Manual Casagrande and Standpipe Piezometer14

Download as pdf or txt
Download as pdf or txt
You are on page 1of 10

_ CASAGRANDE AND STANDPIPE

PIEZOMETERS
USER MANUAL

www.sisgeo.com Casagrandeandstandpipepiezometer_EN_00_14
CASAGRANDE and STANDPIPE PIEZOMETERS

INDEX IntroduCTION Page 4

DescriPTION Page 5

FILTER CELLS SATURATION Page 5

InstallaTION Page 6

TAKING MEASUREMENTS Page 8

Troubleshooting Page 10

maintenance Page 10

www.sisgeo.com Casagrandeandstandpipepiezometer_EN_00_14 Page 2 of 10


CASAGRANDE and STANDPIPE PIEZOMETERS

Notes on the use of product For safe and efficient use of the product, please read carefully the following
instructions before starting any operation.
Any use of the product other than the one described in this manual shall be considered the
user’s full responsibility.
The same applies for any unauthorized modifications.
In addition to the hereby listed standards, the user must comply with the provisions of the
current legislation regarding personal safety and health together with all other persons in the
workplace.
SISGEO is not responsible for any accident, breakdown or other problems due to lack of
knowledge and / or non-compliance with the requirements contained in this manual.
Check that the product has not been damaged during the transport.
Verify that the package includes all items as well as any requested optional accessories; if
anything is missing, please promptly contact SISGEO.
The user must strictly follow all the operations described in this manual.
Maintenance or repair of the device is permitted only by authorized operators.
These operators must be physically and intellectually suitable.
For information about instrument or to order spare parts, always specify the product
information which can be found on the identification label.
When replacing parts, always use ORIGINAL SPARE PARTS.
The manufacturer reserves the right to make either technical and / or commercial changes
without prior notice.
It is our policy to keep manuals continuously updated.

Symbols

Pay particular attention to the following instruction.

Identification Instruments can be identified


• From a production lot number (written on the Compliance Certificate)
• From a serial number (s/n) engraved indelibly on the instrument
• From a label on the instrument
• From a label on the cable

www.sisgeo.com Casagrandeandstandpipepiezometer_EN_00_14 Page 3 of 10


CASAGRANDE and STANDPIPE PIEZOMETERS

IntroduCTION protective cap Standpipe piezometers, also called “open


circuit”, have a wide use in civil and foundation
engineering, for the survey of water-bearing layer
piezometer height, or for the interstitial pressure
in saturated grounds.
They can have a single or a double casing and
they usually consist in:
• a filter unit (or fissured casings covered with
woven-non-vowen fabric);
• 3m long casings and different diameters
(0.5”G, 1”G, 1.5”G, 2”G, 3”G);
• couplings;
• protective caps (optional).
The borehole is filled around the casing with sand
and/or gravel in order to allow the water to enter
the casing through the filter.
After a while (depending on ground permeability),
piezometer
the water stabilizes at the water layer level.
casing

We have a Casagrande piezometer when the


filter cell is insulated from the borehole in order
to measure the interstitial pressure in the layer
where it is installed.

For the measure we use a water level indicator


or an electric pressure transducer (see manual
“Pressure Transducer”).
Sisgeo designed a special pressure transducer
that allows to close the filter cell obtaining a
“close circuit” piezometer: the measure of
the interstitial pressure variations occurs more
quickly also in waterproof grounds such as clay.
This transducer can be removed for the
maintenance (see manual “Pressure
Transducer”).
standpipe filter

unit

www.sisgeo.com Casagrandeandstandpipepiezometer_EN_00_14 Page 4 of 10


CASAGRANDE and STANDPIPE PIEZOMETERS

DescriPTION WATER LEVEL INDICATOR

Water level indicator consists of:


1. a coated steel graduated tape and 2 electric conductors at sides;
2. a probe;
3. some weights;
4. a cable reel. On a side there is a 9V power battery and an electronic circuit with a
ringtone and a led light (2 batteries and a display for the water/temperature level
probe).

3 1

FILTER CELLS Porous filters saturation is needed to:


SATURATION •

minimize the possibility that the filter pores would block;
increase the filter system answer speed when an external pressure is applied.

Usually the filter that has to be saturated is left in a container full of water for 24 h before
the installation.
The saturation is more effective if the water had been previously de-aerated making it boil
for at least 10 minutes and then make it cool.

www.sisgeo.com Casagrandeandstandpipepiezometer_EN_00_14 Page 5 of 10


CASAGRANDE and STANDPIPE PIEZOMETERS

INSTALLATION STANDPIPE PIEZOMETERS INSTALLATION

The installation is expected in a borehole carried out through a continuous core or


destructive drilling.
We recommend to always stabilize the walls with a temporary covering casing, with an
internal diameter not less 85mm, while the use of muds shall be avoided.
To stabilize the bottom hole, without water, you shall keep water level within the borehole a
little higher than piezometer level in the ground.
This precaution has to be taken also during the installation.

The needed tools are: depth meter, bentonite pellets, teflon tape, adhesive tape, jig saw.

Installation scheme without filter

tmeasure casing
covering casing

100cm

sand or gravel
70cm
depth meter

filter cell
70cm

sand or gravel
50cm

www.sisgeo.com Casagrandeandstandpipepiezometer_EN_00_14 Page 6 of 10


CASAGRANDE and STANDPIPE PIEZOMETERS

Casagrande installation scheme with seal

measure casing
covering casing

Bentonite waterproof cap


100cm

100cm
70cm
depth meter

filter cell
70cm

sabbia o ghiaia
50cm

GENERAL RECOMMENDATIONS
• Saturate the cell as explained in the previous paragraph;
• flush accurately the hole with clean water until doesn’t get out clear water;
• use fine sand or gravel (Ø max 0.5 cm);
• if the piezometer is expected to be at an intermediate quote, fill in the empty part with
a mix of water/cement/bentonite in order to recreate the consistency of the original
ground;
• wrap the teflon on the threads to secure the holding. Join two casings using the
couplings;
• be careful when you screw the casings to the filter cell. Handle the cell using the black
upper junction. DO NOT handle it by the filter. Screw without forcing. If there is a
double casing, tie them together with adhesive tape;
• lower the casings in the borehole by hand. Be careful. Don’t use pliers;
• let the casing stick out the ground for 20-30 cm (if necessary);
• recover the coverings ensuring that the instrument will not be lifted;
• use the protective cap to protect the casing.

www.sisgeo.com Casagrandeandstandpipepiezometer_EN_00_14 Page 7 of 10


CASAGRANDE and STANDPIPE PIEZOMETERS

For Casagrande casing

To create the cap, pour the bentonite in more stages; each layer shouldn’t be more than
25÷30 cm thick.
Compact the bentonite pellets with a pestle (not supplied) of suitable dimensions.
Verify with the depth meter the solidity of the cap and the quote reached.

Note: if there is a waterproof ground coat higher then the layer you have been measuring, it
is recommended that the bentonite cap is made at the same height of the layer to restore
the continuity.

If you need to install two piezometers (with two different layers) in the same borehole, it
will be necessary to make two filtering coats and two seals.
Pay attention to the stratigraphy and to the permeable and waterproof ground’s coats
thickness.
In this case the internal diameter of the casings must not be less then 110mm.

TAKING To perform a measurement you have to gather the water level quote within the casing,

MEASUREMENTS using the graduated tape of the water level indicator.


It’s important that the measurement is taken always at the same quote.
Usually the reference is the casing head.
The topographyc survey of the casing height is useful to define the layer level
measurement as absolute value.

1. Drop carefully the probe in the piezometer casing until you hear a sound and the led
switches on;
2. lift the probe until the sound stops;
3. if necessary shake a little the graduated tape to remove any drop of water that would
stick on the probe and create a false signal;
4. drop slowly the graduated tape until you reach precisely the level;
5. gather the value in the chosen reference point;
6. rewind the tape on the reel;
7. clean accurately the probe. Do not use solvents or strong detergents.

www.sisgeo.com Casagrandeandstandpipepiezometer_EN_00_14 Page 8 of 10


CASAGRANDE and STANDPIPE PIEZOMETERS

MEASURE WITH WATER LEVEL INDICATOR/


TEMPERATURE
The water level probe can be supplied with
temperature sensor to gather the temperature at
different heights.
• Turn the switch (ON/OFF) on ON. Level
measurement is active;
• follow the steps as described earlier;
• move the switch in a central position to turn
it off;
• move the switch on temperature ON to
gather the temperature in °C on the display.
Attention: the probe must be underwtater
to work. Wait 10/15 seconds (or the time
needed) in order to adapt the steel probe to
the temperature in the casing;
• move the switch in a central position to turn
it off;
• rewind the tape on the reel;
• clean accurately the probe. Do not use
solvents or strong detergents.

CASAGRANDE PIEZOMETER

With double casing installation, if the


measurements taken in the casing will become
very different, in time, is recommended to clean
the filter cell in order to clean the casings and the
filter from impurities or air bubbles.
Enter, in one of the two casings, clean water at
the minimum pressure needed for a forced flow,
making it leave from the other casing.

www.sisgeo.com Casagrandeandstandpipepiezometer_EN_00_14 Page 9 of 10


CASAGRANDE and STANDPIPE PIEZOMETERS

TROUBLESHOOTING Problem Possible cause Solution

There is no sound Low battery Change the battery


Shake lightly the graduated
False contact caused from
Non-stop sound tape to remove the
waterdrops or sticky liquids.
waterdrops.

Servizio After-sales assistance for calibrations, maintenance and repairs, is performed by SISGEO’s

manutenzione service department.


The authorization for shipment shall be activated by RMA “Return Manufacturer
Authorization”. Create your account and then fill in the RMA module clicking on:

http://www.sisgeo.com/repairs.html

Send back the instrument/equipment with the complete accessories, using suitable
packaging, or, even better, the original ones.
The shipping costs shall be covered by the sender.

Please return to the following address with correct delivery documentation:

SISGEO S.r.l.
Via F.Serpero, 4/F1
20060 MASATE (MI)

On the delivery document it is mandatory to indicate the RMA code received.

Technical assistance e-mail: [email protected]

www.sisgeo.com Casagrandeandstandpipepiezometer_EN_00_14 Page 10 of 10