Conductivity TDS Meter: Users Manual
Conductivity TDS Meter: Users Manual
Conductivity TDS Meter: Users Manual
Users Manual
Mode demploi
Bedienungshandbuch
Manual dUso
Manual de uso
WT-20
English
Conductivity TDS Meter
Users Manual
Repair
All test tools returned for warranty or non-warranty repair or
for calibration should be accompanied by the following: your
name, companys name, address, telephone number, and proof of
purchase. Additionally, please include a brief description of the
problem or the service requested and include the test leads with
the meter. Non-warranty repair or replacement charges should be
remitted in the form of a check, a money order, credit card with
expiration date, or a purchase order made payable to Amprobe
Test Tools.
In USA In Canada
Amprobe Test Tools Amprobe Test Tools
Everett, WA 98203 Mississauga, ON L4Z 1X9
Tel: 877-AMPROBE (267-7623) Tel: 905-890-7600
10
1 9
11
2 8
1) Battery Cover
3 2) Lcd Display
3) Power / Set Key
4 4) Hld / Cal / Down Key
5 5) Mode / Up Key
6) Electrode Cover
7) Conductivity Reading
(unit: uS or mS ) or TDS
Reading (unit: ppt or
ppm)
8) Display temperature in
either degree Celsius or
degree Fahrenheit
9) Calibration Mode
Indicator
6
10) Freeze Display
11) Low Battery Indicator
1
SYMBOLS
Caution! Refer to the explanation in this Manual
1 WT-20 meter
4 LR44 battery
1 Manual
If any of the items are damaged or missing, return
the complete package to the place of purchase for an
exchange.
2
INTRODUCTION
Congratulations on your purchase of WT-20 Conductivity
TDS meter. A convenient instrument to measure water
conductivity, TDS and temperature.
Features
IP65 Waterproof housing.
Dual display with ATC (C / F switchable)
Data hold to freeze display.
Low battery indicator.
Auto power off.
One touch only for calibration.
Operation
1. Remove the electrode cover to expose the
electrode out.
2. Press POWER to power on. LCD will display
parameters (ex: tnr, tCo, tds, rAn) in turns and
then enter normal display.
3. The meter is default in auto-ranging status. Press
MODE key more than 2 seconds can manually
select the range when the meter is in normal
measurement mode.
4. Set the temperature coefficient to the right value.
WT-20 is factory set to 2.1% per C. if needed, see
P1.3 to set the temperature coefficient.
5. Select the normalization temperature. WT-20 is
factory set to 25 C. See P1.2 to change it to 20 C.
6. Rinse the probe with deionized or distilled water
before use to remove any impurities adhering
to the electrode. Soak the probe more than 30
minutes to clear up the lazy effect of probe if the
meter is stored for a long time.
3
7. Dip the probe into the sample container. Make
sure no air bubbles trapped in the slot of the
probe. To remove air bubbles, give the probe
a gentle stir. Making sure the electrode tip is
submerged when you stir it.
8. Stir the probe gently in the sample to create a
homogenous sample. Allow a few seconds to reach
temp. equilibrium. Wait about 15 minutes can get
a stable reading.
9. The unit of conductivity will flash on the LCD to
indicate the meter is in measurement mode. When
the reading is stable, the unit will stop flashing.
(Fig1)
10. Press HLD to freeze display. Hold will appear
on the LCD. (Fig.2) Press HLD again to release.
11. TDS Measurement:
In measurement mode, press MODE to switch
to DS mode. The unit of conductivity are uS or mS.
The unit of TDS are ppm or ppt.
Set the TDS conversion factor to right value. The
factory default is 0.50 To change the TDS factors,
refer Appendix A.
12. Turn off the meter by pressing POWER key.
4
Setup
The advanced setup mode lets you customize your meter.
4 types parameter are available.
P1.0: temperature parameter setting (t)
P1.1: Change temperature unit (tUt)
P1.2: Normalization temp.(tnr)
P1.3: Temp. Coefficient (tCo)
P2.0: TDS factor setting
P2.1: Setting TDS factor (tds)
P3.0: Reset meter (rSt)
P3.1: Reset
P4.0: Review Calibration Info. (CAL)
P4.1: Review range 1 calibration information
P4.2: Review range 2 calibration information
5
P1.3: Temp. Coefficient (tCo)
In temp. coefficient adjustment, press qor p to
change the temperature coefficient from 0.0 to 4.0.
Press SET to confirm or press SET more than 2
sec to return to P1.0 without confirming.
P3.1: Reset
After executing, all parameters will be reset to factor
default. The previous calibration information will be
cleared as well.
In P3.0, press SET to enter P3.1. Press pto select
Y or N. Press SET to confirm or press SET more
than 2 seconds to return without confirming.
6
Calibration Mode (CAL)
Selecting a calibration standard
For best results, select a conductivity or TDS standard
near the sample value you are measuring. Alternatively,
use a calibration solution value which is approximate 2/3
of the full scale of the measurement range you plan to
use. For example, in the 0 to 1999 uS range, use 1413 uS
solution for calibration.
DO NOT reuse the calibration solution. Contaminants in
the solution will affect the calibration and the accuracy
Conductivity calibration
1. Dip the probe into demineralized or distilled
water for about 30 minutes to rinse the probe.
2. Select the conductivity standard for calibration.
3. Pour enough solution into two separate clean
containers.
4. Power on the meter. Select the mode as
conductivity measurement mode.
5. Rinse the probe into one of above containers.
Gently stir the probe.
7
6. Dip the rinsed probe into the second container.
Tap probe on the bottom of container to remove
air bubbles. Wait about 15 mins to make the probe
stabilize to the solution temperature.
7. Press CAL more than 2 seconds to begin
calibration. The conductivity value of solution will
blink on LCD.
8. Press p and q to change the value in
order to match the value to the standard
solution(solution must be referred to
normalization temp. 20oC if P1.2 is adjusted as
20 oC). You can adjust the conductivity reading
for +30%. However, if the measured value and
standard value differs by more than 30%, it is
suggested to clean probe or replace meter.
9. When the CAL stop flashing, you can
pressSET less than 1 second to confirm and
return to conductivity measurement mode. If
CAL always blinks, check if the calibration
solution is stable enough and if the step 8 input
value is correct or not.
10. Repeat 1~9 for other ranges if needed.
11. To exit conductivity calibration mode without
confirming, press SET in step 9 more than 2
seconds.
8
TDS Calibration
Option1: Using TDS standards
1. Insert the probe into demineralized or distilled
water for about 30 minutes to rinse the probe
2. Select the TDS standard for calibration. The
factory default setting of the TDS conversion
factor is 0.50. You can improve the calibration
accuracy by setting the TDS factor before starting
the calibration. Please refer Appendix A for more
information of TDS conversion factor.
3. Pour enough solution into two separate & clean
containers.
4. Turn on the meter. Press MODE to select TDS
mode.
5. Rinse the probe into one of the containers. Gently
stir the probe.
6. Dip the rinsed probe into the second container.
Tap the probe on the bottom of container to
remove air bubbles. Let the probe stabilize to the
solution temperature.
7. Press CAL more than 2 seconds to begin the
calibration. The TDS value will blink on the LCD.
8. Press the q to adjust the value to match the
value to the standard solution.
9. When CAL stop flashing, press SET to confirm
and return to TDS normal mode.
10. Repeat 1~9 for other ranges if needed.
9
SPECIFICATION
Range : 0~1999uS/0~1999ppm
or 0~19.99mS/0~19.99ppt
Resolution : 1uS/ppm or 0.01mS/ppt
Accuracy : 1% Full Scale1digit
TDS factor : 0.40~1.00
Calibration Standard : (0.2~1) * Full Scale
Range
ATC : 0~50C
Temperature Accuracy : 0.5C
Temp. Coefficient : 0~4.0% per degree C
Normalization Temp. : 20 or 25C
Operation temp. : 0~50C
Power Requirements : 4pcs1.5V(Type: A76 or LR44)
10
MAINTENANCE AND REPAIR
If there appears to be a malfunction during the
operation of the meter, the following steps should be
performed in order to isolate the cause of the problem.
1. Check the battery. Replace the battery
immediately when the symbol appears on the
LCD.
2. Review the operating instructions for possible
mistakes in operating procedure.
Except for the replacement of the battery, repair of the
meter should be performed only by a Factory Authorized
Service Center or by other qualified instrument service
personnel. The front panel and case can be cleaned with
a mild solution of detergent and water. Apply sparingly
with a soft cloth and allow to dry completely before
using. Do not use aromatic hydrocarbons or chlorinated
solvents for cleaning.
Make sure the electrode is clean! Between
measurements, rinse the electrode with deionised water.
If the eletrode has been exposed to a solvent immiscible
with water, clean it with a solvent miscible with water
e.g. ethanol or acetone and rinse carefully with water.
Store the electrode carefully! Before storing, rinse it
carefully in deionized water and store DRY.
11
BATTERY REPLACEMENT
1. Turn off the meter and loose the battery cover in
counter-clockwise direction.
2. Replace the old batteries with four new LR44
button cells.
3. Put back the battery cover and turn it tightly. the
Red LED Low Battery indicator illuminates. The
batteries should be replaced as quickly as possible.
12
TROUBLE SHOOTING
Power On But No Display
Make sure you press power key more than 100 mS.
Check the battery conditions and replace if
necessary
Move batteries away for one minute and then
re-install.
Display Disappear
Check whether the low battery icon is appeared
before the display is off. If yes, replace with new
batteries
Air Bubbles Adhere On Electrode
Stir the electrode completely and better to dip
the electrode into solution at oblique angle. After
soaking the electrode for 15~30 minutes, inspect
the electrode carefully to make sure no bubbles
adhere.
If air bubbles still exist, tap the bottom of the
container gently and stir the electrode to remove
the air bubbles. If above method are not working,
remove the electrode out of solution and blow at
the electrode to remove the air bubbles
Error Code
Parameter: Conductivity
- - - Meter is in manual range 1 however the
measured value is higher than 1999uS.
Press p more than 2 seconds to change
the range to range 2 or to auto ranging.
13
E03 Conductivity value is over the range limit
(19.99mS) or meter is damaged.
Put the meter in standard solution. If E03
still appears, send back for repair.
E04 Caused by temp. reading error
Refer to error code of temp. After solving
the error of temp, E04 will disappear.
Parameter: TDS
- - - Meter is in manual range 1 however the
measured value is higher than 1999*TDS factor
ppm
Press p more than 2 seconds to change
the range to range 2 or to auto ranging.
E04 Caused by temp. or conductivity error
Refer to error code of temp. After solving
the error of temp. & conductivity, E04 will
disappear.
Parameter: Temperature
E01 Temperature circuit is damaged
Send back for repair
E02 Temp value is lower than range limit (0C) or
temp. circuit is damaged
Put the meter in room temp. for 5 mins. If
E02 still appears, send back for repair.
E03 Temp value is higher than range limit (50C)
or temperature circuit is damaged
Put the meter in room temp. for 5 mins. If
E02 still appears, send back for repair
14
Appendix A:
Conductivity to TDS Conversion Factors
Conductivity TDS KCL TDS (NaCL) TDS (442)
at 25 C ppm Factor ppm Factor ppm Factor
23S 11.6 0.5043 10.7 0.4652 14.74 0.6409
84S 40.38 0.4807 38.04 0.4529 50.5 0.6012
447S 225.6 0.5047 215.5 0.4822 300 0.6712
1413S 744.7 0.527 702.1 0.4969 1000 0.7078
1500S 757.1 0.5047 737.1 0.4914 1050 0.7
2070S 1045 0.5048 1041 0.5029 1500 0.7246
2764S 1382 0.5 1414.8 0.5119 2062.7 0.7463
8974S 5101 0.5685 4487 0.5 7608 0.8478
12,880S 7447 0.5782 7230 0.5613 11,367 0.8825
15,000S 8759 0.5839 8532 0.5688 13,455 0.897
80mS 52,168 0.6521 48,384 0.6048 79,688 0.9961
Definitions
Actual TDS: Value from the solution bottle label or as a
standard you make using high purity water and precisely
weighed salts.
15
Actual Conductivity: Value measured using a properly
calibration Conductivity/TDS/Temperature meter.
Both the actual TDS and the actual conductivity values
must be in the same magnitude of units. For example, if
the TDS value in is ppm, the conductivity value must be
in S; if the TDS value is in ppt, the conductivity value
must be in mS.
Check this number by multiplying the conductivity
reading by the factor in the above formula and the
result is the TDS in ppm.
16
Linear temperature correction:
In moderately and highly conductive solutions,
temperature correction can be based on a linear
equation involving a temperature coefficient (q). The
coefficient is usually expressed as a conductivity variation
in % / C.
100
kTref kT
100 + q - (T-Tref)
Where:
k Tref = Conductivity at Tref
k T = Conductivity at T
Tref = Reference temperature
T = Sample temperature
q = Temperature coefficient
Note: the correction is accurate only within a limited
temperature range around T1 and T2. The greater the
difference between T and Tref, the higher the risk of
error.
17
T2 should be selected as a typical sample temperature
and should be approximately 10C different from T1.
The temperature coefficients of the following
electrolytes generally fall into the ranges shown below:
18