Tehnic Ventis Mx4
Tehnic Ventis Mx4
Tehnic Ventis Mx4
Product Manual
Set-up
Operation
Service
Contents
►Copyright Notice .........................................................................................................................................................4
►Warnings and Cautionary Statements ........................................................................................................................4
General...................................................................................................................................................... 4
Personnel .................................................................................................................................................. 4
Hazardous Conditions, Poisons, and Contaminants ................................................................................ 4
General Usage .......................................................................................................................................... 5
Agency-issued Conditions of Use and Warnings ...................................................................................... 5
Recommended Practices .......................................................................................................................... 6
►Ventis MX4 Resources ...............................................................................................................................................6
►Ventis MX4 Capabilities ..............................................................................................................................................7
►Unpacking the Monitor................................................................................................................................................7
Contents .................................................................................................................................................... 7
Hardware Features and Functions ............................................................................................................ 8
Display Screen .......................................................................................................................................... 9
Alarms ..................................................................................................................................................... 11
►Monitor Set-up .......................................................................................................................................................... 14
Batteries .................................................................................................................................................. 15
Docking Stations, Chargers, and other Accessories .............................................................................. 15
Battery Charging ..................................................................................................................................... 16
Power-on and -off .................................................................................................................................... 16
Configuration ........................................................................................................................................... 18
Instructions .............................................................................................................................................. 19
►Monitor Use and Service .......................................................................................................................................... 27
Zero, Calibration, and, Bump testing ...................................................................................................... 27
Recommendations .................................................................................................................................. 27
General information ................................................................................................................................. 28
Instructions .............................................................................................................................................. 28
Supplies................................................................................................................................................... 29
Remote Sampling .................................................................................................................................... 35
Cleaning .................................................................................................................................................. 36
Service .................................................................................................................................................... 36
Aspirated Monitor .................................................................................................................................... 37
Diffusion Monitor ..................................................................................................................................... 39
►Copyright Notice
All trademarks and registered trademarks are the property of their respective owners.
These help materials or any part thereof may not, without the written consent of Industrial Scientific Corporation, be
copied, reprinted, or reproduced in any material form including but not limited to photocopying, transcribing,
transmitting, or storing it in any medium or translating it into any language, in any form or by any means, be it digitally,
electronic, mechanical, xerographic, optical, magnetic, or otherwise.
The information contained in this document is proprietary and confidential and all copyright, trademarks, trade names,
patents, and other intellectual property rights in the documentation are the exclusive property of Industrial Scientific
Corporation unless otherwise specified. The information (including but not limited to data, drawings, specification,
documentation, software listings, source or object code) shall not at any time be disclosed directly or indirectly to any
third party without prior written consent.
The information contained herein is believed to be accurate and reliable. Industrial Scientific Corporation accepts no
responsibility for its use by any means or in any way whatsoever. Industrial Scientific Corporation shall not be liable
for any expenses, costs by damage that may result from the use of the information contained within this document.
Although every effort is made to ensure accuracy, the specifications of this product and the content herein are subject
to change without notice.
General
IMPORTANT
Failure to perform certain procedures or note certain conditions may impair the performance of this
product. For maximum safety and optimal performance, please read and understand the Ventis MX4
Product Manual available online at the Ventis MX4 Resource Center at
www.indsci.com/ VentisMX4resources.
Personnel
CAUTION: For safety reasons, this equipment must be operated and serviced by qualified personnel only.
Read and understand the product manual completely before operating or servicing.
WARNING: Substitution of components may impair intrinsic safety and may cause an unsafe condition.
CAUTION: Any rapid up-scale reading followed by a declining or erratic reading may indicate gas
concentration(s) beyond the upper scale limit which may be hazardous.
Silicone compound vapors or other known contaminants may affect the combustible gas sensor and cause
readings of combustible gas to be lower than actual gas concentrations. If the monitor has been used in an
area where silicone vapors were present, always calibrate the monitor before next use to ensure accurate
measurements.
General Usage
Oxygen-deficient atmospheres may cause combustible gas readings to be lower than actual concentrations.
Oxygen-enriched atmospheres may cause combustible gas readings to be higher than actual
concentrations.
Sudden changes in atmospheric pressure may cause temporary fluctuations in the oxygen reading.
Verify the calibration of the combustible gas sensor after any incident where the combustible gas content has
caused the monitor to display an over-range condition.
Sensor openings, water barriers, and the pump inlet must be kept clean. Obstruction of the sensor openings
or pump inlet and/or contamination of the water barriers may cause readings to be lower than actual gas
concentrations.
To avoid the potential of liquid being pulled into the sample tubing and pump assembly, it is recommended
that Industrial Scientific filter (P/N 17027152) be used on the sample tubing when drawing samples using the
aspirated monitor.
WARNING: INSERT THE ALKALINE BATTERIES WITH THE CORRECT POSITIVE “+” AND NEGATIVE “-“
ORIENTATION. WARNING: The Ventis MX4 is only approved for use with AAA battery types Energizer
EN92 and Duracell MN2400. Do NOT mix battery types.
WARNING: The use of leather cases can produce inaccurate readings with diffusion (non-aspirated) gas
detection instruments for specific monitoring applications. Leather cases should be used ONLY as carrying
cases, and NOT for continuous monitoring, with diffusion instruments configured to measure gases other
than O2, CO, CO2, H2S, and combustible gases (LEL/CH4).
Ensure all part-use restrictions (e.g., battery) meet any agency-mandated conditions of use.
Ensure all instrument-configurable settings (e.g., always-on setting) meet any agency-mandated conditions
of use. When using instrument-compatible Industrial Scientific docking stations, maintain mandated settings
through the software (e.g., iNet Control or Accessory Software) or by manually configuring the instrument
settings after docking.
The Ventis MX4 is CSA certified according to the Canadian Electrical Code for use in Class I, Division 1 and
Class I, Zone 1 Hazardous Locations within an ambient temperature range of Tamb: -20°C to +50°C. CSA
has assessed only the %LEL combustible gas detection portion of this instrument for performance according
to CSA Standard C22.2 No. 152. This is applicable only when the monitor is used in the diffusion mode and
has been calibrated to 50% LEL CH4, and when the monitor is used in the aspirated mode with an Extended
range lithium-ion battery and has been calibrated to 50% LEL CH4.
CAUTION: CSA C22.2 No. 152 requires before each day’s usage, sensitivity must be tested on a known
concentration of pentane or methane equivalent to 25% or 50% of full scale concentration. Accuracy must
be within -0% to +20% of actual concentration. Accuracy may be corrected by referring to the
zero/calibration section of the Product Manual.
The equipment complies with the standards IEC 60079-29-1 and EN 60079-29-1 for methane, propane, and
hexane with the following exception: as for the methane (mine) detector, the battery run time was verified to
be seven (7) hours rather than the eight (8) hours recommended by the standards, respectively.
MED-certified instruments may be used only when configured and maintained to disallow power-off when
the unit is in alarm.
The Mine Safety and Health Administration (MSHA) has approved the Ventis MX4 as a Permissible Multi-
Gas Monitor with the following warnings:
• MSHA approved for use with either the P/N 17134453-X2, 3.7 volt, Lithium-ion battery or P/N
17148313-2, 3.7 volt, lithium-ion Extended battery only. The battery pack is not user-replaceable. The
monitor battery and the lithium battery on the main PCB are technician replaceable only. Charge
rechargeable lithium-ion batteries with an ISC battery charger designed for use with this monitor in fresh
air locations only.
• The monitor is to be calibrated according to the procedures in the instruction manual only.
• The aspirated version of the Ventis MX4 is only approved for use with the Extended range battery only.
• The monitor must display methane in the percent-by-volume mode (0-5%) for compliance
determinations required by 30 CFR Part 75, subpart D.
Recommended Practices
Industrial Scientific Corporation recommends the monitor be fully charged (when equipped with a
rechargeable battery), configured, and calibrated before first time use. If the lithium-ion battery is deeply
discharged, it can take up to an hour for the instrument display to indicate that the battery is charging.
Monitors used infrequently should be fully charged every four months.
No part of the unit should be covered by any garment, part of a garment, or other item that would restrict the
flow of air to the sensors or impair the operator's access to the audible, visual, or vibration alarms.
Industrial Scientific Corporation recommends a full monitor calibration be performed monthly (at a minimum),
using a certified concentration(s) of Industrial Scientific calibration gas(es) to help ensure monitor accuracy.
Industrial Scientific Corporation recommends the monitor be zeroed and bump tested before each use with a
certified concentration(s) of Industrial Scientific calibration gas(es).
Battery contacts are exposed on batteries when they are removed from the monitor. Do not touch the battery
contacts and do not stack batteries on top of one another.
When reassembling the instrument or installing a battery pack, maintain ingress protection by tightening
each fastener to its stated torque value (see the “Ventis MX4 Monitor three-Dimensional Diagram” and its
key in this manual).
Contact your service representative immediately if you suspect that the Ventis MX4 is working abnormally.
The Ventis MX4 Product Manual is the primary resource, within a full suite of learning tools, developed for the monitor
user. Its step-by-step “walk through” format covers everything from unpacking to set-up, operation, and service. All
Ventis MX4 users should read and understand the Product Manual prior to unpacking or using the monitor.
Ventis MX4 product-specific resources are part of the organization’s broader training line-up, featuring online training
modules and face-to-face classroom programs for technicians, operators, first responders, trainers, and distributors.
Courses combine theory with hands-on learning, and can be tailored to the customer’s unique requirements and gas
monitoring applications.
The organization’s customer and technical support call centers provide product and order information, how-to product
assistance, and guidance for in-depth technical applications. Its service centers offer comprehensive factory repair
and maintenance services.
Industrial Scientific Corporation provides a full suite of resources to aid customers in the competent and safe use of
its products and services. With 19 manufacturing, support, and service centers and hundreds of distributors
worldwide, Industrial Scientific serves the globe’s gas detection needs.
The Ventis MX4 is a portable multi-gas monitor. Offered as a diffusion monitor, it detects and measures gas(es)
present in open space. To enable monitor use within confined space locations, the Ventis MX4 is also offered as an
aspirated monitor. A pump module and battery accessories enable the conversion of either monitor for dedicated use
in either confined or open spaces.
Based on the customer’s monitor order, up to four sensors are factory installed enabling the monitor to continuously
and simultaneously detect and measure the presence of up to four specific gases.
Equipped with a multi-mode (audible, visual, and vibration) and multi-level alarm system, the Ventis MX4 monitor is
capable of notifying its user of potentially hazardous gas concentrations.
The monitor performs continuous data logging at 10 second intervals. It can store approximately 90 days of data for a
four-sensor configuration. Its date- and time-stamped event log records and stores data for the following: 60 alarm
events, 30 error events, and 250 manually performed calibrations or bump tests. The memory, when full, overwrites
the oldest data as the newest readings and events are logged.
The Ventis MX4 monitor functions as an independent device to monitor the environment for hazardous gas
concentrations. It is compatible with products that charge, calibrate, bump test, read and record instrument data,
protect, and otherwise enable or enhance use of the monitor and its data. For a complete list of these products,
please refer to the manual section, Ventis MX4 Accessories and Parts.
Contents
The monitor box contains the following items including, when ordered, those marked optional. Each item ordered
should be accounted for in the unpacking process.
1 as ordered Ventis MX4 Portable Multi- The monitor type is indicated on the box label.
gas Monitor Options:
• Ventis MX4 Diffusion
• Ventis MX4 Aspirated
• Ventis MX4 Aspirated with Conversion Kit
1 Ventis MX4 Quick start guide Review for important safety information before using
the monitor.
1 installed as ordered Battery One of four battery types are factory installed as
indicated on the box label.
Options:
• Rechargeable Lithium-ion (Li-ion)
• Rechargeable Slim extended lithium-ion (Li-ion)
• Rechargeable Extended range lithium-ion (Li-ion)
• Replaceable Alkaline
1 as ordered Ventis Charger Universal power cord. AC charger products include
interchangeable plugs (US, UK, EU, and AUS).
0 or 1 Calibration Cup Diffusion – 1 included
Aspirated – 0 included
1 Calibration and Bump Test Diffusion – two feet of clear tubing
Tubing
0 or 1 In-field Sampling Tubing Diffusion – 0 included
Aspirated – Ten feet of black tubing
1 Final Inspection & Test Contains the following factory set* information:
Report • Monitor Set-up Date
• Monitor Part Number (P/N)
• Monitor Serial Number (S/N)
• For Each Sensor*:
• P/N
• S/N
• Type
• Location
• Alarm level values
• Span gas values
• Span reserve values
1 Warranty Card --
After unpacking, if any item is missing or appears to have been damaged, contact a local distributor of Industrial
Scientific products or Industrial Scientific Corporation (for contact information, please see the manual’s last page).
►Monitor Overview
Hardware Features and Functions
The monitor’s case top (front of monitor) has two main sections. As shown below, the upper section contains the
sensor ports. The lower section houses the user interface features, a LCD display screen and two buttons. Each
feature’s general functions are noted below. As shown, the diffusion and aspirated monitors differ with respect to the
location of the air intake mechanism and visual alarm indicators.
The instrument may be used in any orientation when clipped to the user or when used with a compatible carrying
case. Normal instrument orientation for measuring gas concentration is hand held with sensors and display facing the
operator.
Display Screen
The Ventis MX4 Boot-up Screen, as shown below, serves to introduce all icons and the alpha-numeric items (e.g.,
8.8.8) that can appear on the display when the monitor is in use, docked, or charging. Each display item is stationary,
communicates unique information, and appears only when relevant to the task being performed.
A sample Gas Monitoring Screen is also shown below, next to the boot-up screen. This illustrates how the icons and
the alpha-numeric characters work together to communicate several points of information to the monitor user.
NOTE: Display screens featured throughout this manual include the “pump” icon.
Similar in appearance to a fan, it indicates an aspirated monitor is in use. For a
diffusion monitor, the pump icon does not appear on the display.
It is helpful to view the boot-up screen in sections. The top and bottom rows each contain icons. The main function of
the middle section, in gas monitoring mode, is to communicate gas concentration readings. Definitions for all icons,
gas name abbreviations, gas measurement units, and other indicators are provided below. Where applicable, display
variations are noted.
Zero: communicates zero status (e.g., zero results, zero in-progress, etc.).
Calendar: communicates overdue warnings for service items (calibration, bump testing,
etc.).
Alarm: indicates an alarm causing condition.
Alpha-numeric Definition
display values
Carbon Monoxide (CO)
Methane (CH4)
Sulfur Dioxide (SO2)
Alpha-numeric Definition
display values
Hydrogen Sulfide (H2S)
CO H2/Low
Over-range: for any sensor in over-range, indicates the measured gas concentration is
greater than the measurement range of the sensor. Display variations:
“Or” (English and German)
“Sup” (French)
Negative Over-range: for any sensor in negative over-range indicates the measured gas
concentration is less than the negative measurement range of the sensor. Display
variations:
“-Or” (English and German)
“InF” (French)
Bottom Row Icons Definition
Battery level indicator; display variations:
Empty battery icon with three dashes in place of each sensor reading = critical battery
warning
Flashing empty battery icon = low battery warning
1 black bar = < 33% charge remaining
2 black bars = 34% – 66% charge remaining
3 black bars = 67% – 100% charge remaining
Security Code: indicates code is set or to be entered.
Alarms
NOTICE
→ All monitor alarms and warnings should be taken seriously and responded to as stated in company safety
standards.
→ Once initiated, an alarm will remain on while the alarm condition is present. For gas-related alarms, once the
detected gas concentration changes, the alarm indicators will change to reflect any new condition such as
low-alarm gas, high-alarm gas, over-range gas, or no gas alarm.
→ When the latch alarm feature is enabled and the monitor goes into alarm, it will remain in alarm until the
alarm condition no longer exists and the monitor user presses the ENTER button for one second. This
applies only to gas-related alarms.
It is practical for the monitor user to be aware of the possible alarms prior to monitor set-up and use. The Ventis MX4
has four alarm and warning levels. A “system level” alarm generates the highest frequency tone and highest level
visual and vibration signals. It is used to indicate such events as a pump, critical battery failure, or sensor failure. The
“high” or “low” level audio alarms, in combination with visual and vibration indicators, turn on when gas concentration
readings are over-range, high, or low. The lowest level indicator is a warning with beep patterns to indicate service
needs (e.g., low battery or calibration due). The beep is also used as a confidence indicator when enabled.
Alarm types and their alarm generating conditions are described below.
Display Description
An over-range condition occurs when the gas
concentration value sensed is above the
sensor’s measuring range.
Display Description
Low Alarm Screen
A flashing gas concentration value* indicates which sensor(s)
reading(s) is the cause for alarm. The low level alarms turn on
and the down arrow icon displays.
A TWA alarm occurs when the calculated time
weighted average reaches the monitor’s
hazardous value for the set time frame.
Display Description
Alarm occurs when, if attached, the pump is not
operating correctly. While in alarm, every ten
seconds the monitor attempts to restart the
pump. If unsuccessful, the monitor remains in
alarm.
►Monitor Set-up
Preparing the monitor for first time use is a “3-C” process: charge (if equipped with a lithium-ion battery), configure,
and calibrate. This manual section covers charging and configuration for set-up purposes and can be consulted for
ongoing instruction thereafter. Immediately following this section, calibration is covered in the manual section, Use
and Service.
Batteries
As shown below, the Lithium-ion and Slim extended lithium-ion batteries are compatible with the diffusion instrument
only. The Extended range battery can be installed for use with a diffusion or aspirated instrument. The battery's
orderable part numbers are supplied in Battery part numbers and options.
Battery Compatibility
Rechargeable Replaceable
(part number*) (part number*)
Lithium-ion battery Slim extended Extended range Alkaline battery
lithium-ion battery lithium-ion battery
Cover Cover
Battery Charging
Power-Off Screen
The screen displays a five-second countdown accompanied by five
beeps and LED flashes.
Configuration
Introduction
Before first time use of the monitor, its settings should be reviewed and, if needed, be adjusted. Qualified safety
personnel should complete the following tasks.
• Review the monitor settings for compliance with company policy and any applicable regulations, laws, and
observed guidelines as issued by regulatory agencies and government or industry groups.
• Determine which settings, if any, require adjustment.
• Make the adjustments or supervise other qualified personnel in the process.
• When using instrument-compatible Industrial Scientific docking station software (e.g., iNet, Docking Station
Server Administrative Console [DSSAC], or Accessory Software), maintain the mandated settings through
the software or by manually configuring the instrument settings after docking.
Monitor settings should be reviewed regularly and adjusted as needed. The following settings are adjustable or
“configurable" for the Ventis MX4.
LEL Type
Calibration Mode Setting
Low Alarm Settings
High Alarm Settings
TWA Alarm Settings
TWA Interval Settings
STEL Alarm Setting
Calibration Gas Settings
Clock Settings
Date Settings
Display Mode Setting
Confidence Indicator (on/off)
Confidence Indicator (type)
Bump Test In-field
Bump Test Due Warning
Bump Test Time Set-point
Bump Test Percentage
Bump Test Response Time
Alarm Latch Set
Zero In-field
Calibration In-field
Calibration Due Alarm
Calibration Due Set-point
Security Code
Language Selection
Always-on Setting
Shutdown In Alarm Setting
Alarm on Dock Setting
The Ventis MX4 can be configured manually as instructed below. Any changes made take effect immediately upon
exiting the configuration mode.
Instructions
NOTICES
→ The configuration mode should be accessed only by safety personnel authorized to change monitor settings
based on company policy.
→ Read ALL requirements and instructions outlined below, including the screen-by-screen process description,
before beginning the configuration process.
The configuration mode can be entered during the 20-second countdown of the power-on process. During the
countdown, simultaneously press ON/OFF/MODE and ENTER, hold for three seconds, and release to enter
configuration mode. (While in the configuration mode, the same button presses cause the monitor to exit
configuration). Each configuration screen times out after 30 seconds and the monitor enters gas monitoring mode. To
re-enter the configuration mode, power-off the monitor, then power-on and repeat the entry process.
Throughout the configuration process, the main functions of the two buttons are as follows.
• The ENTER button is used to edit values. It is also used, where noted, to begin a process or a step in a
process.
• The ON/OFF/MODE button is used to set the value. Where noted, it is also used to bypass a process or
step in a process, or to advance to the next configuration screen.
If the security code setting is 000, the security feature is disabled and the Enter Security Code Screen does NOT
appear. If the security code is NOT 000, the security feature is enabled and the monitor displays the Enter
Security Code Screen.
The monitor next checks for the presence of a China MA mining feature. If this feature is operational, the monitor
displays the Zero Initiate Screen.
If the China MA mining feature is NOT operational, the monitor then checks for an installed LEL sensor. If
installed, the monitor displays the LEL Type Screen. If no LEL sensor is installed, the monitor displays the Zero
Initiate Screen.
Configuration Process
Display and Options Instructions
Press ENTER to edit the value, if needed; press repeatedly or hold
down to speed the increment pace to reach the valid security code.
Press ON/OFF/MODE to enter configuration mode and arrive at the
next applicable screen.
NOTE: If the LEL type is changed, the sensor goes into calibration fail mode. A full calibration
is required before the monitor can be used and is accessible from the next screen in the
configuration process, the Zero Initiate Screen. For complete calibration instructions,
LEL Type Set Screen proceed to the manual section, Zero, Calibration, and Bump Testing.
Options
Configuration Process
Display and Options Instructions
LEL
CH4
Press ON/OFF/MODE to bypass the zero and calibration processes
and advance to one of two screens.
If the installed sensor set includes H2S and NO2, OR, SO2 and NO2, the
monitor is pre-set for standard calibration mode and the Low Alarm Set-
point Screen displays.
Zero Initiate Screen
For all others installed sensor combinations, the Calibration Mode
Options
Selection Screen displays.
Bypass zero and calibration process.
Begin zero and calibration process.
Press ENTER to begin the zero and calibration process. Proceed to the
manual section, Zero, Calibration, and Bump Testing.
The quick calibration option sets the monitor to calibrate all four sensors
simultaneously. The standard calibration option sets the monitor to
calibrate each sensor independently.
After the alarm value is set for each installed sensor, press
ON/OFF/MODE to advance to the High Alarm Set-point Screen.
Configuration Process
Display and Options Instructions
Press ON/OFF/MODE to bypass high alarm value set process and
advance to one of two screens as noted below.
Press ENTER to begin the high alarm value set process.
On the display, the first sensor subject to change flashes.
Press ENTER to edit the value, if needed; press repeatedly or hold down
to speed the increment pace.
High Alarm Set-point Screen Press ON/OFF/MODE to set the value.
Displays the existing high alarm value
for each installed sensor. If any one of The next sensor subject to change flashes. Continue to use the ENTER
the sensors is not installed, its position and ON/OFF/MODE buttons, respectively, to edit and set each sensor’s
on the display is blank. high alarm value.
After the alarm value is set for each installed sensor, press
ON/OFF/MODE and advance to one of two screens.
If at least one toxic sensor is installed, the TWA Alarm Set Screen
displays.
After the alarm value is set for each installed sensor, press
ON/OFF/MODE to advance to the TWA Interval Set Screen.
Press ENTER to edit the value, if needed.
Press ON/OFF/MODE to set the value and advance to the STEL Alarm
Set-point Screen.
Configuration Process
Display and Options Instructions
Press ON/OFF/MODE to bypass the STEL alarm value set process and
advance to the Calibration Gas Set Screen.
Press ENTER to begin the STEL alarm value set process.
On the display, the first sensor subject to change flashes.
Press ENTER to edit the value, if needed; press repeatedly or hold down
to speed the increment pace.
STEL Alarm Set-point Screen Press ON/OFF/MODE to set the value.
Displays the existing STEL values for The next sensor subject to change flashes. Continue to use the ENTER
the toxic sensors installed. No other and ON/OFF/MODE buttons, respectively, to edit and set each sensor’s
sensor readings appear. STEL alarm value.
After the alarm value is set for each installed sensor, press
ON/OFF/MODE to advance to the Calibration Gas Set Screen.
Press ON/OFF/MODE to bypass the calibration gas set process and
advance to the Clock Set Screen.
Press ENTER to begin the calibration gas value set process.
On the display, the first sensor subject to change flashes.
Press ENTER to edit the value, if needed; press repeatedly or hold down
to speed the increment pace.
Calibration Gas Set Screen Press ON/OFF/MODE to set the value.
Displays the existing calibration gas The next sensor subject to change flashes. Continue to use the ENTER
value for each installed sensor. and ON/OFF/MODE buttons, respectively, to edit and set each sensor’s
calibration gas value.
If any one of the sensors is not installed, After calibration gas value is set for each installed sensor, press
its position on the display is blank. ON/OFF/MODE to advance to the Clock Set Screen.
Press ON/OFF/MODE to bypass the clock set process and advance to
the Date Set Screen.
Press ENTER to begin the clock set process.
On the display, the first time value subject to change flashes.
Press ENTER to edit the value, if needed; press repeatedly or hold down
to speed the increment pace.
Clock Set Screen Press ON/OFF/MODE to set the value.
Displays the existing time values using a
24-hour time format. The next value subject to change flashes. Use the ENTER and
ON/OFF/MODE buttons, respectively, to edit the value.
After all values are set, press ON/OFF/MODE and advance to the Date
Set Screen.
Press ON/OFF/MODE to bypass the date set process and advance to
the Display Mode Set Screen.
Press ENTER to begin the date set process
On the display, the first date value subject to change flashes.
Press ENTER to edit the value, if needed; press repeatedly or hold down
to speed the increment pace. Press ON/OFF/MODE to set the value.
Date Set Screen
Displays the existing date. The value The next date value subject to change flashes. Continue to use the
displayed on the far left is the month ENTER and ON/OFF/MODE buttons, respectively, to edit and set each
and to its right the day. The year is value.
displayed beneath the day.
After all values are set, press ON/OFF/MODE and advance to the
Display Mode Set Screen.
Configuration Process
Display and Options Instructions
The display mode selected determines whether the monitor user will see
a numeric or text display (including alarm displays) when the monitor is
in the gas monitoring mode.
Configuration Process
Display and Options Instructions
When enabled, the monitor will sound two beeps every 30 seconds and
its display icons will indicate a bump test is due.
Configuration Process
Display and Options Instructions
When enabled, all monitor users are permitted to zero the monitor from
the gas monitoring mode.
NOTE: The up arrow (▲) will be featured on-screen when the unit is set
to display the number of days before a sensor’s next calibration is due.
The down arrow (▼) will be featured when the unit is set to display the
number of days since the last calibration was performed. A value will be
displayed for each installed sensor.
Calibration Days Set Screen
Press ENTER to edit the value, if needed.
Options
Press ON/OFF/MODE to set the value and advance to the Security
0 = display days since last calibration
Code Set Screen
1 = display days until next calibration
Configuration Process
Display and Options Instructions
A security code value of 000 permits all monitor users to enter
configuration mode and gain access to change the monitor’s settings. A
value other than 000 will restrict access to the configuration mode; it will
also restrict access to the shutdown process for an instrument that is
configured for “always on”.
Press ENTER to edit the value, if needed; press repeatedly or hold
Security Code Set Screen down to speed the increment pace.
Valid values: 000 to 999. Press ON/OFF/MODE to set the value and advance to the Language
Increment value: one Selection Screen.
Configuration Process
Display and Options Instructions
1 = Enable/on
Proper monitor use and service includes everything from bump testing and calibration to keeping the monitor clean,
proper air sampling, and the replacement of parts and components. Beginning with calibration and bump testing, the
following sections provide information and instruction on all use and service tasks.
Procedures
Configuration. The configuration process allows qualified personnel to review and adjust a unit's settings.
Bump Test (or "functional test"). Bump testing checks for sensor and alarm functionality. The installed sensors are
briefly exposed to expected concentrations of calibration gases that are greater than the sensors’ low alarm set
points. When one or more sensors “pass” the test, they are “functional” and the unit will alarm. Each sensor’s “pass”
or “fail” result is indicated on the unit’s display.
Note: a bump test does not measure for sensor accuracy (see “Calibration”).
Zero. Zeroing sets each installed sensor to recognize the ambient air as clean air. If the ambient air is not truly clean
air, any gasses that are present and relevant to the installed sensor types will be measured and displayed as zero.
Readings will be inaccurate until the unit is correctly zeroed in truly fresh air or with a zero air cylinder.
Calibration. All sensors gradually degrade over time. This diminishes a sensor's ability to measure gas concentrations
accurately; however, regular calibrations adjust the instrument to compensate for this decline in sensitivity. During
calibration, the installed sensors are exposed to expected concentrations of calibration gases and, when needed, the
instrument will self-adjust to ensure the accurate measurement and display of gas concentration values.
Note: When a sensor has degraded beyond an acceptable level, no further adjustment is possible and the sensor will no longer
pass calibration.
Peak Readings. The instrument stores the highest detected gas readings, the "peak readings" or "peaks". Bump
testing and calibration will often register new peak readings. Therefore, the clearing of the peak readings should
follow each calibration. The instrument operator may also wish to clear the peak readings after a bump test, before a
change in location, or after an alarm is addressed and cleared.
Note: The peak readings and the data log readings are stored independently of one another; therefore, clearing the peak reading
does not affect the data log. Powering the instrument off or changing its battery does not affect the peak reading. These checks and
balances help promote operator safety, and serve to contain the peak readings in a "black-box" manner. In the event of a gas-
related incident, this black-box record can be useful to the safety team or a prospective investigator.
Recommendations
Industrial Scientific Corporation (ISC) minimum frequency recommendations for each procedure are summarized in
the table below. These recommendations are based on field data, safe work procedures, industry best practices, and
regulatory standards to help ensure worker safety. Industrial Scientific is not responsible for setting safety practices
and policies. These policies may be affected by the directives and recommendations of regulatory groups,
environmental conditions, operating conditions, instrument use patterns and exposure to gas, and other factors.
exposed to an over-range (positive or negative) gas concentration. A calibration is also recommended after the installation of a new
(or replacement) sensor.
b
If conditions do not permit daily testing, bump tests may be done less frequently based on company safety policy.
Note: The use of calibration gases not provided by ISC may void product warranties and limit potential liability claims.
General information
The zero, calibration, and bump testing tasks are in-field enabled or in-field disabled in the configuration process.
This permits or denies access to these functions from the gas monitoring mode. When any of these options is
enabled, it is accessible to all monitor users. In gas monitoring mode, a series of presses on the ON/OFF/MODE
button gives the user access to the following screens and processes in the order shown.
The monitor is capable of performing two types of calibration, and this option is set in configuration mode. The
calibration type selected also determines the monitor’s bump test type. With a “quick” calibration, the monitor is set to
calibrate and bump test all installed sensors simultaneously. With a “standard” calibration setting, these tasks are
completed independently for each installed sensor in the order shown below.
1. Oxygen sensor*
2. Toxic sensor 1
3. LEL sensor
4. Toxic sensor 2
*If set to the default value of 20.9% or 21%, the Oxygen sensor calibrates during the zero process and toxic 1 is the first to calibrate in the
calibration process.
The Ventis MX4 monitor can be calibrated with any of the accessories listed.
• Calibration cup and/or tubing shipped with the monitor (see instructions below)
• V-Cal Calibration Station (consult the calibration station manual for instruction)
• DSX™ Docking Station for Ventis MX4 (consult the docking station manual for instruction)
Instructions
Calibration and Bump Testing with Calibration Cup and/or Tubing
Read all instructions before beginning: notices, supply check-list, gas cylinder preparation, and the complete screen-
by-screen walk-through of the zero, calibrate, and bump test processes. Each process is presented in the order in
which it is accessible from gas monitoring mode.
NOTICES
→ Industrial Scientific recommends that full monitor calibration be performed, using a known certified
concentration(s) of Industrial Scientific calibration gas(es), to prepare the monitor for first time use, and
monthly (at a minimum) thereafter, to help ensure monitor accuracy.
→ Industrial Scientific also recommends that each monitor be zeroed and bump tested before each use with a
known certified concentration(s) of Industrial Scientific calibration gas(es).
→ Read ALL requirements and instructions outlined below, including the screen-by-screen process description,
before beginning the zero, calibration, or bump testing processes.
→ Only qualified personnel should zero, calibrate, or bump test a monitor.
→ Zero, calibration, and bump testing functions should be performed in a fresh air environment known to be
nonhazardous.
→ After calibration or bump testing, or after terminating either process, stop the flow of gas.
Supplies
Item Monitor/Regulator**
Aspirated monitor Aspirated monitor Diffusion monitor
with Demand Flow with Positive Flow with Positive Flow
Regulator** Regulator** Regulator**
Calibration cup* No No Yes
Calibration tubing 2 feet in length* Yes No Yes
Calibration tubing 2 feet in length with No Yes No
integrated “t” fitting
Calibration gas cylinder Yes Yes Yes
Diffusion monitor with positive flow regulator. Aspirated monitor with demand flow regulator.
DO NOT ATTACH THE OTHER END OF THE TUBING TO THE MONITOR BEFORE REACHING THE
“APPLY GAS SCREEN”. Completing the connection of the tubing will cause gas to flow. If gas is
applied before reaching the appropriate screen, the monitor will go into alarm and a failure will be
logged.
B. Aspirated with positive flow regulator
The calibration tubing with the t-fitting (not included) has two different sized openings, a narrow opening
at one end and a wider opening at the other end.
Attach the wider opening to the nipple on the cylinder’s regulator.
Attach the smaller opening to the pump inlet.
DO NOT APPLY THE GAS BEFORE REACHING THE “APPLY GAS SCREEN”. If gas is applied before
that point, the monitor will go into alarm and a failure will be logged.
DO NOT ATTACH THE CALIBRATION CUP TO THE MONITOR OR APPLY THE GAS BEFORE
REACHING THE “APPLY GAS SCREEN”. If gas is applied before that point, the monitor will go into
alarm and a failure will be logged.
NOTE: When zero, calibration, and bump test are ALL in-field enabled, and the user has entered zero from the gas
monitoring mode, the monitor expects to be calibrated following a successful zero.
If the desired task, after zero, is bump testing (or clearing the peaks) and NOT calibration, follow the instructions at
the Calibration Apply Gas Screen to terminate calibration.
Press ENTER to begin the zero process and advance to
the Zero In-process Screen.
Press ON/OFF/MODE to bypass zero and calibration
and advance to one of two screens.
If bump test in-field is enabled, the user advances to the
Bump Test Initiate Screen.
Zero Initiate Screen If bump test in-field is disabled, the user advances to the
Options: Peak Readings Screen.
Enter Zero
Bypass Zero
Zero Results (Pass) Screen …configuration mode, the user advances to the
The check mark displays to indicate a successful zero Calibration Apply Gas Screen.
and a short beep sounds. Each sensor’s numerical
values display at zero except O2. … gas monitoring mode and the calibration in-field option
is enabled, the user advances to the Calibration Apply
Options: Gas Screen.
Repeat zero
Begin calibration …gas monitoring mode and the calibration in-field option
is disabled, the user advances to the Gas Monitoring
Enter gas monitoring mode
Screen in the gas monitoring mode.
To Bump Test
Bump Test Apply Gas Screen
Displays the bump test gas concentrations the monitor From the already prepared gas cylinder, start the flow of
is expecting to receive. The monitor waits up to five gas as noted below for the monitor/regulator combination
in use.
minutes to successfully sense the gas.
NOTE: After a full calibration, the O2 sensor must pass a bump test to
Bump Test Results (Fail) Screen clear the bump test fail status.
The “bF” displays under each gas type to indicate a
bump test failure. The system level alarm turns on and
the gas cylinder icon flashes.
Remote Sampling
Aspirated monitor
In confined space, an air sample should be taken in four foot (1.2192 m) intervals.
Guidelines for using a motorized pump and sampling line
When sampling with a motorized pump and sampling line, Industrial Scientific recommends the following:
• Choose the tubing type based on the target gases. If the target gases are known, use Teflon-lined tubing when
sampling for these gases: chlorine (Cl2), chlorine dioxide (ClO2), hydrogen chloride (HCl), and volatile organic
compounds (VOCs). For other known target gases, urethane tubing or Teflon-lined tubing may be used.
When the target gases are unknown, use Teflon-lined tubing.
• Know the length of the sample line as it is a factor in determining sampling time. A sample line may consist of
tubing, a probe, or a probe and tubing. It should also have a dust filter–water stop installed at the line's end that
will extend into the sample area. Sample-line length is defined as the distance from the dust filter–water stop
opening to the point where the line connects to the pump's inlet. Ensure sample-line length does not exceed the
pump's maximum draw.
• Before and after each air sample, perform a test of the full sampling line.
o Use a thumb to block the end of the sampling line at the water-stop opening. This should cause a pump-fault
alarm.
o Remove the thumb from the water-stop opening. After the alarm cycle completes, the pump should resume
normal operation.
Note: If a pump fault does not occur, check and correct for cracks or other damage, debris, and proper
installation in these areas: all sampling line connections, the pump's inlet cap and inlet barrel, and the dust filter-
water stop items at the end of the sampling line and inside the pump inlet barrel.
• Based on sample-line length, calculate the minimum time recommended for the air sample to reach the
instrument's sensors. As shown below, use a base time of 2 minutes, and add 2 seconds for each 30 cm (1 ') of
line length. Watch the display screen for gas readings and, if present, allow them to stabilize to determine the
reading.
Cleaning
• NEVER use solvents or cleaning solutions of any type.
• When necessary, wipe the outside of the Ventis MX4 with a soft, clean cloth.
• Make sure the sensor diffusion membrane, inside and out, is free of debris; wipe gently with a cloth or brush that
is soft, clean, and dry.
• Make sure the aspirated monitor’s inlet is free of debris.
Service
Instructions are provided for battery service; monitor conversion; sensor, sensor barrier and LCD service; and pump
assembly service. Refer to the three-dimensional view diagrams to identify the parts referenced in the instruction
sets, and for screw torque values.
NOTICES
2. Place AAA battery types Energizer EN92 or Duracell MN2400 so the negative (-)
end of the battery, touches the spring inside the battery housing. Do NOT mix
battery types.
3. When all three AAA batteries are properly installed, close the battery housing. A
“Click” will be heard when the latch catches securely.
4. When replacing batteries, dispose of spent batteries according to company
policy.
Aspirated Monitor
The aspirated monitor can be used with two Ventis battery types.
• The Extended range Li-ion battery
• The Alkaline battery
Power off the instrument before disassembling it or performing any service task.
Pump with battery installation
Unscrew and remove the Unscrew, lift, and remove Loosen the pump door Slide the pump door
belt clip. Store the clip, the battery from the screw. down.
screw, and washer for diffusion instrument; store
future use. it for future use.
Lift door to open. Install a compatible Place the instrument in the Lower the pump door.
Extended range battery— pump case; tighten* the Slide it into its fully
label side up—into the four torx screws on the closed, clicked-shut
lower receptacle of the back of the pump. position.
pump case.
— — —
Loosen the pump door screw. The door is hinged to the pump module with two pegs
Slide the pump door down; lift it to open. that slide into grooves. Angle the door so that one peg
moves to the bottom of its groove and the other moves
the top of its groove. Lift the door to remove it.
Install the new door in the same manner the door was
removed.
— —
Lower the pump door. Slide it into its fully closed, clicked-shut position. Tighten* the pump door screw.
*Torque value is .39 Newton-meters (55 ounce-force inch)
To unscrew and remove Remove the water barrier Place the new water barrier inside the inlet barrel; the
the pump cap, turn it in a from the inlet barrel. side with the larger filter surface should face the user.
counterclockwise
direction.
— — —
Diffusion Monitor
The diffusion monitor can be used with all four Ventis battery types:
• The Li-ion battery
• The Slim extended Li-ion battery
• The Extended range Li-ion battery
• The Alkaline battery
Battery installation
Using a torx screwdriver, loosen all four screws from the Lift the battery (left) or battery cover and Extended
battery (left) or the battery cover (right). range battery (right) away from the instrument.
Note: The Extended range battery will easily fit into the battery cover. If the battery does not easily insert, stop to ensure proper
placement as noted in the instruction.
To install the Extended range battery, first place the To install the battery, Using a torx screwdriver,
battery in the battery cover. When placed correctly, align it with the tighten* each of the four
the battery's label will show. instrument. screws to secure the
battery (shown) or
Next, align the battery cover with the instrument. battery cover to the
instrument.
*Torque value is .39 Newton-meters (55 ounce-force inch).
Suspender clip
When worn, the diffusion monitor should be fastened securely and attached to ensure the sensor portals are exposed
to the air. The monitor should be in full view. No part of the monitor should be covered by any garment or part of a
garment.
Clip replacement
Clip only (use with battery and Slim extended battery)
Lift the clip's cover. To remove the clip, use a To attach the clip, put the washer onto the screw and
Phillips screwdriver to place the screw in the clip's middle hole.
access the clip's screw.
Turn counterclockwise to Turn the screw clockwise to tighten*.
loosen the screw.
Remove the screw,
washer, and clip; set aside
or store for future use.
Clip with spacer (use with Extended battery and battery cover)
To remove the clip, use a To attach the clip and Guide the screw into the
Phillips screwdriver to spacer, cover the case spacer's hole and into the
access the clip's screw. bottom's platform with the instrument case bottom.
Turn counterclockwise to spacer.
loosen the screw. Turn clockwise to tighten*.
Put the washer onto the
Remove the washer, screw and place the screw
screw, clip, and spacer; in the clip's middle hole.
set aside or store for
future use.
*Torque value is .88 Newton-meters (125 ounce-force inch).
Disassembling the Monitor. (For the Diffusion monitor start with Step 6.)
1 Power-off the monitor.
2 Loosen the four captive screws on the lower portion of the pump case module bottom (back of the
module).
3 Loosen the single captive screw on the pump case module top.
4 Loosen the front door screw; slide the case door down; lift the hinged door to reveal and access the
monitor.
5 Lift and remove the monitor from the pump module; set aside the module.
6 Place the monitor display side down. For a diffusion monitor loosen the four captive battery screws to
separate the battery from the case bottom.
7 Loosen the two captive screws on the upper portion of the case bottom.
8 Lift to separate the monitor case top from the monitor case bottom to reveal the circuit board assembly.
9 Remove the circuit board assembly and set aside the monitor case top and bottom.
10 Separate the main circuit board from the sensor board.
Replacing the LCD
1 Grasp the sides of the LCD and lift straight up to remove from the main circuit board.
2 To properly place the new LCD, align the pins on the LCD with their receptacles on the main circuit board.
3 Gently press straight down and into place.
Replacing the Sensor(s)
1 Identify the sensor to be removed.
2 Gently lift and remove the sensor.
3 To add the new sensor, align its pins or connector(s), with the respective receptacles on the sensor board.
4 Press down. A slight click indicates the sensor is securely in place.
Reassembling the Circuit Board Assembly
1 Re-attach the main circuit board to the sensor board, aligning their connectors.
2 Press. A slight click indicates the boards are securely attached.
Replacing the Sensor Barrier or Case Top
To replace the sensor barrier on the inside of the case top, follow steps 1-5 below.
To replace the entire case top, skip to the instruction set, “Reassembling the Monitor”.
1 Observe the placement of the existing sensor barrier. Note that each cut-out is shaped to match the
sensor it protects.
2 Lift and remove the sensor barrier and gasket from inside the monitor case top. Ensure the entire case top
is free of adhesive; gently scrape, if needed. Wipe with a clean, dry, soft cloth or brush.
3 Lift the backing from the new sensor barrier assembly to reveal the adhesive.
4 Carefully position the new barrier. Each shaped opening matches the shape of the sensor it protects.
Press to attach to the inside of the case top.
Replacing the Vibrating Motor
1 Place the monitor’s case top face down.
2 Lift the vibrating motor from its partition. The partition has two sections divided by a ridge. Discard the
used motor.
3 To properly place the new vibrating motor, its contact pins face the user and align with the left edge of the
partition. (The motor’s movable component fits within the small section of the partition.)
4 Press into place.
16 17139262 Screw, Phillips (torque value: 125 oz-in or .88 N.m +/- 10%)
Batteries
Battery Configuration
The base part number that appears on the label of a Ventis battery item uses an eight-digit numeric format
(XXXXXXXX). The corresponding orderable part numbers use the four-letter base reference "VTSB", which is followed
by a three character suffix. The first suffix character is a number that designates the battery type; the second and third
are used to indicate color and approval options, respectively. For example, as shown below in the Battery part numbers
and options, a rechargeable Slim extended lithium-ion battery kit that is black and has a UL approval would have an
orderable part number of VTSB-401 and its label would state a part number of 17157350-01.
a
Color and approval options may vary for each battery item. For more information, contact Industrial Scientific or an authorized distributor of its
products.
b
The battery and cover may be ordered separately using these part numbers 17148313-Y (battery) 17151184-XY (cover).
c
Aspirated instruments only.
Monitor Specifications
Item Description
Display Backlit Liquid Crystal Display (LCD)
Buttons Two (ON/OFF/MODE and ENTER)
Monitor case Polycarbonate with ESD protective rubber overmold
Alarms Ultra-bright LEDs, loud audible alarm (95dB at 30 cm), and vibrating alarm
Size and Weight Diffusion with Lithium-ion Diffusion with Slim Aspirated with Extended
(typical) extended Lithium-ion Range Lithium-ion
(typical) (typical)
Size 103 mm x 58 mm x 30 mm 103 mm x 58 mm x 36 mm 172 mm x 67 mm x 66 mm
(4.1” x 2.3” x 1.2”) (4.1” x 2.3” x 1.4”) (6.8” x 2.6” x 2.6”)
Weight 182 g (6.4 oz) 207 g (7.3 oz) 380 g (13.4 oz)
Battery Specifications
The battery specifications table shown below, includes run time, charge time, charging temperature requirements,
and expected lifetime.
Operating Conditions
Warm-up time 40 seconds (includes stabilization time)
Temperature range -20 °C to +50 °C (-4 °F to +122 °F)
Humidity range 15–95% relative humidity (RH) noncondensing (during continuous operation)
Pressure range 1 atm ±20%
Cold-weather Operation
Use caution when operating the instrument in temperatures below -20 °C (-4 °F), which can diminish display-screen
legibility. To help support functionality and available battery power, the following practices are recommended.
• Do not operate the instrument in temperatures that are not within the temperature ranges of the installed sensors
(see "Sensor specifications").
• Use a compatible, fully charged lithium-ion battery.
• Before using the instrument in the cold-weather environment, power it on a warm-up environment (approximately
20 °C [68 °F]).
• Alternately operate the instrument in the cold-weather and warm-up environments.
• Do not operate the instrument unmanned.
Storage Conditions
Temperature range 0−25 ºC (32−77 ºF)
Humidity range 40–70% relative humidity (RH) noncondensing
Pressure range 0.9–1.1 atm
Maximum time Up to 6 months
Note: Industrial Scientific recommends that infrequently used lithium-ion batteries be
fully charged every four months.
Sensor Specifications
Gas Name Abbr. Measuring Resolution Accuracy at Time and Response Response
Range Temperature of Time Time
Calibration* (typical) (typical)
T50 T90
Oxygen O2 0–30% vol 0.1% vol ± 0.5% 15 30
CO
(H2
CO Low) H2S SO2 NO2 Cl2 ClO2 HCN HCl PH3 NO H2 NH3
CO 100 100 1 1 0 0 0 0 0 0 0 20 0
H2S 5 5 100 1 -40 -3 -25 10 300 25 10 20 25
SO2 0 5 5 100 0 0 0 — 40 — 0 0 -40
NO2 -5 5 -25 -165 100 45 — -70 — — 30 0 -10
Cl2 -10 0 -20 -25 10 100 60 -20 6 -20 0 0 -50
ClO2 — — — — — 20 100 — — — — — —
HCN 15 — 10 50 1 0 0 100 35 1 0 30 5
HCl 3 — 0 5 0 2 0 0 100 0 15 0 0
PH3 — — — — — — -100 425 300 100 — — —
NO 25 40 1 1 5 — — -5 — — 100 30 0
H2 22 3 0.1 0.5 0 0 0 0 0 0 0 100 0
H3 0 0 0 0 0 0 0 0 0 0 0 0 100
The Sensor Cross Sensitivity Table (above) reflects the percentage response provided by the sensor (top row) when exposed to a
known concentration of the target gas (column 1).
The numbers were measured under these environmental conditions: 20 °C (68 °F) , 50% RH and 1 atm.
The specified cross-interference numbers apply to new sensors only, and may vary with time as well as from sensor to sensor.
“—” means no data available.
This table is given as a reference only and is subject to change.
NOTE: The table above provides the LEL for select combustible gases*. It also provides correlation factors that help the safety
technician and instrument operator determine the actual percentage LEL when the sample gas differs from the gas that was used to
calibrate the unit.
For example, if the unit reads 10% LEL in a pentane atmosphere, and was calibrated to methane, the actual percentage LEL is
determined as follows:
1. Locate the table cell where the sample gas (pentane) intersects with the calibration gas (methane).
2. Multiply the cell's value (2.02) by the unit's LEL reading (10%) to calculate the actual concentration of 20.2% LEL.
* The combustible gas list is not a comprehensive list of all combustible gases that can be detected by the Ventis MX4. For
additional information about combustible gas detection and the Ventis MX4, contact the ISC Technical Service department.
Certifications
Certifications for the Ventis™ MX4 Multi-Gas Monitor, at the time of this document's publication, are listed below in
the Hazardous-area certifications table. To determine the hazardous-area classifications for which an instrument is
certified, refer to its label or the instrument order.
Marking Requirements
►Warranty
Industrial Scientific Corporation's Ventis MX4 portable gas monitors are warranted to be free from defects in material and workmanship
under normal and proper use and service for a period of two years from date of purchase. This warranty includes the sensors, the pump,
and the lithium-ion battery as shipped with the Ventis MX4.
Filters are warranted to be free from defects in material and workmanship for 18 months from date of shipment, or one year from
date of first use, whichever occurs first, except where otherwise stated in writing in Industrial Scientific literature.
Limitation of Liability
THE WARRANTY SET FORTH ABOVE IS STRICTLY LIMITED TO ITS TERMS AND IS IN LIEU OF ALL OTHER
WARRANTIES, GUARANTEES, EXPRESS OR IMPLIED, ARISING BY OPERATION OF LAW, COURSE OF
DEALING, USAGE OF TRADE OR OTHERWISE. INDUSTRIAL SCIENTIFIC MAKES NO OTHER WARRANTIES,
EITHER EXPRESSED OR IMPLIED, INCLUDING BUT NOT LIMITED TO THE WARRANTIES OF
MERCHANTABILITY OR FITNESS FOR PARTICULAR PURPOSE.
INDUSTRIAL SCIENTIFIC MAKES NO OTHER WARRANTIES, EITHER EXPRESSED OR IMPLIED, INCLUDING BUT NOT
LIMITED TO THE WARRANTIES OF MERCHANTABILITY OR FITNESS FOR PARTICULAR PURPOSE.
SHOULD THE PRODUCT FAIL TO CONFORM TO THE ABOVE WARRANTY, BUYER’S ONLY REMEDY AND INDUSTRIAL
SCIENTIFIC’S ONLY OBLIGATION SHALL BE, AT INDUSTRIAL SCIENTIFIC’S SOLE OPTION, REPLACEMENT OR REPAIR OF
SUCH NON-CONFORMING GOODS OR REFUND OF THE ORIGINAL PURCHASE PRICE OF THE NON-CONFORMING
GOODS.
IN NO EVENT WILL INDUSTRIAL SCIENTIFIC BE LIABLE FOR ANY OTHER SPECIAL, INCIDENTAL, OR CONSEQUENTIAL
DAMAGES, INCLUDING LOSS OF PROFIT OR LOSS OF USE, ARISING OUT OF THE SALE, MANUFACTURE, OR USE OF
ANY PRODUCTS SOLD HEREUNDER WHETHER SUCH CLAIM IS PLEADED IN CONTRACT OR IN TORT, INCLUDING
STRICT LIABILITY IN TORT.
It shall be an express condition to Industrial Scientific’s warranty that all products be carefully inspected for damage by Buyer upon
receipt, be properly calibrated for Buyer’s particular use, and be used, repaired, and maintained in strict accordance with the
instructions set forth in Industrial Scientific’s product literature. Repair or maintenance by non-qualified personnel will invalidate the
warranty, as will the use of non-approved consumables or spare parts. As with any other sophisticated product, it is essential and a
condition of Industrial Scientific’s warranty that all personnel using the products be fully acquainted with their use, capabilities, and
limitations as set forth in the applicable product literature.
Buyer acknowledges that it alone has determined the intended purpose and suitability of the goods purchased. It is expressly
agreed by the parties that any technical or other advice given by Industrial Scientific with respect to the use of the goods or services
is given without charge and at Buyer’s risk; therefore, Industrial Scientific assumes no obligations or liability for the advice given or
results obtained.
Contact Information