: 53-55. informāciju par ievietoto reaģentu stabilitāti.
GAMINTOJAS PARAUGU IEGŪŠANA/SA
Thermo Fisher Scientific Oy
Clinical Diagnostics Finland Parauga tips
Ratastie 2, P.O. Box 100, FI-01621 Vantaa, Suomija Vienīgi zemāk dotie paraugi bija pārbaudīti un atdzīti par pieņemamiem:
Tel. +358 9 329 100, Faksas +358 9 3291 0300 Serums, Li-heparīna plazma, Na-EDTA plazma un urīns.
www.thermo.com/konelab
PIEZĪME paraugu iegūšanai un sagatavošanai, lieto vienīgi piemērotus stobriņus vai
Peržiūros data (mmmm-mm-dd) savākšanas konteinerus. Urīnu savākt bez konservējošiem līdzekļiem.
2007-08-07
Piesardzības pasākumi
Pokyčiai nuo ankstesnės versijos Ar no cilvēkiem iegūtajiem paraugiem ir jārīkojas un no tiem jāatbrīvojas tā, it kā tie būtu
Atnaujintas kompanijos pavadinimas. iespējami infekciozi.
Uzglabāšana (3)
Seruma un plazmas paraugus var uzglabāt 7 dienas T no 20 līdz 25 °C vai no 4 līdz 8 °C
temperatūrā vai 3 mēnešus –20 °C temperatūrā.
Urīna paraugu var uzglabāt 2 dienas pie T no 20…25 °C, 6 dienas pie T no 4…8 °C
temperatūrā vai 6 mēnešus –minus 20°C temperatūrā.
LV PĀRBAUDES PROCEDŪRA
Vadieties pēc Uzziņas Rokasgrāmatas un Pielietojuma Piezīmēm, atsevišķi priekš
Konelab™ / T Sērijas seruma/plazmas un urīna; lai iegūtu informāciju par Konelab analizatora automātisko procedūru.
Skatieties pēdējo pielietošanu no web (interneta) lapām. Jebkāds pielietojums, kuru nav
CREATININE (Enzymatic) apstiprinājis Thermo Fisher Scientific Oy, nevar sniegt garantētu rezultātu, tādēļ tas jāizvērtē
pašam lietotājam.
Kreatinīns (Enzimātiskais)
981845 4 x 60 ml Piegādātie materiāli
Iepriekš aprakstītie reaģenti.
Kreatinīns (Enzimātiskais)
Kreatinīns
Kreatinīns + H2O ⎯-⎯-⎯-⎯-⎯> Sarkozīns + urea
0.120
Sarkozīna oksidāze 0.100
Sarkozīns + H2O + O2 ⎯-⎯⎯⎯-⎯-⎯-⎯⎯> Glicīns + HCHO + H2O2 0.080
Abs
0.060
Peroksidāze
0.040
H2O2 + 4-aminofenazons + HTIB* -⎯⎯⎯-⎯> Kvinona imīna hromogēns + H2O + HI
0.020
* HTIB = 2,4,6-trijod-3-hidroksibenzoskābe 0.000
0 50 100 150 200 250
Kreatinīns tiek pārvērsts par sarkozīnu ar kreatinināzes un kreatināzes palīdzību. Sarkozīns ar
sarkozīna oksidāzi skābekļa klātbūtnē tiek tad pārvērsts par glicīnu, formaldehīdu ar hidrogēna µmol/l
Template: D01162_4
peroksīdu. Brīvais hidrogēna peroksīds reağē ar 4-aminofenazīnu un HTIB veido kvinona imīna
hromogēnu un reakcijā katalizējas par peroksidāzi. Krāsas intensivitāte ir tieši proporcionāla
klātesošā kreatinīna koncentrācijai. un var būt fotometriski izmērīta pie 540 nm.
Konelab 20/20XT/30/60. Kalibrēšanas līkne ir atkarīga no daudzuma.
REAĢENTU INFORMĀCIJA
Reaģents A 4 x 40 ml
Reaģents B 4 x 20 ml
Page 17 D04282_03_Insert_Creatinine Enzymatic_MU
Precizitātes izpēte tika veikta saskaņā ar noteikumiem CLCI (veidots NCCLS) Dokumentā kreatynaza
EP5-A, mainot kalibrētājus un operatorus lietojot Konelab 60 vismaz 20 dienas, kur mērījumu kreatyna + H2O ⎯-⎯-⎯-⎯-⎯> sarkozyna + mocznik
skaitlis bija n = 80.
oksydaza sarkozynowa
Urīns: sarkozyna + H2O + O2 ⎯-⎯⎯⎯-⎯-⎯-⎯⎯⎯> glicyna + HCHO + H2O2
Vidējais Vidējais
n = 119 Odczynnik B:
Paraugu koncentrācija bija starp 1.4 un 33.2 mmol/l. Bufor TAPS***; pH 8.0 50 mmol/l
Kreatyninaza (mikroorganizmy) > 500 ěkat/l
BIBLIOGRĀFIJA Peroksydaza (chrzan) > 16.7 ěkat/l
1) Burtis, CA and Ashwood, E R (ed.), Tietz Fundamentals of Clinical Chemistry, 5th edition, W 4-aminofenazon 2.0 mmol/l
B Saunders Company, Philadelphia, 2001, pp. 419-422. Heksacjanożelazian potasu (II) 163 ěmol/l
Detergent
Konserwant
Page 18 D04282_03_Insert_Creatinine Enzymatic_MU
UWAGA dotycząca pobierania i przygotowywania próbek: należy używać wyłącznie Przedstawione wielkości należy traktować jedynie jako wartości orientacyjne. Zaleca się, by w
odpowiednich probówek lub pojemników przewidzianych do tego celu. Mocz pobierać bez każdej pracowni analitycznej zweryfikować podane zakresy lub ustalić własny przedział
dodatków. referencyjny dla badanej populacji.
Krzywa kalibracyjna (przykład) Badanie precyzji przeprowadzono zgodnie z wytycznymi zawartymi w dokumencie CLSI
(poprzednio NCCLS) EP5-A, przy użyciu analizatora Konelab 60, wykonując n = 80 pomiarów w
KREATYNINA (enzymatyczna)
ciągu 20 dni, dla różnych kalibracji i różnych operatorów.
Porównanie metod
0.120 Surowica/osocze:
Przeprowadzono badanie porównawcze według wytycznych zawartych w dokumencie CLSI
poziomów kreatyniny. 4) Young, D.S., Effects of Drugs on Clinical Laboratory Tests, Fifth Edition, AACC Press,
Hemoglobina / hemoliza: nie stwierdzono zakłóceń przy poziomie hemoglobiny w hemolizacie Washington, D.C., 2000, pp. 3-240 - 3-261.
do 10 g/l. 5) Mazzachi BC, Peake MJ, Ehrhardt V. Reference range and method comparison studies for
Lipemia: nie stwierdzono zakłóceń w przypadku stężeń Intralipidu™ (znak enzymatic and Jaffé creatinine assays in plasma and serum and early morning urine. Clin
towarowy firmy Fresenius Kabi AB) do poziomu 10 g/l lub 12 mmol/l Lab. 2000; 46: 53-55.
(1062mg/dl) of triglicerydów Istnieje słaba korelacja pomiędzy
stopniem zmętnienia a stężeniem triglicerydów.
Immunoglobuliny monoklonalne (lub ich części) mogą wpływać zakłócająco na test. Wyniki
badań pacjentów, u których podejrzewa się występowanie takich przeciwciał należy uważnie
Page 19 D04282_03_Insert_Creatinine Enzymatic_MU
Precauções
As amostras humanas devem ser manuseadas e eliminadas como se fossem potencialmente
infecciosas.
PT Armazenamento (3)
As amostras de soro e plasma podem ser conservadas durante 7 dias a 20…25 °C ou a
Konelab™ / Série T 4…8 °C ou durante 3 meses a –20 °C.
A amostra de urina pode ser conservada durante 2 dias a 20…25 °C, durante 6 dias a 4…8 °C
CREATININE (Enzymatic) ou durante 6 meses a –20 °C.
CREATININA (Enzimàtica)
PROCEDIMENTO
981845 4 x 60 ml Consulte o Manual de Referência e as Notas de Aplicação, em separado para o soro/plasma e
urina, para obter mais informações sobre o procedimento automático no seu analisador Konelab.
Veja as mais recentes aplicações na página da internet. Qualquer aplicação não validada pela
ESTE FOLHETO INFORMATIVO É APLICÁVEL PARA Thermo Fisher Scientific Oy não pode ter o desempenho garantido e, por isso, tem de ser
avaliada pelo utilizador.
USO FORA DOS E.U.A. QUALQUER REFERÊNCIA
AOS SISTEMAS KONELAB TAMBÉM SE REFERE À Materiais fornecidos
Reagentes conforme descrito acima.
SÉRIE T.
Materiais necessários mas não incluídos
USO PRETENDIDO Fluído de lavagem, código: 981842.
Para a determinação quantitativa in vitro da concentração de creatinina no soro, plasma ou Calibrador e controlos conforme indicado abaixo.
urina humana nos analisadores Konelab utilizando o método enzimático. Todos os resultados
dos testes devem ser interpretados tendo em conta o contexto clínico. Calibragem
Utilize o Calibrador sCal, código 981831, de acordo com as instruções fornecidas para o
RESUMO (1, 2) analisador Konelab. Deixe o calibrador permanecer dentro do analisador durante um período
A creatina é sintetizada nos rins, no fígado e no pâncreas. A creatina é, então, transportada no máximo de uma hora.
sangue para outros órgãos, tais como os músculos e o cérebro. Cerca de 1% a 2% da creatina
muscular é convertida, diariamente, em creatinina. A produção varia com a idade e o sexo dado Rastreabilidade:
que a quantidade de creatinina endógena produzida é proporcional à massa muscular. Dada a Consulte o folheto informativo incluído na embalagem do Calibrador sCal.
mesma taxa de filtração glomerular, os homens possuem uma concentração de creatinina mais A resposta (A) é convertida para unidades de resultado através de um factor de cálculo.
elevada do que as mulheres, os indivíduos musculosos possuem níveis mais elevados do que os
indivíduos menos musculosos e as pessoas mais jovens possuem níveis mais elevados do que Controlo de qualidade
os idosos. Dependendo da ingestão de carne por parte do indivíduo, a alimentação pode Utilize amostras de controlo de qualidade pelo menos uma vez por dia, depois de cada
influenciar o valor em cerca de 10 %. No entanto, de uma forma geral, a flutuação alimentar da calibragem e sempre que utilizar um frasco de reagente novo. Recomenda-se a utilização de
ingestão de creatinina provoca apenas uma pequena variação na excreção diária de creatinina dois controlos de nível.
nesse mesmo indivíduo. Encontram-se valores elevados de creatinina nos casos de Controlos disponíveis:
insuficiência renal aguda e crónica e de desidratação. Soro/plasma:
São utilizados métodos químicos e enzimáticos para medir o nível de creatinina nos fluidos Nortrol, código: 981043
corporais. Em condições de rotina, é possível efectuar a determinação enzimática da creatinina Abtrol, código: 981044
com apenas algumas interferências. Urina:
uTrol, código: 981821
PRINCÍPIO DO PROCEDIMENTO uTrol High, código: 981822
A creatinina é determinada através de ensaio colorimétrico enzimático da seguinte forma: Os intervalos e limites do Controlo devem ser adaptados aos requisitos individuais de cada
laboratório. Os resultados da(s) amostra(s) de controlo de qualidade devem ficar dentro dos
limites predefinidos pelo laboratório.
Creatininase
Creatinina + H2O ⎯-⎯-⎯-⎯⎯-⎯> Creatina CÁLCULO DE RESULTADOS
Os resultados são calculados automaticamente pelo analisador Konelab usando uma curva de
calibragem.
Creatinase
Creatina + H2O ⎯-⎯-⎯-⎯-⎯> Sarcosina + Ureia Curva de calibragem (exemplo)
Para outras substâncias interferentes, consulte a referência 4. Alterações em relação à versão anterior
Nome da empresa actualizado.
Os valores indicados devem servir apenas como referência. Recomendamos que cada
Konelab™ / Séria T
laboratório verifique este intervalo ou obtenha um intervalo de referência para a população que
serve.
CREATININE (Enzymatic)
Kreatinín (enzymatický)
Factor de conversão: 981845 4 x 60 ml
µmol/l x 0.0113 Æ mg/dl
µmol/l x 11.3 Æ mg/dl
INTERVALO DE MEDIÇÃO
Soro/plasma: TENTO PRÍBALOVÝ LETÁK JE URČENÝ PRE
10 – 2500 µmol/l (0.11 – 28 mg/dl) KRAJINY MIMO USA. AKÁKOĽVEK ZMIENKA
Intervalo de medição alargado depois da diluição secundária: 10 -10000 µmol/l (0.11 –
113 mg/dl) SYSTÉMOV KONELAB SA VZŤAHUJE AJ NA SÉRIU T.
Urina:
0.2 - 40 mmol/l (2.3 - 452 mg/dl) POUŽITIE
Intervalo de medição alargado depois da diluição secundária: 0.01 -200 mmol/l (0.11 – Na kvantitatívne in vitro stanovenie kreatinínu v ľudskom sére, plazme alebo moči v
2260 mg/dl) analyzátoroch Konelab použitím enzymatickej metódy. Všetky výsledky testov musia byť
interpretované s ohľadom na klinický kontext.
CARACTERÍSTICAS DE DESEMPENHO ZHRNUTIE (1, 2)
Os resultados obtidos em laboratórios individuais podem diferir dos dados de desempenho Kreatín sa syntetizuje v obličkách, pečeni a v pankrease. Kreatín je potom krvou
fornecidos. transportovaný do ostatných orgánov, ako napr. svalov a mozgu. Asi 1% až 2% svalového
kreatínu sa denne konvertuje na kreatinín. Keďže množstvo endogénne vytvoreného kreatinínu je
Limite de detecção úmerný svalovej hmote, jeho tvorba sa mení podľa veku a pohlavia. Pri rovnakej rýchlosti
Soro/plasma: glomerulárnej filtrácie majú muži vyššiu koncentráciu kreatinínu ako ženy, svalnaté osoby majú
2 µmol/l (0.02 mg/dl) vyššiu hladinu ako menej svalnaté osoby a mladšie osoby majú vyššiu hladinu ako staršie osoby.
O limite de detecção representa a concentração/actividade mensurável mais baixa passível V závislosti na množstve mäsa v strave môže diéta ovplyvniť túto hodnotu asi o 10 %. Celkovo
de ser distinguida de zero. É calculada como a concentração da amostra de zero + 3 DP (intra- však zmeny hladiny kreatinínu v závislosti na strave spôsobujú u tej istej osoby iba menšie zmeny
ensaio, n=24). v dennom vylučovaní kreatinínu. Vysoké hodnoty kreatinínu sa vyskytujú pri akútnej alebo
Urina: chronickej obličkovej nedostatočnosti a pri dehydratácii.
0.002 mmol/l (0.02 mg/dl) Na meranie kreatinínu v telesných tekutinách sa používajú chemické aj enzymatické metódy.
O limite de detecção representa a concentração/actividade mensurável mais baixa passível Pri bežných podmienkach možno kreatinín enzymaticky stanoviť s malým počtom interferencií.
de ser distinguida de zero. É calculada como a concentração da amostra de zero + 3 DP (intra-
ensaio, n=24) PRINCÍP METÓDY
Kreatinín sa stanovuje enzymatickým kolorimetrickým testom nasledovne:
Imprecisão
Soro/plasma:
Kreatinináza
Média Média Média Kreatinín + H2O ⎯-⎯-⎯-⎯⎯-⎯> Kreatín
38 µmol/l 155 µmol/l 484 µmol/l
DP CV% DP CV% DP CV% Kreatináza
Intra-Ensaio 0.56 1.4 0.69 0.4 1.85 0.4 Kreatín + H2O ⎯-⎯-⎯-⎯-⎯> Sarkozín + Urea
Inter-ensaios 0.16 0.4 0.71 0.5 1.30 0.3
Total 0.84 2.1 2.32 1.5 6.76 1.4
Sarkozín oxidáza
Efectuou-se um estudo de precisão de acordo com as directrizes do Documento EP5-A da Sarkozín + H2O + O2 ⎯-⎯⎯⎯-⎯-⎯-⎯⎯> Glycín + HCHO + H2O2
CLSI (antiga NCCLS) mediante a utilização de um analisador Konelab 60 e várias calibragens e
operadores, durante 20 dias, com um número de medições equivalente a n = 80. Peroxidáza
H2O2 + 4-aminofenazón + HTIB* ⎯-⎯⎯⎯-⎯> Chinon iminový chromogén + H2O + HI
Urina:
Média Média * HTIB = 2,4,6-trijodo-3-hydroxybenzoová kyselina
6.7 mmol/l 14.6 mmol/l
DP CV% DP CV% Kreatinín je konvertovaný na sarkozín pomocou kreatininázy a kreatinázy. Sarkozín je potom
Intra-Ensaio 0.06 1.0 0.13 0.9 konvertovaný na glycín, formaldehyd a peroxid vodíku, za prítomnosti kyslíku dodaného sarkozín
Inter-ensaios: 0.05 0.8 0.17 1.2 oxidázou. Uvoľnený peroxid vodíku reaguje s 4-aminofenazónom a HTIB, pričom vzniká chinon
Total 0.24 3.5 0.51 3.4 imínový chromogén v reakcii katalyzovanej peroxidázou. Intenzita zafarbenia je priamo úmerná
koncentrácii prítomného kreatinínu a možno ju merať fotometricky pri 540 nm.
Efectuou-se um estudo de precisão de acordo com as directrizes do Documento EP5-A da
CLSI (antiga NCCLS) mediante a utilização de um analisador Konelab 60 e várias calibragens INFORMÁCIE O ČINIDLÁCH
e operadores, durante 20 dias, com um número de medições equivalente a n = 80. Činidlo A 4 x 40 ml
Činidlo B 4 X 20 ml
Comparação de métodos
Soro/plasma: Koncentrácie
Realizou-se um estudo comparativo de acordo com as indicações contidas no Documento Činidlo A
EP9-A do CLSI (antiga NCCLS), tendo-se utilizado como método de referência um método TAPS** pufer, pH 8.1 30 mmol/l
enzimático comercialmente disponível. Kreatináza (mikroorganizmy) > 333 μkat/l
Regressão linear (unidade dos resultados µmol/l): Sarkozín oxidáza (mikroorganizmy) > 133 μkat/l
y = 0.97 x –0.3 Askorbát oxidáza (mikroorganizmy) > 33 μkat/l
r = 0.999 HTIB 5.9 mmol/l
n = 110 Detergenty
As concentrações das amostras variavam entre 38 e 1951 µmol/l. Konzervačná látka
Varovanie Moč:
Manipulujte a znehodnocujte ľudské vzorky ako potenciálne infekčný materiál. 0.2 – 40 mmol/l (2.3 – 452 mg/dl)
Rozšírený merací rozsah po sekundárnom rozriedení: 0.01 – 200 mmol/l (0.11 – 2260 mg/dl)
Uskladnenie (3)
Vzorky séra a plazmy môžete skladovať 7 dní pri teplote 20…25 °C alebo pri 4…8 °C, prípadne PREVÁDZKOVÉ CHARAKTERISTIKY
3 mesiace pri -20 °C. Výsledky získané v jednotlivých laboratóriách sa môžu líšiť od daných prevádzkových dát.
Vzorku moča môžete skladovať 2 dni pri 20…25 °C, 6 dní pri 4…8 °C alebo 6 mesiacov pri –
20 °C. Hranica citlivosti detekcie
Sérum/plazma:
TESTOVACÍ POSTUP 2 μmol/l (0.02 mg/dl)
Automatický postup použitia analyzátora Konelab nájdete v manuáli a Používateľskej príručke. Hranica citlivosti detekcie reprezentuje najnižšiu merateľnú koncentráciu/aktivitu, ktorá sa dá
Najnovšie informácie o použití nájdete na webovej stránke. Meranie postupom, ktorý nie je odlíšiť od nuly. Vypočíta sa ako koncentrácia nulovej vzorky + 3 SD (počas procesu, n=24).
validizovaný firmou Thermo Fisher Scientific Oy nie je garantované a musí byť vyhodnotené Moč:
užívateľom. 0.002 mmol/l (0.02 mg/dl)
Hranica citlivosti detekcie reprezentuje najnižšiu merateľnú koncentráciu/aktivitu, ktorá sa dá
Poskytnuté materiály odlíšiť od nuly. Vypočíta sa ako koncentrácia nulovej vzorky + 3 SD (počas procesu, n=24)
Činidlá popísané vyššie.
Nepresnosti
Materiály potrebné, ale neposkytované Sérum/plazma:
Premývacia tekutina, kód: 981842.
Kalibrátor a kontroly uvedené nižšie. Priemer Priemer Priemer
38 μmol/l 155 μmol/l 484 μmol/l
Kalibrácia SD CV% SD CV% SD CV%
Použite sCal, kód 981831, v súlade s inštrukciami k vášmu Konelab analyzátoru. Odporúča sa Počas procesu 0.56 1.4 0.69 0.4 1.85 0.4
nechať kalibrátor v analyzátore max. 1 hodinu. Medzi testami 0.16 0.4 0.71 0.5 1.30 0.3
Celkovo 0.84 2.1 2.32 1.5 6.76 1.4
Odvoditeľnosť:
Pozrite si prosím príbalovú informáciu k sCal. Vykonala sa štúdia presnosti v súlade so smernicami CLSI (predtým NCCLS) dokumentu EP5-
Odozva (A) sa konvertuje na výsledné jednotky pomocou prepočtového koeficientu. A, s rôznymi kalibráciami a operátormi s použitím Konelab 60 počas 20 dní, s počtom meraní
n = 80.
Kontrola kvality
Používajte vzorky na kontrolu kvality aspoň raz denne, po každej kalibrácii a vždy, keď Moč:
použijete novú fľašku s činidlom. Odporúča sa použitie dvoch stupňov kontroly.
Dostupné kontroly: Priemer Priemer
Sérum/plazma: 6.7 mmol/l 14.6 mmol/l
Nortrol, kód: 981043 SD CV% SD CV%
Abtrol, kód: 981044 Počas procesu 0.06 1.0 0.13 0.9
Moč: Medzi testami 0.05 0.8 0.17 1.2
uTrol, kód: 981821 Celkovo 0.24 3.5 0.51 3.4
uTrol vysoká, kód: 981822
Intervaly a rozmedzia kontrolných hodnôt sa musia prispôsobiť individuálnym požiadavkám Vykonala sa štúdia presnosti v súlade so smernicami CLSI (predtým NCCLS) dokumentu EP5-
laboratória. Výsledné hodnoty kontrolných vzoriek by mali byť v rozmedzí referenčných hodnôt A, s rôznymi kalibráciami a operátormi s použitím Konelab 60 počas 20 dní, s počtom meraní
laboratória. n = 80.
Abs
Peroxidas 0.060
H2O2 + 4-aminofenazon + HTIB* ⎯-⎯⎯⎯-⎯> Kinoniminkromogen + H2O + HI 0.040
Provmaterial Angivna värden är endast avsedda som vägledning. Vi rekommenderar att varje laboratorium
Endast de prover som anges nedan har testats och befunnits vara acceptabla: verifierar detta område eller fastställer ett referensintervall för populationen som betjänas.
Försiktighetsåtgärder MÄTOMRÅDE
Humana prover ska behandlas som potentiellt smittförande, både vid hantering och kassering. Serum/plasma:
10 – 2500 μmol/l (0.11 – 28 mg/dl)
Förvaring (3) Utökat mätområde efter sekundär spädning: 10 - 10000 μmol/l (0.11 – 113 mg/dl)
Serum och plasmaprov kan förvaras i 7 dagar vid 20…25 °C eller vid 4…8 °C eller i 3 månader Urin:
vid -20 °C. 0.2 - 40 mmol/l (2.3 - 452 mg/dl)
Urinprov kan förvaras i 2 dagar vid 20…25 °C, i 6 dagar vid 4…8 °C eller i 6 månader vid – Utökat mätområde efter sekundär spädning: 0.01 - 200 mmol/l (0.11 - 2260 mg/dl)
20 °C. UTFÖRANDETS KARAKTERISTIKA
TESTUTFÖRANDE Resultaten som erhålls vid varje enskilt laboratorium kan skilja sig från angivna data för
Se Handhavandemanualen och särskild Application Notes för serum/plasma och urin för prestanda.
automatiskt utförande på aktuell Konelab-analysator. Se senaste applikationer från webbplatsen. Detektionsgräns
Varje applikation som ej har validerats av Thermo Fisher Scientific Oy, kan ej garanteras vad
gäller prestanda och måste därför utvärderas av användaren. Serum/plasma:
2 μmol/l (0.02 mg/dl)
Bifogat material Detektionsgränsen representerar lägsta mätbara koncentration/aktivitet som kan skiljas från
Reagenser enligt ovan. noll. Den beräknas som koncentrationen av nollprov + 3 SD (standardavvikelser) (inom serien,
n=24).
Erforderligt material som ej medföljer Urin:
WashFluid, artikelnr.: 981842. 0.002 mmol/l (0.02 mg/dl)
Kalibrator och kontroller enligt nedan. Detektionsgränsen representerar lägsta mätbara koncentration/aktivitet som kan skiljas från
Template: D01162_4
Imprecision
Serum/plasma:
En precisionsstudie har utförts enligt anvisningarna i CLSI (f.d. NCCLS) Dokument EP5-A med
olika kalibreringar och användare, med Konelab 60 under 20 dagar. Antalet mätningar var n = 80.
Urin:
Medel Medel
6.7 mmol/l 14.6 mmol/l
SD CV% SD CV%
Inom serie 0.06 1.0 0.13 0.9
Mellan serier 0.05 0.8 0.17 1.2
Totalt 0.24 3.5 0.51 3.4
En precisionsstudie har utförts enligt anvisningarna i CLSI (f.d. NCCLS) Dokument EP5-A med
olika kalibreringar och användare på Konelab 60 under 20 dagar. Antalet mätningar var n = 80.
Metodjämförelse
Serum/plasma:
En jämförelsestudie har utförts enligt anvisningarna i CLSI (f.d. NCCLS) Dokument EP9-A med
en kommersiellt tillgänglig enzymatisk metod som referens.
Linjär regression (resultatenhet µmol/l):
y = 0.97 x – 0.3
r = 0.999
n = 110
Provens koncentration låg mellan 38 och 1951 µmol/l.
Urin:
En jämförelsestudie har utförts enligt anvisningarna i CLSI (f.d. NCCLS) Dokument EP9-A med
en kommersiellt tillgänglig enzymatisk metod som referens.
Linjär regression (resultatenhet mmol/l):
y = 1.10 x +0.02
r = 0.999
n = 119
Provens koncentration låg mellan 1.4 och 33.2 mmol/l.
REFERENSER
1) Burtis, CA and Ashwood, E R (ed.), Tietz Fundamentals of Clinical Chemistry, 5th edition, W
B Saunders Company, Philadelphia, 2001, pp. 419-422.
2) Thomas, L. (ed.), Clinical Laboratory Diagnostics; Use and Assessment of Clinical
Laboratory Results, 1st edition, TH-Books Verlagsgesellschaft mbH, Frankfurt/Main,
Germany, 1998, pp. 366-370.
3) Guder WG, Narayanan S, Wisser H, Zawta B. List of Analytes; Preanalytical variables.
Brochure in: Samples: From Patient to the Laboratory. GIT Verlag GmbH, Darmstadt, 1996.
4) Young, D.S., Effects of Drugs on Clinical Laboratory Tests, Fifth Edition, AACC Press,
Washington, D.C., 2000, pp. 3-240 - 3-261.
5) Mazzachi BC, Peake MJ, Ehrhardt V. Reference range and method comparison studies for
enzymatic and Jaffé creatinine assays in plasma and serum and early morning urine. Clin
Lab. 2000; 46: 53-55.
TILLVERKARE
Thermo Fisher Scientific Oy
Clinical Diagnostics Finland
Ratastie 2, P.O. Box 100, FI-01621 Vantaa, Finland
Tel. +358 9 329 100, Fax +358 9 3291 0300
www.thermo.com/konelab
Revisionsdatum (åååå-mm-dd)
2007-08-07