Qa Tables Final
Qa Tables Final
Daily Linac QA
Machine-Type Tolerance
Non- IMRT IMRT SRS/SBRT
Dosimetry
X-ray output - Photon (All Energies) 3%
X-ray output - Electron (All Energies) 3%
X-ray output - EDW (6x and 10x) 3%
Mechanical
Laser Localization 2mm 1.5mm 1mm
Distance Indicator (ODI) @iso 2mm 2mm 2mm
Collimator Size Indicator 2mm 2mm 1mm
Safety
Door Interlock (Beam Off) Functional
Door Closing Safety Functional
A/V Monitors Functional
Stereotactic Interlocks (Lockout) N/A N/A Functional
Radiation Area Monitor Functional
Beam On Indicator Functional
Monthly Linac QA
Machine-Type
Tolerance
Non- SRS/SBR
IMRT IMRT T
Dosimetry
X-ray output - Photon (All Energies) 2%
X-ray output - Electron (All Energies) 2%
Backup Monitor Chamber 2%
Typical Dose Rate (for each type) N/A 2% 2%
Photon Beam Profile 1%
Electron Beam Profile 1%
Electron Beam Energy 1%
Mechanical
Light/Radiation Field Coincidence 2mm or 1% on a side
Light/Radiation Field Coincidence
(asymmetric) 1mm or 1% on a side
Distance Check Device for Lasers Compared
with Front Pointer 1mm
Gantry/Collimator Angle Indicators 1.0
Accessory Trays 2mm
Jaw Position Indicators (Symmetric) 2mm
Jaw Position Indicators (Asymmetric) 1mm
Cross-Hair Centering (Walkout) 1mm
1mm/0.5
Treatment Couch Position Indicators 2mm/1 2mm/1
Wedge Placement Accuracy 2mm
Compensator Placement Accuracy 1mm
Latching of Wedges, Blocking Tray Functional
Localizing Lasers 2mm 1mm <2mm
Safety
Laser Guard-Interlock Test Functional
Respiratory Gating
Beam Output Constancy 2%
Phase, Amplitude Beam Control Functional
In-Room Respiratory Monitoring System Functional
Gating Interlock Functional
Annual Linac QA
Machine-Type Tolerance
Non- IMRT IMRT SRS/SBRT
Dosimetry
X-ray Flatness Change from
Baseline 1%
X-ray Symmetry Change from
Baseline 1%
Electron Flatness Change from
Baseline 1%
Electron Symmetry Change 1%
from Baseline
SRS Arc Rotation Mode
(Range: .5-10 MU/deg)
Monitor Units Set vs. Delivered N/A N/A 1 MU or 2%
and Gantry Arc Set
vs. Delivered
X-ray/Electron Output
Calibration (TG-51) 1% (Absolute)
Spot Check of FS Dependent 2% for FS < 4x4 cm2, 1%
Output Factors 4x4 cm2
Output Factors for Electron
Applicators 2% from Baseline
X-ray Beam Quality (PDD10 or
TMR20/10) 1% from Baseline
Electron Beam Quality (R50) 1mm
Physical Wedge Transmission 2%
5% (2-4
MU),2% 5
X-ray MU Linearity 2% 5 MU 5% (2-4 MU),2% 5 MU MU
Electron MU Linearity 2% 5 MU
X-ray Output vs. Dose Rate 2% from Baseline
X-ray Output vs. Gantry Angle 1% from Baseline
Electron Output vs. Gantry
Angle 1% from Baseline
Electron and X-ray off-axis
1% from Baseline
Factor vs. Gantry Angle
Arc Mode (Expected MU,
Degrees) 1% from Baseline
TBI/TSET mode Functional
PDD or TMR and OAF 1% (TBI) or 1mm PDD Shift
Constancy (TSET) from Baseline
TBI/TSET Output Calibration 2% from Baseline
TBI/TSET Accessories 2% from Baseline
Mechanical
Collimator Rotation Isocenter 1mm from Baseline
Gantry Rotation Isocenter 1mm from Baseline
Couch Rotation Isocenter 1mm from Baseline
Electron Applicator Interlocks Functional
Coincidence of Radiation and 2mm from 1mm from
Mechanical Isocenter Baseline 2mm from Baseline Baseline
Table Top Sag 2mm from Baseline
Table Angle 1.0
Table Travel Maximum Range
2mm
Movement in All Directions
Stereotactic Accessories,
Lockouts, etc. N/A N/A Functional
Safety
Follow Manufacturer's Test
Procedures Functional
Respiratory Gating
Beam Energy Constancy 2%
Temporal Accuracy of
Phase/Amplitude Gate On 100ms of expected
Calibration of Surrogate for
100ms of expected
Respiratory Phase/Amplitude
Interlock Testing Functional
Table 4: CT Simulator QA
CT Simulator QA
Frequency Tolerance Limits
Radiation and Patient Safety
Shielding Survey Initially NCRP Guidelines
Anually or after major component 20% of manufacturer
Patient Dose from CT-scan, CTDI
change specs
Electromechanical Components
Alignment of gantry lasers with
Daily 2 mm
center of imaging plane
Orientation of gantry lasers with 2 mm over the length
Monthly and after laser asdjustments
respect to the imaging plane of laser projection
Spacing of lateral wall lasers with
respect to lateral gantry lasers and Monthly and after laser asdjustments 2 mm
scan plane
Orientation of wall lasers with 2 mm over the length
Monthly and after laser asdjustments
respect to the imaging plane of laser projection
Orientation of the ceiling laser with 2 mm over the length
Monthly and after laser asdjustments
respect to the imaging plane of laser projection
Orientation of the CT-scanner 2 mm over the length
Monthly or when daily laser QA tests
tabletop with respect to the imaging and width of the
reveal rotational problems
plane tabletop
Table vertical and longitudinal 1 mm over the range of
Monthly
motion table motion
1 mm over the scan
Table indexing and position Annually
range
1 or 1 mm from
Gantry Tilt Accuracy Annually
nominal position
1 mm over the scan
Scan localization Annually
range
Annually (optional if CTDI accuracy
Radiation Profile Width Manufacturer specs
verified)
Sensitivity Profile Width Semiannually 1 mm of nominal value
After replacement of major generator Manufacturer specs or
Generator tests
component Report No. 39
Imaging Performance
Daily- for water
CT Number accuracy Monthly- 4 to 5 materials For water, 05 HU
Annually- Electron density phantom
Image Noise Daily Manufacturer specs
In plane spacial integrity Daily- x or y directions
1 mm
Monthly- both directions
Monthly- most common kVp
Field uniformity within 5 HU
Annually- other used kVp settings
consistent with
Electron density to CT number
Annually or after scanner calibration commissioning results
conversion
and test phantom
Spatial Resolution Annually Manufacturer specs
Contrast Resolution Annually Manufacturer specs
Annual
Dose Imaging dose Baseline
Imaging system performance X-ray generator performance Baseline
AP, mediolateral and craniocaudal
Geometric Accurate
orientations
support clinical
Long and short term planning of
System operation use and current
rescources (disk space, manpower, etc.)
policies
QA of Imaging Systems
Frequency Procedure non-SRS/SBRT SRS/SBRT
Planar MV imaging
Imaging and treatment coordinate coincidence 2 mm 1 mm
Scaling 2 mm 2 mm
Spatial resolution Baseline Baseline
Contrast Baseline Baseline
Monthly Uniformity and noise Baseline Baseline
Planar kV imaging
Imaging and treatment coordinate coincidence 2 mm 1 mm
Scaling 2 mm 1 mm
Spatial resolution Baseline Baseline
Contrast Baseline Baseline
Uniformity and noise Baseline Baseline
Planar MV imaging
Full range of travel SDD 5 mm 5 mm
Imaging dose Baseline Baseline
Planar kV imaging
Annual
Beam quality/energy Baseline Baseline
Imaging dose Baseline Baseline
Checksum No change
Monthly Subset of reference QA test set 2% or 2mm
I/O System 1mm
MU Calculations 2%
Annual Reference QA test set 2% or 2mm
I/O System 1mm
Table 8: MLC QA
MLC QA
Frequency Test Tolerance
Check of MLC-generated field vs.
simulator film before each field 2mm
treated
Double check of MLC field by
Expected field
Patient Specific therapists for each fraction
On-line imaging verification for
Physician discretion
patient on each infraction
Port film approval before second
Physician discretion
fraction
Dynamic/Universal/Virtual Wedge QA
Frequency Procedure Dynamic Universal Virtual
Morning checkout run
Daily Functional
for one angle
wedge factor for all C.A. axis 45 or 60 C.A. axis 45 or 60 5% from unity,
Monthly
energies WF (within 2%) WF (within 2%) otherwise 2%
check for wedge
Check of offcenter Check of offcenter Check of offcenter
angle for 60, full field
ratios @80% field ratios @80% field ratios @80% field
Annual and spot check for
width @10cm to be width @10cm to be width @10cm to be
intermediate angle,
within 2% within 2% within 2%
field size
Remote Afterloader QA
Frequency Test Tolerance
room safety door
Functional
interlocks, lights, alarms
console functions,
Functional
switches, batteries, printer
Each treatment day
visual inspection of source
Free of kinks and firmly attached
guides
verify accuracy of ribbon
Autoradiograph
preparation
calibration 3%
timer function 1%
At each source change or quarterly check accuracy of source
1mm
guides and connectors
mechanical integrity of
Functional
applicators
1. Klein E, Hanley J, Baymouth J. Task Group 142 report: Quality assurance of medical
accelerators. Med. Phys. 2009;36(9):4197-4212. http://dx.doi.org/10.1118/1.3190392.
2. Mutic S, Palta J, Butker EK, et al. Quality assurance for computed-tomography
simulators and thecomputed-tomography-simulation process: report of the AAPM
radiation therapy committee task group No. 66. Med Phys. 2003; 30(10): 2762-2792. doi:
10.1118/1.16092715.
3. Bissonnette J, Balter P, Dong L, et al. Quality assurance for image-guided radiation therapy
utilizing CT-based technologies: A report of the AAPM TG-179. Med Phys 2012; 39(4): 1946-
1963. http://dx.doi.org/10.1118/1.3690466.
4. Thompson C, Vann AM, Hilsendager L. QA of Imaging systems [Softchalk]. La Crosse, WI: UW-L
Medical Dosimetry Program; 2017.
5. Khan FM, Gibbons JP. Khans The Physics of Radiation Therapy. 5th ed. Philadelphia, PA:
Lippincott Williams & Wilkins; 2014:396-398.
6. Kutcher GL, Coia L, Gillin M, et al. Comprehensive QA for Radiation Oncology. AAPM Report No.
46. TG-40. Med Phys. 1994;21(4):581-618. https://doi.org/10.1118/1.597316.
7. Fraass B, Doppke K, Hunt M, et al. American Association of Physicists in Medicine Radiation
Therapy Committee Task Group 53: Quality assurance for clinical radiotherapy treatment
planning. Med Phys. 1998;25(10):1773-1828. https://doi.org/10.1118/1.598373.