System, US-Defaults F. Exam Sets, F. FD Systems W. SW VB22

Download as pdf or txt
Download as pdf or txt
You are on page 1of 25

released for USA only

2005-04-25 s

AXIOM Artis
AX

s US-Defaults for

- Browser Configuration
- Exam Sets
- Preprogrammed Positions

valid for
AXIOM Artis d BA, d FA, d MP, d TA,
d BC, d FC, d TC, dBCM, dFCM

valid for US Systems only with SW VB22


onwards

Last change April, 2005


Status:released

AX PLM CW ISQ

April 2005
AXA4-000.820.05.01.02 General notes AXIOM Artis valid for
Replaces: n.a. 1 of 25 VB22 onwards
released for USA only
2005-04-25 s
General remarks
This document describes the default settings for the Exam sets, Acquisition- and Fluoro
Programs, Preprogrammed Positions and Browser configuration with AXIOM Artis
systems.
These defaults are based on experiences of the AX applications specialists.
Dependent on the options available on the system, the following OGP's
(organ programs) are applicable:
CARD* (Digital Cine Mode for Cardiac)
DFR* (DR) (Digital ((Fluoro)) Radiography)
DSA* (Digital Subtraction Angiography)
Card Subtraction *
PERIVISION *
DYNAVSION * & 3D DYNA *
DR Scanning * & DR Stepping *
Fluoro
ROADMAP**
*option, dependent on system configuration
** included with DSA option

Important Notes

Not all options are availabe with all systems!

An Exam Set consists of up to 4 Acquisition Programs and 1 Fluoro Triplet.


A Fluoro-Triplet always exists of one program for 'reduced radiation' (Fluoro -), one for
'normal radiation' (Fluoro), and one for 'high radiation' (Fluoro +) program.

For Systems without CAREVISION (framerates <7.5), the here documented


frame rates must be set to 15 f/s.

Measurement field selection is not possible in Cardiac Mode.


Microfocus is only available with a three focal spot tube.

April 2005
AXA4-000.820.05.01.02 General notes AXIOM Artis valid for
Replaces: n.a. 2 of 25 VB22 onwards
released for USA only
2005-04-25 s

According to the users preference, Window values as well as edge enhancement


might need to be adjusted individually.

Artis FD- Systems are only available with 1k matrix. The parameter "Correction" will
shift the gray values before applying the LUT.

PERI and DYNA Programs cannot be placed in first position of an exam set.
In the first position of an exam set must be always a standard acquisition program, such
as DSA, DR or CARD being programmed.

In pediatric work for infants up to 10 kg, the grid should be removed.

The pediatric fluoro triplet is a recommendation, not a default. You have to override
one of the default fluoro programs, if the pediatric triplet is needed.

All settings are recommendations!

Comments, advice and hints are welcome: [email protected]

April 2005
AXA4-000.820.05.01.02 General notes AXIOM Artis valid for
Replaces: n.a. 3 of 25 VB22 onwards
Seite 4
s
General

April 2005
AXA4-000.820.05.01.02 Browser Configuration AXIOM Artis valid for
Replaces: n.a. 4 of 25 VB22 onwards
Seite 5
s
Patient Browser - Tree View
Database – Patient Level

Patient – Study Level

April 2005
AXA4-000.820.05.01.02 Browser Configuration AXIOM Artis valid for
Replaces: n.a. 5 of 25 VB22 onwards
Seite 6
s
Study – Series Level – List Configuration

Study – Series Level – Stamp Configuration

April 2005
AXA4-000.820.05.01.02 Browser Configuration AXIOM Artis valid for
Replaces: n.a. 6 of 25 VB22 onwards
Seite 7
s
Series – Instance Level – List Configuration

Series – Instance Level – Stamp Configuration

April 2005
AXA4-000.820.05.01.02 Browser Configuration AXIOM Artis valid for
Replaces: n.a. 7 of 25 VB22 onwards
Seite 8
s
Patient Browser - Single View

April 2005
AXA4-000.820.05.01.02 Browser Configuration AXIOM Artis valid for
Replaces: n.a. 8 of 25 VB22 onwards
AXIOM Artis for USA only
Defaults Exam Sets s

(1) 1. DSA 1 (2) 2. DSA 2

1. DSA 7.5
1. Single
1. DSA 1

1. DSA 2

1. DSA 4

Venous
1. DSA
1. DSA

1. DR
ACQUIRE

Long
EXPOSURE DSA DSA DSA DR DSA DSA DSA DR
kV 70 70 70 70 70 70 70 70
Pulse Width 160 160 100 250 50 160 100 100
kV Filter 81 81 81 off 81 81 81 off
kV ms 96 96 96 off 96 96 96 off
dose 3600 3600 3600 2400 3600 3600 3600 2400
kV dose 102 102 102 102 102 102 96 102
Correction 0 0 0 0 0 0 0 0
Mfield C C C C C C C C
Focus µ µ µ µ µ µ µ µ
kv Focus µ->S 73 73 73 70 73 73 70 70
kv Focus S->L 81 81 81 81 81 81 81 81
IMAGE
Processing Mode stop(2.1) stop(2.1) stop(2.1) auto stop(2.1) stop(2.1) stop(2.1) auto
Gamma log20 log20 log20 SP2 Blk log20 log20 log20 SP2 Blk
Matrix 1K 1K 1K 1K 1K 1K 1K 1K
Contrast Iodine Iodine Iodine none Iodine Iodine Iodine none
Bones White no no no no no no no no
Auto Window man man man man man man man man
K-Factor off off off off off off off off
EVE off off off off off off off off
Nat Filter 30 30 30 35 30 30 30 35
Center 1600 1600 1600 1400 1600 1600 1600 1400
Width 1800 1800 1800 1800 1800 1800 1800 1800
DDO 0 0 0 35 0 0 0 35
Sub Filter 10 10 10 --- 10 10 10 ---
Brightness 2800 2800 2800 --- 2800 2800 2800 ---
Contrast 50 50 50 --- 50 50 50 ---
ACQUISITION
Scene Time 20 20 20 20 20 80 40 20
Frame Rate 1 fps 2 fps 4 fps single. 7.5 fps 2 fps 4 fps 4 fps
Frame Control Fixed Fixed Fixed Fixed Fixed Fixed Fixed Fixed
No Phases 1 1 1 1 1 1 1 1
Dyna Time
Angul Step
Dyna Control
Washout
Phase Time
Stepping mode
Direction
Number of phases
[01] Rate
[02] Rate
[03] Rate
[04] Rate
[05] Rate
[06] Rate
[07] Rate
[08] Rate
[09] Rate
[10] Rate
[11] Rate
[12] Rate

April 2005
AXA4-000.820.05.01.02 Exam Sets AXIOM Artis valid for
Replaces: n.a. 9 of 25 VB22 onwards
AXIOM Artis for USA only
Defaults Exam Sets s

(3) 3. Neuro (4) 4. DYNA

1. DYNA 6
1. Carotid

Overview
Vertebral

1. Single

1. Single
1. DYNA
1. Arch
ACQUIRE

4.5
1.

1.
EXPOSURE DSA DSA DSA DR DSA DYNA DYNA DR
kV 70 73 73 70 70 70 70 70
Pulse Width 100 100 100 250 160 10 10 250
kV Filter 81 81 81 off 81 off off off
kV ms 96 96 90 off 96 125 125 off
dose 3600 3600 3600 2400 3600 1200 1200 2400
kV dose 102 102 96 102 102 off off 102
Correction 0 0 0 0 0 0 0 0
Mfield C C C C C C C C
Focus µ µ µ µ µ s s µ
kv Focus µ->S 73 73 73 70 73 off off 70
kv Focus S->L 81 81 81 81 81 off off 81
IMAGE
Processing Mode stop(2.1) stop(2.1) stop(2.1) auto stop(2.1) auto auto auto
Gamma log20 log20 log20 SP2 Blk log20 log20 log20 SP2 Blk
Matrix 1K 1K 1K 1K 1K 1K 1K 1K
Contrast Iodine Iodine Iodine none Iodine Iodine Iodine none
Bones White no no no no no no no no
Auto Window man man man man man man man man
K-Factor off off off off off off off off
EVE off off off off off off off off
Nat Filter 30 30 30 35 30 30 30 35
Center 1600 1600 1600 1400 1600 1600 1600 1400
Width 1800 1800 1800 1800 1800 1800 1800 1800
DDO 0 0 0 35 0 0 0 35
Sub Filter 10 10 10 --- 10 0 0 ---
Brightness 2800 2800 2800 --- 2800 2800 2800 ---
Contrast 50 50 50 --- 50 50 50 ---
ACQUISITION
Scene Time 20 20 40 20 20 20
Frame Rate 4 fps 4 fps 4 fps single. 2 fps single.
Frame Control Fixed Fixed Fixed Fixed Fixed Fixed
No Phases 1 1 1 1 1 1
Dyna Time 4s 4s
Angul Step 6 4,5
Dyna Control DYNA aut DYNA aut
Washout 0.5 F/s 0.5 F/s
Phase Time 1s 1s
Stepping mode
Direction
Number of phases
[01] Rate
[02] Rate
[03] Rate
[04] Rate
[05] Rate
[06] Rate
[07] Rate
[08] Rate
[09] Rate
[10] Rate
[11] Rate
[12] Rate

April 2005
AXA4-000.820.05.01.02 Exam Sets AXIOM Artis valid for
Replaces: n.a. 10 of 25 VB22 onwards
AXIOM Artis for USA only
Defaults Exam Sets s

(5) 5. Perivision (6) 6. CARD

Overview

Coro ND

Coro LD
1. PERI

1. PERI
normal
ACQUIRE

fast

LV
1.
EXPOSURE DSA PERI PERI CARD CARD CARD
kV 70 73 73 70 70 70
Pulse Width 160 100 100 6,4 6,4 6,4
kV Filter 81 off off 81 81 81
kV ms 96 off off 96 96 96
dose 3600 3600 3600 240 240 170
kV dose 102 off off 102 102 102
Correction 0 0 0 0 0 0
Mfield C C C C C C
Focus µ s s L L L
kv Focus µ->S 73 off off off off off
kv Focus S->L 81 off off off off off
IMAGE
Processing Mode stop(2.1) auto auto auto stop(0.8) auto
Gamma log20 log20 log20 G6/2 G6/2 G6/2
Matrix 1K 1K 1K 1K 1K 1K
Contrast Iodine Iodine Iodine none none none
Bones White no no no no no no
Auto Window man man man man man man
K-Factor off off off KA1 KA1 KA1
EVE off off off EA1 EA1 EA1
Nat Filter 30 30 30 30 30 30
Center 1600 1600 1600 2100 2000 2100
Width 1800 1800 1800 2300 2700 2300
DDO 0 0 0 50 30 50
Sub Filter 10 10 10 5 5 5
Brightness 2800 2800 2800 2800 2800 2800
Contrast 50 50 50 30 30 30
ACQUISITION
Scene Time 20 10 10 10
Frame Rate 2 fps 15 fps 30 fps 15 fps
Frame Control Fixed Fixed Fixed Fixed
No Phases 1 12 12 1 1 1
Dyna Time
Angul Step
Dyna Control
Washout
Phase Time
Stepping mode 20 20
Direction PERI Man. PERI Man.
Number of phases Feet to Head Feet to Head
[01] Rate 2 fps / 1 2 fps / 1
[02] Rate 1 fps / 1 2 fps / 1
[03] Rate 1 fps / 1 2 fps / 1
[04] Rate 1 fps / 1 2 fps / 1
[05] Rate 0.5 fps / 1 1 fps / 1
[06] Rate 0.5 fps / 1 1 fps / 1
[07] Rate 0.5 fps / 1 0.5 fps / 1
[08] Rate 0.5 fps / 1 0.5 fps / 1
[09] Rate 0.5 fps / 1 0.5 fps / 1
[10] Rate 0.5 fps / 1 0.5 fps / 1
[11] Rate 0.5 fps / 1 0.5 fps / 1
[12] Rate 0.5 fps / 1 0.5 fps / 1

April 2005
AXA4-000.820.05.01.02 Exam Sets AXIOM Artis valid for
Replaces: n.a. 11 of 25 VB22 onwards
AXIOM Artis for USA only
Defaults Exam Sets s

(8) 7. EP (9) 8. Card Sub

Sub 1024
Coro ND

Coro LD
ACQUIRE

LV
EXPOSURE CARD CARD CARD CARD
kV 70 70 70 70
Pulse Width 6,4 6,4 6,4 6,4
kV Filter 81 81 81 81
kV ms 96 96 96 96
dose 240 240 170 240
kV dose 102 102 102 102
Correction 0 0 0 0
Mfield C C C C
Focus L L L L
kv Focus µ->S off off off off
kv Focus S->L off off off off
IMAGE
Processing Mode auto stop(0.8) auto stop(0.8)
Gamma G6/2 G6/2 G6/2 G6/2
Matrix 1K 1K 1K 1K
Contrast none none none none
Bones White no no no no
Auto Window man man man man
K-Factor KA1 KA1 KA1 off
EVE EA1 EA1 EA1 EA1
Nat Filter 30 30 30 20
Center 2100 2000 2100 1600
Width 2300 2700 2300 1800
DDO 50 30 50 0
Sub Filter 5 5 5 5
Brightness 2800 2800 2800 2800
Contrast 30 30 30 30
ACQUISITION
Scene Time 10 10 10 10
Frame Rate 15 fps 30 fps 15 fps 30 fps
Frame Control Fixed Fixed Fixed Fixed
No Phases 1 1 1 1
Dyna Time
Angul Step
Dyna Control
Washout
Phase Time
Stepping mode
Direction
Number of phases
[01] Rate
[02] Rate
[03] Rate
[04] Rate
[05] Rate
[06] Rate
[07] Rate
[08] Rate
[09] Rate
[10] Rate
[11] Rate
[12] Rate

April 2005
AXA4-000.820.05.01.02 Exam Sets AXIOM Artis valid for
Replaces: n.a. 12 of 25 VB22 onwards
AXIOM Artis for USA only
Defaults Exam Sets s

(10) 9. CARD Ped

ACQUIRE

Ped

LV
EXPOSURE CARD CARD
kV 70 70
Pulse Width 4 5
kV Filter 81 81
kV ms 96 96
dose 200 200
kV dose 102 102
Correction 0 0
Mfield C C
Focus s s
kv Focus µ->S off off
kv Focus S->L off off
IMAGE
Processing Mode auto stop(0.8)
Gamma G6/2 G6/2
Matrix 1K 1K
Contrast none none
Bones White no no
Auto Window man man
K-Factor KA1 KA1
EVE EA1 EA1
Nat Filter 30 30
Center 2100 2000
Width 2100 2700
DDO 50 30
Sub Filter 5 5
Brightness 2800 2800
Contrast 30 30
ACQUISITION
Scene Time 10 10
Frame Rate 30 fps 30 fps
Frame Control Fixed Fixed
No Phases 1 1
Dyna Time
Angul Step
Dyna Control
Washout
Phase Time
Stepping mode
Direction
Number of phases
[01] Rate
[02] Rate
[03] Rate
[04] Rate
[05] Rate
[06] Rate
[07] Rate
[08] Rate
[09] Rate
[10] Rate
[11] Rate
[12] Rate

April 2005
AXA4-000.820.05.01.02 Exam Sets AXIOM Artis valid for
Replaces: n.a. 13 of 25 VB22 onwards
AXIOM Artis for USA only
Defaults Exam Sets s

(1) 1. DSA 1 (2) 2. DSA 2

01. Fluoro

02. Fluoro

03. Fluoro

01. Fluoro

02. Fluoro

03. Fluoro
FLUORO

normal

normal
high

high
low

low
EXPOSURE FLUORO FLUORO FLUORO FLUORO FLUORO FLUORO

10R AF

10R AF

10R AF

10R AF

10R AF

10R AF
80kV

80kV

80kV

80kV

80kV

80kV
2kW

2kW

2kW

2kW

2kW

2kW
X-ray Curve
dose 36 36 45 36 36 45
IMAGE
Gamma G5/4 G5/4 G5/4 G5/4 G5/4 G5/4
Contrast none none none none none none
K-Factor KA3 KA3 KA3 KA3 KA3 KA3
EVE off off off off off off
Auto Window man man man man man man
DDO 25 25 25 25 25 25
Nat Filter 10 10 10 10 10 10
Center 2000 2000 2000 2000 2000 2000
Width 2300 2300 2300 2300 2300 2300
Matrix 1K 1K 1K 1K 1K 1K
FLUORO TYPE
Frame Control Pulsed Pulsed Pulsed Pulsed Pulsed Pulsed
Frame Rate 7.5 fps 15 fps 30 fps 7.5 fps 15 fps 30 fps
BiPl Pulse Red no no no no no no

April 2005
AXA4-000.820.05.01.02 Exam Sets AXIOM Artis valid for
Replaces: n.a. 14 of 25 VB22 onwards
AXIOM Artis for USA only
Defaults Exam Sets s

(3) 3. Neuro (4) 4. DYNA

01. Fluoro

02. Fluoro

03. Fluoro

01. Fluoro

02. Fluoro

03. Fluoro
FLUORO

normal

normal
high

high
low

low
EXPOSURE FLUORO FLUORO FLUORO FLUORO FLUORO FLUORO

10R AF

10R AF

10R AF

10R AF

10R AF

10R AF
80kV

80kV

80kV

80kV

80kV

80kV
2kW

2kW

2kW

2kW

2kW

2kW
X-ray Curve
dose 36 36 45 36 36 45
IMAGE
Gamma G5/4 G5/4 G5/4 G5/4 G5/4 G5/4
Contrast none none none none none none
K-Factor KA3 KA3 KA3 KA3 KA3 KA3
EVE off off off off off off
Auto Window man man man man man man
DDO 25 25 25 25 25 25
Nat Filter 10 10 10 10 10 10
Center 2000 2000 2000 2000 2000 2000
Width 2300 2300 2300 2300 2300 2300
Matrix 1K 1K 1K 1K 1K 1K
FLUORO TYPE
Frame Control Pulsed Pulsed Pulsed Pulsed Pulsed Pulsed
Frame Rate 7.5 fps 15 fps 30 fps 7.5 fps 15 fps 30 fps
BiPl Pulse Red no no no no no no

April 2005
AXA4-000.820.05.01.02 Exam Sets AXIOM Artis valid for
Replaces: n.a. 15 of 25 VB22 onwards
AXIOM Artis for USA only
Defaults Exam Sets s

(5) 5. Perivision (6) 6. CARD

01. Fluoro

03. Fluoro
FLUORO

Fluoro +
Fluoro -

Fluoro

Card
Card

Card
high
low
EXPOSURE FLUORO FLUORO FLUORO FLUORO FLUORO

10R AF

10R AF

10R AF

10R AF

20R AF
80kV

80kV

70kV

70kV

70kV
2kW

2kW

3kW

3kW

3kW
X-ray Curve
dose 36 45 23 45 65
IMAGE
Gamma G5/4 G5/4 G6/5 G6/5 G6/5
Contrast none none none none none
K-Factor KA3 KA3 1,8 KA2 KA2
EVE off off EA1 EA1 EA1
Auto Window man man man man man
DDO 25 25 40 50 50
Nat Filter 10 10 10 10 10
Center 2000 2000 2000 2000 2000
Width 2300 2300 2700 2500 2500
Matrix 1K 1K 1K 1K 1K
FLUORO TYPE
Frame Control Pulsed Pulsed Pulsed Pulsed Pulsed
Frame Rate 7.5 fps 30 fps 15 fps 15 fps 15 fps
BiPl Pulse Red no no no no no

April 2005
AXA4-000.820.05.01.02 Exam Sets AXIOM Artis valid for
Replaces: n.a. 16 of 25 VB22 onwards
AXIOM Artis for USA only
Defaults Exam Sets s

(8) 7. EP (9) 8. Card Sub

Fluoro EP
FLUORO

Fluoro +

Fluoro +
Fluoro -

Fluoro -

Fluoro

Card
Card

Card
EP

EP
EXPOSURE FLUORO FLUORO FLUORO FLUORO FLUORO FLUORO

10R AF

10R AF

20R AF

10R AF

10R AF

20R AF
100kV

70kV

70kV

70kV

70kV

70kV
2kW

3kW

3kW

3kW

3kW

3kW
X-ray Curve
dose 15 32 45 23 45 65
IMAGE
Gamma G6/5 G6/5 G6/5 G6/5 G6/5 G6/5
Contrast none none none none none none
K-Factor 1,8 KA2 1,4 1,8 KA2 KA2
EVE EA1 EA1 EA1 EA1 EA1 EA1
Auto Window man man man man man man
DDO 40 50 50 40 50 50
Nat Filter 0 0 0 10 10 10
Center 2100 2100 2100 2000 2000 2000
Width 2700 2700 2700 2700 2500 2500
Matrix 1K 1K 1K 1K 1K 1K
FLUORO TYPE
Frame Control Pulsed Pulsed Pulsed Pulsed Pulsed Pulsed
Frame Rate 7.5 fps 10 fps 10 fps 15 fps 15 fps 15 fps
BiPl Pulse Red no no no no no no

April 2005
AXA4-000.820.05.01.02 Exam Sets AXIOM Artis valid for
Replaces: n.a. 17 of 25 VB22 onwards
AXIOM Artis for USA only
Defaults Exam Sets s

(10) 9. CARD Ped

Fluoro Ped
FLUORO

Fluoro +
Fluoro -
Ped

Ped
EXPOSURE FLUORO FLUORO FLUORO

10R AF

10R AF

10R AF
70kV

70kV

70kV
3kW

3kW

3kW
X-ray Curve
dose 29 29 36
IMAGE
Gamma G6/5 G6/5 G6/5
Contrast none none none
K-Factor KA2 KA2 KA2
EVE EA2 EA2 EA2
Auto Window man man man
DDO 50 50 50
Nat Filter 0 0 0
Center 2200 2200 2200
Width 2500 2500 2500
Matrix 1K 1K 1K
FLUORO TYPE
Frame Control Pulsed Pulsed Pulsed
Frame Rate 7.5 fps 15 fps 15 fps
BiPl Pulse Red no no no

April 2005
AXA4-000.820.05.01.02 Exam Sets AXIOM Artis valid for
Replaces: n.a. 18 of 25 VB22 onwards
AXIOM Artis for USA only
Defaults Exam Sets s

(1) 1. DSA 1 (2) 2. DSA 2


ROADMAP

roadmap

roadmap

roadmap

roadmap

roadmap

roadmap
normal

normal
high

high
low

low
01.

02.

03.

01.

02.

03.
EXPOSURE RDMP RDMP RDMP RDMP RDMP RDMP

10R AF

10R AF

10R AF

10R AF

10R AF

10R AF
80kV

80kV

80kV

80kV

80kV

80kV
2kW

2kW

2kW

2kW

2kW

2kW
X-ray Curve
dose 36 36 36 36 36 36
IMAGE
Gamma G4 G4 G4 G4 G4 G4
Contrast none none none none none none
K-Factor KA4 KA4 KA4 KA4 KA4 KA4
EVE off off off off off off
Auto Window man man man man man man
DDO 0 0 0 0 0 0
Nat Filter 20 20 20 20 20 20
Center 900 900 900 900 900 900
Width 2000 2000 2000 2000 2000 2000
Sub Filter 0 0 0 0 0 0
Brightness 2045 2045 2045 2045 2045 2045
Contrast 25 25 25 25 25 25
Matrix 1K 1K 1K 1K 1K 1K
ROADMAP TYPE
Frame Control Pulsed Pulsed Pulsed Pulsed Pulsed Pulsed
Frame Rate 7.5 fps 15 fps 15 fps 7.5 fps 15 fps 15 fps
BiPl Pulse Red no no no no no no

April 2005
AXA4-000.820.05.01.02 Exam Sets AXIOM Artis valid for
Replaces: n.a. 19 of 25 VB22 onwards
AXIOM Artis for USA only
Defaults Exam Sets s

(3) 3. Neuro (4) 4. DYNA


ROADMAP

roadmap

roadmap

roadmap

roadmap

roadmap

roadmap
normal

normal
high

high
low

low
01.

02.

03.

01.

02.

03.
EXPOSURE RDMP RDMP RDMP RDMP RDMP RDMP

10R AF

10R AF

10R AF

10R AF

10R AF

10R AF
80kV

80kV

80kV

80kV

80kV

80kV
2kW

2kW

2kW

2kW

2kW

2kW
X-ray Curve
dose 36 36 36 36 36 36
IMAGE
Gamma G4 G4 G4 G4 G4 G4
Contrast none none none none none none
K-Factor KA4 KA4 KA4 KA4 KA4 KA4
EVE off off off off off off
Auto Window man man man man man man
DDO 0 0 0 0 0 0
Nat Filter 20 20 20 20 20 20
Center 900 900 900 900 900 900
Width 2000 2000 2000 2000 2000 2000
Sub Filter 0 0 0 0 0 0
Brightness 2045 2045 2045 2045 2045 2045
Contrast 25 25 25 25 25 25
Matrix 1K 1K 1K 1K 1K 1K
ROADMAP TYPE
Frame Control Pulsed Pulsed Pulsed Pulsed Pulsed Pulsed
Frame Rate 7.5 fps 15 fps 15 fps 7.5 fps 15 fps 15 fps
BiPl Pulse Red no no no no no no

April 2005
AXA4-000.820.05.01.02 Exam Sets AXIOM Artis valid for
Replaces: n.a. 20 of 25 VB22 onwards
AXIOM Artis for USA only
Defaults Exam Sets s

(5) 5. Perivision (6) 6. CARD


ROADMAP

roadmap

roadmap

roadmap

roadmap

roadmap
normal
high

high
low

low
01.

03.

01.

02.

03.
EXPOSURE RDMP RDMP RDMP RDMP RDMP

10R AF

10R AF

10R AF

10R AF

10R AF
80kV

80kV

80kV

80kV

80kV
2kW

2kW

2kW

2kW

2kW
X-ray Curve
dose 36 36 36 36 36
IMAGE
Gamma G4 G4 G4 G4 G4
Contrast none none none none none
K-Factor KA4 KA4 KA4 KA4 KA4
EVE off off off off off
Auto Window man man man man man
DDO 0 0 0 0 0
Nat Filter 20 20 20 20 20
Center 900 900 900 900 900
Width 2000 2000 2000 2000 2000
Sub Filter 0 0 0 0 0
Brightness 2045 2045 2045 2045 2045
Contrast 25 25 25 25 25
Matrix 1K 1K 1K 1K 1K
ROADMAP TYPE
Frame Control Pulsed Pulsed Pulsed Pulsed Pulsed
Frame Rate 7.5 fps 15 fps 7.5 fps 15 fps 15 fps
BiPl Pulse Red no no no no no

April 2005
AXA4-000.820.05.01.02 Exam Sets AXIOM Artis valid for
Replaces: n.a. 21 of 25 VB22 onwards
AXIOM Artis for USA only
Defaults Exam Sets s

(8) 7. EP (9) 8. Card Sub


ROADMAP

roadmap

roadmap

roadmap

roadmap

roadmap

roadmap
normal

normal
high

high
low

low
01.

02.

03.

01.

02.

03.
EXPOSURE RDMP RDMP RDMP RDMP RDMP RDMP

10R AF

10R AF

10R AF

10R AF

10R AF

10R AF
80kV

80kV

80kV

80kV

80kV

80kV
2kW

2kW

2kW

2kW

2kW

2kW
X-ray Curve
dose 36 36 36 36 36 36
IMAGE
Gamma G4 G4 G4 G4 G4 G4
Contrast none none none none none none
K-Factor KA4 KA4 KA4 KA4 KA4 KA4
EVE off off off off off off
Auto Window man man man man man man
DDO 0 0 0 0 0 0
Nat Filter 20 20 20 20 20 20
Center 900 900 900 900 900 900
Width 2000 2000 2000 2000 2000 2000
Sub Filter 0 0 0 0 0 0
Brightness 2045 2045 2045 2045 2045 2045
Contrast 25 25 25 25 25 25
Matrix 1K 1K 1K 1K 1K 1K
ROADMAP TYPE
Frame Control Pulsed Pulsed Pulsed Pulsed Pulsed Pulsed
Frame Rate 7.5 fps 15 fps 15 fps 7.5 fps 15 fps 15 fps
BiPl Pulse Red no no no no no no

April 2005
AXA4-000.820.05.01.02 Exam Sets AXIOM Artis valid for
Replaces: n.a. 22 of 25 VB22 onwards
AXIOM Artis for USA only
Defaults Exam Sets s

(10) 9. CARD Ped


ROADMAP

roadmap

roadmap

roadmap
normal

high
low
01.

02.

03.
EXPOSURE RDMP RDMP RDMP

10R AF

10R AF

10R AF
80kV

80kV

80kV
2kW

2kW

2kW
X-ray Curve
dose 36 36 36
IMAGE
Gamma G4 G4 G4
Contrast none none none
K-Factor KA4 KA4 KA4
EVE off off off
Auto Window man man man
DDO 0 0 0
Nat Filter 20 20 20
Center 900 900 900
Width 2000 2000 2000
Sub Filter 0 0 0
Brightness 2045 2045 2045
Contrast 25 25 25
Matrix 1K 1K 1K
ROADMAP TYPE
Frame Control Pulsed Pulsed Pulsed
Frame Rate 7.5 fps 15 fps 15 fps
BiPl Pulse Red no no no

April 2005
AXA4-000.820.05.01.02 Exam Sets AXIOM Artis valid for
Replaces: n.a. 23 of 25 VB22 onwards
AXIOM Artis
Defaults Preprogrammed Positions s
Cardiac monoplane
RAO / LAO cran / caud I.I. Iso height Tilt Collimator Filter
No [°] [°] [cm] [cm] [°]
1 RAO 30 --- 17 15 0 open *1
2 LAO 60 --- 17 15 0 open *2
3 LAO 90 --- 17 15 0 open *2
4 LAO 40 caud 20 17 15 0 open *2
5 LAO 60 cran 25 17 15 0 open *2
6 RAO 10 cran 25 17 15 0 open *1
7 RAO 30 --- 23 15 0 open *1
8 LAO 60 --- 23 15 0 open *2

Cardiac biplane
RAO / LAO cran / caud I.I. Iso height Tilt Collimator Filter
No [°] [°] [cm] [cm] [°]
RAO 30 --- 17 15 0 open *1
1
LAO 60 --- 17 15 0 open *2
LAO 10 --- 17 15 0 open *2
2
LAO 90 --- 17 15 0 open *2
RAO 35 cran 25 17 15 0 open *1
3
LAO 50 caud 30 17 15 0 open *2
RAO 45 caud 30 17 15 0 open *1
4
LAO 60 cran 25 17 15 0 open *2
5 RAO 30 --- 23 15 0 open *1
LAO 60 --- 23 15 0 open *2

*1

*2

April 2005
AXA4-000.820.05.01.02 Preprogrammed Positions AXIOM Artis valid for
Replaces: n.a. 24 of 25 VB22 onwards
AXIOM Artis
Defaults Miscellaneous s
Labelling of CD Drives

1 Reader CD-Lesen CD Read


2 Burner CD-Schreiben CD Write

Configuration for Export and Send

In the Configuration Panel open "Transfer"


Select the "Export" tab card and set the compression for CD-Writer to "JPEG lossless"
Select the "Send" tab card and set the compression to "JPEG lossless"

In Service Mode under System Config check whether "ARCH net" is set to "enabled"

Configuration under "Image System / Acq. Control":

Here in Service Mode the following settings can be configured:


Auto Edge Enhancement for Dose Reduction

April 2005
AXA4-000.820.05.01.02 Miscellaneous AXIOM Artis valid for
Replaces: n.a. 25 of 25 VB22 onwards

You might also like