Manual T500U
Manual T500U
TRADEMARKS
All trademarks, registered trademarks, brand names or product names appearing
in this document are the property of their respective owners and are used herein
for identification purposes only. This product is based on technology licensed
from Aerodyne Research, Inc.
07834D DCN6998 i
This page intentionally left blank.
ii 07834D DCN6998
SAFETY MESSAGES
Important safety messages are provided throughout this manual for the purpose of avoiding personal
injury or instrument damage. Please read these messages carefully. Each safety message is associated
with a safety alert symbol, and are placed throughout this manual; the safety symbols are also located
inside the instrument. It is imperative that you pay close attention to these messages, the descriptions of
which are as follows:
Electrical Ground: This symbol inside the instrument marks the central
safety grounding point for the instrument.
CAUTION
This instrument should only be used for the purpose and in the manner described
in this manual. If you use this instrument in a manner other than that for which it
was intended, unpredictable behavior could ensue with possible hazardous
consequences.
For Technical Assistance regarding the use and maintenance of this instrument or any other Teledyne
API product, contact Teledyne APIs Technical Support Department:
Telephone: 800-324-5190
Email: [email protected]
or access any of the service options on our website at http://www.teledyne-api.com/
MISE EN GARDE
Cet instrument doit tre utilis aux fins dcrites et de la manire dcrite dans ce
manuel. Si vous utilisez cet instrument dune autre manire que celle pour laquelle il
a t prvu, linstrument pourrait se comporter de faon imprvisible et entraner des
consquences dangereuses.
iv 07834D DCN6998
WARRANTY
Coverage
After the warranty period and throughout the equipment lifetime, TAPI stands
ready to provide on-site or in-plant service at reasonable rates similar to those of
other manufacturers in the industry. All maintenance and the first level of field
troubleshooting are to be performed by the customer.
Product Return
All units or components returned to Teledyne API should be properly packed for
handling and returned freight prepaid to the nearest designated Service Center.
After the repair, the equipment will be returned, freight prepaid.
07834D DCN6998 v
This page intentionally left blank.
vi 07834D DCN6998
ABOUT THIS MANUAL
This manual is comprised of multiple documents, in PDF format, as listed below.
Support information such as electrostatic discharge (ESD) prevention and various communications is
presented in our manuals, which are available on the TAPI website http://www.teledyne-api.com in Help
Center>Product Manuals, under Special Manuals.
Note We recommend that all users read this manual in its entirety before
operating the instrument.
07834D DCN6998 ix
This page intentionally left blank.
x 07834D DCN6998
TABLE OF CONTENTS
Safety Messages ...........................................................................................................................................iii
WARRANTY .................................................................................................................................................. v
About This Manual .......................................................................................................................................vii
Revision History ........................................................................................................................................... ix
Table of Contents ......................................................................................................................................... xi
1. INTRODUCTION .................................................................................................................................... 19
2. SPECIFICATIONS, APPROVALS & COMPLIANCE ............................................................................. 21
2.1. Specifications ....................................................................................................................................... 21
2.2. EPA Equivalency Designation ............................................................................................................. 22
2.3. Approvals and Certifications ................................................................................................................ 22
1.1.1 Safety.......................................................................................................................... 22
1.1.2 EMC ............................................................................................................................ 22
3. GETTING STARTED .............................................................................................................................. 23
3.1. Unpacking ............................................................................................................................................ 23
3.1.1. Installation Requirements .............................................................................................. 24
3.2. Instrument Layout ................................................................................................................................ 24
3.2.1. Front Panel .................................................................................................................... 24
3.2.2. Rear Panel ..................................................................................................................... 26
3.2.3. Internal Chassis Layout ................................................................................................. 28
3.3. Connections and Setup ........................................................................................................................ 29
3.3.1. Electrical Connections ................................................................................................... 29
3.3.1.1. Connecting Power ......................................................................................... 29
3.3.1.2. Connecting Analog Inputs (AIN) Option ........................................................ 30
3.3.1.3. Connecting Analog Outputs .......................................................................... 30
3.3.1.4. Current Loop Analog Outputs (Option 41) Setup .......................................... 31
3.3.1.5. Connecting the Status Outputs ..................................................................... 33
3.3.1.6. Connecting the Control Inputs ....................................................................... 34
3.3.1.7. Concentration Alarm Relay (Option 61) ........................................................ 36
3.3.1.8. Connecting the Communications Interfaces ................................................. 37
3.3.2. Pneumatic Connections................................................................................................. 42
3.3.2.1. About Zero Air and Calibration (Span) Gas .................................................. 44
3.3.2.2. Basic Connections from Calibrator, with and Without Span Gas .................. 45
4. OVERVIEW OF OPERATING MODES .................................................................................................. 49
4.1. Sample Mode ....................................................................................................................................... 50
4.1.1. Test Functions ............................................................................................................... 50
4.1.2. Warning Messages ........................................................................................................ 51
4.2. Calibration Mode .................................................................................................................................. 51
4.3. Setup Mode .......................................................................................................................................... 52
4.3.1. Primary Setup Menu ...................................................................................................... 52
4.3.2. Secondary Setup Menu (SETUP >MORE) ................................................................... 53
5. STARTUP, FUNCTIONAL CHECKS, AND INITIAL CALIBRATION...................................................... 55
5.1. Startup .................................................................................................................................................. 55
5.1.1. Warning Messages ........................................................................................................ 55
5.2. Functional Checks................................................................................................................................ 56
5.3. Initial Calibration ................................................................................................................................... 58
5.3.1. Initial Calibration Procedure for T500U Analyzers without VALVE Options .................. 59
5.3.1.1. Verifying the Reporting Range Settings ........................................................ 59
5.3.1.2. Verifying the Expected NO2 Span Gas Concentration .................................. 61
5.3.1.3. Initial Zero/Span Calibration Procedure ........................................................ 62
6. SETUP .................................................................................................................................................... 63
6.1. CFG: Configuration Information ........................................................................................................... 63
07834D DCN6998 xi
6.2. DAS: Internal Data Acquisition System ................................................................................................ 63
6.3. RNGE: Analog Output Reporting Range ............................................................................................. 63
6.3.1. Analog Output Ranges for NO2 Concentration .............................................................. 64
6.3.2. Analog Output Reporting Range Default Settings ......................................................... 65
6.3.3. SETUP > RNGE > MODE ............................................................................................. 65
6.3.3.1. SETUP > RNGE > MODE > SNGL: Single Range Mode Configuration ....... 65
6.3.3.2. SETUP > RNGE > MODE > DUAL: Dual Range Mode Configuration ......... 66
6.3.3.3. SETUP > RNGE > MODE > AUTO: Auto Range Mode Configuration ......... 68
6.3.3.4. SETUP > RNGE > UNIT: Setting the Reporting Range Units of Measure ... 69
6.3.3.5. SETUP > RNGE > DIL: Using the Optional Dilution Ratio Feature .............. 69
6.4. PASS: Password Protection ................................................................................................................ 71
6.5. CLK: Clock, Setting time and Date ...................................................................................................... 74
6.5.1. Setting the Time of Day ................................................................................................. 74
6.5.2. Adjusting the Internal Clocks Speed ............................................................................. 75
6.6. MORE>COMM: Communications Ports .............................................................................................. 76
6.6.1. ID (Machine Identification) ............................................................................................. 76
6.6.2. INET (Ethernet) ............................................................................................................. 77
6.6.3. COM1[COM2] (Mode, Baude Rate and Test Port) ....................................................... 77
6.7. MORE>VARS: Variables Setup and Definition .................................................................................... 77
6.8. MORE>DIAG: Diagnostics Functions .................................................................................................. 80
6.8.1. Signal I/O ....................................................................................................................... 82
6.8.2. Analog Output (DIAG AOUT) ........................................................................................ 82
6.8.3. Analog I/O Configuration (DIAG AIO) ............................................................................ 83
6.8.3.1. Analog Output Voltage / Current Range Selection ........................................ 85
6.8.3.2. Calibration of the Analog Outputs ................................................................. 86
6.8.3.3. Enabling or Disabling the AutoCal for an Individual Analog Output .............. 86
6.8.3.4. Automatic Group Calibration of the Analog Outputs ..................................... 87
6.8.3.5. Automatic Individual Calibration of the Analog Outputs ................................ 89
6.8.3.6. Manual Calibration of the Analog Outputs Configured for Voltage Ranges .. 90
6.8.3.7. Manual Adjustment of Current Loop Output Span and Offset ...................... 92
6.8.3.8. Turning the Analog Output Over-Range Feature On/Off .............................. 95
6.8.3.9. Adding a Recorder Offset to an Analog Output ............................................. 96
6.8.3.10. AIN (Analog Inputs) Calibration ................................................................... 97
6.8.3.11. External Analog Inputs (XIN1XIN8) Option Configuration ....................... 98
6.8.4. Test Chan Output (Selecting a Test Channel Function for Output A4) ......................... 99
6.8.5. NO2 LED Disable ......................................................................................................... 101
6.8.6. Pump Disable .............................................................................................................. 101
6.9. SETUP>MORE>AREF:...................................................................................................................... 101
7. COMMUNICATIONS SETUP AND OPERATION ................................................................................ 103
7.1. Data Terminal/Communication Equipment (DTE DEC) ..................................................................... 103
7.2. Communication Modes, Baud Rate and Port Testing ........................................................................ 103
7.2.1. Communication Modes ................................................................................................ 104
7.2.2. Com Port Baud Rate ................................................................................................... 106
7.2.3. Com Port Testing ......................................................................................................... 107
7.3. RS-232 ............................................................................................................................................... 108
7.4. RS-485 (Option) ................................................................................................................................. 108
7.5. Ethernet .............................................................................................................................................. 109
7.5.1. Configuring Ethernet Communication Manually (Static IP Address) ........................... 109
7.5.2. Configuring Ethernet Communication Using Dynamic Host Configuration Protocol
(DHCP) .................................................................................................................................. 111
7.5.2.1. Changing the Analyzers HOSTNAME ........................................................ 113
7.6. USB Port for Remote Access ............................................................................................................. 114
7.7. Communications Protocols ................................................................................................................ 116
7.7.1. Hessen......................................................................................................................... 116
7.7.1.1. Hessen Com Port Configuration .................................................................. 116
7.7.1.2. Activating Hessen Protocol ......................................................................... 117
07834D DCN6998 xv
LIST OF FIGURES
Figure 3-1: Front Panel Layout .............................................................................................................. 25
Figure 3-2: Display Screen and Touch Control ..................................................................................... 25
Figure 3-3: T500U Rear Panel............................................................................................................... 26
Figure 3-4: Internal Layout..................................................................................................................... 28
Figure 3-5: Analog In Connector ........................................................................................................... 30
Figure 3-6: Analog Output Connector .................................................................................................... 31
Figure 3-7: Current Loop Option Installed on the Motherboard ............................................................. 32
Figure 3-8: Status Output Connector ..................................................................................................... 34
Figure 3-9: Energizing the Control Inputs .............................................................................................. 35
Figure 3-10: Concentration Alarm Relay ................................................................................................. 36
Figure 3-11: Rear Panel Connector Pin-Outs for RS-232 Mode ............................................................. 38
Figure 3-12: Pin Assignments for J11, J12 Connectors on CPU Board. ................................................ 38
Figure 3-13: Jumper and Cables for Multidrop Mode .............................................................................. 40
Figure 3-14: RS-232-Multidrop PCA Host/Analyzer Interconnect Diagram ............................................ 41
Figure 3-15: Gas Line Connections from Calibrator Basic T500U Configuration................................. 45
Figure 3-16: T500U Pneumatics with Zero Span Valves Option ............................................................. 47
Figure 3-17: T500U Pneumatics with Internal Zero Span Valves Option ................................................ 48
Figure 4-1: Front Panel Display ............................................................................................................. 49
Figure 6-1: Analog Output Connector Pin Out ....................................................................................... 64
Figure 6-2: SETUP COMM Menu ....................................................................................................... 76
Figure 6-3: COMM Machine ID ............................................................................................................ 77
Figure 6-4: Accessing the DIAG Submenus .......................................................................................... 81
Figure 6-5: Accessing the Analog I/O Configuration Submenus ........................................................... 84
Figure 6-6: Setup for Checking / Calibrating DCV Analog Output Signal Levels .................................. 90
Figure 6-7: Setup for Checking / Calibration Current Output Signal Levels Using an Ammeter ........... 92
Figure 6-8: Alternative Setup Using 250 Resistor for Checking Current Output Signal Levels .......... 94
Figure 7-1: COMM - LAN /Internet Manual Configuration ................................................................... 110
Figure 7-2: COMM LAN / Internet Automatic Configuration (DHCP) ............................................... 112
Figure 8-1: Example Default DAS Configuration in APICOM Interface ............................................... 128
Figure 8-2: DAS Configuration through a Terminal Emulation Program ............................................. 143
Figure 10-1: Set up for Manual Calibrations/Checks in Base Configuration w/Gas Dilution Calibrator 155
Figure 10-2: Pneumatic Connections for T500U Precision Calibration when IZS Generator Present .. 159
Figure 10-3: Pneumatic Connections for Manual Calibration/Checks with the Internal Span Gas
Generator 160
LIST OF TABLES
LIST OF APPENDICES
APPENDIX A MENU TREES and RELATED SOFTWARE DOCUMENTATION
APPENDIX B - T700 SPARE PARTS LIST
APPENDIX C - REPAIR QUESTIONNAIRE
APPENDIX D - SCHEMATICS
07834D DCN6998 19
Introduction Teledyne API Model T500U CAPS NO2 Analyzer Manual
20 07834D DCN6998
2. SPECIFICATIONS, APPROVALS &
COMPLIANCE
2.1. SPECIFICATIONS
Table 2-1 below presents the T500U specifications.
Table 2-1: Specifications
Parameter Description
Range Min: 0-5 ppb Max: 0-1 ppm NO2 (User-selectable)
3 3
Measurement Units ppb, ppm, g/m , mg/m (User-selectable)
Zero Noise < 20 ppt
Span Noise <0.2% of reading + 20 ppt
Zero Drift <0.1 ppb / 24 hours
Span Drift <0.5% of reading / 24 hours
Lower Detectable Limit <40 ppt
Lag Time ~8 seconds (limited by volumetric flow rate)
Rise/ Fall Time <30 Seconds to 95%
Linearity < 1% Full Scale
Precision 0.5% of reading above 5 ppb
3
Sample Flow Rate 900 cm /min 10%
AC Power Rating Typical Power Consumption
110 - 120 V~ 60 Hz 3.0 A 80 W
220 - 240 V~ 50 Hz 3.0 A 80 W
Analog Output Ranges 10V, 5V, 1V, 0.1V (selectable)
Analog Output Resolution 1 part in 4096 of selected full-scale voltage
Communications
Standard I/O 1 Ethernet: 10/100Base-T
2 RS-232 (300 115,200 baud)
2 USB device ports
8 opto-isolated digital outputs
6 opto-isolated digital inputs (3 defined, 3 spare)
4 analog outputs
Optional I/O 1 USB com port
1 RS485
8 analog inputs (0-10V, 12-bit)
4 digital alarm outputs
Multidrop RS232
2 4-20mA current outputs
o
Operating Temperature 5-40 C
Humidity Range 0-95% RH, Non-Condensing
Dimensions HxWxD 7 x 17 x 23.5 (178 x 432 x 597 mm)
(19 rack mount, 5U, 24 deep)
Weight 33 lbs (15 kg)
Environmental Conditions Installation Category (Over Voltage Category ) II Pollution
Degree 2
Intended for Indoor Use Only at Altitudes 2000m
07834D DCN6998 21
Specifications, Approvals & Compliance Teledyne API Model T500U CAPS NO2 Analyzer Manual
1.1.1 SAFETY
rd
IEC/EN 61010-1:2010 (3 Edition), Safety requirements for electrical equipment
for measurement, control, and laboratory use.
1.1.2 EMC
IEC/EN 61326-1, Class A Emissions/Industrial Immunity
22 07834D DCN6998
3. GETTING STARTED
This section discusses unpacking, connecting, and initializing the instrument.
3.1. UNPACKING
The shipping crate contains:
instrument
power cord
CD-ROM of manuals (and/or paper manual if special-ordered)
Verify that there is no apparent external shipping damage to the unit by removing
the top cover of the instrument and inspecting all internal components. Pay
special attention to the pneumatic components prior to powering the instrument
ON. If damage has occurred, please advise the shipper first, then Teledyne API.
Included with your analyzer is a printed record of the final performance
characterization performed on your instrument at the factory, titled Final Test and
Validation Data Sheet (P/N 07853). It is an important quality assurance and
calibration record and should be placed in the quality records file for this
instrument.
With no power to the unit, carefully open the analyzer and check for internal
shipping damage:
1. Remove the two screws on either side of the chassis toward the back.
2. Slide the cover backward until it clears the analyzers front bezel.
3. Lift the cover straight up.
4. Check that all circuit boards and other components are in good shape
and properly seated.
5. Ensure that the connectors of the various internal wiring harnesses and
pneumatic hoses are firmly and properly seated.
6. Verify that all optional hardware ordered with the unit has been installed
(listed on the paperwork accompanying the analyzer).
7. Reattach the cover and remove the dust plugs at the rear of the analyzer
to the pneumatic ports, retain for use when storing/shipping the analyzer.
WARNING
ELECTRICAL SHOCK HAZARD
Never disconnect any electronic assemblies or subassemblies
while the instrument is under power.
07834D DCN6998 23
Getting Started Teledyne API Model T500U CAPS NO2 Analyzer Manual
24 07834D DCN6998
Teledyne API Model T500U CAPS NO2 Analyzer Manual Getting Started
07834D DCN6998 25
Getting Started Teledyne API Model T500U CAPS NO2 Analyzer Manual
Field Description/Function
Status LEDs indicating the states of Sample, Calibration and Fault, as follows:
Name Color State Definition
Off Unit is not operating in sample mode, DAS is disabled.
On Sample Mode active; Front Panel Display being updated; DAS data
SAMPLE Green being stored.
Blinking Unit is operating in sample mode, front panel display being updated,
DAS hold-off mode is ON, DAS disabled
Off Auto Cal disabled
CAL Yellow On Auto Cal enabled
Blinking Unit is in calibration mode
Off No warnings exist
FAULT Red
Blinking Warnings exist
Displays the actual concentration of the sample gas currently being measured by the analyzer in the
Conc
currently selected units of measure.
Mode Displays the name of the analyzers current operating mode
Displays a variety of informational messages such as warning messages, operational data, test function
Param
values and response messages during interactive tasks.
Control Buttons Displays dynamic, context sensitive labels on each button, which is blank when inactive until applicable.
26 07834D DCN6998
Teledyne API Model T500U CAPS NO2 Analyzer Manual Getting Started
Component Function
cooling fan Pulls ambient air into chassis through side vents and exhausts through rear.
Connector for three-prong cord to apply AC power to the analyzer.
AC power
CAUTION! The cords power specifications (specs) MUST comply with the power
connector
specs on the analyzers rear panel Model label
Model/specs label Identifies the analyzer model number and provides power specs
Connect a gas line from the source of sample gas here.
SAMPLE Calibration gases can also enter here on units without zero/span/shutoff valve options
installed.
Connect an exhaust gas line of not more than 10 meters long here that leads outside the
EXHAUST shelter or immediate area surrounding the instrument. The line must be tubing or
greater.
On units with zero/span valve option installed, connect a gas line to the source of
SPAN
calibrated span gas here.
On units with zero/span valve option installed, but no zero air scrubber, attach a gas line
ZERO AIR to the source of zero air here. If a permeation oven, also known as internal zero/span
valve (IZS), option is installed attach the zero air scrubber here.
DRY AIR Not Used
RX TX LEDs indicate receive (RX) and transmit (TX) activity when blinking.
COM 2 Serial communications port for RS-232 or RS-485 (RS-485.
RS-232 Serial communications port for RS-232 only.
Switch to select either data terminal equipment or data communication equipment during
DCE DTE
RS-232 communication.
STATUS For outputs to devices such as Programmable Logic Controllers (PLCs).
ANALOG OUT (AOUT) For voltage or current loop outputs to a strip chart recorder and/or a data logger.
CONTROL IN For remotely activating the zero and span calibration modes.
ALARM Option for concentration alarms and system warnings.
ETHERNET Connector for network or Internet remote communication, using Ethernet cable
ANALOG IN (AIN) Option for receiving and logging voltage signals from other instrumentation.
USB Connector for direct connection to laptop computer, using USB cable.
Model Label Includes voltage and frequency specifications
07834D DCN6998 27
Getting Started Teledyne API Model T500U CAPS NO2 Analyzer Manual
28 07834D DCN6998
Teledyne API Model T500U CAPS NO2 Analyzer Manual Getting Started
WARNING
ELECTRICAL SHOCK HAZARD
High Voltages are present inside the analyzers case.
Power connection must have functioning ground connection.
Do not defeat the ground wire on power plug.
Turn off analyzer power before disconnecting or connecting
electrical subassemblies.
Do not operate with cover off.
CAUTION
GENERAL SAFETY HAZARD
To avoid damage to your analyzer, ensure that the AC power voltage
matches the voltage indicated on the analyzers model/specs label
located on the rear panel before plugging the T500U into line power.
07834D DCN6998 29
Getting Started Teledyne API Model T500U CAPS NO2 Analyzer Manual
30 07834D DCN6998
Teledyne API Model T500U CAPS NO2 Analyzer Manual Getting Started
To access these signals attach a strip chart recorder and/or data-logger to the
appropriate analog output connections on the rear panel of the analyzer. Pin-outs
for the analog output connector are:
ANALOG OUT
A1 A2 A3 A4
+ - + - + - + -
To change the settings for the analog output channels, see Section 6.8.2.
07834D DCN6998 31
Getting Started Teledyne API Model T500U CAPS NO2 Analyzer Manual
32 07834D DCN6998
Teledyne API Model T500U CAPS NO2 Analyzer Manual Getting Started
07834D DCN6998 33
Getting Started Teledyne API Model T500U CAPS NO2 Analyzer Manual
The status outputs are accessed via a rear panel 12-pin connector labeled
STATUS (Figure 3-8). Pin-outs for this connector are:
STATUS
1 2 3 4 5 6 7 8 D +
ZERO CAL
DIAG MODE
SYSTEM OK
SPAN CAL
TEMP ALARM
HIGH RANGE
CONC VALID
PRESS ALARM
OK
+5V to external device
D Emitter BUS The emitters of the transistors on pins 1 to 8 are bussed together.
SPARE
+ DC Power + 5 VDC, 300 mA source maximum
The ground level from the analyzers internal DC power supplies. This
Digital Ground
connection should be used as the ground return when +5VDC power is used.
34 07834D DCN6998
Teledyne API Model T500U CAPS NO2 Analyzer Manual Getting Started
CONTROL IN CONTROL IN
A B C D E F U + A B C D E F U +
ZERO
SPAN
SPARE
ZERO
SPAN
SPARE
RANGE HIGH
RANGE HIGH
- 5 VDC Power +
Supply
The ground level from the analyzers internal DC Power Supplies (same as chassis
Digital Ground
ground).
External Power
U Input pin for +5 VDC required to activate pins A F.
input
Internally generated 5V DC power. To activate inputs A F, place a jumper
+ 5 VDC output between this pin and the U pin. The maximum amperage through this port is 300
mA (combined with the analog output supply, if used).
07834D DCN6998 35
Getting Started Teledyne API Model T500U CAPS NO2 Analyzer Manual
Alarm 1 System OK 2
Alarm 2 Conc 1
Alarm 3 Conc 2
Alarm 4 Range Bit
Alarm 1 Relay
Alarm 1, which is System OK 2 (system OK 1, is the status bit), is in the
energized state when the instrument is OK and there are no warnings. If there
is a warning active or if the instrument is put into the DIAG mode, Alarm 1
will change states. This alarm has reverse logic meaning that if you put a meter
across the Common & Normally Closed pins on the connector you will find that
it is OPEN when the instrument is OK. This is so that if the instrument should
turn off or lose power, it will change states and you can record this with a data
logger or other recording device.
The Alarm Relay activates any time the concentration set-point is exceeded, and
will return to its normal state when the concentration value returns below the
concentration set-point.
Even though the relay on the rear panel is a NON-Latching alarm & resets when
the concentration falls below the alarm set point, the warning on the front panel
display will remain latched until it is cleared: either push CLR on the front panel
menu or clear the warning through the serial port.
36 07834D DCN6998
Teledyne API Model T500U CAPS NO2 Analyzer Manual Getting Started
Alarm 4 Relay
This relay is connected to the range bit. If the instrument is configured for
Auto Range and the instrument moves into the high range, it activates this
relay.
Ethernet Connection
For network or Internet communication with the analyzer, connect an Ethernet
cable from the analyzers rear panel ETHERNET interface connector to an
Ethernet port. Although the analyzer is shipped with DHCP enabled by default
(Section 7.5.2), it should be manually assigned a static IP address.
Configuration: (manual, i.e., static) Section 7.5.1.
USB Connection
The USB option can be used for direct communication between the analyzer and
a PC; connect a USB cable between the analyzer and computer USB ports. This
USB connection can only be used when the COM2 port is not in use except for
RS-232 Multidrop communication.
Configuration: Section 7.6.
Note If this option is installed, the rear panel COM2 port cannot be used
for anything other than Multidrop communication.
RS-232 Connection
For RS-232 communications with data terminal equipment (DTE) or with data
communication equipment (DCE) connect either a DB9-female-to-DB9-female
cable (Teledyne API part number WR000077) or a DB9-female-to-DB25-male
cable (Option 60A), as applicable, from the analyzers rear panel RS-232 port to
the device. Adjust the rear panel DCE-DTE switch to select DTE or DCE as
appropriate (Section 7.1).
Configuration: Section 7.3
07834D DCN6998 37
Getting Started Teledyne API Model T500U CAPS NO2 Analyzer Manual
Figure 3-12: Pin Assignments for J11, J12 Connectors on CPU Board.
Teledyne API offers either of two mating cables, to select for your use:
P/N WR000077, a DB-9 female to DB-9 female cable, 6 feet long. Allows
connection of the serial ports of most personal computers.
P/N WR000024, a DB-9 female to DB-25 male cable. Allows connection to
the most common styles of modems (e.g. Hayes-compatible) and code
activated switches.
38 07834D DCN6998
Teledyne API Model T500U CAPS NO2 Analyzer Manual Getting Started
Both cables are configured with straight-through wiring and should require no
additional adapters.
To assist in properly connecting the serial ports to either a computer or a modem,
there are activity indicators just above the RS-232 port. Once a cable is
connected between the analyzer and a computer or modem, both the red and
green LEDs should be on.
If the lights are not lit, locate the small switch on the rear panel to switch it
between DTE and DCE modes.
If both LEDs are still not illuminated, ensure that the cable properly
constructed.
Received from the factory, the analyzer is set up to emulate an RS-232 DCE
device.
RS-232 (COM1): RS-232 (fixed) DB-9 male connector
Baud rate: 115200 bits per second (baud)
Data Bits: 8 data bits with 1 stop bit
Parity: None
Note Because the RS-232 Multidrop option uses both the RS232 and
COM2 DB9 connectors on the analyzers rear panel to connect the
chain of instruments, COM2 port is no longer available for separate
RS-232 or RS-485 operation.
07834D DCN6998 39
Getting Started Teledyne API Model T500U CAPS NO2 Analyzer Manual
In each instrument with the Multidrop option there is a shunt that jumpers two
pins on the serial Multidrop and LVDS printed circuit assembly (PCA), as shown
in Figure 3-13. This shunt must be removed from all instruments except that
designated as last in the multidrop chain, which must remain terminated. This
requires powering off and opening each instrument and making the following
adjustments:
1. With NO power to the instrument, remove its top cover and lay the rear
panel open for access to the Multidrop/LVDS PCA, which is seated on the
CPU.
2. On the Multidrop/LVDS PCAs JP2 connector, remove the shunt that jumpers
Pins 21 22 as indicated in Figure 3-13. (Do this for all but the last
instrument in the chain where the shunt should remain at Pins 21 22).
3. Check that the following cable connections are made in all instruments
(again refer to Figure 3-13):
J3 on the Multidrop/LVDS PCA to the CPUs COM1 connector
(Note that the CPUs COM2 connector is not used in Multidrop)
J4 on the Multidrop/LVDS PCA to J12 on the motherboard
J1 on the Multidrop/LVDS PCS to the front panel LCD
40 07834D DCN6998
Teledyne API Model T500U CAPS NO2 Analyzer Manual Getting Started
Ensure jumper is
installed between
JP2 pins 21 22 in
last instrument of
multidrop chain.
7. BEFORE communicating from the host, power on the instruments and check
that the Machine ID code is unique for each (Section 6.6.1).
a. In the SETUP Mode menu go to SETUP>MORE>COMM>ID. The default
ID is typically the model number or 0.
b. to change the identification number, press the button below the digit to
be changed.
c. Press/select ENTER to accept the new ID for that instrument.
8. Next, in the SETUP>MORE>COMM>COM1 menu (do not use the COM2
menu for multidrop), edit the COM1 MODE parameter as follows:
press/select EDIT and set only QUIET MODE, COMPUTER MODE, and
MULTIDROP MODE to ON. Do not change any other settings.
9. Press/select ENTER to accept the changed settings, and ensure that COM1
MODE now shows 35.
07834D DCN6998 41
Getting Started Teledyne API Model T500U CAPS NO2 Analyzer Manual
10. Press/select SET> to go to the COM1 BAUD RATE menu and ensure it
reads the same for all instruments (edit as needed so that all instruments are
set at the same baud rate).
Note The (communication) Host instrument can address only one instrument at
a time, each by its unique ID (see step 7 above).
Note Teledyne API recommends setting up the first link, between the Host and
the first analyzer, and testing it before setting up the rest of the chain.
CAUTION
GENERAL SAFETY HAZARD
Do not vent calibration gas, exhaust gas or sample gas into enclosed areas.
42 07834D DCN6998
Teledyne API Model T500U CAPS NO2 Analyzer Manual Getting Started
CAUTION
NEVER operate the analyzer with the gas ports capped at the rear
of the analyzer: this can damage the internal pump.
NEVER operate the analyzer with the internal pneumatics
disconnected: this will allow contaminants to enter the bench.
07834D DCN6998 43
Getting Started Teledyne API Model T500U CAPS NO2 Analyzer Manual
Note Zero air and span gases must be supplied at twice the instruments
specified gas flow rate. Therefore, the T500U zero and span gases
should be supplied to their respective inlets in excess of 1800
cm3/min (~900 cm3/min x 2).
Zero Air
Zero air or zero calibration gas is defined as a gas that is similar in chemical
composition to the measured medium but without the gas to be measured by the
analyzer.
For the T500U this means zero air should be devoid of NO2 and H2O vapor.
If your application is not a measurement in ambient air, the zero calibration
gas should be matched to the composition of the gas being measured.
Pure nitrogen (N2) could be used as a zero gas for applications where NO2 is
measured in nitrogen.
Note TAPI recommends the use of NO2 permeation tubes for performing
routine calibration checks only, but not for NO2 span calibration.
If a dynamic dilution system such as the Teledyne API Model T700 is used to
dilute high concentration gas standards to low, ambient concentrations, ensure
that the NO2 concentration of the reference gas matches the dilution range of the
calibrator.
Choose the NO2 gas concentration so that the dynamic dilution system operates in
its mid-range and not at the extremes of its dilution capabilities.
EXAMPLE:
A dilution calibrator with 10-1000 dilution ratio will not be able to accurately
dilute a 5000 ppm NO2 gas to a final concentration of 500 ppb, as this would
operate at the very extreme dilution setting.
A 50 ppm NO2 gas in nitrogen is much more suitable to calibrate the T500U
analyzer (dilution ratio of 200, in the mid-range of the systems capabilities).
3.3.2.2. BASIC CONNECTIONS FROM CALIBRATOR, WITH AND WITHOUT SPAN GAS
Source of
VENT here if input
SAMPLE GAS
is pressurized Removed during
calibration
MODEL T700
Calibrated NO2
at HIGH Span
Concentration
Gas Dilution
Enclosure Wall
Calibrator
SAMPLE
MODEL 701
EXHAUST
Zero Gas
Generator Chassis
VENT
Figure 3-15: Gas Line Connections from Calibrator Basic T500U Configuration
For the T500U analyzer in its basic configuration, attach the following pneumatic
lines:
07834D DCN6998 45
Getting Started Teledyne API Model T500U CAPS NO2 Analyzer Manual
Venting
In order to prevent back diffusion and pressure effects, both the span gas and zero
air supply lines should be:
Vented outside the shelter.
Minimum OD .
No less than 2 meters in length.
No greater than 10 meters in length.
Exhaust Outlet
Attach an exhaust line to the EXHAUST outlet fitting. The exhaust line should
be:
Use PTFE tubing; minimum OD .
No greater than 10 meters in length.
Vented outside the shelter.
46 07834D DCN6998
Teledyne API Model T500U CAPS NO2 Analyzer Manual Getting Started
Pneumatic Layouts
07834D DCN6998 47
Getting Started Teledyne API Model T500U CAPS NO2 Analyzer Manual
Figure 3-17: T500U Pneumatics with Internal Zero Span Valves Option
48 07834D DCN6998
4. OVERVIEW OF OPERATING MODES
To assist in navigating the analyzers software, a series of menu trees is available
for reference in Appendix A of this manual.
Note Some control buttons on the touch screen do not appear if they are
not applicable to the operation at hand or if an invalid setting was
input.
The T500U analyzer software has several operating modes, the most commonly
used of which is the SAMPLE mode. In this mode, a continuous read-out of the
NO2 concentration can be viewed on the front panel and output as an analog
voltage from rear panel terminals.
The second most commonly used operating mode is SETUP, which is used to
configure the various subsystems, such as the Data Acquisition System (DAS),
the reporting ranges, or the communication channels. SETUP mode is also used
for performing various diagnostic tests during troubleshooting.
Figure 4-1 shows in the Mode field that the unit is operating in SAMPLE Mode.
07834D DCN6998 49
Overview of Operating Modes Teledyne API Model T500U CAPS NO2 Analyzer Manual
50 07834D DCN6998
Teledyne API Model T500U CAPS NO2 Analyzer Manual Overview of Operating Modes
Pressing the CAL button, switches the analyzer into calibration mode. In this
mode the user can, in conjunction with introducing zero or span gases of known
concentrations into the analyzer cause it to adjust and recalculate the slope (gain)
and offset of its measurement range.
Note This mode is also used to check the current calibration status of the
instrument.
If the instrument includes one of the available zero/span valve options, the
SAMPLE mode display will also include CALZ and CALS buttons. Pressing
either of these buttons also puts the instrument into calibration mode.
The CALZ button is used to initiate a calibration of the analyzers zero point
using zero air.
The CALS button is used to calibrate the span point of the analyzers current
reporting range using span gas.
07834D DCN6998 51
Overview of Operating Modes Teledyne API Model T500U CAPS NO2 Analyzer Manual
52 07834D DCN6998
Teledyne API Model T500U CAPS NO2 Analyzer Manual Overview of Operating Modes
07834D DCN6998 53
Overview of Operating Modes Teledyne API Model T500U CAPS NO2 Analyzer Manual
54 07834D DCN6998
5. STARTUP, FUNCTIONAL CHECKS, AND
INITIAL CALIBRATION
5.1. STARTUP
For accurate results, allow the monitor to run for at least 60 minutes to reach a
stable operating temperature.
07834D DCN6998 55
Startup, Functional Checks, and Initial CalibrationTeledyne API Model T500U CAPS NO2 Analyzer Manual
56 07834D DCN6998
Teledyne API Model T500U CAPS NO2 Analyzer ManualStartup, Functional Checks, and Initial Calibration
07834D DCN6998 57
Startup, Functional Checks, and Initial CalibrationTeledyne API Model T500U CAPS NO2 Analyzer Manual
The method for performing an initial calibration for the T500U nitrogen dioxide
analyzer differs slightly depending on whether valve options are installed.
See Section 5.3.1 for instructions for initial calibration of the T500U analyzers
in their base configuration.
See Sections 10.4 and 10.4.6 for information regarding setup and calibration
of T500U analyzers with Z/S Valve options.
In either case, any time an initial calibration is in order, e.g, starting the analyzer
for the first time or after general maintenance, conducting a manual AREF may
be required (Auto Reference: reference loss measurement). However, note that
the instrument is programmed to automatically run an AREF 20 minutes after
power up. This expedites the AREF_L value to match current measurements. The
AREF_L value is based on the average of the last four readings taken one hour
apart. Repeat the process if large changes in the AREF_L value occur. Both the
AREF_L and MEAS_L readings should be similar while sampling zero air.
To run a manual AREF, press SETUP>MORE>AREF>RUN:
RUN EXIT
58 07834D DCN6998
Teledyne API Model T500U CAPS NO2 Analyzer ManualStartup, Functional Checks, and Initial Calibration
While these are the default settings for the T500U analyzer, it is recommended
that you verify them before proceeding with the calibration procedure, by
pressing the following menu button sequence:
07834D DCN6998 59
Startup, Functional Checks, and Initial CalibrationTeledyne API Model T500U CAPS NO2 Analyzer Manual
60 07834D DCN6998
Teledyne API Model T500U CAPS NO2 Analyzer ManualStartup, Functional Checks, and Initial Calibration
07834D DCN6998 61
Startup, Functional Checks, and Initial CalibrationTeledyne API Model T500U CAPS NO2 Analyzer Manual
Note Once you have completed the above set-up procedures, please fill
out the Quality Questionnaire that was shipped with your unit and
return it to Teledyne API. This information is vital to our efforts in
continuously improving our service and our products for you.
THANK YOU.
62 07834D DCN6998
6. SETUP
The SETUP menu provides access to the instrument parameter settings for
performing configuration, calibration, reporting, and diagnostics operations.
Press NEXT of PREV to move back and forth through SETUP T500U NO2 Analyzer
Press exit at any time
the following list of Configuration information:
PREV NEXT EXIT to return to the SETUP
MODEL NAME menu
PART NUMBER
SERIAL NUMBER
SOFTWARE REVISION
LIBRARY REVISION
BENCH FIRMWARE
CPU TYPE & OS REVISION
07834D DCN6998 63
Setup Teledyne API Model T500U CAPS NO2 Analyzer Manual
A1 A2 A3 A4
+ - + - + - + -
64 07834D DCN6998
Teledyne API Model T500U CAPS NO2 Analyzer Manual Setup
6.3.3.1. SETUP > RNGE > MODE > SNGL: SINGLE RANGE MODE CONFIGURATION
Note Single Range is the default reporting range mode for the analyzer.
When the single range mode is selected (SNGL), all analog concentration outputs
(A1, A2) are slaved together and set to the same reporting range limits (e.g. 500.0
ppb).
Although both outputs share the same concentration reporting range, the
electronic signal ranges of the analog outputs may still be configured for different
values (e.g. 0-5 VDC, 0-10 VDC, etc.; see Section 6.8.3.1).
07834D DCN6998 65
Setup Teledyne API Model T500U CAPS NO2 Analyzer Manual
To select SNGL range mode and to set the upper limit of the range, press:
6.3.3.2. SETUP > RNGE > MODE > DUAL: DUAL RANGE MODE CONFIGURATION
Selecting Dual Range mode allows the A1 and A2 outputs to be configured with
different reporting ranges. The analyzer software calls these two ranges low and
high. The low range setting corresponds with the analog output labeled A1 on the
rear panel of the instrument. The high range setting corresponds with the A2
output. While the software names these two ranges low and high, they do not
have to be configured that way. For example: the low range can be set for a span
of 0-150 ppb while the high range is set for 0-50 ppb.
In DUAL range mode the RANGE test function displayed on the front panel will
be replaced by two separate functions:
Range1: The range setting for the A1 output.
Range2: The range setting for the A2 output.
66 07834D DCN6998
Teledyne API Model T500U CAPS NO2 Analyzer Manual Setup
07834D DCN6998 67
Setup Teledyne API Model T500U CAPS NO2 Analyzer Manual
6.3.3.3. SETUP > RNGE > MODE > AUTO: AUTO RANGE MODE CONFIGURATION
In AUTO range mode, the analyzer automatically switches the reporting range
between two user-defined ranges (LOW and HIGH).
The unit will switch from LOW range to HIGH range when the NO2
concentration exceeds 98% of the low range span.
The unit will return from HIGH range back to LOW range once the NO2
concentration falls below 75% of the low range span.
Also the RANGE test function displayed on the front panel will be replaced by
two separate functions:
RANGE1: The LOW range setting for all analog outputs.
RANGE2: The HIGH range setting for all analog outputs.
The LOW/HIGH range status is also reported through the external, digital status
bits (Section 3.3.1.4). To set individual ranges press the following menu
sequence.
SAMPLE RANGE=500.0 PPB NO2= XXXX
The LOW and HIGH SNGL DUAL AUTO REM ENTR EXIT
ranges have separate
slopes and offsets for
computing the NO2 SETUP X.X RANGE MODE MENU
concentration. MODE SET UNIT EXIT
68 07834D DCN6998
Teledyne API Model T500U CAPS NO2 Analyzer Manual Setup
6.3.3.4. SETUP > RNGE > UNIT: SETTING THE REPORTING RANGE UNITS OF MEASURE
The T500U can display and report concentrations in ppb, ppm, ug/m3, mg/m3
units. Changing units affects all of the COM port values, and all of the display
values for all reporting ranges. To change the units of measure press:
NO2
ppb x 2.05 = g/m3
SETUP X.X CONC UNITS:PPB
ppm x 2.05 = mg/m3
PPB PPM UGM MGM ENTR EXIT
6.3.3.5. SETUP > RNGE > DIL: USING THE OPTIONAL DILUTION RATIO FEATURE
The dilution ratio feature is a software utility option designed for applications
where the sample gas is diluted before being analyzed by the T500U. Typically
this occurs in continuous emission monitoring (CEM) applications where the
quality of gas in a smoke stack is being tested and the sampling method used to
remove the gas from the stack dilutes the gas. Once the degree of dilution is
07834D DCN6998 69
Setup Teledyne API Model T500U CAPS NO2 Analyzer Manual
known, this feature allows the user to add an appropriate scaling factor to the
analyzers NO2 concentration calculations so that the measurement range and
concentration values displayed on the instruments front panel display and
reported via the instruments various outputs reflect the undiluted values.
Using the Dilution Ratio option is a 4-step process:
1. Select the appropriate units of measure (see Section 6.3.3.4).
2. Select the reporting range mode and set the reporting range upper limit (see
Section 6.3.3).
Ensure that the upper span limit entered for the reporting range is the
maximum expected concentration of the UNDILUTED gas.
3. Set the dilution factor as a gain (e.g., a value of 20 means 20 parts diluent
and 1 part of sample gas):
Ensure that the calibration span gas is either supplied through the same
dilution system as the sample gas or has an appropriately lower actual
concentration.
EXAMPLE: If the reporting range limit is set for 1 ppm and the dilution
ratio of the sample gas is 20 gain, either:
a span gas with the concentration of 1 ppm can be used if the span gas
passes through the same dilution steps as the sample gas, or;
a 0.05 ppm span gas must be used if the span gas IS NOT routed
through the dilution system.
70 07834D DCN6998
Teledyne API Model T500U CAPS NO2 Analyzer Manual Setup
07834D DCN6998 71
Setup Teledyne API Model T500U CAPS NO2 Analyzer Manual
72 07834D DCN6998
Teledyne API Model T500U CAPS NO2 Analyzer Manual Setup
Note The instrument still prompts for a password when entering the VARS
and DIAG menus, even if passwords are disabled, but it displays the
default password (818) upon entering these menus. In this case, only
press ENTR to access the password-protected menus.
To disable the PASSWORD feature after it has been turned ON, the SETUP
menu first requires a password; once the password has been input and the ENTR
button pressed, the PRIMARY SETUP MENU appears, and now the PASS menu
can be accessed, where pressing the ON button turns PASSWORD ENABLE
back to OFF, and pressing the ENTR button accepts the change (Table 6-1).
07834D DCN6998 73
Setup Teledyne API Model T500U CAPS NO2 Analyzer Manual
Time is shown in
24-hr cycle, e.g., SETUP X.X TIME-OF-DAY CLOCK
22:30 = 10:30pm.
EXIT returns to
TIME DATE EXIT SAMPLE
MODE.
74 07834D DCN6998
Teledyne API Model T500U CAPS NO2 Analyzer Manual Setup
8 1 8 ENTR EXIT
PREV NEXT JUMP EDIT ENTR EXIT EXIT discards the new
setting.
07834D DCN6998 75
Setup Teledyne API Model T500U CAPS NO2 Analyzer Manual
The default ID is either 0 or the same as the model number, e.g., 0500 for the
Model T500U. Press any button(s) in the MACHINE ID menu (Figure 6-3) until
the Machine ID in the Parameter field displays the desired identifier.
76 07834D DCN6998
Teledyne API Model T500U CAPS NO2 Analyzer Manual Setup
Toggle to cycle
through the available
character set: 0-9 ENTR accepts the new
SETUP X. MACHINE ID: 500 ID
settings
EXIT ignores the new
0 5 0 0 ENTR EXIT settings
The ID can be any unique 4-digit number and can also be used to identify
analyzers in any number of ways (e.g. location numbers, company asset number,
etc.)
Configuring COM1 or COM2 requires setting the DCE DTE switch on the rear
panel. Section 7.1 provides DCE DTE information.
07834D DCN6998 77
Setup Teledyne API Model T500U CAPS NO2 Analyzer Manual
Note There is a 2-sec latency period between the time a VARS value is
changed and its storage into the analyzers memory. DO NOT turn the
analyzer off during this period or the new setting will be lost.
78 07834D DCN6998
Teledyne API Model T500U CAPS NO2 Analyzer Manual Setup
To access and navigate the VARS menu, use the following button sequence:
SAMPLE RANGE=500.0 PPB NO2= XXXX
8 1 8 ENTR EXIT
Toggle these
buttons to enter
the correct
PASSWORD. SETUP X.X 0) DAS_HOLD_OFF=15.0 Minutes
Press EDIT to change settings
PREV NEXT JUMP EDIT PRNT EXIT for any of the VARS
07834D DCN6998 79
Setup Teledyne API Model T500U CAPS NO2 Analyzer Manual
NO2 LED Allows the user to turn off the LED without cycling
6.8.5
DISABLE instrument power.
80 07834D DCN6998
Teledyne API Model T500U CAPS NO2 Analyzer Manual Setup
8 1 8 ENTR EXIT
07834D DCN6998 81
Setup Teledyne API Model T500U CAPS NO2 Analyzer Manual
Access the signal I/O test mode from the DIAG Menu (Figure 6-4), then press:
DIAG SIGNAL I / O Press NEXT & PREV to
move between signal
types.
PREV NEXT JUMP ENTR EXIT
82 07834D DCN6998
Teledyne API Model T500U CAPS NO2 Analyzer Manual Setup
Initiates a calibration of the A1, A2, and A4 analog output channels that
determines the slope and offset inherent in the circuitry of each output.
AOUT CALIBRATED 6.8.3.1
These values are stored and applied to the output signals by the CPU
automatically.
Sets the basic electronic configuration of the A1 output (NO2 Concentration).
There are four options:
RANGE: Selects the signal type (voltage or current loop) and level of the
output
CONC_OUT_1 REC OFS: Allows them input of a DC offset to let the user manually adjust
the output level 6.8.2
AUTO CAL: Enables / Disables the AOUT CALIBRATION Feature
CALIBRATED: Performs the same calibration as AOUT CALIBRATED, but
on this one channel only.
CONC_OUT_2 Same as for CONC_OUT_1 but for analog channel A2 (NO Concentration)
TEST OUTPUT Same as for CONC_OUT_1 but for analog channel A4 (TEST CHANNEL) 6.8.4
AIN CALIBRATED Initiates a calibration of the A-to-D Converter circuit located on the Motherboard. 6.8.3.10
XIN1
. For each of 8 external analog inputs channels, shows the gain, offset,
. engineering units, and whether the channel is to show up as a Test 6.8.3.11
. function.
XIN8
07834D DCN6998 83
Setup Teledyne API Model T500U CAPS NO2 Analyzer Manual
8 1 8 ENTR EXIT
Toggle these
buttons to enter
the correct
PASSWORD. DIAG SIGNAL I/O
84 07834D DCN6998
Teledyne API Model T500U CAPS NO2 Analyzer Manual Setup
To change the output type and range, select the ANALOG I/O
CONFIGURATION submenu (see Figure 6-5) then press:
From the
AIO CONFIGURATION SUBMENU
07834D DCN6998 85
Setup Teledyne API Model T500U CAPS NO2 Analyzer Manual
86 07834D DCN6998
Teledyne API Model T500U CAPS NO2 Analyzer Manual Setup
From the
AIO CONFIGURATION SUBMENU
NOTE:
ANALOG OUTPUTS Continue pressing SET> until you reach the
configured for 0.1V full output to be configured
scale should always be
calibrated manually.
DIAG AIO CONC_OUT_2: 5V, OVR, NOCAL
07834D DCN6998 87
Setup Teledyne API Model T500U CAPS NO2 Analyzer Manual
88 07834D DCN6998
Teledyne API Model T500U CAPS NO2 Analyzer Manual Setup
From the
AIO CONFIGURATION SUBMENU
07834D DCN6998 89
Setup Teledyne API Model T500U CAPS NO2 Analyzer Manual
Note The menu for manually adjusting the analog output signal level will
only appear if the AUTO-CAL feature is turned off for the channel
being adjusted. (See Section 6.8.3.3).
Volt
Meter
+DC Gnd
V OUT + V IN +
V OUT - V IN -
Recording
ANALYZER
Device
Figure 6-6: Setup for Checking / Calibrating DCV Analog Output Signal Levels
90 07834D DCN6998
Teledyne API Model T500U CAPS NO2 Analyzer Manual Setup
To adjust the signal levels of an analog output channel manually, select the
ANALOG I/O CONFIGURATION submenu (see Figure 6-5) then press:
From the
AIO CONFIGURATION SUBMENU
DISPLAYED AS = CHANNEL
CONC_OUT_1 = A1 Continue pressing SET> until you reach the
CONC_OUT_2 = A2 output to be configured
TEST OUTPUT = A4
07834D DCN6998 91
Setup Teledyne API Model T500U CAPS NO2 Analyzer Manual
mADC
Current
Meter
IN OUT
I OUT + I IN +
I OUT - I IN -
Recording
ANALYZER
Device
Figure 6-7: Setup for Checking / Calibration Current Output Signal Levels Using an Ammeter
92 07834D DCN6998
Teledyne API Model T500U CAPS NO2 Analyzer Manual Setup
To adjust the zero and span signal levels of the current outputs, select the
ANALOG I/O CONFIGURATION submenu (see Figure 6-5) then press:
From the
AIO CONFIGURATION SUBMENU
DISPLAYED AS = CHANNEL
CONC_OUT_1 = A1 Continue pressing SET> until you reach the
CONC_OUT_2 = A2 output to be configured
TEST OUTPUT = A4
07834D DCN6998 93
Setup Teledyne API Model T500U CAPS NO2 Analyzer Manual
An alternate method for measuring the output of the Current Loop converter is to
connect a 250 ohm 1% resistor across the current loop output in lieu of the
current meter (see Figure 3-6 for pin assignments and diagram of the analog
output connector). This allows the use of a voltmeter connected across the
resistor to measure converter output as VDC or mVDC.
Volt
Meter
+DC Gnd
V OUT + V IN +
250
V OUT - V IN -
Recording
ANALYZER
Device
Figure 6-8: Alternative Setup Using 250 Resistor for Checking Current Output Signal Levels
In this case, follow the procedure above but adjust the output for the following
values:
94 07834D DCN6998
Teledyne API Model T500U CAPS NO2 Analyzer Manual Setup
From the
AIO CONFIGURATION SUBMENU
07834D DCN6998 95
Setup Teledyne API Model T500U CAPS NO2 Analyzer Manual
0 0 1 0 ENTR EXIT
EXAMPLE
96 07834D DCN6998
Teledyne API Model T500U CAPS NO2 Analyzer Manual Setup
DIAG AIO A/D CALIBRATION ERROR DIAG AIO AIN CALIBRATED: YES
EXIT EXIT
Perform
Troubleshooting or call DIAG AIO AIN CALIBRATED:NO
Teledyne APIs
Technical Support <SET CAL EXIT
Dept.
07834D DCN6998 97
Setup Teledyne API Model T500U CAPS NO2 Analyzer Manual
These parameters can also be captured in the internal Data Acquisition System
(DAS); refer to Appendix A for Analog-In DAS parameters.
To adjust settings for the Analog Inputs option parameters press:
DIAG ANALOG I / O CONFIGURATION
Press to change
< SET SET> EDIT EXIT Gain value
DIAG AIO XIN1 DISPLAY:OFF Pressing ENTR records the new setting
and returns to the previous menu.
Pressing EXIT ignores the new setting and
< SET EDIT EXIT returns to the previous menu.
98 07834D DCN6998
Teledyne API Model T500U CAPS NO2 Analyzer Manual Setup
Table 6-8: Test Channels Functions available on the T500Us Analog Output
Once a function is selected, the instrument not only begins to output a signal on
the analog output, but also adds TEST to the list of test functions viewable in the
front panel display.
07834D DCN6998 99
Setup Teledyne API Model T500U CAPS NO2 Analyzer Manual
8 1 8 ENTR EXIT
Toggle these
buttons to enter
the correct
PASSWORD. DIAG SIGNAL I/O
6.9. SETUP>MORE>AREF:
At 20 minutes after power up and periodically thereafter when in Sample mode,
the analyzer conducts a background measurement, known as an auto reference
(AREF), whereby the sample is routed through an internal charcoal scrubber.
This measurement accounts for drift in the baseline loss. A manual AREF is
recommended prior to initial calibration (Section 5.3).
Please note that AREF becomes disabled during calibration if the instrument
stays in calibration mode (CALZ or CALS) for longer than it would normally
take to run a calibration. To ensure that AREF is enabled, return to Sample mode
after conducting a calibration.
To allow the analyzer to be used with terminals (DTE), modems (DCE) and
computers (which can be either), a switch mounted below the serial ports on the
rear panel labeled DCE DTE (Figure 3-3) allows the user to set the RS-232
configuration for one of these two data devices. This switch exchanges the
Receive and Transmit lines on RS-232 emulating a cross-over or null-modem
cable. The switch has no effect on COM2.
Communication Modes for each COM port being used must be configured
independently. To turn on or off the communication modes for either COM1 or
COM2, access the SETUP>MORE>[COM1 OR COM2] menu, and at the COM1
[2] Mode menu press EDIT.
Combined Mode ID
SETUP COM1 MODE:0 displayed here.
<SET SET> EDIT EXIT
Test runs
SETUP X.X TRANSMITTING TO COM1
automatically.
7.3. RS-232
The RS232 and COM2 communications (COMM) ports operate on the RS-
232 protocol (default configuration). Possible configurations for these two
COM ports are summarized as follows:
RS232 port can also be configured to operate in single or RS-232 Multidrop
mode (Option 62); refer to Section 3.3.1.8.
COM2 port can be left in its default configuration for standard RS-232
operation including multidrop, or it can be reconfigured for half-duplex RS-
485 operation (please contact the factory for this configuration).
Note that the COM2 port and the USB port cannot be used simultaneously except
when COM2 is used for multidrop communication.
A code-activated switch (CAS) can also be used on either RS232 or COM2 port
to connect typically between 2 and 16 send/receive instruments (host computer(s)
printers, data loggers, analyzers, monitors, calibrators, etc.) into one
communications hub. Contact Teledyne API Sales for more information on CAS
systems.
To configure the analyzers communication ports, use the
SETUP>MORE>COMM menu (Section 6.6), and configure per Section 7.2
information.
7.5. ETHERNET
When using the Ethernet interface, the analyzer can be connected to any standard
10BaseT or 100BaseT Ethernet network via low-cost network hubs, switches or
routers. The interface operates as a standard TCP/IP device on port 3000. This
allows a remote computer to connect through the network to the analyzer using
APICOM, terminal emulators or other programs.
The Ethernet connector has two LEDs that are on the connector itself, indicating
its current operating status.
The analyzer is shipped with DHCP enabled by default. This allows the
instrument to be connected to a network or router with a DHCP server. The
instrument will automatically be assigned an IP address by the DHCP server
(Section 7.5.2). This configuration is useful for quickly getting an instrument up
and running on a network. However, for permanent Ethernet connections, a static
IP address should be used. Section 7.5.1 below details static IP address
configuration.
8 1 8 ENTR EXIT Accepts the new setting and returns to the previous
ENTR
menu.
Ignores the new setting and returns to the previous
DHCP: ON is EXIT
menu.
default setting. SETUP X.X DHCP: ON
Skip this step Some buttons appear only when relevant.
if it has been SET> EDIT EXIT
set to OFF.
3000 This number defines the terminal control port by which the
1
TCP PORT instrument is addressed by terminal emulation software, such as
Internet or Teledyne APIs APICOM.
[initially blank] The name by which your analyzer will appear when addressed
HOST NAME from other computers on the LAN or via the Internet. To assign
or change, see Section 7.5.2.1.
1
Do not change the setting for this property unless instructed to by Teledyne APIs Technical Support
personnel.
8 1 8 ENTR EXIT
ON ENTR EXIT
Note If the gateway IP, instrument IP and the subnet mask are all zeroes
(i.e., 0.0.0.0), the DCHP was not successful in which case you may
have to configure the analyzers Ethernet properties manually. See
your network administrator.
3. Connect the USB cable between a USB port on your personal computer to
the rear panel USB port of your analyzer. The USB cable should be a Type A
Type B cable, commonly used as a USB printer cable.
4. Determine the Windows XP Com Port number that was automatically
assigned to the USB connection. (Start>Control
Panel>System>Hardware>Device Manager). This is the com port that should
be set in the communications software, such as APICOM or Hyperterminal.
Refer to the Quick Start (Direct Cable Connection) section of the Teledyne
APICOM Manual, PN 07463.
Note Using the USB port disallows use of the rear panel COM2 port
except for multidrop communication.
7.7.1. HESSEN
The Hessen protocol is a multidrop protocol, in which several remote instruments
are connected via a common communications channel to a host computer. The
remote instruments are regarded as slaves of the host computer. The remote
instruments are unaware that they are connected to a multidrop bus and never
initiate Hessen protocol messages. They only respond to commands from the host
computer and only when they receive a command containing their own unique ID
number.
The Hessen protocol is designed to accomplish two things: to obtain the status of
remote instruments, including the concentrations of all the gases measured; and to
place remote instruments into zero or span calibration or measure mode.
Teledyne APIs implementation supports both of these principal features.
The Hessen protocol is not tightly defined; therefore, while Teledyne APIs
application is completely compatible with the protocol itself, it may be different
from implementations by other companies.
To change the baud rate of the T500Us COM ports, See Section 7.2.2.
To change the remaining COM port parameters listed in the table above, see
Section 7.2.1, Table 7-1.
Note Ensure that the communication parameters of the host computer are
also properly set.
Combined Mode ID
SETUP X.X COM1 MODE:0 displayed here.
<SET SET> EDIT EXIT
WHERE:
GAS TYPE The type of gas to be reported (the T500U has only one gas type:
NO2).
RANGE The concentration range for this gas. This feature permits the
user to select which concentration range will be used. The T500U
analyzer has two ranges: RANGE1 or LOW & RANGE2 or HIGH
(see Section 6.3).
1. 0 - The HESSEN protocol to use whatever range is currently
active.
2. 1 - The HESSEN protocol will always use RANGE1 for this
gas.
3. 2 - The HESSEN protocol will always use RANGE2 for this
gas.
4. 3 - Not applicable to the T500U analyzer.
NO2 213
MGM 01
PPB 10
PPM 11
UNASSIGNED FLAGS
2
MANIFOLD TEMPERATURE FRONT PANEL COMMUNICATION WARNING
BOX TEMP WARNING ANALOG CALIBRATION WARNING
SYSTEM RESET DYNAMIC ZERO WARNING
RELAY BOARD WARNING DYNAMIC SPAN WARNING
REAR BOARD NOT DETECTED IN MULTI-POINT CALIBRATION MODE
AUTOREF WARNING
1
Only applicable if the optional internal span gas generator is installed.
2
Only applicable if the T500U is equipped with the optional manifold.
It is possible to assign more than one flag to the same Hessen status bit. This allows the grouping of
similar flags, such as all temperature warnings, under the same status bit.
Be careful not to assign conflicting flags to the same bit as each status bit will be triggered if any of the
assigned flags is active.
7.7.2. MODBUS
This section provides MODBUS communications protocol setup instructions; see
Appendix A for MODBUS registers. This set of instructions assumes that the user
is familiar with MODBUS communications, and provides minimal information to
get started. For additional instruction, please refer to the Teledyne API MODBUS
manual, PN 06276, downloadable from our website at http://www.teledyne-
api.com/software/apicom/. Also refer to www.modbus.org for MODBUS
communication protocols.
Minimum Requirements
Instrument firmware with MODBUS capabilities installed.
MODBUS-compatible software (TAPI uses MODBUS Poll for testing; see
www.modbustools.com)
Personal computer
Communications cable (Ethernet or USB or RS232)
Possibly a null modem adapter or cable
Table 7-7 MODBUS Setup Instructions
Actions
Set Com Mode parameters
Comm Ethernet: Using the front panel menu, go to SETUP MORE COMM INET; scroll
through the INET submenu until you reach TCP PORT 2 (the standard
setting is 502), then continue to TCP PORT 2 MODBUS TCP/IP; press
EDIT and toggle the menu button to change the setting to ON, then press
ENTR. (Change Machine ID if needed: see Slave ID).
USB/RS232: Using the front panel menu, go to SETUP MORE COMM COM2
EDIT; scroll through the COM2 EDIT submenu until the display shows
COM2 MODBUS RTU: OFF (press OFF to change the setting to ON.
Scroll NEXT to COM2 MODBUS ASCII and ensure it is set to OFF. Press
ENTR to keep the new settings. (If RTU is not available with your
communications equipment, set the COM2 MODBUS ASCII setting to ON
and ensure that COM2 MODBUS RTU is set to OFF. Press ENTR to keep
Slave ID the new settings).
If your analyzer is connected to a network with at least one other analyzer of the same
model, a unique Slave ID must be assigned to each. Using the front panel menu, go to
SETUP MORE COMM ID. The MACHINE ID default is the same as the model
number. Toggle the menu buttons to change the ID.
Reboot analyzer For the settings to take effect, power down the analyzer, wait 5 seconds, and power up
again.
Make appropriate cable Connect your analyzer via either:
connections Ethernet or USB port to a PC (this may require a USB-to-RS232 adapter for your PC; if
so, also install the software driver from the CD supplied with the adapter, and reboot
the computer if required), or
COM2 port to a null modem (this may require a null modem adapter or cable).
A. In the Connection Setup window (see example that follows), select the options
based on your computer.
B. Press OK.
Read the Modbus Poll Use the Register Map to find the test parameter names for the values displayed (see
Register example that follows If desired, assign an alias for each.
Note The DAS can be disabled only by disabling or deleting its individual
data channels.
The principal use of the DAS is logging data for trend analysis and predictive
diagnostics, which can assist in identifying possible problems before they affect
the functionality of the analyzer. The secondary use is for data analysis,
documentation and archival in electronic format.
The green SAMPLE LED on the instrument front panel not only indicates the
analyzer status, but also indicates certain aspects of the DAS status:
OFF: System is in calibration mode. Data logging can be enabled or
disabled for this mode. Calibration data are typically stored at the
end of calibration periods, concentration data are typically not
sampled, diagnostic data should be collected.
BLINKING: Instrument is in hold-off mode, a short period after the system
exits calibrations. DAS channels can be enabled or disabled for
this period. Concentration data are typically disabled whereas
diagnostic should be collected.
These default data channels can be used as they are, or they can be customized
(Section 8.1.3) from the front panel to fit a specific application. They can also be
deleted to make room for custom user-programmed Data Channels.
When editing the data channels, the top line of the display indicates some of the
configuration parameters.
For example, the display line:
0) CONC: ATIMER, 1, 800
Translates to the following configuration:
Channel No.: 0
NAME: CONC
TRIGGER EVENT: ATIMER
PARAMETERS: Five parameters are included in this channel
EVENT: This channel is set up to store 800 records.
C O N C ENTR EXIT
EXIT discards the new
setting.
Note DAS does not keep track of the units (i.e., PPM or PPB) of each
concentration value. Therefore, DAS data files may contain
concentration data recorded in more than one type of unit if the units
of measure were changed during data acquisition
Each data parameter has user-configurable functions that define how the data are
recorded which are listed in Table 8-3:
Data channels can be edited individually from the front panel without affecting
other data channels. However, when editing a data channel, such as during
adding, deleting or editing parameters, all data for that particular channel will be
lost, because the DAS can store only data of one format (number of parameter
columns etc.) for any given channel. In addition, a DAS configuration can only
be uploaded remotely as an entire set of channels. Hence, remote update of the
DAS will always delete all current channels and stored data.
PREV NEXT INS DEL EDIT EXIT EXIT discards the new
setting.
1 ENTR EXIT
Toggle this button to
set from 1 to 4.
Note When the STOR NUM. SAMPLES feature is turned on, the instrument
will store the number of measurements that were used to compute
the AVG, SDEV, MIN or MAX value but not the actual measurements
themselves.
Note In AVG, SDEV, MIN or MAX sample modes (see Section 8.1.3.3), the
settings for the Sample Period and the Report Period determine the
number of data points used each time the parameters are calculated,
stored and reported to the COM ports.
The actual sample readings are not stored past the end of the chosen
report period.
When the STORE NUM SAMPLES feature is turned on, the instrument
will store the number of measurements that were used to compute
the AVG, SDEV, MIN or MAX value but not the actual measurements
themselves.
To define the REPORT PERIOD, follow the instruction shown in Section 8.1.3
then press:
Starting at the EDIT CHANNEL MENU
Use the PREV and PREV NEXT INS DEL EDIT PRNT EXIT
NEXT buttons to
scroll to the DATA
CHANNEL to be SETUP X.X NAME: CONC
edited.
<SET SET> EDIT EXIT
0 0 0 ENTR EXIT
Toggle these buttons to
set the days between
reports (0 366).
SETUP X.X REPORT PERIOD TIME:01:00
EXAMPLE:
Given the above settings, if DAS parameters are activated at 7:57:35 the first
sample would occur at 7:58 and the first report would be calculated at 8:00
consisting of data points for 7:58, 7:59 and 8:00.
During the next hour (from 8:01 to 9:00), the instrument will take a sample
reading every minute and include 60 sample readings.
Use the PREV and PREV MEXT INS DEL EDIT PRNT EXIT
NEXT buttons to scroll
to the DATA
CHANNEL to be SETUP X.X NAME: CONC
edited.
<SET SET> EDIT EXIT
Use the PREV and PREV NEXT INS DEL EDIT PRNT EXIT
NEXT buttons to
scroll to the DATA
CHANNEL to be SETUP X.X NAME: CONC
edited.
<SET SET> EDIT EXIT
Use the PREV and PREV NEXT INS DEL EDIT PRNT EXIT
NEXT buttons to
scroll to the DATA
CHANNEL to be SETUP X.X NAME: CONC
edited.
<SET SET> EDIT EXIT
HOLDOFF also prevents DAS measurements from being made at certain times
when the quality of the analyzers NO2 measurements may be suspect (e.g. while
the instrument is warming up). In this case, the length of time that the
HOLDOFF feature is active is determined by the value of the internal variable
(VARS), DAS_HOLDOFF.
To set the length of the DAS_HOLDOFF period, go to the
SETUP>MORE>VARS menu and at the DAS_HOLDOFF parameter (see Table
6-2), press the Edit button.
Use the PREV and PREV NEXT INS DEL EDIT PRNT EXIT
NEXT buttons to
scroll to the DATA
CHANNEL to be SETUP X.X NAME: CONC
edited.
<SET SET> EDIT PRNT EXIT
See Section 9.2.1 for configuration commands and their strict syntax. Commands
can be pasted in from of an existing text file, which was first edited offline and
then uploaded through a specific transfer procedure.
Where:
X is a command type designator, a single character indicating the
message type, as shown in the Table 9-2.
DDD:HH:MM is the time stamp, the date and time when the message was
issued. It consists of the Day-of-year (DDD) as a number from 1
to 366, the hour of the day (HH) as a number from 00 to 23, and
the minute (MM) as a number from 00 to 59.
[ID] is the analyzer ID, a number with 1 to 4 digits.
MESSAGE is the message content that may contain warning messages, test
measurements, variable values, etc.
<CRLF> is a carriage return / line feed pair, which terminates the
message.
The uniform nature of the output messages makes it easy for a host computer to
parse them into an easy structure. Keep in mind that the front panel display does
not give any information on the time a message was issued, hence it is useful to
log such messages for troubleshooting and reference purposes. Terminal
emulation programs such as HyperTerminal can capture these messages to text
files for later review.
Once this is completed, the appropriate setup command line for your modem can
be entered into the analyzer. The default setting for this feature is:
AT Y0 D0 H0 I0 S0=0
This string can be edited to match your modems initialization and can be up to
100 characters long.
CFG DAS RNGE PASS CLK MORE EXIT SETUP X.X COM1 MODE:0
Test runs
automatically.
SETUP X.X MODEM INITIALIZED
To log on to the T500U analyzer with SECURITY MODE feature enabled, type:
LOGON 940331
940331 is the default password. To change the default password, use the variable
RS-232_PASS issued as follows:
V RS-232_PASS=NNNNNN
CAUTION!
Insufficient gas flow allows gas to build up to levels that will contaminate
the instrument or present a safety hazard to personnel.
In units with a permeation tube option installed, either the tube must be
removed and stored in sealed container (use original container that tube
was shipped in) during periods of non-operation, or vacuum pump must
be connected and powered on to maintain constant gas flow though the
analyzer at all times.
(See Figure 3-4 for location and Section 11.3.2 for instructions on how to remove the
perm tube when the unit is not in operation).
For electronic data recording, the T500U provides an internal data acquisition
system (DAS), which is described in detail in Section 8.
APICOM, a remote control program, is also provided as a convenient and
powerful tool for data handling, download, storage, quick check and plotting (see
Sections 8.2.1, and the APICOM software manual downloadable from:
http://www.teledyne-api.com/manuals ).
NEVER press the ENTR button if you are only checking calibration.
Pressing the ENTR button during the following procedure resets the
stored values for OFFSET and SLOPE and alters the instruments
Calibration. This should ONLY BE DONE during an actual calibration
of the T500U.
Source of
VENT here if input
SAMPLE GAS
is pressurized Removed during
calibration
MODEL T700
Calibrated NO2
at HIGH Span
Concentration
Gas Dilution
Enclosure Wall
Calibrator
SAMPLE
MODEL 701
EXHAUST
Zero Gas
Generator Chassis
Vent here if output of calibrator
is not already vented.
Figure 10-1: Set up for Manual Calibrations/Checks in Base Configuration w/Gas Dilution
Calibrator
Concentration field
displays all of the
available gas Toggle TST> button until ...
measurements
throughout this
procedure.
SAMPLE NO2 STB= XXX.X PPM NO2=XXX.X
NOTE:
In certain instances where low Span
gas concentrations are present ( 50
ppb), both the ZERO & SPAN buttons
may appear simultaneously.
Note If the ZERO or SPAN buttons are not displayed, the measurement
made is outside the allowable range for a reliable calibration.
Note If the ZERO or SPAN buttons are not displayed, the measurement is
outside the allowable range for a reliable calibration.
Source of
VENT here if input
SAMPLE GAS
is pressurized Removed during
CAL calibration
MODEL 700E
Gas Dilution
Calibrated NO2
Concentration
at HIGH Span
Enclosure Wall
Calibrator
SAMPLE
Figure 10-2: Pneumatic Connections for T500U Precision Calibration when IZS Generator Present
SAMPLE
EXHAUST
PUMP
Filter ZERO AIR Chassis
External Zero
FROM DRYER
Air Scrubber
Figure 10-3: Pneumatic Connections for Manual Calibration/Checks with the Internal Span Gas
Generator
Note If the ZERO or SPAN buttons are not displayed, the measurement
made is out of the allowable range for a reliable calibration. See
Section 13.4.4 for troubleshooting tips.
3. Ensure that the reporting range span point is set for a value at least 10%
higher than the measured value of the permeation tube output.
Note If the ZERO or SPAN buttons are not displayed, the measurement
made during is out of the range allowed for a reliable calibration.
For each mode, there are seven parameters that control operational details of the
SEQUENCE as presented next in Table 10-2.
MODE OR
VALUE COMMENT
ATTRIBUTE
SEQUENCE 2 Define Sequence #2
Select Zero and
MODE ZERO-SPAN
Span Mode
TIMER ENABLE ON Enable the timer
Start after
STARTING DATE 31 JUL 2013
July 31, 2013
First Span starts at
STARTING TIME 14:15
2:15PM
Do Sequence #2
DELTA DAYS 2
every other day
Do Sequence #2
DELTA TIME 00:30
-hr later each day
Operate Span valve
DURATION 15.0
for 15 min
Calibrate at end of
CALIBRATE OFF
Sequence
SETUP X.X
Format : DD-MON-YY
CONTINUE NEXT PAGE
with STARTING TIME
Alternately, use the Auto Cal feature described in Section 10.5 with
the CALIBRATE attribute set to OFF.
2. Open the T500Us hinged front panel and disconnect the pneumatic fittings,
using the appropriate wrenches, and remove the disposable sample filter.
5. Remove the old permeation tube and replace it with the new tube.
(Ensure that the tube is placed into the larger of two holes and that the
open permeation end of the tube (plastic) is facing up).
6. Re-attach the cover with three screws (Ensure that the three screws are
tightened evenly).
7. Replace the analyzer cover, plug the power cord back in and turn on the
analyzer.
8. Carry out a span check to see if the new permeation device works
properly (see Section 10.3.2).
9. The permeation rate may need several days to stabilize.
1. Turn OFF power to the instrument and remove the instrument cover.
2. Install a leak checker on the sample inlet at the rear panel.
3. Cap rear panel ports and cap the pump port.
If zero/span valves are installed, disconnect the tubing from the zero
and span gas ports and cap the ports (Figure 3-3).
4. Pressurize the instrument with the leak checker, allowing enough time to
fully pressurize the instrument.
Do not exceed 10 psi pressure.
5. Once the leak has been located and repaired, the leak-down rate of the
indicated pressure should be less than 1 psi in 5 minutes after the
pressure is turned off. Replace the instrument cover and restart the
analyzer.
6. If the leak still cannot be found, check each tube connection (fittings,
hose clamps) with soap bubble solution, looking for fine bubbles.
Pressurize the instrument with the leak checker, allowing enough
time to fully pressurize the instrument.
Do not exceed 10 psi pressure.
Wet the bench last with soap solution.
7. Clean surfaces from soap solution, reconnect the sample and pump lines
and replace the instrument cover.
8. Restart the analyzer.
Sample flow checks are useful for monitoring the actual flow of the instrument.
A decreasing sample flow may point to slowly clogging pneumatic paths, most
likely the sample flow restrictor or the sample filter. To perform a sample flow
check:
1. Disconnect the sample inlet tubing from the rear panel SAMPLE port.
2. Attach the outlet port of a flow meter to the sample inlet port on the rear
panel.
Ensure that the inlet to the flow meter is at atmospheric pressure.
3. Check that the sample flow measured with the external flow meter is
within specification.
4. If sample flow is out of specification, first check for kinks in the tubing,
then check for leaks (Section 11.3.5.1). Once kinks and leaks are ruled
out, first replace the sample flow restrictor (see Figure 3-4 for location),
and if the sample flow is still out of specification, replace the sample filter
(Section 11.3.1).
If sample flow is still out of spec, call Technical Support.
CAUTION
The operations outlined in this section must be performed by qualified
maintenance personnel only.
WARNING
RISK OF ELECTRICAL SHOCK
Some operations need to be carried out with the analyzer open and
running.
Exercise caution to avoid electrical shocks and electrostatic or
mechanical damage to the analyzer.
Do not drop tools into the analyzer or leave those after your procedures.
Do not short or touch electric connections with metallic tools while
operating inside the analyzer.
Use common sense when operating inside a running analyzer.
Note The front panel of the analyzer is hinged at the bottom and may be
opened to gain access to various components mounted on the panel
itself or located near the front of the instrument (such as the sample
filter).
Remove the locking screw located at the right-hand side of the front
panel.
Examine the values of all TEST functions and compare them to factory
values. Note any major deviations from the factory values and take
corrective action.
Use the internal electronic status LEDs to determine whether the electronic
communication channels are operating properly.
Verify that the DC power supplies are operating properly by checking
the voltage test points on the relay PCA.
Note that the analyzers DC power wiring is color-coded and these
colors match the color of the corresponding test points on the relay
PCA.
Technical Support data indicate that the majority of all problems are
eventually traced to leaks in the internal pneumatics of the analyzer
or the diluent gas and source gases delivery systems (refer to
Section 11.3.5).
The analyzer will also alert the user via the Serial I/O COM port(s).
To view or clear the various warning messages press:
NOTE:
If a warning message persists after
SAMPLE SYSTEM RESET
several attempts to clear it, the message
Suppresses the may indicate a real problem and not an
TEST CAL MSG CLR SETUP
warning messages. artifact of the warm-up period.
TEST CAL MSG CLR SETUP Press CLR to clear the current
message.
If more than one warning is
active, the next message will take
SYSTEM CONFIG INITIALIZED its place.
TEST CAL CLR SETUP
Once the last warning has
been cleared, the analyzers
display will return to its
standard Sample Mode STANDBY RANGE=500.0 PPB NO2=XXXX
configuration.
<TST TST> CAL SETUP
13.1.3. DIAG > SIGNAL I/O: USING THE DIAGNOSTIC SIGNAL I/O
FUNCTION
The signal I/O diagnostic mode allows access to the digital and analog I/O in the
analyzer. Some of the digital signals can be controlled through the touchscreen.
These signals are useful for troubleshooting in three ways:
The technician can view the raw, unprocessed signal level of the analyzers
critical inputs and outputs.
Many of the components and functions that are normally under algorithmic
control of the CPU can be manually exercised.
The technician can directly control the signal level Analog and Digital Output
signals.
PREV NEXT JUMP EDIT PRNT EXIT Use the JUMP button to
Use the PREV and go directly to a specific i\
NEXT buttons to cycle I/O signal.
through the I/O signals. (see Appendix A for a list
DIAG I/O JUMPTO: 0 of all I/O SIGNALS)
0 0 JUMP ENTR EXIT
Note Any I/O signals changed while in the signal I/O menu will remain in
effect ONLY until signal I/O menu is exited. The Analyzer regains
control of these signals upon exit.
Motherboard
If D1 is blinking, then the other LEDs can be used in conjunction with DIAG
Menu Signal I/O to identify hardware failures of the relays and switches on the
Relay PCA.
D1 (RED)
Watchdog Indicator
Continuously
D7 Green Internal Zero/Span (option) valve ON or OFF
Valve broken or stuck, valve driver chip broken
Continuously
D8 Green Sample/Cal valve (option) ON or OFF
Valve broken or stuck, valve driver chip broken
Continuously
D9 Green Auto Reference valve ON or OFF
Valve broken or stuck, valve driver chip broken
D10 spare
LED ROW 2
Internal Zero/Span (option) Continuously
D5 Yellow ON or OFF
Heater broken, thermocouple broken
heater
Continuously
D6 Green Oven heater ON or OFF
Heater broken, thermistor broken
D11-15 spares
Continuously
D16 Green Internal pump ON or OFF
Not receiving power; relay board bad
13.4.2. NO RESPONSE
If the instrument shows no response (display value is near zero) even though
sample gas is supplied properly and the instrument seems to perform correctly.
1. Confirm the lack of response by supplying or NO2 span gas of about 80% of
the range value to the analyzer.
2. Check the sample flow rate for proper value.
3. Check for disconnected cables to the sensor module.
Check to ensure that there is no ambient air leaking into the zero air line.
Leaks can add an offset signal to the ambient sample, zero air or span
gas, whose concentration can fluctuate while a linearity test is being
performed.
Check for leaks as described in Section 11.3.5.
Span gas overflow is not properly vented and creates a back-pressure on the
sample inlet port.
Also, if the span gas is not vented at all and does not supply enough
sample gas, the analyzer may be evacuating the sample line.
Ensure to create and properly vent excess span gas.
Note There will not be an AREF warning if the AREF feature was disabled
due to prolonged time in CAL mode. Ensure the instrument is
returned to SAMPLE mode as soon as a calibration has been
completed.
Note The AREF warning displays the value of the Auto Reference reading
when the warning occurs.
If this value is higher than 1100 Mm-1, check that the Auto Reference valve is
operating properly:
In the DIAG menu go to SIGNAL I/O>AUTO_REF_VALVE to toggle the
valve on and off.
Listen to hear whether the valve is switching, and see if the respective
LED on the relay board lights accordingly.
Check the power supply to the valve (12 V to the valve should turn on
and off when measured with a voltmeter).
Note It takes only a small leak across the ports of the valve to show
excessive Auto Reference values when supplying high
concentrations of span gas.
If the Auto Reference valve is working properly, then the problem could be due to
dirty mirrors. Please contact Technical Support to confirm this and to rule out
other possibilities (Section 13.9).
A voltmeter should be used to verify that the DC voltages are correct per the
values in the table below, and an oscilloscope, in AC mode, with band limiting
turned on, can be used to evaluate if the supplies are producing excessive noise (>
100 mV p-p).
13.6.7. MOTHERBOARD
For each of the steps the output should be within 1% of the nominal value listed
except for the 0% step, which should be within 0mV 2 to 3 mV. Ensure you
take into account any offset that may have been programmed into channel (See
Section 6.8.3.9).
If one or more of the steps fails to be within these ranges, it is likely that there has
been a failure of the either or both of the Digital-to-Analog Converters (DACs)
and their associated circuitry on the motherboard. To perform the test connect a
voltmeter to the output in question and perform an analog output step test as
follows:
Choose a parameter in the Signal I/O function list (see Section 13.1.3) such as
EXT_ZERO_CAL
Compare this voltage at its origin (see the interconnect drawing and
interconnect list in Appendix D) with the voltage displayed through the
signal I/O function.
If the wiring is intact but there is a large difference between the
measured and displayed voltage (10 mV) then the motherboard is bad.
+DC Gnd
1 2 3 4 5 6 7 8 D +
SYSTEM_OK
1000
Figure 13-7: Typical Set Up of Status Output Test
13.6.8. CPU
There are two major types of CPU board failures, a complete failure and a failure
associated with the Disk On Module (DOM). If either of these failures occurs,
contact the factory.
For complete failures, assuming that the power supplies are operating properly
and the wiring is intact, the CPU is faulty if on power-on, the watchdog LED on
the motherboard is not flashing.
In some rare circumstances, this failure may be caused by a bad IC on the
motherboard, specifically U57, the large, 44 pin device on the lower right
hand side of the board. If this is true, removing U57 from its socket will allow
the instrument to start up but the measurements will be invalid.
If the analyzer stops during initialization (the front panel display shows a fault
or warning message), it is likely that the DOM, the firmware or the
configuration and data files have been corrupted.
Place it inside the chassis, next to the thermistor labeled XT1 (above
connector J108) on the motherboard.
Compare its reading to the value of the test function BOX TEMP.
Also, there may be more detailed service notes for some of the below
procedures. Contact Teledyne API's Technical Support Department.
Replacing the Disk-on-Module (DOM) will cause loss of all DAS data; it may
also cause loss of some instrument configuration parameters unless the
replacement DOM carries the exact same firmware version. Whenever changing
the version of installed software, the memory must be reset. Failure to ensure that
memory is reset can cause the analyzer to malfunction, and invalidate
measurements. After the memory is reset, the A/D converter must be re-
calibrated, and all information collected in Step 1 below must be re-entered
before the instrument will function correctly. Also, zero and span calibration
should be performed.
1. Document all analyzer parameters that may have been changed, such as
range, auto-cal, analog output, serial port and other settings before replacing
the DOM.
2. Turn off power to the instrument, fold down the rear panel by loosening the
mounting screws.
3. While looking at the electronic circuits from the back of the analyzer, locate
the Disk-on-Module in the right-most socket of the CPU board.
4. The DOM should carry a label with firmware revision, date and initials of the
programmer.
5. Remove the nylon standoff clip that mounts the DOM over the CPU board,
and lift the DOM off the CPU. Do not bend the connector pins.
6. Install the new Disk-on-Module, making sure the notch at the end of the chip
matches the notch in the socket.
7. It may be necessary to straighten the pins somewhat to fit them into the
socket. Press the chip all the way in.
8. Close the rear panel and turn on power to the machine.
If the replacement DOM carries a firmware revision, re-enter all of the setup
information.
Retainer
Mounting
Screws
AC Relay
Retainer Plate
Figure 13-8: Relay PCA with AC Relay Retainer In Place
The Relay retainer plate installed on the relay PCA covers the lower right
mounting screw of the relay PCA. Therefore, when removing the relay PCA, the
retainer plate must be removed first.
Mounting
Screws
QUESTION ANSWER
Why does the ENTR button Sometimes the ENTR button will disappear if you select a setting that
sometimes disappear on the is invalid or out of the allowable range for that parameter, such as
front panel display? trying to set the 24-hour clock to 25:00:00 or a range to less than 1 or
more than 20000 ppb. Once you adjust the setting to an allowable
value, the ENTR button will re-appear.
Why is the ZERO or SPAN The T500U disables certain these buttons expected span or zero value
button not displayed during entered by the users is too different from the gas concentration
calibration? actually measured value at the time. This is to prevent the accidental
recalibration of the analyzer to an out-of-range response curve.
EXAMPLE: The span set point is 400 ppb but gas concentration being
measured is only 50 ppb.
How do I enter or change the Press the CONC button found under the CAL or CALS buttons of the
value of my Span Gas? main SAMPLE display menus to enter the expected NO2 span
concentration.
Can I automate the calibration Any analyzer with zero/span valve can be automatically calibrated
of my analyzer? using the instruments AutoCal feature.
How do I measure the sample Sample flow is measured by attaching a calibrated flow meter to the
flow? sample inlet port when the instrument is operating. The sample flow
should be 900 cm/min 10%.
Can I use the DAS system in Yes. Section 8 describes the setup and operation of the DAS system
place of a strip chart recorder or in detail.
data logger?
How often do I need to change Annually or as needed. Section 11.1 contains a maintenance
the sample filter? schedule listing the most important, regular maintenance tasks. Highly
polluted sample air may require more frequent changes.
QUESTION ANSWER
How long does the sample The sample pump should last one to two years and should be replaced
pump last? when necessary, according to the Maintenance schedule (Table
11-1).,
The baud rate of the analyzers COM port does not match that of
the serial port of your computer/data logger (Section 7.2.2).
How do I make the instruments This most commonly occurs when an independent metering device is
display and my data logger used besides the data logger/recorder to determine gas concentration
agree? levels while calibrating the analyzer. These disagreements result from
the analyzer, the metering device and the data logger having slightly
different ground levels.
Use the data logger itself as the metering device during calibration
procedures.
The CAPS method employed in the T500U is unique in that it applies this
fundamental optical absorption law in the frequency domain, rather than using
relative changes in light intensity as the primary signal. Ultraviolet light (UV)
from the modulating high intensity LED enters a near confocal optical cell
(Figure 14-1) through the rear of mirror A. The intensity of the light, as observed
by the detector, which is also modulating at a slightly different frequency, located
behind Mirror B, builds exponentially in the cell while the LED is ON. The
opposite is true when the LED is OFF. Because both mirrors are highly reflective
at 450 nm, a prominent absorption band for NO2, the light takes a considerable
amount of time to plateau in the absence of the absorbing gas. However, when
NO2 is present, the mean path length traveled by the light is significantly reduced.
This has two effects on the observed intensity as measured by the detector:
The light plateau intensity level is lower, more importantly for the T500U
Thus, an observed phase shift from the modulating LED is detected (Figure
14-2). The phase shift is largest when measuring zero air and decreases when
NO2 is present.
Both the LED and the Detector are modulated ON and OFF such that the
observed signal has a much lower frequency, equal to the difference between the
modulated frequencies and is referred to as a beat frequency. The system
hardware and software take advantage of this, as it makes it easier to post process
the signal using a micro controller. The technique is known as heterodyning.
The instrument translates the phase shift from the presence of absorbing gas into
a concentration measurement. Typical absorption techniques of other analyzers
take a reference and measure value of the light intensity level in order to derive
concentration and compensate for source drift. Using the CAPS technique the
amount of phase shift remains constant for a given concentration, even if the LED
drifts over time. The measurement approach offers many advantages over
traditional (or Chemi) analyzers, such as faster response (single gas stream),
lower noise at span and more importantly greater specificity.
SAMP SAMP
2X 450 nm
BANDPASS MIRRORS
FILTER
T500U OVEN
LED ON
Absorbing Gas
LED OFF
(Black = LED State, Blue = Light build up in the absence of NO 2, Green = phase shifted\attenuated light)
Term Description/Definition
10BaseT an Ethernet standard that uses twisted (T) pairs of copper wires to transmit at
10 megabits per second (Mbps)
100BaseT same as 10BaseT except ten times faster (100 Mbps)
APICOM name of a remote control program offered by Teledyne-API to its customers
ASSY Assembly
CAS Code-Activated Switch
CD Corona Discharge, a frequently luminous discharge, at the surface of a
conductor or between two conductors of the same transmission line,
accompanied by ionization of the surrounding atmosphere and often by a power
loss
CE Converter Efficiency, the percentage of the total amount that is actually
converted (e.g., light energy into electricity; NO2 into NO, etc.)
CEM Continuous Emission Monitoring
Chemical formulas that may be included in this document:
CO2 carbon dioxide
C3H8 propane
CH4 methane
H2O water vapor
HC general abbreviation for hydrocarbon
HNO3 nitric acid
H2S hydrogen sulfide
NO nitric oxide
NO2 nitrogen dioxide
NOX nitrogen oxides, here defined as the sum of NO and NO2
NOy nitrogen oxides, often called odd nitrogen: the sum of NOX plus other compounds such as
HNO3 (definitions vary widely and may include nitrate (NO3), PAN, N2O and other
compounds as well)
NH3 ammonia
O2 molecular oxygen
O3 ozone
SO2 sulfur dioxide
3
cm metric abbreviation for cubic centimeter (replaces the obsolete abbreviation
cc)
CPU Central Processing Unit
DAC Digital-to-Analog Converter
DAS Data Acquisition System
DCE Data Communication Equipment
DFU Dry Filter Unit
Term Description/Definition
DHCP Dynamic Host Configuration Protocol. A protocol used by LAN or Internet
servers to automatically set up the interface protocols between themselves and
any other addressable device connected to the network
DIAG Diagnostics, the diagnostic settings of the analyzer.
DOM Disk On Module, a 44-pin IDE flash drive with up to 128MB storage capacity for
instruments firmware, configuration settings and data
DOS Disk Operating System
DRAM Dynamic Random Access Memory
DR-DOS Digital Research DOS
DTE Data Terminal Equipment
EEPROM Electrically Erasable Programmable Read-Only Memory also referred to as a
FLASH chip or drive
ESD Electro-Static Discharge
ETEST Electrical Test
Ethernet a standardized (IEEE 802.3) computer networking technology for local area
networks (LANs), facilitating communication and sharing resources
FEP Fluorinated Ethylene Propylene polymer, one of the polymers that Du Pont
markets as Teflon
Flash non-volatile, solid-state memory
FPI Fabry-Perot Interface: a special light filter typically made of a transparent plate
with two reflecting surfaces or two parallel, highly reflective mirrors
GFC Gas Filter Correlation
2 a clocked, bi-directional, serial bus for communication between individual
I C bus
analyzer components
Integrated Circuit, a modern, semi-conductor circuit that can contain many basic
IC components such as resistors, transistors, capacitors etc in a miniaturized
package used in electronic assemblies
IP Internet Protocol
IZS Internal Zero Span
LAN Local Area Network
LCD Liquid Crystal Display
LED Light Emitting Diode
LPM Liters Per Minute
MFC Mass Flow Controller
Units of measurement used to express light absorption, expressed in MEGA
Mm-1
(106) per meter.
M/R Measure/Reference
NDIR Non-Dispersive Infrared
Term Description/Definition
the mass, expressed in grams, of 1 mole of a specific substance. Conversely,
one mole is the amount of the substance needed for the molar mass to be the
same number in grams as the atomic mass of that substance.
MOLAR MASS EXAMPLE: The atomic weight of Carbon is 12 therefore the molar mass of
Carbon is 12 grams. Conversely, one mole of carbon equals the amount of
carbon atoms that weighs 12 grams.
Atomic weights can be found on any Periodic Table of Elements.
NDIR Non-Dispersive Infrared
NIST-SRM National Institute of Standards and Technology - Standard Reference Material
PC Personal Computer
PCA Printed Circuit Assembly, the PCB with electronic components, ready to use
PC/AT Personal Computer / Advanced Technology
PCB Printed Circuit Board, the bare board without electronic component
Per-Fluoro-Alkoxy, an inert polymer; one of the polymers that Du Pont markets
PFA
as Teflon
Programmable Logic Controller, a device that is used to control instruments
PLC
based on a logic level signal coming from the analyzer
PLD Programmable Logic Device
PLL Phase Lock Loop
Photo Multiplier Tube, a vacuum tube of electrodes that multiply electrons
PMT
collected and charged to create a detectable current signal
P/N (or PN) Part Number
PPB Parts per billion
PPM Parts per million
PPT Parts per trillion
PSD Prevention of Significant Deterioration
Poly-Tetra-Fluoro-Ethylene, a very inert polymer material used to handle gases
PTFE that may react on other surfaces; one of the polymers that Du Pont markets as
Teflon
PVC Poly Vinyl Chloride, a polymer used for downstream tubing
Rdg Reading
specification and standard describing a serial communication method between
RS-232 DTE (Data Terminal Equipment) and DCE (Data Circuit-terminating Equipment)
devices, using a maximum cable-length of 50 feet
specification and standard describing a binary serial communication method
RS-485 among multiple devices at a data rate faster than RS-232 with a much longer
distance between the host and the furthest device
SAROAD Storage and Retrieval of Aerometric Data
SLAMS State and Local Air Monitoring Network Plan
SLPM Standard Liters Per Minute of a gas at standard temperature and pressure
STP Standard Temperature and Pressure
Term Description/Definition
Transfer Control Protocol / Internet Protocol, the standard communications
TCP/IP
protocol for Ethernet devices
TEC Thermal Electric Cooler
TPC Temperature/Pressure Compensation
Universal Serial Bus: a standard connection method to establish communication
USB between peripheral devices and a host controller, such as a mouse and/or
keyboard and a personal computer or laptop
VARS Variables, the variable settings of the instrument
V-F Voltage-to-Frequency
Z/S Zero / Span
1
TEST1 CAL CALZ2 CALS2 MSG1 CLR SETUP
Press to cycle
through the
ZERO active warning
RANGE=[Value]PPB
PHASE=[Value]DEG messages.
MEAS_L=[Value]Mm-1
AREF_L=[Value]Mm-1 Press to clear
SMP PRS=[Value]InHg an active
SMP TEMP=[Value]C warning
BOX TEMP=[Value]C messages.
SLOPE=[Value]
OFFSET=[Value]PPB
STABIL=[Value]PPB PRIMARY SETUP
OVEN TMP=[Value]C5 MENU
TIME=[HH:MM:SS]
SECONDARY SETUP
MENU
1
Only appears when warning messages are active.
2
3
Only appears if analyzer is equipped with Internal Span Gas option. COMM VARS DIAG AREF ALRM3
Only appears when the Concentration Alarm option is active.
SAMPLE SETUP
ON
<PREV NEXT> PREV NEXT MODE OFF
Go to iDAS TIME DATE
Menu Tree
SEQ 1) Go to SECONDARY SETUP
MODEL NAME SEQ 2) Menu Tree
PART NUMBER SEQ 3)
SERIAL NUMBER
SOFTWARE REVISION
LIBRARY REVISION MODE SET UNIT
iCHIP SOFTWARE
REVISION PREV NEXT
CPU TYPE & OS SNGL IND AUTO PPB PPM UGM MGM
REVISION
DATE FACTORY
CONFIGURATION SAVED DISABLED
SETUP X.X LOW RANGE:500.0 Conc
ZERO
ZERO-SPAN 0 0 5 0 0 .0 ENTR EXIT
SPAN
0 0 5 0 0 .0 ENTR EXIT
2
SET
ON
TIMER ENABLE
<SET SET> OFF
STARTING DATE4
STARTING TIME4
DELTA DAYS4
1
DELTA TIME4
ACAL menu and its submenus only appear if the analyzer is
equipped with calibration valves or the internal span gas DURATION
generator. ON
2
Appears whenever the currently displayed sequence is not set CALIBRATE
for DISABLED. OFF
3
Only appears when reporting range is set to AUTO range mode. RANGE TO CAL3
4 Only Appears if TIME ENABLE is set to ON.
LOW HIGH
SAMPLE SETUP
SAMPLE SETUP
YES NO NO2
NO2, 0, 213, REPORTED GAS TYPE
GAS ID Set/create unique
REPORTED gas ID number
1
When Hessen Protocol turned ON
ON in COM1 or COM2 mode.
OFF
SAMPLE SETUP
PREV NEXT
SIGNAL I/O ANALOG ANALOG I/O TEST CHAN NO2 LED PUMP
OUTPUT CONFIGURATION OUTPUT DISABLE DISABLE
SAMPLE SETUP
VIEW EDIT
CONC
PNUMTC PREV NEXT INS DEL EDIT2 PRNT
CALDAT
CONC
DIAG YES NO
HIRES PNUMTC
CALDAT
VIEW
DIAG
HIRES <SET SET> NEXT NX10
PV10 PREV NEXT NX10 <PRM PRM>
1
Name Message Text Description
3
RANGE RANGE=500.0 PPB D/A range in single or auto-range
modes.
RANGE1 RANGE1=500.0 PPB 3 D/A #1 range in dual range mode.
3
RANGE2 RANGE2=500.0 PPB D/A #2 range in dual range mode.
PHASE PHASE=15.32 DEG Box phase value.
-1
MEASLOSS MEAS_L=300.1 Mm Measurement loss value.
AUTOREF AREF_L=400.3 Mm-1 Autoref value.
SAMPPRESS SMP PRS=29.92 InHg Sample pressure.
SAMPTEMP SMP TEMP=31.2 C Sample temperature.
BOXTEMP BOX TEMP=31.3 C Box internal temperature.
SLOPE SLOPE=1.000 Slope for current range, computed
during zero/span calibration.
OFFSET OFFSET=0.0 PPB 3 Offset for current range, computed
during zero/span calibration.
STABILITY STABIL=0.0 PPB 3 Concentration stability (standard
deviation based on setting of
STABIL_FREQ and
STABIL_SAMPLES).
RESPONSE 2 RSP=3.11(0.00) SEC Instrument response. How
frequently concentration is updated.
Time in parenthesis is standard
deviation.
MANIFOLDTEMP MF TEMP= 50.1 C Manifold temperature.
2
MANIFOLDDTY MF HTR=45.1% Manifold heater duty cycle.
IZSTEMP IZS TEMP= 50.1 C IZS temperature.
2
IZSDTY IZS HTR=45.1% IZS heater duty cycle.
OVENTEMP OVEN TMP= 50.1 C Oven temperature.
2
OVENDTY OVEN HTR=45.1% Oven heater duty cycle.
SIGTHRUPUT SIG=2414.7 MV Bench signal throughput value.
1
Name Message Text Description
SIN_1 SIN 1=-15413 Bench SIN_1/2 value.
COS_1 COS 1=25173 Bench COS_1/2 value.
SIN_2 SIN 2=-15413 Bench SIN_2/2 value.
COS_2 COS 2=25173 Bench COS_2/2 value.
SIN_OV SIN OVP=2 Bench SIN overlap value.
COS_OV COS OVP=3 Bench COS overlap value.
ACCUM_OV ACCUM OV=0 Bench accum overflow value.
TESTCHAN TEST=2753.9 MV Value output to TEST_OUTPUT
analog output, selected with
TEST_CHAN_ID variable.
4
XIN1 AIN1=37.15 EU External analog input 1 value in
engineering units.
4
XIN2 AIN2=37.15 EU External analog input 2 value in
engineering units.
4
XIN3 AIN3=37.15 EU External analog input 3 value in
engineering units.
4
XIN4 AIN4=37.15 EU External analog input 4 value in
engineering units.
4
XIN5 AIN5=37.15 EU External analog input 5 value in
engineering units.
4
XIN6 AIN6=37.15 EU External analog input 6 value in
engineering units.
4
XIN7 AIN7=37.15 EU External analog input 7 value in
engineering units.
4
XIN8 AIN8=37.15 EU External analog input 8 value in
engineering units.
CLOCKTIME TIME=14:48:01 Current instrument time of day
clock.
1
The name is used to request a message via the RS-232 interface, as in T BOXTEMP.
2
Engineering software.
3
Current instrument units.
4
External analog input option.
0 Spare
1
Internal span option.
2
Remote range control option.
3
MODBUS option.
4
Concentration alarm option.
5
High auto range relay option.
6
User-configurable D/A output option.
7
External analog input option.
Name Description
ATIMER Automatic timer expired
EXITZR Exit zero calibration mode
EXITSP Exit span calibration mode
EXITMP Exit multi-point calibration mode
SLPCHG Slope and offset recalculated
EXITDG Exit diagnostic mode
1
CONC1W Concentration limit 1 exceeded
1
CONC2W Concentration limit 2 exceeded
AREFW Auto-ref warning
BASELW Bench baseline loss warning
BCHCOM Bench communication failure warning
MFTMPW Manifold temperature warning
IZSTMPW IZS temperature warning
OVTMPW Oven temperature warning
STEMPW Sample temperature warning
SPRESW Sample pressure warning
BTEMPW Box temperature warning
1
Concentration alarm option.
The command syntax follows the command type, separated by a space character. Strings in
[brackets] are optional designators. The following key assignments also apply.
Option Description
/?, /h Prints command line usage without executing the firmware.
/u CPU-type CPU-type specifies the CPU type to use:
0 = Generic PC (default)
1 = AR-B1320, 386SX, PC/104
2 = AR-B9612A, 386SX, PC/104
/d device Device specifies the display type to use:
0 = PC video display and keyboard (default)
1 = Onyx I/O, parallel display (not currently supported)
2 = DMM-32 I/O, parallel display (not currently supported)
3 = ICA90, 8574-based I2C display
4 = API "E" series rear board, parallel display
5 = API "E" series rear board, 8574-based I2C display
/y Runs the firmware without prompting. This is useful for executing the
firmware automatically from within a batch file. The default is to prompt
for confirmation.
/w Disables the watchdog timer. The default is enabled.
/n file File specifies a different name and/or location for the configuration file. The
default is config.bin in the same directory as the executable file. If
you have configured flash memory as a disk drive, such as B:, then you can
store the configuration information in the flash using the command line
option o b:\config.bin.
/r file File specifies an alternate resource file containing the firmwares default
configuration. The default is the executable file.
/a address Address specifies the Com port for the bench interface. The default is 3.
Valid values are 2-4.
3, 5
/b address Address specifies the I/O address of the I2C interface data register. The
default is 0x32c (hex). This is usually 0x310 for the ICA90 board.
3, 5
/c address Address specifies the I/O address of the I2C interface control register. The
default is 0x32f (hex). This is usually 0x311 for the ICA90 board.
3, 5
/q irq IRQ specifies the IRQ number of the I2C interface. The default is IRQ5.
3, 4, 5
/k irq IRQ specifies the IRQ number of the keyboard (0 = use polling mode
instead of interrupt mode). The default is IRQ12.
Option Description
3, 5 2
/i address Address specifies the I C bus address of the host CPU. The default is 0x57
(hex).
3, 5
/s address Address specifies the I2C bus address of the keyboard/display. The default
is 0x4A (hex).
/l address Address specifies the I2C bus address of the relay board. The default is
0x44 (hex).
3
This option is used with display type 3.
4
This option is used with display type 4.
5
This option is used with display type 5.
Typical command lines
PC keyboard/display: T500U_B7 /y /w
AR-B1320 CPU and 8574-based keyboard/display: T500U_B7 /y /u 1 /d 5
Due to the dynamic nature of part numbers, please refer to the TAPI Website at
Note http://www.teledyne-api.com or call Customer Service at 800-324-5190 for more
recent updates to part numbers.
PLEASE COMPLETE THE FOLLOWING TABLE: (NOTE: DEPENDING ON OPTIONS INSTALLED, NOT ALL TEST
PARAMETERS SHOWN BELOW WILL BE AVAILABLE IN YOUR INSTRUMENT)
*IF OPTION IS INSTALLED
PARAMETER RECORDED VALUE ACCEPTABLE VALUE
RANGE PPB/PPM 5 PPB TO 1000 PPB
PHASE DEG 30 10 WITH ZERO AIR
MEAS_L WITH ZERO AIR Mm-1 400 800 WITH ZERO AIR
Mm-1 400 1800 Mm-1
MEAS_L WITH SPAN GAS
PPB 0-1000 PPB
AREF_L Mm-1 400 800 WITH ZERO AIR
~ 1-5 < AMBIENT
SMP PRS IN-HG-A
(CONSTANT)
SMP TEMP C 45 3C
BOX TEMP C AMBIENT 5C
SLOPE 1.0 0.3
OFFSET PPB 0.0 5
NO2 STAB PPB/PPM 1 PPB WITH ZERO AIR
OVEN TMP C 45 1C
IZS TEMP* C 50 1C
Values are in the Signal I/O
4096mv 2mv and Must be
REF_4096_MV MV
Stable
REF_GND MV 0 0.5 and Must be Stable
2. WHAT IS THE SAMPLE PRESSURE WITH THE SAMPLE INLET ON REAR OF MACHINE CAPPED?
SAMPLE PRESSURE - IN-HG-A
5. WHAT TEST(S) HAVE YOU DONE TRYING TO SOLVE THE PROBLEM? ________________________________
___________________________________________________________________________________________________
___________________________________________________________________________________________________
___________________________________________________________________________________________________
___________________________________________________________________________________________________
___________________________________________________________________________________________________
6. IF POSSIBLE, PLEASE INCLUDE A PORTION OF A STRIP CHART PERTAINING TO THE PROBLEM AND CIRCLE
THANK YOU FOR PROVIDING THIS INFORMATION. YOUR ASSISTANCE ENABLES TELEDYNE API TO RESPOND
W/MD
06950
06881
Xmitter
J4
J1
IZS
Thrm/Heater
J6 PCA, Caps Daq J8
J2 076730000
J3 J9 J5
PS1
SK1 SK2
(+5, 15)
J2 J7 J3
Preamp/Mixr Assy LED
07767 076490000
J9 J14
PS2
Samp Thrm Oven Thrm
SK1 (+12) SK2
019340700 058800000
RELAY BOARD J8
J18 045230100
Oven Fan Assy Hter/sw J1
07713 058820100
ZS Valve
ASSY, CAPS OVEN J11
076990000
042340000
AZ Valve
J14
J15 077120000
J2 LCD w/Touchscreen
J9 LCD Interface
Cntrl Mod J11 06790
07215 06697
J1
KEY:
Printed Documents are Uncontrolled
1. All part numbers in identify cables
The information hereon is the property of
API and is submitted in strictest ADVANCED POLLUTION INSTRUMENTATION
that are refered to in the accompanying confidence for reference only. Everywhere look
Unauthorized use by anyone for any othe
rpurposese is prohibited. This document
document 078480100. or any information contained in it may be
duplicated without proper authorization.