Agilent 1100 PLC Using Chemstation Software: Computer Start All Programs Agilent Chemstation Instrument 1 Online

This document provides instructions for using an Agilent 1100 PLC with Chemstation software to perform liquid chromatography. It details how to [1] turn on the instrument modules and open the software, [2] set up the pump, column thermostat, and other parameters to develop a method, [3] save the method and set up a sequence table to run samples, and [4] run the sequence and generate a report.

Uploaded by

Tendayi Samuriwo
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
88 views14 pages

Agilent 1100 PLC Using Chemstation Software: Computer Start All Programs Agilent Chemstation Instrument 1 Online

This document provides instructions for using an Agilent 1100 PLC with Chemstation software to perform liquid chromatography. It details how to [1] turn on the instrument modules and open the software, [2] set up the pump, column thermostat, and other parameters to develop a method, [3] save the method and set up a sequence table to run samples, and [4] run the sequence and generate a report.

Uploaded by

Tendayi Samuriwo
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 14

Agilent1100PLCusingChemstationSoftware

1.TurnonallthemodulesoftheAgilent1100.Alwaysturnmodulesonbeforeopeningsoftware.

2.TurnonComputer.Nopasswordisneeded

3.GounderStart,AllPrograms,AgilentChemstation,Instrument1online.ThefollowingMethod
andRunControlwindowshouldappear.

4.TurnonPumpbyclickingonGreysquareboxunderinstrumenticon.

Clickonsquarebox

5.AnewboxwillappearaskingTurnonallpumps.ClickYes.

6.TurnonColumnThermostatbyclickingoninstrumentcontrolwhichisgreysquarearrow.

Clickondownwards

arrow.

7.AnewwindowwillappearaskingtoTurnontheThermostat.ClickYes.

8.Ifyoualreadyhaveamethod,UnderMethod,chooseLoadMethodtochooseyourmethod.

9.ChooseyourmethodunderChem32/1/Methods.

9.Ifyouneedtocreateamethod,UnderMethod,chooseEditEntireMethodtocreateanew
method.

7.EditMethodwindowappearswhereyoucancheckoffpartsofmethodtobechanged.HitOK.

8.FillinMethodInformationandhitOK.

9.SetupPumpParametersWindow.FillinFlowrate,Solvents,CreateGradientorrunisocratic.
Youhaveuptofourdifferentsolventstouse.

a.UseInsertorAppendlinetocreategradient.HitOK.

b.UnderDisplay,youcanchangefromTimetabletoSolventstoseeyourgradient.

10.SetupInjectionParameters.Thereisa100lsampleloop.Youcaninjectfrom0.1to100l.
Thereisaneedlewashvialinholder91.Besurethereisorganicinthevial.HitOK.

11.DADSignals:Youcanchooseupto5differentwavelengths.HitOK.BWisforbandwidth.

12.ColumnThermostatparameters.Canprogramfromambientto80CorchooseNotControlled.
HitOK.

13.SignalDetailsWindow:AddtoMethodallyourdifferentwavelengths.HitOK.

14.EditIntegrationEvents.HitOK.

16.SetSpecificationforReport.HitOK.


17.TheInstrumentDataCurvescanbeoverlaidwithyourspectrum,likepercentcomposition.Check
offwhatyouwanttoseeinyourspectrum.HitOK.

18.RunTimeChecklist.HitOK.

19.UnderMethod,ChooseSaveMethodAs.BesuretosaveunderMethods[c:/chem32/1/methods]

20.UnderView,chooseOnlineSignals1.ClickonChangetoaddorremoveOnlinesignals.

21.UnderSequence,ChooseSequenceTable.

a. UnderVial,enterthelocationofthevial.(Youdontneedtotypevial)
b. UnderSample,givenameandcanadddescriptionofsampleinSampleinfojustabove.
c. UnderMethodName,choosenameofmethod.
d. UnderInjection/locationplacea1for1injection.Ifyouwantmultipleinjectionsofsame
sample,youcanadd2,3,4,etc.
e. UnderSampletype,youcanchooseSample,calibrationorcontrol.
f. UseInsertorAppendLinetoaddmoreinjections.

22.UnderSequence,chooseSaveSequenceTemplateas.Saveasyournameandthedate

23.Opensequencetableagain.HitRunSequencewhenready.

RunSequence

24.YoucanalwayspresstoSTOPbuttontostoptherunatanytime.

25.Areportwillautomaticallybegenerated.HitPrintandsavereportaspdf.

26.ShutdownChemstation,thenallthemodules.