Kruss - Polarrimeter - Brochure
Kruss - Polarrimeter - Brochure
Kruss - Polarrimeter - Brochure
www.kruess.com
What is polarimetry?
If we regard light as an electromagnetic wave Optically active are chiral substances whose molecules
propagating through space, it is possible to illustrate the can take up different spatial arrangements that can not
phenomenon of polarized light. be aligned with by a rotation. Hence, this is a form of
The oscillation and the direction of propagation of the configuration isomerism. The different molecules of the
wave define a very specific plane. If you would look substance are called enantiomers.
towards the beam you would see this wave as a line
which is inclined at a certain angle in space. The following two enantiomers of lactic acid are an
Regular light includes waves that are inclined in any example of such a molecule:
direction of the space while polarized light is inclined
only at a defined angle.
This polarization can only be achieved by a very close
meshed grating a polarization filter. This one filters O OH O OH
out all waves from the regular light that do not have C C
the same inclination as the filters grating. Is the light
now directed to a second grating which is exactly 90 to HO C H H C OH
the first grating, no light will fall on the detector or the
human eye located behind it. If you place an optically H C H H C H
active substance between the two filters, light will again H H
pass through the second filter.
Optically active substances change the inclination of the S- (+) lactic acid R- (-) lactic acid
lightwave. Depending on the design of the device, the
second filter is rotated (manually or automatically) until
no light will fall on the detector. This technical setup
explains the terms optical rotation, rotation angle as The chiral center of the lactic acid is the middle carbon
well as the terms clockwise and anticlockwise. atom. Since the two enantiomers have different rotation
The two later terms describe the behavior of the wave angles, the polarimetry will provide information about the
mentioned above during the passage through an molecular structure.
optically active substance. Depending on the molecular In addition to the properties of the substance, the
structure of the substance, the direction of the inclination following factors also have an influence on the strength
of the wave is towards the right or towards the left. of the optical activity: the temperature, the wavelength
In order to be able to measure this change, the second of the light, the concentration of the substance and
filter has to be rotated anticlockwise or clockwise. possibly also the solvent.
Typical substances are sugar, lactic acid, tartaric acid but The following applies as well: The longer the path of the
also many other biologically active substances. light through an optically active substance, the larger the
angle of rotation.
polarizer analyzer
light source sample
2
A.KRSS Optronic Polarimeters
Polarimeter accessories 12
3
P8000 Series | Automatic Polarimeters
Chemical industry
Purity control and determination of
The features at a glance concentrations
Analysis of optically active components
Automatic, digital polarimeter with a high (qualitative and quantitative)
accuracy and resolution
Determination of changes in the configuration
Reliable, robust and easy to operate
Monitoring chemical processes
Reduces the measuring time to 1 second
irrespective of the angle of the sample
Sugar industry
Menu-driven touchscreen operation
Quality control of original and end product
Sample parameters can be customized
Determination of fructose and glucose
Ideal for individual, continuous analyses
RS-232, USB and Ethernet interfaces for PC
and printer Food industry
Simple output of all important measuring data Determination of concentration
including date and time Purity control
Low noise level Quality control
5
Main measuring display
Reading [, Z, g/ml],
optical rotation,
concentration
Tube temperature
Sample number
Tube length
Wavelength
Thermostat temperature
Status information
Parameter selection
Sample designation
Comment
Tube length
Wavelength
Measuring unit [, Z, g/ml]
Specific rotation
Help monitor
6
P8000 P8000-T P8100 P8100-T
Measuring method Optical rotation, Optical rotation, Optical rotation, Optical rotation,
int. sugar scale, int. sugar scale, int. sugar scale, int. sugar scale,
concentration, concentration, concentration, concentration,
spec. rotation, spec. rotation, spec. rotation, spec. rotation,
user-defined user-defined user-defined user-defined
Measuring range 90 90 90 90
259 Z 259 Z 259 Z 259 Z
0...99.9 g/ml 0...99.9 g/ml 0...99.9 g/ml 0...99.9 g/ml
Measuring units Angle [, Z], Angle [, Z], Angle [, Z], Angle [, Z],
conc. [g/100 ml], conc. [g/100 ml], conc. [g/100 ml], conc. [g/100 ml],
user defined user defined user defined user defined
Resolution 0.001 0.001 0.001 0.001
0.01 Z 0.01 Z 0.01 Z 0.01 Z
0.1 g/ml 0.1 g/ml 0.1 g/ml 0.1 g/ml
Accuracy 0.003 0.003 0.002 0.002
0.01 Z 0.01 Z 0.01 Z 0.01 Z
0.5 g/100ml 0.5 g/100ml 0.5 g/100ml 0.5 g/100ml
Reproducibility 0.002 0.002 0.002 0.002
Measuring time 90 1s 1s 1s 1s
Light source 1 LED with filter 589 nm 1 LED with filter 589 nm 1 LED with filter 589 nm 1 LED with filter 589 nm
Wavelength 589 nm 589 nm 589 nm 589 nm
others optional others optional others optional others optional
Wavelength selection One fixed wavelength One fixed wavelength One fixed wavelength One fixed wavelength
Connection for Special tube with Special tube with Special tube with Special tube with
temperature sensor temperature sensor temperature sensor temperature sensor temperature sensor
PT100 required PT100 required PT100 required PT100 required
Temperature measurement 0...99.9 C 0...99.9 C 0...99.9 C 0...99.9 C
Temperature resolution 0.1 C 0.1 C 0.1 C 0.1 C
Temperature accuracy 0.2 C 0.2 C 0.2 C 0.2 C
Temperature Tube Tube Tube Tube
reading point
Thermostat Peltier thermostat Peltier thermostat
with water with water
Range of 15...40.0 C 15...40.0 C
temperature control
Accuracy of 0.2 C 0.2 C
temperature control
Max. length of tube 220 mm 220 mm 220 mm 220 mm
Sensitivity min 0.1 % (OD3) min 0.1 % (OD3) min 0.1 % (OD3) min 0.1 % (OD3)
Calibration Automatic Automatic Automatic Automatic
(menu-driven) (menu-driven) (menu-driven) (menu-driven)
Display LCD 5.7 LCD 5.7 LCD 5.7 LCD 5.7
320x240 pixel 320x240 pixel 320x240 pixel 320x240 pixel
color display color display color display color display
Operation Touchscreen Touchscreen Touchscreen Touchscreen
Measured data storage 999 measurements 999 measurements 999 measurements 999 measurements
Interfaces RS-232 RS-232 RS-232 RS-232
USB USB USB USB
Ethernet Ethernet Ethernet Ethernet
Operating voltage 100 V...250 V~ 100 V...250 V~ 100 V...250 V~ 100 V...250 V~
50/60 Hz 50/60 Hz 50/60 Hz 50/60 Hz
7
PS8000 Series | Automatic Sugar Polarimeters
8
P8000-PT Series | Autom. Polarimeter with Peltier Temperature control
9
P3000 Series | Automatic Polarimeters
Pharmaceutical industry
P3001RS P3002RS
Monitoring chemical processes
Purity control and determination of Measuring range 45 45
concentrations Resolution 0.005 0.001
Examination of substances listed in the
Accuracy <0.01 <0.004
German Pharmacopoeia
Measurement duration 1/ s 1/ s
Minimum light transfer
Chemical industry 10 % 1%
of sample
Purity control and determination of Light source Sodium lamp (589 nm)
concentrations
230 or 110 V, AC,
Analysis of optically active components Power supply 50/60 Hz, 100 W
(qualitative and quantitative)
Determination of changes in the configuration
Monitoring chemical processes
Sugar industry
Quality control of original and end product
Determination of fructose and glucose
10
P1000-LED | Manual Lab Polarimeter
Technical data New LED light: saves current and lasts longer
Measuring range 2 semi-circles (0-180)
Glass tubes 100 and 200 mm
Scale division 1
Reading precision 0.05 (with nonius)
Light source LED
Scope of delivery
Polarimeter tubes 100 and 200 mm
11
Polarimeter Accessories
Custom-made product: If this selection of polarimeter tubes and quartz plates does not meet your requirements, please do not
hesitate to contact us!
12
Polarimeter tubes made of stainless steel
Stainless steel flow-through tube with hose connectors and temperature sensor.
With temperature control
Article No Length in mm Volume in ccm
PRM-100-SDTT 100 12.00
PRM-200-SDTT 200 17.00
13
Quartz Control Plates
14
PT31 | Peltier Thermostat The electronic water bath thermostat with Peltier element is a
versatile and powerful device used to control the temperature of
refractormeters, polarimeters etc.
It is extremely robust, compact and easy to operate.
Saves space in the lab due to its small size.
Specifications PT31
Resolution 0.1 C
Heating output 30 W
Cooling efficiency 15 W
Power supply 115-230 ~
Pump pressure 2000 Pa
Pump capacity 20 l/h
Temperature 8-40 C (continuously variable)
Temperature accuracy 0.2 C
Dimensions L/B/H 140 x 80 x 210 mm
Weight 1.5 kg
15
State: June 2010 | subject to technical modification without notice / errors and omissions excepted