Power:
The Transistor Tester can be powered from 6.8V-12V DC. This can be achieved by a 9V layer-
built battery or two 3.7V Lithium-ion battery in series, or with a 9V DC adapter. When the
power is on, the current is about 30mA at DC 9V and when off the current is about 20nA.
Control:
The Transistor Tester is controlled by a “Rotary Pulse Encoder with switch”, or RPEWS for
short. This component has four modi of operation: Short time press, Long time press , Left
Rotation and Right Rotation. When the Transistor Tester is powered, a short time press of
the RPEWS will switch on the Transistor Tester and start a Test. The Transistor Tester will
wait for user input at the end of a test, before it’s auto switched off. A long time press or
Rotation of left or right the RPEWS will enter the function menu. In the function menu, “>”
at left column shows the Selected menu item. To enter the Specific function, just short press
the RPEWS. Within a Specific function, a long time press will exit and then goes back to the
function menu.
Test:
The Transistor Tester has three Test Points (TP1,TP2,TP3) within the Test socket, these
three are allocated as follows:
At the right side of the Test socket is the SMT test pad, also have number to identify each.
When testing a two lead component (resistor, capacitor, inductor), the two lead can select
any two test point. If TP1 and TP3 are selected, the Test will enter to ”series test mode”
when the test is completed, else the test is started again by a short press the RPEWS.
Attention:
Always be sure to discharge capacitors before connecting them to the Tester!
The Tester may be damaged before you have switched it on. There is only a
little protection at the MCU’s ports. Extra caution is required if you try to test
components mounted in a circuit. In either case the equipment should be disconnected from
power source and you should be sure, that no residual voltage remains in the equipment.
If this is the first time use, the Tester will soon Prompt string “1-||-3 > 100nF”. A capacitor
with any capacity between 100nF and 20μF connected to pin 1 and pin 3 is required for the
last task of calibration. You should connect the capacitor not before this text is shown. With
this capacitor the offset voltage of the analog comparator will be compensated for better
measurement of capacity values.
If you try to measure components in assembled condition, the equipment should be all
ways disconnected from power source. Furthermore you should be sure, that no residual
voltage reside in the equipment, because every electronical equipment has capacitors
inside! If you try to measure little resistor values, you should keep the resistance of plug
connectors and cables in mind. The quality and condition of plug connectors are important,
also the resistance of cables used for measurement. The same is in force for the ESR
measurement of capacitors. With poor connection cable a ESR value of 0.02Ω can grow to
0.61Ω. You should not expect very good accuracy of measurement results, especially the ESR
measurement and the results of inductance measurement are not very exact.
Components with problems:
You should keep in mind by interpreting the measurement results, that the circuit of the
Transistor Tester is designed for small signal semiconductors. In normal measurement
condition the measurement current can only reach about 6 mA. Power semiconductors
often make trouble by reason of residual current with the identification and the
measurement of junction capacity value. The Tester often cannot deliver enough ignition
current or holding current for power Thyristors or Triacs. So a Thyristor can be detected as
NPN transistor or diode. It is also possible, that a Thyristor or Triac is detected as unknown.
Thus the Gate get with the growing of current a negative bias voltage. The Tester reports
the Source current of this circuit and additionally the bias voltage of the Gate. So various
models can be differed. The D-MOS transistors (depletion type) are measured with the same
method. You should know for enhancement MOS transistors (P-E-MOS or N-E-MOS), that
the measurement of the gate threshold voltage (Vth) is more difficult with little gate
capacity values. You can get a better voltage value, if you connect a capacitor with a value of
some nF parallel to the gate/source. The gate threshold voltage will be find out with a drain
current of about 3.5mA for a P-E-MOS and about 4mA for a N-E-MOS.
Function menu descriptions:
1. Switch off
Enter this Function Tester will shut down immediately.
2. Transistor
Transistor test, it is also the default Function at switch on.
3. Frequency
Measurement of frequency, For frequencies below 25kHz the normal measurement is
followed by a measurement of period time. This additional measurement is only followed
after a normal frequency measurement.
4. f-Generator
Signal generation, this Function can output square wave with various frequencies.
5. 10-bit PWM
The function ”10-bit PWM” (Pulse Width Modulation) generates a fixed frequency
(7812.5Hz) with selectable pulse width at the pin TP2. With a short key press (< 0.5 s) the
pulse width is increased by 1%, with a longer key press the pulse width is increased by 10%.
If 99% is overstepped, 100% is subtracted from the result. The function can be exit with a
very long key press (> 1.3 s).
6. C+ESR@TP1:3
The additional function ”C+ESR@TP1:3” selects a stand-alone capacity measurement with
ESR (Equivalent Series Resistance) measurement at the test pins TP1 and TP3. Capacities
from 2μF up to 50mF can be measured. Because the measurement voltage is only about
300mV , in most cases the capacitor can be measured ”in circuit” without previous
disassembling. The series of measurements can be finished with a long press of RPEWS.
7. Rotary encoder
With the function ”rotary encoder” a rotary encoder can be checked. The three pins of the
rotary encoder must be connected in any order to the three probes of the transistor tester
before the start of the function. After starting the function the rotary knob must be turned
not too fast. If the test is finished successfully, the connection of the encoder switches is
shown symbolic in display row 2. The tester finds out the common contact of the two
switches and shows, if the indexed position has both contacts in open state (’o’) or in closed
state (’C’). A rotary encoder with open switches at the indexed positions is shows in row 2
for two seconds as ”1-/-2-/-3 o”.
This type of encoder has the same count of indexed positions as count of pulses for every
turn. Of course the pin number of the right common contact is shown in the middle instead
of ’2’. If also the closed switches state is detected at the indexed positions, the row 2 of the
display is also shown as ”1—2—3 C” for two seconds. I don’t know any rotary encoder,
which have the switches always closed at any indexed position. The interim state of the
switches between the indexed positions is also shown in row 2 for a short time (< 0.5s)
without the characters ’o’ or ’C’.
8. Selftest
With the menu function ”Selftest” a full self test with calibration is done. With that call all
the test functions T1 to T7 (if not inhibited with the NO TEST T1 T7 option) and also the
calibration with external capacitor is done every time.
9. Voltage
Voltage measurement, Because a 10:1 (180K:20K) voltage divides is connected, the
maximum external voltage can be 50V, The measurement can also be exit with a extra long
key press (> 4 seconds).
11. FrontColor
This function can change the color of the font, the 16bit color is encoded in RGB(565)
format, meaning: Red = 31 maximum, Green = 63 maximum and Blue = 31 max. respectively.
Within the function, a short time press can change the color. Turn left decreases it value and
turn right increases it value. A long time press will save the Result and exit the function.
Please keep in mind that the FrontColor and BackColor cannot be the same, this will cause
the LCD to show nothing. If this happens, you need to do a Selftest , how to enter the
Selftest is described in the manual at page 2. The Selftest will change the back Color to black
and font color to white automatically. When Selftest is finished, you will have the chance to
modify the color again.
12. BackColor
This is function is the same as the FrontColor except it’s change the background color.
13. 1-||-3
This function can series Measurement the capacitance at TP1 and TP3, this function can
measure very small capacitors. A long time press will exit the function.
14. 1- - 3 This function can series Measurement the Resistance and Inductance at
TP1 and TP3. A long time press will exit the function.