PS70299 ICS 1100 Product Specifications

Download as pdf or txt
Download as pdf or txt
You are on page 1of 7

C H R O M ATO G R A P H Y

Produ c t Spe ci f i cat i ons


Thermo Scientific
Dionex ICS-1100
Ion Chromatography System

The preconfigured Thermo Scientific Dionex ICS-1100


and ICS-1600 are the first integrated Reagent-Free IC
systems with Eluent Regeneration (RFIC-ER systems). The
Dionex ICS-1000 system is designed to perform isocratic
ion chromatography (IC) separations using conductivity
detection and standard bore (4 mm) and microbore (2mm)
columns. Multiple available valving configurations support
many modes of automated sample preparation. When
configured as an RFIC-ER system, the Dionex ICS-1100 allows
continuous operation for up to four weeks using a single
eluent preparation. Thermo Scientific Dionex Chromeleon
Chromatography Data System provides full control and
digital data collection from a PC using USB high-speed
communication protocol. Available options include column
heating, in-line vacuum degassing, and RFIC-ER installation kit.

Versatility Flexible flow rate supports 2, 3, 4, and


Performs isocratic IC separations using 5 mm column formats.
conductivity detection. RFIC-ER system technology provides up to
Integrated, preconfigured, factory 28 days of chromatography results using
plumbed, and tested for immediate only a single eluent preparation.
productivity. Automated sample preparation capabilities
Streamlined design with small footprint enable techniques such as on-line
occupies minimal bench space. filtration, concentration, and matrix
Dual-piston pump design reduces elimination.
pulsations, allowing high-sensitivity
detection and excellent flow-rate accuracy
and precision.
2 Simple and Precise Control curves remain valid over the entire period. chromatography.
Built-in control for Thermo Scientific For improved reproducibility, the Eluent Regeneration
Dionex SRS Self-Regenerating thermostated high-performance When configured as a Reagent-Free IC
Suppressor and Atlas electrolytic conductivity detector permits system with Eluent Regeneration (RFIC-ER
suppressors. Electrolytic suppression with measurements that are unaffected by system), the Dionex ICS-1100 allows the
an AutoSuppression device eliminates temperature variation. use of a single preparation of eluent for up
the need to hand-prepare acid or base
Advanced single-range digital output with to four weeks.
regenerants. Suppression reduces
operating range to 15,000 S full scale, The RFIC-ER system uses the Dionex
background conductivity and provides
with autoranging to provide accurate SRS300 self-regenerating suppressor to
high signal-to-noise ratios.
detection of major and minor constituents simultaneously regenerate returning eluent
Full control and digital data collection in a single run. Single-range analog signal as it suppresses eluent before detection.
available with Windows-based output is also standard.
Chromeleon SE Chromatography Trap and catalytic columns purify returning
Optional column heater provides eluent, assuring consistent, high quality
Workstation software using USB
day-to-day consistency, ensuring eluent for separations.
high-speed communication protocol.
reproducibility and stability. Preheating of
Chromeleon eWorkflows preload all Because it is a closed loop, the always
the eluent prior to the column maintains
instrument parameters for fast and easy on, always ready RFIC-ER system remains
the column temperature set by the user.
operation and data analysis. equilibrated and calibrated between eluent
A transparent cover allows viewing of the
changes; up to four weeks.
Application templates preload all instrument column without temperature disruption.
parameters for fast and easy operation. Less frequent eluent preparation reduces
Optional built-in vacuum degas provides
unintentional variations in concentration,
Chromeleon software control includes an in-line degassing of eluents, ensuring
increasing reliability and reproducibility.
electronic logbook for monitoring of nearly reproducibility and protection of eluents from
unlimited user-selectable operational contamination and decomposition. Control RFIC-ER systems are designed for high
parameters. of the degas operation can be automated to throughput analyses of anions or cations
sense when degassing is required. in low- to moderate-concentration
High Performance matrices without sample preparation,
Eluent regeneration provides extremely Inert, nonmetallic PEEK components
or in high-concentration matrices after
stable baselines, day to day, for up to 28 throughout the system ensure
days of continuous operation. Calibration compatibility with corrosive eluents
and provide metal-contamination-free

Column: Thermo Scientific Dionex IonPac AS22, 4 mm 20 F


Eluent: 4.5 mM carbonate/ 1.4 mM bicarbonate Cl
Eluent Source: Regenerated carbonate/bicarbonate NO2
Flow Rate: 1.2 mL/min Br
Inj. Volume: 50 L NO3
10
Temperature: 30 C PO4
Detection: Suppressed conductivity, Thermo Scientific Dionex SO4

}
5
ASRS 300, 4 mm AutoSuppression recycle mode, 31 mA
Peak Area Change (%)

Sample: Seven anion standard


0 Pass

8
Peaks: -5
2
1. Fluoride 0.5 mg/L
2. Chloride 2.5 -10
3. Nitrite 2.5
4. Bromide 2.5
5. Nitrate 2.5
6. Phosphate 5.0 -20
7. Sulfate 2.5 1 6 11 16 21 26 28
3
Days
S 1
5 Figure 2. Graph of peak area changes for a seven-anion calibration check
4
7 standard run daily for 28 days on an RFIC-ER system using a single 4 L
6
preparation of eluent. The system passed for the entire four weeks without
reequilibration or recalibration.
0

-3
0 5 10 15
Minutes

Figure 1. Overlay of chromatograms from a representative week of the seven


anion calibration check standard runs on an RFIC-ER system using a single 4 L
preparation of eluent. The peak retention times demonstrate high reproducibility.
matrix elimination. 3

Convenience Small Loop


RFIC-ER system technology reduces
eluent preparation to once every 28 days.
Versatile eluent organizer tray
accommodates 1, 2, or 4 liter 6
eluent bottles. Waste
Standard 7 5
Electrically actuated six-port Rheodyne
8 4
PEEK injection valve for precise sampling. 9
3
Ergonomically placed injection port for
easy manual sampling. 10 1
2
Eluent valve provides positive shut- off
of eluent flow prior to the pump for easy
servicing.
Easy-access door to chromatography
components.
Leak detection and management allow Large Loop
fast response to system leaks.
TTL controls for external pump, injection
valve, range selection, and signal offset
Sample
for stand-alone operation.

Automated Sample Preparation Detector


Optional 6- or 10-port valves support Waste
automated sample preparation.
The 6-port valve supports techniques
such as matrix elimination, sample
concentration, and on-line filtration. CRD
Waste
The 10-port valve supports Auto-Dilution
using a small loop and large loop to
reinject out-of-range samples Pump Suppressor

6- or 10-port valves can be used for


matrix diversion prior to MS detection.

Key Features 4
5
Eluent regeneration functionality built in Analytical
Column
Dual-piston pump
6
Electrolytic suppression 3
Digital conductivity detection
Guard
USB connectivity, plug-n-play 1
2 Column

Optical leak detector Concentrator

Electronic logbook and trending


Optional 6- or 10-port valve for automated
sample preparation
Figure 3. Optional 6- or 10-port valves support automated, on-line sample preparation techniques, such
as sample concentration and matrix elimination, or AutoDilution using large and small injection loops with a
10-port valve, shown above.
4
Dionex ICS-1100 IC System Specifications
Analytical Pump and Fluidics
Type: Serial dual-reciprocating pistons, microprocessor-controlled constant stroke, variable speed
Construction: Chemically inert, metal-free PEEK pump heads and flow paths compatible with aqueous eluents of
pH 014 and reversed-phase solvents
Pump Operating Pressure: 035 MPa (05000 psi)
Flow Rate Range: 0.005.00 mL/min without changing pump heads
Flow Precision: <0.1%, typically
Flow Accuracy: <0.1%, typically
Pressure Ripple: <1% at 13.8 MPa (2000 psi) and 1.0 mL/min
Eluent On-Off Valve: Standard
Piston Seal Wash: Dual-pump head, wash can be continuous when connected to rinse solution supply
Pressure Alarm Limits: Upper limit 035 MPa or 05000psi in one unit (MPa or psi) increments; lower limit can be set up
to one unit lower than upper limit
Vacuum Degas: Yes, optional, automatic control
Eluent Bottles: Polypropylene, up to 4 L volume
Eluent Bottle Pressure: Not required
Injection Valve: 6-port, 2-position Rheodyne valve, electrically activated
Columns Supported: 2, 3, 4, and 5 mm i.d.; maximum length 250 mm analytical column with 50mm guard column
Column Heater (Optional)
Operating Temperature Range: 30 to 60 C (86 to 140 F); minimum 5 C above ambient; Settable range is equal to working range
Temperature Accuracy: 0.5 C at sensor, at 40 C
Auxiliary Valve (Optional)
Available Valves 2-position, 6- or 10-port high-pressure Rheodyne valve, fully inert PEEK construction, electrically activated
Eluent Generation
Optiona Thermo Scientic Dionex RFC-30
Eluent Regeneration
Eluent Regeneration Support: Yes, with optional RFIC-ER kit
Eluents: Carbonate and carbonate/bicarbonate up to 20 mM MSA up to 34 mM
Flow Rates: 0.012.00 mL/min
Continuous Operation (4 L of Eluent): Up to 28 days or 2000 samples, typically
Always On, Always Ready Capable: Yes, standard feature
Remains Fully Calibrated for Extended Yes, standard feature. results are traceable to a single calibration
Periods (<28 days):
System Wellness: Consumables usage monitoring for predictive maintenance
Maximum Operating Pressure: 21 MPa (3000 psi)
Operating Temperature Range: 440 C
5
Dionex ICS-1100 IC System Specifications (CONTD)
Suppressors and Control
Chemical Suppression: 2 mm and 4 mm anion and cation, membrane suppression bed types
Displacement Chemical Regeneration: 2 mm and 4 mm anion and cation membrane suppression bed types
Electrolytic SuppressionSelf-Regenerating: 2 mm and 4 mm anion and cation; both membrane and MonoDisk suppressionbed types available
Electrolytic SuppressionSelf-Regenerating 2 mm and 4 mm anion and cation; both membrane and MonoDisk suppression bed types available
with External Water Mode:
Current ControlRange: Dionex SRS: 4 mm, 0300 mA in 1 mA increments
2 mm, 0100 mA in 1 mA increments
Dionex AES Atlas Electrolytic Suppressor: 0150 mA in 1 mA increments
Thermo Scientific Dionex CMD Carbohydrate Membrane Desalter: 0500 mA in 1 mA increments
Thermo Scientific Dionex SRN Self-Regenerating Neutralizer: 0500 mA in 1 mA increments
Salt Converter: Available in 2 and 4 mm versions
Thermo Scientific Dionex AMMS-ICE: Available in 2 and 4 mm versions
Carbonic Acid Removal for Anions: ASRS 300 with Thermo Scientific Dionex CRD 200 Carbonate Removal Device for hydroxide eluents
Dionex ASRS 300 with Dionex CRD 300 for carbonate eluents
Non-Suppressed Chromatography: Yes, supported
Suppressor Wear Parts: None; peristaltic pump and inline filters not required
Suppression Capacity: Anions:
Dionex ASRS 300 (4 mm): 200 eq/min
Dionex ASRS 300 (2 mm): 50 eq/min
Thermo Scientific Dionex AMMS 300 Anion MicroMembrane Suppressor (4 mm): 150 eq/min
Dionex AMMS 300 (2 mm): 37.5 eq/min
Dionex AAES: 25 eq/min
Cations:
Thermo Scientific Dionex CSRS 300 Cation Self-Regenerating Suppressor (4 mm): 110 eq/min
Dionex CSRS 300 (2 mm): 37.5 eq/min
Thermo Scientific Dionex CMMS 300 Cation MicroMembrane Suppressor (4 mm): 150 eq/min
Dionex CMMS 300 (2 mm): 37.5 eq/min
Thermo Scientific Dionex Cation Atlas Electrolytic Suppressor: 25 eq/min
Void Volumes:
Dionex SRS 300 (4 mm): <50 L
Dionex SRS 300 (2 mm): <15 L
Dionex MMS 300 (4 mm): <50 L
Dionex MMS 300 (2 mm): <15 L
Dionex AMMS-ICE 300 (4 mm): <50 L
Dionex AMMS-ICE 300 (2 mm): <15 L
Dionex AAES: <35 L
Dionex CAES: <35 L
Conductivity Detector Electronics and Flow Cell
Type: Microprocessor-controlled digital signal processor
Cell Drive: 8 kHz square wave
Linearity: 1% up to 1 mS
Resolution: 0.00238 nS/cm
Full-Scale Output Ranges: Digital signal range 015000 S
Analog signal range 015000 S
Electronic Noise: 0.1 nS when background conductivity is 0150 S/cm
2 nS when background conductivity is 1513200 S
Filter: Rise times from 0 to 10 s, user selectable
Temperature Compensation: Fixed at 1.7% per 1 C at cell temperature
Temperature Range: Ambient +7 C, 30 to 55 C
Cell Electrodes: Passivated 316 stainless steel. Compatible with MSA
Cell Body: Chemically inert polymeric material
Cell Volume: <1 L
Heat Exchanger: Inert, tortuous path for low axial dispersion
Maximum Cell Operating Pressure: 10 MPa (1500 psi
6
Dionex ICS-1100 IC System Specifications (CONTD)
Autosampler
Automation Using Autosampler: Thermo Scientific Dionex AS-DV, AS-AP, AS-HV, or third-party autosamplers
Sequential/Simultaneous Injection: Yes, depending on autosampler capabilities
Automated Dilution: Yes, available with Dionex AS-AP Autosampler
Dilution Factor, Dionex AS-AP Autosampler: 1:1 to 1:1000
Dilution Time, Dionex AS-AP Autosampler: 15 seconds with sample overlap
Inline Sample Degassing: Yes, optional with Dionex CRD 300/200
Inline Filtration: Yes, AS-DV Autosampler or inline filter
High Automation Flexibility: Conditionals using Chromeleon and post run features
System Software
Chromeleon Chromatography Management Software, supports Windows XP or Vista:
Automated Procedure Wizards
System Wellness and Predictive Performance
Data trending plots (numerical device parameters)
Virtual Column Simulator (evaluation mode standard, isocratic and gradient optional)
Application templates
Multivendor automation support of 3rd party instruments (fully controls over 300 instruments from more than 30 manufacturers, including GC,
HPLC, and MS)
3-D Software for photodiode array, mass spectrometer, and electrochemical detectors (optional)
Customizable System Control Panels
System Status Virtual Channels
Power Failure Protection
Sequential Injection
System Trigger Commands and Conditionals
Daily Audit Trail
Sample Audit Trail
Multiple Network Control and Network Failure Protection (optional)
System Calibration Storage (factory, present, and previous; completely user selectable)
Customized Reporting (unlimited report workbooks)
Automated System Qualification (detailed, comprehensive qualification reports)

Physical Specifications
Power Requirements: 100240 V ac, 50-60 Hz autoranging
Operating Temperature: 440 C (40104 F); cold-room-compatible (4 C) as long as system power remains on
Operating Humidity Range: 595% relative, noncondensing
Control Modes: Full control through Chromeleon software; alternative control through TTL or relay closures; two
relay outputs, two TTL outputs, four programmable inputs
USB Communication Protocol: One USB input; one built-in two-output USB hub
Leak Detection: Built-in, optical sensor
Dimensions (h w d): 56.1 cm 22.4 cm 53.3 cm
(22.1 in 8.8 in 21 in)
Weight: 24.5 kg (54 lb)
Ordering Information
To order in the U.S., call (800) 346-6390 or contact the Thermo Scientific office nearest you. Outside the U.S., order through your local

Produ c t Spe ci f i cat i ons


Thermo Scientific office or distributor. Refer to the following part numbers.

Dionex ICS-1100 Ion Chromatography System with Software and PC Part Number
Dionex ICS-1100 Ion Chromatography System with Chromeleon SE and Windows XP Workstation, without Degas: 069648
Dionex ICS-1100 Ion Chromatography System with Chromeleon SE,Windows XP Workstation, and Degas: 069649
Dionex ICS-1100 Ion Chromatography System with Chromeleon SE, without Windows XP Workstation, or Degas: 069650
DionexICS-1100 Ion Chromatography System with Chromeleon SE and Degas, without Windows XP Workstation: 069651
Optional Column Heater: 069564
RFIC-ER Anion Startup Kit. Includes anion installation kit and anion consumables: 069570
RFIC-ER Cation Startup Kit. Includes cation installation kit and cation consumables: 069569

A Dionex ICS-1100/Chromeleon SE/Windows Workstation bundled package includes: a Dionex ICS-1100 with isocratic dual-piston pump,
injection valve, heated conductivity cell, USB cable, Chromeleon SE , computer (with Windows XP), and USB dongle. Chromeleon SE comes
with one SE timebase controlling one Dionex ICS-1100 system. The Dionex ICS-1100 is supplied without a front control panel, and must be
controlled through Chromeleon software. Consumables must be ordered separately.

www.thermoscientific.com/dionex
2013 Thermo Fisher Scientific Inc. All rights reserved. ISO is a trademark of the International Standards Organization. PEEK is
Thermo Scientific Dionex products are
a trademark of Victrex PLC. Windows is a registered trademark of Microsoft Corporation. All other trademarks are the property designed, developed, and manufactured
of Thermo Fisher Scientific Inc. and its subsidiaries. Specifications, terms and pricing are subject to change. Not all products are under an ISO 9001 Quality System.
available in all countries. Please consult your local sales representative for details.

Australia +61 3 9757 4486 Denmark +45 70 23 62 60 Japan +81 6 6885 1213 Switzerland +41 62 205 9966
Austria +43 1 333 50 34 0 France +33 1 60 92 48 00 Korea +82 2 3420 8600 Taiwan +886 2 8751 6655
Belgium +32 53 73 42 41 Germany +49 6126 991 0 Netherlands +31 76 579 55 55 UK/Ireland +44 1442 233555
Brazil +55 11 3731 5140 India +91 22 2764 2735 Singapore +65 6289 1190 USA and Canada +847 295 7500
China +852 2428 3282 Italy +39 02 51 62 1267 Sweden +46 8 473 3380

PS70299_E 01/13S