V 770 UV Visible Spectrophotometer

The document describes the specifications of a V-770 UV-Visible Spectrophotometer. It has a Czerny-Turner grating mount optical system with a halogen lamp and deuterium lamp light source. It has a wavelength range of 190 to 2700 nm with options to 3200 nm and provides high accuracy and repeatability in wavelength and photometric measurements. It also lists various scanning and slew speeds, stray light performance, photometric range and accuracy, and standard functions and programs.

Uploaded by

caochien
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
541 views2 pages

V 770 UV Visible Spectrophotometer

The document describes the specifications of a V-770 UV-Visible Spectrophotometer. It has a Czerny-Turner grating mount optical system with a halogen lamp and deuterium lamp light source. It has a wavelength range of 190 to 2700 nm with options to 3200 nm and provides high accuracy and repeatability in wavelength and photometric measurements. It also lists various scanning and slew speeds, stray light performance, photometric range and accuracy, and standard functions and programs.

Uploaded by

caochien
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 2

V-770 UV-Visible Spectrophotometer

Czerny-Turner grating mount


Optical System Single monochromator
Fully symmetrical double-beam
Light Source Halogen lamp, Deuterium lamp
Wavelength range 190 to 2700 nm (3200 nm, option)
+/-0.3 nm (at 656.1 nm)
Wavelength accuracy
+/-1.5 nm (at 1312.2 nm)
Wavelength repeatability +/-0.05 nm (UV-Vis), +/-0.2 nm (NIR)
UV-Visible: 0.1, 0.2, 0.5, 1, 2, 5, 10 nm
L2, L5, L10 nm (low stray light mode)
M1, M2 nm (micro cell mode)
Spectral bandwidth (SBW)
NIR: 0.4, 0.8, 1, 2, 4, 8, 20, 40
L8, L20, L40 nm (low stray light mode)
M4, M8 nm (micro cell mode)
1 % (198 nm KCL)
0.005 % (220 nm NaI)
0.005 % (340 nm NaNO2)
0.005 % (370 nm NaNO2)
Stray light SBW: L2 nm

0.04 % (1420 nm: H O)2

0.1 % (1690 nm: CH2Br2)


SBW: L8 nm
UV-Visible: -4~4 Abs
Photometric range
NIR: -3~3 Abs
+/-0.0015 Abs (0 to 0.5 Abs)
Photometric accuracy +/-0.0025 Abs (0.5 to 1 Abs)
+/-0.3 %T Tested with NIST SRM 930D
+/-0.0005 Abs (0 to 0.5 Abs)
Photometric repeatability +/-0.0005 Abs (0.5 to 1 Abs)
Tested with NIST SRM 930D
Scanning speed 10~4000 nm/min (8000 nm/min in preview mode)
UV-Vis: 12,000 nm/min
Slew speed
NIR: 48,000 nm/min
0.00003 Abs
RMS noise
(0 Abs, wavelength: 500 nm, measurement time: 60 sec, SBW :2 n
Baseline stability 0.0003 Abs/hour (Wavelength: 250 nm, response: slow and SBW:
Baseline flatness +/-0.0002 Abs (200 - 2500 nm)
Detector PMT, Peltier cooled PbS
Standard functions IQ Accessory, Start Button, Analog output
Abs/%T meter, Quantitative analysis, Spectrum measurement
Standard programs Time course measurement, Fixed wavelength measurement
Validation, Daily check and Dual wavelength time course measurem
Dimensions and weight 460(W) x 602(D) x 268(H) mm, 29 kg
Power requirements 150 VA
Installation requirements Room temperature: 15-30 Celsius, humidity: below 85%