N1996A-01C
S E R V I C E N O T E
Supersedes:
N1996A-01B
(Refer this Service Note if the A4 Digital IF assembly has not been applied with the Aluminum
Foil Taping)
Parts Required:
Part# Description Quantity
ADMINISTRATIVE INFORMATION
SERVICE NOTE CLASSIFICATION:
MODIFICATION RECOMMENDED
ACTION ON SPECIFIED FAILURE STANDARDS
LABOR: 3 Hours
CATEGORY: [[]] AGREEABLE TIME
LOCATION [[]] CUSTOMER INSTALLABLE SERVICE RETURN USED [[]] RETURN
CATEGORY: [[]] ON-SITE PARTS: SCRAP
INVENTORY: [[]] SCRAP
SERVICE CENTER [[]] SEE TEXT [[]] SEE TEXT
[[]] CHANNEL PARTNER
AVAILABILITY: PRODUCT’S SUPPORT LIFE NO CHARGE AVAILABLE UNTIL: Aug 2012
ADDITIONAL INFORMATION:
September 2, 2009
Rev. 15 Page 1 of 7
Page 2 of 7 N1996A-01C
Situation:
N1996A spectrum analyzers may exhibit a high-density of residual responses caused by switching
power supply signals leaking into the circuitry on the A4 Digital IF (DIF) Assembly.
Follow this procedure to determine if the N1996A exhibits the problem described above:
Solution/Action:
1 Remove the analyzer’s outer case as described in the Instrument Outer Case, Removal procedure
in the N1996A Service Guide.
2 Remove the analyzer’s board cover as described in the Board Cover, Removal procedure in the
N1996A Service Guide.
3 Remove A4 DIF assembly as described in the A4 IF Board, Removal procedure in the N1996A
Service Guide.
4 Tape aluminum foil to the 9 large shields (Agilent Part Number: 5022-7816), the details of taping
instruction is described as below:
Page 4 of 7 N1996A-01C
Page 5 of 7 N1996A-01C
Page 6 of 7 N1996A-01C
Page 7 of 7 N1996A-01C