EN
technical manual
10030406_12
Table of contents
Chapter F: ADJUSTMENTS
1 MECHANICAL ADJUSTMENTS ....................................................................................... 163
1.1 Adjusting vertical position of OP delivery arm column ............................................................. 163
1.2 Adjusting friction of OP delivery arm joints .............................................................................. 164
1.3 Adjusting balance of OP delivery arm ...................................................................................... 165
1.4 Adjusting angle of OP delivery arm instrument console .......................................................... 171
1.5 Adjusting side delivery arm base joint friction .......................................................................... 172
1.6 Adjusting friction of side delivery arm joint ............................................................................... 173
1.7 Adjusting side delivery arm instrument console friction ........................................................... 173
1.8 Adjusting lifting friction of side delivery arm ............................................................................. 174
1.9 Adjusting tension of side delivery arm gas spring .................................................................... 175
1.10 Adjusting friction of adjustable suction arm ............................................................................. 176
1.11 Removing play of lifting adapter .............................................................................................. 177
2 ELECTRICAL ADJUSTMENTS ........................................................................................ 178
2.1 Calibrating touch panel ............................................................................................................ 178
2.2 Calibrating backrest motor position sensor .............................................................................. 179
2.3 Calibrating lift motor position sensor ........................................................................................ 181
2.4 Rewiring mains voltage selection jumper ................................................................................ 183
3 PNEUMATICAL ADJUSTMENTS ..................................................................................... 184
3.1 Adjusting internal water & air pressures .................................................................................. 184
3.2 Adjusting syringe water and air flow rates ............................................................................... 185
3.3 Adjusting air and water to instruments ..................................................................................... 185
4 BLEEDING OF SPRAY WATER SYSTEM ........................................................................ 186
5 CALIBRATING SUCTION TUBE CLEANING SYSTEM FEEDING TIME ............................ 187
6 ADJUSTING WATERLINE CLEANING SYSTEM WATER PRESSURE .............................. 188
Chapter I: DIAGRAMS
The manufacturer, assembler and importer are responsible for the safety, reliability and
performance of the unit only if:
- installation, calibration, modification and repairs are carried out by qualified author-
ised personnel
- electrical installations are carried out according to the appropriate requirements such
as IEC 60364
- equipment is used according to the operating instructions.
COPYRIGHT PLANMECA
Publication number 10030406 Revision 12
Released 12 December 2014
1 INTRODUCTION
Planmeca Compact i dental unit is an electrically controlled dental device that consists of a
patient chair, cuspidor, delivery arm, dental instruments and an operating light. Planmeca
Compact i is meant to be used for dental treatment by dental care professionals. Service
and maintenance must always be performed by a qualified Planmeca service technician.
This manual contains instructions for operation, maintenance, adjustments, as well as
parts replacement and repair. Please read this manual carefully before servicing the unit.
Type BF applied parts for specified dental instruments (Standard IEC 60878).
Separate collection for electrical and electronic equipment according to Directive 2002/96/
EC (WEEE).
To avoid risk of electric shock, this equipment must only be connected to a supply mains
with protective earth.
Note that the mains voltage is always present at the mains terminal under the cover, when
the unit is switched on. Do NOT open the cover.
3 ASSOCIATED DOCUMENTATION
Planmeca Compact i Touch comes with the following manuals:
Users manual, publication number 10030405
For dental care professionals. Describes the dental unit and its different parts as well
as instructs how to operate and clean the dental unit.
Installation manual, publication number 10030407
For service personnel. Describes how to install the dental unit.
Technical manual, publication number 10030406
For service personnel. Describes the touch panel functions, the various service
modes as well as help and error messages. Contains instructions for preventive
maintenance, calibrations, adjustments and parts replacement and repair. Wiring and
pneumatical diagrams are found at the end of the manual.
4 TRAINING
A hands-on users training is given in connection with the dental unit installation.
WARNING
The unit contains live mains voltage parts. Always first turn off the power to the unit
externally before attempting to service the lift-motor, backrest motor, or any parts
inside the electronics control box.
The power supply must be disconnected externally from, for example, a fuse or a
mains switch. The fuse or main switch must be lockable into off-position.
Turning off the unit from its own mains switch DOES NOT cut off the mains voltage
from all internal nodes.
WARNING
No modification of this dental unit is allowed.
WARNING
Do not modify this equipment without authorisation of the manufacturer.
If this equipment is modified, appropriate inspection and testing must be conducted
to ensure continued safe use of equipment.
CAUTION
ELECTRICAL SAFETY
Always replace the fuses with fuses recommended by Planmeca in section 4.3
"Replacing fuses" on page 211. Otherwise patient, operator or equipment safety
cannot be guaranteed. Please see the fuse label on the left side of the control box
for fuse ratings.
CIRCUIT BOARDS
The circuit boards in the unit can be damaged by static electricity and require spe-
cial handling and antistatic packaging materials to be used both during handling
and shipping. See section 8 "HANDLING PRECAUTIONS FOR PCBS AND SD-
CARDS" on page 28 for handling details.
CAUTION
ELECTRICAL SAFETY
Medical electrical equipment needs special precautions regarding electromagnetic
compatibility (EMC) and needs to be installed and put into service according to the
EMC information provided in the accompanying documents. Portable and mobile
RF communications equipment can affect medical electrical equipment.
CAUTION
GENERAL SAFETY
The unit must be serviced by qualified personnel only, trained by PLANMECA or
approved dealers. Repairs and parts replaced by unqualified personnel carry no
warranty.
Periodic maintenance as described in this manual must be performed on a regular
basis.
CAUTION
Do not use the equipment in close conjunction with anaesthetic gas or in highly oxy-
genated environments (oxygen content >25%).
NOTE
The PC installed to the system must fulfil IEC 60950 requirements.
NOTE
The monitor installed to the system must fulfil IEC 60950 requirements.
6 MANUAL VERSIONS
Planmeca pursues a policy of continual product development. Although every effort is
made to produce up-to-date product documentation, this publication should not be
regarded as an infallible guide to current specifications. We reserve the right to make
changes without prior notice.
NOTE
This manual is only valid for the Main Control PCB software version 7.0 and later.
The installed software version determines the available functions and behaviour of the
unit. If the unit is reporting error or help messages not listed in this manual, please refer to
a newer revision of the manual.
If you find service modes described in this manual that does not exist in the unit, please
update the unit software to the latest version.
7 TECHNICAL SPECIFICATIONS
Original manufacturer
PLANMECA Oy, Asentajankatu 6, 00880 Helsinki, FINLAND
Phone: +358 20 7795 500, fax: +358 20 7795 555, www.planmeca.com
Colours
Painted parts: RAL-9016
Upholstery colours: Please consult your dealer for availability
Mechanical dimensions
Installed: (H x D x W) 1210mmx963mmx1990mm
(see template for details)
Weight
130 kg (287 lbs)
Environmental conditions
Transportation conditions
Accepted limits for unit transportation are the following:
Temperatures: -20 C to +60 C (-4 F to 140 F)
Relative humidity: 5% RH to 95% RH; non-condensing humidity
Air pressure: 700 hPa to 1060 hPa (10 psi to 15 psi)
Storage conditions
Accepted limits for unit storage are the following:
Temperatures: -5 C to +60 C (23 F to 140 F)
Relative humidity: 5% RH to 95% RH; non-condensing humidity
Air pressure: 700 hPa to 1060 hPa (10 psi to 15 psi)
NOTE
If the unit has been stored at temperatures below +10 C (50 F) for more than few hours,
time must be allowed for the unit to reach the room temperature in its own package, before
connecting the dental unit to mains voltage.
Power consumption
Idle unit: 60 VA (unit not in use, OP-light turned on)
Typical average: 350 VA (during patient treatment)
Maximum: 1450 VA (at maximum rated mains voltage, both chair motors
running)
Electrical classification
Class I, type B
Water supply
Pressure range: min. 300kPa (45 psi), max. 900 kPa (130 psi)
Flow rate: 4 l/ min (maximum consumption at any instance)
Quality: hardness; 8 dH (1dH= 20mg Ca/ 3 liter water)
Connection: 1/4
Air supply
Pressure range: min. 550 kPa (80 psi), max. 900 kPa (130 psi)
Flow rate: 55 litres / minute (maximum consumption at any instance)
Air quality: medical grade, dry and oil-free
Connection: 1/4
Suction connection
Vacuum: 150 mbar
Flow rate: 550 l/ min
Connection: 25 mm (3/4)
Drain connection
Capacity: min. 10 l/min
Connection: 25 mm (3/4)
7.1 Dimensions
2 3 4 5 9 6 7
11
10
7.3 Labels
The following figures show all the labels and their positions.
Headrest
The following detachable components are not critical to operation. The user can perform
dental treatment even if a wrong, similar part is fitted.
Hygienic membrane
Armrests
Flushing holder
SC_UM_33.eps
Suction hoses
SC_UM_47.eps
9 EMC INFORMATION
WARNING
Use of any accessories and cables other than those specified in Planmeca Com-
pact i dental unit documentation, with the exception of cables sold by Planmeca as
replacement parts for internal components, may result in increased emission or
decreased immunity of the dental unit.
WARNING
Planmeca Compact i dental unit should not be used adjacent to or stacked with
other equipment. If adjacent or stacked use is necessary, the Planmeca Compact i
dental unit should be observed to verify normal operation in the configuration in
which it will be used.
IEC 61000-3-3
NOTE 1: At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects and people.
a Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless)
telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast
cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due
to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured
field strength in the location in which Planmeca Compact i dental unit is used exceeds the
applicable RF compliance level above, Planmeca Compact i dental unit should be observed to
verify normal operation. If abnormal performance is observed, additional measures may be
necessary, such as re-orienting or relocating Planmeca Compact i dental unit.
b
Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.
1 COMPACT i TOUCH V2
When you make changes to, for example, the dental unit or instrument settings, you are
required to save the changes by pressing OK. If you do not want to save the changes, you
can close the window without saving the changes by pressing Cancel.
In case of malfunction, an error code or a help message is displayed on the touch panel,
see Chapter D: "HELP & ERROR MESSAGES" on page 123.
The picture below shows the main functions of the touch panel when the dental unit is in
idle state and no instrument is activated.
1 9
10
2
3
4
11
12
7
CP4_display_Touch_v1_num_002.eps
1 2
flexy_panel.eps
3
1 Chair buttons
2 Unit buttons
3 Flexy button
The following information is displayed in the About this unit window. Each item is
described below.
NOTE
The following is an example only and does not necessarily portray the actual situation.
1.5 Service
Service contains information for service situations.
Press > to display the following items:
Yearly Maintenance
GUI Diagnostics
NOTE
In the Planmeca Compact i Classic v2 dental unit, the information about the yearly
maintenance can be configured in service modes 400, 401 and 402.
Touchpanel test
Lets you test where the screen registers your touch and if there are any areas that do
not register your touch.
A black screen with buttons Drag, Draw and Quit opens.
Press Draw to draw something on the screen.
Press Drag to drag the drawing on the screen.
Press Quit to quit the test.
Touch the screen to see where your finger registers on the screen.
Display ghosting test
Lets you test screen burn.
Screen test
Lets you test how the touch screen functions by displaying different test images.
NOTE
The following is an example only and does not necessarily portray the actual situation.
You can scroll the list either from the list itself, from the scroll bar to the right, or by using
the Up and Down buttons
Romexis connection
Toggle the arrow button to enable/disable the connection to Planmeca Romexis.
When the button is grey, the connection is disabled. When the button is blue, the
connection is enabled.
When the connection is enabled, the Romexis symbol can be seen on the control
panel.
Those settings that can be changed have an Edit button next to them. Press the Edit
button to open a new window where you can edit the setting. You will be prompted to enter
the PIN-code 1701.
When you have made all necessary changes, press OK to save the changes and reboot
the unit.
For more information on Planmeca Romexis Clinic Management, see the Planmeca
Romexis users manual.
The message history window can be filtered by message type. In the picture above, all
types are listed:
Help messages
Error messages
You can delete the message history by pressing the Delete message history button. A
confirmation message is displayed. Confirm the deletion by pressing OK.
You can save the message history to a USB memory stick. Insert a USB memory stick to
the USB port on the cuspidor marked user settings and service and press the Save to
USB button.
1.5.6 Licences
Licences displays the licences used by the dental unit when you press >.
NOTE
All programming is performed from the touch panel.
There are several adjustments the user can perform to fulfil individual needs by using the
user programming mode.
Automatic chair positions, light modes for the operating light and instrument settings are
saved to the user profile. Before you start programming, make sure that you have signed
in to the dental unit with the correct user number. The current user is displayed on the
screen.
The local users whose settings are stored in the dental unit have numbers 1 - 4. The
Romexis user is indicated by a tooth symbol.
NOTE
If you have signed in as a Romexis user and want to save the changes you have made to
dental unit or instrument settings, you must save the changes to the Romexis user profile
from the Planmeca Romexis software before shutting down the dental unit or before
another user signs in to the dental unit. For instructions, see chapter Clinic module in the
Planmeca Romexis users manual.
NOTE
Alternatively, you can change the settings with the foot control. To increase the
parameter value, push the pedal to the right. To decrease the parameter value, push
the pedal to the left. When the pedal is pushed halfway (right or left) the parameter
changes slowly. When the pedal is pushed to its extreme position (right or left) the
parameter changes quickly.
When you change instrument settings with the foot control, the instrument runs
normally during programming.
5. Press the OK button to store the new setting into memory. The pop-up window closes.
NOTE
The maximum air and water flow rate of the syringe can not be changed via the touch
panel. However, there are screwdriver-adjustable valves inside the console that can
be adjusted if desired. Refer to section 3.2 "Adjusting syringe water and air flow rates"
on page 185.
1. Press Program.
2. Press the desired chair position button where to store the chairs position (rinsing
position, A, B, C or D).
If you want the operating light to be on (off) in this position, turn it on (off). Note that if
you are programming the rinsing position, you can not select the operating light
setting.
3. Press OK to confirm that you want to save the current position as an automatic
position.
If the dental unit is equipped with the Planmeca Sterile Water System or Steripump, the
sterile water function is selected in this mode.
1. Press Program.
2. Press Instrument spray. A pop-up window opens.
3. Press Sterile water.
4. Press OK to exit and store the new settings.
The Sterile water symbol indicates that sterilised water is fed to the selected instrument.
The indicator light of the instrument spray button flashes green when the spray is on. For
more information, refer to the Sterile Water System Users Manual.
NOTE
The manual chip blow is set as default to be dry (air only), independent of the automatic
chip blow programming. However, if desired, this can be altered, please refer to "n.43 Set
mode of foot controlled chip-blow" on page 88.
1. Press Program.
2. Press RPM limit. A pop-up window opens.
3. Change the micromotor speed rate (rpm) with the plus (+) and minus (-) buttons.
The speed/power limit can be altered as follows:
5. Press Program.
6. Press Torque. A pop-up window opens.
7. Change the torque limit rate with the plus (+) and minus (-) buttons.
The adjustment range is 10% - 95%.
8. Press OK to exit and store the new settings.
9. Press Program.
10. Press Auto forward to adjust the counter-clockwise rotation time of the micromotor in
the Auto forward mode. A pop-up window opens.
11. Change the Auto forward time with the plus (+) and minus (-) buttons.
The maximum time is 6 seconds and the minimum 0 seconds, and the time can be
changed in steps of 0.2 seconds.
12. Press OK to exit and store the new settings.
Press Instrument spray to switch the instrument spray on or off. The amount of water and
air can also be programmed, see 1.7.3 "Instrument spray flow rates" on page 46.
Press Chip blow to switch the automatic chip blow on or off. The amount of water and air
can also be programmed, see 1.7.4 "Chip blow flow rates" on page 46.
When you have modified the settings of a preset, but have not saved them, a dash ()
replaces the preset number on the display.
Polishing water
1. Press Program.
2. Press Polishing water. A pop-up window opens.
3. Change the water flow rate with the plus (+) and minus (-) buttons.
The minimum value is 0 (no flow) and maximum 100. The adjustment step is 1 in the
value range 0 - 35% and 5 in the value range 35 - 100%.
4. Press OK to exit and store the new settings.
Cleaning water
1. Press Program.
2. Press Cleaning water. A pop-up window opens.
3. Change the water flow rate with the plus (+) and minus (-) buttons.
The minimum value is 0 (no flow) and maximum 100. The adjustment step is 1 in the
value range 0 - 25% and 5 in the value range 25 - 100%.
4. Press OK to exit and store the new settings.
NOTE
Cup filling is not activated unless the cup is positioned in its place under the cup fill tube.
Light modes
Enable and disable the light modes as follows:
1. Press Program.
2. Press Operating light. A pop-up window opens.
3. Press Light modes to enable/disable the light modes.
Blue button = light modes are enabled, press to disable.
Grey button = light modes are disabled, press to enable.
4. Press OK to exit and store the new settings.
NOTE
When the light modes are disabled, the intensity of the operating light can also be adjusted
as follows:
You can adjust the intensity of the operating light during treatment by pressing and holding
the Operating light button for longer than one second.
You can also adjust the intensity during treatment by keeping your hand in front of the
infrared light sensor of the operating light for longer than one second.
After one second, the intensity will increase until you release the button/knob. When you
press and hold the button/knob again, the intensity will start to decrease.
The intensity starts to decrease/increase after the maximum/minimum value has been
reached. When the limit value is reached, you will hear a short signal tone. The adjustment
direction is automatically toggled each time the button is released.
Both methods have the same effect when adjusting the intensity of the operating light.
However, the former is implemented only for the sake of symmetry of the functions, the
later method being the easier to use.
1.7.14 Language
Change the language of the touch panel as follows:
1. Press Program.
2. Press the flag to open a list of available languages.
3. Select a language by pressing it.
Move up and down on the list by pressing the up/down buttons.
The list of languages closes and the language of your choice is displayed on the touch
panel.
select language
NOTE
If the Planmeca Romexis Clinic Management module is used together with the Planmeca
Compact i Touch v2 dental unit, you do not need to adjust the clock settings. The dental
unit clock is synchronised with the Planmeca Romexis clock every time the dental unit is
booted up.
The button will turn blue and all items in the window will turn black (which means that
programming is enabled).
5. Select the date format.
Press the arrow to scroll through the different format options.
6. Change the time with the plus (+) and minus (-) buttons.
7. Press OK to exit and store the new settings.
2 COMPACT i CLASSIC V2
A 3
(3) Patient chair control buttons;
S A - backrest down / go to position A
B B - backrest up / go to position B
C - seat up / go to position C
D - seat down / go to position D
S - run to / return from rinsing position
D
(4) Bowl rinse/glass fill button
(5) Extra function button (door open / assistant call
4 5 6 etc.)
(6) Operating light control button
(7) Instrument spray button and indicator light
(8) Micromotor reverse button and indicator light
(9) Automatic chip-blow button and indicator light
7 8 9
3. Change the setting of the function with the foot control. Push the pedal halfway right to
increase the parameter value and halfway left to decrease it. If the pedal is pushed
more than 0.4 sec., the value will change as long as the pedal is pushed.
4. Press the Program button to store the new setting into the memory. The indicator light
will go out.
When programming instrument settings; remove the instrument that you wish to program
from its holder. After identifying the instrument it can be either held out from or returned to
its holder. If returned you have 3 seconds time to enter the programming mode.
NOTE
The maximum air and water flow rate of the syringe can not be changed via the control
panel. However, there are screwdriver-adjustable valves inside the console that can be
adjusted if desired. Please refer to 3.2 "Adjusting syringe water and air flow rates" on page
185.
instrument
instrument
instrument
661100__vvooss
sov_016
58
661100__vvooss
sov_016
Alternative method: press and hold
duration page 60
PoS.A
A
S
B
SA.35
661100__vvooss
sov_016
+
1. Press Program.
2. Press Instrument spray. The text SA.nn or SH.nn will appear on the display
indicating which of the parameters will be adjusted (A for air and H for water).
3. Press Instrument spray repeatedly (or the foot control pedal briefly down) to toggle
between air and water adjustment.
4. Change the flow rate with the help of the foot control pedal. Bringing the pedal to the
right will increase the flow, taking it to the left side will decrease the flow. Please note
that the instrument will run normally during this operation. Turbines will start
immediately to full speed.
5. Press Program to exit and store the new spray settings.
NOTE
If the dental unit is equipped with the Planmeca Sterile Water System, the sterile water
function is selected in this mode. Letters St indicate that sterilised water is fed to the
selected instrument. The indicator light of the instrument spray button flashes when the
spray is on. For more information, refer to the Sterile Water System Users Manual.
NOTE
The manual chip blow is set as default to be dry (air only), independent of the automatic
chip blow programming. However, if desired, this can be altered, please refer to "n.43 Set
mode of foot controlled chip-blow" on page 88.
1. Press Program.
2. Press Chip blow. The text CA.nn or CH.nn will appear on the display indicating
which of the parameters will be adjusted (A for air and H for water). The factory setting
is full air pressure and water off.
3. Press Chip blow repeatedly (or the foot control pedal briefly down) to toggle between
air and water adjustment.
4. Change the flow rate with the help of the foot control pedal. Bringing the pedal to the
right will increase the flow, taking it to the left side will decrease the flow. Please note
that the instrument will not run during this programming operation.
5. Press Program to exit and store the new automatic chip blow settings.
661100__vvooss
sov_016
+ CA.nn
+ L. 90
-
1. Press Program.
2. Press Reverse rotation. The text L. nn or L. oFF will appear on the display.
3. Change the light intensity with the help of the foot control pedal. Bringing the pedal to
the right will increase the light, taking it to the left side will decrease the light. Please
note that the instrument will not run during this programming operation. The light can
be programmed to be always off: after reaching the minimum value the light is
switched off, indicated by L.oFF on the display.
4. Press Program to exit and store the new light intensity into the memory.
b. 30
661100__vvooss
sov_016
+
c. 6.0
-
661100__vvooss
sov_016
o
661100__vvooss
sov_016
+ 82
NOTE
The intensity of the operating light can also be programmed as follows:
Press and hold the operating light button (or the button in the light handle) for longer than
0,5 seconds. As long as the button is pressed, the light intensity slowly decreases (or
increases) until it comes to the minimum (or maximum) value. A short beep sounds at the
end limit. The adjustment direction is automatically toggled each time the button is
released.
Both methods have the same effect when adjusting the brightness of the operating light.
However, the former is implemented only for the sake of symmetry of the functions, the
later method being the easier to use.
d. 12
661100__vvooss
sov_016
+
CAUTION
Some of the service mode functions can be used to alter the operation of the unit.
Never use service modes whose operation you are not familiar with.
CAUTION
The chair lift mechanism can be driven at any time (even if the service door is open)
in the service mode. Make sure to remove tools and fingers from the mechanism
prior to running the chair from its controls.
You must enter a PIN-code to enter the service mode. The service mode number is shown
on the display and you can scroll through the service modes with the up and down buttons
on the touch panel.
The main service mode window summarises the contents of the service mode. If it
contains much information, you can view the rest of the text by scrolling the window. A
more detailed description and possible instructions can be accessed by pressing the i-
symbol in the top right-hand corner. The new window is closed by pressing OK.
NOTE
In the service mode, only service mode information (no error or help codes) is displayed on
the touch panel.
NOTE
Even if listed in this manual, some service modes are reserved for future use and might not
yet be implemented in the current software version.
PIN code
NOTE
To ensure that changes made in the service mode are stored, you must exit the service
mode by using the Program button. All altered settings, even if they have an immediate
effect, are lost if the power is turned off before pressing the Program button.
Table 6: Instrument
Table 6: Instrument
3.3 Instrument
n.39 Set maximum fibre optic and Intracam bulb voltage for
selected hose type
This service mode can be used to fine adjust the maximum allowed fibre-optic light bulb
voltage for the selected hose type. The display shows four dashes (----) if no valid hose is
selected. When a hose (instrument) that contains a fibre-optic light is selected, the display
shows the setting for this hose type. The factory default is 170. Positive values increase,
negative decrease the factory default. Each step equals about 50mV. Determine the bulb
voltage by measuring it directly at the bulb socket when the bulb is on. Never exceed the
maximum allowed voltage.
Satelec Newtron LED: Set value to the maximum value = 170.
Note: The dental unit can be equipped with 2 or more scalers, but they must be of the
same type. Two different types of scaler can not be supported simultaneously.
Note: If the Satelec LED scaler is replaced by the Satelec Lux (halogen scaler) the service
mode value must be set to 0 (halogen light) while the Satelec LED is still attached to the
dental unit.
n.41 Set fibre optic light turn-off delay for all instruments
This service mode can be used to alter the turn-off delay for instruments with an internal
fibrelight. The factory default is 30 seconds.
n.49 Display test for Planmeca Compact i 2009 control panel. Not
applicable for touch panel.
This service mode is not applicable to the touch panel.
The service mode turns on all individual LEDs and display segments on the control
panel(s) (Planmeca Compact i 2009 control panel). Can be used to check that all displays
work properly, as well as to check the control panel for dead pixels. Shows also the
version number of the control panel software.
Note: For scalers the momentary spray is always with water regardless of this service
mode setting.
0 = Momentary spray enabled, momentary spray water off, instrument spray setting can
be toggled with foot control
1 = Momentary spray enabled, momentary spray water on, instrument spray setting can
be toggled with foot control
2 = Momentary spray enabled, momentary spray water off, instrument spray setting can
not be toggled with foot control
3 = Momentary spray enabled, momentary spray water on, instrument spray setting can
not be toggled with foot control
4 = Momentary spray disabled, instrument spray setting can be toggled with foot control
(default)
5 = Factory use
6 = Momentary spray disabled, instrument spray setting can not be toggled with foot
control
7 = Factory use
n 85 Output signals
1st vertical segment (counting from right): WMS use, see service mode 126
2nd: Not in use
3rd: Saliva suction selective valve
4th: High-volume suction selective valve
5th: Intracam activated
6th: Microvac pulse valve
7th: Main water valve
8th: Cup fill valve
n 86 Output signals
1st vertical segment (counting from right): Assistant syringe water valve
2nd: Assistant syringe heater
3rd: Metasys/Drr error acknowledge signal or Drr VSA rinse valve
4th: Metasys/Microvac/Drr start signal or Drr VSA/Wet suction valve
5th: Water flush to suction system
6th: Ejector control
7th: Bowl rinse control
8th: Main air valve
n 87 Driver signals
1st vertical segment (counting from right): Power source control
2nd: Beeper control
3rd: Boiler control
4th: Assistant call / door open / Intracam control relay
5th: Suction motor relay
6th: Backrest movement reversion relay
7th: Lift motor movement reversion relay
8th: Instrument reverse rotation
n 88 Driver signals
1st vertical segment (counting from right): Scaler electronics activated
2nd: Status LED / Run LED control
3rd: Factory use
4th: Factory use
5th: Not in use
6th: Not in use
7th: Not in use
8th: Not in use
n 89 Input signals
1st vertical segment (counting from right): Microvac overfill error
2nd: Metasys/Drr functional
3rd: Metasys/Drr separator alarm or Drr VSA float reed relay signal
4th: Factory use
5th: Operating light switch
6th: Cuspidor door switch
7th: Metasys separator beeper
8th: Factory use
n 90 Input signals
1st vertical segment (counting from right): Scaler electronics selection 1
2nd: Scaler electronics selection 2
3rd: Scaler electronics selection 3
4th: Separator selection 1
5th: Separator selection 2
6th: Separator selection 3
7th: WCS container recognition
8th: Patient cup fill button
n 91 Outputs
1st vertical segment (counting from right): Not in use
2nd: Not in use
3rd: Backrest motor control
4th: Lift motor control
5th: Chair control
n 92 Inputs
1st vertical segment (counting from right): Factory use
2nd: Safety switch
3rd: Factory use
3.11 Other
n.109 Unit type setting to enable or disable the IMUX and the door
features
This service mode is used to enable or disable the IMUX and the door safety features, if
needed. The factory default is either 0 or 3, depending on the unit type selection in service
mode n.100.
0 = Compact i (nor)
3 = Compact e (Co.E). Door switch is ignored.
28 = Cart (CArt). No error messages, but the text "CArt" is displayed if the IMUX is
disconnected.
31 = Independent cart (In.CA). The door switch is ignored and no error messages are
displayed, but the text "CArt" is displayed if the IMUX is disconnected.
44 = No instrument console (noI.C). Neither error messages nor the text "CArt" are
displayed if the IMUX is disconnected. Cup filling and bowl rinsing get water, the assistant
syringe gets water and air, the syringe heater can be powered on, and the water heater
can be powered on even if the IMUX is disconnected.
n 110 Outputs
1st vertical segment (counting from right): WMS use
2nd: WMS use
3rd: Factory use
4th: Factory use
5th: Factory use
6th: Factory use
7th: Factory use
8th: Factory use
n 111 Inputs
1st vertical segment (counting from right): WMS use, see service mode n 126
2nd: WMS use, see service mode n 126
3rd: WMS use, see service mode n 125
4th: Bowl above patient chair
5th: Cup recognition
6th: WMS use, see service mode n 125
7th: Chair type input 0
8th: Chair type input 1
n 112 Inputs
For factory use.
3.19 Instrument
3.20 Suction
3.21 Flexy
n.203 Processor watchdog test. Restarts the dental unit with error
E 75.1.
Restarts the dental unit with error E 75.1.
n.204 Processor reset test. Restarts the dental unit without errors.
Restarts the dental unit without errors.
3.23 Chair
3.24 Headrest
3.26 Network
3.28 Headrest
n.319 As service mode n.18 but chair and headrest endurance tests
For factory use.
3.29 Instrument
3.30 Maintenance
n.402 How many days left until next yearly maintenance or confirm
yearly maintenance
View how many days is left until next yearly maintenance. Confirm yearly maintenance by
pressing SET button.
1 HELP MESSAGES
2 ERROR MESSAGES
ERROR EXPLANATION
CODE
E1.01 Main water valve short circuit.
E1.02 Main air valve short circuit.
E1.03 Bowl rinse valve short circuit.
E1.04 Cup filling valve short circuit.
E1.05 Water flush valve short circuit.
E1.06 Separator pulse valve short circuit.
E1.07 Ejector valve short circuit.
E1.08 Extra out 1 short circuit.
E1.09 Extra out 2 short circuit.
E1.10 Extra out 3 short circuit.
E1.11 Separator start signal / Microvac's start valve short circuit.
E1.12 Assistant syringe water valve short circuit.
E1.13 Separator alarm reset signal short circuit.
E1.14 PLANET output (ELMP) short circuit.
E1.15 Not used or not in the manuals.
E2.01 Incoming water valve open circuit.
E2.02 Incoming air valve open circuit.
E2.03 Bowl rinse valve open circuit.
E2.04 Cup filling valve open circuit.
E2.05 Main Control PCB output open circuit warning.
E2.06 Separator pulse valve open circuit.
E2.07 Main Control PCB output open circuit warning.
E2.08 Main Control PCB output open circuit warning.
E2.09 Main Control PCB output open circuit warning.
E2.10 Main Control PCB output open circuit warning.
E2.11 Separator start signal open circuit / Microvac valve open circuit.
E2.12 Assistant syringe water valve open circuit.
E2.13 Separator alarm reset signal open circuit.
E2.14 Main Control PCB output open circuit warning.
E2.15 Not used or not in the manuals.
E3.1 Incoming air pressure compared to water pressure is too low.
E3.2 General pressure problem.
E3.3 General pressure problem.
E3.4 Internal air pressure (after regulator) is too low.
ERROR EXPLANATION
CODE
E3.5 Internal air pressure (after pres. regulator) is too high.
E4.1 Short circuit in cable from MCB to control panel / suction holder in branch 1.
E4.2 Short circuit in cable from MCB to control panel / suction holder in branch 2.
E5 Unsupported device attached.
E6.01 Cuspidor up -safety circuit error.
E6.02 Cuspidor down -safety circuit error.
E6.03 Chair down -safety circuit error.
E6.04 Bowl up -safety circuit error.
E6.05 Bowl down -safety circuit error.
E6.07 Console up -safety circuit error.
E6.08 Console down -safety circuit error.
ERROR EXPLANATION
CODE
E7 Power supply related error messages. IPS voltage failure on MCB.
E8 Power supply related error messages.
E9 Internal operating voltage too low.
E10 Internal operating voltage too high.
E11.1 OP-light fuse (F5) and/or Water heater fuse (F7) has blown on the MCB.
E11.2 Separator fuse (F6) and/or Syringe heater fuse (F8) has blown on the MCB.
E11.3 Either of the +24V electronics fuses (F9 or F10) has blown on the MCB.
E11.4 Rectifier D5 or D6 short circuit or either of the +24V electronics fuses (F3 or
F4) has blown on the MCB. Chair, instrument and headrest control not allowed.
E11.5 Separator fuse (F5) has blown on the MCB. Chair, instrument and headrest
control not allowed.
E11.6 Water heater fuse (F6) has blown on the MCB. Chair, instrument and headrest
control not allowed.
E11.7 Syringe heater fuse (F7) has blown on the MCB. Chair, instrument and
headrest control not allowed.
E11.8 Polymerisation light fuse (F8) has blown on the MCB. Chair, instrument and
headrest control not allowed.
E12.1 SELV voltage too low (below 30V). Chair, instrument and headrest control not
allowed.
E12.2 SELV voltage too high (above 39V)
E13 Mains frequency outside acceptable range.
E14 IPS temperature measurement sensor error.
E15 IPS heatsink is running too hot on MCB.
ERROR EXPLANATION
CODE
E16 Attached scaler electronics is of new type.
E16.1 New type of scaler electronics is attached to MCB.
E16.2 New type of scaler electronics is attached to IMUX.
E16.3 Scaler electronics slot does not match with scaler hose connection.
E16.4 Other instrument than scaler is connected to slot where scaler electronics is.
E17 Power Supply related error messages.
E18 Power Supply related error messages.
ERROR EXPLANATION
CODE
E19.1 Control panel in branch 1 is of new type.
E19.2 Control panel in branch 2 is of new type.
E20.1 Control panel 1 button stuck.
E20.2 Control panel 2 button stuck.
E21 Control panel related error messages.
E22 Control panel related error messages.
ERROR EXPLANATION
CODE
E23 Polymerisation light bulb has blown or is not properly attached to socket.
E24 Control panel related error messages.
E25 Satelec mini LED polymerisation light requires newer IMUX PCB version.
E26 Instrument related error messages.
E27 Instrument related error messages.
ERROR EXPLANATION
CODE
E28.1 Reference resistor signal out of bounds when unit is switched on.
E28.2 Active instrument signal out of bounds.
E28.3 Syringe signal out of bounds.
E28.4 Reference resistor signal out of bounds when instrument is activated.
E28.5 Syringe signal changes during use (blinking).
E29 Voltage error in fibre light power supply in IMUX.
ERROR EXPLANATION
CODE
E30 Instrument multiplexer related error messages.
E31 Instrument multiplexer related error messages.
E32.1 IMUX is not responding. IMUX cable disconnected or failure in cable or IMUX.
E32.2 IMUX software is not compatible (IMUX is of new type) with MCB software.
E32.3 IMUX data error (cable short circuit).
E32.4 IMUX communication error.
E32.5 Wrong IMUX type or wrong unit type.
E33.1 Syringe select valve short circuit.
E33.2 Instrument 1 select valve short circuit.
E33.3 Instrument 2 select valve short circuit.
E33.4 Instrument 3 select valve short circuit.
E33.5 Instrument 4 select valve short circuit.
E33.6 Instrument multiplexer related error messages.
E33.7 Drive/Cooling air valve short circuit.
E33.8 Air coolant valve short circuit.
E33.9 Water coolant valve short circuit.
E34 Not used or not in the manuals.
E34.1 Syringe select valve open circuit.
E34.2 Instrument 1 select valve open circuit.
E34.3 Instrument 2 select valve open circuit.
E34.4 Instrument 3 select valve open circuit.
E34.5 Instrument 4 select valve open circuit.
E34.6 Instrument multiplexer related error messages.
E34.7 Drive/Cooling air valve open circuit.
E34.8 Air coolant valve circuit.
E34.9 Water coolant valve circuit.
E35.1 Drive/Cooling air pressure sensor output <0.2 V.
E35.2 Air coolant pressure sensor output <0.2 V.
E35.3 Water coolant pressure sensor output <0.2 V.
E35.4 Differential pressure sensor output <0.2 V.
E36.1 Drive/Cooling air measurement over range.
E36.2 Spray air measurement over range.
E36.3 Spray water measurement over range.
E36.4 Differential pressure sensor output >5 V.
E37 Pressure sensor cable not properly attached to IMUX.
E38 Instrument multiplexer related error messages.
ERROR EXPLANATION
CODE
E39 Drive (cooling) air valve control servo cannot maintain pressure to instrument.
E40 Instrument multiplexer related error messages.
E41 Instrument multiplexer related error messages.
E42 Air coolant (spray air) control servo cannot maintain required pressure.
E43 Instrument multiplexer related error messages.
E44 Instrument multiplexer related error messages.
E45 Water coolant (spray water) control servo cannot maintain required pressure.
E46 Instrument multiplexer related error messages.
ERROR EXPLANATION
CODE
E47 Suction holder is of new type.
E47.1 Suction holder in branch 1 is of new type.
E47.2 Suction holder in branch 2 is of new type.
E48.1 No suction holders detected.
E48.2 Suction holder related error messages.
E48.3 Suction holder related error messages.
E48.4 Suction holder related error messages.
E49.1 Suction holder PCB or its cable is faulty in branch 1.
E49.2 Suction holder PCB or its cable is faulty in branch 2.
E50 Unidentified STCS jumper configuration.
E50.1 WMS must be set to Bottle water off -mode when starting STCS cleaning
sequence.
ERROR EXPLANATION
CODE
E51.1 Foot control related error messages.
E51.2 Foot control is not responding.
E51.3 Foot control pedal was pressed down when dental unit was switched on.
E51.4 Foot control pedal was not in home pos. when dental unit was switched on.
E51.5 Foot ctrl chair ctrl knob was not in home pos. when dent. unit was switched on.
E51.6 Foot control left knob is either stuck or active at power-up or after pedal use.
E51.7 Foot control right knob is either stuck or active at power-up or after pedal use.
E52 Foot control is of new type, update MCB software.
E52.1 Incompatible wireless foot control receiver software
ERROR EXPLANATION
CODE
E52.2 Incompatible wireless foot control software
E52.3 Incompatible foot control software
E53 Foot control data error (cable short circuit).
E54.0 Foot control error.
E54.1 Foot control EEPROM write failure.
E54.2 Foot control EEPROM read failure.
E54.3 Pedal too far from sensors, channel A.
E54.4 Pedal too far from sensors, channel B.
E54.5 Sensor plate is tilted too much to either side, checksum error.
E54.6 Sensor plate is too close to PCB, channel A.
E54.7 Sensor plate is too close to PCB, channel B.
E54.8 Foot control re-trigger error.
E54.9 Foot control calibration error.
E55 Wireless foot control connection lost
ERROR EXPLANATION
CODE
E56 Operating light erroneously connected or short circuit in push button.
E57 Incompatible SingLED software
E58 Operating light bulb burned out.
ERROR EXPLANATION
CODE
E59.1 Separator reports functional error.
E59.2 Separator reports functional error.
E60 Separator is not responding or separator cable is faulty.
E61 Separator is flooded with too much water.
E62 Newly attached separator is of new type and cannot be controlled by MCB.
E63.1 WMS seems to be installed, although it should not be (according to service
mode).
E63.2 No WMS is installed, although it should be according to service mode.
ERROR EXPLANATION
CODE
E64 Seat pos. potentiometer or its cable is faulty or disconnected.
E65 Lift motor does not run or no signal from position potentiometer.
E65.1 Lift motor does not run. EmeStop relay PCB/cables or position sensor/cable
may be faulty.
E66 Lift motor or its position potentiometer cable has a wrong connection.
E67 Backrest position potentiometer or its cable is faulty or disconnected.
E68 Backrest motor does not run or no signal from position potentiometer.
E68.1 Backrest motor does not run. EmeStop relay PCB/ cables or position sensor/
cable may be faulty.
E69 Backrest motor or its position potentiometer cable has a wrong connection.
E70
E70.1 Lift motor error.
E70.2 Backrest motor error.
E71 Patient chair -related.
ERROR EXPLANATION
CODE
E72.1 Program (Flash-EPROM) error. Main software is corrupted. Replace Flash-
EPROM.
E72.2 Program (Flash-EPROM) error. Boot software is corrupted. Replace Flash-
EPROM.
E72.3 Program (Flash-EPROM) error.
E72.4 Data transfer error while updating main PCB SW from control panel memory
device.
E73.1 EEPROM error, CPU EEPROM erase/program failure, replace CPU.
E73.2 CPU configuration register contents are wrong, replace CPU.
E73.3 CPU EEPROM checksum error.
E73.4 CPU EEPROM checksum programmed. Replace CPU if error occurs often.
E74.1 RAM error, External RAM test failure, replace RAM.
E74.2 CPU RAM test failure, replace CPU.
E75.1 CPU error, WD reset.
E75.2 Illegal OP-code.
E75.3 Illegal vector.
E75.6 Mains voltage failure.
E75.7 Clock monitor reset.
ERROR EXPLANATION
CODE
E76 Main PCB hardware test failed.
E77 Error in communicating with external device reserved for future use.
E79 IMUX water pressure cannot be released during WCS cycle.
ERROR EXPLANATION
CODE
E80.1 Motorised headrest is not found (according to service mode configuration).
E80.2 Motorised headrest is found, but is not set in service mode configuration.
E80.3 Headrest joystick (dentist side) stuck or cable broken.
E80.4 Headrest joystick (assistant side) stuck or cable broken.
E81 Motorised headrest communication error (may be too old or wrong software or
loose cables or short circuit).
E81.1 Motorised headrest communication error.
E82 Headrest SW is incompatible with MCB SW.
E82.1 Headrest SW is incompatible with MCB SW.
E82.2 Incompatible headrest software.
E90 Water leak detected.
ERROR EXPLANATION
CODE
E91 Yearly maintenance overdue.
ERROR EXPLANATION
CODE
E99 Software upgrade failure.
E99.1 MCB software upgrade failure.
E99.2 IMUX software upgrade failure.
E99.3 Control panel software upgrade failure.
E99.4 Operating light software upgrade failure.
E99.5 Motorised headrest software upgrade failure.
E99.6 Foot control software upgrade failure.
E99.7 Wireless foot control software upgrade failure.
E99.8 Wireless foot control receiver software upgrade failure.
ERROR EXPLANATION
CODE
EP19.3 Control panel version mismatch.
EP19.5 Control panel software upgrade has failed or memory card is faulty.
EP21.1 Control panel display is faulty.
EP21.2 Control panel data error.
EP21.5 Unable to read Main PCB software version.
EP21.6 Control panel software not installed or software error.
EP21.7 Touch panel is not working.
E11 OP-light fuse (F5) and/or Water heater fuse (F7) has blown on
the MCB
The conditions of the secondary fuses are indicated also by green LEDs at each fuse
holder position. Look at the inner side of the PCB edge. If a green LED is dark, then the
respective fuse at that position is blown. The following list gives more details about the
blown fuse. Remember to replace the fuse with the same type and rating indicated, to
guarantee safety of the equipment.
Always turn off the power during fuse replacement. If a fuse blows immediately again, try
to disconnect the respective load (OP-light, water heater, separator or syringe) to check
whether the load itself is causing the problem.
E12 SELV voltage too low (below 30V). Chair, instrument and
headrest control not allowed
SELV voltage outside acceptable range (below 30V or above 39V) Check first the mains
voltage at the input terminals on the MCB. If the error only occurs when running the patient
chair, check that the mains impedance is not too high so that the incoming mains voltage
does not drop under the minimum allowed. The following voltage variations are allowed to
guarantee normal operation;
230 volt setting (Europe) mains can be from 187VAC to 265VAC
115 volt setting (USA) mains can be from 98VAC to 132VAC
100 volt setting (Japan) mains can be from 80VAC to 110VAC
E42 Air coolant (spray air) control servo cannot maintain required
pressure
The proportional valve gets full power, but still the measured pressure does not reach the
required level. The most probable reason is that there is something reducing the air flow
so that the pressure cannot reach the required level even if the control valve is fully open.
Check for air tube kinks inside the unit and console arm. Check the condition of the
incoming air pressure regulator and also the air filter. Check that the incoming air pressure
is within recommendations. The control valve in the IMUX could also cause this problem.
Replace the valve as a last remedy.
E65 Lift motor does not run or no signal from position potentiometer
There can be several reasons for this error, but they can easily be divided into 3
categories to be more easily understood;
a) If the motor runs and the chairs lift mechanism moves: The error must be in the
potentiometer, its cable or on the MCB. Visually check the condition of the potentiometer.
If the potentiometers axle is rotating together with the worm screw, then check the cable
from the potentiometer all the way to the MCB. Turn off the power and use and ohmmeter
to do the checking. If the potentiometer and its cable seem to be ok, you can additionally
check the reading from the potentiometer with the help of service mode n.76 that shows
the potentiometer signal on the unit display. Connect an external potentiometer directly to
the MCB and use the service mode to check that the MCB is OK.
b) The motor does not run, but there is a faint humming sound from the motor. Check that
the movement has not reached either of the mechanical limits. Is the chair overloaded
(>150 kg load)? Is there any other mechanical obstruction? Check that the motor capacitor
is properly connected and that the capacitor is OK. Check that the motor cable is OK (you
should be able to measure the resistance of both windings at the connector that connects
to the MCB).
c) The motor is completely dead. Is the motor very hot? It is possible to overheat the motor
if you run it continuously for more than about 2 minutes, or if the motor runs more that 13%
of the time. When the motor heats up, the built in over-temperature switch disconnects the
common wire, resulting in a temporarily dead motor. When the motor cools down (this can
take 10 - 20 min.) it returns back to normal operation. Check also the connector and
cables to the motor, if it is not overheated, but still dead.
NOTE
The preventive maintenance should be performed once a year by a qualified service
technician. After performing the yearly maintenance you must confirm the completion, see
section 1.5.1 "Yearly maintenance" on page 36.
1 DELIVERY ARMS
2 SUCTION ARMS
3 INSTRUMENT CONSOLE
4 INSTRUMENTS
4.1 Syringe
Tip is clean.
4.2 Turbine
Drive air pressure. Service mode 30.
Bulb voltage.
O-rings. Change automatically twice a year.
Lubricate.
4.3 Micromotor
Bulb voltage.
O-rings. Change automatically twice a year.
Lubricate.
5 FOOT CONTROL
5.1 Pedal
Operation check. If the pedal does not move smoothly, remove the cover and clean
and oil the pedal joint and all the surfaces which the pedal touches.
6 CUSPIDOR
6.4 Fuses
Condition check
6.5 Bowl
Bowl filter condition. Replace if needed.
Glass filling/bowl rinsing is working.
Check the operation of the bowl microswitch.
7 CHAIR
7.3 Upholsteries
Condition check.
Upholsteries are firmly in place. The screws of the seat upholstery are tightened.
8 SUCTION SYSTEM
9 OPERATING LIGHT
Clean the protective fascia.
Operating light is in focus. Applies only to the Delight operating light.
Light is not turning with the arm.
13 ELECTRICAL SAFETY
These measurements must be performed after installation, electrical parts replacement
and during preventive maintenance.
All the measurements shall be performed using a IEC62353 compliant measurement
device. The unit shall be separated from the supply mains (live, neutral and protective
earth) during the measurements.
NOTE
General tip: if a measurement point is in a visible part, avoid scratching by using magnet or
metal foil between the measurement point and test clip.
Measurement points
1. Instrument console cover (in balanced arm console, attachment screws).
6. Base casting
7. Backrest motor
8. Chair frame
WARNING
Risk of electric shock due to high voltage! Do not touch the unit during testing.
Applied parts needed in the measurements are instruments in the instrument console and
suction holder that contains metallic conductive parts (scaler, micromotor, syringe, curing
light, turbine, for instance).
Do not measure non-earthed accessible conductive part, because there are not any.
Measurements
1. Between the mains part (L and N together) and protective earth
2. Between the mains part (L and N together) and the applied parts (all the instruments
with conductive metallic parts)
3. Between the protective earth and the applied parts (all the instruments with
conductive metallic parts)
Measurement
1. Measure with all the applied parts (dentist and assistant instruments with conductive
metallic parts), L, N and PE connected to measurement device.
WARNING
Risk of electric shock due to high voltage! Do not touch the unit during testing.
Do not measure non-earthed accessible conductive part, because there are not any.
Alternative method
Measurement
1. Measure with all the applied parts (dentist and assistant instruments with conductive
metallic parts), L, N and PE connected to measurement device.
1 MECHANICAL ADJUSTMENTS
Console arm
1 2
Adjustment screws
Comp9_20.eps
Adjusting hook
Spring support
y-spring
Adjusting hook
varsikuvaI-1.eps
2. Insert a screwdriver into the arm (through the opening at the underside of the arm)
and push the adjusting hook (inside the arm) into a groove of the adjustment plate.
Make sure that the screwdriver goes through the hole in the y-spring. Insert the
screwdriver into the arm perpendicularly towards the arm.
NOTE
Do not let the screwdriver go through the groove on the adjustment plate to avoid
damaging the tubes above the adjustment plate.
Adjusting hook
varsikuvaI-2.eps
Adjusting plate
90
Hole in the y-spring
3. Hold the screwdriver inside the arm and lift the arm to its highest position, while
keeping the screwdriver in the same position compared to the arm.
varsikuvaI-3.eps
varsikuvaI-4.eps
6. Repeat the procedure if necessary. There are totally seven steps to increase the
tension of the gas spring (from the fully decreased spring).
varsikuvaD-1.eps
2. Insert a screwdriver into the arm (through the opening at the under side of the arm)
and push the adjusting hook (inside the arm) into a groove of the adjustment plate.
Make sure that the screwdriver goes through the hole in the y-spring. Insert the
screwdriver into the arm at the right angle (90) compared to the arm.
90
varsikuvaD-2.eps
3. Hold the screwdriver inside the arm in the same position compared to the arm and lift
the arm to its highest position.
varsikuvaD-3.eps
90
4. Turn the handle of the screwdriver towards the instrument console and hold it in that
position. Do not turn the screwdriver more than 30 to avoid damaging the y-spring.
varsikuvaD-4.eps
~30
The spring support is released from the groove.
5. Lower the arm to horizontal position while holding the screwdriver at the same
position compared to the arm.
varsikuvaD-5.eps
~30
varsikuvaD-7.eps
8. Repeat the procedure if necessary. There are totally seven steps to decrease the
tension of the gas spring (from the fully increased spring).
Locking ring
8 DIN 471
Lower joint
adjustment screw
Lower joint
pin (short)
Lifting arm cover
Horizontal arm
2. Adjust the lifting friction of the delivery arm with the 5 mm Allen key. Adjust the two
screws of the level arm friction block equally. Tightening the screws increases the
friction.
2. Before adjusting the tension of the gas spring decrease the lifting friction of the side
delivery arm with the 5 mm Allen key. Adjust the two screws of the level arm friction
block equally. Loosening the screws decreases the friction.
3. Adjust the tension of the gas spring by rotating the adjustment screw (located in the
lower end of the gas spring) with the 6mm Allen key as shown on the figure below.
Turning the screw clockwise will increase the tension, i.e, the weight-carrying
capacity.
4. Adjust the lifting friction of the delivery arm with the 5 mm Allen key.
5. Attach the level arm cover back to its position.
CAUTION
When removing the play of the lifting adapter, tightening the screws that adjust the
sliding support and the colliding support is recommended to be done by hand. Over-
tightening the screws will increase friction and cause wearing to the sliding support
and the colliding support.
2
Comp_lift_adap_110313.eps
2 ELECTRICAL ADJUSTMENTS
NOTE
When you hear the beep, stop tapping the start-up display.
Calibration marker
4. Small calibration markers appear on the display in order: first one in each corner and
then one in the middle. Tap each marker to calibrate the touch panel.
When the calibration is finished, the dental unit will continue the start-up.
WARNING
During the calibration of the backrest motor position sensor all software limits are
ignored. Special care must be taken for not to damage the unit by driving the back-
rest to its mechanical limits.
1. Enter service mode 80. The display will show letter b (for backrest) followed by two
horizontal lines.
2. Enable backrest motor movements by holding down the SET button for 10 seconds
(until you hear a beep).
3. Using the minus (-) and plus (+) buttons on the touch panel, drive the backrest into a
position where the calibration mark on the chair casting and the lug of the spindle nut
assembly are aligned.
Spindle nut
assembly
4. Loosen the two M3x6 DIN 916 screws at the side of the cog wheel of the sensor (1).
5. Rotate the axle of the sensor with a screwdriver (2) until the two horizontal lines on
the display are in line.
b _ b b
The sensor presumes the back- The backrest position sen- The sensor presumes the back-
rest to be at a too low position. sor is calibrated correctly. rest to be at a too high position.
Rotate the sensor axle counter Rotate the sensor axle clockwise
clockwise (seen from the axle). (seen from the axle).
6. Tighten the two M3x6 DIN 916 screws to secure the cog wheel into position.
7. Using the minus (-) button, drive the backrest downwards and after the lowest position
is reached continue pressing the minus (-) button for approx. 2 seconds to drive the
backrest to horizontal position. Check that the spindle nut does not touch the cog
wheel or potentiometer.
8. Drive the guiding nut past the calibration mark a couple of times to ensure that the
sensor is now calibrated correctly (the horizontal lines are aligned).
9. Exit the service mode.
WARNING
During the calibration of the lift motor position sensor all software limits are ignored.
Special care must be taken for not to damage the unit by driving the lifting adapter
to its mechanical limits.
1. Enter service mode 79. The display will show letter L (for lift motor) followed by two
horizontal lines.
2. Enable lift motor movements by holding down the SET button for 10 seconds (until
you hear a beep).
3. Using the minus (-) and plus (+) buttons on the touch panel, drive the lifting adapter to
position where the upper edge of the adapter and the calibration mark at the side of
the lifting column are aligned.
4. Loosen the two M3x6 DIN 916 screws holding the cog wheel of the sensor in position.
5. Rotate the axle of the sensor with a screwdriver until the two horizontal lines on the
display are in line.
L _ L L
The sensor presumes the lifting The sensor is calibrated The sensor presumes the lifting
adapter to be at a too low position. correctly. adapter to be at a too high position.
Rotate the axle counterclockwise Rotate the axle clockwise (seen from
(seen from the axle). the axle).
6. Tighten the two M3x6 DIN 916 screws to secure the cog wheel into position.
7. Drive the lifting adapter past the calibration mark a couple of times to ensure that the
sensor is now calibrated correctly.
8. Exit the service mode.
WARNING
Erroneously wired voltage selection jumper may cause serious damage to the
equipment.
The mains voltage selector jumper (connector P21 on the Main control PCB) is used to
select the transformer winding to correspond with the mains voltage. Since the transformer
is connected using the back-to-back method, the same transformer can be used for
different mains voltages. Voltage selection jumpers for different mains voltages can also
be ordered as spare parts.
1. Select the mains voltage by connecting pin 3 and one of the pins 4 - 6 together.
The same voltage setting jumper is also used to bypass the neutral line fuse. In some
areas it is not allowed to have a fuse in the neutral wire, thus the neutral line fuse
must be bypassed.
2. Bypass the neutral line fuse by connecting the pins 1 and 2 together.
10
5
11 0V 3&6 100V~
4
5 ~
22 V~
3
0 - 3&5 115V~
CO
2
2
MM 40V~
1
ON 3&4 220-240V~
N
FU EUT
SE RA
BY L L
PA IN
SS E
3 PNEUMATICAL ADJUSTMENTS
Lift to adjust,
push down to lock.
+
-
AdjBurk_Inal_press.eps
NOTE
Make sure that the hoses are properly connected to the console to avoid leakage.
610_vos
4. Loosen the bleeding screw until the water bleeds out (while the instrument is running).
NOTE
If the water is not bleeding out properly then loosen the two nuts a little. When the
bleeding is done then tighten the nuts again.
3. Wait until the water surface is just above the overflow opening and stop the water flow
by pressing the yellow cancel button.
4. Exit the service mode. The water flow is now calibrated.
PIN code
4. Enter service mode n 38 by using the up and down buttons on the touch panel.
The current service mode number (symbol n followed by the current service mode
number) is shown on the display when moving around in the list.
The service mode number disappears after 1 second, after which the current service
mode setting or value is displayed.
Select the water pressure measurement mode by pressing the plus (+) button. The
letter H followed by the current setting is shown on the display.
5. Decompress the water pressure with the syringe instrument.
6. Pull the pressure adjustment knob downwards and adjust the pressure by turning the
knob until the service mode setting is between 1.2-1.5 bar. In case the pressure
exceeds the value 1.7 bar, decompress the water pressure with the syringe
instrument (or with bowl flushing/glass filling) and readjust the pressure.
Pressure adjustment
knob
NOTE
Ensure that one bottle of Planosil is enough for the waterline cleaning cycle. The bottle
must not be completely emptied! If one bottle is not enough, you can, for example,
decrease the glass fill flow rate.
1 REMOVING COVERS
1
Cuspidor door Side cover plate
Cuspidor cover
2 UPHOLSTERIES REPLACEMENT
4. Remove the cover plugs. Unscrew the screws shown in the figure below.
5. Slide the seat upholstery towards the footrest so that the attachment screw comes out
from the groove of the upholsterys backplate. Remove the seat upholstery.
6. Slide the attachment plates to their positions at the new seat upholstery backplate as
shown in the figure below.
7. Unscrew the M4x50 DIN 912 and M4x20 DIN 912 screws from sides of the
multiplexer assembly.
3.3 Attaching extra voltage feed cables to the Instrument multiplexer PCB
and to the Control connector board PCB
There are four separate voltage feed outputs in the two PCBs inside the console.
Following three are in the right corner of the multiplexer PCB (115-10-37):
P25 outputs 24 Vac (marked CONN1 in warning label)
P29 outputs +32 Vdc (marked CONN2 in warning label)
P28 outputs +15 Vdc (marked CONN3 in warning label)
One voltage feed is located in the control connector board PCB (115-10-38) left from the
multiplexer:
These voltage feeds are used with certain instruments or with their possible electronic
boxes.
NOTE
Check the reference designator for connecting the cable to the right connector!
CAUTION
Care must be taken when removing and installing the multiplexer block for not to
damage the quick connector pins that are attached to the Instrument multiplexer
PCB.
NOTE
Avoid bending the Instrument multiplexer PCB when removing and installing the
multiplexer block.
Spindle screw
Blocking flange
NOTE
Care must be taken to position the new multiplexer control membrane correctly.
NOTE
Carefully tighten the screws equally and crosswise when repositioning the multiplexer block
cover. Using too much force might damage the threads in the multiplexer block.
Block cover
Spring
Membrane press
Multiplexer control
membrane
NOTE
Do not turn off the unit until the upgrade is finished! A power cut during the software
upgrade may break the touch panel.
NOTE
The touch panel software must be compatible with the main software. Do not use
incompatible software versions. If you have installed an incompatible software version,
error message EP 19.3 is displayed. You can update the main software to match the GUI
software without an upgrade USB memory stick by following the steps below starting from
step 4.
NOTE
If you have installed a peripheral device with an old software, you can upgrade the software
without an upgrade USB memory stick by following the steps below starting from step 4.
NOTE
If the unit is equipped with a wireless foot control, you have to connect it to the charger and
wake it up by pressing the handle before starting the software upgrade.
NOTE
If E82.1 appears after the main and touch panel software is upgraded, the headrest
software must be upgraded.
Enter service mode 303 and press the SET button to upgrade headrest software on the
headrest PCB into a version that is on the MCB software chip. The unit will turn off and on
again and the headrest software is upgraded.
1. Check the unit type by pressing Program > About this unit.
NOTE
The following is an example only and does not necessarily portray the actual
situation.
2. Make sure that the unit type is Compact i Touch v2 and that the software version is
7.0.4 or newer.
NOTE
If the unit type is Compact i Touch and the unit SW version 6.x.x.x, follow the upgrade
instructions in the Compact i Touch Technical Manual, Chapter G Parts replacement
and repair, section 5.1 Upgrading software.
3. Make sure you have the USB memory stick with the correct Compact i software
version.
4. Open the instrument console cover.
5. Insert the USB memory stick to the USB port in the touch panel assembly.
6. Press Program.
8. The following is displayed while the unit searches for an upgrade package.
9. The following is displayed after the upgrade package has been found. Only the
software that is shown in the view is upgraded.
Start the upgrade by pressing OK.
If you wish, you can cancel the upgrade by pressing Cancel.
NOTE
After this, the update cannot be cancelled.
NOTE
The following is an example only and does not necessarily portray the actual situation.
Cancel OK
10. The following is displayed after starting the upgrade. The progress of the update is
shown in a new window.
11. When the upgrade is completed, the dental unit reboots automatically.
12. Check the software versions by pressing Program > About this unit.
NOTE
The following is an example only and does not necessarily portray the actual situation.
NOTE
In software versions 7.1.0 or newer you can check all the peripheral software versions
by pressing Unit SW Version in the About this unit -window.
OK
13. Ensure that the correct unit software version and peripheral software versions have
been installed.
14. Press OK to close the window.
NOTE
Depending on the updated devices, the display will show either the text PCC or the
progress of the update.
NOTE
The screen may flicker during the upgrade.
NOTE
EUsb is shown on the display, if there is an error with the USB stick.
WARNING
Always replace fuses with the same type and rating.
NOTE
The mains fuse ratings depend on the selected voltage setting.
WARNING
The mains voltage is always present at the mains terminal on the main control PCB,
even if the unit is switched off from its own mains switch.
1. Turn off the power supply line or disconnect the unit from the mains power source.
2. Unscrew two attachment screws at the side of the electronics control box and open
the cover.
4. Disconnect all the cables that are connected to the main control PCB.
6. Install the new main control PCB in reverse order. All the cable connectors, except the
ones for the touch panel, suction arm and foot control cables, are unique and no other
cable can be connected to the connector than the one in question. All the cable
connectors are also labelled.
NOTE
After replacing the Main control PCB the dental unit automatically enters service mode
n.100 when the unit is switched on (see section "n.100 Select dental unit type" on page 99).
The unit type must be selected before you can use the dental unit.
NOTE
If the dental unit software version is updated when replacing the main control PCB, you
must upgrade the GUI software to the same level.
NOTE
When you have set the factory presets for all service modes and the dental unit
configuration differs from the software factory preset values, you must edit the service
modes to suit the dental unit configuration. Check, for example, the values for service
modes n.83, n.84, n.100 and n.133.
6. Unscrew five screws that attach the Instrument multiplexer PCB to the multiplexer
blocks.
7. Remove the Instrument multiplexer PCB by pulling it away from the multiplexer.
NOTE
The touch panel PCB must not be disassembled in any case!
1. Unscrew the M6x16 DIN 7984 screw from under the instrument console and open the
console frame.
2. Disconnect the Instrument multiplexer cable from the connector P1.
3. Disconnect the Suction arm cable from the connector P3.
4. Unscrew the four TORX WN1451 3x6 PT screws and remove the Touch panel PCB.
5. Install the new Touch panel PCB in reverse order.
Console frame
5. Replace the suction sensor PCB with a new one and assemble the suction arm in
reverse order.
Console frame
D
B
88
88 A
S
Screw M4x8 GE
RIN
SY OR R
TO
EC
EJ
IVA
AL
SCTION
SU
IVA
AL
SCTION
SU
Screws M4x10
Suction arm
WARNING
The unit must be switched off prior to transformer replacement.
6. Unscrew three attachment screws that hold the cuspidor on the floor.
7. Disconnect two transformer cables and water leakage sensor cable from the Main
PCB.
8. Turn the dental unit to its side as shown on the figure below.
9. Remove the water leakage sensor from the transformer casing (black arrow).
Unscrew four transformer attachment screws (white arrows). The transformer
assembly can now be removed from the base.
10. When attaching the new transformer assembly to its position, tentatively attach the
right side screws first to the base (1). Do not tighten the screws yet. Place the right
side of the casing below the screws and tighten the screws slightly. Attach the left side
screws (2) and tighten the screws equally.
2 1
11. Attach the water leakage sensor to the transformer metal frame.
NOTE
Never try to force the position sensor into the cog wheel, as it might cause the position
sensor to break.
NOTE
Care must be taken when assembling the position sensor for positioning the cog wheel
correctly compared to the threaded axle of the motor (see illustration below).
6. Calibrate the new position sensor as described in section 2.2 "Calibrating backrest
motor position sensor" on page 179 or in 2.3 "Calibrating lift motor position sensor" on
page 181 section respectively.
Position sensor
Washer
M4x4 DIN 916
Nut
Cog wheel
Cog wheel too close to the threaded axle Cog wheel too far from the threaded axle of
of the motor. The eccentricity of the axle the motor. The eccentricity of the axle causes
damages the position sensor. the cog wheel to slip and loose position.
Cog wheel
G G
N N Threaded axle
O O
W
R
W
R of the motor
C O R R E C T
5 MOTORS REPLACEMENT
4. Open the cable clamp located on the upholstery support plate (1). Remove the
upholstery support plate by unscrewing the three M6x10 DIN 912 screws (2).
5. Detach the legrest support bar from the spindle fork assembly by unscrewing the
attachment screw using the 6mm Allen key (1).
Spindle fork
assembly
6. Detach the spindle fork assembly from the seat casting by unscrewing the attachment
screw using the 6mm Allen key (1).
7. Cut the cable tie located on the spindle fork assembly (1) and unscrew the two
attachment screws of the backrest safety microswitch attachment plate using the
3mm Allen key and detach the attachment plate from the backrest motor assembly
(2).
8. Disconnect the Backrest motor cable from the seat cable adapter (1).
9. Detach the position sensor plate by unscrewing the two Ejot Torx DG 50x10 A2
attachment screws.
11. Unscrew the two M6x14 ISO 7380 screws at the backrest motor attachment plate
using the 4mm Allen key.
12. Detach the backrest motor mechanical limit from the seat casting by unscrewing the
M6x16 DIN 7984 attachment screw using 4mm Allen key.
14. Detach the mechanical limit and bearing from the threaded motor axle.
15. Install the new backrest motor assembly in reverse order. Assemble new cable ties in
the place of the ones cut previously.
WARNING
Care must be taken when replacing the chair lift motor. Removing parts of the lifting
assembly may enable the patient chair to fall causing a dangerous situation and/or
damage to the equipment. The downward movement of the chair must be elimi-
nated before the motor is detached.
Do not remove the worm screw securing nut. It must always be attached to the worm
screw.
Worm screw
securing nut
1. In the case that the motor is somewhat operational, drive the patient chair upwards
until you gain access to the opening on the lifting column. Insert a robust steel bar
through the lifting column. Carefully drive the chair downwards until the steel bar
supports the lifting adapter (and the chair).
If the motor is completely jammed, position a robust box etc. under the patient chair to
eliminate the downward movement of the chair.
NOTE
Ensure that the steel bar supports the whole weight of the patient chair by trying to pull
out the bar. The steel bar must not move.
NOTE
Pay attention when reconnecting the lift motor position sensor, refer to section 4.9
"Replacing position sensors" on page 221.
NOTE
Calibrate the lift motor position sensor as described in section 2.3 "Calibrating lift motor
position sensor" on page 181.
Electronics
control box
lifted aside Remove the nut attachment plate
and the lifting frames
WARNING
Special care must be taken when disconnecting the water and air inlet tubes.
Ensure that the water and air supply lines are turned off.
NOTE
Any splashed water must be carefully removed prior to unit turn-on.
1. Decompress the water and air pressures as described in section 6.1 "Decompressing
water and air pressures" on page 232.
2. Loosen the two attachment screws (1) and lift off the magnetic valve/pressure
regulator assembly from the cuspidor plate (2).
3. Disconnect all the magnetic valve control cables that are connected to the valves.
4. Detach all the water and air tubes that are connected to the valves.
5. Install the new magnetic valve/pressure regulator assembly in reverse order.
6. Apply pressure to test that there are no air or water leaks.
CAUTION
Ensure that the water and air supply lines are turned off.
1. Decompress the water and air pressures as described in section 6.1 "Decompressing
water and air pressures" on page 232.
2. Loosen the two screws attaching the magnetic valve assembly to the cuspidor plate
and lift off the magnetic valve/pressure regulator assembly.
3. If necessary for easier access, disconnect the control cable from the main valve you
are working on.
4. Detach the water (N108-DS02) or air (N108-DS01) filter/regulator by unscrewing the
two bolts which hold it in position.
If necessary, detach the water or air filter by unscrewing it from the water or air
regulator.
5. Install the new filter/regulator into position in reverse order.
NOTE
Be careful not to overtighten the bolts. (Max. Torque 7 Nm)
6. Adjust the pressure regulator as described in section 3.1 "Adjusting internal water &
air pressures" on page 184:
water 280 kPa (2.8 bar) (41 psi)
air 550 kPa (5.5 bar) (80 psi)
7. Apply pressure to test that there are no air or water leaks.
Water main
valve
Air main valve
Water filter/regulator
N108-DS02 I
ZZ
In P. max. 16 bar
MO
C AMade in Italy
Valve cable
Out P. 0,5-10 bar
Air filter/regulator
Bolts
Water filter I
Air filter N108-DS01
MO
ZZ
NOTE!
In P. max. 16 bar
Out P. 0,5-10 bar C AMade in Italy
Max. Torque 7 Nm
6.4 Replacing water and air filters and water regulator plug
1. Decompress the water and air pressures as described in section 6.1 "Decompressing
water and air pressures" on page 232.
2. Open the filter cup. Use the 17mm fork spanner if needed.
3. Remove the filter by unscrewing it from the regulator.
Filter cup
Filter
Plug
5. Attach the new filter to its position and attach the filter cup. Be careful not to
overtighten the filter cup, you should be able to open it manually.
6. Adjust the pressure regulator as described in section 3.1 "Adjusting internal water &
air pressures" on page 184:
water 280 kPa (2.8 bar) (41 psi)
air 550 kPa (5.5 bar) (80 psi)
7. Apply pressure to test that there are no air or water leaks.
Lock ring
Regulator cap
Membrane
CAUTION
Ensure that the water and air supply lines are turned off.
1. Decompress the water and air pressures as described in section 6.1 "Decompressing
water and air pressures" on page 232.
2. Disconnect the control cable from the main valve you are working on.
3. Detach the locking nut from the main valve. Remove the old main valve. Detach the
valve bottom plate.
4. Remove the valve support and its attachment plate by unscrewing the four screws.
Remove the O-ring, spring and core.
5. Install the new main valve assembly in reverse order.
Valve cable
(green leads)
Screws
Valve support
Bmainvalves1202.eps
Core
4. Open the electronics control box. Detach the water heater cable from the Main PCB
connector P14.
5. Unscrew two attachment screws using a 2.5 mm Allen key (white arrows).
6. Cut the instrument console water tubes (one from the magnetic valve assembly, the
other to the instrument console; white arrows).
7. Detach the cup fill water tubes (one from the magnetic valve assembly, other to the
bowl assembly; white arrows).
8. Cut approx. 1 cm (0.4 in.) piece from the tube ends before attaching them to the new
water heater.
9. Install the new water heater in reverse order.
5. Lift up the valve assembly and detach the arm water tube from the nipple.
6. Disconnect the cables and tubes of the console arm cable from the instrument
multiplexer .
7. Remove the friction adjuster by unscrewing two side and one top screws (black
arrows 4 mm Allen key, white 3 mm Allen key).
8. Lift off the whole instrument console (black arrows) and remove the bearing assembly
from the lower joint axle (white arrows).
NOTE
Please note that the console arm cable need not to be removed during this operation.
1. Release the tension of the gas spring (i.e. decrease the tension as much as possible)
as described in section 1.3 "Adjusting balance of OP delivery arm" on page 165.
2. Remove the instrument console and the lower joint axle as described in section 8.1
"Detaching instrument console from OP delivery arm" on page 240.
4. Loosen the long lower joint pin locking screw using a 2 mm Allen key..
5. Unscrew two screws from under the lifting arm using a 3 mm Allen key.
10. Lift off the lower joint axle and the lower joint cover.
11. Slide the arm joint cover from over the upper joint.
12. Lift the console arm to its upmost position. Remove the securing ring from the upper
joint pin.
13. Tap off the upper joint pin from the upper joint axle.
14. Remove the whole lifting arm assembly from inside the lifting arm.
15. Check the condition of the lifting arm assembly. If it is twisted or otherwise damaged,
replace the whole assembly.
16. Remove securing ring from the gas spring pin and remove the gas spring pin.
17. Remove the securing ring from the upper joint pin at the other end of the gas spring
and remove the upper joint pin.
NOTE
Check the condition of the arm cable. Replace also the cable, if needed.
NOTE
Be careful not to break the bowl collar when you replace the console arm cable.
1. Release the tension of the gas spring (i.e. decrease the tension as much as possible)
as described in section 1.3 "Adjusting balance of OP delivery arm" on page 165.
2. Detach the instrument console from the console arm as described in section 8.1
"Detaching instrument console from OP delivery arm" on page 240.
3. It is easier to replace the cable if you remove the gas spring, see section 8.2
"Replacing OP delivery arm gas spring" on page 243.
4. Unscrew four attachment screws of the rotating arm decorative cover and remove the
cover.
5. Pull the console arm cable out from the rotating arm opening.
10. Pull the new console arm cable to inside the delivery arm and into the cuspidor using
the draw cord. Ensure that the cable is not stretched or rubbed by the arm castings in
any position of the arm.
11. Pull the other end of the cables through the rotating arm.
12. Assemble the protective membrane carefully around the console arm joint.
13. Assemble and attach the removed parts in reverse order.
14. Adjust the balance of the console arm as described in section 1.3 "Adjusting balance
of OP delivery arm" on page 165.
15. If necessary, adjust the angle of the instrument console as described in section 1.4
"Adjusting angle of OP delivery arm instrument console" on page 171.
NOTE
Be careful not to break the bowl collar when you replace the console arm.
1. In the case that the dental light / Planmeca Intra / monitor is (are) installed to the
console arm, disconnect the cables from the main control PCB and remove the
equipment with posts from the delivery arm.
2. Disconnect the cables and tubes of the console arm from inside the cuspidor.
3. Rotate the console arm to over the patient chair. Unscrew two of the M8x30 DIN 912
screws that attach the console arm to the cuspidor. Loosen the M8x30 DIN 912 screw
inside the groove of the console arm adapter plate. Lift off the console arm (see
section 8.1 Attaching the OP delivery arm to the cuspidor of the Installation manual).
4. Install the new console arm in reverse order.
5. Adjust the vertical position of the console arm column as described in section 1.1
"Adjusting vertical position of OP delivery arm column" on page 163.
Console arm
Loosen this screw
Comp9_CP3_24.eps
Console arm column
3. Cut the cable ties that attach arm cables to the instrument console.
5. Detach the water nipple and potential equalizer cable from the console.
6. Unscrew the IMUX attachment screws and move the IMUX slightly away from its
position.
10. Remove the upper joint cover ball and level arm cover.
16. Pull the arm downwards and unscrew the level arm nut.
18. Measure the adjustment screw length from the adjustment nut to the upper end.
When attaching the adjustment screw, install the screw to this length.
19. Rotate the adjustment screw out from the adjustment nut.
23. Lift the joint cover and loosen two axle attachment screws.
25. Remove the level arm profile. Support the upper end, the level arm friction guide pins
are unfastened.
NOTE
When attaching the level arm profile, insert the rubber O-rings to the level arm axle to
between the joint and profile as shown on the figure below.
Electrical function:
No control voltage is connected to any of the valves, suc-
tion motor stopped (if controlled by the unit).
Pneumatical function:
All the membranes are open.
Electrical function:
Suction closing The suction motor relay connects on the Main control PCB
membrane and the suction motor starts (if controlled by the unit).
Control voltage is connected to the Microvac and flushing
valves.
Pneumatical function:
The Microvac valve enables the suction closing mem-
brane to connect the vacuum to the upper chamber (and
to the suction hose). The intermediate membrane is
forced to expand by the vacuum in the upper chamber.
The pressure balancing membrane over the intermediate
Pressure balancing membrane moves upwards and blocks the opening
membrane between the upper chamber and the lower chamber.
Saliva and water begins to gather into the upper chamber.
Intermediate
membrane
Pulse valve
Electrical function:
Control voltage is connected to the microvac, pulse and
flushing valves.
Pneumatical function:
The vacuum is connected to the membrane assembly
(drain and intermediate membranes). The intermediate
Breather membrane is forced to shrink by the vacuum. The pres-
valve sure balancing membrane stays in position (due to the
pressure difference between the chambers) preventing
water from flowing to the lower chamber. The drain mem-
brane blocks the opening from the lower chamber to the
drain. The vacuum also closes the breather valve.
Intermediate
membrane
Membrane
assembly
Drain membrane
Pressure
balancing tube
Electrical function:
Control voltage is no longer connected to the pulse valve.
Pneumatical function:
The vacuum in the upper chamber forces the intermediate
membrane to expand, thus the pressure balancing mem-
brane blocks the opening between the upper and lower
chambers. The breather valve opens, since the pulse
valve no longer connects the vacuum to the membrane of
the breather valve. The drain membrane opens allowing
water to flow into the drain.
Electrical function:
The safety switch connects and cuts the control voltage
from all the valves related to the separating tank.
Safety
switch Pneumatical function:
The liquid level in the upper chamber reaches the critical
point and the float pushes the microswitch. All the mem-
Float branes are open and water flows to the drain from both
the chambers. Since the main valve is closed, the vacuum
is not led to the tank nor the suction tubes. The suction will
start again after five seconds delay, supposing that the liq-
uid level has fallen below the critical level.
WARNING
Used separating tank includes saliva and bacteria. Always wear protective clothing
when handling the separating tank.
1. Turn off the unit. In the case that suction motor is not controlled by the unit, turn off the
suction motor.
2. Detach the suction line and the suction filter tubes from the separating tank.
3. Disconnect the Microvac cable from the Microvac valve and disconnect the overfill
switch cables.
4. Disconnect the pulse valve tube from the front of the separating tank.
5. Lift the separating tank from the socket.
6. Install the new separating tank in reverse order.
Suction
line tube
Overfill
switch cables
Microvac
Microvac II socket
Upper chamber
assembly
Fastening clip
Pressure
balancing tube
(upper part)
Quick connector Intermediate section
assembly
Pressure balancing
membrane
Intermediate
membrane
MVac31298.eps
Lower chamber
assembly
MVac61298.eps
Pressure
balancing tube
(lower part)
O-ring 10x3.5
Drain membrane
NOTE
Care must be taken when positioning the suction closing membrane.
NOTE
Tighten the lid assembly screws crosswise.
Screw M3x16
Note the positions of the grooves!
Lid assembly
Suction closing
membrane
MVac41298.eps
1. Press Program.
Cancel OK
NOTE
If the Foot control channel and Transmitter power fields are grey, you must press the foot
control handle to enable the adjusting of the settings.
3. Select the foot control channel by pressing the minus (-) and plus (+) buttons.
You can freely choose between the channels 1 and 16, but, to minimise radio signal
interference, no two foot controls within 10 meters of each other should have the
same channel.
4. Adjust the transmitter power by pressing the minus (-) and plus (+) buttons.
Select the lowest power level and increase it, if needed. The value range is 1 (lowest)
to 6 (highest).
5. Press Battery to show/hide the battery power level symbol to the user on the display.
When the Battery button is blue, the battery power level symbol is shown to the user,
and when it is grey, it is not.
6. Press OK to confirm your selections and exit the programming mode.
To exit the programming mode without confirming, press Cancel.
S78.eps
2. Unscrew the four attachment screws of the cover assembly using a T20 Torx key and
remove the cover assembly.
S79.eps
6. Ensure that the microswitch works by pressing down the foot control handle.
7. Attach the foot control PCB and foot control cover to their positions.
NOTE
After applying Loctite 2700 you have 15 minutes to complete the pedal assembly
replacement.
10. Place the bearing bar to its position and secure it with the two attachment screws.
Apply a drop of Loctite 2400 sealer (or corresponding) to the attachment screws.
Bearing
S81.eps
11. Adjust the pedal height as described in section 3.2 "Adjusting height of foot control
pedal" on page 286.
12. Put the foot control PCB back in position and secure it with the four screws. Connect
the foot control cable to connector J2.
13. Attach the foot control cover.
14. Calibrate the foot control as described in section 3.3 "Calibrating foot control" on page
289.
CAUTION
RISK OF EXPLOSION IF BATTERY IS REPLACED BY AN INCORRECT TYPE.
DISPOSE OF USED BATTERIES ACCORDING TO THE INSTRUCTIONS.
1. Unscrew the eight attachment screws of the bottom plate using a 2.5 mm Allen key
(1).
2. Remove the plate (2).
3. Replace three batteries (3).
3
fcntrl1_120313.eps
3 ADJUSTMENTS
S82_1.eps
Locking nut
Angle adjusting screw
3. Judging by eye, adjust the slope of the sector plate so that it is parallel with the four
PCB supports by tightening the angle adjusting screw.
4. Adjust the angle of the pedal by tightening/loosening the angle adjusting screw. The
front end of the pedal should move 4-5mm vertically (up/down) for convenient use of
the foot control. If the pedal is tilted too much, the front end of the pedals sector plate
could touch the foot control PCB and damage it quickly.
S84_1.eps
4. Loosen the pedal height adjustment on the underside of the foot control approx. 1/4
turn.
S85_1.eps
5. Apply a drop of Loctite 2400 sealer (or corresponding) to the upper adjustment screw.
Do not allow Loctite to get to the pedal axle. Tighten the upper adjustment screw
completely (the foot control pedal does not move freely).
clockwise
Upper adjustment screw
S86_1 .eps
6. Then loosen the upper adjustment screw approx. 1/4 turn so that the foot control
pedal can move freely.
S87.eps
7. Check that the sector plate is approx. 0.5mm - 0.7mm away from the foot control PCB
and that the sector plate is visually estimated straight (0.1mm).
0.5-0.7mm
Sector plate
S88_1.eps
8. Apply a drop of Loctite 2400 sealer to the locking nut (1) and the angle adjusting
screw (2). Loosen the locking nut (1) with a 2 mm screwdriver. Adjust the angle of the
pedal by tightening/loosening the angle adjusting screw (2). The front end of the pedal
should move approx. 4 mm vertically (up/down) for convenient use of the foot control
(3). If the pedal is tilted too much, the front end of the pedals sector plate could touch
the foot control PCB (4) and damage it quickly.
4 3
1
S89_1.eps
4 mm
If you want to exit the foot control calibration mode, press the calibration switch briefly with
the screwdriver.
Remain the foot control pedal at the centre position (1). While in centre position, move the
side knob briefly up (2).
Move the foot control pedal to the rightmost position and hold (1). While holding, move the
side knob briefly up (2). Release the foot control pedal.
Move the foot control pedal to the rightmost position (1). Press the pedal down and hold
(2). While holding, move the side knob briefly up (3). Release the foot control pedal.
3
2
1
Remain the foot control pedal at the centre position (1). Press the foot control pedal
downwards and hold (2). While holding, move the side knob briefly up (3). Release the
foot control pedal.
1
2
Move the foot control pedal to the leftmost position (1). Press the pedal down and hold (2).
While holding, move the side knob briefly up (3). Release the foot control pedal.
1
2
The foot control calibration is now performed and the dental unit will automatically exit the
calibration mode.
Switch the dental unit off and again on.
Test the calibration by driving micromotor instrument and verifying that 100% strength is
reached before mechanical limit in first preventing step.
Check that the foot control side knobs change the GUI displays and the centre knob
activates the functions.
P P P
From cuspidor
P
WATER
A
AIR
B
1 2 3 4 5
Steripump
Syringe water system
control
valve
A
B
C
To instrument
Instr.1
control
valve
A
B
C
Instr.2
control
valve
A
B
C
Instr.3
control
valve
A
B
C
Instr.4
control
valve
air exhaust
1 2 3 4 5
PLANMECA OY
Date
Rev. ECO Description of changes Designer Date 19.11.2012 PNEUMATICAL DIAGRAM
Sorvaajankatu 7 Designed
E Diagram updated V.Poutiainen 19.11.12 00880 Helsinki Finland PLANMECA Compact i Touch V2
Tel +358 20 7795 500 instrument console
Drawn
Fax +358 20 7795 555
www.planmeca.com S.Jnkvaara
[email protected] Checked
V.Poutiainen Rev.E
air air water
drive spray spray
P P P
From cuspidor
P
WATER
A
AIR
B
1 2 3 4 5
Steripump
Syringe water system
control
valve
A
B
C
To instrument
Instr.1
control
valve
A
B
C
Instr.2
control
valve
A
B
C
Instr.3
control
valve
A
B
C
Instr.4
control
valve
air exhaust
1 2 3 4 5
PLANMECA OY
Date
Rev. ECO Description of changes Designer Date 19.11.2012 PNEUMATICAL DIAGRAM
Sorvaajankatu 7 Designed
E Diagram updated V.Poutiainen 19.11.12 00880 Helsinki Finland PLANMECA Compact i Touch V2
Tel +358 20 7795 500 instrument console with Steripump
Drawn
Fax +358 20 7795 555 water system
www.planmeca.com S.Jnkvaara
[email protected] Checked
V.Poutiainen Rev.E
To instrument
Spittoon
console
Instrument OPTIONAL ASSISTANT
Flushing SYRINGE
A B
Suction tubes
Water
Air
Water collector
Coarse filters
OPTIONAL CLEAN /
STERILE WATER 2 cm
SYSTEM
SUCTION RINSE
Drr
suction
valve
BOWL RINSE
separator
Drr station
selector valve
Deposit cup
OPTIONAL
WATER
HEATER
OPTIONAL
AIR
QUICK CONNECTOR
OPTIONAL
WATER
QUICK CONNECTOR
Water collector overflow tube
Suction 19 mm
Drain 19 mm
AIR WATER
PLANMECA OY
Date
Rev. ECO Description of changes Designer Date 19.11.2012 PNEUMATICAL DIAGRAM
Sorvaajankatu 7 Designed PLANMECA Compact i Touch V2
G Diagram updated V.Poutiainen 19.11.12 00880 Helsinki Finland
with Drr CAS1 amalgam
Tel +358 20 7795 500
Fax +358 20 7795 555 Drawn separating system ,
www.planmeca.com S.Jnkvaara Non-Selective suction system
[email protected] Checked
V.Poutiainen Rev.G
To instrument
Spittoon
console
Instrument OPTIONAL ASSISTANT
Flushing SYRINGE
A B
Suction tubes
Water
Air
Water collector
Coarse filters
SUCTION RINSE
Drr Drr
suction suction
valve valve
BOWL RINSE
separator
Drr station
selector valve
Deposit cup
OPTIONAL
WATER
HEATER
OPTIONAL
AIR
QUICK CONNECTOR
OPTIONAL
WATER
QUICK CONNECTOR
Water collector overflow tube
Suction 19 mm
Drain 19 mm
AIR WATER
PLANMECA OY
Date
Rev. ECO Description of changes Designer Date 19.11.2012 PNEUMATICAL DIAGRAM
Sorvaajankatu 7 Designed PLANMECA Compact i Touch V2
G Diagram updated V.Poutiainen 19.11.12 00880 Helsinki Finland
with Drr CAS1 amalgam
Tel +358 20 7795 500
Fax +358 20 7795 555 Drawn separating system ,
www.planmeca.com S.Jnkvaara Selective suction system
[email protected] Checked
V.Poutiainen Rev.G
To instrument
Spittoon
console
Instrument OPTIONAL ASSISTANT
Flushing SYRINGE
A B
Suction tubes
Water
Air
Water collector
Coarse filters
OPTIONAL CLEAN /
STERILE WATER 2 cm
SYSTEM
SUCTION RINSE
Drr
suction
valve
CUP FILLER
BOWL RINSE
separator
Breather
Drr station
selector valve
Deposit cup
Outlet plug
OPTIONAL
WATER
HEATER
OPTIONAL
AIR
QUICK CONNECTOR
OPTIONAL
WATER
QUICK CONNECTOR
Water collector overflow tube
Suction 19 mm
Drain 19 mm
AIR WATER
PLANMECA OY
Date
Rev. ECO Description of changes Designer Date 19.11.2012 PNEUMATICAL DIAGRAM
Sorvaajankatu 7 Designed PLANMECA Compact i Touch V2
00880 Helsinki Finland
with Drr CS1+GF1 amalgam
Tel +358 20 7795 500
Fax +358 20 7795 555 Drawn separating system ,
www.planmeca.com S.Jnkvaara Non-Selective suction system
[email protected] Checked
V.Poutiainen Rev.A
To instrument
Spittoon
console
Instrument OPTIONAL ASSISTANT
Flushing SYRINGE
A B
Suction tubes
Water
Air
Water collector
Coarse filters
OPTIONAL CLEAN /
STERILE WATER 2 cm
SYSTEM
SUCTION RINSE
Drr
suction
valve
CUP FILLER
BOWL RINSE
separator
Breather
Drr station
selector valve
Deposit cup
Outlet pipe
OPTIONAL
WATER
HEATER
OPTIONAL
AIR
QUICK CONNECTOR
OPTIONAL
WATER
QUICK CONNECTOR
Water collector overflow tube
Suction 19 mm
Drain 19 mm
AIR WATER
PLANMECA OY
Date
Rev. ECO Description of changes Designer Date 19.11.2012 PNEUMATICAL DIAGRAM
Sorvaajankatu 7 Designed PLANMECA Compact i Touch V2
B Diagram updated V.Poutiainen 19.11.12 00880 Helsinki Finland
with Drr CS1 amalgam
Tel +358 20 7795 500
Fax +358 20 7795 555 Drawn separating system ,
www.planmeca.com S.Jnkvaara Non-Selective suction system
[email protected] Checked
V.Poutiainen Rev.B
To instrument
Spittoon
console
Instrument OPTIONAL ASSISTANT
Flushing SYRINGE
A B
Suction tubes
Water
Air
Water collector
Coarse filters
SUCTION RINSE
Drr Drr
suction suction
valve valve
CUP FILLER
BOWL RINSE
separator
Breather
Drr station
selector valve
Deposit cup
Outlet pipe
OPTIONAL
WATER
HEATER
OPTIONAL
AIR
QUICK CONNECTOR
OPTIONAL
WATER
QUICK CONNECTOR
Water collector overflow tube
Suction 19 mm
Drain 19 mm
AIR WATER
PLANMECA OY
Date
Rev. ECO Description of changes Designer Date 19.11.2012 PNEUMATICAL DIAGRAM
Sorvaajankatu 7 Designed PLANMECA Compact i Touch V2
B Diagram updated V.Poutiainen 19.11.12 00880 Helsinki Finland
with Drr CS1 amalgam
Tel +358 20 7795 500
Fax +358 20 7795 555 Drawn separating system ,
www.planmeca.com S.Jnkvaara Selective suction system
[email protected] Checked
V.Poutiainen Rev.B
To instrument
Spittoon
console
Instrument OPTIONAL ASSISTANT
Flushing SYRINGE
A B
Suction tubes
Water
Air
Water collector
Coarse filters
OPTIONAL CLEAN /
STERILE WATER 2 cm
SYSTEM
Drr
SUCTION RINSE
suction
valve
Drr VSA
CUP FILLER
BOWL RINSE
system
OPTIONAL
WATER
HEATER
OPTIONAL
AIR
Suction 25 mm
QUICK CONNECTOR
OPTIONAL
WATER
QUICK CONNECTOR
Water collector overflow tube
To VSA suction pump
AIR WATER
PLANMECA OY
Date
Rev. ECO Description of changes Designer Date 19.11.2012 PNEUMATICAL DIAGRAM
Sorvaajankatu 7 Designed PLANMECA Compact i Touch V2
G Diagram updated V.Poutiainen 19.11.12 00880 Helsinki Finland
with Drr VSA system ,
Tel +358 20 7795 500
Fax +358 20 7795 555 Drawn Non-Selective suction system
www.planmeca.com S.Jnkvaara
[email protected] Checked
V.Poutiainen Rev.G
To instrument
Spittoon
console
Instrument OPTIONAL ASSISTANT
Flushing SYRINGE
A B
Suction tubes
Water
Air
Water collector
Coarse filters
SUCTION RINSE
Drr Drr
suction suction
valve valve
Drr VSA
CUP FILLER
BOWL RINSE
system
OPTIONAL
WATER
HEATER
OPTIONAL
AIR
Suction 25 mm
QUICK CONNECTOR
OPTIONAL
WATER
QUICK CONNECTOR
Water collector overflow tube
To VSA suction pump
AIR WATER
PLANMECA OY
Date
Rev. ECO Description of changes Designer Date 19.11.2012 PNEUMATICAL DIAGRAM
Sorvaajankatu 7 Designed PLANMECA Compact i Touch V2
H Diagram updated V.Poutiainen 19.11.12 00880 Helsinki Finland
with Drr VSA system ,
Tel +358 20 7795 500
Fax +358 20 7795 555 Drawn Selective suction system
www.planmeca.com S.Jnkvaara
[email protected] Checked
V.Poutiainen Rev.H
To instrument
Spittoon
console
Instrument OPTIONAL ASSISTANT
Flushing SYRINGE
A B
Suction tubes
Water
Air
Water collector
Coarse filters
OPTIONAL CLEAN /
STERILE WATER 2 cm
SYSTEM
SUCTION RINSE
Drr
suction
valve
Separator
Metasys Typ1
CUP FILLER
BOWL RINSE
separator
Amalgam container
OPTIONAL
WATER
HEATER
OPTIONAL
AIR
QUICK CONNECTOR
OPTIONAL
WATER
QUICK CONNECTOR
Water collector overflow tube
Suction 19 mm
Drain 19 mm
AIR WATER
PLANMECA OY
Date
Rev. ECO Description of changes Designer Date 19.11.2012 PNEUMATICAL DIAGRAM
Sorvaajankatu 7 Designed PLANMECA Compact i Touch V2
G Diagram updated V.Poutiainen 19.11.12 00880 Helsinki Finland
with Metasys Typ1 amalgam
Tel +358 20 7795 500
Fax +358 20 7795 555 Drawn separating system,
www.planmeca.com S.Jnkvaara Non-Selective suction system
[email protected] Checked
V.Poutiainen Rev.G
To instrument
Spittoon
console
Instrument OPTIONAL ASSISTANT
Flushing SYRINGE
A B
Suction tubes
Water
Air
Water collector
Coarse filters
OPTIONAL CLEAN /
STERILE WATER 2 cm
SYSTEM
separator
CUP FILLER
BOWL RINSE
MICROVAC
PULSE
Outlet pipe
Deposit cup
OPTIONAL
WATER
HEATER
OPTIONAL
AIR
QUICK CONNECTOR
OPTIONAL
WATER
QUICK CONNECTOR
Water collector overflow tube
Suction 19 mm
Drain 19 mm
AIR WATER
PLANMECA OY
Date
Rev. ECO Description of changes Designer Date 19.11.2012 PNEUMATICAL DIAGRAM
Sorvaajankatu 7 Designed PLANMECA Compact i Touch V2
G Diagram updated V.Poutiainen 19.11.12 00880 Helsinki Finland
with Microvac amalgam separating system,
Tel +358 20 7795 500
Fax +358 20 7795 555 Drawn Non-Selective suction system
www.planmeca.com S.Jnkvaara
[email protected] Checked
V.Poutiainen Rev.G
To instrument
Spittoon
console
Instrument OPTIONAL ASSISTANT
Flushing SYRINGE
A B
Water
Air
OPTIONAL CLEAN /
STERILE WATER
SYSTEM
No Suction
CUP FILLER
BOWL RINSE
Outlet pipe
Deposit cup
OPTIONAL
WATER
HEATER
OPTIONAL
AIR
QUICK CONNECTOR
OPTIONAL
WATER
QUICK CONNECTOR
Drain 19 mm
AIR WATER
PLANMECA OY
Date
Rev. ECO Description of changes Designer Date 19.11.2012 PNEUMATICAL DIAGRAM
Sorvaajankatu 7 Designed PLANMECA Compact i Touch V2
F Diagram updated V.Poutiainen 19.11.12 00880 Helsinki Finland
No suction system,
Tel +358 20 7795 500
Drawn
Fax +358 20 7795 555
www.planmeca.com S.Jnkvaara
[email protected] Checked
V.Poutiainen Rev.F
To instrument
Spittoon
console
Instrument OPTIONAL ASSISTANT
Flushing SYRINGE
A B
Suction tubes
Water
Air
Water collector
Coarse filters
OPTIONAL CLEAN /
STERILE WATER 2 cm
SYSTEM
Drr
SUCTION RINSE
suction
valve
Wet system
CUP FILLER
BOWL RINSE
Outlet pipe
Deposit cup
OPTIONAL
WATER
HEATER
OPTIONAL
AIR
QUICK CONNECTOR
OPTIONAL
WATER
QUICK CONNECTOR
Water collector overflow tube
Suction 19 mm
Drain 19 mm
AIR WATER
PLANMECA OY
Date
Rev. ECO Description of changes Designer Date 19.11.2012 PNEUMATICAL DIAGRAM
Sorvaajankatu 7 Designed PLANMECA Compact i Touch V2
G Diagram updated V.Poutiainen 19.11.12 00880 Helsinki Finland
with Wet suction system ,
Tel +358 20 7795 500
Fax +358 20 7795 555 Drawn Non-Selective suction system
www.planmeca.com S.Jnkvaara
[email protected] Checked
V.Poutiainen Rev.G
To instrument
Spittoon
console
Instrument OPTIONAL ASSISTANT
Flushing SYRINGE
A B
Suction tubes
Water
Air
Water collector
Coarse filters
SUCTION RINSE
Drr Drr
suction suction
valve valve
Wet system
CUP FILLER
BOWL RINSE
Outlet pipe
Deposit cup
OPTIONAL
WATER
HEATER
OPTIONAL
AIR
QUICK CONNECTOR
OPTIONAL
WATER
QUICK CONNECTOR
Water collector overflow tube
Suction 19 mm
Drain 19 mm
AIR WATER
PLANMECA OY
Date
Rev. ECO Description of changes Designer Date 19.11.2012 PNEUMATICAL DIAGRAM
Sorvaajankatu 7 Designed PLANMECA Compact i Touch V2
I Diagram updated V.Poutiainen 19.11.12 00880 Helsinki Finland
with Wet suction system ,
Tel +358 20 7795 500
Fax +358 20 7795 555 Drawn Selective suction system
www.planmeca.com S.Jnkvaara
[email protected] Checked
V.Poutiainen Rev.I
Centralized STCS
Suction
Valve
Vacuum
Suction Tubes
Over
Leaking
STCS STCS
Clean LQD
Water Valve Valve
Vacuum
Suction
Vacuum
Suction
Valve
Suction
DrainValve
PLANMECA OY
Date
Rev. ECO Description of changes Designer Date 19.11.2012 PNEUMATICAL DIAGRAM
Sorvaajankatu 7 Designed PLANMECA Compact i Touch V2
C Diagram updated V.Poutiainen 19.11.12 00880 Helsinki Finland
Centralized Suction Tube Cleaning System
Tel +358 20 7795 500
Fax +358 20 7795 555 Drawn (STCS)
www.planmeca.com V.Poutiainen
[email protected] Checked
V.Poutiainen Rev.C
Manual STCS
Orotol
Suction rinsing water
Suction
Valve
Vacuum
Suction Tubes
Over
Leaking
STCS
Water Valve
Vacuum
Vacuum
Suction
Suction
DrainValve
Suction
Valve
Suction Water
PLANMECA OY
Date
Rev. ECO Description of changes Designer Date 19.11.2012 PNEUMATICAL DIAGRAM
Sorvaajankatu 7 Designed PLANMECA Compact i Touch V2
C Diagram updated V.Poutiainen 19.11.12 00880 Helsinki Finland
Manual Suction Tube Cleaning System
Tel +358 20 7795 500
Fax +358 20 7795 555 Drawn (STCS)
www.planmeca.com V.Poutiainen
[email protected] Checked
V.Poutiainen Rev.C
Air regulator
1.5bar
Waterline cleaning
solution Cleaning
or Main water in system
switch adapter
Sensor
to Planmeca Compact
dental unit
container
cleaning
Air Water / cleaning solution solution
AIR WATER
PLANMECA OY
Date
Rev. ECO Description of changes Designer Date 19.11.2012 PNEUMATICAL DIAGRAM
Sorvaajankatu 7 Designed PLANMECA Compact i Touch V2
B Diagram updated V.Poutiainen 19.11.12 00880 Helsinki Finland
with water line cleaning system (WCS)
Tel +358 20 7795 500
Drawn
Fax +358 20 7795 555
www.planmeca.com S.Jnkvaara
[email protected] Checked
V.Poutiainen Rev.B
Planmeca Oy | Asentajankatu 6 | 00880 Helsinki | Finland
tel. +358 20 7795 500 | fax +358 20 7795 555 | [email protected] | www.planmeca.com