Sonoline G20: System Display Specifications

Download as pdf or txt
Download as pdf or txt
You are on page 1of 10

SONOLINE G20

US

Configuration
System
Display Specifications

08647831
08647815

© Siemens AG 2004
The reproduction, transmission or use
of this document or its contents is not
permitted without express written
Siemens Medical
Ultrasound
ejt DivisionSolutions
authority. Offenders will be liable for
damages. All rights, including rights
created by patent grant or registration
of a utility model or design, are
reserved.

Print No.: US06-101.843.01.02.02 English


Replaces: US06-101.843.01.01.02 Doc. Gen. Date: 01.05
n.a.

Part No.: n.a.


sd
2 Revision / Disclaimer
1Revision / Disclaimer

Copyright
Copyright © 2004 by Siemens Corporation. All rights reserved. No part of this publication
may be reproduced, transmitted, transcribed, stored in retrieval systems, or translated into
any language or computer language, in any form or by any means, electronic, mechanical,
magnetic, optical, chemical, manual, or otherwise, without the prior written permission of
Siemens Corporation. Siemens Corporation reserves the right to change its products and
services at any time. In addition, this manual is subject to change without notice. Siemens
Corporation welcomes customer input on corrections and suggestions for improvements
to this manual.Although Siemens Corporation has attempted to ensure accuracy through-
out this manual, Siemens Corporation assumes no liability for any errors or omissions, nor
for any damages resulting from the application or use of this information.
Trademarks
ACUSON, Sequoia, Cypress, AEGIS, Aspen, 128XP, XP, AcuNav, CV70, SONOLINE,
Adara, Antares, G20, G40, G50, G60S, Omnia and KinetDx are trademarks of Siemens
Corporation registered in the U.S. Patent and Trademark Office. CWS3000, DS3000,
WS3000, and WebPro are trademarks of Siemens Corporation.
Windows XP, 2000 and NT are registered trademarks of Microsoft Corporation. Internet
Explorer is a trademark of Microsoft Corporation. Netscape and Netscape Navigator are
registered trademarks of Netscape Communications Corporation. Java and the JavaScript
name are registered trademarks of Sun Microsystems, Inc. Adobe Acrobat is a registered
trademark of Adobe Systems, Inc.
Other products and brand names are trademarks of their respective owners.
Disclaimer
The service of equipment described herein is to be performed by qualified personnel who
are employed by Siemens or one of its affiliates or who are otherwise authorized by Sie-
mens or one of its affiliates to provide such services.
Assemblers and other persons who are not employed by or otherwise directly affiliated
with or authorized by Siemens or one of its affiliates are directed to contact one of the
local offices of Siemens or one of its affiliates before attempting installation or service pro-
cedures.
License Agreement
All computer programs copyright 1990-2004 by Siemens Corporation or its suppliers. Pro-
grams are licensed under the following agreement:
Siemens or its suppliers retain(s) ownership of and title to any computer program supplied
with the equipment and to the trade secrets embodied in such computer programs. Sub-
ject to the Buyer’s acceptance and fulfillment of the obligations in this paragraph, Siemens
grants the Buyer a personal, non-transferable, perpetual, non-exclusive license to use any
computer program supplied with the Equipment that is necessary to operate the Equip-
ment solely on the medium in which such program is delivered for the purpose of operat-
ing the equipment in accordance with the instructions set forth in the operator’s manuals
supplied with the Equipment and for no other purpose whatsoever. Buyer may not
reverse-assemble, reverse-compile or otherwise reverse-engineer such computer pro-
grams nor may Buyer make a copy of such program or apply any techniques to derive the
trade secrets embodied therein. In the event of a failure by Buyer to comply with the terms

SONOLINE G20 US06-101.843.01.02.02 Page 2 of 10 Siemens


01.05 sd Medical Solutions USA, Inc.
Revision / Disclaimer 3

of this license, the license granted by this paragraph shall terminate. Further, because
unauthorized use of such computer programs will leave Siemens without an adequate
remedy at law, Buyer agrees that injunctive or other equitable relief will be appropriate to
restrain such use, threatened or actual. Buyer further agrees that (i) any of Siemens"s
suppliers of software is a direct and intended beneficiary of this end-user sublicense and
may enforce it directly against Buyer with respect to software supplied by such supplier,
and (ii) No supplier of Siemens shall be liable to buyer for any general, special, direct, indi-
rect, consequential, incidental or other damages arising out of the sublicense of the com-
puter programs supplied with the equipment.

Siemens US06-101.843.01.02.02 Page 3 of 10 SONOLINE G20


Medical Solutions USA, Inc. 01.05 sd
4 Table of Contents
1- 0Table of Contents

1 _______ Display Specifications____________________________________________ 5

Overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Display Formats and Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Monitor Display Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8

SONOLINE G20 US06-101.843.01.02.02 Page 4 of 10 Siemens


01.05 sd Medical Solutions USA, Inc.
Display Specifications 5
2-
Overview
1Display Specifications

The Video Format and Specification section supplies information about the G20 system
video formats used during imaging. This information is divided into the following sections:

Display Formats and Specifications 0

Display Formats
Display formats are displayed while imaging in 2-D or M-mode. 2-D mode uses three
video display formats and M-mode uses five video formats.
The following table describes the 2-D mode display formats:

2-D Mode Display 2-D Mode Display Description


Single mode 2-D image displays a full-screen image.

Dual 2-D images display side-by-side in one of three ways:


• Dual 2-D images display that have identical imaging parame-
ter settings. The left-side displays is the actual image and
right-side image displays with post-processing applied.
• Dual 2-D images display that have independent imaging pa-
rameter settings.
• Activate Pan or Zoom on either of the dual displays.The track-
ball controls the position of the Zoom and Pan functions. The
Depth/Zoom knob actives and controls the Zoom function for
a zoom factor of up to five times the image size.
Quad 2-D images display independent imaging parameter sets
on each image, but uses identical modality for the quad images.
Only one image will be real-time.

The following table describes the M-mode display formats:

Siemens US06-101.843.01.02.02 Page 5 of 10 SONOLINE G20


Medical Solutions USA, Inc. 01.05 sd
6 Display Specifications

M-mode Display M-mode Display Description


M-mode images displays a full-screen variable-speed sweep.

Displays a 2-D image and an M-mode sweep simultaneously.


When zooming either the 2-D image or the M-mode image can
be zoomed.

Organize the position of both the 2-D and M-mode screens as


desired for individual preferences.

Zooming either the 2-D image or the M-mode sweep will zoom
both the image and the sweep simultaneously.

Fig. 1:

Simultaneously display a zoomed M-mode and a 2-D image and


the original 2-D image.

Fig. 2:

Video Specifications
The 2-D and M-modes are available on all G20 transducers. 2-D mode is available in all
display formats; M-mode is available in the full screen display format, but not available
during Dual or Quad 2-D mode display formats.

Sweep Speed The M-mode image sweep speed is selectable. Choose from the
following sweep speeds: 2.0 s, 4.0 s, 8.0 s, or 16.0 s. The sweep
speed is displayed in seconds (s) when a full screen is selected;
the sweep speed displays in millimeters per second (mm/s)
when Dual or Quad screen displays are selected.
Acoustic Line Den- Acoustic line density supports single and high resolution and are
sity selectable. The acoustic line density, frame rate, depth, and
transmit focus are all affected when any of these settings are
changed. The G20 system automatically optimizes these setting
each time a change to one of them is made.

SONOLINE G20 US06-101.843.01.02.02 Page 6 of 10 Siemens


01.05 sd Medical Solutions USA, Inc.
Display Specifications 7

Imaging Depth or Imaging depth is transducer dependent and can be adjusted


Field of View (FOV) within the transducer dependencies.
(For specification details see the transducer catalog at url,
http://www.acuson.com/transducers/catalog/index.htm)
Transmit Focus Transmit Focus is available during 2-D imaging and has three
selectable modes: Single, Dual and Quadruple.
• Single focus is transducer frequency dependent. The default
setting for single focus is set for the maximum depth and has
eight selectable focal positions. Dual focus, depending on the
depth selection, reduces the frame rate. The default setting
for dual focus is the maximum depth for each transducer fre-
quency. There are eight selectable focal positions that can be
positioned in the image. Quadruple focus, depending on the
depth selection, reduces the frame rate. Each transducer fre-
quency has eight selectable focal positions which can be po-
sitioned in the image.
Dynamic Focus Dynamic focus is supported for linear and convex transducers
for 2-D and M-mode imaging. This dynamic focus is applicable
to focusing in scanning direction.
Transmit Aperture The transmit aperture control changes the transmit aperture size
Control according to the focal depth.
Receive Aperture During 2-D and M-mode imaging with an array transducer, the
Control receive aperture control increases the aperture up to 48 chan-
nels.
Gain Control The gain control increases or decreases gain (from 0 to 40 dB in
increments of 1 dB) of the entire 2-D or M-mode image. The gain
control for 2-D and M-mode are independent from one another.
The gain control can be used in combination with the DGC slid-
ers.
Note: The scan width is selectable; when the scan width is
decreased the frame rate increases.
DGC Control The DGC controls increase or decrease 2-D gain at the eight
focal points of the image (from 0 to 30 dB in 1 dB increments).
One DGC slider is assigned to each focal point. The DGC curve
is displayed on the monitor; the duration of the monitor display
can be set to: off, time out, or always.
Persistence For 2-D images, select from up to five persistence levels includ-
ing 0 (OFF). The exact value depends on the difference between
current frame image and previous frame image. The levels are:
Previous / Current: OFF(0/100), 15/85, 30/70, 40/60, 50/50, and
60/40.

Siemens US06-101.843.01.02.02 Page 7 of 10 SONOLINE G20


Medical Solutions USA, Inc. 01.05 sd
8 Display Specifications

Edge Enhancement Select from four edge enhancement levels including OFF. Dur-
ing simultaneous mixed 2-D and M-modes, echo enhancement
is individually selectable for both modes.
Dynamic Range Select a dynamic range setting from 35 to 70 dB in 1 dB incre-
ments. During simultaneous mixed 2-D and M-modes, dynamic
range is individually selectable for both modes.
Gray Scale Pro- • Seven gray scale maps are available for postprocessing. The
cessing selected postprocessing map will display on the monitor.
• Seven postprocessing map curves are available to optimize
gray scale mapping for 2-D images.
Zoom The amount a 2-D image can be zoomed is dependent on the
transducer frequency. Zooming a 2-D image optimizes the
image quality and the acoustic condition. Select a Region of
Interest (ROI) on a 2-D image, and then enlarge the ROI.

Monitor Display Specifications 0

Input Condition

Supply Voltage: +12 V +3%


Ripple Voltage: <10 mVp-p
Power Consumption: < 3.0 A
Rush Current: < 40 A
Protection for Circuit: Fuse (3.15 A/250 V: Time Lag type)

Power Supply Video Amplifier

Supply Voltage: +12 V +3%


Ripple Voltage: <10 mVp-p
Power Consumption: < 3.0 A
Rush Current: < 40 A
Protection for Circuit: Fuse (3.15 A/250 V: Time Lag typ

SONOLINE G20 US06-101.843.01.02.02 Page 8 of 10 Siemens


01.05 sd Medical Solutions USA, Inc.
Display Specifications 9

Video Amplifier

Frequency Band: 30 MHz / -3 dB


Amplifier: Linear
Brightness: Calibrated Luminance
• Both brightness and contrast pots =
center detent position
• VGA:3.0 +- 1.5 cd/m^2
(input: 137 mV)
Max Luminance
• Both brightness and contrast pots =
Max position
Video input = 700 mV, Center -window
(40 mm x 40 mm)
• VGA: > 400 cd/ m2
Imaging: < 15 seconds
Set up time: 30 minutes

Imaging Performance

Image Size: Horizontal 210 mm +/- 4 mm 23.37 us


Vertical 160 mm +/- 4 mm 13.91 ms
Raster 439 line
Linearity Horizontal < 8%
Vertical < 8%
Inclination |h-h"| < 2.5 mm
Static Regulation Horizontal < 4 mm
Vertical < 4 mm

Siemens US06-101.843.01.02.02 Page 9 of 10 SONOLINE G20


Medical Solutions USA, Inc. 01.05 sd
10 Display Specifications

Monitor Screen Specification

Fig. 3: Monitor Specifications


Pos. A CRT area
Pos. B Display area
Pos. C Image area
Pos. X1 245 mm (min.)
Pos. X2 207.8 +/- 5 mm
Pos. X3 169 +/- 5 mm
Pos. Y1 181 mm (min.)
Pos. Y2 167+/- 5 mm
Pos. Y3 145 +/- 5 mm

SONOLINE G20 US06-101.843.01.02.02 Page 10 of 10 Siemens