General Monitors S5000 Manual-En

Download as pdf or txt
Download as pdf or txt
You are on page 1of 63

Operating Manual

General Monitors S5000


Gas Monitor

Order No.: MANS5000/01

MSAsafety.com
WARNING!
Read this manual carefully before using the device. The device will perform as designed only if it is used
and maintained in accordance with the manufacturer's instructions. Otherwise, it could fail to perform as
designed and persons who rely on this device for their safety could sustain serious personal injury or
death.
The warranties made by GM with respect to the product are voided if the product is not installed, used
and serviced in accordance with the instructions in this manual. Please protect yourself and your
employees by following the instructions.
Please read and observe the WARNINGS and CAUTIONS inside. For any additional information relative
to use or repair, call +1-949-581-4464 during regular working hours.

26776 Simpatica Circle


Lake Forest, CA 92630 USA
Contents

1 Safety Regulations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
1.1 Correct Use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
1.2 Product Warranty . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6

2 Description. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
2.1 Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
2.2 No Tool Interface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
2.3 Bluetooth® Wireless Technology . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
2.4 Dual Sensing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
2.5 Retrofit Installations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
2.6 XCell Sensors Optimized for Fixed Gas Applications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
2.7 TruCal Sensing Technology for CO and H2S Electrochemical Sensors . . . . . . . . . . . . . . . . . . . . . 9
2.8 Safe Swap . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
2.9 Housing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
2.10 Component Overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10

3 Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
3.1 Installation Warnings - Read before Installation. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
3.2 Reviewing Shipment and Identifying Product Model . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
3.3 Product Installation Check List . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
3.4 Mounting. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
3.4.1 Sensor Mounting Location . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
3.4.2 Transmitter Mounting Location. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
3.4.3 Sensor Orientation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
3.4.4 Connecting Sensor to Transmitter Housing or Remote Junction Box . . . . . . . . . . . . . . . . . . . . . . 14
3.4.5 Integrated Mounting Points . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
3.4.6 Adjustable Pipe Mount . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
3.4.7 Duct Mount . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
3.4.8 Mounting with a Sunshield . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
3.5 Installing a Remote Sensor Junction Box. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
3.6 Electrical Power Connections. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
3.6.1 Electrical Warnings - Read before Connecting Power . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
3.6.2 Electrical Hardware Requirements. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
3.6.3 Power Load Requirements and Maximum Mounting Distances. . . . . . . . . . . . . . . . . . . . . . . . . . . 21 US
3.6.4 Instructions for Power and Analog Output . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
3.6.5 Relay Electrical and Power Connections . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23

4 Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
4.1 Startup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
4.1.1 Initial Startup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
4.1.2 Startup after Power Failure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24

General Monitors S5000 3


4.2 Settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
4.2.1 Instrument Settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
4.2.2 Sensor Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
4.2.3 Relay Settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32
4.2.4 HART Settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34
4.2.5 HART AO . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34
4.2.6 Bluetooth® . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34
4.2.7 Modbus Settings. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35
4.2.8 EZ Touch Button . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36
4.2.9 Swap Delay . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37
4.2.10 UI Password . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37
4.2.11 Transmitter Setting Reset. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38
4.3 Info Menu - Viewing Device Status . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38
4.3.1 View or Change System Time . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39
4.3.2 Last Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39
4.3.3 Non-Critical Fault Log. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39
4.3.4 Device Tag . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39
4.3.5 Bluetooth® FCC/IC ID . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 40
4.3.6 Sensor Life and Health Status (only displayed if XCell Sensor is connected) . . . . . . . . . . . . . . . . 40
4.4 Setting only configurable via Bluetooth®, Modbus, or HART . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 41
4.4.1 AO Custom Levels . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 41
4.4.2 Unit Device ID/Tag . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 41
4.4.3 Alarm Direction. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42
4.4.4 Alarm Enable/Disable. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42

5 Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42
5.1 Calibration Equipment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42
5.2 Calibration Frequency . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 43
5.3 Calibration Frequency for XCell Sensors with TruCal (H2S & CO only). . . . . . . . . . . . . . . . . . . . . 44
5.4 Calibration Types: Zero vs. Span . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 44
5.5 How to Zero Calibrate XCell and IR400 Sensors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 45
5.6 How to Calibrate XCell Sensors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 46
5.7 How to Calibrate an Oxygen XCell Sensor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 47
5.8 How to Calibrate an IR400 Sensor. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 47
5.9 How to Calibrate a Passive Sensor (Catalytic Bead or MOS) . . . . . . . . . . . . . . . . . . . . . . . . . . . . 47
5.10 XCell Catalytic Bead Failsafe . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 48
US
5.11 Calibration Confirmation. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 48

6 Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 48
6.1 IR400 Cleaning Procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 49
6.2 Replacing an XCell Sensor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 50
6.3 Replacing a Passive Sensor (Cat Bead or MOS). . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 51
6.4 Troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 51

7 Ordering Information. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 55
7.1 Replacement Parts. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 55

8 Appendix: Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 56

9 Appendix: General Certification Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 59

10 Appendix: HART Specific Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 62

General Monitors S5000 4


Safety Regulations

1 Safety Regulations
1.1 Correct Use
The S5000 Gas Monitor, hereafter also called device, is a gas monitor for measuring toxic and
combustible gases as well as oxygen. Using sensors, the device tests the ambient air and triggers
the alarm as soon as the gas exceeds a specific concentration level.

WARNING!
Read this manual carefully before servicing the device. The device will perform as designed only
if it is used and maintained in accordance with the manufacturer's instructions. Otherwise, it could
fail to perform as designed and persons who rely on this device for their safety could sustain
serious personal injury or death.

WARNING!
Do not use silicone-type lubricants in assembling the device and do not allow silicone vapors to
be drawn into the flow system while in operation. Silicone can desensitize the combustible gas
sensor, thereby giving erroneously low readings.
Use only genuine MSA replacement parts when performing any maintenance procedures on the
device. Substitution of components can seriously impair performance.
Failure to follow the above can result in serious personal injury or loss of life.

This device complies with Part 15 of the FCC Rules. Operation is subject to the
following two conditions:
• this device may not cause harmful interference, and
• this device must accept any interference received, including interference that may
cause undesired operation.
This equipment has been tested and found to comply with the limits for a Class A digital
device, pursuant to Part 15 of the FCC Rules. These limits are designed to provide
reasonable protection against harmful interference when the equipment is operated in
a commercial environment. This equipment generates, uses, and can radiate radio
frequency energy and, if not installed and used in accordance with the instruction
manual, may cause harmful interference to radio communications. Operation of this
equipment in a residential area is likely to cause harmful interference in which case the
user will be required to correct the interference at his own expense.

US
NOTICE
This is a Class A product in accordance with CISPR 22. In a domestic environment, this product
may cause radio interference, in which case the user may be required to take adequate measures.

NOTICE
The XCell sensor refers to the sensor portion of the Digital Sensor throughout this manual.

FCC Warning Statements


Changes or modifications not expressly approved by the manufacturer could void the user's
authority to operate the equipment.
Industry Canada (IC) Warning Statements
The installer of this radio equipment must ensure that the antenna is located or pointed such that
it does not emit RF field in excess of Health Canada limits for the general population; consult
Safety Code 6, obtainable from Health Canada's website www.hc-sc.gc.ca.

General Monitors S5000 5


Safety Regulations

1.2 Product Warranty


ITEM WARRANTY PERIOD
S5000 Gas Monitor MSA warrants that this product will be free from mechanical
defects and faulty workmanship for the period specified in this
table for each component, provided it is maintained and used in
accordance with MSA’s instructions and/or recommendations.
Warranty shall not exceed.
Main Transmitter Housing 2 years from date of shipment.
and PCBA Shall not exceed 2 years and 6 months from date of manufacture.
XCell Sensors 3 years from date of shipment.
Shall not exceed 3 years and 6 months from date of manufacture.
IR400 2 years from date of shipment.
Shall not exceed 2 years and 6 months from date of manufacture.
Passive Catalytic Bead 2 years from date of shipment.
Shall not exceed 2 years and 6 months from date of manufacture.
Passive MOS 2 years from date of shipment.
Shall not exceed 2 years and 6 months from date of manufacture.

This warranty does not cover filters, fuses, etc. Certain other accessories not specifically listed
here may have different warranty periods. This warranty is valid only if the product is maintained
and used in accordance with Seller's instructions and/or recommendations. The Seller shall be
released from all obligations under this warranty in the event repairs or modifications are made by
persons other than its own or authorized service personnel or if the warranty claim results from
physical abuse or misuse of the product. No agent, employee or representative of the Seller has
any authority to bind the Seller to any affirmation, representation or warranty concerning this
product. Seller makes no warranty concerning components or accessories not manufactured by
the Seller, but will pass on to the Purchaser all warranties of manufacturers of such components.
THIS WARRANTY IS IN LIEU OF ALL OTHER WARRANTIES, EXPRESSED, IMPLIED OR
STATUTORY, AND IS STRICTLY LIMITED TO THE TERMS HEREOF. SELLER
SPECIFICALLY DISCLAIMS ANY WARRANTY OF MERCHANTABILITY OR OF FITNESS
FOR A PARTICULAR PURPOSE.
Exclusive Remedy
It is expressly agreed that Purchaser's sole and exclusive remedy for breach of the above
warranty, for any tortious conduct of Seller, or for any other cause of action, shall be the
replacement at Seller's option, of any equipment or parts thereof, which after examination by US
Seller is proven to be defective. Replacement equipment and/or parts will be provided at no cost
to Purchaser, F.O.B. Seller's Plant. Failure of Seller to successfully replace any nonconforming
equipment or parts shall not cause the remedy established hereby to fail of its essential purpose.
Exclusion of Consequential Damage
Purchaser specifically understands and agrees that under no circumstances will seller be liable to
purchaser for economic, special, incidental or consequential damages or losses of any kind
whatsoever, including but not limited to, loss of anticipated profits and any other loss caused by
reason of non-operation of the goods. This exclusion is applicable to claims for breach of warranty,
tortuous conduct or any other cause of action against seller.

General Monitors S5000 6


Description

2 Description
2.1 Display
The S5000 utilizes a dot matrix LED display, capable of displaying four alphanumeric characters
at a time. The display will scroll words that exceed four letters. Most of the messages scroll twice
across the screen before moving onto next selection.

Fig. 1 S5000 Main Display


In addition to the red LED display, the S5000 uses six icons to indicate status. Green LED
indicates power supply status. A yellow triangle and red bell indicate fault and warning or alarm
conditions respectively. The Bluetooth® icon indicates that the Bluetooth® wireless technology is
enabled on the device. Yellow "1" and "2" icon indicate which sensor gas reading is being
displayed, or during configuration which sensor's options are being accessed.
US

General Monitors S5000 7


Description

2.2 No Tool Interface


The S5000does not require any tools or third party devices to change settings, reset alarms or
perform any maintenance operation. The EZ touch button works through the glass and does not
require opening the explosion proof enclosure. The EZ touch button works with bare fingers or
with gloved hands, so long as the gloves are not black. See section 4 for more information on
navigating the menu with the EZ touch button.

Fig. 2 Interface

The user menu can also be accessed using the round GM magnet on the General Moni-
tors logo.

2.3 Bluetooth® Wireless Technology


The S5000 comes by default with Bluetooth® communication. Using the X/S Connect App on an
appropriate smart phone or tablet, users are able to interface with the S5000 menu options in a
larger and more user friendly setting. The Bluetooth® communication can interact with the device
within a maximum transmission distance of 70 feet (21 m).

WARNING!
Bluetooth® Operation is dependent upon signal availability of the wireless service(s) necessary to
maintain the communication link. Loss of wireless signal will prevent communication of alarms and
other information to linked devices. Take appropriate precautions in the event a loss of wireless
signal occurs.

If ordered with Bluetooth®, the device will be shipped with Bluetooth® enabled. See
section 4 for instructions on disabling Bluetooth®. US
If the device was not ordered with Bluetooth® it cannot be upgraded.

The S5000 and user provided communication device will need to be paired. This requires both
devices to be in range and for a pairing sequence inputting a 6 digit pairing code. The instructions
will be displayed on both the S5000 and communication device.

There are communication devices capable of being used in classified areas. Please
contact your MSA representative for additional information.

General Monitors S5000 8


Description

2.4 Dual Sensing


The S5000 supports two Digital Sensors, or one IR400 point IR detector and one Digital Sensor
simultaneously with four wire connections. However, the device will only support one passive
sensor, either combustible catalytic bead or metal oxide semiconductor (MOS) sensors, based on
the ATO configuration. Passive catalytic bead uses three wires, passive MOS uses four wires.
The S5000 Gas Monitor generates two independent analog outputs; one for each sensor
connected to the transmitter. The analog output associated with Sensor 1 also has the digital
HART (Highway Addressable Remote Transducer) communication superimposed on the analog
signal. If two sensors are connected, the digital HART communication carries information for both
sensors.
2.5 Retrofit Installations
The S5000 has the conduit entries located in the exact same orientation and distance from the
wall and the mounting holes for attaching to a wall are identical to the S4000 Series detectors.
Users will be able to re-use the existing wiring for the remotely mounted sensors.
2.6 XCell Sensors Optimized for Fixed Gas Applications
XCell toxic and combustible catalytic bead sensors are developed and manufactured by MSA to
be the most shelf-stable and longest lasting sensors available. The XCell non-consuming reaction
chemistry is tried-and-true, with over 600,000 used in MSA’s portable gas detectors. Now
optimized for fixed gas applications, the XCell sensor platform is available in the S5000 and
provides multiple benefits, including a standard 3-year warranty on all XCell sensors.
One important optimization for fixed gas was incorporating the GM catalytic bead into the XCell
sensor. The GM cat bead is the largest in the industry, providing market leading expected life time
and stability.
By using a non-consuming reaction chemistry, the XCell Oxygen sensor is expected to last well
over 3 years and can be safely stored on the shelf for at least 1 year without sensor performance
degradation.
2.7 TruCal Sensing Technology for CO and H2S Electrochemical Sensors
Using patented pulse check technology and proprietary Adaptive Environmental Compensation
(AEC) algorithms, XCell sensors with TruCal verify operation.
Automated pulse checks are performed every six hours to verify operation. An electrical pulse
stimulates the XCell sensor similar to having actual calibration gas applied, providing a snapshot
of the sensor’s sensitivity at the time of the pulse. Using this sensitivity snapshot, the sensor can
diagnose sensor failures like electrode poisoning, electrolyte leaking, or electrical connectivity
issues. The sensitivity snapshots are mapped over time to provide an overall Life & Health status
as either “Good” or “Fair” if the sensor is approaching end of life.
The proprietary Adaptive Environmental Compensation (AEC) algorithms use the sensitivity
snapshots provided by the pulse check to adjust sensor output, compensating for environmental
impacts on sensor accuracy. If the AEC adjustment is greater than expected based on typical
US
environmental impact variations, the corresponding sensor channel LED will slow flash GREEN,
indicating that the sensor should be calibrated to reset the AEC cycle. The result is a sensor that
actively self-monitors for operation and accuracy, with far fewer manual calibrations.
Actual TruCal sensor performance will depend on the application, background gas exposure, and
environment. To validate XCell sensors with TruCal, it is recommended that users follow their
regular calibration cycle and record the “as found” and “as left” values. This data can be used to
extend calibrations depending on the required specification of the application.
2.8 Safe Swap
The S5000 comes with patented SafeSwap technology, which allows users to change or replace
XCell sensors without needing to power down the instrument. Swap delay is enabled on the
S5000 by default; a feature that gives users a 2 minute window to change sensors without
triggering a fault condition. SafeSwap and Swap Delay are only applicable for XCell sensors. For
more information on SafeSwap and Swap Delay, see section 6.

General Monitors S5000 9


Description

2.9 Housing
The S5000 comes in 316 Stainless Steel for the highest corrosion resistance. All housings have
¾” NPT conduit entries. Custom tags are available and easily attach to an integral ring.
2.10 Component Overview

Fig. 3 Exploded View


US

General Monitors S5000 10


Installation

3 Installation
3.1 Installation Warnings - Read before Installation
WARNING!
S5000 Digital Sensors labeled for Class I Hazardous (Classified) locations and installed in a
Class I Hazardous (Classified) location only may be changed under power. To change, unscrew
sensor at least three full turns (but no more than four full turns from its fully engaged position),
wait 10 seconds, and then remove the sensor completely. Failure to follow this warning can
result in the ignition of a hazardous atmosphere.
Do not paint the device. Avoid painting in areas where the S5000 and remote sensor junction box
are located. If painting is required in an area where an S5000 or remote sensor has been installed,
exercise caution to ensure paint is not deposited on the sensor inlet fitting. Paint solvents can also
cause an alarm condition to occur or potentially poison electrochemical sensors.
Protect the device from extreme vibration.
Do not mount the sensing head in direct sunlight without a sunshield (Part Number 10178977,
IR400 Sunshield Part Number 10178977).
The S4000 IR400 Gas Monitor contains no user- or field-serviceable parts and must be returned
to the factory for repair. Any attempt to open the monitor will damage the unit and void the
warranty.

CAUTION!
When installing the IR400 sensor, under no circumstances should a pry-bar be applied to the two
legs that support the unit's reflectors during installation or removal of the sensor. Applying force
to the legs can permanently damage the IR400 sensor.

3.2 Reviewing Shipment and Identifying Product Model


To determine the sensor type and options, check the shipping carton.

US

Fig. 4 Shipping Label


Digital and IR400 sensors are not shipped attached to the S5000. The IR400 is a one piece
sensor, but all other Digital Sensors are comprised of two parts; the sensor body and the sensor
puck. Digital Sensors are shipped unattached from the S5000 housing and must be attached.
Check the sensor details before attaching to the S5000 housing. The sensor details are listed on
the inside of the sensor puck. Unscrew the sensor puck from the sensor body and check the label
on the inside for gas type, range, configuration ordered, serial number, and firmware revision
number.
Passive sensors are shipped attached to the housing.

General Monitors S5000 11


Installation

3.3 Product Installation Check List


Before Installation
• Review national electrical codes
• Review local procedural and building codes
• Determine optimum transmitter placement
• Determine wire requirements
• Determine mounting hardware requirements
Mounting
• Attach appropriate sensor to housing or junction box
• Mount transmitter or junction box using appropriate mounting hardware
• Confirm free air flow around the sensor
3.4 Mounting
3.4.1 Sensor Mounting Location
The best location for the transmitter and the sensor may not be the same location. Sensors should
be placed in a location where a gas leak is most likely to be detected. When the best sensor
placement would not allow the transmitter display to be easily viewed or accessed, a remote
junction box can be used to mount the sensor remotely from the transmitter, allowing both to be
installed in the optimum location.
Two main factors should be considered when choosing a sensor location. The first is the density
of the target gas relative to the air. Gases, such as propane, that are heavier than air should be
placed near ground level while gases that are lighter than air should be placed above potential
leak sources.
Optimum sensor placement will depend on the surrounding processing equipment, such as pipes,
valves, or turbines. MSA offers a gas and flame mapping service that systematically evaluates
potential sources of leaks and recommends detector quantity and placement to create the most
effective detection system.
3.4.2 Transmitter Mounting Location
The transmitter display should be mounted so that the screen is visible and easily accessed after
installation. The electronics assembly inside the enclosure have one orientation inside the
cylindrical housing. Take care to position the conduit entries and display so that the display is read
in the correct orientation.

US

General Monitors S5000 12


Installation

3.4.3 Sensor Orientation


Sensor orientation will depend on the sensor type. If mounting an IR400 point IR sensor, whether
locally on the transmitter or via remote junction box, the sensor should be mounted horizontally.
If the IR400 sensor is not mounted horizontally, the sensor will be prone to more frequent beam
blocking issues due to accumulated dust and condensation on the surface of the IR400 sensor.
Fig. 5 shows the correct and incorrect mounting orientations for the IR400.

Fig. 5 Correct and Incorrect Mounting Orientations for IR400 Point IR Sensor
All other sensors, including electrochemical, catalytic bead, oxygen, passive catalytic bead, and
passive MOS should be mounted vertically with the gas inlet pointed downward. If the sensor is
not mounted with the gas inlet facing down, it is more likely to become clogged with particulate
matter or liquids. Fig. 6 shows the correct and incorrect mounting orientation for XCell sensors.
Passive catalytic bead and MOS sensors come already installed on the transmitter housing.

US

Fig. 6 Correct and Incorrect Mounting Orientation for XCell Sensors

General Monitors S5000 13


Installation

3.4.4 Connecting Sensor to Transmitter Housing or Remote Junction Box


Digital and IR400 sensors are not shipped attached to the main enclosure. The main sensor input
is provided via a four-terminal connection that provides a digital interface for all sensor modules.
Up to two sensors (excluding passive sensors) can be connected to a single transmitter with two
analog outputs capable of representing the readings of the individual sensors.
If planning to install the sensor attached to the transmitter or remote junction box, connect the
sensor before selecting the final mounting location so that the dimensions of the sensor can be
considered.
Passive sensors are shipped already attached and electrically wired to the device. Only one
passive sensor can be used on a single S5000, and they are not interchangeable with other
passive sensors or Digital Sensors. Consider the sensor dimensions when choosing a
mounting location for the transmitter or junction box.
To connect the sensor:
(1) Loosen the set screw located on the lid using a 1.5 mm Allen wrench.
(2) Turn the lid counter-clockwise to remove.
(3) Pull out display module to expose terminal connections.

Fig. 7 Terminal Connections


US
(4) Route the cable from the sensor through a conduit entry hole in the enclosure so that the
sensor is oriented in the correct position (see section 3.6 for details).
(Repeat to attach a second sensor to the S5000 transmitter).
(5) Connect the sensor to the "Sensor 1" position on the electronics assembly.
a) If using a second sensor, connect it to the second sensor position.

CAUTION!
If only using one sensor and it is connected to “Sensor 2” position, the S5000 will enter Sensor
Missing fault.

General Monitors S5000 14


Installation

Fig. 8 Connecting Sensor to the Stack and Grounding Terminal


(6) Verify the sensor connector is firmly seated on the terminal board.
(7) Attach the sensor's ground to either of the grounding screws inside the S5000 housing.
(8) Replace the display module. Push firmly on the board stack.

NOTICE
Ensure that the electronics assembly is fully engaged in the mounting holes. If not fully seated,
the user interface button may not function properly.

(9) Replace the cover by turning clockwise.


(10) Tighten the set screw located on the lid using a 1.5 mm Allen wrench.

US

General Monitors S5000 15


Installation

3.4.5 Integrated Mounting Points


The S5000 transmitter can be surface mounted without any additional brackets using the
integrated mounting tabs.

Fig. 9 Integral Mounting Tabs (not compatible with S4000 IR400 sensors)
3.4.6 Adjustable Pipe Mount
A Universal Pipe Mount Kit (Part Number 10176946) can be used to mount the S5000 on pipes
ranging from 20-150 mm in diameter. Two brackets are mounted over the top of the integrated
mounting tabs and fitted with an adjustable pipe band (not included).

US

Fig. 10 Adjustable Pipe Mount

General Monitors S5000 16


Installation

3.4.7 Duct Mount


Duct mount kits are available for monitoring atmosphere inside flat or round ducts. Round duct
mount kits are available for small ducts 12-20" in diameter (Part Number 10179323) and large
ducts 20-40" in diameter (Part Number 10179324). The flat duct mount (Part Number 10179322)
is universal for flat ducts.

NOTICE
Consider the sensor type before choosing a duct mount location. IR400 sensors should be
mounted horizontally and all other sensors should be mounted vertically.

NOTICE
Air flow in the duct must be zero to ensure proper calibration.

Fig. 11 Flat Duct Mount

US

Fig. 12 Round Duct Mount

General Monitors S5000 17


Installation

3.4.8 Mounting with a Sunshield


A sunshield is recommended to protect the S5000 from direct sun light (Part Number 10180254).
The sunshield can be used in any of the mounting configurations.

Fig. 13 Sunshield with Surface Mount

Fig. 14 Sunshield with Universal Pipe Mount

US

General Monitors S5000 18


Installation

3.5 Installing a Remote Sensor Junction Box


Sensors mounted remotely must use the S5000 junction box. The junction box housing is the
same construction as the transmitter. The mounting options and instructions for connecting the
sensor are the same for sensors connected directly to the S5000 transmitter housing. The junction
box is available in 316 Stainless Steel.
Sensors can be remote mounted up to 100 m from the transmitter housing, as long as the S5000
transmitter is mounted within maximum distance from the power supply, as indicated in Tab. 17.
The junction box does not have an illuminated display and has two connectors for attaching a
single sensor input and an output connects to the transmitter. A 16 AWG 4 element cable with a
braided shield should be used for the electrical connection between the junction box and the
S5000 transmitter. Specific cable recommendations are available upon request.

Fig. 15 Junction Box

US

Fig. 16 Junction Box Electrical Connections

General Monitors S5000 19


Installation

3.6 Electrical Power Connections


3.6.1 Electrical Warnings - Read before Connecting Power
WARNING!
Before wiring the S5000 transmitter, disconnect the power source supplying the transmitter and
ensure no hazardous atmosphere present; otherwise, electrical shock or ignition of hazardous
gases could occur.
• Install wiring in accordance with the electrical code of the country in use, the local authority
having jurisdiction and these installation instructions, as applicable.
• Do not make any connections to the S5000 main board or junction box input, output, and relay
connections while under power. Failure to do so could lead to electrical shock or ignition of a
hazardous atmosphere.
• Ensure that water and dirt are not able to enter the unit via the wire or conduit. If the unit is
installed in a location known to be wet or damp, it is good practice to loop or bend the entry
into the unit that prevents water incursion.
• The internal grounding terminal (located on the interior metal board stack plate) must be used
for equipment grounding. The external grounding terminal is only to be used as a supplemental
bonding connection where local authorities permit or require such a connection.
• S5000 Digital Sensors marked for Class I hazardous (classified) locations and installed in a
Class I hazardous (classified) location may be changed under power. To change, unscrew
three full turns from its fully engaged position, wait 10 seconds, and then remove the cap
completely. Failure to follow this warning can result in the ignition of a hazardous atmosphere.
• For S5000 Digital Sensors marked for Class II or III and installed in a Class I or III hazardous
location atmosphere must be free of dust or fibers filings and the power removed from the unit
before the sensor cap can be removed from the housing. Failure to follow this warning can
result in the ignition of a hazardous atmosphere.
3.6.2 Electrical Hardware Requirements
Braided shielded, twisted pair, instrument quality wire or cable should be used to minimize the
possibility of noise interference and contact with other voltages. Selection of shielded cable must
comply with local requirements.
Conduit, in addition to braided shielded wire, may also be needed in areas where a large amount
of electrical noise is expected. All cable shields should be terminated to earth ground at one end
only.
The S5000 has a four-wire power terminal, one four-wire communication terminal, and three four-
wire sensor terminals. Relays can be added as an option. Terminals for power and relays can take
wires up to 12 AWG while all other terminals take wires up to 14 AWG. Four conductors are also US
required for the S5000 remote junction boxes.
Incoming power and signal cables should be a braided shield cable such as Alpha Wire 3248 or
equivalent. The braided shield must be terminated to the instrument housing with a 360 degree
connection to earth ground as shown in Fig. 37, or alternatively, the earth ground at the user’s
power source location.
An external Class 2 power supply is required to supply 12-30 VDC to the S5000. Incoming power
and signal cables should be a braided shield cable such as Alpha Wire 3248 or equivalent.

General Monitors S5000 20


Installation

3.6.3 Power Load Requirements and Maximum Mounting Distances


Consider future needs when selecting cable size and power supply. The maximum distance
between the S5000 transmitter and the power supply depends on the sensor configuration
(sensing technology and one or two sensors), wire gauge, and the power supply voltage. The
table below outlines the maximum transmitter mounting distances. First determine if the sensor(s)
will be locally or remotely mounted. Then choose sensor type(s). The corresponding nominal
maximum power and mounting distances by wire gauge are shown.

Nominal Max Distances (ft)


Average Maximum
Max
Sensor Configuration Current Current
Power3 16 AWG 14 AWG 12 AWG
Draw Draw
(W)
Passive CB 220 mA 250 mA 6.0 2030 3230 5130
Passive MOS 180 mA 450 mA 10.8 1130 1800 2850
IR400 340 mA 370 mA 8.9 1370 2180 3470
Digital CB 220 mA 250 mA 6.0 2030 3230 5130
Digital Toxic 140 mA 150 mA 3.6 3380 5390 8560
IR400 & Digital CB 450 mA 490 mA 11.8 1040 1650 2620
IR400 & Digital Toxic 370 mA 400 mA 9.6 1270 2020 3210
Dual Digital Toxic 170 mA 180 mA 4.3 2820 4490 7130
Dual Digital CB 320 mA 350 mA 8.4 1450 2310 3670
Digital CB & Toxic 250 mA 280 mA 6.7 1810 2890 4580

Tab. 17 Power Load Requirements and Maximum Mounting Distances for 24 V Power Supply1,2

1 Values listed are only for one transmitter and the indicated sensor options. The total number
of transmitters needs to be considered when specifying a power supply
2
Assumes transmitter was ordered with relays
3 When sizing a system's 24 V supply, a 1 A inrush current with a 1 ms duration should be
considered for each device on the power supply
3.6.4 Instructions for Power and Analog Output
CAUTION!
Read all electrical warnings and wiring requirements before connecting power to the S5000.
Connector for HART analog output and power are provided to increase ease of connecting power.
Connect power and remote sensor cable shields to shield terminals on the main PC board. US
Provide shield terminations inside he sensor housing as indicated on the installation outline
drawing.
(1) Remove the cover by turning counter-clockwise.
(2) Remove the display module to expose the wiring terminations and sensor connections.
(3) Remove the 5.08 mm pitch connector for power supply.
The power connector is larger than other 3.81 mm pitch connectors.
(4) Use a small flat head screw driver to open wire entries on connector.
(5) Insert wires to connector so that when installed each wire is in the correct location.
1. +DC
2. -DC
3. mA1 - analog output of sensor 1
4. mA2 - analog output of sensor 2
(6) Tighten screws on connector and tug gently on wires to ensure they are secure.
(7) Attach the connector to the board stack.

General Monitors S5000 21


Installation

(8) Make sure the appropriate wires are in the correct terminals.
(9) Remove enough of the wire housing to expose the 3-4 inches of the cable shielding, but do
not expose so much that it goes beyond the cable entry.
(10) Attach the cable, shielding exposed, to the grounding point.
(11) Replace the display module. Push firmly on the board stack where indicated.

Fig. 18 Connecting Power and Grounding Cable

CAUTION!
Ensure that the electronics assembly is fully engaged in the mounting holes. If not fully seated,
the touch interface performance can be negatively affected.
Care must be taken to insure the S5000 inside glass surface glass is free of smudges/dirt and
grease. Dirt and grease can interfere with the touch interface of the display. US
NOTICE
Avoid pressing on the left and right areas where the LEDs are located. Pressing directly on the
display will damage the display and will void the warranty. Replace the S5000 cover by turning
clockwise. Be sure to align threads to avoid cross-threading.

S5000 Installation Outline Drawings

Model Document No.


S5000 324102

Tab. 19 Installation Outline Drawings

General Monitors S5000 22


Installation

3.6.5 Relay Electrical and Power Connections


Relay Board Stack Overview
The S5000 can be purchased with three relays. Two of the relays can be configured for either de-
energized (default) or energized and latching or non-latching. The third relay is a dedicated fault
relay.
All electrical connections to internal relays can be made directly on the PC board. The board is
labeled for Normally Open and Normally Closed de-energized state.

Fig. 20 PC Board with Relays


Relay Specifications

Temperature Range -40 to +85 °C (-40 to 185 °F)


Relays SPDT (Single Pole Double Throw)
Fault Normally Energized
Warning Configurable US
Alarm Configurable
Relay Rating
5A, 100K Cycles
125 or 250 VAC (Resistive)
1.6 HP @ 250VAC
30 VDC (Resistive) 5A, 100K cycles

Tab. 21 Relay Specifications


If using AC power, the relay wires should not be run within the same conduit or cable tray as the
DC power supplied to the S5000 or the S5000 junction box. A separate wire entry on the device
should be used for AC power connected to the relays. The S5000 is built with an additional wire
entry to allow this.

Exceeding the volt-amp rating of the relay can cause damage to the switching contacts.

General Monitors S5000 23


Operation

Relay Connections to Inductive Loads


If connecting the relays to motors, fluorescent lighting, or other inductive loads, it is necessary to
suppress any sparks or inductive feedback that may occur at the relay contact. These effects may
render the unit inoperative.
One way to reduce these effects is to install a Quencharc® (Part Number 630413) across the load
being switched.
Fault Relay Wiring and Configurations
The Fault relay state in non-fault operating condition is Energized and terminal connections are
supplied for Normally Closed and Normally Open.
The energized fault relay setting provides an electrical path for fail-safe relay operation. In the
event of any failure, including loss of power, the relay will change to the de-energized state to indi-
cate a fault condition.
The Fault relay state cannot be reconfigured.
Relay Energy State and Terminal Connections
The S5000 relay states are labeled for the default De-Energized state. The alarm/warning relay
energy state can be changed on the device, which will exchange the normally open and normally
closed terminals. The preferred relay energy state should be determined before making connec-
tions. Tab. 22 shows the terminal connections by energy state and is applicable to both relay 1
and relay 2.

Energy State NC NO
De-Energized (default) Normally Closed Normally Open
Energized Normally Open Normally Closed

Tab. 22 Relay Terminal Connections by Energy State


4 Operation
The S5000 Series gas monitor is factory calibrated and shipped with the most common default
options to minimize set up effort. Any of the default settings can be changed to meet the user’s
individual needs via the EZ touch button on the display of the S5000 Controller application
installed on the mobile device if equipped with Bluetooth®.
4.1 Startup
4.1.1 Initial Startup
The first time the S5000 is powered on, the following will appear on the display:
• S5000
• Software Version No.
• Sensor Warm-up US
• Wait
The S5000 will remain in Start-up mode in which the fault relay is de-energized and the analog
output is 3.5 mA by default. The time the S5000 stays in Start-up mode depends on the sensor.
Oxygen and Carbon Monoxide sensors require a 30 minute warm-up time before being fully
functional. The device will be in fault during the 30 minute start-up time. All other sensor start-up
times vary, the unit will be in fault for the first 2 minutes of the start-up time and the analog output
signal will be at the maintenance level (3.5 mA default).
A full calibration is recommended after one hour of a sensor being installed and acclimated to the
environmental conditions. See section 5 for calibration details.
4.1.2 Startup after Power Failure
If the S5000 loses power, all of the settings are saved to the internal memory. When power is
restored to the device, it will go back to the same settings as before the loss of power. To check
the settings, go through the menu or view on the X/S Connect App.

General Monitors S5000 24


Operation

4.2 Settings
The S5000 is a tool free transmitter. The infrared EZ touch button on the face of the display can
be used to navigate through the menu structure. The button is designed for use with fingers with
a “press” and “release” action, and works best without gloves.
The button works the same as the magnet does with the S4000 menu.
Changing a value
(1) Press and hold EZ Touch button.
(2) Wait for relevant menu to scroll (each menu scrolls twice).
(3) Release to enter menu while it is scrolling.
(4) Use Press and Release function to change values.
The EZ touch interface can be disabled, but will require a password. Menu settings can also be
activated using a magnet on the General Monitors logo.
Values that are changed in the menu are saved after main setup loop “Finished”, except for sensor
range selection under sensor setup. Each menu ends with "Finished?" scrolling twice on the
display. If the button is not touched during "Finished?", the menu will begin at the start again and
scroll through the options and values again. The new values that were entered will be displayed
on the first pass.
When user exits a menu by touching "Finished?" and there isn't a second sub menu, the previous
menu will be displayed starting at the menu that had just been used.

Some instrument settings are only configurable through Bluetooth® X/S Connect App,
Modbus, or HART. See section 4.4.

4.2.1 Instrument Settings


The following settings are saved to the device memory and will not change if the sensor type is
changed.
(1) Scroll to Settings.
(2) Select Instrument.
(3) Select to enter the menu.

US

General Monitors S5000 25


Operation

Setting Menu 1 Default Range of options


Sensor 1/Sensor 2
Disable Enabled Enable/Disable
Gas Label and Unit see Tab. 24 see Tab. 24
Range see Tab. 24 see Tab. 24
Sensor Setup Cal Level see Tab. 24 see Tab. 24
Warning Level see Tab. 24 see Tab. 24
Alarm Level see Tab. 24 see Tab. 24
Alarm Action see Tab. 24 see Tab. 24
Sensor Reset No Yes/No
Relay Zone/Mapping Common Common/Discrete
Relay Settings Relay 1/Relay 2 Relay 1/Relay 2
Energized/De-energized De-energized Energized/De-energized
HART Enabled/Disabled Enabled Enable/Disable
HART AO Analog Output Enabled 1.25 or 3.5
Bluetooth® Enable/Disable Enabled Enable/Disable
Baud Rate 19200 2400-115200 BPS
Modbus Format 8-n-1
Address 1 1-247
EZ Touch Enable/Disable Enabled
Swap Delay Enabled Enabled Enable/Disable
Enabled/Disabled Disabled Enable/Disable
UI Password
New Password No Yes/No
Transmitter Reset No No Yes/No

Tab. 23 Default Device Settings

US

General Monitors S5000 26


Operation

Alarm Cal Cal Cal


Range Unit Warn Alarm Alarm Alarm Range Range
Gas (Code) Action Level Level Level
Default3 Default Default Default Min Max Min max
Default Default Min Max
Carbon Monoxide Increasing
0-100 PPM 10 30 60 10 1000 0-10 0-1000 5 FS1
(D10) Non-Latch
Carbon Monoxide Increasing
0-500 PPM 50 150 300 10 1000 0-10 0-1000 5 FS1
(D11) Non-Latch
Carbon Monoxide Increasing
0-1000 PPM 100 300 400 10 1000 0-10 0-1000 5 FS1
(D12) Non-Latch
Carbon Monoxide Increasing
0-100 PPM 10 30 1000 0-10 0-1000 5 FS1
H2 Resistant (D14) Non-Latch
Catalytic Bead Increasing
0-100 % LEL 10 30 50 5 60 0-202 0-100 10 100
5% Methane (D60) Non-Latch
Catalytic Bead
Increasing
4.4% Methane 0-100 % LEL 10 30 57 5 60 0-202 0-100 10 100
Non-Latch
(D65)
Catalytic Bead
Increasing
2.1% Propane 0-100 % LEL 10 30 29 5 60 0-202 0-100 10 100
Non-Latch
(D61)
Catalytic Bead
Increasing
1.7% Propane 0-100 % LEL 10 30 36 5 60 0-202 0-100 10 100
Non-Latch
(D66)
Hydrogen Sulfide Increasing
0-10.0 PPM 1.0 3.0 5.0 1.0 100 0-10 0-100 5 FS1
(20) Non-Latch
Hydrogen Sulfide Increasing
0-50.0 PPM 5.0 15.0 40.0 1.0 100 0-10 0-100 5 FS1
(21) Non-Latch
Hydrogen Sulfide Increasing
0-100.0 PPM 10.0 30.0 40.0 1.0 100 0-10 0-100 5 FS1
(22) Non-Latch
Decreasing
Oxygen (16) 0-25.0 %VOL 19.5 18.0 20.8 15.0 25.0 5.0-25 5.0-25 15.0 25.0
Non-Latch
IR400 Increasing
0-100 % LEL 30 60 50 5 60 0-50 0-100
5% Methane (R31) Non-Latch
IR400
Increasing
4.4% Methane 0-100 % LEL 30 60 50 5 60 0-50 0-100
Non-Latch
(R43)
IR400
Increasing
2.1% Propane 0-100 % LEL 30 60 50 5 60 0-50 0-100
Non-Latch
(R32)
IR400
US
Increasing
1.7% Propane 0-100 % LEL 30 60 50 5 60 0-50 0-100
Non-Latch
(R44)

Tab. 24 Digital and IR400 Sensor Default Settings


1
FS = Full Scale range
2
The maximum range value on catalytic bead cannot be set below 20%.

General Monitors S5000 27


Operation

Alarm Cal
Range Unit Warn Alarm Alarm Alarm Range Range
Gas (Code) Action Level
Default3 Default Default Default Min Max Min max
Default Default
Increasing
Passive CB (C01) 0-100 % LEL 20 60 50 5 60 0-90 0-100
Non-Latch
Increasing
Passive MOS (M04) 0-100 PPM 30 60 50 5 95 0-50 0-100
Non-Latch
Increasing
Passive MOS (M05) 0-50 PPM 15 30 25 3 48 0-25 0-00
Non-Latch
Increasing
Passive MOS (M06) 0-20 PPM 6 12 10 1 19 0-10 0-20
Non-Latch

Tab. 25 Passive Sensor Default Settings

Display Warn Warn Alarm Alarm


Gas (Code) Range
Resolution1 Default Default AO2 Default Default AO2
Carbon Monoxide (D10) 0-100 1 10 5.6 30 8.8
Carbon Monoxide (D11) 0-500 1 50 5.6 150 8.8
Carbon Monoxide (D12) 0-1000 1 100 5.6 300 8.8
Carbon Monoxide
0-100 1 10 5.6 30 4
H2 Resistant (D14)
Catalytic Bead
0-100 1 10 5.6 30 8.8
5% Methane (D60)
Catalytic Bead
0-100 1 10 5.6 30 8.8
4.4% Methane (D65)
Catalytic Bead
0-100 1 10 5.6 30 8.8
2.1% Propane (D61)
Catalytic Bead
0-100 1 10 5.6 30 8.8
1.7% Propane (D66)
Hydrogen Sulfide (D20) 0-10.0 0.1 1 4.16 3 4.48
Hydrogen Sulfide (D21) 0-50.0 0.1 5 4.16 15 4.48
Hydrogen Sulfide (D22) 0-100 0.1 10 4.16 30 4.48
Oxygen (D16) 0-25.0 0.1 19.5 5.248 18 5.152
IR400
0-100 1 10 5.6 20 7.2
5% Methane (AA)
IR400
0-100 1 10 5.6 20 7.2 US
4.4% Methane (AC)
IR400
0-100 1 10 5.6 20 7.2
2.1% Propane (AB)
IR400
0-100 1 10 5.6 20 7.2
1.7% Propane (AD)

Tab. 26 Digital and IR400 Sensor Default Alarm Analog Outputs


1
Display resolution is not a configurable option
2
Analog Output for default range and alarms

General Monitors S5000 28


Operation

Display Warn Warn Alarm Alarm


Gas (Code) Range
Resolution1 Default Default AO2 Default Default AO2
Passive CB (C01) 0-100 20 8.8 60 13.6
Passive MOS (M04) 0-100 30 8.8 60 13.6
Passive MOS (M05) 0-50 15 8.8 30 13.6
Passive MOS (M06) 0-20 6 8.8 12 13.6

Tab. 27 Passive Sensor Default Alarm Analog Outputs


4.2.2 Sensor Setup
Configure gas unit, range, calibration level (i.e. span value), warning and alarm levels and whether
they are latching or non-latching.
The Sensor Setup menu will go through each sub menu before exiting Sensor Setup.
(1) Hold finger on button while menu selections scroll across screen.
(2) Remove finger when Setup menu appears.
(3) Touch button when Sensor Setup appears (first option).
(4) The following menus are available under Sensor Setup.
When "Finished?" is selected under each of the following menus, the next menu will start.
a) Sensor 1/Sensor 2
b) Disable
c) Gas Label & Unit
d) Range
e) Cal Level (Span Value)
f) Warning Level
g) Warning Action
h) Alarm Level
i) Alarm Action
j) Reset
(5) Select "Finished?" after the Reset menu to go back to the main Setup menu.
US

General Monitors S5000 29


Operation

Disable Sensor
When removing a sensor from the transmitter while under power, the S5000 will enter a Sensor
Missing fault condition after the two minute Swap Delay period has expired (if enabled). If
Swap Delay is disabled, the transmitter will go into Sensor Missing fault immediately. If the system
is off at the time a sensor is removed, the transmitter will go into fault after its startup sequence.
This fault condition can be removed by disabling the affected second sensor position.
Disabling a sensor removes the fault and stops communications with the sensor, the sensor’s
reading on the display is removed, and the mA channel for that sensor position is set to 0 mA. By
default, the S5000 has the Sensor 2 position disabled. If at any time a sensor is connected to a
position that is disabled, the device will automatically enable that sensor position.
To disable the sensor after removal:
(1) Go to Sensor Setup menu.
(2) Touch button when Disable appears on screen.
(3) Select Sensor 1 or Sensor 2.
Current status (enabled/disabled) displays.
(4) Touch button to toggle to desired status.
(5) Select “Finished?”.

To remove a second sensor that is already installed, the sensor must first be removed.
The second sensor cannot be disabled while it is installed on the transmitter.
The device only allows a sensor to be disabled after the transmitter has gone into
Sensor Missing fault.
Only one sensor can be disabled at a time. The transmitter will not allow.

Gas Label and Unit


Change the gas unit displayed. For toxic sensors, select from ppm, mg/m3, and µm/mol.
(1) Go to Sensor Setup menu.
(2) Touch button when Gas Label & Unit appears on screen.
The Gas type and unit will appear on the screen.
(3) Touch button while the units scroll to change to another unit.
(4) Touch button when “Finished?” is displayed to exit.
US
Sensor Range
Set the range of the sensor. To see what ranges are possible for the sensor type, see Tab. 24
"Digital and IR400 Sensor Default Settings".
(1) Go to Sensor Setup menu and select “Finished?” until Range displays.
Range will scroll across the screen and the current Range will be displayed.
(2) Hold finger on button to scroll quickly through values or touch button repeatedly to move
through values more slowly.
(3) When the desired value is displayed, wait until "Finsihed?" scrolls across screen.
(4) Touch button when “Finished?” is displayed to exit.

General Monitors S5000 30


Operation

Cal Level (i.e. Span Value)


Set gas concentration used during calibration.
(1) Go to Sensor Setup menu and select “Finished?” until Cal Level displays.
Cal Level will scroll across the screen and the current Cal Level will be displayed.
(2) Hold finger on button to scroll quickly through values or touch button repeatedly to move
through values more slowly.
(3) When the desired value is displayed, wait until "Finsihed?" scrolls across screen.
(4) Touch button when “Finished?” is displayed to exit to save the sensor parameter.
Warning Settings
Set Warning level and whether warning is latching or non-latching.
(1) Go to Sensor Setup menu and select “Finished?” until Warning displays.
Warning will scroll across the screen and the current Warning level will be displayed.
(2) Hold finger on button to scroll quickly through values or touch button repeatedly to move
through values more slowly.
(3) When the desired value is displayed, wait until Latching option appears.
(4) To change warning to latching or non-latching, touch button when the option is displayed.
(5) Touch button when “Finished?” is displayed to exit.
Alarm Settings
Set Alarm level and whether warning is latching or non-latching.
(1) Go to Sensor Setup menu and select “Finished?” until Alarm displays.
Alarm will scroll across the screen and the current Alarm level will be displayed.
(2) Hold finger on button to scroll quickly through values or touch button repeatedly to move
through values more slowly.
(3) When the desired value is displayed, wait until Latching option appears.
(4) To change alarm to latching or non-latching, touch button when the option is displayed.
(5) Touch button when “Finished?” is displayed to exit.

Alarm Relay Settings is the last Sensor Setup menu. “Finished?” will save changes and US
take the user back to Setup menus.

General Monitors S5000 31


Operation

Sensor Reset - Last option in Sensor Setup

CAUTION!
The sensor must be calibrated after a sensor reset. There will be displayed four dashes where the
current reading should be.
All settings, including Alarm Set Point and Calibration Values, will be returned to factory defaults.
The sensor default values can be restored by resetting the sensor. During sensor reset, the analog
output displays the sensors current gas value (i.e. 0 % LEL = 4 mA) during the 120 s count down.
To reset sensor to factory default:
(1) Go to Sensor Setup menu.
(2) Go through menus selecting “Finished?” at the end of each option.
(3) Touch button when Reset Setting scrolls across screen.
(4) Touch button to change to yes.
(5) Touch button when “Finished?” is displayed to exit.

This is the last Sensor Setup menu. “Finished?” will take the user back to Setup Menus
and will save any changes made.

4.2.3 Relay Settings


Relay Setup is used to change the relay zone mapping and energized/de-energized settings.
Relay Zone/Mapping
(1) Hold finger on button while menu selections scroll across screen.
(2) Remove finger when Setup menu appears.
(3) Touch button when Relay Setup is displayed.
Current setting of Zone/Mapping will be displayed (Common/Discrete).
(4) Touch button again will toggle the setting.
(5) If setting is changed, touch button when “Finished?” is displayed to save the change and
move on to next setup.
If no change is made, the menu will move on to next setup after two screen scroll rotation.
US

General Monitors S5000 32


Operation

Relay 1 and Relay 2 can be configured for common and discrete modes via the device display
menu or X/S Connect app.
Common mode is the default relay mapping setting. In Common mode, Relay 1 is actuated by
Alarm 1 on either sensor, and Relay 2 is actuated by Alarm 2 on either sensor.

Fig. 28 Common Mode Relay Map and Alarm Actuation


Discrete mode allows a separate action for each sensor. Relay 1 is actuated by Sensor 1 alarms
and Relay 2 is actuated by Sensor 2 alarms

US

Fig. 29 Discrete Mode Relay Map and Alarm Actuation

Energized/De-energized
(6) Relay 1 and Relay 2 will scroll across screen alternatively.
(7) Touch button when Relay 1 or Relay 2 is scrolling.
(8) Touch button to show the current setting (energized/de-energized).
(9) Touch button again will toggle the setting.
(10) Touch button when “Finished?” is displayed to exit from Relay setting.

General Monitors S5000 33


Operation

4.2.4 HART Settings


There are two HART menus under Setup. HART Setup is used to Enable/Disable HART.
HART AO Setup is used to configure the analog output signal for HART.
Enable/Disable HART
(1) Hold finger on button while menu selections scroll across screen.
(2) Remove finger when Setup menu appears.
(3) Touch button when HART Setup is displayed.
The current HART setting is displayed (Enabled or Disabled).
(4) To change the availability of HART, touch button to change to the opposite state.
(5) Touch button when "Finished?" is displayed to exit.
4.2.5 HART AO
(1) Hold finger on button while menu selections scroll across screen.
(2) Remove finger when Setup menu appears.
(3) Touch button when HART AO Setup is displayed.
The current HART analog output signal is displayed.
(4) To change the HART output signal, touch button to change to either 1.25 or 3.5.
(5) Touch button when "Finished?" is displayed to exit.
4.2.6 Bluetooth®
Enable/Disable Bluetooth®
If the device is ordered with Bluetooth®, it will come enabled by default. If unit is not ordered with
Bluetooth®, it cannot be retroactively added and there will not be this menu option. This menu will
not be displayed if it the device was ordered without Bluetooth®.
(1) Hold finger on button while menu selections scroll across screen.
(2) Remove finger when Setup menu appears.
(3) Touch button when Bluetooth® Setup is displayed.
The current Bluetooth® setting is displayed (Enabled or Disabled).
(4) To change the Bluetooth® availability, touch button to change to the opposite state.
(5) Touch button when "Finished?" is displayed to exit.
Bluetooth® Pairing US
The device memory has the ability to store up to 25 mobile devices in its memory. If a 26th device
is paired, the memory will erase the first device stored in its memory. Each time a device is paired,
it is logged as an event that is viewable in the X/S Connect App.
The S5000 blue LED in the middle left of the display will flash when a device is paired with it as a
visual indication of the pairing. Once paired with an S5000, the user will be able to connect to the
same S5000 remotely and without a password, unless over 25 other devices are paired with the
same S5000 afterwards.
To pair with the S5000:
(1) Download the X/S Connect App from the Google Play Store.
(2) Open the X/S Connect App.
(3) Select “Pair” for the S5000 that you would like to connect with.
(4) (First Time Only) Enter Pairing Code on S5000 display.
(5) (First Time Only) Accept pairing by touching button on S5000 display.

General Monitors S5000 34


Operation

Bluetooth® Security
The Bluetooth® connection is encrypted and secured with a unique six digit pin that must be
confirmed on the mobile device and acknowledged on the detector display. All of the previously
paired devices can be erased from the S5000 to provide additional security and control.
To reset all device pairings:
(1) Scroll to Settings.
(2) Select Instrument.
(3) Scroll and select Bluetooth®.
(4) Scroll and select “Reset All”.
(5) Select “Continue”.

NOTICE
Reset All will delete all paired device memory. All devices will have to re-initiate pairing at the
device.

Bluetooth® Tag ID
See section 4.3.5 to view Bluetooth® Tag ID.
4.2.7 Modbus Settings
The Modbus Setup menu consists of three configurable sub menus: Baud Rate, Format and
Address.
(1) Hold finger on button while menu selections scroll across screen.
(2) Remove finger when Setup menu appears.
(3) Touch button when Modbus Setup is displayed.
Modbus Baudrate
(1) Go to Modbus Setup menu. Baudrate option is the first submenu.
Baudrate will scroll across display followed by the current Baudrate setting.
(2) Touch the button to change to one of the following baudrate options:
a) 2400
b) 4800
c) 9600 US
d) 19K2
e) 38K4
f) 115K
(3) Touch button when “Finished?” is displayed to exit.

General Monitors S5000 35


Operation

Modbus Data Format


(1) Go to Modbus Setup menu and select “Finished?” until Format is displayed.
Format will scroll across display followed by the current Format setting.
(2) Touch the button to change to one of the following format options:
a) 8N1
b) 8N2
c) 8E1
d) 8O1
(3) Touch button when “Finished?” is displayed to exit.
Modbus Address
(1) Go to Modbus Setup menu and select “Finished?” until Address is displayed.
Address will scroll across display followed by the current Address setting.
(2) Hold finger on button to scroll quickly through values or touch button repeatedly to move
through values more slowly. Address options range from 1-247.
(3) Touch button when “Finished?” is displayed to exit.
4.2.8 EZ Touch Button
The device has a single button that allows the user to change configurations without the use of a
magnet or any other tool. Disabling the EZ Touch button requires the user to access menus with
a magnet or via HART, Modbus, or Bluetooth®.
(1) Hold finger on button while menu selections scroll across screen.
(2) Remove finger when Setup menu appears.
(3) Touch button when EZ Touch is displayed.
The current EZ Touch setting is displayed (Enabled or Disabled).
(4) To change the EZ Touch availability, touch button to change to the opposite state.
After disabling and selecting “Finished?”, the user will need a magnet to enable the setting.
Otherwise it will not be possible to navigate the S5000 menus.
(5) Touch button when "Finished?" is displayed to exit.

US

General Monitors S5000 36


Operation

4.2.9 Swap Delay


Swap Delay allows the user a brief window to change an XCell sensor without the device going
into a fault condition. Once a sensor is disconnected from the transmitter, the user will have
2 minutes to reconnect a sensor. During this time, the device display and analog output will go to
its Maintenance level. If a sensor is reconnected or replaced during the 2 minute window, the new
sensor’s countdown sequence will begin and the analog output will remain at the Maintenance
level. After the sensor countdown is complete, the analog output will return to reporting a live gas
reading. If a sensor is not reconnected after the 2 minute window, the S5000 will enter a fault
condition. All XCell Sensors have Safe Delay and do not need to be disconnected from power
while changing sensors. For more details on how to change sensors, see section 6.2. Swap Delay
is enabled on all S5000 transmitters by default.

NOTICE
The transition to maintenance mode during the 2 minute Swap Delay window and sensor count-
down will not trigger the Fault Relay. The Fault Relay will only be triggered when the device enters
a fault condition.

To change Swap Delay status:


(1) Hold finger on button while menu selections scroll across screen.
(2) Remove finger when Setup menu appears.
(3) Touch button when Swap Delay is displayed.
The Swap Delay status is displayed (enabled or disabled).
(4) To change the Swap Delay status, touch button to change to either enabled or disabled.
(5) Touch button when "Finished?" is displayed to exit.
4.2.10UI Password
Enable/Change Password
Enabling and changing the password is within the same menu. Enabling password will require the
user to enter the password before entering any of the settings menus. The password entry screen
defaults to 0000 and is disabled by default.
If the password is lost, call GM Customer Service at +1-949-581-4464.
(1) Hold finger on button while menu selections scroll across screen.
(2) Remove finger when Setup menu appears.
(3) Touch button when UI Password Setup is displayed. US
The current UI Password setting is displayed (Enabled or Disabled).
(4) Touch button to change to the opposite state.
(5) Touch button when “Finished?” is displayed to save the enable/disable setting.
Next menu "New Password?" is displayed is for setting password. Default password is 0000
if enabled.
(6) To change password, touch button when “No” appears on screen to change option to “Yes”.
(7) Touch button while “Finished?” is displayed to save change.
(8) The password entry screen will display a current password and each digit will blink for
4 seconds starting from left.
(9) While blinking, touch button to change desired numbers. The range is from 0 to 9 for each
digit and 0000 to 9999 for all 4 digits.
(10) After changing the password, select “Finished?” to store the new password in the system.

General Monitors S5000 37


Operation

Enter Password (if password is enabled)


If a password is enabled, the user must enter the password before entering any of the settings
menus. The password entry screen defaults to 000 and is disabled by default
(1) Holding finger on button will display rotating left arrow then “password”.
(2) Remove finger when UI Password appears.
(3) The password entry screen will display with 0 and each digit will blink for 4 seconds starting
from left.
(4) While blinking, touch button to change the desired numbers. The range is from 0 to 9 for one
digit and 0000 to 9999 for all 4 digits.
(5) If the password is correct, holding finger on the button again will display rotating left arrow
then go into the menu items.

As long as the setup menu is active, no password is required to enter again. If menu
was not active for 30 seconds, the password will be required again.

4.2.11Transmitter Setting Reset


The transmitter default values can be restored by resetting the main board. The unit will restart
after the reset and the unit may go into sensor fault. During transmitter reset, the analog output
displays 1.25 mA.

CAUTION!
Verify all sensor settings (calibration level and alarm values) after resetting transmitter. Sensors
may require recalibration to clear a fault after resetting transmitter.

4.3 Info Menu - Viewing Device Status


This section describes how to view the following options:
• Display Current System Time
• Change System Time
• Sensor Life and Health Status
• Last Calibration Dates
• Non-critical Faults
• Device Tag
• Bluetooth® ID
US

General Monitors S5000 38


Operation

4.3.1 View or Change System Time


(1) Go to Info menu and select System Time.
(2) Hold finger on button while menu selections scroll across screen.
(3) Remove finger when Info is displayed.
(4) Touch button when System Time is displayed.
The current date followed by the current time will scroll across the screen twice.
(5) Touch button when Change Time displays if the current date and time are not correct.
(6) If selected, change the following values using the button and selecting "Finished?" after each
entry:
a) Month
b) Day
c) Year
d) Hour
e) Minutes
(7) Touch button when "Finished?" is displayed to exit.
4.3.2 Last Calibration
(1) Go to Info menu and select Last Calibration.
The last Calibration and Zero Calibration dates will scroll across the screen twice.
(2) Touch button when "Finished?" is displayed to exit.
4.3.3 Non-Critical Fault Log
(1) Go to Info menu and select Non-Critical Fault.
Each of the last 10 faults will scroll across the screen twice.
(2) At the end of the fault log, menu goes back to Info menu options.

The user will not be asked to exit with “Finished?” at the end of this menu.

4.3.4 Device Tag


US
(1) Go to Info menu and select Device Tag.
The current Device Tag name will scroll across screen.
• DEFAULT - GM_S5K_0
• The device tag name can be changed via Modbus or HART and is limited to 8 characters.
(2) After the Device Tag name scrolls across screen twice, menu goes back to Info menu
options.

The user will not be asked to exit with “Finished?” at the end of this menu.

General Monitors S5000 39


Operation

4.3.5 Bluetooth® FCC/IC ID


(1) Go to Info menu and select Bluetooth FCC/IC ID.
The current Bluetooth FCC/IC ID will scroll across screen.

The Bluetooth® FCC/IC ID cannot be changed and is unique to the device.

(2) Touch button when "Finished?" is displayed to exit.


4.3.6 Sensor Life and Health Status (only displayed if XCell Sensor is connected)
(1) Select Info.
(2) Select Sensor Status.
If only one sensor is installed, the device displays the Life & Health status for Sensor #1.
If two sensors are installed, "Sensor Selection" will be displayed.
(3) Select the desired sensor (Sensor 1 or Sensor 2) and touch button when "Finished?" is
displayed.
The Life and Health information reports the general health of the sensor. The sensor Life and
Health can either be "Good" or "Fair". When sensor status turns to "Fair", the remaining sensor
life is approximately 2-3 months, and a replacement sensor should be ordered to ensure
continuity of gas detection.
XCell sensors with TruCal (Hydrogen Sulfide and Carbon Monoxide) calculate remaining sensor
life using automated pulse checks. The pulse stimulates the sensor with a response similar to
having actual calibration gas applied. The stimulated response is compared to the last calibration
and will make adjustments to sensitivity to match the last calibration. When the required
adjustment is greater than the accuracy of the algorithm's adjustment, the sensor will call for a
calibration. The pulse check tracks the overall degradation in sensitivity and will adjust the Life
and Health status from "Good" to "Fair" when loss of sensitivity is due to overall sensor health.
TruCal pulse checks will also detect when the sensor is no longer capable of sensing gas,
resulting in a Fault condition.

US

General Monitors S5000 40


Operation

4.4 Setting only configurable via Bluetooth®, Modbus, or HART


The following settings are only configurable through Bluetooth®, Modbus, or HART.
• AO Custom Levels
• Relay Zone – Horn Mode
• Unit Device ID/Tag
• Alarm Direction
• Alarm Enable/Disable
4.4.1 AO Custom Levels
S5000 offers three factory defined analog outputs (AO) level sets, i.e., 0 mA with HART disabled,
or 3.5 mA and 1.25 mA with HART enabled. Customer can also specify their own custom
AO levels for each transmitter mode. The Maintenance AO level is used during start up,
Reset Main Unit, and Transmitter Settings Reset. Output setting for oxygen sensor calibration is
not configurable. All changeable settings can be modified through Modbus/Bluetooth®/HART.

HART HART Default Enabled: Custom 1 Custom 2


Mode
Disabled (3.5 mA) HART (1.25 mA) Default Default
Fault 0 mA 3.5 mA 1.25 mA 1.25 mA*
1.5 mA*
Calibration & (All Other Sensors)
1.5 mA 3.5 mA 1.5 mA 1.5 mA*
Gas Check 21.0 mA
(Oxygen Sensor)
IR Cleaning
2.0 mA 2.0 mA*
Needed
Maintenance 3.5 mA 3.5 mA*
Zero Reading 4.04 mA
Over Range 21.7 mA

Tab. 30 Default AO Values


The range for the changeable AO settings is 0.000 mA to 3.750 mA. The change step is
0.025 mA. Please note the HART communication may not work reliably if the current level is set
to below 1.25 mA.
Horn mode is designed to allow local acknowledgment of a relay-triggered horn, while the alarm
state is still present. All alarms on both sensors trigger both relays, however the first relay can be
acknowledged over Modbus/Bluetooth®/HART.
4.4.2 Unit Device ID/Tag US
This is device tag/name used to identify the unit during HART and Bluetooth® communication.
Default is “GM_S5K_0 “. User can change this string to any characters from the following char-
acter set. The string length is limited to eight characters.

General Monitors S5000 41


Calibration

4.4.3 Alarm Direction


Relays can be triggered by an alarm threshold as gas readings increase or decrease. Most appli-
cations require increasing alarm thresholds, except for Oxygen monitoring, which is most often a
decreasing alarm.
The default alarm direction for Oxygen sensor is decreasing, or downward.
The default alarm direction for all other sensors is increasing, or upward.
The alarm direction can be set to increasing/upward or decreasing/downward.
Alarm direction for Oxygen sensor can be changed through UI directly, or through Modbus/Blue-
tooth®/HART.
The alarm direction setting for all other sensors can only be changed through Modbus/Bluetooth®/
HART.
4.4.4 Alarm Enable/Disable
User can disable alarms by changing this setting.
Default value is Enabled for this setting.
User can change this setting to Enabled or Disabled.
5 Calibration
Calibration is the process of applying a known quantity of gas to the transmitter so that the
transmitter can adjust the precision and accuracy of the measurements made in normal operating
mode. This process ensures that gas measurements are as accurate as possible.
Calibration Warnings - Read before Calibrating
Although S5000 sensors are factory calibrated, another calibration is recommended once the unit
is installed in its final environmental destination.

WARNING!
Use zero gas when zeroing the S5000 transmitter if there is any possibility of background gas.
Otherwise, improper calibration could occur.
Always calibrate for the least sensitive gas or vapor (higher number category) expected to be
measured. Otherwise, instrument readings may be incorrect.

CAUTION!
Before attempting a calibration, power the unit at least one full hour.
Perform a calibration at initial start-up and at regular intervals to ensure a fully functional sensor.

5.1 Calibration Equipment


A gas cylinder with a known concentration of gas appropriate for the range of measurement is
needed. Sensors come with preset span gas values appropriate for the measurement range.
US
All of the calibration hardware for XCell sensors is included in a convenient calibration kit. It
includes a carrying case that can fit two cylinders, either a 0.5 L/min regulator for passive sensors
or a 1 L/min regulator for IR400 and Digital Sensors, tubing, and any other accessories required
for each sensor type. The calibration kit can be ordered separately (CALKIT1) or as a single line
item with the gas cylinder. See Tab. 31 for a complete list of calibration cylinders.
The Digital Toxic, Oxygen, and Catalytic bead sensors share the same sensor body design and
use the same calibration hardware. A black plastic sensor guard should be screwed to the bottom
of the sensor body before calibration and removed after calibration. The calibration tubing should
be thread through a green rubber boot and then attached to the inlet of the sensor guard. The
edge of the green rubber boot should be flush against the sensor guard while calibrating.
Passive sensors use different calibration hardware to accommodate the different sized sensor
housing. Passive MOS sensors can be calibrated with either calibration gas cylinders or ampules.
See Tab. 31 for part number details.

General Monitors S5000 42


Calibration

The IR400 sensor has different calibration hardware, but is still included in CALKIT1.
The sensor guard and rubber boot should be removed after calibration. The sensor
guard actually makes the sensor more sensitive to gas in windy conditions and could
result in a false alarm.

P/N with
Gas Type Range Concentration Part Number
Calibration Kit
0-100 PPM 60 PPM 710882 710882-KIT1
Carbon
0-500 PPM 300 PPM 10027938 10027938-KIT1
Monoxide
0-1000 PPM 400 PPM 10028048 10028048-KIT1
0-10 PPM 5 PPM 710414 710414-KIT1
Hydrogen
0-50 PPM
Sulfide 40 PPM 10028062 10028062-KIT1
0-100 PPM
Oxygen 0-25 % 20.8 % 10028028 10028028-KIT1
0-100 % LEL 2.5 % Methane
5 % Methane (50 % LEL)
10028032 10028032-KIT1
0-100 % LEL 2.5 % Methane
Combustible 4 % Methane (57 % LEL)
(IR400 or
Cat Bead) 0-100 % LEL 0.6 % Propane
2.1 % Propane (29 % LEL)
10028035 10028035-KIT1
0-100 % LEL 0.6 % Propane
1.7 % Propane (35 % LEL)
Passive Cat
0-100 % LEL 2.5% Methane 1400155-M 1400150-M
Bead
Ampules Gas Ampules Gas
0-100 PPM 50 PPM 50004-13 1400255-5 50009-9 1400250-5
Passive MOS
0-50 PPM 25 PPM 5004-21 1400255-3 5009-16 1400250-3
0-20 PPM 10 PPM 50004-3 1400255-1 50009-10 1400250-1

Tab. 31 Calibration Kits


5.2 Calibration Frequency US
The frequency of calibration gas testing depends on the operating time, chemical exposure, and
type of sensor. Especially in new installations or applications, it is recommended that the first
sensors be calibrated more often to establish the sensor performance in this particular
environment.
For this, record the "as found" and "as left" values and track the percent adjustment over time.
Then, gradually extend calibration intervals until the percent adjustment is greater than the
expected accuracy of the sensor.

General Monitors S5000 43


Calibration

5.3 Calibration Frequency for XCell Sensors with TruCal (H2S & CO only)
XCell sensors with TruCal actively self-monitor for operation and accuracy with fewer manual cali-
brations required in order to maintain the equivalent level of performance. When the sensor
detects a change in sensitivity outside what is typical for a six hour time interval, the sensor will
call for a calibration. The transmitter LED status indicators slowly pulse green, alerting the user
that a calibration is recommended to maintain optimum performance. There are no changes to the
analog output when TruCal calls for a manual calibration.
Actual TruCal sensor performance will depend on the application, background gas exposure, and
environment. To validate XCell sensors with TruCal, it is recommended that users follow their
regular calibration cycle and record the “as found” and “as left” values, tracking the percent adjust-
ment over time. Once a baseline is established, the calibration intervals can be extended until the
percent adjustment is greater than the expected accuracy of the sensor.
5.4 Calibration Types: Zero vs. Span
There are two types of calibration; zero and span calibration.
The zero calibration resets the baseline level reading to zero. If the target gas is suspected to be
occasionally present, it is best to also use a zero gas cylinder during the zero calibration. If the
target gas is not present in the atmosphere, an additional calibration cylinder is not required.
The span calibration is a process that involves first zeroing the baseline reading, then applying a
known amount of target gas so that the accuracy and precision of the transmitter can be adjusted
to the known value. The known concentration of gas is called a "Span Value" because it
represents the span or reach of the calibration curve from zero to that value.

US

Fig. 32 Calibration Curve


The Sensor Span Value (i.e. Cal Level) in the device menu should be the same as the
concentration listed on the calibration gas cylinder.

General Monitors S5000 44


Calibration

5.5 How to Zero Calibrate XCell and IR400 Sensors


NOTICE
If a password is enabled, you will not be able to proceed with the calibration without the password.

To abort, press the EZ touch button or apply magnet over GM logo before applying
target gas.
In the event that a calibration cannot be completed, the user can acknowledge the FAIL
by pressing both EZ touch buttons simultaneously for 5 seconds. The unit will revert to
the settings of the last successful calibration.

WARNING!
The regulator used with the zero gas cylinders should not be the same as the regulator used for
the target gas. The regulator can be contaminated with the target gas over time, and thus raise
the detection baseline and make the sensor less sensitive to the target gas.

If there is no target gas in the atmosphere around the sensor, using a zero gas cylinder
is optional.

To zero calibrate the sensor:


(1) Screw the sensor guard to the bottom of the sensor housing.
(2) Place the green calibration cap over the sensor guard inlet so that it is flush with the bottom
of the sensor guard and completely covers the sensor guard inlet.
(3) Attach the tubing to the plastic stem protruding through the green calibration cap.
(4) Screw the regulator onto the top of the zero gas cylinder.

If a password is enabled, you will need to enter it to access the menu system.

(5) Press EZ touch until Zero is displayed, and release.


(6) Once the screen displays Zero in Progress, turn on zero gas flow by turning knob on the
US
regulator.
(7) Wait while the device displays Zero in Progress.
(8) Once the zero calibration is complete, a Zero PASS or FAIL is displayed.
If PASS displays, the procedure is complete.
If FAIL displays, the procedure was unsuccessful (see section 6.4).
(9) Remove the zero calibration equipment from the device including the sensor guard and
calibration cap.
(10) Remove the calibration equipment from the device, including the sensor guard and
calibration cap.

General Monitors S5000 45


Calibration

5.6 How to Calibrate XCell Sensors


See section 5.7 for calibrating oxygen sensors.

NOTICE
If a password is enabled, you will not be able to proceed with the calibration without the password.

To abort, press the EZ touch button or apply magnet over GM logo before applying
target gas.
In the event that a calibration cannot be completed, the user can acknowledge the FAIL
by pressing both EZ touch button until “Reset” is displayed and then release. The device
will revert to the settings of the last successful calibration.

If it is suspected that gases are present, it will be necessary to purge the sensor envi-
ronment with zero air.

To calibrate the sensor:


(1) Attach a regulator to the zero gas cylinder (if using) and the calibration cylinder.
(2) Attach the plastic sensor guard to the bottom of the sensor housing if not already installed.
(3) Place the green calibration cap over the sensor guard inlet so that it is flush with the bottom
of the sensor guard and completely covers the sensor guard inlet.
(4) Attach the tubing to the plastic stem protruding through the green calibration cap.
(5) Push the other end of the tubing over the zero cylinder regulator. Ensure the tubing
completely covers the gas outlet.
(6) Press EZ touch until Calibration is displayed, and release.

If a password is enabled, you will need to enter it to access the menu system.

(7) Select Sensor #1 or Sensor #2, and then select “Finished?” to start the calibration process.
(8) Once the screen displays Zero in Progress, turn on zero gas flow by turning knob on the US
regulator (if using).
(9) Wait while the device displays Zeroing.
(10) Once the Zero Calibration is complete, remove the tubing from the sensor guard inlet.
(11) The display will show Apply Gas. Attach tubing for calibration gas and turn on the regulator
Display will show “Calibration in Progress”.
“Calibration Complete Remove Gas” will show when span is complete.
If “PASS” displays, the procedure is complete.
If “FAIL” displays, the procedure was unsuccessful. See section 6.4.
(12) Remove the calibration equipment from the device, including the sensor guard and
calibration cap.

The green calibration cap should always be removed after calibration.

General Monitors S5000 46


Calibration

5.7 How to Calibrate an Oxygen XCell Sensor


An oxygen span gas cylinder is not needed if the sensor is in an area that maintains ambient air
conditions. Follow the same process for XCell sensors as described in section 5.6. When the
display prompts "Apply Span Gas", simply allow the countdown to occur without applying gas.
If the sensor is located in an area of normally low or enriched oxygen, then a 20.8% oxygen
sample must be applied.
5.8 How to Calibrate an IR400 Sensor
A full span calibration is not required for the IR400 sensors. Any degradation of the sensor's
performance is associated with slight drifts in its zero response. Restoring the sensor's zero is
typically sufficient.
See section 5.5 for zero calibration instructions.

If a full calibration of the IR400 sensor is required, a larger calibration cap is used and
covers the entire sensor guard during calibration. The calibration cap must be removed
from the XIR sensor guard after completing the zeroing and/or spanning procedure.

5.9 How to Calibrate a Passive Sensor (Catalytic Bead or MOS)

If a password is enabled, you will need to enter it to access the menu system.

If the sensor is fitted with a sensor guard, calibration must be performed with the sensor
guard in place.

The sensor calibration cup should be removed after calibration.

If it is suspected that gases are present, it will be necessary to purge the sensor envi-
ronment with zero air.

To calibrate sensor: US
(1) Attach a regulator to the zero air cylinder (if using) and the calibration cylinder.
(2) Attach the plastic calibration cup (P/N 1400152-1 or 1400154) to the bottom of the sensor if
not already installed.
(3) Attach the tubing to the plastic stem protruding through the calibration cup.
(4) Push the other end of the tubing over the zero cylinder regulator. Ensure the tubing
completely covers the gas outlet.
(5) Press EZ touch until Calibration is displayed, and release the touch.
(6) Turn on zero gas flow by turning knob on the regulator (if using).
(7) Unit will start flashing the sensor life. Tap the EZ Touch if you want to reset the sensor life.
(8) Wait while the device displays “Zero in Progress”.
(9) Once the Zero Calibration is complete, remove the tubing from the sensor guard inlet.
(10) The display will show “Apply Gas”. Attach tubing for calibration gas and turn on the regulator.
(11) Display will show “Calibration in Progress”.

General Monitors S5000 47


Maintenance

(12) “Calibration Complete. Remove Gas” will show when span is complete.
(13) Turn off the calibration gas regulator. Remove the calibration equipment from the device,
including the sensor guard and calibration cap. Unit will return to normal operation.
If “Cal Fail” and fault code displayed, the procedure was unsuccessful. See section 6.4.
5.10 XCell Catalytic Bead Failsafe
Catalytic bead sensors require the presence of oxygen in order to sense combustible gas. In the
event of very large combustible gas leaks that exceed 100 % LEL, enough oxygen can be
displaced so that the sensor's response to gas is no longer proportional to the calibration profile.
The XCell catalytic bead has a locking fail safe mechanism that prevents the false reporting of a
safe condition while the % LEL concentration is still above 100 % LEL. When the gas
concentration exceeds 100 % LEL, the sensor will go into LOC over range.
To clear the LOC over rage, the user needs to acknowledge and calibrate the sensor. To
acknowledge LOC over range, place one finger over each of the EZ touch buttons and hold for
1 second before releasing. This will allow a recalibration of the sensor to clear the LOC condition.

CAUTION!
Ensure that the area has been cleared of gas before acknowledging the LOC Over Range and
recalibrating the sensor.

5.11 Calibration Confirmation


The S5000 Series Monitor records the date of the last successful calibration. This date can then
be displayed via the Status Menu.
6 Maintenance
WARNING!
Repair or alteration of the device, beyond the procedures described in this manual or by anyone
other than a person authorized by MSA, could cause the unit to fail to perform properly. Use only
genuine MSA replacement parts when performing any maintenance procedures described in this
manual.
Substitution of components can seriously impair performance of the device, alter intrinsic safety
and flameproof/explosion proof characteristics or void agency approvals. Failure to follow this
warning can result in serious personal injury or death.
The S5000 Gas Monitor is constantly performing a self-check. When a problem is found, it
displays the appropriate error message (see section 6.4). When a critical error is detected within
the device, the 4-20 mA output signal goes to a fault condition of 0 mA, 1.25 mA, 3.5 mA or user
defined level.
US

General Monitors S5000 48


Maintenance

6.1 IR400 Cleaning Procedure


The presence of particulate matter, oil films, liquid water, or the residue from water drops on the
two monitor windows can adversely affect its performance. The sensor guard is designed to
prevent foreign solids or liquids from reaching the monitor's optical system. Additionally, heating
elements are incorporated into the device to prevent water condensation. Under severe condi-
tions, however, some material may collect on these surfaces and it may be necessary to occa-
sionally check and clean the windows.
While both windows are made of a highly durable material that is not easily scratched, avoid
excessive pressure when cleaning them. Clean, cotton-tipped applicators are the most convenient
tool to remove material collected on the windows.
(1) Remove the sensor guard.
(2) Use a dry applicator or one moistened with distilled water to wipe the window and remove
dust.
(3) Use an additional clean, dry applicator to remove any residual water.
(4) Use an applicator moistened with isopropyl alcohol to remove heavy deposits of solids,
liquids or oil films. Clean the window again with a second applicator moistened with distilled
water; then, dry the window with a final applicator.
Avoid using excessive amounts of water or alcohol in the cleaning procedure, and inspect
the window to ensure that the entire surface is clean.
(5) The device will go into a “Cleaning required” fault with an analog output of 2.0 mA when
cleaning is needed.
The S5000 analog output may go down to 1.25 mA while the optical path is blocked, and
display will indicate “Beam Block”.

While in "Beam Block" fault, the sensor will not respond to the presence of gas.

(6) When cleaning is done and the objects are removed from the sensor window, the device
returns to normal operation. If water or isopropyl alcohol was used, allow the device to
operate for 15 minutes to completely dry before replacing the sensor guard continuing to
monitor for combustible gas.
(7) Replace the sensor guard.
US
When the cleaning process is complete, be sure to remove all objects from the light
path. It is recommended to check the sensor’s response to zero and calibration gas after
cleaning.

WARNING!
Do not place foreign objects in the sensor's analytical region (except per the “S5000 IR400
Cleaning Procedure” as described above); otherwise, the infrared beam can be partially blocked,
causing the sensor to generate false readings. All objects must be removed from the sensor's
analytical region for it to function properly.

General Monitors S5000 49


Maintenance

6.2 Replacing an XCell Sensor


The only routine maintenance item is the sensing element itself, which has a limited lifetime. The
S5000 sensors with TruCal technology will indicate when the sensor is near end of life through the
Status Menu. When the Sensor Life & Health status is "Fair", there is approximately 2 month time
frame to replace the sensor before it will no longer function. Actual time between “Fair” status and
sensor end of life depends on exact environmental conditions. When a TruCal sensor is no longer
capable of sensing it will go into fault and LEDs will flash yellow. It is good practice to obtain a
replacement sensing element before the sensing element within the unit becomes inoperative.

WARNING!
Handle the sensor carefully; the electrochemical version is a sealed unit which contains a corro-
sive electrolyte.
Any leaked electrolyte that come in contact with skin, eyes or clothes can cause burns.
If any contact with the electrolyte does occur, immediately rinse with a large quantity of water. On
contact with the eyes, rinse thoroughly with water for 15 minutes and consult a doctor.

CAUTION!
Do not install a leaking sensor in the sensing head assembly. The leaking sensor must be
disposed of in accordance with local, state and federal laws.
There is no need to open the main enclosure. Simply unscrew the sensor assembly located on
the bottom of the main assembly.

WARNING!
XCell Digital Sensors marked for Class I hazardous (classified) locations and installed in a Class I
hazardous (classified) location only may be changed under power. To change, unscrew sensor at
least three full turns (but no more than four full turns from its fully engaged position), wait
10 seconds, and then remove the sensor completely. Failure to follow this warning can result in
the ignition of a hazardous atmosphere.
For XCell Digital Sensors marked Class II or III and installed in a Class II or III hazardous
location, the atmosphere must be free of dust or fibers and filings and the power removed from
the unit before the sensor cap can be removed from the housing. Failure to follow this warning
can result in the ignition of a hazardous atmosphere.
Identify the sensor assembly needed via the A-5K-SENS code on the interior sensor label and
obtain the appropriate sensor assembly; replace sensor assembly.
US
Alarm set points, span value, full scale limit and alarm direction will not change when
replacing a sensor with the same gas type.
Alarm set points, span value, full scale limit and alarm direction will change to the new
sensor’ settings when replacing a sensor with a different gas type.

The S5000 Gas Monitor is shipped with the Sensor Swap Delay enabled. IF enabled, the 4-20 mA
output signal goes to 3.5 mA and the FAULT relay will hold off a fault indication for 2 minutes. This
setting allows the operator to exchange sensor modules without a FAULT indication.

It is recommended that all other maintenance be performed at an MSA factory-autho-


rized service center.

General Monitors S5000 50


Maintenance

6.3 Replacing a Passive Sensor (Cat Bead or MOS)


When the passive Catalytic Bead or MOS sensor is damaged or gradually losing sensitivity and
eventually fails to respond to gas adequately or fails to pass calibration, the sensor needs to be
replaced.
The mounting threads of the sensor head may require lubrication. Grease must not be used. As
an alternate, PTFE (Teflon) tape may be used on sensor threads.
To replace the passive sensors:
(1) Declassify the area and make sure there is no hazardous gas/vapor present.
(2) Power down the S5000 transmitter.
(3) Unscrew the lid.
(4) Remove the User Interface Module.
(5) Refer to the sensor wiring diagram label inside the S5000 transmitter to locate the sensor
connector and unplug it from the transmitter.
(6) Unscrew the sensor head until it is totally off. Guide the sensor wires so they will not get
caught by the internal parts of the S5000 while sensor head is turning.
(7) Guide the connector and wires of the new sensor head through the conduit entry for sensor.
Do not plug in the sensor connector at this time.
(8) Screw in the new sensor head. Guide the sensor wires so they will not get caught by the
internal parts of the S5000 while sensor head is turning.
(9) Verify the sensor connector wiring against the wiring diagram.
(10) Plug in the connector to the indicated header and make sure they are mated correctly.
6.4 Troubleshooting
Fault Code Description Action
F001 Low Supply Voltage This fault occurs if the supply Ensure that the supply voltage is
voltage at the S5000 is below within the range of 12-30 VDC at the
+10.5 VDC or higher than S5000.
31.5 VDC.
F002 RAM Checksum Error This fault indicates that the Cycle the power to the unit. If the fault
contents of the S5000's internal persists, the unit must be returned to
RAM memory have corrupted. the factory or authorized service
center for repair.
F003 Flash Checksum Error This fault indicates that the The unit must be returned to the
US
contents of the S5000's factory or authorized service center
program memory have for repair.
changed. This usually occurs
when powering the unit up after
a lightning strike or large
voltage transient on the power
or signal lines to the unit.

General Monitors S5000 51


Maintenance

Fault Code Description Action


F004 EEPROM Error In the event of an EEPROM (1) Unit returns to normal.
error, the user must recycle the
a) This indicates the EEPROM
power to potentially clear the
writing did not keep up to
error. After power reset, the
changing events or the write
following may occur:
cycle is too fast.
(2) Unit goes to sensor faults. The
user must recalibrate after 1-2
minutes sensor warm up.
a) This means the non-critical
part of the EEPROM was
corrupted.
b) This is probably caused by an
event-logging problem.
c) There is a possibility any of
the following may be
corrupted: Event logging data,
Modbus settings, HART
settings, Calibration informa-
tion.
(3) Unit returns to F004.
a) This is a critical error. This
fault occurs when an attempt
to verify the setup/calibration
parameters just written to the
EEPROM memory fails.
b) Reset transmitter settings to
reload the factory default to
the EEPROM. Sensor recali-
bration is required after this
operation. All settings
including, but limited to, alarm/
warning, setpoint, relay,
Modbus, AO output level
US
should be verified and repro-
grammed.

F005 Internal Circuit Error An internal error has occurred. The


unit must be returned to the factory for
service. The possible errors are
internal voltages are not at their
proper values.
F006 Relay Fault This fault indicates that either The unit must be returned to the
the relays or the relay driver is factory or authorized service center
not working properly. for repair.
F007 Invalid Sensor Configu- This fault occurs when the Verify at least one sensor channel is
ration Fault sensor channel configuration is enabled. Cycle the power to the unit. If
invalid. the fault persists, the unit must be
returned to the factory or authorized
service center for repair.

General Monitors S5000 52


Maintenance

Fault Code Description Action


F008 Other System level Fault This fault occurs when other The unit must be returned to the
fault conditions not covered factory or authorized service center
above happen. for repair.
F101/F201 Sensor Missing This fault occurs when the Check the wiring and sensor connec-
Fault sensor is disconnected or wiring tion.
is faulty or damaged.
F102/F202 Sensor Supply This fault occurs when the Check the wiring and sensor connec-
Voltage Fault power supply to the sensor is tion.
faulty.
F103/F203 Invalid Sensor This fault occurs when the Check the sensor parameters,
Parameters in Main EEPROM sensor parameters stored in the including Calibration Level, Full Scale,
Fault main EEPROM is incorrect. Warning and Alarm set points. Correct
any mis-configured parameters.
F104/F204 Sensor Element This fault occurs when the Check the wiring and sensor connec-
Fault sensor sensing element is faulty tion. Replace the sensor if needed.
or damaged.
F105/F205 Sensor Heater This fault occurs when the Check the wiring and sensor connec-
Fault sensor heater is faulty or tion. Replace the sensor if needed.
damaged.
F106/F206 Other Sensor This fault occurs when the Check the wiring and sensor connec-
Internal Fault sensor has other internal faults. tion. Replace the sensor if needed.
F107/F207 Sensor Internal This fault occurs when the Replace the sensor.
Data CRC Error Fault internal data of the sensor has
CRC error.
F108/F208 Sensor EOL Fault This fault occurs when the Replace the sensor.
sensing element of the sensor is
at the end of life state.
F109/F209 Sensor Blockage This fault occurs when the Clean out any debris or foreign
Fault sensor gas sensing path and/or objects blocking the sensing path/
opening is blocked. opening.
F110/F210 Negative Drift Fault This fault occurs when the Re-calibrate the sensor. Replace the
sensor has severe negative sensor if this fault persists.
drift.
F111/F211 Cal Line Shortage This fault occurs when the Check the wiring and sensor connec- US
Fault IR400/IR700 sensor Cal Line tion.
(brown wire) is shorted to
ground.
F112/F212 Zero Calibration This fault occurs when last zero Use the “Reset Sensor” entry in main
Failed Fault calibration has failed. menu to reset the fault. Re-calibrating
the sensor will also clear the fault.
Replace the sensor if this fault
persists.
113/F213 Full Span Calibra- This fault occurs when last full Use the "Reset" entry in UI menu to
tion Failed Fault span calibration has failed reset the fault. Re-calibrate the sensor
will also clear the fault. Replace the
sensor if this fault persists.
F114/214 Cal Check Timeout This fault occurs when the last Check whether test gas is left on for
Fault Cal Check has failed due to Calibration Check mode. Remove the
timeout. test gas if it is left on. Once it is
cleared, the unit will return to normal
operation.

General Monitors S5000 53


Maintenance

Fault Code Description Action


F115/F215 Sensor Configura- This fault occurs when the Check the sensor parameters,
tion Reset Fault configuration parameters for including calibration level, Range,
sensor have been reset. Warning and Alarm set points. Verify
all these parameters are configured
as intended. Recalibrate the sensor to
clear the fault if needed. Running
through the Calibration Check is
recommended.
F116/F216 Calibration This fault occurs when the Recalibrate the sensor.
Required Fault sensor is in need of calibration.
F117/F217 Bead Off Fault This fault occurs when the bead Make sure the environment is free of
of the Catalytic Bead sensor is dangerous combustible gas and then
turned off when over use the “Reset” entry in UI menu to
100 % LEL gas has been reset the fault.
detected by the sensor.
F118/F218 Analog Output This fault occurs when the 4- Check the 4-20 mA analog output
Mismatch Fault 20 mA analog output loop is loop is closed properly with nominal
malfunctioning. 250 ohm load resistor installed. The
fault will be cleared automatically
once the loop is closed and wires are
properly connected.

Tab. 33 Troubleshooting

US

General Monitors S5000 54


Ordering Information

7 Ordering Information
7.1 Replacement Parts
See Tab. 34 for replacement sensor kits. To obtain a replacement sensor, address the order or
inquiry to:
General Monitors
26776 Simpatica Circle
Lake Forest, CA 92630
or call, toll-free, 1-800-446-4872.
Inquiries can also be e-mailed to [email protected].

WARNING!
Use only genuine MSA replacement parts when performing any maintenance procedures
provided in this manual. Failure to do so may seriously impair sensor and gas monitoring
performance. Repair or alteration of the S5000, beyond the scope of these maintenance
instructions or by anyone other than authorized MSA service personnel, could cause the product
to fail to perform as designed and persons who rely on this product for their safety could sustain
serious personal injury or loss of life.

Type Description P/N


PCBA Assemblies S5000-PCBA
Sensor Body, XCell A-5K-SENS
Sensors, XCell A-5K-SENS
Sensor Guard, XCell SENSOR GUARD ASSEMBLY, S5000 10177624
Sensor Guard, IR400 SENSOR GUARD, IR400, WITH SCREEN 32545-1
CB 0-100% (C04 – Screened, Div Approvals) 10058-1
CB 0-100% (C07 – Sintered, Zone Approvals) 11159-1L
MOS 0-100 (M04 – Screened, Div Approvals) 50448-1
MOS 0-100 (M11 – Sintered, Zone Approvals) 51457-1L
Sensors, Passive
MOS 0-50 (M05 – Screened, Div Approvals) 50448-5
MOS 0-50 (M12 – Sintered, Zone Approvals) 51457-5L
MOS 0-20 (M06 – Screened, Div Approvals) 50448-9
MOS 0-20 (M06 – Sintered, Zone Approvals) 51457-9L
Sensor, IR400 IR400 ATO US
RECTANGULAR DUCT MOUNT KIT, S5000 10176947
Duct Mount Kit ROUND DUCT MOUNT KIT, SMALL, S5000 10179124
ROUND DUCT MOUNT KIT, LARGE, S5000 10179321
Calibration Kit (XCell) CALKIT1
Pipe Mount Kit, Universal PIPE MOUNT, X5000/S5000 10176655
JUNCTION BOX, SST, IIB+H2 324240-1
Junction Box
JUNCTION BOX, SST, IIC 324240-3
Calibration Cap, XCell CALIBRATION CAP, X5000/S5000, PCKGD 10181450
Calibration Cap, IR400 CALIBRATION CAP, IR400 PCKGD 10181460
Sun Shield SUNSHIELD, X5000/S5000 10178977

Tab. 34 Replacement Parts

General Monitors S5000 55


Appendix: Specifications

8 Appendix: Specifications
Digital XCell IR
Sensor Options
Toxics Oxygen Combustible IR400
Operating Range -40 °C to +60 °C -40 °C to +60 °C -40 °C to 60 °C -40 °C to +75° C
Storage -40 °C to +60 °C -40 °C to +60 °C -40 °C to 60 °C -50 °C to +85 °C
Zero Drift1 ≤ 1 % FS ≤ 0.2 % Vol < 5 % LEL/mo < 2 % per year
Span Drift 1
< 2 %/mo ≤ 0.2 %  5 % LEL
Noise < 2 display units < 2 display units 1 % LEL 1 % LEL
±3 % LEL at
≤ 50 % LEL
Repeatability 1
±10 % ≤ 0.3 % Vol ≤ 3 % LEL
±5 % LEL at
> 50 % LEL
1 ppm, CO
Resolution 0.1 % Vol 1 % LEL 1 % LEL
0.1 ppm, H2S
T90 < 25 s < 15 s < 30 s ≤3s
Humidity 10-95 % RH 10-95 % RH 5-100 % RH 10-95 % RH
Sensor Life 5 years 5 years 5 years > 5 years
Warranty 3 years 3 years 3 years 2 years
Enabling XCell - XCell - XCell -
IR
Technology non-consuming non-consuming GM cat bead
TruCal Yes No No No
SafeSwap Yes Yes Yes No
Remote Mount
100 m 100 m 100 m 1100 m
Distance
Power:
3.6 W 3.6 W 6.0 W 4.8 W
Single Sensor
Power:
4.3 W 4.3 W 8.4 W NA
Dual Sensing

Tab. 35 Digital Sensor Specifications


1 not guaranteed for extended temperature range
US

General Monitors S5000 56


Appendix: Specifications

Passive Catalytic Bead Passive MOS


Sintered Screened
Sensor Options Sintered Screened
(Zone Approvals) (Division Approvals) (Zone Approvals (Division Approv
) als)
Operating Range -40 °C to +70 °C -40 °C to +75 °C -40 °C to +70 °C -40 °C to +75 °C
Storage -50 °C to +85 °C -50 °C to +85 °C -50 °C to +85 °C -50 °C to +85 °C
< 5 % of full scale < 5 % of full scale
Zero Drift1 None None
per year per year
Noise 1 % LEL 1 % LEL 1 ppm 1 ppm
±3 % LEL at ±3 % LEL at
≤ 50 % LEL ≤ 50% LEL ±2 ppm or 10 % of ±2 ppm or 10 % of
Repeatability1
±5 % LEL at ±5 % LEL at applied gas applied gas
> 50 % LEL > 50% LEL
Resolution 1 % LEL 1 % LEL 1 ppm 1 ppm
T50 < 10 s < 10 sec < 30 s < 14 s
T90 < 30 s < 30 s NA NA
Humidity 10-90 % RH 10-90 % RH 0-95 % RH 0-95 % RH
Sensor Life 3-5 years 3-5 years 3-5 years 3-5 years
Warranty 2 years 2 years 2 years 2 years
Enabling
Catalytic Bead Catalytic Bead MOS MOS
Technology
TruCal No No No No
SafeSwap No No No No
Remote Mount
307 m (12 AWG wiring) 1200 m (14 AWG wiring)
Distance
Power:
6.0 W 10.8 W
Single Sensor
Transmitters purchased with passive sensors are not capable of dual sensing
Power:
and cannot be changed to any other sensor type than the one with which it
Dual Sensing
was originally purchase.

Tab. 36 Passive Sensor Specifications


1
US
not guaranteed for extended temperature range

General Monitors S5000 57


Appendix: Specifications

Fig. 37 Housing Dimensions

Material Weight
7.30 lbs
316 Stainless Steel
7.92 lbs w/ relays

Tab. 38 Housing Material Specifications

US

General Monitors S5000 58


Appendix: General Certification Information

9 Appendix: General Certification Information


S5000 Transmitter
U.S. Division Hazardous Locations as defined by the NEC
Class I, Division 1, Groups A, B, C, D; T5
Class I, Division 2, Groups A, B, C, D; T4
Class II, Division 1,2 Groups E, F, G; Class III, Division 1,2; T6
Note: See CSA Certificate for detailed certification information including possible limita-
tions to the above information.

U.S. Zone Hazardous Locations as defined by the NEC


Class I, Zone 1, AEx db IIC T5 Gb
Class I, Zone 2, AEx nA nC IIC T4 Gc
Zone 21, AEx tb IIIC T85°C Db
Note: See CSA Certificate for detailed certification information including possible limita-
tions to the above information.

Canadian Division Hazardous Locations as defined by the CEC


Class I, Division 1, Groups A, B, C, D; T5
Class I, Division 2, Groups A, B, C, D; T4
Class II, Division 1,2 Groups E, F, G; Class III, Division 1,2; T6
Note: See CSA Certificate for detailed certification information including possible limita-
tions to the above information.

Canadian Zone Hazardous Locations as defined by the CEC


Ex db IIC T5 Gb
Ex nA nC IIC T4 Gc
Ex tb IIIC T85°C Db
Note: See CSA Certificate for detailed certification information including possible limita-
tions to the above information.

ATEX/IECEx Hazardous Locations


(Sira 17ATEX1048X, Sira 17ATEX4052X, IECEx SIR 17.0016X)
Ex db IIC T5 Gb
Ex nA nC IIC T4 Gc
Ex tb IIIC T85°C Db US
Note: See ATEX/IECEx Certificates for detailed certification information including possible
limitations to the above information.

General Monitors S5000 59


Appendix: General Certification Information

Digital Sensor
U.S. Division Hazardous Locations as defined by the NEC
Class I, Division 1, Groups A, B, C, D; T5
Class II, Division 1, Groups E, F, G; Class III; T5

U.S. Zone Hazardous Locations as defined by the NEC


Class I, Zone 1, AEx db IIC T5 Gb
Zone 21, AEx tb IIIC T85°C Db

Canadian Division Hazardous Locations as defined by the CEC


Class I, Division 1, Groups A, B, C, D; T5
Class II, Division 1, Groups E, F, G; Class III, T5

Canadian Zone Hazardous Locations as defined by the CEC


Ex db IIC T5 Gb
Ex tb IIIC T85°C Db

ATEX/IECEx Hazardous Locations


(Sira 17ATEX1048X, Sira 17ATEX4052X, IECEx SIR 17.0016X)
Ex db IIC T5 Gb
Ex tb IIIC T85°C Db
IR400
See the IR400 instruction manual, P/N MANIR400 or certificates Sira 07ATEX1253 and
IECEx SIR 07.0080.
Passive Sensors
See certificates Sira 00ATEX1039U and IECEx SIR 07.0007U.

US

General Monitors S5000 60


Appendix: General Certification Information

Cautions/Warnings
• READ AND UNDERSTAND ALL INSTRUCTIONS, WARNINGS AND CAUTIONS PRIOR TO
INSTALLATION OF ANY COMPONENTS OF THIS SYSTEM
• CAUTION – OPEN CIRCUIT BEFORE REMOVING COVER
• SEALING REQUIREMENTS -
• Aluminum S5000 transmitter and Junction Box:
A SEAL SHALL BE INSTALLED WITHIN 18 in (450 mm) OF THE ENCLOSURE
• Stainless Steel S5000 transmitter and Junction Box:
SEAL NOT REQUIRED
• WARNING FOR TRANSMITTER AND JUNCTION BOX (CEMENTED OR NON-CEMENTED
COVERS) – DO NOT OPEN WHEN ENERGIZED OR WHEN AN EXPLOSIVE ATMO-
SPHERE IS PRESENT
• WARNING FOR TRANSMITTER AND JUNCTION BOX (CEMENTED OR NON-CEMENTED
COVERS) – LIVE PARTS BEHIND COVER, OPEN CIRCUIT BEFORE REMOVING COVER
– DO NOT CONTACT
• WARNING FOR TRANSMITTER AND JUNCTION BOX (CEMENTED OR NON-CEMENTED
COVERS) – POTENTIAL ELECTROSTATIC CHARGING HAZARD. CLEAN EQUIPMENT
ONLY WITH DAMP CLOTH. FAILURE TO FOLLOW THIS WARNING CAN RESULT IN
SEVERE PERSONAL INJURY OR LOSS OF LIFE.
• WARNING – NO INGRESS PROTECTIONS CLAIMS ARE MADE FOR COMBUSTIBLE GAS
DETECTION PERFORMANCE
• WARNING – FOR SAFETY REASONS THIS EQUIPMENT MUST BE OPERATED AND
SERVICED BY QUALIFIED PERSONNEL ONLY. DO NOT OPERATE THIS EQUIPMENT
UNTIL AFTER THE INSTRUCTION MANUAL IS READ AND UNDERSTOOD FOR PROPER
INSTALLATION AND OPERATION
• CAUTION: HIGH OFF-SCALE READING MAY INDICATE EXPLOSIVE CONCENTRATION
• WARNING – WIRING TO OR FROM THIS DEVICE, WHICH ENTERS OR LEAVES THE
SYSTEM ENCLOSURE, MUST UTILIZE WIRING METHODS SUITABLE FOR CLASS I,
DIVISION 2/ZONE 2 HAZARDOUS LOCATIONS, AS APPROPRIATE FOR THE INSTALLA-
TION
• WARNING – EXPLOSION HAZARD. DO NOT CONNECT OR DISCONNECT WHEN ENER-
GIZED
• CAUTION: USE WIRING SUITABLE TO AT LEAST 99 °C WHEN WIRING THE S5000 MAIN
TRANSMITTER OR TO AT LEAST 75 °C WHEN WIRING THE S5000 JUNCTION BOX
Additional Certification Information
US
As part of this Approval, it was verified that optional communication functions of this gas detection
instrument while operating at the maximum transaction rate do not adversely affect the gas
detection operation and functions of the instrument. This Approval, however, does not include or
imply Approval of the “SafeSwap” feature, communications protocol or functions provided by the
software of this instrument or of the communications apparatus or software connected to this
instrument.
Use of some accessories are not covered under the performance approval. Contact
General Monitors for details.

General Monitors S5000 61


Appendix: HART Specific Information

10 Appendix: HART Specific Information


The S5000 Gas Monitor is available with an optional HART (Highway Addressable Remote
Transducer) output communications protocol. With this option, the S5000 complies with HART
Protocol Revision 7 and uses the 16-bit manufacturer and device codes. This document specifies
all the device specific features and documents HART Protocol implementation details (e.g., the
Engineering Unit Codes supported). These specifications assume the reader is somewhat familiar
with HART Protocol requirements and terminology.
This specification is a technical reference for HART-capable HOST Application Developers,
System Integrators and knowledgeable End Users. It also provides functional specifications (e.g.,
commands, enumerations and performance requirements) used during Field Device deployment,
maintenance, testing, and operations. It is recommended that the 4-20 mA output be the primary
gas monitoring signal. The HART signal can be the secondary method.

Manufacturer Name General Monitors Model Name S5000 Transmitter


HART ID Code 227 (0xE3) Device Type Code 115 (0x73)
HART Protocol Revision 7 Device Revision 1
Number of Device Variables 1/2
Physical Layers Supported FSK, Wireless
Physical Device Category Transmitter

Tab. 39 Device Identification

US

General Monitors S5000 62


For local MSA contacts, please visit us at MSAsafety.com

Because every life has a purpose...

You might also like