Injectomat 2000 MCM
Injectomat 2000 MCM
Injectomat 2000 MCM
Injectomat 2000
Technical Manual
Revision table
TM Injectomat 2000_Int : ref NT 0871 / from serial 17730124 till ... .
Injectomat 2000_IntTDM.fm 3
4 Preventive maintenance................................................................................. 39
4.1 Recommendations............................................................................... 39
4.2 Maintenance schedule......................................................................... 39
4.2.1 Preventive maintenance........................................................................................... 39
4.2.2 Sicherheitstechnische Kontrolle................................................................................ 39
4.3 Checks .................................................................................................. 41
4.3.1 Test access .............................................................................................................. 41
4.3.2 Visual check ............................................................................................................. 41
4.3.3 Running time and last servicing inspection date ...................................................... 42
4.3.4 Indicator lights check................................................................................................ 42
4.3.5 Keyboard check ....................................................................................................... 43
4.3.6 Checking the battery voltage.................................................................................... 44
4.3.7 Display of the last 10 alarms .................................................................................... 44
4.3.8 Total operation time check ....................................................................................... 46
4.3.9 TTL serial link test .................................................................................................... 46
4.3.10 RS 232 serial link check ........................................................................................... 47
4.3.11 Checking the software version. ................................................................................ 47
4.3.12 Checking the ADC.................................................................................................... 48
4.3.13 Checking the position sensor ................................................................................... 48
4.3.14 Buzzer test ............................................................................................................... 49
4.3.15 Display of the calibration values............................................................................... 49
4.3.16 Checking the syringe clamp ..................................................................................... 49
4.3.17 Checking the syringe group number ........................................................................ 50
4.3.18 Checking the list of syringes .................................................................................... 50
4.3.19 Checking the disengagement................................................................................... 50
4.3.20 Checking the flanges detection system.................................................................... 50
4.3.21 Checking the syringe head detection system........................................................... 51
4.3.22 Checking backpressure............................................................................................ 52
4.3.23 Checking the pre-alarm and end of infusion alarm .................................................. 52
4.3.24 Checking the linearity ............................................................................................... 53
4.3.25 Checking mains/battery operation............................................................................ 53
4.3.26 Battery autonomy test .............................................................................................. 54
4.3.27 Continuity test .......................................................................................................... 54
4.3.28 Stk-Protokoll ............................................................................................................. 55
4.4 Flow rate control .................................................................................. 57
4.4.1 Measurement with a computer ................................................................................. 57
4.4.2 Measurement with scales......................................................................................... 59
4.4.3 Measurement using a test tube................................................................................ 61
4.5 Cleaning and disinfecting ................................................................... 63
4.6 Storage.................................................................................................. 64
5 Diagnosis......................................................................................................... 65
5.1 Troubleshooting guide ....................................................................... 65
5.2 Error messages.................................................................................... 67
44 Injectomat 2000_IntTDM.fm
6 Intervention procedures ................................................................................ 71
N°1, Procedure: Display and central unit boards .................................... 73
N°2, Procedure: Syringe clamp ................................................................. 75
N°3, Procedure: Syringe detection system .............................................. 77
N°4, Procedure: Motor + Opto + Disk ........................................................ 83
N°5, Procedure: Dynamometer sensor ..................................................... 85
N°6, Procedure: Plunger advance control potentiometer ....................... 91
N°7, Procedure: Plunger cover and/or disengagement lever + anti-siphon arm 93
N°8, Procedure: Power supply board........................................................ 95
N°9, Procedure: Battery door and battery ................................................ 97
N°10, Procedure: Rear plug support ......................................................... 99
N°11, Procedure: Ribbon cable winding kit............................................ 101
N°12, Procedure: Syringe head detection plunger kit ........................... 105
N°13, Procedure: Centering ring kit ........................................................ 109
N°14, Procedure: Flex circuit and tube kit .............................................. 113
N°15, Procedure: Upper and lower cases ............................................... 119
7 Calibration..................................................................................................... 123
7.1 Calibration procedure ........................................................................ 123
7.1.1 Calibration access .................................................................................................. 123
7.1.2 EtA.4 Calibration of the 3 battery voltage levels................................................ 124
7.1.3 EtA.6 Calibration of the position sensor. ........................................................... 124
Injectomat 2000_IntTDM.fm 5
66 Injectomat 2000_IntTDM.fm
1 Overview
1.1 General
The Injectomat 2000 is a syringe pump intended for the infusion of intravenous agents at a
accurate, low flow rate. The wide choice of syringes, the use of the universally recognised
control symbols and the synoptic display of the alarms contribute to making the
Injectomat 2000 easy to use. The adjustable occlusion detection, the correct positioning
detection and the overall syringe protection system guarantee optimum safety.
Its technical characteristics, the flow range from 0.1 to 400 ml/hr (configured at
200 ml/hr) and its excellent accuracy (+1% on the device) contribute to making the
Injectomat 2000 the ideal instrument for medical services.
The Injectomat 2000 is equipped with a 16ko-EEPROM enabling the memorization of 896
events. These events are available for consultation through the ISCTRL maintenance
software, 4.1 and upper version.
Overview
Inj2_01_008b_en.fm 7
1.2 Overview diagram
Ext
12/15 V
15 W
Step by
DC-DC Motor
230 V Power supply ON / OFF step
converter driver motor
Battery
EPROM
128K x 8
RAM
8K x8 Motor speed
sensor
EEPROM
512
Syringue body
sensor
BUS Antisiphon
sensor
SPI
CPU
Watch
dog Displacement
transducer
Interface Opto
UART
bus interface
Nurse call
(option)
ADC
Occlusion
switch
Disengagement
switch
LCD LCD
RS 232 driver display
Keyboard
Flanges switch
LED LED
Master
Buzzer driver display
Overview
8 Inj2_01_008b_en.fm
1.3 Precautions to be taken before use
The symbol ! in the concise instrument instructions guide of the device recommends that
the operator’s guide should be read complelety in accordance with standard EN 60601-1.
Fresenius Kabi may in no case be held responsible for medical problems or any other
problems resulting from inadequate use of the equipment.
Refer to the User’s instructions for further details.
Inj2_01_008b_en.fm 9
1.5.4 Conformity and norms
10 Inj2_01_008b_en.fm
2 Description and operation
2.1 Physical description
Anti-siphon arm
The Injectomat 2000 is fitted with an upper case and a lower case.
! The upper case holds the syringe clamp and contains:
" A display board associated with the front control panel.
" A CPU board.
! The lower case contains:
" A power supply board and a battery.
" A mechanical base unit.
" A plunger unit.
Description and operation
Inj2_02_008b_en.fm 11
2.1.1 The display board and the front panel
The display board is mounted under the front control panel and is fitted with all the organs
required for man-machine interaction.
! Keyboard interface.
! Control lamps and overview diagrams.
! 7-segment display units.
J1
J3
J2
Pin Description
1 SEG1 display matrix Line 1
2 SEG2 display matrix Line 2
3 SEG3 display matrix Line 3
4 SEG4 display matrix Line 4
5 SEG5 display matrix Line 5
6 SEG6 display matrix Line 6
7 SEG7 display matrix Line 7
8 SEG8 display matrix Line 8
9 COL1 display matrix Column 1
10 COL2 display matrix Column 2
11 COL3 display matrix Column 3
12 FAIL LED control Fail
13 COL/DIG 9 LED type control "
14 LIG1 keyboard interface Line 1
15 LIG2 keyboard interface Line 2
16 LIG3 keyboard interface Line 3
17 LDSECT lighting control Mains LED
Description and operation
12 Inj2_02_008b_en.fm
J2 connector to keyboard
Pin Description
1 Column 1
2 Column 2
3 Column 3
4 Column 4
5 Column 5
6 Column 6
7 Line 1
8 Line 2
9 Line 3
10 Ton
11 Toff
12 Gnd power supply
Pin Description
1 Ton ON key
2 Toff OFF key
3 SI SPI bus
4 Clk SPI bus
5 CSLCD SPI bus
6 Buzz BUZZER control
7 Vbat power supply
8 Gnd power supply
Inj2_02_008b_en.fm 13
2.1.2 CPU board
The CPU board holds an 80C32 microprocessor. It is mounted and connected to the display
board through J4 and J5 connectors.
A ribbon cable connects this to a power supply board by means of a connector J3.
J2
J4
J3 J5
CPU board
J2 connector: to flanges detection switch and syringe detection opto-electronic sensor
Pin Description
1 Ground
2 Flanges contact
3 Opto anode diode +5V
4 Common points between cathode LED , opto 1 and opto 2 transistor emitters.
5 Opto 1 transistor collector
6 Opto 2 transistor collector
Pin Description
1 +5V
2 Gnd
3 +Vbat
4 Gnd
5 Phase A Motor control
6 Phase B Motor control
7 Phase C Motor control
8 Phase D Motor control
9 I signal Motor control
10 Boost signal Motor control
11 Sopt1 Opto rotation motor output
12 Sopt2 Opto anti-siphon
Description and operation
14 Inj2_02_008b_en.fm
J3 connector to power supply board
Pin Description
19 Ldsect Mains LED control
20 Cts Clear to send
21 Deb/off Disengagement signal
22 Rts request to send Request to send
23 Occ/off Occlusion signal
24 Buz Nurse call relay control
25 Eoc End Of adc conversion
26 Csadc Selection spi adc bus
27 Clk Clock spi adc bus
28 Si Data in spi adc bus
29 So Data out spi adc bus
30 Cdana Analogue sensor power supply control
31 Rx2 Receive data TTL
32 Tx2 Transmit data TTL
33 Txd1 Transmit data TTL
34 Rxd1 Receive data TTL
35 Ton ON key
36 Toff OFF key
37 +Vbat Power supply
38 Gnd
39 +5V
40 Gnd
Pin Description
1 Seg1 display matrix Line 1
2 Seg2 display matrix Line 2
3 Seg3 display matrix Line 3
4 Seg4 display matrix Line 4
5 Seg5 display matrix Line 5
6 Seg6 display matrix Line 6
7 Seg7 display matrix Line 7
8 Seg8 display matrix Line 8
9 Col1 display matrix and keyboard Column 1
10 Col2 display matrix and keyboard Column 2
11 Col3 display matrix and keyboard Column 3
12 Fail diode fail control
13 Rdcrt current reduction control
14 Lig1 keyboard interface Line 1
Description and operation
Inj2_02_008b_en.fm 15
J5 connector to display board
Pin Description
1 Ton ON key
2 Toff OFF key
3 Si spi bus
4 Clk spi bus
5 Cslcd spi bus
6 Buzz buzzer control
7 Vbat power supply
8 Gnd power supply
Description and operation
16 Inj2_02_008b_en.fm
2.1.3 The power supply board and the battery
The power supply board is mounted on the lower case. It allows to supply the electronic parts
using the network 230 V AC or the external 12 / 15 DC. It also charges the 1.1 or 1.3 Ah
battery.
J6
J7
J8
J9
J5 F2
J3
J4 J1
J2 F1
J1 connector to mains
Pin Description
1 Neutral
2 Phase
Pin Description
1 +5V controlled power supply
2 Gnd power supply
3 +Vbat power supply
4 Gnd power supply
5 Phase A Motor control
6 Phase B Motor control
7 Phase C Motor control
8 Phase D Motor control
9 I signal Motor control
10 BOOST signal
11 Sopt1 Opto rotation module output
12 Sopt2 Opto anti-siphon module output
13 N.U
14 Cdopt1 Opto rotation module control
Description and operation
Inj2_02_008b_en.fm 17
J2 connector to CPU board
Pin Description
22 RTS Request to send
23 OCC/OFF Occlusion signal active at 0
24 BUZ Nurse call relay control
25 EOC End of ADC conversion
26 CSADC Selection bus SPI ADC
27 CLK Clock bus SPI ADC
28 SI Data IN bus SPI ADC
29 SO Data out bus SPI ADC
30 CDANA Analogue sensor power supply control
31 RX2 receive data TTL
32 TX2 transmit data TTL
33 TXD1 transmit data TTL
34 RXD1 receive data TTL
35 Toff OFF key
36 Ton ON key
37 +Vbat Power supply
38 Gnd
39 +5V
40 Gnd
J3 connector to potentiometer
Pin Description
1 Vref
2 Centre point
3 Gnd
Pin Description
1 + battery
2 - battery
J5 connector to motor
Pin Description
1 +Vbay
2 +Vbat
3 Phase D
4 Phase C
5 Phase B
Description and operation
6 Phase A
7 Opto rotation anode diode /+5V
8 Opto rotation cathode diode
9 Opto rotation transistor collector
10 GND/ opto rotation transistor emitter
18 Inj2_02_008b_en.fm
J6 connector to RS232 and Master plugs
Pin Description
1 TX1
2 +5V
3 RX1
4 Gnd
5 Interface validation
6 Nurse call relay common point
7 Nurse call relay normally open
8 Nurse call relay normally closed
9 CD ON external on
10 CD OFF external off
11 I-OPTON Motor control output
12 I-SECT Mains led
13 +Vbat External power supply plug
14 RX2
15 TX2
16 Gnd
17 CTS
18 RTS
19 BUZ
20 NC
Pin Description
1 ± external power supply
2 ± external power supply
Pin Description
1 Not used
2 Micro-switch input/output
3 Micro-switch input/output
4 Not used
5 Opto anti-siphon cathode diode
6 Opto anti-siphon anode diode/+5V
7 Opto anti-siphon transistor collector
8 Disengagement micro-switch on
9 Disengagement micro-switch off
Description and operation
10 Gnd
Do not forget to dismount the ribbon cable holder on the power supply board before
extracting the mechanical assembly from the housing (risk of breaking the ribbon
cable).
Inj2_02_008b_en.fm 19
J9 connector, test points
Pin Description
1 GND
2 Position sensor output
3 Battery discharge control output
4 Amplified force sensor output
5 Cd coupler power supply 0-5V
6 Motor control opto output
7 Force and position sensor reference voltage
8 Piston head detection opto output
9 Control/APIN F
10 Control/APIN F
20 Inj2_02_008b_en.fm
2.2 Functional description
From a functional point of view, the Injectomat 2000 is composed of three sub-assemblies:
! A syringe position control and maintenance assembly.
! A motorisation assembly.
! An external connection assembly.
Inj2_02_008b_en.fm 21
Description and operation
22
Inj2_02_008b_en.fm
3 Description of the menus
3.1 Configuration menu of the current operation parameters
The configuration menu enables users to adapt the Injectomat 2000 to the specific needs of
each department. It provides access to the menus allowing for customisation of the
parameters associated with current operation modes.
Fresenius Kabi recommends users to implement the selected configuration procedures in
the presence of a member of its qualified personnel or a member of the technical department.
It is possible to exit the configuration mode at any time by pressing the OFF key.
Inj2_03.2_008b_en.fm 23
3.1.1 Menu access
Useful keys
Key Function
ON, is used to switch the machine ON.
OFF, is used to switch the machine off when pressed for over three
seconds.
SILENCE ALARM, is used to access the configuration mode of the
current operation parameters.
The selection keys allow to scroll the figures and letters on the tenths,
units, tens segments etc.
24 Inj2_03.2_008b_en.fm
3.1.2 PAr1, configuration of the memorisation type
This configuration enables users to choose whether or not to memorise the infusion flow rate
when the Injectomat 2000 is shut down.
! PAr1, press "CONFIRM"
" If MEM, the flow rate is memorised when the Injectomat
2000 is shut down, this value will be displayed by default
when the machine is next switched on.
" If noME, the flow rate is not memorised, the default
value is 00.0 each time the machine is switched on.
# Select the memorisation type using the selection
keys.
When mode SEL3 is selected, and if there is a choice of more than one syringe, the
Injectomat 2000 automatically moves onto the configuration of the list of syringes that
may be selected PAr4 when the machine is next switched on.
Description of the menus
Inj2_03.2_008b_en.fm 25
3.1.4 PAr3, configuration of the maximum flow rate that may be selec-
ted on the keyboard
This configuration enables users to choose the maximum flow rate that may be selected
using the keyboard for each type of syringe.
Syringe type Min. flow rate (ml/hr) Max. flow rate (ml/hr)
50/60 cc 0,1 400
20 cc 0,1 275
26 Inj2_03.2_008b_en.fm
Details of the syringe are displayed
when the "tenths" keys are pressed (see chapter 3.1.17
"Typical syringe/details correspondence table").
Inj2_03.2_008b_en.fm 27
" By validating once again, the configuration is memorised
and it is possible to select another configuration .
28 Inj2_03.2_008b_en.fm
3.1.10 PArb, configuration of the frequency of preventive checks
This configuration enables users to select the maintenance frequency which lies between 1
and 9999 hours.
! PArb, press "CONFIRM"
" The current value is displayed.
It is possible to exit the configuration mode at any time by pressing the "OFF"
key.
Inj2_03.2_008b_en.fm 29
3.1.12 PArd, configuration of the Flanges detection mode
This configuration enables users to choose whether or not to activate the syringe flanges
position check.
! PArd, press "CONFIRM"
" If AiLE, detection activated, inappropriate positioning
of the flanges is signalled by means of an alarm.
" If noAL, no detection, the inappropriate positioning of
the flanges is not checked.
If the "not memorised" mode is selected, the bolus default value must be defined.
30 Inj2_03.2_008b_en.fm
" The 20 cc LED lights up and the bolus flow is displayed.
# Using the keys, enter the bolus value to be defined by
default for a 20 cc syringe.
Inj2_03.2_008b_en.fm 31
3.1.15 PArJ, configuration of the mains disconnection signal
This configuration enables users to choose whether or not to activate the power cut detection
beep.
! PArJ, press "CONFIRM"
" If SEC.t, detection activated, a power cut is signalled
by a beep.
" If noS;E, no detection.
" Make your choice using the keys.
H : hour m: minute
" Make your choice using the keys.
32 Inj2_03.2_008b_en.fm
3.1.17 Typical syringe/details correspondence table
Inj2_03.2_008b_en.fm 33
Description of the menus
34
Inj2_03.2_008b_en.fm
Configuration, calibration and inspection
Inj2_03.3_002a_en.fm 35
Configuration, calibration and inspection
36
Inj2_03.3_002a_en.fm
Configuration, calibration and inspection
The Service tests may also be performed more easily and more quickly using a PC
with installed maintenance software (consult our After Sales Service).
The Service test menu enables users to perform a series of 16 tests or checks:
! tSt1: Display of the running time and the maintenance date.
! tSt2: Indicator lights test.
! tSt3: Keyboard test.
! tSt4: Display of the battery voltage.
! tSt5: Display of the codes of the last 10 alarms.
! tSt6: Display of the total running time.
! tSt7: TTL serial link test.
! tSt8: RS 232 serial link test.
! tStA: Display of the software version.
! tStb: Display of the analogue input.
! tStC: Display of the plunger position.
! tStd: Buzzer test.
! tStE: Display of the calibration values.
! tStf: Display of the syringe type.
! tStG: Display of the syringe group.
! tStH: Display of the list of syringes.
Inj2_03.4_002a_en.fm 37
Configuration, calibration and inspection
38
Inj2_03.4_002a_en.fm
4 Preventive maintenance
4.1 Recommendations
The Injectomat 2000 syringe pump can only be checked, serviced or repaired by
Fresenius Kabi or by a certified and approved maintenance service. Any abnormal
functioning of the pump must be brought to the attention of your in-house qualified technical
personnel or our After-Sales Service.
Should you need to return the pump to our After-Sales Service, it should be cleaned,
disinfected and very carefully packaged, preferably in its original packaging, before being
sent.
For all information concerning the repair and use of the pump, kindly contact our After-Sales
Service or our Sales department.
Fresenius Kabi is not liable for loss or damage to the pump during its transport to our After-
Sales Service.
Important:
If these maintenance procedures are not observed, the pump’s correct operation will
be impaired.
The STK (not included in the guarantee) consists of different inspection procedures as
defined in the pump’s Technical Manual. Only a qualified technician may perform the STK
which must be performed using Fresenius Kabi software.
For more information, kindly contact our After-Sales Service.
Inj2_04.1&2_001b_en.fm 39
Preventive maintenance
40 Inj2_04.1&2_001b_en.fm
Preventive maintenance
4.3 Checks
A checking certificate is available at the end of this chapter.
To ensure the check procedure is carried out efficiently, recharge the battery
beforehand (16 hours).
Keyboard description.
Key Function
ON, to switch the machine "On".
OFF, to switch the machine "Off" when pressed for over three seconds.
SILENCE ALARM, to access the test mode.
The selection keys allow to scroll the figures and letters on the tenths,
units, tens segments etc.
Inj2_04.3_002b_en.fm 41
4.3.3 Running time and last servicing inspection date
This test allows for ! tSt.1, press the "CONFIRM" key.
" If xxx.H: number of hours of use, 999 hours max.
display of the
Injectomat 2000
running time since " If xxx.J: number of days of use, 999 days max.
its last servicing " If xxx.M : number of months, 999 months max.
inspection. It also (average duration of one month considered as 30 days).
allows for display " If CtrL is displayed alternately, the result exceeds the
and modification of
the "last" servicing max. servicing frequency memorised (see "PArb,
inspection date. Configuration of the frequency of preventive checks"):
When the servicing carry out preventive maintenance procedures.
inspection date is # By keeping the "tenths" keys pressed down, the
modified, the number of times the equipment has been switched on
running time is is displayed.
reset.
! Press "CONFIRM" again to display the date of the last
servicing inspection.
" xx.d, day of the servicing inspection date.
" Validate once again,
# xx.m, month of the servicing inspection date.
" By validating once again,
# xxx.x, year of the servicing inspection date.
Each time this information is read, the month, day and year of
the servicing inspection date may be modified by using the
tens and units keys. This date will be stored in the EEPROM
and the running time will be reset.
The test may be stopped at any time by pressing the "STOP" key, and a
different test may be selected.
The test may be stopped at any time by pressing the "STOP" key, and a
different test may be selected.
42 Inj2_04.3_002b_en.fm
4.3.5 Keyboard check
This test allows to ! tSt.3, press "CONFIRM".
check that all " The message CLAV is displayed.
keyboard keys
" Keep each key pressed down, one by one,
function correctly.
# Check the name of the key displayed on the display
unit.
StoP STOP
VAL CONFIRM
boLu BOLUS
S-.. + tens
S.-. + units
S..- + tenths
S_.. - tens
S._. - units
S_.. - tenths
If two or more keys are pressed simultaneously, the display unit shows "Err"
and an audio alarm sounds "beep!beep!beep!" indicating an error.
The "OFF" key is not included in the keyboard check.
The test may be stopped at any time by pressing the "CONFIRM" key for
over 3 seconds, and a different test may be selected.
Preventive maintenance
Inj2_04.3_002b_en.fm 43
4.3.6 Checking the battery voltage
This test enables ! tSt.4, press "CONFIRM".
users to display " The voltage is displayed in Volts.
the battery voltage
in Volts and tenths
" By validating once again, a different test may be selected.
of a Volt.
The test may be stopped at any time by pressing the "STOP" key, and a
different test may be selected.
The test may be stopped at any time by pressing the "STOP" key, and a
different test may be selected.
44 Inj2_04.3_002b_en.fm
Error codes description
Error Description
10 Internal RAM anomaly
20 External RAM anomaly
30 Eprom anomaly
40 Access (E2prom) Eeprom anomaly
50 Adc anomaly
60 Syringues parameters anomaly
70 Motor frequency anomaly
80 Keyboard matrix (key press non functioning) fault
Electromagnetic disturbance exeeding standards
01 Rotation (absence top turn) error
E32 Error concerning segment advance control
E52 Error concerning advance during compensation for play.
E72 Error concerning advance over the whole length.
E82 Error concerning advance flow rate
E03 Error concerning communication
14 Period frequency anomaly
24 Rotation direction anomaly
34 Flow rate / motor period calculation anomaly
44 Incoherence quartz µC principal and µC secondary
55 µC secondary access fault
16 Date and time in µC secondary verification
56 Association condition output/step number
66 LCD Display time-out
18 Maximum infused volume reached
28 Language file incompatible
CFPC Anomaly duringthe self set
COM Communication error
CTRL Preventive maintenance request
Inj2_04.3_002b_en.fm 45
4.3.8 Total operation time check
This test enables ! tSt.6, press "CONFIRM".
" If xxx.H: number of hours of use, 999 hours max.
users to display
the total running
time of the " If xxx.J: number of days of use, 999 days max.
The test may be stopped at any time by pressing the "STOP" key, and a
different test may be selected.
The test may be stopped at any time by pressing the "STOP" key, and a
different test may be selected.
46 Inj2_04.3_002b_en.fm
4.3.10 RS 232 serial link check
This test enables To perform this test, use plug on which lines Rx and Tx, RTS and CTS are short-circuited
users to check the (2 and 3, 7 and 8).
efficiency of the
RS 232 (2691) ! tSt8, press "CONFIRM".
serial link. " NoUA is displayed if the UART is missing, in which
case the test cannot be carried out.
" RTS andCTS line operation test.
# If the test is unsuccessful, NoRC is displayed, in
which case the test cannot be continued.
" Lr is displayed.
# If LrEr is displayed, the test is not successful,
otherwise, the test is successful.
The test may be stopped at any time by pressing the "STOP" key, and a
different test may be selected.
The test may be stopped at any time by pressing the "STOP" key, and a
different test may be selected.
Preventive maintenance
Inj2_04.3_002b_en.fm 47
4.3.12 Checking the ADC
Run this test to ! tSt.b, press "CONFIRM".
read the result of " The result displayed is the first of the five analogue inputs
the conversion of
(from 0 to 4), example 318.0.
the five analogue
inputs and three " The second type of result displayed corresponds to the
test inputs of the converter test according to the channel number,
converter. xxx.Lis the 0 test input, xxx.M is the mid-scale
test input, and xxx.H is the ADC full scale test input.
Input Name
0 Battery voltage
1 N.U.
2 Force sensor
3 N.U.
4 Potentiometer displacement
L Converter zero test, between 0000 and 0004 if correct
M Converter mid-scale test, between 01FB and 204 if correct
H Converter full-scale test, between 03B and 3FF if correct
If the value is out of limits, recalibrate the position sensor (see "EtA.6 Position sensor
calibration.").
48 Inj2_04.3_002b_en.fm
4.3.14 Buzzer test
Run this test to ! tSt.d, press "CONFIRM".
check that the
" The buzzer sounds continuously and the bEEp
buzzer is working.
message is displayed.
" By validating once again, a different test may be selected.
" Press one of the unit or tenth keys to move onto another
value.
Display Name
bat.1 Alarm and pre-alarm battery voltage 6.3 V
Capacities which are non-existent or non-configured in the EEPROM are displayed in the
form of ---.-.
Inj2_04.3_002b_en.fm 49
4.3.17 Checking the syringe group number
Run this test to ! tSt.G, press "CONFIRM".
determine for " The syringe group number configured in EEPROM is
which list of
displayed, example G01.
syringes the
instrument has
been configured. " By validating once again, a different test may be selected.
! Press "ON".
" Lift the disengagement lever.
# Check for presence of the mechanical disengagement
alarm (red Led at the end of the syringe diagram).
" Fit the device with the 50 cc syringe, ensuring the flanges
and plunger are in position.
" Release the disengagement lever.
# Ensure there is no mechanical lever release alarm.
# Check plunger locking.
! Press "ON".
" Fit the device with the 50 cc syringe, ensuring the flanges
and plunger are in position.
Preventive maintenance
50 Inj2_04.3_002b_en.fm
4.3.21 Checking the syringe head detection system
This check must be carried out for every syringe provided with the instrument.
Inj2_04.3_002b_en.fm 51
4.3.22 Checking backpressure
To carry out this operation, exit the test mode and press "OFF",
! Press "ON".
" Fit the instrument with the 50 cc syringe and lock it into
position using the syringe clamp, ensuring the flanges and
plunger are correctly positioned.
" Place the manometer (or any other pressure
measurement device) at the syringe outlet.
" Select a 50 ml "B-D PLASTIPAK" syringe by pressing
"CONFIRM".
" Check the position of the backpressure adjustment
button.
" Select a maximum flow rate and initiate the infusion by
pressing the "CONFIRM" key.
# Ensure there is no acoustic and visual alarm
(backpressure Led off).
# Check that the infusion Leds are flashing.
# Ensure the alarm is triggered for a value of
0,7 bar + 0,1 bar.
" Stop the infusion cycle by pressing the "STOP
INFUSION" key.
" Repeat the test setting the backpressure level to the
minimum and maximum levels.
# Min. backpressure = 0,4 bar + 0,1 bar.
# Max. backpressure = 1.2 bar + 0.2 bar.
! Press "ON".
" Fit the device with the syringe, ensuring the flanges and
plunger are in position.
" Ensure the syringe is detected appropriately eg.
50c.c.
" Select a "B-D PLASTIPAK" syringe filled to 20 ml.
" Select a flow rate of 120 ml/h.
# For normal flow rates, the pre-alarm is activated 5
minutes before end of infusion.
# Flow rate example: > 50 ml/h, the pre-alarm is
activated when the remaining volume equals 10% of
the total syringe capacity.
# Ensure the end of infusion pre-alarm is present.
" Press the "SILENCE ALARM".
# The acoustic alarm is silenced and the visual
Preventive maintenance
signal is maintained.
" Measure the "hard height" at "end of infusion".
# 18,5 < x < 19,5.
Top view
52 Inj2_04.3_002b_en.fm
For accurate checking of the "hard height", do not move the plunger when measuring.
If the "hard height" reading is out of limits, recalibrate the position sensor (see "EtA.6
Position sensor calibration.").
! Press "ON".
" Fit the device with the "B-D PLASTIPAK" 50 ml syringe,
ensuring the flanges and plunger are in starting position.
" Measure the distance X1 in mm.
For accurate checking of the linearity do not move the plunger when measuring.
Inj2_04.3_002b_en.fm 53
4.3.26 Battery autonomy test
! Recharge the battery for 16 hours.
54 Inj2_04.3_002b_en.fm
Preventive maintenance
Preventive maintenance
In order to purchase the ISDébit flow rate control software, contact the Fresenius Kabi
customer service.
The test procedure defined below must be carried out using a 50 ml or 20 ml syringe.
Equipment required
! Scales coupled to a microcomputer:
Flow rate value Scales sensitivity
x < 5 ml/hr 1/10000th
5 ml/hr < x < 30ml/hr 1/1000th
x > 30 ml/hr 1/100th
! Multi scales acquisition programme.
! Test tube or beaker with 1 ml graduating.
! Liquid: distilled water and oil.
! Luer lock type plastic syringe (50 or 20 ml).
! Catheter extension with Luer Lock (length 100 cm, inside diameter 2.5 mm).
! Needle:
Flow rate value Needle type
x < 30ml/hr G26
x > 30 ml/hr G18 or G21
Installation
! According to the installation drawings shown below.
1 cm
Preventive maintenance
Cord
Anti vibration measurement table
RS232
Inj2_04.4_001b_en.fm 57
! Fill the syringe with 50 ml of distilled water.
" Prime to eliminate any air bubbles.
! Secure the female Luer Lock end piece of the catheter
extension onto the syringe and the male Luer lock end piece
onto the needle.
! Fit the syringe onto the device.
! Fill the test tube ensuring the needle is dipped in the liquid
(> 1 cm).
" Add several drops of oil to create a greasy film on the
surface of the liquid. This way the user will avoid any
measurement error due to evaporation of the liquid.
! Place the test tube in the centre of the scales platform.
! Place the needle inside the test tube.
The infusion line (needle/catheter extension) must not rest on the scales/test tube
assembly.
Operating mode
The software works following the operating mode described in the EN-60-601-2-24
standard for infusion pumps.
58 Inj2_04.4_001b_en.fm
4.4.2 Measurement with scales
Equipment required
! Stop clock.
! Scales:
Flow rate value Scales sensitivity
x < 5 ml/hr 1/10000th
5 ml/hr < x < 30ml/hr 1/1000th
x > 30 ml/hr 1/100th
! Test tube or beaker with 1 ml graduating.
! Liquid: distilled water and oil.
! Luer lock type plastic syringe (50 or 20 ml).
! Catheter extension with Luer Lock (length 100 cm, inside diameter 2.5 mm).
! Needle:
Flow rate value Needle type
x < 30ml/hr G26
x > 30 ml/hr G18 or G21
Installation
! According to the installation drawings shown below.
1 cm
(> 1 cm).
! Add several drops of oil to create a greasy film on the
surface of the liquid. This way the user will avoid any
measurement error due to evaporation of the liquid.
! Place the test tube in the centre of the scales platform.
! Place the needle inside the test tube.
Inj2_04.4_001b_en.fm 59
The infusion line (needle/catheter extension) must not rest on the scales/test tube
assembly.
Operating mode
! Select a flow rate.
For low flow rates (< 5 ml/hr), validate and wait for the infusion to stabilise for 1 hour.
For higher flow rates, wait for 10 to 30 minutes after infusion.
1 gram = 1 ml.
! The error percentage can be calculated from this difference :
( Real value – Design value )
--------------------------------------------------------------------------- × 100 = Error percentage
Design value
Preventive maintenance
60 Inj2_04.4_001b_en.fm
4.4.3 Measurement using a test tube
Equipment required
! Stop clock.
! Test tube or beaker with 1 ml graduating.
! Liquid: distilled water and oil.
! Luer lock type plastic syringe (50 or 20 ml).
! Catheter extension with Luer Lock (length 100 cm, inside diameter 2,5 cm).
! Needle:
Flow rate value Needle type
x < 30ml/hr G26
x > 30 ml/hr G18 or G21
Installation
! According to the installation drawings shown below.
1 cm
Inj2_04.4_001b_en.fm 61
Operating mode
! Select a flow rate.
For low flow rates (< 5 ml/hr), validate and wait for the infusion to stabilise for 1 hour.
For higher flow rates, wait for 10 to 30 minutes after infusion.
62 Inj2_04.4_001b_en.fm
Preventive maintenance
Do not use:
TRICHLOROETHYLENE-DICHLOROETHYLENE
"
AMMONIA.
"
" AMMONIUM CHLORIDE.
" CHLORINE AND AROMATIC HYDROCARBON.
" ETHYLENE DICHLORIDE-METHYLENE.
" CHLORIDE-CETONE.
These aggressive agents could damage the plastic parts and cause device malfunction.
Take car also with ALCOHOL based SPRAYS (20% -40% alcohol). They lead to
tarnishing and small cracks in the plastic, and do not provide the neceeeary
cleaning prior to disinfection.
Please contact the appropriate service, responsible for cleaning and disinfecting products, in
your establishment for further details. Preventive maintenance
Inj2_04.5&6_001b_en.fm 63
4.6 Storage
The device should be stored in a dry, cool place. In case of prolonged storage, the battery
should be disconnected via the battery access flap situated underneath the device. This
should be done by a qualified technician
! Storage conditions and carrying:
" Temperature: -10°C to 60°C / 14°F to 140°F.
" Atmospheric pressure: 500 hPa to 1060 hPa.
" Humidity: 10% to 90%, no condensation.
! Use conditions:
" Temperature: 5°C to 40°C / 41°F to 104°F.
" Atmospheric pressure: 700 hPa to 1060 hPa.
" Humidity: 20% to 90%, no condensation.
Fully recharge the battery before using the equipment to avoid any risks caused by micro
power cuts in the mains supply and to ensure maximum autonomy.
Preventive maintenance
64 Inj2_04.5&6_001b_en.fm
5 Diagnosis
5.1 Troubleshooting guide
Inj2_05.1_001b_en.fm 65
Problems Causes Solutions
The power supply LED doesn’t ! Power supply fuse is out of order. ! Replace the power supply
lit. fuse.
Battery alarm while it has been ! Voltage regulator MAX 652 is ! Check the battery charge
completely charged. defective (charging voltage greater than voltage and replace the regulator
10 V or lower than 5 V). if necessary.
! Wrong calibration of the battery. ! Recalibrate the 3 battery
voltage levels with IS contol
software.
! Battery is totally discharged. ! Replace the battery.
Battery alarm while the Pilot is ! The battery voltage calibration was ! Recalibrate the 3 battery
connected to the mains power made while the Pilot was connected to voltage levels with IS contol
supply. the mains power supply. software.
! Recalibrate the 3 battery
voltage levels with a stabilisad
power supply adjusted at 9 volts
in order to set the device to
CFPC or ETAL modes.
! Replace the CPU board and
carry on a complete
configuration of the device.
The Pilot turns "ON" or turns ! Keyboard is defective. ! Check and replace the
"OFF" alone. keyboard if necessary.
! Power supply board is defective. ! Check that there was no liquid
infiltration under the power
supply board relay.
! Check the relay and replace it
if necessary.
! Check the relay transistors
control and replace them if
necessary.
! Replace the power supply
board.
Some keys of the keyboard keys ! Keyboard is defective. ! Check the keyboard (test3)
don’t work. Replace the keyboard if
necessary.
After a fall ! Mechanical elements damaged. ! Check that the centering ring,
the mechanical assembly and
the cases are intact.
Diagnosis
66 Inj2_05.1_001b_en.fm
Diagnosis
Inj2_05.2_001b_en.fm 67
Error code Description Recommended action
Er32 ! Displacement error on a ! Check the position sensor attachement.
short distance. ! Check the position sensor calibration.
! Check the ADC.
! Check the position sensor and replace if
necessary.
! Reconfigure the Injectomat 2000 with
IS Control sofware.
Er34 ! Flow rate / Motor period ! Reconfigure the Pilot with IS Control
calculation anomaly software.
! Replace the CPU board and carry on a
complete configuration of the device.
CFPC ! Anomaly during the self- ! Reconfigure the Injectomat 2000 with
test IS Control software.
! Replace the CPU board and carry on a
complete configuration of the device.
Master.
68 Inj2_05.2_001b_en.fm
Error code Description Recommended action
CTRL ! Preventive maintenance ! Perform the regular servicing checks
request and update the maintenance date (test1)
! This error doesn’t jam the Injectomat
2000. Press confirm to leave
*: When the device is switched off and the EEPROM is rewritten, the checksum is rewritten
in the memory to save the settings.
If the hardware downtime is longer than the software downtime, the device is switched off
before the EEPROM is completely written: Non-conforming checksum.
Er(-)0 or CFPc: When the device goes into CFPc, you must reconfigure it: WATCH DOG
faulty.
Diagnosis
Inj2_05.2_001b_en.fm 69
Diagnosis
70 Inj2_05.2_001b_en.fm
6 Intervention procedures
This chapter details all the dismantling and re-assembling procedures.
Intervention procedures
Inj2_06.0_001a_en.fm 71
Intervention procedures
72
Inj2_06.0_001a_en.fm
Intervention procedures
Safety:
For safety reasons, the technician should not carry out any maintenance when the device is
connected to the 230 V mains supply voltage.
Disconnect the mains power supply cable.
Equipment required:
! 1 Posidriv Z1 screwdriver.
! 1 antistatic wriststrap.
Maintenance level:
Level 3, specialised technician (see documentation on biomedical training).
Procedure:
Access
! Turn the Injectomat over onto the upper case.
! Unscrew the 5 Phillips head screws (ref. 1) located at the bottom of the lower case, which
link this to the upper case.
! Keep the Injectomat assembled and turn it over onto the lower case.
! Remove the upper case making sure the CPU board flat jumper is not accidentally
removed in the process.
! Disconnect the CPU board flat jumper.
Hands must not come into contact with the CPU boards.
1
1
1
Intervention procedures
Inj2_06.0_0010b_en.fm 73
Dismounting
2
2
3 2
Remounting
When mounting the display board, it is important to reduce the torque in the plastic inserts
so as to avoid causing damage to these.When fitting the upper case, ensure the joint is
perfectly positioned in its slot after reassembly.
! When a remplacement of the CPU board by a new one, a complete configuration is
necessary using IS control maintenance software.
! Perform the regular servicing checks (see "Quality Control Certificate").
74 Inj2_06.0_0010b_en.fm
Intervention procedures
Safety:
For safety reasons, the technician should not carry out any maintenance when the device is
connected to the 230 V mains supply voltage.
Disconnect the mains power supply cable.
Equipment required:
! 1 Posidriv Z1 screwdriver.
! 1 Posidriv Z1 flat screwdriver.
Maintenance level:
Level 2, specialised technician (see documentation on biomedical training).
Procedure:
Access
! Turn the Injectomat over onto the upper case.
! Unscrew the 5 Phillips head screws (ref. 1) located at the bottom of the lower case, which
link this to the upper case.
! Keep the Injectomat assembled and turn it over onto the lower case.
! Remove the upper case.
! Disconnect the CPU board flat cable.
Hands must not come into contact with the CPU boards.
1
1
1
Intervention procedures
Inj2_06.0_0020b_en.fm 75
Dismounting
! Remove the spring retaining ring (ref. 2) located on the syringe clamp shaft, which holds
this to the CPU support.
! Remove the obturator and the spring (ref. 3).
! Remove the syringe clamp.
Reassembly
Carry out the same procedures in reverse to remount the unit.
When fitting the upper case, ensure the joint is perfectly positioned in its slot after
reassembly.
76 Inj2_06.0_0020b_en.fm
Intervention procedures
Safety:
For safety reasons, the technician should not carry out any maintenance when the device is
connected to the 230 V mains supply voltage.
Disconnect the mains power supply cable.
Equipment required:
! 1 Posidriv Z1 screwdriver.
! 1 antistatic wriststrap.
! 1 soldering iron
! "RADIEL Sn60Pb RI 1" welding wire (cleaning not required for rewelding) or equivalent.
! Silicone ref. 161249.
Maintenance level:
Level 2, specialised technician (see documentation on biomedical training).
Access
! Turn the Injectomat over onto the upper case.
! Unscrew the 5 Phillips head screws (ref. 1) located at the bottom of the lower case, which
link this to the upper case.
! Keep the Injectomat assembled and turn it over onto the lower case.
! Remove the upper case.
! Disconnect the CPU board flat cable.
Hands must not come into contact with the CPU boards
1
1
1
Intervention procedures
Inj2_06.0_0030b_en.fm 77
Dismounting
! Disconnect the connector (ref. 4).
! Unscrew the Phillips head screw (ref. 2) which holds the conductor to the upper case.
! Unscrew the 2 Phillips head screws (ref. 3) which hold the switch support in position.
! Remove the switch assembly.
3
Welding and unwelding operations are carried out using a soldering iron fitted with a tip in
good condition, constantly tapered and clean kept. The temperature of the iron should be
between 315°C and 340°C.
78 Inj2_06.0_0030b_en.fm
Reassembly
! Place the new switch on the support.
! Bend the mounting lugs towards the inside in order to lock thel into position.
! Weld the wires of the new switch:
" Tin-plate the 2 wires to be soldered and the switch support.
" Place the wires under the bendt lugs.
" Place the welding iron tip on the welding surface.
" Place the welding wire on the pin surface of the switch to be welded.
" Remove the welding wire and then the wilding iron tip.
# Check that the weling has been carried out correctly. It bshould not form a ball,
but should lie against the lengh of the wire.
! Screw the 2 Phillips head screws (ref. 3) which hold the "switch support" assembly in
position.
When mounting the "switch support", it is important to reduce the torque in the plastic
inserts so as to avoid causing damage to these.
When fitting the upper case, ensure the seal is perfectly positioned in its slot after
reassembly.
Intervention procedures
Inj2_06.0_0030b_en.fm 79
Opto replacement procedure:
Access
! Turn the Injectomat over onto the upper case.
! Unscrew the 5 Phillips head screws (ref. 1) located at the bottom of the lower case, which
link this to the upper case.
! Keep the Injectomat assembled and turn it over onto the lower case.
! Remove the upper case.
! Disconnect the CPU board flat cable.
Hands must not come into contact with the CPU boards.
1
1
Dismounting
! Disconnect the connector (ref. 2).
! Unscrew the Phillips head screw (ref. 3) which holds the conductor to the upper case.
! Unscrew the 2 Phillips head screws (ref. 5) which hold the opto PCB in position.
! Remove the opto assembly.
2
Intervention procedures
3
Green
Black
White
5 Red
80 Inj2_06.0_0030b_en.fm
Reassembly
! Place the new opto assembly on the PCB support.
! Screw the 2 Phillips head screws (ref. 5) which hold the "PCB and opto" assembly in
position.
When mounting the "PCB and opto", it is important to reduce the torque in the plastic
inserts so as to avoid causing damage to these.
When fitting the upper case, ensure the joint is perfectly positioned in its slot after
reassembly.
Intervention procedures
Inj2_06.0_0030b_en.fm 81
Intervention procedures
82 Inj2_06.0_0030b_en.fm
Intervention procedures
Safety:
For safety reasons, the technician should not carry out any maintenance when the device is
connected to the 230 V mains supply voltage.
Disconnect the mains power supply cable.
Equipment required:
! 1 Posidriv Z1 screwdriver.
! 1 antistatic wriststrap.
Maintenance level:
Level 2, specialised technician (see documentation on biomedical training).
Procedure:
Access
! Turn the Injectomat over onto the upper case.
! Unscrew the 5 Phillips head screws (ref. 1) located at the bottom of the lower case, which
link this to the upper case.
! Keep the Injectomat assembled and turn it over onto the lower case.
! Remove the upper case making sure the CPU board flat jumper is not accidentally
removed in the process.
! Disconnect the CPU board flat jumper.
Hands must not come into contact with the CPU boards.
1
1
1
Intervention procedures
Inj2_06.0_0040b_en.fm 83
Dismounting
! Press the disengagement lever towards the back of the Injectomat as far as possible.
! Maintain this position and slide the whole unit to the back left.
! Remove the plunger guide.
! Disconnect the black connectors (ref. 2).
! Unscrew the flat jumper mounting lug (ref. 3) without removing it.
! Remove the mechanical plunger unit from its slot.
! Disconnect the blue connector from the flat jumper (ref. 4).
! Remove the mechanical plunger unit.
2
3
Reassembly
Intervention procedures
When fitting the upper case, ensure the joint is perfectly positioned in its slot after
reassembly.
84 Inj2_06.0_0040b_en.fm
Intervention procedures
Safety:
For safety reasons, the technician should not carry out any maintenance when the device is
connected to the 230 V mains supply voltage.
Disconnect the mains power supply cable.
Equipment required:
! 1 Posidriv Z1 screwdriver.
! 1 hexagon socket key (2.5).
! 1 soldering iron.
! "RADIEL Sn60Pb RI 1" welding wire (cleaning not required for re-welding) or equivalent.
Maintenance level:
Level 2, specialised technician (see documentation on biomedical training).
Procedure:
Access
! Turn the Injectomat over onto the upper case.
! Unscrew the 5 Phillips head screws (ref. 1) located at the bottom of the lower case, which
link this to the upper case.
! Keep the Injectomat assembled and turn it over onto the lower case.
! Remove the upper case making sure the CPU board flat jumper is not accidentally
removed in the process.
Hands must not come into contact with the CPU boards.
1
1
1
Intervention procedures
Inj2_06.0_0051b_en.fm 85
Dismounting
! Press the disengagement lever towards the back of the Injectomat as far as possible.
! Maintain this position and slide the whole unit to the back right.
! Remove the plunger guide.
! Unscrew the 3 Phillips head screws (ref. 2) located at the bottom of the plunger end shield
which links this to the plunger cover.
! Remove the plunger end shield.
6 5 4
3
Intervention procedures
86 Inj2_06.0_0051b_en.fm
! Unscrew the 2 Phillips head screws (ref. 7 and 8) which link the plunger support to the
plunger cover.
! Unscrew and remove the Phillips head screw and the washer (ref. 9) which attach the anti-
siphon cam to the anti-siphon arm.
! Remove the cam/arm assembly.
! Remove the plunger cover fitted with the bonding pad.
7
9
! Remove the retaining ring that holds the anti-siphon opto in place (ref. 10).
! Unscrew the 6 hexagon socket screws (ref. 11).
11
10
Intervention procedures
Inj2_06.0_0051b_en.fm 87
! Remove the backpressure micro support/spring leaf (ref. 12).
! Pull off the ribbon cable from the double-sided board
12
Welding and unwelding operations are carried out using a soldering iron fitted with a tip in
good condition, constantly tapered and kept clean. The temperature of the iron should be
between 315°C and 340°C.
! Unweld the sensor wires on the flex circuit.
When dismounting the force sensor, ensure the welding pellets are not damaged.
88 Inj2_06.0_0051b_en.fm
Reassembly
! Weld the ribbon cable to the new pressure switch, ensuring the mounting procedure is
carried out in order:
" Taper the 3 welding points.
" Weld the 3 points.
" Check that the welding has been carried out correctly. It should not form a ball, but
should lie against the length of the component snubbers.
! Repeat the operation for the anti-siphon opto (4 welding points).
! Stick the bottom side of the ribbon cable to the double-sided board.
Anti-siphon opto
w
Backpressure switch
Lower side
(connection specific for device equiped with a force sensor)
stuck to the double-sided board on the plunger support
! Impregnate with weak loctite and screw the 6 hexagon socket screws (ref. 11), which hold
the spring leaf/backpressure micro support in position.
! Position the opto and its retaining ring (ref. 10).
! Mount the plunger cover onto the plunger support. Impregnate the screw with weak loctite
(rep. 9).
! Position and screw the cam/anti-siphon arm unit using the Phillips head screw and the
washer (ref. 9).
! Screw the 2 Phillips head screws (ref. 8 and 7) which link the plunger support to the
plunger cover.
! Position the plunger cover clip (ref. 6).
9
Intervention procedures
Inj2_06.0_0051b_en.fm 89
! Cover the plunger with the protective film, mount the disengagement lever, its spring and
the adjustment button. Screw the plunger end shield back on (ref. 5, 4 and 3).
! Position the plunger cover and screw it back on (ref. 2).
When fitting the upper case, ensure the joint is perfectly positioned in its slot after
reassembly.
! Recalibrate the pressure sensor (see "Calibrations") then Perform the regular servicing
checks (see "Quality Control Certificate").
Intervention procedures
90 Inj2_06.0_0051b_en.fm
Intervention procedures
N°6, Procedure:
Plunger advance control potentiometer
Safety:
For safety reasons, the technician should not carry out any maintenance when the device is
connected to the 230 V mains supply voltage.
Disconnect the mains power supply cable.
Equipment required:
! 1 Posidriv Z1 screwdriver.
! 1 flat screwdriver.
! 1 flat key (12).
! 1 potentiometer dismounting tool ref. T300869.
Maintenance level:
Level 2, specialised technician (see documentation on biomedical training).
Procedure:
Access
! Turn the Injectomat over onto the upper case.
! Unscrew the 5 Phillips head screws (ref. 1) located at the bottom of the lower housing
case, which link this to the upper case.
! Keep the Injectomat assembled and turn it over onto the lower case.
! Remove the upper case.
! Disconnect the CPU board flat cable.
Hands must not come into contact with the CPU boards.
1
1
1
Intervention procedures
Inj2_06.0_0060b_en.fm 91
Dismounting
! Press the disengagement lever towards the back of the Injectomat as far as possible.
! Maintain this position and slide the whole unit to the back left.
! Remove the guide rail.
! Disconnect the black connectors (ref. 2).
! Unscrew the flex circuit brackets (ref. 3) and remove it.
! Remove the mechanical plunger unit from its slot.
! Disconnect the blue connector from the flex circuit (ref. 4).
! Remove the mechanical plunger unit.
! Unscrew the potentiometer lock nut (ref. 5).
! Remove the tooth lockwasher and the potentiometer.
4
5
Reassembly
! Dismount the end shield from the reducer frame.
! Mount the potentiometer onto the end shield (screw the nut onto a thread).
! Position the end shield in equipment T 300 869 and lock it into position using the knurled
screw.
! Position the potentiometer in the tool and bring it up against the end shield.
! Tighten the potentiometer.
! Extract the end shield from device T 300 869.
! Mount the pinion on the potentiometer (match up the indexing plane).
" The larger diameter of the pinion must be flattened against the potentiometer.
! With the potentiometer facing you, turn the pinion anticlockwise until it blocks, then turn it
¼ of a turn in the opposite direction.
! Mount the moving mechanical assembly onto the reducer frame.
! Slide the end shield onto the guides and rack.
" Check the position of the input bearing which must be on the plunger side.
" Ensure the flex circuit is not damaged when mounting (folding).
Intervention procedures
When fitting the upper case, ensure the seal is perfectly positioned in its slot after
reassembly.
! Recalibrate the position sensor (see "Calibrations") then perform the regular servicing
checks (see "Quality Control Certificate").
92 Inj2_06.0_0060b_en.fm
Intervention procedures
Safety:
For safety reasons, the technician should not carry out any maintenance when the device is
connected to the 230 V mains supply voltage.
Disconnect the mains power supply cable.
Equipment required:
! 1 Posidriv Z1 screwdriver.
Maintenance level:
Level 2, specialised technician (see documentation on biomedical training).
Procedure:
Access
! Press the syringe pump disengagement lever towards the back of the Injectomat as far as
possible.
! Maintain this position and slide the whole unit to the back right.
Dismounting
! Unscrew the 3 Phillips head screws (ref. 1) located at the bottom of the plunger end shield
which links this to the plunger cover. Remove the plunger end shield (ref. 2).
! Remove the backpressure adjustment button (ref. 3), then the disengagement lever and
its spring (ref. 4) as well as the protective plunger film (ref. 5).
! Unscrew and remove the Phillips head screw and the washer (ref. 6) which attach the anti-
siphon cam to the anti-siphon arm.
! Remove the cam/anti-siphon arm assembly (ref. 7).
4
6-7
2
3 1
5
1 1
Intervention procedures
Inj2_06.0_0071b_en.fm 93
Reassembly
Carry out the same procedures in reverse to remount the unit.
! Impregnate the screw with weak loctite (ref. 6) before mounting the cam and arm
assembly.
94 Inj2_06.0_0071b_en.fm
Intervention procedures
Safety:
For safety reasons, the technician should not carry out any maintenance when the device is
connected to the 230 V mains supply voltage.
Disconnect the mains power supply cable.
Equipment required:
! 1 Posidriv Z1 screwdriver.
! 1 small flat screwdriver.
! 1 antistatic wriststrap.
Maintenance level:
Level 2, specialised technician (see documentation on biomedical training).
Procedure:
Access
! Turn the Injectomat over onto the upper case.
! Unscrew the 5 Phillips head screws (ref. 1) located at the bottom of the lower case, which
link this to the upper case.
! Keep the Injectomat assembled and turn it over onto the lower case.
! Remove the upper case making sure the CPU board flat jumper is not accidentally
removed in the process.
! Disconnect the CPU board flat jumper.
Hands must not come into contact with the CPU boards.
1
1
1
Intervention procedures
Inj2_06.0_0081b_en.fm 95
Dismounting
! Press the disengagement lever towards the back of the Injectomat as far as possible.
! Maintain this position and slide the whole unit to the back left.
! Remove the syringe pump guide.
! Disconnect the black connectors (ref.4, ref.5).
Reassembly
Carry out the same procedures in reverse to remount the unit.
Intervention procedures
When fitting the upper case, ensure the joint is perfectly positioned in its slot after
reassembly.
! Recalibrate the force sensor (see "Calibrations") then perform the regular servicing checks
(see "Quality Control Certificate")
96 Inj2_06.0_0081b_en.fm
Intervention procedures
Safety:
For safety reasons, the technician should not carry out any maintenance when the device is
connected to the 230 V mains supply voltage.
Disconnect the mains power supply cable.
Equipment required:
! 1 Posidriv Z1 screwdriver.
Maintenance level:
Level 2, specialised technician (see documentation on biomedical training).
Procedure:
Access
! Turn the Injectomat over onto the upper case.
Dismounting
! Unscrew the 2 Phillips head screws (ref. 1) which link the battery holder to the lower case.
! Swivel the battery holder and remove the battery.
! Disconnect this.
Reassembly
Intervention procedures
It is recommended to perform a full loading and unloading cycle to ensure the battery is in
working condition.
! Perform the regular servicing checks (see "Quality Control Certificate").
Inj2_06.0_0090b_en.fm 97
Intervention procedures
98 Inj2_06.0_0090b_en.fm
Intervention procedures
Safety:
For safety reasons, the technician should not carry out any maintenance when the device is
connected to the 230 V mains supply voltage.
Disconnect the mains power supply cable.
Equipment required:
! 1 Posidriv Z1 screwdriver.
Maintenance level:
Level 2, specialised technician (see documentation on biomedical training).
Procedure:
Access
! Turn the Injectomat over onto the upper case.
! Unscrew the 5 Phillips head screws (ref. 1) located at the bottom of the lower case, which
link this to the upper case.
! Keep the Injectomat assembled and turn it over onto the lower case.
! Remove the upper case making sure the CPU board flat jumper is not accidentally
removed in the process.
! Disconnect the CPU board flat jumper.
Hands must not come into contact with the CPU boards.
1
1
1
Intervention procedures
Inj2_06.0_00101b_en.fm 99
Dismounting
! Disconnect the three black connectors (ref. 2).
! Unscrew the 2 Phillips head screws (ref. 3) located on the cable plug support, which link
this to the lower case.
Reassembly
Carry out the same procedures in reverse to remount the unit.
100 Inj2_06.0_00101b_en.fm
Intervention procedures
Safety:
For safety reasons, the technician should not carry out any maintenance when the device is
connected to the 230 V mains supply voltage.
Disconnect the mains power supply cable.
Equipment required:
! 1 Posidriv Z1 screwdriver.
Maintenance level:
Level 2, specialised technician (see documentation on biomedical training).
Procedure:
Access
! Turn the Injectomat over onto the upper case.
! Unscrew the 5 Phillips head screws (ref. 1) located at the bottom of the lower case, which
link this to the upper case.
! Keep the Injectomat assembled and turn it over onto the lower case.
! Remove the upper case making sure the CPU board flat jumper is not accidentally
removed in the process.
Hands must not come into contact with the CPU boards.
1
1
1
Intervention procedures
Inj2_06.0_00110b_en.fm 101
Dismounting
! Press the disengagement lever towards the back of the Injectomat as far as possible.
! Maintain this position and slide the whole unit to the back left.
! Remove the plunger guide.
! Disconnect the black connectors (ref. 2 and ref. 3).
2&3
Remounting
Fitting the flexible tongue:
! Remove the clip (ref. 6).
! Reposition the clip making sure the ribbon cable and the flexible tongue are centred on the
mechanical unit sprocket.
! Lift both ends of the flexible tongue slightly.
Intervention procedures
102 Inj2_06.0_00110b_en.fm
Fitting the guiding flange:
There are several types of flange tongues. If you have an older version, replace it with
that provided with the kit.
! Perform a few movement of the plunger to check the ribbon cable winds up correctly.
! Connect the black connectors (ref. 2 and ref. 3).
! Replace the plunger guide.
Carry out the same procedures in reverse to remount the upper case.
When fitting the upper case, ensure the joint is perfectly positioned in its slot after
reassembly.
Inj2_06.0_00110b_en.fm 103
Intervention procedures
104 Inj2_06.0_00110b_en.fm
Intervention procedures
Safety:
For safety reasons, the technician should not carry out any maintenance when the device is
connected to the 230 V mains supply voltage.
Disconnect the mains power supply cable.
Equipment required:
! 1 Posidriv Z1 screwdriver.
! 1 hexagon socket key (2.5).
Maintenance level:
Level 2, specialised technician (see documentation on biomedical training).
Procedure:
Access
! Turn the Injectomat over onto the upper case.
! Unscrew the 5 Phillips head screws (ref. 1) located at the bottom of the lower case, which
link this to the upper case.
! Keep the Injectomat assembled and turn it over onto the lower case.
! Remove the upper case making sure the CPU board flat jumper is not accidentally
removed in the process.
Hands must not come into contact with the CPU boards.
1
1
1
Intervention procedures
Inj2_06.0_00121b_en.fm 105
! Press the disengagement lever towards the back of the Injectomat as far as possible.
! Maintain this position and slide the whole unit to the back right.
! Remove the plunger guide.
Dismounting
! Unscrew the 3 Phillips head screws (ref. 2) located at the bottom of the plunger end shield
slots which link this to the plunger cover
! Remove the plunger end shield.
! Remove the backpressure adjustment button (ref. 3)
! Remove the disengagement lever and its spring as well as the protective plunger film.
! Remove the plunger cover clip.
2
2
2
! Unscrew the 2 Phillips head screws (ref. 4 and 5) which link the plunger support to the
plunger cover.
! Unscrew and remove the Phillips head screw and the washer (ref. 6) which attach the anti-
siphon cam to the anti-siphon arm.
! Remove the cam/arm unit as well as the plunger cover.
! Remove the plunger cover fitted with the bonding pad (ref. 7).
4 7
5
6
Intervention procedures
106 Inj2_06.0_00121b_en.fm
! Remove the Phillips head screw to separate the bonding pad (ref. 8), the removable stop
plate (ref. 9), the spring leaf (ref. 10), the stop plate (ref. 11).
9 10
11
8
Reassembly
! Tighten the Phillips head screw which holds the stop plate (ref. 11), the spring leaf
(ref. 10), the removable stop plate (ref. 9) to the bonding pad (ref. 8).
! Clip the bonding pad assembly to the plunger cover (ref. 7)
! Position the plunger cover.
1) Do not forget to replace the anti-siphon cam with the new one supplied with the kit.
2) Weak loctite must be used to mount the new arm and the new anti-siphon arm supplied
with the kit.
! Position and screw the cam/anti-siphon arm unit using the Phillips head screw and the
washer (ref 6).
! Screw the 2 Phillips head screws (ref. 5 and 4) which link the plunger support to the
plunger cover.
! Position the plunger cover clip.
! Cover the plunger with the protective film, and mount the disengagement lever and its
spring, and screw the plunger end shield back on.
When fitting the upper case, ensure the joint is perfectly positioned in its slot after
reassembly.
Inj2_06.0_00121b_en.fm 107
Intervention procedures
108 Inj2_06.0_00121b_en.fm
Intervention procedures
Safety:
For safety reasons, the technician should not carry out any maintenance when the device is
connected to the 230 V mains supply voltage.
Disconnect the mains power supply cable.
Equipment required:
! 1 Posidriv Z1 screwdriver.
! 1 cutter.
! GEB type 100% silicone grease.
Maintenance level:
Level 2, specialised technician (see documentation on biomedical training).
Procedure:
Access
! Turn the Injectomat over onto the upper case.
! Unscrew the 5 Phillips head screws (ref. 1) located at the bottom of the lower case, which
link this to the upper case.
! Keep the Injectomat assembled and turn it over onto the lower case.
! Remove the upper case making sure the CPU board flat cable is not accidentally removed
in the process.
Hands must not come into contact with the CPU boards.
1
1
1
Intervention procedures
Inj2_06.0_00130b_en.fm 109
! Press the disengagement lever towards the back of the Injectomat as far as possible.
! Maintain this position and slide the whole unit to the back left.
! Remove the rail guide.
! Disconnect the black connectors (ref. 2).
! Unscrew the flex circuit brackets (ref. 3) and remove it.
! Remove the mechanical plunger unit from its slot.
! Disconnect the blue connector from the flex circuit (ref. 4).
! Remove the mechanical plunger unit..
2
3
! Press the disengagement lever towards the back of the Injectomat as far as possible.
! Maintain this position and slide the whole unit to the back right.
Dismounting
! Unscrew the 2 slotted head screws (ref. 5) which link the centering ring to the mechanical
end shield.
! Using the cutter, cut the input bearing and remove it from the plunger tube.
5
Intervention procedures
110 Inj2_06.0_00130b_en.fm
Remounting
! Mount the new centering ring onto the plunger tube by twisting it.
! Laying it flat on a table, cut the O-ring using a cutter.
! Place the O-ring around the tube.
! Fit the O-ring into the centering ring slot (horizontal cut, so as to avoid lining it up with the
centering ring cut).
! Mount the stainless steel slotted plate onto the pin, by twisting it.
! Place the silicone grease cord onto the plate around the passage of the tube.
! Place the plate on the centering ring.
! Position the whole assembly against the mechanical end shield and screw it on using the
two slotted screws (ref. 5).
! Check the plunger slides correctly.
It should slide uniformly across the entire centering ring and slight friction is due to the O-
ring scraping against the tube.
Intervention procedures
Inj2_06.0_00130b_en.fm 111
Intervention procedures
112 Inj2_06.0_00130b_en.fm
Intervention procedures
Safety:
For safety reasons, the technician should not carry out any maintenance when the device is
connected to the 230 V mains supply voltage.
Disconnect the mains power supply cable.
Equipment required:
! 1 Posidriv Z1 screwdriver.
! 1 antistatic wriststrap.
Maintenance level:
Level 2, specialised technician (see documentation on biomedical training).
Procedure:
Access
! Turn the Injectomat over onto the upper case.
! Unscrew the 5 Phillips head screws (ref. 1) located at the bottom of the lower case, which
link this to the upper case.
! Keep the Injectomat assembled and turn it over onto the lower case.
! Remove the upper case making sure the CPU board flat jumper is not accidentally
removed in the process.
! Disconnect the CPU board flat jumper.
Hands must not come into contact with the CPU boards.
1
1
1
Intervention procedures
Inj2_06.0_00141b_en.fm 113
Dismounting
! Press the disengagement lever towards the back of the Injectomat as far as possible.
! Maintain this position and slide the whole unit to the back left.
! Remove the plunger guide.
! Disconnect the black connectors (ref. 2).
! Unscrew the flat jumper mounting lug (ref. 3) without removing it.
! Remove the mechanical plunger unit from its slot.
! Disconnect the blue connector from the flat jumper (ref. 4).
! Remove the mechanical plunger unit.
4
2
3
! Press the syringe pump disengagement lever towards the back of the Injectomat as far as
possible.
! Maintain this position and slide the whole unit to the back right.
Welding and unwelding operations are carried out using a soldering iron fitted with a tip in
good condition, constantly tapered and kept clean. The temperature of the iron should be
between 315°C and 340°C.
! Unweld the flex circuit (ref. 5) ensuring the disengagement switch is not damaged in the
process.
! Unscrew the 2 slotted head screws (ref. 6) which link the centering ring to the mechanical
end shield.
! Unscrew the 2 Phillips head screws (ref. 7) which link the centering ring to the mechanical
end shield.
! Remove the flex circuit clip (ref. 8).
5 7
Intervention procedures
6
8
114 Inj2_06.0_00141b_en.fm
! Remove the plunger guide.
! Unscrew the 3 Phillips head screws (ref. 9) located at the bottom of the plunger end shield
which links this to the plunger cover.
! Remove the plunger end shield.
9
9
11
13 12
10
Intervention procedures
Inj2_06.0_00141b_en.fm 115
! Unscrew the 2 Phillips head screws (ref. 14 and 15) which link the plunger support to the
plunger cover.
! Unscrew and remove the Phillips head screw and the washer (ref. 16) which attach the
anti-siphon cam to the anti-siphon arm.
! Remove the cam/arm unit as well as the plunger cover.
14
15
16
! Remove the retaining ring that holds the anti-siphon opto in place (ref. 17).
! Unscrew the 6 hexagon socket screws (ref. 18).
18
17
Intervention procedures
116 Inj2_06.0_00141b_en.fm
! Remove the backpressure micro support/spring leaf (ref. 19).
! Pull off the ribbon cable from the double-sided board
19
Welding and unwelding operations are carried out using a soldering iron fitted with a tip in
good condition, constantly tapered and kept clean. The temperature of the iron should be
between 315°C and 340°C.
When dismounting the force sensor, ensure the welding blocks are not damaged and the
square is not removed in the process.
" Add more weld to facilitate the unwelding process.
" Heat and pull on the wires one by one.
! Remove the flex circuit and tube kit from the plunger support.
Intervention procedures
Inj2_06.0_00141b_en.fm 117
Reassembly
! Remove the sticky tape holding the flex circuit to the tube.
! Mount the new flex circuit onto the plunger support.
! Weld the ribbon cable to the new pressure switch, ensuring the mounting procedure is
carried out in order:
" Taper the 3 welding points.
" Weld the 3 points.
" Check that the welding has been carried out correctly. It should not form a ball, but
should lie against the length of the component snubbers.
ube
Switch
w disengagement
Switch
backpressure
Anti-siphon opto
Lower side
(connection specific for device equiped with a force sensor)
stuck to the double-sided board on the plunger support
Do not forget to impregnate the cam/arm clamping screw (rep. 16) with weak loctite.
When fitting the upper case, ensure the joint is perfectly positioned in its slot after
reassembly.
Intervention procedures
! Recalibrate the force sensor (see "Calibrations") then perform the regular servicing checks
(see "Quality Control Certificate").
118 Inj2_06.0_00141b_en.fm
Intervention procedures
Safety:
For safety reasons, the technician should not carry out any maintenance when the device is
connected to the 230 V mains supply voltage.
Disconnect the mains power supply cable.
Equipment required:
! 1 Posidriv Z1 screwdriver.
! 1 antistatic wriststrap.
! 1 soldering iron.
! "RADIEL Sn60Pb RI 1" welding wire (cleaning not required for rewelding) or equivalent.
Maintenance level:
Level 2, specialised technician (see documentation on biomedical training).
Procedure:
Access
! Turn the Injectomat over onto the upper case.
! Unscrew the 5 Phillips head screws (ref. 1) located at the bottom of the lower case, which
link this to the upper case.
! Keep the Injectomat assembled and turn it over onto the lower case.
! Remove the upper case making sure the CPU board flat jumper is not accidentally
removed in the process.
! Disconnect the CPU board flat jumper.
Hands must not come into contact with the CPU boards.
1
1
1
Intervention procedures
Inj2_06.0_00150b_en.fm 119
Dismounting the upper cover
! Dismount the central unit and display boards (see corresponding sheet).
! Dismount the syringe clamp (see corresponding sheet).
! Dismount the syringe detection system (see corresponding sheet).
Welding and unwelding operations are carried out using a soldering iron fitted with a tip in
good condition, constantly tapered and kept clean. The temperature of the iron should be
between 315°C and 340°C.
" Unweld the snubbers from the buzzer:
# add more weld to facilitate the unwelding process.
# Heat and pull on the buzzer pins one by one.
Clean the buzzer slot surfaces using a cloth soaked in rubbing alcohol.
" Weld the snubbers of the new buzzer:
+
Buzzer with snubbers
Display board
# taper the 2 buzzer pins to be welded.
# place the buzzer onto the board following the
mounting diagram.
# weld to buzzer onto the board.
# check that the welding has been carried out correctly.
It should not form a ball, but should lie against the
length of the wire.
Intervention procedures
Continue with the remounting procedure, referring to the “central unit and display board”
sheet.
120 Inj2_06.0_00150b_en.fm
Dismounting the lower cover
! Dismount the power supply board (see corresponding sheet).
! Dismount the battery holder (see corresponding sheet).
! Dismount the rear plug support (see corresponding sheet).
Intervention procedures
Inj2_06.0_00150b_en.fm 121
Intervention procedures
122 Inj2_06.0_00150b_en.fm
7 Calibration
7.1 Calibration procedure
The calibration menu is reserved for authorised personnel only, its access is
protected by a secret code.
Keyboard description.
Key Function
ON, is used to switch the machine ON.
OFF, is used to switch the machine off when pressed for over three
seconds.
SILENCE ALARM, is used to access the calibration mode.
The select keys allow to scroll the figures and letters on the LCD
screen, on the units, tens, tenths segments etc.
Activate calibration
! Press the "SILENCE ALARM" and
"BOLUS" keys simultaneously.
Inj2_07.1_002b_en.fm 123
7.1.2 EtA.4 Calibration of the 3 battery voltage levels.
This menu enables ! EtA.4, press "CONFIRM".
" Bat.1: supply the device with 6.3 V+ 0,05 using a
the user to store
the three Bat1,
stabilised power supply.
Bat2 and Bat3
# Press "CONFIRM".
battery voltage
The voltage is read and stored in the EEPROM.
values in an
" Press "CONFIRM".
EEPROM.
" Bat.2: supply the device with 5.9 V+ 0,05 using a
stabilised power supply.
# Press "CONFIRM".
The voltage is read and stored in the EEPROM.
" Press "CONFIRM".
" Bat.3: supply the device with 5,7 V+ 0,05 using a
stabilised power supply.
# Press "CONFIRM".
The voltage is read and stored in the EEPROM.
Once both high and low values have been stored in the EEPROM, the device indicates
the number of LSB in decimals between the two measurement points.
This value should be 776 + 10 LSB.
124 Inj2_07.1_002b_en.fm
8 Spare parts catalogue
8.1 Upper case
99
100
101
.....
.....
.....
104
.....
..... .....
.....
.....
..... .....
103
.....
102
..... : see next page
Inj2_08.1_008b_en.fm 125
.....
.....
.....
113
111
112
.....
125
126 114
127
123
128 129
.....
124
.....
..... : see previous page
126 Inj2_08.1_008b_en.fm
Spare parts catalogue
109
115
108
116
117
118
107
119
120
121
Inj2_08.2_008b_en.fm 127
Spare parts catalogue
128 Inj2_08.2_008b_en.fm
Spare parts catalogue
.....
..... .....
.....
..... .....
200
201 ..... ..... .....
202 ..... .....
224
225
203
204
205
206 .....
207 ..... ..... .....
208
220
221 223 .....
209 .....
..... .....
222 .....
.....
226
..... .....
210
Inj2_08.3_009b_en.fm 129
249
229 260
261
..... 253 250
..... 230 254 251
..... ..... 231 252
.....
.....
.....
.....
..... 270
..... 232 271 272
..... .....
..... ..... 255
..... ..... 256
..... 257
258
259
.....
..... 227
.....
273
..... : see previous page
130 Inj2_08.3_009b_en.fm
298
Inj2_08.3_009b_en.fm 131
Spare parts catalogue
132 Inj2_08.3_009b_en.fm
Spare parts catalogue
304
296
294
300
301
302
303
Inj2_08.4_008b_en.fm 133
Spare parts catalogue
134 Inj2_08.4_008b_en.fm
Useful addresses
TRAINING / SERVICE DEPARTMENT
Fresenius Kabi Tél. : 00 33 (0) 4 76 67 10 76
Le Grand Chemin, Tél. : 00 33 (0) 4 76 67 11 40
38590 Brézins Fax : 00 33 (0) 4 76 67 11 22
France
This document may not be reproduced in part or in whole without the written permission of
Fresenius Kabi