Prosthetic-driven planning
The tooth positions for the planned restoration should be
determined preoperatively, allowing for the selection of the
most appropriate position and angulation for each planned
implant. The existing or the newly fabricated full denture must
be in the correct VDO (vertical dimension of occlusion) and
have the proper AP (anterior-posterior) tooth position in order
to allow fabrication of a surgical guide.
Surgical guide
The prosthetic team should ensure that the surgical team
understands the tooth positions required for the final prosthe-
sis. An appropriate method can be to provide a surgical guide.
A quick and simple way of fabricating such a guide is to make
a replica in clear acrylic resin, either of the existing or the
newly fabricated denture.
NobelClinician
The NobelClinician Software helps the treatment team to
achieve optimized esthetic outcomes, as they can plan the
implant treatment based on the patient’s anatomy and pros-
thetic requirements.
For implants with internal conical connection, Multi-unit Abutments Plus and Temporary Snap Copings
Multi-unit Plus are available. Their snap functionality is designed to simplify the workflow during the try-in
and adjustment of the provisional prosthesis.
Notes:
–– Alternatively, the copings can be attached with the
prosthetic screws during try-in by manually tightening the
screws with the Screwdriver Manual Unigrip.
–– Until the copings are fully secured with the prosthetic
screws, care should be taken that they do not detach from
the Multi-unit Abutment Plus (e.g., through pressure from
the tongue).
–– The Temporary Snap Copings Multi-unit Plus are not
compatible with the conventional Multi-unit Abutments.
Notes:
–– The following illustrations show the conversion of either an existing or a previously fabricated denture
into a fixed provisional full-arch bridge on abutment level following the Immediate Function protocol.
–– This chairside procedure can be applied if limited adjustments have been made to the tissues during
surgery and the denture is forseen to be stable.
–– The fixed provisional bridge can be used for immediate (Immediate Function) or delayed loading of
four or more implants in either jaw with an existing opposing dentition.
Notes:
–– For straight Multi-unit Abutments, tighten the abutment
screw to 35 Ncm using Screwdriver Machine Multi-unit and
Manual Torque Wrench Prosthetic.
–– For angulated Multi-unit Abutments, tighten the abutment
screw to 15 Ncm using Screwdriver Machine Unigrip and
Manual Torque Wrench Prosthetic.
Notes:
–– For implants with internal conical connection, snap the
Temporary Snap Copings Multi-unit Plus onto the Multi-unit
Abutments Plus or manually tighten the prosthetic screws
with the Screwdriver Manual Unigrip.
–– For implants with external hex and internal tri-channel
connection, attach the Temporary Copings Multi-unit by
manually tightening the prosthetic screws with the Screw-
driver Manual Unigrip.
–– Ensure that no soft tissue is trapped between coping and
abutment.
10 Denture conversion manual // Conversion procedure for Immediate Function
11 Remove denture
–– Manually unscrew the prosthetic screws of the temporary
copings with the Screwdriver Manual Unigrip.
–– Remove denture together with luted temporary copings
from patient’s mouth.
–– Finish luting procedure extraorally and polish.
13 Trim denture
In order to complete the conversion of the denture into a fixed
provisional bridge, remove the palatal portion and recontour
the buccal flange. In addition, remove cantilevers that exist
distal to the position of the most posterior implants.
12 Denture conversion manual // Conversion procedure for Immediate Function
ü
14 Finalize provisional bridge
Make sure that the palatal surface of the bridge is convex and
smoothly polished to avoid food impaction and bacteria
accumulation.
Notes:
–– The following illustrations show the conversion of either an existing or a previously fabricated denture
into a fixed provisional full-arch bridge on abutment level with the support of the dental laboratory
following the Immediate Function protocol.
–– This procedure can be applied if adjustments have been made to the tissues and the denture therefore
needs to be adjusted to fill the voids.
–– The fixed provisional bridge can be used for the immediate (Immediate Function) or delayed loading of
four or more implants in either jaw with an existing opposing dentition.
Notes:
–– For straight Multi-unit Abutments, tighten the abutment
screw to 35 Ncm using Screwdriver Machine Multi-unit and
Manual Torque Wrench Prosthetic.
–– For angulated Multi-unit Abutments, tighten the abutment
screw to 15 Ncm using Screwdriver Machine Unigrip and
Manual Torque Wrench Prosthetic.
Notes:
–– For implants with internal conical connection, snap the
Temporary Snap Copings Multi-unit Plus onto the Multi-unit
Abutments Plus or manually tighten the prosthetic screws
with the Screwdriver Manual Unigrip.
–– For implants with external hex and internal tri-channel
connection, attach the Temporary Copings Multi-unit by
manually tightening the prosthetic screws with the Screw-
driver Manual Unigrip.
–– Ensure that no soft tissue is trapped between coping and
abutment.
12 Remove denture
–– Manually unscrew the prosthetic screws of the temporary
copings with the Screwdriver Manual Unigrip.
–– Remove denture together with luted temporary copings
from patient’s mouth.
18 Denture conversion manual // Conversion procedure for Immediate Function
Notes:
–– For implants with internal conical connection, snap the
Temporary Snap Copings Multi-unit Plus onto the Multi-unit
Abutments Plus or manually tighten the prosthetic screws
with the Screwdriver Manual Unigrip.
–– For implants with external hex and internal tri-channel
connection, attach the Temporary Copings Multi-unit by
manually tightening the prosthetic screws with the Screw-
driver Manual Unigrip.
–– Ensure that no soft tissue is trapped between coping and
abutment.
Denture conversion manual // Conversion procedure for Immediate Function 19
Notes:
–– For implants with internal conical connection, snap the
Temporary Snap Copings Multi-unit Plus onto the Multi-unit
Abutments Plus or manually tighten the prosthetic screws
with the Screwdriver Manual Unigrip.
–– For implants with external hex and internal tri-channel
connection, attach the Temporary Copings Multi-unit by
manually tightening the prosthetic screws with the Screw-
driver Manual Unigrip.
–– If needed, reduce the length of the posterior temporary
copings to eliminate any occlusal interferences.
17 Remove denture
–– Manually unscrew the prosthetic screws of the temporary
copings with the Screwdriver Manual Unigrip.
–– Remove denture together with luted temporary copings
from patient’s mouth.
20 Denture conversion manual // Conversion procedure for Immediate Function
Notes:
–– Make sure that the palatal surface of the bridge is convex
and smoothly polished to avoid food impaction and bacteria
accumulation.
–– Avoid or minimize any cantilever on the provisional bridge.
ü
Denture conversion manual // Conversion procedure for Immediate Function 21
Multi-unit Abutments
External hex Internal conical Internal tri-channel
connection connection connection
Brånemark System® and NobelActive®, NobelParallel™ and NobelReplace®, Replace Select™,
NobelSpeedy® Groovy NobelReplace® Conical Connection NobelSpeedy® Replace,
NobelReplace® Platform Shift
NP RP WP NP RP WP NP RP WP 6.0
1 mm
1 mm
Multi-unit Abutment 1
mm 29176 29179 29184 – – – 29196 29199 29204 –
1.5 mm
2 mm
2 mm
Multi-unit Abutment 2
mm 29177 29180 29185 – – – 29197 29200 29205 –
2.5 mm
3 mm
3 mm
Multi-unit Abutment 3
mm 29178 29181 29186 – – – 29198 29201 29206 –
3.5 mm
4 mm
4 mm
Multi-unit Abutment 4
mm – 29182 – – – – – 29202 – –
4.5 mm
5 mm
5 mm
Multi-unit Abutment 5
mm – 29183 – – – – – 29203 – –
0.5 mm
NP RP WP NP RP WP NP RP WP 6.0
2 mm
2 mm
2.5 mm
3 mm
3 mm
3.5 mm
4 mm
4 mm
3.5 mm
4 mm
4 mm
4.5 mm
5 mm
5 mm
NP RP WP NP RP WP NP RP WP 6.0
Abutment Screw Multi-unit Angled 29194 29195 – 36892 37893 37893 29242 29243 – –
Prosthetic Screw Multi-unit 29285 29285 29286 29285 29285 29285 29285 29285 29285 –
Healing Cap Multi-unit (1/pkg) 31145 31145 29066 31145 31145 31145 31145 31145 31145 –
Healing Cap Multi-unit (5/pkg) 29064 29064 – 29064 29064 29064 29064 29064 29064 –
Healing Cap Wide Multi-unit (1/pkg) 31146 31146 29067 31146 31146 31146 31146 31146 31146 –
Torque guide for clinical screws for Nobel Biocare implant systems
Prosthetic components
External hex Internal conical Internal tri-channel
connection connection connection
Brånemark System® and NobelActive®, NobelParallel™ and NobelReplace®, Replace Select™,
NobelSpeedy® Groovy NobelReplace® Conical Connection NobelSpeedy® Replace,
NobelReplace® Platform Shift
NP RP WP NP RP WP NP RP WP 6.0
Abutment Replica Multi-unit (1/pkg) 38918 38918 31162 38918 38918 38918 38918 38918 38918 –
Abutment Replica Multi-unit (5/pkg) 38919 38919 – 38919 38919 38919 38919 38919 38919 –
NP RP WP NP RP WP NP RP WP 6.0
Lab Screw Multi-unit (5/pkg) 29287 29287 – 29287 29287 29287 29287 29287 29287 –
Lab Screw Multi-unit Angled (1/pkg) 37896 37897 – 37896 37897 37897 31166 31167 – –
Protection Analog Multi-unit (5/pkg) 29123 29123 29124 29123 29123 29123 29123 29123 29123 –
Kit includes
Screwdriver Machine Unigrip™ 20 mm 29151
Screwdriver Machine Unigrip™ 30 mm 29153
Screwdriver Machine Multi-unit 21mm 29158
Omnigrip™ Screwdriver Machine 20 mm 37379
Omnigrip™ Screwdriver Machine 30 mm 37381
Manual Torque Wrench Prosthetic 29165
Screwdrivers
Screwdriver Manual Unigrip 36 mm 29150
Nobel Biocare Canada Nobel Biocare Belgium Phone: +27 11 802 0112
Dental Biocare Nobel Biocare Denmark Phone: +34 900 850 008
Nobel Biocare Mexico Nobel Biocare France Phone: 0800 211 424
Nobel Biocare USA Nobel Biocare Germany Phone: +44 208 756 3300
Hungary
Distributor markets
Nobel Biocare Hungary
Asia Pacific Phone: +36 1 279 33 79 Algeria, Bulgaria, Croatia, Cyprus,
nobelbiocare.com/contact
v 16.1
nobelbiocare.com
80567 GB 1609 © Nobel Biocare Services AG, 2016. All rights reserved. Nobel Biocare, the Nobel Biocare logotype and all other trademarks are, if nothing else is stated or is evident from the context in a certain
case, trademarks of Nobel Biocare. Please refer to nobelbiocare.com/trademarks for more information. Product images are not necessarily to scale. Disclaimer: Some products may not be regulatory cleared/
released for sale in all markets. Please contact the local Nobel Biocare sales office for current product assortment and availability. For prescription use only. Caution: Federal (United States) law restricts this device
to sale by or on the order of a licensed dentist. See Instructions for Use for full prescribing information, including indications, contraindications, warnings and precautions.