ABB Navigator 600
ABB Navigator 600
D
The Company
We are an established world force in the design and manufacture of measurement products for industrial process control, flow
Navigator 600 Silica measurement, gas and liquid analysis and environmental applications.
Multi-stream As a part of ABB, a world leader in process automation technology, we offer customers application expertise, service and support
worldwide.
We are committed to teamwork, high quality manufacturing, advanced technology and unrivalled service and support.
The quality, accuracy and performance of the Company’s products result from over 100 years experience, combined with a
continuous program of innovative design and development to incorporate the latest technology.
Introduction
The Navigator 600 Silica is an advanced colorimetric
analyzer used to measure the level of silica in the steam
water cycle in power plants. This multi-stream version can
sample up to six independent streams sequentially.
This manual applies only to Navigator 600 Silica model
numbers AW641/xxxxx9xx
Configuration Level
Common Configuration
Measurement
Logging Level*
Logging
Contents
1 Safety ............................................................................... 3 5 Configuration ................................................................ 25
1.1 Health & Safety ........................................................ 3 5.1 Front Panel Controls .............................................. 25
1.2 Electrical Safety – CEI / IEC 61010-1:2001-2 ........... 3 5.2 Navigation and Editing ........................................... 25
1.3 Symbols – CEI / IEC 61010-1:2001-2 ...................... 4 5.2.1 Text Editing ................................................ 25
1.4 Product Recycling Information ................................. 5 5.2.2 Numeric Editing .......................................... 25
1.5 Product Disposal ...................................................... 5 5.2.3 Other Methods of Editing ........................... 26
1.6 Restriction of Hazardous Substances (RoHS) ........... 5 5.2.4 Menus ........................................................ 26
1.7 Chemical Reagents .................................................. 5 5.3 Common ............................................................... 28
1.8 Safety Precautions ................................................... 5 5.3.1 Setup ......................................................... 28
1.9 Safety Conventions .................................................. 6 5.3.2 Screen ....................................................... 28
1.10 Safety Recommendations ........................................ 6 5.3.3 Time ........................................................... 29
1.11 Service and Repairs ................................................. 6 5.3.4 Security ...................................................... 30
1.12 Potential Safety Hazards .......................................... 6 5.3.5 User ........................................................... 32
5.3.6 Messages .................................................. 33
2 Overview .......................................................................... 7 5.4 Measurement ....................................................... 34
5.4.1 Setup ......................................................... 34
5.4.2 Streams ..................................................... 35
3 Principle of Operation ..................................................... 8
5.4.3 Diagnostics ................................................ 35
3.1 Chemical Principle .................................................... 8 5.5 Alarm Relays ......................................................... 36
3.2 Liquid Handling ...................................................... 10 5.6 Current Outputs ..................................................... 38
3.3 Multi-Stream Operation .......................................... 12 5.7 Logging ................................................................. 39
3.4 Single Stream Operation for Maintenance .............. 14 5.7.1 Recording .................................................. 39
3.5 Out of Sample ........................................................ 14 5.7.2 Chart .......................................................... 39
5.7.3 Archive ....................................................... 40
4 Installation ..................................................................... 15 5.8 Communications ................................................... 41
4.1 Accessories ........................................................... 15 5.8.1 Ethernet ..................................................... 41
4.2 Sampling Requirements ......................................... 16 5.8.2 email 1 and email 2 .................................... 42
4.3 Location ................................................................. 16
4.4 Reagent Bottle Location ......................................... 17 6 Start-up ......................................................................... 43
4.4.1 Behind Mounting Plate or Wall .................... 17 6.1 Commissioning the Analyzer .................................. 43
4.4.2 To Side of Analyzer ..................................... 17 6.1.1 Start-up and Configuration ......................... 43
4.4.3 Below Analyzer ........................................... 17 6.2 Performing a Baseline Calibration .......................... 44
4.5 Mounting ................................................................ 18
4.5.1 Mounting Dimensions ................................. 18
7 Calibration & Maintenance ........................................... 45
4.5.2 Mounting the Analyzer ................................ 19
4.5.3 Mounting the Reagent Shelf (Optional) ........ 19 7.1 Zero Calibration ..................................................... 46
4.6 Electrical Connections ............................................ 20 7.2 Secondary Calibration ........................................... 47
4.6.1 Connections Overview ................................ 21 7.3 Cleaning ................................................................ 47
4.6.2 Accessing and Making Connections ........... 22 7.4 Setup Calibration / Clean ....................................... 48
4.6.3 Ethernet Connections ................................. 23 7.5 Calibration ............................................................. 49
4.6.4 Alarm Relay Contact Protection and
Interference Suppression ............................ 23 8 Maintenance ................................................................. 50
4.7 Preparing the Analytical Section ............................. 24 8.1 Scheduled Maintenance ........................................ 50
4.7.1 Connecting Sample Inlet and Drain Line(s) .. 24 8.1.1 Chemical Solutions ..................................... 50
4.7.2 Connecting the Contaminated Drain Tube .. 24 8.1.2 Solution Replacement ................................ 51
8.1.3 Annual Service ........................................... 51
8.2 Periodic Maintenance ............................................ 53
8.2.1 Sample Filter Replacement ......................... 53
8.2.2 Reagent Tube Filter Replacement ............... 53
8.3 DC Fuse Replacement ........................................... 54
8.4 Manual Test Settings ............................................. 54
8.5 Start / Stop Pumps ................................................ 54
IM/NAV6S/MS–EN Rev. D 1
Navigator 600 Silica
Multi-stream Contents
13 Specification ..................................................................64
2 IM/NAV6S/MS–EN Rev. D
Navigator 600 Silica
Multi-stream 1 Safety
1 Safety
Information in this manual is intended only to assist our customers in the efficient operation of our equipment. Use of this manual for
any other purpose is specifically prohibited and its contents are not to be reproduced in full or part without prior approval of the
Technical Publications Department.
Safety advice concerning the use of the equipment described in this manual or any relevant Material Safety Data Sheets (where
applicable) may be obtained from the Company, together with servicing and spares information.
IM/NAV6S/MS–EN Rev. D 3
Navigator 600 Silica
Multi-stream 1 Safety
The equipment is protected through double This symbol indicates the need for protective hand
insulation. wear.
This symbol, when noted on a product, indicates a Electrical equipment marked with this symbol may
potential hazard that could cause serious personal not be disposed of in European public disposal
injury and / or death. systems. In conformity with European local and
national regulations, European electrical equipment
The user should reference this instruction manual
users must now return old or end-of-life equipment
for operation and / or safety information.
to the manufacturer for disposal at no charge to
This symbol, when noted on a product enclosure the user.
or barrier, indicates that a risk of electrical shock
Products marked with this symbol indicates that
and / or electrocution exists and indicates that only
the product contains toxic or hazardous
individuals qualified to work with hazardous
substances or elements. The number inside the
voltages should open the enclosure or remove the
symbol indicates the environmental protection use
barrier.
period in years.
This symbol indicates that the marked item can be
hot and should not be touched without care.
4 IM/NAV6S/MS–EN Rev. D
Navigator 600 Silica
Multi-stream 1 Safety
Note. For return for recycling, please contact the equipment manufacturer or supplier for instructions on how to return end-of-life
equipment for proper disposal.
ABB is committed to ensuring that the risk of any environmental damage or pollution caused by any of its products is
minimized as far as possible. The European Waste Electrical and Electronic Equipment (WEEE) Directive (2002/96/EC)
that came into force on August 13 2005 aims to reduce the waste arising from electrical and electronic equipment; and
improve the environmental performance of all those involved in the life cycle of electrical and electronic equipment.
In conformity with European local and national regulations (EU Directive 2002/96/EC stated above), electrical equipment
marked with the above symbol may not be disposed of in European public disposal systems after 12 August 2005.
Warning. To familiarize yourself with handling precautions, dangers and emergency procedures, always review the Material
Safety Data Sheets prior to handling containers, reservoirs, and delivery systems that contain chemical reagents and standards.
Protective eye wear and protective hand wear. is always recommended when contact with chemicals is possible.
IM/NAV6S/MS–EN Rev. D 5
Navigator 600 Silica
Multi-stream 1 Safety
Warning. In this manual, a warning is used to indicate a condition that, if not met, could cause serious personal injury and / or
death. Do not move beyond a warning until all conditions have been met.
If a warning sign appears on the instrument itself, refer to Precautionary Labels – UL Certification and Electrical Safety – CEI/IEC
61010-1:2001-2 for an explanation.
Caution. A caution is used to indicate a condition that, if not met, could cause minor or moderate personal injury and / or
damage to the equipment. Do not move beyond a caution until all conditions have been met.
Note. A note is used to indicate important information or instructions that must be considered before operating the equipment.
Warning. The installation of the instrument must be performed exclusively by personnel specialized and authorized to work on
electrical installations, and in accordance with relevant local regulations.
6 IM/NAV6S/MS–EN Rev. D
Navigator 600 Silica
Multi-stream 2 Overview
2 Overview
The measurement of silica in a sample involves the addition of To maintain optimum measurement accuracy, the analyzer
various chemical reagent solutions* to the sample, in a specific performs a zero and secondary calibration automatically, at
order, under constant temperature conditions. The result is a predetermined intervals, using solenoid valves.
chemical complex, in solution, that has a characteristic color.
Data is stored in the analyzer's internal memory and can be
The absorbance of this colored complex is proportional to the
archived either to an SD Card or via an internet connection. The
concentration of the silica in the original sample, thus making it
SD card can also be used to upgrade the analyzer's software –
possible for the measurement to be made optically. During
see Appendix F, page 85.
operation, a signal generated from the sensing system is
converted by the analyzer into data and this information is *For information about reagent solutions, contact your local ABB
presented on the display. representative.
Cable Entries
(Both Sides
of the Case)
Hinged Door to Liquid
Handling Section
Display Electronics Section
Keypad
Peristaltic Pump
Reaction Block
*Date Label
Constant-head Unit
*A second (identical) Date Label is fixed to the underside of the Electronics Section
IM/NAV6S/MS–EN Rev. D 7
Navigator 600 Silica
Multi-stream 3 Principle of Operation
3 Principle of Operation
3.1 Chemical Principle
The chemical method used by the analyzer is based on the 6. The reduction solution is added to reduce the yellow
reaction of ammonium molybdate with silicate species, in acidic molybdosilicic acid complex to the more
conditions, to form a yellow molybdosilicic acid complex. To optically-absorbent blue form. The solution enters the third
improve the sensitivity of the method the yellow complex is reaction coil that provides a 1 minute delay.
reduced to the blue form. The absorbance of the blue complex
7. The amount of blue color formed is directly proportional to
is measured spectroscopically in the optical system.
the concentration of silica in the sample. The fully
Referring to Fig. 7.3, the sequence of events in the analyzer is as developed solution passes to the measuring cuvette
follows: where the intensity of the color is measured.
1. Sample flows into a reaction coil that is maintained at a 8. During the zero calibration the analyzer generates a zero
temperature of 45 °C (113 °F). solution by diverting the sample to the point where the
second acid is added. The pH at this point is too low for
2. The first acid reagent is added to the sample to reduce the
any silica in the sample to react with the molybdate, so a
pH to a value between 1.4 and 1.8.
zero solution is produced (refer also to Section 7.1, page
3. Ammonium molybdate is added to the acidified sample. 46.
4. The first reaction coil provides a 2-minute delay where the During the zero calibration the system allows any silica
yellow molybdosilicic acid is developed. present in the first acid and molybdate solutions
5. The second acid reagent is then added to reduce the pH (Reagents 1 and 2) to form the yellow molybdosilicic acid
value further to between 0.8 and 1.0. This is the value complex. This is then converted to the blue form when it
required to stop the formation of the yellow complex. The reacts with the reduction solution.
solution enters the second delay coil that provides a
further 2 minutes for the reaction to take place.
8 IM/NAV6S/MS–EN Rev. D
Navigator 600 Silica
Multi-stream 3 Principle of Operation
1 Minute
0.8 to 1.0 pH providing reducing conditions
to stop formation of yellow complex Reduction Solution
2 Minutes
1.4 to 1.8 pH providing conditions for the Second Acid
production of yellow ß-molybdosilicic acid
2 Minutes
Molybdate
First Acid
Sample
Auto-zero valve
IM/NAV6S/MS–EN Rev. D 9
Navigator 600 Silica
Multi-stream 3 Principle of Operation
10 IM/NAV6S/MS–EN Rev. D
Navigator 600 Silica
Multi-stream 3 Principle of Operation
Secondary Debubbler
6
To Drain
Cuvette Valve
2 Peristaltic Pump
Cuvette
Reaction
4 Block
Auto-zero Valve
5
Primary To Drain
3
Debubbler
Secondary Pre-heater
Calibration Valve Assembly
Cleaning Valve
Sample
Filter
1
Sample(s)
Cleaning Calibration Reduction 2nd Acid Molybdate 1st Acid
Solution Solution
IM/NAV6S/MS–EN Rev. D 11
Navigator 600 Silica
Multi-stream 3 Principle of Operation
12 IM/NAV6S/MS–EN Rev. D
Navigator 600 Silica
Multi-stream 3 Principle of Operation
IM/NAV6S/MS–EN Rev. D 13
Navigator 600 Silica
Multi-stream 3 Principle of Operation
If all enabled streams are out of sample, but there are disabled
streams with sample present, the analyzer continues to run on
the available disabled stream. The diagnostic message 'No valid
streams available' is displayed in the diagnostic window. Stream
values are not updated, but the analyzer continues to run and
perform scheduled calibrations and cleans.
As soon as an enabled stream has sample present the
multistream sequence initiates and the readings update
accordingly.
14 IM/NAV6S/MS–EN Rev. D
Navigator 600 Silica
Multi-stream 4 Installation
4 Installation
4.1 Accessories
The following accessories are supplied with the analyzer:
4 x 2.5 l reagent bottle and cap
1 x 500 ml cleaning solution bottle and cap
1 x 500 ml standard solution bottle and cap
1 x DataManager software CD
IM/NAV6S/MS–EN Rev. D 15
Navigator 600 Silica
Multi-stream 4 Installation
4.3 Location
For general location requirements refer to Fig. 4.1. Install indoors in a clean, dry, well ventilated and vibration-free location giving easy
access and where short sample lines can be used. Avoid rooms containing corrosive gases or vapors – for example, chlorination
equipment or chlorine gas cylinders.
Select a location away from strong electrical and magnetic fields. If this is not possible, particularly in applications where mobile
communications equipment is expected to be used, screened cables within flexible, earthed metal conduit must be used.
It is also advisable to ensure drains are adjacent and near ground level, so that the analyzer waste outlet can be as short as possible,
together with maximum fall.
Mount reagent solutions less than 1 m (3.28 ft) from the bottom of the analyzer housing – see Fig. 4.2, page 17.
Ensure that the power supply and power isolation switch are adjacent to the analyzer.
16 IM/NAV6S/MS–EN Rev. D
Navigator 600 Silica
Multi-stream 4 Installation
Protective Cover
(not supplied)
Maximum 1 (3.28)
IM/NAV6S/MS–EN Rev. D 17
Navigator 600 Silica
Multi-stream 4 Installation
4.5 Mounting
4.5.1 Mounting Dimensions
Dimensions in mm (in.)
314 (12.36) 182 (7.16)
271 (10.67)
638 (25.12)
770 (30.31)
Maximum 1000 (39.4)
122 (4.80)
289 (11.34)
184 (7.25)
Reagent bottles mounted on optional brackets (two bottles per bracket) Example – two brackets mounted side-by-side
– see Sections 4.4.1, 4.4.2 and 4.4.3, page 17 for reagent bottle mounting options with end plates attached to wall
18 IM/NAV6S/MS–EN Rev. D
Navigator 600 Silica
Multi-stream 4 Installation
4.5.2 Mounting the Analyzer 4.5.3 Mounting the Reagent Shelf (Optional)
82
(3.23)
525 75 76 75
(20.7) (2.95) (3.00) (2.95)
81
(3.18)
IM/NAV6S/MS–EN Rev. D 19
Navigator 600 Silica
Multi-stream 4 Installation
20 IM/NAV6S/MS–EN Rev. D
Navigator 600 Silica
Multi-stream 4 Installation
Note. Field wiring terminal screws must be tightened to a torque of 0.6 to 0.8 Nm (5 to 7 lbf.in).
Socket for
Optional Digital
Communications
NO NO Out of Service
TB2
Hold/Maintenance C
C Relay
Relay NC
NC
Ethernet RJ45
Connector NO NO
TB3 Calibration in
Calibration Failed C
C Progress Relay
Relay NC
O/P 6 NC
O/P 5 NO NO
C Alarm 1 Relay
TB4
C
Alarm 2 Relay NC
O/P 4 NC
TB7
Current Outputs
O/P 3 NO
NO Alarm 3 Relay
TB5
C
Alarm 4 Relay C
O/P 2 NC
NC
O/P 1
NO NO
Alarm 5 Relay
TB6
C C
Alarm 6 Relay
NC NC
TB8
Optional Digital AC Fuse 1 A Type T or **DC Fuse 12.5 A Type T
Communications
L +
N –
Use fuse rated:
E TB1
E
AC Supply 1 A (max.), Type T, 250 V
DC Supply* 12.5 A 125 V DC Type T 100 V to 240 V AC ±10 % 18 to 36 V DC
Manufacturer* SCHURTER, (90 V min. to 264 V max.) **see Fig. 8.4, page 54, for
Model SPT 5 x 20 Series 50 / 60 Hz DC fuse replacement details
IM/NAV6S/MS–EN Rev. D 21
Navigator 600 Silica
Multi-stream 4 Installation
Note.
Cable entry holes are located on both sides of the enclosure.
Application board connection labels for the terminal blocks are identified in Fig. 4.8, page 21.
Warning. Remove all power from supply, relay and any powered control circuits and high common mode voltages before
accessing or making any connections.
D
E
J I
G F H
B C
Note. Cable glands are supplied with single- and 12. Close the electronics section door and turn door retaining
twin-holed bushes. Use the single-holed bush for the screws A 1/4 turn clockwise to secure.
mains power cable.
22 IM/NAV6S/MS–EN Rev. D
Navigator 600 Silica
Multi-stream 4 Installation
4.6.3 Ethernet Connections 4.6.4 Alarm Relay Contact Protection and Interference
Suppression
Warning. Remove all power from supply, relay and any
powered control circuits and high common mode voltages NC C NO
Relay Contacts
before accessing or making any connections.
R C
E H G C Load
D External L N
AC Supply
A – AC Applications
I Relay Contacts
NC C NO
Diode
F
B A Load
External
DC Supply
+ –
B – DC Applications
The Ethernet gland is different from the other connections to If the relays are used to switch loads on or off, the relay contacts
accommodate an RJ45 plug: can become eroded due to arcing. Arcing also produces RFI
that can cause analyzer malfunctions and incorrect readings. To
1. Referring to Fig. 4.9: minimize the effects of RFI, arc suppression components are
a. Turn door retaining screws A 1/4 turn required; these are resistor / capacitor networks for AC
counter-clockwise and open the electronics section applications or diodes for DC applications. These components
door. are connected across the load.
b. Using a cross-head screwdriver, remove 4 screws B Maximum relay ratings are:
and remove transparent cover plate C. 250 V, 5 A AC, 1250 VA (non-inductive)
2. Referring to Fig. 4.10: 30 V, 5 A DC 150 W
a. Slide retaining clip A off blanking plug B and For AC applications, the value of the resistor / capacitor network
remove plug. depends on the load current and inductance that is switched.
b. Fit cable gland C and secure using nut D. Initially, fit a 100R / 0.022 µF RC suppressor unit. If the analyzer
malfunctions the value of the RC network is too low for
c. Remove gland nut E and feed cable F through it.
suppression and an alternative value must be used.
d. Fit split-bush G and split-washer H to cable F.
For DC applications fit a diode – see Fig. 4.11. For general
e. Feed the cable through cable gland C and into the applications use an alternative IN5406 type (600 V peak inverse
electronics section enclosure. voltage at 3 A).
f. Plug RJ45 connector I into the RJ45 socket on the
application board (see Fig. 4.8, page 21 for location Note. The minimum voltage must be >12 V and the
details) and tighten gland nut E. minimum current >100 mA for reliable switching.
IM/NAV6S/MS–EN Rev. D 23
Navigator 600 Silica
Multi-stream 4 Installation
9 mm (1/3 in.)
PVC Tubing
A
A
6 mm (1/2 in.)
B
PVC Tubing
9 mm (1/3 in.)
PVC Tubing
Fig. 4.13 Connecting the Contaminated Drain Tube
Referring to Fig. 4.12: If samples are to be recycled, use separate drains for
contaminated and sample outlets.
1. Connect sample inlet tube to constant-head unit sample
inlet connector A.
2. Connect drain tube to constant-head unit outlet connector
B.
24 IM/NAV6S/MS–EN Rev. D
Navigator 600 Silica
Multi-stream 5 Configuration
5 Configuration
5.1 Front Panel Controls 5.2 Navigation and Editing
Depending on the type of field to be edited, the software
provides a variety of methods for entering values.
5.2.1 Text Editing
If the field to be edited requires text, a keyboard is displayed:
IM/NAV6S/MS–EN Rev. D 25
Navigator 600 Silica
Multi-stream 5 Configuration
26 IM/NAV6S/MS–EN Rev. D
Navigator 600 Silica
Multi-stream 5 Configuration
Invalid Password
Entered
Displayed only if 'Security
system' parameter is set to
Displayed only if 'Security
'Basic' – see on-line help
system' parameter is set to
system
'Advanced' – see on-line
help system
Configuration Level
Protected
Valid Password
Entered
Note.
1. If 'New Configuration' or 'Open Configuration' is selected and the modified configuration file is saved, new data files for all
log files are created and any unarchived data is lost.
2. Existing security configuration parameters are retained when a configuration is opened from file or when a new
configuration is loaded (the security remains as currently configured). Check 'Load security configuration from file' to
overwrite the current configuration with data from the file to be loaded.
The option to load or retain the security configuration applies only to Advanced Security mode and is available only to the
System Administrator (User 1). If a new or existing configuration file is opened by a user other than the System
Administrator, existing security settings are retained.
IM/NAV6S/MS–EN Rev. D 27
Navigator 600 Silica
Multi-stream 5 Configuration
5.3 Common
There are nine Common Configuration screens:
5.3.1 Setup
Fields Description
The instrument (analyzer) tag text is displayed in the top-left corner of the operator views. Up to 20
Instrument Tag characters can be used.
The instrument (analyzer) tag is also displayed on the analyzer on configuration and audit log files.
5.3.2 Screen
Fields Description
Screen saver wait time The time delay for the screen-saver. The screen dims after the time set.
28 IM/NAV6S/MS–EN Rev. D
Navigator 600 Silica
Multi-stream 5 Configuration
5.3.3 Time
Fields Description
Warning. Changing the time can result in the permanent loss of data. Once it is changed a warning
Date and Time
is displayed stating that recording is disabled until the configuration has been saved.
If 'Daylight Saving – Enable' is set to USA or Europe, the start date is displayed but cannot be edited.
Daylight Saving – Start
If 'Daylight Saving – Enable' is set to 'Custom' the date and time can be edited.
If 'Daylight Saving – Enable' is set to USA or Europe, the end date is displayed but cannot be edited.
Daylight Saving – End
If 'Daylight Saving – Enable' is set to 'Custom' the date and time can be edited.
IM/NAV6S/MS–EN Rev. D 29
Navigator 600 Silica
Multi-stream 5 Configuration
5.3.4 Security
'User 1' is the System Administrator and is the only user that has access to the 'Security type' parameter. User 1 details cannot be
changed by any other user.
There are two types of security:
Basic
Enables up to four users to have access to the 'Configuration' menu. Each user has a password that can have up to four digits.
A separate password can be set to gain access to the 'Calibration & Maintenance' and 'Logging' menus. All four users use this
password.
Advanced
Enables up to twelve users to have password-protected access to any of the 'Configuration', 'Calibration & Maintenance' or
'Logging' menus. Each user can be assigned a unique 20-digit (alphanumeric) case-sensitive password. A minimum password
length can be set.
Note. A separate password for access to the 'Calibration & Maintenance' and 'Logging' menus is not required.
User 1 can set permissions for each user to access the 'Calibration & Maintenance' and / or 'Logging' menus. Access to the
'Configuration' menu can also be set – see Section 5.3.5, page 32.
– Passwords are set by User 1. If permissions are set, users can subsequently change their own passwords.
– Password expiry dates can be set.
– Inactive user accounts can be disabled after a set time.
– Password failure limits can be set.
– A minimum password length can be set.
30 IM/NAV6S/MS–EN Rev. D
Navigator 600 Silica
Multi-stream 5 Configuration
Fields Description
Operator level security If set to 'On' and 'Security type' is set to 'Basic' an additional 'Operator level password' field is
displayed.
If set to 'On' and 'Security type' is set to 'Advanced', all users are required to enter their user
password to gain access to the 'Calibration & Maintenance' and 'Logging' menus.
Displayed only if 'Security system' is set to 'Basic' and 'Operator level security' is set to 'On'. All
Operator level password users are required to enter this password to gain access to the 'Calibration & Maintenance' and
'Logging' menus.
The following tabs are displayed only if 'Security System' is set to 'Advanced'.
Passwords are set initially by User 1 (System Administrator) but any user can subsequently change
their own password.
Reconfigure preset
When set to 'Yes' each user must change their password after it is used for the first time following
initial configuration.
Select the number of days that the password is valid for. When a password expires, the user is
Password expiry
prompted to provide a new password.
Inactive user disabling Select the number of days after which an inactive user's access privileges are de-activated.
Enter the number of consecutive incorrect password entries allowed by a user. If the number of
Password failure limit incorrect entries exceeds this limit, the user's access privileges are de-activated and can be
reinstated only by the System Administrator (User 1).
Min password length Sets the minimum length required for user's passwords.
IM/NAV6S/MS–EN Rev. D 31
Navigator 600 Silica
Multi-stream 5 Configuration
5.3.5 User
Fields Description
If 'Security system' is set to 'Basic' this tab lists the four users, User 1 to User 4. Selecting a user opens a new page with two
fields:
Name – the user’s name, up to 20 characters.
Password – each user can be assigned a unique 4-digit security code for Configuration level access.
If 'Security System' is set to 'Advanced' and User 1 (administrator) is logged on, the 'User' tab shows additional fields:
A page opens with two checkboxes to select whether User 1 has 'Calibration and Maintenance' and
User 1 Access
/ or 'Logging access'.
User 1's password – a unique 20-character (alphanumeric) security code. A minimum password
User 1 Password
length applies.
View / Edit Other Users Selects the other user's access levels and passwords. If selected additional fields appear:
User X Name Where X is the user number (2 to 12) – up to 20 characters can be used.
User X Access Where X is the user number (2 to 12). A dialog box is displayed listing the access available for the
user:
Calibration & Maintenance
Logging
Configuration (No access)
Configuration (Load)
Configuration (Limited)
Configuration (Full)
User X Password Where X is the user number (2 to 12). The password for User X.
If 'Security System' is set to 'Advanced' and a user other than User 1 is logged on, the 'User' tab has three fields.
These fields can be edited only if User 1 has set the security field 'Reconfigure preset' to 'Yes' – See page 31.
Where X is the user number (2 to 12).
Where X is the user number (2 to 12). A dialog box is displayed listing the access available for the
user:
User X Access
Calibration & Maintenance
Logging
User X's password – a unique 20-character (alphanumeric) security code. A minimum password
User X Password
length applies.
32 IM/NAV6S/MS–EN Rev. D
Navigator 600 Silica
Multi-stream 5 Configuration
5.3.6 Messages
Fields Description
Up to 24 messages can be defined to indicate a particular event or action has occurred. These are
Messages
displayed on the chart when the relevant annotation is enabled.
IM/NAV6S/MS–EN Rev. D 33
Navigator 600 Silica
Multi-stream 5 Configuration
5.4 Measurement
There are three tabs in the 'Measurement' screen:
5.4.1 Setup
Fields Description
The display units for silica concentration. Three options are available:
ppb
Display Units
µg/l
µg/kg
The reaction temperature range is 37 to 50 °C (98 to 122 °F) in 0.1 °C (0.18 °F) increments. The
analyzer is optimized to run at a reaction temperature of 45 °C (113 °F).
Reaction / Pre-Heat Any changes to the reaction temperature may affect analyzer performance.
Temp The pre-heater temperature range is 37 °C to 80 °C (98 to 176 °F) in 0.1 °C (0.18 °F) increments. The
analyzer is optimized for operation at a pre-heater temperature of 75 °C (167 °F).
Any changes to the pre-heater temperature may affect analyzer performance.
Up Scale Time The time taken for the analyzer to measure a full step change from a low concentration sample to a
high concentration sample (12 to 60 minutes).
Down Scale Time The time taken for the analyzer to measure a full step change from a high concentration sample to a
low concentration sample (14 to 60 minutes).
34 IM/NAV6S/MS–EN Rev. D
Navigator 600 Silica
Multi-stream 5 Configuration
5.4.2 Streams
Fields Description
If the difference between the low and high value is too small, the trace displayed appears very noisy.
The difference between the low and high value must be at least 50 ppb.
5.4.3 Diagnostics
Fields Description
The Reagent Bottle Capacity can be set between 0.1 and 10.0 liters. This should be changed only if
Reagent Bottle Capacity the initial volume of each reagent installed is not 2.5 liters.
This value is used to predict the level of reagent remaining in the bottle.
The early warning alarm point for the reagent solutions. Can be set between 1 and 30 days.
Reagent Limit
Set to a suitable interval to allow new reagents to be ordered.
IM/NAV6S/MS–EN Rev. D 35
Navigator 600 Silica
Multi-stream 5 Configuration
Concentration
Hysteresis
Trip Point
Hysteresis
Concentration
Alarm On Alarm On
Alarm Off
Concentration
Hysteresis
Trip Point
Hysteresis
Concentration
Alarm On Alarm Latched Alarm On Alarm Latched
Concentration
Hysteresis
Trip Point
Hysteresis
Concentration
Alarm On Alarm On
Alarm Off
36 IM/NAV6S/MS–EN Rev. D
Navigator 600 Silica
Multi-stream 5 Configuration
There are six tabs in the Alarm Relays screen, one for each alarm:
Fields Description
Each of the six alarms can be configured independently to one of the following sources:
None – no other fields are visible
Alarm Source
Stream 1 to Stream 6
Cleaning in progress
If 'Alarm Source' is set to Stream X, the alarm type can be set to:
High / Low process – see Fig. 5.3, page 36.
High / Low latch – see Fig. 5.4, page 36.
Alarm Type
High / Low annunciate– see Fig. 5.5, page 36.
Out of sample – the alarm state is active if an out-of-sample condition occurs in the selected
stream source.
When an alarm trip value is exceeded, the alarm does not become active until the time hysteresis
value has expired. If the signal goes out of the alarm condition before the time hysteresis has expired,
Hysteresis the hysteresis value is reset – see page 36 for hysteresis actions.
The hysteresis value is set in concentration units (0 to 5000 ppb), and the hysteresis time is set in
seconds (0 to 5000 s).
If set to 'Yes' the alarm relay is normally energized and is de-energized when an alarm condition
occurs.
Fail Safe
If set to 'No' the alarm relay is normally de-energized and is energized when an alarm condition
occurs.
If set to 'On' all changes in the alarm state in the Alarm Event log are recorded – see Section 9.5.2,
Log Enable
page 59.
IM/NAV6S/MS–EN Rev. D 37
Navigator 600 Silica
Multi-stream 5 Configuration
Fields Description
The high and low limits for the output range. Both values can be set independently (0 to 5000 ppb).
Output Range
If the difference between the zero and span is too small, the output is very noisy.
Out of sample indicator. If set to 'Yes', the output goes to the default output value when an out of
Out of Sample Ind.
sample condition occurs for the selected stream source.
Default Output The output value used when an out of sample condition occurs and 'Out of Sample Ind.' is set to
'Yes' (0 to 22 mA).
38 IM/NAV6S/MS–EN Rev. D
Navigator 600 Silica
Multi-stream 5 Configuration
5.7 Logging
There are three tabs in the 'Logging' screen:
5.7.1 Recording
Fields Description
The frequency (up to 720 minutes) at which stream data is recorded to the analyzer's internal
Sample Rate
memory and the SD card (if fitted).
5.7.2 Chart
Fields Description
IM/NAV6S/MS–EN Rev. D 39
Navigator 600 Silica
Multi-stream 5 Configuration
5.7.3 Archive
Used to configure the data that is to be recorded on the SD card – see Section 9, page 55.
When external archive media contains approximately 300 files, its read / write performance becomes too slow, archiving is stopped
automatically and the icon is displayed alternating with the icon. In this condition, data continues to be recorded on the
internal memory. Replace the SD card with an empty card to prevent loss of unarchived data.
Fields Description
Archive file format The file format that archive files saved as.
Toggles between 'Text Format' and 'Binary Format'.
Archive file enables At least one of these options must be selected for data to be archived automatically to an SD Card.
A dialog box is displayed showing the log files that are to be recorded:
Text format file containing the stream data.
Binary format file containing stream data.
Alarm event log file enable (*.e)
Audit log file enable (*.a)
New file interval Available only if 'Wrap' is set to 'Off' and text format is selected.
The interval that text format stream data files are created. Options are:
Off
Hourly
Daily
Monthly
Wrap If set to 'On', the oldest archived data on the SD card is deleted automatically when the SD card
approaches its maximum capacity.
If set to 'Off', archiving stops when the SD card is full. The analyzer continues to store data internally
– see Section 9, page 55. When an empty SD card is inserted, archiving continues from the point
that the last archive was made.
40 IM/NAV6S/MS–EN Rev. D
Navigator 600 Silica
Multi-stream 5 Configuration
5.8 Communications
There are three tabs in the I/O modules screen:
5.8.1 Ethernet
Configures the way that the analyzer can be accessed via an Ethernet network – see Appendix E, page 81.
Note. Changes to the 'IP-address', 'Subnet mask' and 'Default Gateway' parameters are implemented only after the analyzer
has been restarted. Change the addressing parameters, exit and save the configuration, then power down and restart the
analyzer.
Fields Description
The IP-address to be assigned to the analyzer. The IP address is used by the TCP/IP protocol to
IP-address distinguish between different devices. The address is a 32 bit value expressed with four values (0 to
255), each separated by a period (.).
The subnet mask is used to indicate which part of the IP address is for the network ID and which is
Subnet mask for the host ID. Set each section that is part of the network ID to 255.
For example, 255.255.255.0 indicates first 24 bits are for the network ID.
The IP address for the 'Default gateway' (for example, router, switch) required to communicate with
Default Gateway other networks.
The default setting is '0.0.0.0'
Enables up to four users to access the analyzer via the internet. A dialog box is displayed with four
options:
User name – the name of user granted FTP access (up to 12 characters).
Password – the password required for FTP login (up to 12 characters).
FTP user 1 to FTP user 4
Access Level – toggles between 'Full' or 'Read-only' access.
Remote Operation – toggles between 'None', 'Operator' or 'Configuration'.
Note. If a user is given full access via FTP, that user is able to select from the saved
configuration files in the analyzer.
IM/NAV6S/MS–EN Rev. D 41
Navigator 600 Silica
Multi-stream 5 Configuration
The analyzer can be configured to send emails to a maximum of 6 recipients in response to certain events. The addressees can all
subscribe to the same SMTP server or the analyzer can be configured to send emails via 2 different SMTP servers to a maximum of 3
addressees per server.
Up to 10 independently configurable triggers may be enabled to generate an email when the selected source becomes active. When
a trigger source becomes active, an internal 1 minute delay timer is started. At the end of that minute, an email is generated that
includes, not only the event that initiated the delay timer, but every other event that occurred during the delay period together with any
enabled reports. The data returned in the email therefore reflects the real-time alarm state at the time the email was generated, not
the state when the first trigger source became active.
Each email sent includes a link to the analyzer's embedded web server, enabling the analyzer's data and status to be viewed remotely
using an internet browser on a PC.
Fields Description
SMTP Server IP Address The IP address of the SMTP server through which emails are to be routed.
The trigger for an email message to be sent. A dialog box is displayed with three options:
None – no triggers are set.
Archive state – a dialog box is displayed with six options:
– Archive media not present
– Too many files on the archive media
– Archive media 100% full
– Archive media 80% full
– Archive media present
– Archive on-line
42 IM/NAV6S/MS–EN Rev. D
Navigator 600 Silica
Multi-stream 6 Start-up
6 Start-up
6.1 Commissioning the Analyzer
7. Purge the zero calibration line and the secondary and
Warning. clean solution lines as follows:
Protective eye wear and protective hand wear must be worn a. Press the key, select 'Calibration & Maintenance'
when contact with chemicals is possible. Take appropriate and select the 'Manual Test Settings' button.
Health & Safety precautions.
b. Select 'Test Pumps' and increase the sample pump
speed to 5 rpm using the key pad.
Refer to Appendix A for details of reagents, calibration solutions
and cleaning solutions. c. Exit the 'Test Pumps' page and enter the 'Test Valves'
page.
Before operating the analyzer:
d. Select 'Cleaning ON' to energize the clean valve.
1. Ensure the analyzer has been installed correctly – see Leave on for at least 1 minute to flush the clean
Section 4, page 15. solution line.
2. Ensure all reagent, calibration and cleaning solution e. De-energize the clean valve and energize the
bottles are filled with the appropriate solutions. secondary calibration valve in the 'Test Valves' page.
3. Ensure the reagent bottles are mounted safely and Leave on for a least 1 minute to flush the secondary
securely in the trays* (if supplied) and the bottle tops are solution line.
fitted securely. f. De-energize the secondary calibration valve and
energize the zero cal valve. Leave on for at least 30
*These accessories are not included with all models. seconds.
6.1.1 Start-up and Configuration g. Once all lines have been flushed, select 'Normal Mode'
The following steps describe briefly how to start up and to exit 'Manual Test Settings'. The analyzer returns to
configure the analyzer prior to operation: the correct pump speed and de-energizes the valves.
1. Clip the pump plattens into place. h. A recovery period is initiated (this can be dismissed to
2. Ensure at least one sample stream is connected and access the menu).
flowing.
8. Press the key, select 'Calibration & Maintenance' and
3. Ensure the drain tube from the analyzer is: select 'Solution Replacement'. Select 'Replace Cleaning
a. connected and directed to a drain Solution' and then 'Replace Secondary Cal Solution' – this
updates the dates they were installed.
b. as short as possible and routed as vertically as
possible. 9. Select 'Replace All Reagent Solutions' – this initiates a
purge sequence followed by a recovery sequence that
4. Switch on power to the analyzer. enables the analyzer to stabilize.
5. After the initial power up procedure a message is 10. Initiate a Manual Clean procedure.
displayed asking if the 'Power Up Sequence' should be
aborted. Select 'Stop' to abort the automatic start up. At this stage a baseline calibration must be performed.
6. Check that the failure icon ( ) is not displayed in the
diagnostic window. If it is displayed, resolve the cause of Note. The analyzer reading takes some time to settle after
the failure before proceeding. the initial Commissioning procedure – this is due mainly to
the gradual wetting of the internal plastic flow path.
Therefore, if an accurate calibration is required it is advisable
to wait several hours before initiating a calibration.
IM/NAV6S/MS–EN Rev. D 43
Navigator 600 Silica
Multi-stream 6 Start-up
Note.
Before starting the initial Baseline calibration, it is
important to check the reagent lines are fully purged
and there is no air flowing through the reaction block.
Check there are no air bubbles in the cuvette by
checking the stability of the 'Measure Detector mV'
signal in the 'Measurement Status' / 'Diagnostic
Information' page.
If the signal is moving by more than 20 mV over a 30
second period, air may still be present in the cuvette.
Initiate a manual clean sequence.
44 IM/NAV6S/MS–EN Rev. D
Navigator 600 Silica
Multi-stream 7 Calibration & Maintenance
Invalid
Password
Password
Required
(if set)
Valid
Password
No Password
Required
A full two-point calibration of the analyzer is performed by running an internally-generated zero calibration and replacing the sample
automatically with a secondary calibration solution of a known concentration.
IM/NAV6S/MS–EN Rev. D 45
Navigator 600 Silica
Multi-stream 7 Calibration & Maintenance
1 Minute
0.8 to 1.0 pH providing reducing conditions
to stop formation of yellow complex Reduction Solution
2 Minutes
1.4 to 1.8 pH providing conditions for the Second Acid
production of yellow ß-molybdosilicic acid
2 Minutes
Molybdate
Small quantity of
SiO2 in reagents
First Acid
Sample
Auto-zero valve
46 IM/NAV6S/MS–EN Rev. D
Navigator 600 Silica
Multi-stream 7 Calibration & Maintenance
Calibration Factor 1.0 Table 7.1 shows the timings for the full calibration sequence.
Secondary Recovery
Zero Normal
Activity Cal (introduce
(if required) Operation
(if required) sample)
Timing
(minutes) 36 24 18 N/A
Default
0%
Absorbance Calibration
Zero Offset – Energized Energized Energized De-energized
Baseline Relay
2 ppb
Table 7.1 Calibration Sequence
Fig. 7.5 Secondary Calibration
7.3 Cleaning
The analyzer can be programmed to introduce an alkaline
During the Secondary calibration a known value silica
cleaning solution through the normal solution flowpath
(secondary) standard is passed through the normal sample
automatically.
flowpath, allowing a full chemical reaction to take place. The
silica chemistry is so reproducible that this response can be The alkaline solution has 2 main functions:
compared to a theoretical response built into the analyzer. The
to remove the build up of precipitation or deposits in the
theoretical response has an ideal calibration factor of 1.0.
system due to the acidic chemistry conditions inside the
If the response is not perfect, compensation is applied to make reaction block and optical unit. If cleaning is carried out
the readings correct and the resulting Calibrating factor provides frequently, build up, and its effects on the zero offset, are
an indication of how much compensation is required – see minimized.
Fig. 7.5.
to keep all the surfaces very clean and wetted, reducing
If this is within acceptable limits, the calibration passes and the the effects of bubble formation due to sample and reagent
necessary compensation is applied. If the results are outside degassing.
acceptable limits the analyzer shows a calibration fail.
Cleaning can be initiated either manually or automatically at
Calibration options comprise:
pre-programmed intervals with an adjustable duration.
Baseline (2-point, resets zero) – runs both a zero and a
The ideal frequency and duration of cleaning is determined by
secondary calibration. Perform this routine only after the
monitoring the zero offset parameter at the end of each
reagents have been replaced or after an annual service.
calibration sequence. If the zero offset is not increasing by more
This routine can only be initiated manually.
than 1 ppb from one zero calibration to the next, cleaning can
Zero (1-point) – required to compensate for any zero drift. be less frequent.
Performed weekly or as required to ensure accurate silica
The recommended initial starting point is to clean frequently for
measurement.
a short period, for example, weekly for 30 minutes.
Secondary (1-point) – to verify sensitivity (as required) or if
the secondary calibration solution has been replaced.
Zero & Secondary (2-point) – can be performed instead of
a weekly zero calibration to compensate for drift and verify
sensitivity.
IM/NAV6S/MS–EN Rev. D 47
Navigator 600 Silica
Multi-stream 7 Calibration & Maintenance
Fields Description
The automatic calibration schedule for the analyzer. A dialog box is displayed with the following options:
Zero Frequency – the frequency zero calibrations are performed (off to 7 days).
Next Zero – if 'Zero Frequency' has been enabled, this displays / sets the scheduled date for the
zero calibration. Once a scheduled calibration has been performed the date is incremented by the
number of days set in 'Zero Frequency'.
Important. A warning is displayed if the Next Zero is scheduled 6 hours (or less) after an
automatic clean. It is recommended that an automatic clean and zero are separated by 6 hours to
ensure the best zero accuracy; this also applies to manually triggered functions.
Secondary Frequency – the frequency secondary calibrations are performed (off to 4 weeks).
Next Secondary – if 'Secondary Frequency' has been enabled, this displays / sets the scheduled
Automatic Schedule date for the secondary calibration. Once a scheduled calibration has been performed the date is
incremented by the number of days set in 'Secondary Frequency'.
Important. A warning is displayed if the 'Next Secondary' is scheduled 6 hours (or less) after an
automatic clean. It is recommended that they are separated by 6 hours to ensure the best
calibration accuracy; this also applies to manually-triggered functions.
Clean Frequency – the frequency the analyzer is cleaned (off to 4 weeks).
Next Clean – if 'Clean Frequency' has been enabled, this displays / sets the scheduled date for
analyzer cleaning. Once a scheduled clean has been performed the date is incremented by the
number of weeks set in 'Clean Frequency'.
Manual Clean – to ensure zero accuracy, perform a zero at least 6 hours after the clean
sequence has completed.
The value of the secondary calibration solution used (10 to 5000 ppb) – see Appendix A.2, page 66.
Secondary Calibration
Value Important. This is the same value that is used in the 'Calibration' options – see Section 6.2. Changing
the value here also changes the value in the 'Calibration' options.
Automatic Scheduled The duration used in scheduled automatic cleans (20, 30, 60 minutes, 2, 4, 6, 8, 12 hrs). See Note.
Clean Duration
Manual Clean The duration used when a manual clean is initiated (20, 30, 60 minutes, 2, 4, 6, 8, 12 hrs). See Note.
Duration
Reset Cal. Factor & Immediately resets the calibration factor and the calibration offset to the default factory setting. This
Offset option can be used when a calibration has not been performed correctly.
A non-adjustable clean duration is used automatically when the analyzer is returned to service following out-of-service conditions
such as excessive zero offset, power loss for more than 24 hours and loss of sample for more than 24 hours.
Note. All Cleans are followed by an 18 minute recovery period.
48 IM/NAV6S/MS–EN Rev. D
Navigator 600 Silica
Multi-stream 7 Calibration & Maintenance
7.5 Calibration
Enables manual calibrations to be performed.
Fields Description
The value of the secondary calibration solution used (10 to 5000 ppb)
Secondary Calibration – see Appendix A.2, page 66.
Value Note. This is the same value used in the 'Setup Calibration / Clean' options. Changing the value here
also changes the value in the 'Setup Calibration / Clean' options.
If the analyzer does not have any active alarms or warnings that can affect a calibration, the following options are available:
Both a zero and a secondary calibration are performed. A new calibration factor is calculated and the
Zero Offset is reset to 0.0 ppb. A recovery period follows.
Baseline (2 point, resets
A baseline calibration must be performed when the reagents have been replaced or an annual
zero offset)
service has been performed.
A progress bar is displayed indicating the maximum time required for the operation.
A zero calibration is performed and a new zero offset calculated. The recovery period follows.
Zero (1 point)
A progress bar is displayed indicating the time required for the operation.
A secondary calibration only is performed and a new calibration factor calculated. A recovery period
Secondary (1 point) follows.
A progress bar is displayed indicating the time required for the operation.
Both a zero and secondary calibration are performed. A new zero offset and calibration factor are
Zero & Secondary calculated. A recovery period follows.
(2 point)
A progress bar is displayed indicating the time required for the operation.
IM/NAV6S/MS–EN Rev. D 49
Navigator 600 Silica
Multi-stream 8 Maintenance
8 Maintenance
8.1 Scheduled Maintenance
Warning.
8.1.1 Chemical Solutions
Ensure personal protective equipment (PPE) such as Reagent, calibration and cleaning solutions are necessary to
gloves and eye protection are worn during any maintain the analyzer in operation. Use the bottles supplied. If
maintenance and that any spillages are cleaned-up solutions are stored, keep them in polythene bottles with tightly
using clean water. fitting tops.
To familiarize yourself with handling precautions,
dangers and emergency procedures, always review Caution. Avoid contamination of the solutions.
the Material Safety Data Sheets prior to handling When replacing solutions, the reagent and standard
containers, reservoirs, and delivery systems that solution bottles must be emptied and rinsed with high
contain chemical reagents and standards. purity water before re-filling.
Take care if cleaning any spillages and observe all Do not top solution bottles up, always refill them.
relevant safety instructions – see Section 1, page 3.
Bottle tops must be fitted to keep out dust (dust can
Clean the analyzer using a damp cloth only; mild be a source of silica).
detergent can be used as a cleaning aid. Do not use
Acetone or any organic solvents. Analyzer performance relies heavily on the integrity of
these solutions, so it is very important to prepare,
Isolate electrical components before maintenance or store and handle them with care.
cleaning.
Observe all health and safety procedures for handling Take care when storing the bottles. Ensure they are date
chemicals – see Section 1, page 3. stamped, used in strict rotation and not used after the expiry
date. Take care to avoid spillages if solutions are transferred
from bulk containers to the analyzer's own bottles. Any spillages
Note. Due to the specialized nature of this type of analyzer, must be diluted with water and flushed to drain.
maintenance must be performed only by qualified
personnel. Training can be provided by the Company via a For information about reagent solutions refer to Appendix A,
network of local or overseas agents, or may be performed page 66.
on the users' premises.
50 IM/NAV6S/MS–EN Rev. D
Navigator 600 Silica
Multi-stream 8 Maintenance
Fields Description
IM/NAV6S/MS–EN Rev. D 51
Navigator 600 Silica
Multi-stream 8 Maintenance
Referring to Fig. 8.1: 11. Ensure the tubing is aligned correctly over the rollers.
1. Unlock catch A, use handle B to slide the analytical 12. Refit the platens C and pump end plate F.
section forward and lock it into place using catch A.
2. Release the pump tube clips on the downstream side to Note. When fitting the pump end plate, ensure the
prevent back-siphoning of reagents (and possible shaft / bearing adaptor engages with the central hole
contamination of solutions in the bottles) when platen C in the pump shaft and that the pivot pin aligns with the
is unlatched. mating hole in the pump end plate F.
4. Remove pump end plate F (pull off squarely), ensuring Tighten screw D and clip each platen C into place:
that the shaft-to-bearing adaptor is not lost.
5. Unlatch and rotate each platen C, to access the pump
tubing. Slide the platens off the pivot pin.
6. Unlatch the other side of the pump tubing G and discard
the tubing.
7. Slide off and discard each capstan H.
8. Replace each capstan. Ensure the capstans are aligned to
maximize mixing effectiveness.
9. If not already fitted, clip cover I to the outer capstan.
10. Clip the new tubing into place. Ensure the correct tubing is
used for each pump and the seals are in place:
Rear 14. Unlock the analytical section and slide it back into the
Green Brown normal operation position. Lock the analytical section into
Blue Orange place.
Front Violet
Gray Red For part numbers, refer to the Spares section – see Appendix H,
page 87.
Wet the mating surfaces with demineralized water and Once the service has been completed, the annual service date
squeeze between the two points shown below to ensure is reset and a purge routine is run automatically to purge the
that tubing clips into place correctly: system of air. After the purge sequence, perform a Baseline
calibration.
52 IM/NAV6S/MS–EN Rev. D
Navigator 600 Silica
Multi-stream 8 Maintenance
E
D
IM/NAV6S/MS–EN Rev. D 53
Navigator 600 Silica
Multi-stream 8 Maintenance
54 IM/NAV6S/MS–EN Rev. D
Navigator 600 Silica
Multi-stream 9 Logging
9 Logging
Invalid
Password Password
Required
(if set)
Valid
Password
No Password
Required
Switches internal logging on / off (Available only from the chart view)
Note.
The amount of time that data remains in the analyzer's internal
memory depends on the sample rate and the number of To avoid loss of archive data, always place the media
streams – see Table 9.1. Sample data can be saved to card off-line before removing it.
removable media as either binary-encoded or The analyzer accepts SD cards with a maximum
comma-separated files. capacity of up to 4 Gb.
Additional files can also be archived: The analyzer complies with approved industry
Alarm event log data standards for memory cards. SanDisk Standard
Grade memory cards have been fully tested and are
Audit log data recommended for use with ABB analyzers. Other
Configuration files brands may not be fully compatible with this device
and therefore may not function correctly, leading to
Screen capture images irretrievable loss of data.
IM/NAV6S/MS–EN Rev. D 55
Navigator 600 Silica
Multi-stream 9 Logging
56 IM/NAV6S/MS–EN Rev. D
Navigator 600 Silica
Multi-stream 9 Logging
When an SD card is inserted and there is <1 day (Binary format) 9.2 Reset Archiving
or <1 hour (Text format) of data in internal memory, a dialog box If 'Reset Archiving' is selected, all data in the internal memory is
is displayed giving the user the choice of putting the media re-archived to external media.
on-line or remaining offline. If no selection is made within 10
seconds, the media card is placed on-line automatically: Note. Ideally, insert a blank media storage card before
selecting this function.
To re-archive data:
1. Insert an SD card, with sufficient free space, into the
analyzer.
2. Select 'Off-line' in the menu.
3. Select 'Reset archiving' in the menu.
4. Select 'On-line' in the menu.
5. Select the data to be archived if >1 hour (Text format) or
>1 day (Binary format) of data in internal memory and
press the key.
Note. Data stored in the internal memory buffer can still be
transferred to the archive media when the archive media is
9.3 File Viewer
A prompt is displayed providing the option to view either internal
placed on-line again (providing it is not off-line so long that
or external files (if an SD card is present).
the un-archived data in the internal memory is overwritten).
IM/NAV6S/MS–EN Rev. D 57
Navigator 600 Silica
Multi-stream 9 Logging
Text files – *.A** extension The analyzer's power is lost then restored.
Binary files – *.AE* extension The analyzer is taken offline and the archive media
removed, replaced or refitted.
9.5 Data Files The analyzer's configuration is changed.
Text format archived data is stored in a comma-separated value
format and can be imported directly into a standard One of the current files exceeds the maximum permissible
spreadsheet, for example, Microsoft® Excel. The files can also size.
be saved in ASCII text format. When the daylight saving period starts or ends.
Alternatively, you can carry out detailed graphical analysis of the
data on a PC using our DataManager data analysis software. Note. The analyzer's internal clock can be configured to
adjust automatically at the start and end of 'Daylight Saving
Time' periods.
58 IM/NAV6S/MS–EN Rev. D
Navigator 600 Silica
Multi-stream 9 Logging
Summertime starts at 2:00am on 30th March 2011 and the clock Monthly files start at 00:00:00 on the first of the month.
changes automatically to 3:00am.
The existing file is closed and a new file is created filename:
30Mar11AW6SIL~DS.D00
IM/NAV6S/MS–EN Rev. D 59
Navigator 600 Silica
Multi-stream 10 Chart Functions
10 Chart Functions
60 IM/NAV6S/MS–EN Rev. D
Navigator 600 Silica
Multi-stream 11 Alarm Acknowledge
IM/NAV6S/MS–EN Rev. D 61
Navigator 600 Silica
Multi-stream 12 Diagnostic Information
12 Diagnostic Information
Ensure that a note is made of all the information in the 12.2 System Information
Diagnostic screens when contacting the Company regarding The system information screen provides details regarding the
spares or enquires. calibration, maintenance and software release.
Last / Next Zero Calibration – the date of the last
12.1 Reagent and Solution Levels
Displays the predicted levels of the reagent solutions, along with calibration and the scheduled date for the next – see
the approximate number of days remaining and the last date Section 7.4, page 48
that the reagents were replaced. Reagent consumption is Last / Next Secondary Calibration – the date of the last
monitored by the analyzer and a diagnostic message appears calibration and the scheduled date for the next – see
once a lower limit (customer configurable) is reached. Also Section 7.4, page 48
displays whether the clean and calibration solutions are LOW or
Zero Offset – a baseline calibration sets this offset to zero.
OK and the last date these solutions were replaced.
The result of a subsequent zero calibration is then
compared with the baseline to give the zero offset.
If the zero offset goes outside pre-configured limits the
calibration is not accepted and an alarm is triggered.
Calibration Factor – the calculated value determined
during the last secondary calibration carried out. This
gives an indication of the sensitivity of the analyzer to
silica.
If the calibration factor goes outside pre-configured limits,
the calibration is not accepted, an alarm is triggered and
the actual calibration is shown here in red.
Next Service – the due date for the next annual service –
see Section 8.1.3, page 51.
Software Version – the software release number.
LL App / OS / HMI App – the low level application,
operating system and human-machine interface numbers.
IP address / MAC address – communication settings.
62 IM/NAV6S/MS–EN Rev. D
Navigator 600 Silica
Multi-stream 12 Diagnostic Information
IM/NAV6S/MS–EN Rev. D 63
Navigator 600 Silica
Multi-stream 13 Specification
13 Specification
Silica Measurement Display
Range Color*, passive matrix, liquid crystal display (LCD) with built-in
Fully user programmable 0 to 5000 ppb SiO2, minimum range 0 to 50 backlight and brightness adjustment
ppb Diagonal display area 144 mm (5.7 in.)
64 IM/NAV6S/MS–EN Rev. D
Navigator 600 Silica
Multi-stream 13 Specification
DS/NAV6S-EN Rev. N
IM/NAV6S/MS–EN Rev. D 65
Navigator 600 Silica
Multi-stream Appendix A – Solutions
Appendix A – Solutions
A.1 Reagent Solutions Choose a secondary solution concentration close to the point
where the greatest accuracy is required. Do not use solutions
Warning. containing less than 50 ppb silica because solution preparations
below this level can cause significant inaccuracy.
Before handling containers, reservoirs and delivery
systems that contain chemical reagents and A stock solution of 1000 mg l–1 silica, SiO2, can be obtained in 1
standards, review the Material Safety Data Sheets to of the following 3 ways:
familiarize yourself with handling precautions, dangers 1. The preferred method:
and emergency procedures.
purchase a 1000 mg l–1 silica, SiO2 stock solution from a
Take care if cleaning any spillages and observe all proprietary chemical supplier,
relevant safety instructions – see Section 1, page 3.
or
General cleaning of the instrument must be carried
out using a damp cloth only; mild detergent can be 2. From sodium fluorosilicate:
used as a cleaning aid. Do not use Acetone or any a. Dissolve 3.133 (±0.001) g of sodium fluorosilicate
organic solvents. (Na2SiF6 – the purest grade available) in approximately
Observe all Health & Safety procedures for handling 900 ml of high purity water. Ensure all solids have
chemicals. dissolved fully by stirring the solution for several hours.
b. Transfer the solution to a 1 liter volumetric flask and
A volume of 2.5 l of each reagent solution maintains the analyzer make up to the mark with more high purity water.
in operation for a minimum of three months. The bottles and c. Store the solution in a polythene bottle.
associated tubing are color-coded to identify each reagent.
Contact your local ABB representative for further information or
about reagent solutions. 3. From sodium metasilicate
66 IM/NAV6S/MS–EN Rev. D
Navigator 600 Silica
Multi-stream Appendix A – Solutions
A.3 Cleaning Solution A.3.1 Regular Cleaning Solution (for Automatic Cleaning
It is important that the internal pipework of the analyzer is Operation)
cleaned using the automatic cleaning schedule – see To prepare 1 liter of regular cleaning solution:
Section 7.4, page 48. This prevents gradual fouling by 1. Dissolve 5 g of analytical grade sodium hydroxide pellets,
molybdate precipitation in the pipework which can introduce NaOH, in approximately 600 ml of high purity water in a
errors and cause noise. plastic container. Allow the solution to cool to ambient
Regular cleaning is normally performed using a regular cleaning temperature.
solution – see Appendix A.3.1. If the pipework in the analyzer is 2. Add to the solution 1 g analytical grade di-sodium EDTA
extremely dirty or the tubing has become blocked, extra and stir to dissolve.
strength cleaning solution can be used – see Appendix A.3.2.
The analyzer may take some time to stabilize after the clean if 3. Transfer the solution to a 1 liter measuring cylinder and
the stronger solution is used. make to the mark with more high purity water. Mix well
and store in a tightly stoppered plastic bottle.
Warning. Sodium hydroxide is extremely caustic and must A.3.2 Extra Strength Cleaning Solution (for Troubleshooting
be handled with great care. Wear gloves and eye protection. and More Intense Wet Section Cleaning)
To prepare 1 liter of extra strength cleaning solution:
1. Dissolve 50 g of analytical grade sodium hydroxide pellets,
NaOH, in approximately 600 ml of high purity water in a
plastic container. Allow the solution to cool to ambient
temperature.
2. Add to the solution 5 g analytical grade di-sodium EDTA
and stir to dissolve.
3. Transfer the solution to a 1 liter measuring cylinder and
make to the mark with more high purity water. Mix well
and store in a tightly stoppered plastic bottle.
IM/NAV6S/MS–EN Rev. D 67
Navigator 600 Silica
Multi-stream Appendix B – Diagnostics and Alarms
Monitor in Service
ON
The analyzer is operating correctly.
A / D Error (1, 2 or 3)
A hardware error exists on the main board.
Power the analyzer down, wait for 10 seconds and power back up.
All calibrations
Notes. OFF ON
and cleans
If the error persists contact your local ABB representative.
Replacement board (see Appendix H, page 89 – part number
AW600065).
68 IM/NAV6S/MS–EN Rev. D
Navigator 600 Silica
Multi-stream Appendix B – Diagnostics and Alarms
IM/NAV6S/MS–EN Rev. D 69
Navigator 600 Silica
Multi-stream Appendix B – Diagnostics and Alarms
If the temperature is low, check the heater LED on the main board is
lit. If it is not, contact your local ABB representative.
70 IM/NAV6S/MS–EN Rev. D
Navigator 600 Silica
Multi-stream Appendix B – Diagnostics and Alarms
If the temperature is low, check the heater LED on the main board is
lit. If it is not, contact your local ABB representative.
IM/NAV6S/MS–EN Rev. D 71
Navigator 600 Silica
Multi-stream Appendix B – Diagnostics and Alarms
Cleaning In Progress
Scheduled
A clean routine is being run.
OFF ON calibrations
A progress bar on the cleaning page indicates the time to and cleans
completion.
Calibration In Progress
A calibration routine is being run. A progress bar indicates the time
to completion. Scheduled
ON calibrations
1. Wait for the routine to be completed. and cleans
2. Cancel the routine and allow the analyzer to complete the
recovery routine.
In Annual Service
The analyzer has been taken off-line for the annual service. Scheduled
OFF ON calibrations
Complete the servicing tasks and press the key when finished
and cleans
to reset the annual service alarm and place the analyzer on-line
(after a purge and recovery period).
In Purge
This routine is initiated at power up and after solution replacement
or an annual service.
ON
The analyzer has been taken off-line while the system is primed and
the reaction block temperature is stabilized. When the routine is
completed, the analyzer is ready for service.
In Recovery
The analyzer is in recovery mode.
This recovery routine introduces sample back into the system to
prepare it for return to run mode.
72 IM/NAV6S/MS–EN Rev. D
Navigator 600 Silica
Multi-stream Appendix B – Diagnostics and Alarms
Out of Sample n
The sample indicated in stream 'n' is not present.
The indicated stream is excluded from the sampling sequence until
the analyzer has detected that sample has returned.
IM/NAV6S/MS–EN Rev. D 73
Navigator 600 Silica
Multi-stream Appendix B – Diagnostics and Alarms
Date / time or daylight saving start / end changed Inactive low annunciate alarm
74 IM/NAV6S/MS–EN Rev. D
Navigator 600 Silica
Multi-stream Appendix C – Troubleshooting
Appendix C – Troubleshooting
C.1 Analyzer Malfunction C.2 Single-Stream Mode for Maintenance
For the analyzer to function correctly and to specification, the To carry out maintenance, switch to single-stream mode
following conditions must be met: (provided that a single stream is available) by disabling all
streams except one – see Section 5.4.1, page 34. If no streams
The correct reagents and calibration solutions must be
are available, introduce a solution via the secondary calibration
connected.
valve.
The order that the reagents are introduced into the
Single-stream mode stops multi-stream sequencing and
reaction block must be correct.
enables the display and the current output to respond with 3
The flowrate of all solutions must be correct. minute updates to color changes in the cuvette. This mode is
The analyzer's electronics (transmitter and optical used to check the basic performance of the analyzer, such as
measuring device) must be functioning correctly. response or drift, without waiting for the normal stream update.
Failing calibration Check the system for visible air bubbles – see Section C.3.1, page 76
(calibration
Check that reagents enter the reaction block in the correct order – see Section C.3.3, page 78
parameters
consistent from one Check reagents are correct – see Section C.3.3, page 78
calibration to the Check solution flowrates – see Section C.3.3, page 78
next)
Check operation of zero and secondary calibration valves – see Section C.3.4, page 79
Check mV values from the optical system – see Section C.3.2, page 77
Incorrect readings Check each stream reading independently. Switch off the sample flow to all streams except the one
(Calibrations pass being checked and remove the inlet tube from those constant head units, draining down the constant
with consistent head chamber. Measure each stream for at least 20 minutes or until stable readings are obtained. Ensure
results) the analyzer is configured for single stream operation – see Section C.2.
If readings are correct for individual streams but incorrect when the analyzer is running in multi-stream
mode, the problem is likely to be a leaking stream select valve. Replace the valve (see Appendix H,
page 88 – part number AW600034).
If all readings are the same as when run in full multistream mode:
1. Disconnect the cleaning solution tube from the clean valve manifold and check if the reading
changes after 30 minutes to 1 hour. If it does, the problem is likely to be a leaking clean valve.
Replace the valve (see Appendix H, page 88 – part number AW600034).
2. Disconnect the secondary calibration solution tube from the calibration valve manifold and check if
the reading changes after 30 minutes to 1 hour. If it does, the problem is likely to be a leaking
calibration valve. Replace the valve (see Appendix H, page 88 – part number AW600034).
If no problems can be found, it suggests that the laboratory method could be inaccurate.
Noisy / unstable Check the system for visible air bubbles – see Section C.3.1, page 76
readings
Check mV values from the optical system for stability – see Section C.3.2, page 77
Check solution flowrates – see Section C.3.3, page 78
IM/NAV6S/MS–EN Rev. D 75
Navigator 600 Silica
Multi-stream Appendix C – Troubleshooting
Symptom Action
Air in system Check for air bubbles in the reaction block and in the tube feeding the optical unit.
If bubbles are found, check if they are present in the sample line or originate from one or more of the
reagents.
Note: Air can enter the system only from the suction side of the 2 pump assemblies.
76 IM/NAV6S/MS–EN Rev. D
Navigator 600 Silica
Multi-stream Appendix C – Troubleshooting
Symptom Action
Unstable mV values The optical system receives mV signals from the cuvette and cuvette reference sensors. For accurate
readings both these values must be stable.
Note. When viewing these readings, be aware that if the silica level in the analyzer changes gradually,
the cuvette voltage also changes gradually.
To check mV values:
1. Press and select 'Diagnostic Information', 'Measurement Status'. The 'Measurement Status'
window is displayed and a mV signal value is displayed for the cuvette and the cuvette reference
sensors.
2. If the mV values are unstable and fluctuate by more than ± 10 mV over a short period of time, there
could be bubbles developing within the cuvette.
3. If bubbles are confirmed in the cuvette, check the tube connections – see Section C.3.1 above. If
the problem persists, perform a manual system clean – see Section 7.4, page 48.
Calibration failure or If there is no response or calibrations fail, check the output of the optical sensors as follows:
no response from
1. Empty the cuvette of reacted chemicals:
optical system
a. Press the key.
b. Select 'Calibration & Maintenance'.
c. Select 'Manual Test Settings'.
d. Select 'Test Pumps'.
e. Switch the reagent pump off and increase the speed of the sample pump to 5 rpm.
f. Run the sample pump at the increased speed for 5 minutes.
g. Set the sample pump speed back to its normal (previous) speed.
4. If the Cuvette Ref signal is low, replace the optical unit (see Appendix H, page 88 – part number
AW601110).
5. If the Cuvette signal is low, perform a manual system clean – see Section 7.4, page 48. If the
problem persists, replace the optical unit (see Appendix H, page 88 – part number AW601110)
IM/NAV6S/MS–EN Rev. D 77
Navigator 600 Silica
Multi-stream Appendix C – Troubleshooting
Symptom Action
Reagent order For correct chemistry and color generation to take place in the reaction block area, the reagents must
enter the reaction block in the correct order:
1. Check that each reagent is contained in the correct color-coded reagent bottle – see Fig. A.1,
page 67
2. Check that each reagent tube is inserted in the correct reagent bottle (see Fig. 3.1, page 9)
Chemistry and color For correct chemistry and color generation to take place in the reaction block area, the reagents must
generation errors flow correctly.
To check flowrates:
1. Remove all 4 reagent tubes from the reagent bottles.
2. Increase the reagent pump speed to 10 rpm – see Section 8.4, page 54.
3. Check the air / reagent interface is moving up each of the reagent tubes at the same rate. If one or
more is moving more slower than the others:
a. Check the pump capstan for damage.
b. Check the pump tube assembly is sealed correctly. If necessary, remove and refit the relevant
assembly, ensuring the O-rings are wetted.
c. Initiate a manual clean sequence to remove any possible partial blockages.
For correct chemistry and color generation to take place in the reaction block area, the reagents must
produce the correct color changes.
To quickly check that the reagents can produce the required color changes:
1. To 100 ml of a 1000 µgl-1 silica solution or even drinking water:
a. Add 10 ml of the first reagent (red channel – first acid)
b. Add 10 ml of the second reagent (violet channel – molybdate)
Note. Reagent 3 is not used in this check as it is used to fine tune the chemistry in the analyzer for
accurate measurements
78 IM/NAV6S/MS–EN Rev. D
Navigator 600 Silica
Multi-stream Appendix C – Troubleshooting
Zero Valve
IM/NAV6S/MS–EN Rev. D 79
Navigator 600 Silica
Multi-stream Appendix D – Online Help
80 IM/NAV6S/MS–EN Rev. D
Navigator 600 Silica
Multi-stream Appendix E – Web Server
The analyzer is fitted with an ethernet card as standard and, d Language Selection – selects the language for the web
when the appropriate configuration settings have been made, pages.
users can access the analyzer's data via an Ethernet network.
e Auto-Refresh – updates the analyzer view automatically.
In addition, files can be transferred to and from the analyzer via
f View Selection – selects the required analyzer view:
an FTP connection.
– Alarm / Event Log
a Log On – click to log on to the analyzer to enable
configuration access. – Audit Log
b Analyzer View – displays the current screen of the – Chart View
analyzer. If the analyzer has a screen-saver set, it does not – Bar View
affect this view.
c Access Buttons – provides access to the analyzer's data
if the user has the appropriate access permission.
– Stream Values – see Appendix E.1, page 82.
– Reagent & Solution Levels – see Appendix E.2,
page 82
– Measurement Status – see Appendix E.3, page 82
– Valve & Pump Status – see Appendix E.4, page 82
– Statistics – see Appendix E.5, page 82
– System Information – see Appendix E.6, page 82
– Logging Status – see Appendix E.7, page 82
– Operator Messages – see Section E.8, page 83
– Configuration – see Appendix E.9, page 83
IM/NAV6S/MS–EN Rev. D 81
Navigator 600 Silica
Multi-stream Appendix E – Web Server
82 IM/NAV6S/MS–EN Rev. D
Navigator 600 Silica
Multi-stream Appendix E – Web Server
Change or retain
security configuration
An 'Operator Message' is displayed on the chart view. For Note. A user must be logged-on for the configuration
example: button to be enabled.
IM/NAV6S/MS–EN Rev. D 83
Navigator 600 Silica
Multi-stream Appendix E – Web Server
84 IM/NAV6S/MS–EN Rev. D
Navigator 600 Silica
Multi-stream Appendix F – Updating the Software
IM/NAV6S/MS–EN Rev. D 85
Navigator 600 Silica
Multi-stream Appendix G – Analyzing a Grab Sample
Appendix G –
Analyzing a Grab Sample
To analyze a grab sample:
1. Disable all streams apart from one:
a. at the 'Main Configuration' menu, select 'Measurement'
and scroll down to 'Enable Streams'.
b. press 'Enter' and untick each stream, leaving 1 ticked.
c. press the 'Menu' key and exit by 'Saving Current
Configuration' to internal memory.
5. Finish:
a. remove the level sensor form the grab sample, rinse
thoroughly with high purity water, place back in the
secondary calibration solution.
b. leave running for a 5 minutes to allow the sample line
to flush through.
86 IM/NAV6S/MS–EN Rev. D
Navigator 600 Silica
Multi-stream Appendix H – Spare Parts
IM/NAV6S/MS–EN Rev. D 87
Navigator 600 Silica
Multi-stream Appendix H – Spare Parts
AW600710 Out of sample switch new type AW600046 Reaction block temperature sensor
Blanking plug
88 IM/NAV6S/MS–EN Rev. D
Navigator 600 Silica
Multi-stream Appendix H – Spare Parts
IM/NAV6S/MS–EN Rev. D 89
Navigator 600 Silica
Multi-stream Appendix H – Spare Parts
90 IM/NAV6S/MS–EN Rev. D
Navigator 600 Silica
Multi-stream Index
Index
A I
Access Illustrated spares list ....................................................... 87
User access .............................................................. 32 Inactive user disabling ..................................................... 31
Accessories .................................................................... 15 Installation ...................................................................... 15
Alarms Location .................................................................... 16
Fail safe ..................................................................... 37 Instrument diagnostic information ................................... 68
Relay contact protection ............................................ 23 Instrument tag ................................................................ 28
Relays ....................................................................... 36 Interference suppression ................................................. 23
Annotation
Chart ......................................................................... 60 L
Archive Language ....................................................................... 28
Configuration ............................................................. 40 Logging
File types ................................................................... 58 Configuration ............................................................ 39
Automatic chemical cleaning ........................................... 47
Automatic shutdown ....................................................... 14 M
Maintenance ................................................................... 50
B Calibration ................................................................ 45
Brightness ....................................................................... 28 Cleaning ................................................................... 47
Sample filter replacement .......................................... 53
C Scheduled ................................................................ 50
Cable connections .......................................................... 22 Measurement status ....................................................... 62
Chart Messages ....................................................................... 33
Annotation ................................................................. 60 Operator ................................................................... 60
Functions .................................................................. 60 Minimum password length .............................................. 31
Chemical solutions .......................................................... 50 Mounting ........................................................................ 18
Cleaning Multi-stream operation .................................................... 12
Automatic chemical cleaning ..................................... 47
Cleaning solution ....................................................... 67 N
Setup ........................................................................ 48 Navigation ...................................................................... 25
Communications ............................................................. 41
Connections O
Cable connections ..................................................... 22
External electrical connections ................................... 20 Operation – Calibration and maintenance ....................... 45
Operator
Controls .......................................................................... 25 Messages ................................................................. 60
Current outputs ............................................................... 38
Passwords ................................................................ 31
Security .................................................................... 31
D
Outputs .......................................................................... 38
Date and time ................................................................. 29
Daylight saving .......................................................... 29, 59
Diagnostic information ..................................................... 68 P
Dimensions ..................................................................... 18 Password expiry ............................................................. 31
Display Password failure limit ...................................................... 31
Brightness ................................................................. 28 Passwords ............................................................... 30, 32
Pumps status ................................................................. 63
E
R
Editing ............................................................................. 25
Email ............................................................................... 42 Reagent levels ................................................................ 62
Ethernet .......................................................................... 41 Reconfigure preset ......................................................... 31
Recording ....................................................................... 39
Relays ............................................................................ 36
F Contact protection and interference suppression ...... 23
Fail safe ........................................................................... 37
File viewer ....................................................................... 57
Filters .............................................................................. 53
Flow schematic ................................................................. 8
Front panel controls ........................................................ 25
H
Historical review .............................................................. 60
Hysteresis ....................................................................... 37
IM/NAV6S/MS–EN Rev. D 91
Navigator 600 Silica
Multi-stream Index
S
Sample filter replacement ................................................53
Sampling requirements ....................................................16
Schematic .........................................................................8
Screen capture ................................................................28
Screen interval .................................................................60
Screen saver wait time ....................................................28
SD Card ..........................................................................56
Security ...........................................................................30
Security type ...................................................................31
Shutdown
Automatic ..................................................................14
Software
Updating ....................................................................85
Solution levels .................................................................62
Solutions ...................................................................50, 66
Spares
Illustrated spares list ..................................................87
Statistics .........................................................................63
Status
Measurement .............................................................62
Valves and pumps .....................................................63
System information ..........................................................62
T
Time ................................................................................29
Troubleshooting ..............................................................75
Tubing
Schematic ...................................................................8
U
Unpacking .......................................................................15
Updating the software .....................................................85
User settings ...................................................................32
V
Valve status .....................................................................63
W
Web server ......................................................................81
92 IM/NAV6S/MS–EN Rev. D
Products and customer support
Automation Systems Customer support
For the following industries: We provide a comprehensive after sales service via a
— Chemical & Pharmaceutical Worldwide Service Organization. Contact one of the following
— Food & Beverage offices for details on your nearest Service and Repair Centre.
— Manufacturing
— Metals and Minerals UK
— Oil, Gas & Petrochemical
ABB Limited
— Pulp and Paper
Tel: +44 (0)1453 826661
Drives and Motors Fax: +44 (0)1453 829671
— AC and DC Drives, AC and DC Machines, AC Motors to
1kV
USA
— Drive Systems
— Force Measurement ABB Inc.
— Servo Drives Tel: +1 215 674 6000
Fax: +1 215 674 7183
Controllers & Recorders
— Single and Multi-loop Controllers
— Circular Chart and Strip Chart Recorders
— Paperless Recorders
— Process Indicators
Flexible Automation
— Industrial Robots and Robot Systems
Flow Measurement
— Electromagnetic Flowmeters
— Mass Flowmeters
— Turbine Flowmeters
— Wedge Flow Elements
Marine Systems & Turbochargers
— Electrical Systems
— Marine Equipment
— Offshore Retrofit and Refurbishment
Process Analytics
— Process Gas Analysis
— Systems Integration
Transmitters
— Pressure
— Temperature
— Level
— Interface Modules
Valves, Actuators and Positioners
— Control Valves
— Actuators
— Positioners
Water, Gas & Industrial Analytics Instrumentation
— pH, Conductivity and Dissolved Oxygen Transmitters and
Sensors
— Ammonia, Nitrate, Phosphate, Silica, Sodium, Chloride,
Fluoride, Dissolved Oxygen and Hydrazine Analyzers Client Warranty
— Zirconia Oxygen Analyzers, Katharometers, Hydrogen Prior to installation, the equipment referred to in this
Purity and Purge-gas Monitors, Thermal Conductivity manual must be stored in a clean, dry environment, in
accordance with the Company's published specification.
Periodic checks must be made on the equipment's
condition. In the event of a failure under warranty, the
following documentation must be provided as
substantiation:
— A listing evidencing process operation and alarm logs
at time of failure.
— Copies of all storage, installation, operating and
maintenance records relating to the alleged faulty unit.
Contact us
ABB Limited
09.2012
Note
Process Automation We reserve the right to make technical changes or
modify the contents of this document without prior
Oldends Lane
IM/NAV6S/MS–EN Rev. D
notice. With regard to purchase orders, the agreed
Stonehouse particulars shall prevail. ABB does not accept any
Gloucestershire GL10 3TA responsibility whatsoever for potential errors or
UK possible lack of information in this document.