General Monitors S5000 Manual-En
General Monitors S5000 Manual-En
General Monitors S5000 Manual-En
MSAsafety.com
WARNING!
Read this manual carefully before using the device. The device will perform as designed only if it is used
and maintained in accordance with the manufacturer's instructions. Otherwise, it could fail to perform as
designed and persons who rely on this device for their safety could sustain serious personal injury or
death.
The warranties made by GM with respect to the product are voided if the product is not installed, used
and serviced in accordance with the instructions in this manual. Please protect yourself and your
employees by following the instructions.
Please read and observe the WARNINGS and CAUTIONS inside. For any additional information relative
to use or repair, call +1-949-581-4464 during regular working hours.
1 Safety Regulations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
1.1 Correct Use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
1.2 Product Warranty . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
2 Description. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
2.1 Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
2.2 No Tool Interface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
2.3 Bluetooth® Wireless Technology . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
2.4 Dual Sensing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
2.5 Retrofit Installations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
2.6 XCell Sensors Optimized for Fixed Gas Applications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
2.7 TruCal Sensing Technology for CO and H2S Electrochemical Sensors . . . . . . . . . . . . . . . . . . . . . 9
2.8 Safe Swap . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
2.9 Housing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
2.10 Component Overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
3 Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
3.1 Installation Warnings - Read before Installation. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
3.2 Reviewing Shipment and Identifying Product Model . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
3.3 Product Installation Check List . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
3.4 Mounting. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
3.4.1 Sensor Mounting Location . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
3.4.2 Transmitter Mounting Location. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
3.4.3 Sensor Orientation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
3.4.4 Connecting Sensor to Transmitter Housing or Remote Junction Box . . . . . . . . . . . . . . . . . . . . . . 14
3.4.5 Integrated Mounting Points . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
3.4.6 Adjustable Pipe Mount . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
3.4.7 Duct Mount . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
3.4.8 Mounting with a Sunshield . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
3.5 Installing a Remote Sensor Junction Box. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
3.6 Electrical Power Connections. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
3.6.1 Electrical Warnings - Read before Connecting Power . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
3.6.2 Electrical Hardware Requirements. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
3.6.3 Power Load Requirements and Maximum Mounting Distances. . . . . . . . . . . . . . . . . . . . . . . . . . . 21 US
3.6.4 Instructions for Power and Analog Output . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
3.6.5 Relay Electrical and Power Connections . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
4 Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
4.1 Startup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
4.1.1 Initial Startup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
4.1.2 Startup after Power Failure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
5 Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42
5.1 Calibration Equipment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42
5.2 Calibration Frequency . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 43
5.3 Calibration Frequency for XCell Sensors with TruCal (H2S & CO only). . . . . . . . . . . . . . . . . . . . . 44
5.4 Calibration Types: Zero vs. Span . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 44
5.5 How to Zero Calibrate XCell and IR400 Sensors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 45
5.6 How to Calibrate XCell Sensors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 46
5.7 How to Calibrate an Oxygen XCell Sensor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 47
5.8 How to Calibrate an IR400 Sensor. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 47
5.9 How to Calibrate a Passive Sensor (Catalytic Bead or MOS) . . . . . . . . . . . . . . . . . . . . . . . . . . . . 47
5.10 XCell Catalytic Bead Failsafe . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 48
US
5.11 Calibration Confirmation. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 48
6 Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 48
6.1 IR400 Cleaning Procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 49
6.2 Replacing an XCell Sensor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 50
6.3 Replacing a Passive Sensor (Cat Bead or MOS). . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 51
6.4 Troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 51
7 Ordering Information. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 55
7.1 Replacement Parts. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 55
8 Appendix: Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 56
1 Safety Regulations
1.1 Correct Use
The S5000 Gas Monitor, hereafter also called device, is a gas monitor for measuring toxic and
combustible gases as well as oxygen. Using sensors, the device tests the ambient air and triggers
the alarm as soon as the gas exceeds a specific concentration level.
WARNING!
Read this manual carefully before servicing the device. The device will perform as designed only
if it is used and maintained in accordance with the manufacturer's instructions. Otherwise, it could
fail to perform as designed and persons who rely on this device for their safety could sustain
serious personal injury or death.
WARNING!
Do not use silicone-type lubricants in assembling the device and do not allow silicone vapors to
be drawn into the flow system while in operation. Silicone can desensitize the combustible gas
sensor, thereby giving erroneously low readings.
Use only genuine MSA replacement parts when performing any maintenance procedures on the
device. Substitution of components can seriously impair performance.
Failure to follow the above can result in serious personal injury or loss of life.
This device complies with Part 15 of the FCC Rules. Operation is subject to the
following two conditions:
• this device may not cause harmful interference, and
• this device must accept any interference received, including interference that may
cause undesired operation.
This equipment has been tested and found to comply with the limits for a Class A digital
device, pursuant to Part 15 of the FCC Rules. These limits are designed to provide
reasonable protection against harmful interference when the equipment is operated in
a commercial environment. This equipment generates, uses, and can radiate radio
frequency energy and, if not installed and used in accordance with the instruction
manual, may cause harmful interference to radio communications. Operation of this
equipment in a residential area is likely to cause harmful interference in which case the
user will be required to correct the interference at his own expense.
US
NOTICE
This is a Class A product in accordance with CISPR 22. In a domestic environment, this product
may cause radio interference, in which case the user may be required to take adequate measures.
NOTICE
The XCell sensor refers to the sensor portion of the Digital Sensor throughout this manual.
This warranty does not cover filters, fuses, etc. Certain other accessories not specifically listed
here may have different warranty periods. This warranty is valid only if the product is maintained
and used in accordance with Seller's instructions and/or recommendations. The Seller shall be
released from all obligations under this warranty in the event repairs or modifications are made by
persons other than its own or authorized service personnel or if the warranty claim results from
physical abuse or misuse of the product. No agent, employee or representative of the Seller has
any authority to bind the Seller to any affirmation, representation or warranty concerning this
product. Seller makes no warranty concerning components or accessories not manufactured by
the Seller, but will pass on to the Purchaser all warranties of manufacturers of such components.
THIS WARRANTY IS IN LIEU OF ALL OTHER WARRANTIES, EXPRESSED, IMPLIED OR
STATUTORY, AND IS STRICTLY LIMITED TO THE TERMS HEREOF. SELLER
SPECIFICALLY DISCLAIMS ANY WARRANTY OF MERCHANTABILITY OR OF FITNESS
FOR A PARTICULAR PURPOSE.
Exclusive Remedy
It is expressly agreed that Purchaser's sole and exclusive remedy for breach of the above
warranty, for any tortious conduct of Seller, or for any other cause of action, shall be the
replacement at Seller's option, of any equipment or parts thereof, which after examination by US
Seller is proven to be defective. Replacement equipment and/or parts will be provided at no cost
to Purchaser, F.O.B. Seller's Plant. Failure of Seller to successfully replace any nonconforming
equipment or parts shall not cause the remedy established hereby to fail of its essential purpose.
Exclusion of Consequential Damage
Purchaser specifically understands and agrees that under no circumstances will seller be liable to
purchaser for economic, special, incidental or consequential damages or losses of any kind
whatsoever, including but not limited to, loss of anticipated profits and any other loss caused by
reason of non-operation of the goods. This exclusion is applicable to claims for breach of warranty,
tortuous conduct or any other cause of action against seller.
2 Description
2.1 Display
The S5000 utilizes a dot matrix LED display, capable of displaying four alphanumeric characters
at a time. The display will scroll words that exceed four letters. Most of the messages scroll twice
across the screen before moving onto next selection.
Fig. 2 Interface
The user menu can also be accessed using the round GM magnet on the General Moni-
tors logo.
WARNING!
Bluetooth® Operation is dependent upon signal availability of the wireless service(s) necessary to
maintain the communication link. Loss of wireless signal will prevent communication of alarms and
other information to linked devices. Take appropriate precautions in the event a loss of wireless
signal occurs.
If ordered with Bluetooth®, the device will be shipped with Bluetooth® enabled. See
section 4 for instructions on disabling Bluetooth®. US
If the device was not ordered with Bluetooth® it cannot be upgraded.
The S5000 and user provided communication device will need to be paired. This requires both
devices to be in range and for a pairing sequence inputting a 6 digit pairing code. The instructions
will be displayed on both the S5000 and communication device.
There are communication devices capable of being used in classified areas. Please
contact your MSA representative for additional information.
2.9 Housing
The S5000 comes in 316 Stainless Steel for the highest corrosion resistance. All housings have
¾” NPT conduit entries. Custom tags are available and easily attach to an integral ring.
2.10 Component Overview
3 Installation
3.1 Installation Warnings - Read before Installation
WARNING!
S5000 Digital Sensors labeled for Class I Hazardous (Classified) locations and installed in a
Class I Hazardous (Classified) location only may be changed under power. To change, unscrew
sensor at least three full turns (but no more than four full turns from its fully engaged position),
wait 10 seconds, and then remove the sensor completely. Failure to follow this warning can
result in the ignition of a hazardous atmosphere.
Do not paint the device. Avoid painting in areas where the S5000 and remote sensor junction box
are located. If painting is required in an area where an S5000 or remote sensor has been installed,
exercise caution to ensure paint is not deposited on the sensor inlet fitting. Paint solvents can also
cause an alarm condition to occur or potentially poison electrochemical sensors.
Protect the device from extreme vibration.
Do not mount the sensing head in direct sunlight without a sunshield (Part Number 10178977,
IR400 Sunshield Part Number 10178977).
The S4000 IR400 Gas Monitor contains no user- or field-serviceable parts and must be returned
to the factory for repair. Any attempt to open the monitor will damage the unit and void the
warranty.
CAUTION!
When installing the IR400 sensor, under no circumstances should a pry-bar be applied to the two
legs that support the unit's reflectors during installation or removal of the sensor. Applying force
to the legs can permanently damage the IR400 sensor.
US
US
Fig. 5 Correct and Incorrect Mounting Orientations for IR400 Point IR Sensor
All other sensors, including electrochemical, catalytic bead, oxygen, passive catalytic bead, and
passive MOS should be mounted vertically with the gas inlet pointed downward. If the sensor is
not mounted with the gas inlet facing down, it is more likely to become clogged with particulate
matter or liquids. Fig. 6 shows the correct and incorrect mounting orientation for XCell sensors.
Passive catalytic bead and MOS sensors come already installed on the transmitter housing.
US
CAUTION!
If only using one sensor and it is connected to “Sensor 2” position, the S5000 will enter Sensor
Missing fault.
NOTICE
Ensure that the electronics assembly is fully engaged in the mounting holes. If not fully seated,
the user interface button may not function properly.
US
Fig. 9 Integral Mounting Tabs (not compatible with S4000 IR400 sensors)
3.4.6 Adjustable Pipe Mount
A Universal Pipe Mount Kit (Part Number 10176946) can be used to mount the S5000 on pipes
ranging from 20-150 mm in diameter. Two brackets are mounted over the top of the integrated
mounting tabs and fitted with an adjustable pipe band (not included).
US
NOTICE
Consider the sensor type before choosing a duct mount location. IR400 sensors should be
mounted horizontally and all other sensors should be mounted vertically.
NOTICE
Air flow in the duct must be zero to ensure proper calibration.
US
US
US
Tab. 17 Power Load Requirements and Maximum Mounting Distances for 24 V Power Supply1,2
1 Values listed are only for one transmitter and the indicated sensor options. The total number
of transmitters needs to be considered when specifying a power supply
2
Assumes transmitter was ordered with relays
3 When sizing a system's 24 V supply, a 1 A inrush current with a 1 ms duration should be
considered for each device on the power supply
3.6.4 Instructions for Power and Analog Output
CAUTION!
Read all electrical warnings and wiring requirements before connecting power to the S5000.
Connector for HART analog output and power are provided to increase ease of connecting power.
Connect power and remote sensor cable shields to shield terminals on the main PC board. US
Provide shield terminations inside he sensor housing as indicated on the installation outline
drawing.
(1) Remove the cover by turning counter-clockwise.
(2) Remove the display module to expose the wiring terminations and sensor connections.
(3) Remove the 5.08 mm pitch connector for power supply.
The power connector is larger than other 3.81 mm pitch connectors.
(4) Use a small flat head screw driver to open wire entries on connector.
(5) Insert wires to connector so that when installed each wire is in the correct location.
1. +DC
2. -DC
3. mA1 - analog output of sensor 1
4. mA2 - analog output of sensor 2
(6) Tighten screws on connector and tug gently on wires to ensure they are secure.
(7) Attach the connector to the board stack.
(8) Make sure the appropriate wires are in the correct terminals.
(9) Remove enough of the wire housing to expose the 3-4 inches of the cable shielding, but do
not expose so much that it goes beyond the cable entry.
(10) Attach the cable, shielding exposed, to the grounding point.
(11) Replace the display module. Push firmly on the board stack where indicated.
CAUTION!
Ensure that the electronics assembly is fully engaged in the mounting holes. If not fully seated,
the touch interface performance can be negatively affected.
Care must be taken to insure the S5000 inside glass surface glass is free of smudges/dirt and
grease. Dirt and grease can interfere with the touch interface of the display. US
NOTICE
Avoid pressing on the left and right areas where the LEDs are located. Pressing directly on the
display will damage the display and will void the warranty. Replace the S5000 cover by turning
clockwise. Be sure to align threads to avoid cross-threading.
Exceeding the volt-amp rating of the relay can cause damage to the switching contacts.
Energy State NC NO
De-Energized (default) Normally Closed Normally Open
Energized Normally Open Normally Closed
4.2 Settings
The S5000 is a tool free transmitter. The infrared EZ touch button on the face of the display can
be used to navigate through the menu structure. The button is designed for use with fingers with
a “press” and “release” action, and works best without gloves.
The button works the same as the magnet does with the S4000 menu.
Changing a value
(1) Press and hold EZ Touch button.
(2) Wait for relevant menu to scroll (each menu scrolls twice).
(3) Release to enter menu while it is scrolling.
(4) Use Press and Release function to change values.
The EZ touch interface can be disabled, but will require a password. Menu settings can also be
activated using a magnet on the General Monitors logo.
Values that are changed in the menu are saved after main setup loop “Finished”, except for sensor
range selection under sensor setup. Each menu ends with "Finished?" scrolling twice on the
display. If the button is not touched during "Finished?", the menu will begin at the start again and
scroll through the options and values again. The new values that were entered will be displayed
on the first pass.
When user exits a menu by touching "Finished?" and there isn't a second sub menu, the previous
menu will be displayed starting at the menu that had just been used.
Some instrument settings are only configurable through Bluetooth® X/S Connect App,
Modbus, or HART. See section 4.4.
US
US
Alarm Cal
Range Unit Warn Alarm Alarm Alarm Range Range
Gas (Code) Action Level
Default3 Default Default Default Min Max Min max
Default Default
Increasing
Passive CB (C01) 0-100 % LEL 20 60 50 5 60 0-90 0-100
Non-Latch
Increasing
Passive MOS (M04) 0-100 PPM 30 60 50 5 95 0-50 0-100
Non-Latch
Increasing
Passive MOS (M05) 0-50 PPM 15 30 25 3 48 0-25 0-00
Non-Latch
Increasing
Passive MOS (M06) 0-20 PPM 6 12 10 1 19 0-10 0-20
Non-Latch
Disable Sensor
When removing a sensor from the transmitter while under power, the S5000 will enter a Sensor
Missing fault condition after the two minute Swap Delay period has expired (if enabled). If
Swap Delay is disabled, the transmitter will go into Sensor Missing fault immediately. If the system
is off at the time a sensor is removed, the transmitter will go into fault after its startup sequence.
This fault condition can be removed by disabling the affected second sensor position.
Disabling a sensor removes the fault and stops communications with the sensor, the sensor’s
reading on the display is removed, and the mA channel for that sensor position is set to 0 mA. By
default, the S5000 has the Sensor 2 position disabled. If at any time a sensor is connected to a
position that is disabled, the device will automatically enable that sensor position.
To disable the sensor after removal:
(1) Go to Sensor Setup menu.
(2) Touch button when Disable appears on screen.
(3) Select Sensor 1 or Sensor 2.
Current status (enabled/disabled) displays.
(4) Touch button to toggle to desired status.
(5) Select “Finished?”.
To remove a second sensor that is already installed, the sensor must first be removed.
The second sensor cannot be disabled while it is installed on the transmitter.
The device only allows a sensor to be disabled after the transmitter has gone into
Sensor Missing fault.
Only one sensor can be disabled at a time. The transmitter will not allow.
Alarm Relay Settings is the last Sensor Setup menu. “Finished?” will save changes and US
take the user back to Setup menus.
CAUTION!
The sensor must be calibrated after a sensor reset. There will be displayed four dashes where the
current reading should be.
All settings, including Alarm Set Point and Calibration Values, will be returned to factory defaults.
The sensor default values can be restored by resetting the sensor. During sensor reset, the analog
output displays the sensors current gas value (i.e. 0 % LEL = 4 mA) during the 120 s count down.
To reset sensor to factory default:
(1) Go to Sensor Setup menu.
(2) Go through menus selecting “Finished?” at the end of each option.
(3) Touch button when Reset Setting scrolls across screen.
(4) Touch button to change to yes.
(5) Touch button when “Finished?” is displayed to exit.
This is the last Sensor Setup menu. “Finished?” will take the user back to Setup Menus
and will save any changes made.
Relay 1 and Relay 2 can be configured for common and discrete modes via the device display
menu or X/S Connect app.
Common mode is the default relay mapping setting. In Common mode, Relay 1 is actuated by
Alarm 1 on either sensor, and Relay 2 is actuated by Alarm 2 on either sensor.
US
Energized/De-energized
(6) Relay 1 and Relay 2 will scroll across screen alternatively.
(7) Touch button when Relay 1 or Relay 2 is scrolling.
(8) Touch button to show the current setting (energized/de-energized).
(9) Touch button again will toggle the setting.
(10) Touch button when “Finished?” is displayed to exit from Relay setting.
Bluetooth® Security
The Bluetooth® connection is encrypted and secured with a unique six digit pin that must be
confirmed on the mobile device and acknowledged on the detector display. All of the previously
paired devices can be erased from the S5000 to provide additional security and control.
To reset all device pairings:
(1) Scroll to Settings.
(2) Select Instrument.
(3) Scroll and select Bluetooth®.
(4) Scroll and select “Reset All”.
(5) Select “Continue”.
NOTICE
Reset All will delete all paired device memory. All devices will have to re-initiate pairing at the
device.
Bluetooth® Tag ID
See section 4.3.5 to view Bluetooth® Tag ID.
4.2.7 Modbus Settings
The Modbus Setup menu consists of three configurable sub menus: Baud Rate, Format and
Address.
(1) Hold finger on button while menu selections scroll across screen.
(2) Remove finger when Setup menu appears.
(3) Touch button when Modbus Setup is displayed.
Modbus Baudrate
(1) Go to Modbus Setup menu. Baudrate option is the first submenu.
Baudrate will scroll across display followed by the current Baudrate setting.
(2) Touch the button to change to one of the following baudrate options:
a) 2400
b) 4800
c) 9600 US
d) 19K2
e) 38K4
f) 115K
(3) Touch button when “Finished?” is displayed to exit.
US
NOTICE
The transition to maintenance mode during the 2 minute Swap Delay window and sensor count-
down will not trigger the Fault Relay. The Fault Relay will only be triggered when the device enters
a fault condition.
As long as the setup menu is active, no password is required to enter again. If menu
was not active for 30 seconds, the password will be required again.
CAUTION!
Verify all sensor settings (calibration level and alarm values) after resetting transmitter. Sensors
may require recalibration to clear a fault after resetting transmitter.
The user will not be asked to exit with “Finished?” at the end of this menu.
The user will not be asked to exit with “Finished?” at the end of this menu.
US
WARNING!
Use zero gas when zeroing the S5000 transmitter if there is any possibility of background gas.
Otherwise, improper calibration could occur.
Always calibrate for the least sensitive gas or vapor (higher number category) expected to be
measured. Otherwise, instrument readings may be incorrect.
CAUTION!
Before attempting a calibration, power the unit at least one full hour.
Perform a calibration at initial start-up and at regular intervals to ensure a fully functional sensor.
The IR400 sensor has different calibration hardware, but is still included in CALKIT1.
The sensor guard and rubber boot should be removed after calibration. The sensor
guard actually makes the sensor more sensitive to gas in windy conditions and could
result in a false alarm.
P/N with
Gas Type Range Concentration Part Number
Calibration Kit
0-100 PPM 60 PPM 710882 710882-KIT1
Carbon
0-500 PPM 300 PPM 10027938 10027938-KIT1
Monoxide
0-1000 PPM 400 PPM 10028048 10028048-KIT1
0-10 PPM 5 PPM 710414 710414-KIT1
Hydrogen
0-50 PPM
Sulfide 40 PPM 10028062 10028062-KIT1
0-100 PPM
Oxygen 0-25 % 20.8 % 10028028 10028028-KIT1
0-100 % LEL 2.5 % Methane
5 % Methane (50 % LEL)
10028032 10028032-KIT1
0-100 % LEL 2.5 % Methane
Combustible 4 % Methane (57 % LEL)
(IR400 or
Cat Bead) 0-100 % LEL 0.6 % Propane
2.1 % Propane (29 % LEL)
10028035 10028035-KIT1
0-100 % LEL 0.6 % Propane
1.7 % Propane (35 % LEL)
Passive Cat
0-100 % LEL 2.5% Methane 1400155-M 1400150-M
Bead
Ampules Gas Ampules Gas
0-100 PPM 50 PPM 50004-13 1400255-5 50009-9 1400250-5
Passive MOS
0-50 PPM 25 PPM 5004-21 1400255-3 5009-16 1400250-3
0-20 PPM 10 PPM 50004-3 1400255-1 50009-10 1400250-1
5.3 Calibration Frequency for XCell Sensors with TruCal (H2S & CO only)
XCell sensors with TruCal actively self-monitor for operation and accuracy with fewer manual cali-
brations required in order to maintain the equivalent level of performance. When the sensor
detects a change in sensitivity outside what is typical for a six hour time interval, the sensor will
call for a calibration. The transmitter LED status indicators slowly pulse green, alerting the user
that a calibration is recommended to maintain optimum performance. There are no changes to the
analog output when TruCal calls for a manual calibration.
Actual TruCal sensor performance will depend on the application, background gas exposure, and
environment. To validate XCell sensors with TruCal, it is recommended that users follow their
regular calibration cycle and record the “as found” and “as left” values, tracking the percent adjust-
ment over time. Once a baseline is established, the calibration intervals can be extended until the
percent adjustment is greater than the expected accuracy of the sensor.
5.4 Calibration Types: Zero vs. Span
There are two types of calibration; zero and span calibration.
The zero calibration resets the baseline level reading to zero. If the target gas is suspected to be
occasionally present, it is best to also use a zero gas cylinder during the zero calibration. If the
target gas is not present in the atmosphere, an additional calibration cylinder is not required.
The span calibration is a process that involves first zeroing the baseline reading, then applying a
known amount of target gas so that the accuracy and precision of the transmitter can be adjusted
to the known value. The known concentration of gas is called a "Span Value" because it
represents the span or reach of the calibration curve from zero to that value.
US
To abort, press the EZ touch button or apply magnet over GM logo before applying
target gas.
In the event that a calibration cannot be completed, the user can acknowledge the FAIL
by pressing both EZ touch buttons simultaneously for 5 seconds. The unit will revert to
the settings of the last successful calibration.
WARNING!
The regulator used with the zero gas cylinders should not be the same as the regulator used for
the target gas. The regulator can be contaminated with the target gas over time, and thus raise
the detection baseline and make the sensor less sensitive to the target gas.
If there is no target gas in the atmosphere around the sensor, using a zero gas cylinder
is optional.
If a password is enabled, you will need to enter it to access the menu system.
NOTICE
If a password is enabled, you will not be able to proceed with the calibration without the password.
To abort, press the EZ touch button or apply magnet over GM logo before applying
target gas.
In the event that a calibration cannot be completed, the user can acknowledge the FAIL
by pressing both EZ touch button until “Reset” is displayed and then release. The device
will revert to the settings of the last successful calibration.
If it is suspected that gases are present, it will be necessary to purge the sensor envi-
ronment with zero air.
If a password is enabled, you will need to enter it to access the menu system.
(7) Select Sensor #1 or Sensor #2, and then select “Finished?” to start the calibration process.
(8) Once the screen displays Zero in Progress, turn on zero gas flow by turning knob on the US
regulator (if using).
(9) Wait while the device displays Zeroing.
(10) Once the Zero Calibration is complete, remove the tubing from the sensor guard inlet.
(11) The display will show Apply Gas. Attach tubing for calibration gas and turn on the regulator
Display will show “Calibration in Progress”.
“Calibration Complete Remove Gas” will show when span is complete.
If “PASS” displays, the procedure is complete.
If “FAIL” displays, the procedure was unsuccessful. See section 6.4.
(12) Remove the calibration equipment from the device, including the sensor guard and
calibration cap.
If a full calibration of the IR400 sensor is required, a larger calibration cap is used and
covers the entire sensor guard during calibration. The calibration cap must be removed
from the XIR sensor guard after completing the zeroing and/or spanning procedure.
If a password is enabled, you will need to enter it to access the menu system.
If the sensor is fitted with a sensor guard, calibration must be performed with the sensor
guard in place.
If it is suspected that gases are present, it will be necessary to purge the sensor envi-
ronment with zero air.
To calibrate sensor: US
(1) Attach a regulator to the zero air cylinder (if using) and the calibration cylinder.
(2) Attach the plastic calibration cup (P/N 1400152-1 or 1400154) to the bottom of the sensor if
not already installed.
(3) Attach the tubing to the plastic stem protruding through the calibration cup.
(4) Push the other end of the tubing over the zero cylinder regulator. Ensure the tubing
completely covers the gas outlet.
(5) Press EZ touch until Calibration is displayed, and release the touch.
(6) Turn on zero gas flow by turning knob on the regulator (if using).
(7) Unit will start flashing the sensor life. Tap the EZ Touch if you want to reset the sensor life.
(8) Wait while the device displays “Zero in Progress”.
(9) Once the Zero Calibration is complete, remove the tubing from the sensor guard inlet.
(10) The display will show “Apply Gas”. Attach tubing for calibration gas and turn on the regulator.
(11) Display will show “Calibration in Progress”.
(12) “Calibration Complete. Remove Gas” will show when span is complete.
(13) Turn off the calibration gas regulator. Remove the calibration equipment from the device,
including the sensor guard and calibration cap. Unit will return to normal operation.
If “Cal Fail” and fault code displayed, the procedure was unsuccessful. See section 6.4.
5.10 XCell Catalytic Bead Failsafe
Catalytic bead sensors require the presence of oxygen in order to sense combustible gas. In the
event of very large combustible gas leaks that exceed 100 % LEL, enough oxygen can be
displaced so that the sensor's response to gas is no longer proportional to the calibration profile.
The XCell catalytic bead has a locking fail safe mechanism that prevents the false reporting of a
safe condition while the % LEL concentration is still above 100 % LEL. When the gas
concentration exceeds 100 % LEL, the sensor will go into LOC over range.
To clear the LOC over rage, the user needs to acknowledge and calibrate the sensor. To
acknowledge LOC over range, place one finger over each of the EZ touch buttons and hold for
1 second before releasing. This will allow a recalibration of the sensor to clear the LOC condition.
CAUTION!
Ensure that the area has been cleared of gas before acknowledging the LOC Over Range and
recalibrating the sensor.
While in "Beam Block" fault, the sensor will not respond to the presence of gas.
(6) When cleaning is done and the objects are removed from the sensor window, the device
returns to normal operation. If water or isopropyl alcohol was used, allow the device to
operate for 15 minutes to completely dry before replacing the sensor guard continuing to
monitor for combustible gas.
(7) Replace the sensor guard.
US
When the cleaning process is complete, be sure to remove all objects from the light
path. It is recommended to check the sensor’s response to zero and calibration gas after
cleaning.
WARNING!
Do not place foreign objects in the sensor's analytical region (except per the “S5000 IR400
Cleaning Procedure” as described above); otherwise, the infrared beam can be partially blocked,
causing the sensor to generate false readings. All objects must be removed from the sensor's
analytical region for it to function properly.
WARNING!
Handle the sensor carefully; the electrochemical version is a sealed unit which contains a corro-
sive electrolyte.
Any leaked electrolyte that come in contact with skin, eyes or clothes can cause burns.
If any contact with the electrolyte does occur, immediately rinse with a large quantity of water. On
contact with the eyes, rinse thoroughly with water for 15 minutes and consult a doctor.
CAUTION!
Do not install a leaking sensor in the sensing head assembly. The leaking sensor must be
disposed of in accordance with local, state and federal laws.
There is no need to open the main enclosure. Simply unscrew the sensor assembly located on
the bottom of the main assembly.
WARNING!
XCell Digital Sensors marked for Class I hazardous (classified) locations and installed in a Class I
hazardous (classified) location only may be changed under power. To change, unscrew sensor at
least three full turns (but no more than four full turns from its fully engaged position), wait
10 seconds, and then remove the sensor completely. Failure to follow this warning can result in
the ignition of a hazardous atmosphere.
For XCell Digital Sensors marked Class II or III and installed in a Class II or III hazardous
location, the atmosphere must be free of dust or fibers and filings and the power removed from
the unit before the sensor cap can be removed from the housing. Failure to follow this warning
can result in the ignition of a hazardous atmosphere.
Identify the sensor assembly needed via the A-5K-SENS code on the interior sensor label and
obtain the appropriate sensor assembly; replace sensor assembly.
US
Alarm set points, span value, full scale limit and alarm direction will not change when
replacing a sensor with the same gas type.
Alarm set points, span value, full scale limit and alarm direction will change to the new
sensor’ settings when replacing a sensor with a different gas type.
The S5000 Gas Monitor is shipped with the Sensor Swap Delay enabled. IF enabled, the 4-20 mA
output signal goes to 3.5 mA and the FAULT relay will hold off a fault indication for 2 minutes. This
setting allows the operator to exchange sensor modules without a FAULT indication.
Tab. 33 Troubleshooting
US
7 Ordering Information
7.1 Replacement Parts
See Tab. 34 for replacement sensor kits. To obtain a replacement sensor, address the order or
inquiry to:
General Monitors
26776 Simpatica Circle
Lake Forest, CA 92630
or call, toll-free, 1-800-446-4872.
Inquiries can also be e-mailed to [email protected].
WARNING!
Use only genuine MSA replacement parts when performing any maintenance procedures
provided in this manual. Failure to do so may seriously impair sensor and gas monitoring
performance. Repair or alteration of the S5000, beyond the scope of these maintenance
instructions or by anyone other than authorized MSA service personnel, could cause the product
to fail to perform as designed and persons who rely on this product for their safety could sustain
serious personal injury or loss of life.
8 Appendix: Specifications
Digital XCell IR
Sensor Options
Toxics Oxygen Combustible IR400
Operating Range -40 °C to +60 °C -40 °C to +60 °C -40 °C to 60 °C -40 °C to +75° C
Storage -40 °C to +60 °C -40 °C to +60 °C -40 °C to 60 °C -50 °C to +85 °C
Zero Drift1 ≤ 1 % FS ≤ 0.2 % Vol < 5 % LEL/mo < 2 % per year
Span Drift 1
< 2 %/mo ≤ 0.2 % 5 % LEL
Noise < 2 display units < 2 display units 1 % LEL 1 % LEL
±3 % LEL at
≤ 50 % LEL
Repeatability 1
±10 % ≤ 0.3 % Vol ≤ 3 % LEL
±5 % LEL at
> 50 % LEL
1 ppm, CO
Resolution 0.1 % Vol 1 % LEL 1 % LEL
0.1 ppm, H2S
T90 < 25 s < 15 s < 30 s ≤3s
Humidity 10-95 % RH 10-95 % RH 5-100 % RH 10-95 % RH
Sensor Life 5 years 5 years 5 years > 5 years
Warranty 3 years 3 years 3 years 2 years
Enabling XCell - XCell - XCell -
IR
Technology non-consuming non-consuming GM cat bead
TruCal Yes No No No
SafeSwap Yes Yes Yes No
Remote Mount
100 m 100 m 100 m 1100 m
Distance
Power:
3.6 W 3.6 W 6.0 W 4.8 W
Single Sensor
Power:
4.3 W 4.3 W 8.4 W NA
Dual Sensing
Material Weight
7.30 lbs
316 Stainless Steel
7.92 lbs w/ relays
US
Digital Sensor
U.S. Division Hazardous Locations as defined by the NEC
Class I, Division 1, Groups A, B, C, D; T5
Class II, Division 1, Groups E, F, G; Class III; T5
US
Cautions/Warnings
• READ AND UNDERSTAND ALL INSTRUCTIONS, WARNINGS AND CAUTIONS PRIOR TO
INSTALLATION OF ANY COMPONENTS OF THIS SYSTEM
• CAUTION – OPEN CIRCUIT BEFORE REMOVING COVER
• SEALING REQUIREMENTS -
• Aluminum S5000 transmitter and Junction Box:
A SEAL SHALL BE INSTALLED WITHIN 18 in (450 mm) OF THE ENCLOSURE
• Stainless Steel S5000 transmitter and Junction Box:
SEAL NOT REQUIRED
• WARNING FOR TRANSMITTER AND JUNCTION BOX (CEMENTED OR NON-CEMENTED
COVERS) – DO NOT OPEN WHEN ENERGIZED OR WHEN AN EXPLOSIVE ATMO-
SPHERE IS PRESENT
• WARNING FOR TRANSMITTER AND JUNCTION BOX (CEMENTED OR NON-CEMENTED
COVERS) – LIVE PARTS BEHIND COVER, OPEN CIRCUIT BEFORE REMOVING COVER
– DO NOT CONTACT
• WARNING FOR TRANSMITTER AND JUNCTION BOX (CEMENTED OR NON-CEMENTED
COVERS) – POTENTIAL ELECTROSTATIC CHARGING HAZARD. CLEAN EQUIPMENT
ONLY WITH DAMP CLOTH. FAILURE TO FOLLOW THIS WARNING CAN RESULT IN
SEVERE PERSONAL INJURY OR LOSS OF LIFE.
• WARNING – NO INGRESS PROTECTIONS CLAIMS ARE MADE FOR COMBUSTIBLE GAS
DETECTION PERFORMANCE
• WARNING – FOR SAFETY REASONS THIS EQUIPMENT MUST BE OPERATED AND
SERVICED BY QUALIFIED PERSONNEL ONLY. DO NOT OPERATE THIS EQUIPMENT
UNTIL AFTER THE INSTRUCTION MANUAL IS READ AND UNDERSTOOD FOR PROPER
INSTALLATION AND OPERATION
• CAUTION: HIGH OFF-SCALE READING MAY INDICATE EXPLOSIVE CONCENTRATION
• WARNING – WIRING TO OR FROM THIS DEVICE, WHICH ENTERS OR LEAVES THE
SYSTEM ENCLOSURE, MUST UTILIZE WIRING METHODS SUITABLE FOR CLASS I,
DIVISION 2/ZONE 2 HAZARDOUS LOCATIONS, AS APPROPRIATE FOR THE INSTALLA-
TION
• WARNING – EXPLOSION HAZARD. DO NOT CONNECT OR DISCONNECT WHEN ENER-
GIZED
• CAUTION: USE WIRING SUITABLE TO AT LEAST 99 °C WHEN WIRING THE S5000 MAIN
TRANSMITTER OR TO AT LEAST 75 °C WHEN WIRING THE S5000 JUNCTION BOX
Additional Certification Information
US
As part of this Approval, it was verified that optional communication functions of this gas detection
instrument while operating at the maximum transaction rate do not adversely affect the gas
detection operation and functions of the instrument. This Approval, however, does not include or
imply Approval of the “SafeSwap” feature, communications protocol or functions provided by the
software of this instrument or of the communications apparatus or software connected to this
instrument.
Use of some accessories are not covered under the performance approval. Contact
General Monitors for details.
US