Astm F90-14
Astm F90-14
Astm F90-14
Copyright by ASTM Int'l (all rights reserved); Wed Sep 21 17:33:39 EDT 2016
2
Downloaded/printed by
(UFSC) Universidade Federal de Santa Catarina ((UFSC) Universidade Federal de Santa Catarina) pursuant to License Agreement. No further reproductions authorized.
F90 − 14
TABLE 3 Mechanical Requirements, Bar and WireA
Condition
Ultimate Tensile Strength, Yield Strength (0.2 % Offset), ElongationB ,C in 4D or 4W,
MPa (psi), min MPa (psi), min min, %
Annealed 860 (125 000) 310 (45 000) 30
Cold-workedD 1250 (180 000) 760 (110 000) 15
A
Annealed wire less than 0.063 in. (160 mm) diameter is covered in Specification F1091.
B
4D = 4 × diameter; 4W = 4 × width.
C
Elongation of material 0.062 in. (1.575 mm) or greater in diameter (D) or width (W ) shall be measured using a gauge length of 2 in. or 4D or 4W. The gauge length shall
be reported with the test results. The method for determining elongation of material under 0.062 in. (1.575 mm) in diameter or thickness may be negotiated. Alternately,
a gauge length corresponding to ISO 6892 may be used when agreed upon between supplier and purchaser; 5.65 times the square root of So, where So is the original
cross-sectional area).
D
Cold worked and aged condition may be ordered in accordance with mechanical requirements agreed upon between supplier and purchaser.
TABLE 5 Microcleanliness Requirements unit in the last right-hand digit used in expressing the specifi-
Inclusion A B C
D cation limit, in accordance with the rounding method of
(Globular Practice E29.
Type (Sulfide) (Alumina) (Silicate)
Oxides)
Thin 1.5 1.5 1.5 1.5 11. Certification
Heavy 1.0 1.0 1.0 1.0
11.1 The supplier shall provide a certification that the
material was tested in accordance with this specification and
met all requirements. A report of the test results shall be
furnished to the purchaser at the time of shipment.
of SI units. Annex A of IEEE/ASTM SI 10 provides conversion
tables and Annex B of IEEE/ASTM SI 10 provides rules for 12. Quality Program Requirements
conversion and significant digits. 12.1 The supplier shall maintain a quality program, such as
defined in ISO 9001, or similar.
10. Significance of Numerical Limits
10.1 The following applies to all specified numerical limits 13. Keywords
in this specification. To determine conformance to these limits, 13.1 cobalt alloys (for surgical implants); cobalt chromium;
an observed or calculated value shall be rounded to the nearest L-605 alloy; metals (for surgical implants)—cobalt alloys
APPENDIXES
(Nonmandatory Information)
X1. RATIONALE
X1.1 The primary reason for this specification is to charac- X1.3 ISO standards are listed for reference only. Use of an
terize composition and properties to ensure consistency in the ISO standard, in addition to, or instead of a preferred ASTM
starting material used, directly or as modified by forging, standard may be negotiated between the purchaser and sup-
casting, or shape forming, in the manufacturing of medical plier.
devices.
X1.4 This cobalt base alloy, UNS R30605, is known generi-
X1.2 The minimum mechanical properties for sheet and cally as “L-605” and has been used extensively in the aero-
strip supplied in the annealed condition are generally recog- space industry since the early 1980s. Aerospace Material
nized as commercially attainable for the thickness range cited Specification AMS 5759 includes the chemical and mechanical
in this specification. properties for the UNS 30605, Cobalt Alloy, Corrosion and
Copyright by ASTM Int'l (all rights reserved); Wed Sep 21 17:33:39 EDT 2016
3
Downloaded/printed by
(UFSC) Universidade Federal de Santa Catarina ((UFSC) Universidade Federal de Santa Catarina) pursuant to License Agreement. No further reproductions authorized.
F90 − 14
Heat Resistant, Bars, Forgings, and Rings, 52Co - 20Cr - 10Ni and 2018. In the first phase of this transition, running to 2013,
15W, Solution Heat-Treated. ISO standard 5832-5 for Wrought the specifications will be structured to allow the use of either SI
Cobalt-Chromium-Tungsten-Nickel Alloy also describes co- or inch - pound units. The choice of primary units in each
balt base alloy UNS R30605. specification will be determined by the industry using the
specification. The change to SI units during this period may be
X1.5 Units of Measure initiated by the purchaser through his purchase documentation.
X1.5.1 ASTM Policy—ASTM is promoting the use of ratio- In the second phase of this transition the specifications will be
nalized SI (metric) units in their standards. The F12.04 written with SI as the primary units. Harmonization with
Committee has modified this specification to facilitate the corresponding ISO documents will be considered when assign-
transition by the medical materials industry to SI between now ing the SI values.
X2. BIOCOMPATIBILITY
X2.1 The material composition covered by this standard has X2.2 No known surgical implant material has ever been
been employed successfully in human implant applications in shown to be completely free of adverse reactions in the human
contact with soft tissue and bone for over a decade. Due to the body. However, long term clinical experience has shown an
well characterized level of local biological response estab- acceptable level of biological response can be expected, if the
lished by this material, it has been used as a control material in material is used in appropriate applications.
Practice F981.
SUMMARY OF CHANGES
Committee F04 has identified the location of selected changes to this standard since the last issue (F90 – 09)
that may impact the use of this standard. (Approved Nov. 1, 2015)
(1) Editorial corrections have been made in order to meet (4) Cold worked and aged condition was added to 5.1, Table 3,
terminology and formatting guidelines established for implant and new 7.3.
material standards within F04.12. (5) Units of measure information was added as new Section 9.
(2) Section 1.2 was changed to make SI units the primary and new X1.5.
units.
(3) Designation for Test Methods E8 was corrected to Test
Methods E8/E8M.
ASTM International takes no position respecting the validity of any patent rights asserted in connection with any item mentioned
in this standard. Users of this standard are expressly advised that determination of the validity of any such patent rights, and the risk
of infringement of such rights, are entirely their own responsibility.
This standard is subject to revision at any time by the responsible technical committee and must be reviewed every five years and
if not revised, either reapproved or withdrawn. Your comments are invited either for revision of this standard or for additional standards
and should be addressed to ASTM International Headquarters. Your comments will receive careful consideration at a meeting of the
responsible technical committee, which you may attend. If you feel that your comments have not received a fair hearing you should
make your views known to the ASTM Committee on Standards, at the address shown below.
This standard is copyrighted by ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959,
United States. Individual reprints (single or multiple copies) of this standard may be obtained by contacting ASTM at the above
address or at 610-832-9585 (phone), 610-832-9555 (fax), or service@astm.org (e-mail); or through the ASTM website
(www.astm.org). Permission rights to photocopy the standard may also be secured from the Copyright Clearance Center, 222
Rosewood Drive, Danvers, MA 01923, Tel: (978) 646-2600; http://www.copyright.com/
Copyright by ASTM Int'l (all rights reserved); Wed Sep 21 17:33:39 EDT 2016
4
Downloaded/printed by
(UFSC) Universidade Federal de Santa Catarina ((UFSC) Universidade Federal de Santa Catarina) pursuant to License Agreement. No further reproductions authorized.