AccuPyc II Series Operator Manual
AccuPyc II Series Operator Manual
AccuPyc II Series Operator Manual
OPERATOR MANUAL
134-42800-01
Jan 2016
(Rev A)
Trademarks
Copyright
The software described in this manual is furnished under a license agreement and may be used or
copied only in accordance with the terms of the agreement.
WARRANTY
MICROMERITICS INSTRUMENT CORPORATION warrants for one year from the date of shipment each
instrument it manufactures to be free from defects in material and workmanship impairing its usefulness under normal
use and service conditions except as noted herein.
Our liability under this warranty is limited to repair, servicing and adjustment, free of charge at our plant, of any
instrument or defective parts when returned prepaid to us and which our examination discloses to have been
defective. The purchaser is responsible for all transportation charges involving the shipment of materials for warranty
repairs. Failure of any instrument or product due to operator error, improper installation, unauthorized repair or
alteration, failure of utilities, or environmental contamination will not constitute a warranty claim. The materials of
construction used in MICROMERITICS instruments and other products were chosen after extensive testing and
experience for their reliability and durability. However, these materials cannot be totally guaranteed against wear
and/or decomposition by chemical action (corrosion) as a result of normal use.
Repair parts are warranted to be free from defects in material and workmanship for 90 days from the date of shipment.
No instrument or product shall be returned to MICROMERITICS prior to notification of alleged defect and
authorization to return the instrument or product. All repairs or replacements are made subject to factory inspection of
returned parts.
MICROMERITICS shall be released from all obligations under its warranty in the event repairs or modifications are
made by persons other than its own authorized service personnel unless such work is authorized in writing by
MICROMERITICS.
The obligations of this warranty will be limited under the following conditions:
1. Certain products sold by MICROMERITICS are the products of reputable manufacturers, sold under their
respective brand names or trade names. We, therefore, make no express or implied warranty as to such
products. We shall use our best efforts to obtain from the manufacturer, in accordance with his customary prac-
tice, the repair or replacement of such of his products that may prove defective in workmanship or materials.
Service charges made by such manufacturer are the responsibility of the ultimate purchaser. This states our
entire liability in respect to such products, except as an authorized person of MICROMERITICS may oth-
erwise agree to in writing.
2. If an instrument or product is found defective during the warranty period, replacement parts may, at the dis-
cretion of MICROMERITICS, be sent to be installed by the purchaser, e.g., printed circuit boards, check
valves, seals, etc.
3. Expendable items, e.g., sample tubes, detector source lamps, indicator lamps, fuses, valve plugs (rotor) and
stems, seals and O-rings, ferrules, etc., are excluded from this warranty except for manufacturing defects.
Such items which perform satisfactorily during the first 45 days after the date of shipment are assumed to be
free of manufacturing defects.
Purchaser agrees to hold MICROMERITICS harmless from any patent infringement action brought against
MICROMERITICS if, at the request of the purchaser, MICROMERITICS modifies a standard product or
manufactures a special product to the purchaser’s specifications.
MICROMERITICS shall not be liable for consequential or other type damages resulting from the use of any of its
products other than the liability stated above. This warranty is in lieu of all other warranties, express or implied,
including but not limited to, the implied warranties of merchantability or fitness for use.
CONTACT US
Micromeritics Instrument Corporation
4356 Communications Drive
Phone: 1-770-662-3636
Fax: 1-770-662-3696
www.Micromeritics.com
---------------
U.S. Inquiries
U.S. Sales Phone: 1-770-662-3636
---------------
---------------
CAUTION - Cautions contain information to help prevent actions that may damage the
analyzer or components.
WARNING - Warnings contain information to help prevent actions that may cause per-
sonal injury.
KEYPAD FUNCTION- - The keypad icon indicates the section is applicable only when
using the keypad.
Warranty i
Contact Us ii
6 Perform an Analysis 6 -1
Prepare and Load a Sample 6 -1
Perform an Analysis 6 -3
QuickStart Analysis 6 -7
Sequence Analysis 6 -8
Review Analysis 6 -9
7 About Reports 7 -1
Open and Close Reports 7 -1
Start Reports 7 -1
SPC Report 7 -2
8 Selected Reports 8 -1
Combined Report 8 -1
Density and Volume Table 8 -1
Density vs Cycle Number Trend Plot 8 -2
Density vs Time Trend Plot 8 -3
Equilibration Report 8 -4
Options Report 8 -5
Pressure and Volume Table 8 -5
Sample Log Report 8 -6
Summary Report 8 -6
Tabular Report 8 -7
Total Pore Volume vs Temperature 8 -8
Volume vs Cycle Number Trend Plot 8 -8
10 About Calibration 10 - 1
11 Hardware 11 - 1
Add Analysis Module to Control Module 11 - 1
Analytical Balance 11 - 3
Printer Setup 11 - 3
A Calculations A-1
Analysis A -1
Asphalt Density A -3
Calibration A -4
FoamPyc Methods A -5
Geometric Volume and Active Area A -7
Percent Porosity A -7
Resin Volume A -8
Run Precision A -8
SPC Report Variables A -9
Specific Gravity A - 10
Total Pore Volume A - 10
Total Solids Concentration A - 11
G TEC Module G -1
I Transmitted Data I -1
Index Index - 1
The AccuPyc II is an easy to use, fully automatic gas displacement pycnometer. It determines
density and volume by measuring the pressure change of helium within calibrated volumes. It also
reports the chamber temperature at the end of the requested cycles. The AccuPyc II can be operated
using the analysis program installed on an attached computer, or if no computer is attached to the
pycnometer, the keypad can be used to start and run analyses. If using the keypad, the analyzer may
be operated in English, French, German, Italian, or Spanish. The application is available in English
only.
The AccuPyc II consists of a control module and an analysis module. When ordering a single unit,
the control module is built into the same unit as the analysis module for convenience. The analysis
modules are also available in a standalone configuration, allowing multiple analysis modules to be
attached to a single controlling unit. A maximum of five analysis modules can be installed to one
controller (six total). Each module has its own gas connection. The pycnometer can be controlled by
either:
l a 32-bit Windows interface. The operational status of the pycnometer can be continually mon-
itored in the application. If the Windows interface is used, the keypad is disabled.
l the keypad if not running the 32-bit Windows interface. A separate keyboard and printer
Density measurements at specific temperatures are required for certain applications. The AccuPyc II
Thermo Electric Cooling (TEC) Module maintains an accurate temperature control during analysis
using the principle of Peltier cooling.
TEC Modules, also known as Peltier coolers, are solid-state heat pumps that utilize the Peltier effect
to move heat. By passing a current through the TEC Module, heat is transferred from one side of the
module to the other, typically producing a heat differential of approximately 40° C.
The transferred heat is then removed from the system through the combination of a heat sink and
fan, cooling the system to maintain a set temperature for accurate analysis.
This solution can be closely correlated (< 0.15% difference) to results obtained with ASTM Test
Method D70-09. The AccuPyc asphalt density measurement is completed much faster than the
ASTM method while virtually eliminating operator error. The Peltier thermoelectric cooling control
(15 to 36 °C) provides stability for handling of samples.
Performing an asphalt density measurement requires the AccuPyc II Asphalt application and
disposable sample cups.
l Method A. Measures the closed cell fraction and corrects for the cells damaged while cutting the
sample to the necessary size and shape. This is accomplished by using either the average cell dia-
meter or the cell chord length (as defined in ASTM method D-3576) and the measurements of the
sample to determine the volume of the cut cells. This volume is deducted from the total volume of
the open cells measured by the pycnometer.
l Method B. Corrects for the cut cells by using two separate measurements. For the 2nd meas-
urement, the sample is recut to double the amount of cut surface. The observed difference in cut
open cell volume is applied as a correction to the initial measured volume. This method offers the
distinct advantage that no assumptions are needed about the relative amounts of open and
closed cells.
l Method C. Does not correct for cut cells. It is used for materials with predominantly open cells
where good accuracy can be achieved without correction. The accuracy level deteriorates as the
percentage of closed cells increases.
l Method D. Is a compressibility test. The fill pressure is increased incrementally over the sample
with each repeat of the P1,P21 ) cycle. The apparent variation of the measured sample volume
1 ) Where P1 is the initial pressure to which the sample is charged, and P2 the final pressure after
expansion.
with the average pressure is determined. This test is an approximate indication. It is not intended
to be an exact measure of the volume compressibility.
l Method E. Is a cell fracture test; a perfectly rigid foam is assumed. First, a P1,P2 cycle is per-
formed at the lower of two specified P1 pressures, and the results stored. A second cycle is per-
formed at a higher specified value of P1, then a third cycle identical to the first cycle is performed.
The difference between the volume of the sample on the first measurement and on the third meas-
urement is reported as the volume of fractured cells. It is assumed that cells fracture by exposure
to the highest pressure (2nd cycle) so that when the third measurement is made, the measured
sample volume has decreased from the first cycle by the amount of the closed cell volume which
was fractured.
The 100 cm3 AccuPyc II has been designed and tested to follow the procedure in ASTM Method D
6226 for FoamPyc analyses. The 10 cm3 and 350 cm3 units can be used for some types of foam
materials; however, analyses on these AccuPycs will not conform to the ASTM method. The 1 cm3
AccuPyc II cannot be used to perform FoamPyc analyses.
The FoamPyc software also provides the capability to perform analyses in the Standard mode of
operation. This is especially convenient for users who have upgraded to the FoamPyc method from
the standard method. For example, a 100 cm3 AccuPyc on which FoamPyc analyses are performed
and a 1 cm3 analysis module attached on which standard analyses are performed.
INSTRUMENT COMPONENTS
FRONT
The sample chamber, located on the top panel, is where the sample container is placed for analysis,
or the standard for calibration.
The sample chamber should remain capped except when inserting or removing a sample. When left
uncapped, water vapor adsorbs on the inner surface of the chamber or the chamber temperature
becomes unstable; either condition can affect analysis results. If water vapor accumulates in the
chamber, the pycnometer must be purged.
The standalone control module, often used for the Glove Box pycnometer, does not contain a
sample chamber.
Component Description
Active indicator A blue LED which illuminates when an operation is in progress. It is included
on all analysis modules. It is not included on standalone control modules.
Display window Displays Remote Operation when the Windows application is being used.
The AccuPyc can be operated in the keypad mode when the Windows applic-
ation is closed.
Keypad Not used with AccuPyc software application.
Power indicator A green LED which illuminates when the analyzer is powered on.
REAR
Component Description
Analysis module Used to connect up to five analysis modules to the initial control / analysis mod-
connector ule. Each module contains analysis module input and output connectors.
Brightness con- Used to adjust the brightness of the display. See Brightness Control on
trol page J - 18.
Dust filter (not (For 2000 cm3 units and high pressure units only.) Traps sample dust to pro-
shown) tect internal valves.
Ethernet port Provides for setting up for e-mailing reports, sending data to a web browser,
or interfacing with the AccuPyc Windows application.
Gas inlet port Use to connect the analysis gas; helium is recommended. If multiple analysis
modules are attached, each module contains a gas inlet port. Multiple gases
can be attached using the Multigas option. See Multigas Option on page C
- 1.
On/Off switch For powering the analyzer on and off.
Power connector For connecting the power source to the instrument.
RS-232 port Used to transmit data only when using the keypad function. Also used to con-
nect to the analytical balance.
USB connectors Use for connecting a keyboard, printer, or balance to the instrument.
EQUIPMENT OPTIONS
l Operates at a pressure of up to 500 psi to provide a better diffusion of the gas into the rock.
l Has a stainless steel sample chamber with a volume of 3100 cm .
l Has a sample chamber that can accommodate a 48 mm (1.85 in.) diameter core of up to 60 mm
MULTIVOLUME INSERTS
The MultiVolume option is used to analyze smaller sized samples. Options are:
3 3 3
l 100 cm nominal cell volume; contains 10 cm and 35 cm cups
3 3 3
l 2000 cm nominal cell volume, contains 650 cm and 1300 cm cups
MULTIGAS OPTION.
The Multigas Option allows connection of up to four gases to one analyzer. See Multigas Option
on page C - 1.
CONFIGURATION OPTIONS
For best fit with a sample, the AccuPyc II is available in multiple configurations.
l 1 cm3sample chamber
3
l 10 cm sample chamber
3
l 100 cm sample chamber
3
l 350 cm sample chamber
3
l 2000 cm sample chamber
3
l 100 cm High Pressure sample chamber
Best fit means the sample nearly fills the sample chamber and, therefore, optimizes the precision of
the results.
GLOVE BOX
Provides for analysis of samples in which a controlled environment is required.
Options Description
Configuration l 1 cm3 sample chamber
3
l 10 cm sample chamber
3
l 100 cm sample chamber
3
l 350 cm sample chamber
l 2000 cm
3 sample chamber
l 100 cm
3 High Pressure sample chamber
This unit consists of two separate modules. The controller is placed outside the glove box, while the
analysis module is placed inside the glove box. Up to six analysis modules may be placed in the glove
box, all controlled by one external control module.
Options Description
Configuration l 10 cm3 sample chamber
l 100 cm3 sample chamber
The temperature controlled unit is specifically designed for temperature sensitive materials. This unit
is ideally suited for laboratories in which ambient temperature varies during normal work hours, or in
which subambient cooling is required.
Do not power off the analyzer while initialization is in progress. Doing so may damage
the instrument.
When the analyzer is powered on, after a few seconds, the system vents automatically and the
green indicator light on the front panel illuminates. Allow approximately 30 minutes for the analyzer
to warm before performing analyses. For analyses that require very precise results, allow the
analyzer to warm a minimum of two hours. If running the analyzer with a computer:
1. Computer
2. Monitor
3. Printer
4. Analyzer
1. Analyzer
1. Allow any analyses in progress to complete prior to powering off the analyzer.
2. Printer
3. Analyzer
Characteristic Specification
Analysis Module Height. 43 cm (17 in.)
(2000 cm3) Width. 27 cm (10.6 in.)
Depth. 36.2 cm (14.3 in.)
Weight. 26 kg (57 lbs)
Control Module Height. 17.9 cm (7.0 in.)
Width. 27.3 cm (10.7 in.)
Depth. 36.2 cm (14.3 in.)
Weight. 3.6 kg (8.0 lbs)
Control Analysis Mod- Height. 17.9 cm (7.0 in.)
ule Width. 27.3 cm (10.7 in.)
Depth. 36.2 cm (14.3 in.)
Weight. 9.3 kg (20.5 lbs)
TEC Module Height. 27.3 cm (10.7 in.)
Width. 28.7 cm (11.3 in.)
Depth. 36.1 cm (14.2 in.)
Weight. 13.8 kg (30.5 lbs.)
Printers
Requirements l must be USB 2.00 (or newer)
lmust have a printer language supported by one of the printer drivers
l cannot be host based
Characteristic Specification
Computer Requirements
with shared access. Multiple users cannot operate the application at the same time.
l 10 Base T Ethernet Port. If the computer is to be connected to a network, two Ethernet ports
are required. If more than one Ethernet based unit is connected to the same computer, an eth-
ernet switch will also be required.
l Read/Write Permissions. All users of the application will need Read / Write permission to all
In keeping with a policy of ongoing product improvement, specifications are subject to change
without notice.
l Keypad (or a keyboard). Used if a computer with the analyzer software installed is not con-
nected to the analyzer. In this case, any section of this manual referring to software can be omit-
ted. All instructions specific to the keypad are designated with the icon specified in the section
About this Manual on page iii. Also see Configure the Pycnometer using the Keypad on
page J - 9.
l Computer. Used if the computer with analysis software is installed and connected to the analysis
Line Description
First Displays the unit number, serial number, and nominal cell volume of
the selected unit. Up to five analysis modules may be attached; though
the initial module, containing the keypad and display, will control all
units.
Second At the Reload prompt, manual control, or an automatic operation, this
line contains three characters, separated with dashes. The first
character represents the Fill valve, the second represents the Expand
valve, and the third represents the Vent valve.
During user interface, this line shows the command currently in use.
Third Displays status of the current operation, or Reload when in an idle
state. When the Reload prompt is displayed, this line may also contain
an asterisk (*), indicating there is a message in the queue. During cer-
tain functions (such as Setup), this line contains a prompt for additional
information.
Line Description
Fourth At the Reload prompt, shows the current pressure and temperature
alternately, where:
l P = Pressure
l T = Temperature
This line is also used to choose options, enter information, or provide
information about the current operation.
A keyboard can be connected to the USB port on the rear panel of the AccuPyc II. Com-
mands can be entered from either the keypad or the keyboard.
Most keys on the keypad perform one primary and one alternate function. The primary function of
any key is indicated by the number or command on the face of the key. The alternate function is
indicated by the command above the key. The top right corner of the display screen will show a +
(plus sign) when the Alt key is pressed and the alternate function mode is entered. See About the
Display on the previous page.
To select the:
lPrimary function. Press only the key. Primary functions are labeled on the key face. For
example; press 4 to enter the number 4.
l Alternate function. Press Alt, then press the key. Alternate functions have blue labels above
the key. For example; press Alt + 4 to start an analysis. If the Alt key is pressed accidentally,
pressing it a second time will cancel its function.
The alternate functions for keys 7 (FILL), 8 (EXPAND), and 9 (VENT) are executed differently.
Press Alt + 1 first to enter Manual mode then press the alternate key. For example: press Alt + 1 +
7 to manually control the Fill valve. Press Alt + CLEAR to exit Manual mode.
l Fill (Key 7)
l Expand (Key 8)
l Vent (Key 9)
l Analysis parameters
l Report options
l Calibration data
l Data transmission parameters
l Unit types
l Operating language
l Date and time
CANCEL AN ANALYSIS
To cancel an analysis, press Alt + CLEAR.
Press ENTER within five seconds to cancel the operation. A notification that the automatic operation
has been canceled will be displayed.
When canceling an operation, messages are displayed indicating that termination is in progress. The
termination process, which vents the system, takes about 30 seconds and returns to the Reload
prompt.
MANUAL MODE
Press Alt + 1 to enter manual mode, allowing manual control of the Bypass, Fill, Expansion, and
Vent valves.
The Bypass valve is only available on the 2000 cm3 units and the High Pressure Core
Pycnometer units.
Prolonged opening of all three valves at one time may cause excessive loss of helium.
TRANSMIT
The AccuPyc RS-232 interface transmits report data to a computer using the standard ASCII file
format. Reports may be transmitted in a single column format or a spreadsheet format. See
Transmitted Data on page I - 1.
ANALYSIS PARAMETERS
This option enables specific analysis and calibration parameters. Although the Setup Type is shown
as Analysis Parameters, these same prompts are used to specify calibration parameters. The
pycnometer is shipped with default values; however, these settings may be modified to meet
laboratory requirements.
After each selection, press Enter to display the next prompt. If data results are to be e-mailed
automatically after analysis, specify e-mail parameters in Setup Reports on page J - 14.
Enter appropriate analysis and report parameters for the current analysis. The chosen or specified
parameters in Set Up determine what prompts and information are displayed for the analysis.
A high rate will produce faster results, but results may not be as
precise as desired. The lowest rates may cause errors when some
materials (such as those with appreciable vapor pressures or
organics) are analyzed.
REVIEW DATA
The Review function allows review of and editing of the results of the last operation – analysis or
calibration – along with its entered parameters. It is important to review and print the report before
starting another operation. Data are no longer available for review on the display when another
operation is started; however, data results are saved in the control module and can be reviewed
using a web browser.
Data results for the last five analyses are saved in the control module for each attached unit. These
results can be viewed by accessing a web browser. To use this feature, connect to a network.
1. Press Alt + 2 to access Setup. Then press CHOICE until Communications is displayed.
2. Press ENTER to display a prompt showing that the IP Address mode is DHCP.
The display may show Specify, depending on how the IP address was assigned.
3. Press ENTER to view the assigned address on the display. If the address was specified, the
display shows IP Address? and the entered address. This line is editable since it is specified
and not assigned automatically. An assigned address cannot be edited.
4. Make a note of the IP address.
5. Press Alt + CLEAR to return to the Reload prompt.
6. Access the web browser and type in the IP address.
7. Press Enter to display the serial number(s) of the attached unit(s).
8. Click on the serial number of the unit containing the data to review. A window displaying the
data files for the last five analyses, and for the zero (slope) and volume offset files is displayed.
The files are in .HTML format and are assigned the name used as the Sample ID. If sample
identifications are not used, Untitled is assigned and appended for each file.
9. Click the file to view the data results. The progress of a current operation can be monitored by
reloading the page.
10. Use the File > Print option to print a copy of the results.
Reports are generated after analysis and calibration, and remain available for viewing or printing
until another automatic operation (other than zero) is performed. When an automatic operation is
performed, data from the previous operation are no longer available for viewing on the display.
A display report is always generated regardless of the specified destination and contains:
l Analysis. The average density (or volume) and the deviation from the mean.
l Calibration. Average cell (or expansion volume) and the deviation from the mean.
PRINT A FILE
Press Alt + 6 to print a report of the last operation; analysis or calibration. If Alt + 6 is pressed during
an automatic operation, a partial report is printed.
Reports are generated after analysis and calibration, and remain available for viewing or printing
until another automatic operation (other than zero pressure) is performed. Previously generated
reports are deleted whenever a new automatic operation begins.
For each attached analysis module, the data for up to five analyses are stored in the control module.
These data sets can be viewed using a web browser when connected to a network.
An asterisk next to a cycle number indicates that it has not been included in calculations.
l Analysis. Volume, density, deviation for each cycle, an average of all cycles, the date and time
the analysis was started and completed, and the temperature of the cell chamber.
l Calibration. Cell and expansion volumes and deviations for each cycle, an average of all cycles,
and the date and time the calibration started and completed.
Analysis provides the measurement of sample volume, from which density can be derived
automatically if the sample mass is entered. The unit comes preprogrammed with default conditions
and ready to perform analyses. It can be reprogrammed to meet specific needs. Analyses can be
modified by entering the number of purges and purge fill pressure, number of cycles and cycle fill
pressure, equilibration rate, and run precision. Modifying these parameters allow control of the
primary portions of the analysis; purge and run.
When the PC software is not running, the AccuPyc II can be operated in keypad mode. For example;
if the computer is inoperative for some reason, analyses can be performed using the keypad.
If running the FoamPyc module, an option to select a Standard method is available so that analyses
on other types of sample materials can be performed.
See Installation Instructions on page J - 1 for hardware installation, software installation, and
software upgrade instructions.
MENU STRUCTURE
All program functions use standard Windows menu functionality. The title bar contains a Unit
Number. If multiple units (analyzers) are installed, ensure the appropriate unit is selected before
continuing.
Copies. Select the number of copies to print. This field is only enabled
when Print is selected.
File. Select the destination directory. Enter a new file name in the File
Start Date. Displays a calendar to select the start date for the report.
Copies. Select the number of copies to print. This field is only enabled
when Print is selected.
File. Select the destination directory. Enter a new file name in the File
name field, or accept the default. Select to save the file as a report
system (.REP), a spreadsheet (.XLS), a portable document format
(.PDF), or an ASCII text (.TXT) file format.
Save Saves changes to the active window.
Save As Saves a file in the active window under a different file name.
Start Starts the report, test, analysis, or operation.
Table buttons Use to modify the table contents.
Clear. Clears all table entries and displays only one default value.
OPTION PRESENTATION
Options > Option Presentation
Use to change the way sample files and parameter files display: Advanced, Basic, or Restricted.
Each display option shows sample information and options differently.
If the AccuPyc II TEC application is installed, additional fields for inserting values for the sample cup
can be displayed. Go to Options > Option Presentation > Show Cup Properties to enable
these fields.
Specify or change the default option presentation by selecting Options > Option
Presentation, or select Basic or Advanced from the drop-down list at the bottom of the
window.
The following examples show the same sample information file in Advanced and Basic display. Basic
and Restricted displays will look the same.
A sample information file must be created for each analysis. The file can be created prior
to or at the time of analysis. The sample information file identifies the sample, guides the
analysis, and specifies report options.
KEYBOARD SHORTCUTS
Shortcut keys can be used to activate some menu commands. Shortcut keys or key combinations
(when applicable) are listed to the right of the menu item.
Certain menus or functions can also be accessed using the Alt key plus the underlined letter in the
menu command. For example, to access the File menu, press Alt + F, then press the underlined
letter on the submenu. For example, Alt + F opens the File menu, then press O to access the File
Selector for opening files.
If the underscore does not display beneath the letter on the menu or window, press the
Alt key on the keyboard.
When Sample Default files are created and saved, use the Replace All button when on a sample
file to replace values in the open file with values from the Sample Default file. Sample defaults can be
created and saved for sample files (.SMP), analysis conditions files (.ANC), and report options files
(.RPO).
ANALYZER STATUS
SHOW INSTRUMENT LOG
Unit [n] > Show Instrument Log
Use to display a log of recent analyses, calibrations, errors, or messages.
For fields and buttons not listed in this table, see Common Fields and But-
tons on page 3 - 2.
Closed Valve. Yellow indicates a closed valve with manual control enabled.
Closed Valve. White indicates a closed valve with manual control not enabled
Sample Chamber.
Expansion Chamber.
Each manually controlled schematic component has a shortcut menu displaying the operations
available for that particular component. To access the shortcut menu, hover the mouse pointer over
the component and right click.
SHOW STATUS
Unit [n] > Show Status
Use to show the current status for each port. If multiple units are attached to the computer, go to Unit
[n] > Show Status for the indicated unit.
For fields and buttons not listed in this table, see Common Fields and But-
tons on page 3 - 2.
Analyzing. Sample information files that are currently being used for
analysis.
For fields and buttons not listed in this table, see Common Fields and But-
tons on page 3 - 2.
OPEN A FILE
File > Open > [File Type]
When working with an existing file, it is recommended that a copy of the file be used
rather than the original.
Fields and buttons will vary depending on the file type selected.
analysis.
l Complete. Sample information files used in an analysis that has been
completed.
l No Analysis. Sample information files which have not been used to per-
form an analysis.
For fields and buttons not listed in this table, see Common Fields and But-
tons on page 3 - 2.
A sample information file consists of parameter sets; however, parameter sets can also stand alone.
A sample information file may be created either prior to or at the time of analysis.
Parameter files allow for repeated use of parameter sets. For example, if the same analysis
conditions exist for multiple analyses, an Analysis Conditions file containing the recurring conditions
can be created. When the sample file is created, the Analysis Conditions file can be selected for the
analysis conditions. Once it becomes part of the new sample file, the new file can be edited as
needed without affecting the original Analysis Conditions file.
The analysis software contains a default method. A method is a template for sample files that
contains the parameters to be used for an analysis. When a new sample information file is created,
all the parameters are filled with the values in the default method.
Specify or change the default option presentation by selecting Options > Option
Presentation, or select Basic or Advanced from the drop-down list at the bottom of the
window.
Each analysis must be linked with a sample information file before the analysis can proceed. A
sample information file can consist of parameter files; however, parameter files can also stand alone.
See About Parameter Files on page 5 - 1.
The values specified in the parameter portions of the default sample file are saved as the defaults for
new sample files. To navigate from one set of parameters to another, select the parameter tab
across the top of the window.
The Basic and Restricted formats use predefined parameter files to create a sample information file.
When using the Basic option presentation, switch to Advanced to edit parameter file
values. When using the Restricted option presentation, parameter files cannot be
edited.
Compute. Select the shape of the sample and enter the dimensions.
For fields and buttons not listed in this table, see Common Fields and But-
tons on page 3 - 2.
SAMPLE CUP
Options > Option Presentation > Show Cup Properties
File > Open > [.CUP File]
(or click the Sample Information tab when in Advanced option presentation)
For fields and buttons not listed in this table, see Common Fields and But-
tons on page 3 - 2.
MATERIAL PARAMETERS
File > Open > [.SMP File]
Material parameters are used to report porosity or total solids concentration when using the
Standard analysis method.
For fields and buttons not listed in this table, see Common Fields and But-
tons on page 3 - 2.
The following file types can exist as part of the sample information file as well as individual parameter
files:
The system default directory is the recommended location for saving report options files.
The default directory is ..\param, if accessed from the top-level installation directory.
ANALYSIS CONDITIONS
File > Open > [.ANC File]
(or click the Analysis Conditions tab when in Advanced presentation option)
The default is 500 psig for the High Pressure sample chamber.
Record P1 equilibration Record first. Enter the number of cycles to be recorded at the
data [check box] beginning of the analysis.
Record P2 equilibration
data [check box] Record last. Enter the number of points to be recorded at the end
of equilibration.
Use run precision Select to use run precision. This feature provides early termination
[check box] of the analysis when certain criteria are met. The analysis ter-
minates after five consecutive cycles are within the specified tol-
erance. Typically, a large number (50 to 99) of cycles is requested. If
a small number of runs is selected, the analysis stops when the
entered number is reached even though the specified tolerance has
not been met. If fewer than five cycles are requested, this feature is
disabled.
For fields and buttons not listed in this table, see Common Fields and But-
tons on page 3 - 2.
FOAMPYC APPLICATION
File > Open > [.ANC File]
Analysis Methods
Selection Required Action
Standard Select Record equilibration data then enter the criteria (if recording equilibration
data).
A, B, or C Enter the Resin density.
D Enter the initial and final pressure, and the pressure increment.
Enter Resin density.
E Enter the Low pressure and the Fracture pressure.
Enter Resin density.
REPORT OPTIONS
File > Open > [.RPO File]
(or click the Report Options tab when in Advanced presentation display)
If using the FoamPyc application, reports for all FoamPyc Methods are fixed
and cannot be created or selected from within the application.
Report options specify the types of reports to be generated from an analysis and allow for
customization of reports such as axis scale, axis range, and column headings.
Report Options files can be customized to accommodate standard analysis requirements. Reports
can be generated automatically after each analysis or at any time during or after an analysis. A report
generated during an analysis only includes data collected up to the time of the report.
Report Options files may be defined to include overlay options. This system allows the overlay of up
to 8 plots of different samples onto a plot of the same type or overlay one plot type onto a different
plot type from the same analysis.
For fields and buttons not listed in this table, see Common Fields and But-
tons on page 3 - 2.
COLLECTED DATA
The Collected Data tab displays after an analysis has been completed.
For fields and buttons not listed in this table, see Common Fields and But-
tons on page 3 - 2.
6 PERFORM AN ANALYSIS
An analysis can be performed from either the computer application or directly from the keypad on the
analyzer.
FOAMPYC METHODS
FoamPyc Methods are intended to conform to the procedures detailed in ASTM Test Methods D-
6226 and D-3576. Prepare samples according to the methods outlined in these procedures. Once
the sample is prepared and placed into the sample chamber, see Perform an Analysis above.
STANDARD METHOD
Samples must be properly prepared to obtain accurate results. Samples must be free of moisture in
order to obtain true sample mass and to avoid the distorting effect of water vapor on the volume
measurement. The following procedures are recommended; however, modifications may be
necessary for some materials:
lHeat-sensitive materials may have to be dried by long-time exposure to silica gel, freeze drying,
etc.
l Materials having a low melting point may be dried using the purge process. In this case, do not
weigh the sample and cup until after the purge and analysis have been completed.
Avoid exposure of the dried sample to atmospheric moisture during each step of the preparation
process. This means weighing as rapidly as possible and installing in the instrument without
unnecessary delay.
Keep the cap on the cell chamber except when actually inserting or removing a sample.
If the chamber remains uncapped, water vapor will adsorb on the inner surface of the
chamber and temperature instability will occur; either condition can affect analysis res-
ults.
4. Remove the sample cup from the oven and transfer it to a desiccator provided with active
desiccant. Allow it to cool until near room temperature. Minimize air exposure of the sample.
5. Use this equation to determine the sample mass. (If volume only is to be measured, skip this
step.)
6. Remove the cell chamber cap, place the sample cup (with sample) into the cell chamber, then
replace the cap.
It is best not to lay the cap down while loading the sample. Immediately replace the cap when
the sample is loaded. This will prevent particles from accumulating on the greased surface
(see Handling System Components on page J - 1 for guidelines on handling system
components).
7. If using a temperature-controlled AccuPyc, perform the following steps before proceeding,
see Perform an Analysis on the previous page.
a. Power on the bath circulator and specify the temperature for analysis. (Refer to the man-
ufacturer manual for instructions on operating the bath circulator.)
b. Allow the temperature to stabilize to the set temperature. Observe the temperature
reading on instrument schematic.
8. If using a TEC module, ensure the temperature reading has stabilized at the specified point.
PERFORM AN ANALYSIS
1. Select the sample file to run and click OK. To select a different sample file, click Browse and
locate the correct file. If sample file parameters exist in another file, rather than re-enter them
again, click Replace to locate and replace parameters in the current file with those of the selec-
ted file.
2. Edit or enter information for the analysis.
3. Click Report after analysis to generate reports automatically when the analysis is complete.
On the Report Settings window, select the report destination. Click OK to return to the pre-
vious window.
4. Click Export after Analysis analysis to have data exported automatically after the analysis.
Or, if exporting equilibration or cycle data, data can be exported after the analysis using the
File > Export.
5. Ensure the correct Chamber Insert is selected.
6. Click Start to start the analysis. A window displays data as they are collected. A short delay is
encountered before the port status at the bottom of the window changes from the Idle state.
7. If running a FoamPyc analysis, a message displays indicating the sample should be cut.
a. Remove the sample chamber cap, then remove the cap before performing the next
step.
b. Cut the sample and return it to the cup.
c. Remove the cap, place the cup back into the sample chamber. Replace the cap.
d. Click OK to complete the second analysis.
1. Press Alt + 4 to display the Sample ID prompt. Enter an appropriate identification. Press .
(decimal) to insert a dash.
2. Press ENTER to display the Description prompt.
Most users display this prompt only when a keyboard is attached to the pycnometer.
3. Press ENTER to display the Sample Mass prompt. At the Anls Display Mode prompt in the
Setup > Report Options choose Density. Enter the sample mass.
4. If an analytical balance is connected to the module, press the appropriate button on the bal-
ance to transfer the sample mass while this prompt is displayed.
5. Press ENTER to display the Chamber Insert prompt if using an insert. Press CHOICE until
the appropriate insert is displayed, then press ENTER.
6. Press ENTER to begin the analysis. Operational status messages display during analysis.
7. When the analysis has finished, the pycnometer beeps three times and the display returns to
the Reload prompt.
Do not remove the cell chamber cap when the pycnometer is pressurized. The sample
may be discharged from the chamber.
8. Ensure the pressure reads approximately 0 (zero), then remove the cell chamber cap and
remove the sample from the chamber.
9. Replace the cell chamber cap or load another sample.
Display Text Description
Analyze Sample ID: Enter sample identification for the current analysis. Sample ID can be
up to 20 numbers and dashes. Press . (decimal) to insert a dash.
Analyze Enter the description for the current sample. A prompt for Line 2 of the
Description Line 1: description is displayed.
This field will also accept input from a connected analytical balance.
While this prompt is displayed, press the appropriate button on the
analytical balance to transfer the mass (refer to the manufacturer’s
manual for the appropriate command).
Analyze Chamber Select the insert to be used.
insert?
l None, 0.1 cm3 (for 1 cm3 unit)
l None, 3.5 cm3, 1.0 cm3 (for 10 cm3 unit)
l None, 35 cm3, 10.0 cm3 (for 100 cm3 unit)
l None, 650 cm3, 1300 cm3 (for 2000 cm3 pycnometer)
Press CHOICE until the appropriate insert is displayed or None if an
insert is not being used.
Analyze Cup Mass? Enter the mass of the sample cup.
Analyze ENTER. Starts the analysis. The analysis begins and operational
[ENTER] to start status messages are continually displayed during analysis
[ESCAPE] to cancel
ESCAPE. Cancels the analysis.
For keypad and/or keyboard functions or commands not listed in this table,
see About the Keypad on page 2 - 2.
QUICKSTART ANALYSIS
Unit [n] > QuickStart Analysis
Use to analyze a series of samples of the same type and same analysis conditions. Sample files will
be created using default values.
For fields and buttons not listed in this table, see Common Fields and But-
tons on page 3 - 2.
SEQUENCE ANALYSIS
Unit [n] > Sequence Analysis
Use to analyze the same sample using up to eight different sample files.
For fields and buttons not listed in this table, see Common Fields and But-
tons on page 3 - 2.
REVIEW ANALYSIS
Keypad Function
The Review function allows review of the results of the last analysis operation. Press Alt + 5 to
review or edit analysis results. The analysis information can also be edited through the prompts. The
prompts that display during the analysis review depend on the options selected in Setup > Analysis
Parameters and Report Options.
1. Press Alt + 5 to display the Sample ID prompt containing the sample identification.
2. Press ENTER to display the Description prompt containing an additional description. This
prompt typically is not used unless a keyboard is connected to the analyzer.
3. Press ENTER to display the Sample Mass prompt containing the sample’s mass.
4. Press ENTER. The Start prompt containing the start date and time of the analysis is dis-
played.
5. Press ENTER. The End prompt containing the ending date and time of the analysis is dis-
played.
6. Press ENTER. The first of the prompts containing data is displayed. The [n] displayed in the
example prompts represents the cycle number.
7. Continue pressing ENTER to view data. Press CHOICE to exclude data from report cal-
culations. An asterisk will display indicating that it will be excluded. Press CHOICE again to
remove the asterisk.
8. Press SAVE to return to the Reload prompt.
9. Press Alt + 6 to print report results.
Display Text Description
Sample ID: Identification entered for the analysis.
(sample identification)
Sample Mass: Sample mass for the analysis (if Density was chosen as Analysis
(mass) display mode).
Temperature: Sample chamber temperature.
(sample chamber temp)
V[n] = (volume) Press CHOICE to exclude the displayed density (or volume) from the
Dv[n] = (deviation) calculated average. An asterisk indicates that it has been excluded.
Press CHOICE again to remove the asterisk. Each time CHOICE is
pressed to exclude (or include) the value, a new deviation is calculated
and displayed.
For keypad and/or keyboard functions or commands not listed in this table,
see About the Keypad on page 2 - 2.
7 ABOUT REPORTS
Review this section for information on the Reports menu options as well as customizing and running
reports.
START REPORTS
Reports > Start Report
Starts the selected report. Select a file from the Files list. Ensure the selected file has a status of
either Complete or Analyzing.
SPC REPORT
Reports > SPC Report Options
Use to generate reports with various SPC (Statistical Process Control) options. All selected
variables must be computed for each sample file used in an SPC report; therefore, it is more efficient
to select only the necessary variables.
The selected items display as graph variable selections in Reports > Regression Report and
graph selections in Reports > Control Chart. If additional report options are required, click More.
REGRESSION REPORT
Reports > Regression Report
Use to generate a Statistical Process Control (SPC) Regression report to determine the
interdependency between two variables. Up to three dependent variables (y-axis) may be plotted
against a single independent variable (x-axis). The degree of correlation between the variables is
also reported.
The first time this option is used, the time it takes to generate the report
is lengthened. The second time the report is generated, if using the
same sample files used in the initial calculation, it is recommended that
this option not be selected since the data was recalculated previously.
If a sample file is added or removed from the report after the initial
recalculation, this option should be selected again to ensure the data
from the newly added or removed sample file is recalculated.
Samples [button] Select additional sample files to add to the report.
Save as Default [button] Click to save selected report options as default report settings.
Show graphic Use to show a graphic on the report header.
[check box] Height / Width. Enter the height and width of the selected graphic.
These values determine the graphic appearance on the generated
report.
Show report title Select and enter a report title to appear on the report header.
[check box]
Tabular report Generates a tabular report of the included samples. A tabular report
[check box] contains the numeric values contributed by each sample.
For fields and buttons not listed in this table, see Common Fields and But-
tons on page 3 - 2.
Center Line. Displays placement options for the center line in the
graph. Choose Entered to specify placement of the line.
Limit Lines group box. Displays limiting lines options. Lines can be
placed at some multiple of the standard deviation or at specified
positions (Entered). When Entered is selected, enter the High limit and
Low limit fields with appropriate values.
The first time this option is used, the time it takes to generate the report
is lengthened. The second time the report is generated, if using the
same sample files used in the initial calculation, it is recommended that
this option not be selected since the data was recalculated previously.
If a sample file is added or removed from the report after the initial
recalculation, this option should be selected again to ensure the data
from the newly added or removed sample file is recalculated.
Report [button] Generates the report.
Samples [button] Select additional sample files to add to the report.
Show graphic Use to show a graphic on the report header.
[check box] Height / Width. Enter the height and width of the selected graphic.
These values determine the graphic appearance on the generated
report.
Show report title Select and enter a report title to appear on the report header.
[check box]
Tabular report Generates a tabular report of the included samples. A tabular report
[check box] contains the numeric values contributed by each sample.
Minutes. Minutes elapsed from the first file placed on the list, which is
the earliest-analyzed file.
Days. Number of days elapsed from the first file placed on the list,
which is the earliest-analyzed file.
For fields and buttons not listed in this table, see Common Fields and But-
tons on page 3 - 2.
l After analysis, reports can be viewed, printed, and/or copied and pasted into other documents.
l The report zoom feature provides the viewing of fine graph details and the ability to shift the axes.
l All reports contain a header displaying file statistics.
If configured, the report header can also contain a graphic and/or a title.
l Tabular and graphical reports contain sample and analyzer statistics such as analysis date / time,
analysis conditions, etc.
l The headers contain notes of sample file changes occurring after analysis.
l Summary report headers contain the same information as tabular and graphical reports with the
exception of notes.
If configured, the report header can also contain a graphic and a title.
l Tabular and graphical reports contain sample and instrument statistics — such as analysis date,
analysis time, analysis conditions, etc.
l The headers contain notes of sample file changes occurring after analysis.
l Summary report headers contain the same information in tabular and graphical reports with the
exception of notes.
Report header Shortcut Field and Button Table
Field or Button Description
Copy header as text Use to copy the report header as text. Text is copied to the clipboard
and then can be pasted into other documents.
Edit Use to edit the report title and/or graphic in the report header.
REPORT TOOLBAR
The Report window has a toolbar on the right portion of the window and selectable tabs at the top of
the report header. To view a specific report, either select the tab or the report in the Reports list box,
then click Show.
For fields and buttons not listed in this table, see Common Fields and But-
tons on page 3 - 2.
For fields and buttons not listed in this table, see Common Fields and But-
tons on page 3 - 2.
Scale font. Use to modify the font for the scale label. Deselect Use
default font to enable font options.
Title font. Use to modify the font for the selected axis label. Deselect
the Use default font to enable font options. Select new font attributes
for the report data. To return to the default fonts, enable Use default
font.
Edit curve Use to edit selected curve properties.
l Curve group box. Use to change the interpolation, point style and
pen style for the selected curve. These options are disabled if Use
default fill style is selected in the Histogram group box.
l Histogram group box. Enabled only if Histogram is selected in
the Style drop-down list. Use to specify the type of fill, fill color and
label position for the selected curve.
Label. Select where the graph point labels will display (left, right,
center, etc.) on the SPC report.
Style. Use to select another style for the collected data curve.
Edit title Use to change the graph title and font. Click Font to font attributes.
Deselect the Use default font to enable font options. Select new font
attributes for the report data. To return to the default fonts, enable Use
default font.
Hide curve Hides selected curve.
Redraw Sets the axis boundaries to the original view. This option also removes
cross-hairs.
Show curve Displays a list of all curves. Select the curve to display.
For fields and buttons not listed in this table, see Common Fields and But-
tons on page 3 - 2.
AXIS CROSS-HAIR
Left click on the graph to view the cross-hair coordinates.
Use to select how grid lines appear on reports. This menu option is not available if using Restricted
option presentation.
For fields and buttons not listed in this table, see Common Fields and But-
tons on page 3 - 2.
ZOOM FEATURE
Use the zoom feature to examine graph details. Click, hold, and drag the left mouse button on the
graphical area to be enlarged. A box will display in the area to be enlarged.
l Multiple Graph Overlays. Overlay two different types of graphs from one sample.
l Multiple Sample Overlays. Overlay graphs of the same type with that of the current plot.
This feature is available only when using Advanced option presentation. Go to Options
> Option Presentation > Advanced.
4. Click the down arrow to the right of the Overlay field and select a graph. This selection will over-
lay the graph selected in the Y-axis field.
Use a copy of a file rather than the original file when performing a sample overlay.
1. Go to Options > Option Presentation and ensure there is a checkmark to the left of
Advanced.
2. Go to File > Open > Sample Information and open the sample file.
3. On the Report Options tab, in the Selected Reports list box, select the reports to be used by
double clicking the report name. This places a checkmark to the left of each selected report.
4. Highlight the graph to be used for multiple overlays, then click Edit.
5. On the overlay window, click the down arrow to the right of the Overlay field and select the
Overlay Samples Select other options as needed, then Click OK.
7. Om the Graph Overlays window, click Browse to locate and select which graphs are to be
overlaid. Click OK.
8. Go to Reports > Start Report. Select the report destination in the Settings group box. Select
the overlay file name in the Files list box, then click OK.
9. In the pop-up Selected Reports list box, reports can be added or removed by double clicking
the entry. Selected reports as indicated by a checkmark to the left. Click OK.
10. Reports display in a tabbed format. Click each tab to view each report.
REPORT EXAMPLES
COMBINED REPORT
SUMMARY REPORT
TABULAR REPORT
GRAPH REPORT
OVERLAY REPORT
EQUILIBRATION REPORT
8 SELECTED REPORTS
COMBINED REPORT
The Combined Reports option creates a single report from all selected reports. The Combined
Report contains no breaks between reports and no headings. Highlight the report, then click Edit to
edit report details.
For fields and buttons not listed in this table, see Common Fields and But-
tons on page 3 - 2.
There are four types of graphical reports available. Report titles are based on the parameters
selected for the X-axis and Y-axis.
Density vs Cycle Number Trend Plot Fields and Buttons Table (continued)
Field or Button Description
Plot curve Select to display how data are plotted.
Plot points
[check box]
Show +/- Select to place two horizontal lines on the graph. The top line will be
[check box] placed at a point relative to the average and standard deviation of the
data. The bottom line will be placed at the point entered in the adjacent
field.
Show Average Select to show the average of all cycles. A horizontal line is placed at
[check box] the average point for the cycles.
Show Min Select to show the average of all cycles. Horizontal lines are placed at
Show Max the entered limits.
[check box]
Show Reference Select to show a horizontal line placed at a reference point specified in
[check box] the adjacent field.
X-Axis [check box] Variables available for the X-axis.
Y-Axis [check box] Variables available for the Y-axis.
For fields and buttons not listed in this table, see Common Fields and But-
tons on page 3 - 2.
The Density vs Cycle Number Trend Plot and the Volume vs Cycle Number are identical unless
otherwise noted. See Density vs Cycle Number Trend Plot on the previous page.
EQUILIBRATION REPORT
This report is available only if the Record equilibration data option was selected on the Analysis
Conditions tab.
For fields and buttons not listed in this table, see Common Fields and But-
tons on page 3 - 2.
OPTIONS REPORT
The Options report provides information on the sample, its material properties, and analysis
conditions. This report cannot be edited.
The Pressure and Volume Table and the Density and Volume Table are identical unless otherwise
noted. See Density and Volume Table on page 8 - 1.
l Automatic operations
l Information entered using Add Log Entry on the Sample Information tab
l Warnings and/or errors which occurred during analysis
SUMMARY REPORT
The Summary Report provides a condensed listing of data results. Select the type of information to
include in the report. If Pass/Fail is selected, the Min and Max fields are enabled. Enter the minimum
and maximum values for the pass / fail.
TABULAR REPORT
For fields and buttons not listed in this table, see Common Fields and But-
tons on page 3 - 2.
The Total Pore Volume vs Temperature and the Density vs Cycle Number Trend Plot are identical
unless otherwise noted. See Density vs Cycle Number Trend Plot on page 8 - 2.
The Volume vs Cycle Number and the Density vs Cycle Number Trend Plot are identical unless
otherwise noted. See Density vs Cycle Number Trend Plot on page 8 - 2.
For fields and buttons not listed in this table, see Common Fields and But-
tons on page 3 - 2.
10 ABOUT CALIBRATION
The pycnometer volume should be calibrated every time it is restarted. Cell volume and expansion
volume in the set up parameters are updated automatically upon calibration. Run an analysis with an
empty cup to see how close the average volume is to 0. It should be ±0.05% of full scale. If it is not
within ±0.05% of full scale, calibration is required.
If calibrating using the keypad, see Calibrate using the Keypad on page 10 - 12.
Calibrates the temperature offset. Use an external reference device to determine the sample cell
temperature. Enter the value in the field, then click Done.
A Calibration report is not generated for the temperature offset. The entry is recorded in the
Instrument Log. To view the log, go to Unit [n] > Show Instrument Log.
For fields and buttons not listed in this table, see Common Fields and But-
tons on page 3 - 2.
This option can be used when the pressure appears not to be reported correctly by the analyzer. For
example: a zero, negative, or unreasonably high reading is being consistently reported. Resetting to
nominal may restore normal pressure readings but a proper calibration should be performed.
A popup warning indicates that when this option is used, the previous pressure calibration is lost and
the nominal calibration is used.
Temperature and volume calibrations are not affected by a reset of the pressure cal-
ibration.
Click OK. The nominal calibrations may be sufficient. However, it is recommended to contact your
Micromeritics Service Representative and schedule a proper pressure calibration as soon as
possible since data accuracy may be compromised.
Although the instrument is calibrated prior to shipment, the pycnometer should be recalibrated to
laboratory atmospheric and environmental conditions.
1. On the pressure regulator, turn the pressure control knob clockwise to set the regulator pres-
sure at 21.500 psig (low pressure gauge). For a High Pressure sample chamber con-
figuration, set regulator pressure 20 psi above cycle and fill pressure.
2. Remove the chamber cap and place an empty cup into the sample chamber, then replace the
cap. If an insert is being used, place the insert sample cup into the chamber.
Wear latex or nitrile, powder-free gloves to prevent transfer of oil from hands when
handling sample cups, calibration standards, and inserts. See Handling System
Components on page J - 1.
4. Enter the volume of the calibration standard (located on the outside of the calibration stand-
ards case shipped with the instrument) in the Volume of Calibration Standard field.
The value recorded on the case for a 10 cm3 pycnometer is the sum of the two standards, and
on a 350 cm3 pycnometer, the sum of the three standards.
5. Select the Measured option.
6. Click Analysis Conditions to specify calibration parameters. Click Replace and select the
file calstd.anc. The values in the current file will be replaced with those contained in the selec-
ted file. Click OK, then click OK again to return to the Calibrate Volume Scale window.
7. Click Start to begin the first phase of the calibration (volume offset). Status messages display
until this phase is complete. When the phase has finished, a prompt display indicates to insert
the calibration standard.
8. Remove the sample chamber cap, then remove the sample cup from the chamber. Place the
cap back onto the chamber while placing the standard into the cup.
9. Tilt the sample cup and allow the standard to roll gently into the cup to prevent denting of the
cup.
l one standard is shipped with the 1-, 100-, and 2000 cm3 pycnometers
l two are shipped with the 10 cm3 pycnometer
l three are shipped with the 350 cm3 pycnometer
10. Remove the sample chamber cap, place the sample cup back into the chamber, and replace
the cap. Turn the cap clockwise to tighten.
11. Allow the standards to reach thermal equilibrium with the pycnometer.
l Wait 10 minutes for 1- and 10 cm3 pycnometer
3
l Wait 15 minutes for the 100 cm pycnometer l
Entered. Enables the fields for the cell and expansion volumes.
For fields and buttons not listed in this table, see Common Fields and But-
tons on page 3 - 2.
VERIFY OPERATION
Verify operation of the pycnometer by performing an analysis on the calibration standard to ensure
that the pycnometer is operating properly.
4. Click Report After Analysis. Select the Report after analysis option, then click OK.
Example: calculate the tolerance for a 10 cm3 pycnometer with a value of 6.372242
recorded on the case as :
Tolerance = (10 * 0.0003) + (6.372242 * 0.0003) = 0.003 + 0.0019116726 = 0.005
If the values do not agree, a leak may be indicated. Check the system for leaks. Then repeat
steps 1 through 6. If the values continue to disagree, contact your Micromeritics service
representative. See Diagnostics using the Software on page 9 - 1.
7. Remove the standards and place them back into the case.
It is recommended that the current calibration settings be saved using Unit [n] > Calibration >
Save to File prior to loading another calibration file. When loading a previously saved calibration file,
a backup of the current file is created and saved as [SN]last.cal. The backup file is overwritten each
time a new one is created.
The default file naming convention for calibration files can be used or the file name can be changed.
The default file name of 0217-2013-04-25.CAL is interpreted as:
CALIBRATION DATA
Calibration Data Displays the cell volume determined during calibration Cell volumes
Cell Volume: can also be entered at this prompt. Press ENTER.
CALIBRATE VOLUME
Place an empty cup into the sample chamber for Cal1 (volume offset). If using an insert, place the
insert and its appropriate sample cup into the sample chamber.
Calibrate Volume Enter the volume of the calibration standard to be used in the
Volume of Cal Std: calibration (located on the outside of the calibration standards case).
Some pycnometers require and are shipped with multiple standards.
In this instance, the value recorded on the case is the sum for all
standards and is the one that should be entered at the prompt. Press
ENTER.
This option can be used when the pressure is suspected of not being reported correctly by the
analyzer. For example: a zero, negative, or unreasonably high reading is being consistently
reported. Resetting to nominal may restore normal pressure readings but a proper calibration should
be performed.
REVIEW CALIBRATION
The Review function reviews the results of the last calibration operation.
The sample cell volume and the expansion cell volume are used for calculating the sample volume.
The cell volume and expansion volume are updated automatically when the pycnometer is
calibrated.
The prompts that display during a review of calibration data depend on the options specified in Setup
> Analysis Parameters and Report Options.
1. Press Alt + 5 to display the Chamber Insert prompt from which the insert can be selected (if
used).
2. Press ENTER to display the volume of the calibration standard. This volume can be edited
from this prompt.
3. Press ENTER to display the Which Chamber prompt.
4. Press CHOICE to select whether to view the results for the sample chamber or the expansion
chamber.
5. Press ENTER to display a prompt showing the starting time and date.
6. Press ENTER to display a prompt showing the ending time and date.
7. Press ENTER to display the first of the prompts containing data. The [n] in this example rep-
resents the cycle number.
8. Continue pressing ENTER to view all of the data. Press CHOICE to exclude data from report
calculations. An asterisk will display indicating that it will be excluded.
9. Press SAVE to return to the Reload prompt
10. Press Alt + 6 to print report results.
Display Text Description
Chamber insert? Displays the insert selected for the calibration or None if an insert was
(insert) not used. If the selection is incorrect, press CHOICE until the correct
one is shown, then press ENTER.
The chamber type options are Cell volume and Expansion volume.
Start: Displays the time and date the calibration began. Press ENTER.
(time) (date)
Time: HH:MM:SS
Date: DD/MM/YY
End: Displays the time and date the calibration completed. Press ENTER.
(time) (date)
Time: HH:MM:SS
Date: DD/MM/YY
Cell[n] = (volume) Displays the Volume, depending on the chamber type selected (Cell
Dv[n] = (deviations) volume or Expansion volume).
Exp[n] = (volume) Press CHOICE to exclude the displayed density (or volume) from the
Dv[n] = (deviations) calculated average. An asterisk next to the density indicates it has
been excluded. Press CHOICE again to remove the asterisk and have
it included.
TEMPERATURE
This operation typically is performed by a Micromeritics service representative.
Does not display for l None, 0.1 cm3 (for 1 cm3 unit)
Pressure 3 3 3
l None, 3.5 cm , 1.0 cm (for 10 cm unit)
l None, 35 cm3, 10.0 cm3 (for 100 cm3 unit)
l None, 650 cm3, 1300 cm3 (for 2000 cm3 pycnometer)
Zero Volume Press ENTER to begin the zero operation or Alt + CLEAR to cancel
[ENTER] to start the operation.
[ESCAPE] to cancel
If zeroing the Volume, the purges and cycles specified in the current
or analysis parameters are used to calculate the volume offset.
Zero Pressure After the operation is complete, the pycnometer returns to the Reload
[ENTER] to start prompt.
[ESCAPE] to cancel
11 HARDWARE
Diagram Components
Component Description
A Separate power cord for external power supply. Connect cable to wall outlet.
B External power supply
C Hardwired 8-pin mini-DIN power cable for external power supply.
D 8 pin mini DIN cable.
E Power adapter DIN port.
F Power adapter.
G Module Connection - 8-pin mini-DIN OUT port to connect next analysis module.
H Module Connection - 8-pin mini-DIN IN port to connect the control module or pre-
viously connected analysis module (not shown).
I Hardwired power mini-DIN cable for power adapter.
If installing an external TEC module, it is suggested that the TEC system be installed
first, to allow time for temperature to stabilize at the desired temperature, while the
AccuPyc is being installed. The TEC module has a separate power cord and external
power supply.
10. Calibrate the pycnometer. See Calibrate using the Keypad on page 10 - 12 or Calibrate
using the Software on page 10 - 1.
ANALYTICAL BALANCE
The analytical balance tested by Micromeritics is Scientec® model #ZSA120.
An analytical balance can be connected to a serial port on the computer and used to transfer the
sample’s mass via a third-party software directly to the AccuPyc II application. Refer to the balance
manufacturer’s operator’s manual for transmission instructions.
1. Connect the analytical balance to the RS-232 port on the rear panel of the analyzer.
2. Go to Options > Configure Balance to connect and set up an RS-232 balance to the ana-
lyzer.
PRINTER SETUP
If using the Keypad only, see Printer Setup on page J - 10. This step is not required if using the
software.
Power indicator did not illuminate when analyzer was powered ON.
Fine fibers and particles between the O-ring and its sealing surfaces can cause leaks, as can
scratches or cuts in the O-ring or in the metal surfaces.
Do not to scratch the metal surface of the chamber cap. Scratches could result in an
imperfect seal.
2. Clean the groove in the chamber cap using a small brush or lint-free tissue moistened with iso-
propyl alcohol.
3. Allow the chamber cap to dry thoroughly.
4. Use a small drop of Dow Corning high vacuum grease (or equivalent).
5. Grasp the O-ring with the two greased fingers. Distribute the grease evenly and completely
around the O-ring.
Apply the grease sparingly. Too much grease may alter cell volume while too little
grease results in an imperfect seal and leaks.
6. Place the O-ring back into the groove on the cap and, with the greased index finger, gently
press it back into position.
7. Ensure the O-ring groove is properly greased. See Grease the Chamber Cap O-Ring on
the previous page.
8. Replace the chamber cap.
Before performing this procedure, check the chamber cap to ensure that it is not the
source of leaks. It should be free from particles, the O-ring should be properly seated,
and it should not contain excessive grease.
Before performing this procedure, check the chamber cap to ensure that it is not the
source of leaks. It should be free from particles, the O-ring should be properly seated,
and it should not contain excessive grease.
psig (0.14 kPag) in a two minute span, temperature instability or a leak may be indicated.
Vent the system, then repeat steps 4 through 7 several times to verify that a leak is indic-
ated. If a leak is indicated, call a Micromeritics Service Representative.
8. Press 9 to close the Vent valve, 8 to open the Expansion valve, and 7 to open the Fill valve;
O - O - X.
9. Fill the chambers to 19.5 psig.
10. Press 7 to close the Fill valve; X - O - X.
11. Observe the pressure display. After an equilibration period (about 20 to 30 seconds), the pres-
sure should not decrease more than 0.007 psig (0.048 kPag) or increase more than 0.02 psig
(0.14 kPag) in a two minute span.
If the pressure decreases more than 0.007 psig (0.048 kPag) or increases more than 0.02
psig (0.14 kPag) in a two minute span, temperature instability or a leak may be indicated. Vent
the system, then repeat steps 8 through 11 several times to verify that a leak is indicated. If a
leak is indicated, call a Micromeritics Service Representative.
Applicable only to 2000 cm3 units and 100 cm3 high pressure units.
Dust from the sample may be carried into the instrument plumbing. A dust filter protects the valves,
minimizing valve leaks.
Do not immerse the pycnometer or the power cord in any liquids. Doing so could result
in electrical shock to personnel or damage to the unit.
Do not allow liquid to penetrate the casing of the pycnometer. Doing so could result in
damage to the unit.
l Always use the gas lines provided with the analyzer. It is very important that proper gas lines are
l Long gas lines, such as those used with gas cylinders placed in remote areas, must be evacuated
for an extended period of time to remove ambient gases. When possible, avoid placing gas cyl-
inders in remote locations. It is always better to have gas cylinders located near the analyzer.
l Carefully route the gas lines from the cylinder to the analyzer avoiding overlapping or entangling
gas lines. This will help avoid confusion when maintenance is required.
l Label the gas line at the analyzer inlet for proper identification and maintenance.
l Replace gas cylinders before gas is depleted. It is best to replace a gas cylinder when the pres-
sure reads approximately 200 psi (1500 kPa) on the high pressure gauge. Contaminants
absorbed to the walls of the cylinder will desorb as the pressure decreases.
l Ensure the gas cylinder is closed before connecting to the analyzer.
These instructions apply to working with inert gases only. When working with hazardous
gases, follow the safety procedures established by your lab.
A power failure can result in dangerous pressures in the sample chamber. When using
toxic or flammable gases, additional venting of the cabinet may be required.
Move the replacement bottle close to the analyzer and tether it into place. It is not necessary to
disconnect the gas line from the regulator or the analyzer.
1. Use an appropriate cylinder wrench to remove the protective cap from the replacement gas
cylinder. Place the protective cap in a secure location. It will be needed to recap the gas cyl-
inder when it is depleted and replaced.
2. Attach the gas regulator to the gas cylinder connector. Hand tighten the nut, then use an
appropriate wrench to tighten an additional 3/4 turn.
3. Check for leaks at the high-pressure side of the regulator and in the connector.
a. Turn the regulator control knob fully counter-clockwise.
b. Slowly open the gas cylinder shut-off valve, then quickly close it.
c. Observe the pressure on the high-pressure gauge for approximately one minute:
l If the pressure is stable, proceed with the next step.
l If the pressure decreases, tighten the regulator connector nut until it becomes stable.
If the pressure does not remain stable, remove the regulator and clean all contacts at
the regulator connection, then reinstall the regulator.
4. Purge the lines and set the regulator pressure.
a. If the gas line is not connected to the regulator and the instrument connect it now.
b. Fully open the gas bottle shut-off valve and the gas regulator shut-off valve.
c. Under manual control, open the fill, expand, and vent val
d. Adjust the regulator pressure control knob until the Purge fill or Cycle fill pressure,
whichever is higher, is shown on the regulator display.
e. Allow gas to flow for 30 seconds.
f. Close the fill, expand, and vent valves.
g. Turn the regulator control knob clockwise to increase pressure by 2 psi (15 kPa).
REGULATOR PRESSURE
Tank pressure should be at least 200 psi above regulator pressure. Pressures less than
200 psig indicate the tank is low on gas. Analyses are terminated automatically if gas is
depleted.
For a High Pressure sample chamber configuration, set regulator pressure 20 psi above
cycle and fill pressure.
The Purge fill and Cycle fill pressures are specified in Analysis Conditions. Set the regulator
pressure at the higher of the two, plus 2.0 psig. For example; if the Purge fill pressure is set to 19.500
psig (134.445 kPag) and the Cycle fill pressure as 19.000 psig (131.000 kPag), set the regulator
pressure at 21.500 psig (148.237 kPag).
1. Go to Unit [n] > Enable Manual Control. Ensure a checkmark displays to the left of the
menu item. If the instrument schematic does not display, go to Unit [n] > Show Instrument
Schematic.
2. On the schematic, right click the Expansion and Vent valves and select Open for both. Then
open the Fill valve.
3. Adjust the regulator pressure control knob until the Purge fill or Cycle fill pressure, whichever
is higher, is shown on the regulator display. Tank pressure should be at least 200 psi above
regulator pressure.
4. On the schematic, close the Fill valve, then increase the regulator pressure by 2.0 psig (13.8
kPag).
5. Allow the pressure in the pycnometer to drop below 2.0 psig, then close the Expansion and
Vent valves.
Ensure the tank pressure for the gas regulator is at least 200 psig. Pressures less than
200 psig indicate the tank is low on gas. Analyses are terminated automatically if gas is
depleted.
The Purge fill pressure and cycle fill pressure are specified using Setup > Analysis Parameters. Set
the regulator pressure at the higher of the two (plus 2.0 psig). For example; if the purge filll pressure
is specified as 19.500 psig and the cycle fill pressure as 19.000 psig, then set the regulator pressure
at 21.500 psig.
Even though the pycnometer saves data during a power failure, any operation should be
restarted to ensure complete results.
Keypad Function
l Hold the . key on the keypad during power ON. This erases data and setup parameters except for
printer and network settings.
l Hold 5 on keypad during power ON. This resets setup parameters, as well as printers and net-
work settings.
Depth listed is for solid materials. There is a small hole near the top of the cup to allow gas to
*
flow. This feature reduces the useful depth for powders by approximately 2 to 3 mm.
Depth listed is for solid materials. There is a small hole near the top of the cup to allow gas to
**
flow. This feature reduces the useful depth for powders by approximately 1 mm.
Depth listed is for solid materials. There is a lifting notch near the top of the cup. This feature
***
reduces the useful depth for powders by approximately 1.5 cm.
A CALCULATIONS
l Calculations for the 1340 AccuPyc
http://www.micromeritics.com/Repository/Files/Calculations_-_AccuPyc_II_Series.pdf
ANALYSIS
For derivation, see the Sample Volume Equation Derivation section of the AccuPyc II Series
Operator Manual.
ms
ρs=
Vs
where
where
ASPHALT DENSITY
Sample volume at 60 °F is found by multiplying the measured volume by the asphalt volume
correction factor.
where
Sample density at 60 °F is
p(60) = m / V (60 )
Specific gravity at 60 °F is calculated by dividing the adjusted sample density by the density of water
at 60 °F.
where
3
ρH20 = 0.990170 g / cm
Note that the density of water at the analysis temperature is not required for this calculation.
CALIBRATION
Volume calibration uses the ultra-precise method of separate adjustments for the offset and scale
factor.
VOLUME OFFSET
Vc = Vcelprev − Vsampempty
−Vsampempty
is reported as the offset in volume calibration reports.
VOLUME SCALE
Vc = Vcelprev Vcalib
V
sampball
Vx = Vexpprev Vcalib
V
sampball
Vcalib
V
sampball
is reported as the scale factor in volume calibration reports.
where
FOAMPYC METHODS
ResinPlusClosedCellVol
ResinPlusClosedCellPct = × 100
GeomVol
OpenCellVol
OpenCellPct = × 100
GeomVol
where
OpenCellVol
OpenCellPct = × 100
GeomVol
AvgVolChangeP2
PctAvgVolChangeP2 = × 100
GeomVol
∆Vol
PctFracturedCells = × 100
GeomVol
where
CUBE
3
GeomVol = EdgeLength × NumPieces
2
ActiveArea = EdgeLength × (6 − NumSk ins ) × NumPieces
CYLINDER
Diam2
GeomVol = π × × Height × NumPieces
4
π
ActiveArea = π × Diam × Height +
2
× (2 − NumSk ins ) × NumPieces
× Diam
4
RECTANGLE
GeomVol = LongEdge × ShortEdge × RemainEdge × NumPieces
PERCENT POROSITY
percent porosity = (ρs − ρ ) ÷ ρs
RESIN VOLUME
ms
ResinVol =
ResinDensity
ResinVol
ResinVolPct = × 100
GeomVol
where
ms = sample mass
RUN PRECISION
Run precision requires at least five runs. Run precision criterion is met when the sample volumes
calculated for the four most recent previous runs fall within the specified error band for the current
run’s sample volume. The error band is a specified percentage of the nominal volume of the sample
chamber (1, 10, 100, 350 and/or 2000 cm3).
Vs[0] to Vs[4] are the five most recent sample volumes, Vs[4] being the most recent.
V
ErrBand = PctFullScale × nominal
100
where
If
( Vs[i ] − Vs[4 ] ≤ ErrBand) for i = 0 to 3, run precision is achieved.
Σχi
χ =
N
Σ yi
y =
N
Slope =
(
Σ χi − χ )(yi − y )
(
Σ χi − χ 2 )
Intercept = y − Slope x
The coefficient of correlation for this line is also calculated in the usual way. 2 )
σχ =
σy =
Cov(x, y ) =
( )(
Σ x i − x yi − y )
N
Cov(x, y )
Correlation Coeff =
σxσ y
1 ) BASIC Scientific Subroutines Vol II, by F.R. Ruckdeschel, Copyright 1981 BYTE
Publications/McGraw Hill, p. 16.
2 ) Mathematical Handbook for Scientists and Engineers, G.A. Korn and T.M. Korn, McGraw
Hill, Sec. 18.4. (1968)
Standard Deviation =
N−1
C. V.= StdDev
Mean
SPECIFIC GRAVITY
SG = ρs ÷ ρw
where
where
where
Cause: Start Report failed to execute the report subsystem (which is a separate
process).
Action: Restart the computer. If the problem persists, reinstall the application (this will
not affect any of your sample files). Contact a Micromeritics service rep-
resentative if this error message continues.
2204 | Communication attempt with [n] failed [n]. Please check application directory for
file.
Cause A: User logged onto a computer that does not have Write permissions.
Action A: Contact your IT department to change user permissions.
2433 | New entries have been found in this directory. Refresh the directory information?
Cause: Several analyzer files (sample information, analysis conditions, adsorptive prop-
erties, or report options) have been added to this directory by some function
other than the analyzer program.
Action: Click Yes to update the directory information with data from each new file. This
operation may take a minute.
Click No to locate the file manually. This option may be feasible if a large
number of files have been copied into the directory and the file name is known.
2434 | File [n] — Subset [n] wrote wrong [n] of data, expected [n] bytes.
Cause: The overlay file selected in the report option does not exist.
Action: Enter an existing file specification, or select a file name from the list box.
Cause: An unexpected error occurred when you tried to access a data file.
Action: Contact your Micromeritics service representative.
2448 | Basic Mode default parameter file directory [n] is invalid. Resetting to the install-
ation directory.
Cause: A working directory specified in the .INI file is invalid, has been moved, or has
been deleted.
Action: The installation directory will be substituted. The next time a file is opened, use
the directories list to move to the correct directory.
2448 | Default Adsorptive Properties directory [n] is invalid. Resetting to the installation
directory.
Cause: A working directory specified in the .INI file is invalid, has been moved, or has
been deleted.
Action: The installation directory will be substituted. The next time a file is opened, use
the directories list to move to the correct directory.
2448 | Default parameter file directory [n] is invalid. Resetting to the installation directory.
Cause: A working directory specified in the .INI file is invalid, has been moved, or has
been deleted.
Action: The installation directory will be substituted. The next time a file is opened, use
the directories list to move to the correct directory.
2448 | Default report options directory [n] is invalid. Resetting to the installation dir-
ectory.
Cause: A working directory specified in the .INI file is invalid, has been moved, or has
been deleted.
Action: The installation directory will be substituted. The next time a file is opened, use
the directories list to move to the correct directory.
2448 | Default sample file directory [n] is invalid. Resetting to the installation directory.
Cause: A working directory specified in the .INI file is invalid, has been moved, or has
been deleted.
Action: The installation directory will be substituted. The next time a file is opened, use
the directories list to move to the correct directory.
2448 | Default script test file directory [n] is invalid. Resetting to the installation directory.
Cause: A working directory specified in the .INI file is invalid, has been moved, or has
been deleted.
Action: The installation directory will be substituted. The next time a file is opened, use
the directories list to move to the correct directory.
2448 | The reference file directory [n] cannot be used. Resetting to the installation dir-
ectory.
Cause: A working directory specified in the .INI file is invalid, has been moved, or has
been deleted.
Action: The installation directory will be substituted. The next time a file is opened, use
the directories list to move to the correct directory.
2448 | The z-table file directory [n] cannot be used. Resetting to the installation directory.
Cause: A working directory specified in the .INI file is invalid, has been moved, or has
been deleted.
Action: The installation directory will be substituted. The next time a file is opened, use
the directories list to move to the correct directory.
2450 | Sample Defaults may not be edited while this operation is in progress. Do you
wish to save and close the Sample Defaults edit session?
Cause: An automatic analysis (an analysis in which sample files are created using the
defaults) was processing while editing the defaults.
Action: Finish the edit session of the defaults, close the window, then restart the
automatic analysis.
Cause: The Read-Only attribute is turned on in the application’s MICATTR.DIR file (this
file exists in each folder containing sample or parameter files).
Action: Use Windows Explorer to access the folder containing the MICATTR.DIR file,
then disable the Read-Only option.
Cause: The Read-Only attribute is turned on in the application’s MICATTR.DIR file (this
file exists in each folder containing sample or parameter files).
Action: Use Windows Explorer to access the folder containing the MICATTR.DIR file,
then disable the Read-Only option.
2454 | Too many selections for a print-to-file operation. Only the first (number) selections
will be processed. Please reselect the remainder.
2455 | Too many selections for an export-to-file operation. Only the first (number) selec-
tions will be processed. Please reselect the remainder.
2457 | Results cannot be displayed. More than [n] windows are currently displaying or
printing results.
2458 | An instrument is performing a critical operation. Wait a few moments before exit-
ing the application.
Cause: An attempt was made to exit the application while the instrument was per-
forming a critical operation. This operation must be completed before the applic-
ation can be exited.
Action: Wait a short time and attempt to exit the application again.
2459 | An instrument is busy. A delay in restarting this application could result in loss of
new data. Continue with program Exit?
Cause: An attempt was made to exit the application while an analysis was in progress.
While this is possible, the data collected when the application is inactive will not
be permanently recorded until the application is re-started. A power failure to
the instrument could cause some data to be lost.
Action: If not concerned with the potential for loss of data should a power failure occur,
click Yes to continue; otherwise, click No.
Cause: There was a fatal error in communication between the application and the
software in the instrument. All displays for that instrument will be closed.
Action: Ensure that the analyzer is connected to the computer on the communications
port configured in the Setup program. Stop and restart the analyzer software.
Contact your Micromeritics service representative.
Cause: At least one instrument must be active for the application to operate. The
initialization of all of the instruments configured with the Setup program has
failed. The application stops.
Action A: Usually this message is preceded by another message giving the reason for the
Cause: The software was unable to initialize this instrument. This is usually caused by
one of the conditions listed in the previous error messages.
Action A: Run the Setup program and ensure that a valid port is specified; if not, specify a
valid one when prompted.
Action B: Reinstall the software, then restart application.
Action C: Contact a Micromeritics service representative if this error message continues.
Cause: During initialization of the application, the status of the analyzer was found to be
in a different state than expected.
Action: Click Yes to cancel the operation in process and allow the analyzer to reset,
then continue with initialization. Click Noto cancel the initialization process.
If this error message continues, verify that files in the application directory
structure are not being changed or removed.
Cause: An attempt was made to start an analysis using a file that is open for editing.
Action: Finish editing the file, save and close it, then start the analysis.
2483 | An analysis cannot be performed on [n]. It is open for editing and contains errors.
Cause: An attempt was made to use a sample file containing errors that is currently
open.
Action: Go to the window containing the file, correct the errors, then save it.
2484 | The edit session for [n] must be saved before the analysis. Save changes and con-
tinue with the analysis?
Cause: An attempt was made to start an analysis using a file that contains unsaved
changes and is open for editing.
Action: Click Yes to save the changes, then proceed with the analysis. Click No to
cancel the analysis, then continue editing the Sample Information file.
2485 | The service test file has an invalid status and cannot be used for this analysis.
2486 | Could not construct [n] report type. Program will terminate.
2487 | Could not start report generator. Error code [n]. Program will terminate.
Cause: The specified file in a Save As operation is already open for edit.
Action: Select a different file for the Save As operation.
Cause: The ASCII .INI file containing initialization information and system options
information used during program startup does not exist.
Action: Restore the .INI file from back-up, or re-install the software.
2491 | Highlighted fields contain errors. Please correct the errors before dialog box.
Cause: The fields highlighted in red contain invalid entries. The window cannot be
closed until all errors are corrected.
Action: Check the entries, correct the errors, then close the window.
Cause: The entered value in the highlighted field is outside the valid range of values.
Action: Check the entry, then either enter or select an appropriate value.
2495 | Value is out of the valid range. Enter a value between [n] and [n].
Cause: The entered value in the highlighted field is outside the valid range of values.
Action: Check the entry, then either enter or select an appropriate value.
2498 | The requested change to the Sample’s status is invalid at this time.
Cause: A request to change the file’s status, for example, from automatically collected
to manually entered could not be done.
Action: Contact your Micromeritics service representative. Record the name of the
sample file in which the problem occurred.
Cause: An attempt was made to enter a sequence number that did not contain at least
three digits.
Action: Enter a sequence number that contains at least three digits.
2500 | All sample file names that can be created using the sequence number pattern
already exist. You may want to modify the next sequence number.
Cause: No more sample information files can be created using the currently entered file
name sequence number.
Action: Go to Options > Default Method, then enter another sequence number.
2501 | System resources have reached a dangerously low level. Please close some win-
dows to avoid the loss of data.
Cause: A large number of windows are open and consuming the system resources
available to all applications.
Action: Close one or more windows. Contact your Micromeritics service representative.
2502 | Error writing to file [n] during print. Error code [n].
Cause: An error occurred in the file being written to during a print operation.
Action: Ensure there is sufficient space on the drive containing the file.
2503 | Error converting file [n]. Cannot create DIO intermediate file.
Action: An attempt was made to replace a file that is read only or the computer login
does not have permission to do so. Contact your IT Department.
2505 | Error logger cannot be initialized. Error code [n]. Program will exit.
2508 | Overlay [n] was not found. It will not be included in the reports.
Cause: An error occurred while the program was opening a file necessary to the report
operation.
Action: Use the name given in the error message to investigate. Contact your Micro-
meritics service representative.
2510 | Error parsing reports from file [n]. Reports cannot be produced.
Cause A: One or more data entry fields in the sample file may contain an invalid character
(such as a single quote or double quotes).
Action A: Review the data entry fields (for example, the Sample field), then remove the
invalid character.
Cause B: The system was unable to create the usual temporary files during the report,
possibly due to insufficient disk space.
Action B: Check the space available on the hard disk.
Cause C: An internal processing error occurred.
Action C: Contact your Micromeritics service representative.
2511 | Print job [n] has been canceled due to insufficient disk space. Delete unnecessary
files and restart the report.
Cause: The disk drive does not have required space for the temporary file.
Action: Delete unnecessary files from the disk. At least five megabytes of free space is
required for normal operation.
Cause: An attempt to save calibration data has failed due to possible media problems.
Action A: Ensure the destination location has no problems.
Action B: Choose an alternate media to save the calibration data.
2515 | Warning: Changing the calibration information will affect the performance of the
instrument. Only qualified service personnel should do this. Do you wish to proceed?
2516 | Warning: Keeping a backup copy of the calibration data is recommended by Micro-
meritics. Would you like to do so now?
2517 | Canceling this dialog will reset the calibration state to what it was when this dialog
was first opened. Are you sure you want to cancel?
Cause: The selected sample file does not have collected data and cannot be used for
reporting.
Action: Select a different sample file.
2532 | The instrument contains a different software version. Do you want to reset it?
Cause: The application has discovered a different version of software operating in the
analyzer.
Action: If there are no analyzers other than the one connected to the computer, click
Yes, then allow the updated software to load.
Action B: Copy the contents of the manual folder from the setup media to the application
directory.
Cause: The Micromeritics web page for DFT models cannot be accessed. This could be
caused by an ISP problem of high internet traffic.
Action: Try the operation later.
6000 | An error occurred while loading the application control information. Data entry
cannot be performed. (Code [n]).
Cause: An error occurred accessing the control information disk file required by this
application.
Action: The disk drive may have failed or be corrupt. Run diagnostics on the disk drive.
Cause: A communications problem between the computer and the analyzer has
occurred.
Action A: Check the cable connection between the computer and the analyzer
Action B: Go to Unit [n] > Unit Configuration and verify that the TCP/IP configuration
for the computer and the analyzer are correct.
Action C: Exit the application, and turn off the analyzer. Then turn on the analyzer and
restart the application. Contact a Micromeritics service representative if this
error message continues.
Cause: An incorrect serial number was entered when installing the software.
Action Exit the application. Access the Setup program and remove the incorrect ana-
lyzer. Then use the Add an Analyzer option in the Setup program to add the cor-
rect serial number for the controlling unit. Contact a Micromeritics service
representative if this error message continues.
Cause: Calibration data for the selected unit was not found
Action: Go to Unit > Calibration. Select the appropriate Calibration menu item or load
a saved calibration.
6112 | File [n] cannot be opened. It is currently being setup for analysis.
Cause: An open file was selected that is currently being used for analysis.
Action: Select a different file.
6113 | An analysis cannot be performed on [n]. It is open for editing and contains errors
Cause: The selected analysis file contains errors and is open for editing.
Action: Finish editing the file, save, and close it, or choose a different file. Then restart
the analysis.
6114 | The edit session for [n] must be saved before the analysis. Save changes and
proceed with the analysis?
Cause: An attempt was made to start an analysis using a file that contains unsaved
changes and is still open.
Action: Click Yes to save the changes and proceed with the analysis. Click No to can-
cel the analysis and continue editing the file, save and close it; or click Browse
and choose a different file.
Cause: An attempt to save calibration data has failed due to possible media problems.
Action A: Confirm the media containing the calibration file has no problems.
Action B: Choose an alternate media to save the calibration data.
6123 | The device has no calibration data for the selected insert.
Cause: A calibration report was produced for an insert that has not been calibrated.
Action: Perform a volume calibration for the insert or produce a report for a different
insert.
6126 | The cell and expansion volumes must be positive. Enter positive values and
recalibrate.
Cause: According to the calibration data, the cell volume or expansion volume for the
selected insert is zero or a negative value
Action: Perform these steps:
6127 | The entered calibration volume is < 10% of the cell volume. The resulting
calibration may be inaccurate.
Cause: The application was closed during an analysis. When the application was restar-
ted, the analysis had been canceled.
Action: None required. The data collected up to the point of cancellation is saved in the
sample file.
Cause: An attempt was made to insert an existing file into the Sample File list.
Action: None required. The same file cannot be inserted into the list multiple times.
Cause: An attempt was made to add another file to the Sample File list for sequenced
analyses when the list already contains eight entries.
Action A: Run the sequence analysis with the current list and include the preferred file in
the next sequence analysis.
Action B: Remove one of the files from the list, then add the preferred file.
6163 | The sample file has an invalid status and cannot be used for analysis.
Cause: A sample file with a status other than No Analysis was selected.
Action: Select a sample file that has not been used in an analysis.
Cause: An attempt was made to perform a FoamPyc analysis on a unit that does not
support it. FoamPyc analyses cannot be performed on a unit with a 1 cm3 or 2L
sample cell.
Action A: Start the analysis on a different unit.
Action B: Close the analysis window, open the sample file, and select Standard Method.
Then restart the analysis.
Cause: There was insufficient pressure to allow filling within five minutes during an ana-
lysis
Action: Increase the regulator pressure or, if there is insufficient helium in the tank,
obtain a new tank of helium.
Cause A: A pressure overrange occurred during an analysis because the regulator pres-
sure is set too high.
Action A: Adjust the pressure so that it is equal to the higher of the two fill pressures spe-
cified in the analysis parameters plus 2.0 psig.
Cause B: A pressure overrange occurred during an analysis because an error occurred in
the pressure measurement electronics.
Cause: A power failure occurred and when power resumed, the automatic operation
was canceled.
Action: Restart the automatic operation.
6170 | The measured volume is < 10% of the cell volume. Results may be inaccurate.
Cause: Not enough sample was placed in the sample cup to give an accurate reading.
Action: Rerun the sample using enough material to occupy the maximum amount of the
nominal cell chamber volume
6200 | At least one report item must be selected. Press Cancel if you do not want a report.
Cause: No reports were selected in the Selected Reports list. No output could be pro-
duced.
Action: Ensure that at least one report is selected for the sample and re-submit the
sample for reporting, or click Cancel.
Cause: The requested report file does not contain collected data.
Action: Use the file in an analysis before requesting reports.
6202 | Total solids concentration is invalid; liquid density is greater than or equal to solids
density.
Cause: A report was requested on a file containing invalid densities for Total Solids Con-
centration.
Action A: Specify appropriate densities for liquid and solids.
Action B: Do not request that total solids concentration be reported.
Cause A: A new analysis module has been attached to the instrument where previously a
different module had been attached.
Action A: Press Choice and select Yes to add the new analysis module.
Cause B: An incorrect serial number was read by the analysis module.
Action B: Press Choice and select No to use the serial number and calibration inform-
ation stored in the command module. Contact a Micromeritics service rep-
resentative if this error message continues.
Cause: There was insufficient pressure to allow filling within five minutes during an ana-
lysis
Action: Increase the regulator pressure or, if there is insufficient helium in the tank,
obtain a new tank of helium.
Cause A: A pressure overrange occurred during an analysis because the regulator pres-
sure is set too high.
Action A: Adjust the pressure so that it is equal to the higher of the two fill pressures spe-
Cause: The calibration information could not be read from the command or analysis
module.
Action A: Select Yes and press Enter to reset the pressure and temperature calibration
to nominal values. Resetting to nominal may restore normal readings but a
proper calibration should be performed by a service technician.
Action B: Select No to avoid resetting the calibration. The instrument will be unre-
sponsive.
Cause: The empty cell chamber failed to equilibrate in 1000 seconds during the first
pass of calibration.
Action: Check the system for leaks.
Cause: There was insufficient pressure to allow filling within five minutes during the first
pass of calibration.
Action: Open the tank valves if shut, increase the regulator pressure or, if there is insuf-
ficient helium in the tank, obtain a new tank of helium. Tank pressure should be
at least 200 psi above regulator pressure.
Cause A: A pressure overrange occurred during the first pass of calibration because the
regulator pressure is set too high.
Action A: Adjust the pressure so that it is equal to the higher of the two fill pressures
specified in the analysis parameters plus 2.0 psig.
Cause B: A pressure overrange occurred during the first pass of calibration because an
error occurred in the pressure measurement electronics.
Action B: Contact your Micromeritics service representative.
Cause: A pressure underrange occurred during the first pass of calibration because an
error occurred in the pressure measurement electronics.
Action: Contact your Micromeritics service representative.
Cause: The calibration standard failed to equilibrate in 1000 seconds during the second
pass of calibration.
Action: Check the system for leaks.
Cause: There was insufficient pressure to allow filling within five minutes during the
second pass of calibration.
Action: Open the tank valves if shut, increase the regulator pressure or, if there is insuf-
ficient helium in the tank, obtain a new tank of helium. (Tank pressure should be
at least 200 psi above regulator pressure.)
Cause A: A pressure overrange occurred during the second pass of calibration because
the regulator pressure is set too high.
Action A: Adjust the pressure so that it is equal to the higher of the two fill pressures spe-
cified in the analysis parameters plus 2.0 psig.
Cause B: Adjust the pressure so that it is equal to the higher of the two fill pressures spe-
cified in the analysis parameters plus 2.0 psig.
Action B: Contact your Micromeritics service representative.
Cause: A pressure underrange occurred during the second pass of calibration because
an error occurred in the pressure measurement electronics.
Action: Contact your Micromeritics service representative.
Cause: According to the calibration data, the cell volume or expansion volume for the
selected insert is zero or a negative value
Action: Perform these steps:
Cause A: An automatic operation was canceled before a complete cycle could be com-
pleted.
Action A: Restart the automatic operation.
Cause B: All the cycles have been excluded in review mode.
Action B: Return to review mode and include some cycles.
Cause A: You entered a value for the calibration standard that was less than 10% of the
nominal cell chamber volume. Resulting data may be inaccurate.
Action A: Be sure you enter the value recorded on the Calibration Standard case.
Cause B: You used an inappropriate calibration standard for the nominal cell chamber.
Action B: Use a calibration standard of sufficient size (calibration standards are available
from Micromeritics). The calibration standard should occupy at least 10% of the
nominal cell chamber volume and the more nearly filled the cell is, the better the
calibration.
Cause: There was not enough sample placed in the sample cup to give an accurate
reading.
Action: Rerun the sample using enough material to occupy the maximum amount of the
nominal cell chamber volume.
Cause: The calibration information could not be read from the command or analysis
module.
Action: Close the application and cycle power on the instrument. Contact a Micro-
meritics service representative if this error message continues.
Cause: The fill valve was left open until the maximum system pressure was exceeded.
Action: Close the fill valve and open the vent and expansion valves. Allow the pressure
to stabilize.
No A.C. Found
Cause: You requested a report in which there is either no data available or data have
been excluded via review mode.
Action: Initiate an automatic operation or return to review mode and include at least one
cycle.
Cause: The . (period) key was pressed at startup and the instrument was reset. Data
files and HTML reports have been erased. Setup options have been returned to
default values.
Action: Review setup options.
Cause: A power failure occurred and when power resumed, the automatic operation
was canceled.
Action: Restart the automatic operation if desired.
Cause: The receiving device took longer than five seconds to acknowledge receipt of
data from the pycnometer.
Action: Make sure the receiving device is properly connected to the pycnometer RS-
232 port and is turned on. Verify that the serial I/O parameters controlling the
receiving device correspond with the data transmission parameters in the set up
mode.
Cause: The receiving device stopped transmission by sending an Xoff, and hasn’t
resumed the transmission by sending an Xon.
Action: None; when the receiving device is ready for more data, it should send the
pycnometer an Xon.
Cause: The receiving device took longer than 10 seconds to acknowledge receipt of
data from the pycnometer.
Action: Make sure the receiving device is properly connected to the pycnometer RS-
232 port and is turned on. Verify that the serial I/O parameters controlling the
receiving device correspond with the data transmission parameters in the set up
mode.
Cause: You tried to enter a calibration standard volume that is less than 10% of the nom-
inal full-scale volume.
Action: Enter a volume that represents at least 10% of the nominal cell chamber
volume.
Cause: You tried to review data for an automatic operation when there were no data to
review.
Action: Abandon request.
Cause: You tried to enable run precision without increasing the number of runs to at
least five.
Action: Increase the number of runs to five or abandon request.
Cause: You tried to enter a value that is out of the valid range.
Action: Enter a value in the specified range.
Cause: You tried to enter a calibration standard volume that is less than 10% of the nom-
inal full-scale volume.
Action: Enter a volume that represents at least 10% of the nominal cell chamber
volume.
Cause: You tried to review data for an automatic operation when there were no data to
review.
Action: Abandon request.
Cause: You tried to enable run precision without increasing the number of runs to at
least five.
Action: Increase the number of runs to five or abandon request.
Cause: You tried to enter a value that is out of the valid range.
Action: Enter a value in the specified range.
C MULTIGAS OPTION
The Multigas option enables the AccuPyc to use up to four different gases. The Multigas assembly
consists of:
If a gas is attached to the gas inlet on the rear panel of the analyzer, close the gas cyl-
inder valve and remove the inlet tubing from the analyzer before installing the multigas
assembly. The gas can then be attached to one of the inlet valves on the assembly.
1. Remove the two retaining screws on the rear panel of the AccuPyc.
2. Position the valve assembly against the rear panel so that the retaining screw holes on the
valve assembly align with those on the rear panel of the analyzer.
3. Attach the valve assembly to the rear panel using the two screws included in the multigas kit.
4. Carefully cut and remove the cable ties holding the gas entrance tubing in position and remove
the ties.
5. Remove the protective caps from the ends of the gas entrance tubing, then remove the fer-
rules from the upper end of the tubing.
6. Remove the nut from the center port and slide it onto the end of the gas entrance tubing, then
replace the front and rear ferrules.
7. Insert the tubing into the center port of the multigas assembly. Use the nut to secure the tubing
to the assembly.
8. Insert the other end of the tubing into the gas inlet. Use the self contained swage nut to secure
the tubing to the inlet.
CONNECT GASES
1. Connect one end of the gas supply tubing to the gas supply. See Instrument Components
on page 1 - 5.
2. Insert the other end of the tubing into one of the valve inlets on the multigas assembly.
3. The faceplate of the valve assembly contains four sections; one for each gas inlet on the
assembly. Write the name of the gas in the applicable section.
SELECT GASES
After installing the gases, use the knob to select the gas. Purge the gas lines when changing gases.
Wear latex gloves when handling inserts and sample cups. Oils from skin may con-
taminate the surface and affect analysis results. See Handling System Components
on page J - 1.
Equipment Description
AccuPyc 1 cm3 Includes a 0.1 cm3 insert/sample cup combination.
AccuPyc 10 cm3 Includes 1 and 3.5 cm3 inserts with corresponding sample cups and
appropriate calibration standards. Fritted filter lids are included for both
inserts.
AccuPyc 100 cm3 Includes 10 and 35 cm3 inserts with corresponding sample cups and
appropriate calibration standards. A fritted filter lid, which prevents the
escape of sample particles under rapid gas flow, is included for the 10
cm3 insert.
AccuPyc 2000 cm3 Includes 650 and 1300 cm3 cups, supporting inserts and calibration
standards. A tool is also included to allow removal of the supporting
inserts.
An insert changes the size of the sample chamber and requires its own sample cup. All inserts are
shipped with appropriate cups, with the exception of the 0.1 cm3 insert for the 1 cm3 unit. Because of
its size, the sample cup for the 0.1 cm3 insert is built-in. The appearance of the inserts and cups
varies.
The inserts for the 2000 cm3 systems are installed and removed with a special tool.
1 CM3 INSERT
The MultiVolume kit contains a special tool required to install the 1 cm3 inserts.
1. Remove the sample chamber cap and place on a clean work surface with the greased side
facing upward.
Contamination from airborne particles can occur rapidly. The chamber cap should be
left off the sample chamber for as little time as possible.
4. Place the cup into the insert, then tilt the tool to one side and remove it from the sample cup.
5. If using a fritted filter cap, place it onto the top of the insert.
6. Replace the sample chamber cap.
To Remove the Insert
1. Remove the sample chamber cap and place on a clean work surface with the greased side
facing upward.
2. Remove the fritted filter cap (if used).
3. Insert the handling tool into the sample cup until the sample cup is gripped by the O-ring.
4. Lift the sample cup out of the insert.
5. Pull the insert from the sample chamber.
6. Replace the sample chamber cap.
Do not force the insert or cup into the openings. This may damage the instrument, insert,
or sample cup.
Install the fritted filter cap (if used) before closing the sample chamber. A fritted filter cap is used to
constrict gas flow and is included with some inserts. The fritted filter cap for the 10 cm3 insert fits on
the top of the sample cup. The fritted filter caps for the 1 and 3.5 cm3 inserts fit on the top of the
insert. Fritted filter caps are not available for the 0.1 and 35 cm3 inserts.
CALIBRATE INSERT
An insert must be calibrated and its operation verified. Appropriate calibration standards are
included in the MultiVolume kits. For example, a 100 cm3 capacity AccuPyc with a 10 cm3 insert
must be calibrated with the 10 cm3 insert, cup, and calibration standards.
If the 1 cm3 pycnometer has not been calibrated recently, perform a calibration before
calibrating the 0.1 cm3 insert.
1. Remove the sample chamber cap and place on a clean work surface with the greased side
facing upward.
2. Place the insert into the sample chamber (the sample cup is built into the insert). Replace the
chamber cap.
3. Press Alt + 4 to access the Analyze function. Press ENTER until the Chamber Insert prompt
is displayed. Ensure that None is displayed.
4. Press ENTER to display the Analyze prompt. Press ENTER to begin the calibration.
5. After the calibration is complete, the Reload prompt is displayed.
6. Press CHOICE until the average measured volume is displayed. Record this value.
7. Press Alt + 2 to access Setup.
8. Press CHOICE until Calibration Data is displayed.
9. Press ENTER to display the Chamber Insert prompt.
10. Ensure that None is selected, then press ENTER to display the Cell Volume prompt.
11. Subtract the average volume of the 0.1 cm3 insert/cup (recorded in Step 6) from the cell
volume displayed in the prompt. Record this value.
12. Press ENTER to display the Expansion Volume prompt.
13. Record the value displayed for the expansion volume, then press ENTER to return to the Cal-
ibration Data prompt.
14. Press ENTER. The Chamber Insert prompt is displayed.
15. Press CHOICE until 0.1 cm3 is displayed, then press ENTER to display the Cell Volume
prompt. Enter the value recorded in Step 11.
16. Press ENTER to display the Expansion Volume prompt. Enter the value recorded in Step 12.
17. Press SAVE to save the information and return to the Reload prompt.
The pycnometer is now ready for analyses using the 0.1 cm3 insert/cup.
5. Press SAVE to save the changes and return to the Reload prompt.
A Chamber Insert prompt will display during all normal operating procedures. For example, a
Chamber Insert prompt will display when starting a calibration or analysis.
The AccuPyc uses the DTR and DSR signals for hardware flow control. Ensure that the serial
device provides these signals. For example, if attaching to a computer (also a DTE device), use a
null modem cable which includes the designated signals. If transmission problems occur, ensure that
the signals are set up properly. If the signals are correctly configured, contact the receiving device
manufacturer for assistance.
Sample volume is determined from the known cell and expansion volumes and from measured
pressures.
The cell and expansion chamber are initially at ambient pressure and the valve is closed. The cell is
then charged to an elevated pressure.
P1(Vc − Vs ) = nc RT (1)
where
Expansion volume:
PaVx = nx RT (2)
where
Pa = ambient pressure
nx = moles of gas in the expansion volume
When the valve is opened, the pressure falls to an intermediate value, P2, and mass balance yields:
P2( Vc − Vs + Vx ) = nc RT + nx RT (3)
or
then
Pa − P2 (6)
Vc − Vs = V
P2 − P1
x
Pa − P2 (7)
−Vs = −Vc + V
P2 − Pa − P1 + Pa x
Vx (8)
Vs = Vc −
−1 − P1 − Pa
Pa − P2
or
Vx (9)
Vs = Vc −
P1 − Pa
P2 − Pa − 1
Pig = Pi − Pa (10)
Vx (11)
Vs = Vc − P1g
−1
P2g
For the 1 cm3 and 2000 cm3 pycnometers, equation (1) becomes
P1Vx = nx RT
Pa(Vc − Vs ) = nc RT
A derivation similar to that above yields the working equation for the 1 cm3 and 2000 cm3
pycnometers
V=V−V P1g − 1
s c x P
2g
G TEC Module
G TEC MODULE
This AccuPyc system has been modified to provide temperature control from 15 to 36 °C. The
system uses a thermoelectric control (TEC) system to heat or cool the AccuPyc. The temperature is
controlled using a digital temperature controller and a separate power system from the AccuPyc.
It is suggested that the TEC system be installed first, to allow time for temperature to stabilize at the
desired temperature, while the AccuPyc is being installed. See Add Analysis Module to Control
Module on page 11 - 1.
2. Press P. SP 1 displays followed by the current setpoint (SP 1). Use the Up or Down key to
set the temperature.
3. Press P to accept the temperature. The controller will now start heating or cooling to achieve
the setpoint
The TEC module uses a fan to remove heat. The fan is located on the underside of the
instrument. The fan pulls air into the instrument and the air is exhausted through
ventilation slots on the rear panel. The exhaust slots must remain unobstructed.
Do not allow papers or other debris to be pulled across the fan as this will stop the
necessary air flow. The TEC system will be unable to control temperature, and may
become damaged.
The TEC controller may be adjusted to compensate for the small difference. If the AccuPyc
temperature is a little high, then reduce the setpoint of the TEC controller by the difference. Wait for
stabilization. Readjustment may need to be repeated until the correct AccuPyc temperature is
achieved.
The AccuPyc volume must be calibrated at the analysis temperature. Perform the volume calibration
in the normal way until it is time to insert the reference volume (sphere or spheres). After installing
the reference volume, wait at least 20 minutes to allow the volume to achieve the same temperature
as the AccuPyc. Then continue as normal.
When it is necessary to perform AccuPyc tests at a different temperature, re-calibrate the system
volume at the new temperature.
For example, if the ball is 6.37182 cm3 at 20 °C but run at 50 °C, multiply the Volume × the Factor in
the table.
Every ball is a slightly different size therefore the table provides the factor to be used if the value used
during calibration needs to be adjusted. Multiply the volume provided with the reference sphere by
the value in the right column.
3
Temperature (°C) Typical Volume (cm ) Factor
15 51.089712 0.99995
20 51.092292 1.000000
25 51.094872 1.000051
30 51.097452 1.000101
35 51.100032 1.000152
For example, if the ball is 51.092292 cm3 at 20 °C but run at 50 °C, multiply the Volume × the Factor
in the table.
If using the Windows version of the AccuPyc II, exit the application before powering off
the analyzer.
5. Power ON the analyzer. If using the Windows version of the AccuPyc II, start the application .
6. Calibrate the analysis module using the calibration kit shipped with the system. See About
Calibration on page 10 - 1.
I TRANSMITTED DATA
Keypad Function
Analysis and calibration data can be transmitted in a single column or spreadsheet with data in
ASCII delimited format. Units are displayed as:
Data Units
Type Measurement
Date DD/MM/YY
Time HH:MM:SS
Pressure psig
Temperature °C
Elapsed Time seconds
ANALYSIS REPORT
Analysis Report - Spreadsheet
Record Description Form
Number
1 Version Number 20 characters
2 Serial Number 1 integer
3 Report type = analysis 8 characters
4 Start (reported on one line as ASCII comma delimited 8 characters (each)
data)
l 4a. Date
l 4b. Time
l 5a. Date
l 5b. Time
l 0 = None
l 1 = 10 cm3 (100 cm3 unit)
l 0 = Disabled
l 1 - Enabled
l 0 = Excluded
l 1 = Included
CALIBRATION REPORT
Calibration Report - Single Column
Record Description Form
Number
1 Version Number 20 characters
2 Serial Number 1 integer
3 Report type = calibration 11 characters
4 Start (reported on one line as ASCII comma delimited 8 characters (each)
data)
l 4a. Date
l 4b. Time
l 5a. Date
l 5b. Time
l 0 = None
l 1 = 10 cm3 (100 cm3 unit)
l 0 = Excluded
l 1 = Included
l 4a. Date
l 4b. Time
l 5a. Date
l 5b. Time
l 0 = None
l 1 = 10 cm3 (100 cm3 unit)
l 0 = Excluded
l 1 = Included
J INSTALLATION INSTRUCTIONS
These instructions assume that the shipping cartons have been opened and their contents checked
and verified .
CALIBRATION STANDARD
l Wear latex or nitrile, powder-free gloves to prevent transfer of oil from hands.
l Do not drop the standard into the sample cup. Roll the standard into the cup to prevent damage.
l Always return the standards to the case. Standards are unit specific.
When left uncapped, the sample chamber temperature may become unstable and/or
water vapor will adsorb on the inner surface of the chamber. Either of these conditions
can affect analysis results.
l Avoid laying the chamber cap on a work surface. Debris may collect on the greased surface of the
O-ring which can also affect analysis results.
l When multiple analysis modules are connected, never interchange sample chamber caps. Caps
GAS REQUIREMENTS
The pycnometer uses helium or nitrogen (99.995% pure or better) to provide rapid, accurate
analyses. The cylinder containing helium must be fitted with a gas regulator set for 19-23 psig (131-
159 kPag). The pressure input to the pycnometer should never be greater than 25 psig (172 kPag).
Excessive pressures waste gas due to a protection device contained in the pycnometer that vents
the pycnometer to atmospheric pressure if pressure exceeds 35 psig (241 kPag).
Other inert, dry gases can also be used and may be more applicable for some applications.
PYCNOMETER SET UP
PYCNOMETER CONFIGURATION
Diagram Components
Component Description
A Separate power cord for external power supply. Connect cable to wall outlet.
B External power supply
C Hardwired 8-pin mini-DIN power cable for external power supply.
D 8 pin mini DIN cable.
E Power adapter DIN port.
Set Up Procedure
1. Unpack and inventory all boxed parts. Contact your Micromeritics Service Representative if
parts are missing or broken.
2. Place module(s) on work surface.
3. Connect separate power cable(s) to external power supply(s).
4. Connect external power supply(s) to power outlet on the module(s).
5. Connect power cable to wall outlet.
6. Toggle the module power switch to the ON position.
7. Check the power indicator on the front panel of the analyzer. The indicator will blink green to
indicate that power is applied to the analyzer. Once the application is running, the indicator will
stop blinking and remain illuminated.
8. On the keypad, press Alt + CHOICE + 2. Unit [n] appears in the upper left corner of the dis-
play.
9. Adjust the display brightness control using the Brightness Control button on the rear of the
analyzer.
See Pycnometer Configuration on page J - 2 for setup instructions. When complete, connect
the circulating bath.
The rear panel of the temperature controlled AccuPyc contains connections for a circulating bath.
Provided in the accessory kit are two (2) hose fittings and a four (4) foot length of tubing.
CONNECT KEYBOARD
An optional USB keyboard can be connected to a USB port on the rear panel of the analysis module.
The keyboard is not provided by Micromeritics.
GAS CONNECTION
These instructions are for connecting inert gases only. Follow the safety procedures
used by your laboratory for connecting hazardous gases.
1. Attach an appropriate regulator to the gas cylinder or gas line, then tighten with a wrench.
Leave the gas bottle shutoff valve closed until instructed otherwise.
2. Attach the gas inlet line to the regulator, reducer fitting, or regulator expansion:
It is important to use the inlet tubing supplied with the analyzer. Gas lines made of mater-
ials other than copper or stainless steel can cause operational problems as well as inac-
curate data.
Do not overtighten the fittings. Doing so can collapse the brass fitting and cause a leak.
Purge the regulator and inlet tubing to prevent contamination of the analysis gas.
If the pressure decreases, tighten the nut connecting the regulator to the gas cylinder. If
the pressure is stable, proceed to the next step.
Ensure the gas supply equipment is adequately vented before performing the next step.
e. Turn the pressure regulator control knob clockwise until the low-pressure gauge indicates
19.500 psig (134.445 kPag).
f. Turn the regulator shutoff valve counter-clockwise to open.
g. Turn the gas cylinder shutoff valve counter-clockwise to open. Flow gas for 10 to 30
seconds, then close the valve.
h. Turn the gas regulator shutoff valve clockwise to close
i. Attach the other end of the copper tubing to the gas inlet on the rear panel of the pycno-
meter. Remove the inlet plug that shipped with the pycnometer.
3. Turn the nut on the tubing clockwise until it is finger-tight. Then use a 7/16 in.(1 mm) wrench to
tighten the nut an additional 1/4 turn.
4. Open the gas cylinder and regulator shutoff valves.
5. Turn the regulator control knob until the low pressure gauge indicates 21.5 psig (148 kPag).
For a High Pressure sample chamber configuration, set regulator pressure 20 psi above cycle
and fill pressure.
An Ethernet switch is required if connecting a different type of analyzer to the same computer as the
AccuPyc. If configuring multiple AccuPyc modules. they are connected together with only the
Control Module connected to the Ethernet switch. See Connect an Ethernet Switch on page J -
18.
If using the keypad to control analyses, follow these instructions for configuring the mod-
ule. If using the Windows application, skip this section.
SYSTEM SETUP
Press Alt + 2 to access the set-up function.
Display Description
Setup. Choose or enter parameters for analysis,
calibration, reporting, and data transmission.
The parameters for each set up type are stored and used
to control the pycnometer until new ones are specified.
At the completion of each set up, the display returns to the Setup type? prompt.
l Press SAVE to save the information and return to the Reload prompt.
l Choose another set up type. Do not press SAVE until all parameters have been entered. The dis-
PRINTER SETUP
Printers are connected to the AccuPyc II via a USB port. See Specifications for AccuPyc II on
page 1 - 11 for a list of supported printers. Using the keypad:
COMMUNICATIONS
Use these instructions for keypad installations only. If a computer will be attached to the
module, skip this section. All setup functions will be done in the analysis software applic-
ation.
Communications enables specification of communications criteria, such as baud rate, parity, etc.
Ensure the Ethernet cable is plugged into the Ethernet port on the
analyzer and to the network. Choose DHCP to have an IP address
assigned automatically, or Specify to specify an address. Press
CHOICE until the option is displayed.
Press Alt + CLEAR to cancel the operation and return to the Reload
prompt. Contact your IT department, or choose the Specify option and
enter the appropriate information.
Communications Displays when the Specify option is selected. Enables entry of an
IP Address * appropriate IP address. This is a numerical field in the following format:
(user entered) nnn.nnn.nnn.nnn. Enter the address.
Communications Enables entry of the subnet mask number. Enter the subnet mask
Subnet Mask * number.
(user entered)
Communications The Gateway address is used for communicating outside the local
Gateway * network. The Gateway address is often the same as the IP address,
(user entered) with a “1” after the last dot instead of the last three characters
(nnn.nnn.nnn.1). Contact your IT department with any questions.
Communications When connected to a network, analysis or calibration results can be e-
E-mail Address: mailed automatically upon their completion. This prompt enables entry
of the e-mail address where results are to be sent.
If not using DHCP and the e-mail address is outside the local network,
specify a gateway address (see previous prompt). A keyboard is
required to type in the e-mail address.
required:
meritics-1340
UNIT TYPES
If not using the analyzer software, these keypad prompts enable selection or entry of:
l units of measurement
l operating language
l current date and time
Options are:
The time is entered in a two digit format for the hour, minutes, and
seconds. For example; to set the time to 1:05 pm, enter 13:05:00
(hour:minutes:seconds). Use the . (decimal) key to enter a colon (:).
SETUP REPORTS
The Report Options command enables the report mode (density or volume for analysis) and
specification of the report destination. After analysis or calibration, the display returns to the Reload
prompt. An asterisk (*) in the third line of the display indicates an error message exists.
A report is generated automatically to the destination selected at the Setup > Report Options as
follows:
To view report data by points, use the Review function. See Review Data on page 2 - 11.
Request Cup Mass? is only necessary if different size cups are used for cal-
ibration and analysis.
CALIBRATION
See Calibrate Volume on page 10 - 13.
ANALYTICAL BALANCE
The analytical balance tested by Micromeritics is Scientec® model #ZSA120.
An analytical balance can be connected to a serial port on the computer and used to transfer the
sample’s mass via a third-party software directly to the AccuPyc II application. Refer to the balance
manufacturer’s operator’s manual for transmission instructions.
1. Connect the analytical balance to the RS-232 port on the rear panel of the analyzer.
2. Go to Options > Configure Balance to connect and set up an RS-232 balance to the ana-
lyzer.
BRIGHTNESS CONTROL
SOFTWARE INSTALLATION
The following prerequisites must be completed prior to installing the software:
l Operating System. Windows 7 Professional or higher operating system is recommended for the
best user experience.
l Desktop Installation Required. The application should not be installed on a network drive with
shared access. Multiple users cannot operate the application at the same time.
l 10 Base T Ethernet Port. If the computer is to be connected to a network, two Ethernet ports
are required. If more than one Ethernet based unit is connected to the same computer, an eth-
ernet switch will also be required.
l Read/Write Permissions. All users of the application will need Read / Write permission to all dir-
1. Configure an Ethernet port on the computer to communicate with the analyzer. The IP
address must be 192.168.77.100. Reference the computer operator manual for instructions
on changing the IP address.
2. Insert the installation media into the media drive. The program automatically starts the install-
ation. If the installation does not immediately start, navigate to the media drive, then locate and
double click the setup.exe file.
3. On the Ethernet Introduction window, verify that all prerequisites have been met.
a. If not, click Cancel and complete the steps, then restart the installation program. Do not
proceed with installation until these tasks have been completed.
b. If the prerequisites have been met, click OK.
4. On the New Installation window, verify that sufficient disk space is available to continue the
installation. If not, click Exitand free up the minimum required disk space. If the disk space is
OK, click Next to accept the default installation directory, or click Browse to change the install-
ation location.
a. To add a shortcut to the desktop, select the Add the application icon to the Desktop.....
checkbox.
b. To allow all computer users access to the software application , select Install this applic-
ation for All Users.all users of the computer.
c. Click Next.
4. On the Analyzer Configuration window, select 1 for the number of analyzers to be installed
and enter the analyzer serial number in the text box.
10. On the Ethernet Ports window, select one option from the following table:
If... Then...
the configured port is listed Select the port, then click Next.
the configured port is not listed Click Specify and enter the remaining portion of the
IP Address, then click Next.
11. If prompted to configure a balance, select either 0 or 1. The balance should be connected to
Com Port 1. The default parameters should be OK. If not prompted for a balance, skip this
step.
The Setup program is located on the installation media. After initial software installation, if changes
need to be made to the analyzer setup, reinsert the Setup media and follow the prompts.
If the IP address needs to be changed on the computer connected to the analyzer, refer
to the computer's operating system manual or the internet for instructions. The
IP address for the computer and the IP address specified in the setup program must
match. The IP address must be 192.168.77.100.
Keypad Function
1. Turn the On/Off switch on the rear panel in the Off ( O ) position.
2. Insert the USB stick containing the software update into one of the USB connectors on the
rear panel.
3. Place the Power switch in the On ( | ) position.
4. After approximately 1-2 minutes of initialization, the software starts to load. The following
prompt is displayed:
INDEX B
Brightness Control J - 18
0
0.1 cm3 Insert C
calibrate D - 5
Calculations A - 1, A - 1
Active Area A - 7
1 Analysis A - 1
Asphalt Density A - 3
1, 10, 3.5, 35, 650, and 1300 cm3 Inserts Calibration A - 4
calibrate D - 6 for FoamPyc Methods A - 5
Geometric Volume A - 7
Percent Porosity A - 7
A Resin Volume A - 8
Run Precision A - 8
About this manual iii SPC report variables A - 9
AccuPyc, about 1 - 1 Total Pore Volume A - 10
Analysis Total Solids Concentration A - 11
Cancel 2 - 5 Volume Offset A - 4
Parameter Files 2 - 7 Volume Scale A - 4
Perform 6 - 1, 6 - 3 Calibrate
QuickStart 6 - 7 Insert
review 6 - 9 0.1cm3 D - 5
sequence 6 - 8 1, 10, 3.5, 35, 650, and 1300 cm3
Analysis Conditions Inserts D - 6
create parameter file 5 - 2 temperature offset 10 - 2
defined 5 - 2 using the keypad 10 - 12
FoamPyc Application 5 - 4 using the software 10 - 1
Analysis Module volume 10 - 13
Add to Temperature Controlled Unit H - 2 volume scale 10 - 6
Analytical Balance 11 - 3, J - 17 zero cell volume 10 - 3
Analyzer zero pressure 10 - 4
connect to computer J - 8 Calibration 10 - 1
schematic 3 - 10 Calculations A - 4
show status 3 - 11 copy data 10 - 18
status 3 - 9 data, keypad 10 - 12
Analyzing, file status 3 - 7 load data 10 - 18
Application. start J - 22 load from file 10 - 11
Asphalt Density Application 1 - 3 report 10 - 1
Asphalt Density Calculations A - 3 reset pressure 10 - 5, 10 - 14
Asphalt Density Measurement, TEC review 10 - 15
Modules G - 4 save to file 10 - 11
Axis Cross-hair 7 - 16 Temperature 10 - 17
verify operation 10 - 10
I N
Inserts No Analysis, file status 3 - 7
10, 3.5, and 35 cm3 D - 4
Operating Parameters D - 6
Install O
software J - 19
software upgrade J - 24 Option presentation 3 - 5
Installation instructions J - 1 Options Report
Instrument components 1 - 5 report options 8 - 5
Instrument log, show 3 - 9 Overlays
generate 7 - 18
multiple graph 7 - 18
K multiple sample 7 - 19
Report Example 7 - 25
Keyboard, connect J - 5
Keypad
about 2 - 2 P
calibrate using 10 - 12
Parameter files
about 5 - 1
L analysis 2 - 7
analysis conditions 5 - 2
Leaks directory 5 - 1
check for 12 - 5 FoamPyc application 5 - 4
check using software 12 - 5 Parts and Accessories 13 - 1
check using the keypad 12 - 5 Power Analyzer On and Off 1 - 9
List files 3 - 12 Power failure, recover from 12 - 14
Pressure and Volume Table
report options 8 - 5
M Pressure calibration, reset 10 - 5
Print files 2 - 13, 3 - 12
Maintenance Pycnometer
guidelines for connecting gases 12 - 9 clean 12 - 8
replace gas cylinder 12 - 10 configure, with Keypad J - 9
Manual mode 2 - 6 High Pressure Core 1 - 7
Manual, about this iii Large Volume Core 1 - 7
Material parameters 4 - 5 reset 12 - 15
Menu structure 3 - 1 setup J - 2
Modules, connect 11 - 1 Pycnometer Configuration Diagrams J - 2
Multigas Option 1 - 8, C - 1
MultiVolume Inserts 1 - 7, D - 1
Q
QuickStart Analysis 6 - 7
R S
Regression Report 7 - 3 Sample
Regulator Pressure 12 - 13 Load 6 - 1
Report examples 7 - 21 Prepare 6 - 1
Combined Report 7 - 21 Sample cups, properties 4 - 4
Graph 7 - 24 Sample Defaults 3 - 9
Overlay 7 - 25 Sample file
Sample Log 7 - 27 create 4 - 1
Summary 7 - 22 defined 4 - 1
Tabular 7 - 23 open 3 - 14
Report header shortcuts 7 - 10 Sample log
Report Options 5 - 5 report 8 - 6
Density and Volume Table 8 - 1 Report Example 7 - 27
Density vs Cycle Number 8 - 2 Sample overlays 7 - 19
Density vs Time 8 - 3 Sample Volume Equation Derivation F - 1
Equilibration Report 8 - 4 Schematic
Options Report 8 - 5 instrument 3 - 10
Pressure and Volume Table 8 - 5 shortcuts 3 - 11
Sample Log Report 8 - 6 Selected Reports 8 - 1
Summary Report 8 - 6 Sequence Analysis 6 - 8
Total Pore Volume vs Temp. 8 - 8 shortcuts
Volume vs Cycle 8 - 8 application 3 - 8
Reports report header 7 - 10
about 7 - 1 schematic 3 - 11
axis cross-hair 7 - 16 Show instrument log 3 - 9
calibration 10 - 1 Show instrument schematic 3 - 10
Combined Report 8 - 1 Software
Control Chart 7 - 6 about 3 - 1
features and shortcuts 7 - 9 installation J - 19
graphs 7 - 13 reinstall or modify J - 23
reports 7 - 9 uninstall 3 - 15, J - 25
tabular report 7 - 12 SPC report 7 - 2
generate graph overlays 7 - 18 calculations A - 9
generate sample overlays 7 - 19 Specific Gravity A - 10
graph grid lines 7 - 17 Specifications 1 - 11
open and close 7 - 1 Status
Regression 7 - 3 analyzer 3 - 9, 3 - 11
SPC report options 7 - 2 Summary
start 7 - 1 Report Example 7 - 22
Tabular 8 - 7 Summary Report 8 - 6
toolbar 7 - 11
zoom feature 7 - 17
Review, data 2 - 11
RS-232 Pin Assignment E - 1
Run Precision A - 8
T Z
Tabular reports 8 - 7 Zero Cell Volume, calibrate 10 - 3
features and shortcuts 7 - 12 Zero Pressure, calibrate 10 - 4
Report Example 7 - 23
TEC Module 1 - 2, G - 1
Asphalt Density Measurement G - 4
operate G - 2
set temperature G - 1, J - 5
Temperature Controlled Module 1 - 9, H - 1
Add an Analysis Module H - 2
hardware setup J - 5
Temperature offset, calibrate 10 - 2
Toolbar, Report 7 - 11
Total Pore Volume A - 10
vs Temperature
report options 8 - 8
Total Solids Concentration A - 11
Trademarks 2
Transmit 2 - 7
Transmitted Data, data units I - 1
Troubleshooting and maintenance 12 - 1
U
Unit selections J - 23
V
Volume Offset A - 4
Volume Scale A - 4
calibrate 10 - 6
Volume vs Cycle
report options 8 - 8
Volume, calibrate 10 - 13
W
Warranty i
Web Browser, Review Data with 2 - 11