NUEN 605 Laboratory #7: Uranium Mass Measurement Using He Neutron Detectors
NUEN 605 Laboratory #7: Uranium Mass Measurement Using He Neutron Detectors
Introduction
The purpose of this lab is to get introduced to different types of neutron detectors and see
how detector efficiency got affected by moderation. In addition, the lab was intended to measure
the enrichment of several unknown uranium samples using the comparator method. Moreover,
the mass of an unknown uranium sample was determined using the total neutron counting
method.
To begin, the detector was connected to the preamplifier (Pre-Amp), which was
connected the amplifier (Amp). The Amp was connected to the Single Channel Analyzer (SCA)
that was connected to the Timer/Counter. In order to check the pulse height, the oscilloscope (O-
Scope) was connected to the MCA. The testing arrangement was as seen in Figure 1. The serial
Single
Pre- Timer/
Detector Amplifier Channel
Amplifier Counter
Analyzer
High
O-scope
Voltage
1
Table 1. Nuclear instrumentation modules and their serial numbers
A 252Cf sample (metallic form) was used as a neutron source during the first half of the
experiment; see Table 2 for its radioactive properties. The source current activity, A(t), was
calculated using Eq. 1, taking into account the decay of the source,
A(t) = Ao et (Eq. 1)
ln(2)
= (Eq 2.)
T12
where,
Ao: The original source activity and equals (2.3 0.23) 106 neutron/second.
Nuclide T1/2 (s) (s-1) Ao (n/s) Decay Time (s) A(t) (n/s)
252
Cf 8.34107 8.26106 (2.3 0.23) 106 3.85108 (8.86 0.88) 104
The power supply High Voltage (HV) was adjusted at 1150 V using the voltage dial. The
background counts were recorded by the 3He detector for 10 s, in which the Lower Level
Discriminator (LLD) level was gradually increased to eliminate background noise. The detector
2
efficiency was calculated for the 3He and BF3 detectors at different Polyethylene (PE)
thicknesses see Eq. 3. The PE slabs were placed in between and behind the detectors. The testing
R
= YA (Eq. 3)
where,
Figure 2. Test configuration. The PE slabs were place between the detector and the neutron
source.
3
Figure 3. Test configuration. The 3 PE slabs were placed behind the detector.
An optimum configuration was sat based on the detector efficiency, in which the PE slabs
were reconfigured to get the detector maximum efficiency (see Figures 4 & 5).
Figure 4. Moderation PE slabs Optimum configuration. One 2 PE slab were placed in all
directions except in front where two 2 inches were added.
4
Figure 5. Moderation PE slabs optimum configuration. One more 2 PE slab was placed
on the top
For the second half of the experiment, natural enriched uranium metal rods were used for
measuring the sample mass, using the Total Neutron Counting (TNC) method. The testing
optimum configuration was adapting (see Figure 6), in which two 3He detectors were using. The
first sample rod counts recorded for 45 minutes, while 30 minutes for each of the other two rods.
Figure 6. TNC testing configuration. The first measurement using one uranium rod.
5
The uranium mass was calculated according to Eq .4, as follows,
= Y (Eq. 4)
where,
The 252Cf decays (3% of the time) by spontaneous fission, in which 3.7 neutrons
(average) are emitted. The produced fast neutrons have to be thermalized in order to be detected
by 3He and BF3 detectors. This is because 3He and 10B have high neutron absorption cross
section (a) in the thermal energy region (Eth ~ 0.035 eV). The counts were recorded for 10 s
Table 3. 252Cf data at different PE thicknesses between the detector and the neutron source.
calculating the detector efficiency at different moderation thicknesses (Eq. 3). The calculations
were performed using Excel software; the results are as shown in Table 3 and plotted as in
Figure 5.
6
0.40%
0.35%
0.30%
Efficiency (%)
0.25%
0.20%
y = -1E-04x3 + 7E-05x2 + 0.0009x + 0.002
0.15%
R = 1
0.10%
0.05%
0.00%
0 0.5 1 1.5 2 2.5 3 3.5
PE Thickness (in)
polyethylene, which means that the efficiency is increasing with moderation. However, the
efficiency values at 1 and 2 inches thick are statistically the same. This means that the efficiency
was not increasing upon PE addition. Moreover, the efficiency tends to decrease for larger
amount of moderation, because the neutron absorption increase, which decreases the number of
The same measurement was conducted, where the PE slaps were placed behind the
detector as seen in Figure 3. The moderating material behind detector reflects the unscattered
neutrons, which increase the population of the neutrons near the detector, and hence increase the
detection efficiency (see Figure 8). It was noted that the detector efficiency was affected by the
thickness of moderation material. The 1 inch-thick material, was not enough to get the neutrons
thermalized, which result in lower detector efficiency. However, the detector efficiency
7
0.70%
0.60%
0.50%
Efficiency (%)
0.40%
0.20%
0.10%
0.00%
0 0.5 1 1.5 2 2.5 3 3.5
PE Thickness (in)
The 3He has higher absorption (a) than 10B, which reflects the higher efficiency of the
3
He detector over the BF3 detectors. Figure 9 shows that both of the two detectors shares the
same behavior of the moderation thickness effect on the efficiency. However, it should be
mentioned that the BF3 detectors are cheaper and hence more affordable. The 3He production is
very small compared to the demand, which justifies the high prices of 3He detectors.
8
0.40%
y = -1E-04x3 + 7E-05x2 + 0.0009x + 0.002
0.35% R = 1
0.30%
0.25%
Effeciency (%)
0.20%
He Detector
0.10%
y = -7E-06x3 - 0.0003x2 + 0.001x + 0.0007
R = 1
0.05%
0.00%
0 0.5 1 1.5 2 2.5 3 3.5
PE Thickness (in)
Figure 9. The efficiency of the BF3 detector vs 3He detector at different PE thicknesses.
The new configuration (Figures 4 and 5) was meant to reflect as many neutron as it could
and minimizes the neutron leakage, which in turn would increase the neutron population near the
detector. It was found that the recorded number of counts increased by order of five to six, see
Table 5. Moreover, the detector efficiency increased to be 2.12% compared to 0.3 0.55% for
For the TNC experiment, the background counts were recorded for 10 min, in which the
LLD was set at 0.30 V to eliminate the noise while the E adjusted to 10 V. The background
reading was 88 counts. Another 252Cf source (10Ci) was used for detector efficiency calculation
(Eq. 3) and was 2.9 0.1 %. The uranium mass was calculated according to the Eq. 4 and the
results as shown in Table 6
9
Number Number of Recording time Count rate Uranium mass
of Rods Counts (min) (s-1) (kg)
1 6259 45 2.3 1.756
2 5868 30 3.26 2.45
3 6071 30 3.37 2.55
Conclusion
It was concluded that the number of slabs affects the detector efficiency, which indicated the
importance of the moderation in the neutron detector. An optimum configuration was decided
based on the detection efficiency in which the PE slabs were rearranged to get the maximum
efficiency. On the other hand, the mass of a uranium samples were measured, in which a natural
uranium samples were recorded for 45 min, 30 min, and 30 min, consecutively. The uranium
mass were found different from the actual one (1.8 kg/rod) due to the recording time. The
Uranium metal neutron yield is relatively small which require long time recording to get the
accurate value. Moreover, the errors associated with taking measurement has to be considered in
10
11