Olympus 45MG Manual
Olympus 45MG Manual
Olympus 45MG Manual
45MG
Ultrasonic Thickness Gage
Users Manual
DMTA-10022-01EN Revision A
September 2012
Olympus NDT, 48 Woerd Avenue, Waltham, MA 02453, USA
2012 Olympus NDT, Inc. All rights reserved. No part of this publication may be
reproduced, translated, or distributed without the express written permission of
Olympus NDT, Inc.
This document was prepared with particular attention to usage to ensure the
accuracy of the information contained therein, and corresponds to the version of
the product manufactured prior to the date appearing on the title page. There
could, however, be some differences between the manual and the product if the
product was modified thereafter.
All brands are trademarks or registered trademarks of their respective owners and
third party entities.
DMTA-10022-01EN, Rev. A, September 2012
Table of Contents
iv Table of Contents
DMTA-10022-01EN, Rev. A, September 2012
Table of Contents v
DMTA-10022-01EN, Rev. A, September 2012
10. About Custom Setups for Single Element Transducers ................... 159
10.1 Creating a Custom Setup for a Single Element Transducer .............................. 160
10.2 Quickly Adjusting Waveform Parameters For Single Element Transducers . 162
10.3 About the Detection Modes ................................................................................... 164
10.4 About the First Peak ............................................................................................... 166
10.5 About the Pulser Power ......................................................................................... 167
10.6 About the Time-Dependent Gain Curve ............................................................. 168
10.6.1 About the Maximum Gain .......................................................................... 169
10.6.2 About the Initial Gain .................................................................................. 169
10.6.3 About the TDG Slope ................................................................................... 170
10.7 About the Main Bang Blank .................................................................................. 170
vi Table of Contents
DMTA-10022-01EN, Rev. A, September 2012
List of Abbreviations
List of Abbreviations ix
DMTA-10022-01EN, Rev. A, September 2012
x List of Abbreviations
DMTA-10022-01EN, Rev. A, September 2012
Serial number
label yynnnnnmm
Where:
yy: last two digits of the production year
nnnnn: five-digit non duplicated incrementing number
representing the nth production unit of this product
mm: production month
Instruction label:
Contains:
The C-Tick label indicates that the product complies with the
applicable standard, and establishes a traceable link between
the equipment and the manufacturer, importer, or agent
responsible for compliance, and for placing it on the
Australian market.
Intended Use
Do not use the 45MG for any purpose other than its intended use. It must never be
used to inspect or examine human or animal body parts.
Instruction Manual
This instruction manual contains complete and essential information on how to use
this Olympus product safely and effectively. Before use, thoroughly review this
instruction manual, and use the product as instructed. Keep this instruction manual
in a safe, accessible location.
Apart from the batteries, the 45MG instrument does not contain any user-serviceable
parts.
Safety
This instrument has been tested according to IEC Publication 61010: Safety
Requirements for Electronic Measuring Apparatus. This instruction manual contains
warnings and safety rules that must be observed by the user to ensure safe operation
of the instrument, and maintain it in safe condition. Please read through these
operating instructions before using the instrument.
Safety Symbols
The following safety symbols might appear on the instrument and in the instruction
manual:
The following safety symbols might appear in the documentation of the instrument:
The following safety symbols could appear in the documentation of the instrument:
The IMPORTANT signal word calls attention to a note that provides important
information, or information essential to the completion of a task.
The NOTE signal word calls attention to an operating procedure, practice, or the like,
which requires special attention. A note also denotes related parenthetical
information that is useful, but not imperative.
The TIP signal word calls attention to a type of note that helps you apply the
techniques and procedures described in the manual to your specific needs, or
provides hints on how to effectively use the capabilities of the product.
Warnings
General Warnings
Carefully read the instructions contained in this instruction manual prior to
turning on the instrument.
Battery Precautions
Before disposing of a battery, check your local laws, rules, and regulations, and
follow them accordingly.
When lithium metal batteries are used, transportation of the batteries are
regulated by the United Nations under the United Nations Recommendations on the
Transport of Dangerous Goods. It is expected that governments, intergovernmental
organizations, and other international organizations shall conform to the
principles laid down in these Regulations, thus contributing to worldwide
harmonization in this field. These international organizations include the
International Civil Aviation organization (ICAO), the International Air Transport
Association (IATA), the International Maritime Organization (IMO), the US
Department of Transportation (USDOT), and others.
Please contact the transporter and confirm current regulations before
transportation of lithium metal battery.
Equipment Disposal
Before disposing of this product, make sure that you follow the regulations and rules
of your local government.
WEEE Directive
China RoHS
This equipment generates and uses radio-frequency energy and, if not installed and
used properly (that is, in strict accordance with the manufacturers instructions), may
cause interference. The 45MG has been tested and found to comply with the limits for
an industrial device in accordance with the specifications of the EMC directive.
This equipment has been tested and found to comply with the limits for a Class A
digital device, pursuant to Part 15 of the FCC Rules. These limits are designed to
provide reasonable protection against harmful interference when the equipment is
operated in a commercial environment. This equipment generates, uses, and can
radiate radio frequency energy, and if not installed and used in accordance with the
instruction manual, might cause harmful interference to radio communications.
Operation of this equipment in a residential area is likely to cause harmful
interference, in which case you will be required to correct the interference at your own
expense.
Regulatory Information
The 45MG may display a regulatory screen that lists the specific regulation with
which it complies.
3. Use the up and down arrow keys to scroll through the different REGULATORY
screens.
4. Press [MEAS] to return to the measurement screen.
Warranty Information
Olympus guarantees your Olympus product to be free from defects in materials and
workmanship for a specific period, and in accordance with the conditions specified in
the Olympus NDT Terms and Conditions available at http://www.olympus-
ims.com/en/terms/.
The Olympus warranty only covers equipment that has been used in a proper
manner, as described in this instruction manual, and that has not been subjected to
excessive abuse, attempted unauthorized repair, or modification.
This instruction manual explains the proper operation of your Olympus product. The
information contained herein is intended solely as a teaching aid, and shall not be
used in any particular application without independent testing and/or verification by
Olympus reserves the right to modify any product without incurring the
responsibility for modifying previously manufactured products.
Technical Support
Olympus is firmly committed to providing the highest level of customer service and
product support. If you experience any difficulties when using our product, or if it
fails to operate as described in the documentation, first consult the users manual, and
then, if you are still in need of assistance, contact our After-Sales Service. To locate the
nearest service center, visit the Service Centers page at:
http://www.olympus-ims.com.
1. Instrument Description
This chapter describes the main features and hardware components of the 45MG
instrument.
The PDF file for the 45MG Ultrasonic Thickness Gage Users Manual (P/N:
DMTA-10022-01EN) is included on the documentation CD (P/N: 45MG-MAN-CD
[U8147024]) that is shipped with the 45MG.
The 45MG by Olympus is a handheld ultrasonic thickness gage designed for a wide
variety of thickness-measurement applications. With the 45MG, you only need access
to one side of a part in order to obtain nondestructive measurements of the thickness
of corroded, pitted, scaled, granular, and other difficult materials.
The 45MG displays a thickness readout and offers an optional A-scan view for
waveform verification. The microprocessor of the 45MG continuously adjusts the
receiver setup so that every measurement is optimized for reliability, range,
sensitivity, and accuracy. An optional advanced internal datalogger can store up to
475000 thickness measurements and 20000 waveforms.
Instrument Description 15
DMTA-10022-01EN, Rev. A, September 2012
With the Single Element option, the 45MG operates with a full line of single element
transducers. Operation with dual element transducers is a standard feature.
Depending on the software options, the 45MG is capable of measuring material
thicknesses between 0.003 in. and 25.0 in. (0.08 mm and 635.0 mm). The temperature
range of measured materials may vary between 4 F and 932 F (20 C and 500 C),
depending on the material characteristics, the transducer, and the measurement
mode.
Basic features
Measurement-related status flags and alarms
Quarter VGA color transflective LED back-lite display
Automatic probe recognition for the standard D79X and MTD705 series
transducers
Warning against calibration doubling (for dual element transducers)
Calibration for unknown material sound velocity and/or transducer zero
Fast scan mode with 20 readings per second
Hold or blank thickness display during loss-of-signal (LOS) conditions
Hold minimum and maximum functions
Differential thickness display relative to the set point in absolute values or
percentage ratios
Selection of password-protected lockout functions
Selectable resolution: low of 0.01 in. (0.1 mm), standard of 0.001 in. (0.01 mm), or
high (optional) of 0.0001 in. (0.001 mm); [option not available for all transducers]
Optional features
Single element transducers
Echo-to-Echo and THRU-COAT measurements
High Penetration software for low-frequency single element transducers
A-scan or waveform display
Real-time A-scan waveform display for verification of critical measurements
Manual freeze mode with post processing
Manual zoom and range control of waveform display
Auto hold on loss of signal (LOS) and auto zoom (measured echo centering)
Extended blank
Blank after first received echo in Echo-to-Echo mode
16 Chapter 1
DMTA-10022-01EN, Rev. A, September 2012
The 45MG is a rugged and durable instrument that can be used in harsh
environments. The 45MG was designed to meet the requirement of the IP67 rating
(Ingress Protection rating).
Instrument Description 17
DMTA-10022-01EN, Rev. A, September 2012
Olympus cannot guarantee any level of ingress protection rating once the instrument
seals have been manipulated. You must use sound judgment, and take proper
precautions before exposing the instrument to harsh environments.
To maintain the original level of ingress protection, you are responsible for the proper
care of all routinely exposed membrane seals. Additionally, you are responsible for
returning the instrument to an authorized Olympus service center on an annual basis
to ensure that the instrument seals are properly maintained.
The 45MG front panel features a color display and a keypad. The instrument comes
with a wrist strap. An optional protective rubber boot includes a dust flap seal for the
USB communication connector, strap rings at the four corners, and a stand at the back
of the instrument (see Figure 1-1 on page 18).
Keypad
Stand
Power key
Figure 1-1 The 45MG hardware components Front, top, and side views
18 Chapter 1
DMTA-10022-01EN, Rev. A, September 2012
1.4 Connectors
Figure 1-2 on page 19 illustrates the possible connections between the 45MG and
external devices.
Transducer
USB connector
MicroSD memory
card
The USB and Transmit/Receive transducer connectors are located on the top of the
45MG (see Figure 1-3 on page 19).
Transmit/Receive transducer
connector 2
Instrument Description 19
DMTA-10022-01EN, Rev. A, September 2012
The external MicroSD memory card slot is located behind the battery door (see
Figure 1-4 on page 20).
Battery compartment
The 45MG comes either with the English or the international keypad (see Figure 1-5
on page 21). The functions are the same for both keypads. On the international
keypad, the text labels on many keys are replaced by pictograms. In this document,
keypad keys are referred to using the English label in bold and within brackets (ex.:
[MEAS]).
20 Chapter 1
DMTA-10022-01EN, Rev. A, September 2012
Each key is labelled according to its primary function. The area immediately above
certain keys contains a secondary key function that can be activated by first pressing
[2nd F]. Throughout this document, references to a secondary function are written as
follows: [2nd F], [Primary] (Secondary). For example, the instruction to activate the
REF Value function is written as follows: Press [2nd F], [SAVE] (REF Value).
The [], [], [], and [] keys, along with the [ENTER] key, are used to select menu
items or screen parameters, and to change parameter values. Use the [MEAS] key at
any time to return to the measurement screen. The yellow keys are related to
calibration.
Table 2 on page 21 lists the key functions available on the 45MG keypad. Many
functions are optional, and may not be available depending on which software
options have been purchased.
Instrument Description 21
DMTA-10022-01EN, Rev. A, September 2012
Up arrow
In a screen or a list, moves to the previous
element.
For some parameters, a numerical entry
increases the value.
Down arrow
In a screen or a list, moves to the next
element.
For some parameters, a numerical entry
decreases the value.
22 Chapter 1
DMTA-10022-01EN, Rev. A, September 2012
Instrument Description 23
DMTA-10022-01EN, Rev. A, September 2012
24 Chapter 1
DMTA-10022-01EN, Rev. A, September 2012
This chapter describes how to power the 45MG using different power options.
The power indicator is always present on the left side of the screen, and it shows the
type of power source used. The 45MG can be powered by three AA-size batteries, by a
computer through its USB connector, or by a commercially available 5-volt USB
power supply.
When using batteries, the vertical green bar in the power indicator indicates the
remaining battery level (see Figure 2-1 on page 25). Each graduation mark represents
20 % of the level.
When using a computer or a 5-volt USB power supply, the power indicator is
represented by the USB logo, or by the AC logo, respectively (see Figure 2-2 on
page 26).
Figure 2-2 The power indicator when using a computer or an AC power supply
The 45MG can also be operated using three AA-size nickel-metal hydride (NiMH)
rechargeable batteries. The 45MG does not recharge NiMH batteries. To recharge the
batteries, you must use a commercially available external battery charger (not
included).
26 Chapter 2
DMTA-10022-01EN, Rev. A, September 2012
When the batteries are full (100 % level), the battery power indicator will show full
bars.
Batteries
Battery compartment
door
Battery compartment
gasket
Battery compartment
door lock
Do not replace the batteries while the instrument is on. Dispose of used batteries
promptly. Keep batteries out of reach of children. The batteries used in this device
may present a risk of fire or chemical burn if mistreated. Do not disassemble, heat
above 50 C, or incinerate the batteries.
28 Chapter 2
DMTA-10022-01EN, Rev. A, September 2012
When replacing batteries, make sure that they are full in order to ensure the accuracy
of the estimated remaining battery life shown by the power indicator.
30 Chapter 2
DMTA-10022-01EN, Rev. A, September 2012
The following sections describe the main elements of the 45MG software screens and
menus.
The first screen (see Figure 3-1 on page 31) displays when the Waveform option is
not activated, or when the Waveform option feature is turned off.
OR
The second screen (see Figure 3-2 on page 32) displays when the Waveform
option is purchased and turned on.
Measurement units
Delay Range
Thickness value
Measurement units
Mode and update rate
Gain level
The measurement screen is the main screen of the 45MG software. From anywhere in
the 45MG software, simply press [MEAS] to return to the measurement screen. The
power indicator is always present on the left side of the 45MG screen (see section 2.1
on page 25 for details).
The optional waveform trace, called the A-scan, allows a skilled operator to verify that
the signal used to make a thickness measurement is the correct back-wall echo, and
not noise, material anomaly, or the second multiple echo. The A-scan also enables you
to observe indications that may be too small to be measured by the instrument.
The ID bar (available with the Datalogger option), located at the top of the
measurement screen, contains the ID for the actual thickness-measurement location
and the previously stored value (see Figure 3-3 on page 32). The download indicator
( ) appears when the previously stored thickness measurement comes from a file
rather than a newly acquired value.
32 Chapter 3
DMTA-10022-01EN, Rev. A, September 2012
The indicator ( ) of the optional external MicroSD memory card appears at the top
left side of the screen when a MicroSD memory card is inserted in its slot, which is
located in the battery compartment (see Figure 1-4 on page 20). The 45MG will
recognize an external MicroSD memory card, with a maximum capacity of 2 GB,
upon instrument start-up.
Depending on the context, and on the available functions and options, various
indicators and numeric values appear around the waveform display and around the
main measurement value (see Figure 3-4 on page 33 and Figure 3-5 on page 34). A
help text bar will appear upon certain key presses in order to indicate the keys that
can be used to navigate and make selections in the menu structure.
Rectification indicator
The loss of signal (LOS) appears and the thickness value is cleared when the 45MG no
longer detects ultrasonic echoes (see Figure 3-6 on page 34).
Loss-of-signal (LOS)
indicator (or SAT in the
case of saturation)
In the rare case that the returning signal from the transducer is beyond the upper
receiver-voltage limit, the LOS indicator will be replaced by the SAT flag for
indication of saturation. This situation can normally be corrected by lowering the
pulser voltage.
34 Chapter 3
DMTA-10022-01EN, Rev. A, September 2012
The 45MG displays menus and submenus when you press certain front panel keys.
The menu appears at the top-left corner of the screen (see Figure 3-7 on page 35). If
applicable, a submenu conveniently showing the parameters available for the
highlighted menu command also appears.
The 45MG parameters are logically grouped in parameter screens that can be accessed
using front panel keys or menu commands. Figure 3-8 on page 36 shows the MEAS
parameter screen as an example.
Parameters
The title bar, located at the top of the parameter screen, indicates the parameter
subject. When you access a parameter screen from a menu, a menu button appears on
the left side of the title bar. This menu button can be used to easily return to the
original menu. One or two help text bars appear at the bottom of the screen to indicate
the keys required to select a parameter and edit its value.
36 Chapter 3
DMTA-10022-01EN, Rev. A, September 2012
In the remainder of this document, the above procedure is summarized by the simple
instruction to select a specific parameter or list, and its value. For example:
In the MEAS screen, set MEAS MODE to THICKNESS.
The 45MG offers two methods of editing the value of alphanumeric parameters.
Either the virtual keyboard or traditional method can be used. The virtual keyboard
appears on the screen to display all the available characters that can be used (see
section 3.4.1 on page 37 for details). With the traditional method, you must select each
character from a hidden list of standard sorted letters, numbers, and special
characters (see section 3.4.2 on page 39 for details).
Title bar
Parameter value text
box
Virtual keyboard
38 Chapter 3
DMTA-10022-01EN, Rev. A, September 2012
When editing a multiple line parameter value, highlight DONE and press [ENTER] to
move the cursor to the next line. You can also press [2nd F], [] to accept the text and
move the cursor to the next line.
[]
Cycling begins at A when the
original character is a letter.
Figure 3-10 The character cycle of the traditional text edit method
2. Use the [] and [] keys to select the character you want to enter. Hold down the
key to quickly cycle through the letters, numbers, and special characters.
3. Use the [] keys to move to the next character.
4. Repeat step 2 and step 3 to enter other characters.
5. If you need to move the position of the cursor in the value text box, use the [] or
[] key.
6. To insert a character at the cursor position, press [CAL ZERO].
The character at the cursor, and all those to the right, move one position to the
right, making space for one new character.
7. To delete the character at the cursor position, press [CAL VEL].
The character at the cursor is deleted, and any characters to the right move to the
left by one position.
8. Press [ENTER] to accept the character string and move to the next parameter.
40 Chapter 3
DMTA-10022-01EN, Rev. A, September 2012
4. Initial Setup
4.1 Setting the User Interface Language and Other System Options
The 45MG can be configured to display the user interface in the following languages:
English, German, French, Spanish, Japanese, Chinese, Russian, Swedish, Italian,
Norwegian, Portuguese, or Czech. You can also set the character delimiting the radix
of a number.
The 45MG includes a beep tone generator that confirms when a key is pressed and
notifies you of an alarm condition. The beeper can be set to On or Off.
To save battery power while the instrument is not in use, enable the inactive time
function so that the instrument automatically turns off when no key has been pressed
and no measurement has been made within a period of about six minutes.
Initial Setup 41
DMTA-10022-01EN, Rev. A, September 2012
The 45MG has a built-in date and time clock. You can set the date and the time and
select their format. The 45MG saves all measurement values with their acquisition
date.
42 Chapter 4
DMTA-10022-01EN, Rev. A, September 2012
3. In the CLOCK screen (see Figure 4-3 on page 43), set the parameters to the
current date and time.
4. Set the desired DATE MODE and HOUR MODE, and then select SET.
Initial Setup 43
DMTA-10022-01EN, Rev. A, September 2012
Many of the display parameters are only visible when the Waveform or Datalogger
options have been activated.
2. In the DISPLAY screen (see Figure 4-4 on page 44), select the desired parameter
and value for the following parameters:
WAVEFORM ENABLE allows you to activate or deactivate the waveform
display (see section 6.4 on page 76 for details).
ZOOM OPTION allows you to activate or deactivate the zoom function (see
section 6.4.3.3 on page 82 for details).
RECTIFICATION allows you to select one of the rectification modes (see
section 6.4.1 on page 78 for details).
WAVEFORM TRACE is used to select one of the trace types (see section 6.4.2
on page 80 for details).
COLOR SCHEME is used to select either the INDOOR or OUTDOOR
optimized visibility (see section 4.4.1 on page 45 for details).
DISPLAY BRIGHTNESS is used to select one of the predefined brightness
levels (see section 4.4.2 on page 46 for details).
ID REVIEW LINE is used to activate or deactivate the feature.
44 Chapter 4
DMTA-10022-01EN, Rev. A, September 2012
Figure 4-5 Example of the indoor and the outdoor color schemes
The indoor scheme gives the best visibility when the instrument is used indoors or in
low-light conditions. The indoor scheme displays green characters and waveform
trace on a black background.
The outdoor scheme provides the best visibility when the instrument is used in direct
sunlight. The outdoor mode displays black characters and waveform trace on a white
background. For best readability in this document, most screen captures are shown in
the outdoor color scheme.
Initial Setup 45
DMTA-10022-01EN, Rev. A, September 2012
Reducing the display brightness percentage increases the battery life. Battery life
specifications are based on the backlight brightness set to 20 %.
46 Chapter 4
DMTA-10022-01EN, Rev. A, September 2012
using a single element transducer, it is possible to select 4 Hz, 8 Hz, 16 Hz, or Max (up
to 20 Hz). The measurement update rate indicator is permanently displayed on the
left side of the thickness measurement (see Figure 4-6 on page 47).
Measurement update
rate indicator
The Fast update rate can be as high as 20 Hz, and is dependent on the measurement
type. This option is useful when making high-temperature thickness measurements in
order to help limit transducer contact time, or for applications where the transducer is
scanned over an area in order to determine the minimum thickness.
The 45MG automatically uses the fastest update rate when put into the Minimum or
Maximum mode (see section 7.2 on page 125).
Initial Setup 47
DMTA-10022-01EN, Rev. A, September 2012
The resolution can be reduced in certain applications where the extra precision of the
last digit is not required, or where extremely rough outside or inside surfaces make
the last thickness display digit unreliable.
The Single Element including High Resolution software option (P/N: 45MG-SE
[U8147022]) makes it possible to increase the resolution to either 0.0001 in. or
0.001 mm. High resolution is available for measured thickness smaller than 4 in.
(102 mm). High resolution is not active with low-frequency transducers, or when the
High Penetration software option is activated.
48 Chapter 4
DMTA-10022-01EN, Rev. A, September 2012
5. Basic Operation
The following sections describe the basic operation for the 45MG ultrasonic thickness
gage.
The 45MG operates with a full line of single element (optional software) and dual
element transducers. The 45MG automatically recognizes standard D79X dual
element transducers, and automatically loads the appropriate predefined setup. The
predefined setup contains ultrasonic velocity for the stainless steel step block
supplied with the instrument. With dual element transducers, you need to perform a
transducer zero compensation.
For Single Element or High Penetration software options and single element
transducers, you need to manually recall an appropriate setup. The 45MG is factory
shipped with default conditions for the purchased transducers set up using an
approximate sound velocity for the stainless steel test block provided with the
instrument. The default conditions are selected to facilitate instrument usage for your
applications.
Basic Operation 49
DMTA-10022-01EN, Rev. A, September 2012
Figure 5-2 Initial screen with a standard D79X dual element transducer
50 Chapter 5
DMTA-10022-01EN, Rev. A, September 2012
4. For the Single Element software option and a single element transducer, load an
appropriate setup:
a) Press [2nd F], [FREEZE] (XDCR RECALL).
b) In the menu, select the default choice for the probe type that you use (ex.:
DEFAULT SINGLE ELEMENT).
c) In the DEFAULT screen for the type of probe you are using (see the example
shown in Figure 5-3 on page 51), highlight the setup in the list of available
default setups for the transducer that you are using.
The setups listed as USER-1 through USER-35 can be renamed for special
applications. Refer to chapter 10 on page 159 for more details on setups.
Basic Operation 51
DMTA-10022-01EN, Rev. A, September 2012
d) Press [MEAS] to automatically recall the setup parameters for the chosen
setup and return to the measurement screen.
52 Chapter 5
DMTA-10022-01EN, Rev. A, September 2012
The following procedure is illustrated using a dual element probe and a five-step test
block. Refer to section 5.2 on page 52 for more details on the calibration process.
0.500 in.
Figure 5-4 Sound velocity calibration on a five-step test block Thick part
Basic Operation 53
DMTA-10022-01EN, Rev. A, September 2012
Figure 5-5 Performing the sound velocity calibration on a five-step test block
0.100 in.
Figure 5-6 Sound velocity calibration on a five-step test block Thin part
54 Chapter 5
DMTA-10022-01EN, Rev. A, September 2012
3. Press [MEAS] to complete the calibration and return to the measurement screen.
If you turn off the instrument before pressing [MEAS], the velocity will not be
updated to the new value. Instead, the instrument retains the previous value.
When the 45MG detects an error in the calibration procedure, it successively displays
the following messages in the help text bar before returning to the measurement
screen:
Potential wrong echo detected!
Invalid calibration results!
In such case, the velocity is not changed. The probable cause is that an incorrect
thickness value was entered, or the incorrect back-wall echo was detected.
Basic Operation 55
DMTA-10022-01EN, Rev. A, September 2012
Precision step test blocks are also often used when more than two known thicknesses
are needed (see Figure 5-8 on page 56).
When performing the material sound velocity and zero calibrations, you must use a
test block with the following characteristics:
56 Chapter 5
DMTA-10022-01EN, Rev. A, September 2012
When measurements are made on surfaces that are significantly above room
temperature, the zero point should be recalibrated on a regular basis. This is less
important for transducers with part numbers D790-SM, D791-RM, D797-SM, and
D798 than for other transducers with various types of resin delay lines.
5.2.4 About the Material Sound Velocity and the Zero Calibrations
The 45MG performs a calibration doubling verification to help prevent mis-
calibration on thin samples. Doubling occurs when the instrument measures the time
of flight to the second back-wall echo rather than measuring the time of flight to the
first back-wall echo. The 45MG compares the measured time of flight to the expected
time of flight based on the current sound velocity. The 45MG displays a warning
message if doubling is suspected. Doubling can occur when measuring a thickness
that is below the minimum range of the transducer, or when using a transducer that is
worn out, or getting low in sensitivity.
You can also achieve a material sound velocity and zero calibration procedure by
performing the operations in the reverse order, starting with the zero calibration,
followed by the material sound velocity calibration.
Basic Operation 57
DMTA-10022-01EN, Rev. A, September 2012
58 Chapter 5
DMTA-10022-01EN, Rev. A, September 2012
Message
Calibration
The accuracy of any ultrasonic measurement is only as good as the accuracy and
care with which you calibrate the instrument. The 45MG ships from the factory
with standard setups for a number of transducers and applications. In certain
cases, it may be desirable to optimize these setups for specific measurement
situations. In all cases, it is essential to perform the velocity and zero calibrations
whenever the test material or transducer is changed. Periodic checks with
samples of known thicknesses are recommended to verify that the gage is
operating properly.
Surface roughness of the test piece
The best measurement accuracy is obtained when both the front and back
surfaces of the test piece are smooth. When the contact surface is rough, the
minimum thickness that can be measured will be increased as a result of sound
reverberating in the increased thickness of the couplant layer. Additionally, when
the two test-piece surfaces are rough, the slightly different multiple sound paths
seen by the transducer may cause distortion in the returning echo, resulting in
measurement inaccuracies.
Coupling technique
In mode 1 (contact transducer) measurements, the couplant layer thickness is part
of the measurement, and is compensated by a portion of the zero offset. If
maximum accuracy is to be achieved, the coupling technique must be consistent.
In order to accomplish consistent measurements, use a couplant of reasonably
low viscosity, employ only enough couplant to achieve a reasonable reading, and
apply the transducer with uniform pressure. Practice will determine the degree of
moderate to firm pressure needed to produce repeatable readings. In general,
smaller-diameter transducers require less coupling force to squeeze out excess
Basic Operation 59
DMTA-10022-01EN, Rev. A, September 2012
60 Chapter 5
DMTA-10022-01EN, Rev. A, September 2012
Velocity variations:
An ultrasonic thickness measurement is accurate only to the degree that
material sound velocity is consistent with the instrument calibration. Some
materials exhibit significant variations in sound velocity from point to point.
This happens in certain cast metals due to the changes in grain structure that
result from varied cooling rates, and the anisotropy of sound velocity with
respect to grain structure. Fiberglass can show localized velocity variations
due to changes in the resin/fiber ratio. Many plastics and rubbers show a
rapid change in sound velocity with temperature, and as such, velocity
calibration must be performed at the temperature of the location where the
measurements are made.
Phase reversal or phase distortion
The phase or polarity of a returning echo is determined by the relative acoustic
impedances (density velocity) of the boundary materials. The 45MG performs
computation based on the customary situation, where the test piece is backed by
air or a liquid, both of which have a lower acoustic impedance than metals,
ceramics, or plastics. However, in some specialized cases, such as measurement of
glass or plastic liners over metal, or copper cladding over steel, this impedance
relationship is reversed, and the echo appears phase-reversed. In such cases, it is
necessary to change the appropriate echo detection polarity in order to maintain
accuracy (see section 10.8.1 on page 173). A more complex situation can occur in
anisotropic or in homogeneous materials, such as coarse-grain metal castings or
certain composites, where material conditions result in the existence of multiple
sound paths within the beam area. In such cases, phase distortion can create an
echo that is neither cleanly positive nor negative. Careful experimentation with
reference standards is necessary in such cases in order to determine the effects on
measurement accuracy.
To measure thicknesses
1. Apply couplant to the measurement spot on the test block or tested part.
Basic Operation 61
DMTA-10022-01EN, Rev. A, September 2012
In general, use a thinner couplant (such as propylene glycol, glycerin, or water) for
smooth material surfaces. Rough surfaces require a more viscous couplant, such as
gel or grease. Special couplants are required for high-temperature applications.
2. Using moderate to firm pressure, couple the tip of the transducer to the surface of
the test material, and keep the transducer as flat as possible on the material
surface (see Figure 5-11 on page 62).
0.500 in.
3. Read the measured thickness value for the tested part (see Figure 5-12 on
page 63).
62 Chapter 5
DMTA-10022-01EN, Rev. A, September 2012
The 45MG optional datalogger is a file-based system, in which one file is opened at a
time. The active file stores a measurement at a thickness-measurement location ID.
Each time you press [SAVE], the displayed value is saved to the active file at the
current ID. The ID is automatically incremented for the next measurement. When you
press [FILE], the name of the active file appears in the ID bar above the menu (see
Figure 5-13 on page 63).
Basic Operation 63
DMTA-10022-01EN, Rev. A, September 2012
The NONAME00 increment-type file, starting with the 001 ID, is the active file by
default when you first use the 45MG, or after resetting the instrument memory.
Various types of files can be created, and IDs can be defined to represent various 1-D,
2-D, or 3-D thickness-measurement locations. When the instrument is restarted, it
automatically opens the last file used.
64 Chapter 5
DMTA-10022-01EN, Rev. A, September 2012
6. Software Options
The available software options can be used to increase the capability of the already
versatile 45MG (see Table 3 on page 65).
Software Options 65
DMTA-10022-01EN, Rev. A, September 2012
When you order a software option with your original 45MG purchase, the option
comes preactivated. Software options can also be purchased at a later date. The
software options can be activated by entering an activation code in the instrument,
without the need to return the instrument to the factory. (see section 6.1 on page 66).
Contact your local Olympus representative for more information about ordering
software options. Refer to Table 3 on page 65 for the software-option part numbers.
Each 45MG has a unique serial number code. A provided option key specific to a
particular 45MG activates the purchased software options only on that specific 45MG
unit. A single-option key is capable of activating one, several, or all software options.
66 Chapter 6
DMTA-10022-01EN, Rev. A, September 2012
The OPTION list indicates the actual status of the software options, where the
check mark ( ) indicates that the option is activated.
Alphanumeric serial
number
With dual element transducers, the 45MG offers three echo detection modes that
allow you to measure thicknesses in various material conditions. A description of
each of the three echo detection modes (STANDARD, optional AUTO E-TO-E, and
MANUAL E-TO-E) follows:
STANDARD
The standard echo detection mode measures the thickness based on the time of
flight between the main bang and the first back-wall echo. Use this mode for
uncoated materials.
Software Options 67
DMTA-10022-01EN, Rev. A, September 2012
The DE-STD indicator appears to the left of the thickness reading, and a
triangular echo detection marker appears at the back-wall echo, below the
waveform display (see Figure 6-2 on page 68).
Echo detection marker Dual element (DE) standard (STD) echo detection mode
68 Chapter 6
DMTA-10022-01EN, Rev. A, September 2012
MANUAL E-TO-E (optional and only available when Echo-to-Echo and the
Waveform options are activated)
The manual Echo-to-Echo detection mode also measures the thickness using the
time of flight between two consecutive back-wall echoes. However, the gain and
blanking parameters can also be manually adjusted in this mode. This mode is
useful when material conditions produce noisy signals that could render the
automatic mode less effective.
The DE-MEtoE indicator appears to the left of the thickness reading. The Echo-to-
Echo detection bar is similar to the automatic Echo-to-Echo mode, but includes
the adjustable E1 blank bar that indicates the area to exclude for echo detection
(see Figure 6-4 on page 70). Following the E1 blank, the instrument detects the
next echo with an amplitude of at least 20 % of the waveform display height. In
this mode, press [GAIN/WAVE ADJ], and then use the arrow keys to adjust the
EXT BLANK, the E1 BLANK, and the GAIN parameters.
Software Options 69
DMTA-10022-01EN, Rev. A, September 2012
Echo-to-Echo detection
bar
Dual element (DE)
manual Echo-to-Echo
(MEtoE) detection mode
In severe corrosion situations where valid multiple echoes are not present, you must
use either the standard or optional THRU-COAT mode in order to be able to measure
thicknesses.
Dual element transducers can be used in all three modes, and all measurement,
display, and datalogger functions can be used with the echo-to-echo modes. The
internal datalogger recognizes all necessary echo-to-echo information, and saves it in
order to upload and download thickness, waveform, and setup data.
It is not necessary to switch between echo detection modes when measuring both
coated and uncoated areas, because it is possible to measure uncoated wall
thicknesses using an echo-to-echo mode.
70 Chapter 6
DMTA-10022-01EN, Rev. A, September 2012
2. In the MEAS screen, set the MEAS MODE to the desired echo detection mode:
STANDARD, AUTO E-TO-E (optional), or MANUAL E-TO-E (only available
with the Waveform option).
EXT BLANK
The extended blank creates a blanked zone that begins at the left edge of the
waveform display, and in which no signals are detected. In situations where the
second or third pair of back-wall echoes are stronger or cleaner than the first pair,
use the extended blank to control which pair of echoes are used for measurement.
E1 BLANK
The echo 1 (E1) blank runs for a selected interval following the first detected echo.
Use the E1 blank to exclude any unwanted peaks occurring between the first and
second back-wall echoes. Unwanted peaks may be trailing edges of a large first
echo, or shear-wave reflections on thick test pieces. The E1 blank parameter is
only available in the manual Echo-to-Echo detection mode.
Software Options 71
DMTA-10022-01EN, Rev. A, September 2012
3. Use the [] and [] keys to select either the EXT BLANK or the E1 BLANK
parameter.
4. Use the [] and [] keys to adjust the value to exclude unwanted peaks and to
detect the desired echoes.
72 Chapter 6
DMTA-10022-01EN, Rev. A, September 2012
In some cases, an error can occur if you are using a D790 transducer to measure
thicknesses above 0.7 in. (18 mm). Typically, this error is due to a mode-converted
shear-wave echo that may appear before the second back-wall echo. If this unwanted
echo is larger than the second back-wall echo, the gage measures the distance to it,
which produces a thinner reading.
You can usually distinguish the unwanted shear-wave echo from the correct back-wall
echo by examining the waveform display. The distance between the first and the
second back-wall echoes is the same as the distance between the zero thickness point
and the first back-wall echo. If there is an echo between the first two back-wall echoes,
it is probably a mode-converted shear-wave echo. Use the manual echo-to-echo
detection mode techniques, and manually adjust the E1 blank to eliminate this error
(see section 6.2.1 on page 71). Using the D797 transducer beyond 0.7 in. (18 mm) helps
eliminate the possibility of this error.
In some cases, the second or third back-wall echo is smaller in amplitude than
subsequent echoes, which causes the instrument to give a double or triple reading. If
you are using a D790 transducer, this effect may occur around 0.2 in. (5 mm) on flat,
smooth steel samples. If this occurs, it is clearly visible on the waveform display, and
you can work around it using the manual Echo-to-Echo detection mode, or by moving
the extended blank beyond the previously detected first echo.
When the 45MG cannot make an echo-to-echo reading, the LOS flag appears on the
screen. In this case, the waveform display shows that either no echoes are large
enough to be detected, or that only one echo is detectable. In the latter case, the Echo-
to-Echo detection bar begins at the detected echo, but extends indefinitely to the right.
Software Options 73
DMTA-10022-01EN, Rev. A, September 2012
Increase the gain value to make a good echo-to-echo reading. If this does not help, you
can still obtain an approximate measurement by returning to the standard echo
detection mode.
THRU-COAT is an optional function that measures the true metal thickness of coated
or painted parts. This function only requires a single back-wall echo, and is
recommended for heavy corrosion applications where the outside of the material is
coated or painted. If necessary, the measurement of the coating/paint layer can also be
calibrated to precisely measure the thickness of the coating/paint.
In order for the THRU-COAT function to work properly, the thickness of the coating
or paint must be at least 0.005 in. (0.125 mm) thick. The maximum coating/paint
thickness that THRU-COAT can measure is dependent on the type of coating, but
typically greater then 0.080 in. (2 mm).
If the THRU-COAT function does not display a coating thickness, or if the coating
thickness does not seem realistic, THRU-COAT may not be able to properly resolve
the coating thickness. In such cases, the user should try a different measurement
mode, like Echo-to-Echo or standard.
74 Chapter 6
DMTA-10022-01EN, Rev. A, September 2012
Software Options 75
DMTA-10022-01EN, Rev. A, September 2012
Pressing [2nd F], [CAL VEL] (VEL) opens the VELOCITY screen, in which you can
see and edit the calibrated metal sound velocity. Pressing [2nd F], [CAL VEL] (VEL)
again opens the VELOCITY screen for the calibrated sound velocity through the
coating.
The live Waveform option for the 45MG allows the user to view the live ultrasonic
waveform in order to aid alignment of the transducer during difficult applications.
When this option is activated, the user can switch between the standard thickness
display (see Figure 6-7 on page 77) and the optional waveform thickness display (see
Figure 6-8 on page 78).
[GAIN/WAVE ADJ] allows the user to make setup adjustments that are not
available in standard mode.
With dual element transducers (see chapter 9 on page 143):
Manually adjust the GAIN in 1 dB increments.
Set an extended blank.
Set an Echo1 blank in Manual Echo-to-Echo mode.
Adjust the waveform range and delay.
With the Single Element transducer option (see chapter 10 on page 159):
Adjust the pulse-receive parameters (TVG gain and blanks)
76 Chapter 6
DMTA-10022-01EN, Rev. A, September 2012
Software Options 77
DMTA-10022-01EN, Rev. A, September 2012
78 Chapter 6
DMTA-10022-01EN, Rev. A, September 2012
Full Half
Half+ RF
FULL
This mode shows the negative portion of the echo folded around the baseline, so
that both positive and negative waveform lobes are displayed. This mode
provides the best overall representation of position and magnitude for most
thickness-measurement applications. FULL is the default mode for dual element
transducers.
HALF (NEG indicator)
This mode shows negative waveform lobes as positive, and does not show the
positive lobes.
HALF+ (POS indicator)
This mode shows positive waveform lobes, and does not show the negative
waveform lobes.
RF
This mode shows negative and positive lobes on either side of the baseline. RF is
the default mode for single element transducers.
Software Options 79
DMTA-10022-01EN, Rev. A, September 2012
A filled-in waveform trace is only possible when the waveform rectification is set to
FULL, HALF+, or HALF.
80 Chapter 6
DMTA-10022-01EN, Rev. A, September 2012
Units
Software Options 81
DMTA-10022-01EN, Rev. A, September 2012
When the [GAIN/WAVE ADJ] parameters are active, the range and delay become
items in the parameter list. Use the [] and [] keys to highlight the range parameter,
and the [] and [] keys to adjust the range. Press [MEAS] to stop adjusting the
range.
When the [GAIN/WAVE ADJ] parameters are active, the range and delay become
items in the parameter list. Use the [] and [] keys to highlight the delay parameter,
and the [] and [] keys to adjust the delay. Press [MEAS] to stop adjusting the delay.
6.4.3.3 Activating the Zoom Function (Available Only with the Waveform Option)
The zoom function automatically and dynamically sets the delay and range values to
optimally track and show the detected echo in the waveform display.
82 Chapter 6
DMTA-10022-01EN, Rev. A, September 2012
The resulting zoomed waveform depends on the current measurement mode. The
zoom for D79X dual element transducers and mode 1 single element transducers
centers the first back-wall echo on the screen (see Figure 6-12 on page 83).
The zoom with single element transducers in mode 2 adjusts the waveform range and
delay so that the interface echo and the first back-wall echo appear on the waveform
display (see Figure 6-13 on page 83).
Software Options 83
DMTA-10022-01EN, Rev. A, September 2012
The zoom with single element transducers in mode 3 adjusts the waveform range and
delay so that the interface echo and the second measured back-wall echo appear on
the waveform display (see Figure 6-14 on page 84).
The optional Single Element High Resolution software allows the 45MG to use single
element direct contact, delay line, and immersion transducers. This enables the 45MG
to be used for precision thickness gaging applications. When this option is activated,
the user can select from a set of default single element transducer setups, or create and
store/recall customers single element setups.
The 45MG is not able to automatically recognize which single element transducer is
connected. As such, the appropriate default or customer setup must be recalled for
the single element transducer being used.
84 Chapter 6
DMTA-10022-01EN, Rev. A, September 2012
For single element transducers, the high-resolution software option provides the
added ability to display thickness readings with a high resolution of either 0.0001 in.
or 0.001 mm. High resolution is not available for all transducers and measurement
modes, and is also limited to the maximum thickness. Although the 45MG is able to
display thickness readings in high resolution, the measurement accuracy is highly
dependent on the material, geometry, surface condition, and temperature, and also
needs to be determined on a case-by-case sample evaluation basis.
Single element transducer in the frequency range of 2.25 MHz to 30.0 MHz
Thickness measurement below 4.00 in. or 100 mm
High resolution is not available for the following transducers or measurement
conditions:
Software Options 85
DMTA-10022-01EN, Rev. A, September 2012
With a M2008 transducer only, press [2nd F], [CAL ZERO] (Do ZERO) at any time to
auto-adjust the zero offset and compensate for temperature changes in the delay line.
This section describes how to use the 45MG internal datalogger to organize your data.
86 Chapter 6
DMTA-10022-01EN, Rev. A, September 2012
The file also contains header parameters that can be defined to better describe the
content of the file. In the file, you can organize the ID range, select the data format,
and select the saved data. Table 5 on page 88 summarizes the contents of a file, and
indicates where you can find more information.
Software Options 87
DMTA-10022-01EN, Rev. A, September 2012
You can identify the datalogger parameters in the ID bar at the top of the
measurement screen (see Figure 6-16 on page 88).
With each measurement, the 45MG also stores a complete description of the
measurement conditions. Table 6 on page 89 describes the additional data stored with
each thickness measurement and with each waveform.
88 Chapter 6
DMTA-10022-01EN, Rev. A, September 2012
With the datalogger, you can easily create a data file (see section 6.7.2 on page 90),
perform a number of file operations (see section 6.7.4 on page 103), and perform data
operations (see section 6.7.5 on page 112).
Software Options 89
DMTA-10022-01EN, Rev. A, September 2012
It is also possible to create a 45MG data file from a computer using the GageView
interface program. Refer to the GageView Interface Program Users Manual (P/N:
910-259-EN [U8778347]) for details.
90 Chapter 6
DMTA-10022-01EN, Rev. A, September 2012
At any time, you can press [2nd F], [] or [2nd F], [] to scroll between parameters on
the screen.
Incremental
Sequential
Sequential with Custom Point
2-D Matrix Grid
Boiler
Increments only digits and letters, not punctuation marks or other special
characters.
Software Options 91
DMTA-10022-01EN, Rev. A, September 2012
To configure the gage to increment through a range of numbers several digits wide
while beginning with a single-digit ID number, the maximum number of digit
positions must be entered initially using leading zeroes (see examples in Table 7 on
page 92).
92 Chapter 6
DMTA-10022-01EN, Rev. A, September 2012
Figure 6-18 The CREATE screen for the incremental data file type
Software Options 93
DMTA-10022-01EN, Rev. A, September 2012
Figure 6-19 Selecting the ID range for the sequential file type
94 Chapter 6
DMTA-10022-01EN, Rev. A, September 2012
6.7.2.4 About the Sequential with Custom Point Data File Type
The sequential with custom point (SEQ+CUSTOM PT) data file type is defined by a
starting and an ending ID number, plus a series of custom points. The resulting file is
inclusive of the start and end points, and all points in between. In addition, multiple
thickness readings per ID number location are assigned using the assigned custom
points.
Use the sequential with custom point data file type when you want, for example, to
measure along a pipe or tube on which, at each ID number location, you can take
measurements at the top, bottom, left, and right of the pipe (see an example in Table 9
on page 95).
The allowable number of characters for each custom point depends on the number of
ID characters defined in the start and end ID values. The total number of characters of
the ID value plus the custom points cannot exceed 20. For example, when the start
and end ID values are seven characters long, as shown in the example in Table 9 on
page 95, the maximum allowable length for each custom point is thirteen (20 7 = 13).
Software Options 95
DMTA-10022-01EN, Rev. A, September 2012
Figure 6-20 Configuring ID range for a sequential with custom points data file type
A 2-D (two-dimensional) sequence begins with the ID number that refers to the first
column and the first row (see Figure 6-21 on page 97). The column (or row) then
increments one value at a time until the sequence reaches the last column (or row)
value, while the other dimension value stays constant. At this point, the other
dimension increments from its first to last value. This continues until the ID number
that refers to the last column and the last row is reached. You have the option to select
whether the columns or the rows will be incremented first.
96 Chapter 6
DMTA-10022-01EN, Rev. A, September 2012
A grid structure may associate one dimension of the grid with the physical parts
whose wall thickness is to be measured. The particular measurement points on each
part are then associated with the other dimension of the grid (see the example in
Figure 6-22 on page 98).
Software Options 97
DMTA-10022-01EN, Rev. A, September 2012
Measurement location
ID
Row
Column
Alternatively, the rows and columns of a grid may refer to a two-dimensional map of
measurement points on the surface of one part. In this case, a different grid is created
for each part (see the examples in Figure 6-23 on page 99).
98 Chapter 6
DMTA-10022-01EN, Rev. A, September 2012
Name: Elbow
Rows: 01 through 10
Columns: A through E
IDs: Elbow/A0 through
Elbow/E10
Name: Tee
Rows: 1 through 4
Columns: 1 through 3
IDs: Tee/11 through Tee/34
Software Options 99
DMTA-10022-01EN, Rev. A, September 2012
Figure 6-24 Configuring the ID range for a 2-D grid data file type
The 45MG has the capability to add a row or a column, and to change the
incrementation direction after a grid file is created (see section 6.7.4.4 on page 106 for
details).
Elevation
The first dimension refers to the physical distance from the bottom of the boiler to
the top.
100 Chapter 6
DMTA-10022-01EN, Rev. A, September 2012
Tube number
The second dimension refers to the number of the specific boiler tube to be
inspected.
Custom points
The third dimension refers to the actual thickness reading location at the specified
elevation on the specified tube.
The three dimensions are combined in a single ID number to precisely identify the
exact location of each thickness reading. Table 10 on page 101 shows an example in
which incrementation for the custom points was chosen first, the tube number second,
and the elevation third.
Figure 6-25 Configuring the ID range for a boiler data file type
102 Chapter 6
DMTA-10022-01EN, Rev. A, September 2012
The default file data mode can be set to the most frequently used.
To open a file
1. Press [FILE], and then select OPEN.
2. In the OPEN screen (see Figure 6-27 on page 104):
a) In SORT BY, select the order in which the files appearing on-screen are sorted
(by NAME or by DATE CREATED).
b) In the list of files, select the file you want to open.
The descriptive header for the highlighted file name appears on the lower
section of the display.
c) Select OPEN to return to the measurement screen with the selected file set as
the active file, and the ID number set to the first ID number in the file.
104 Chapter 6
DMTA-10022-01EN, Rev. A, September 2012
2. Press [2nd F], [FILE] (ID#) to open the file review screen (see section 6.7.6 on
page 112).
The file copy function can only be used to copy an existing file in the internal memory
to the internal memory. Use the file import and export functions to copy data to and
from the internal memory and the external MicroSD card.
To copy a file
1. In the measurement screen, press [FILE], and then select COPY.
2. In the COPY screen (see Figure 6-28 on page 106):
a) In the list, select the source file.
b) In COPY NAME, enter the file name for the destination file.
c) Set COPY THICKNESS DATA? to YES when you also want to copy the
thickness readings from the original file into the new file.
d) Select COPY.
3. Open the newly created file in order to make it the active file (see section 6.7.4.1
on page 104.)
File name
File description
Inspector ID
Location note
Delete protection (on/off)
End row, column, or point of a grid file
Incrementing order of a grid file
The edit function does not allow you to edit the file type, and cannot be used to edit
individual measurement identifiers (ID), or actual thickness readings.
106 Chapter 6
DMTA-10022-01EN, Rev. A, September 2012
When scrolling through the file names, a descriptive header for the highlighted file
name appears on the lower section of the display. This information can assist in
selecting the proper file if you are uncertain of the exact file name.
3. For a grid file, select CONTINUE, and then perform the following actions in the
second page of the EDIT screen (see Figure 6-30 on page 108):
a) Increase the END COLUMN and END ROW values as needed. These values
cannot be decreased.
b) If needed, change the INC 1ST BY value.
c) Select UPDATE.
108 Chapter 6
DMTA-10022-01EN, Rev. A, September 2012
OR
Set DELETE MODE to FILE if you want to completely erase the file from the
memory.
f) Press ENTER to confirm the operation.
If you select multiple files for deletion, and some of those files are delete-protected,
the 45MG will only delete the files that are not delete-protected.
3. Edit the STARTING ID and the ENDING ID values to define the range of IDs
you want to delete from the file.
4. Select DELETE.
Using the internal memory reset or master reset erases all files, and the data contained
in those files. The deleted files and the data they contain cannot be recovered. The
datalogger will be completely empty after this procedure.
110 Chapter 6
DMTA-10022-01EN, Rev. A, September 2012
When the ID overwrite protection is enabled, a message appears (see Figure 6-35 on
page 112) on the help text bar when you perform a save to inform you that the existing
thickness readings/waveforms will be overwritten. Select YES to replace the previous
reading with the new one, or NO to leave the original value.
112 Chapter 6
DMTA-10022-01EN, Rev. A, September 2012
Figure 6-36 on page 113 shows an example of the ID review screen and describes its
contents. The area under the waveform is reserved for status flags describing the
displayed stored thickness values. The flags are the same single letter abbreviations
for status words transmitted by the gage using the send commands (see section 11 on
page 181).
Active ID
m: Minimum value
m: MIN / MAX M: Maximum value
Change the active ID to quickly jump to an existing ID. This is useful when you
are using a large file, and it would take too long to locate the desired ID using the
arrow keys.
Change the active ID to a new ID that does not yet exist in the file. This mode is
useful when you want to include additional measurement points in the active file.
Additional IDs can be added anywhere in the database (beginning, middle, and
end).
114 Chapter 6
DMTA-10022-01EN, Rev. A, September 2012
5. Press [MEAS] to return to the measurement screen with the new active ID.
6. When the edited ID is not in the database, the help text bar message shown in
Figure 6-38 on page 115 appears. Select INSERT to insert the new ID in front of
the active ID.
OR
Select APPEND to add the new ID to the end of the file.
Figure 6-38 The message when the edited ID is not in the database
The easiest way to replace a thickness reading is to save a new measurement at the
desired ID in the measurement screen. If you do not want to save a measurement at a
specific ID, press [SAVE] in the measurement screen when you are not taking a
measurement. This stores an LOS condition and -.- at the specific ID number.
116 Chapter 6
DMTA-10022-01EN, Rev. A, September 2012
File summary
Shows basic statistics for the file (minimum thickness and location, maximum
thickness and location, and high and low alarm conditions along with the mean,
median, and standard deviation).
Minimum/Maximum summary
Shows a list of ID number locations that have the minimum and maximum
thickness in a file.
Alarm summary
Shows a list of all the ID number locations where a low alarm and high alarm
occurred.
File comparison
Allows you to select two files and compare them. The first file contains the
previous inspection data, and the second file contains the current inspection data.
The report indicates the maximum wall loss and any area of greater wall
thickness (growth), and their ID number locations.
Minimum Review
Allows you to select a file, and then review all the minimum thickness locations in
the file. In addition to verifying the thickness at all minimum locations, you can
replace these minimum thickness locations if necessary.
To generate a report
1. In the measurement screen, press [FILE], and then select REPORT.
2. In the submenu, select the desired type of report. If you select:
FILE SUMMARY, go to step 3
MIN/MAX SUMMARY, go to step 4
FILE COMPARISON, go to step 5
ALARM SUMMARY, go to step 6
MIN REVIEW, go to step 7
3. In the FILE SUMMARY screen (see Figure 6-39 on page 118):
a) Select the file for which you want to create the report.
b) Select REPORT.
The FILE SUMMARY report result screen opens (see Figure 6-40 on
page 118).
118 Chapter 6
DMTA-10022-01EN, Rev. A, September 2012
b) Select REPORT.
The MIN/MAX SUMMARY report result screen opens with the first min ID #
highlighted (see Figure 6-41 on page 119).
c) Press [2nd F], [] or [2nd F], [] to move between the #MINS and #MAXS
lists.
d) Select CANCEL to return to the measurement screen, or NEW REPORT to
generate another report.
5. In the FILE COMPARISON screen (see Figure 6-42 on page 120):
a) In the upper list, select the reference file you want to use in the comparison.
b) In the lower list, select the comparison file (containing newer data for the
same measurement points).
c) Select REPORT.
The FILE COMPARISON report result screen opens with the first maximum
wall-loss ID highlighted (see Figure 6-43 on page 120).
d) Review the maximum wall-loss location list, and the maximum wall growth
location list.
e) Select CANCEL to return to the measurement screen, or NEW REPORT to
generate another report.
6. In the ALARM SUMMARY screen:
a) Select the file for which you want to generate the report.
b) Select REPORT.
The report page of the ALARM SUMMARY screen opens with the first low
alarm location ID highlighted (see Figure 6-44 on page 121).
120 Chapter 6
DMTA-10022-01EN, Rev. A, September 2012
d) Couple the probe back to the minimum ID location to verify the thickness,
and then press [SAVE] to store the new measurement.
e) Use the [] and [] keys to move to other minimum review list IDs.
Press [MEAS] to exit the minimum review.
122 Chapter 6
DMTA-10022-01EN, Rev. A, September 2012
This chapter describes how to use special 45MG functions and modes. The 45MG has
many thickness-measurement features. Although the features outlined in this section
are not required for basic thickness operation, they can make the gage a more versatile
instrument.
Activating and Configuring a Differential Mode (see section 7.1 on page 123).
Using the Minimum, Maximum, or Min/Max Thickness Mode (see section 7.2 on
page 125).
Preventing False Minimum/Maximum Thickness Readings (see section 7.3 on
page 127).
Using Alarms (see section 7.4 on page 128).
Locking the Instrument (see section 7.5 on page 132).
Freezing the Measurement or Optional Waveform (see section 7.6 on page 134).
The 45MG includes differential modes that can be used to easily compare the actual
measurement with an entered reference value. The actual thickness measurement
appears on the thickness display, and the differential value appears in the differential
display area (see Figure 7-1 on page 124).
Actual thickness
Differential value
Figure 7-1 Normal differential mode (shown with Waveform option activated)
The units and resolution of the differential thickness are the same as those selected for
the thickness measurement.
When you press [SAVE] (with the optional datalogger) while in the NORMAL or
% RATIO differential modes, the 45MG saves the actual thickness value along with
the D flag indicating that the Differential mode is active.
% RATIO: Shows the actual thickness along with the percent difference
from the REF VALUE entered.
124 Chapter 7
DMTA-10022-01EN, Rev. A, September 2012
minimum or maximum thickness value replaces the main thickness display when the
transducer is uncoupled, or when a loss of signal (LOS) occurs. The replacement value
appears in an outlined font.
Figure 7-3 Displaying the minimum and/or maximum thickness (shown with the
Waveform option activated)
126 Chapter 7
DMTA-10022-01EN, Rev. A, September 2012
The fastest display update rate is automatically activated when entering the minimum
or maximum thickness mode. When leaving the mode, the display update rate is
restored to its previous state.
The minimum and maximum thickness modes display the smallest and largest
thickness value measured from the time you activate or reset the minimum mode. The
modes are useful when determining the thinnest/thickest reading obtained while
making a series of readings on a test piece.
A false minimum or maximum reading can occur when you lift the transducer from
the test piece. This is due to excess couplant (particularly on smooth surfaces), which
causes the instrument to read the thickness of a couplant drop when you lift the
transducer.
Any one of the 45MG alarm modes can be activated to help you identify when the
actual thickness measurement is above or below editable reference values.
The HIGH or LOW alarm indicator flashes with a red background at the bottom
right corner of the measurement screen (see Figure 7-4 on page 128).
The thickness value appears in red.
When the beeper is active (see section 4.1 on page 41), the 45MG emits a long
beep.
The thickness value and alarm indicator only appear in color when the indoor color
scheme is active (see section 4.4.1 on page 45 to change the color scheme).
Previous thickness
Thickness reading
(turns red during an
alarm condition)
High alarm indicator
Figure 7-4 Example of a HIGH alarm indicator (shown with the Waveform option
activated)
128 Chapter 7
DMTA-10022-01EN, Rev. A, September 2012
When the Datalogger option has been purchased, the datalogger records an alarm
condition in the second status box for all stored measurements. An A indicates the
alarm mode, an L indicates a low alarm condition, and an H indicates a high alarm
condition.
There are three different alarm modes (STANDARD, B-SCAN, and REDUCTION
RT):
STANDARD
The standard alarm warns you when the actual measured thickness is below a
low reference value, or above a high reference value. The reference values are
thickness set points using the current instrument units and resolution.
The B-SCAN and REDUCTION RT alarms are only available when the B-scan or
reduction rate (DIFF mode) are activated. These functions must be turned on first
before the B-SCAN and REDUCTION RT alarms can be activated.
B-SCAN
The B-scan alarm mode is similar to the standard alarm mode, except that in the
B-scan alarm mode, lines are shown in the B-scan grid when the reference values
fall within the B-scan thickness range (see Figure 7-5 on page 129). In addition, the
alarms are operational while you review B-scan thicknesses in the B-scan freeze
review mode.
The thickness value and the alarm indicator only appear in color when the indoor
color scheme is active (see section 4.4.1 on page 45 to change the color scheme).
REDUCTION RT
The REDUCTION RT option only appears when the active file is configured with
FILE DATA MODE set to REDUCTION RT. You set the reduction percentage for
the low (YELLOW ALARM) and high (RED ALARM) limits, following which the
instrument displays (see Figure 7-6 on page 130):
The RED indicator for reduction rates that are greater than or equal to the
RED ALARM value.
The YEL indicator for reduction rates that are between the YELLOW ALARM
and the RED ALARM values.
The GRN indicator for reduction rates that are below the YELLOW ALARM
value.
Figure 7-6 YEL (yellow) and RED alarm indicators (shown with the Waveform
option activated)
130 Chapter 7
DMTA-10022-01EN, Rev. A, September 2012
The B-SCAN option only appears when the B-scan mode is active (see section 9.3.1 on
page 150).
The REDUCTION RT option only appears when the active file is configured with the
FILE DATA MODE parameter set to REDUCTION RT (see section 6.7.3 on page 102).
3. When ALARM MODE is set to STANDARD or B-SCAN, set the LOW ALARM
and the HIGH ALARM values.
OR
When ALARM MODE is set to REDUCTION RT, set the YELLOW ALARM and
the RED ALARM values.
4. Press [MEAS] to return to the measurement screen.
Alarm reference values that were entered in one unit system are displayed as the
equivalent value when the alternate units are selected.
The 45MG is equipped with an instrument lock that can be used by a supervisor to
restrict access to selected functions. The supervisor can also enter a password to
prevent other users from unlocking the functions. Once a password has been set, you
must reenter the password before you can lock or unlock any function.
When locking [CAL VEL] and [CAL ZERO], the function ([2nd F], [CAL ZERO]
(Do ZERO) is still available.
Anytime a user attempts to use a locked function, a message appears in the help bar to
indicate that the function is locked (see Figure 7-8 on page 133).
132 Chapter 7
DMTA-10022-01EN, Rev. A, September 2012
If you forget your password, you can unlock the instrument and deactivate the
password by entering the master password: OLY45MG.
To change the password, you must first use the master password to deactivate the
password, and then set a new password.
3. Select DONE to activate the instrument lock and return to the measurement
screen.
Pressing [FREEZE] stops updating the display and optional waveform, and keeps the
waveform and thickness on the display even if you move or uncouple the transducer.
The freeze indicator ( ) appears on the lower right side of the display when the
freeze function is active.
The freeze function is convenient for easily setting the gain parameters, reviewing a
B-scan, or performing a high-temperature measurement without having to keep the
transducer coupled to the test piece.
In order to prevent recording wrong minimum or maximum values, you may also
want to use the freeze function to pause measurements before uncoupling the
transducer from the test piece.
134 Chapter 7
DMTA-10022-01EN, Rev. A, September 2012
Pressing [MEAS] or [SAVE] (when the Datalogger option has been activated) also
unfreezes the display.
136 Chapter 7
DMTA-10022-01EN, Rev. A, September 2012
The MEAS setup is the most commonly used setup menu screen, and is used to access
global parameters concerning the instrument measurement features.
In the MEAS screen, some parameters differ depending on whether a single or dual
element transducer is connected (see Figure 8-1 on page 138).
2. In the MEAS screen for single element transducers, in the MEAS MODE text box,
select the instrument measures and displays from the following options:
THICKNESS: The thickness of the inspected part.
VELOCIMETER: The sound velocity in the material of the inspected part.
TIME OF FLIGHT: The round trip time of flight (TOF) of the sound in the
inspected part
OR
In the MEAS screen for dual element transducers, in the MEAS OPTION text
box, select the measurement method to be used by the instrument from the
following options:
STANDARD: Measure to the first back-wall echo.
AUTO ECHO-TO-ECHO (optional): Measure between the successive back-
wall echoes.
MANUAL ECHO-TO-ECHO (optional with the Waveform): Measure
between the multiple back-wall echoes with the manual control.
3. In the UNIT TYPE text box, select between the INCH (English) or MILLIMETER
(metric) units.
Time-of-flight measurements are always expressed in microseconds.
4. In the RESOLUTION text box, select between STANDARD, LOW, or HIGH (see
section 4.6 on page 48 for details).
5. In the MIN/MAX text box, select either MIN (minimum), MAX (maximum), or
BOTH to activate a mode (see section 7.2 on page 125 for details).
138 Chapter 8
DMTA-10022-01EN, Rev. A, September 2012
6. In the HOLD BLANK text box, configure the instrument to either continue to
show (HOLD), or not hold (BLANK) the last measured thickness and waveform
while a loss of signal (LOS) occurs.
The MIN/MAX and HOLD BLANK functions are mutually exclusive. You must set
MIN/MAX to OFF in order to change the HOLD BLANK function. Similarly, you
must set HOLD BLANK to BLANK in order to change the MIN/MAX function.
7. In MEASURE RATE, adjust the measurement update rate (see section 4.5 on
page 46 for details).
8. For single element transducers only, set AVERAGING to OFF in order to turn
thickness averaging off, set AVERAGING to ON to perform a running average of
the last five thickness readings, or set AVERAGING to On-QBar to display a
quality measurement Q-bar below the measurement screen, indicating the
stability of the averaged reading.
9. Set ID OVERWRITE PROT to ON (with the Datalogger option only) if you want
to see a confirmation message in the help bar when attempting to save a
measurement reading in an ID that already contains a value (see section 6.7.5 on
page 112 for details).
10. For single element transducers only, set QUICK SETUP RECALL to ON in order
to activate the quick recall of the first four custom setups by using a combination
of the [2nd F] and arrow keys (see section 10.10 on page 178 for details).
11. For single element transducers only, set AGC to ON to configure the automatic
gain control (AGC) function to automatically bring all measured back-wall echoes
to the same amplitude.
The AGC function works well for most standard thickness gage applications, and is
turned on by default. In some thickness applications, the receiver gain is set at or near
its maximum value. In such cases, turn off the AGC function to prevent reading
instability.
The SYSTEM screen, allows you to configure many 45MG system parameters.
The calibration and setup parameters are also always saved/sent along with the
thickness measurement.
140 Chapter 8
DMTA-10022-01EN, Rev. A, September 2012
7. Set DEFAULT FILE MODE to the desired default file data mode when you create
a file (see section 6.7.3 on page 102 for details).
8. Set TEXT EDIT MODE to the appropriate option for entering alphanumeric
values. The two available options are the virtual keyboard (VIRTUAL) and the
legacy character wheel selection (TRADITIONAL) [see section 3.4 on page 37 for
details].
9. Set SCR TO SD CARD to ON to enable the 45MG to create a BMP image file on
the external MicroSD card for the actual screenshot when you press [2nd F],
[SETUP] (see section 11.4.2 on page 189 for details).
10. Press [MEAS] to return to the measurement screen.
The 45MG features a USB port that can be used to connect the instrument to a
computer. When connected to a computer, the 45MG can send and receive data, or be
remotely controlled by the computer (when the Datalogger option has been
activated).The remote command document and the FTP (file transfer protocol)
document are available upon request.
142 Chapter 8
DMTA-10022-01EN, Rev. A, September 2012
Adjusting the Gain with Dual Element Transducers (see section 9.1 on page 143)
Adjusting the Extended Blank with Dual Element Transducers (see section 9.2 on
page 145)
About the B-Scan (see section 9.3 on page 147)
About the DB Grid (see section 9.4 on page 153)
With D79X series dual element transducers, you can manually adjust the gain by
pressing [GAIN/WAVE ADJ]. The 45MG offers two types of gain adjustments:
The standard feature allows the user to set the gain to HIGH (+10 dB),
standard (default), and LOW (6 dB).
When the waveform (option) is activated, the user can adjust the gain in 1 dB
increments.
The gain value is expressed in decibels (dB), and appears near the lower-right corner
of the display (see Figure 9-1 on page 144).
When activated, the manual gain adjustment also modifies the way in which echoes
are shown on the waveform display (optional). With the default automatic gain, the
measured echo peak is visible on the display, enabling you to readily observe the
measured echo position, independent of its strength or the gain. When you manually
adjust the gain, the displayed echo height changes proportionally to the actual echo
amplitude at the receiver output, enabling you to readily observe gain changes.
1. Press [GAIN/WAVE ADJ] to change the gain from default gain to HIGH (+10 dB).
2. Press [GAIN/WAVE ADJ] again to change the gain to LOW (6 dB).
3. Press [GAIN/WAVE ADJ] a third time to change the gain back to default gain (no
gain showing).
When the optional waveform is activated:
144 Chapter 9
DMTA-10022-01EN, Rev. A, September 2012
The gain value appears near the lower-right corner of the display. The waveform
display changes to the amplitude proportional to the gain mode.
2. Press [] or [] to adjust the gain in 1 dB steps.
The gain value and the height of the echoes change accordingly.
Increase the gain value when the peak amplitude of the echo that should be
detected is below 20 % of the waveform display height.
This prevents the gage from reading too high a value when skipping one
back-wall echo, and from measuring twice the correct thickness (doubling).
Reduce the gain value when noise peaks are above 20 % of the waveform
display height.
This prevents the gage from detecting a noise peak rather than the correct
back-wall echo.
3. Press [2nd F], [GAIN/WAVE ADJ] again to restore the default gain value.
4. Press [GAIN/WAVE ADJ] again to return to the automatically adjusted echo
height mode.
The default automatic gain mode is indicated by a blank Gain field.
Olympus recommends using the extended blank parameter only if you are an
experienced operator who thoroughly understands the acoustic properties of the
material being measured. Incorrect use of the extended blank can cause the gage to
misread areas of thin material.
Normally, the 45MG searches for echoes down to nearly zero thickness. However,
some special circumstances, such as a high degree of near-surface corrosion,
aluminum material, enclosed flaws, or laminations, can generate echoes that the
instrument may falsely detect as the low thickness. When these echoes are larger than
the sought after back-wall echo, the manual gain adjustment (see section 9.1 on
page 143) cannot prevent this false detection. However, the extended blank parameter
allows you to define an early period, during which the 45MG will not perform echo
detection, thereby preventing erroneous measurements.
Waveform adjustment
parameter and value
4. To turn off the extended blank, use the [] key to bring the extended blank value
back down to zero.
146 Chapter 9
DMTA-10022-01EN, Rev. A, September 2012
If the measurement point changes when the extended blank is moved, echoes can
change in height. This is because in the normal waveform display mode, the 45MG
attempts to adjust the height.
The instrument also attempts to make the most accurate measurement by identifying
the beginning of an echo. When the extended blank is positioned within an echo
instead of to its left, the gage is unable to make an accurate detection.
A B-scan is a cross-sectional image of thickness readings. The 45MG can acquire and
display B-scan data (see Figure 9-3 on page 147). When you activate the B-scan, the
thickness reading profile builds up and scrolls on the screen. Once a B-scan is
acquired, you can freeze the screen and review the recorded thickness values.
B-scan
While the B-scan is running, you can save individual thickness readings, the current
B-scan screen (with all thickness values), or the entire scan (up to 10000 readings) [see
section 9.3.3 on page 151].
The B-scan can be actively configured from the B-SCAN screen (see Figure 9-4 on
page 148), which is accessible by pressing [SETUP], and then selecting B-SCAN in the
menu.
DIRECTION:
Choose the B-scan direction to match the direction of transducer movement. A
scan direction arrow appears below the left corner of the B-scan display to
indicate the transducer scan direction (see Figure 9-5 on page 149). The data starts
appearing on-screen in the opposite direction.
LEFT TO RIGHT
The transducer scans the part from left to right, and the data appears on the
right of the screen, scrolling right to left.
RIGHT TO LEFT
The transducer scans the part from right to left, and the data appears on the
left of the screen, scrolling left to right.
148 Chapter 9
DMTA-10022-01EN, Rev. A, September 2012
Loss-of-signal (LOS)
indicator
Scan direction
(left to right) Min/Max marker B-scan range
LOS MODE
Determines how the B-scan behaves when a loss of signal (LOS) occurs.
STOP ON LOS
The B-scan stops scrolling when an LOS occurs. When the instrument regains
the measurements, a thin blank vertical line is inserted into the B-scan to
indicate that an LOS has occurred (see Figure 9-5 on page 149).
CONTINUE ON LOS
The B-scan continues scrolling when an LOS occurs.
FREEZE MODE
Determines which thickness readings are displayed when [FREEZE] is pressed
while a B-scan is active.
SHOW MINIMUM
Displays the thickness reading for the minimum reading collected during the
scan.
SHOW MAXIMUM
Displays the thickness reading for the maximum reading collected during the
scan.
SHOW CURRENT
Displays the last thickness reading made before pressing [FREEZE].
FREEZE REVIEW
When this feature is enabled and a B-scan is active, pressing [FREEZE] freezes the
B-scan image in review mode. In this mode, a vertical line (review marker)
appears, indicating the location of the displayed thickness (see Figure 9-6 on
page 150). The displayed thickness is either the minimum, maximum, or current
thickness, depending on which B-SCAN FREEZE MODE option is selected. Use
the [] and [] keys to move the review marker and read the thickness at the
review marker location.
If the minimum or maximum value moves off the B-scan screen, press [FREEZE] to
center the B-scan and the review marker on the minimum or maximum thickness.
Review marker
Minimum location
marker
150 Chapter 9
DMTA-10022-01EN, Rev. A, September 2012
The B-scan begins drawing the cross section of the material as soon as the first
non-LOS reading is collected.
4. To stop the collection of B-scan data:
Press [FREEZE].
OR
With B-SCAN LOS MODE set to STOP ON LOS, simply uncouple the
transducer from the test material.
5. While the B-scan is frozen, if B-SCAN FREEZE REVIEW is set to ON, use the []
and [] keys to move the review marker, and then read the thickness value
corresponding to the marker position.
6. Press [MEAS] again to reset the B-scan and begin a new B-scan.
7. Refer to section 9.3.3 on page 151 for details on how to save B-scan data.
152 Chapter 9
DMTA-10022-01EN, Rev. A, September 2012
When a B-scan screen is saved to the datalogger, the gage saves the thickness values
for the data points appearing on the display. All thickness values on a saved B-scan
can be reviewed during ID Review. Recall the saved B-scan, and use the [] and []
keys to review each thickness reading.
You can save a maximum of 10000 thickness readings to a B-scan. When the
maximum number of thickness points reaches 10000, the gage prompts you to save
the B-scan history, or to reset the B-scan without saving.
The database grid (DB grid) is a table representation of 2-D data. This representation
enables you to move freely in any direction on a grid, rather than follow a preset list of
IDs. Instead of automatically incrementing to the next ID location, you can use the
arrow keys to move to a location that is more convenient. You can simultaneously
display the A-scan, the DB grid, and the thickness reading (see Figure 9-7 on
page 154). You can configure the grid cells to show a data cell flag and a background
color corresponding to the range in which their thickness reading falls.
Half-size DB grid
154 Chapter 9
DMTA-10022-01EN, Rev. A, September 2012
The ID number increments in the order assigned when the file was initially set up
independent of the TRANSPOSE GRID value.
5. Set LINEARIZE GRID to ON to display the grid IDs in the linear form (see
Figure 9-10 on page 155).
6. Set DATA CELL FLAG to one of the available options in order to display a single
data flag with each data cell in the DB grid. The data cell flag is a letter that
appears in a small box to the right of the thickness value in the data cell (see
Figure 9-7 on page 154). The available options are:
NONE
No data cell flag appears.
MIN/MAX
m indicates a minimum thickness.
M indicates a maximum thickness.
ALARM
L indicates any kind of low alarm condition, including a standard low
alarm condition.
H indicates any high alarm condition.
7. Set GRID COLOR OPTION to ON to activate usage of low, mid, and high-range
colors for the grid cell background.
8. Set LO RANGE COLOR to the desired cell background color (RED, YELLOW, or
GREEN) when the cell thickness value is smaller than the LO RANGE VALUE.
9. Set MID RANGE COLOR to the desired cell background color (RED, YELLOW,
or GREEN) when the cell thickness value is between the LO RANGE VALUE and
the HI RANGE VALUE.
10. Set HI RANGE COLOR to the desired cell background color (RED, YELLOW, or
GREEN) when the cell thickness value is higher than the HI RANGE VALUE.
156 Chapter 9
DMTA-10022-01EN, Rev. A, September 2012
Highlighted cell
4. Press [MEAS] to return to the measurement screen with the current ID number
changed to the ID location selected in the ID review screen.
Inserted cell
158 Chapter 9
DMTA-10022-01EN, Rev. A, September 2012
Once the Single Element High Resolution option has been activated, the 45MG
includes predefined setups for standard single element transducers. In some cases, a
45MG ships from the factory preprogrammed with one or more custom setups to
meet special customer requirements. You can create your own custom setups to meet
the need of a particular single element transducer, or for a particular application.
Predefined and custom setups allow you to quickly change between transducer and
application setups by selecting a previously saved setup.
Creating a Custom Setup for a Single Element Transducer (see section 10.1 on
page 160).
Quickly Adjusting Waveform Parameters For Single Element Transducers (see
section 10.2 on page 162).
About the Detection Modes (see section 10.3 on page 164).
About the First Peak (see section 10.4 on page 166).
About the Pulser Power (see section 10.5 on page 167).
About the Time-Dependent Gain Curve (see section 10.6 on page 168).
About the Main Bang Blank (see section 10.7 on page 170).
About the Echo Window (see section 10.8 on page 172).
Saving Setup Parameters (see section 10.9 on page 177).
Quickly Recalling a Custom Setup for Single Element Transducers (see
section 10.10 on page 178).
Many of the adjustments that can be made in a setup are interactive. All adjustments
have an effect on the measurement range and/or measurement accuracy of the 45MG.
In most cases, do not attempt adjustments without monitoring the waveform.
Additionally, when establishing a custom setup for a specific application, it is
essential to verify the performance on reference standards representing the materials
and thickness range to be measured.
To reduce the number of changes needing to be made to parameter values, you can
also select an existing single element setup with parameter values that are close to the
desired setup.
160 Chapter 10
DMTA-10022-01EN, Rev. A, September 2012
Figure 10-1 The ACTIVE screen for the setup of a single element transducer
b) Enter a SETUP NAME that describes the transducer and/or the application to
be used to create the setup.
c) Set MEAS TYPE to the desired measurement type. The available choices are:
STANDARD: For normal mode 1, 2, or 3 positive or negative peak
measurement.
FIRST PEAK: For detection of the first of several peaks of similar
amplitudes (see section 10.4 on page 166 for details).
d) Set PROBE TYPE to indicate the transducer type to be used with this setup.
The selected probe type must match the frequency of the transducer being
used for proper pulser/receiver performance.
e) Set VELOCITY to the value of the sound velocity in the material that is to be
tested with this setup (see section 5.2.4 on page 57).
f) Set ZERO VALUE to the calibrated zero-offset value (time of flight of the echo
that is not traveling through the material) for this transducer (see section 5.2.1
on page 53).
g) If needed, increase the PULSER POWER to increase the ultrasound wave
penetration in the material. Reduce the value for better near-surface
resolution (see section 10.5 on page 167 for details).
h) Set MAX GAIN to the desired maximum gain value (see section 10.6.1 on
page 169 for details).
i) Set INIT GAIN to the desired initial gain value (see section 10.6.2 on page 169
for details).
j) Set TDG SLOPE to the desired time-dependent gain slope value (see
section 10.6.3 on page 170 for details).
k) Set MB BLANK to the desired main bang blank time interval (see section 10.7
on page 170 for details).
l) Set ECHO WINDOW to the desired time interval (see section 10.8 on
page 172 for details).
m) Set ECHO 1 DETECT to SLOPE to detect the negative peak of the first echo,
or to +SLOPE to detect the positive peak of the first echo (see section 10.8.1 on
page 173 for details).
6. Press [SAVE].
7. In the SAVE SETUP screen:
a) If needed, in the SAVE AS dialog box, edit the setup name.
b) In the SAVE TO list, select the desired custom setup location in which you
want to save the setup.
c) Select SAVE.
8. Press [MEAS] to return to the measurement screen.
The saved setup becomes the active setup.
With single element transducers, you can quickly adjust individual waveform
parameters using the [GAIN/WAVE ADJ] key.
162 Chapter 10
DMTA-10022-01EN, Rev. A, September 2012
Waveform parameter
Waveform parameter value
3. Use the [] or [] key to select the parameter that you want to adjust. The choices
are as follows:
DET MODE (see section 10.3 on page 164 for details)
M3 BLANK in mode 3 only (see section 10.8.3 on page 176 for details)
IF BLANK in modes 2 and 3 only (see section 10.8.2 on page 174 for details)
ECHO 2 DETECT in modes 2 and 3 only (see section 10.8.1 on page 173 for
details)
ECHO 1 DETECT (see section 10.8.1 on page 173 for details)
ECHO WINDOW (see section 10.8.1 on page 173 for details)
MB BLANK (see section 10.7 on page 170 for details)
TDG SLOPE (see section 10.6.3 on page 170 for details)
INIT GAIN (see section 10.6.2 on page 169 for details)
MAX GAIN (see section 10.6.1 on page 169 for details)
PULSER POWER (see section 10.5 on page 167 for details)
PROBE TYPE
MEAS TYPE (see step 5.c in section 10.1 on page 160 for details)
4. Use the [] or [] keys to select the value for the selected parameter.
5. Repeat steps 3 and 4 to adjust other parameters.
6. Press [GAIN/WAVE ADJ] again to hide the waveform adjustment parameter.
There are three detection modes (Mode 1, Mode 2, and Mode 3):
Mode 1
Uses a direct contact transducer to measure the time of flight between the main
bang and the first back-wall echo (see Figure 10-3 on page 164).
Main bang
Detection marker
Mode indicator
Mode 2
Measures the time of flight between the interface (or delay line) echo and first
back-wall echo using a delay line or an immersion transducer (see Figure 10-4 on
page 165).
164 Chapter 10
DMTA-10022-01EN, Rev. A, September 2012
Interface echo
Main bang
Detection bar
Mode indicator
Mode 3
Uses a delay line or an immersion transducer to measure the time of flight
between one back-wall echo and the next back-wall echo (see Figure 10-5 on
page 165).
Second back-wall
echo
Interface echo
Detection bar
Mode indicator
Refer to section 10.8 on page 172 for information on the modes in relation with the
echo window.
With single element transducers, the 45MG normally detects peaks on either the
highest positive or highest negative peak of the RF waveform. This feature works well
for most precision thickness applications.
This normal peak detection can be unstable for applications in which the back-wall
echo is irregular and contains several negative or positive peaks that are close in
amplitude. In such cases, the thickness reading can fluctuate as the gage switches
detection from one peak to another. This can happen, for example, in applications
measuring the length of bolts, or the thickness of the gel coat over fiberglass (see
Figure 10-6 on page 166). In such cases, help stabilize the echo detection and thickness
measurements by selecting the first peak algorithm to detect the first of several peaks
of similar amplitudes (see step 5.c in section 10.1 on page 160).
166 Chapter 10
DMTA-10022-01EN, Rev. A, September 2012
The excitation pulse (main bang) voltage can be set to one of the following values:
60 V, 110 V, 150 V, and 200 V.
Higher voltages may provide greater penetration at the expense of a lower near-
surface resolution, especially in mode 1. Inversely, lower voltages may provide better
near-surface resolution at the expense of a lower penetration.
For most applications, the 110 V value provides the best signal-to-noise ratio for
returning echoes. The pulser power indicates the voltage used to excite the
transducer, and therefore affects the size of the initial pulse (see Figure 10-7 on
page 167), and the amount of energy transmitted into the material.
When the 45MG displays the SAT flag below the thickness reading, the input voltage
from the transducer is above the maximum range, and proper measurements cannot
be made. This can normally be corrected by lowering the PULSER POWER until the
SAT flag no longer appears.
With single element transducers, the 45MG either uses the automatic gain control
(AGC) [see section 8.1 on page 137 for details on how to activate the AGC] or the time-
dependent gain (TDG) functions to automatically adjust receiver gain to an optimum
level when an echo is detected.
The 45MG offers three parameters for drawing a time-dependent gain curve: INIT
GAIN, TDG SLOPE, and MAX GAIN (see Figure 10-8 on page 168). From the initial
gain level, receiver gain slopes up to the maximum gain level at the rate determined
by the TDG SLOPE setting. When you adjust any of the received gain parameters, a
black time-dependent gain curve is displayed that clearly indicates the zones of initial
gain, slope, and maximum gain.
MB BLANK
Time-dependent gain
curve
The TDG curve can be used to optimize near-surface resolution while providing a
higher maximum gain for thicker samples. You can also use the TDG curve in
measurements of highly scattering materials, such as cast metals and fiberglass, in
order to minimize detection of scattered echoes occurring ahead of the back-wall
echo.
168 Chapter 10
DMTA-10022-01EN, Rev. A, September 2012
The maximum available receiver gain can be adjusted from 0.0 dB to 99.0 dB. When
no echo is detected (LOS prompt), the gain rises to the maximum level set by the
initial gain, slope, and maximum gain. When the maximum gain is set too high, the
gage may hang up on transducer noise or other spurious signals; when it is set too
low, returning echoes may not be high enough for detection.
The max gain can never be lower than the initial gain, and has a maximum value of
99.0 dB.
The INIT GAIN setting is most critical in applications where minimum thickness
measurements must be optimized. This setting should always be set with the aid of a
reference standard representing that minimum. In cases where the minimum
thickness capability is less important than penetration, and scatter echoes are not a
problem, the initial gain can be set to be equal to the maximum gain.
The main bang blank is effectively a blank zone that protects the receiver from false
readings generated by the main bang. This blank or dead zone (up to 18 microseconds
from the excitation pulse) prevents the trailing edge echoes of the excitation pulse
from being detected as back-wall or interface echoes. The end of the main bang blank
indicates the point in time at which the gage begins to search for echoes.
In general, set the main bang blank just beyond the point at which the gage hangs up,
and then proceed to test with the transducer both coupled to and uncoupled from the
test material in order to ensure accurate measurements.
In mode 1, however, the length of the main bang blank determines the minimum
thickness that can be measured, and must therefore be positioned with care after
selecting the initial gain level (see Figure 10-9 on page 171). When the main bang
blank is too short, the instrument hangs up on the excitation pulse and readings are
not possible. When the main bang blank is too long, the minimum measurable
thickness is unnecessarily restricted. When using immersion transducers, ensure that
the main bang blank is always set before the interface echo from the shortest water
path.
170 Chapter 10
DMTA-10022-01EN, Rev. A, September 2012
Initial pulse
In mode 2 and mode 3, the setting of the main bang blank is not critical if set to some
point between the end of the excitation pulse and the interface echo (see Figure 10-10
on page 171).
Interface echo
Figure 10-10 Main bang blank position for mode 2 and mode 3
The echo window is the time interval after each main bang, during which the
instrument is enabled to detect echoes. The echo window interval begins at the end of
the main bang blank. The echo window ends at x s following the main bang in
mode 1, or at x s following the interface blank in mode 2 and mode 3.
In mode 1, you can usually set the echo window to any value greater than the round-
trip pulse transit time in the thickest or slowest piece of material to be measured (see
Figure 10-11 on page 172). The exact setting is not critical as long as it is long enough
to include the farthest echo of interest.
Echo window
In mode 2 and mode 3, the echo window is limited to the time interval between
successive interface echoes (see Figure 10-12 on page 173). The end of the echo
window must be set ahead of the second interface echo in order to prevent an
incorrect detection, which in turn determines a maximum measurable thickness. In
applications involving mode 2 and mode 3 immersion measurements, the echo
window placement must apply throughout the range of the water paths to be used.
172 Chapter 10
DMTA-10022-01EN, Rev. A, September 2012
Echo window
Interface echo
Interface blank
Refer to Table 12 on page 174 for assistance in selecting the appropriate detection
polarity for a given application.
174 Chapter 10
DMTA-10022-01EN, Rev. A, September 2012
In mode 2, the interface blank prevents the detection of trailing lobes or cycles of the
interface echo which might otherwise be detected as back-wall echoes, which results
in a hang-up condition (see Figure 10-14 on page 175). The interface blank should be
set as short as possible to avoid unnecessarily restricting minimum measurable
thicknesses. The initial gain parameter often helps reduce the interface echo
amplitude, and permits the use of a shorter interface blank. Check the interface blank
settings with the transducer both coupled to and uncoupled from the test material.
In mode 3, the interface blank selects which pair of back-wall echoes are measured
(see Figure 10-15 on page 176). In most conditions, the interface blank would be set
just short of the first back-wall echo. However, as a practical matter, the first back-wall
echo from thin materials is often distorted or lost in the interface echo. With some
challenging geometries (such as tight radii), later pairs of back-wall echoes may be
cleaner than early ones. In such cases, set the interface blank to a length that ensures
detection of a clean and well-defined pair of back-wall echoes, even if the echoes are
not the first two.
Gage reading between back- Interface blank adjusted so the gage measures
wall echoes 1 and 2 between back-wall echoes 2 and 3
176 Chapter 10
DMTA-10022-01EN, Rev. A, September 2012
Because the mode 3 echo blank limits minimum measurable thicknesses, set the
controls to as short a period as possible; typically no longer than a few hundred
nanoseconds. Special cases may be an exception where mode conversion effects on
curved samples cause significant spurious signals between legitimate echo peaks. In
such cases, set the mode 3 echo blank to as long a period as necessary in order to
prevent detection of the spurious signals.
After adjusting selected waveform parameters, it is possible to store the settings for
quick and easy recall. The 45MG can store up to 35 custom setups in its internal
memory.
Setups named SE-USER-1 through SE-USER-35 are empty locations. Use caution
when selecting a location that already contains a custom setup, because the new setup
will replace the old information.
Normally, you can change the setup for a custom application by pressing
[RECALL XDCR] to select the appropriate setup in the list of available setups), and
then pressing [MEAS]. This process is adequate for typical setups that are changed
infrequently. However, when it is necessary to change between two or more custom
setups on a more frequent basis, use the quick setup recall function instead.
When the quick setup recall function is activated, a simple keypad shortcut can be
used to quickly switch to any of the first four custom setups for single element
transducers.
178 Chapter 10
DMTA-10022-01EN, Rev. A, September 2012
This feature only works when a single element transducer is plugged into the 45MG,
and the single element transducer option has also been purchased.
180 Chapter 10
DMTA-10022-01EN, Rev. A, September 2012
This section describes the process through which the 45MG communicates with a
computer to send, receive, import, and export files. The 45MG comes standard with a
USB cable for communication using the USB 2.0 protocol.
GageView is compatible with Windows XP, Windows Vista, and Windows 7. For
more details, refer to the GageView Interface Program Users Manual (P/N: 910-259-
EN [U8778347]).
For more information on installing GageView, refer to the GageView Interface Program
Users Manual (P/N: 910-259-EN [U8778347]).
2. If you use a communication program other than GageView, refer to the program
documentation for proper configuration of the program for USB communication.
3. Turn the 45MG on.
4. Connect one end of a USB cable to the USB client connector on the top of the
45MG, and connect the other end to a USB port of the computer (see Figure 11-1
on page 182).
USB cable
The first time the 45MG is connected to this computer, the computer alerts you
that a new hardware device has been detected, and asks you if you want to install
the driver. Refer to the GageView Interface Program Users Manual (P/N: 910-259-
EN [U8778347]) for more information.
The driver loads, after which you can start using a program like GageView to
communicate with the 45MG.
182 Chapter 11
DMTA-10022-01EN, Rev. A, September 2012
Data transmitted from the 45MG remains in the 45MG internal memory.
184 Chapter 11
DMTA-10022-01EN, Rev. A, September 2012
At the next survey date, months or years later, retrieve the previously saved
survey thickness data stored in a computer file with ID numbers. This retrieved
data can be used for the following purposes:
To guide you through the measurement sequence using the ID numbers.
To compare current values at the measurement site with previously measured
thickness values.
To manually or automatically verify that the current measurement setup is
identical to the previously used setup.
Create ID number sequences in a computer, and then download them to the
instrument. This externally created sequence can guide you through the
prescribed measurement location path. The ID sequence created in the computer
must have setup information. The setup can be the gage default setups, or any
other desired sequence of setups.
The data downloaded to the 45MG must be in the exact same format as the data
transmitted. Olympus recommends using the GageView interface program to
handle all functions of interfacing, storing, and creating 45MG data. Please
contact Olympus for information on additional software data management
programs.
186 Chapter 11
DMTA-10022-01EN, Rev. A, September 2012
It is possible to save a screenshot of the full 45MG screen contents to an image file.
This function is useful when you need an exact replica of the display for reporting or
documentation purposes, and can be performed using one of the following two
methods:
Refer to the GageView Interface Program Users Manual (P/N: 910-259-EN [U8778347])
for information on how to install and use GageView.
188 Chapter 11
DMTA-10022-01EN, Rev. A, September 2012
Figure 11-5 The Device Tools dialog box with a screen capture
190 Chapter 11
DMTA-10022-01EN, Rev. A, September 2012
192 Chapter 11
DMTA-10022-01EN, Rev. A, September 2012
This section describes how to maintain your 45MG by carrying out routine care and
maintenance.
The 45MG case is sealed to prevent ingress of environmental liquids and dust when
the battery door is closed. However, the instrument should never be immersed in any
fluid.
The 45MG case is designed to withstand normal field use. However, as with any
electronic instrument, damage is possible if the instrument is not handled with
adequate care. Observe the following instructions in particular:
When connecting cables to the instrument, first align the connector with the
mating receptacle on the instrument (dual element D79X series transducer center
pin protruding downward), and then carefully insert the connector straight into
the receptacle.
When disconnecting cables from the instrument, first grasp the connector (not the
cable), and then gently pull it out.
Do not throw or drop the instrument.
Do not use strong solvents or abrasives to clean the rubber boot, case, keypad, or
display screen.
First, clean the instrument with a dry damp cloth. If necessary, use a wet damp cloth
with a mild detergent to wash the instrument. Dry the instrument before using it.
The ultrasonic transducers used with the 45MG are rugged devices that need little
care. However, they are not indestructible, and some attention to the following items
result in a longer lifespan.
194 Chapter 12
DMTA-10022-01EN, Rev. A, September 2012
The 45MG includes reset functions that can be used to quickly restore the gage to its
default parameters. Resets are useful shortcuts to known configurations. The reset
functions are as follows:
MEASUREMENT RESET
Changes the measurement parameters to the default factory values listed in
Table 14 on page 195.
The internal memory reset permanently deletes all thickness readings/waveforms that
are stored in the memory card. However, this reset does not delete any stored
transducer setups.
CLOCK RESET
Resets the date to 01/01/2010 in the MM/DD/YYYY format, and the time to 12:00
A.M. in the 12-hour format.
COMMUNICATION RESET
Changes the communication settings to the default factory values listed in
Table 15 on page 196.
MASTER RESET
Performs the measurement reset and the internal memory reset in a single step.
The master reset permanently deletes all stored thickness readings/waveforms that
are stored in the internal memory card of the 45MG.
196 Chapter 12
DMTA-10022-01EN, Rev. A, September 2012
The 45MG includes a function that can be used to perform several diagnostic self
tests. The tests can help to localize a suspected hardware problem, or to verify correct
hardware operation. Some tests are designed for Olympus internal testing procedures
during the manufacturing process.
Keypad test
Video test
Internal MicroSD card test (Pass or Fail)
External MicroSD card test (Pass or Fail)
Dual element transducer test
ESS test (Electronic Stress Screening) [for Olympus internal use only]
B-scan test (for Olympus internal use only)
Battery test (for Olympus internal use only)
One-wire test (for Olympus internal use only)
198 Chapter 12
DMTA-10022-01EN, Rev. A, September 2012
FAIL: Indicates that there is a problem with the memory card. When the
external card fails, reinstall or replace the card, and then restart the
instrument. When the internal card fails, contact Olympus for service.
b) Press [ENTER] to terminate the SD CARD test.
7. When DUAL XDCR TEST is selected, in the DUAL XDCR TEST screen (see
Figure 12-3 on page 199):
a) Press [MEAS] to start the test that measures the time of flight through each
side of a dual element transducer.
The measured parameter values appear.
b) Interpret the Tx and the Rx values as follows:
Similar values indicate a normal dual element transducer.
A difference between the values indicates that the delay line wear is not
the same for each element.
A missing value indicates that a cable is broken, or that the element is not
functioning.
c) Take note of the calculated ZERO VALUE.
d) Press [ENTER] to terminate the dual element transducer test.
The software diagnostic (SW DIAG) function generates an error log documenting all
errors that occurred during instrument operation. Olympus uses this information to
troubleshoot the 45MG operating software.
The STATUS screen lists important information about the instrument. The status
screen displays the following information:
200 Chapter 12
DMTA-10022-01EN, Rev. A, September 2012
When operating the gage, certain error messages may appear. These messages usually
indicate a problem with the operating procedure, but certain messages might indicate
a physical problem with the gage itself. If you do not understand an error message,
contact Olympus for assistance.
The power indicator (in the upper-left corner of the display) contains bars that
indicate the percentage of remaining battery charge level. When the charge level
becomes low, the gage automatically turns off. If the gage turns off immediately after
you turn it on, or if you are unable to turn it on, the batteries are probably depleted.
202 Chapter 12
DMTA-10022-01EN, Rev. A, September 2012
204 Appendix A
DMTA-10022-01EN, Rev. A, September 2012
206 Appendix A
DMTA-10022-01EN, Rev. A, September 2012
a.The maximum thickness measuring capability depends on the transducer type, material con-
ditions, and temperature.
208 Appendix A
DMTA-10022-01EN, Rev. A, September 2012
210 Appendix A
DMTA-10022-01EN, Rev. A, September 2012
The following table lists the sound velocities in a variety of common materials. Use
this table only as a guide. The actual velocity in these materials may vary significantly
due to a variety of causes, such as composition, preferred crystallographic orientation,
porosity, and temperature. Therefore, for maximum accuracy when establishing the
sound velocity in a given material, first test a sample of the material.
References
Fredericks, J. R. Ultrasonic Engineering, John Wiley & Sons, Inc., New York, 1965.
Handbook of Chemistry and Physics Chemical Rubber Co., Cleveland, Ohio, 1963.
Mason, W. P. Physical Acoustics and the Properties of Solids, D.Van Nostrand Co., New
York, 1958.
212 Appendix B
DMTA-10022-01EN, Rev. A, September 2012
214 Appendix B
DMTA-10022-01EN, Rev. A, September 2012
Table 32 Couplants
Description Part Number
Glycerin, 2 oz. (0.06 l) Couplant B-2 [U8770023]
Gel type, 12 oz. (0.35 l) Couplant D-12 [U8770026]
Ultratherm-high temperature up to 1000 F Couplant E-2 [U8770274]
(538 C), 2 oz. (0.06 l)
Medium temperature up to 500 F (260 C), 2 oz. Couplant F-2 [U8770141]
(0.06 l)
216 Appendix C
DMTA-10022-01EN, Rev. A, September 2012
218 Appendix C
DMTA-10022-01EN, Rev. A, September 2012
List of Figures
Figure 6-32 Deleting the data of an ID range in the active file ...................................... 110
Figure 6-33 Warning message when resetting measurements ....................................... 111
Figure 6-34 The MEMORY status screen .......................................................................... 111
Figure 6-35 The ID overwrite protection message ........................................................... 112
Figure 6-36 Identifying the ID review screen ................................................................... 113
Figure 6-37 Editing the ID # edit mode ............................................................................. 115
Figure 6-38 The message when the edited ID is not in the database ............................ 115
Figure 6-39 The FILE SUMMARY report screen .............................................................. 118
Figure 6-40 The FILE SUMMARY report result screen ................................................... 118
Figure 6-41 The MIN/MAX SUMMARY report screen ................................................... 119
Figure 6-42 The FILE COMPARISON report screen ....................................................... 120
Figure 6-43 The FILE COMPARISON report result screen ............................................ 120
Figure 6-44 The ALARM SUMMARY report result screen ............................................ 121
Figure 6-45 The MIN/MAX SUMMARY report result screen ........................................ 121
Figure 6-46 Returning to the measurement screen .......................................................... 122
Figure 7-1 Normal differential mode (shown with Waveform option activated) ..... 124
Figure 7-2 The DIFF screen ............................................................................................... 125
Figure 7-3 Displaying the minimum and/or maximum thickness (shown with the
Waveform option activated) ........................................................................... 126
Figure 7-4 Example of a HIGH alarm indicator (shown with the Waveform option
activated) ........................................................................................................... 128
Figure 7-5 Example of a B-scan alarm mode .................................................................. 129
Figure 7-6 YEL (yellow) and RED alarm indicators (shown with the Waveform option
activated) ........................................................................................................... 130
Figure 7-7 Setting up the STANDARD alarm ................................................................ 131
Figure 7-8 Example of a locked function message in the help bar .............................. 133
Figure 7-9 The PASSWORD screen .................................................................................. 133
Figure 7-10 The LOCKS screen ........................................................................................... 134
Figure 8-1 The MEAS screen ............................................................................................. 138
Figure 8-2 The SYSTEM screen ......................................................................................... 140
Figure 8-3 The COMM screen ........................................................................................... 142
Figure 9-1 Manually adjusting the gain .......................................................................... 144
Figure 9-2 Adjusting the extended blank length ........................................................... 146
Figure 9-3 B-scan example ................................................................................................. 147
Figure 9-4 Changing B-scan parameters ......................................................................... 148
Figure 9-5 B-scan elements ................................................................................................ 149
Figure 9-6 B-scan freeze review elements ....................................................................... 150
Figure 9-7 Example of the half-size DB grid ................................................................... 154
Figure 9-8 Changing DB grid parameters ....................................................................... 154
Figure 9-9 Grid transposition example ............................................................................ 155
Figure 9-10 Example of a linearized DB grid ................................................................... 155
Figure 9-11 The highlighted DB grid cell in the ID review screen ................................ 157
List of Tables
Index
Index 225
DMTA-10022-01EN, Rev. A, September 2012
C compliance
calibration 52, 59 C-Tick (Australia) 4
instrument 53 EMC directive 12
locked 58 FCC (USA) 12
material sound velocity 57 ICES-001 (Canada) 12
test block 56 regulatory information 12
THRU-COAT 75 configuring
transducer zero compensation 56 beeper 41
Canada, ICES-001 compliance 12 communications 141
capturing screen with GageView 187 DB grid 154
caution notes differential mode 124
battery replacement 28 inactive time 41
cannot recover data after measurement reset instrument 137
110 measurement parameters 137
cannot recover deleted file content 108 radix type 41
cannot recover erased data 116 system parameters 140
experienced operator only 145, 160 connections 19
ingress protection guarantee 18 copying
internal memory reset deletes all data 196 file 105
master reset deletes all data 196 corrosion applications 74
replacing setup deletes original data 177 couplant types 62
CE marking 4 coupling technique 59
changing creating
active ID 114 custom setups for SE transducers 160
display settings 44 data file 90
echo detection mode 70 C-Tick mark 4
file data mode 103 curvature of test piece 60
highlighted cell in DB grid 156 custom setups 160
language 41 D
thickness-measurement resolution 48
danger notes
character
electric shock 1
cycle in traditional method 39
data
deleting 40
erasing in active file 116
inserting 40
exchange 183
China RoHS 4, 11
saving 64
cleaning, instrument 194
data file
clock
2-D grid 96
reset 196
boiler 100
setting 42
creating 90
coated material 74
incremental 91
color scheme 45
sequential 93
changing 44
sequential with custom points 95
communication
datalogger
configuring 141
about the 87
reset 190, 196
file based system 87
226 Index
DMTA-10022-01EN, Rev. A, September 2012
Index 227
DMTA-10022-01EN, Rev. A, September 2012
228 Index
DMTA-10022-01EN, Rev. A, September 2012
Index 229
DMTA-10022-01EN, Rev. A, September 2012
P communication 190
painted material 74 resolution, changing 48
parameter review screen 112
screens 36 reviewing stored data 114
selecting 36 RF rectification 79
password RoHS symbol 4, 11
setting 133 S
performing
safety
diagnostic tests 198
installation category 6
THRU-COAT calibration 75
instrument class 6
phase
labels and symbols 1
distortion 61
signal words 7
reversal 61
symbols 7
polarity of detection 174
saving
positive detection 173
B-scan 151
power
data 64
indicator 25
DB grid readings 157
preventing false min/max readings 127
setup parameters 177
product description 15
secondary key function 21
protective rubber boot 18
selecting
pulse power 167
menu command 35
Q parameter and value 36
quick setup recall, activating 178 range value 81
text edit mode 37
R sending
radix type 41 screen capture to GageView 187
range 80 screen capture to memory card 189
deleting 109 sequential data file 93
selecting value 81 sequential with custom points data file 95
recalling serial number
quick setup 179 label location 2
receiving serial number label
files from a computer 186 content 3
rectification setting
changing setting 44 alarms 130
indicator 32 clock 42
mode 78 ID overwrite protection 112
regulatory information 12 password 133
renaming a file 107 transducer 49
repair and modification 6 units 42
replacing the battery 28 USB communications 181
report, generating 117 setups
reports 117 creating for SE transducers 160
resets 195 custom for SE transducers 159
resetting loading 51
230 Index
DMTA-10022-01EN, Rev. A, September 2012
Index 231
DMTA-10022-01EN, Rev. A, September 2012
232 Index