Data Sheet Dusthunter t100
Data Sheet Dusthunter t100
Data Sheet Dusthunter t100
DUSTHUNTER T100
A Ordering information
Type Part no.
The exact device specifications and performance data of the product may deviate from the information provided here, and
depend on the application in which the product is being used and the relevant customer specifications.
C Our regional sales organization will help you to select the optimum device configuration.
D
Product description
E The DUSTHUNTER T100 is a measuring device for dust at medium to high concentrations. Basic
measuring value is transmittance. Opacity and extinction can be calculated and displayed as well
as the dust concentration after gravimetric calibration. The device has an integrated contamina-
At a glance
• For medium to high dust concentrations
H
• On-board contamination check
• Automated thorough check of zero and reference point
• For small to large measuring distances
I
Your benefits
J • Easy installation, commissioning and operation
• Measurement independent of gas velocity, humidity and particle charge
• Low maintenance due to self-monitoring
15267 14181
L
certified certified
M
N
O
P
Q
R
S
T 2 DUST MEASURING DEVICES | SICK Online data sheet | 2016-08-15 09:21:28
Subject to change without notice
DUSTHUNTER T100
TRANSMITTANCE DUST MEASURING DEVICES
Fields of application
• Emissions monitoring at power plants and waste incineration plants
• Monitoring of filter systems
• Monitoring of dust load in factory halls
• Control of fresh air and exhaust air units
Measuring ranges
Transmittance 100 ... 80 % / 100 ... 0 %
Opacity 0 ... 20 % / 0 ... 100 %
Relative opacity 0 ... 20 % / 0 ... 100 %
Extinction 0 ... 0.1 / 0 ... 2
Dust concentration 0 ... 200 mg/m³ / 0 ... 10,000 mg/m³
The measurement depends on measuring distance and dust properties
Certified measuring ranges
Dust concentration (transmittance) 0 ... 0.1 Ext / 0 ... 0.05 Ext / 0 ... 0.2 Ext / 0 ... 0.5 Ext / 0 ... 1 Ext
Response time (t90) 1 s ... 600 s
Freely adjustable
Accuracy
± 2 %
Process temperature –40 °C ... +600 °C
A Ambient temperature
Enclosure rating
–40 °C ... +60 °C
IP 66
B
Dimensions (W x H x D) 212 mm x 241 mm x 429 mm (for details see dimensional drawings)
Weight ≤ 6.5 kg
Power supply
C Voltage 24 V
Supply via control unit
E Description
Enclosure rating
Reflector unit with triple reflector
IP 66
H Ambient temperature
Enclosure rating
–40 °C ... +60 °C
IP 66
I
Analog outputs 3 outputs:
0/2/4 ... 20 mA, 750 Ω
Electrically isolated; two additional outputs if using I/O modules (option)
Analog inputs 2 inputs:
J 0 ... 20 mA
Not electrically isolated; two additional inputs if using I/O modules (option)
Digital outputs 5 relay contacts:
K 48 V, 1 A
Potential-free; for status signals
Digital inputs 4 potential-free contacts
O Indication
USB SOPAS ET
LC display
Status LEDs: "Power", "Maintenance" and "Failure"
P Operation
Dimensions (W x H x D)
Via LC-display or software SOPAS ET
Q Weight
Power supply
≤ 3.7 kg
R
Voltage 90 ... 250 V
Version with 24 V DC available as an option
Frequency 47 ... 63 Hz
S
T 4 DUST MEASURING DEVICES | SICK Online data sheet | 2016-08-15 09:21:28
Subject to change without notice
DUSTHUNTER T100
TRANSMITTANCE DUST MEASURING DEVICES
Weight ≤ 13.5 kg
Power supply
Voltage 90 ... 250 V
Version with 24 V DC available as an option
Frequency 47 ... 63 Hz
Power consumption ≤ 70 W
Auxiliary gas connections Purge air
18 kg
B
Power supply
Three-phase current 3-phase, Δ: 200 ... 240 V, 50 Hz, 2.6 A, 350 W
3-phase, Δ: 220 ... 275 V, 60 Hz, 2.3 A, 450 W
C
3-phase, Y: 345 ... 415 V, 50 Hz, 1.5 A, 350 W
3-phase, Y: 380 ... 480 V, 60 Hz, 1.3 A, 450 W
Auxiliary gas connections Purge air: 40 mm
D Test functions
Integrated components
Pressure switch (switching point –35 hPa)
E
Dimensional drawings (Dimensions in mm (inch))
F DHT-T10 sender/receiver unit
429 (16.89)
401 (15.79)
104 (4.09)
309 (12.17)
H
(2.56)
13 (.51)
65
Ø 126
205 (8.07)
200 (7.87)
(4.96)
241 (9.49)
I
Ø 100
(3.94)
Ø 11 (0.43)
J 125 (4.92)
K
208 (8.19) 168 (6.61)
355 (13.98)
L
Ø 60 (2.36)
M
212 (8.35)
208 (8.19)
202 (7.95)
N
O
P
Q
R
S
T 6 DUST MEASURING DEVICES | SICK Online data sheet | 2016-08-15 09:21:28
Subject to change without notice
DUSTHUNTER T100
TRANSMITTANCE DUST MEASURING DEVICES
364 (14.33)
336 (13.23)
104 (4.09)
244 (9.61)
(2.56)
65
Ø 126
13 (.51)
200 (7.87)
205 (8.07)
(4.96)
241 (9.49)
Ø 100
(3.94)
Ø 11 (0.43)
60 (2.36)
208 (8.19) 103 (4.06)
290 (11.42)
Ø 60 (2.36)
212 (8.35)
208 (8.19)
202 (7.95)
MCU-N control unit; wall-mounting enclosure, compact version (for non-hazardous areas only)
210 (8.27)
160 (6.30) 120 (4.72)
340 (13.39)
320 (12.60)
300 (11.81)
Meas
A MCU-P control unit; wall-mounting enclosure, compact version (for non-hazardous areas only)
B
C
D
455 (17.91)
420 (16.54)
400 (15.75)
Meas
E
F
300 (11.81) 220 (8.66)
I Ø 10 (0.39)
J
K
550 (21.65)
470 (18.50)
L
M
40
(1.57)
N 40
(1.57)
470 (18.50)
550 (21.65)
128 (5.04)
258 (10.16)
O
P
Q
R
S
T 8 DUST MEASURING DEVICES | SICK Online data sheet | 2016-08-15 09:21:28
Subject to change without notice
SICK at a glance
Online data sheet
SICK is one of the leading manufacturers of intelligent sensors and sensor solutions for industrial applica-
tions. A unique range of products and services creates the perfect basis for controlling processes securely
and efficiently, protecting individuals from accidents and preventing damage to the environment.
We have extensive experience in a wide range of industries and understand their processes and require-
ments. With intelligent sensors, we can deliver exactly what our customers need. In application centers in
Europe, Asia and North America, system solutions are tested and optimized in accordance with customer
specifications. All this makes us a reliable supplier and development partner.
Comprehensive services complete our offering: SICK LifeTime Services provide support throughout the ma-
chine life cycle and ensure safety and productivity.
Worldwide presence:
Contacts and other locations – www.sick.com