Manual Smartcell
Manual Smartcell
ENGINEERING LTD
109080
INSTRUCTION MANUAL
BILGE OIL ALARM
Section 1
Introduction
1.1 Important Notes 3
1.2 SMART CELL - BILGE Overview 4
1.3 Specification Listing 5-6
Section 2
Installation
2.1 Construction details 7
2.2 Mounting 8
2.3 Pipe work and Connections 9
2.4 Electrical Connections 10 - 13
Section 3
Commissioning
3.1 Piping 14
3.2 Electrical 14
3.3 Functional testing 14
3.4 Throttle 14
Section 4
Operating Instructions
4.1 Getting Started 15 - 16
4.2 Clean Water Calibration 17
4.3 Set Alarm Points 18
4.4 Set Alarm Delays 19
4.5 Set Analogue out 20
4.6 View IMO Data 21
4.7 Diagnostics & Clean Water Flushing 22
4.8 Test Procedure 23
4.9 Monitor Verification 23
Section 5
Maintenance and Spares
5.1 Maintenance (weekly) 24
5.2 Maintenance (Yearly) 24
5.3 Maintenance (5 Years) 24
5.4 Spares 25
Section 6
Trouble Shooting & Contact
6.1 Trouble Shooting Guide 26 - 27
6.2 Faultfinding 28
6.3 Additional Features and Functions 29
6.4 Tech Support registration 29
Annexe
Certificates
EC MED & USCG Certificates 30 – 34
RIVERTRACE ENGINEERING LIMITED Page 2 of 34
SMART CELL - BILGE - Instruction Manual
Doc. 109080 Rev. S 21.10.11 ECN 4102
Section 1
Introduction
1.1 Important Notes
All National or local codes of practice or regulations must be observed and, where
applicable, are deemed to take precedence over any directive or information contained in
this manual.
This equipment must be installed and operated in strict accordance with the instructions
contained in this manual. Failure to do so will impair the protection provided.
Installation and retro fitting of parts listed in Section 5.3 must be undertaken by a competent
and suitably skilled person.
The unit must be isolated from the electrical supply before any maintenance of the
equipment is attempted.
Hazardous voltages are present within the monitor enclosure and solenoid valve assembly.
It is essential that maintenance is carried out as described in Section 5. Failure to do this will
impair the functionality of the equipment and could lead to permanent damage to the
monitor.
The SMART CELL enclosure must not be opened. If opened a cell breach error will be
displayed on the screen (this error cannot be reset) and the “VOID” tamper proof labels on
either side of the enclosure lid will be destroyed. Forcing the throttle to its extents will also
damage the cell and if unscrewed will distort the cell lid thereby causing a cell breach error.
Analysis of the oily water sample is undertaken by a unique detector array arrangement. A
narrow beam of light is emitted across a glass cell tube which the oily water sample passes
through. Light scattered and transmitted is detected across the array and passed through a
complex oil content calculation algorithm. The resulting oil content is then displayed on the LCD
screen.
Two oil alarm relays and a 0-20mA / 4-20mA output are provided to control the overboard
discharge and to view / record oil content levels remotely.
To meet the new requirements of MEPC 107(49) an on board flash memory, Real time clock,
dual switch input and clean water / sample selection valve are fitted. These are described in
more detail in the operating instructions (Section 4).
Alarms
Oil alarm 1 setpoint: 1 – 15 ppm user adjustable
Oil alarm 2 setpoint: 1 - 15 ppm user adjustable
Oil alarm 1 delay: 0 – 5 seconds user adjustable (For Overboard Valve)
Oil alarm 2 delay: 0 – 600 seconds user adjustable in 10 sec increments
(For Annunciation only)
Alarm hysteresis: 0.5 ppm
Alarm contacts: 2 SP Alarm Relays 10A @ 250VAC (NC in alarm)
User Interface
LCD display: 4 x 16 alphanumeric back lit LCD display
Control: 3 button keypad
Input / Output
Analogue output: Loop powered 4-20mA / 0 - 20mA 24V
Communications: RS232 - output
Cable terminals: Accept cores of 2.5mm² (HV) and 1.5mm² (LV)
Cable glands: 4 x to accept cable diameters 10 – 14mm dia.
Switch inputs: 2 x switch inputs for separator & flow status
Alarm 1: 15 ppm
Alarm 1 delay period: 5 seconds
Alarm 2: 15 ppm
Alarm 2 delay period: 70 seconds
Current loop: 4-20mA
Autoclean: Not set
Flowswitch: Not set
Section 2
Installation
2.1 Construction Details
The SMART CELL - BILGE Monitor is split into two logical modules. The display / control module
(2) (left) contains the membrane keypad, LCD display and the Smart Bilge Control PCB.
Contained on this PCB are all the control and data storage associated electronics in addition to
all the customer terminations which can be accessed via the cable glands(5). This PCB links a
5 volt supply and RS485 communications to the Cell electronics via a 4 wire interface cable.
The SMART CELL (1), contains the optoelectronics and mechanics used in the calculation of the
sample oil content. UNDER NO CIRCUMSTANCE IS THIS ENCLOSURE TO BE OPENED. Sample (7)
and clean water (8) lines are connected to the solenoid valve to allow automatic selection of
clean water for clean water calibration and diagnostics. Sample/Clean water is exhausted
from the monitor via the outlet (6) after passing through the measuring sensor. A manual cell
clean unit (3) is provided which must be used in accordance with the maintenance
requirements in Section 5. The throttle (9) can be used to increase or decrease the back
pressure in the cell. This is especially useful if air is present in the sample.
Fig 2.1
Construction details
Mounting
Mount the SMART CELL - BILGE by means of 4 x M6 bolts on to a rigid vertical surface and
preferably with the display panel of the monitor at operator eye level.
Fig 2.2
Mounting Details
Fig 2.3
Pipe Fittings
Please read the following points before attempting any electrical work.
User terminations are located by removing the lid from the SMART Display/Control module via
the four screws. A protection plate is situated above the terminals which must be removed via
4 M3 screws.
Fig 2.4.1
User Terminations
Table 2.4.2
User Terminal Connections
J9 - 1 5V DC from PCB
Separator Status Switch Input 1 5V DC 20mA
J9 - 2 Switch In
J11 - 1 Tx
J11 - 2 GND RS232 Connection +/-12V DC 10mA
J11 - 3 Rx
Precise wiring details will vary dependent upon the control system to be employed but Alarm
relay 1 must be used for control of the overboard discharge valve and Alarm relay 2 for alarm
annunciation only. The separator status switch input should be connected to a suitable volt-free
contact in the separator. A separate screened cable should be used for this connection.
Fig 2.4.3
User Wiring Connections
Mains Input
Overboard To Engine
Discharge Room Alarm
Valve
Smart Bilge Electrical Connections
(if there is a separate discharge and return to bilge valve)
J1 J2 J3 J4 J5
L E N L E N L E N 1 2 3 1 2 3
Mains Input
To Engine
Room Alarm
Return to Overboard
Bilge Discharge
Valve Valve
If the separator does not have a “Separator Status” volt-free contact then a suitable volt-free
contact will have to be wired in parallel with the motor contactor for the separator pump. In all
cases the wiring between the volt-free contact and the separator status input SW1 on the bilge
monitor should be a screened cable.
Section 3
Commissioning
3.1 Piping
Check all piping for leaks and rectify where appropriate
Ensure all piping and fittings conform to specifications set in Section 2.3.
3.2 Electrical
Check that correctly rated isolator is connected to the mains input (see Section 2.4.).
Check the SMART CELL - BILGE is correctly earthed.
Check all cabling conforms to specifications set out in Table 2.4.2
3.4 Throttle
The throttle is used to regulate the flow through the measuring cell and to provide backpressure
to decrease any aeration in the sample flowing from the separator.
The throttle should not be forced beyond its end stops – if rotated clockwise beyond its stop it is
possible to strip its thread thereby causing the cell to leak. If the throttle is rotated anticlockwise
beyond its stop then it is possible to distort the cell case lid which in turn will activate a cell
breach error. This will require intervention by RTE
IF you experience fluctuating ppm readings then firstly try rotating the throttle clockwise to increase
the backpressure. Monitor the flow from the cell. You should ensure that at the flow rate does not fall
below 0.5 litres/min.
Section 4
Operating instructions
Fig 4.1.1
User interface
4 x 16 Alphanumeric LCD
Display
Fig 4.1.2
Screen Shot – Start Screen
After initialisation the user will be presented with a screen as shown in Fig 4.1.3.
At the top of the display the user will see the oil concentration in Parts Per Million (PPM).
The status of the two oil alarms will be displayed.
The date and time is displayed. (Time is set to GMT/UTC and can only be adjusted by RTE
approved engineer).
If either of the alarm set points is passed the LCD backlight will flash.
Fig 4.1.3
Screen Shot - Main Information Screen
Shows Oil
Concentration
in PPM
Date & Time
Alarm 1 state Display
Alarm 2 state
To interrogate the SMART CELL - BILGE Monitor settings the user can press any of the Interface
Keys which brings up the Menu Options Screen as shown in Fig 4.1.4
Fig 4.1.4
Screen Shot - Menu Screen Options
A clean water calibration can be performed to account for differences in clean water between
the factory calibrated clean water and the ships own clean water supply. On activation of the
clean water calibration the clean water / sample solenoid will select clean water. The SMART
CELL - BILGE will then average the clean water reading over the period of 30 seconds after
which a further 10 seconds will lapse for the user to verify the calibration in monitor mode before
the solenoid selects sample once more.
To set the clean water calibration enter the Menu screen as shown in Fig. 4.1.3 and scroll down
until the cursor reaches "Clean wtr cal" then press the ENTER key. The screen will change as
shown in Fig 4.2.2
Fig 4.2.2
Screen Shot - Menu Screen Options
Note: The alarms have a 0.5 ppm hysteresis included to prevent relay chatter. e.g if an alarm is
set at 15 ppm and the oil concentration exceeds 15 ppm, the alarm will be activated. The
alarm will not deactivate until the oil concentration drops below 14.5 ppm.
To set the required Alarm points enter the Menu screen as shown in Fig. 4.1.3 and scroll down
until the cursor reaches "Set alarm point" then press the ENTER key. The screen will change as
shown in Fig 4.3.1.
Fig 4.3.1
Screen Shot - Alarm Set Options
Fig 4.3.2
Screen Shot - Alarm PPM Value
Note: The main information display screen will show if an oil alarm is less or more than the
current oil reading regardless of the alarm delay.
To set the required alarm delays go the menu screen as shown in Fig. 4.1.3 and scroll down until
the "Set alarm delay" then press the ENTER key. The screen will change as shown in Fig 4.4.1
Fig 4.4.1
Screen Shot - Alarm Delay Options
Fig 4.4.2
Screen Shot - Alarm Set Delay Value
Fig 4.4.3
Screen Shot – Version Info
Fig 4.5.1
Screen Shot - Analogue Output Options
The separator status is recorded via the SW1 and SW2 inputs. The following configuration is
used.
SW1 Separator ON/OFF status.
SW2 Separator Back-flush (If fitted)
Note. SW2 will select clean water and activate alarms permanently when closed. To be
connected to Separator backflush control circuit via a potential free contact.
The configuration of the switches must be recorded on board and date stamped.
To view or download the IMO required data go to the Menu screen as shown in Fig. 4.1.3 and
move the cursor down until "View IMO data" then press the ENTER key. The screen will change as
shown in Fig 4.6.1.
Fig 4.6.1
Screen Shot - View IMO Data
Fig 4.6.2
Screen Shot - IMO Data
Fig 4.6.3
Screen Shot - Dump IMO Data
NOTE: Dump IMO data command will NOT erase any data
Fig 4.7.1
Screen Shot - Diagnostics
Fig 4.7.2
Screen Shot – View detectors
Fig 4.7.3
Screen Shot – View disp PPM
Fig 4.7.4
Screen Shot – Temperature
The following procedure is to be adopted to test the Smart Cell Bilge Monitor
Section 5
Maintenance and Spares
5.1 Weekly Maintenance
At yearly intervals (or where necessary) the wiper ring on the manual clean assembly should be
replaced. This can be ordered from RTE directly as part no. 109039.
To remove wiper assembly, electrically isolate the monitor. Have some rags available to wipe up
any spills.
Ensure that the separator is switched off and there is no flow through the monitor cell.
Isolate clean water supply and remove manualclean by using the knurled cap to unscrew it.
Pull the manualclean from the cell in a vertical motion and wipe with a rag. Locate the wiper
and prise off with a small flat bladed screwdriver. Fit the new wiper by pushing it against a flat
surface and rotating the manualclean at the same time.
Refit the manualclean into the cell and de-isolate power and water supplies.
IF required by your environmental or quality policy, it is possible to verify the accuracy of the
monitor by purchasing a Calibration Verification kit. This can be purchased directly from
Rivertrace or by one of our agents or service centres. If the calibration verification is carried out
by one of our service centres then a Calibration Verification Certificate will be issued by them on
behalf of Rivertrace.
5.4 Spares
When ordering spares, it is important to supply details of the type of monitor, serial number of
monitor, part number of each spare required, its description and any other relevant information.
OTHER SPARES
Section 6
Trouble Shooting Guide
Monitor stays in alarm or Dirty measuring cell. Clean cell and perform Clean
reading is inaccurate. water calibration. Section 5.1
Display shows SYS ERROR, Air bubbles are present in Turn throttle clockwise until air is
Oil content Is under / over the Sample. removed from sample (taking
range. care not to exceed the maximum
operating pressure.
Monitor ppm remains high Measuring Cell is empty. Check water flowrate through
and does not change measuring cell is >0.5l/min.
Display shows Warning, flow Flow through measuring cell Increase flow through monitor
Error detected not sufficient. cell, Remove pipework from
**Only If flow switch fitted** sample inlet to solenoid and
check that there is no debris
blocking the port.
Display shows SYS ERROR, Current Ambient Move equipment to an area
Temperature Is under / over temperature inside the Cell with ambient temp between
range. is not within 0 - 60°°C. 0 - 60°°C.
Display shows SYS ERROR, RS 485 communications Ensure 4-pin plug is tightly
Communications failure cable is loose / not inserted into J15. Fig 2.4.1
connected
Display shows SYS ERROR, Smart Bilge electronics Contact RTE for a replacement
RTC – I2C failure failure. Smart Bilge PCB (Section 5.4).
Display shows SYS ERROR, Smart Bilge electronics Contact RTE for a replacement
EEPROM Failure failure. Smart Bilge PCB (Section 5.4).
Display shows SYS ERROR, Measuring Cell has been Contact RTE for a replacement
Cell breach contact RTE opened or overuse of the measuring cell (Section 5.4).
throttle has damaged the
cell.
Display shows WARNING Manual Clean has become Press the Manual clean UP &
PLUNGER DOWN DETECTED stuck in the down position. DOWN a few times to un stick.
Check to ensure Manual Clean
The PPM value will rise to the Operates as normal. Screen
MAX setting, thus energising message will disappear.
the Alarms as set in 4.3.1. If system persists remove Manual
Once the Manual Clean Clean and check Wiper.
has been released the PPM Replace if wear evident or wipe
value will count down. damaged.
Once it has gone below the
Alarm set point the Alarms
will be de-energised.
This function can also be
used to simulate a high PPM
reading thus triggering the
Alarms, and then monitoring
the Alarms when the PPM
rate falls to within the set
limits.
6.2 Faultfinding
There are additional features that can be added to the Smart Bilge monitor which increases security
and lessens maintenance. They are the flowswitch module and the autoclean module.
6.2.1.1 Flowswitch
6.2.1.2 Autoclean
If Autoclean has been invoked in the software then
the user menu will include “Autoclean” as a sub
menu.
(Note: this is an option and not available on the
standard monitor).