Cobas c311 Check List

This document provides a maintenance checklist for a cobas 311 analyzer. It lists various components that should be checked, cleaned, replaced, adjusted, lubricated or have tasks done as part of routine maintenance. The checklist is comprehensive and covers components like reaction baths, pipettors, syringes, ISE unit, pumps, electronics and more.
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
100% found this document useful (4 votes)
2K views2 pages

Cobas c311 Check List

This document provides a maintenance checklist for a cobas 311 analyzer. It lists various components that should be checked, cleaned, replaced, adjusted, lubricated or have tasks done as part of routine maintenance. The checklist is comprehensive and covers components like reaction baths, pipettors, syringes, ISE unit, pumps, electronics and more.
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 2

Maintenance Checklist - cobas® 311

Site: .................................................. Site Contact: ……............................................

S/N: .................................................. SW Version: ...................................................


U C R A L T Remarks Done
Cover / housing X
Reaction bath / disc
Reaction bath filling , circulation X
Reaction bath temperature X
Reaction bath X
Reaction bath filter ( metal) X
Reaction cuvette X
Ultra Sonic Mixer
Check USM X
Photometer
Lamp cooling X
Photometer lens clean X
Photocurrent monitor
Heat Cut Filter X
Photocurrent values (specially
for 340 X
nm) blank
Cell X
Sample Pipettor
Needle X
Wash bath Contamination X
Rinse water adjustment
Drying cylinder contamination X
Sample probe adjustment X
Shaft X X X
Function and movement X
Pressure sensor (Clot- X
detection)
Reagent Pipettor
Needle contamination X
Wash bath Contamination X
Rinse water adjustment X
Drying cylinder contamination X
Probe adjustment X
Shaft, linear guides X X
Function and movement X
Pressure sensor ( Init. Vol. X
check)
Piercer
Contamination and condition X
Reagent disk
Reagent disk contamination X
check disk sensors X

Version 1.0_07.12.2012
Rinse unit
Tubing, connections + photo
interrupter X
Rinse nozzle X
Nozzle tip X
Linear guides X
Rinse water X
Sampling Rotor
Contamination + photo X
interrupter
Barcode reader X
Belt tension X
Syringes: Sample, Reagent,
ISE
Syringe seals X
Gear, pinion, shaft, steel plate X
Bubbles, leakages X
Syringe movement + X
photointerrupter
ISE Unit
Sipper Nozzle X
Sipper Tube and Pinch Valve X
Tube Nozzle / contamination /
Sipper
function X
Sipper waste pipe X
Airbubbles in tubing X
ISE electrodes X
Water pumps
Gearpump pressure a(300 X
kPa) pump (80-100 kPa)
Main X
Power supplies, electronics
Fan, contamination X
Boards X
Adjustments power supply
(5V, 12V, 24V) X X
Vacuum system
Diaphragm a X
Tubings / vacuum bottles X
Tank X
Hepa Filter X
Cooling unit
Reagent cooling function X
Fan, dust filter X
Condenser X
Miscellaneous
Greasing of probe guard X
Mechanism check
(incl. Photointerrupter) X
Calibration & QC Checks X

C: Check/ Clean R: Replace A: Adjust L: Lubricate T: To do

Comments: ……………………………………………………………………………………
…………………………………………………………………………………………………
…………………………………………………………………………………………………
FSR: ................................................................ Date: ................................................................

Version 1.0_07.12.2012