Lab 4: Uni-Axial Tension Testing: Written By: Ryan Betz
Lab 4: Uni-Axial Tension Testing: Written By: Ryan Betz
1
Introduction:
When examining the correct material to use for a biomedical application, precision is necessary.
There are many different aspects of a material to examine from biocompatibility so the body
does not reject the material, to seeing if the material is strong enough to perform the desired task.
If the material is not compatible to the body, then the surgeons will have to perform surgery to
remove the material before serious harm comes about. If the material is too strong, there could be
a limitation formed by the metal such as the individual not having as much mobility after the
material is placed in the body. On the other hand, if a material is too weak then the material has a
chance to break and damage internal organs. Another risk associated with breaking of the
material is more surgery needed to fix the problem. Biomedical engineers have to figure out the
correct balance of stiffness along with ductility in order to allow the individual with the implant
the ability to perform functions that they could without the material. If the material cannot
mimic the normal functions of the body, injury could come about as the material will hinder
normal movement.
The goal of the lab was to examine the elastic modulus of the aluminum rod along with other
properties of the material. An aluminum rod (Aluminum 6061-T6) was used in the experiment to
measure the properties that the metal exhibited. The main device that helped to record data for
this experiment was the Tinius Olsen Testing Machine. This machine performs a uniaxial tension
test by locking one end of the rod down and pulling with an increasing force on one end. This
force along with the displacement from the starting position is vital in trying to solve for the
modulus of elasticity, along with other factors on the stress strain curve. One other gadget used
was an extensometer. The extensometer was used to measure the elastic deformation region as it
measures the small range of change of length. This small measurement of elastic deformation
2
allows for the elastic region of material to be determined. This is due to the Tinius Testing
Machine not having tension right at the beginning of the test and therefore the linear elastic line
cannot be found. The line instead of being straight had a curve due to the lag of not being taut to
The relevance to biomedical engineering, in specific, biomechanics is that the axial load applied
can be analyzed. Through this, the forces that the metal can withstand are able to be seen. One
way to get even better results is to place the metal in an environment that mimics the body. This
change in composition of the air and change in the heat that the material is exposed could help to
gain more accurate results. These alterations in comparison to not testing as if in the body could
change the properties of the material. If the properties were to change even the slightest this
could alter the functionality of the material which is bad. If the metal broke easier under ideal
body conditions this would be something that could not be used for an implant in the body as
This test even under not ideal body conditions is extremely useful to a biomedical engineer. This
is due to the relevant data collected not only about the stresses and strains, but how this compares
to biological tissue such as bone. If the testing of bone and a different material occurred and the
findings of the tests were similar, the material would be a good candidate for the body. The next
aspect to examine is seeing if the material is biocompatible. If so, the material such as titanium
The first part of the experiment dealt with creation of the LabVIEW code. This code was an
integral part to the lab as it helped to take the data from recording along with some user’s inputs
about the material and show not only stress versus strain curves, but what the ultimate strength
and tensile strength were among many things. The code had to take the data file of the results
3
from the UTM and sort the components by every third component equaling a different aspect of
what was being tested. From there the program performs many calculations to give the user all
the information about the material that is being tested. This type of program helps to reduce time
that the lab spends on acquiring results along with the fact that it could be used for more than one
thing. One key component before testing the metal and using the UTM is to zero the data so less
Results:
Table 1: These are the table of the variables used in the lab setting.
Table 2: This table shows the diameters found through the use of an electronic caliper and the
associated cross sectional area that can be calculated as well based on the diameter.
Original Cross Sectional Area
Diameter (mm) Area (mm2)
1 8.55 57.41
2 8.53 57.15
3 8.51 56.87
4 8.55 57.41
5 8.55 57.41
Average 8.54 57.25
4
Table 3: This table shows the measured lengths of the gage before and after testing. Also shown
is the final diameter measured at the fracture point of the material and the corresponding cross
sectional area. The last part of the table is the speed at which the UTM was operating when
pulling the Al bar.
Initial Gage Length 111.4 mm
Final Gage Length 118.9 mm
Final Diameter 6.05 mm
Final Cross Sectional Area 28.74 mm2
UTM Speed 2.5 mm/min
Table 4: This table shows the values when running the experiment using the LabVIEW to import
the data from extensometer and the Tinius Olsen Testing Machine. All of this data like fracture
strength could be generated when putting in initial diameter and initial gage length and final
diameter and gage length.
Ultimate Strength (Us) 298 MPa
Fracture Strength (Uf) 224 MPa
Modulus of Elasticity (E) 69.8 GPa
Yield Strength (σy) 274 MPa
Modulus of Resilience (Ur) 540 KPa
Ductility (D) .067
Area Reduction (%A) 49.7%
Table 5: This table represents the values found about the material and all their values. This is in
the middle column followed by the percent error in comparison to the data collected during the
lab. These percent error values can be found in the right hand column
Property Literature Values Percent Error
Ultimate Strength (Us) 310 MPa 3.87%
Fracture Strength (Uf) 209 MPa 6.69%
Modulus of Elasticity (E) 68.9 GPa 1.31%
Yield Strength (σy) 276 Mpa 0.72%
Modulus of Resilience (Ur) 530 KPa 1.89%
Ductility (D) Not given N/A
Area Reduction (%A) Not given N/A
.
Equations:
1) Force= ( Lcmax
30000 )
∗Number generated by UTM
5
0.1
3) Extensometer Strain= ( 30000 )∗Number generated by UTM
These are the three equations used for lab. For the first equation, the force in Pa was found by
taking the loading cell max and dividing that by 30000. For the experiment the load cell max was
50000 and this value was multiplied by every third value generated by the UTM. For the second
equation, this was jaw displacement. This is when the jaws exerted a greater force on the metal
increasing the strain. The calculation was 0.001mm* every third value generated by the UTM.
Finally, the extensometer strain. The equation took the number generated by the UTM and
multiplied it by the first value in the equation. Again the program took every third value and
multiplied it by this value. The values that were given by the UTM in order were force,
Discussion:
Through analyzing the results collected from the experiment in comparison to the literature
values some interesting conclusions can be made. The first part that is interesting is the ultimate
strength of the material was found to be around 3.87% less than the literature value. This means
the max force that the material can withstand is smaller than the literature value for Al 6061-T6.
Another component that was interesting was the fracture strength of the material tested in the lab
was 6.7% greater. This means that the fracture occurred at a higher force than the literature
values. Some other components to analyze between the literature values and the experimental
data are Modulus of Resilience and Modulus of Elasticity. Both of the values determined in lab
were slightly bigger, 1.89% and 1.31% respectively, which means that the material tested in lab
had a greater elasticity region as the slope was larger. Finally, the yield strength was compared
6
between literature values and the experimental. The experimental yield strength was found to be
During the lab there were many sources of error. One of the sources dealt with measuring the
initial gage length along with the diameter. The tool used to measure these components was an
electronic caliper, but there is still human error associated with this type of measuring. As a lab
group, the goal was to minimize error by taking five readings of the diameter and averaging
them. If the caliper was not zeroed then the measurements would be off. For the gage length only
one measurement was taken so there could be a good amount of error associated with the value
and this is not a good thing. Finally, in measuring the diameter and gage length after testing there
is a decent amount of error as the smallest diameter measured might not have been the actual
smallest. This along with the gage length could have been off so there might have been error in
measurement of the items. These are all human errors associated with the measuring.
Some other source of error which is key can be seen in figure 1 in the appendix, but the top
graph. The error in the graph is associated to the starting position which was -0.05. The graph is
correct other than this one error. This error was associated with the strain gauge of the UTM not
being reset fully. This means that instead of starting the experiment after setup at 0, the
experiment started in the negative. For other groups their total strain to fracture was about 0.15
which is exactly what the value in the graph below shows, but the only difference is the starting
value. This is an error associated with the lab except the error is not detrimental only shows a
flaw in the procedure. If the experiment was to be completed again, the thing to make sure of
before testing the metal bar would be to zero all the recording mechanisms. This will give the
best values for the experiment and show a graph that is correct instead of starting in negative
7
The effects of the error in recording has no effect on the calculations just the data presented as
the graph is off. This is in comparison to the errors in the measuring of the bar which might
affect calculations and therefore mess up some of the data of the lab. Error is associated in most
labs, but the goal is to do everything necessary to make sure that the error is minimized. Any
examples can range from a procedural component like not zeroing the machine to a
The Tinius Olsen is a great machine when it comes to testing uni-axial loads on materials. This
can allow biomedical engineers to try and examine if different metals and materials that are
biocompatible and have the ability to withstand the loads that the body undergoes in day to day
life. Through testing biomedical engineers can not only choose the metal, but also design
materials that will help replace or aid with bones. Another test that can occur in the Tinius Olsen
machine is the use of soft tissue and polymers to replace tissues in the body and see what loads
they can undergo. The Tinius Olsen machine is accurate in allowing biomedical engineers the
ability to test different substances in an effort to replace bone or tissue depending on what is
References:
[1]"ASM material data sheet," in ASM Aerospace. [Online]. Available:
http://asm.matweb.com/search/SpecificMaterial.asp?bassnum=MA6061t6. Accessed: Oct. 21,
2016.
[2] Gielo-Perczak, Krystyna, Ph.D. “Lab#4 Uni-Axial Tension Testing.” BME 3600 Lab.
Bronwell, Storrs. 28 Sept. 2016. Lab
Appendix:
8
Figure 1: This figure shows the front panel of the LABVIEW program used to run the data
collected from the extensometer and the Tinius Olsen Testing Machine. Different values found in
the program were Modulus of Elasticity to ultimate strength of the material.
9
Figure 2: This figure shows the block diagram of the LABVIEW program used to run the data
collected from the extensometer and the Tinius Olsen Testing Machine.
10