GF 9900 Manual

Download as pdf or txt
Download as pdf or txt
You are on page 1of 72

English

Signet 9900 Transmitter


*3-9900.090*
3-9900.090 Rev. H 08/17
Operating Instructions

Quick Start

Your new Signet 9900 Transmitter needs to be calibrated and the


sensor needs to be initialized prior to use. The following steps
outline the recommended procedure to start up a new system.
1. Module Installation (page 3)
2. Installation (page 7)
Look for the 3. Wiring (page 8)
Quick Start 4. Sensor Wiring (page 11)
icon to quickly 5. Power Wiring (page 17)
set up your 6. Relay and Open Collector Wiring (page 18)
new 9900. 7. Relay Functions (page 19)
8. Operation (page 23)
9. Menu System (page 25)

Panel Mount Description

The 9900 Transmitter, a member of Signet's line of SmartPro® instruments,


provides a single-channel interface for all Flow, pH/ORP, Conductivity/Resistivity,
Salinity, Pressure, Temperature, Level, Dissolved Oxygen, Turbidity, Batch and
other applications.
The 9900 is available in either Panel or Field Mount. Both versions run on
10.8 to 35.2 VDC power (24 VDC nominal), and can power certain sensors on
loop power (see NOTE on page 11).
The 9900 Transmitter, also allows third-party 4 to 20 mA signals to be used as an
input (optional Signet 8058 i-Go® Signal Converter required, sold separately).

Compatibility Flow
515*/8510*, 525*, 2000,
Field Mount The 9900 is compatible with all GF Signet 2100, 2507, 2536*/8512*,
products listed in the column to the right. 2537, 2540*, 2551, 2552
pH/ORP
● pH and ORP electrodes require the 2724-2726 with 2750*/2751
Signet 2750/2751 DryLoc® Sensor 2734-2736 with 2750*/2751
Electronics (sold separately). 2756-WTx–2757-WTx with
● Conductivity/Resistivity or Salinity 3719 and 2750*/2751
• English 2764-2767 with 2750*/2751
measurement requires either the optional
• Deutsch 2774-2777 with 2750*/2751
Direct Conductivity/Resistivity Module
• Français (part number 3-9900.394) or the Signet Conductivity/Resistivity, Salinity
• Español 2850 Conductivity/Resistivity Sensor 2819-2823 with
• Italiano 2850 or Cond/Res Module
Electronics (sold separately). 2839-2842 with
• 中文 NOTE: If using the 2850, use the one- 2850 or Cond/Res Module
channel Digital (S3L) models. The two- Level, Temperature, Pressure
channel model 3-2850-63 may be used 2250*, 2350*, 2450*
with only one channel connected. Do Turbidity
not use with both channels connected. 4150 requires 8058
The 4 to 20 mA models 3-2850-52 and Dissolved Oxygen
3-2850-62 are incompatible with the 9900. 2610-41 direct to 9900
● Turbidity measurement using Signet 4150 2610-31 requires 8058
or Dissolved Oxygen measurement using * Can be run on Loop Power
Signet 2610-31 requires Signet 8058 i-Go (see NOTE on page 11)
Signal Converter (sold separately).
Warranty Information Table of Contents
Refer to your local Georg Fischer Sales office for the General Information .........................................................2
most current warranty statement. Dimensions ......................................................................3
Module Installation ...........................................................3
All warranty and non-warranty repairs being returned Plug-In Modules .........................................................4
must include a fully completed Service Form and Relay Module .............................................................4
goods must be returned to your local GF Sales office or Direct Conductivity/Resistivity Module .......................5
distributor. H COMM Module .......................................................5
Product returned without a Service Form may not be Batch Module .............................................................6
warranty replaced or repaired. 4 to 20 mA Output Module .........................................6
Installation ........................................................................7
Signet products with limited shelf-life (e.g. pH, ORP, Wiring ...............................................................................8
chlorine electrodes, calibration solutions; e.g. pH buffers, Signal Type: Frequency .............................................9
turbidity standards or other solutions) are warranted out Signal Type: Digital (S3L) ...........................................9
of box but not warranted against any damage, due to Signal Type: 4 to 20 mA ...........................................10
process or application failures (e.g. high temperature, Terminal Identification ..............................................10
chemical poisoning, dry-out) or mishandling (e.g. broken Sensor Wiring...........................................................11
glass, damaged membrane, freezing and/or extreme Power Wiring ............................................................17
temperatures). Relay and Open Collector Wiring ............................18
Relay Functions .............................................................19
Relay and Open Collector Outputs ................................19
Product Registration Operation .......................................................................23
Thank you for purchasing the Signet line of Georg Menu System .................................................................25
Fischer measurement products. System Setup Menu.................................................25
Common Menus.............................................................27
If you would like to register your product(s), you can now LOOP Menu .............................................................27
register online in one of the following ways: RELAY Menu............................................................28
• Visit our website www.gfsignet.com. OPTION Menu .........................................................30
Under Service and Support click Product Sensor-Specific Menus ..................................................31
Registration Form Flow ..........................................................................31
• If this is a pdf manual (digital copy), click here pH .............................................................................33
ORP..........................................................................35
Conductivity/Resistivity ............................................37
Safety Information Pressure ...................................................................39
Level/Volume............................................................40
● Follow instructions carefully to avoid personal injury. Temperature .............................................................42
● This unit is designed to be connected to equipment 4 to 20 mA ................................................................43
which can be hazardous to persons and property if Salinity ......................................................................45
used incorrectly. Dissolved Oxygen ....................................................47
● Read and understand all associated equipment Troubleshooting .............................................................49
manuals and safety warnings before using with this Appendix ........................................................................51
product. Averaging .................................................................51
● Remove power to unit before wiring connections. LOG Current Loop Output .......................................51
Custom Measurements............................................52
● Wiring connections to this product should only be Calibration Procedures - pH.....................................56
performed by qualified personnel. Calibration Procedure - ORP ...................................58
● Do not use unit if front panel is cracked or broken. Calibration Procedure - Conductivity/Resistivity ......60
Calibration Procedure - Flow ...................................61
Warning / Caution / Danger Calibration Error Messages .....................................62
Indicates a potential hazard. Failure to follow all warnings USP Limits................................................................63
may lead to equipment damage, injury, or death. H COMM Module Overview .....................................64
H COMM Module Installation ...................................65
Electrostatic Discharge (ESD) / Electrocution Danger
Alerts user to risk of potential damage to product by ESD,
H COMM Module Wiring ..........................................65
and/or risk of potential of injury or death via electrocution. H COMM Module Operation ....................................67
HART Commands ....................................................68
Personal Protective Equipment (PPE)
Unit Codes ...............................................................69
Always utilize the most appropriate PPE during installation
and service of Signet products. Specifications .................................................................70
Maintenance...................................................................71
NOTE / Technical Notes Ordering Information ......................................................72
Highlights additional information or detailed procedure.

2 9900 Transmitter
Dimensions Module Installation
Panel Mount 99.06 mm
3-9900-1P (3.90 in.) Base Unit H COMM
Module
er
DC Pow

ltage
Loop Vo

99.06 mm
(3.90 in.)

3-9900.393 Relay Module

5A 30 VDC ‫׽׽‬
5A VAC
Rating:
~
H COM

29.97 mm 4
3-9900.39 dule
(1.18 in.) Cond/Re
s Mo

Relay Module Conductivity/


(Panel Mount Resistivity
Only) Module
91.44 mm
(3.60 in.) If the 9900 Base Unit will be mounted in a panel, the
54.10 mm
(2.13 in.) plug-in modules may be installed either before or after
the base unit is mounted. If the 9900 Base Unit will be
mounted using the wall mount accessory kit (3-9900.392),
8.13 mm install plug-in modules first.
(0.32 in.) If installing both the Conductivity/Resistivity
(Cond/Res) and the H COMM Modules, install the
Side View Top View H COMM Module first, then the Cond/Res Module on
top of it (see illustration on page 5).

CAUTION
Field Mount Exercise care when installing modules.
Do not bend connecting pins.
3-9900-1
44.45 mm
(1.75 in.)
107 mm To install modules:
(4.21 in.) Remove power from the 9900. Carefully align pins and
connectors (do not bend connecting pins) and push
module firmly into place, then attach with screw(s)
(except H COMM Module).
24.13 mm
(0.95 in.)
CAUTION
88.90 mm LOOP as well as DC power MUST be removed
(3.50 in.) BEFORE installing H COMM Module.
(3-8051-X shown)
NOTE: 3-8051-X Integral Mounting Kit sold separately. To remove modules:
Remove power from the 9900.
For Relay, Cond/Res, and 4 to 20 mA Output Modules,
unplug connectors, remove screw(s), and carefully pull
3-9900.396 Angle Adjustment Adapter Kit module straight out from the base unit. Do not bend the
connecting pins.
For H COMM Module, squeeze the tabs on the bottom
88.90 mm edge, grasp the module and pull straight out. Do not bend
(3.50 in.) the connecting pins.
25° 0.68 mm For Batch Module, remove the Relay module. Loosen
Ref. (0.27 in.) bottom screw of Batch module. Carefully grip and squeeze
the tabs at the top of the module to release. Pull module
away from the unit. Do not bend the connecting pins.

3.50
88.90 mm WARNING
(3.50 in.) Relays may be connected to external
high-voltage power sources or multiple power
sources creating an electrocution hazard.

9900 Transmitter 3
Plug-In Modules
Optional modules and accessories are available for the 9900:
a. Base Unit (required)
b. Slot for optional H COMM Module
wer
DC Po

c. Slot for optional Conductivity/Resistivity, Batch, or 4 to 20 mA Output Module


ltage
Loop Vo

d. Slot for optional Relay Module (not available on field mount) a.


Each item is ordered separately.
Modules are field-replaceable at any time.
5
3-9900.39dule
See Module Installation (pg. 3) and Ordering Information (pg. 72) sections for more details. H COMM
Mo

9900 Generation b.
9900 Module
I II III IV CAUTION c.
Avoid Electrostatic d.
H COMM* X X X X
Discharge (ESD).
Relay X X X X ● Minimize handling of the plug-in modules to
Conductivity/ reduce the possibility of damage due to ESD.
X X X X ● Handle modules by the edges.
Resistivity
Never touch any exposed circuitry or contacts.
Batch X X X
● Wear an anti-static wristband or stand on an
4-20mA anti-static mat, or keep one hand touching a
X X properly grounded pipe or other piece of properly
Output*
grounded metal when handling modules.
*Able to run on Loop power

Relay Module
(Panel Mount installations only)
Mfr. Part No. Code Description wer
DC Po

3-9900.393 159 001 698 Relay Module - Two dry-contact relays ltage
Loop Vo

In addition to the standard programmable Open Collector output in the base unit, the
Panel Mount version of the 9900 has a slot for an optional Relay Module, which adds two
programmable dry-contact relays. The Open Collector output in the base unit uses the Relay
1 setting in the menus. If the optional Relay Module is installed, these are assigned to relays
2 and 3 in the menus.
Dry-contact relays are electromechanical switches with a moving contact armature.

3-9900.393 Relay Module


They are suitable for many general-purpose applications, AC or DC, including AC loads up to

5A 30 VDC ‫׽׽‬
5A VAC
Rating:
250 V. Install RC Filter Kits, 3-8050.396, on relays used to switch motor or inductive loads.

~
● Two (2) SPDT dry-contact relay (DCR) inputs
● User programmable
● 250 V, 5 A maximum resistive loading (AC).
● Can switch line voltage (typically 120 to 240 VAC)
● Can switch DC voltage (< 30 VDC @ 5A)
● Larger voltage and current ratings than Open Collector outputs.
For wiring information, refer to the Relay and Open Collector Wiring section, page 18.

NOTE:
The Relay Module requires 10.8-35.2 VDC, 300 mA power connection to DC PWR Terminals.
The Relay Module cannot be used with loop power. CAUTION
● The two red Mechanical Relay Indicator LEDs on the front panel of the 9900 show the
status of relays 2 and 3. (Status of all relays and Open Collector is available at all times
DO NOT bundle
in a single screen in View mode.)
Relay Module wiring
● Hysteresis and time delay are adjustable for each relay. with other wiring.
Doing so may cause
injury and/or damage
CAUTION Switching active loads (usually inductive) can cause contact arcing sufficient
to 9900 Transmitter,
to damage the relays.
Relay Module, and
The RC Filter Kit or “snubber” (part number 3-8050.396) is available as an Batch Module.
accessory to reduce or eliminate these damaging effects. Recommended for
inductive loads greater than 50 VAC (remote relays, solenoids, pumps, etc.)

4 9900 Transmitter
Direct Conductivity/Resistivity Module
Mfr. Part No. Code Description

SHIELD (SILVER)
3-9900.394 159 001 699 Direct Conductivity/Resistivity Module

ISO GND (BLK)


SIGNAL (RED)

TEMP (WHT)
The Direct Conductivity/Resistivity (Cond/Res) Module interfaces Signet 2818-2823 and
2839-2842 Conductivity electrodes directly to the 9900. (Conductivity/Resistivity and Salinity
measurements may also be performed via the 2850 Sensor Electronics connected through
the 9900 Digital (S3L) input.)
● Provides filtering and conditioning.
● Sensor cable length can be extended to 30 m (100 ft).
● 2839 - 2842 sensors come with a cell constant certificate to improve the accuracy of
the sensor measurements (see page 38).
Signet Conductivity Module cannot operate when using loop power and requires an Red
external DC power supply to be used with the 9900 Transmitter (see Power Wiring, pg. 17). Black
White
Shield

wer
DC Po

Voltage
Loop

SHIELD
TEMP (WHT)
odule

ISO GND (BLK)


SIGNAL (RED)
MM M
H CO

2818 Sanitary 2819- 2822 2823 2839-


2821 2842 Cond
/Res
Modu
le

H COMM Module
Mfr. Part No. Code Description
3-9900.395 159 001 697 H COMM Module
wer
DC Po

The H COMM Module enables communication between the 9900 and a HART®-enabled Volta
ge
Loop
device. The HART (Highway Addressable Remote Transducer) Protocol superimposes digital
signals on top of the 4 to 20 mA analog signal.
Refer to the 9900 H COMM Module Instruction Sheet 3-9900.094 for further details.
NOTE:
With H COMM Module installed, a minimum of 24 V is required for loop-powered systems.
.395
3-9900 odule
MM M
H CO
NOTE:
The black rubber jumper adjacent to the power terminal should only be removed when both
the H COMM Module is utilized and the required sensor cable length is over 304 m (1000 ft).
NOTE:
See Appendix for H COMM Module Overview, Installation, Wiring, Operation, The H COMM Module
HART Commands, and Unit Code information. (See page 64-69) can run on Loop Power.

HART® is a registered trademark of the HART Communication Foundation, Austin, Texas, USA.
Any use of the term HART hereafter in this document implies the registered trademark.

9900 Transmitter 5
Batch Module
Mfr. Part No. Code Description
3-9900.397 159 310 163 Batch Module
Convert a 9900-1P Transmitter (Generation II** or later) into a Batch Controller system by
utilizing a Batch Module (3-9900.397) and a Relay Module (3-9900.393).
Optional Module Wiring:
● Wire an external button or keypad (customer supplied) to stop, start or resume
a batch remotely.
● *Wire an external input that can inhibit a batch from starting.
Full 9900-1BC Batch Controller System manual available at www.gfsignet.com

3-9900.397
**NOTE:
Batch Module Start Verify 9900 Transmitter
generation in the
Stop OPTIONS Menu.
Start
Stop/Inhibit
Inhibit*
Resume
Resume
Common

Common

4 to 20 mA Output Module
Mfr. Part No. Code Description
3-9900.398-1 159 001 784 4 to 20 mA Output Module
The optional 4 to 20 mA Output Module adds an additional 4 to 20 mA loop output to a
Signet 9900 Transmitter (Panel or Field Mount). Requires the 9900 Generation III** or later.
Features:
● 4 to 20 mA Output Module can be powered using Loop Power or DC Power in the
9900 base unit.
● Independent settings for the loop currents of the Output Module and 9900 base unit
(Error, Loop Adjustments, etc.).
● The Primary or Secondary (if applicable) measurement of a compatible sensor can
be selected as the loop source.
● The same source (Primary or Secondary measurement) can be used for Loop 1 in
the 9900 base unit and Loop 2 in the Output Module.

NOTE:
The 4 to 20 mA Output,
PWR+ Red
+ Power Supply Direct Conductivity, and
PWR– Black
– 10.8 to 35.2 VDC Batch Modules share the
Loop+
same installation site on
Loop–
the 9900 base units.
Loop2+ PLC or Recorder Only one of these
Loop2– Blue modules can be used
+ Loop 1 Input
– 4 to 20 mA per 9900 Transmitter.

NOTE:
PLC or Recorder The Output Module can
Blue
+ Loop 2 Input run on Loop Power.
– 4 to 20 mA

6 9900 Transmitter
Installation
System Start-up: Step 1
Prepare the transmitter installation location. If the back of the transmitter is difficult to
access when installed, wire the removable terminal blocks first, then install it completely.

Next step: Wiring (see page 8).

For future reference, for each installation, it is recommended to record the part number
and serial number of each of the components listed here:
Panel Cutout Panel Cutout

Facility Tag Number or System ID (user assigned):___________________


92 x 92 mm 92 x 92 mm
(+ 0.8, - 0 mm) (+ 0.8, - 0 mm)

3.6 x 3.6 in. 3.6 x 3.6 in.

Base unit 3-9900-____ S/N ___________________ (+0.031, -0 in.) (+0.031, -0 in.)

Relay Module 3-9900.393 S/N ___________________ minimum


clearance
Cond/Res Module 3-9900.394 S/N ___________________ 25 mm
(1 in.)
H COMM Module 3-9900.395 S/N ___________________ Panel Cutout Panel Cutout

92 x 92 mm 92 x 92 mm
Batch Module 3-9900.397 S/N ___________________ (+ 0.8, - 0 mm) (+ 0.8, - 0 mm)

3.6 x 3.6 in. 3.6 x 3.6 in.


4 to 20 mA Output Module 3-9900.398-1 S/N ___________________ (+0.031, -0 in.) (+0.031, -0 in.)

gasket on
front side
Do not mount in direct sunlight. of panel
panel

terminal
blocks
r
DC Powe

ge
Loop Volta

Panel Mount Installation


Tools and Equipment Required 3-9900.395 le
H COMM
Modu

● Fine-tooth file
● ¼ DIN punch or jigsaw suitable for cutting panel opening to within 1 mm (0.04 in) tolerance.
● ¼ DIN punches are available and recommended for creating clean, precise openings
quickly and easily in most instrument panels. Quick-clip
● If a punch is not available, a jigsaw or other cutting tool can be used. mounting
An adhesive template is provided to help guide the cutting process. bracket

De-burr and smooth the opening with a file.

1. The panel mount transmitter is designed for installation using a ¼ DIN punch.
Recommended clearance on all sides between instruments is 25 mm (1 in).
2. Place gasket on instrument, and install in panel.
3. Slide mounting bracket over back of instrument until bracket snaps into latches on
sides of instrument.

To remove:
1. Secure instrument temporarily with tape from front or grip from rear of instrument. Field Mount with
DO NOT RELEASE. 3-8050 Universal Mount Kit
2. Press bracket clips outward and remove.

Field Mount Installation


Field mounting requires a separate mounting kit. The 3-8050 Universal Mount Kit, the
3-8051 or 3-8052 Integral Mount Kits, and the 3-9900.396 Angle Adjustment Adapter Kit
enable the transmitter to be installed virtually anywhere. Detailed instructions for field
installation options are included with the 3-8050, 3-8051 and 3-8052 adapter kits
(see Ordering Information section).

For Field Mount installations with a Cond/Res Module, the 3-9900.396 Angle Adjustment
Adapter is required along with a 3-8050 or 3-8052 adapter kit to allow for sufficient Field Mount with
clearance for the wiring. 3-8051 Integral Mount Kit
and Angle Adjustment Adapter

9900 Transmitter 7
Wiring
System Start-up: Step 2
Wire the transmitter for all connections with the power off. Keep any 4 to 20 mA and
relay-actuated output devices that are connected to it offline at this time.
Connect the sensors (page 11), power (page 17) and relay(s) (page 18).
Next step: Relay Functions (see page 19).

Wiring Tips:
● Do not route the sensor, DC power, or 4 to 20 mA cables in conduit containing Tools Required
AC power wiring. Electrical noise may interfere with sensor signal. ● Phillips screwdriver
● Routing the sensor cable in grounded metal conduit can help prevent electrical noise ● Flat-head screwdriver
and mechanical damage. ● Wire strippers
● Seal the cable entry points to prevent moisture damage.
● Only one wire should be inserted into a terminal.
● Splice double wires outside the terminal or use appropriate wire ferrule,
not to exceed 2 mm (0.08 in) diameter. For Field Mount
installations, refer to
the wiring diagram
inside the Field
All wiring connections to the 9900 are made via removable terminals.
Mount housing.
In general:
● The Power, Loop and Open Collector plugs and the Relay Module plug accept
12 to 28 AWG wire.
The S3L/Freq plug and the Cond/Res Module plug accept 16 to 28 AWG wire.
● Strip 7 mm (0.28 in.) of insulation from wire tips and tin bare ends to eliminate fraying.
● Insert wire tip or ferrule completely into the terminal and secure with the screw.
● Do not allow any AC leads that may be connected to the internal relays to come in
contact with low voltage wiring.

CAUTION:
Avoid Electrostatic Discharge (ESD)
● Minimize handling of plug-in modules to reduce the
possibility of damage due to ESD.
● Handle modules by the edges. Never touch any exposed
circuitry or contacts.
● Wear an anti-static wristband or stand on an anti-static mat,
or keep one hand touching a properly grounded pipe or other
properly grounded piece of metal when handling modules.

8 9900 Transmitter
Signal Type: Frequency
Signet flow sensors 515/8510, 525, 2000, 2100, 2507, 2536/8512 and 2540 provide a
Flow sensor Maximum
frequency output. (Flow sensors 2551 and 2552 can be configured with either Digital (S3L)
models with Cable Length
or Frequency outputs, see page 13.)
Frequency 60 m 305 m
The maximum allowable cable length for sensors with frequency output is dependent upon Output (200 ft) (1000 ft)
the output signal strength of the sensors themselves, and the degree to which the signals
are susceptible to EMI or "noise." This is largely a function of whether the sensors are 515/8510 X
self-powered (515/8510 and 525), or powered by an external source. 525 X
● The input terminals on the 9900 carry frequency data signals from the sensor. 2000 X
● Do not route sensor or output cables in conduit containing AC power wiring.
2100 X
Electrical noise may interfere with sensor signal.
● Routing cable in grounded metal conduit will help prevent electrical noise and 2507 X
mechanical damage. 2536/8512 X
● Seal cable entry points to prevent moisture damage.
2537 X
● Only one wire should be inserted into a terminal. Splice double wires outside the terminal.
● In case of noise interference, ground the sensor SHIELD wire to a local earth ground 2540 X
at a point near the sensor. 2551 X
● Consult the sensor manual for additional wiring information.
2552 X

Signal Type: Digital (S3L)


● The input terminals on the 9900 carry Digital (S3L) serial data from the sensor.
● Do not route sensor or output cables in conduit containing AC power wiring. In case of noise
Electrical noise may interfere with sensor signal. interference, connect
● Routing cable in grounded metal conduit will help prevent electrical noise and the cable shield to
mechanical damage. earth ground.
● Seal cable entry points to prevent moisture damage.
● Only one wire should be inserted into a terminal. Splice double wires outside the terminal. Maximum total cable length
● The TOTAL cable length from I/O devices to the transmitter must not exceed 305 m (1000 ft). of the Digital (S3L) Bus:
● In case of noise interference, ground the sensor SHIELD wire to a local earth ground
at a point near the sensor. The quality of the cable
● Consult the sensor manual for additional wiring information. used in the bus determines
● The maximum cable length of the Digital (S3L) bus varies depending on the types the maximum length of all
of sensors connected and the size of the conductors in the cable. For best results, branches combined.
determine the maximum cable length for the system before routing cables. The maximum cable length
● There are several methods that can help route the digital cables and remain within the may not exceed
distance limitations. 305 m (1000 ft),
regardless of current
requirements.

9900 Transmitter 9
Signal Type: 4 to 20 mA
When connecting a non-Signet sensor to the 9900, the sensor’s 4 to 20 mA signal must be
converted to Digital (S3L). The 8058 i-Go Signal Converter accepts any 4 to 20 mA signal
and converts it into Digital (S3L).

1. Wire the 8058 between the 4 to 20 mA loop source and the 9900 Digital (S3L) input
terminals (see page 14).
2. In the 9900 INPUT TYPE menu (see System Setup Menu discussion, page 25),
specify 4 to 20 mA INPUT.
3. Set additional labels and abbreviations as described on pgs. 42-43.

AX

1
CM

2
35 VD

Loop1 +
3

Loop1 -
For the 8058-2,

Loop2+ 4

N/C
Re +G

Loop2 - 5
F

6
3 +
d

4-20 mA
Input
L Converter
2

3-8058.610D
8058 i-Go™
6 1

N/C
5

GND
d 4

iel

+5VDC
connect Channel 1

S3 L
+GF+
4-to-20 mA

Sh White
Lo
Signet 8058 i-Go™
op

9
PW 1

8
4-20mA to S
R

4-20 mA to S3L Converter

7
Output
Output
Lo
Input

S L
op

Black S3L
ONLY
PW 2

3
R

+GF+
Red Sign
S3
DA L
TA
et

4-20 mA input S3L Output


Bla N /C
80
58

ck 7
d 8
iel
9

Sh

8058-1 8058-2

Terminal Identification
The 9900 requires regulated 10.8 to 35.2 VDC (24 VDC nominal)
from an external power supply (not supplied).
Maximum current draw is:
200 mA = 9900 without Relay module
300 mA = 9900 with Relay module Power

Terminals 1-2: DC Power 1 PWR+


Required by the instrument
● 10.8 to 35.2 VDC input power to sensors, relays and the LCD backlight 2 PWR–
3 Loop+
Terminals 3-4: Loop Power (may also be used for system power)
● 10.8 to 35.2 VDC 4 Loop–
NOTE: Backlight, LEDs and optional Relay Module do not operate on
loop power. Any connected sensors or sensor electronics that cannot 5 6
operate on loop power will also be inoperative.
OC- OC+
Terminals 5-6: Open Collector
● Software selectable for Normally Open or Normally Closed.
● May be disabled (Off) if not used.
Digital (S3L)/Freq
Terminals 7-10: Digital (S3L)/Frequency Input
SHLD
DATA

7: V+: +5 VDC out to sensor (black wire)


GND

8: DATA: Input signal from sensor (red wire)


V+

9: GND: Sensor ground (white wire)


10: SHLD: Cable shield

7 8 9 10

10 9900 Transmitter
Sensor Wiring

er
DC Pow

oltage
Connect sensor wires here as
Loop V
shown in the following figures.

95
3-9900.3
Module
H COMM
Loop powered systems
cannot power both an
H COMM Module and
pH sensor.
If using both an H COMM
Module and pH sensor,
Connect power and open DC power is required.
collector wires here as
shown on pages 17 and 18.

Freq Digital (S3L) Run on 9900 Generation


Sensor model
Output Output Loop Power I II III IV
515/8510 X X X X X X
525 X X X X X X
2000 X X X X X
2100 X X X X X Loop Power can be
used ONLY for the
2250 X X X X X X
following sensors:
2350 X X X X X X
2450 X X X X X X 515/8510, 525,
2250, 2350, 2450,
2507 X X X X X
2536/8512, 2540, 8058
2536/8512 X X X X X X and pH/ORP sensors
2537-5 X X X X X with 2750; all other
2540 X X X X X X measurement sensors
2551 X X X X X X require DC power.
2552 X X X X X X
2610-41 X X X
2610 + 8058 X X X X X
2724-2726 X X X X X
2734-2736 X X X X X
2750 X X* X X X X
2751 X X X X X Advanced 2751
2756-2757 X X X X X features require
Gen 4 or greater
2764-2767 X X X X X
2774-2777 X X X X X
2819-2823 X X X X X
2839-2842 X X X X X
2850 X X X X X
4150 + 8058 X X X X X
U1000 X X X X X

*A minimum of 24 VDC Loop Power is required for the 2750.

9900 Transmitter 11
Sensor Wiring

NOTE: Loop Power cannot be used to power Signet models 2000, 2100, 2507, 2537,
2551 or 2552 Flow sensors.

Wiring for:

Technical Notes:
515/8510 525 2536/8512 2540
• See corresponding product
FLOW
Frequency manuals for maximum cable
length.
• Maintain cable shield through

SHLD
DATA
GND
cable splice.

V+
• Route sensor cable away
from AC power lines.
• 515/8510 and 525
installations, connect the
silver (shield) wire to earth

No connection
ground in case of EMI noise
2000 2100 2507 interference.
Black
Red
FLOW
Shield

Wiring for:
Technical Notes:
2537-5 • Wiring terminals on the 2537
are rated for 16 to 22 AWG
9900 S3L Inputs wires.
• The cable must be
SHLD
DATA
GND
V+

7 mm to 10 mm in diameter
(0.275 in. to 0.394 in.) to seal
properly in the liquid-tight
connector.
• The conduit ports have
-
½-inch NPT threads.
Blk Red Shld +

S1 S2
After routing the cables,
seal the port with a liquid-
Black tight conduit connector
(3-9000.392-1) or with
Red conduit.
White • The 2537 models can
connect to the 9900 via a
relay frequency signal or with
a Digital (S3L) output. Signet
recommends installing the
Digital (S3L) output model
(2537-5) because Digital
(S3L) is more accurate.
• See 2537 instruction manual
for additional installation
information.

12 9900 Transmitter
Sensor Wiring

Wiring for:

Frequency

4 2551 Technical Notes:

SHLD
DATA
GND
V+
• When the blue jumper
3 illustrated here is placed over
both pins, the 2551-XX-11
(Blind Magmeter) outputs
2551-XX-11
an open collector frequency
Blind Magmeter
signal. When the jumper is
S3 L removed (or placed over
one pin for storage) the
Black 2551-XX-11 outputs a digital
4
Red (S3L) signal (recommended).
White 3
Shield

Only in case of EMI interference.


See Frequency Output Technical Frequency Output Technical
Notes (2551 & 2552) at right Notes (2551 & 2552):
• The frequency output will
be displayed as positive
flow regardless of the flow
direction.
Input Wiring for 2551 and 2552 sensors
• 5 VDC power required by
• Either Frequency or Digital (S3L) may be used. the 2551 & 2552 is supplied
• Signet recommends configuring these sensors with the Digital (S3L) output by the 9900. No additional
because it is more accurate and will also display reverse flow (negative numbers). power is required.
• Input type is selected by choosing between “SENSOR FREQ” and "SENSOR S3L" in • Connect the silver wire
the FLOW sensor type INPUT menu (see page 32). (shield) to earth ground
• Loop Power cannot be used to power these sensors. in case of EMI noise
interference.
• If EMI noise interference
continues, disconnect silver
wire (shield) from 9900
Wiring for:

2552

9900 S3L Inputs Frequency 2552 Technical Notes:


• The 2552 outputs an open
SHLD

SHLD
DATA

DATA
GND

GND

collector frequency signal that


V+

V+

can be connected to the 9900.

Black Black
Brown Brown
White White
Blue Blue X No
connection
Shield Shield

9900 Transmitter 13
Sensor Wiring

Wiring for:

2250 2350 2450


Technical Notes:
9900 S3L Inputs • Use three conductor shielded
cable for sensor cable splices

SHLD
DATA
GND
V+
up to 305 m (1000 ft) max.
• Maintain cable shield through
cable splice.
• Route sensor cable away
2551* 2750/2751 2850 from AC power lines.
• Connect the silver (shield)
wire to earth ground in case
Black
of EMI noise interference.
Red
White Technical Notes:
Shield • The 2850 has no SHIELD wire.
* 2551-XX-21, -41 • To work correctly with the
Display Magmeter 9900, the 2850 must be set
for the custom cell constant or
NOTE: The 2850 has no SHIELD wire. the actual probe cell constant
and the 9900 set for a 1.0 cell
constant.

Wiring for:

9900 S3L Inputs


SHLD
DATA
GND
V+

+GF+
Technical Notes:
4-to-20 mA

Signet 8058 i-Go™


Output

8058-1
4-20 mA to S3L Converter
Input

S3L

4-20 mA input S3L Output

• The cable length from the


8058 to the 9900 must not
exceed 60 m (200 ft).
Black • When using the 8058-2,
8058-1 Signet 8058 i-Go™ Red connect the loop source to
Output

4-20 mA to S3L Converter


S3L

close-up White Channel 1 input ONLY.


• See the 8058 manual for
S3L Output
Shield
more information.

9900 S3L Inputs


S3 L
SHLD
DATA

DATA
GND
V+

X
1

MA
2

DC
Loop1 +
3

35V
Loop1 -
Loop2+ 4

N/C

+G
Loop2 - 5

F+
6
4-20 mA

3
Input
L Converter

3-8058.610D

2
8058 i-Go™

1
N/C

6
5
GND

4
+5VDC

S3 L

8058-2
Loo
9

p1
3

PW

N/C
8
4-20mA to S

R
7
Output

Loo
S3 L

p2
PW
R
+GF+

S3
L
DA
Sig TA
ne
t 805
8
N/
C

7 8 9
7
8
9

8058-2 WHITE

close-up RED

BLACK

14 9900 Transmitter
Sensor Wiring

Wiring for:
9900 S3L Inputs

SHLD
DATA
GND
V+
2610-41
9900 Power Inputs
Technical Notes:
The wiring of the 3-2610-41
RED
PWR+ is non-standard:
BLACK
PWR– • RED is 12 to 24 VDC
V+ RED
LOOP+ • WHITE is Data

WHITE
24 VDC
LOOP– • BLACK is VDC Ground
Power Supply
7310-XXXX
RED • A jumper MUST be installed
V-
BLACK between PWR- and S3L GND.

SHIELD Blue
Green
No connection
Brown

3-2610-41

2610-31 via 8058-1


9900 Power Inputs 9900 S3L Inputs Technical Notes:
• 3-2610-31 Dissolved Oxygen

SHLD
PWR+
DATA
GND
Sensor can be connected
V+
PWR–
LOOP+ to the 9900 only via a
LOOP– 3-8058-1 or 3-8058-2 i-Go
Signal Converter.
V–
BLACK
• Program the 9900 for the
24 VDC BLACK
Power Supply 2610 DO sensor via the
+GF+
4 to 20 mA sensor settings
4-20 mA

7310-XXXX Signet 8058 i-Go™ RED


Output
Input

4-20 mA to S3L Converter


S3L

SHIELD
V+
(no connection) on the 9900 (see page 44).
RED RED
4-20 mA input S3L Output WHITE • See the 2610 manual for
8058-1 SHIELD
more information.

DO Sensor BROWN
3-2610-31 GREEN
BLUE No connection
WHITE Technical Notes:
• The cable length from the
8058 to the 9900 must not
exceed 60 m (200 ft).

2610-31 via 8058-2


Technical Notes:
9900 S3L Inputs
• When using the 8058-2 with
No connection

SHLD
DATA
GND

24 VDC the 9900, connect the loop


V+

DO
Sensor 1 Power Supply power supply to 8058-2
3-2610-31 7310-XXXX
Loop 1 inputs ONLY as
9900 Power Inputs shown in the figure.
+

PWR+
RED
• See the 8058 manual for
more information.
BLACK

PWR–
WHITE
RED

BLACK
LOOP+
BROWN

LOOP–
+GF+ SIGNET

Loop1

Loop2

DATA
PWR

PWR

S3 L

7 8 9
3 2 1
6 5 4

N/C

9900 Transmitter 15
Sensor Wiring
9900
Wiring for: 4150 S3L Inputs

SHLD
DATA
GND
via 8058-1

V+
Q2

R2
+ D2
D1
+

Black
Red
White
Shield
R3
EMCF1 R4 Q1
Signet Turbidimeter

ENTER

R1

RLY2 RLY1
TB2

TB1 Connect 4 to 20 mA wires to


S3L
terminal board TB3 as shown in
+ 24 VDC ALARM2 ALARM1
TB3
+
the figure.
COM

COM

Output
NO

NO
NC

NC

+ –
B
SHLD
A Refer to the latest revision of
4-20mA/

S3L Output
Signet 8058 i-Go™
4-20 mA to S3L Converter
L7
RS-485 L1 the Signet 4150 Turbidimeter
L2

RED WHT Manual (3-4150.090) for further


TB4
L5
ORG BRN information.
L3
BLK GRN
L7
SHLD YEL

4-20 mA input
ISL 1

+ +
L4

+GF+
Input
4-20 mA

Black
Technical Notes:
Shield • All of the electrical
Power Red connections to the
Connection instrument are made
through the terminal box.
The connections are labeled
via 8058-2 within the terminal box and
are self-descriptive.
Q2 Technical Notes: • A plug is inserted into the
R2
When using the 8058-2, connect the RS-485 cable bulkhead
+ D2
+ when shipped to ensure a
R3
D1
loop source to Channel 1 input ONLY.
EMCF1 R4 Q1
watertight seal. Remove
R1
and discard this plug when
RLY2 RLY1
cabling to this connection.
TB2
• The bulkhead will accept
TB1
cable diameters from
TB3
5.8 mm (0.23 in.) up to
+ 24 VDC ALARM2 ALARM1 + 10 mm (0.34 in.).
COM

COM
NO

NO

The terminals are designed to


NC

NC

+ SHLD –
B A
4-20mA/
+GF+ SIGNET
accept wires in the range of
RS-485 L1
L7
L2 3 2 1 14-28 AWG. All wires should
RED
L5
WHT
6 5 4 be stripped to a length of
TB4 ORG
L3
BRN
Loop1 6 mm (0.25 in.).
BLK GRN
L7
SHLD YEL
PWR
9900 • Ensure that the watertight
ISL 1 Loop2
S3L Inputs seal is maintained after the
+ +
L4
PWR

terminal box has been wired


for operation. If any of the
SHLD
DATA

S3 L
GND
V+

DATA

bulkheads are not tightened


N/C properly around a cable
7 8 9 or plug, the ratings of the
instrument will be jeopardized
and there is a possibility of
Black White (Gnd) creating a shock hazard.
Red (S3L) • Do not run 4 to 20 mA cables
Red
in the same conduit as power.
Black (5 VDC)
Power
Connection

16 9900 Transmitter
Power Wiring

CAUTION!
DO NOT connect your 9900 to AC power.
The 9900 MUST be powered by
10.8 to 35.2 VDC ONLY.

Stand-alone application, no current loop used

ProcessPro
(for reference only)
9900

Transmitter 9900
Terminals Terminals
Black Red
System Pwr Loop - 4 – Power
PWR + + Power Supply
Supply
System Pwr Loop + 3 + PWR – – 10.8 to 35.2 VDC
12 to 24 VDC Black
Red Loop +
AUX Power - 2 Loop –
AUX Power +
1

Connection to a PLC/Recorder, separate supply

ProcessPro
(for reference only)
9900

9900
Transmitter Terminals
Terminals PLC or Recorder
Blue
+ Red
Loop Input
– PWR + + Power Supply
4 to 20 mA
System Pwr Loop - 4 PWR – – 10.8 to 35.2 VDC
Loop +
System Pwr Loop + 3 Loop –
Black Blue
AUX Power - 2 – Power PLC or Recorder
Red Supply
AUX Power + 1 + 12 to 24 VDC + Loop Input
– 4 to 20 mA
Black
AUX power required for all 8750-2 systems

Loop Powered
ProcessPro
9900
(for reference only)

9900
Transmitter Terminals
Terminals
+
Black Power PWR + Power Supply
System Pwr Loop - 4 + 10.8 to 35.2 VDC
Supply PWR – Red

System Pwr Loop + 3 – 12 to 24 VDC Loop +
Loop –
AUX Power - 2 Black
PLC or Recorder
AUX Power + 1 PLC or Recorder

Red Loop Input –
+ Loop Input
4 to 20 mA + 4 to 20 mA
NOTE: Loop Power
cannot be used to power
certain Signet sensors.
See table on page 11.

9900 Transmitter 17
Relay and Open Collector Wiring

Open Collector wiring

NPN Style Wiring PNP Style Wiring


Pull-Up Resistor
..ȍ: 3:5
9WR9 3:5±
PWR + Loop +
PWR – /RRS± Input
Power
Loop + Supply PLC
Loop – Gnd 2&± OC+
Gnd
OC – OC+
Input 9WR9
PLC Power
Gnd Supply
Gnd
Pull-Down Resistor
If PLC needs 0 logic input when relay is
..ȍ:
not energized, set NORMAL to CLOSED
in the RELAY menu when using the Open
Collector (R1) with NPN style wiring. with NORMAL set to OPEN.

Alarm
PWR +
PWR –
Loop + V+
Loop – Power
Supply
OC – OC+ Gnd

• The 9900 Open Collector (R1) output provides high-speed switching capability.
Signal frequencies can reach 400 pulses per minute.
• The 9900 Open Collector (R1) output connection is dependent upon the type of circuit being controlled by the output.
• Most indicating instruments or control system inputs require a signal voltage of 0 to 5 V (TTL or CMOS logic levels) or 0 to 24 V.
Therefore, the 9900 Open Collector output circuits must be equipped with a pull-up or pull-down resistor (not supplied), and a
quality regulated 5 to 24 V (depending on the application) power supply (not supplied) is recommended to function properly.

Relay Module wiring


RELAY 2

The alarm is OFF during normal operation, NO


ALARM!
and will go ON when relay energizes NC

according to 9900 Relay settings. C

AC or DC
power
RELAY 3

The valve is ON during normal operation, NO


and will go OFF when relay energizes NC
C
according to 9900 Relay settings.

AC or DC
power

Valve
NO = normally open (closes when energized) Flow
NC = normally closed (opens when energized)

18 9900 Transmitter
Relay Functions

System Start-up: Step 3


Set your relay functions to your own application requirements.
Next step: System Setup (see page 25).
Once a setting is saved it becomes immediately active.

1. Go to the Relay Menu (RELAY flashing on screen, press ENTER). Example: Set a relay R1 to turn
2. If prompted, select desired source. on at a low setpoint of 5.5 gpm
3. Press ▼ to relay MODE selection screen. with a time delay of 15 seconds
4. If necessary, press ► and then ▼ or ▲ to select R1 MODE LOW. and turn off at 8.0 gpm.
Press ENTER to confirm. Remember,
5. Press ▼ to R1 SET LOW. Press ► to enter GPM value of 5.5. SET LOW + hysteresis = OFF point:
6. Press ENTER to save. 5.5 + 2.5 = 8.0
7. Scroll ▼ to the R1 HYSTERESIS menu.
8. Press ► to edit.
9. Set the hysteresis for this relay. This affects the turn off only: 2.5 gpm.
10. Press ENTER.
11. Scroll down ▼ to the R1 ON DELAY menu.
12. Press ► to edit.
13. Set the turn-on delay in seconds for the relay: 15.0.
14. Press ENTER.
15. Exit to View Mode.

● Relay function can be tested in the RELAY menu.

Relay and Open Collector Outputs

RELAY HIGH and LOW Settings


Depending on the desired function of the circuit attached to the Open Collector (R1) output, it may be necessary to have the
Open Collector turned “on” or “off” when the criteria for the activation of this output are met.
If the 9900 is set to operate in RELAY LOW mode, when the user-defined condition for the activation is met (e.g. exceeding an
alarm limit) the Open Collector switch is turned “on.” If wired as standard “NPN-style” output (see previous page) the logic level of
the attached control system or PLC input consequently becomes “low” logic level (when NORMAL is set to OPEN).
If a high input logic level is required for activation, it can be accomplished in one of three ways.
In order of preference,
1. Change the Open Collector (Relay 1) output function to “high” in the instrument's RELAY menu, or
2. Wire the Open Collector (R1) output “PNP” style as described on the previous page, or
3. Set the Open Collector (R1) to NORMAL CLOSED in the RELAY menu.
Fail-Safe Behavior
No matter the setting, the Open Collector output turns off if the 9900 loses power. This must be taken into account when evaluating
system failure consequences. If the system layout requires a “closed” or “on” condition for the output in case of power loss, a
mechanical dry-contact relay (NC contacts) must be used instead of the Open Collector (R1) output.
Voltage and Current Limitation
The supply voltage in the Open Collector output circuit MUST be limited to the specified maximum Open Collector voltage
(see operating manual for specific instrument). The use of a quality 5 to 24 V (depending on the application) regulated power supply
(not supplied) is recommended.
The current through the Open Collector switch also must be limited. Typical Open Collector outputs allow only for 10 to 50 mA
switch current. Exceeding this current limit can burn out the Open Collector output components immediately.
Load and Pull-Up/Down Resistor Considerations
By utilizing basic arithmetic and Ohm’s Law, the safe limits of load resistance can be determined. When the Open Collector switch is
closed, almost the entire supply voltage is applied to the load (e.g., the pull-up or pull-down resistor, the alarm horn input, a potential
power relay coil or annunciator lamp).The resulting current through the load and through the Open Collector switch, as well, can be
calculated as:
(Current) = (Supply Voltage) / (Load Resistance)

9900 Transmitter 19
Relay Outputs
The 9900 open collector and relays are selectable and Open Collector Output
configurable and can be used as switches that respond when ● Longer life than a mechanical relay
the process value moves above or below a user-defined ● No moving parts
setpoint or it can be used to generate a pulse at a rate ● Faster ON/OFF switching capabilities than mechanical relays
proportional to the process value. ● Can switch DC voltage only (< 30 VDC)
They can be used for Low Alarm, High Alarm or Proportional ● Not recommended for use with inductive loads.
Pulse triggering related to the process value. All relay
functions are set up in the RELAY menus.

Relay energized
Relay de-energized

Process Low Setpoint:


Relay is on when the measured value is less than the setpoint.

Hysteresis

Low Setpoint

Time

Process High Setpoint:


Relay is on when the measured value is higher than the setpoint.
High Setpoint

Hysteresis

Time

Cycle Time Cycle High/Low:


Process The relay can stay energized for a set length of time after the
process value goes above (or below) the setpoint.
High Setpoint
The relay will stay on for the CYCLE TIME and then turn off,
Hysteresis even if the process value is still above (or below) the setpoint.
The cycle will not repeat until the process value goes below
(or above) the setpoint minus the hysteresis after the relay
Time times out.

In FLOW, Cycle High activates the relay each time the volume
Relay energized
reaches the SET VOLUME setpoint (see page 28).
Relay de-energized

NOTE: To reset the timer (or volume in Flow): in the RELAY


menu, select TEST RELAY function. The timer will reset to 0
if the condition no longer exists when the TEST is performed.
The timer will restart if the condition still exists.

CAUTION!

If power is lost to the 9900 Transmitter during a


cycle, the Cycle Time will reset. If the condition
still exists after power is restored, the relay will be
energized for the complete Cycle Time.

20 9900 Transmitter
Relay Outputs
Window In/Out: Process
Relay is on when the value is higher or lower than the high or
low setpoint. High Limit
Hysteresis
WINDow IN = relay on if measurement is inside the window Window
of two setpoints. Measurement inside the two setpoints is
Low Limit
abnormal condition. Hysteresis
WINDow OUT = relay on if measurement is outside the
window of two setpoints. Time

Window IN example

Relay energized
Relay de-energized

Process

High Limit
Hysteresis

Window

Low Limit

Time

Window OUT example

Proportional Pulse Operation:


The transmitter can output a pulse at the rate defined by the

:
in. ate
settings in the CAL menu and the sensor input. The maximum 100 puls
pulses
s/m ut R
100
pulse output from the relays is 300 pulses per minute. lse utp
Example usage would be to control solenoid-operated
Pu O
0 or

dosing pumps.
10 ct

Pulse rate
to lle
0 Co

For example: As the process value drops below the setpoint,


n
pe

the output will start pulsing in relation to the process value,


O

the maximum pulse endpoint and the programmed


pulses/minute. The pulse rate will change as the process
value changes and approaches the programmed endpoint. 0 pulses
This functionality can be used to precisely control the process.
5 10
Starting point Endpoint
The starting point, endpoint and maximum pulse rate are
selectable in the RELAY menus.

NOTE: Relay LEDs are not lit in PULSE mode.

In the example:
• The output will be 0 pulses/min. when value is less
than 5.
• The output will be 50 pulses/min. when value is 7.5.
• The output will be 100 pulses/min. when value is
greater than 10.

9900 Transmitter 21
Relay Outputs

PWM • Pulse Width Modulation


Maximum
Range PWM automatically varies the ratio of ON time to OFF time
100% proportional to minimum and maximum range settings.
Relay is
(as a % of Relay period)

always ON The relay period is the sum of the time a relay is ON and the
Relay Pulse Width

80%
time it is OFF.
60% Relay pulse width is the time the relay is ON.

40% The 9900 must be programmed with the relay period, and with
the low and high setpoints.
20% NOTE: The PWM mode is not used for Pressure applications.
0% Process value NOTE: Relay LEDs are not lit in PWM mode.
Relay is PWM
always OFF Minimum

Example:
• The pulse width will be 0% of the relay period
(relay always OFF) when the process value is
less than the minimum range.
• The pulse width will be 100% of the relay period
(relay always ON) when the process value is
greater than the maximum range.
• The pulse width will be 60% of the relay period when
the process value is at 60% of the span between the
minimum and maximum range.

• Volumetric Pulse
A pulse is generated each time a specified volume of fluid is
registered. For flow inputs only.

NOTE: Relay LEDs are not lit in VOLUMETRIC PULSE mode.

• Totalizer Volume
Relay activates and latches when a specified volume of fluid is
registered. For Flow inputs only.
Total Volume mode counts the TOTALIZER Units until the
setpoint volume is reached, then turns on the relay until the
resettable totalizer is reset.
If the Resettable Totalizer reading is greater than the setpoint,
the relay will be turned on immediately. The relay will be off
when the totalizer is reset to zero.
This mode is useful to trigger a reminder when a process is
due, as for a backwash cycle or filter change.

22 9900 Transmitter
Operation

Open Collector (R1) Backlight Sensor


Indicator LED (do not block)
Mechanical Relay (R2) Mechanical Relay (R3)
Indicator LED Indicator LED
Warning LED

Bar Graph
Units of Measure
(GPM, pH, sec, %, etc.)

Value

Label

Menu Indication

Menu Navigation Keys

ENTER

All possible segments shown in this illustration. The instrument's software controls which segments are shown at
any particular time. Only the bar graph and GF logo are visible when the unit is turned off.

Warning LED will be lit when No Sensor or Wrong Sensor is detected in Digital (S3L) mode.

UP, DOWN keys


Scroll through Menu options or adjust values during editing
Press both together to exit a menu or escape without saving

RIGHT key
Select item or character to edit

ENTER key
ENTER Access menus
Save changes

9900 Transmitter 23
Operation

Keypad Functions
3s The four buttons of the keypad are used to navigate display modes
according to the descriptions in this table. Notice that the function
ENTER of each button may change depending on the display mode.

Edit

System Setup: Menu Navigation


This basic operating procedure repeats throughout the 9900 program:
1. Press ENTER for 3 seconds to enter MENU mode.
2. Press ► to move to the desired menu then press ENTER to
select it. (Password may be required.)
or or
3. Press ▲ or ▼ to select the desired menu item for editing.
4. Press ► to edit the value/selection.
ENTER (Password may
5. Press ENTER to store the new value/selection.
be required) 6. Press ▲ or ▼ to select another menu item if desired.
Repeat steps 3-5 as required.
7. Press ▲+▼ to select a different menu to edit.
Input Edit Repeat steps 2-5 as required.
8. When finished editing all menus, press ▲+▼ again to return to
normal operation.

The menu is constructed in a loop, so you can move forward and


backward to select an item. After any item is saved (by pressing
ENTER), the display will return to the previous menu.
Input Edit Choices

ENTER
or

Saves Changes
+
+
2x Or Select another
Menu Item

Return to
View Mode

24 9900 Transmitter
Menu System

System Setup Menu

All of the basic system setup functions are automated in the


9900 for many sensors and sensor electronics. This includes
identifying the sensor connected to the 9900, and configuring
the display for the sensor. After installation and wiring is
completed, apply power to the 9900.

When the 9900 is first powered on, it will attempt to determine


the sensor type connected when ENTER is pressed (unit will
display LOOKING FOR).
If no sensor is attached to the 9900, the words “TYPE” and ENTER
“FLOW” are displayed.
If the 9900 does not identify your sensor type correctly,
use the ▲ and▼ keys to select your sensor type.

As you scroll through the available sensor types, press ► to


select the desired sensor and then press ENTER. Sensor Choices

You may change sensor type after initial


power-on (if the sensor type is changed
after your 9900 is already in service).
Enter the INPUT menu, scroll to TYPE,
press ►, and scroll to select the desired sensor type
(you may be prompted for your password).
Press ENTER. The bottom line will display ALL SETTINGS
WILL BE RESET. ARE YOU SURE? The top line of the
display will blink NO (unless switching from Factory mode).
Press ▼ or ▲ to select YES.
Press ENTER again to finalize your selection.

NOTE: User is strongly discouraged from changing the


sensor type away from the correct sensor.

ENTER

NOTE:
The 9900 displays the BATCH instrument type
ONLY if the Batch Module is installed.

9900 Transmitter 25
Menu System

VIEW Mode Overview Error Handling


The top level of menus is referred to as the VIEW Mode. This view displays measurement Errors occurring while in the
values as well as current outputs and relay status. The radial bar graph represents the VIEW Mode show a specific
measurement value that is also displayed in the 7-segment numeric field below the bar message (e.g., CHECK SENSOR).
graph. The bar graph is primarily used to display the full scale range of the sensor, but can This message is displayed every
be scaled via a menu item. 10 seconds and stays on for 5
seconds. Once the error is
During normal operation, the 9900 displays the VIEW mode. resolved or cleared, the error
● To select a display, press the ▲ or ▼ arrow keys. message stops.
The display selections scroll in a continuous loop.
● Changing the display selection does not interrupt system operations.
● No password is necessary to change display selection.
● Output settings cannot be edited from the View Mode. Scrolling
● The display will return to the VIEW mode if no button is pressed for 10 minutes. In some cases, more than one
message or measurement may
need to be displayed. This is
accomplished by alternating the
MENU Mode Overview message portions across the
The MENU mode enables the user to view and configure all menu items. screen.
The five menus available are: CAL, INPUT, LOOP, RELAY, and OPTION.

MENU Mode is entered by pressing and holding ENTER for three seconds.

The ► button is used to change the position of the blinking cursor.


When the desired menu is blinking, press ENTER.

In the selected menu, use the ▲ and ▼ keys to navigate through the menu.
Use the ▲, ▼ and ► keys to edit the selected item
(see Menu Navigation discussion, page 24).

To save the new selection, press the ENTER key. A message displaying “Saving…” will be
displayed for 3 seconds. After this message is displayed, the newly selected value will be
displayed, if applicable.

Password Overview
The password is often required to start editing. Once entered correctly, this password will In the MENU mode, if the
not be needed for subsequent edits. However, once the menu system is exited, wrong code or password is
the password will again be required when edit mode is re-entered. entered, an ERROR message
is displayed.
Your choice of password (STD or CODE) is selected in the Options Mode.

● STD
The standard (STD) password is ▲▲▲▼, pressed in sequence. This password is
designed to protect the 9900 from unintentional changes. It is best suited for systems
where a group of people need to be able to change settings.
● CODE
The CODE default setting is 0000, adjustable to any 4-digit numerical code up to
9999. Using a personal code provides the maximum degree of security.
This code can be modified in the Options mode. To change your CODE, go
to OPTIONS mode, enter
your desired code and press
ENTER. (The STD password
cannot be changed.)

26 9900 Transmitter
Common Menus

System Start-up: Step 4


Customize your 9900 to your own installed sensors.

Common Menus
The menu system shares certain modes between sensor types.
The following describes the EDIT Mode menus found in common between most sensor types.

NOTE:
Menu and Mode displays shown are examples only.
Your displays may vary.

INPUT Menu

(ALL) Manually select Sensor Type (See page 25 for further instruction).
Allows user to reset 9900 Transmitter to Factory settings.
Note: User is strongly discouraged from changing the sensor type away from the correct sensor.

The following can individually be set for each current loop


LOOP Menu (Loop1 = 9900 base unit, Loop2 = Output Module)

(pH, COND/RES, LEVEL, SALINITY, DO only)


Set LOOP output source; select between Primary and Secondary measurements of applicable sensor.
Secondary measurements: DO, pH, COND/RES, and SALINITY = TEMP; LEVEL = VOL.

(COND/RES only) Select LIN/LOG. Default = LIN.


See LOG Current LOOP Output discussion in Appendix.

(ALL) Set value corresponding to desired 4 mA output. 5 digits max. Default = 0 (ORP = -999).

(ALL) (Not shown in COND/RES LOG Mode)


Set value corresponding to desired 20 mA output. 5 digits max.
Defaults = 100 (Flow, Cond/Res, Temp), 14 (pH), 1000 (ORP), 10 (Lvl/Prs), 5 (4 to 20 mA), 80 (Sal).

(ALL) Set desired LOOP output value when sensor error (e.g., bad sensor, broken wire) is detected.
Select (3.6 mA, 22 mA). Default = 22.

(ALL) Allows fine-tuning to compensate for errors in other equipment connected to the 9900.
Adjust the minimum and maximum current output. The display value represents the precise current output.
Adjustment limits: from 3.80 mA minimum to 5.00 mA maximum. Default = 4.00 mA.

(ALL) Allows fine-tuning to compensate for errors in other equipment connected to the 9900. Adjust
the minimum and maximum current output. The display value represents the precise current output.
Adjustment limits: from 19.00 mA minimum to 21.00 mA maximum. Default = 20 mA.

(ALL) Press ▲ or ▼ to manually order any output current value from 3.8 mA to 21.00 mA to test the
output of LOOP.

9900 Transmitter 27
Common Menus

RELAY Menu The following can individually be set for each relay (R1 = open collector, R2/R3 = Relay Module)

(pH, LEVEL/VOL, COND/RES, SALINITY and DO only.) Select source for each of R1, R2 and R3 outputs.
Choose pH/TEMP, LEVEL/VOLUME, COND/TEMP, SAL/TEMP, (DO)PPM/TEMP.
Defaults = pH, COND, LEVEL, SAL, PPM.

(ALL) Set Open Collector (R1) as Normally Open or Normally Closed. Default = OPEN.

(ALL) Select the desired mode of operation for the open-collector (R1) output (OFF, LOW, HIGH,
WINDow IN, WINDow OUT, CYC LOW (except FLOW), CYC HIGH, PROP PuLSe, VOL PuLSe, PWM,
TOTAL, USP, ERROR mode) (See chart on page 29). Default = OFF. Continue stepping through to
select R2 and R3 output modes. When MODE is set to ERROR, delays energizing relay until after
ON DELAY time expires if sensor problem is detected. See Cycle High/Low discussion on page 20.

(ALL) (Shown if LOW, WIND IN/OUT or CYC LOW mode)


Relay turns on if process measurement goes lower than this value. Set desired value.
NOTE: The corresponding indicator lights do not light up in PROP PLS and PWM modes.
The LEDs light up only when the Test Relay options are selected.

(ALL) (Shown if HIGH or WIND IN/OUT mode)


Relay turns on if process measurement goes higher than this value. Set desired value.
NOTE: The corresponding indicator lights do not light up in PROP PLS and PWM modes.
The LEDs light only when the Test Relay options are selected.

(FLOW only) (Shown if CYC HIGH or VOL PLS mode)


Amount of accumulated flow that must be counted before a pulse is sent out.
Relay turns on if flow volume exceeds this value. Set desired value. Default = 100.00.

(ALL) (Shown if LOW, HIGH, WIND IN/OUT, CYC LOW/HIGH or USP mode)
Hysteresis prevents the system from chattering around the set point.
Set amount (in units of measure from INPUT Mode) to add to SET LOW or SET HIGH values.

(COND/RES only) (Shown only in USP mode)


Relay turns on if USP value drifts by this value away from USP limit.
See USP Limits discussion in the Appendix.

(ALL) (Shown if Low, High, WIND IN/OUT, CYC LOW/HIGH or Error mode)
Set seconds (up to 9999.9) to wait before activating relay.

(ALL except PRESSURE) (Shown only if PROP PLS mode)


Set minimum setpoint value for proportional pulsing.

(ALL except PRESSURE) (Shown only if PROP PLS mode)


Set maximum setpoint value for proportional pulsing.

(ALL except PRESSURE) (Shown only if PROP PLS mode)


Set desired maximum pulse rate (300 max)
NOTE: Pulse width fixed at 100 ms.

(ALL except PRESSURE and FLOW) (Shown only if PWM mode)


Set minimum value for pulse width modulation.

(ALL except PRESSURE and FLOW) (Shown only if PWM mode)


Set maximum value for pulse width modulation.

NOTE: Defaults for most relay functions are dependent upon sensor type and are not listed here.

28 9900 Transmitter
Common Menus

RELAY Menu The following can individually be set for each relay (R1 = open collector, R2/R3 = Relay Module)

(ALL) (Shown only if CYC LOW/HIGH mode)


Set time in seconds (up to 99999) for relay to remain on. See discussion on page 20.

(FLOW only) (Shown only if VOL PULS mode)


Amount of accumulated flow that must be counted before a pulse is sent out. Set value.

(FLOW only) (Shown only if VOL PULS mode) Set time value for one pulse width.

(ALL except PRESSURE and FLOW) (Shown only if PWM mode)


Set time value for one complete pulse cycle. (relay ON time + relay OFF time).

(FLOW only) (Shown only if TOTAL) Resettable value that, when exceeded, turns relay on.
Must reset Totalizer (in VIEW Mode) to clear relay. Set maximum value.

(ALL) Press ▲ or ▼ to turn relay on or off for testing purposes.


Can also be used to reset or latch/unlatch the relay. Does NOT reset the Totalizer.

Available Relay Modes by Sensor Type

Flow pH ORP Cond/Res Pressure Lvl/Vol Temp 4 to 20 mA Salinity DO


Off X X X X X X X X X X
Low X X X X X X X X X X
High X X X X X X X X X X
Wind In X X X X X X X X X X
Wind Out X X X X X X X X X X
Cyc Low X X X X X X X X X
Cyc High X X X X X X X X X X
Prop Pulse X X X X X X X X X
Vol Pulse X
PWM X X X X X X X X
Total X
USP X*
Error X X X X X X X X X X

* In USP Relay Mode in Conductivity, Relay Source must be set to COND, TEMP COMP must be set to NONE
and Unit Of Measure must be set to μS.

9900 Transmitter 29
Common Menus

OPTION Menu

Adjust the LCD contrast for best viewing for your environment.
A setting of 1 is lowest contrast, 5 is highest. Default = 3.

Select backlight level (OFF, LOW, HIGH, AUTO). Default = AUTO.


NOTE: No backlight when operating on loop power.

Enter 5 digit value to represent bar at minimum.


Default = 0 (ORP = -999).

Enter 5 digit value to represent bar at maximum.


Defaults = 100 (Flow, Cond/Res, Temp), 14 (pH), 1000 (ORP), 10 (Lvl/Prs), 5 (4 to 20 mA),
80 (Sal), 20 (DO)

(ALL) Set the decimal to the best resolution for your application.
The display will automatically scale up to this resolution.
Select -----., ----.-, ---.--, --.---, or -.---- (varies by parameter). Default = ----.-.

(pH, COND/RES, TEMP, SAL, DO only)


Set the Temperature decimal to the best resolution for your application.
The display will automatically scale up to this resolution. Select -----., ----.-, or ---.--. Default = ----.-.

(FLOW only) Set the decimal to the best resolution for the Permanent Totalizer display.
The display will automatically scale up to this resolution. Select -----., ----.-, ---.-- or --.---.
Default = ----.-

(COND/RES only) Displays mS or μS as set in COND UNITS in INPUT Mode.


Set ON/OFF. Default = OFF.

(FLOW only) Locks the TOTALIZER output. Select OFF, ON (Does not affect Permanent Totalizer).
Default = OFF.

(ALL) Select STD, CODE.


Default = STD.

(ALL) Enter desired password code. 4-character entry not displayed, ---- displayed instead.
(Shown if type = CODE.)

(ALL) Enter 13-character string, if desired.


Default = Blank.

Enables Remote Setup to configure the 9900 via a computer and the optional 0252 Configuration tool.
Press ► and select YES to enable. REMOTE SETUP flashes when mode is enabled.
NOTE: Communication with 0252 tool is automatic when 9900 is in FACTORY state (Enter flashing).
Refer to the 0252 Configuration Tool manual, 3-0252.090, included with your configuration tool.

Displays Transmitter Generation Version.

30 9900 Transmitter
Sensor-Specific Menus

The following pages list the sensor-specific settings for each sensor type.

Flow

FLOW Setup Checklist


1. Make sure FLOW sensor type is selected
(see System Setup Menu, page 25).
2. Set the Units of Measurement.
3. Set Sensor Type (Freq or S3L).
4. If LOOP is used, set the minimum and maximum 4 to 20 mA setpoints.
5. Set K-Factor (pulses per Unit Volume) from Flow Sensor manual.
6. Set Totalizer factor.
7. Set Last Cal Date and initials.
8. If desired, set up relay functions for your own application.
This is the normal display
and does not time out.

VIEW Mode Menu


Display the flow rate and the resettable totalizer. Press ► to reset the totalizer. (If Reset is locked, enter
the password first.) Lock or Unlock the totalizer in the OPTIONS menu.
This is the resettable totalizer View display.
Display the Permanent Totalizer value (note the "P" indicating Permanent).
Pressing ► displays units of measure.

Displays the 4 to 20 mA LOOP output.

Bottom line shows one of three states (OFF, ON, PLS) for each of the three relays.
Displays remaining time for CYC LOW or CYC HIGH mode. The relay(s) will remain ON while counting down.
NOTE: (Reset the Timer) In the RELAY menu, select TEST RELAY function.
The timer will reset to 0 if the condition no longer exists when the TEST is performed.
The timer will restart if the condition still exists.

CAL Menu

YES prevents relays from activating while making adjustments, and relays in PULSE mode will suspend
pulsing. Output is held until the user exits the CAL menu. Select YES/NO.
Default = NO.

Set K-Factor (pulses per unit volume) from Flow Sensor manual.
Min: 0.0001, max 999999. Cannot be zero. Default = 60.0000.

Sets the volume of each count of the Totalizer as a multiple of the volume unit of the K-Factor.
Min: 0.0001, max 999999. Cannot be zero. Default = 1.0000.

Select to calibrate using Rate method (see Appendix).

Select to calibrate using Volume method (see Appendix).

Enter date of calibration (mm-dd-yyyy) and initials of calibrator (ii).

9900 Transmitter 31
Flow

INPUT Menu

If desired, a custom name can be entered.


Enter 13-character string.
Default = FLOW.

If your flow sensor is configured for frequency output, select FREQ.


If configured for Digital (S3L) output (recommended), select (S3L).
Default = FREQ.

Set the units of measure.


The last character sets the timebase: S (seconds) M (minutes) H (hours) D (days).
Default = GPM.

Identifies the Totalizer units. It has no effect on any calculation.


Default = GALLONS.

(Only shown if HART Module is installed. 9900 Gen IV, v.24-00 or later)
Select Permanent or Resettable Totalizer (as secondary value).
NOTE: Totalizer units are transmitted as unit code 240 via HART communication.

Dampens display, output and relay response rates.


Select Low, Med, High, OFF. (See discussion in Appendix.)
Default = OFF.
The Sensitivity setting determines how the 9900 responds to sudden changes in the flow rate.
The value is expressed in units of measurement.
If the setting is exceeded, it "overrides" the Averaging function breifly to allow for the actual change in
flow rate to be displayed. Averaging resumes shortly after.
The result is a smooth flow display and a quick response to large shifts in the flow rate.
(See discussion in Appendix, page 51.)

32 9900 Transmitter
pH pH

pH Setup Checklist
1. Make sure pH sensor type is selected
(see System Setup Menu, page 25).
2. Set the Temperature Units (°C or °F).
3. If LOOP is used, set the minimum and maximum 4 to 20 mA setpoints.
4. Perform calibration (EasyCal, Standard or Standard and Slope).
5. Set Last Cal Date and initials.
6. Select source for Open Collector and Relay output (pH or Temp).
7. If desired, set up relay functions for your own application.

This is the normal display


and does not time out.

VIEW Mode Menu

Displays temperature at the sensor.

Displays the millivolt input from the electrode. Use this display to determine the relative condition of your
electrode during periodic calibration. (7 pH buffer = 0 mV, ± 50 mV)
(2751 preamp only, 9900 Generation IV or greater) Displays last AUTOMATED or MANUAL GLASS
IMPEDANCE measurement.
Press ► to MANUALLY measure GLASS IMPEDANCE.
AUTOMATED GLASS IMPEDANCE update time is set under the pH INPUT menu (see next page).
NOTE: While AUTOMATED/MANUAL GLASS IMPEDANCE measurement is in progress, all outputs are
held and an "UPDATING" message will be displayed until the measurement is completed.

(2751 preamp with memory chip enabled sensors only, 9900 Generation IV or greater)
Press ► to access sensor data stored in sensor memory chip.
NOTE: Pressing ▲+▼, or changing to a sensor without a memory chip, while viewing SENSOR DATA,
will return the 9900 to the top VIEW mode display (current pH measurement).

(2751 preamp with memory chip enabled sensors only, 9900 Generation IV or greater)
Sub-menu of stored sensor data, accessed by pressing ► on the above SENSOR DATA menu.
SENSOR DATA includes:
Sensor Serial Number (S/n), Sensor Model Number (MOD), pH Slope, pH Offset, Temperature Offset,
Factory Reference Impedance, Total Usage Time (HRS), Minimum pH value measured (MIN PH),
Maximum pH value measured (MAX PH), Minimum Temperature value measured (MIN TEMP),
Maximum Temperature value measured (MAX TEMP).

Displays the 4 to 20 mA LOOP output.

Bottom line shows one of three states (OFF, ON, PLS) for each of the three relays.
Displays remaining time for CYC LOW or CYC HIGH mode. The relay(s) will remain ON while counting down.
NOTE: (Reset the Timer) In the RELAY menu, select TEST RELAY function.
The timer will reset to 0 if the condition no longer exists when the TEST is performed.
The timer will restart if the condition still exists.

9900 Transmitter 33
pH

CAL Menu

Select AT SENSOR to perform calibration using the Signet 2750/2751 sensor electronics.
Select AT INSTRUMENT to perform calibration at the 9900 via EasyCal or manual calibration.
(See pH Calibration procedures in the Appendix.) Default = AT INSTRUMENT.

YES prevents relays from activating while making adjustments, and relays in PULSE mode will
suspend pulsing. 4 to 20 mA output is held until the user exits the CAL menu. Select YES/NO. Default = NO.

(CAL AT INSTRUMENT only) Press ► to start the EasyCal process.


You will be prompted to enter your password. (See pH EasyCal procedure in the Appendix).

(CAL AT INSTRUMENT only) Applies a linear offset to the pH measurement. The ideal value is the average
pH of your application. (A sample of your application at process temperature is recommended) (See pH
Calibration procedures in the Appendix) Shows error message if offset too high.

(CAL AT INSTRUMENT only) Applies a slope to the pH measurement. The slope value and the standard
value must be at least 2 pH units apart. The ideal values are the minimum and maximum values of your
process. (See pH Calibration procedures in the Appendix.) Shows error message if slope is too low or high.

(CAL AT INSTRUMENT only) Applies a linear offset to the temperature measurement.


The ideal value is the average temperature of your application. "SAVING" will appear if offset is acceptable,
"ERR TOO LARGE TO CALIBRATE" if offset is outside of range.

(CAL AT INSTRUMENT only) Press ► to reset pH Calibration to factory default.

(CAL AT INSTRUMENT only) Press ► to reset temperature calibration to factory default.

Enter date of calibration (mm-dd-yyyy) and initials of calibrator (ii).

INPUT Menu

Enter string up to 13 characters (optional).


Default = MEAS TYPE PH.

Select °F or °C.
Default = °C.

Dampens display, output and relay response rates.


Select Low, Med, High, OFF. Default = OFF. (see discussion in Appendix).
NOTE: Signet strongly recommends leaving averaging OFF for pH and Pressure measurements

(2751 preamp only, 9900 Generation IV or greater) Set AUTOMATED GLASS IMPEDANCE update time
from 0 to 1440 minutes.
Setting update time to 0 turns off automatic glass impedance updates. Default = 60 minutes.

(2751 preamp with memory chip enabled sensors only, 9900 Generation IV or greater)
High Impedance Warning is activated when the AUTOMATED GLASS IMPEDANCE measurement is
greater than the Factory Glass Impedance by selected multiplier.
Select multiplier of 3, 4, 5, or OFF. Default = OFF.
NOTE: Factory Glass Impedance can be found under the VIEW mode by pressing ► on SENSOR DATA.

34 9900 Transmitter
ORP

ORP Setup Checklist


1. Make sure ORP sensor type is selected
(see System Setup Menu, page 25).
2. If LOOP is used, set the minimum and maximum 4 to 20 mA setpoints.
3. Set Averaging.
4. Perform calibration or set Standard (and Slope if desired).
5. Set Last Cal Date and initials.
6. If desired, set up relay functions for your own application.

This is the normal display


and does not time out.

VIEW Mode Menu

Displays the millivolt input from the electrode.


Use this display to determine the relative condition of your electrode during periodic calibration.

(2751 preamp with memory chip enabled sensors only, 9900 Generation IV or greater)
Press ► to access sensor data stored on sensor memory chip.
NOTE: Pressing ▲+▼, or changing to a sensor without a memorychip, while viewing SENSOR DATA,
will return the 9900 to the top VIEW mode display (current ORP measurement).

(2751 preamp with memory chip enabled sensors only, 9900 Generation IV or greater)
Sub-menu of stored sensor data, accessed by pressing ► on the above SENSOR DATA menu.
SENSOR DATA includes: Sensor Serial Number (S/n), Sensor Model Number (MOD), ORP Slope,
ORP Offset, Total Usage Time (HRS), Minimum ORP value measured (MIN mV),
Maximum ORP value measured (MAX mV)

Displays the 4 to 20 mA LOOP output

Bottom line shows one of three states (OFF, ON, PLS) for each of the three relays.
Displays remaining time for CYC LOW or CYC HIGH mode. The relay(s) will remain ON while counting down.
NOTE: (Reset the Timer) In the RELAY menu, select TEST RELAY function.
The timer will reset to 0 if the condition no longer exists when the TEST is performed.
The timer will restart if the condition still exists.

9900 Transmitter 35
ORP

CAL Menu

Select AT SENSOR to perform calibration using the Signet 2750/2751 sensor electronics.
Select AT INSTRUMENT to perform calibration at the 9900 via EasyCal or manual calibration.
(See ORP Calibration procedures in the Appendix.). Default = AT INSTRUMENT.

YES prevents relays from activating while making adjustments, and relays in PULSE mode will
suspend pulsing. Output is held until the user exits the CAL menu. Select YES/NO.
Default = NO.

(CAL AT INSTRUMENT only) Press ► to start the EasyCal process.


You will be prompted to enter your password. (See ORP EasyCal procedure in the Appendix).

(CAL AT INSTRUMENT only) Applies a linear offset to the ORP measurement. For single point calibrations,
assign the average value of your process to ORP STANDARD. For two-point calibrations, assign the min or
max value of your process to ORP STANDARD. (See ORP Calibration procedures in the Appendix).

(CAL AT INSTRUMENT only) Applies a slope to the ORP measurement. The ORP SLOPE is used for two-point
calibration along with the ORP STANDARD. If you applied the min value of your process to the ORP STANDARD,
then apply the max value to the ORP SLOPE. Else, apply the min value to the ORP SLOPE. The slope value and
the standard value must be at least 30 mV apart. (See ORP Calibration procedures in the Appendix)

(CAL AT INSTRUMENT only) Resets calibration to factory settings. After pressing ►, select YES/NO.

Enter date of calibration (mm-dd-yyyy) and initials of calibrator (ii).

INPUT Menu

Enter string up to 13 characters (optional).


Default = ORP.

Dampens display, output and relay response rates. Select Low, Med, High, OFF.
(See discussion in Appendix.) Default = OFF.

36 9900 Transmitter
Conductivity / Resistivity

Cond/Res Setup Checklist


1. Make sure COND/RES sensor type is selected
(see System Setup Menu, page 25).
2. Set Cell Constant.
3. Set the Temperature Units (°C or °F).
4. Set Conductivity units.
5. If LOOP is used, set the minimum and maximum 4 to 20 mA setpoints.
6. Set Temperature Compensation.
7. Set Last Cal Date and initials.
8. Select source for Open Collector and Relay output (COND or TEMP).
This is the normal display 9. If desired, setup relay functions for your own application.
and does not time out.

VIEW Mode Menu

Same as above with temperature, does not time out.

Displays the 4 to 20 mA LOOP output.

Bottom line shows one of three states (OFF, ON, PLS) for each of the three relays.
Displays remaining time for CYC LOW or CYC HIGH mode. The relay(s) will remain ON while counting down.
NOTE: (Reset the Timer) In the RELAY menu, select TEST RELAY function.
The timer will reset to 0 if the condition no longer exists when the TEST is performed.
The timer will restart if the condition still exists.

CAL Menu

YES prevents relays from activating while making adjustments, and relays in PULSE mode will suspend
pulsing. Output is held until the user exits the CAL menu. Select YES/NO.
Default = NO.

Shows real-time value and selected standard. “PLACE SENSOR IN STANDARD”.


Unit waits until reading is stable; if bad cal, returns “ERROR, CANNOT DETERMINE STANDARD”.
Refer to buffer values and AUTO CAL Procedure in the Appendix.
Shows “CONDUCTIVITY” on bottom line; when user presses any button the live value is frozen and the
user edits that value. If bad cal, returns “ERR TOO LARGE TO CALIBRATE”.
See Manual Cal procedure in Appendix.

Shows “TEMPERATURE” on bottom line; when user presses any button the live value is frozen and the
user edits that value. If bad cal, returns “ERR TOO LARGE TO CALIBRATE”.

Resets Conductivity calibration. After pressing ►, select YES/NO.

Resets Temperature calibration. After pressing ►, select YES/NO.

Enter date of calibration (mm-dd-yyyy) and initials of calibrator (II).

9900 Transmitter 37
Conductivity / Resistivity

INPUT Menu

Enter string up to 13 characters (optional).


Default = COND/RES

Enter cell constant of sensor. Select 20.0, 10.0, 1.0, 0.1, 0.01, or CUSTOM.
Default = 1.0 (See NOTE below)

Enter the precise cell constant from the certificate provided with your sensor, or from the information
label on the sensor. Shown if CELL CONSTANT = CUSTOM. (See NOTE below)

Select °C, °F.


Default = °C

Select μS, mS, PPM, PPB, KOhm, or MOhm.


Default = μS.
NOTE: In USP Relay Mode, TEMP COMP must be set to NONE and Unit Of Measure must be set to μS.

If the COND UNITS selection is PPM or PPB, set the ratio of Total Dissolved Solids to μS.
Default = 0.50

Dampens display, output and relay response rates.


Select Low, Med, High, or OFF. (See discussion in Appendix)
Default = OFF

Select temperature compensation (NONE, LINEAR, PURE H2O). Default = LINEAR.


NOTE: In USP Relay Mode in Conductivity, Relay Source must be set to COND, TEMP COMP must be
set to NONE and Unit Of Measure must be set to μS.

For LINEAR or PURE H2O temperature compensation, select a % per ºC slope.


Maximum slope setting is 9.99 % per ºC. Default = 2.0
(If Temperature Compensation setting is NONE, this item will not be displayed)

Factory-Set Span:
0.01 cell (2819, 2839) ......... 0 to 100 μS
0.10 cell (2820, 2840) ......... 0 to 1000 μS
1.0 cell (2821, 2841) ........... 0 to 10,000 μS
10.0 cell (2822, 2842) ......... 0 to 200,000 μS
20.0 cell (2823) ................... 0 to 400,000 μS

NOTE: If using a 2850 Conductivity/Resistivity Sensor Electronics in conjunction with your 9900, the 2850 must
be set for the custom cell constant or the actual probe cell constant and the 9900 set for a 1.0 cell constant.

38 9900 Transmitter
Pressure

PRESSURE Setup Checklist


1. Make sure PRESSURE sensor type is selected
(see System Setup Menu, page 25).
2. If LOOP is used, set the minimum and maximum 4 to 20 mA setpoints.
3. Set Units of Measurement (PSI, BAR, KPa).
4. Set Last Cal Date and initials.
5. If desired, set up relay functions for your own application.

This is the normal display


and does not time out.

VIEW Mode Menu

Displays the 4 to 20 mA LOOP output

Bottom line shows one of three states (OFF, ON, PLS) for each of the three relays. Displays remaining
time for CYC LOW or CYC HIGH mode. The relay(s) will remain ON while counting down.
NOTE: (Reset the Timer) In the RELAY menu, select TEST RELAY function. The timer will reset to 0 if
the condition no longer exists when the TEST is performed.
The timer will restart if the condition still exists.

CAL Menu
ON prevents relays from activating while making adjustments, and relays in PULSE mode will suspend
pulsing. Output is held until the user exits the CAL menu. Select OFF/ON.
Default = OFF.

With process pressure at zero, set zero point for measurement.

Calibrate pressure reading to external reference. Provides a maximum 5 psi offset.

Resets calibration to factory default. After pressing ►, select YES/NO.

Enter date of calibration (mm-dd-yyyy) and initials of calibrator (II)

INPUT Menu

Enter string up to 13 characters (optional).


Default = PRESSURE.

Enter units of pressure measurement. Select PSI, BAR, or KPa.


Default = PSI.

Dampens display, output and relay response rates.


Select: Low, Med, High, OFF (see discussion in Appendix.)
Default = OFF. Signet strongly recommends leaving averaging OFF for pH and pressure measurements
(see discussion in Appendix).

9900 Transmitter 39
Level / Volume

LEVEL/VOLUME Setup Checklist


1. Make sure LEVEL/VOLUME sensor type is selected
(see System Setup Menu, page 25).
2. Select Main Measurement (Level or Volume).
3. Set Units of Measurement for LEVEL display (FT, IN, M, CM).
4. If desired, set Units of Measurement for VOLUME display.
5. Set the minimum and maximum 4 to 20 mA setpoints.
6. Set Specific Gravity.
7. Set Sensor Offset.
8. If VOLUME is used, set Shape.
This is the normal display 9. Set Last Cal Date and initials.
and does not time out. 10. If desired, set up relay functions for your own application.

VIEW Mode Menu

Displays the Volume value on the bottom line of the screen when LVL is the MAIN MEAS selection in
INPUT menu.
Displays the Level value on the bottom line of the screen when VOL is the MAIN MEAS selection in
INPUT menu.
Displays the 4 to 20 mA LOOP output
Bottom line shows one of three states (OFF, ON, PLS) for each of the three relays. Displays remaining
time for CYC LOW or CYC HIGH mode. The relay(s) will remain ON while counting down.
NOTE: (Reset the Timer) In the RELAY menu, select TEST RELAY function. The timer will reset to 0 if
the condition no longer exists when the TEST is performed.
The timer will restart if the condition still exists.

CAL Menu

ON prevents relays from activating while making adjustments, and relays in PULSE mode will suspend
pulsing. Output is held until the user exits the CAL menu. Select OFF/ON.
Default = OFF.

Shows SET LEVEL on bottom line. When user presses any key, the live value is frozen and the user
edits that value. Returns either GOOD CAL or LEVEL OFFSET TOO LARGE.

Resets calibration to factory default. After pressing ►, select YES/NO.

Enter date of calibration (mm-dd-yyyy) and initials of calibrator (ii).

40 9900 Transmitter
Level / Volume

INPUT Menu

Enter 13-character string (optional). Default = LEVEL/VOLUME.

Select between Level or Volume.


Default = LVL.

Select unit of measure for LEVEL display (FT, IN, M, CM).


Default = FT

ON = Measurement will be displayed as a percentage of full scale.


OFF = Measurement will be displayed in unit of measure selected in previous setting.
Default = OFF.

If PERCENT LEVEL = ON, set the desired full scale (100%) value in units of measure.
Default = 10.00.

Select unit of measure for VOLUME display (GAL, LIT, Lb, KG, FT3,, in3, M3, cm3).
Default = GAL.

Select ON = Measurement will be displayed as a percentage of full scale. OFF = Measurement will be
displayed in unit of measure selected in previous setting.
Default = OFF.

If PERCENT VOLUME = ON, set the full scale value (100%) in units of measure.
Default = 100.00.

Enter the specific gravity of the fluid at normal operating temperature. This setting is required only if the
level measurement is made by a pressure sensor or if kg or lb volume units are selected.
Default = 1.0000 (water).

Enter the distance from sensor location to the Zero reference point in the vessel (see discussion in
Appendix). Displayed in units of measure chosen in LEVEL UNITS.
Default = 0.

Dampens display, output and relay response rates.


Select Low, Med, High, OFF (see discussion in Appendix). Default = OFF
Select the shape of the vessel where the level sensor is located. VERT CYLINDER, HORIZ
CYLINDER, RECTANGLE, or CUSTOM.
(To define a custom tank shape, see Appendix page 52, Defining a Custom Tank.)
Default = VERT CYLINDER.

If VERT CYLINDER or HORIZ CYLINDER is selected, enter the diameter of the cylinder. Displayed in
units of measure chosen in LEVEL UNITS.
Default = 2.0000

9900 Transmitter 41
Temperature

TEMPERATURE Setup Checklist


1. Make sure TEMPERATURE sensor type is selected
(see System Setup Menu, page 25).
2. If LOOP is used, set the minimum and maximum 4 to 20 mA setpoints.
3. Set Units of Measurement (°C or °F).
4. Set Last Cal Date and initials.
5. If desired, set up relay functions for your own application.

This is the normal display


and does not time out.

VIEW Mode Menu

Displays the 4 to 20 mA LOOP output


Bottom line shows one of three states (OFF, ON, PLS) for each of the three relays.
Displays remaining time for CYC LOW or CYC HIGH mode. The relay(s) will remain ON while counting down.
NOTE: (Reset the Timer) In the RELAY menu, select TEST RELAY function.
The timer will reset to 0 if the condition no longer exists when the TEST is performed.
The timer will restart if the condition still exists.

CAL Menu

ON prevents relays from activating while making adjustments, and relays in PULSE mode will suspend
pulsing. Output is held until the user exits the CAL menu. Select OFF/ON.
Default = OFF.

Provides a maximum 20 °C offset to match to a known standard (external reference).

Resets Temperature Calibration to factory settings. After pressing ►, select YES/NO.

Enter date of calibration (mm-dd-yyyy) and initials of calibrator (ii).

INPUT Menu

Enter string up to 13 characters (optional).


Default = “TEMPERATURE”.

Select °C or °F.
Default = °C.

Dampens display, output and relay response rates.


Select Low, Med, High, OFF. (See discussion in Appendix.)
Default = OFF.

42 9900 Transmitter
4 to 20 mA

4 to 20 mA Setup Checklist
1. Make sure 4-20 mA INPUT sensor type is selected
(see System Setup Menu, page 25).
2. Set 4 mA value (refer to your 3rd-party sensor manual).
3. Set 20 mA value (refer to your 3rd-party sensor manual).
4. If LOOP is used, set the minimum and maximum 4 to 20 mA setpoints.
5. Set Last Cal Date and initials.
6. If desired, set up relay functions for your own application.

This is the normal display


and does not time out.

VIEW Mode Menu

Displays the 4 to 20 mA LOOP output

Diagnostic display showing raw input from 4 to 20 mA sensor.

Bottom line shows one of three states (OFF, ON, PLS) for each of the three relays. Displays remaining
time for CYC LOW or CYC HIGH mode. The relay(s) will remain ON while counting down.
NOTE: (Reset the Timer) In the RELAY menu, select TEST RELAY function. The timer will reset to 0 if
the condition no longer exists when the TEST is performed.
The timer will restart if the condition still exists.

CAL Menu

ON prevents relays from activating while making adjustments, and relays in PULSE mode will suspend
pulsing. Output is held until the user exits the CAL menu. Select OFF/ON.
Default = OFF.

Applies a linear offset to the measurement.


For single point calibrations, assign the average value of your process to STANDARD.
For two-point calibrations, assign the min or max value of your process to STANDARD.
Applies a slope to the measurement. The SLOPE is used for two-point calibrations along with the
STANDARD above. If you assigned the min value of your process to the STANDARD, then assign the
max value to the SLOPE. Else, assign the min value to the SLOPE. The slope and standard values must
be at least 0.1 units apart.

Resets Standard and Slope calibration to factory settings. After pressing ►, select YES/NO.

Enter date of calibration (mm-dd-yyyy) and initials of calibrator (ii).

9900 Transmitter 43
4 to 20 mA

INPUT Menu

Enter string up to 13 characters (optional).


Default = 4-20 mA INPUT.

Enter up to 4 characters describing unit of measure.


Default = UNIT.

Measurement value of your sensor when its output is 4.00 mA.

Measurement value of your sensor when its output is 20.00 mA.

Dampens display, output and relay response rates.


Select Low, Med, High, OFF. (See discussion in Appendix.)
Default = OFF.

To program the 9900 for Dissolved Oxygen measurement using the 3-2610-31 sensor:
From the 4 to 20 mA View Mode display:
1. Press and hold the ENTER key for 2 seconds.
2. Press the ▼ key to select the INPUT menu.
3. The first item is NAME. Press the ► key to change the displayed name from “4-20 mA INPUT” to a more descriptive name
(e.g., DO) and press ENTER when done.
4. Press ▼ to select SENSOR UNIT menu item.
5. Press ► to change the label from UNIT to MG/L and press ENTER.
6. Press ▼ and ensure the 4 mA VALUE is set to 0.0000.
7. Press ▼ and change the 20 mA VALUE from 5.0000 to 20.000 and press ENTER.
8. Press both ▲ and ▼ simultaneously to return to the Menu.
9. Press ▼ to select the LOOP menu and press ENTER.
10. Set the 4 mA SETPOINT to your desired value. The 2610 is factory set for a 0 to 20 mg/L output. Press ENTER when
done.
11. Press the ▼ key to select the 20 mA SETPOINT and set to your desired value. The 2610 is factory set for a 0 to 20 mg/L
output. Press ENTER when done.
12. Press both ▲ and ▼ simultaneously to return to the Menu.
13. Press ▼ twice to select the OPTION menu and press ENTER.
14. Press ▼ twice to select SET BAR MIN. Change this option if desired. The 2610 is factory set for a 0 to 20 mg/L output.
Press ENTER when done.
15. Press ▼ to select SET BAR MAX. Change this option if desired. The 2610 is factory set for a 0 to 20 mg/L output. Press
ENTER when done.
16. Press both ▲ and ▼ simultaneously to return to the Menu.
17. ENTER the other menus and set the unit as desired for your application.
18. Press both ▲ and ▼ simultaneously to return to the View Menu.

44 9900 Transmitter
Salinity

SALINITY Setup Checklist


1. Make sure SALINITY sensor type is selected
(see System Setup Menu, page 25).
2. Set Cell Constant.
3. Set the Temperature Units (°C or °F).
4. If LOOP is used, set the minimum and maximum 4 to 20 mA setpoints.
5. Set Last Cal Date and initials.
6. Select source for Open Collector and Relay output (SAL or TEMP).
7. If desired, set up relay functions for your own application.

This is the normal display


and does not time out.

VIEW Mode Menu

Displays temperature at the sensor.

Displays the 4 to 20 mA LOOP output.

Displays the equivalent conductivity value in milliSiemens.

Bottom line shows one of three states (OFF, ON, PLS) for each of the three relays. Displays remaining
time for CYC LOW or CYC HIGH mode. The relay(s) will remain ON while counting down.
NOTE: (Reset the Timer) In the RELAY menu, select TEST RELAY function. The timer will reset to 0 if
the condition no longer exists when the TEST is performed.
The timer will restart if the condition still exists.

CAL Menu

YES prevents relays from activating while making adjustments, and relays in PULSE mode will suspend
pulsing. Output is held until the user exits the CAL menu. Select YES/NO.
Default = NO.

Manually set salinity value to match to a known standard (external reference).

Provides a maximum 20 °C offset to match to a known standard (external reference).

Resets Salinity calibration to factory settings. After pressing ►, select YES/NO.

Resets Temperature calibration to factory settings. After pressing ►, select YES/NO.

Enter date of calibration (mm-dd-yyyy) and initials of calibrator (II).

9900 Transmitter 45
Salinity

INPUT Menu

Enter string up to 13 characters (optional).


Default = SALINITY.

Enter cell constant of sensor. Select 20.0, 10.0, 1.0 or CUSTOM.


Default = 20.

Enter the precise cell constant from the certificate provided with your sensor, or from the information
label on the sensor. Shown if CELL CONSTANT = CUSTOM.

Select °C or °F.
Default = °C.

Dampens display, output and relay response rates.


Select Low, Med, High, OFF. (See discussion in Appendix.)
Default = OFF.

Select temperature compensation (NONE, LINEAR).


Default = LINEAR.

For LINEAR temperature compensation, select a % per ºC slope. Maximum slope setting is 9.99 % per
ºC. (If Temperature Compensation setting is NONE, this item will not be displayed.)

46 9900 Transmitter
Dissolved Oxygen

DISSOLVED O2 Setup Checklist (3-2610-41)


2610 wiring on page 15.
3-2610-31 setup on page 44.
1. Make sure DISSOLVED O2 sensor type is selected
(see System Setup Menu, page 25).
2. Set Units of Measurement (PPM, %SAT, TOR).
3. Set the Temperature Units (°C or °F).
4. Set Salinity reference value.
5. Set Barometric reference value.
6. If LOOP is used, set the minimum and maximum 4 to 20 mA setpoints.
This is the normal display 7. Select source for Open Collector and Relay output (PPM or TEMP).
and does not time out. 8. If desired, set up relay functions for your own application.

VIEW Mode Menu

Displays temperature at the sensor.

Displays the 4 to 20 mA LOOP output.

Displays Cap Expiration Date MM-YYYY. If sensor cap is missing, - - - - - - - will be displayed.

Bottom line shows one of three states (OFF, ON, PLS) for each of the three relays. Displays remaining
time for CYC LOW or CYC HIGH mode. The relay(s) will remain ON while counting down.
NOTE: (Reset the Timer) In the RELAY menu, select TEST RELAY function. The timer will reset to 0 if
the condition no longer exists when the TEST is performed.
The timer will restart if the condition still exists.

CAL Menu

YES prevents relays from activating while making adjustments, and relays in PULSE mode will suspend
pulsing. Output is held until the user exits the CAL menu. Select YES/NO.
Default = NO.

Allows user to initiate the optional calibration process.


NOTE: Dissolved Oxygen sensors are calibrated at the factory and do not require regular calibration.
Press ► to begin the calibration process (password required). User will be prompted to place sensor in
100% Solution standard. Press ENTER to save value and establish a calibration point.
Allows user to establish an optional second calibration point.
NOTE: Dissolved Oxygen sensors are calibrated at the factory and do not require regular calibration.
Press ► to begin the calibration process. User will be prompted to place sensor in 0% Solution
standard. Press ENTER to save value and establish an optional second calibration point.
This option is only available immediately after a successful 100% Solution calibration.

Resets Dissolved Oxygen calibration to factory settings. After pressing ►, select YES/NO.

Enter date of calibration (mm-dd-yyyy) and initials of calibrator (II).

9900 Transmitter 47
Dissolved Oxygen

INPUT Menu

Enter string up to 13 characters (optional).


Default = DISSOLVED O2.

Set the units of measurement:


PPM = DO in mg/L; %SAT = DO % saturation; TOR = Oxygen partial pressure. Default = PPM.

Manually set Salinity value to match application Salinity (0 - 42 PSU). Units in Practical Salinity Unit (PSU).
Fresh water = 0.00 PSU. Default = 0.00

mBAR
Manually set Barometric value to match application altitude above or below sea level (506.62 - 1114.7 mBAR).
1013.2
Default = 1013.2 (sea level)
BAROMETRIC

Select °C or °F.
Default = °C.

Dampens display, output and relay response rates.


Select Low, Med, High, OFF. (See discussion in Appendix.)
Default = OFF.

48 9900 Transmitter
Troubleshooting

Condition Possible Causes Suggested Solution

Incorrect sensor installed Connect correct sensor


Wrong Sensor
Set correct sensor TYPE in INPUT menu
Sensor Type set incorrectly in 9900
(see page 25)

Wrong Code Wrong password entered Enter correct password (see page 26)

K-Factor Out Of Range K-Factors cannot be set to 0 Enter K-Factor from 0.0001 to 99999

9900 operating on loop power Connect 9900 to 10.8 to 35.2 VDC power.

Backlight inoperative Backlight turned OFF


Set BACKLIGHT to LOW, HIGH or AUTO
(NOTE: Backlight can turn off
in OPTION menu.
automatically in AUTO mode)

9900 operating on loop power Connect 9900 to 10.8 to 35.2 VDC power.

Relays 2 and 3 Relay Module installed incorrectly Remove and reseat relay module
inoperative
Use test relay to verify relay operation then
Wrong settings in RELAY menu
check relay settings.

Use DC power. Check relay states in VIEW


Relay LEDs inoperative 9900 operating in Loop Power
mode for status.

Open Collector (R1) or Hysteresis value too large Change the hysteresis value
Relay (R2 or R3)
always on Defective Relay Module Replace Relay Module

Relay pulse rate exceeds maximum of Increase volume pulse setting


300 pulses per minute. Reduce system flow rate
OVR relay state
(Pulse Overrun) Pulse width set too wide Decrease pulse width

(NOTE: Max pulse rate = 300; max pulse width = 100 mS.

Increase Flow units time base


– – – – – Flow rate exceeds display capability
Change unit of measure

9900 Transmitter 49
Troubleshooting

Condition Possible Causes Suggested Solution

9900 cannot "talk" to sensor


Check Sensor • Check wiring
(pH/ORP only) • Install or replace sensor
Missing sensor or bad temperature element.

No Sensor
• Check wiring
(Flow, Cond/Res, Press, Level, 9900 cannot "talk" to sensor
Temp, 4-20 mA, • Install or replace sensor
Sal, Batch, DO)

Check Preamp 9900 cannot "talk" to the preamp Check wiring or replace preamp

Warning LED lit Look for error message Correct error condition

Dissolved Oxygen sensor is missing the


Missing Cap Reinstall Dissolved Oxygen sensor cap
sensor cap.

Replace Cap Dissolved Oxygen sensor cap has expired. Install new Dissolved Oxygen sensor cap

Visually inspect pH sensor for cracked and/or chipped


glass. After inspection or replacement of pH Sensor wait
pH sensor glass has been damaged,
Broken Glass (2751 Only) 30 seconds and then manually start a glass impedance
causing very low impedance.
measurement, page 33. This will clear the Broken Glass
alarm.

The measured pH sensor impedance is Visually inspect the pH electrode and


above the high impedance level. clean if necessary.
Hi Impedance (2751 Only)
Electrode could be in air. Ensure electrode is submersed at all times.

Perform pH EasyCal (pg. 34 & 56)

Perform ORP EasyCal (pg. 36 & 58)

Check Cal Slope and/or Offset are out of range Set pH Slope or Standard (pg. 34 & 57)
(possibly due to memory failure in sensor or
(2751 pH/ORP only) preamp) Set ORP Slope or Standard (pg. 36 & 59)

Reset pH CAL (pg. 34)

Reset ORP CAL (pg. 36)

Older GF Signet electrodes, such as the 2754, or


electrodes with BNC type connectors do not have
memory chips. When the 9900 detects a non-memory
chip electrode, the Sensor Data display will not be
shown.
If a memory chip enabled sensor is installed, a
Sensor Data Screen missing pH/ORP Sensor installed does not have a communication error between the memory chip and
(2751 Only) memory chip. or communication error 2751, or 2751 and 9900, may have occurred. To clear
the error and restore the Data Screen, you may do one
of the following:
• Disconnect the (S3L) connection between the 9900
and the 2751, wait 5 seconds, reconnect the (S3L)
connection between the 9900 and the 2751.
• Cycle power to the 9900.

50 9900 Transmitter
Appendix

Averaging
NO AVERAGING, NO SENSITIVITY
With SENSITIVITY set to a large value and AVERAGING set
to OFF (0 seconds), the 9900 responds immediately to every
shift in the process. The dashed red line represents the actual
output of the sensor in varying conditions.

AVERAGING ONLY
With SENSITIVITY still set to a large value and AVERAGING
set to MED or HIGH the rate is stabilized, but a sharp change
in rate is not represented for 8 to 32 seconds or longer.
AVERAGING AND SENSITIVITY
With SENSITIVITY at 50 and AVERAGING set to MED or
HIGH, the rate is stabilized, while a sudden shift in flow rate 0s 5s 10 s 15 s 20 s 25 s 30 s 35 s
exceeding 50 units of measure will be displayed immediately.
NOTE: The SENSITIVITY function applies only to FLOW.
The SENSITIVITY function has no effect if the AVERAGING Averaging is different depending on the measurement type.
function is set to OFF. Averaging will not function if the Seconds to 99.5% of Final Value for Low, Med, and High are:
Sensitivity is set to 0 (Zero).
Sensor Type Low Medium High
Flow 10 40 120
Important
pH 2 4 12
With Averaging set to Off the Sensitivity function is
disabled. With Sensitivity set to 0 (zero) the Averaging ORP 2 4 12
function is disabled. Cond/Res 4 6 12
To set the unit to Averaging with no Sensitivity set
Pressure 4 10 30
Sensitivity to a large value, 99999.
Level/Volume 4 10 30
Temperature 3 10 30
4 to 20 mA 4 10 30
Salinity 4 6 12

LOG Current Loop Output

In Conductivity/Resistivity, the logarithmic (LOG) mode can be used when a very large measurement range is required,
yet high resolution is needed at the low end (e.g. in a clean-in-place application where a high-resolution conductivity
reading is needed at the low end, while a very high conductivity reading is needed when a cleaning cycle is in progress).
Only two parameters need to be set up, the starting or base conductivity value (4 mA SETPNT) and the ending or
maximum conductivity value (20 mA SETPNT). The 4 mA setpoint may be larger than the 20 mA point (reverse span).

What equation should be put in the PLC?

Conductivity = 10n
(Log10 20 mA setpnt – Log10 4 mA setpnt)
n = (mA input – 4.0) × + Log10 4 mA setpnt
16 mA

If only fixed thresholds are of required, they can be calculated in mA. Then the mA value can be checked directly.
Inside the 9900 the following equation is used:
16
mA = (Log10 Conductivity – Log10 4 mA setpnt) × +4
(Log10 20 mA setpnt – Log10 4 mA setpnt)
NOTE:
If ADJUST 4 mA or ADJUST 20 mA is used, the mA value can be affected. To prevent any problems the adjust function
should only be used to get exactly 4.0 and 20.0 at the PLC. The 9900 is accurate and the adjust functions are only needed
to compensate for an offset due to noise or a not-so-accurate PLC input card.
The error value of either 3.6 mA or 22 mA should be tested first before applying the conductivity equation.

9900 Transmitter 51
Custom Measurements

Defining a Custom Tank


1. Determine where the level measurement
should start. This is the zero reference point (Z).
Review the diagram to help select the best option.
2. Determine where you will mount the sensor.
This is SLOC. Consult the Sensor manual for
information regarding the best location for
the sensor.
3. Measure the distance between Z and SLOC.
This is O(ffset).
Z 4. Enter the Offset into the INPUT Mode menu.

Zero reference point (Z):


Z
The point in the vessel where you want the 9900 to
display zero (0 ft, 0 gal. etc.).
• If Z is located below the fluid surface, the 9900
For most vessels, the zero reference point (Z) may be will display a positive level measurement.
designated as any height in the vessel.
• If Z is located above the fluid surface, the 9900
For horizontal cylinders only, the zero reference point
will display a negative level measurement.
MUST be the lowest point in the vessel.

Sensor Location point (SLOC):


The point on the level sensor where the
measurement is taken.
• The pressure sensor measures from the
centerline of the diaphragm.
Z(ero)

L<0 Offset (O):


O(ffset) < 0
The distance from Z to SLOC.
L>0
L>0 • Enter a positive value in the Calibrate menu
if the sensor is located above Z.
SLoc

O(ffset) > 0
Z(ero) SLoc • Enter a negative value in the Calibrate menu
O(ffset) < 0
SLoc if the sensor is located below Z.
Z(ero)
• Enter 0 in the Calibrate menu if the sensor is
located at Z.

Level (L):
The distance from Z to surface of fluid
(displayed as “Level” by 9900).

52 9900 Transmitter
Custom Measurements

Level and Volume Calculation in Custom Shaped Vessels


In the LEVEL/VOLUME menu, if Custom Shape is selected in the INPUT menu, you can define
from three to ten Custom Points to establish the relationship of level to volume in the vessel.
● Select Manual Level Measurement mode to edit both level and volume data (dry configuration).
● Select Automatic Level Measurement mode to accept the sensor measurement of the Level,
while you assign a volumetric value to each custom point (wet configuration).
● Enter from 3 to 10 custom points to link level and volume values.
● The first custom point must be the lowest fluid level in the vessel. Each successive point must
be greater than the preceding point.
● The last point must be equal to or greater than the highest fluid level in the vessel.
● A custom point should be located at all transition points in the vessel shape (for example, at
custom point #9, where the shape changes from a cylinder to a cone).
● The more complex sections should be defined with more points.
NOTE: The conical section of the illustration has been defined by custom points 1 through 9.
● Simpler sections require fewer defining points.
NOTE: A cylinder requires only custom points 9 and 10.

Custom point #10 3000 gal.

Custom point #9 1300 gal.

Custom point #8 1000 gal.

Custom point #7 700 gal.

Custom point #6 380 gal.

Custom point #5 250 gal.

Custom point #4 160 gal.

Custom point #3 118 gal.

Custom point #2 45 gal.

Custom point #1 0 gal.

9900 Transmitter 53
Custom Measurements

In the LEVEL/VOLUME INPUT menu (see page 41), if SHAPE is set to HORIZ CYLINDER, RECTANGLE or CUSTOM,
the tank shape can be defined with the following screens:

If Horiz Cylinder or Rectangle shape is selected, enter the length of the vessel in
LEVEL UNITS. 0.0000 to 99999.
If Rectangle shape is selected, enter the width of the vessel in LEVEL UNITS.
0.0000 to 99999.
If Custom shape is selected, enter the number of measurement points to be used to define
the vessel shape (see Level and Volume Calculation in Custom Shaped Vessels discussion).
Minimum 3 points, maximum 10 points. A larger number of points improves accuracy.
Select (AUTO, MAN). Manual allows you to edit both the Level and the corresponding
Volume for your custom tank. Automatic allows you to edit the Volume measurement
(while displaying an automatically calculated Level value). See example below.
Enter the Level (if MAN measurement is selected) at each custom point in your vessel.
If AUTO is selected, indication will read actual tank level in LEVEL UNITS at that point
in your tank.
Set the Volume (if manual measurement is selected) at each custom point in your vessel.
Where (X) is number of custom points
Where (X) is number of custom points

To set AUTO LEVEL MEAS value:


1. Pour a known quantity of fluid into a tank.
2. POINT 1 LEVEL indicates actual tank level.
3. Press ▼ for POINT 1 VOL. Press ► to enter quantity of fluid (in VOLUME UNITS)
you poured into the tank in step 1. Press ENTER.
4. Repeat for each point set in NUM CUST PNTS.

For example, in a 25-gallon conical tank set for three custom points:

1. Pour 10 gallons of fluid into the conical tank. POINT 1 LEVEL will indicate actual tank level.
2. In POINT 1 VOL, enter 10.
3. Pour another 10 gallons into the tank. POINT 2 LEVEL will indicate actual tank level.
4. In POINT 2 VOL, enter 10.
5. Pour the final 5 gallons into the tank. POINT 3 LEVEL will indicate actual tank level.
6. In POINT 3 VOL, enter 5.

54 9900 Transmitter
Custom Measurements

Technical Reference for


Level, Volume, and Mass Measurement Pressure to level conversion:
Level = P ÷ (SG × D)
The 9900 can automatically perform where P = Pressure
level, volume and mass calculations: SG = Specific Gravity of fluid
D = Density of water
● Pressure-to-level
● Mass With pressure in psi:
● Volume Level (meters) = 0.703069 × (P/SG)
With pressure in bar:
Level (meters) = 1.019715 × (P/SG)

Mass Conversion
m = D × SG × V

where m = mass of fluid


D = density of water = 1000 kg/m3
SG = Specific Gravity of fluid
V = Volume of fluid (m3)

m (kg) = 1000 × SG × V

Volume Calculations

Vertical cylinder:
V = π × r2 × h
r where r = radius of cylinder
h = height of fluid

A h Rectangular vessel:
V=w×l×h
where w = width
l = length
h = height

Horizontal cylinder:
V=A×L
where A = area of segment
L = length of vessel

⎡⎛ ⎛ r −h⎞ ⎞ 2 ⎤
A = ⎢⎜⎜ ⎜ r 2 × cos −1 × ⎟ − (r − h )⎟⎟ × 2rh − h ⎥
⎣⎝ ⎝ r ⎠ ⎠ ⎦
Length
where r = radius of vessel
h = height of segment

9900 Transmitter 55
Calibration Procedures - pH

EasyCal Procedure - pH
EasyCal is the fastest and simplest periodic calibration method.
Requires prepared 4, 7 or 10 pH buffers (any two).

To Calibrate: Response: To Accept: • This procedure simplifies


pH calibration using
standard 4.0, 7.0, 10.0 pH
buffers only. If these pH
buffers are not available,
use MANUAL CAL and
calibrate the system using
the STANDARD and
ENTER to accept SLOPE settings.
Place electrode tip in 30s
first pH buffer. • Set sensor temperature
Allow for stabilization
pH 4.0 = 177 mV 30 seconds in the CAL Mode before
pH 7.0 = 0 mV performing EasyCal for new
pH 10 = -177 mV
Limit ± 50 mV 1 electrode installations.

Theoretical mV values
to accept second
pH @ 25 °C mV
Place electrode tip in
30s ENTER
buffer calibration 2 ................... +296
second pH buffer. Allow for stabilization 3 ................... +237
30 seconds
4 ................... +177
2
5 ....................+118
6 ..................... +59
7 ....................... +0
To exit menus and return to + 8 ......................–59
9,(:SUHVVŸDQGź Display returns to VIEW
buttons at the same time. Mode in 10 minutes 9 ....................–118
10 ..................–177
11 ..................–237
12 ..................–296

NOTE: The solutions can be used for calibrating more than one sensor;
however, the solution must remain free of debris and must not be diluted by
rinse water from previous calibrations.

56 9900 Transmitter
Calibration Procedures - pH

Manual Calibration Procedure - pH


Requires prepared buffers. System calibration is possible with two known pH
solutions within 0 to 14 pH (buffers of pH 4.01, 7, or 10 are recommended,
but use a buffer close to your own process value.)

To Calibrate: To Change Reading: Single-point calibration sets


STANDARD only; Signet
recommends a two-point
calibration to set SLOPE in
addition to STANDARD.

Quick Manual
Calibration Procedures:
ENTER
to accept
1-Point Calibration:
30s ENTER
1. Set solution standard.
Allow for stabilization
30 seconds to 2-Point Calibration
several minutes
to accept (recommended):
Place electrode tip in
pH buffer. 1. Set solution standard.
2. Set solution slope.

To Set Slope: To Change Reading:

to accept ENTER

30s ENTER

Allow for stabilization


30 seconds to
several minutes
to accept
Place electrode tip in
pH buffer two pH units
different than standard.

To Set Calibration Date:

+
Display returns to
VIEW Mode

9900 Transmitter 57
Calibration Procedures - ORP

EasyCal Procedure - ORP (one-point calibration)


EasyCal is the fastest and simplest periodic calibration method.
Requires a prepared quinhydrone solution or Light's Solution:
Saturate 50 mL of pH 7 (87 mV) or pH 4 (264 mV) buffers with 1/8 g quinhydrone.
Premixed Light's Solution (476 mV) can be used instead of pH buffers with quinhydrone.

1. Go to CAL menu.
To Calibrate: Response: To Accept:
2. Press ▼ twice to display
EASY CAL setting.
3. Press ► to begin
one-point EasyCal.
4. Place sensor in solution:
ENTER to accept
• 87 mV
30s (7 pH + quinhydrone)
Place electrode tip in
saturated pH 7.0 buffer. Allow for stabilization • 264 mV
pH 7.0 = 87 mV 30 seconds (4 pH + quinhydrone)
• 476mV
1 (Light's Solution)
5. Press ENTER.
+ 6. After 30 seconds,
To exit menus and return to
9,(:SUHVVŸDQGź
Display returns to VIEW 9900 will recognize
buttons at the same time. Mode in 10 minutes current buffer ±80 mV.
7. Press ENTER to accept
mV value.

NOTE: ORP solutions made with quinhydrone are very unstable and
may not read properly once exposed to air for a prolonged time.
These solutions must be disposed of within an hour.
The solution can be used for calibrating more than one sensor.
However, the solution must remain free of debris and must not be
diluted by rinse water from previous calibrations.

Acceptable ranges for the readings are ± 80 mV (i.e., 87 ± 80 mV).

58 9900 Transmitter
Calibration Procedures - ORP

Manual Calibration Procedure - ORP


Requires prepared buffers and a prepared quinhydrone solution:
Saturate 50 mL of pH 4 (264 mV) and 7 (87 mV) buffers with 1/8 g quinhydrone or
Light's Solution (476 mV).
(System calibration is possible with two known ORP solutions,
but use a buffer close to your own process value).

To Calibrate: To Change Reading: Single-point calibration


sets STANDARD only;
Signet recommends a
two-point calibration to
set SLOPE in addition
to STANDARD.

Quick Manual Calibration


to accept
ENTER Procedures:

30s ENTER 1-Point Calibration:


Allow for stabilization
30 seconds to
1. Set solution standard.
several minutes to accept
Place electrode tip in 2-Point Calibration
saturated pH 7.0 buffer. (recommended):
pH 7.0 = 87 mV
1. Set solution standard.
2. Set solution slope.

To Set Slope: To Change Reading:

to accept ENTER

30s ENTER

Allow for stabilization


30 seconds to
several minutes to accept
Place electrode tip in
saturated pH buffer two
pH units different than
standard.
pH 4.0 = 264 mV

To Set Calibration Date:

+
Display returns to
VIEW Mode

NOTE: ORP solutions made with quinhydrone are very unstable and
may not read properly once exposed to air for a prolonged time.
These solutions must be disposed of within an hour.
The solution can be used for calibrating more than one sensor.
However, the solution must remain free of debris and must not be
diluted by rinse water from previous calibrations.
Acceptable ranges for the readings are ± 80 mV (i.e., 87 ± 80 mV).

9900 Transmitter 59
Calibration Procedures - Conductivity / Resistivity

Calibration Procedure - Conductivity / Resistivity


AutoCal is the fastest and simplest periodic calibration method.
Requires prepared buffer of a value appropriate to your process.

AutoCal Procedure Conductivity units are


displayed as selected in
AutoCal is a single-point calibration system. During this procedure, if the measured
the CALIBRATE menu.
value is within ± 10% of any of the test values listed below, the 9900 will automatically
Resistivity displayed
recognize the test value and calibrate the output to that value.
when KΩ or MΩ ranges
are selected.
NOTE: The first step (Reset) is recommended each time an electrode is replaced, but is
NOT necessary upon initial installation or periodic calibration.

NOTE: Ensure that the buffer solution is within ± 5 °C of 25 °C. Available buffer values are:
● 10
1. Reset the sensor to factory calibration (refer to sensor manual for procedure). ● 100
2. On the 9900, select AUTO CAL from the CAL menu. Press ►. ● 146.93
3. Place the electrode/sensor assembly into the conductivity test solution appropriate ● 200
to your operating range. Shake the electrode to dislodge any air bubbles visible on ● 500
the surface of the electrode. ● 1000
4. Allow at least 2 minutes for the electrode response to stabilize. ● 1408.8
5. When the display stabilizes, press ENTER. ● 5000
6. If calibration is successful, 9900 will display "SAVING". If error is too large, ● 10,000
"OUT OF RANGE USE MANUAL CALIBRATION" will display. ● 12856
● 50,000
Calibration is complete. Return the system to service. ● 100,000
(all values in μS)

Manual Cal Procedure

NOTE: The first step (Reset) is recommended each time an electrode is replaced,
but is NOT necessary upon initial installation or periodic calibration.

NOTE: Ensure that the buffer solution is within ± 5 °C of 25 °C.

1. Reset the sensor to factory calibration (refer to sensor manual for procedure).
2. On the 9900, select MANUAL CAL from the CAL menu. Press ►.
3. Place the electrode/sensor assembly into the conductivity test solution appropriate
to your operating range. Shake the electrode to dislodge any air bubbles visible on
the surface of the electrode.
4. Allow at least 2 minutes for the electrode response to stabilize.
5. When the display stabilizes, enter the value of the buffer solution using the ▼, ▲
and ► buttons.
6. Press ENTER.
7. 9900 will display "SAVING". If error is too large, "ERR TOO LARGE TO
CALIBRATE" will display.

Calibration is complete. Return the system to service.

60 9900 Transmitter
Calibration Procedures - Flow

Calibration Procedure - Flow


Select RATE CALIBRATION to match the dynamic flow rate to an external reference.
Entering a rate will modify the existing K-Factor.

Select VOLUME CALIBRATION if the flow rate can be determined by filling a vessel
of known volume. The 9900 will count the number of pulses generated as the known
volume of fluid passes through the sensor, and then use the information to calculate
a new K-Factor.

Rate Calibration Procedure


1. Use ▲, ▼ and ► to set the flow rate in the flashing display to match the reference meter.
Press ENTER when completed.
2. The 9900 displays the newly calculated K-Factor for your reference.
(If the calculated K-Factor is less than 0.0001 or greater than 999999 (out of range at
either extreme), the 9900 displays "ERROR NEW KF OUT OF RANGE" and returns to
RATE CAL. If flow is too low to accurately calibrate, the 9900 displays "ERROR FLOW
RATE TOO LOW and returns to RATE CAL.
3. Press ENTER to accept the new K-Factor (9900 displays "SAVING") or press ▲+▼
keys simultaneously to escape without saving and return to Enter Volume.
NOTE: You may enter your own calculated K-Factor in the INPUT menu.

Volume Calibration Procedure


1. Press ENTER to start the volumetric calibration period.
The 9900 starts counting pulses from the flow sensor.
2. Press ENTER to stop the volumetric calibration period.
The 9900 stops counting pulses from the flow sensor.
3. Enter the volume of fluid known to have flowed past the sensor during the volumetric
calibration period. This will modify the existing Flow K-Factor.
4. The 9900 displays the newly calculated K-Factor for your reference.
(If the calculated K-Factor is less than 0.0001 or greater than 999999
(out of range at either extreme), the 9900 displays "ERROR VOLUME TOO HIGH"
(or LOW) and returns to VOLUME CAL.)
5. Press ENTER to accept the new K-Factor (9900 displays "SAVING") or press ▲+▼
keys simultaneously to escape without saving and return to Enter Volume.
NOTE: You may enter your own calculated K-Factor in the INPUT menu.

9900 Transmitter 61
Calibration Error Messages

Message Cause Solution


(Cond/Res) Error > 10% in AutoCal Use manual calibration method
(pH) Buffer not found; Use 4, 7, 10 pH buffers (with quinhydrone for
Out Of Range Use Manual
Error > ±1.5 pH units ORP calibration)
Calibration
(ORP) No quinhydrone in buffer Clean sensor and retry EasyCal
Error greater than ±80 mV Use manual calibration method
(Cond/Res) Manual cal when error > Inspect sensor and wiring for damage
100% Clean sensor
(pH) Offset > 1.3 pH units;
Err Too Large To Calibrate Slope error > 100%
Check reference
(Press) Slope must be < ±50% or offset Clean sensor
must be < 2.75 PSI or equivalent. Replace sensor
(Sal) Slope error > 1000%
User-entered volume too small to Correct volume entry
Error Volume Too Low
calibrate Use longer calibration period
Verify volume or rate entered
Error New KF Out Of Range The calculated K-Factor too low or high
Verify flow is present
(Rate Cal) Flow too low to accurately
Error Flow Rate Too Low Increase flow
calibrate
(4 to 20 mA) Slope error > 1000% Check input at 4 mA and 20 mA settings
Cal Error Out Of Range Check sensor range
(Temp) Offset must be < ±20 °C or
Check reference
equivalent.
Replace sensor
(4 to 20 mA) Difference in calibration
values must be > 0.1 units
Check sensor
(pH) Difference in calibration values Use fresh buffer
Slope Too Close To Standard
must be > 2 pH units Use two different buffer values
Clean sensor
(ORP) Difference in calibration values
must be > 30 mV
(4 to 20 mA) Difference in calibration
values must be > 0.1 units
Clean sensor
(pH) Difference in calibration values
Standard Too Close To Slope Use fresh 4, 7, 10 pH buffers
must be > 2 pH units
Use two different buffer values
(ORP) Difference in calibration values
must be > 30 mV
Decrease offset
Level Offset Too Large Offset must be < 1.0 meter
Replace sensor
Pressure must be lower than 2.5 PSI
Pressure Too High Decrease pressure
or equivalent to do zero cal.
Pressure must be higher than 3 PSI or Increase pressure
Pressure Too Close To Zero
equivalent to do slope calibration. Check reference
Perform pH EasyCal (pg. 34 & 56)
Perform ORP EasyCal (pg. 36 & 58)
Check Cal Slope and/or Offset are out of range Set pH Slope or Standard (pg. 34 & 57)
(possibly due to memory failure in
(2751 pH/ORP only) sensor or preamp) Set ORP Slope or Standard (pg. 36 & 59)
Reset pH CAL (pg. 34)
Reset ORP CAL (pg. 36)

62 9900 Transmitter
USP Limits

USP (United States Pharmacopoeia) has defined a set of conductivity values (limits)
to be used for pharmaceutical water. The standard requires that conductivity
measurement without temperature compensation be used for these applications. Temperature USP limit
Range (ºC) (μS)
The limits vary according to the temperature of the sample. The 9900 has the USP
limits stored in memory. It will automatically determine the proper USP limit based 0 to < 5 0.6
on the measured temperature. 5 to < 10 0.8
10 to < 15 0.9
Using the USP function
15 to < 20 1.0
USP setpoints are defined as a percentage below the USP limit, so a USP alarm
is always a HIGH alarm. The 9900 can be set to warn you if the conductivity 20 to < 25 1.1
approaches within a set percentage of the USP limit. 25 to < 30 1.3
30 to < 35 1.4
The following settings and conditions are required for a USP relay function: 35 to < 40 1.5
40 to < 45 1.7
1. In the RELAY menu:
45 to < 50 1.8
• RELAY MODE must be set to USP.
50 to < 55 1.9
2. In the INPUT menu: 55 to < 60 2.1
• COND UNITS must be set to μS. 60 to < 65 2.2
• TEMP COMP must be set to None. 65 to < 70 2.4
70 to < 75 2.5
Example:
75 to < 80 2.7
• The water temperature is 19 ºC, so the USP limit is 1.0 μS.
80 to < 85 2.7
• The USP PERCNT is set to 40%. 85 to < 90 2.7
• The relay will be activated when the conductivity value reaches 40% below the 90 to < 95 2.7
1.0 USP limit, or 0.6 μS. 95 to < 100 2.9
• If the water temperature drifts to more than 20 ºC, the 9900 will automatically 100 to < 105 3.1
adjust the USP limit to 1.1. The relay will now be activated when the
conductivity value reaches 40% below 1.1 μS (0.66 μS).

9900 Transmitter 63
H COMM Module Overview

How HART® Works


The HART® (Highway Addressable Remote Transducer)
Protocol uses Frequency Shift Keying (FSK) to 20 mA - Digital
superimpose digital signals on top of the analog Signal

4 to 20 mA current loop. This allows two-way digital


communication to occur and allows additional information
beyond the normal process data to be communicated
to the 9900. This digital signal can contain data such as
device status, diagnostics, etc.
Analog
The HART protocol provides two simultaneous Signal
4 mA -
communication channels: a 4 to 20 mA analog signal
and a digital signal.The analog signal communicates the Time
primary measured value using the 4 to 20 mA current loop.
Time
Note: Drawing not to scale
Additional information is communicated using a digital
signal superimposed on the 4 to 20 mA signal. Frequency Shift Keying
Digital over Analog

Communication occurs between two HART-enabled devices, in this application a Signet 9900 Transmitter and a PLC
or handheld device, using standard wiring and termination practices. The HART Protocol communicates at 1200 bits
per second without interfering with the 4 to 20 mA signal and allows the PLC or handheld device to communicate two or
more updates per second to and from the 9900.

The HART protocol operates according to the master-slave method. Any communication activity is initiated by the
master, usually a programmable logic controller (PLC) or a data acquisition system. HART accepts two masters:
the primary master - usually the control system (PLC) - and the secondary master - a PC laptop or handheld
terminal used in the field.

HART field devices - the slaves - never send without being requested to do so. They respond only when they
have received a command message from the master. Once a transaction (i.e., a data exchange between the
control station and the field device) is complete, the master will pause for a fixed time period before sending
another command, allowing the other master to break in. The two masters observe a fixed time frame when
taking turns communicating with the slave devices.

As deployed in the 9900 application, HART allows remote verifying, testing, adjusting and monitoring of primary
and secondary device variables. Features available in the 9900 Transmitter with H COMM Module installed:
• Adjust 4 mA: Allows fine-tuning to compensate for errors in other equipment connected to the 9900.
Adjust the minimum and maximum current output.
• Adjust 20 mA: Allows fine-tuning to compensate for errors in other equipment connected to the 9900.
Adjust the minimum and maximum current output.
• Supports Multi-Drop Mode: Allows up to four 9900 Transmitters be installed in Multidrop mode.
• Supports all Universal HART Protocol Revision 7.2 commands
• Supports many Common Practice Commands
• Makes Primary and Secondary values available at PLC. Secondary values are sensor-dependent and
are available with Flow, pH, Conductivity, Resistivity, Salinity and Level sensors.

HART® is a registered trademark of the HART Communication Foundation, Austin, Texas, USA.
Any use of the term HART hereafter in this document implies the registered trademark.

64 9900 Transmitter
H COMM Module Installation

If the 9900 Base Unit will be mounted in a panel, plug-in modules ower
DC P
may be installed either before or after the base unit is mounted.
ge
If the 9900 Base Unit will be mounted using the accessory wall Loop
Volta

mount kit (3-9900.392), install plug-in modules first.


If the Direct Conductivity/Resistivity Module will be included in
your unit, install the H COMM module first and then install
Conductivity/Resistivity Module over the H COMM Module.
To install the H COMM module, carefully align the module pins
into its plug (see illustration) and push the module straight in until
the tabs on the bottom edge snap into place.
0.395
3-990 odule
To uninstall, squeeze tabs, grasp the module and pull straight out. H CO
MM M

Be careful not to bend the pins when installing or


removing the module to or from the base unit.

NOTE: The black rubber jumper adjacent to the power terminal should only be removed when both
the H COMM Module is utilized and the required sensor cable length is over 304 m (1000 ft).

H COMM Module Wiring

Connecting HART with a Loop-powered sensor

9900 Device
Connections
(Typical installation)

(Primary HART Master)


Black –
Loop + PLC
Device
+
Loop –
Red

Optional ammeter Power


(recommended for – Supply +
4 to 20 mA trim only)

Hand-held device
(Secondary HART Master)

Connecting HART to a Hand-Held Master Device


9900 Device
Connections
(Typical installation)

Loop +
Red + Power
Loop – Supply
Black 250 Ÿ

Load Resistor

Hand-held device
(HART Master)

9900 Transmitter 65
H COMM Module Wiring

NOTE: On pages 64-69, the term “9900 Transmitter” or “Transmitter” will assume the H COMM Module is installed
unless otherwise noted.

Multi-Drop Mode

Up to four 9900 Transmitters can be connected in Multi-Drop mode using the H COMM Module.

To ensure proper operation in Multi-Drop mode, configure each 9900 Transmitter with its own poll address using a
configuration tool (laptop or hand-held device).

After configuring the 9900 Transmitter for multi-drop function, reset the Transmitter (remove power for five seconds
then apply power) before use.

Connecting HART in Multi-Drop mode

9900 (HART Master)


Connections
PLC
Device
Device 1
Red
+ –
Loop +
Loop –
Black
(Typical four-unit installation)

Red

Device 2 Black
Red
Loop +
Loop –
Black
Hand-held device
(Secondary HART Master)

Device 3
Red
Loop +
Loop –
Black

Device 4
Red
Loop +
Loop –
Black

66 9900 Transmitter
H COMM Module Operation

Loop Powered systems require a minimum of 24 VDC. If connecting with DC, nominal 12 VDC is acceptable.
(See Power Wiring, pg. 17)
• In LVL/VOL mode, the primary variable will always represent Level, the secondary variable will always
represent Volume.
• In pH, Conductivity, Resistivity and Salinity systems the secondary variable represents Temperature.
• In Flow, the secondary variable represents the Totalizer. Select Permanent or Resettable Totalizer in the
Input menu (pg. 32).

Loop Current Trim Procedure

The Loop Current HART commands allow a Master HART device to update a loop current value in the 9900
Transmitter and to perform a two-point calibration (zero and span) of the loop current.
1. Use Command 40 (Enter/Exit Fixed Current Mode) to update the 4.00 mA current.
2. Using the measured value of your reference instrument (either a digital multimeter or the HART Master device),
set the zero trim using Command 45 (Trim Loop Current Zero). The transmitter will then trim its calibration and
return the loop current value in the response message. The response value may differ slightly from the value
sent by the Master due to rounding.
3. Use Command 40 (Enter/Exit Fixed Current Mode) to update the 20.00 mA current.
4. Using the measured value of your reference instrument (either a digital multimeter or the HART master device),
set the span trim using Command 46 (Trim Loop Current Gain). The transmitter will then trim its calibration and
return the loop current value in the response message. The response value may differ slightly from the value
sent by the Master due to rounding.
5. Repeat steps 1 through 4 as needed to gain the accuracy desired. Once the loop current is calibrated to your
satisfaction, return the device to normal operation by issuing Command 40 (Enter/Exit Fixed Current Mode)
with a value of 0.0. This will take the 9900 out of fixed current mode.

Note: With the H COMM Module installed, the following functions are not accessible via the 9900 keypad:
• Trim Loop Current
• Test Loop Current
These functions are only accessible via the HART interface.

Changes to Units of Measure in Transmitter

HART devices can be used to change the units of measure in a 9900 Transmitter. After an update, you must
cycle power to the 9900 Transmitter (remove power for 5 seconds, then restore power). In a flow system,
the units update automatically and it is not necessary to cycle power to the 9900 Transmitter.

9900 Transmitter 67
Hart Commands

Universal Commands
All HART Rev. 7.2 Universal Commands are supported:

CMD Command 0 – Read Unique Identifier


Function
ID Returns device type, device and software revision levels, device
0 Read Unique Identifier status, and codes for the manufacturer and product information.
1 Read Primary Variable Command 1 – Read Primary Variable
2 Read Loop Current And Percent Of Range Returns the numeric value of the Primary Variable
3 Read Dynamic Variables And Loop Current (the 4 to 20 mA current loop) and the unit code for that value
6 Write Polling Address
(e.g. ’45.3’ and ‘Degrees Celsius’).
7 Read Loop Configuration Command 2 – Read Loop Current and Percent of Range
8 Read Dynamic Variable Classification Returns the loop current value of the 4 to 20 mA current loop and
the percent of range (e.g. ’12.0’ and ‘50%’).
9 Read Device Variable With Status
11 Read Unique Identifier Associated With Tag Command 3 – Read Dynamic Variables and Loop Current
12 Read Message
Returns the loop current value of the 4 to 20 mA current loop, as
well as the numeric value of the Secondary Variable (if present)
13 Read Tag, Descriptor, Date
and the Secondary Value’s unit code.
14 Read Primary Variable Transducer Information
Command 6 – Write Polling Address
15 Read Device Information
Enables (or disables) Multi-Drop mode. While in Multi-Drop
16 Read Final Assembly Number
mode, loop current is held at a fixed value and is no longer
17 Write Message available for signaling. Also sets the polling address of the
18 Write Tag, Descriptor, Date device for Multi-Drop mode.
19 Write Final Assembly Number Command 7 – Read Loop Configuration
20 Read Long Tag Reads the polling address of the device and the loop
21 Read Unique Identifier Associated With Long Tag configuration (see Command 6).
22 Write Long Tag
Command 8 – Read Dynamic Variable Classifications
38 Reset Configuration Changed Flag
Returns the classification code for the Primary Variable and
48 Read Additional Device Status Secondary Variable (if present).

Command 9 – Read Device Variable with Status


Returns the value, status, variable code, variable classification and unit code of up to four device variables.
Command 11 – Read Unique Identifier Associated with Tag
Returns all identity information associated with the device, i.e., the device type, device revision level and Device ID.
Issued using the ‘tag’.
Command 12 – Read Message
Read back the message stored in the device. See Command 17.
Command 13 – Read Tag, Descriptor, Date
Reads the tag, descriptor and date values contained within the device. See Command 18.
Command 14 – Read Primary Variable Transducer Information
Reads transmitter serial number, unit code, upper and lower limits and minimum span for primary variable.
Command 15 – Read Device Information
Returns the alarm selection code, transfer function code, upper and lower range values, write protect code and
unit code.

68 9900 Transmitter
Hart Commands

Universal Commands - Continued

Command 16 – Read Final Assembly Number


Returns the assembly number of the device. This will be defined by the customer. See Command 19.
Command 17 – Write Message
Write a message to be stored in the device. See Command 12.
Command 18 – Write Tag, Descriptor, Date
Writes the tag, descriptor and date values into the device. See Command 13.
Command 19 – Write Final Assembly Number
Writes the final assembly number of the device. See Command 16.
Command 20 – Read Long Tag
Read the 32-byte long tag. The ‘long tag’ is separate from the ‘tag’ that is used in Commands 13 & 18.
Command 21 – Read Unique Identifier Associated with Long Tag
Returns all identity information associated with the device - the device type, device revision level and Device ID.
Issued using the long tag.
Command 22 – Write Long Tag
Write the 32-byte long tag. See Command 20.
Command 38 – Reset Configuration Changed Flag
Resetting the device’s configuration changes counter back to 0.
Command 48 – Read Additional Device Status
Returns extended device status information.

Supported HART Common Practice Commands


The following Common Practice Commands are supported.

CMD Command 40 - Enter/Exit Fixed Current Mode


Function
ID The loop current of the 9900 is set to the value transmitted in the command
40 Enter/Exit Fixed Current Mode (in milliamperes). Setting a level of ‘0’ exits Fixed Current Mode.
45 Trim Loop Current Zero If the device is in Multi-Drop mode, Error Code 11 will be returned.
46 Trim Loop Current Gain Command 45 – Trim Loop Current Zero
54 Read Device Variable Information The 9900 will trim its offset of the loop current to match the loop current
value sent to it. This is typically performed at 4.00 milliamperes to optimize
calibration.
Command 46 – Trim Loop Current Gain
The 9900 will trim the gain of the loop current to match the loop current value sent to it.
This is typically performed at 20.00 milliamperes to optimize calibration.
Command 54 – Read Device Variable Information
Returns serial number, limits, damping value and minimum span for a selected device variable.

Unit Codes

The H COMM module uses standard HART Foundation Protocol 7.2 unit codes. The unit code allows the
HART Master to interpret and display the units of measure (e.g., GPM, PPB, °F, etc.) with two exceptions.
The following Unit Codes will not be interpreted by the HART Master:
Code Measurement Unit
240 Cubic Centimeters A HART Master will display these unit codes instead
244 Parts per Thousand
of the units of measure that the code represents.

9900 Transmitter 69
Specifications
General Performance Specifications
Input channels ............One System Accuracy
Enclosure and Display ● Primarily dependent upon the sensor.
Case Material .............PBT System Response
Window .......................Shatter-resistant glass ● Primarily dependent upon the sensor. Controller adds
Keypad .......................4 buttons, injection-molded a maximum of 150 ms processing delay to the sensor
silicone rubber seal electronics.
Display ........................Backlit, 7- and 14-segment ● Minimum update period is 100 ms
Indicators ....................“Dial-type” digital bar graph ● System response is tempered by the display rate,
Update rate ................. 1 s output averaging and sensitivity feature.
LCD Contrast .............. 5 settings
Enclosure
Size .....................¼ DIN Electrical Requirements
Color ....................Black (Panel Mount), Power to Sensors
Yellow and black (Integral Mount) Voltage............................... +4.9 to 5.5 VDC @ 25 °C, regulated
Mounting Current............................... 1.5 mA max in loop power mode;
Panel ...................¼ DIN, ribbed on four sides for panel 20 mA max when using DC power
mounting clip inside panel, silicon Short Circuit .......................Protected
gasket included Isolation ............................. Low voltage (< 48 V AC/DC)
Field.....................Mounts to standard Signet field to loop with DC power connected
mount junction boxes. Optional angle No isolation when using loop power only
adjustment adapter is available
Wall ...................... Large enclosure (sold as an accessory)
that encases the panel mount transmitter Input Power Requirements
Terminal Blocks DC (preferred) ................... 24 VDC; input range:
Pluggable screw type: use minimum 105 °C rated wire 10.8 to 35.2 VDC regulated
Torque ratings 9900 w/o Relay Module ......200 mA *
Cond/Res, Open Collector, 9900 w/ Relay Module........300 mA *
Power/Loop..........0.33 Nm (3.0 lb-in.) *The current draw of the other modules and the sensors are minimal
Freq/S3L ...............0.24 Nm (2.2 lb-in.) Loop................................... 10.8 to 35.2 VDC
Relay....................0.49 Nm (4.4 lb-in.) 4 to 20 mA (30 mA max.)
Connector wire gauge: Overvoltage protection ....... 48 Volt Transient Protection
Power, Loop .........12 to 28 AWG Device (for DC ONLY)
Open Collector .....12 to 28 AWG Current limiting for circuit protection
Freq/S3L ...............16 to 28 AWG Reverse-Voltage protection
Module connector wire gauge:
Relay....................12 to 28 AWG
Cond/Res .............16 to 28 AWG
Batch....................up to 14 AWG Loop Characteristics
4-20 mA Output....up to 14 AWG DC Powered System (preferred)
Max. loop impedance:
@ 12 V loop power ........ 250 Ω max.
Environmental Requirements @ 18 V loop power ........ 500 Ω max.
Ambient operating temperature: @ 24 V loop power ........ 750 Ω max.
Backlit LCD .............-10 °C to 70 °C (14 °F to 158 °F) Loop Powered System
Storage Temp .........-15 °C to 70 °C (5 °F to 158 °F) Max. loop impedance:
Relative Humidity........0 to 100% condensing for Field and @ 12 V loop power .......... 50 Ω max.
Panel Mount (front only); @ 18 V loop power ........ 325 Ω max.
0 to 95% non-condensing for Panel @ 24 V loop power ........ 600 Ω max.
Mount back side
Maximum Altitude .......4,000 m (13,123 ft); use only
DC power supply to maintain Standards and Approvals
UL safety standard up to this altitude ● CE, UL, CUL, WEEE
Enclosure Rating ..........Designed to meet NEMA 4X/IP65 ● RoHS Compliant
(front face only on panel mount; ● Lloyd's Register
field mount is 100% NEMA 4X/IP65) ● Manufactured under ISO 9001 for Quality,
Installation Category Cat II ISO 14001 for Environmental Management and
Pollution Degree 2 OHSAS 18001 for Occupational Health and Safety.
China RoHS (Go to gfsignet.com for details)
Shipping Weights Declaration of Conformity according to FCC Part 15
Base Unit ....................0.63 kg (1.38 lb) This device complies with Part 15 of the FCC rules.
H COMM Module ........0.16 kg (0.35 lb) Operation is subject to the following two conditions:
(1) This device may not cause harmful interference, and
Conductivity Module ...0.16 kg (0.35 lb) (2) This device must accept any interference received,
Relay Module..............0.19 kg (0.41 lb) including interference that may cause undesired operation.
Batch Module..............0.16 kg (0.35 lb)
Output Module ............0.16 kg (0.35 lb)

70 9900 Transmitter
Specifications
Relay Specifications Display Ranges:
Hysteresis ............................. Adjustable (absolute in pH ...............................-1.00 to 15.00 pH
Engineering Units) pH Temp..................-39.99 °C to 149.99 °C
Latch ..................................... Reset in test screen only (-40 °F to 302 °F)
On Delay............................... 9999.9 seconds (max) ORP ............................-1999 to +1999 mV
Cycle Delay .......................... 99999 seconds (max) Flow Rate ...................-9999 to 99999 units per second,
Test Mode ............................. Set On or Off minute, hour or day
Maximum Pulse Rate ........... 300 pulses/minute Totalizer...................0.00 to 99999999 units
Proportional Pulse ................ 400 pulses/minute Conductivity ................0.0000 to 99999 μS, mS, PPM and
Volumetric Pulse Width......... 0.1 to 3200 s PPB (TDS), kΩ, MΩ
PWM period .......................... 0.1 to 320 s Cond. Temp.............-99 °C to 350 °C (-146 °F to 662 °F)
Open Collector Temperature ...............-99 °C to 350 °C (-146 °F to 662 °F)
Type ...................................... NPN Pressure .....................-40 to 1000 psi
Max. Voltage Rating ............. 30 VDC Level ...........................-9999 to 99999 m, cm, ft, in, %
Max. Current Rating ............. 50 mA Volume ......................... 0 to 99999 cm3, m3, in3, ft3, gal, L, lb, kg, %
Dry-Contact Relays Salinity ........................0 to 99.97 PPT
Type ...................................... SPDT Dissolved O2 ...............0 to 20 mg/L, 0 to 200%
Form ..................................... C
Max. Voltage Rating ............. 30 VDC or 250 VAC
Max. Current Rating ............. 5 A resistive Output Specifications
Current Loop Out ........ANSI-ISA 50.00.01 Class H
Span ...........................3.8 to 21 mA
Input Types Zero ............................4.0 mA factory set; user programmable
● Digital (S3L) or AC frequency from 3.8 to 4.2 mA
● 4 to 20 mA input via the 8058 Full Scale ....................20.00 mA factory set; user
● Open collector programmable from 19.0 to 21.0 mA
● pH/ORP input via the Digital (S3L) output from the Accuracy .....................± 32 μA max. error @ 25 °C
2750/2751 pH/ORP Sensor Electronics @ 24 VDC
● Raw Conductivity/Resistivity input directly from Signet Resolution...................6 μA or better
Conductivity/Resistivity electrodes via Direct Conductivity/ Temp. Drift ..................± 1 μA per °C
Resistivity Module or via the Digital (S3L) output from the Pwr Sply Rejection .....± 1 μA per V
2850 Conductivity/Resistivity Sensor Electronics Isolation ......................Low voltage (< 48 VAC/DC)
Voltage........................10.8 to 35.2 VDC
Sensor Types:
Max. Impedance: ........250 Ω @ 12 VDC
Flow, pH/ORP, Conductivity/Resistivity, Salinity, Pressure, 500 Ω @ 18 VDC
Temperature, Level/Volume, Dissolved Oxygen, Other (4-20 mA) 750 Ω @ 24 VDC
Update Rate ...............100 mS nominal
Actual update rate determined by sensor type
Input Specifications
Digital (S3L)...................... Serial ASCII, TTL level, 9600 bps Short circuit and reverse polarity protected
Frequency Type Sensors: Adjustable span, reversible
Sensitivity ................... (for coil type sensors): Error Condition ...........Selectable error condition 3.6 or 22 mA.
80 mV @ 5 Hz, gradually increasing Test Mode ...................Increment to desired current
with frequency to 2.5 V (range 3.8 to 21.00 mA)
Freq. Range ............... (for square wave type sensors): Open Collector Output ..1
0.5 Hz to 1500 Hz @ TTL level Analog Outputs ...........1 passive
input or open collector
Accuracy .................... ± 0.5% of reading max error @ 25 °C
Range ........................ 0.5 to 1500 Hz
Resolution .................. 1 μs
Repeatability .............. ± 0.2% of reading
Power Supply
Rejection .................... No Effect ± 1 μA per volt
Short Circuit ............... Protected
Reverse Polarity......... Protected (no isolation when using
loop power only) Maintenance
Update Rate .................... (1/frequency) + 150 ms

● Clean the instrument case and front panel with a soft


Current Outputs cotton cloth dampened with a mild liquid soap solution.
● One 4 to 20 mA output in base unit (additional 4 to 20 mA
● Never wipe the front window with static retentive
output available via 3-9900.398-1 Output Module)
● Linear scaling cloths such as wool or polyester which may induce
● Logarithmic scaling for Conductivity a static charge. If a static charge develops on the
● Reverse span window, you may notice temporary blotches form on
● Selectable error mode: 3.6 mA or 22 mA the screen. When this occurs, clean the front window
● Test Output mode: allows testing of the current output with an anti-static cloth, or a soft cotton cloth and anti-
● Adjustable 4 to 20 mA end points static spray or a mild liquid soap solution to remove
● HART communication via optional H COMM Module the static charge.

9900 Transmitter 71
Ordering Information

9900 Transmitter Base Unit:


Single Channel, Multi-Parameter, 4 to 20 mA, Open Collector, DC Power
Mfr. Part No Code Description
3-9900-1P 159 001 695 9900 Base Unit, Panel Mount
3-9900-1 159 001 696 9900 Base Unit, Field Mount
3-9900-1BC 159 001 770 Batch Controller System

Optional Modules
3-9900.393 159 001 698 Relay Module - 2 DCR (Dry Contact Relays)
3-9900.394 159 001 699 Direct Conductivity/Resistivity Module
3-9900.395 159 001 697 H COMM Module
3-9900.397 159 310 163 Batch Module
3-9900.398-1 159 001 784 4 to 20 mA Output Module

Accessories
3-9900.090-CD 159 900 094 9900 Transmitter product manual CD
6682-0204 159 001 709 Conductivity Module Plug, 4 Pos, Right Angle
6682-1102 159 001 710 Open Collector Plug, 2 Pos, Right Angle
6682-1103 159 001 711 Relay Module Plug, 3 Pos, Right Angle
6682-1104 159 001 712 Power/Loop Plug, 4 Pos, Right Angle
6682-3104 159 001 713 Freq/S3L Plug, 4 Pos, Right Angle
6682-3004 159 001 725 Freq/S3L Plug, In-Line
7310-1024 159 873 004 24 VDC Power Supply, 10 W, 0.42 A
7310-2024 159 873 005 24 VDC Power Supply, 24 W, 1.0 A
7310-4024 159 873 006 24 VDC Power Supply, 40 W, 1.7 A
7310-6024 159 873 007 24 VDC Power Supply, 60 W, 2.5 A
7310-7024 159 873 008 24 VDC Power Supply, 96 W, 4.0 A
3-0252 159 001 808 0252 Configuration Tool
3-8050 159 000 184 Universal Mount Kit
3-8050.396 159 000 617 RC Filter Kit (for relay use), 2 per kit
3-8051 159 000 187 Flow Sensor Integral Mounting Kit, NPT, Valox
3-8051-1 159 001 755 Flow Sensor Integral Mounting Kit, NPT, PP
3-8051-2 159 001 756 Flow Sensor Integral Mounting Kit, NPT, PVDF
3-8052 159 000 188 ¾ in. Integral Mount Kit
3-8058-1 159 000 966 i-Go Signal Converter, wire-mount
3-8058-2 159 000 967 i-Go Signal Converter, DIN rail mount
3-9900.390 159 001 714 Standard Connector Kit, Right Angle, (Included with 9900 Transmitter)
3-9900.391 159 001 715 Connector Kit, In-Line, 9900 Transmitter
3-9900.392 159 001 700 Wall Mount Accessory for 9900
3-9000.392-1 159 000 839 Liquid Tight Connector Kit, NPT (1 pc.)
3-9900.396 159 001 701 Angle Adjustment Adapter Kit (for Field Mounting)

Georg Fischer Signet LLC, 3401 Aero Jet Avenue, El Monte, CA 91731-2882 U.S.A. • Tel. (626) 571-2770 • Fax (626) 573-2057
For Worldwide Sales and Service, visit our website: www.gfsignet.com • or call (in the U.S.): (800) 854-4090
For the most up-to-date information, please refer to our website at www.gfsignet.com

3-9900.090 Rev. H 08/17 English © Georg Fischer Signet LLC 2017

You might also like