Manual CSI Ultraspec 8000
Manual CSI Ultraspec 8000
Manual CSI Ultraspec 8000
Introduction
The UltraSpec 8215/8225 Laser Alignment Fixtures are the latest in a com-
plete line of alignment equipment provided by CSI. The 8215/8225 system
is used in conjunction with the UltraSpec Alignment Analyzer. Although
the UltraSpec analyzer can be purchased separately, it must be used to pro-
cess the data from the fixtures. The Model 8215/8225 system attaches
sensor heads to shafts that are typically coupled together. Laser beams are
emitted from each sensor head and targeted on the opposite sensor head.
Each shaft position is measured in relation to the other shaft by rotating the
two shafts. Shaft misalignment is calculated from the position of each laser
beam on its respective target.
The position of each laser beam on its target is transmitted from each sensor
head to the UltraSpec analyzer through radio frequency (RF) or direct
cable communication. The UltraSpec analyzer processes the data and cal-
culates the required machine moves necessary to bring the machine com-
ponents into alignment.
With most alignment systems available today, the sag of the alignment fix-
tures must be considered to accurately calculate machine moves. With the
use of laser beams, there is no fixture sag to consider! The RF communica-
tion between the sensor heads and the UltraSpec analyzer simplifies data
acquisition and reduces operator error. The direct cable connect provides
you with an alternate method to transfer data when the RF link is not prac-
tical.
Since each head has a self-contained angle sensor and additional memory,
the data can be acquired in a variety of ways. Choices range from the tra-
ditional top, bottom, left, and right locations (minimum of three) where the
fixtures interact directly with the analyzer to the patented partial sweep
method (the sensor heads memorize the data and later dump it to the ana-
lyzer).
1-1
When the 8215/8225 fixtures and the UltraSpec analyzer are used with the
UltraMgr PC software program, a complete alignment program can be
managed from a centralized location. UltraMgr provides job tracking, con-
trol of alignment tolerances, alignment history and much more.
Some advantages of the UltraSpec 8215/8225 Laser Alignment Fixtures
are:
• Easy setup
• No fixture sag
• Visible laser to assist with rough alignments
• Large diameter mounting range
• Large span range between sensor heads
• Large vertical expansion to clear larger couplings
• Large misalignment measurement range
• Partial revolutions (down to 35°)
• Dual Pass method for uncoupled alignments
• Real time machine move monitoring, vertical and horizontal
• Soft foot locator
• Radio frequency communication
• Direct cable connect, if needed
• Automatic data acquisition
• Dual beam technology
• Internal angle sensors
• Internal memory
• 15-minute charge cycle for sensor heads
• Simultaneously charging, sensor heads and analyzer
• Best ambient light protection
• Straightness measurement
Note
The note paragraph indicates special comments or instruc-
tions.
Caution!
The caution paragraph alerts you to actions that may have a major
impact on the analyzer, stored data, etc.
Warning!
The warning paragraph alerts you to actions that may
have extremely serious consequences for equipment
and/or personnel.
Note
The 8215/8225 Laser Fixtures are only compatible with the
ProAlign Plus application and not compatible with the
ProAlign, Easy Align, Quick Align, Vertical Align, and Train
Align applications.
Note
The UltraSpec Thermal Growth application will not commu-
nicate with the UltraMgr software.
Note
UltraMgr requires some planning and set up before jobs can be
downloaded into the UltraSpec analyzer. See the UltraMgr
User’s manual for more information.
Mounting
Chains
Tape
Measure
8000RF Mounting
Posts
** Optional
1” Vertical
Extension
Blocks
Note
Other UltraSpec analyzer models may have a different type of
leather case and vinyl sheath. These may require a different
method of inserting the analyzer into the leather case and vinyl
sheath.
HJP #4635891
To Com Port
Model 639
Communications Cable
If the host computer uses a 25-pin COM port, you must use an optional
9-to-25 pin adapter cable between the computer and the Model 639 com-
munications cable as shown in the following illustration.
Multifunction Input
Connector
HJP #4635891
Model 639
Communications Cable
Note
For more information about transferring stored jobs to and
from UltraMgr, see “Communications” on page 2-21.
Note
Firmware and application version 2.40 require version 6.0 or
later ROM to load into the UltraSpec analyzer. For more infor-
mation, contact CSI Customer Support.
Note
Firmware and application versions others than 2.40 and later
are not interchangeable with 2.40 and later firmware and appli-
cations and cannot be loaded into the UltraSpec analyzer
together.
Note
The UltraSpec firmware and applications are serialized for the
UltraSpec analyzer and can only be loaded into the UltraSpec
analyzer with that serial number.
Download Procedure
The following flowchart is an abbreviated version of the download proce-
dure. For a thorough explanation, refer to the “DOS Download Procedure”
on page 1-16 or the “Windows Download Procedure” on page 1-22.
Use communication cable to connect computer Use communications cable to connect computer
COM1 to UltraSpec analyzer 25-pin connector COM1, COM2, COM3, or COM4 to UltraSpec
analyzer 25-pin connector
On analyzer: select No, press the Enter key, On analyzer: select Yes
and then select the active application and press the Enter key
End
Note
When you are loading either firmware or applications, begin
with the firmware diskette first.
4. ··· Press the F6 key to bring up the following screen which shows the
communication configuration of the computer.
5. ··· Change the computer settings if necessary. Normally, you should use
one stop bit and a baud rate of 19.2k or 38.4k. After all changes have
been made, press the Enter key.
Note
Select this option whether you are replacing the base firmware
or loading an application or program.
Loading Firmware
Caution!
Reloading firmware in an UltraSpec 8000 analyzer resets the calibra-
tion factors. Always review calibration factors after loading firmware
to verify the settings are the same as in the factory supplied calibration
card. Refer to “Review Calibration” on page 2-32 for information
about calibrating the UltraSpec analyzer.
Loading Applications
A list of programs on the diskette will be shown.
12.·· Select a program and press the Enter key.
14. · When the last application program has been loaded, answer No and
press Enter.
After the desired application(s) are loaded, you must use Select Appli-
cation to designate the active application.
15. · Exit the Download program. At the computer, press the F10 key to
exit the DOS Download program.
Note
Depending on the amount of memory available, no more than
six applications can be loaded into the analyzer at the same
time. In an analyzer with only 0.5 Mb of memory available, no
more than one or two applications can be loaded at the same
time, depending on the application.
Note
Although individual firmware programs are shipped on sepa-
rate diskettes, all programs serialized for the same instrument
can be copied to a single diskette (for downloading) if desired.
To do this, copy all *.DLH and *.TSK files to the base firmware
diskette.
Note
The Calibration values can be automatically loaded into the
analyzer when the firmware is loaded. For more information,
see the “Calibration Write Program” on page 2-39.
Loading Firmware
Caution!
Reloading firmware in an UltraSpec 8000 analyzer may reset calibra-
tion factors. Always review calibration factors after loading firmware
to verify the settings are the same as in the factory supplied calibration
card. Refer to “Review Calibration” for information about calibrating
the UltraSpec analyzer.
There may be up to a 20-30 seconds pause before loading begins. After the
firmware has been loaded, a list of application programs to be loaded will
be shown (if available on the same diskette). An error message will appear
if no application programs are located.
10. · If the error message occurred, insert the Application Diskette into the
computer’s drive A (or B). On the analyzer, press the Reset/Main
key. Otherwise, go directly to Step 11.
13.·· When the last application program has been loaded, answer No and
press Enter.
After the desired application(s) are loaded, you must use Select Appli-
cation to designate the active application.
Note
When the ProAlign Plus application is first selected after
loading into the analyzer a message is displayed telling you
that all DLP (Downloadable Program) memory is cleared and
displaying the amount of free memory (in bytes) left in the ana-
lyzer. From this screen press Enter to continue and enter the
ProAlign Plus application.
14.·· Exit the Download program. At the computer, select the Exit button
to exit the Windows Download program.
Note
Although individual firmware programs are shipped on sepa-
rate diskettes, all programs serialized for the same instrument
can be copied to a single diskette (for downloading) if desired.
To do this, copy all *.DLH and *.TSK files to the base firmware
diskette.
Note
After loading firmware and application(s) in an UltraSpec 8000
analyzer, check the calibration factors (see “Review Calibra-
tion” on page 2-32).
Note
The Calibration values can be automatically loaded into the
analyzer when the firmware is loaded. For more information,
see the “Calibration Write Program” on page 2-39.
16
Note
The baud rate and stop bit set in the analyzer and in the
UltraMgr module must match each other.
Note
There is no communication between UltraMgr and the
UltraSpec Thermal Growth application.
Note
Where applicable, differences in the UltraSpec analyzer
models will be discussed.
2-1
UltraSpec Analyzer Top Panel
When viewing the analyzer from the top, the features in this figure are vis-
ible. Descriptions are listed below.
Caution!
Do not connect any adapters or accessories to the RS232 serial port
while the analyzer is turned on.
Charger Input
• Provides an input for the Model 8211 or Model 8212 battery
chargers. In the UltraSpec 8000 analyzer, it provides an input for the
Model A2115-C battery charger. In the UltraSpec 8117 analyzer, it
provides an input for the Model 93140 power supply, 8211 battery
charger, or 8212 battery charger.
Power Output
• Provides power for the Model 404 series phototach.
ON
_____
OFF On/Off
UTILITY OPTIONS PRINT NOTES HELP
Command
Keys
ENTER ENTER
Control
Keys
Activity ACTIVITY CHARGING
LED Charging
PAGE
UP
DELETE
LED
PAGE
Reset/Main RESET
DOWN EXP
RESET
Key INSERT
DEC
Keypad
MARK
Control
Keys
Activity
LED Charging
LED
Reset/Main
Key
Keypad
Enter Main
Control
Keys
Activity
LED GHI
Page ABC DEF
Delete
Up 7 8 9
JKL MNO PQR /#&
Page
Down 4 5 6 Exp Keypad
STU VWX YZ* - +
Insert
1 2 3 Dec
Space
Alt . ,
0
ON
On/Off
OFF Press the On/Off key once to turn on the analyzer, press it again to turn it
off.
Activity LED
This LED, located on the left side of the front panel, flashes to provide
ACTIVITY
visual confirmation of keyboard entries and certain software operations (i.e.
8000 8117 starts and stops of data acquisitions).
Charging LED
This LED, located on the right side of the UltraSpec 8117 front panel, is illu-
N/A CHARGING
minated when the battery charger is connected to the UltraSpec 8117 ana-
8000 8117 lyzer and the battery pack is being charged (see “UltraSpec Analyzer
Battery Maintenance” on page 9-4).
Alt Key
Alt Alt
The Alt Key (lower left corner) enables a second function on some keys.
Keys with second functions are: Print, Notes, and Delete. Other keys with
second functions may be application dependent. These keys and functions
will be discussed in greater detail where applicable.
Note
Pressing the Alt key along with the Notes key allows you to
enter the user defined notepad area (for adding additional
notes).
Utility
Utility
The Utility function key is used to select the application you want to work
with. Although the analyzer normally defaults to the last application that
was used, you must use this key to change applications. Additional func-
tions available from the Utility Functions menu are Communications,
System Setup, and Battery Test (discussed in later sections).
Note
All of the functions accessed with these keys are described in
detail in later sections.
Options
Options
The Options key enables you to store, recall, and delete jobs. Certain appli-
cation-specific options and specified functions may also be accessed using
the options key.
Screen/Print
The Screen/Print key is a dual purpose key. You can:
Screen • Print a job report for the application being run, or
PRINT
Print
• When used with the Alt key, you can get a printout of the current
8000 8117 screen.
Notes
Notes
The Notes key is used to access the notepad screen which provides a list of
preprogrammed or user-defined observations concerning the status of the
equipment. By pressing Alt/Notes, you can access the user-defined notepad
area (to create a new note).
Help
? Help
The Help key can be used to access help screen information at appropriate
locations in the various menus. By pressing Alt/Help while in the Utility
Function, you an access the help screen information for special language
characters.
Control Keys
Enter
Enter
• Accepts menu selections and screen inputs.
• Returns you to the active application Main menu.
Up/Down Arrows
• Moves the cursor through menu selections.
• Expands and contracts the scale of graphic tolerance charts by a
factor of two with each press.
• Scrolls the text screen up or down when more selections are avail-
able than will fit on the screen.
• Scrolls to the next reading (Up Arrow) or previous reading (Down
Arrow) required under the Alignment Data Screen when using the 4
Point Manual method. (Laser Alignment program only.)
• Moves the cursor in increments of ten points or angles at a time to
the next point (Up Arrow) or previous point (Down Arrow) when
viewing or editing a sine curve. (ProAlign Plus program only.)
8000 8117
Left/Right Arrows
• Scrolls fields (Notes, Machine Description, Station Description, etc.)
that extend off of the display.
• Controls cursor movement on data plots.
• Moves one space per press to the right or left in alphanumeric fields.
• Moves the cursor across the set of offset and angle points in the tol-
erance graphs. (Laser Alignment program only.)
• Steps through the alternate solutions in the Machine Move screens
(Laser Alignment program only.)
• When viewing or editing a sine curve, pressing the Alt/Left Arrow
key moves the cursor to the first point or angle. Pressing the Alt/
Right Arrow key moves the cursor to the last point or angle (Laser
Alignment program only.)
• Moves the cursor one point or angle at a time to the next point (Right
Arrow) or to the previous point (Left Arrow) when viewing or editing
a sine curve. (Laser Alignment program only.)
Page Up
Page
Up • Formerly known as the Home/Back function.
• The Page Up function enables you to back up to the previous screen
within a section.
• In Notes, the Page Up function moves backward six notes at a time
within a section.
• In the Tolerance Charts, pressing the Alt/Page Up key shows the
acceptable/excellent ranges. (Laser Alignment program only.)
Page Down
Page
Down • Formerly known as the End/Back function.
Insert
Insert
• Formerly known as the Insert/Delete function.
• Inserts alphanumeric characters.
• Undeletes a point or angle and refits curve when editing a sine curve.
(Laser Alignment program only.)
Delete
Delete
• Formerly known as the Clear function.
• Deletes alphanumeric characters.
• Alt/Delete clears alphanumeric entries from text fields and the
notepad.
• Pressing the Alt/Delete key while in the Main menu allows deletion
of the current job or reading and notes in the current job.
• Pressing Alt/Delete while acquiring data clears magnitude and phase
averages and begins a new average. (Balance and Analyze programs
only.)
• Deletes a point or angle and refits curve when editing a sine curve.
(Laser Alignment program only.)
• When editing a sine curve, press the Alt/Delete key undeletes all
points or angles (36 maximum) and refits curve. (Laser Alignment pro-
gram only.)
Mark
. , • Provides Special Functions when displaying spectra with the cursor
. , on screen (required with the Analyze program). Pressing the key the first
MARK
8000 8117
time more precisely locates the frequency and amplitude of the peak.
Pressing the key the second time places boxes at multiples of the
located peak.
Expand (Exp)
/#& /#&
Exp • Expands the horizontal scale of a spectral or waveform plot.
• Pressing the Alt/Exp key replaces the bar graphs with numbers and
visa-versa under Laser Head Status (Laser Alignment program only.)
Decrease (Dec)
- + – +
Dec • Decreases the horizontal scale of a spectral or waveform plot.
Response Fields
Many functions of the UltraSpec analyzer require that you enter data;
response fields are provided for this purpose. Some response fields provide
a predefined set of selections. Pressing any key on the keypad will step
through these selections as shown in the following example.
76.8K 1
38.4K
19.2K
9600
4800
2400
1200
300
ONE
TWO
P S T U M P
PQR STU STU STU MNO PQR
6 , 1 1 1 , 5 , 6 ,
P Q R 6
(_) PQR PQR PQR PQR
0 , 6 6 6 6
Note
The Delete key will delete an individual character as well as the
space it occupies. The Alt/Delete key will clear the entire field.
Numerical Values
For fields that require numerical entry only, the alphabetic characters of the
keys are disabled. Only one key press is required to enter a number. The
“-” key can be used to enter a negative number and can be entered before or
after the number(s).
• /
,
In the Machine Dimensions screen, these keys are used to enter fractional
values. For instance, to enter 5 5/8, press keys in the following order:
(1)5
(2).
(3)5
(4)/
(5)8
Key Special
Sequence Character
Alt/1 Ä
Alt/2 Å
Alt/3 Ö
Alt/4 Ñ
Alt/5 Ü
Alt/6 É
Alt/7 Ç
Note
To view the special language characters in UltraMgr, a special
table will need to be installed. Call CSI Customer Support for
more information.
Press the Utility key to access the Utility Functions menu shown below.
Select Application
Although the analyzer defaults to the last application used, you can use this
menu to select another application (when multiple applications have been
installed).
Note
Applications must be loaded from a computer to the UltraSpec
analyzer before they can be selected.
Communications
The Communications function menu is used to establish communications
with the host computer and to provide utilities for controlling communica-
tion parameters of the analyzer, as well as an external modem.
Host PC Load/Dump
The Load function is used to transfer one or more jobs from the computer
database into the analyzer. The Dump function is used to transfer the mod-
ified jobs in storage to the UltraMgr database.
Communications Setup
The Communications Setup menu is used to configure the communications
port of the analyzer.
7
Select the Baud Rate from 76.8K, 38.4K, 19.2K, 9600, 4800, 2400, 1200, or
300 baud. For rapid data transfer, select the highest baud rate possible, con-
sistent with the capabilities of the host computer equipment.
Set the Stop Bit option at One or Two.
The Timeout option controls the amount of time (selectable from 0 through
99.9 seconds) that the analyzer will wait for the computer to respond before
aborting the procedure and trying again. Entering a value of zero will dis-
able this feature.
Note
CSI recommends a value of “one” for the Stop Bits setting. A
suggested timeout value of 10 seconds, and five retries should
allow sufficient time for the computer to process any commu-
nication errors and multiple jobs being loaded from UltraMgr.
The CTS Enable option should be set to “Yes” when using a modem that
supports hardware handshake via CTS signal (this includes most 9600 baud
modems).
The example Set Modem screen above shows a modem command line for
use with a Hayes-compatible modem which will:
• Use tone dialing ATDT (ATDP for pulse dialing)
• Dial 1 and the area code 423
• Then dial the number 675-2400
Underneath “Type in New Commands,” use the alphanumeric keypad to
type in the desired letters/numbers. Press the Enter key to accept the new
modem number.
Note
You must have RBMware’s CSICOM program to transfer data
via modem.
To Establish Communications
1.···· On the PC, access RBMware. From the RBMware main menu, select
the Portable Vibration button, select the Setup/Communications tab,
select Analyzer Data Transfer, then select Setup Comm.
2. ··· Specify the COM port the modem is connected to.
3. ··· In the Modem Initialization String field, enter the initialization string
for your modem and click OK. Refer to the manual supplied with
your modem for recommended initialization strings.
If you are unable to locate the correct initialization string, CSI recom-
mends trying one of the following generic strings.
System Setup
From the Utility Functions menu, press System Setup to access the fol-
lowing screen. This screen shows setup parameters from which various
operational control parameters of the analyzer may be specified.
9
Note
The Units mode is the only parameter under Units Function
used by the Alignment application.
The <Data Units> column lists the available types of converted data. The
<Data> column specifies how the acquired data will be displayed for each
data unit. Press any key in the lower section of the keypad to step through
and select RMS, Peak, and Pk-Pk.
11
Note
If the analyzer displays the “Warning - Check GMT & DST
Setting” screen, it will continue to do so with power-up and
reset until you have entered, made the necessary changes, and
accepted the date and time. In the UltraSpec 8000 analyzer,
you will be required to set the date and time every time you
reload the base firmware. In the UltraSpec 8117 analyzer, the
date and time with GMT offset and D.S.T. are automatically
saved in non-volatile memory when the date and time are
accepted; therefore, you will not be required to set the date and
time when the base firmware is reloaded.
12
UltraSpec 8000
UltraSpec 8117
Key Beeper: - This field is used to toggle the audible indicator (beep) func-
tion between On and Off. When on, a single beep indicates that a key has
been pressed.
Automatic Poweroff: - The Automatic Poweroff feature will turn the ana-
lyzer off if there is no user activity within a specified time period. This time
period is defined in minutes from 0 to 10. Entering 0 will disable the Auto-
matic Poweroff feature.
Contrast: – alters the viewing angle of the analyzer display screen. Any key
on the keypad will advance the contrast values from 1 (maximum contrast)
through 41 (minimum contrast), and then the sequence repeats.
Note
To display and acquire data simultaneously a splitter adapter
(e.g., model 8003) is required.
Note
Switching printer types with the 720A attached might require
letting the 720A printer interface sit unconnected for about five
minutes. If you are using a printer interface with a reset button,
it should be pressed after changing the printer selection.
Load Firmware
This feature is used to load firmware and applications from the PC into the
analyzer. See “Installing the Application” on page 1-10 for complete
instructions on this procedure.
Review Calibration
To ensure maximum measurement accuracy, each UltraSpec analyzer is
thoroughly calibrated by CSI before shipping, and the results of the cali-
bration are recorded on the calibration tag that is shipped with the analyzer.
The values entered into the analyzer’s calibration routine allow the firm-
ware to compensate for any minor errors in measurement that may be
caused by component tolerances within the analyzer.
Normally, these values do not have to be modified. However, the analyzer
calibration values should be checked against the calibration tag under the
following conditions:
• The analyzer produces invalid or unreasonable readings.
• The battery has been allowed to discharge to the point where ana-
lyzer memory has been lost. (This typically would require several
months of inactivity.)
• New firmware has been loaded into the analyzer (UltraSpec 8000
analyzer only).
• The analyzer displays the “Warning – Check Calibration Factors”
screen.
Normally, the calibration values can simply be entered into the analyzer
using the values recorded on the calibration tag. It is recommended, how-
ever, that the analyzer be returned to CSI once a year for re-calibration in
order to maintain the instrument’s accuracy.
Note
In the UltraSpec 8117 analyzer, the calibration values can be
saved into non-volatile memory. For more information, see
“Save Calibrations” on page 2-37.
UltraSpec 8000
UltraSpec 8117
13
Integration Factors
These four parameters apply to the gain corrections used for the analog
integrators which convert sensor units to display units.
14
UltraSpec 8000
UltraSpec 8117
Note
Battery Offsets is the only factor used in the Alignment appli-
cation.
15
If pole corrections are enabled (Apply Poles field set to “Yes”), the specified
lower frequency for correction will be used not only for analog integration
but also for digital integration. Do not enable pole corrections for normal
use unless there are special requirements for measurements.
Filter Calibration
These two parameters define the break frequency intervals of the internal
pre- and post-filter stages of the anti-aliasing filter section.
16
Save Calibrations
This feature is only available in the UltraSpec 8117 analyzer. It it used to
store the calibration values into non-volatile memory. This must be done to
save new calibration values. Otherwise, the values will be lost once power
to the analyzer is turned off.
Note
If the analyzer displays the “Warning – Check Calibration Fac-
tors” screen, it will continue to do so with each power-up and
reset until you have entered the Set Calibration function. This
procedure is intended to ensure that the calibration factors are
at least checked for accuracy after downloading firmware or
inadvertently losing analyzer memory data.
17
The factors are displayed in two columns. The first column contains the
Default calibration factors. The second column contains the editable
(Actual) calibration values.
The Current Drive defaults to the first drive found to exist starting with
Drive A. When the program is first run, it automatically attempts to read
the factors from the file containing the firmware on the selected drive. To
change drives, click on the Previous or Next button. The program will auto-
matically attempt to read the factors from the file containing the firmware
on the newly selected drive.
Note
The Calibration Write program is a Windows program. There-
fore, the calibration factors can only be entered or modified
from within Windows. Once the calibration factors have been
successfully written into the firmware file, the firmware can be
downloaded in the analyzer either using the DOS Downloader
or Windows Downloader program. Refer to the pages under
“Download Procedure” on page 1-13 for more information.
Note
Even after loading firmware with the modified calibration fac-
tors into the UltraSpec 8000 analyzer, you are prompted to
check them to make sure they are correct. This prompt mes-
sage will continue appearing until you look at the calibration
factors.
This display presents approximate values and should only be used as a guide-
line in determining the amount of battery charge remaining. When the per-
cent value first reaches zero, a built-in safety margin allows approximately
ten minutes of additional use before the analyzer turns itself off in order to
protect the memory.
Utility Key
The Utility key provides functions that are generic to all applications. Refer
to “UltraSpec Analyzer Utility Functions” on page 2-19 for an explanation
of these functions.
3-1
Options Key
The Options key provides access to those functions that are specific to this
application (see the following screen).
Note
Depending on the amount of memory available in the analyzer
and the type of alignment data (e.g., 4 point or Auto Sweep)
being stored in the alignment job it is possible to store up to a
maximum of 100 alignment jobs in the analyzer.
Note
Follow these steps carefully to ensure that you do not overwrite
data that you really want to keep
Note
If there are more jobs than can be displayed on the analyzer
screen, use the Page Down key to page to the remaining selec-
tions on the list.
After selecting the job, the analyzer returns to the program’s Main menu.
If All Jobs is selected, the following warning screen will be displayed. Press
any key to toggle Yes or No. Press Enter to continue. Answering Yes deletes
all jobs stored (including the active job) in the analyzer and displays a
screen stating that they were deleted successfully.
Caution!
Use extreme caution when using the Delete Job(s) function if the ana-
lyzer contains job data. After accepting or answering Yes to the
warning screen, the data cannot be retrieved unless the data have been
dumped to UltraMgr.
Alignment Setup
Selecting Alignment Setup displays the following screen.
In all bar graphs, the bar should be between the lines. Definitions are:
A. ·· A bar graph of the present battery condition of each sensor head.
B.··· Bar graph of the laser diode current in each head. This is for system
troubleshooting.
C. ·· Bar graph of the laser intensity on each target. This is for system
troubleshooting.
D. ·· The present rotational angle of each sensor head.
E & F. The position of each laser beam on each target. This position is
located on the vertical and horizontal axes (read in mm or mils).
G. ·· The number of averages taken before each target data value is sent to
the analyzer. Pressing the Up or Down Arrow key increases or
decreases the number of averages by one. A minimum of one and a
maximum of 25 averages can be set.
H. ·· The version number of the sensor heads.
1 ···· (If applicable), you are using the proper direct connect cable (Model
821510) and it has been properly connected to the analyzer.
2 ···· If applicable, you are using the proper wireless adapter (Model
8000RF) and it has been properly connected to the analyzer
3 ···· Both laser heads are turned on.
4 ···· The RF communication addresses for both the laser heads and
analyzer are set up properly.
5 ···· You are in the proper alignment application (ProAlign Plus) for the
laser heads (Models 8215 and 8225).
Pressing the Alt/0 (Alt/Space) key displays a screen with a complete status
and error code report from laser heads. If a “W” or an “X” appears under
a laser head column for a condition, as shown in the following figure, then
that condition exists for that laser head.
Error Report
Low Battery: A “W” in this column indicates the 1st of two low battery
warnings. From the time a “W” appears in the column the laser head
will have approximately thirty minutes of power left before it powers
off. Any data acquired when the head is in this state is considered
good and can be used, but you need to find be deciding on a stopping
point. An “X” in this column indicates the 2nd and final low battery
warning before the laser head powers off. From the time an “X”
appears in the column the laser head will power off in approximately
five minutes. Any data acquired when the head is in this state is
questionable and is not used. With both indications the laser head
needs to be recharged. Refer to “Battery Charging” on page 4-9 and
“Battery Usage - Laser Heads” on page 4-16 for more information.
Note
The amount of time (life) left in the battery is only an approxi-
mation. As battery age and usage increases, battery life can
decrease; therefore, the amount of time left in the battery
decreases.
Note
The circuit driving input current for the laser signal being
transmitted and the laser power for the laser signal being trans-
mitted are not the same. It is possible for the laser signal power
being transmitted to be low and still have the circuit driving
input current for the laser signal being transmitted be within
factory specified limits.
Note
Contact CSI Customer Support if you are unable to resolve
any problem or have any questions about any of the parame-
ters under Laser Head Status.
Caution!
When connecting any device to the RS232 serial port located on the top
panel of the analyzer, the analyzer must be turned off.
With the direct connect cable connect, select Read/Set 82X5 Addressing
from under the Addressing Selection menu and the following menu will be
displayed with step-by-step instructions:
The second screen will display the address received from the A-head. If the
address is not correct or you wish to change the address type in the new
address from 1 to 10 using the number keys on the keypad and press Enter
to continue to the next screen. If the address is correct then press Enter to
continue to the next screen.
10
The forth screen will display the address received from the B-head. If the
address is not correct or you wish to change the address type in the new
address (different then the A-head address) from 1 to 10 using the number
keys on the keypad and press Enter to continue to the Alignment Options
menu. If the address is correct then press Enter to continue back to the
Alignment Options menu.
Note
If you enter a number outside the range of 1 to 10 or try to
enter two addresses which are the same you will get a warning
telling you of the error and to repeat the procedure.
11
This screen will display the address set in the analyzer. If the address is not
correct or you wish to change the address type in the new address from 11
to 15 using the number keys on the keypad and press Enter to continue to
the Alignment Options menu. If the address is correct then press Enter to
continue back to the Alignment Options menu.
Note
If you enter a number outside the range of 11 to 15 you will get
a warning telling you of the error. If this occurs, press any key
to continue and repeat the procedure.
12
Calculator Mode
Calculator mode serves as an aid in performing calculations on collected
alignment data. The Predict Mode allows you to estimate the resulting mis-
alignment based on the current values needed to align the machine
(required move) and operator specified values that the machine can be
moved. This is very useful if the machine cannot be dropped any further in
the vertical direction or if the machine is bolt bound horizontally. The
Face/Rim Correlation converts current misalignment laser position read-
ings to corresponding face/rim readings. This is useful to users familiar with
the face/rim alignment method who want to correlate the alignment condi-
tion, as measured by the lasers, to indicator values.
13
Predict Mode
Press the Enter key to view the Predict-Vertical screen.
14
This screen displays the movement currently required to align the machine
vertically. The target indicates how close the machine vertical position is to
being in tolerance. The center (bullseye) indicates that the alignment con-
dition is in the excellent range. The middle ring represents an acceptable
range. The outer ring represents >1 times and <2 times the acceptable
range. No ring highlighted indicates >2 times acceptable range.
Similar to the “Machine Moves” section, one of six screens showing an
alternate solution to bring the equipment into alignment may be selected.
If no foot is selected (no value is highlighted), the Left/Right Arrow can
select alternate solutions.
15
16
Face/Rim Correlation
Press the Enter key to view the Sensor Location screen.
17
Use the Up/Down Arrow or any alphanumeric key to select which side of
the coupling the sensor is located. Actual face/rim measurements are to be
taken on the coupling half pointed to by the arrow. The sensor measure-
ment location determines the direction (sign) of the “rim” total indicator
runout (TIR).
Press the Enter key again to view the Sensor Targets screen.
18
Print Report
Report Type: – Select one of two entries (use any number key on the
keypad to toggle the choices).
• Detail – allows every set of alignment readings and machine moves
(maximum 20) for the job(s) selected to be printed.
• Summary – allows only the first and the last set of alignment readings
and machine moves for the job(s) selected to be printed.
Job To Print: – Select one of three entries by pressing any numbered key
on the keypad.
• All – allows alignment data on all stored jobs to be printed.
Caution!
DO NOT connect the printer directly into the RS232 port located on
the top panel of the analyzer. Read the steps on pages 3-27 to 3-29
before printing directly from the analyzer.
Note
When printing via the Model 720A Printer Interface, the ana-
lyzer is not compatible with printers that use a serial interface.
19
5. ·· From the Capture menu select a capture mode: Screen, File, or File
and Screen.
If you select Screen, the information from the analyzer is sent directly
to the computer screen.
Notepad Screen
Note
The Notes key always returns you to the Alignment Notepad
screen. Press <Alt> <Note> to access the new user-defined
notes area (see “Creating Notes” on page 3-42 for more infor-
mation.)
Predefined Notes
The 70 predefined notes are separated into four categories:
• Feet/Base
• General Observations
• Coupling/Shaft
• Movement Related
Note
In the following tables, “fixed” will be replaced by the name of
the machine on the left and “move” will be replaced by the
name of the machine on the right.
Note # Description
Note # Description
Note # Description
Note # Description
56 Equipment problem
57 Both machines moved
58 Bolt bound horz, fixed IB
59 Bolt bound horz, fixed OB
60 Bolt bound horz, move IB
61 Bolt bound horz, move OB
62 Vert move limited, fixed IB
63 Vert move limited, fixed OB
64 Vert move limited, move IB
65 Vert move limited, move OB
66 Smaller hold down bolt used
67 Thermal growth inconsistent
68 Thermal growth unknown
69 Piping strain present, inlet
70 Piping strain present, outlet
Storing Notes
To store a note or notes for a machine, press the Notes key, then select the
category where the note exists. If you do not know the category or just want
to review existing job notes, select the (6) All Notes section. Using the up/
down arrow keys to scroll through the list, highlight the desired note
number and press the Enter key. The note text will then highlight automat-
ically, indicating that the note is stored.
Note
If no User Defined notes are present in memory, the Notepad
Data Entry screen will automatically be displayed after the
User Defined section is selected.
Note
Pressing the Alt key, then pressing the Notes key at any time
allows you to immediately create a User Defined note.
Enter the note text using the alphanumeric keypad (32 characters max-
imum) and then press the Enter key to log the note into the User Defined
section. The note will be placed at the end of the section and automatically
highlighted.
Note
By pressing Alt/Help while in the Utility Functions, you can
access the help screen information for the special language
characters (see “Special Language Characters” on page 2-18 for
more information).