Perfusore Space Manual
Perfusore Space Manual
Perfusore Space Manual
and Accessories
P E R F U S O R ® S PA C E O V E R V I E W
Arrow up and -down
Scroll through menus, change setting of numbers from Press to reset single values Drive head with
0-9, answer Yes/No questions. to zero and switch back to claws to hold the
Arrow left and -right the previous screen/menu syringe plunger
Select data from a scale and switch between digits when level. plate and
emergency re-
c
numbers are entered. Open a function while pump is
running or stopped with the left arrow key. lease button.
q Press to
initiate Press to turn
Yellow LED: Pre-alarm, reminder alarm
bolus. pump on/off.
Green / Red LED: Infusion occuring / device alarm,
B o
n
operating alarm
Blue LED: Currently connected to SpaceControl
k m
Open certain functions Press to link the pump to
f
S Syringe holder locks syringe
and press to confirm SpaceControl and to assign Press to Start/Stop in position. The drive will
values/settings/alarms. a barcode after scanning. infusion. automatically move back.
3
PERFUSOR® SPACE OVERVIEW
Syringe Fixation
Pull and turn the syringe holder to the right
to open the green axial fixation (see red
arrow). Syringe must be fixed with wings
upright in the slot (found to the left hand
side of the axial fixation) before closing
syringe holder. Make sure that syringe is
properly inserted.
Caution: Don't touch piston brake when
moving forward.
Transport
A maximum of three pumps (Perfusor®
Space or Infusomat® Space) plus one
SpaceControl may be stacked together (in
ambulance cars or helicopters only one
pump). Avoid external mechanical influence.
Locking Devices Together
Line up the bar of the lower pump with the
bar of the pump above and slide the lower
pump backwards until the lock clicks and
the green buttons are above each other.
To disconnect, push green locking buttons
of top pump device and slide bottom pump
forward.
Pole Fixation
Push the opening of PoleClamp against the
vertical pole and lock the screw tightly.
Unscrew to release.
For vertical fixation of PoleClamp push
lever down and rotate either way until lever
clicks into notch. Push lever for rotation.
Caution: Do not lean on pump when
attached to pole!
4
SYMBOLS ON PRODUCT
SYMBOLS ON PRODUCT
Symbol Explanation
Caution, general warning symbol
W Caution, see documentation supplied
with the product
Type CF unit with defibrillation
protection
5
PATIENT SAFETY
PAT I E N T S A F E T Y
w
Intended use
The Perfusor® Space Infusion Syringe Pump System includes an external trans-
portable electronic infusion syringe pump and pump accessories. The system is
Read Instructions
intended for use on adults, pediatrics, and neonates for the intermittent or con-
for Use prior to use.
tinuous delivery of parenteral and enteral fluids through clinically accepted
The infusion device
should only be routes of administration. These routes include, but are not limited to intra-
used by specially venous, intra-arterial, subcutaneous, epidural, and enteral. The system is used
trained staff. for the delivery of medications indicated for infusion therapy including but not
limited to drugs like anesthetics, sedatives, analgesics, catecholamines, antico-
agulants etc.; blood and blood components; Total Parenteral Nutrition (TPN);
lipids, and enteral fluids. The Perfusor® Space Infusion Syringe Pump System is
intended to be used by trained healthcare professionals in healthcare facilities,
home care, outpatient, and medical transport environments.
Using TCI the scope of patients is:
Minimum Maximum
Weight [kg] 30 200
Height [cm] 130 220
Age [Yrs] 16 100
Some parameter sets are using the Lean Body Mass (LBM) to individualize the
parameterization. The LBM calculation may furthermore restrict the scope of
patients as it will not allow TCI for obese patients.
Using TCI the scope of procedures is:
• Propofol: Anaesthesia and Conscious Sedation
• Remifentanil: Anaesthesia
The medical specialist must decide on suitability for application on the basis of
the warranted properties and the technical data.
For further details please refer to the Instructions for Use.
Operation
• The initial training of the Perfusor® Space is to be performed by B. Braun
sales personnel or other authorized persons.
After each software update, the user is required to inform himself of the
changes to the device and accessories by referring to the Instructions for Use.
• Ensure the unit is properly positioned and secured. Do not position pump
unit above patient.
• Prior to administration, visibly inspect the pump and the accessories
(especially the axial fixation) for damage, missing parts or contamination and
check audible and visible alarms during selftest.
6 6
PATIENT SAFETY
• Only connect to patient once the syringe has been inserted correctly and
there is proper fixation of the syringe pressure plate by the claws of the drive
head. Interrupt connection during syringe change to prevent incorrect dose
delivery.
• Select syringe/catheter suitable for use with the intended medical application.
• Position the infusion line free of kinks.
• Recommended change of disposable each 24 h (or as per national hygiene
regulations).
• Installation in medically used rooms must comply with the appropriate
regulations (e.g. VDE 0100, VDE 0107 or IEC-publications). Observe national
specifications and deviations.
• Do not operate the pump in the presence of flammable anaesthetics to
prevent explosion.
• Compare the displayed value with the entered value. Start infusion only if
the values are corresponding.
• If staff call is used we recommend checking the equipment once after
connecting the pump.
• Protect the device and the power supply against moisture.
• Do not carry the pump device by it's drive mechanism during transportation.
• If the pump device falls or is exposed to force it needs to be examined by the
service department.
• The displayed data must always be checked by the user prior to making
further medical decisions.
• During mobile use (homecare, patient transport inside and outside the
hospital): Make sure the device is securely fixed and positioned. Positioning
changes and severe shock can lead to minor changes in the delivery
accuracy.
• A supplemental patient monitoring must be carried out if life-saving
medication is performed.
• Avoid applying external force on the drive mechanism during administration.
• In case high potent drugs are given be sure to have a second infusion pump
for that drug at hand. The therapy documentation should be suitable to
continue the therapy at the second infusion pump.
• Independant of the soft limits the selected values have to be the medically
correct ones for the given patient.
• In case values relevant for the dose rate calculation (e.g. body weight) are
changing always the flow rate will be updated and the dose rate will be fix.
7
PATIENT SAFETY
Other components
• Only use pressure resistant tubes (min. 2 bar/1500 mmHg).
• Where several infusion lines are connected on one single vascular access, the
possibility of the lines exerting a mutual influence over each other cannot be
excluded.
• Refer to the according manufacturer’s information for possible
incompatibilities of equipment with respect to drugs.
• Use only compatible combinations of equipment, accessories, working parts
and disposables with luer lock connectors.
• The use of incompatible disposables may influence the technical
specifications of the device.
• Connected electrical equipment must comply with the relevant IEC/EN-
specifications (e.g. IEC/EN 60950 for data-processing equipment).
The user/operator is responsible for the system configuration if additional
equipment is connected. The international standard IEC/EN 60601-1-1 has to
be taken into account.
Safety Standards
Perfusor® Space satisfies all safety standards for medical electrical devices in
compliance with IEC/EN 60601-1 and IEC/EN 60601-2-24.
• The EMC-limits (electro-magnetic compatibility) according to
IEC/EN 60601-1-2 and IEC/EN 60601-2-24 are maintained. If the equipment
is operated in the vicinity of other equipment which may cause high levels of
interference (e.g. HF surgical equipment, nuclear spin tomography units,
mobile telephones etc.) maintain the recommended protective distances from
these devices.
• The Perfusor® Space fulfils the applicable requirements of EN 13718 to be
used in the air, on the water and in difficult terrain. During transport the
Perfusor® Space needs to be fixed on a suitable restraint system by means of
SpaceStation or Pole Clamp SP. When stored under temperature conditions
beyond the defined operating conditions the Perfusor® Space needs to
remain under room temperature at least one hour before usage.
All display screen shots are examples and may be different when related to an
individual patient and individualized therapy.
Display Meaning
10
MENU STRUCTURE / NAVIGATION
Display Meaning
11
OPERATION
Chapter 1
OPERATION
1.1 Start of Infusion
• Ensure correct installation of the pump device. If the pump is connected to mains,
the display states information such as the battery status, the mains connection
symbol and the last therapy.
• Press o to switch on unit. Note the automatic selfcheck: “Selftest active” and the
software version are displayed, two audible tones sound and all three LEDs (yellow,
green/red and blue) flash once. Information on power supply (battery or mains
connection), the set pressure level and the syringe (if syringe already inserted) are
displayed. Hence the drive moves backwards.
The pump offers the possibility to load up to four languages into the pump
(depending on the number of the language specific characters), among which the
user can choose during the operation of the pump. During the first ever start-up of
the device, the user is requested to select the languages and to mark them with l.
After that, the selection has to be confirmed by choosing the last menu item at the
bottom of the list and pressing k. Then the desired language must be selected with
t and confirmed with k. Answer the following question with d in order to
activate the selected language.
• Press c to start with direct entry of therapy parameters or open pump cover and
syringe holder to start with syringe insertion.
• Insert syringe with wings of the syringe upright in the slot to the right of the
housing. Close syringe holder and pump door. Piston brake moves forward.
Caution: Never leave the pump unattended during syringe loading.
• Confirm syringe type with k. Type of syringe indicated must coincide with syringe
inserted.
• Drive will advance and grip pressure plate of syringe.
Caution: Keep your hands away from advancing device.
Note: Make sure that the piston brake moves back into the syringe holder.
• If the prime function is activated, press U to prime infusion set at 1200 ml/h
(pressing key once = 1 ml). Interrupt prime function with k. Repeat procedure
until infusion line is fully primed. Then press d to proceed.
• Connect with patient.
• Respectively answer questions in Start Up Menu with u and d, until the rate is
displayed in the Main Menu.
Enter infusion rate:
• Press l and set rate using q.
12
OPERATION
Chapter 1
• Press s
f to commence infusion. Running arrows on display and green LED
above display indicate pump is infusing.
Note: Stop the infusion at any time by pressing s
f. The pump can be turned off
at any time by pressing o for 3 sec (Exception: Data lock level 2) and as long a
disposable is inserted.
13
OPERATION
Chapter 1
Note: A new therapy can be started at any time during a stopped infusion. Press
c (repeatedly) when the pump is in the Main-, Status- or Options Menu and
proceed to follow instructions as described.
15
OPERATION
Chapter 1
16
ADVANCED OPERATIONS
Chapter 2
ADVANCED OPERATIONS
2.1 Status Request of Pump when Infusion is Running
Press c to switch between run display and Main Menu while the device is infusing.
Navigate through the menu using t to check parameters. In order to check the menu
parameters in the Status-/Options Menu, select "Status" respectively "Options" in the
Main Menu, open menu with l and scroll through menu with t.
17
SPECIAL FUNCTIONS
Chapter 3
SPECIAL FUNCTIONS
Caution: A change of the patient weight or height will alter the flow rate.
19
SPECIAL FUNCTIONS
Chapter 3
Loading a drug (including the according parameters) from the Main Menu:
• Go to Special Functions Menu and press l.
• Open the drug library by pressing l.
• Navigate through the list with t and select the Care Area with l. If you
have already set the Care Area once on your pump this step will be skipped for
the next time.
• Change the Care Area by navigating through the list until "Change Care
Area" will be displayed. Press k to change the Care Area.
• Navigate through the list with t and select the patient profile with l.
If no profile is set, this step will be skipped.
• Navigate through the list with t and select in alphabetical order (all drugs)
or within a category with l.
• If different therapies are related to a drug, choose therapy type with t and
confirm with l.
• Confirm the displayed drug information with l.
• Check if the drug short name in the Main Menu is the same as the selected
drug. Check the parameter in the Main Menu with t and start infusion
with s
f.
Note: Care Area and Patient Profile can not be changed within a therapy.
Initial Bolus:
Initial Bolus has to be configured in the Drug List Manager.
• Use the drug library according to the instructions for use.
• Select the desired drug with t and press l.
Before the initial bolus begins, the bolus menu is displayed to allow editing
the bolus with q.
• Check the parameter and start infusion with s
f.
Hard Limits:
If the set rate/dose/bolus volume and bolus rate exceed the values stored in the
drug library (hard limits), the drug will be rejected, a hint will be displayed and
the pump will fall back into the drug selection. If this occurs while the pump is
infusing the pump will continue to administrate.
Soft Limits:
For the same parameters so called soft limits can be preset via the Drug List
Editor. These can be exceeded without any constraint. The following symbols
that describe the status with regard to the soft limits are being displayed:
The infusion is within the range of the minimum and maximum
soft limits =
20
SPECIAL FUNCTIONS
Chapter 3
Remote Drug Library update from Upload Manager (Space Online Suite)
The file icon blinks every 2 s. An update is available.
The Drug Library Upload starts as soon as the pump is in Passive mode.
Note: You can cancel the upload by pressing c.
Please contact your local sales represantative in case you like to use Remote
Drug Library update.
21
OPTIONS
Chapter 3
After the selection the pump
offers additional drug related
information which are confirmed
by l.
Select profile PCA by using and
confirm with t .The therapy set-
tings stored in the drug list are
displayed *.
The therapy can be started now with s
f in case all values are defined.
Depending on the pre-defined settings the therapy is started with an initial
bolus and a basal rate or not.
Before leaving the patient the pump should be put into DataLock level 3 with
q in Menu “Options”. This is necessary especially in case non-authorised access
to the settings can be anticipated.
The code is entered with q and confirmed with k.
The pump display now may look
like this.
In this state the patient is allowed to demand boli. Depending on the status of
the therapy these are either administered or denied. Changing the syringe is also
possible by using the code for level 1 or level 2. Altering the settings for PCA or
other therapies however is only possible with the code for level 3.
The status of the therapy can be checked in the menu „Status“ .
Enter the „Main Menu“ with c and select the “Status“ with q.
22
SPECIAL FUNCTIONS
Chapter 3
23
SPECIAL FUNCTIONS
Chapter 3
24
SPECIAL FUNCTIONS
Chapter 3
TCI with B. Braun Space is possible with two drugs: Propofol and Remifentanil.
For Propofol the user can choose between two parameter sets. The parameter
sets used for these drugs are (Not all parameter sets allow effect-site targeting):
25
SPECIAL FUNCTIONS
Chapter 3
Drug List
The pre-installed drug list offers the following values:
Propofol Remifentanil
Available Concentrations 5 mg/ml 20 µg/ml
10 mg/ml 50 µg/ml
20 mg/ml
Short name TCIProp TCIRemi
Default Max. Rate 1.200 ml/h 1.200 ml/h
Hard Limit Rate Max of pump Max of pump
Plasma Limit Default 400 % 400 %
Plasma Limit Hard Low 100 % 100 %
Plasma Limit Soft Max 450 % 450 %
Default Target 0.0 µg/ml 0.0 ng/ml
Target Soft Max 8.0 µg/ml 8.0 ng/ml
Target Hard Max 15.0 µg/ml 20.0 ng/ml
Decrement Concentration 1.0 µg/ml 1.0 ng/ml
Default
Default Parameter Set Marsh Minto
Important note: Before installing an additional drug list please contact your local
B. Braun representative!
Selecting a drug
Select drug list, category (the TCI drugs need to be selected from the category “TCI”)
and desired drug by using q .
26
SPECIAL FUNCTIONS
Chapter 3
These steps are only necessary in case there are different options for that drug.
The editing window appears with the initial setting “0” to make sure editing a value
takes place (exemption: initial setting for gender is “male”).
Using effect-site targeting the weight may be limited due to the constraints of the
LBM calculation.
Important notes:
• Be sure to enter the data corresponding to the respective patient.
• Once the TCI is started patient data can not be altered!
27
SPECIAL FUNCTIONS
Chapter 3
Mode-Indicator
Drug short name
In the top line there is an icon indicating the parameter set and the mode (Mode
Indicator) with following meaning:
• “TCI Ma P”: TCI Marsh plasma targeting
• “TCI Sc P”: TCI Schnider plasma targeting
• “TCI Sc E”: TCI Schnider effect-site targeting
• “TCI Mi P”: TCI Minto plasma targeting
• “TCI Mi E”: TCI Minto effect-site targeting
In the bottom line the status parameters like flow rate, Cp/Ce, infused volume etc. can
be displayed. The desired parameter can be selected by using t . It is recommended to
select Cp/Ce.
In case it is necessary to change the target press l to edit the value.
By pressing r additional
information can be requested.
The line describes the course of Cp over the time and the area describes the course of
Ce over the time. The time window is 20 min (15 min past, 5 min future).
Additional information is left with C.
28
Finishing TCI
There are two possibilities to finish the TCI Therapy (reversion of anaesthesia or
sedation):
• Set Target= 0
• Stop pump
3.6 Barcoding
Software version J has the barcoding functionality included but initially not activated.
Please contact your local sales representative in case you like to use
barcoding.
29
SPECIAL FUNCTIONS
Chapter 3
Example:
30
SPECIAL FUNCTIONS
Chapter 3
Starting Ramp and Taper via Drug Library:
Note: Ramp and Taper settings have been configured in the Drug List Manager
before and have been uploaded into the pump.
• Switch on pump with o and wait until self-check is finished.
• Insert disposable and use the drug library according to the Instructions for
Use.
• Select the desired drug with t and press l.
The pump now lits the possible therapy profiles.
• Select “Ramp and Taper Mode” with t and press l.
The therapy settings for “Ramp and Taper Mode” are shown on the display.
• To change the values, press l to change and k to confirm.
The pump can be started now by pressing s
f.
Ramp phase
The pump now linearly
increases the rate in the
predefined time until it
reaches the plateau rate and
then automatically switches
to continuous phase.
Continuous phase
The pump continuously infuses
the same rate for a predefined
time and then automatically
switches to taper phase.
31
SPECIAL FUNCTIONS
Chapter 3
Taper phase
The pump linearly decreases the
rate in the predefined time
until it reaches the KVO rate
Note: After starting infusion it is only possible to change rates, time and VTBI in
the continuous phase.
By editing (increasing/decreasing) the plateau rate, the therapy is recalculated. With
the increase/decrease of the plateau rate the volumes in the ramp phase, the conti-
nuous phase and the taper phase are increased/decreased. The continuous phase is
shortened/prolonged to infuse the VTBI still completely with the end of the taper
phase.
By editing the Ramp/Taper-Time, the therapy is recalculated. The Continuous Phase is
extended/shortened to infuse the VTBI still completely until the end of the Taper
phase.
By increasing/decreasing the VTBI, the continuous phase is prolonged/shortened to in-
fuse the new entered VTBI completely with the end of the taper phase.
Note: The delivery of drugs can be stopped and started again in Ramp and Taper
Mode at any time by pressing s f. Ramp and Taper is stopped immediately
without Taper phase and started without a new Ramp phase. This will not have
any effect on the settings of the therapy.
32
SPECIAL FUNCTIONS
Chapter 3
Program Mode should only be performed by an experienced user being familiar with
the principles of the Program Mode function and properly trained in using the present
device.
Note: The active Program Mode function always displays this icon in the Display
( ).
Note: Bolus function is disabled for Program Mode.
Set Profile Parameters: The therapy can be started directly via the drug library or via
the Main Menu/Special functions.
Starting Program Mode via Drug Library:
Note: Program Mode settings have been configured in the Drug List Manager
before and have been uploaded into the pump.
• Switch on pump with o and wait until self-check is finished.
• Insert disposable and use the drug library according to Instructions for Use.
• Select the desired drug with t and press l.
• Select Program Mode with l.
33
SPECIAL FUNCTIONS
Chapter 3
In the following screen the user has to confirm the number of steps for the
therapy with k.
The settings for the steps of the infusion are shown on the display. These
settings, configured in the Drug List Editor, need to be confirmed with r.
• To change the values, press l to change and k to confirm.
• Adjust VTBI with q.
The pump can be started now by pressing s
f.
In the upper part of the display the icon for “Program Mode” appears.
The screen shows the following:
The pump infuses the
predefined rate in the
predefined time for the current
step.
34
Number of cycles is defined by VTBI. Take care to set the VTBI in the correct
relation to the volume of one Cycle. VTBI may needs to be adjusted after
changing the intervals.
The Main menu informs about the current interval. The configerd paramters can
be checked by Program Parameter Menu in Main.
Example:
35
SPECIAL FUNCTIONS
Chapter 3
In the upper part of the display the icon for “Intermittent Mode” appears.
After the bolus phase the pump switches to rate phase and the screen shows
the following:
Note: To cancel bolus infusion in the Intermittet Bolus therapy at any time it is
only possible with s
f.
Note: The delivery of drugs can be stopped and started again in the Intermittent
Mode at any time by pressing sf. During infusion it is possible to change the
bolus volume, amount, VTBI as well as the time interval.
• Press C.
36
SPECIAL FUNCTIONS
Chapter 3
• Use t to navigate through the parameter list and select the parameter to
be changed with l.
• Enter the new value and press k.
The pump continues infusion.
Note: The KVO function and Bolus function are disabled during Dose Over Time.
Note: The feature Dose Over Time always requires the usage of dosing units (i.e.,
mg or mg/kg patient weight).
Before using Dose Over Time contact your local B. Braun representative!
Starting Dose Over Time via Drug Library:
Note: Dose Over Time settings have been configured in the Drug List Manager
before and have been uploaded into the pump.
• Switch on pump with o and wait until self-check is finished.
• Insert disposable and use the drug library according to the Instructions for
Use.
• Select a drug by using t and press l.
The pump now offers the possible therapy profiles. Select “Dose over Time” with t and
press l.
The editor for Total Dose is shown if a drug with therapy Dose over Time is sel-
ected from drug library and no default value for Total Dose was entered in libr-
ary. The editor is also shown if the Total Dose is edited in the Main menu.
Enter the total dose, if necessary,
and confirm with k.
The editor for Time is shown if a drug with therapy Dose over Time is selected
from drug library and no default value for Time was entered in library. The editor
is also shown if the Time is edited in Main Menu.
Enter the time, if necessary,
and confirm with k.
38
SPECIAL FUNCTIONS
Chapter 3
Note: It is always possible to press the key C in the Run Menu and edit or
check values in the Main Menu while the pump is delivering.
39
OPTIONS
Chapter 4
OPTIONS
The options functions may be selected and changed while the pump is infusing or
stopped. To edit a menu item, select “Options” in the Main Menu and press l.
Then select desired function with t and follow the Instructions for Use as described.
40
OPTIONS
Chapter 4
4.4 KVO-Mode
After reaching a preselected VTBI/time, the pump can continue the infusion with a
predefined KVO-rate (see "Technical Data"). The duration of the KVO-infusion is set via
the service program.
41
OPTIONS
Chapter 4
42
OPTIONS
Chapter 4
4.9 Language
This function enables a change of the pump language.
• Open language in the Options Menu with l.
• Select language with t, then press l.
• Confirm Yes/No with u.
43
ALARMS
Chapter 5
ALARMS
The Perfusor® Space is equipped with a audible and optical alarm signal.
When a device alarm occurs the infusion is immediately stopped. Press O to switch
off the device. Then switch the device on again. In case of a repeated device alarm you
must disconnect from the patient, open the front door of the pump and take out the
disposable. The device needs to be handed to the service department.
44
ALARMS
Chapter 5
45
ALARMS
Chapter 5
46
ALARMS
Chapter 5
“Data Lock" An attempt was made to stop or switch the pump off
without entering the code.
Enter the correct code in order to continue the
therapy or in order to turn the pump off.
The red LED doesn’t extinguish until the administration is started again respectively the
pump is turned off.
Caution: If a wrench is displayed and/or a yellow, red and blue LED blink then the
pump is in the service mode and is not permitted to be used on a patient. The pump is
then to be checked by a service technician.
47
BATTERY OPERATION AND MAINTENANCE
Chapter 6
48
BATTERY OPERATION AND MAINTENANCE
Chapter 6
Caution: Batteries may explode or leak if they are opened or incinerated. Consider
disposal directions!
Battery maintenance:
To accurately balance the battery capacity a cyclical battery maintenance is necessary.
The pump asks the user to perform a battery maintenance every 30 days. The battery
maintenance mode detects a possible capacity loss (e.g. through ageing of the battery
pack) and then the capacity/running time will be calculated anew. After a longer stora-
ge time or a longer operation without battery maintenance it can happen that the
battery pre-alarm time can no longer be maintained. In this case it is necessary to per-
form a battery maintenance.
To initiate the discharge process the message „Battery maintenance“ and the k-key
will be displayed after switching the pump off. By pressing k and u the discharge
process will start. The process is interrupted by switching the pump on again. If the
battery maintenance is to be continued a new activation is necessary. After completely
discharging the battery it will be completely charged again. The total duration of the
battery maintenance process takes approx. twelve hours.
Caution: Please take into account that, if the battery maintenance has not been
completed there is a possibility of a reduced battery operating time.
49
COMPATIBLE SYRINGES
Chapter 7
COMPATIBLE SYRINGES
The syringe types listed in the following tables can be used with the Perfusor® Space.
Please refer to the listed material number (Mat. No.1)) to ensure specific syringe brand
compatibility.
The Time to Occlusion2) alarm has been measured at 5 ml/h. The measured data are
typical values which may vary because of possible syringe tolerances.
B. Braun has no influence on the quality of disposables from other suppliers. Changes
in quality may have an impact on the technical data of the pumps. B. Braun is not
liable for such changes. In these cases please contact the respective supplier.
Manufacturer:
B. Braun
Syringe Type Omnifix Omnifix Omnifix Omnifix Omnifix
B. Braun 2 ml 5 ml 20 ml 30 ml 50 ml
Mat. No.1) 461 7029V 461 7053V 461 7207V 461 7304F 461 7509F
Time to Occl.2) typ. typ. typ. typ. typ.
P1 [mm:ss] 0:39 0:58 1:04 1:13 1:16
P9 [mm:ss] 1:05 1:32 3:26 6:07 13:46
Manufacturer: Manufacturer:
B. Braun B. Braun
Syringe Type OPS OPS Syringe Type Omnifix Omnifix
B. Braun 20 ml 50 ml B. Braun 3ml3) 10ml LL
Mat. No.1) 872 8615 872 8810F Mat. No.1) 4617022V 4617100V
Time to Occl.2) typ. typ. A/P 4617022V-03 A/P 4617100V-03
P1 [mm:ss] 1:08 1:34 US 4610303V-02 US 4617100V-02
P9 [mm:ss] 4:35 15:27 Time to Occl.2) typ. typ.
P1 [mm:ss] 0:25 0:53
P9 [mm:ss] 1:43 3:50
Manufacturer:
TYCO EU
Syringe Type Monoject Monoject Monoject Monoject Monoject Monoject
TYCO EU 3 ml 6 ml 12 ml 20 ml 35 ml 50/60 ml
Mat. No.1) 1100- 1100- 1100- 1100- 1100- 1100-
603495 606159 612173 620036 635430 650090
Time to Occl.2) typ. typ. typ. typ. typ. typ.
P1 [mm:ss] 0:51 0:56 1:04 1:19 1:32 2:23
P9 [mm:ss] 1:16 1:41 3:27 5:27 12:05 15:58
50
COMPATIBLE SYRINGES
Chapter 7
Manufacturer:
TYCO USA
Syringe Type Monoject Monoject Monoject Monoject Monoject Monoject
TYCO USA 3 ml 6 ml 12 ml 20 ml 35 ml 50/60 ml
Mat. No.1) 8881- 8881- 8881- 8881- 8881- 8881-
513934 516937 512878 520657 535762 560125
8881- 8881- 8881- 8881-
713005 716008 712023 760089
Time to Occl.2) typ. typ. typ. typ. typ. typ.
P1 [mm:ss] 0:41 0:50 1:07 1:13 1:27 1:35
P9 [mm:ss] 1:17 2:07 3:45 4:49 11:50 15:46
Manufacturer:
Becton
Dickinson
Syringe Type Plastipak Plastipak Plastipak Plastipak Plastipak Plastipak
B-D EU/USA 3 ml 5 ml 10 ml 20 ml 30 ml 50/60 ml
Mat. No.1) 309585 309603 309604 309661 309662 309663
300910 300911 300912 300913 300863 300865
300134 300869
300629
Time to Occl.2) typ. typ. typ. typ. typ. typ.
P1 [mm:ss] 0:53 0:55 1:15 2:05 2:14 2:53
P9 [mm:ss] 1:15 1:34 3:27 6:30 6:36 15:34
Manufacturer:
Becton
Dickinson
Syringe Type Plastipak BD BD BD BD BD
B-D EU/USA BD 30 ml Luer Lock Luer Lock Luer Lock Luer Lock Luer Lock
3 ml A/P3) 5 ml A/P3) 10 ml A/P3) 20 ml A/P3) 50 ml A/P3)
Mat. No.1) 301229 302113 302135 300149 300141 300144
Time to Occl.2)
P1 [mm:ss] 1:25 0:24 0:28 0:50 1:11 3:17
P9 [mm:ss] 8:50 1:04 1:22 2:36 5:03 16:36
51
COMPATIBLE SYRINGES
Chapter 7
Manufacturer:
TERUMO
Syringe Type 3 ml 5 ml 10 ml 20 ml 30 ml 50 ml 60 ml
TERUMO
EU/USA/JAP
Mat. No.1) 3SS*03L 3SS*05L 3SS*10L 3SS*20L 1SS*30LZ1 2BS-50LG 3SS*60L
1SS*05LZ1 1SS*10LZ1 SS*20ES
Time to Occl.2) typ. typ. typ. typ. typ. typ. typ.
P1 [mm:ss] 0:43 0:35 0:55 2:12 2:25 3:01 3:34
P9 [mm:ss] 1:17 1:16 4:48 7:53 8:18 16:55 17:03
Manufacturer:
Codan
Syringe Type Codan Codan Codan Codan Codan Codan
Codan 2 ml3) 5 ml3) 10 ml3) 20 ml3) 30/35 ml3) 50/60 ml3)
Manufacturer:
Fresenius
Syringe Type Injectomat
Fresenius 50 ml
Mat. No.1) 9000701
Time to Occl.2) typ.
P1 [mm:ss] 4:37
P9 [mm:ss] 21:09
52
START UP GRAPHS AND TRUMPET CURVES
Chapter 8
The graphs show the accuracy/uniformity of flow in relation to time. They allow
for the following:
The delivery behaviour or delivery precision is essentially influenced by the type
of (disposable syringe) used. Deviations from the technical data of the pump
cannot be excluded if syringes (disposables) other than those stated in the order
data are used.
Trumpet Curves
Measured values for second hour in each case.
Measurement interval ∆t = 0.5 min
Observation interval p x ∆t [min]
Start Up Graphs
Measurement interval ∆t = 0.5 min
Measurement duration T = 120 min
Flow Qi (ml/h)
53
TECHNICAL DATA
Chapter 9
T E C H N I C A L D ATA
55
WARRANTY / TSC* / SERVICE / TRAINING / CLEANING / DISPOSAL
Chapter 10
Training
B. Braun offers a training for version J. Please ask your local representative for
further details.
56
WARRANTY / TSC* / SERVICE / TRAINING / CLEANING / DISPOSAL
Chapter 10
Check regularly
Check for cleanliness, completeness and damage. Use only according to
Instructions for Use. During an exchange interval of the disposable the pump
has to perform a self-test. Check the following items each time the pump is
switched on: self-check, audible alarm, process- and alarm control indication.
Cleaning
Clean external surface of pump using mild soap suds. Do not use spray
disinfectants at the mains power connection. Recommended: disinfectant for
wiping available from B. Braun: Meliseptol® Foam pure, Melsitt 10% and
Melsept SF 10%. After cleaning, allow the device to vent for at least 1 min prior
to use. Do not spray into openings in the device. Be sure to observe the instruc-
tions provided concerning waste
disposal and hygiene for batteries and disposables. Wipe magnifying- and dis-
playglas on front of pump door only with a soft cloth. Do not use Hexaquart® or
other alkylamine containing disinfectants.
Disposal
The pumps as well as battery packs can be returned to B. Braun for further dis-
posal. When taking care of disposing of disposables as well as infusion
solutions, please consider the applicable hygiene and disposal regulations.
Inspection on Delivery
Despite careful packaging, the risk of damage during transport cannot be
entirely prevented. Upon delivery, please check that all items are present. Do not
use a damaged device. Contact the service department.
Testing the proper funciton of the device should be performed before initial use.
This is even ruled by law in several countries. A respective form can be obtained
from B. Braun.
Included in Delivery
Perfusor® Space, Battery-Pack SP, Instructions for Use-Set.
57
INSTRUCTIONS FOR USE ACCESSORY
Chapter 11
PoleClamp SP (8713130)
A maximum of three B. Braun Space pumps and one SpaceControl can be
stacked together when used with the PoleClamp SP. For detailed instructions on
secure fixation of the PoleClamp SP please refer to "Overview Perfusor® Space"
and "Patient Safety".
Note: A maximum of three plugs can be stacked upon each other in socket P2.
59
INSTRUCTIONS FOR USE ACCESSORY
Chapter 11
dynamic 1 sec
without Alarm
Off–Alarm
Operation
*) in the mode static without Off-Alarm, the staff call can be surpressed with k
Caution: The user should always closely observe the local pump alarms as well.
Note: A maximum of three plugs can be stacked upon each other in socket P2.
Technical Data
Connecting Wire
white and green white and brown
Alarm disconnected connected
Operation connected disconnected
60
INSTRUCTIONS FOR USE ACCESSORY
Chapter 11
PCA-Accessories
• Space PCA-Kit (REF 8713554) consisting of: :
- Demand button
- Hook and loop tape
for fixation of the demand button at
the patient s arm
- Line fixation
connection between hook and loop tape
and demand button
- Metal clip
alternatively for fixation at the bed
sheet
- Cable strap
for wrapping the cable of the demand
button
- PCA-Key
for locking the syringe holder or the Syringe Anti Removal Cap
Art. No.
B. Braun Perfusor® Space (100 – 240 V) ...................................................871 3030
62
ORDERING
63
Manufactured by: B. Braun Melsungen AG
B. Braun Melsungen AG Sparte Hospital Care
34209 Melsungen 34209 Melsungen
Germany Germany
Tel +49 (0) 56 61 71-0 Tel.: +49 (0) 56 61 71-0
38916517 • Drawing No. M688 70 02 01 F04 Fax: +49 (0) 56 61 71-20 44
Printed on pulp bleached 100 % chlorine-free
1210 • Date of last revision: November 2010 www.bbraun.com