0% found this document useful (0 votes)
25 views19 pages

Medical Imaging Techniques

This document discusses medical imaging techniques using radioisotopes. It describes how small amounts of radioactive tracers are injected into patients and their uptake and distribution in organs is then measured using radiation detectors. Specifically, it discusses how radioactive iodine uptake by the thyroid gland can indicate hyperthyroidism or hypothyroidism. It also describes the commonly used radioisotope technetium-99m which emits gamma rays that can be used to examine organs like the brain, liver, lungs and more.

Uploaded by

Adal Arasu
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
25 views19 pages

Medical Imaging Techniques

This document discusses medical imaging techniques using radioisotopes. It describes how small amounts of radioactive tracers are injected into patients and their uptake and distribution in organs is then measured using radiation detectors. Specifically, it discusses how radioactive iodine uptake by the thyroid gland can indicate hyperthyroidism or hypothyroidism. It also describes the commonly used radioisotope technetium-99m which emits gamma rays that can be used to examine organs like the brain, liver, lungs and more.

Uploaded by

Adal Arasu
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 19

4/10/2018

Medical Imaging
1 2
Textbook and Materials
Rafael C. Gonzalez, Richard E. Woods,
Techniques “Digital Image Processing”, 2nd Edition,
Pearson Education, 2003
Digital Image Processing by Jayaraman,
Veerakumar, 2012
Khandpur R.S, Handbook of Biomedical
Instrumentation, 3/e, Tata McGraw
Hill,New Delhi, 2014
Dr. K. Adalarasu
KA – MIT – Unit III – March, 2018, Sastra, Deemed University KA – MIT – Unit III – March, 2018, Sastra, Deemed University

3 4

Reference
William K. Pratt, “Digital Image Processing” ,
John Willey ,2001
Steve Webb, The physics of medical imaging,
Adam Hilger, Bristol, England, Philadelphia, Radio Isotope
USA, 1988
Jain A.K., “Fundamentals of Digital Image
Imaging
Processing”, PHI, 1995.

KA – MIT – Unit III – March, 2018, Sastra, Deemed University KA – MIT – Unit III – March, 2018, Sastra, Deemed University

1
4/10/2018

5 6
Radio Isotopes in Medical Diagnosis Radio Isotopes in Medical Diagnosis
 Used in medicine both for therapeutic as well as  Detector
diagnostic applications  Amount of radioactivity can be measured within parts of
organs as well as within the whole organ
 Diagnostic practice
 Small amounts of radioactive chemicals, called ‘tracers’ (or  Measurement of the uptake of radioactive iodine by
radio-pharmaceuticals), the thyroid gland
 Injected into an arm vein or administered through ingestion or  Rate of uptake of iodine by the thyroid
inhalation
 Greatly increased in patients with disease characterized
 Amount of radioactivity at different points within the by increased production of thyroid hormone
patient’s body, or in body fluids (hyperthyroidism)
 Then examined by radiation detectors  Disease that led to nervousness, tremor, weight loss
and, in extreme cases, even death

KA – MIT – Unit III – March, 2018, Sastra, Deemed University KA – MIT – Unit III – March, 2018, Sastra, Deemed University

7 8
Radio Isotopes in Medical Diagnosis Radio Isotopes in Medical Diagnosis
 Decreased iodine uptake (hypothyroidism) by the  Radio-nuclide imaging depends on
thyroid  Obtaining a suitable distribution of the radio-nuclide in the
 Had symptoms and signs of diminished thyroid function patient
 Thyroid gland continues to be the organ most  Radio-nuclide is labelled to a compound
frequently examined by nuclear medicine  Which will be taken up or metabolized in some way by the
 Imaging of organ functions is carried out non- human tissue to be studied
invasively  Patient receives the material, usually by intravenous
 Nuclear medical examination approach is primarily injection
function-oriented  After a suitable delay, which may be minutes or hours to
 Vital processes such as blood circulation, metabolism and  Allow uptake in the target tissues and clearance from the
vitality of organs and tumours can be displayed as blood, imaging can commence
functional images  Each taking 2–10 minutes

KA – MIT – Unit III – March, 2018, Sastra, Deemed University KA – MIT – Unit III – March, 2018, Sastra, Deemed University

2
4/10/2018

Radio Isotopes in Medical Diagnosis


9 10
Radio Isotopes in Medical Diagnosis
Dynamic studies can be performed with the  Tecnetium-99m (Tc-99m) - radio-nuclide used to
gamma camera  Examine the brain, liver, lungs, bones, thyroid, kidney and
heart
Moment of injection and capturing frames of the
data either photographically or digitally  Tecnetium-99m (Tc-99m)
 Metastable nuclear isomer of technetium-99 (itself an
At times ranging from minutes down to fractions of isotope of technetium), symbolized as 99m Tc
a second  That is used in tens of millions of medical diagnostic procedures
Data can produce useful information on organ annually

function, blood flow, clearance rates, etc  Emits readily detectable gamma rays with a photon
energy of 140 keV (these 8.8 pm photons are about the
same wavelength as emitted by conventional X-ray
diagnostic equipment)
 Half-life for gamma emission is 6.0058 hours
KA – MIT – Unit III – March, 2018, Sastra, Deemed University KA – MIT – Unit III – March, 2018, Sastra, Deemed University

11 12
Radio Isotopes in Medical Diagnosis
 TC-99m cannot be coupled with all required
biologically active substances
 So that with this radio-nuclide the spectrum of radio-
pharmaceuticals for examinations of the organ metabolism Radiation Detectors
is limited
 Iodine-123 labelled substances
 Which are used in many clinical examinations
 Iodine-123 (123I or I-123)
 Half-life is 13.22 hours
 Predominant energy of 159 keV (this is the gamma
primarily used for imaging)

KA – MIT – Unit III – March, 2018, Sastra, Deemed University KA – MIT – Unit III – March, 2018, Sastra, Deemed University

3
4/10/2018

13 14
Radiation Detectors Radiation Detectors
Depending upon the radiation emitted by the
radio isotope of the radiopharmaceutical Ionization chamber
A suitable detector is selected and operated under Geiger Muller Counter
optimum conditions
Proportional Counter
Radiation falls on a photographic plate
Semiconductor detectors
It would cause darkening when developed after
exposure Solid state detectors
Photographic method is useful for measuring the
total exposure to radiation of workers

KA – MIT – Unit III – March, 2018, Sastra, Deemed University KA – MIT – Unit III – March, 2018, Sastra, Deemed University

15 16

Ionization Chamber Ionization Chamber


 When an atom is ionized  Chamber which is filled with gas and is provided with
 It forms an ion pair
two electrodes
 If the electrons are attracted towards a positively  Very high insulation resistance used as the insulation
charged electrode between the inner and outer electrodes of the ion
chamber
 Positive ions to a negatively charged electrode, a
 Polytetrafluoroethylene
current would flow in an external circuit
 Potential difference of a few hundred volts is applied
 Principle of the ionizing chamber
between the two electrodes
 Magnitude of the current would be proportional to the
amount of radioactivity present between the electrodes  Radioactive source is placed inside or very dose to
the chamber

KA – MIT – Unit III – March, 2018, Sastra, Deemed University KA – MIT – Unit III – March, 2018, Sastra, Deemed University

4
4/10/2018

17 18
Ionization Chamber Ionization Chamber
 Charged particles moving through the gas undergo  Current 10–10 A or less
inelastic collisions to form ion pairs  Measured by using a very high input impedance
 Electrodes is sufficiently high to collect all the ion voltmetre
pairs
 Chamber current will then be proportional to the
amount of radioactivity in the sample
 Ionization chambers are operated
 Counting mode
 They respond separately to each ionizing current
 Integrating mode
 Collection of ionization current over a relatively long period

KA – MIT – Unit III – March, 2018, Sastra, Deemed University KA – MIT – Unit III – March, 2018, Sastra, Deemed University

Ionization Chamber
19 20
Scintillation Detector
 100,000 ion pairs produced  Scintillator
 Single alpha particle  Crystalline substance which produces minute flashes of
 Traversing approximately 1 cm in air light in the visible or near ultraviolet range
 Would be around 3 x10–14 coulomb  When it absorbs ionizing radiation

 Average charge is made to pass through a resistance  Number of fluorescent photons is proportional to the
of 3 x 1010 Ω in 1s energy of the radioactive particle
 Difference of approximately 1 mV potential would develop  Flashes occur
across the high resistance  Due to the recombination and de-excitation of ions
 Voltage is a function of the rate of ionization in the  Excited atoms produced along the path of the radiation
chamber  Light flashes are of very short duration
 Portable ionization chambers are also used to  Detected by using a photo-multiplier tube
monitor personnel radiation doses
KA – MIT – Unit III – March, 2018, Sastra, Deemed University KA – MIT – Unit III – March, 2018, Sastra, Deemed University

5
4/10/2018

21 22
Scintillation Detector Scintillation Detector
 Which produces a pulse for each particle  Scintillating material which is generally used as
 Scintillation counter gamma ray detector
 Scintillator along with the photo-multiplier tube  A crystal of sodium iodide activated with about 0.5%
 Gamma radiations cannot be detected directly in a of thallium iodide
scintillating material  Counting beta particles
 Because gamma rays possess no charge or mass  Scintillator crystals of anthracene are employed
 Gamma ray energy must be converted into kinetic  Detector must be able to absorb a high proportion of
energy of electrons present in the scintillating the incident radiation
material  Convert this energy rapidly into suitable electronic signals
 Conversion power will be proportional to the number of  Thallium-activated sodium-iodide NaI(TI)
electrons (electron density) available for interaction with the
gamma rays
scintillation crystal is used in all commercial
cameras
KA – MIT – Unit III – March, 2018, Sastra, Deemed University KA – MIT – Unit III – March, 2018, Sastra, Deemed University

23 24

Scintillation Detector Semiconductor Detectors


Conversion efficiency  Detectors can be made very small and robust
Ratio of light output to incident photon energy  Silicon and germanium crystals
Typically 10% for 140 keV radiation  Counting alpha and beta particles
 On absorption of radiation in the crystal
40 keV gamma ray absorbed in the crystal
 Electrons and positive holes are formed
Produces about 4200 light photons  Which move towards opposite electrodes under the
In the blue-green region of the spectrum where each influence of applied potential
light photon has an energy of around 3 eV
 Resulting current is proportional to the energy of the
Decay time of this light flash ionizing radiation
Half-life of approximately 0.2 ms
Sufficiently fast for most clinical applications

KA – MIT – Unit III – March, 2018, Sastra, Deemed University KA – MIT – Unit III – March, 2018, Sastra, Deemed University

6
4/10/2018

25 26
Semiconductor Detectors Solid State Detectors
Advantage
 Miniature form and can be utilized as in-vivo probes
Low energy (3 to 3.5 eV) required to produce an for clinical and experimental applications in medicine
electron-hole pair relative to that of a gas (30 to 35
 Silicon and cadmium telluride detectors (1 mm
eV to produce one ion pair)
diameter)
Or scintillation detector (200 to 300 eV to produce one
 Which can be encapsulated
photoelectron)
 Used as catheter tubes for studying blood circulation and
Operate lithium-drifted silicon or germanium regional pulmonary functions
detectors at a low temperature (at the liquid
 Detectors are highly sensitive to gamma rays
nitrogen boiling point of 77°K) in order to reduce
because of the high atomic number of cadmium (48)
the leakage current
and telluride (52)

KA – MIT – Unit III – March, 2018, Sastra, Deemed University KA – MIT – Unit III – March, 2018, Sastra, Deemed University

27 28
Pulse Height Analyser
 Radioactivity measurements
 Individual particles are detected as single electrical
impulses in the detector

Pulse Height Analyser  Measurement of pulse height is thus a useful tool for
energy determination
 To sort out the pulses of different amplitudes and to
count them
 Electronic circuits are employed
 Instrument which accomplishes this is called a ‘pulse
height analyser
 Analysers are either single or multiple-channel instruments

KA – MIT – Unit III – March, 2018, Sastra, Deemed University KA – MIT – Unit III – March, 2018, Sastra, Deemed University

7
4/10/2018

29 30
Pulse Height Analyser Pulse Height Analyser
 Output pulses from the photo-multiplier are amplified
in a high input impedance low-noise pre-amplifier
 Amplified pulses are fed into a linear amplifier of
sufficient gain to produce output pulses in the
amplitude range of 0–100 V
 Pulses are then given to two discriminator circuits
 Schmitt trigger circuit
 Which can be set to reject any signal below a certain voltage
 Required for excluding scattered radiation and
amplifier noise

KA – MIT – Unit III – March, 2018, Sastra, Deemed University KA – MIT – Unit III – March, 2018, Sastra, Deemed University

31 32
Pulse Height Analyser Pulse Height Analyser
 Upper discriminator circuit  Anti-coincidence circuit
 Rejects all but signal 3  Output pulse when there is an impulse in only one of the
 Lower discriminator circuit input channels
 Rejects signal 1 only  Cancels all the pulses which trigger both the Schmitt
triggers
 Transmits signals 2 and 3
 Only signal reaching the counter is the one lying in
 Two discriminator circuits give out pulses of
the window of the pulse height analyser
constant amplitude
 Window can be manually or automatically adjusted to
 Pulses with amplitudes between the two triggering
cover the entire voltage range with a width of 5–10 V
levels are counted
 Difference in two levels is called the window width, the
 Scaler and counter
channel width or the acceptance slit  Scaling unit counts down the pulses from the analyser
 So that they are digitally displayed

KA – MIT – Unit III – March, 2018, Sastra, Deemed University KA – MIT – Unit III – March, 2018, Sastra, Deemed University

8
4/10/2018

Pulse Height Analyser


33 34

 A decade system of counting is employed, which


displays units, tens, hundreds
 Multi-channel pulse height analysers are often used
to measure a spectrum of nuclear energies Uptake Monitoring
Equipment
 May contain several separate channels
 Schmitt trigger discriminators are adjusted to be
triggered by pulses of successively longer amplitude
 Simultaneous counting and recording of an entire spectrum
 Parallel array of discriminators is generally used
 Provided the number of channels is ten or less

KA – MIT – Unit III – March, 2018, Sastra, Deemed University KA – MIT – Unit III – March, 2018, Sastra, Deemed University

35 36
Uptake Monitoring Equipment Uptake Monitoring Equipment
 Medical investigations depends  Fraction of the chemical present in the organ at any
 On obtaining a suitable distribution of the radio-nuclide in time
the patient  Would indicate the functional status or what is called
 Achieved by administering a suitable chemical the uptake of the organ
substance tagged with a radio-nuclide  Gamma energy range for uptake monitoring studies
 Emitting gamma radiations is from 100 keV to 500 keV
 Gamma ray gets transmitted through the body  NaI (TI) scintillation crystal for detection of gamma
tissues rays
 An external monitoring system can be used to detect  Photo-multiplier which converts the scintillations into an
them electrical signal
 Provide the measurement of the chemical substance

KA – MIT – Unit III – March, 2018, Sastra, Deemed University KA – MIT – Unit III – March, 2018, Sastra, Deemed University

9
4/10/2018

37 38
Uptake Monitoring Equipment Uptake Monitoring Equipment
 Output from the photo-multiplier is given to a
 Pre-amplifier followed
 By a pulse shaping circuit
 A pulse height analyser
 Output of the analyser drives a counter/timer
 Which displays the information on a digital counter and a
strip chart recorder
 Vital component in the system is the collimator
 Simple collimator
 Consists of a single tapered hole in a cylindrical lead block

KA – MIT – Unit III – March, 2018, Sastra, Deemed University KA – MIT – Unit III – March, 2018, Sastra, Deemed University

39 40

Uptake Monitoring Equipment


 Sides of the lead block must be sufficiently thick to
absorb the majority of the gamma rays impinging
obliquely
 Detector must be surrounded by an adequate Radio-isotope Rectilinear
thickness of lead to effectively shield it from all
gamma rays except those entering through the Scanner
aperture
 Shielded and collimated detector is called a probe
 Probe is mounted on an adjustable support allowing it to
be appropriately positioned in relation to the patient

KA – MIT – Unit III – March, 2018, Sastra, Deemed University KA – MIT – Unit III – March, 2018, Sastra, Deemed University

10
4/10/2018

41 42
Radio-isotope Rectilinear Scanner Radio-isotope Rectilinear Scanner
 Distribution of radioactive material within an organ or  Scanner is a moving detector imaging system
part of the body is studied by using radio-isotope  Heart of the system is the detector-collimator
rectilinear scanners assembly
 Rectilinear scanner is a device for imaging the  Detector
distribution of radioactive material within the body  Three or five inch diameter NaI crystal
 Situated behind a focusing collimator
 Travel in a regular scanning pattern back and forth
across the area of interest
 Detected and amplified signals can be plotted to
give a picture or contour map of radioactivity within
the organ

KA – MIT – Unit III – March, 2018, Sastra, Deemed University KA – MIT – Unit III – March, 2018, Sastra, Deemed University

43 44
Radio-isotope Rectilinear Scanner Radio-isotope Rectilinear Scanner
 Detector-collimator assembly
 Photo-multiplier
 Pre-amplifier are housed in a single unit
 Which is attached to a motor-driven device
 Single probe scanner makes use of one detector
that scans the area of interest
 Dual probe scanners that have two synchronously
moving
 Axially opposite detectors with the patient between
the two detectors

KA – MIT – Unit III – March, 2018, Sastra, Deemed University KA – MIT – Unit III – March, 2018, Sastra, Deemed University

11
4/10/2018

45 46
Radio-isotope Rectilinear Scanner Radio-isotope Rectilinear Scanner
 Scanning can be linear or one-dimensional  Photographic recorder
 In the whole body counting applications  Light flashes can be photographed on a film
 Detector is moved continuously over the body and the  Face of a cathode ray tube
counts are integrated over the entire scan
 Dot recorder is most commonly used
 Recording may be done either by
 Photographic recorder or by dot recorders

KA – MIT – Unit III – March, 2018, Sastra, Deemed University KA – MIT – Unit III – March, 2018, Sastra, Deemed University

47 48
Radio-isotope Rectilinear Scanner Radio-isotope Rectilinear Scanner
 It produces a map of the distribution of activity within
Scaling factor of 16
the area of interest by recording dots or slit-like
marks on paper Every 16 counts arriving at the input of the
scaling circuit from the pulse height analyser
 Dot recording mechanism consists of an electrically
heated stylus to burn a small spot on a sheet of One dot appears on the paper
electrically conducting paper Reduction in counting rate is necessary
 Each time a pulse passes through the stylus Because extremely high counting rates will drive
 Pulses to the stylus are delivered from the pulse the stylus wild
height analyser Count-rate metre
 After scaling down the counts by an adjustable scaling
factor from 1 to 256 Display or record the average count rate

KA – MIT – Unit III – March, 2018, Sastra, Deemed University KA – MIT – Unit III – March, 2018, Sastra, Deemed University

12
4/10/2018

49 50
Radio-isotope Rectilinear Scanner
Some scanners make use of colour printing
Maximum count rate is first established
Divided into six ranges
Each being associated with a different colour print Gamma Camera
Coloured map showing the distribution of the
isotope is built up

KA – MIT – Unit III – March, 2018, Sastra, Deemed University KA – MIT – Unit III – March, 2018, Sastra, Deemed University

51 52
Gamma Camera Gamma Camera
 Used to produce images of the radiation generated
by radio-pharmaceuticals within a patient’s body in Gamma camera was developed by Anger
order (1958)
 To examine organ anatomy and function, and to visualize Large circular area of thin scintillation crystal
bone abnormalities An array of closely packed photo-multiplier tubes
 Stationary gamma camera systems to amplify
 Whole-body images (single head-to-toe skeletal profiles) Locate the gamma ray interactions in the crystal
 Tomographic images (cross-sectional slices of the body
To display the scintillations instantly on a cathode
acquired at various angles around the patient and
displayed as a computer-reconstructed image) ray tube
Study the rapidly changing distribution of activity

KA – MIT – Unit III – March, 2018, Sastra, Deemed University KA – MIT – Unit III – March, 2018, Sastra, Deemed University

13
4/10/2018

53 54
Gamma Camera Gamma Camera
 Modern-day gamma cameras constitute extremely  When a photon of the radiation leaves the patient’s
complex electronic equipment body
 Detector  It passes through the collimator
 Collimator, crystal  Interacts with a crystal
 Photo-multiplier tubes  Wherein its energy is converted into light

 Position localization circuitry  Light from the crystal is received by photo-multiplier tubes
 Camera Electronics – includes  Converted into an electrical signal
 Correction circuitry  Electrical signal passes through the position
 Energy analysis circuitry localization circuitry
 Counting circuit  Whose output consists of X and Y positional signals, and
a Z or energy signal
 Image display
 Image recording device
KA – MIT – Unit III – March, 2018, Sastra, Deemed University KA – MIT – Unit III – March, 2018, Sastra, Deemed University

55 56
Gamma Camera Gamma Camera
 Special correction circuits
 Which compensate for errors in the detection and
localization of photon
 Z or energy signal is then analysed in the pulse
height analyser circuit
 To determine if the detected photon is within a user-
specified energy range
 If it is registered in the counter
 X and Y signals are then sent to an image recording
device
 Where they are used to position the beam of a cathode ray
tube

KA – MIT – Unit III – March, 2018, Sastra, Deemed University KA – MIT – Unit III – March, 2018, Sastra, Deemed University

14
4/10/2018

57 58
Gamma Camera Gamma Camera
 Z pulse then turns the beam on Collimator
 Causing a bright dot to appear at a location on the face of Heavy metal absorber such as lead, with some
the CRT corresponding to the location in the crystal
where the photon deposited its energy
tungsten or platinum parts
 Hundreds of thousands of photons leave the Basic types of collimators used in conventional
patient’s body and strike crystal gamma camera imaging
 Each causing a black spot to be formed on the film Pinhole
 Collimator is used to selectively absorb unwanted Parallel-hole
radiation Diverging
 Only photons traveling along the desired path are allowed
to pass through to the detector
Converging

KA – MIT – Unit III – March, 2018, Sastra, Deemed University KA – MIT – Unit III – March, 2018, Sastra, Deemed University

59 60
Gamma Camera
Crystal of up to 500 mm diameter
Typically 6.4 mm or 9.6 mm thick with an array of
61, 75 or 93 photo-multiplier tubes
Number of gamma rays received by any region
Multicrystal Gamma
of the crystal is directly proportional to the
amount of nuclide located directly below the Camera
region
Polaroid camera is mounted on the
oscilloscope for photographing the build-up of
about 50,000 dots on the screen

KA – MIT – Unit III – March, 2018, Sastra, Deemed University KA – MIT – Unit III – March, 2018, Sastra, Deemed University

15
4/10/2018

61 62
Multicrystal Gamma Camera Multicrystal Gamma Camera
 Major limitation in using a scintillation camera for  Z-pulse analyser accepts only those events which
rapid dynamic studies is the counting losses that fall within the selected window
occur at the high count rates  Two pulses occurring within a short span of each
 Actual count rate is always greater than the other are piled up and the resulting summation
registered count rate that passes the window of the pulse is rejected
analyser  Because its amplitude exceeds the upper window level
 Scintillation  Results in the loss of two valid pulses
 Output of all photomultipliers is summed to give a Z-signal  With subsequent total count loss
 Proportional to the total amount of light emitted during the  Principle of operation differs slightly for multi-crystal
scintillation cameras

KA – MIT – Unit III – March, 2018, Sastra, Deemed University KA – MIT – Unit III – March, 2018, Sastra, Deemed University

63 64
Multicrystal Gamma Camera Multicrystal Gamma Camera
 Rectangular array of many small crystal detectors If two or more events interact at the same time
 Each surrounded by reflective or shielding material to in different crystals
minimize scattering
Both events are discarded
 Modified parallel-hole collimator
This is not possible with a mono-crystal
 Bisecting lead septa within the hole of each detector
 Used to increase sensitivity while maintaining overall
system
spatial resolution Computerized multi-crystal gamma camera
 Multi-crystal data accumulation matrix system can accumulate high count rates
 Every gamma event coming from the patient and (200,000 Hz)
interacting in any crystal
At rapid time intervals (20 s) making possible both
 Detected as a separate event at a unique location
clinical and research applications

KA – MIT – Unit III – March, 2018, Sastra, Deemed University KA – MIT – Unit III – March, 2018, Sastra, Deemed University

16
4/10/2018

65 66
Multicrystal Gamma Camera Multicrystal Gamma Camera
 Detector system of this camera consists of 294
discrete crystals arranged in 14 rows and 21 columns
 Each column and row is optically coupled by a lucite
light guide to a photo-multiplier tube
 35 photo-multiplier tubes
 Pulses from 35 pre-amplifiers and amplifiers serving
photo-multiplier tubes pass to 35 low-discriminators
 Which eliminate events observed simultaneously in
adjacent crystals
 Row and column photo-tube are uniquely identified

KA – MIT – Unit III – March, 2018, Sastra, Deemed University KA – MIT – Unit III – March, 2018, Sastra, Deemed University

67 68
Multicrystal Gamma Camera Multicrystal Gamma Camera
 Scintillations occur in the front part of the detector  Movable bed which performs a scanning motion is
 Thin crystals provide better resolution by bringing the light used
flashes closer to the PMTs
 To improve spatial resolution in static imaging
 Crystals of such units are 9.5 mm thick and 28 to 61 measurements
cm in diameter  Bed motion permits each detector crystal to scan a
 Rectangular crystals are of similar thickness and range from square area within 1.11 cm sides in 16 programmed
20 X 20 cm to 46 X 66 cm
movements of 2.78 mm each
 Random-access, solid-state buffer memory
 Bed motion cannot be used during a dynamic study
 Valid events are stored
 Spatial resolution may be somewhat less in these studies
 Information representing data from the entire detector
assembly passes into the memory of the computer

KA – MIT – Unit III – March, 2018, Sastra, Deemed University KA – MIT – Unit III – March, 2018, Sastra, Deemed University

17
4/10/2018

Multicrystal Gamma Camera


69 70
Multicrystal Gamma Camera
 Light pipe array that is required to localize each event must  Y-co-ordinates are obtained by gathering the 11 rods
transmit as much light as possible in order to optimize the to a 2-inch diameter multiplier photo-tube
energy resolution of the system  Y-co-ordinates for the entire array require 14 groups of rods
gathered
 X-co-ordinates are derived from 21 spatulas that are
shaped and bent to fit on to the end of 2-inch diameter
photo-tubes
 Main advantage of the addressing scheme is that
 Detection and positioning are made independent
 Disadvantage of the light-pipe scheme is that
 Light is attenuated in the light guides
 Leading to a degradation of the gamma ray energy
resolution
KA – MIT – Unit III – March, 2018, Sastra, Deemed University KA – MIT – Unit III – March, 2018, Sastra, Deemed University

71 72
Multicrystal Gamma Camera Multicrystal Gamma Camera
 System ensures that only events that occur in one
and only one detector at the correct energy range
will be stored in the memory
 Modern gamma camera employs a crystal of up to
500 mm diameter
 Typically 6.4 mm or 9.6 mm thick
 With an array of 61,75 or 93 photo-multiplier tubes
 Prime general purpose instrument for radio-nuclide
imaging in routine nuclear medicine investigations

2 X 2 crystal array

KA – MIT – Unit III – March, 2018, Sastra, Deemed University KA – MIT – Unit III – March, 2018, Sastra, Deemed University

18
4/10/2018

73 74

Multicrystal Gamma Camera Multicrystal Gamma Camera


 New solid-state detectors constructed of cadmium  Image output
zinc telluride (CZT) are under development to  Either black and white laser printers with a resolution of at
replace the crystal/PMT structure currently used least 1200 dpi (dots per inch)
 These solid-state CZT detectors would be used to  Colour printers (thermal transfer) or X-ray film printers are
used
directly convert gamma rays to electrical pulses
 Use of dry laser images is preferred in nuclear medicine in
 Requirements for the computer system used with modern cameras
gamma camera are speed and stability in acquisition
 Gantry and the table are operated by
of the data
microprocessor-controlled motors
 Modern camera computers are at least 64 Mbytes
RAM, 20” Monitor with a resolution of at least 1024 X
786 for 256 colours, at least 2 Gbytes hard disk and
network connection
KA – MIT – Unit III – March, 2018, Sastra, Deemed University KA – MIT – Unit III – March, 2018, Sastra, Deemed University

19

You might also like