Radiation Therapy - External Beam Operations Manual
Radiation Therapy - External Beam Operations Manual
Radiation Therapy - External Beam Operations Manual
Contents ..................................................................................................................................... 1
Introduction................................................................................................................................ 7
Scope .......................................................................................................................................... 7
Background................................................................................................................................. 7
Key Objectives ............................................................................................................................ 7
Section 1 – Treatment Area General Information ..................................................................... 7
1.1 Introduction ................................................................................................................. 7
1.2 Radiation Therapists Staff Organisation ...................................................................... 8
1.3 Radiation Therapist Responsibilities ......................................................................... 10
1.4 Linac Start-up and Handover ..................................................................................... 10
1.5 Daily Machine Geometry Quality Assurance ............................................................. 11
1.6 Treatment Delivery .................................................................................................... 11
1.6.1 Starting a New Case ................................................................................................ 12
1.6.2 Team Work Guidelines ............................................................................................ 12
1.6.3 Notes on Patient Set-Ups ........................................................................................ 13
1.6.4 Billing Procedures .................................................................................................... 14
1.6.5 Verification Images.................................................................................................. 14
1.6.6 Weekly Chart Checks ............................................................................................... 15
1.6.7 Patient Reviews ....................................................................................................... 15
1.6.8 Course Completions ................................................................................................ 16
1.6.9 Close of Business ..................................................................................................... 16
1. 6.10 Quality Assurance ................................................................................................. 16
1.6.11 Personal Stabilisation Devices ............................................................................... 16
1.6.12 Other Matters........................................................................................................ 17
1.7 Annual Business ......................................................................................................... 17
1.8 Fault Documentation ................................................................................................. 17
1.8.1 LINAC Faults preventing the progression of treatment .......................................... 18
1.8.2 LINAC Faults not preventing the progression of treatment ................................... 19
Introduction
This manual describes the procedures and guidelines associated with the External Beam
Treatment Radiation Therapy using the Linear Accelerator.
Scope
This Manual applies to Radiation Therapists working within Radiation Oncology at the
Canberra Hospital.
Background
The clinical practice of megavoltage radiation therapy is complex and dynamic. Radiation
Therapists must have the appropriate qualifications and training to practice radiation
therapy in a safe, effective and efficient manner. The RT must exercise professional
judgement, skill and due care at all times.
The External Beam Radiation Therapy Operations Manual is a general guide for procedures
carried out in the Linear Accelerator section of Radiation Oncology, hereafter referred to as
LINAC.
Key Objectives
The key objective is the safe operation of the Radiation Therapy Service at the Canberra
Hospital.
1.1 Introduction
Each Radiation Therapist (RT) working on the LINAC is expected to:
be familiar with LINAC Procedure Manual contents
clarify any uncertainty with appropriately skilled staff related to equipment operation,
treatment instructions or image verification, prior to proceeding with treatment delivery
follow all related procedures and policies described at:
o Radiation Oncology department level
o Capital Region Cancer Service level
o Health Directorate level
Radiation Therapist's must also be familiar with the contents of Vendor supplied Manuals,
located in the LINAC Control Room. These manuals include, but are not limited to:
Varian - CLINAC Instructions for Use Guide
Varian - Treatment Delivery Instructions for Use Guide
On Board Imager (OBI) Advanced Imaging Reference Guide
ARIA Radiation Oncology Applications Manuals
The LINAC Procedure Manual will be revised annually by the Radiation Therapist 4.2 (Head of
Treatment) in consultation with LINAC Supervisors. However, in between such revisions any
changes in procedures will be communicated via:
Distribution of hard copy memorandum to each work area (LINAC 1, LINAC 2, LINAC 3,
LINAC 4, Brachytherapy, SXRT, Planning, CT1, CT2)
Each hard copy memorandum is to be filed in the respective LINAC Memorandum File
located in the LINAC Control Room. An electronic copy of the memorandum will be filed
in the Radiation Therapist Memo's Folder on the Radiation Oncology G Drive.
Memoranda emailed to Radiation Therapists.
Upon commencing a rotation to the LINAC, each Radiation Therapist must familiarise or re-
familiarise themselves with the contents of the LINAC Procedure Manual and relevant
memoranda.
Head of Treatment
(RT 4.2)
Unit Supervisor
(RT 4.1)
Senior Radiation
Therapist (RT 3)
Radiation Radiation
Therapist (RT 2) Therapist (RT 2)
NPDP Radiation
Therapist (RT 1)
Under the current RT organisational structure, the Radiation Therapist 4.2, referred to locally
as Head of Treatment, is responsible for the overall management of the treatment service
related to LINAC 1, LINAC 2, LINAC 3, LINAC 4, Stereotactic Radiation Therapy, Brachytherapy
Treatment and Superficial X-Ray Therapy.
Doc Number Version Issued Review Date Area Responsible Page
CHHS16/204 1 02/11/2016 01/09/2019 CACHS - Radiation Oncology 8 of 117
Do not refer to a paper based copy of this policy document. The most current version can be found on the ACT Health Policy Register
CHHS16/204
On days when staffing deviates from the four fully accredited person model outlined above
due to unplanned staff sick leave, the LINAC Supervisor or their delegate will inform the
Head of Treatment who will assess the need to provide an RT to backfill. Where possible,
Treatment RT's will be limited to working with three fully accredited team members for no
more than two consecutive days.
At the beginning of each day, the LINAC Supervisor will organise staff into one team of two
and one team of three (including the RT Intern). The ideal combination of staff in each team
will be one experienced RT and one less experienced RT (+/- Intern). The team combinations
will also take into account new case commencement, tea breaks and lunch breaks.
delegation of duties
From time to time, the LINAC Supervisor will be required to perform other professional
duties resulting in their absence from the LINAC. In his/her absence, the RT3 and RT2's will
continue to ensure the safe, effective and efficient operation of the LINAC. During any
period of absence, the level of responsibility normally held by the LINAC Supervisor will
default to the RT3 working within the team.
Their routine duties as a team include, but are not limited to, undertaking:
confirmation of patient identity and treatment site prior to treatment delivery
accurate treatment set-up
treatment delivery
acquisition of imaging
analysis of imaging
continuous visual monitoring of the patient on CCTV
accurate documentation and sign off to acknowledge treatment delivery
update any written/electronic instructions/records
determination of correct Medicare billing codes
referral of patients to nursing/medical team for appropriate medical care
restocking supplies
administrative duties
other duties as requested by the LINAC Supervisor
The RT who switches on the LINAC is responsible for activating the door interlocks upon exit
of the LINAC bunker and ensuring no other member of staff or public is within the LINAC.
To indicate the successful completion of the LINAC start-up and OBI warm-up, the MPRE
Officer will tick the electron and photon modalities at the top of the Morning Geometry
Quality Assurance Log Book along with their signature and date.
If the Morning Geometry Quality Assurance Log Book has both radiation modalities ticked
and MPRE Officers initials, the RTs can assume the LINAC and OBI are safe for Clinical Use
with no further checking, other than Geometry and OBI Quality Assurance, required from
the RT’s.
If the LINAC is in the ON condition1, but the refer to fault book box is ticked or MPRE
signature is missing from the appropriate section of the Morning Geometry Quality
Assurance Book, the RT must confirm with MPRE staff that the LINAC is ready for Clinical Use
prior to commencing clinical treatment.
It is departmental procedure, if for any reason, the MPRE Staff member does not want the
LINAC to be used, they will:
record the relevant information in the LINAC Fault Book
stick a yellow “OUT OF SERVICE" notice on the control console key of the machine
indicating that the machine is not to be used
inform the LINAC Radiation Therapists personally (if present)
When the LINAC is handed over from MPRE, the Radiation Therapists must complete the
Daily Machine Geometry and OBI Quality Assurance Checks.
The RT who completes the Morning Geometry and OBI QA must sign the LINAC Daily
Geometry Quality Assurance Record and notify the LINAC Supervisor or their delegate of any
failure, which must be discussed with the MPRE Officer prior to proceeding with any clinical
treatment.
1. When the LINAC is expected to be in the ON condition but is not, RE should be immediately informed. If unable to
contact RE, a Medical Physicist should be contacted.
Doc Number Version Issued Review Date Area Responsible Page
CHHS16/204 1 02/11/2016 01/09/2019 CACHS - Radiation Oncology 11 of 117
Do not refer to a paper based copy of this policy document. The most current version can be found on the ACT Health Policy Register
CHHS16/204
The LINAC Supervisor is responsible for the maintenance and monitoring of LINAC
patient bookings. The LINAC Supervisor should:
o Adjust the patient bookings in between treating patients. The LINAC Supervisor
holds a large clinical service delivery responsibility and should not regularly utilise
large blocks of time for changing/adjusting patient appointments, completing
project work or responding to emails. Provisions for project work time are in place
and must be discussed in advance with the Head of Treatment.
o Make necessary adjustments to the time allotted for each patient. Patients may
require more or less time than originally allocated by the Radiation Oncology Liaison
Officer. The LINAC supervisor must adjust the allocated treatment time as necessary
to ensure the even distribution of patient load. Booking one patient over the top of
another patient must be avoided in all but exceptional circumstances as this will
impact on efficiency.
o Arrange staggered starts for staff at least one week in advance (where possible) as
required. In circumstances where less than one weeks’ notice is provided,
cooperation amongst the RT staff in determining shift rosters is required.
All patients treated within the department will be checked in to ARIA and their name
will appear in the 4DTC Appointment Schedule.
Inpatient treatment will be facilitated by the Wards pool service. The inpatient check
box must be ticked in ARIA to allow nursing staff to accurately organise the wards pool
service schedule in advance for the following day. Where possible, all Inpatient
treatments should be performed during core business hours so as ensure medical
and/or nursing cover. After hours, coverage should be arranged in advance with the
Ward Services Supervisor. The afterhours CNC may also be contacted through the
Switchboard if necessary.
Details of the procedure for commencing a new case are found in Section 5 of the External
Beam Operations Manual.
Escorting next patient to the LINAC Bunker entrance as the previous patient treatment
nears completion. Do not bring the patient to the bunker entrance too soon as there is
no suitable seating.
Ensure the next patient is fully prepared for with any pre-treatment prerequisites, e.g.
Full/empty bladder, premed etc
Organising patient from the ward with or without prior medication in advance ensuring
that others are aware it has been done.
During quiet periods the LINAC Supervisor may delegate cleaning, stock replenishment,
quality assurance or administrative tasks to any member of the team.
On the first treatment, the responsible RT is required to read out all the beam
parameters as well as wedge orientation from the treatment sheet and check against
the LINAC computer while the other RT cross checks on ARIA 4DTC.
Before switching the beam on, also check the box on the treatment sheet where the
current day's MU will be recorded, as there may be a modification from planning.
Radiation Therapists must keep a close watch of the patient via the CCTV and ensure
that the audio intercom is switched on during every treatment.
IMPORTANT: The treatment RTs are not to be interrupted or distracted by other team
members during treatment delivery. During rotation of the gantry from outside the room,
the Responsible RT must monitor the CCTV to ensure collision is avoided.
Images are to be reviewed throughout the day prior to the delivery of the next
treatment fraction.
Two accredited RT's must review all offline images. The results need to be noted in ARIA
and confirmed prior the following treatment.
Two accredited RT's must review all Online Images. One accredited RT is required to
perform a retrospective, offline assessment of the online match.
All changes to a treatment arising from a reviewed image are to be clearly documented
in the Treatment Sheet by the RT observing the change at the image review. A second
RT must countersign any change to indicate their agreement with magnitude and
direction
Images must be made available for the Radiation Oncologist to review.
between Linacs, all shells should be transported directly from bunker to bunker. If a
trolley is required for transport or additional storage, this should be stored within
the bunker only.
o If a patient has been identified as contracting an infectious condition, all patients
treatment equipment will now be placed inside a linen bag.
No yellow or clear garbage bags are to be used. This is to eliminate any
misunderstandings and to isolate the patients equipment from others shells. The
bag should be clearly tagged and identified with the patient’s information.
Please continue to use universal precautions when handling these bags.
Equipment Disposal:
For details of equipment disposal at the completion of a course of treatment see Section
8 of the External Beam Operations Manual.
determining the implications of machinery and equipment faults and failures and whether a
fault will prevent the progression of treatment.
As a rule, all faults and interlocks must be recorded in the LINAC Logbook. It is essential that
information in relation to the machine parameters such as gantry, collimator, couch and
energy are recorded to determine whether there is a consistency in fault or trends.
All logged faults must have the clinical release or referral to MPRE section completed. The
following summarises the expectations of all Radiation Therapist if they are releasing a linac
back into clinical use:
Linacs can only be released for clinical use by fully accredited Radiation Therapists (Level
2.1 or greater)
Only minor faults not preventing the progression of treatment or Dosimetry Interlocks
described in Section 18 of the External Beam Operations Manual should be released by
Radiation Therapist, all other faults should be referred to MPRE
Dosimetry interlocks can be cleared according to the Dosimetry Interlocks Standard
Operating Procedure. Staff need to make themselves familiar with the limitations,
documentation requirements and alerts specified in this protocol.
All hardware/software/other faults must be recorded in the fault book. If an accredited
Radiation Therapist using professional judgement feels this fault is able to be cleared
and will not affect the safe delivery of treatment they may release the machine back
into clinical use by signing this section.
If a staff member is not confident making the decision to release the linac for clinical
use, the linac senior or Head of Treatment should be consulted if available. If these staff
members are not available or unable to safely resolve the issue Medical Physics and
Radiation Engineering (MPRE) staff must be consulted prior to treatment resuming.
If a staff member releases a linac for clinical use, they are responsible for notifying the
linac senior of this fault in a timely manner. This is to enable recurring faults and/or
patterns of faults to be communicated to the Head of Treatment and MPRE, and the
liaising of requirements and time frames for repairs.
In the event that the RE officer is not contactable, Radiation Therapists are to contact a
Medical Physicist on duty or the Chief Medical Physicist by Mobile phone.
IMPORTANT*
The mechanical counter is affected if the "FIXED" or "READY/COMPLETE" console buttons
are pushed. Therefore, DO NOT push any buttons until the mechanical counter value has
been recorded.
In general:
the patients currently waiting in the department for treatment will be informed of the
situation and will be asked to either wait, or go away and to call the LINAC in about one
hours time.
patients due for treatment in the next hour will be contacted by phone, informed of the
situation, asked to hold off coming in for treatment and to call the LINAC in about one
hour’s time.
the patient’s decision and any details will be recorded on the front of the treatment
folder.
If a patient is transferred from LINAC 1 to LINAC 2 or 3, no recalculation of the plan is
required as the machines have been beam matched.
If a patient is transferred to LINAC 4, care should be taken to ensure the difference in
the BrainLab couch top is not dosimetrically significant. If in doubt, discuss with Head of
Planning. Patients with posterior or oblique fields may be transferred to LA4 in
emergency circumstances only, refer to section 23 of the External Beam Operations
Manual.
If treatment is to recommence, the Head of Treatment will judge if the full day’s
workload can be treated up to 7pm. If not the Head of Treatment will estimate the
number of patients that can be treated in the remaining time (up to 7pm) and obtain a
patient priority list from each Radiation Oncologist for the available treatment time
slots. Staff will be required to work overtime as necessary.
The LINAC Supervisor will:
o Produce a revised treatment appointment list (either from the Appointment Booking
System or hand written).
o Inform patients of their revised appointment times/the departmental response.
o Redistribute patients and staff to other LINACs
o Notify the ward. If ward patients are coming late.
The LINAC Supervisor will update the appointments as required, i.e. change patient
finishing dates if patients have been cancelled and as necessary relay these changes to
other LINACs.
The LINAC Unit Supervisor will provide a shift roster for staff on LINAC for all
breakdowns beyond one day. At least one RT3.1 or above will be required to work a late
shift.
Community Transport
Community Transport is usually organised by the Radiation Oncology Liaison Officer or
Social Worker.
599 p286.
o Slide the treatment couch away from the tube head so that the whole of the patient
(including the patient's head) is outside the area which is vertically underneath the
shadow tray.
o Insert the appropriate shadow tray plate and put the shielding blocks on to the
template which has already been aligned to the centre cross of the plate and taped
to the plate.
o After the shielding blocks are aligned in position, slide the couch back to the position
underneath the tube head.
o Check that the field centre and borders are still correct according to the light field.
o After the treatment of the field, again slide the couch out before removing the
shielding blocks from the tray.
Be conscientious about patient/staff safety when moving equipment, whether manually
or by electric device.
Be conscientious about safeguards in working with low-melting-point alloy. In particular,
o Keep the work area clean and prohibit eating and drinking in the work area.
o Wash hands with soap and cold water after each work session to prevent ingestion
or absorption of the toxic metals.
o Never dispose of the alloy yourself. Return any unrequired alloy to the mould room.
Do not allow public access to alloy.
The following steps that must be undertaken to ensure the correct patient, correct site and
the correct procedure for patients receiving treatment in the Radiation Oncology
department.
2.1. Contraindications/Alerts
Do not proceed with treatment delivery until all pre-procedure documentation is
accurate and complete.
For patients who are transferred from locations within the hospital who are incapable of
personally participating in the verification process, and with no authorised
representative present, a member of staff from the preceding location must act as the
patient’s representative for the verification.
If a discrepancy arises at any point during the Correct Patient, Correct Site, Correct
Procedure verification, the procedure will not commence until the discrepancy is
resolved. All team members must agree on the resolution of the discrepancy. If
disagreement continues to occur then the Radiation Oncologist in charge of the
treatment carries ultimate responsibility for the care of the patient and should decide
the most appropriate course of action.
When a patient refuses skin marking or tattoos, documentation must be recorded in the
treatment record.
Correct Patient
The Radiation Therapist (RT) who will switch on the Linear Accelerator shall check the
patient’s identification, patient identification photo (if available), full name, address, date of
birth and armband to confirm the correct patient.
The RT must always ensure the correct patient by asking the patient to state their name and
date of birth.
Verification of the correct patient must occur upon entry to the linear accelerator bunker or
as soon as practicable after entering the linear accelerator bunker. Aim to avoid correct
patient verification immediately before leaving the bunker as patient movement has been
observed and treatment accuracy may be compromised. If correct patient verification is
overlooked until the patient is in the set-up position, isocentre positioning must be carefully
re-checked prior to delivering the radiation therapy treatment to ensure the patient has not
moved.
Prior to leaving the Linac Bunker, the RT who will switch on the Linear Accelerator must
verbally state the setup parameters including:
patient name
anatomic site being irradiated with particular attention to left/right
isocentre location in relation to standard anatomic position
other key treatment information
For example: “John Jones on the bed, treating a left hip, isocentre is 10cm left of anterior
midline, anterior field 10 x 10cm, no MLC”
The second Radiation Therapist present shall check these instructions and parameters on the
Clinac NOT ARIA.
3.1 Linac 1, Linac 3 and Linac 4 Geometry QA and Imaging System Checks
Daily Linear Accelerator (Linac) Geometric Quality Assurance checks must be performed
each morning prior to commencement of treatment delivery and following the repair of
equipment to ensure operation within accepted tolerances. It is to be undertaken by an
accredited Radiation Therapist.
The aims of the Daily Linac Geometry Quality Assurance checks are to ensure:
correct daily handover procedures are observed prior to treatment delivery
Linac geometry meets acceptable limits
the simulated treatment geometry can be accurately reproduced on a daily basis at the
Linac
Success is entered on the chart as a tick (√). Failure is entered as a cross (x). This assumes
IEC1217 scales for all machine mechanical parameters.
3.1.1 Responsibilities
3.1.1.1 Radiation Engineering (RE) & Medical Physics (MP) Staff
Prior to the daily release/handover of the Linear Accelerator, the RE or MP staff will perform
morning run-up to ensure the Linac is free from faults and dose output, for all beam
energies, and is within acceptable limits.
The success of the daily dose output checks will be recorded by the RE or MP staff member
on the Linac Daily Geometry Quality Assurance Record. In the absence of ticking, refer to
fault book box the RE or MP staff member is signing to denote the Linear Accelerator is fit
for clinical use.
When the Linac is handed over from RE or MP, the Radiation Therapists must complete the
Daily Machine Geometry Checks to determine the accuracy of the treatment geometry
system.
The following documentation contains procedural instruction and the current acceptable
limits for daily geometric quality assurance checks.
If any geometry of functionality failures are detected during this process, the RT staff must
contact the RE or MP staff to have the issues rectified prior to clinical use. RT staff are
required to log any failures in the Linac Log/Fault Book. RT staffs are only to take
responsibility for the Linac after notation is dated and recorded in the Linac Log/Fault Book,
from either RE or MP, stating the Linac is “Fit for Clinical Use”
3.1.2 Contraindications/Alerts:
If the RE or MP section of the chart is not completed, RE or MP staff must be contacted prior
to commencement of clinical treatment.
If any failures in the Daily Geometry QA are detected, RE or MP staff must be contacted to
make any required adjustments prior to clinical use.
If any RE or MP Staff member does not want the Linac to be used clinically, they will:
Tick the refer to fault book box on the chart.
record the relevant information in the ‘Linac Fault Book'
place a yellow ‘OUT OF SERVICE” notice on the control console key of the Linac
indicating that the machine is not to be used inform the Radiation Therapists personally
(if present)
2. If the refer to fault book box has been ticked, the fault book must be referred to and the
released for clinical use box MUST be ticked prior to the commencement of clinical
treatment.
Success is entered on the chart as a tick (√). Failure is entered as a cross (x) with a comment
to describe the nature and extent of the variation e.g. (x) 3mm sup.
All failures are to be discussed with RE or MP staff and entered in the Linac Fault/Log Book
found in the Linac Control Room.
PROCEDURE: 2
a) Digital Couch Vertical Readout Correspondence.
The limit for variation of the digital couch vertical readout is +/-0.2cm.
This part of the procedure does not require the Iso-Align Jig
1. Using the ODI set 100cm FSD to surface of the solid carbon fibre couch top.
2. Ensure digital couch vertical readout equals 0 (+/-0.2cm).
3. Enter success or failure on the chart.
Overhead Laser
1. Ensure collimator is at 0 degrees
2. Rotate the gantry off zero degrees until the overhead laser can be visualised on the
Iso-Align Jig
3. Check that the overhead laser aligns with the crosses of the Iso-Align Jig.
4. Enter success or failure on the chart.
Lateral Lasers
1. Carefully rotate the Iso-Align faceplate to a face the 90 degree
2. Confirm the lateral lasers align with the crosshairs marked on each side of Iso-Align
Jig
3. Rotate the gantry to 270 degrees, then 90 degrees
4. Check the Linac collimator crosshairs align with the Iso-Align Baseplate crosshairs at
each gantry angle
5. Enter success or failure on the chart.
Sagittal Lasers
1. Check that the sagittal laser aligns with the scribed mark on the foot end of the Iso-
Align Jig.
2. Enter success or failure on the chart.
f) MV Functionality/Geometry Test
The limit for misalignment of the field size is 0.2cm.
Inside Bunker
1. Ensure the Iso-Align Jig is rotated to 90 degree vertical position
2. Rotate the Gantry to 90 degrees
3. Ensure the Linac crosshairs correspond with the crosshairs marked on the Iso-Align
Jig
4. Maintain 100cm SSD to the surface of the Iso-Align Jig
5. Set 15cm x 15cm field size on digital readout
6. Move the MV Imaging arm into acquisition position
Outside Bunker
1. Open the patient named “T-LA_ Morning QA”
2. Mode up the planned image for “Field 1 – Gantry 90”
3. Acquire both the planned and open exposures
4. Ensure the 15cm x 15cm irradiated field size corresponds with the 15cm x 15cm
markers on the Iso-Align Jig, as in Figure 2. If the red irradiated field falls beyond the
outermost or innermost white radio opaque markers delineating the 15cm x 15cm
field size, notify RE and/or Medical Physics
5. Enter success or failure on the chart.
g) Parallelism of the patient positioning/set-up laser system (green) with isocentric lateral
lasers (red laser).
The limit for misalignment of the laser system is the width of the laser or 0.2cm
whichever is less.1
1. Rotate the Gantry to 0 degrees
2. Confirm the lateral isocentre (red) lasers correspond with the central axis markings
on the Iso-Align Jig
3. Using the Digital Couch Readout, lower the couch 20cm
4. Check that the lateral (green) lasers are at central axis of the Iso-Align Jig.
5. Check the opposing lateral (green) laser on the opposite side of the Iso-Align Jig
6. Enter success or failure on the chart.
3. Align the couch so that the lasers and cross hairs run through the centre crosses on the
QA Cube
4. Move the couch so that the lasers and cross hairs align with the offset marker on the QA
cube
5. Remove the Bow Tie Filter
6. Leave the room activating the door interlock on the way out.
At the 4 D Console
1. Open the QA Patient
2. Select the Ant kV setup field
3. Click Mode up
Success is entered on the chart as a tick (√). Failure is entered as a cross (x)
At the 4 D Console
1. Select the CBCT setup field
2. Click Mode up
3. Select 3D/3D Match Icon
4. Follow the On-screen instructions to acquire a CBCT image of the QA Cube using the Low
Dose Head Protocol (refer to CBCT General Instructions for use)
5. Accept and export the CBCT dataset
6. Wait for the CBCT dataset to be saved, as indicated by ‘saved 100%” message
7. Click Save Match
8. Select Cancel
9. Close Patient from 4DTC
10. Log into Offline review and ensure the acquired CBCT dataset is available to review
11. Set the acquired CBCT image to Reviewed. It is not necessary to analyse the image
match.
DATE
Photon Electron Photon Electron Photon Electron Photon Electron Photon Electron
Refer to fault book Refer to fault book Refer to fault book Refer to fault book Refer to fault book
Time
RE or MP Signature
PARAMETER √/x Comment √/x Comment √/x Comment √/x Comment √/x Comment
(a) Digital Couch with 100cm SSD (+/-0.2cm)
Readout = 0
(b) ODI Readout (+/-0.2cm)
ODI @ Gantry = 0
ODI @ Gantry = 50
ODI @ Gantry = 310
(c) Field Size (+/- 0.2cm)
Asymmetrical
10 x 10cm
Symmetrical
10 x 10cm
Symmetrical
15 x 15cm
(d) Collimator Angle (+/-1degree)
Coll. Rtn. +/-1 degree
(e) Isocentre Lasers [Red] (+/-0.2cm)
Lateral
Sagittal
Overhead
f) MV Imager Geometry (+/-0.2cm)
15cm x 15cm
(g) Set-up Lasers [Green] (+/- 0.2cm)
Parallel to Red
20cm below Red
h) OBI Couch Translations (+/-0.2cm)
Longitudinal
Lateral
Vertical
(i) CBCT
Functional
Exports to ARIA
RT SIGNATURE
RT NAME
DESIGNATION
The aims of the Daily Linac Geometry Quality Assurance checks are to ensure:
correct daily handover procedures are observed prior to treatment delivery
Linac geometry meets acceptable limits
the simulated treatment geometry can be accurately reproduced on a daily basis at the
Linac
Success is entered on the chart as a tick (√). Failure is entered as a cross (x). This procedure
assumes IEC1217 scales for all machine mechanical parameters.
3.2.1 Responsibilities
3.2.1.1 Radiation Engineering Staff (RE)
Prior to the daily release/handover of the Linear Accelerator, the RE staff will perform
morning run-up to ensure the Linac is free from faults and dose output, for all beam
energies, is within acceptable limits.
The RE staff member will complete the checkmark(s), time and sign the appropriate section
of the Linac Daily Geometry Quality Assurance Log to indicate the Linac is fit for clinical use.
Details of any failures, faults or issues will be documented in the Linac Fault book.
If any RE Staff member does not want the Linac to be used clinically, they will:
record the relevant information in the ‘Linac 3 Fault Book'
place a yellow ‘OUT OF SERVICE” notice on the control console key of the Linac
indicating that the machine is not to be used inform the Radiation Therapists personally
(if present)
The following documentation contains procedural instruction and the current acceptable
limits for daily geometric quality assurance checks.
If any geometry or functionality failures are detected during this process, the RT staff must
contact the RE staff to have the issues rectified prior to clinical use. RT staff are required to
log any failures in the Linac Log/Fault Book. RT staff are only to take responsibility for the
Linac after notation is dated and recorded in the Linac Log/Fault Book, from either RE or
Medical Physics, stating the Linac is “Fit for Clinical Use”
Check the Daily Geometry Quality Assurance Record; identify the RE staff’s entry and
signature indicating Dose Output Checks have been successfully performed.
Success is entered on the chart as a tick (√). Failure is entered as a cross (x).
A failure must be immediately followed up with the RE staff member on duty and
clinical treatment must not proceed until this criterion is met.
12. Rotate Gantry to 50 degrees and 310 degrees to check the 100cm ODI reading is
stable at both angles
13. Enter success or failure of above on the chart
Lateral Lasers
1. Carefully rotate the Iso-Align faceplate to a face the 90 degree
2. Confirm the lateral lasers align with the crosshairs marked on each side of Iso-Align
Faceplate
3. Rotate the gantry to 270 degrees, then 90 degrees
4. Check the Linac collimator crosshairs align with the Iso-Align Baseplate crosshairs at
each gantry angle
5. Enter success or failure on the chart.
Sagittal Lasers
1. Check that the sagittal laser aligns with the scribed mark on the foot end of the Iso-
Align Faceplate.
2. Enter success or failure on the chart.
m) MV Functionality/Geometry Test
The limit for misalignment of the field size is 0.2cm.
Inside Bunker
1. Ensure the Iso-Align Faceplate is rotated to 90 degree vertical position
2. Rotate the Gantry to 90 degrees
3. Ensure the Linac crosshairs correspond with the crosshairs marked on the Iso-Align
Faceplate
4. Maintain 100cm SSD to the surface of the Iso-Align Faceplate
5. Set 15cm x 15cm field size on digital readout
6. Move the MV Imaging arm into acquisition position
Outside Bunker
1. Open the patient named “T-LA2 Morning QA”
2. Mode up the planned image for “Field 1 – Gantry 90”
3. Acquire both the planned and open exposures
4. Ensure the 15cm x 15cm irradiated field size corresponds with the 15cm x 15cm
markers on the Iso-Align Faceplate, as in Figure 2. If the red irradiated field falls
beyond the outermost or innermost white radio opaque markers delineating the
15cm x 15cm field size, notify RE and/or Medical Physics
5. Enter success or failure on the chart.
o) Correlation of longitudinal, lateral and vertical physical couch motion with digital
readout.
Limit = +/- 0.2cm
1. With the Gantry on 0 degrees
2. Raise the couch so that the lateral isocentre (red) lasers correspond with the central
axis markings on the Iso-Align Faceplate
3. Using the Digital Couch Readout, raise or lower the couch 10cm
4. Confirm the lateral isocentre (red) laser now runs through the 20cm markings on the
upper or lower surface of Iso-Align Faceplate.
5. Move the couch back so that the lateral isocentre (red) lasers correspond with the
central axis markings on the Iso-Align Faceplate.
6. Using the Digital Couch Readout, move the couch 10cm in or out from the gantry.
7. Confirm the lateral isocentre (red) laser now runs through the 20cm markings on the
outer or inner surface of Iso-Align Faceplate.
8. Move the couch back so that the lateral isocentre (red) lasers correspond with the
central axis markings on the Iso-Align Faceplate.
9. Carefully rotate the Iso-Align faceplate to a horizontal position.
10. Using the Digital Couch Readout, move the couch 10cm left or right.
11. Confirm the lateral isocentre (red) laser now runs through the 20cm markings on the
lateral surface of Iso-Align Faceplate.
12. Enter success or failure on the chart.
DATE
Photon Electron Photon Electron Photon Electron Photon Electron Photon Electron
Refer to fault book Refer to fault book Refer to fault book Refer to fault book Refer to fault book
Time
RE Signature
PARAMETER √/x Comment √/x Comment √/x Comment √/x Comment √/x Comment
(a) Digital Couch with 100cm SSD (+/-0.2cm)
Readout = 0
(b) ODI Readout (+/-0.2cm)
ODI @ Gantry = 0
ODI @ Gantry = 50
ODI @ Gantry = 310
(c) Field Size (+/- 0.2cm)
Asymmetrical
10 x 10cm
Symmetrical
10 x 10cm
Symmetrical
15 x 15cm
(d) Collimator Angle (+/-1degree)
Coll. Rtn. +/-1 degree
(e) Isocentre Lasers [Red] (+/-0.2cm)
Lateral
Sagittal
Overhead
f) MV Imager Geometry (+/-0.2cm)
15cm x 15cm
g) Set-up Lasers [Green] (+/- 0.2cm)
Parallel to Red
20cm below Red
h) Couch Translations (+/-0.2cm)
Longitudinal
Lateral
Vertical
RT SIGNATURE
FULL NAME
DESIGNATION
This section describes the correct shutdown sequence for all Linear Accelerators.
Doc Number Version Issued Review Date Area Responsible Page
CHHS16/204 1 02/11/2016 01/09/2019 CACHS - Radiation Oncology 41 of 117
Do not refer to a paper based copy of this policy document. The most current version can be found on the ACT Health Policy Register
CHHS16/204
4.1 Procedure
This procedure is to be followed when routine shutdown is performed at the end of clinical
operation each day.
NOTE LA3+4 Only: The Emergency Door Release Key is kept on the Linac key ring. It is not to
be kept in the lock itself as there is a risk of breakage.
LA4 Only:
8. Attach the charging power cable to the robotics couch charging outlet. Ensure the green
battery charging indicator on the side of the couch is flashing.
9. Turn on the ExacTrac Robotics using the Emergency Robotics Red button
10. Ensure the battery charger LED is illuminated green
11. Ensure all three bards of the battery LED are blinking
14. Remove the key from the Electronics Cabinet and store both Linac keys and the
Emergency Door Release key on the LA4 key ring in the Radiation Therapy Key Cupboard
in the Planning Room
15. Shutdown all other ARIA and ACTGov computer applications and PC’s
16. Switch off the Night Switch which is located above the CCTV monitors
This section describes the requirements for undertaking Pre-Treatment Verification Checks,
hereafter referred to as V1 and Day 2 Treatment Verification Checks using a plan PDF
uploaded in ARIA11.
5.1. Alerts
Check marking the items within this Pre-Treatment Verification Template indicates you have
verified each item in its entirety in accordance with each criterion described below.
Enter the Course, Site and Phase details in the Template Name section of the Dynamic
Document in the following format “ C1_Site_V1”
Repeat this step for the second window you want to view, but by dragging it to the
opposite of the screen.
Both windows will now be displayed side by side allowing easy comparison of
parameters and any other relevant information.
NOTE: Some of the smaller computer monitors may not split the screens evenly, leaving
some overlap of the two windows in the middle. You may choose continue like this or
manually drag the inner borders of each window to fit to your liking.
Ensure the Plan PDF has been approved by the RO. This electronic password protected
approval serves as the Radiation Oncologists signature. Treatment must not proceed until
this item has been check marked.
Fractionation indicated
Ensure the Radiation Oncologist has indicated the number of treatments per fortnight. If 10
per fortnight has this been noted on the treatment sheet add a yellow sticker to the front of
the treatment sheet and the blue file.
Tattoos indicated
Ensure all tattoos, both new and previous, are clearly indicated. Ensure their location in
relation to the isocentre is clear.
Machine ID
Ensure the Machine ID for each field in ARIA corresponds with the machine in which that
patient will be predominantly treated on.
Energy/Mode
Using the Plan PDF as reference, ensure the Energy/Mode in ARIA and the Treatment Sheet
are correct.
Dose Rate
Ensure the dose rate correctly selected/populated in ARIA for each field.
MU
Using the Plan PDF as reference, ensure the MUs for each field in ARIA and on the treatment
sheet are correct.
Time
Ensure the treatment time documented on the treatment sheet is correctly calculated on
the treatment sheet and matches in ARIA. The correct computation of time for:
IMRT is MU’s divided by 80
Forward Planned is MU’s divided by 300
all static beams is MU’s divided by 450
Tolerance Table
Ensure each beam has the correct Tolerance Table attached in both the Treatment Sheet
and in ARIA. Set-up fields should have the same Tolerance table as the treatment fields.
Refer to Section 26 of the External Beam Operations Manual for the tolerance table
parameters.
Planned SSD
Using the Plan PDF as reference, ensure the FSDs for each field match the Treatment Sheet
and ARIA.
Gantry Rotation
Using the Plan PDF as reference, ensure the Gantry Angle for each beam matches in the
Treatment Sheet and in ARIA.
Collimator Rotation
Using the Plan PDF as reference, ensure the Collimator angle for each beam matches the
Treatment Sheet and ARIA.
Jaw Settings
Using the Plan PDF as reference, ensure the X1, X2, Y1 and Y2 jaws for each beam match the
Treatment Sheet and ARIA.
MLC
Doc Number Version Issued Review Date Area Responsible Page
CHHS16/204 1 02/11/2016 01/09/2019 CACHS - Radiation Oncology 50 of 117
Do not refer to a paper based copy of this policy document. The most current version can be found on the ACT Health Policy Register
CHHS16/204
Are the MLCs (If any) are indicated on the Treatment Sheet and in ARIA for the each beam.
Wedge Angle/Code/Orientation
Using the Plan PDF as reference, ensure the Wedge Angle, Code and Orientation match the
Treatment Sheet and ARIA.
Electron Applicator/Insert
Using the Plan PDF as reference, ensure the Electron Applicator matches the Treatment
Sheet and ARIA. Ensure the Electron Insert is entered on the Treatment Sheet in Red Ink.
Bolus
Ensure bolus (if any) is indicated on the Treatment Sheet field parameters section in green
ink. Ensure a Bolus Interlock has been activated in ARIA for each beam containing Bolus.
Floor Rotation
Using the Plan PDF as reference, ensure the Floor/Couch Rotation for each beam matches
the Treatment Sheet and ARIA.
Daily Dose
Ensure the daily dose to each Reference Point corresponds with the daily dose recorded on
the prescription page.
Total Dose
Ensure the Total Dose for each Reference Point equals the total prescribed dose(s).
Breakpoint
Ensure there are Breakpoints for all Reference Points at the correct doses/fractions in the
Treatment Sheet and in ARIA.
Imager Offset
Ensure the imager (kV or MV) offset is calculated to capture the appropriate anatomy
needed for matching. Enter imager offset using the Imager Vrt, Lng and Lat Values in Plan
Parameters (or Treatment Preparation). The imager offset program located on the G drive
can be used for MV offset for asymmetrical fields.
5.2.8 Other
Plan Treatment Approved
Ensure that every plan that has been validated and treatment approved in the plan
parameters tab.
requirements for the course of treatment. Refer to section 11 for RT billing codes of the
External Beam Operations Manual.
Gown Organised
Has a gown (if necessary) been organised? Consider the patient’s mobility/ability to getting
changed and the site being treated.
V/V1 Signature
Are the initials of staff member who entered the plan data into ARIA on the Treatment
Sheet in the ‘By’ column of the ARIA Entries section? Have you placed your initials next to
this in the ‘V1’ column?
NOTE: If there are any additional comments that need to be documented, they may be
done so in the grey text box next to ‘Notes’ found underneath the V1 section of the
document.
This will stamp the document with your name, the date, and the time whilst allowing the
document to stay open for the staff member performing the Day 2 Verification Check.
NOTE: By ticking any item within this Dynamic Document indicates you have verified each
item in its entirety as described for each one below. V1 ‘Signed Off’
Ensure the staff member who completed the V1 has performed a ‘Sign Off’ in ARIA. If so,
their name and the date/time of when this was done will be stamped at the bottom of the
V1 section of the document.
MLC’s Checked
Ensure a qualified RT’s signatures or initials are present on each of the MLC printouts to
indicate that they have been checked. Ask the staff who treated Day 1 if unsure. MLC
templates should be disposed of in the confidential waste bins after the V2 check.
If there are any additional comments that need to be documented, they may be done so in
the grey text box next to ‘Notes’ found underneath the V2 section of the document.
To verify that the treatment parameters transcribed to the treatment sheet are consistent
with the actual machine settings for a patients treatment, to ensure safe and accurate
delivery of radiation therapy treatment to patients.
For patients who have daily imaging or pre treatment imaging on the first fraction the
following parameters should be checked at the treatment console after imaging has been
performed, but prior to treatment field delivery. For patients who do not have pre
treatment imaging on the first fraction the following parameters should be read out in the
treatment bunker prior to leaving.
Collimator Angle
Gantry Angle
Each Jaw Setting
Couch Rotation (if applicable)
Wedge Angle
Wedge Orientation
MLC placement
Treatment Site
Isocentre Location
Set-up FSD
ATT (if applicable)
Wedge Angle (if applicable)
Electron Applicator and Insert (if applicable)
Bolus + Thickness (if applicable)
Other personalised accessory equipment
NOTE: 1) The treatment sheet is regarded as the primary source of treatment information
2) The treatment sheet must be taken into the bunker for all patient set-ups
This section describes the requirements for undertaking a Weekly Chart Check using the
Chart QA module of ARIA11.
7.1 Alerts
Each RT must ensure they are logged into ARIA using their own login account.
Placing a Checkmark against each of the criteria indicates that the item has been
completed and/or follow-up action initiated.
RT Notes
Read through all the RT Journal entries and ensure all documented items have been
appropriately actioned.
Patient Alerts
Review all patient alerts and ensure they have been appropriately actioned.
An occasion of service needs to be added in CAS for all Radiation Oncologist, and
registrar reviews. The Radiation Therapist will write a note in the treatment sheet
indicating the date the patient was seen, who saw the patient and sign the "Action By"
column so that other Radiation Therapist's do not have to follow them up again. i.e.
12/06/2015 Seen by Dr XXX....... RTs sig.
Once Journal entries have been reviewed, the menu item must be ticked off in the RT-
Weekly Chart Check Encounter. Ticking items off indicates that you have read all the
relevant entries and initiated any required follow-up action.
NOTE: If the correct type (i.e. Radiation Oncologist) is not selected when the journal note is
created then the note will not be correctly displayed in the Encounters tab, please ensure
you navigate to the Journal tab to ensure that all journal notes are reviewed.
Review each column to ensure that the planned data (blue) matches the actual data for
the following fields:
o MU
o Energy
o Tolerance Table
o Gantry Rotation
o Collimator Rotation
o Field Size
Repeat the process of checking field parameters for each field using the Data Filter.
Arrange for repeat FSD checks to be performed if anomalies are identified. A patient
alert may be used to request repeat FSD’s.
Return file to planning for re-calculation/replanning if 3 days of FSD changes outside
department tolerance.
Return to the Encounters tab
Tick the box to indicate this item has been completed
Analyse the last 3 images to identify the need for correction of any systematic shifts in
the longitudinal, lateral and vertical directions.
Ensure the systematic corrections to the isocentre have been correctly transcribed to
the treatment record.
Select Close.
Select Back
7.9 Other
7.9.1 Document Review
Select Document Review from the Encounter Menu
Review the current Course documents
Ensure that Treatment related documents have been approved.
o Initiate any follow-up action
If this is the first Weekly Chart Check, review the V1/V2 documents and ensure that any
comments made have been actioned
Tick the box to indicate this item has been completed
Search for the patient using surname, first name and ID1
Ensure the following columns are displayed:
o Patient o Hospital Procedure Code
o Activity o Staff/Resources
o In Patient Status o Diagnosis
o Course ID o Reviewed
o Date of Service o Reviewed By
o Procedure Code o Primary
o Oncologist
Check the:
o Inpatient status is correct
o Correct Medicare Codes have been billed.
o Play close attention to the commencement of new phases etc, where codes may
have been incorrectly auto filled.
o Ensure the Patient Status is set to New Course, or Retreatment Course, for the Initial
Fraction
o Ensure the “N”(New Course) or “S”(Subsequent/Retreatment Course) suffix is
attached to the site code for the first fraction.
o Correct number of fields for each treatment have been billed
o Treatment Linac is specified
o Imaging code is correct and attached to the correct fraction
DEFINITIONS: For definitions of Inpatient status, Patient Status, Procedure Codes, Imaging
Codes please refer to Section 11 of the External Beam Operations Manual
NOTE: When changing or updating a billing code ensure the date of Service is adjusted as the
software defaults the date to the current date.
NOTE: Whilst the paper based treatment sheet is in use, all electronic journal entries must
be duplicated in the treatment sheet
This section outlines the steps involved in performing a Course Completion Check at the
conclusion of a patient’s course of External Beam Radiation Therapy
Signature Check
Ensure all required signatures are recorded on the treatment sheet
Ensure any signatures missing are followed up
Task Review
Select the Tasks Menu Item and ensure:
o Review all tasks to ensure they have been set to complete.
o Follow up any anomalies and delete any excess or unrequired tasks
Patient Alerts
Select Patient Alerts from the Encounter
Ensure all Patient Alerts have been signed off
Follow-up an outstanding patient alerts
Templates
Ensure any treatment templates have been scanned into ARIA prior to disposal
Ensure the template, once scanned in, is disposed of in the confidential waste bin
Disposal of Accessories
Ensure all electron cut-outs are clearly marked with an “F” or “Finished” and returned to
the Radiation Engineering Mould Room
Ensure all masks and accuforms are de-identified and disposed of in the designated bin.
The bin will be stored in the treatment storage area. Arrangements will be made for the
bin to be cleared as required.
Doc Number Version Issued Review Date Area Responsible Page
CHHS16/204 1 02/11/2016 01/09/2019 CACHS - Radiation Oncology 69 of 117
Do not refer to a paper based copy of this policy document. The most current version can be found on the ACT Health Policy Register
CHHS16/204
Ensure any custom bolus or other treatment accessories are de-identified and disposed
of correctly
This section describes the day to day use of the Appointment Scheduling Workspace within
Radiation Oncology Treatment section.
For a detailed overview of application functionality and icons refer to the Radiation Oncology
Appointment Scheduling Reference Manual
9.1 Alerts
The Read Only Account for each Linac is to be used for viewing/tracking purposes. Making,
changing appointments adding notes and completing tasks in this account is disabled. Switch
Users to log into personalised accounts with adequate rights to perform required editing
functions.
Leaving applications logged in under an individual’s account for all staff to utilise full
functionality is not permitted as it breaches ACT Health IT Security Policy. Under no
circumstances are accounts to be left logged in for others to use.
Tasks are not allocated to an individual; rather the entire team will work together to
complete all of the Available Tasks
To evenly spread the administrative duties throughout the working week, tasks are to
be completed as they become available. Any overdue task will be escalated to the RT
Management Team
A Weekly Chart Check element has been created to ensuring the sequencing and
frequency of tasks is appropriate. If a task needs to be added into the Carepath it is
essential to review the Visual Carepath to ensure the order of activities is correctly
sequenced and the link is placed in the correct place
9.6 Documentation
Any note or documentation that would be normally recorded in a patient’s medical file
is to be recorded as a Journal Note as well as on the paper based treatment sheet
(whilst in use)
Any urgent notes or situations affecting dosimetry are to be entered as a Patient Alert
Appointment notes should be used as described above
RO notes will be documented on the treatment sheet as is current practice, unless
formally advised otherwise by memorandum
If the Radiation Oncologist or their delegate advises that the patient is not for treatment
o the patient’s appointment status must be set to cancelled
o an appropriate note attached to the appointment (e.g. As per Dr.X, patient not for
treatment- see paper based treatment sheet for note”)
o an additional appointment is to be scheduled
o review the Carepath and ensure the sequencing of appointments and tasks is
correct.
If the Radiation Oncologist or their delegate advises that the patient is for no further
treatment
o the patient’s appointment status on the day of notification must be set to cancelled
o an appropriate note attached to the appointment (e.g. As per Dr.X, patient not for
further treatment- see paper based treatment sheet for note”)
o all remaining appointments should be deleted
In the window that appears enter the patient’s MRN. If the MRN is not known, search
for the patient using the Patient Explorer tool
o Click the button to the right to bring up the Patient Explorer Window
o Adjust the date range, ensuring it is adequate enough to cover the full length of the
patient’s course
This section guides the clinical staff in the workflow related to the Treatment Review booked
in ARIA
The patients name for which the review is scheduled is clearly displayed within the
appointment
The appointment time and duration is indicated both on the side of the workspace and
within each appointment.
If it is their appointment time and this symbol is not yet present, the patient may not
have arrived yet. Staff are advised to check all patient waiting areas before calling the
patient. If the patient is in the waiting room and has forgotten to check in, re-education
may be required.
If the check-in symbol is present but the patient cannot be found in the waiting area, the
patient may be attending an unscheduled appointment such as a dietician. Alternatively
the appointment may have been accidently checked-in by a staff member.
All information regarding a patient’s treatment progress including fraction number, dose
delivered and diagnosis can be found in the Patient Manager workspace of ARIA.
Upon completion of the patient review, select ‘Back’ at the top of the Patient Manager
workspace. This will return the user to the Appointment Scheduling workspace.
Select the patient’s appointment then select the Complete button at the top.
10.3.2.5 Non-Attendance
If the patient does not attend for their Treatment Review, the appointment must be
cancelled in ARIA.
At the completion of the day, ensure all appointments have either been completed or
cancelled.
The RO Journal Note Review Tab will display patients from all machines. It is therefore the
responsibility of the machine staff to regularly check this space for their patients as they
become available throughout the day.
Radiation Therapists will regularly review the RO Journal Note Review Tab from their User
Home workspace to follow-up/note Journal entries added by the Radiation Oncologist as
part of the treatment review.
IMPORTANT NOTE: It is imperative that each user is logged in to their own ARIA account to
complete this task. By Selecting ‘Done’ this serves as a formal signoff that all notes have
been actioned and the task is complete
NOTE: All patient journal entries must be reviewed and signed off prior to the subsequent
treatment fraction
11.1 Alerts
Radiation Therapists are responsible for reviewing billing codes entered in ARIA.
Administration staff will be responsible for checking accuracy of data translated from ARIA
to the third party billing interface.
Patient Status: for the first fraction ONLY select either New Course or Retreatment Course
from the drop down list using the summarised NSW Health Definitions for Radiotherapy
Data Collection below. For subsequent fractions in the same course the patient status should
be left as Course.
New Course: the FIRST course of Radiation Therapy delivered to a patient for a new Primary
diagnosis or related metastasis, to which neither have been previously treated with
Radiotherapy in ANY facility.
Retreatment Course: a SECOND course of Radiation Therapy after the patient has received
Radiation Therapy to any site related to the same primary diagnosis. May include:
metastatic disease if the primary has already been treated with a course of Radiation
Therapy.
metastatic disease if the primary has not been treated but previous metastatic disease
related to the same diagnosis has been treated.
Important:
‘A patient can receive more than one course of radiotherapy treatment at the same time
(i.e. courses which are simultaneous or which overlap). These courses may have the same or
different ready-for-care dates and the same or different radiotherapy start dates’.
If there is any uncertainty as to the appropriate classification of the course, refer to the
examples cited in the NSW Health Definitions for Radiotherapy Data Collection
documentation.
NOTE: An Inpatient is a patient admitted to The Canberra Hospital only or Medi hotel only.
Patients admitted and transferred from other hospitals e.g. National Capital or Calvary are
considered outpatients for the purposes of our billing. This section will default to Outpatient
unless the Inpatient check box is ticked in Patient Summary in which case the default option
becomes Inpatient.
Date of Service: confirming the date as being the correct date of service provision
Qty: confirm or edit to ensure this value is set to “1” for each entry. This field does not
indicate the quantity of additional fields treated. The hospital procedure code indicates the
number of fields being billed
Initial Field Procedure Code: confirming or attaching the correct site code for the Initial
treatment field. On the first treatment fraction for each course of the appropriate ‘N’ (new
course) or ‘S’ (subsequent/retreatment course) site code must be selected. Refer to Table 2
for all site specific MBS codes.
Additional Field Procedure Codes: confirming or attaching the correct site code and number
of subsequent fields
Image Verification Codes: selecting the appropriate image verification code (if applicable)
according to the summary of the Medicare Benefits Schedule Book listed below in table 1:
Table 1:
MBS
Code: Suffix: Subgroup 7 RADIATION ONCOLOGY TREATMENT VERIFICATION
15700 Single projection with single or double exposures
Multiple projection Images are taken in more than one plane e.g. orthogonal views
15705 acquisition Treatment must involve three or more fields
TABLE 2
Medicare Benefit Schedule- RADIATION ONCOLOGY TREATMENT
MBS
Code: Site: Subgroup 3 Megavoltage:
15248N Prostate New Course
each attendance at which treatment is given - 1 field - treatment delivered to primary site (prostate)
15248S Prostate Subsequent Course
each attendance at which treatment is given - 2 or more fields up to a maximum of 5 additional fields
15263 Prostate Additional Field/s (rotational therapy being 3 fields)
15245N Lung New Course
each attendance at which treatment is given - 1 field - treatment delivered to primary site (lung)
15245S Lung Subsequent Course
each attendance at which treatment is given - 2 or more fields up to a maximum of 5 additional fields
15260 Lung Additional Field/s (rotational therapy being 3 fields)
15251N Breast New Course
each attendance at which treatment is given - 1 field - treatment delivered to primary site (breast)
15251S Breast Subsequent Course
each attendance at which treatment is given - 2 or more fields up to a maximum of 5 additional fields
15266 Breast Additional Field/s (rotational therapy being 3 fields)
15254N Other New Course each attendance at which treatment is given - 1 field - treatment delivered to primary site for diseases
15254S Other Subsequent Course and conditions not covered by items 15245, 15248 or 15251
each attendance at which treatment is given - 2 or more fields up to a maximum of 5 additional fields
15269 Other Additional Field/s (rotational therapy being 3 fields)
15257N Secondary New Course each attendance at which treatment is given - 1 field - treatment delivered to secondary site
15257S Secondary Subsequent Course
each attendance at which treatment is given - 2 or more fields up to a maximum of 5 additional fields
15272 Secondary Additional Field/s (rotational therapy being 3 fields)
Applies the initial Field and all
additional fields. with IGRT imaging facilities undertaken:
N/ S suffix should be applied at the (a) to implement an IMRT dosimetry plan prepared in accordance with item 15565;
15275 IMRT commencement of a course
The aim of the Monthly Accessory Equipment Condition Check is to ensure that accessory
equipment is clean, functional and in good repair to minimise the potential risk of injury to
patients, staff and other stakeholders.
12.1 Alert
All equipment, including that not specifically identified in this procedure is subject to
ongoing inspection on each occasion of use. Any employee who observes a hazard, such as
equipment damage or malfunction, which may lead to injury or incident, must take
necessary action to avoid, eliminate or minimise hazards.
12.2 Procedure
The procedure is to be performed by the Linac Supervisor or delegate on a monthly basis
for all accessory equipment identified in the Accessory Equipment Monthly Condition
Checklist (Checklist).
Each accessory item must be visually inspected for cleanliness, general wear, tear and
damage that may affect function. Inspection must also consider the risk of harm, injury for
patients, staff members and other stakeholders.
Radiation Therapists (RT) must use professional judgement in determining and rating the
condition of accessory equipment as either:
Good (G) - defined as no obvious damage identified.
Reasonable (R) - defined as minor wear and tear only.
Needing Repair (NR) - physical damage detected.
Any wear and tear or damage must be recorded on the Checklist and reported to the Linac
Supervisor who will initiate any necessary follow-up action, either immediate or proposed.
Where an item is rated (R) or (NR) a description of the damage must be recorded in the
Notes Section of the Checklist.
Where an item is rated (NR), the item must be removed from service until repair has been
undertaken or a replacement has been sourced.
At the end of each calendar year the Linac Supervisor will forward all completed Checklists
to the Head of Treatment for their collation and storage. All records will be stored for three
years before being destroyed.
Condition
Accessory Items Notes
G R NR
Med Tec Pronefix (Red)
Includes inspection for damage and general wear & tear
White Base Plate (WBP)
Iinspection for damage/ general wear & tear - incl locking clips
Breast Board
Includes inspection for damage and general wear & tear -
particularly locking clips
15° Styrox Wedge (2)
Includes inspection for damage and general wear & tear +
includes inspection of covering
Rice Bags
Includes inspection for damage and general wear and tear
Bed Ladder Strap
Includes inspection for damage and general wear and tear;
includes inspection of attachment and security
Yoga Mats
Includes inspection for damage and general wear and tear
Red/Green handgrip/donut
Condition
Accessory Items Notes
G R NR
30cm x 30cm 1.0cm thick
Includes inspection for 2 @ LA3
damage and 30cm x 30cm 0.5cm thick
general wear and tear 10cm x 10cm 1.0cm thick
Clam Shells
Includes inspection for damage and general wear and tear
LA1 (one set only)
Iso-Align QA Box and CT Fans
Includes inspection for damage and general wear and tear of
QA box, spirit level, frontpointer and mount & OBI QA cube,
and full and half CT Fans
Magnet
Camera
Med Tec Baseplate & > B’plate
Inserts B’Plate
5
LA 1, 2
LA 3, 4 10
30
Headrests B
C
D
E
Includes inspection for damage and general
wear & tear F
------------------------------------------------------------------------------------------------------------------------
-------------------------------------------------------------------------------------------------------------------------
-------------------------------------------------------------------------------------------------------------------------
-------------------------------------------------------------------------------------------------------------------------
-------------------------------------------------------------------------------------------------------------------------
-------------------------------------------------------------------------------------------------------------------------
This section provides guidance in the situation that a treatment session is unable or should
not be completed as planned in ARIA to ensure the safe delivery and correct accumulation
of dose for each treatment site.
The patient should be closed down from the 4D Treatment Console (4DTC).
The option the resume should be selected.
Figure 1
The plan should be opened in ARIA
THE STATUS OF THE PARTIAL TREATMENT MUST BE RESOLVED AS A FIRST PRIORITY.
Figure 2
NOTE: Any alterations to the plan should be reviewed by a second fully accredited
Radiation Therapist prior to approval of the plan.
The revised plan then needs to be planning approved and treatment approved.
The patient should be closed down from ARIA
The partial treatment can then be loaded from the 4DTC again.
o The fields that were treated in the original plan/s should still be ticked as
completed.
o The untreated fields on the original plan should now be locked.
o The already treated fields on the plan revision should be locked.
o The only available fields should be those fields in the plan revision that were not
previously treated in the original plan.
Figure 3
Once all fields have been delivered the patients should be able to be closed down as
usual.
Activity capture should auto load as per normal
The patient should be closed down from the 4D Treatment Console (4DTC).
The option the resume partial treatment and leave on queue should be selected.
The plan should be opened in ARIA
Figure 4
Figure 5
Activity capture should auto load as per normal, and the treatment session should now
be seen as complete
13.3 Appendix 1
A BrainLab couch top has been fitted to LA4 which is different from the Varian IGRT couch
tops on LA1, LA2 and LA3. Medical Physics have advised that the attenuation of the beam
through the BrainLab couch needs to be accounted for in the delivery of the treatment
plan.
14.1 Procedure
Calculation must be completed for any beams where 50% or more of the beam’s light field
is passing through the BrainLab couch top to account for the attenuation.
Example: If monitor units from dosimetric plan = 100 for 6x beam angle 185, then
new monitor units accounting for attenuation through the BrainLab couch
= 100 x 1.043
= 104
14.1.3 ARIA:
Current patients on treatment:
Complete a plan revision of the original plan
For the plan including the BrainLab couch top attenuation factor, rename the plan to
include the prefix LA4.
Change the MU’s on the LA4 plan to account for attenuation through the BrainLab
couch.
Attach this LA4 plan to the original reference point to ensure the dose accumulation
equals the prescription.
Planning approve as per normal.
A second Radiation Therapist must verify the MUs that require correction have been
calculated and correctly transposed in Aria than treatment approve.
On treatment setup, ensure the attenuation factor has been applied appropriately, i.e.:
applied for beams where 50% or more of the beam is passing through the couch top.
Action any identified corrections prior to treatment delivery. This may include the
application or removal of the attenuation factor.
Ensure that if the patient is to be treated on another machine, that another plan
revision is created and the MU’s corrected.
Where there are more than 4 cases that need correction consult with planning to take
over this task.
Cases that may not be treated on LA4 with the BrainLab couch top:
Single fractions – when beams are entering the patient through the couch top
Supine rectum
Supine chest patients with posterior lesions
SABR
Cases that can have treatment on LA4 BrainLab couch with the attenuation factor applied
for required beams:
Supine Cervix/ Endometrium/Gynae
Chest cases – (Note: excluding those with majority of weight and beams passing
through the couchtop.)
Three field breast
Tangent fields breast
4 field breast
Breast boost
Prostate
Rectum- prone
Cervix on belly board
Head and Neck
Brain
Ant and post palliative- not single fraction
Post palliative – not single fraction
This section describe the steps to deliver treatment to a left breast patient using the Deep
Inspiration Breath Hold (DIBH) technique
2. Switch on the power to the RPM camera (inside the internal cupboard with HDMI
cable)
3. Connect the Gating Goggles into the extension HDMI cable
4. Plug in the micro USB into the battery pack for the goggles (not available yet)
IMPORTANT: The goggles must be switched on when the PC shows the loading windows
screen during boot-up. To switch goggles on, press the menu button on the goggles
controller box/battery (see figure 1). A blue light on the controller box will indicate that the
video goggles are switched on. The goggles will switch off after 1 minute if no input is
detected. See trouble shooting for more information.
IMPORTANT: Verify patient I.D. between 4DCT, RPM computer, and patient treatment
sheet
3. The RPM system will automatically enter “planning mode”. To enter treatment mode,
press ‘Stop’ on RPM/gating computer.
4. Ensure gating key is turned on, and beam hold is glowing yellow.
Figure 1
6. Keeping the couch lateral at 0.0, move the couch long to the Up The Bed measurement
listed in the setup instructions. Check that the Reference Tattoo is within a 0.5cm
Tolerance Sup or Inf.
a) If the Reference Tattoo is outside this tolerance, have the patient move up or down
the bed as required.
b) Having the patient take in a deep breath, then exhale and relax can help achieve
this after they have wriggled.
7. Keeping the couch lateral at 0.0, straighten the patient through midline to the sagittal
laser
a) If the patients midline is not running through the sagittal laser, use the sheet
underneath the patient to move them to the centre of the Breast Board
b) If both therapists are unable to move the patient using this method, ask the patient
to move themselves in the appropriate direction
c) Check that midline of the KneeFix is also aligned through the laser
8. At the Reference Tattoo, set the Rotation FSD. Ensure that the Post Edge Tattoo is
aligning in BOTH the Sup/Inf and Ant/Post Directions. If this is not the case, roll the
patient accordingly.
a) Rotate the patient to match ipsilateral tattoo with the laser, re-setting the AP
rotational FSD.
b) Ensure that after rolling, midline in the centre of the bed has not been affected. If
so, move the patient back to midline before continuing.
c) Shoulder position can often cause a discrepancy in the Sup/Inf positioning of the
Post Edge Tattoo
Figure 2
9. If the patient is having their Supraclavicular Fossa or Axilla treated, turn their head
according to the measurements in the setup instructions. Ensure that BOTH the
Reference and Post Edge Tattoos still align.
10. With the patient straight and level, mark on the Marker Box location in midline from
Reference Tattoo. Place the INFERIOR EDGE of the Marker Box on this mark in midline.
11. Move the couch long Sup, according to the Junction tattoo measurement in the setup
instructions, to assess Junction Tattoo location. The tattoo should align in the Sup/Inf
Direction.
a) If this is not the case, the shoulder position and/or rotation will need to be
assessed and altered as required.
b) Ensure that when adjusting the shoulder position that you are not just moving the
skin to match.
12. Move back to Reference Tattoo.
Figure 3
2. The session’s settings should be set to auto record, however the record button can
also be selected.
3. Select “2 displays, Clone mode”, then restart the computer with the goggles switched
on
4. After restarting, the goggles will show a copy of the main monitor view
5. Again, open Matrox “Multi-Display Setup” and select “1 display + 1 feature display”
then restart the computer again
6. Ensure the goggles are switched on during computer boot up, especially when “Varian
loading” is being displayed (the computer must recognise the goggles as a second
monitor)
7. Then press Ctrl + alt + Z simultaneously, then click and drag to select a region to be
visualised through the goggles
8. The region selected needs to meet a minimum/maximum resolution. If this is not met,
then occasionally nothing will display on the goggles. If this occurs, simply reselect an
area to display, ensuring that a smaller or larger region is chosen
9. This process can be repeated at any stage to modify what the patient is able to see
through the goggles, but it is important to ensure the goggles are working before
getting the patient into the treatment room
10. If the goggles still don’t work, repeat steps 1-7 and consult a DIBH super user
YES – NO –
Treat patient and consult with Can the patient be sent away for a period
RadEng/RadPhys for fault resolution of time whilst the fault is resolved and
timeframe on the affected Linac treated later ON THE SAME DAY?
Appendix A
This Daily QA checklist is used to unsure the RPM system and video goggles are functioning
correctly, and should be completed daily before simulation or treatment of any DIBH
patients.
Task Checks Completed Comments
(tick)
Hardware Video Goggles
Locate all DIBH equipment
(including adequate battery HDMI cable in-room
charge)
Breathing phantom & marker box
This section provides guidelines to ensure Radiation Therapists are able to deliver services
to patients efficiently without undue delays caused by common dosimetry Interlocks during
treatment delivery.
16.1 Background
After review of common dosimetry interlocks recorded for LA2 and LA3, dosimetry
interlocks that require medical physics intervention and those that can be cleared by
qualified RTs were determined and documented using the following methodology:
1. The log books from LA2 and LA3 were reviewed to determine the most common
dosimetry interlocks that have occurred. Data of LA-1 and LA-4 were not collected as
LA-1 is out service and LA-4 recently commenced clinical operation and did not have
sufficient data.
2. There were six main combinations of dosimetry interlocks identified as shown in Table
1 (see Appendix 1).
3. Most dosimetry interlocks are minimally recorded in the Log Book as “Dosimetry
interlocks” and do not capture required interlock information.
4. During data collection, the XDRS interlock was found to appear with photon energies at
gantry angles between 240o-280o.
5. Dosimetry Interlocks, their description and options which can be cleared by qualified
Radiation Therapists were documented in Table 2 (see attachment 2).
16.2 Alerts
Qualified Radiation Therapists are authorised to clear dosimetry interlocks as described in
Table 2. Interlocks other than those specified in Table 2 are to be addressed by a Radiation
Engineer in the first instance. The Radiation Engineer will contact a Medical Physicist as
appropriate.
1. If any dosimetry interlock or combination of dosimetry interlocks described in Table 2
appears more than once (>1) during treatment beam-on then do not clear it and refer
to MPRE in the first instance.
2. If a dosimetry interlock described in Table 2 occurs more than twice (>2) in any one
day, MPRE must be contacted prior to proceeding with treatment delivery.
3. DOS2 and DSFA are the most critical Dosimetry Interlocks (which are highlighted with
red in Table 2). If DOS2 or DSFA Interlocks appear and terminate the beam, report to
MPRE and do not proceed with any treatment until the machine is cleared for use by
Medical Physics.
4. If the DOS2 interlock appears during the calibration cycle, it may be overridden by
using the dosimetry password. If the DOS2 interlock occurs more than two times for
the same patient, you must contact Medical Physics before proceeding with the
treatment.
16.3 Procedures
All dosimetry interlocks must be recorded in the Linac Fault/Log Book. As a minimum the
following must be recorded for each entry.
1. The dosimetry interlock or combination of dosimetry interlocks;
2. Collimator angle,
3. Gantry angle,
4. Beam energy.
An appropriately experienced Radiation Therapist at Level RT2.1 and above can clear the
dosimetry interlock or a combination of dosimetry interlocks described in Table 2 with the
current dosimetry password.
A Dosimetry Interlock described in Table 2 may only cleared by the designated Radiation
Therapist in reference to detailed alerts.
Table 1 Common Dosimetry Interlocks: Log book recorded events (LA-2 and LA-3)
2008 to 2012
2 XDRS 122 69
3 XDRS-EXT-RVBP-IPSN-MLC-DS12* 24 29
4 CLFA-ION1-ION2-EXQ1-EXQ2-EXQT-LVPS-XDR1 15 70
5 Dosimetry Interlocks (excluding DOS2 and DSFA) 19 12
6 DS12 5 11
DOSIMETRY AUTHORITY TO
SR. # DESCRIPTION
INTERLOCKS CLEAR
1 CKFA Calibration Cycle Interlock Yes
2 CKTO Check Time Out Interlock Yes
3 CLFA Calibration Failed Yes
4 CLTO Calibration Cycle Took Too Long Yes
5 DOS2* Dosimeter Interlock No
6 DS12 MU1 and MU2 are not Matching Yes
7 DSFA Actual Dose is Greater than set Dose No
8 ENSW Energy Switch Interlock Yes
9 EXQ1H Radial Symmetry Tolerance Interlock (2%) Yes
10 EXQ2H Transverse Symmetry Tolerance Interlock (2%) Yes
11 EXQTH Total Asymmetry (2%) Yes
12 ION1 Radial Ion Chamber Interlock Yes
13 ION2 Transverse Ion Chamber Interlock Yes
14 LVPS Low Voltage Power Supply Interlock Yes
15 SYM1L Radial Symmetry Yes
16 SYM2L Transverse Symmetry Yes
17 XDP1 Radial Dose Per Pulse Interlock Yes
18 XDP2 Transverse Dose Per Pulse Interlock Yes
19 XDR1 Radial Dose Rate Yes
20 XDR2 Transverse Dose Rate Yes
21 XDRS Servoed Dose Rate Yes
Peak Current Interlock (Target and Ion
22 YLDL Yes
Chambers)
Note:
* If the DOS2 interlock appears during the calibration cycle then it can be overridden
by using the dosimetry password.
This section includes the acceptable variance tolerance tables for the treatment couch and
Linac treatment parameters specific for each treatment technique.
17.1 Alert
Professional judgment should be used to guide decision making in relation to the
appropriate tolerance table on a case by case basis.
There are circumstances that an auto gantry movement is deemed unsafe. If there is any
uncertainty between the couch and Linac clearance, during the V1 check stage (Section 5) of
the External Beam Operations Manual the checker should ensure a Manual Gantry
movement is applied until after the initial treatment and all angles have been physically
checked. During the V2 verification check or any subsequent treatment the treatment
tolerance table maybe amended.
Policies
CHHS Patient Identification and Procedure Matching Policy
Work Health and Safety Policy DGD 12-036
Standards
Australian Commission on Safety and Quality in Health Care (2006). Ensuring correct
patient, correct site, correct procedure; http://www.safetyandquality.org/
Standard 12, Radiation Oncology Practice Standards 2011
Planning for the Best, Tripartite National Strategic Plan for Radiation Oncology 2012-2022
Radiation Oncology Practice Standards 2011
Planning for the Best, Tripartite National Strategic Plan for Radiation Oncology 2012-2022
Legislation
Radiation Protection ACT 2006
Radiation Safety Policy for ACT Government Health Directorate
References
1. ACPSEM Position Paper Recommendations for the Safe Use of External Beams and
Sealed Brachytherapy sources in Radiation Oncology, Ser No 157, ISSN 0158-9938
2. Iso-Align: Reference Guide, Civco, USA, 2011
3. Varian Medical Systems:
a. Trilogy Tx Instructions for Use, P/N 1000028159-08 September 2009
b. On-Board Imager (OBI) Reference Guide Advanced Imaging, P/N 13501007RO1B; Vs
1.4, April 2008
4. Radiation Therapy – Treatment – Use of LA4’s Robotic Couch Top for Standard
Treatment
5. NSW Health –New Definitions for radiotherapy data collection in NSW
www.health.nsw.gov.au
6. Varian Medical Systems Activity Capture Reference guide (June 2006)
7. Bell, L, Shakespeare, T & Willis, A 2008, ‘Importance of daily electronic portal imaging in
radiotherapy’, Journal of Medical Imaging and Radiation Oncology, vol. 52, no. 4, pp.
414-418.
8. eviQ Treatments Online 2014, Radiation Oncology, Breast, Invasive Breast Cancer,
Adjuvant Nodal Irradiation, EBRT, Cancer Institute NSW, viewed 31 July 2014,
https://www.eviq.org.au/Protocol/tabid/66/categoryid/178/id/1282/Radiation+Oncolo
gy%2c+Breast%2c+Invasive+Breast+Cancer%2c+Adjuvant+Nodal+Irradiation%2c+EBRT.a
spx
9. eviQ Treatments Online 2014, Radiation Oncology, Palliation, Spinal Cord Compression
(SCC), External Beam Radiotherapy, Cancer Institute NSW, viewed 31 July 2014,
https://www.eviq.org.au/Protocol/tabid/66/categoryid/190/id/296/Radiation+Oncology
%2c+Palliation%2c+Spinal+Cord+Compression+(SCC)%2c+External+Beam+Radiotherapy.
aspx
10. Fein, D, McGee, K, Schultheiss, T, Fowble, B & Hanks, G 1996, ‘Intra- and interfractional
reproducibility of tangential breast fields: a prospective on-line portal imaging study’,
International Journal of Radiation Oncology Biology and Physics, vol. 34, no. 3, pp. 733-
740.
11. Michalski, A, Atyeo, J, Cox, J & Rinks, M 2012, ‘Inter- and intra-fraction motion during
radiation therapy to the whole breast in the supine position: A systematic review’,
Journal of Medical Imaging and Radiation Oncology, vol. 56, no. 5, pp. 499-509.
12. Middleton, M, Rolfo, A, Medwell, S, See, A, Wong, J, Joon, M & Lynton-Moll, M 2006,
‘Online versus offline corrections: operation or evolution? A comparison of two
electronic portal imaging approaches for locally advanced prostate cancer’, The
Radiographer, vol. 53, no. 1, pp. 24-28.
13. Russo, G, Qureshi, M, Truong, M, Hirsch, A, Orlina, L, Bohrs, H, Clancy, P, Willins, J &
Kachnic, L 2012, ‘Daily orthogonal kilovoltage imaging using a gantry-mounted on-board
imaging system results in a reduction in radiation therapy delivery errors’, International
Journal of Radiation Oncology Biology Physics, vol. 84, no. 3, pp. 596-601.
14. Smith, R, Bloch, P, Harris, E, McDonough, J, Sarkar, A, Kassaee, A, Avery, S & Solin, L
2005, ‘Analysis of interfraction and intrafraction variation during tangential breast
irradiation with an electronic portal imaging device’, International Journal of Radiation
Oncology Biology Physics, vol. 62, no. 2, pp. 373-378.
15. On target: ensuring geometric accuracy in radiotherapy. The royal college of
Radiologists/ Institute of Physics and Engineering in Medicine, Society and College of
Radiographers, 2008
16. Using Portal Vision With 4D Treatment Console, Training Solutions 2008, Varian Medical
System, USA, 2008
17. OBI 1.5 Therapist Operations rev 3.1.0 , Varian Medical Systems, USA 2006
18. OBI 1.5 Advanced Imaging Reference Guide, Varian Medical Systems, USA, March 2013
19. OBI 1.5 Therapist Operations rev. 3.1.0, Varian Medical Systems, USA 2006
20. Varian Medical Systems. (2012, November). ARIA 11 Offline Review Workbook.
Retrieved from MyVarian.
21. Christopher F Njeh, Jason Parker, Joseph Spurgin and Elizabeth Rhoe. A validation of
carbon fiber imaging couch top modelling in two radiation therapy treatment planning
systems: Philips Pinnacle3 and BrainLAB iPlan RT dose, , Rad. Onc 2012, 7:190
22. Jan K.H Seppala, Jarmo A.J Kulmala. Increased beam attenuation and surface dose by
different couch inserts of treatment tables used in megavoltage radiotherapy. J Appl
Clin Med Phys. 2011: Vol 12, No 4.
23. Christopher F Njeh, Timothy W Raines, Mark W Saunders. Determination of the photon
beam attenuation by the BrainLAB imaging couch:angular and field size dependence. J
Appl Clin Med Phys. 2009: Vol 10, No 3
24. Memo from Medical Physics and Radiation Engineering - LA4 MU Correction Factors for
the BrainLab Couch. Dated 28th August 2013
25. The dosimetry interlock information was collected from the Log Books provided by the
Radiation Engineering Department.
26. Appendix F, “Trilogy Tx Instructions for Use”, Varian Medical Systems, P/N 100028159-
08
Definition of Terms
Search Terms
Disclaimer: This document has been developed by ACT Health, Canberra Hospital and Health Service specifically
for its own use. Use of this document and any reliance on the information contained therein by any third party
is at his or her own risk and Health Directorate assumes no responsibility whatsoever.